Fluke Portable Multimedia Player 8842a User Manual

®
8842A  
Digital Multimeter  
Instruction Manual  
PN 879309  
Date December 1991  
Rev.3 7/96  
© 1999 Fluke Corporation, All rights reserved. Printed in USA  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
Chapter  
1
Title  
Page  
Introduction and Specifications........................................................ 1-1  
1-1.  
1-2.  
1-3.  
1-4.  
INTRODUCTION................................................................................ 1-2  
THE 8842A DIGITAL MULTIMETER.............................................. 1-2  
OPTIONS AND ACCESSORIES........................................................ 1-2  
SPECIFICATIONS .............................................................................. 1-3  
2
Operating Instructions....................................................................... 2-1  
2-1.  
2-2.  
2-3.  
2-4.  
2-5.  
2-6.  
2-7.  
2-8.  
INTRODUCTION................................................................................ 2-2  
INSTALLATION................................................................................. 2-2  
Installing the Power-Line Fuse ........................................................ 2-2  
Connecting to Line Power................................................................ 2-2  
Adjusting the Handle........................................................................ 2-3  
Rack Mounting Kits......................................................................... 2-3  
OPERATING FEATURES .................................................................. 2-4  
Power-Up Features........................................................................... 2-4  
Front and Rear Panel Features ......................................................... 2-4  
Display Features............................................................................... 2-8  
Error Messages................................................................................. 2-8  
Overrange Indication........................................................................ 2-10  
Diagnostic Self-Tests ....................................................................... 2-10  
Ranging ............................................................................................ 2-10  
AUTORANGE............................................................................. 2-10  
MANUAL RANGE ..................................................................... 2-11  
Triggering......................................................................................... 2-11  
CONTINUOUS TRIGGER MODE ............................................ 2-11  
EXTERNAL TRIGGER MODE ................................................. 2-11  
Reading Rates and Noise Rejection ................................................. 2-11  
Automatic Settling Time Delay........................................................ 2-12  
External Trigger Input (Option -05 Only)........................................ 2-12  
Sample Complete Output (Option -05 Only) ................................... 2-12  
2-9.  
2-10.  
2-11.  
2-12.  
2-13.  
2-14.  
2-15.  
2-16.  
2-17.  
2-18.  
2-19.  
2-20.  
2-21.  
2-22.  
2-23.  
2-24. MAKING MEASUREMENTS............................................................ 2-13  
2-25.  
2-26.  
2-27.  
Input Overload Protection Limits..................................................... 2-13  
Measuring Voltage and Resistance .................................................. 2-13  
Measuring Current ........................................................................... 2-13  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
2-28.  
2-29.  
Current Fuse Protection.................................................................... 2-13  
Offset Measurements ....................................................................... 2-14  
2-30. EXTERNAL CLEANING.................................................................... 2-15  
3
Remote Programming........................................................................ 3-1  
3-1.  
3-2.  
3-3.  
3-4.  
3-5.  
3-6.  
3-7.  
3-8.  
INTRODUCTION................................................................................ 3-3  
CAPABILITIES ................................................................................... 3-3  
BUS SET-UP PROCEDURE............................................................... 3-3  
AN OVERVIEW OF REMOTE OPERATION................................... 3-4  
A NOTE ABOUT EXAMPLES........................................................... 3-6  
DEVICE-DEPENDENT COMMAND SET ........................................ 3-6  
Bn (Offset Commands) .................................................................... 3-9  
Cn (Calibration Commands) ............................................................ 3-9  
Dn (Display Commands).................................................................. 3-9  
Fn (Function Commands) ................................................................ 3-10  
Get Commands................................................................................. 3-10  
G0 (Get Instrument Configuration).................................................. 3-12  
G1 (Get SRQ Mask)......................................................................... 3-12  
G2 (Get Calibration Prompt)............................................................ 3-12  
G3 (Get User-Defined Message)...................................................... 3-13  
G4 (Get Calibration Status).............................................................. 3-13  
G5 (Get IAB Status)......................................................................... 3-14  
G6 (Get YW Status)......................................................................... 3-14  
G7 (Get Error Status) ....................................................................... 3-14  
G8 (Get Instrument Identification)................................................... 3-14  
N (Numeric Entry Command).......................................................... 3-15  
Put Commands ................................................................................. 3-15  
P0 (Put Instrument Configuration)................................................... 3-15  
P1 (Put SRQ Mask).......................................................................... 3-16  
P2 (Put Calibration Value).............................................................. 3-16  
P3 (Put User-Defined Message)...................................................... 3-17  
Rn (Range Commands).................................................................... 3-17  
Sn (Reading Rate Commands)......................................................... 3-17  
Tn (Trigger Mode Commands) ........................................................ 3-17  
Wn (Terminator Commands) ........................................................... 3-19  
X0 (Clear Error Register Command) ............................................... 3-19  
Yn (Suffix Commands).................................................................... 3-19  
Z0 (Self-Test Command) ................................................................. 3-19  
(Device-Clear Command) ................................................................ 3-20  
? (Single-Trigger Command) ........................................................... 3-20  
3-9.  
3-10.  
3-11.  
3-12.  
3-13.  
3-14.  
3-15.  
3-16.  
3-17.  
3-18.  
3-19.  
3-20.  
3-21.  
3-22.  
3-23.  
3-24.  
3-25.  
3-26.  
3-27.  
3-28.  
3-29.  
3-30.  
3-31.  
3-32.  
3-33.  
3-34.  
3-35.  
3-36. INPUT SYNTAX................................................................................. 3-20  
3-37.  
3-38.  
3-39.  
Definitions........................................................................................ 3-20  
Input Processing............................................................................... 3-21  
Syntax Rules .................................................................................... 3-23  
3-40. OUTPUT DATA .................................................................................. 3-24  
3-41.  
3-42.  
3-43.  
3-44.  
3-45.  
3-46.  
3-47.  
3-48.  
Loading Output Data........................................................................ 3-24  
Types of Output Data....................................................................... 3-25  
Numeric Data and Error Messages................................................... 3-25  
MEASUREMENT DATA........................................................... 3-25  
OVERRANGE INDICATION .................................................... 3-26  
ERROR MESSAGES .................................................................. 3-26  
Status Data ....................................................................................... 3-26  
Output Priority ................................................................................. 3-26  
3-49. SERVICE REQUESTS ........................................................................ 3-27  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
3-50.  
3-51.  
The Serial Poll Register ................................................................... 3-27  
The SRQ Mask................................................................................. 3-28  
3-52. INTERFACE MESSAGES .................................................................. 3-29  
3-53.  
3-54.  
3-55.  
Address Messages............................................................................ 3-29  
Universal Commands....................................................................... 3-29  
Addressed Commands...................................................................... 3-30  
3-56. TALK-ONLY MODE .......................................................................... 3-30  
3-57. REMOTE CALIBRATION.................................................................. 3-31  
3-58. TIMING CONSIDERATIONS ............................................................ 3-31  
3-59. IMMEDIATE MODE COMMANDS.................................................. 3-31  
3-60. EXAMPLE PROGRAMS .................................................................... 3-32  
4
Measurement Tutorial........................................................................ 4-1  
4-1.  
4-2.  
4-3.  
4-4.  
4-5.  
4-6.  
4-7.  
4-8.  
4-9.  
4-10.  
4-11.  
4-12.  
INTRODUCTION................................................................................ 4-2  
DC VOLTAGE MEASUREMENT ..................................................... 4-2  
Circuit Loading Error....................................................................... 4-2  
Input Bias Current Error................................................................... 4-3  
RESISTANCE MEASUREMENT ...................................................... 4-4  
2-Wire Ohms.................................................................................... 4-4  
Correcting for Test Lead Resistance in 2-Wire Ohms ..................... 4-5  
4-Wire Ohms.................................................................................... 4-5  
Applications of the Ohms Functions................................................ 4-7  
TESTING DIODES ..................................................................... 4-7  
TESTING ELECTROLYTIC CAPACITORS ............................ 4-7  
A PRECISION CURRENT SOURCE......................................... 4-8  
4-13. DC CURRENT MEASUREMENT ..................................................... 4-8  
4-14. REDUCING THERMAL VOLTAGES............................................... 4-9  
4-15. AC VOLTAGE AND CURRENT MEASUREMENT........................ 4-10  
4-16.  
4-17.  
4-18.  
4-19.  
4-20.  
4-21.  
4-22.  
True RMS Measurement.................................................................. 4-10  
Waveform Comparison.................................................................... 4-10  
Crest Factor...................................................................................... 4-12  
AC-Coupled AC Measurements ...................................................... 4-12  
Combined AC and DC Measurements............................................. 4-13  
Bandwidth ........................................................................................ 4-13  
Zero-Input VAC Error...................................................................... 4-13  
4-23. MAKING ACCURATE MEASUREMENTS ON THE 20 mV  
AND 20RANGES ............................................................................ 4-14  
4-24. MAKING ACCURATE HIGH-RESISTANCE MEASUREMENTS . 4-15  
5
Theory of Operation........................................................................... 5-1  
5-1.  
5-2.  
5-3.  
5-4.  
5-5.  
5-6.  
5-7.  
5-8.  
5-9.  
5-10.  
5-11.  
5-12.  
5-13.  
INTRODUCTION................................................................................ 5-3  
OVERALL FUNCTIONAL DESCRIPTION...................................... 5-3  
DETAILED CIRCUIT DESCRIPTION .............................................. 5-4  
DC SCALING ...................................................................................... 5-4  
VDC Scaling .................................................................................... 5-6  
VDC Protection................................................................................ 5-7  
mA DC Scaling ................................................................................ 5-7  
Analog Filter .................................................................................... 5-7  
TRACK/HOLD CIRCUIT ................................................................... 5-8  
Track Configuration......................................................................... 5-10  
Settling Configuration...................................................................... 5-11  
Hold Configuration .......................................................................... 5-11  
Pre-Charge Configuration................................................................ 5-11  
5-14. PRECISION VOLTAGE REFERENCE.............................................. 5-11  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
5-15. OHMS CURRENT SOURCE .............................................................. 5-12  
5-16. OHMS PROTECTION......................................................................... 5-13  
5-17. OHMS FUNCTIONS ........................................................................... 5-13  
5-18.  
5-19.  
2-Wire Ohms.................................................................................... 5-13  
4-Wire Ohms.................................................................................... 5-15  
5-20. A/D CONVERTER .............................................................................. 5-15  
5-21.  
5-22.  
5-23.  
5-24.  
Timing/Data Control........................................................................ 5-17  
Precision DAC ................................................................................. 5-18  
A/D Amplifier.................................................................................. 5-18  
Bootstrap Supplies ........................................................................... 5-18  
5-25. DISPLAY ............................................................................................. 5-19  
5-26. KEYBOARD........................................................................................ 5-19  
5-27. DIGITAL CONTROLLER................................................................... 5-19  
5-28.  
5-29.  
5-30.  
5-31.  
5-32.  
5-33.  
5-34.  
In-Guard Microcomputer ................................................................. 5-21  
Function and Range Control ............................................................ 5-23  
A/D Control and Computation......................................................... 5-23  
Calibration Correction...................................................................... 5-23  
Keyboard/Display Control ............................................................... 5-24  
Troubleshooting Modes ................................................................... 5-24  
Guard-Crossing Communication ..................................................... 5-24  
5-35. GUARD CROSSING ........................................................................... 5-24  
5-36. POWER SUPPLY ................................................................................ 5-25  
5-37. IEEE-488 INTERFACE (OPTION -05)............................................... 5-26  
5-38.  
5-39.  
5-40.  
5-41.  
5-42.  
Out-Guard Microcomputer............................................................... 5-26  
Guard Crossing................................................................................. 5-26  
Bus Interface Circuitry..................................................................... 5-26  
Signal Conditioning ......................................................................... 5-27  
IEEE-488 Interface Power Supply................................................... 5-27  
5-43. TRUE RMS AC (OPTION -09)........................................................... 5-27  
5-44.  
5-45.  
5-46.  
5-47.  
VAC Scaling .................................................................................... 5-27  
mA AC Scaling ................................................................................ 5-28  
Frequency Response Trimming ....................................................... 5-28  
True RMS AC-to-DC Conversion ................................................... 5-28  
6
Maintenance ....................................................................................... 6-1  
6-1.  
6-2.  
6-3.  
6-4.  
6-5.  
6-6.  
6-7.  
6-8.  
INTRODUCTION................................................................................ 6-3  
PERFORMANCE TEST...................................................................... 6-4  
Diagnostic Self-Tests ....................................................................... 6-4  
DC Voltage Test............................................................................... 6-5  
AC Voltage Test (Option -09 Only)................................................. 6-7  
Resistance Test................................................................................. 6-8  
DC Current Test............................................................................... 6-9  
AC Current Test (Option -09 Only)................................................. 6-10  
CALIBRATION................................................................................... 6-10  
Basic Calibration Procedure............................................................. 6-11  
INITIAL PROCEDURE.............................................................. 6-11  
A/D CALIBRATION .................................................................. 6-12  
OFFSET AND GAIN CALIBRATION ...................................... 6-14  
HIGH-FREQUENCY AC CALIBRATION................................ 6-15  
Advanced Features and Special Considerations............................... 6-16  
STORING VARIABLE INPUTS................................................ 6-16  
CALIBRATING INDIVIDUAL RANGES................................. 6-17  
VERIFYING CALIBRATION.................................................... 6-17  
ERASING CALIBRATION MEMORY..................................... 6-18  
6-9.  
6-10.  
6-11.  
6-12.  
6-13.  
6-14.  
6-15.  
6-16.  
6-17.  
6-18.  
6-19.  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
6-20.  
6-21.  
6-22.  
6-23.  
6-24.  
6-25.  
6-26.  
TOLERANCE CHECK............................................................... 6-18  
AC CALIBRATION AT OTHER FREQUENCIES ................... 6-19  
OPTIMIZING USE OF THE 5450A........................................... 6-19  
Remote Calibration .......................................................................... 6-20  
TIMING CONSIDERATIONS ................................................... 6-23  
REMOTE ERASURE.................................................................. 6-24  
EXAMPLE CALIBRATION PROGRAM.................................. 6-24  
6-27. DISASSEMBLY PROCEDURE.......................................................... 6-25  
6-28.  
6-29.  
6-30.  
6-31.  
6-32.  
Case Removal................................................................................... 6-26  
True RMS AC PCA Removal (Option -09 Only)............................ 6-30  
IEEE-488 Interface PCA Removal (Option -05 Only) .................... 6-30  
Main PCA Removal ......................................................................... 6-30  
Front Panel Disassembly.................................................................. 6-32  
6-33. REASSEMBLY PROCEDURE........................................................... 6-32  
6-34. INTERNAL FUSE REPLACEMENT ................................................. 6-36  
6-35. EXTERNAL TRIGGER POLARITY SELECTION (Option -05 Only)6-36  
6-36. TROUBLESHOOTING ....................................................................... 6-36  
6-37.  
6-38.  
6-39.  
6-40.  
6-41.  
6-48.  
6-59.  
6-62.  
6-63.  
6-64.  
6-65.  
6-66.  
6-67.  
6-68.  
6-69.  
6-70.  
6-71.  
6-72.  
6-73.  
6-74.  
6-75.  
Initial Troubleshooting Procedure.................................................... 6-36  
Diagnostic Self-Tests ....................................................................... 6-41  
Self-Test Descriptions...................................................................... 6-42  
Digital Controller Troubleshooting.................................................. 6-45  
IN-GUARD MICROCOMPUTER SYSTEM............................. 6-45  
DISPLAY SYSTEM.................................................................... 6-49  
ANALOG CONTROL SIGNALS............................................... 6-51  
DC Scaling Troubleshooting............................................................ 6-54  
Track/Hold Troubleshooting............................................................ 6-56  
Ohms Current Source Troubleshooting............................................ 6-56  
Precision Voltage Reference Troubleshooting................................. 6-58  
A/D Converter Troubleshooting....................................................... 6-59  
Power Supply Troubleshooting........................................................ 6-61  
IEEE-488 Interface Troubleshooting (Option -05) .......................... 6-64  
SERVICE POSITION.................................................................. 6-64  
DIAGNOSTIC PROGRAM ........................................................ 6-64  
True RMS AC Troubleshooting (Option -09).................................. 6-66  
SERVICE POSITION.................................................................. 6-66  
MAJOR PROBLEMS.................................................................. 6-67  
MORE OBSCURE PROBLEMS ................................................ 6-69  
Guard Crossing Troubleshooting ..................................................... 6-70  
6-76. INTERNAL CLEANING..................................................................... 6-70  
6-77.  
6-78.  
Cleaning Printed Circuit Assemblies ............................................... 6-70  
Cleaning After Soldering ................................................................. 6-71  
7
8
List of Replaceable Parts................................................................... 7-1  
7-1.  
