		 Maintenance   
					® 
					Performance Test   
					Verify the probe accuracy by measuring a 25 kV dc ±0.25%   
					voltage source. When used with a compatible dc voltmeter, the   
					probe should measure the source with ±1% accuracy. No   
					calibration adjustments are provided.   
					80K-40   
					High Voltage Probe   
					Cleaning   
					Instruction Sheet   
					Use a soft cloth dampened in distilled water to clean the 80K-40.   
					Never use solvents or abrasive cleaners. Make sure the 80K-40 is   
					dry before reuse.   
					Introduction   
					The Model 80K-40 is a high voltage accessory probe designed to   
					extend the voltage measuring capability of an ac/dc voltmeter up   
					to 40,000 volts Overvoltage Category I. This means the probe can   
					only be used to make measurements on energy limited circuits   
					within equipment. Examples include high voltage within   
					televisions or photo copy machines. DO NOT use this probe to   
					measure high voltages on power distribution systems. The probe   
					is a precision 1000:1 voltage divider formed by two matched   
					resistors. The unusually high input impedance offered by these   
					resistors minimizes circuit loading and thereby, optimizes   
					measurement accuracy. A special plastic body houses the divider   
					and provides the user with isolation and protection from the   
					voltage being measured.   
					LIMITED WARRANTY & LIMITATION OF LIABILITY   
					Each Fluke product is warranted to be free from defects in material and   
					workmanship under normal use and service. The warranty period is one   
					year and begins on the date of shipment. Parts, product repairs and   
					services are warranted for 90 days. This warranty extends only to the   
					original buyer or end-user customer of a Fluke authorized reseller, and   
					does not apply to fuses, disposable batteries or to any product which, in   
					Fluke’s opinion, has been misused, altered, neglected or damaged by   
					accident or abnormal conditions of operation or handling. Fluke warrants   
					that software will operate substantially in accordance with its functional   
					specifications for 90 days and that it has been properly recorded on non-   
					defective media. Fluke does not warrant that software will be error free or   
					operate without interruption.   
					Fluke authorized resellers shall extend this warranty on new and unused   
					products to end-user customers only but have no authority to extend a   
					greater or different warranty on behalf of Fluke. Warranty support is   
					available if product is purchased through a Fluke authorized sales outlet   
					or Buyer has paid the applicable international price. Fluke reserves the   
					right to invoice Buyer for importation costs of repair/replacement parts   
					when product purchased in one country is submitted for repair in another   
					country.   
					Specifications   
					The 80K-40 will achieve rated accuracy when used with a   
					voltmeter (ac or dc) having an input impedance of 10 MΩ ±1.0%.*   
					Specifications for the probe are as follows:   
					†Voltage Range: 1 kV to 40 kV dc or peak ac, 28 kV rms ac   
					Input Resistance: 1000 MΩ   
					Division Ratio: 1000: 1 (1000X attenuator)   
					Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the   
					purchase price, free of charge repair, or replacement of a defective   
					product which is returned to a Fluke authorized service center within the   
					warranty period.   
					Accuracy DC:   
					20 kV to 35 kV: ±1% at 20°C to 30°C; add 1% at 10°C   
					<20°C and >30°C to 45°C. (For total measurement accuracy   
					add accuracy specification of voltmeter being used.)   
					To obtain warranty service, contact your nearest Fluke authorized service   
					center or send the product, with a description of the difficulty, postage and   
					insurance prepaid (FOB Destination), to the nearest Fluke authorized   
					service center. Fluke assumes no risk for damage in transit. Following   
					warranty repair, the product will be returned to Buyer, transportation   
					prepaid (FOB Destination). If Fluke determines that the failure was   
					caused by misuse, alteration, accident or abnormal condition of operation   
					or handling, Fluke will provide an estimate of repair costs and obtain   
					authorization before commencing the work. Following repair, the product   
					will be returned to the Buyer transportation prepaid and the Buyer will be   
					billed for the repair and return transportation charges (FOB Shipping   
					Point).   
					0 kV to <20 kV and >35 kV to 40 kV: ±2%.   
					Accuracy AC: 60 Hz, ±5%.   
					Safety: Meets IEC 1010-2-031:1993, Type B, 40 kV dc or, peak   
					ac, 28 kV rms ac, Overvoltage Category I (voltages derived from   
					limited energy transformer).   
					* The input impedance of Autoranging Fluke handheld digital   
					multimeters varies as a function of range. The only range that   
					deviates significantly from 10 MΩ is the 3V (Models 21, 23, 25,   
					27, 70, 73, 75, 77) or 4V (Models 10, 11, 12, 29, 79, 83, 85, 86,   
					87, 88) range where the impedance is 11.11 MΩ. To enhance the   
					measurement accuracy when using this range, apply a correction   
					factor of 0.99, i.e. multiply the displayed reading by .99.   
					THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND   
					IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,   
					INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF   
					MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.   
					FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,   
					INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,   
					INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF   
					WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY   
					OTHER THEORY.   
					† This probe is intended for low energy applications such as CRT   
					and similar circuits. Above 2000 meters altitude, and up to 5000   
					meters, derate linearly the working voltage from 40 kV peak to 28   
					kV peak, and derate linearly the transient overvoltage from 80 kV   
					peak to 57 kV peak. Transient overvoltage refers to micro-second   
					duration impulses caused by lightning or load switching. See   
					International Electrotechnical Commission Publication 664-1980,   
					Clause 3.9 Table II, and Appendix A.   
					Since some countries or states do not allow limitation of the term of an   
					implied warranty, or exclusion or limitation of incidental or consequential   
					damages, the limitations and exclusions of this warranty may not apply to   
					every buyer. If any provision of this Warranty is held invalid or   
					unenforceable by a court of competent jurisdiction, such holding will not   
					affect the validity or enforceability of any other provision.   
					Fluke Corporation   
					P.O. Box 9090   
					Everett, WA 98206-9090   
					U.S.A.   
					Fluke Europe B.V.   
					P.O. Box 1186   
					Measurement Considerations   
					Before attempting to use the 80K-40, the following paragraphs   
					should be read and understood. Particular attention should be   
					given to Operator Safety.   
					5602 BD Eindhoven   
					The Netherlands   
					PN 481978   
					January 1978, Rev.9, 3/97   
					© 1978, 1996, 1997 Fluke Corporation, All rights reserved. Printed in U.S.A.   