7-2.  
7-3.  
7-4.  
7-5.  
INTRODUCTION................................................................................ 7-3  
HOW TO OBTAIN PARTS................................................................. 7-3  
MANUAL STATUS INFORMATION................................................ 7-3  
NEWER INSTRUMENTS................................................................... 7-4  
SERVICE CENTERS........................................................................... 7-4  
Options and Accessories .................................................................. 8-1  
8-1.  
8-2.  
8-3.  
8-4.  
INTRODUCTION................................................................................ 8-3  
ACCESSORIES ................................................................................... 8-4  
Rack-Mount Kits (Y8834, Y8835 and Y8836)................................ 8-4  
Shielded IEEE-488 Interface Cables (Y8021, Y8022, and Y8023). 8-4  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
8-5.  
8-6.  
8-7.  
8-8.  
Replacement Test Leads (TL70A)................................................... 8-4  
Deluxe Test Lead Kits (Y8134) ....................................................... 8-4  
Slim-Flex Test Leads (Y8140)......................................................... 8-4  
Temperature Probes (80T-150U, and 80TK.).................................. 8-4  
RF Probes (85RF and 83RF)............................................................ 8-4  
Current Shunt (80J-10)..................................................................... 8-5  
Current Probes (Y8100, Y8101, 80i-400 and 80i-600).................... 8-5  
High Voltage Probes (80K-6 and 80K-40) ...................................... 8-5  
8-9.  
8-10.  
8-11.  
8-12.  
805  
Option –05 IEEE-488 Interface........................................................... 805-1  
805-1. INTRODUCTION................................................................................ 805-3  
805-2. CAPABILITIES ................................................................................... 805-3  
805-3. EXTERNAL CONTROLS................................................................... 805-3  
805-4. INSTALLATION................................................................................. 805-3  
805-5. PROGRAMMING INSTRUCTIONS.................................................. 805-4  
805-6. MAINTENANCE................................................................................. 805-4  
805-7. LIST OF REPLACEABLE PARTS..................................................... 805-4  
809  
Option –09 True RMS AC................................................................... 809-1  
809-1. INTRODUCTION................................................................................ 809-3  
809-2. INSTALLATION................................................................................. 809-3  
809-3. OPERATING INSTRUCTIONS.......................................................... 809-4  
809-4. MAINTENANCE................................................................................. 809-4  
809-5. LIST OF REPLACEABLE PARTS..................................................... 809-4  
9
Schematic Diagrams.......................................................................... 9-1  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Tables  
Table  
Title  
Page  
2-1. Error Codes............................................................................................................ 2-9  
2-2. Input Overload Limits............................................................................................ 2-13  
3-1. Status Data............................................................................................................. 3-11  
3-2. Numeric Output Data Format ................................................................................ 3-25  
3-8. Serial Poll Register ................................................................................................ 3-28  
3-3. Immediate-Mode Commands for Various Controllers .......................................... 3-32  
3-4. ASCII/IEEE Std 488-1978 Bus Codes .................................................................. 3-49  
4-1. Ohms Test Current................................................................................................. 4-6  
5-1. Sample Rates and Reading Rates........................................................................... 5-23  
6-1. Recommended Test Equipment............................................................................. 6-3  
6-2. DC Voltage Test .................................................................................................... 6-6  
6-3. Low- and Mid-Frequency AC Voltage Test .......................................................... 6-7  
6-4. High-Frequency AC Voltage Test ......................................................................... 6-8  
6-5. Resistance Test ...................................................................................................... 6-9  
6-6. DC Current Test..................................................................................................... 6-9  
6-7. AC Current Test..................................................................................................... 6-10  
6-8. A/D Calibration Steps............................................................................................ 6-13  
6-9. A/D Calibration Verification Test ......................................................................... 6-13  
6-10. Offset and Gain Calibration Steps ......................................................................... 6-14  
6-11. High-Frequency AC Calibration Steps .................................................................. 6-16  
6-12. Prompts When Calibrating Individual Ranges....................................................... 6-17  
6-13. Tolerance Limits.................................................................................................... 6-19  
6-14. Commands Used During remote Calibration......................................................... 6-21  
6-15. Error Numbers Which Are Displayed When Commands Are Not Valid .............. 6-23  
6-16. Overall State Table ................................................................................................ 6-38  
6-17. Circuitry Tested by the Analog Self-Tests............................................................. 6-40  
6-18. Self-Test Voltages.................................................................................................. 6-42  
6-19. Keyboard Wiring ................................................................................................... 6-51  
6-20. Analog Control Devices......................................................................................... 6-52  
6-21. Analog Control Logic States.................................................................................. 6-53  
6-22. DC Scaling and Track/Hold Supply Voltages ....................................................... 6-55  
6-23. Power Supply Voltages.......................................................................................... 6-63  
6-24. Diagnostic Modes .................................................................................................. 6-65  
6-25. I/O Port Configurations ......................................................................................... 6-66  
6-26. Isolating a Defective AC Stage.............................................................................. 6-67  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
6-27. AC Signal Tracing ................................................................................................. 6-68  
6-28. Truth Table for U804 and K2 ................................................................................ 6-69  
7-1. 8842A Digital Multimeter...................................................................................... 7-5  
7-2. A1 Main PCA ........................................................................................................ 7-10  
7-3. A2 Display PCA .................................................................................................... 7-14  
8-1. Accessories ............................................................................................................ 8-3  
8-2. Options................................................................................................................... 8-3  
805-1. Option -05A IEEE-488 Interface PCA .................................................................. 805-7  
809-1. Option -09 True RMS AC PCA............................................................................. 809-6  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Figures  
Figure  
Title  
Page  
1-1. External Dimensions.............................................................................................. 1-11  
2-1. Line Voltage Selection Settings............................................................................. 2-2  
2-2. Adjusting the Handle ............................................................................................. 2-3  
2-3. Rack-Mount Kits.................................................................................................... 2-3  
2-4. Installing the Single Rack Mount Kit .................................................................... 2-4  
2-5. Front Panel Features .............................................................................................. 2-5  
2-6. Rear Panel Features ............................................................................................... 2-7  
2-7. Typical Error Messages ......................................................................................... 2-8  
2-8. Overrange Indication ............................................................................................. 2-12  
2-9. Measuring Voltage and Resistance........................................................................ 2-15  
2-10. Measuring Current ................................................................................................. 2-15  
3-1. IEEE-488 Address Selection.................................................................................. 3-4  
3-2. Remote Operation Block Diagram......................................................................... 3-5  
3-3. Typical Command String....................................................................................... 3-6  
3-4. Commands Which Correspond to the Front Panel................................................. 3-7  
3-5. Device-Dependent Command Set.......................................................................... 3-8  
3-6. Output Data Format ............................................................................................... 3-10  
3-6. Trigger Selection Logic Diagram .......................................................................... 3-18  
3-7. Interpretation of Messages..................................................................................... 3-22  
3-9. Example Program................................................................................................... 3-33  
3-10. Example Program: Taking Readings with Local Control...................................... 3-34  
3-11. Example Program: Using the Serial Poll Register................................................. 3-35  
3-12. Example Program: Record Errors During Selftest................................................. 3-36  
3-13. Example Programs: Using the IBM PC................................................................. 3-37  
4-1. Circuit Loading Error Calculation ......................................................................... 4-2  
4-2. Measuring Input Bias Current Error ...................................................................... 4-3  
4-3. Wire Ohms Measurement ...................................................................................... 4-4  
4-4. Wire Ohms Measurement ...................................................................................... 4-6  
4-5. Burden Voltage Error Calculation ......................................................................... 4-9  
4-6. Waveform Comparison Chart................................................................................ 4-11  
4-7. Typical Crest Factors for Various Waveforms ...................................................... 4-12  
4-8. Combined AC and DC Measurement .................................................................... 4-13  
4-9. Reduction of Zero-Input Error............................................................................... 4-14  
4-10. Shielding for Low Voltage Measurements ............................................................ 4-15  
4-11. Shielding for Low Resistance Measurements........................................................ 4-15  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
4-12. Leakage Resistance in High Resistance Measurement.......................................... 4-16  
5-1. Overall Functional Block Diagram........................................................................ 5-5  
5-2. DC Scaling (VDC and mA DC)............................................................................. 5-6  
5-3. Track/Hold Amplifier ............................................................................................ 5-8  
5-4. Track/Hold Circuit Configurations........................................................................ 5-9  
5-5. Timing Diagram for One A/D Cycle ..................................................................... 5-10  
5-6. Precision Voltage Reference.................................................................................. 5-11  
5-7. Ohms Current Source............................................................................................. 5-12  
5-8. Ohms Scaling......................................................................................................... 5-14  
5-9. Analog-to-Digital Converter.................................................................................. 5-16  
5-10. First Remainder-Store Period ................................................................................ 5-17  
5-11. Autozero Period ..................................................................................................... 5-18  
5-12. Vacuum Fluorescent Display................................................................................. 5-19  
5-13. Digital Controller Block Diagram.......................................................................... 5-20  
5-14. Read/Write Timing Diagrams for Internal Bus...................................................... 5-22  
5-15. Guard Crossing Circuit .......................................................................................... 5-24  
5-16. IEEE-488 Interface Block Diagram....................................................................... 5-26  
5-17. True RMS AC Option Block Diagram .................................................................. 5-27  
5-18. True RMS AC-to-DC Converter............................................................................ 5-28  
6-1. DC Calibration Connections.................................................................................. 6-6  
6-2. First A/D Calibration Prompt................................................................................. 6-11  
6-3. Calibration Functions............................................................................................. 6-12  
6-4. Optimizing Use of the 5450A................................................................................ 6-20  
6-5. Example A/D Calibration Program........................................................................ 6-25  
6-6. 8842A Disassembly ............................................................................................... 6-26  
6-7. Front Panel Disassembly........................................................................................ 6-34  
6-8. Removing the Display Window............................................................................. 6-35  
6-9. U202 Pin Diagram ................................................................................................. 6-46  
6-10. Waveforms for In-Guard Troubleshooting Mode.................................................. 6-47  
6-11. Waveforms for Display Logic ............................................................................... 6-49  
6-12. Typical Dynamic Control Signals.......................................................................... 6-54  
6-13. Typical Output Waveforms for Track/Hold Circuit (TP103) ................................ 6-57  
6-14. Output of A/D Amplifier (TP101)......................................................................... 6-59  
6-15. Waveforms at U101-24 and U101-25.................................................................... 6-60  
6-16. Typical Bus Data Line Waveform......................................................................... 6-60  
6-17. Waveforms at TP102 for Several Inputs on 2V DV Range................................... 6-62  
6-18. Calculating the A/D Reading From TP102 Waveform.......................................... 6-63  
6-19. Option -05 Service Position................................................................................... 6-65  
6-20. Option -09 Service Position................................................................................... 6-68  
6-21. Guard Crossing Test Waveforms........................................................................... 6-71  
7-1. 8842A Digital Multimeter...................................................................................... 7-6  
7-2. A1 Main PCA ........................................................................................................ 7-13  
7-3. A2 Display PCA .................................................................................................... 7-15  
7-4. Service Centers ...................................................................................................... 7-18  
7-4. Service Centers (cont)............................................................................................ 7-19  
805-1. Installing Option -05.............................................................................................. 805-5  
805-2. IEEE-488 Interface PCA........................................................................................ 805-8  
809-1. Installing Option -09.............................................................................................. 809-5  
809-2. True RMS AC PCA............................................................................................... 809-7  
9-1. Main PCA, DC Scaling and F/R Switch................................................................ 9-3  
9-2. Main PCA, A/D Converter .................................................................................... 9-5  
9-3. Main PCA, Ohms Current Source ......................................................................... 9-7  
9-4. Main PCA, Digital ................................................................................................. 9-9  
9-5. Main PCA, Power Supply...................................................................................... 9-11  
9-6. Display PCA .......................................................................................................... 9-13  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
9-7. IEEE-488 Interface PCA, Option -05 .................................................................... 9-15  
9-8. True RMS AC PCA, Option -09............................................................................ 9-17  
xi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
xii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1  
Introduction and Specifications  
Title  
Page  
1-1.  
1-2.  
1-3.  
1-4.  
INTRODUCTION................................................................................1-2  
SPECIFICATIONS ..............................................................................1-3  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
1-1. INTRODUCTION  
This manual provides complete operating instructions and service information for the  
8842A. If you want to get started using your 8842A right away, proceed to the operating  
instructions in Section 2. If you intend to use the 8842A with the IEEE-488 Interface  
(Option -05), read Sections 2 and 3. This meter has been designed and tested according to  
IEC publication 348, Safety Requirements for Electronic Measuring Apparatus. This  
manual contains information and warnings which must be followed to ensure safe  
operation and retain the meter in safe condition.  
1-2. THE 8842A DIGITAL MULTIMETER  
The Fluke 8842A Digital Multimeter is a high-performance 5-1/2 digit instrument  
designed for general-purpose bench or systems applications. The 8842A is the top-of-the-  
line DMM in the 8840A family. Using proprietary thin film resistor networks, a stable  
reference amplifier and stable active components, the 8842A offers superior  
measurement performance and stability. It also offers additional 20 mV, 20 ohm, and  
200 mA dc ranges. Features of the 8842A include:  
Highly legible vacuum fluorescent display  
Intuitively easy front panel operation  
Basic dc accuracy of 0.003% for 1 year  
2-wire and 4-wire resistance measurement  
DC current measurement  
Up to 100 readings per second  
Closed-case calibration (no internal adjustments)  
Built-in self-tests  
1-3. OPTIONS AND ACCESSORIES  
A number of options and accessories are available for the 8842A which can be easily  
installed at any time. The options include:  
IEEE-488 Interface (Option -05), featuring:  
Full programmability  
Simple and predictable command set  
Fast measurement throughput  
External Trigger input connector  
Sample Complete output connector  
Automated calibration  
Low cost  
True RMS AC (Option -09), featuring:  
AC voltage measurement  
AC current measurement  
Accessories include a variety of rack mounting kits, probes, test leads, and cables. Full  
information about options and accessories can be found in Section 8.  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Introduction and Specifications  
SPECIFICATIONS  
1
1-4. SPECIFICATIONS  
Specifications for the 8842A are given in Table 1-1. External dimensions are shown in  
Figure 1-1.  
Table 1-1. Specifications  
DC VOLTAGE  
Input Characteristics  
FULL SCALE 5ñ  
RESOLUTION  
INPUT  
DIGITS  
RESISTANCE  
RANGE  
5ñ DIGITS  
4ñ DIGITS*  
1 µV  
20 mV  
200 mV  
2V  
19.9999 mV  
199.999 mV  
1.99999V  
19.9999V  
199.999V  
1000.00V  
0.1µV  
1µV  
10,000 MΩ  
10,000 MΩ  
10,000 MΩ  
10,000 MΩ  
10 MΩ  
10 µV  
10 µV  
100 µV  
1 mV  
100 µV  
1 mV  
20V  
200V  
1000V  
10 mV  
100 mV  
10 mV  
10 MΩ  
*4ñ digits at the fastest reading rate.  
Accuracy  
NORMAL (S) READING RATE ............. ±(% of Reading + Number of Counts)  
RANGE  
24 HOURî 23±1°C  
90 DAY 23±5°C  
1 YEAR 23±5°C  
2 YEAR 23±5°C  
20 mV2  
0.0050 + 203  
0.0030 + 2  
0.0070 + 303  
0.0045 + 3  
0.0100 + 303  
0.0070 + 3  
0.0120 + 403  
0.0100 + 4  
200 mV2  
2V  
0.0015 + 2  
0.0015 + 2  
0.0015 + 2  
0.0020 + 2  
0.0025 + 2  
0.0030 + 2  
0.0030 + 2  
0.0035 + 2  
0.0030 + 2  
0.0035 + 2  
0.0035 + 2  
0.0045 + 2  
0.0050 + 3  
0.0060 + 3  
0.0060 + 3  
0.0070 + 3  
20V  
200V  
1000V  
1. Relative to calibration standards.  
2. Within one hour of dc zero, using offset control.  
3. When offset control is not used the number of counts are 50, 70, 90 and 90 for 24 hours, 90 day, 1  
year, and 2 year respectively.  
4. When offset control is not used the number of counts are 5, 7, 9 for 24 hours, 90 day, 1 year, and 2  
year respectively.  
MEDIUM AND FAST RATES: ................In medium rate, add 3 counts (20 counts on 20 mV Range) to  
number of counts. In fast rate, use 2 (4½ digit mode) counts  
(30 counts on 20 mV range) for the number of counts  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Operating Characteristics  
TEMPERATURE COEFFICIENT ...........<(0.0006% of reading + 0.3 Count) per °C from 0°C to 18°C  
and 28°C to 50°C.  
MAXIMUM INPUT...................................1000V dc or peak ac on any range.  
NOISE REJECTION................................Automatically optimized at power-up for 50, 60, or 400 Hz.  
RATE  
READINGS/  
SECONDî  
FILTER  
NMRRï  
PEAK NM  
SIGNAL  
CMRRð  
S
M
F
2.5  
20  
Analog & Digital  
Digital  
>98 dB  
>45 dB  
_
20V or 2x FS  
1x FS  
>140 dB  
>100 dB  
>60 dB  
100  
None  
1x FS  
1. Reading rate with internal trigger and 60 Hz power line frequency. See “reading rates” for more detail.  
2. Normal Mode Rejection Ratio, at 50 or 60 Hz ±0.1%. The NMRR for 400 Hz ±0.1% is 85 dB in S rate  
and 35 dB in M rate.  
3. Common Mode Rejection Ratio at 50 or 60 Hz ±0.1%, with 1 kin series with either lead. The CMRR  
is >140 dB at dc for all reading rates.  
4. 20 volts or 2 times full scale whichever is greater, not to exceed 1000V.  
5. Reading rate-1/3 rdg / sec. in the 20 mV, 20, 200 mA dc ranges  
6. Reading rate-1.25 rdg / sec. in the 20 mV, 20, 200 mA dc ranges  
TRUE RMS AC VOLTAGE (OPTION 8842A-09)  
Input Characteristics  
RANGE  
FULL SCALE 5ñ  
RESOLUTION  
INPUT  
DIGITS  
IMPENDANCE  
5ñ DIGITS  
4ñ DIGITS*  
200 mV  
2V  
199.999 mV  
1.99999V  
19.9999V  
199.999V  
700.00V  
1 µV  
10 µV  
100 µV  
1 mV  
10 µV  
100 µV  
1 mV  
1 MΩ  
Shunted  
By  
20V  
200V  
700V  
10 mV  
100 mV  
<100 pF  
10 mV  
*4ñ digits at the fastest reading rate  
1-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction and Specifications  
SPECIFICATIONS  
1
Accuracy  
NORMAL (s) READING RATE .............. ±(% of Reading + Number of Counts).  
For sinewave inputs 10,000 counts1.  
FREQUENCY  
24 HOURS2 23±1°C  
90 DAY 23±5°C  
1 YEAR 23±5°C  
2 YEARS ±5°C  
20-45  
45-200  
1.2 + 100  
0.3 + 100  
1.2 + 100  
1.2 + 100  
0.4 + 100  
1.2 + 100  
0.5 + 100  
0.35 + 100  
200-20k  
(200 mV range)  
(2V-200V range)  
(700V range)  
20k-50k  
0.06 + 100  
0.05 + 80  
0.06 + 100  
0.15 + 120  
0.4 + 300  
0.08 + 100  
0.07 + 80  
0.08 + 100  
0.19 + 150  
0.5 + 300  
0.10 + 100  
0.08 + 80  
0.10 + 100  
0.21 + 200  
0.5 + 400  
0.20 + 100  
0.15 + 80  
0.20 + 100  
0.25 + 250  
0.5 + 500  
50k-100k  
1. For sinewave inputs between 1,000 and 10,000 counts, add to number of counts 100 counts for  
frequencies 20 Hz to 20 kHz, 200 counts for 20 kHz, and 500 counts for 50 kHz to 100 kHz.  
2. Relative to calibration standards.  
MEDIUM AND FAST READING RATES........In medium rate, add 50 counts to number of counts. In the  
fast rate the specifications apply for sinewave inputs  
1000 (4½ digit PRGHꢀ FRXQWV and >100 Hz.  
NONSINUSOIDAL INPUTS ...........................For nonsinusoidal inputs 10,000 counts with frequency  
components 100 kHz, add the following % of reading to  
the accuracy specifications.  
FUNDAMENTAL  
FREQUENCY  
CREST FACTOR  
1.5 TO 2.0  
1.0 TO 1.5  
2.0 TO 3.0  
45 Hz to 20 kHz 20 Hz  
0.05  
0.2  
0.15  
0.7  
0.3  
1.5  
20 Hz to 45 Hz and 20  
kHz to 50 kHz  
Operating Characteristics  
MAXIMUM INPUT...................................700V rms, 1000V peak or 2 x 107 Volts-Hertz product  
(whichever is less) for any range.  
TEMPERATURE COEFFICIENT............±(% of reading + Number of Counts) per °C, to 18°C and 28°C  
to 50°C.  
FREQUENCY IN HERTZ  
FOR INPUTS  
20-20k  
20k-50k  
50k-100k  
10,000 counts  
1,000 counts  
0.019 + 9  
0.021 + 9  
0.027 + 10  
0.027 + 21  
0.019 + 12  
0.021 + 15  
COMMON MODE REJECTION ..............>60 dB at 50 or 60 Hz with 1 kin either lead.  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
CURRENT  
Input Characteristics  
RANGE  
FULL SCALE 5½  
DIGITS  
RESOLUTION  
5½ DIGITS  
4½ DIGITS1  
200 mA 2  
2000 mA  
199.999 mA  
1999.99 mA  
1 µA  
10 µA  
10 µA  
100 µA  
1. 4½ digits at the fastest reading rate.  
2. The 200mA range is available for dc current only.  
DC Accuracy  
NORMAL (S) READING RATE...............±(% of reading + number of counts).  
RANGE  
90 DAYS 23±5°C  
1 YEAR 23±5°C  
2 YEARS 23±5°C  
200 mA  
0.04 + 40  
0.05 + 40  
0.08 + 40  
2000 mA  
1A  
0.04 + 4  
0.1 + 4  
0.05 + 4  
0.1 + 4  
0.08+4  
0.15+4  
>1A  
MEDIUM AND FAST READING RATES  
In medium reading rate, add 2 counts (20 counts on 200 mA  
range) to number of counts. In fast reading rate, use 2 (4½  
digit mode) counts (20 counts on 200 mA range) for number  
of counts.  
AC Accuracy (Option –09)  
NORMAL (S) READING RATE...............±(% of Reading + Number of Counts).  
23±5°C, for sinewave inputs 10,000 counts1.  
FREQUENCY IN HERTZ  
20-45  
45-100  
100-5K*  
0.4 + 200  
0.6 + 300  
ONE YEAR  
TWO YEAR  
2.0 + 200  
3.0 + 300  
0.5 + 200  
0.7 + 300  
*Typically 20 kHz  
1. For sinewave inputs between 1,000 and 10,000 counts, add to number of counts 100 counts for  
frequencies 20 Hz to 5 kHz (typically 20 kHz).  
1-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction and Specifications  
SPECIFICATIONS  
1
MEDIUM AND FAST READING RATES........In medium rate, add 50 counts to number of counts. In fast  
reading rate, for sinewave inputs 1000 (4½ digit mode)  
counts and frequencies >100 Hz, the accuracy is ±(0.4%  
of reading +30 (4½ digit mode) counts).  
NONSINUSOIDAL INPUTS ...........................For nonsinusoidal inputs 10,000 counts with frequency  
components 100 kHz, add the following % of reading to  
the accuracy specifications  
FUNDAMENTAL  
FREQUENCY  
CREST FACTOR  
1.5 TO 2.0  
1.0 TO 1.5  
2.0 TO 3.0  
45 HZ to 5 kHz  
20 Hz to 45 Hz  
0.05  
0.2  
0.15  
0.7  
0.3  
1.5  
Operating Characteristics  
TEMPERATURE COEFFICIENT............Less than 0.1 x accuracy specification per °C to 18°C and  
28°C to 50°C.  
MAXIMUM INPUT...................................2A dc or rms ac. Protected with 2A, 250V fuse accessible at  
front panel, and interval 3A, 600V fuse.  
BURDEN VOLTAGE...............................1V dc or rms ac typical at full scale.  
RESISTANCE  
Input Characteristics  
RANGE  
FULL SCALE  
RESOLUTION  
CURRENT  
4½ DIGITS1  
5½ DIGITS  
THROUGH UNKNOWN  
5½ DIGITS  
2
20Ω  
19.999Ω  
199.999Ω  
0.1 mΩ  
1 mΩ  
10 mΩ  
100 mΩ  
1Ω  
1 mΩ  
10 mΩ  
100 mΩ  
1Ω  
1 mA  
1 mA  
200Ω  
2 kΩ  
1.99999 kΩ  
19.9999 kΩ  
199.999 kΩ  
1999.99 kΩ  
19.9999 MΩ  
1 mA  
20 kΩ  
100 µA  
10 µA  
5 µA  
200 kΩ  
2000 kΩ  
20 MΩ  
10Ω  
10Ω  
100Ω  
1 kΩ  
100Ω  
0.5 µA  
1. 4½ digits at the fastest reading rate.  
2. Four-wire ohms only.  
1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
Accuracy  
NORMAL (S) READING RATE...............±(% of Reading + Number of Counts)1.  
RANGE  
24 HOURS 23±1°C  
90 DAY 23±5°C  
1 YEAR 23±5°C  
2 YEARS 23±1°C  
3
20Ω  
0.007 + 304  
0.0040 + 35  
0.0025 + 2  
0.0025 + 2  
0.0025 + 2  
0.023 + 3  
0.009 + 404  
0.007 + 45  
0.005 + 3  
0.005 + 3  
0.006 + 3  
0.025 + 3  
0.040 + 4  
0.012 + 404  
0.010 + 45  
0.008 + 3  
0.008 + 3  
0.010 + 3  
0.027 + 3  
0.042 + 4  
0.015 + 404  
0.012 + 45  
0.010 + 3  
0.010 + 3  
0.012 + 3  
0.030 + 3  
0.050 + 4  
3
200Ω  
2 kΩ  
20 kΩ  
200 kΩ  
2000 kΩ  
20 MΩ  
0.023 + 3  
1. Within one hour of ohms zero, using offset control.  
2. Relative to calibration standards.  
3. Applies to 4-wire ohms only.  
4. When offset control is not used the number of counts are 50, 70, 90 and 90 for 24 hours,90 day, 1  
year, and 2 year respectively.  
5. When offset control is not used the number of counts are 5, 7, 9 and 9 for 24 hours, 90 day, 1 year,  
and 2 year respectively.  
MEDIUM AND FAST READING RATES........In medium rate, add 2 counts to the number of counts for  
the 200through 200 kranges, 3 counts for the 2000  
kand 20 Mranges, and 20 counts for the 20range.  
In fast reading rate, use 3 (4½ digit mode) counts for the  
number of counts for the 200range, 20 (4½ digit mode)  
counts for the 20range and 2 (4½ digit mode) counts for  
all other ranges.  
Operating Characteristics  
TEMPERATURE COEFFICIENT............Less than 0.1 x accuracy specification per °C from 0°C to 18°C  
and 28°C to 50°C.  
MEASUREMENT CONFIGURATION .....2-wire or 4-wire in all ranges accept 20range. Only 4-wire  
configuration is allowed in the 20range.  
OPEN CIRCUIT VOLTAGE ....................Less than 6.5V on the 20through the 200 kranges. Less  
than 13V on the 2000 kand 20 Mranges.  
INPUT PROTECTION.............................To 300V rms.  
1-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction and Specifications  
SPECIFICATIONS  
1
Reading Rates  
READING RATES WITH INTERNAL TRIGGER (readings per second)  
RATE  
POWER LINE FREQUECNCY1  
50 Hz  
2.08 (.26)2  
16.7 (1.04)2  
100  
60 Hz  
2.5 (.31)2  
400 Hz  
2.38 (.30)2  
19.0 (1.19)2  
100  
S
M
F
20 (1.25)2  
100  
1. Sensed automatically at power-up.  
2. In 20 mV, 20 ohm, and 200 mA DC ranges.  
AUTORANGING  
The 8842A autoranges up to the highest ranges in all funtions, down to the 200 mV range in the VDC  
and VAC funtions, and down to the 200 ranges in the ohms funtions. To select the 20 mV dc, 20, or  
200 mA dc range, press the respective range button (or send the respective range command, if using  
the IEEE-488 option).  
AUTOMATIC SETTLING TIME DELAY  
Time in milliseconds from single trigger to start of A/D conversion, Autorange off.  
FUNCTION  
RANGE  
READING RATE  
M
NUMBER OF COUNTS  
FROM FINAL VALUE1  
S
F
VDC  
20 mV  
200 mV  
2V-1000V  
All  
342  
342  
342  
551  
342  
342  
551  
395  
395  
322  
342  
141  
141  
1020  
342  
61  
9
9
30  
5
17  
9
9
VAC  
551  
342  
17  
551  
9
30 (Note 2)  
MA DC  
200 mA  
2000 mA  
2000 mA  
20Ω  
9
9
5
MA AC  
Ohms  
551  
395  
106  
17  
551  
17  
17  
13  
13  
21  
81  
723  
30 (Note 2)  
40  
5
200Ω  
2 kΩ  
5
20 kΩ  
17  
5
200 kΩ  
2000 kΩ  
20 MΩ  
121  
101  
964  
5
10  
10  
1. Difference between first reading and final value for an in-range step change coincident with trigger.  
For slow reading rate. 50 counts for medium rate; 10 counts for fast rate.  
1-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
EXTERNAL TRIGGER TIMING CHARACTERISTICS  
The following diagram shows the nominal timing for the various processes which take place between an  
external trigger and data sent out on the IEEE-488 interface. Delays will vary if a second trigger comes  
before the data handshake is complete.  
t1-1.wmf  
NOTES:  
1. Time for single trigger to start of A/D conversion.(See “Automatic Settling Time  
Delay” on previous page.) If the delay is disabled by using the T3 or T4 command,  
then the delay is 1 ms±150 µs. When the 8842A is triggered with an IEEE-488  
command (GET or ?), the automatic settling time delay begins after the trigger  
command has been processed and recognized.  
2. A/D conversion time is dependent on the reading rate and power-line frequency:  
RATE  
A/D CONVERSION TIME (ms)  
50 Hz  
60 Hz  
400 Hz  
S
M
F
472 (3800)*  
52 (960)*  
7
395 (3195)*  
45 (795)*  
7
414 (3300)*  
47 (840)*  
7
*In 20 mV DC, 20and 200 mA DC ranges.  
3. Sample complete is a 2.5 µs pulse which indicates that the analog input may be  
changed for the next reading.  
4. When talking to a fast controller.  
1-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction and Specifications  
SPECIFICATIONS  
1
GENERAL  
COMMON MODE VOLTAGE..................1000V dc or peak ac, or 700V rms ac from any input to earth.  
TEMPREATURE RANGE .......................0 to 50°C operating, -40 to 70°C storage.  
HUMIDITY RANGE.................................80% RH from 0 to 35°C, 70% to 50°C.  
WARMUP TIME......................................1 hour to rated specifications.  
POWER ..................................................100, 120, 220, or 240V ac ±10% (250V ac maximum), switch  
selectable at rear panel. 50, 60, or 400 Hz, automatically  
sensed at power-up. 20 VA maximum.  
VIBRATION.............................................Meets requirements of MIL-T- 28800C for Type III, Class 3,  
Style E equipment.  
PROTECTION ........................................ANSI C39.5 AND IEC 348, Class I.  
SIZE........................................................8.9 cm high, 21.6 cm wide, 37.1 cm deep(3.47 in high, 8.5 in  
wide, 14.6 in deep).  
WEIGHT..................................................Net, 3.4 kg (7.5 lb); shipping, 5.0 kg (11 lb).  
INCLUDED..............................................Line cord, test leads, Instruction/Service Manual, IEEE-488  
Quick Reference Guide, (Option –05 only), and instrument  
performance record.  
IEEE-488 INTERFACE FUNTION ..........Option allows complete control and data output capability, and  
supports the following interface funtion subsets: SH1,AH1, T5,  
L4, SR1, RL1, DC1, DT1, E1, PP0, AND C0.  
ELECTROMAGNETIC COMPATIBILITYSpecifications apply when used in an environment with fields  
strengths 1 V/m, (0.8 V/m for DC Current.) For fields  
strengths up to 3 V/m, multiply floor adder by 12 for VDC and  
Resistance and 200 for DC current. VAC and AC Current have  
no adders up to 3 V/m.  
f1-01.wmf  
Figure 1-1. External Dimensions  
1-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 2  
Operating Instructions  
Title  
Page  
2-1.  
2-2.  
2-3.  
2-4.  
2-5.  
2-6.  
2-7.  
2-8.  
INTRODUCTION................................................................................2-2  
Installing the Power-Line Fuse ........................................................2-2  
Connecting to Line Power................................................................2-2  
Adjusting the Handle........................................................................2-3  
Rack Mounting Kits.........................................................................2-3  
Power-Up Features...........................................................................2-4  
Front and Rear Panel Features .........................................................2-4  
Display Features...............................................................................2-8  
Error Messages.................................................................................2-8  
Diagnostic Self-Tests .......................................................................2-10  
Ranging ............................................................................................2-10  
Triggering.........................................................................................2-11  
Reading Rates and Noise Rejection .................................................2-11  
Automatic Settling Time Delay........................................................2-12  
External Trigger Input (Option -05 Only)........................................2-12  
Sample Complete Output (Option -05 Only) ...................................2-12  
2-9.  
2-10.  
2-11.  
2-12.  
2-13.  
2-14.  
2-15.  
2-16.  
2-17.  
2-18.  
2-19.  
2-20.  
2-21.  
2-22.  
2-23.  
2-25.  
2-26.  
2-27.  
2-28.  
2-29.  
Input Overload Protection Limits.....................................................2-13  
Measuring Voltage and Resistance ..................................................2-13  
Measuring Current ...........................................................................2-13  
Current Fuse Protection....................................................................2-13  
Offset Measurements .......................................................................2-14  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
2-1. INTRODUCTION  
This section provides instructions for installing and operating the 8842A. Refer to  
Section 4 for measurement considerations.  
2-2. INSTALLATION  
2-3.  
Installing the Power-Line Fuse  
WARNING  
FOR POWER-LINE VOLTAGES OF 198V TO 250V, THE POWER-  
LINE FUSE MUST BE REPLACED WITH A 1/8A, 250V SLO-BLO  
FUSE FOR FIRE PROTECTION. TO AVOID ELECTRIC SHOCK,  
REMOVE THE POWER CORD BEFORE REPLACING THE  
EXTERNAL LINE FUSE.  
The 8842A has a rear-panel power-line fuse in series with the power supply. A 1/4A,  
250V slow-blow fuse is installed in the factory for operation from 90V to 132V. For  
operation with power-line voltages of 198V to 250V, the fuse must be replaced with a  
1/8A, 250V slo-blow fuse.  
To replace the power-line fuse, first remove the power cord. Then turn the rear-panel fuse  
cover 1/4-turn counterclockwise with a screwdriver.  
For power-line voltages of 198V to 250V, use only a 1/4 x 1 1/4 (6.3mm x 32mm) fuse  
with at least a 100A breaking capacity.  
2-4.  
Connecting to Line Power  
WARNING  
TO AVOID SHOCK HAZARD, CONNECT THE INSTRUMENT  
POWER CORD TO A POWER RECEPTACLE WITH EARTH  
GROUND. TO AVOID INSTRUMENT DAMAGE, CHECK THAT  
THE REAR PANEL LINE VOLTAGE SELECTION SWITCHES ARE  
SET TO THE POWER-LINE VOLTAGE IN OUR AREA.  
The 8842A can be configured to accept line power of 100, 120, 220, or 240V ac (+/-10%,  
250V maximum) at 50, 60, or 400 Hz. The voltage must be selected by setting the rear  
panel LINE SET switches as shown in Figure 2-1. The 8842A automatically senses the  
power-line frequency at power-up, so that no adjustment for frequency is necessary.  
f2-01.wmf  
Figure 2-1. Line Voltage S election Settings  
2-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
INSTALLATION  
2
2-5.  
2-6.  
Adjusting the Handle  
The handle provides two viewing angles for bench-top use. To adjust its position, pull the  
ends out to a hard stop (about 1/4 inch on each side) and rotate it to one of the four stop  
positions shown in Figure 2-2. To remove the handle, adjust it to the vertical stop  
position and pull the ends all the way out.  
Rack Mounting Kits  
You can mount the 8842A in a standard 19-inch rack panel using the accessory rack  
mounting kits shown in Figure 2-3. To install the Single Rack Mount Kit, remove the  
handle and handle mounting plates, and attach the rack ears with the screws provided  
(Figure 2-4). The Dual Rack Mount Kit is installed similarly. (Both kits include mounting  
instructions.)  
The rear feet may be rotated 180 degrees to clear a narrow rack space.  
f2-02.wmf  
Figure 2-2. Adjusting the Handle  
f2-03.wmf  
Figure 2-3. Rack-Mount Kits  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f2-04.wmf  
Figure 2-4. Installing the Single Rack Mount Kit  
2-7. OPERATING FEATURES  
2-8.  
Power-Up Features  
When the 8842A is turned on, all display segments light up for about 2 seconds while the  
instrument performs an internal self-test of its digital circuitry. The 8842A then assumes  
the following configuration:  
VDC function  
Autorange, starting in the 1000V range  
Slow reading rate  
Continuous, internal trigger  
OFFSET off  
Local (front panel) control  
While all display segments are lit during the power-up self-test, you can freeze the  
display by pressing the SRQ button. All display segments will then remain lit until you  
press any button.  
2-9.  
Front and Rear Panel Features  
Front panel features are explained in Figure 2-5. Rear panel features are explained in  
Figure 2-6.  
The alternate functions embossed below the front panel range buttons and the special  
feature buttons are enabled by the CAL ENABLE switch. These functions are for use  
only when calibrating the instrument. See the Maintenance section for further  
explanation.  
CAUTION  
To avoid accidentally uncalibrating the 8842A, do not press the  
CAL ENABLE switch unless calibrating the instrument. Never  
cycle power on or off while the CAL ENABLE switch is on.  
2-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
OPERATING FEATURES  
2
Note that the VAC and mA AC functions are available only with the True RMS AC  
option. If this option is absent, pressing the VAC and mA AC function buttons causes the  
8842A to briefly display an error message (ERROR 30).  
FUNCTION BUTTONS:  
4-Wire Ohms  
DC Current  
AC Current*  
DC Volts  
AC Volts*  
CAL ENABLE switch  
enables calibration  
2-Wire Ohms  
HIGH and LO SENSE  
mode. (CAUTION! See text.)  
Terminals for 4-Wire Ohms Only  
Display  
HI and LO  
!
INPUT Terminals  
SPECIAL  
2A INPUT Terminal  
(Houses 2A fuse)  
FEATURES  
!
M
RANGE BUTTON:  
FRONT/REAR switch selects  
either front or rear inputs.  
200V, 200 k or  
200 mA dc  
20 mV or 20  
1000V dc, 700V ac,  
2 M or 2000 mA  
200 mV or 200  
Calibration Functions  
(embossed)  
POWER switch turns 8842A  
on or off. Also initiates  
power-up self-test and  
resets instrument to:  
20 M  
2V or 2 k  
20V or 20 k  
Autorange On/Off **  
VDC function  
Autorange  
Slow reading rate  
Continuous trigger  
OFFSET off  
Local (front panel) control  
** For a description of the autorange feature,  
see paragraph 2-14.  
*Available with 8842A-09 True RMS AC only.  
Gives error message if option not installed.  
f2-05_1.wmf  
Figure 2-5. Front Panel Features  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
TRIG triggers a new reading.  
EXT TRIG toggles between internal  
and external trigger modes  
Enabled in external trigger mode.  
RATE cycles between slow, medium and  
fast reading rates. Automatically  
selects the optimum filter for each  
reading rate.  
SRQ generates a service request over the IEEE-  
488 bus if enabled by the SRQ mask (IEEE-488  
Interface option only). When pressed for 3  
seconds, SRQ initiates diagnostic self-tests.  
NOTE: Leave inputs disconnected during self-  
tests or the 8842A may indicate errors.  
M
OFFSET stores the displayed reading as an  
offset, which the 8842A subtracts from all  
subsequent readings in the function  
If the 8842A is in remote, LOCAL returns it to  
local control. If the 8842A is in local, the  
LOCAL button causes the 8842A to display its  
bus address for two and one half seconds.  
Ignored if the IEE-488 Interface is not  
installed.  
presently selected. Readings in the other  
functions remain unaffected. Pressing  
OFFSET again cancels the offset, or stores  
a new offset if in a different function.  
Error Condition  
Reading Rate Slow, Medium  
Overrange  
Calibration Mode Enabled  
Autorange On  
Offset On  
and Fast. Blinks off when a  
reading is triggered.  
External Trigger  
Mode Enabled  
Function  
Self-Test  
Enabled  
and Units  
Annunciators  
IEEE-488 Interface  
Annunciators  
f2-05_2.wmf  
Figure 2-5. Front Panel Features (cont)  
2-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
OPERATING FEATURES  
2
SAMPLE COMPLETE output  
TTL-level. Normally high,  
pulsed low when samples for  
a reading are completed*  
IEEE-488 Interface  
Bus Connector*  
EX TRIG input*  
TTL-level,  
falling-edge triggered  
(internally selectable)  
HIGH and LO SENSE  
Terminals for 4-Wire  
Ohms Only  
TALK ONLY and IEEE-488  
ADDRESS Selection Switches*  
ADDRESS  
A5 A3 A1  
TTL LEVELS  
IEEE STD-488 PORT  
IN  
OUT  
SHELL NOT  
GROUNDED  
15V MAX  
INPUT  
SENSE  
4 WIRE  
TALK  
ONLY  
EXT TRIG  
10V MAX  
SAMPLE  
COMPLETE  
SH1, AH1, T5, L4, SR1, RL1,  
DC1, DT1, PP0, C0, E1  
HI  
240V  
220V  
120V  
100V  
JOHN FLUKE MFG. CO., INC.  
EVERETT, WA MADE IN U.S.A.  
IEEE-05  
AC-09  
LINE  
SET  
1000V  
700V  
MAX  
300V MAX  
PATENTS PENDING  
!
LINE FUSE  
250V  
!
CAUTION  
WARNING  
GROUNDING CONDUCTOR  
IN POWER CORD MUST BE  
CONNECTED TO ENSURE  
PROTECTION FROM  
FOR FIRE PREVENTION  
REPLACE ONLY WITH  
LO  
1/4 A SLOW FUSE, 100/120v  
1/8 A SLOW FUSE, 220/240V  
ALL  
INPUTS  
20 VA 50/60/400 Hz  
1000V  
700V  
ELECTRICAL SHOCK  
MAX  
!
REMOVE GROUNDING SCREW  
BEFORE REMOVING COVER  
HI and LO INPUT  
Terminals  
Power-Line Cord  
Connector  
Line Voltage  
Selection Switches  
Rear feet rotate  
for rack mounting  
Serial Number  
Power-Line Fuse  
*Available with IEEE-488 Interface only.  
Otherwise, the upper portion of the rear panel is  
covered with an insert as shown at right.  
CAUTION: The rear panel insert is attached from  
inside the case. Refer to Section 8 for instructions  
on removing it.  
Rear Panel Insert  
INPUT  
SENSE  
4 WIRE  
HI  
240V  
220V  
120V  
100V  
JOHN FLUKE MFG. CO., INC.  
EVERETT, WA MADE IN U.S.A.  
IEEE-05  
AC-09  
LINE  
SET  
1000V  
700V  
MAX  
300V MAX  
PATENTS PENDING  
!
LINE FUSE  
250V  
!
CAUTION  
WARNING  
GROUNDING CONDUCTOR  
IN POWER CORD MUST BE  
CONNECTED TO ENSURE  
PROTECTION FROM  
FOR FIRE PREVENTION  
REPLACE ONLY WITH  
LO  
1/4 A SLOW FUSE, 100/120v  
1/8 A SLOW FUSE, 220/240V  
ALL  
INPUTS  
20 VA 50/60/400 Hz  
BEFORE REMOVING COVER  
1000V  
700V  
ELECTRICAL SHOCK  
MAX  
!
REMOVE GROUNDING SCREW  
f2-06.wmf  
Figure 2-6. Rear Panel Features  
2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
2-10. Display Features  
The 8842A features a vacuum fluorescent display with a numeric field and annunciators.  
The annunciators are explained in Figure 2-5.  
2-11. Error Messages  
If the 8842A detects an operator error or an internal failure, it displays an error message  
for about 2-1/2 seconds and then resumes normal operation. During this time, the front  
panel buttons are ignored. The error message consists of the ERROR annunciator and a  
two-digit error code. (See Figure 2-7.) Error codes are explained in Table 2-1.  
If the FRONT/REAR switch is set to the REAR position while the mA DC or mA AC  
function is selected, ERROR 31 is displayed. In this case the error message is displayed  
until you return the switch to the FRONT position or select another function.  
f2-07.wmf  
Figure 2-7. Typical Error Messages  
2-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
OPERATING FEATURES  
2
Table 2-1. Error Codes  
ERROR CODE  
ERROR  
CODE  
MEANING  
MEANING  
ANALOG SELF-TEST ERRORS  
200 VAC, Zero  
OPERATION ERRORS  
1
2
3
4
30 AC funtions availible only with 8842A-09 True  
RMS AC option.  
700 VAC, Zero  
mA AC, Zero  
mA DC, Zero  
31 mA AC or mA DC funtion selected while  
REAR inputs selected.  
32 OFFSET selected with reading unavailible for  
overrange.  
40 Computed calibration constant out of  
range.(Previous cal may be wrong or there  
may be a hardware problem.)  
5
200 VDC, Zero  
41 Calibration input out of acceptable range.  
Check that input is correct. (Previous cal may  
be wrong or there may be a hardware  
problem.)  
6
7
8
1000 VDC, Zero  
1000 VDC, Zero  
20 VDC + 20M  
42 Calibration memory write error. (Probably a  
hardware problem.)  
50 Guard crossing error detected by In-Guard  
µC.  
51 Calibration command not valid unless  
calibration mode is enabled.  
9
20 VDC + 2000 kΩ  
2 VDC + 2000 kΩ  
52 Command not valid at this time.  
10  
53 Invalid calibration value in Put command.  
(Example: Sending a negative value during  
ac calibration.)  
11  
12  
200, Overrange  
2 k, Overrange  
54 Command not valid in calibration verification.  
56 Variable inputs not allowed during A/D  
calibration.Use prompted value.  
13  
14  
15  
16  
17  
20 k, Overrange  
60 Device-dependent commands not valid  
during self-tests.  
200 k, Overrange  
1000 VDC + X10 T/H + 20 MΩ  
200 VDC + 200 kΩ  
71 Syntax error in device-dependent command  
string.  
72 Guard crossing error detected by out Guard  
µC.  
73 Guard crossing error detected at power on or  
CAL ENABLE switch on at power on.  
200 VDC + 20 kΩ  
77 IEEE-488 Interface self-test error.  
DIGITAL SELF-TEST ERRORS  
25 In-Guard µC Internal RAM  
26 Display RAM  
27 In-Guard µC Internal Program Memory  
28 External Program Memory  
29 Calibration Memory  
NOTE: See the Maintenance section for a detailed description of self-tests.  
2-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
2-12. Overrange Indication  
An input is overrange if it exceeds the full scale of the selected range. In most ranges, the  
8842A indicates an input is overrange by lighting the OVER annunciator and showing a  
"1" on the display. (See Figure 2-8.) The sign, the position of the decimal point, and the  
other annunciators are not affected.  
As a safety feature, the 8842A treats the 1000V dc and 700V ac rangesdifferently. In  
these ranges, the 8842A indicates when the input exceeds the input overload limit of  
1000V dc or 700V ac, respectively, by lighting the OVER annunciator and flashing the  
display. Readings are still displayed.  
2-13. Diagnostic Self-Tests  
The 8842A features diagnostic self-tests which check both the digital and analog circuitry  
in the instrument. The self-tests consist of 21 analog tests followed by the in-guard  
program memory, calibration memory, and display self-tests. To initiate the self-tests,  
press the SRQ button for 3 seconds. The instrument can be stopped in any of the test  
configurations by pressing the SRQ button while the test number is displayed. Press any  
button to continue the tests.  
During the test, the TEST annunciator lights, and the numeric field displays the number  
of each analog test as it is performed. Then all display segments light up while the  
instrument performs the in-guard program memory, calibration memory, and display self-  
tests. The 8842A then returns to the power-up configuration. The self-tests are described  
in greater detail in the Maintenance section.  
NOTE  
The inputs must be left disconnected while the self-tests are performed or  
the 8842A may indicate that errors are present.  
If the 8842A detects an error, it displays an error message for about 2-1/2 seconds. (Error  
codes 01 through 29 correspond to the self-tests.) If self-test errors are displayed even  
when the input terminals are disconnected, there may be a hardware problem in your  
8842A. In that event, refer to the Maintenance section or contact your local Fluke  
representative.  
2-14. Ranging  
Measurement ranges can be selected using either autorange (by pressing the AUTO  
button) or manual range (by pressing another range button). The 8842A displays explicit  
units in every range, so that the display may be read directly.  
2-15. AUTORANGE  
In autorange, the 8842A goes to a higher range when the input exceeds full scale (199999  
counts), and goes to a lower range when the input falls below 9% of full scale (18000  
counts). While the instrument changes range, the numeric field on the display is blanked  
until a new reading is completed. However, the decimal point and units annunciators  
always indicate what range the instrument is in.  
Pressing the AUTO button when the instrument is already in autorange toggles the  
8842A from autorange to manual range. This causes the instrument to remain locked in  
the present range.  
The 8842A autoranges up to the highest ranges in all functions, down to the 200 mV  
range in the VDC and VAC functions, and down to the 200range in the ohms  
functions. To select the 20 mV dc, 20, or 200 mA dc range, press the respective range  
button (or send the respective range command, if using the IEEE-488 option).  
2-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
OPERATING FEATURES  
2
2-16. MANUAL RANGE  
In manual range, the 8842A remains fixed in the selected range until you select another  
range or press AUTO. If you select a range which is not valid for the present function, or  
select a function which is not valid for the present range, the 8842A selects the nearest  
valid range. For example, if the 8842A is in the VDC function and you press the 20 MΩ  
button, the 8842A selects the 1000V range.  
The range buttons have no effect in the mA AC functions, since all measurements in  
these functions are made in the 2000 mA range.  
2-17. Triggering  
Triggering causes the 8842A to execute a measurement cycle and display the result.  
During each measurement cycle, the instrument samples the input a number of times and  
then averages the samples to compute a reading. The number of samples averaged for  
each reading depends upon the reading rate.  
Each time a reading is triggered, the rate annunciator (S, M, or F) blinks off. In the fast  
reading rate, the F annunciator flashes so rapidly it appears to be almost constant.  
How the 8842A is triggered depends on whether the continuous trigger mode or external  
trigger mode is selected. Pressing the EX TRIG (external trigger) button toggles the  
8842A between the two modes.  
2-18. CONTINUOUS TRIGGER MODE  
In the continuous trigger mode, readings are triggered by a continuous, internal trigger.  
The rate of the trigger is set by the RATE button.  
2-19. EXTERNAL TRIGGER MODE  
In the external trigger mode, readings are triggered by pressing the TRIG button. If the  
IEEE-488 Interface option is installed, readings can also be triggered by remote  
commands or by using the rear panel external trigger (EXT TRIG) connector. (See the  
Options and Accessories section.)  
In the external trigger mode, pressing any front panel button blanks the numeric field on  
the display until a new measurement is triggered. This ensures that all readings  
correspond to the instrument configuration indicated by the display annunciators. The  
blanking also occurs in the continuous trigger mode, but usually isnt noticed because  
new measurements are triggered automatically.  
The TRIG button does not trigger readings in the continuous trigger mode.However, it  
does blank the last reading to acknowledge a button was pressed.  
2-20. Reading Rates and Noise Rejection  
The RATE button allows you to optimize either measurement speed or noise rejection.  
The 8842A uses both analog and digital filtering to allow measurements in the presence  
of unwanted environmental noise (especially line-related noise). However, since filtering  
introduces a delay in response to a change in the input signal, there is an inherent trade-  
off between noise rejection and measurement speed.  
The instrument has three reading rates: slow (S) and medium (M), with a 5-1/2 digit  
display, and fast (F), with a 4-1/2 digit display. To provide optimum combinations of  
measurement speed and noise rejection, the RATE button allows control of both the  
internal trigger rate and the degree of filtering. The same degree of filtering is used in  
both the continuous and external trigger modes. In the 20 mV, 20, and 200 mA dc  
ranges, use of slow (S) filter provides maximum noise rejection.  
2-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
In the continuous trigger mode, the actual number of readings displayed per second for  
each reading rate is determined by the line-power frequency. At power-up, the 8842A  
senses the line-power frequency and adjusts the analog-to-digital converter timing  
characteristics for optimum normal-mode noise rejection. The resulting reading rates are  
shown in the specifications in Section 1.  
2-21. Automatic Settling Time Delay  
When the external trigger mode is selected, the 8842A automatically inserts a delay after  
receiving a trigger signal, but before starting the first input sample. The delay is just long  
enough so that the reading will be correct (within a specified number of counts of the  
final value) even if the trigger signal occurs as the input makes a step change between  
zero and full scale (10,000 counts and full scale in the ac functions).The length of the  
delay depends on the range, function, and reading rate, as shown in the specifications in  
Section 1. The delay is enabled only in the external trigger mode. It can be turned off  
with a remote command over the IEEE-488 interface bus to accommodate special timing  
considerations.  
f2-08.wmf  
Figure 2-8. Overrange Indication  
2-22. External Trigger Input (Option -05 Only)  
The rear panel EXT TRIG input is a TTL-level input which can be used to trigger  
measurements when the 8842A is in the external trigger mode. A measurement is  
triggered on the falling edge of the input. Since the EXT TRIG input is pulled high  
internally, it can also be controlled by a normally open switch. A measurement is  
triggered when the switch is closed. For special applications using the IEEE-488  
Interface, the automatic setting time delay can be disabled using remote commands. (See  
Section 3.) Refer to Section 1 for timing details.  
The polarity of the EXT TRIG input can be reversed by changing internal jumpers. Refer  
to the Maintenance section for instructions.  
2-23. Sample Complete Output (Option -05 Only)  
The SAMPLE COMPLETE output indicates when analog input sampling for a reading is  
completed. The output is a TTL-level signal which is pulsed low for approximately 2.5  
µs when the input-sampling portion of the A/D conversion is completed. The signal is  
useful for interfacing with other equipment when the 8842A is used in external trigger  
mode in an instrumentation system. For example, the SAMPLE COMPLETE output  
could be used to advance a scanner to the next channel.  
2-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
MAKING MEASUREMENTS  
2
2-24.MAKING MEASUREMENTS  
2-25. Input Overload Protection Limits  
WARNING  
TO AVOID SHOCK HAZARD AND/OR INSTRUMENT DAMAGE,  
DO NOTAPPLY INPUT POTENTIALS THAT EXCEED THE INPUT  
OVERLOAD LIMITS SHOWN IN TABLE 2-2.  
The 8842A is protected against input overloads up to the limits shown in Table 2-2.  
Exceeding these limits may damage the instrument and/or pose a shock hazard.  
Table 2-2. Input Overload Limits  
FUNTION  
CONNECTORS  
INPUT HI and LO:  
MAXIMUM INPUT  
1000 dc  
VDC  
MA DC  
2A INPUTand INPUT LO:  
INPUT HI and LO:  
2000mA  
300V rms  
300V rms  
2 WIRE/4 WIRE kΩ  
SENSE HI and LO:  
INPUT HI and LO:  
VAC  
700V rms, 1000V peak, or 2 x 107  
V-Hz (whichever is less)  
MA AC  
2A INPUT and INPUT LO:  
Any terminal to earth:  
2000 mA rms  
All Funtions  
1000V dc or peak ac  
2-26. Measuring Voltage and Resistance  
To measure voltage or resistance, select the desired function and connect the test leads as  
shown in Figure 2-9. Resistance can be measured in either the 2-wire or 4-wire  
configuration.  
2-27. Measuring Current  
To measure current, select the desired function and connect the test leads as follows:  
1. Turn off power in the circuit to be measured (Figure 2-10).  
2. Break the circuit (preferably on the ground side to minimize the common mode  
voltage), and place the 8842A in series at that point.  
3. Turn on power in the circuit, and read the display.  
4. Turn off power in the circuit, and disconnect the 8842A.  
2-28. Current Fuse Protection  
The 2A input terminal is protected from overloads by a 2A, 250V fuse which is  
accessible from the front panel, and by an internal 3A, 600V fuse. If either fuse blows,  
the 8842A will respond as though the input were zero.  
WARNING  
TO AVOID ELECTRIC SHOCK, REMOVE THE TEST LEADS  
BEFOREREPLACING THE FRONT PANEL FUSE.  
2-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
To replace the front panel fuse, first remove the test leads. Then press in the lip of the 2A  
input terminal slightly and rotate it 1/4-turn counterclockwise. Spring tension will force  
the fuse and fuse holder out of the front panel. The internal 3A fuse should be replaced  
only by qualified service personnel.  
2-29. Offset Measurements  
WARNING  
WHEN THE OFFSET FEATURE IS IN USE, DISPLAYED  
READINGS ARE RELATIVE AND MAY NOT INDICATE THE  
PRESENCE OF DANGEROUS POTENTIALS AT THE INPUT  
CONNECTORS OR TEST LEADS. USE CAUTION TO AVOID  
ELECTRIC SHOCK OR INSTRUMENT DAMAGE.  
The OFFSET feature allows you to store a reading as a relative reference value. When the  
OFFSET button is pressed, the 8842A stores the present reading and displays subsequent  
measurements as the difference between the measured value and the stored reading. The  
OFFSET annunciator is lit whenever an offset is in use.  
The OFFSET feature may be used in all functions. Since the display represents a numeric  
difference, it always has a sign, even in the resistance and ac functions.  
The offset can be canceled by pressing the OFFSET button again, in which case the  
OFFSET annunciator disappears from the display. The offset can also be canceled by  
storing an offset in another function. If a reading is overrange or unavailable when the  
OFFSET button is pressed, the 8842A indicates ERROR 32 and does not store the offset.  
If you change functions while an offset is stored, the OFFSET annunciator disappears and  
the offset temporarily disappears. However, when you return to the original function, the  
offset is restored (and the OFFSET annunciator reappears) unless a new offset was  
established in another function. Note that the input overload limits are not changed by the  
use of the offset feature. However, the display flashes if the 8842A is in the 1000V dc or  
700V ac ranges and the input exceeds 1000V dc or 700V ac, respectively.  
While an offset is enabled, the 8842A indicates an overrange condition if either of the  
following conditions occur:  
The input signal is overrange  
The calculated reading is overrange  
For example, suppose the instrument is in the 20V range of the VDC function and you  
store an offset of +15V. The maximum positive voltage reading that can be displayed  
without overranging is +4.9999V, which is actually a +19.9999V input signal. The  
maximum negative voltage reading that can be displayed without overranging is -  
19.9999V, which is actually a -4.9999V input signal. You can measure a greater range of  
voltages by selecting a higher range.  
When in autorange, the 8842A selects the range appropriate for the input signal,  
regardless of any stored offset. If, for example, a +10V offset is stored, and a +1V input  
is applied, the 8842A will autorange to the 2V range and display an overrange condition  
since it cannot display -9V on the 2V range. Manual range control could be used to lock  
the 8842A into the 20V range in this case.  
Applications of the offset feature include correcting for test lead resistance in 2-wire  
resistance measurements, nulling offset currents or voltages, measuring voltage  
deviations, and matching resistors.  
2-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
EXTERNAL CLEANING  
2
f2-09.wmf  
Figure 2-9. Measuring Voltage and Resistance  
f2-10.wmf  
Figure 2-10. Measuring Current  
2-30.EXTERNAL CLEANING  
WARNING  
TO AVOID ELECTRIC SHOCK OR INSTRUMENT DAMAGE,  
NEVER GET WATER INSIDE THE CASE. TO AVOID  
INSTRUMENT DAMAGE, NEVER APPLY SOLVENTS TO THE  
INSTRUMENT.  
Should the 8842A case require cleaning, wipe the instrument with a cloth that is lightly  
dampened with water or a mild detergent solution.  
2-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
2-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3  
Remote Programming  
Title  
Page  
3-1.  
3-2.  
3-3.  
3-4.  
3-5.  
3-6.  
3-7.  
3-8.  
INTRODUCTION................................................................................3-3  
CAPABILITIES ...................................................................................3-3  
AN OVERVIEW OF REMOTE OPERATION...................................3-4  
DEVICE-DEPENDENT COMMAND SET ........................................3-6  
Bn (Offset Commands) ....................................................................3-9  
Dn (Display Commands)..................................................................3-9  
Fn (Function Commands) ................................................................3-10  
G0 (Get Instrument Configuration)..................................................3-12  
G2 (Get Calibration Prompt)............................................................3-12  
G3 (Get User-Defined Message)......................................................3-13  
G4 (Get Calibration Status)..............................................................3-13  
G5 (Get IAB Status).........................................................................3-14  
G6 (Get YW Status).........................................................................3-14  
G7 (Get Error Status) .......................................................................3-14  
G8 (Get Instrument Identification)...................................................3-15  
N (Numeric Entry Command)..........................................................3-15  
P1 (Put SRQ Mask)..........................................................................3-16  
P2 (Put Calibration Value)..............................................................3-16  
P3 (Put User-Defined Message)......................................................3-17  
Sn (Reading Rate Commands).........................................................3-17  
Tn (Trigger Mode Commands) ........................................................3-18  
Wn (Terminator Commands) ...........................................................3-19  
X0 (Clear Error Register Command) ...............................................3-19  
Yn (Suffix Commands)....................................................................3-19  
Z0 (Self-Test Command) .................................................................3-19  
(Device-Clear Command) ................................................................3-20  
3-9.  
3-10.  
3-11.  
3-12.  
3-13.  
3-14.  
3-15.  
3-16.  
3-17.  
3-18.  
3-19.  
3-20.  
3-21.  
3-22.  
3-23.  
3-24.  
3-25.  
3-26.  
3-27.  
3-28.  
3-29.  
3-30.  
3-31.  
3-32.  
3-33.  
3-34.  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3-35.  
? (Single-Trigger Command) ...........................................................3-20  
3-37.  
3-38.  
3-39.  
Definitions........................................................................................3-21  
Syntax Rules ....................................................................................3-23  
3-41.  
3-42.  
3-43.  
3-44.  
3-45.  
3-46.  
3-47.  
3-48.  
Loading Output Data........................................................................3-24  
Types of Output Data.......................................................................3-25  
Numeric Data and Error Messages...................................................3-25  
Status Data .......................................................................................3-26  
Output Priority .................................................................................3-26  
3-50.  
3-51.  
The Serial Poll Register ...................................................................3-27  
3-53.  
3-54.  
3-55.  
Address Messages............................................................................3-29  
Universal Commands.......................................................................3-29  
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
INTRODUCTION  
3
NOTE  
This section contains programming instructions for use with the IEEE-488  
Interface (Option -05). For installation instructions, refer to the Options  
and Accessories section.  
3-1. INTRODUCTION  
The IEEE-488 Interface turns the 8842A into a fully programmable instrument for use  
with the IEEE Standard 488-1978 interface bus (IEEE-488 bus). With the IEEE-488  
Interface, the 8842A can become part of an automated instrumentation system. The  
8842A can be under complete, interactive control from a remote bus controller; or it can  
be set to the talk-only mode, connected to a data logger or printer, and dedicated to a  
single task.  
This manual assumes you know the basics of the IEEE-488 interface bus. For an  
introduction to the bus, request Fluke Application Bulletin AB-36, "IEEE Standard 488-  
1978 Digital Interface for Programmable Instrumentation."  
3-2. CAPABILITIES  
The IEEE-488 Interface provides remote control of all front panel controls except for the  
POWER, CAL ENABLE, and FRONT/REAR switches. Other features include:  
A simple and predictable command set  
Fast measurement throughput  
Full talk/listen capability, including talk-only operation  
Full serial poll capability, with bit-maskable SRQ  
Full remote/local capability, including local lockout  
EXTERNAL TRIGGER and SAMPLE COMPLETE connectors  
Remote calibration  
Programmable trigger sources, including two bus triggers  
Informative output suffix (suppressible)  
Selectable output terminators  
The 8842A supports the following interface function subsets: SH1, AH1, T5, L4, SR1,  
RL1, DC1, DT1, E1, PP0, and C0.  
3-3. BUS SET-UP PROCEDURE  
To set up the 8842A on the IEEE-488 bus, proceed as follows:  
1. Turn the 8842A POWER switch OFF and set the 8842A IEEE-488 address using the  
rear panel IEEE-488 address switches shown in Figure 3-1.  
2. With the 8842A POWER switch OFF, plug the IEEE-488 cable into the 8842A rear  
panel IEEE-488 connector.  
3. Switch on the 8842A.  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
A
D
D
R
E
S
S
T
A
L
K
A5  
A4 A3 A2 A1  
A
D
D
R
E
S
S
T
A
L
K
A5 A4 A3 A2 A1  
A
D
D
R
E
S
S
T
A
L
A5 A4 A3 A2 A1  
K
O
N
L
O
N
L
O
N
L
Y
Y
Y
00  
01  
02  
03  
04  
05  
06  
07  
08  
09  
10  
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
0
0
0
0
1
1
1
1
0
0
0
0
0
1
1
0
0
1
1
0
0
1
0
1
0
1
0
1
0
1
0
1
0
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
1
1
1
0
0
0
0
0
0
0
1
1
1
1
0
0
0
0
1
1
1
0
0
1
1
0
0
1
1
0
0
1
0
1
0
1
0
1
0
1
0
1
22  
23  
24  
25  
26  
27  
28  
29  
30  
31  
0
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
1
0
0
1
1
1
1
1
1
1
1
1
0
0
0
0
1
1
1
1
1
0
0
1
1
0
0
1
0
1
0
1
0
1
0
1
0
Not allowed  
TALK  
ONLY  
1
X
X
X
X
X
X = setting does not matter  
f3-01.wmf  
Figure 3-1. IEEE-488 Address Selection  
Whenever the 8842A is in the local state, the IEEE-488 address can be displayed on the  
front panel by pressing the LOCAL button.  
3-4. AN OVERVIEW OF REMOTE OPERATION  
An overview of remote operation is presented in the block diagram in Figure 3-2. Each  
block represents a register, buffer, etc., contained in the 8842A. The status registers in the  
center column indicate the instruments status, including its function, range, reading rate,  
etc. The input buffer receives data from the IEEE-488 bus. The output buffer receives  
data from the blocks to its left, and sends data on to the IEEE-488 bus.  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
AN OVERVIEW OF REMOTE OPERATION  
3
f3-02.wmf  
Figure 3-2. Remote Operation Block Diagram  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Information is transferred between blocks by device-dependent commands. Each  
command is shown next to an arrowhead which indicates the resulting information  
transfer. For example, Put command P0 takes a number from the input buffer and stores it  
in the primary status registers. Likewise, Get command G0 gets the content of the  
primary status registers and copies it into the output buffer.  
3-5. A NOTE ABOUT EXAMPLES  
In the examples in this manual, device-dependent commands are shown enclosed within  
quotation marks, as they would be entered in Fluke BASIC. For clarity, the commands  
are also separated by spaces. However, the spaces are are not necessary and may be  
omitted.  
Example  
Explanation  
"* F3 R1 S1 T2"  
This example is equivalent to "*F3R1S1T2" or  
"*,F3,R1,S1,T2".  
Using the Fluke 1722A Instrument Controller, these commands might be written into a  
BASIC program as shown in Figure 3-3. Examples using other controllers are given at  
the end of this section.  
f3-03.wmf  
Figure 3-3. Typical Command String  
Examples of 8842A output data show the terminators CR and LF. The terminator EOI is  
not shown because it is a uniline message. However, the terminators CR, LF, and EOI are  
all selectable using the Write commands.  
For reference, the ASCII and IEEE Std 488-1978 bus codes are shown at the back of this  
section.  
3-6. DEVICE-DEPENDENT COMMAND SET  
Device-dependent commands are the heart of 8842A remote control. They tell the 8842A  
how and when to make measurements, when to put data on the bus, when to make service  
requests, etc. Commands which correspond directly to the front panel controls or display  
are shown in Figure 3-4. The complete set of device-dependent commands is listed in  
Figure 3-5. The commands may be entered using either upper- or lower-case letters. See  
Table 6-15 for conditions under which certain commands are not valid.  
3-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
TRIGGER COMMANDS  
?
Trigger Measurement  
GET Trigger and Execute  
TRIGGER MODE COMMANDS  
T0  
T1-T4  
Continuous Trigger  
External Trigger  
DISPLAY COMMANDS  
READING RATE  
COMMANDS  
SUFFIX COMMANDS  
D0 Normal Display  
D1 Blank Display  
Y0 Disable Suffix  
Y1 Enable Suffix  
S0 Slow  
S1 Medium  
S2 Fast  
M
RANGE COMMANDS  
SELF-TEST  
Sensed  
by G5  
R0 Autorange On  
R1 200 mV, 200  
R2 2V, 2 kΩ  
COMMANDS  
Z0 Begins Self-Tests  
R3 20V, 20 kΩ  
FUNCTION COMMANDS  
R4 200V, 200 k, 200 mA  
R5 1000V dc, 700V ac,  
2 M, 2000 mA  
R6 20 MΩ  
R7 Autorange Off  
R8 20 mV, 20Ω  
GTL Go To Local  
F1 VDC  
F2 VAC  
F3 2 Wire kΩ  
F4 4 Wire kΩ  
F5 mA DC  
F6 mA AC  
DEVICE-CLEAR  
COMMANDS  
*Reset 8842A to  
power-up state  
OFFSET COMMANDS  
B0 Offset Off  
B1 Offset On  
f3-04.wmf  
Figure 3-4. Commands Which Correspond to the Front Panel  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
FUNCTON COMMANDS  
TERMINATOR COMMANDS  
F1  
F2  
F3  
F4  
F5  
F6  
VDC (Default)  
VAC  
W0  
W1  
W2  
W3  
W4  
W5  
W7  
Enable CR LF EOI (Default)  
Enable CR LF Only  
Enable CR EOI Only  
Enable CR Only  
2 WIRE k  
4 WIRE kΩ  
MA DC  
Enable LF EOI Only  
Enable LF Only  
MA AC  
RANGE COMMANDS  
Disable All Output Terminators  
R0  
R2  
Autorange  
CLEAR COMMANDS  
2V, 2 kΩ  
*
Device Clear (Resets 8842A to default  
conditions)  
R3  
R5  
R6  
R8  
20V, 20 kΩ  
X0  
Clear Error Register  
1000V dc, 700V ac, 2 M, 2000 mA  
SINGLE-TRIGGER COMMAND  
Trigger Measurement  
GET COMMANDS  
20 mΩ  
?
20mV, 20Ω  
READING RATE COMMANDS  
G0  
Get Instrument Configuration (F,R,S, and  
T)  
S0  
S1  
S2  
Slow (Default)  
Medium  
G1  
G2  
G3  
G4  
G5  
Get SRQ Mask  
Get Calibration Input Prompt  
Get User-Defined Message  
Get Calibration Status  
Fast  
TRIGGER MODE COMMANDS  
COMMAND  
TRIGGER  
MODE  
REAR PANEL  
TRIGGER  
SETTLING  
DELAY  
Get IAB Status (input F/R, Autorange  
On/Off, Offset On/Off)  
TO (Default)  
Internal  
Disabled  
-
G6  
Get YW Status (Suffix  
Enabled/Disabled, Terminator  
Selection)  
T1  
T2  
T3  
T4  
External  
External  
External  
External  
Enabled  
Disabled  
Enabled  
Disabled  
On  
On  
Off  
Off  
G7  
Get Error Status  
G8  
Get Instrument Identification  
G2 valid only in calibration mode.  
Note:  
PUT COMMANDS  
Note: Delay is enabled by entering EX TRIG mode while in local.  
P0  
Put Instrument Configuration (F,R,S,  
and T)  
OFFSET COMMANDS  
P1  
Put SRQ Mask  
B0  
P2  
Put Variable Calibration Value  
Put User-Defined Message  
B1  
P3  
DISPLAY COMMANDS  
Note:  
P2 and P3 valid only in calibration  
mode.  
D0  
D1  
Offset Off (Default)  
Offset On  
PUT COMMAND FORMAT  
N
<value> P0  
SUFFIX COMMANDS  
N
<value> P1  
Y0  
Normal Display (Default)  
Blank Display  
N
<value> P2  
Y1  
P3  
< 16 ASCII characters>  
Figure 3-5. Device-Dependent Command Set  
3-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
Device-dependent commands are device-dependent messages. For the 8842A to receive  
them, they must be sent over the IEEE-488 bus when the 8842A is in remote and has  
been addressed as a listener.  
The following paragraphs describe the device-dependent commands in alphabetical order.  
Special characters (* and ?) are described last.  
3-7.  
3-8.  
Bn (Offset Commands)  
The Offset commands duplicate the function of the front panel OFFSET button. When  
the 8842A receives the B1 command, the 8842A stores the present reading as an offset  
for the present function. The B0 command cancels the offset. As with front panel  
operation, only one offset is allowed at a time.  
The offset status (not the offset value) can be read using the G5 command. The 8842A  
defaults to B0 on both power-up and on any device-clear command (*, DCL, or SDC).  
Cn (Calibration Commands)  
CAUTION  
The command string "C3 C0" erases the entire calibration  
memory. A complete calibration must then be performed.  
The Calibration commands allow the 8842A to be calibrated under remote control.  
Commands C0, C1, and C2 duplicate the front panel calibration functions STORE, A/D,  
and HF AC, respectively. For a complete description of remote calibration, see the  
Maintenance section of this manual.  
For the 8842A to accept these commands, the 8842A must be in the calibration mode  
(enabled by pressing the front panel CAL ENABLE switch). Otherwise, the commands  
generate an error message.  
3-9.  
Dn (Display Commands)  
The Display commands allow the user to blank the numeric field in the 8842A front  
panel display. The D0 command causes the display to operate normally, and is the default  
on power-up and upon any device-clear command (*, DCL, or SDC).  
The D1 command blanks the numeric field in the display. The annunciators remain  
active, and all of the annunciators still flash if the input exceeds 1000V dc or 700V ac in  
the respective ranges. The D1 command is used for best performance when high IEEE-  
488 Interface Data rates are required.  
3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3-10. Fn (Function Commands)  
The function commands duplicate the front panel function buttons. The 8842A defaults to  
F1 on power-up and on any device-clear command (*, DCL, or SDC). If F0 is sent to the  
8842A, it is internally converted to F1. The function setting can be read using the G0  
command.  
As with the front panel commands, selecting F6 automatically selects the 2000 mA range  
(R5). If the instrument is in range R8, commanding F5 automatically selects the 200 mA  
range (R4). If the instrument is in R1 through R6, commanding F5 automatically selects  
the 2000 mA range (R5). If the 8842A is in a resistance function (F3 or F4) and in R6,  
selecting any other function automatically selects R5. If the 8842A is in range R8 and F2  
or F3 is commanded, range R1 is selected.  
Example  
"F3"  
Explanation  
Selects 2 WIRE kfunction; it does not affect any other settings.  
"* F6"  
Selects mA AC function and 2000 mA range. Resets all other settings to  
default.  
3-11. Get Commands  
The Get commands "get" information from the 8842A for the controller. Each Get  
command loads the output buffer with an output string in the format shown in Figure 3-6.  
Status data (the output from Get commands G0, G1, G3, G4, G5, G6, G7 and G8) is  
interpreted as shown in Table 3-1. The Get commands should not be confused with the  
interface message GET (Group Execute Trigger).  
f3-06.wmf  
Figure 3-6. Output Data Format  
3-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
Table 3-1. Status Data  
OUTPUT  
STRING  
COMMAND  
MEANING  
G0  
Frst  
F =  
1 – 6 as in Funtion commands (Fn)  
9 for Self-Test  
r =  
s =  
t =  
1 – 6 and 8 in Range commands (Rn)  
0 – 2 as in Reading Rate commands (Sn)  
0 – 4 as in Trigger Mode commands (Tn)  
G1  
nn  
nn =  
00 for SRQ disabled (default)  
01 for SRQ on overrange  
04 for SRQ on front panel SRQ  
08 for SRQ on cal step complete  
16 for SRQ on data availible  
32 for SRQ on any error  
Note: SRQ mask values may be added for combinations.Example:  
33 for SRQ on overrange or any error.  
G3  
G4  
aaaaaaaaaaaaaaaa 16 user-defined ASCII characters  
10vm  
V =  
0 Not in cal verification  
1 Cal verification  
m =  
0 Not in calibration mode  
1 A/D calibration  
2 Offset and gain calibration  
4 HF AC calibration  
G5  
1iab  
I =  
0 FRONT inputs selected  
1 REAR inputs selected  
a =  
0Autorange on  
1 Autorange off (manual range)  
b =  
Y =  
w =  
0 OFFSET off  
1 OFFSET on  
G6  
10yw  
10nn  
0 output suffix disabled  
1 output suffix enabled  
0 – 7 as in Terminator commands (Wn)  
G7  
G8  
nn represents error code (See Table 2-1)  
FLUKE,  
mmmmm,  
Mmmmm = 8842A  
Vn.n = IEEE-488 Interface software version number  
0, Vn.n  
The output data from some Get commands starts with a leading 1 or 10. This prevents the  
controller from suppressing leading zeroes and gives a uniform four-character length to  
all instrument configuration data (the data from Get commands G0, G4, G5, G6, and G7).  
The Get commands are described further in the following paragraphs. For more  
information about output data, see paragraph 3-39.  
3-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3-12. G0 (Get Instrument Configuration)  
The G0 command copies the 8842A function, range, reading rate, and trigger mode into  
the output buffer in the format shown in Figure 3-6. The four digits returned represent the  
arguments for the equivalent F, R, S, and T commands, as shown in Table 3-1. An  
example output string follows.  
Example  
Explanation  
3410 CR LF  
3: F3 (2 WIRE kfunction)  
4: R4 (200 krange)  
1: S1 (Medium reading rate)  
0: T0 (Continuous trigger)  
The second digit, which can vary from 1 to 6, indicates what measurement range the  
8842A is in regardless of whether the 8842A is in autorange or manual range.  
The output string from a G0 command is acceptable as an argument for an "N" command.  
This allows you to configure the 8842A from the front panel and then record the  
configuration over the bus for future use with a P0 command. However, 9mmm (meaning  
self-test) can not be used with the P0 command.  
3-13. G1 (Get SRQ Mask)  
The G1 command copies the present SRQ mask into the output buffer in the format  
shown in Figure 3-6. The SRQ mask values are explained in Table 3-1. An example  
output string follows. For more about the SRQ mask, see paragraph 3-47.  
Example  
Explanation  
33 CR LF  
Enable SRQ on any error or overrange  
3-14. G2 (Get Calibration Prompt)  
The G2 command is used when calibrating the 8842A under remote control. The  
command loads the output buffer with a calibration prompt that represents the input  
expected at the analog inputs. The calibration prompt is formatted as a signed decimal  
with exponent, as shown in Figure 3-6. The suffix may be enabled with the Y1 command.  
Example output strings follow.  
Examples  
Explanation  
+1.00000E+0 CR LF  
+190.000E-3 CR LF  
+1.90000E+0, VDC CR LF  
Calibration prompt  
Calibration prompt  
Calibration prompt (Suffix enabled)  
If an error has occurred, the G2 command loads the output buffer with an error message  
instead of the prompt. (See paragraph 3-39.)  
The G2 command is valid only when the calibration mode is enabled by pressing the  
front panel CAL ENABLE switch. If the 8842A is not in the calibration mode, the G2  
command generates an error message.  
3-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
3-15. G3 (Get User-Defined Message)  
The G3 command loads the output buffer with the user-defined message stored in  
calibration memory during the calibration procedure. The message consists of 16 ASCII  
characters, as shown in Figure 3-6.  
The message is stored in calibration memory during calibration using the P3 command. If  
fewer than 16 characters have been stored, the remaining characters returned are spaces.  
If no message has ever been stored, a string of 16 null characters (hex 00) will be  
returned. Some example output strings follow.  
Example  
Explanation  
FL8842A.12-17-83 CR LF  
01-25-84 CR LF  
Identifies instrument and gives cal date.  
Gives cal date. The last eight characters are  
blank.  
3-16. G4 (Get Calibration Status)  
The G4 command is used when calibrating the 8842A under remote control. The  
command loads the output buffer with the instruments calibration status in the format  
shown in Figure 3-6. The status is represented by a four-digit integer which is interpreted  
in Table 3-1.  
The first two digits are always 1 and 0. The third digit indicates whether or not the  
calibration verification mode is enabled. (This mode is enabled only when the calibration  
mode is enabled.) The fourth digit indicates whether or not the calibration mode is  
enabled, and if so, which part of the calibration procedure the 8842A is in. Example  
output strings follow.  
Example  
Explanation  
1000 CR LF 1: Leading 1  
0: Leading 0  
0: Not in cal verification  
0: Cal mode disabled  
1001 CR LF 1: Leading 1  
0: Leading 0  
0: Not in cal verification  
1: Cal mode enabled; A/D cal selected  
3-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3-17. G5 (Get IAB Status)  
The G5 command loads the output buffer with the IAB status in the format shown in  
Figure 3-6. As Table 3-1 explains, the IAB status is a four-character string which  
indicates the status of the FRONT/REAR switch (front or rear analog inputs selected), the  
autorange feature (autorange on or off), and the OFFSET feature (OFFSET on or off).  
The first digit is always 1. An example output string follows.  
Example  
Explanation  
1011 CR LF  
1: Leading 1  
0: FRONT inputs  
1: Autorange off  
1: OFFSET feature on  
It is useful to know whether autorange is on or off because this information is not  
available from the G0 command. For example, the G0 command could indicate that the  
8842A was in the 200 mV range, but it would not indicate whether the 8842A was in  
autorange or manual range.  
3-18. G6 (Get YW Status)  
The G6 command loads the output buffer with the YW status in the format shown in  
Figure 3-6. The YW status is a four-character string which indicates which terminators  
are selected and whether the output suffix is enabled or disabled, as shown in Table 3-1.  
The first two digits are always 1 and 0. An example output string follows.  
Example  
Explanation  
1015 LF CR 1: Leading 1  
0: Leading 0  
1: Y1 (enable output suffix)  
5: W5 (enable LF only)  
3-19. G7 (Get Error Status)  
The G7 command copies the error status register into the output buffer in the format  
shown in Figure 3-6. The first two digits are always 1 and 0. The second two digits  
represent the appropriate error code, if an error has occurred. (Error codes are listed in  
Table 2-1, Section 2). If an error has not occurred, the second two digits are 00. An  
example output string follows.  
Example  
Explanation  
1071 CR LF  
1: Leading 1  
0: Leading 0  
71: Syntax error in device-dependent command string  
The G7 command gives the error status as it exists when the command is executed at its  
position in the input string. The G7 command does not clear the error status register. For  
more information about error messages, see paragraph 3-40.  
3-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
3-20. G8 (Get Instrument Identification)  
The G8 command copies the 8842A instrument identification into the output buffer in the  
format shown in Figure 3-6. The identification is represented by four comma separated  
fields that are interpreted in Table 3-1.  
The first field indicates the manufacturer, the second indicates the instrument model  
number, the third is always zero, and the fourth indicates the version number of the  
IEEE-488 interface software.  
Example  
Explanation  
FLUKE,8842A,0,V4.0 CR LF  
This instrument is a Fluke 8842A with IEEE-  
488 interface software version 4.0.  
3-21. N (Numeric Entry Command)  
Format  
Explanation  
N<numeric entry>  
Where <numeric entry> is one of the following:  
<signed integer>  
<signed real number without exponent>  
<signed real number>E<signed exponent>  
Explanation  
Example  
"N12001"  
Enters the five-digit integer 12001  
Enters -1.23 x 102  
Enters 1.5433 x 101  
"N-1.23E2"  
"N+154.33E-1"  
The N command enters numeric values for use with subsequent Put commands. The  
interpretation of the numeric value depends on which Put command it is used with.  
The E can be used within an N command for entering an exponent of 10. The N can be  
used without an E, but an E requires a prior N. The exponent can be any integer from -9  
to +9.  
The mantissa may exceed 5-1/2 digits. The 8842A accurately calculates the appropriate  
exponent and then disregards all but the first 5-1/2 digits of the mantissa. However, a  
syntax error will occur if the numeric entry overflows the input buffer.  
Example  
Explanation  
"N123456789"  
Enters +1.23456 x 108  
3-22. Put Commands  
The Put commands P0 through P3 set up the 8842As configuration and operating modes  
by entering ("putting") information in the appropriate registers. The put commands are  
described further in the following paragraphs.  
3-23. P0 (Put Instrument Configuration)  
Format  
Explanation  
N<frst>P0  
Where <frst> is a four-digit integer interpreted as arguments for the F,  
R, S, and T commands.  
Example  
Explanation  
3-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
"N3120 P0"  
Identical to F3 R1 S2 T0. Selects 2 WIRE kfunction, 200range,  
fast sample rate, continuous trigger.  
The P0 command allows broadside loading of the Function, Range, Reading Rate, and  
Trigger Mode commands (F, R, S, and T). The codes for these commands are listed in  
Figure 3-5.  
A numeric entry for P0 must be within +1000 and +6824. Each of the four digits must not  
exceed its maximum allowed value (6, 8, 2, and 4, respectively) or an error message will  
occur and the instrument configuration will remain unchanged. The entry may be  
expressed as an integer, real number, or real number with exponent, as described under  
the N command. Any fractional part is ignored.  
Example  
Explanation  
"N3112 P0"  
Sets the 8842A to F3, R1, S1, and T2.  
3-24. P1 (Put SRQ Mask)  
Format  
Explanation  
N<SRQ mask>P1  
Where <SRQ mask> is a two-digit integer from 00 to 63.  
The P1 command is used to program the 8842A to make service requests on user-  
specified conditions. The two-digit code for the SRQ mask is interpreted in Table 3-1  
under the G1 command. For more about the SRQ mask, see paragraph 3-51.  
Numeric entries for the P1 command must be between 0 and +63 (inclusive), or an error  
will occur and the SRQ mask will remain unchanged. The entry may be expressed as an  
integer, real number, or real number with exponent, as described under the N command.  
Any fractional part is ignored.  
Example  
Explanation  
"N0.17E+2 P1" Sets SRQ mask to 17. Enables SRQ on data available or overrange.  
"N1 P1"  
Sets SRQ mask to 01. (A leading zero is assumed.) Enables SRQ on  
overrange.  
3-25. P2 (Put Calibration Value)  
Format  
Explanation  
N<value>P2  
Where <value> can be an integer, real number, or real number with  
exponent, as described under the N command.  
Example  
Explanation  
"N1 P2"  
If the 8842A is in VDC, the next calibration input expected is  
1.00000V dc.  
The P2 command is used to enter variable input calibration values just like the front panel  
VAR IN button. To accept the P2 command, the 8842A must be in the calibration mode  
(enabled by pressing the front panel CAL ENABLE switch). Otherwise, the P2 command  
will generate an error message.  
The variable input is a measurement value that is to be used as the calibration value for  
the next calibration step. Its format is the same as a measurement value. But since it is  
coming from the controller, the value can be specified using any valid format (signed  
interger, real number, or real number with exponent). For example, if the 8842A prompts  
for an input value of 100for the next calibration step, but the available source is  
98.97, the variable input can be specified as "N+9.897E+1", "N0.9897E2", N9897E-2",  
3-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
etc. All of these strings result in the same value being used for the next calibration step.  
For complete information about remote calibration, refer to the Maintenance section.  
Numeric values exceeding full scale and negative values for ohms and AC generate error  
messages.  
3-26. P3 (Put User-Defined Message)  
Format  
Explanation  
P3<value>  
Example  
Where <value> is a string of up to 16 ASCII characters.  
Explanation  
"P3FL.8842.12-17-83" Loads the message "FL.8842.12-17-83" into calibration  
memory.  
"P3HIMOM"  
Loads the message "HIMOM" into calibration memory. The  
remaining eleven characters are assumed to be blank.  
The P3 command stores a user-defined message in the internal calibration memory during  
remote calibration. The message may be read with a subsequent G3 command.  
The message may consist of up to 16 ASCII characters, and typically represents the  
instruments identification, calibration date, calibration facility, etc. If fewer than 16  
characters are specified, spaces are appended to fill the message to 16 characters. Spaces  
and commas in the 16-character input string are suppressed. Lower-case letters are  
converted to upper-case.  
NOTE  
If fewer than 16 characters are specified, the P3 command must not be  
followed by other commands in the same input command string. Otherwise,  
the subsequent commands will be misinterpreted as part of the 16-character  
string.  
To accept the P3 command, the 8842A must be in the calibration mode (enabled by  
pressing the front panel CAL ENABLE switch). Otherwise, the P3 command will  
generate an error message.  
3-27. Rn (Range Commands)  
The Range commands duplicate the front panel range buttons. For example, R0 selects  
autorange, and R3 selects the 20V/20 kQ range.  
The R7 command turns autorange off, just as the AUTO button does when it is toggled.  
Command R7 puts the 8842A into manual range, selecting whatever range the instrument  
is in when the command is received.  
The 8842A defaults to R0 on power-up and any device-clear command (*, DCL, or  
SDC). The range setting can be read using the G0 command.  
3-28. Sn (Reading Rate Commands)  
The Reading Rate commands duplicate the front panel RATE button. Like the RATE  
button, the reading rate command also selects the number of digits displayed and the  
filter setting. (Filter settings are shown in the specifications in Section 1).  
The 8842A defaults to S0 on power-up and any device-clear command (*, DCL, or  
SDC). The reading rate can be read using the G0 command.  
3-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3-29. Tn (Trigger Mode Commands)  
The Trigger Mode commands duplicate the front panel EX TRIG button. In addition, the  
commands can enable or disable the rear panel trigger and the automatic settling time  
delay.  
Figure 3-7 illustrates how to select among the five types of triggers: continuous trigger,  
front panel trigger, rear panel trigger, and two bus triggers. Note that the front panel  
TRIG button is enabled only while the instrument is under local control.  
f3-07.wmf  
Figure 3-6. Trigger Selection Logic Diagram  
In the continuous trigger mode (T0), triggers are initiated at the selected reading rate.  
Each new reading is loaded into the output buffer as it becomes available, unless the  
instrument is busy sending previous output data.  
3-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
DEVICE-DEPENDENT COMMAND SET  
3
The trigger mode can be read using the G0 command. The 8842A defaults to T0 on both  
power-up and any device-clear command (*, DCL, or SDC).  
3-30. Wn (Terminator Commands)  
The Terminator commands select what terminators the 8842A appends to every output  
string. The available terminators are: Carriage Return (CR), Line Feed (LF), and End Or  
Identify (EOI).  
CR and LF are ASCII control codes, sent over the data lines just like output data. EOI is a  
uniline message which is sent simultaneously with the last character in the output string.  
Normally, each output string is terminated with CR followed by LF and EOI.  
The terminator selection can be read using the G6 command. The 8842A defaults to W0  
on power-up and any device-clear command.  
3-31. X0 (Clear Error Register Command)  
The X0 command clears the 8842As error status register. After an X0 command is  
executed, a G7 command (Get Error Status) would return 1000 (no errors).  
Note that the error status register is also cleared when any device-clear command (*,  
DCL, or SDC) is executed. However, X0 is useful for clearing the error status register  
without forcing a complete instrument clear (as do the device-clear commands).  
3-32. Yn (Suffix Commands)  
The Suffix commands enable or disable a suffix which the 8842A can append to all  
numeric data (the data in response to G2 or trigger commands). The suffix includes a  
comma, an overrange indicator (>), and a function indicator (VDC, VAC, OHM, IDC, or  
IAC). The suffix is illustrated in Figure 3-6. An example of suffixed data is given in  
paragraph 3-43.  
To read suffixed data with a controller using BASIC, one can read the whole line into a  
string variable and then convert the numeric part into a numeric variable. However, it is  
much easier to read the numeric part directly into a numeric variable and the suffix into a  
string variable. The leading comma of the suffix serves as a convenient delimiter. For  
example, a BASIC program statement might be:  
INPUT @1,A, B$  
The suffix status can be read using the G6 command. The 8842A defaults to Y0 on  
power-up and any device-clear command (*, DCL or SDC), unless in talk-only mode.  
3-33. Z0 (Self-Test Command)  
The Z0 command initiates the diagnostic self-tests as does pressing the front panel SRQ  
button for 3 seconds. The 8842A then runs through the tests in sequence. (For a  
description of the self-tests, see the Maintenance section.)  
If the 8842A detects an error, an error message is loaded into the output buffer and  
displayed on the front panel. After the last test, the 8842A is reset to the power-up  
configuration, and it begins taking readings.  
It is an error to send the 8842A device-dependent commands during the self-tests.  
However, the controller can still make the 8842A a talker to read the output buffer during  
the test, and thus record each error that occurs, except that only the last of the digital self-  
test errors can be read. After the tests, only the last error is stored in the output buffer if  
more than one error occurred.  
3-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Error messages are indicated by an exponent of +21. For more about error messages, see  
paragraph 3-40.  
Since the 8842A is reset at the end of the self-tests, the Z0 command should be the last  
command in a given command string. The 8842A will ignore any subsequent commands  
in the same command string.  
When the self-tests are complete and no errors have occurred, the serial poll register will  
have bit 5 (Data Available) true and bit 6 (Any Error) false. See paragraph 3-50 for more  
about the serial poll register.  
3-34. (Device-Clear Command)  
The asterisk command (*) is a device-dependent message which resets the 8842A to the  
power-up default settings and clears all registers and buffers except for the input buffer.  
The remote/local status remains unchanged. The asterisk command performs the  
following:  
1. Implements the default settings F1, R0, S0, T0, D0, B0, Y0, W0.  
2. Clears the error status register (equivalent to X0).  
3. Zeros the SRQ mask, prohibiting service requests (equivalent to N0 P1).  
4. Zeros the numeric entry register (equivalent to N0).  
5. Zeros the serial poll register.  
6. Sets the SRQ line false.  
The asterisk command is executed in its proper turn in a string, just like any other  
command, without affecting the contents of the input buffer. All commands which  
precede the asterisk command are performed.  
The asterisk command is useful to ensure that the 8842A is initialized to the same state  
each time a program is run. By contrast, the similar interface messages DCL (Device  
Clear) and SDC (Selected Device Clear) cause the entire input buffer to be cleared  
immediately.  
DCL, SDC, and the asterisk command are all considered to be device-clear commands  
because the results are so similar; however, DCL and SDC are not identical to the asterisk  
command described here. DCL and SDC are discussed further in the paragraph on  
interface messages.  
3-35. ? (Single-Trigger Command)  
The Single-Trigger command (?) causes the 8842A to take a reading and place the result  
into the output buffer. To accept this command, the 8842A must be in external trigger  
mode (selected by the T1, T2, T3, or T4 command).  
The Single-Trigger command is one of five ways to trigger a reading. (See Figure 3-7.)  
Of these, only the Single-Trigger command (?) and the Group Execute Trigger command  
(GET) are loaded into the input buffer.  
3-36.INPUT SYNTAX  
The following paragraphs describe how to construct groups of commands for the 8842A.  
A few definitions are presented first, followed by a description of how the 8842A  
processes input commands. Guidelines are then summarized in four syntax rules.  
3-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
INPUT SYNTAX  
3
3-37. Definitions  
Output commands: Commands which load data into the output buffer. The output  
commands are: the Get commands (G0 through G8); the Single-Trigger Command  
(?); the Continuous Trigger command (T0); and Group Execute Trigger (GET), not to  
be confused with the Get commands.  
Input terminator: An ASCII control code sent by the controller which tells the 8842A  
to execute all device-dependent commands since the previous terminator.  
Terminators are CR (Carriage Return), LF (Line Feed), EOI (End Or Identify), and  
GET (Group Execute Trigger).  
Input command string: One or more device-dependent commands followed by a  
terminator.  
3-38. Input Processing  
When the 8842A receives commands from the bus, it stores them in a 31-character input  
buffer as a continuous string of characters. Commands in the input buffer are not  
executed or checked for syntax until an input terminator is received or the input buffer  
becomes full. The only valid input terminators are CR, LF, GET (Group Execute  
Trigger), and/or EOI.  
When the 8842A receives an input terminator, it executes the previous commands in the  
order in which they were received. As input characters are processed and executed, space  
is made available in the input buffer for new characters.  
If the input buffer becomes full, the 8842A stops accepting characters from the bus until  
all complete command strings currently in the input buffer have been executed. In this  
way, characters sent to the 8842A are never lost due to buffer overflow.  
In some instances, a terminator is automatically transmitted at the end of the controllers  
output string. For example, in Fluke BASIC, the PRINT statement always finishes with a  
CR LF pair. If a controller does not have this feature, the programmer must transmit a  
terminator explicitly.  
The 8842A accepts alphabetic characters in either upper or lower case. Spaces, commas,  
and control codes are ignored and are not placed in the input buffer. If the 8842A  
receives a group of terminators (such as CR LF or CR LF EOI), only a single terminator  
is loaded into the input buffer. Numeric values used in PUT commands may be in NR1,  
NR2, or NR3 format as described in the IEEE-488 Codes and Formats Recommended  
Practice. (These correspond to the signed integer, real number, and real-number-with-  
exponent formats described under the N command.) For reference, Figure 3-8 shows how  
the 8842A interprets messages.  
3-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
DEVICE-DEPENDENT MESSAGES  
Single-character Commands  
Bn Cn Dn Fn Gn Pn Rn Sn Wn Xn Yn Zn  
Each of these commands requires the single  
numeric digit (n).  
Numeric-entry Characters  
NE. + - 0123456789  
These characters are used for entering numbers  
Terminators  
CR  
Carriage Return  
LF  
Line Feed  
GET  
Group Execute Trigger  
End Or Idenity (used as END/DAB)  
EOI  
INTERFACE MESSAGES  
Address Messages  
MLA  
My Listen Address  
My Talk address  
Unlisten  
MTA  
UNL  
UNT  
Untalk  
Universal Commands  
ATN  
Attention  
DCL  
Device Clear  
IFC  
Interface Clear  
Local Lockout  
Remote Enable  
Serial Poll Disable  
Serial Poll Enable  
LLO  
REN  
SPD  
SPE  
Addressed Commands  
GET  
Group Execute Trigger  
Go to Local  
GTL  
SDC  
Selected Device Clear  
Ignored Characters  
, comma  
These characters may be inserted anywhere in a  
character string without affecting the string.  
Space  
Character string without affecting the string.  
All Other ASCII non-printing characters (except CR They carry no special meaning and are ignored by  
and LF)  
the 8842A. They are not placed in the input buffer.  
ERROR-PRODUCING CHARACTERS  
! “ # $ ‘ ( ) / : < = > ; @ [ / ]  
HIJKLMOQUV  
~
The error annunciator is displayed on the 8842A  
front panel when one of these characters is  
encountered (ERROR 71).  
Figure 3-7. Interpretation of Messages  
3-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
INPUT SYNTAX  
3
Illegal commands (e.g., F9) generate an error message, but are otherwise ignored, and do  
not affect the instruments configuration.  
Example  
Explanation  
"* F9"  
This would load the output buffer with an error message and select F1  
(established by the * command).  
3-39. Syntax Rules  
Four syntax rules should be followed when constructing input command strings. They  
are:  
RULE 1: Read output data only once.  
To prevent old (previously read) data from being read a second time by mistake, the  
output buffer is always cleared after it has been read. If the output buffer is read twice  
without an intervening output command, the 8842A will not respond to the second  
attempt to read the output buffer. (However, if the 8842A is in T0, no intervening  
command is necessary.)  
RULE 2: Use no more than one output command per input command string.  
Because the 8842A has only one output buffer, it writes new data over old. If an input  
command string contains more than one output command, only the data from the last  
command can be read.  
Example  
Explanation  
"F1 T3 ? F2 ?" Improper construction. The second trigger writes over the first. To  
obtain two readings, send two complete command strings (separated by  
terminators).  
"F2 R3 S0 T3 ?"  
"F2 R3 S0"  
Correct construction. The string contains only one output command.  
Correct construction. It is permissible for a string not to contain an  
output command.  
RULE 3: Read the output data generated by one input command string before sending  
the next input command string.  
Output data remains available in the output buffer until it is read, or until the next  
input command string is received. As soon as the controller finishes reading the output  
buffer, or as soon as the 8842A receives a new input terminator, the Data Available bit  
in the serial poll register is set false. When this bit is false, data can no longer be read  
from the output buffer. Therefore, an output string which is available must be read by  
the controller before, rather than after, the next input command string is sent.  
Rule 3 is most evident in the external trigger mode, and is best demonstrated by a  
programming example. The following program is written first incorrectly, and then  
correctly, in Fluke BASIC using the 1722A Instrument Controller.  
Incorrect example  
100  
200  
300  
PRINT @3, "T1 ?"  
PRINT @3, "F4"  
INPUT @3, A  
3-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
In this incorrect example, the INPUT statement is located incorrectly for reading the  
measurement data from line 100. The new input command string "F4" disallows the  
reading of data from the output buffer.  
Correct example  
100  
200  
300  
PRINT @3, "T1 ?"  
INPUT @3, A  
PRINT @3, "F4"  
In this example, the reading taken at line 100 is read at line 200. Then the F4 command is  
sent. Note that in the external trigger mode, the reading from line 100 flashes on the  
8842A display too briefly to see. This is because the function change at line 300 blanks  
the display until the next trigger.  
The previous example could also be correctly programmed as follows:  
100  
200  
PRINT @3, "T1 ? F4"  
INPUT @3, A  
RULE 4: If an input command string contains a trigger, enter the commands in the  
following order:  
a. Commands to configure the instrument (if any).  
b. The trigger command.  
c. Commands to re-configure the instrument (if any).  
d. Terminator(s).  
The principle behind this rule is that the 8842A executes all commands in the exact order  
they are received, from left to right as written.  
Example  
"F3 F4 ?"  
"F3 ? F4"  
Explanation  
Improper construction. F3 is effectively discarded.  
Correct construction. The 8842A is configured in F3, and the trigger is  
executed. Then the 8842A is left in F4.  
3-40.OUTPUT DATA  
The following paragraphs describe the data that can be loaded into the 8842A output  
buffer and sent to the IEEE-488 bus. The paragraphs describe how and when data is  
loaded into the output buffer, the types of output data, and their relative priority.  
Note that the 8842A can also send data to the IEEE-488 bus from the serial poll register.  
For a description of this data, see paragraph 3-47.  
3-41. Loading Output Data  
The output buffer is loaded when the 8842A receives an output command, or when an  
error occurs. Output commands are those device-dependent commands which load the  
output buffer with data:  
Get commands (G0 through G8)  
Single-trigger command (?)  
Group execute trigger (GET)  
Continuous Trigger (T0)  
3-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
OUTPUT DATA  
3
Because the 8842A gives priority to input processing, it completely processes all  
characters in the input buffer before it loads the output buffer. When the output buffer is  
loaded, the Data Available bit in the serial poll register is set true.  
Data from the output buffer is not actually loaded onto the IEEE-488 bus until the  
controller addresses the 8842A as a talker. This is done by sending the interface message  
MTA (My Talk Address).  
3-42. Types of Output Data  
The types of data that can be loaded into the output buffer are shown in Figure 3-6. Each  
type has its own format. Error messages, which are loaded into the output buffer if an  
error occurs, are formatted as numeric data.  
3-43. Numeric Data and Error Messages  
Numeric data is loaded into the output buffer in response to the G2 command or an  
instrument trigger, and has the format shown in Table 3-2. The exponent is always a  
multiple of 3, as in engineering notation. The position of the decimal point matches the  
front panel display.  
Numeric data is of constant length, It is 11 characters (plus terminators) when the suffix  
is disabled, and 16 characters (plus terminators) when the suffix is enabled.  
The suffix is enabled by the Y1 command, and consists of five ASCII characters as  
shown in Figure 3-6. The suffix is appended only to numeric data, never to status data.  
The terminators are determined by the Terminator commands. The default is CR LF EOI.  
There are three types of numeric data: measurement data, overrange indication, and error  
messages.  
3-44. MEASUREMENT DATA  
Measurement data has the numeric data format shown in Table 3-2, and is always in the  
units of volts, amps, or ohms.  
Table 3-2. Numeric Output Data Format  
RANGE  
MEASUREMENT DATA  
2-, 4-WIRE k  
±1x.xxxxE+0**  
±1xx.xxxE+0  
OVERRANGE  
INDICATION  
ERROR  
MESSAGES  
VDC, VAC  
±1x.xxxxE-3*  
±1xx.xxxE-3  
±1.xxxxxE+0  
±1x.xxxxE+0  
±1xx.xxxE+0  
±1xxx.xxE+0  
--  
MA DC, mA AC  
R8  
R1  
R2  
R3  
R4  
R5  
R6  
--  
±9.99999E+9  
±9.99999E+9  
±9.99999E+9  
±9.99999E+9  
±9.99999E+9  
±9.99999E+9  
±9.99999E+9  
+1.00xxE+21  
+1.00xxE+21  
+1.00xxE+21  
+1.00xxE+21  
+1.00xxE+21  
+1.00xxE+21  
+1.00xxE+21  
--  
±1.xxxxxE+3  
--  
±1x.xxxxE+3  
--  
±1xx.xxxE+3  
±1xx.xxxE-3***  
±1xxx.xxE-3  
--  
±1xxx.xxE+3  
±1x.xxxxE+6  
* VDC only  
** 4-wire ohms  
*** mA DC only  
3-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
NOTE  
In the fast (F) reading rate, the least significant digit is always zero, and  
should be disregarded when interpreting accuracy specifications.  
3-45. OVERRANGE INDICATION  
If a reading is overrange (200,000 counts), the measurement data has the following  
format:  
±9.99999E+9 <suffix> <terminators>  
Overvoltage readings (> 1000V dc or 700V ac) do not result in this display.  
3-46. ERROR MESSAGES  
If the 8842A detects an error, it loads an error message into the output buffer in the  
following numeric format:  
+1.00xxE+21 <terminators>  
The digits xx represent a two-digit error code. (Error codes are listed in Table 2-1,  
Section 2.) The suffix is always suppressed for error messages.  
Example  
Explanation  
+1.0071E+21 CR LF ERROR 71: Syntax error in device-dependent command string.  
As with local operation, none of the errors are latching except for ERROR 31. If the mA  
DC or mA AC function is requested while the FRONT/REAR switch is in the REAR  
position, ERROR 31 will persist until the switch is set to FRONT or another function is  
selected.  
To check for an error condition, test whether the output buffer data is greater than or  
equal to +1E+21, or test the Any Error bit (bit 6) in the serial poll register.  
3-47. Status Data  
Status data is the output in response to G0, G1, G3, G4, G5, G6, G7 and G8, commands.  
The data is formatted as shown in Figure 3-2, and is interpreted in Table 3-1. Examples  
of status data can be found in the description of the Get commands.  
The user-defined message loaded by the G3 command consists of 16 characters plus  
terminators. The SRQ mask loaded by the G1 command consists of two integers plus  
terminators. All other status data is always a four-digit integer plus terminators. The  
terminators LF (Line Feed) and CR (Carriage Return) each add an extra character when  
enabled.  
The 8842A begins some status data with a leading ASCII one (1) or a one and a zero  
(10). This prevents the controller from suppressing any leading zeros present in the  
8842As output string. It also gives a uniform four-character length to all instrument  
configuration data.  
Status data from the Get commands reflects the status of the 8842A at the time the  
command is executed at its place in the input command string.  
3-48. Output Priority  
Since only one output string is allowed per input command string, the 8842A gives  
priority to some types of data over others. An input command string may call for more  
than one output string. (For example, an input string may contain a Get command but  
also cause an error message.) However, the output buffer is loaded with only one output  
string. That string is selected according to the following priority:  
3-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
SERVICE REQUESTS  
3
1. Status data (from G0, G1, G3, G4, G5, G6, G7 and G8)  
2. Error messages (if an error exists)  
3. Numeric data (from G2 or a trigger)  
This means that an error message always overrides numeric data, but status data is sent  
even in the presence of an error. However, the status data does not clear the error  
message; the error message is sent the next time numeric data is requested.  
3-49.SERVICE REQUESTS  
Service requests let bus instruments get the attention of the system controller. The  
requests are sent over the SRQ line (one of the IEEE-488 bus lines). If more than one  
instrument on the bus is capable of sending service requests, the controller can learn  
which one made the request by taking a serial poll. Each device (including the 8842A)  
responds to the poll by sending the contents of its serial poll register. The serial poll  
register indicates whether or not the device requested service, and if so, the reason for the  
request.  
The 8842A may be programmed to make a service request on user-specified conditions.  
The conditions are specified by entering a value for the service request mask (SRQ mask)  
with the P1 command. The SRQ mask works by monitoring the serial poll register, which  
in turn monitors various conditions in the 8842A.  
3-50. The Serial Poll Register  
The serial poll register is a binary-encoded register which contains eight bits, as  
illustrated in Figure 3-9. The controller can read the 8842A serial poll register at any time  
by taking a serial poll. Because serial poll register data is loaded directly onto the bus  
(instead of being loaded into the output buffer first), reading the serial poll register leaves  
data in the output buffer intact.  
Service requests may also be initiated using the front panel SRQ button if it has been  
enabled by the SRQ mask.  
The eight bits of the serial poll register are described in Figure 3-9. Note that the SRQ  
mask uses bits 1 through 6 to set bit 7 (the RQS bit). When the RQS bit is set true, the  
8842A sets the SRQ line true, which generates a service request. A bit is considered true  
when it is set to 1.  
3-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
BIT: 8  
7
6
5
4
3
2
1
0
RQS  
ANY  
DATA  
CAL STEP  
FRONT  
0
OVER-  
ERROR  
AVAILIBLE COMPLETE PANEL SRQ  
RANGE  
DECIMAL  
VALUE:  
64  
32  
16  
8
4
2
1
Bit  
Name  
Set  
Cleared  
1
Overrange  
An overrange condition occurs  
Device command received, or  
Bus or Rear Panel Trigger, or  
Output buffer is read  
2
3
Not used  
Never  
Always  
Front panel SRQ  
Cal Step Complete  
Data Availible  
Front panel SRQ button  
pressed  
Device command received  
4
5
Completion of store command  
(C0)  
Device command received  
Output buffer loaded with any  
data (Readings, Error  
Messages. Get Responses)  
Device command received, or  
Bus or Rear Panel Trigger, or  
Output buffer is read  
6
Any Error  
An error condition occurs. At  
the same time the output buffer  
is loaded with an error  
Device command received, or  
Output buffer is read  
message. This sets bit 5.  
7
8
RQS  
Any SRQ mask-enabled bit is  
set.  
All SRQ mask-enabled bits are  
cleared  
Not used  
Never  
Always  
Figure 3-8. Serial Poll Register  
For example, the serial poll register reads 00010000 (binary) when data is available. This  
value is read in binary by the controller, which might numerically reformat the value to  
16 (decimal) or 10 (hexadecimal).  
The serial poll register is cleared whenever the 8842A receives a new input command  
string.  
3-51. The SRQ Mask  
The SRQ mask is a two-digit integer that specifies which conditions will generate a  
service request. The SRQ mask is entered using the P1 command and can be read with  
the G1 command. The conditions corresponding to each SRQ mask value are listed under  
G1 in Table 3-1.  
3-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
INTERFACE MESSAGES  
3
The SRQ mask can enable any combination of serial poll register bits 1 through 6. Its six-  
bit binary representation is ANDed bit-for-bit with serial poll register bits 1 through 6  
whenever the output buffer is loaded. If any mask-enabled bit in the serial poll register  
comes true, the RQS bit (bit 7) is set true, generating a service request.  
Example  
Explanation  
"* N4 P1 ?" An SRQ is generated if the front panel SRQ button is pressed. The string  
sets the SRQ mask to 04, which is 000100 in binary. This mask is ANDed  
with the lower six bits of the serial poll register. The mask thus enables bit  
3, the Front Panel SRQ bit.  
The SRQ mask codes can be added to select combinations of conditions.  
Example  
Explanation  
"* N5 P1 ?" An SRQ is generated if the SRQ button is pressed or if the trigger results  
in an overrange reading. The SRQ mask is set to 05, which is 000101 in  
binary.  
At power-up or on any device-clear command, the SRQ mask is set to 00 (decimal). This  
prevents service requests by holding the RQS bit false under all conditions. For other  
examples of the SRQ mask, see the description of the P1 command.  
3-52.INTERFACE MESSAGES  
The interface messages understood by the 8842A can be separated into the three main  
classes described in the IEEE-488 Standard: address messages (AD), universal  
commands (UC), and addressed commands (AC). All interface messages described here  
originate at the controller.  
3-53. Address Messages  
Address messages are used by the controller to communicate talk and listen control to  
other devices on the bus. Address messages are sent over the eight data lines of the bus  
while the controller holds ATN true. Address messages are processed immediately and  
are not placed in the input buffer. The address messages are:  
MLA My Listen Address -- Addresses a device to listen  
MTA My Talk Address -- Addresses a device to talk  
UNL Unlisten -- Addresses all listeners to unlisten  
UNT Untalk -- Addresses all talkers to untalk  
3-54. Universal Commands  
Universal commands are accepted and interpreted by all devices on the bus. The  
commands are of two types, multiline messages and uniline messages. Multiline  
messages are sent over the eight parallel data lines in the IEEE-488 bus. Uniline  
messages are sent over one of the individual interface management lines in the IEEE-488  
bus. All universal commands except DCL are processed immediately by the 8842A,  
ahead of any device-dependent commands. Only DCL enters the 8842A input buffer.  
The 8842A responds to the following universal messages:  
ATN Attention -- A uniline message which causes the 8842A to interpret multiline  
messages as interface messages (AD, AC, or UC). When false, multiline  
messages are interpreted as device-dependent messages.  
3-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
IFC  
Interface Clear -- A uniline message which clears only the interface (not the  
8842A) by placing it in a known quiescent state.  
REN Remote Enable -- A uniline message which, when received with MLA, switches  
the 8842A to remote. When REN is set false, the 8842A switches to local and  
removes local lockout.  
DCL Device Clear -- A multiline message which is loaded into the input buffer as a  
special device-clear command. DCL performs the same operation as the device-  
dependent command *, except that it is read before any other characters that are  
already present in the input buffer, and clears the entire input buffer. Processing  
then continues normally. The action of DCL is not immediate; if the 8842A is  
taking a reading when DCL is received, the DCL command is executed after the  
measurement is finished.  
LLO Local Lockout -- A multiline message which disables the front panel LOCAL  
button. The result is that the local mode is not accessible by front panel control.  
SPD Serial Poll Disable -- A multiline message which removes the serial poll enable  
state.  
SPE  
Serial Poll Enable -- A multiline message which causes the serial poll data (rather  
than output buffer data) to be transferred on the bus once ATN becomes false.  
3-55. Addressed Commands  
Addressed commands are multiline messages which are accepted and interpreted by only  
those devices currently addressed to listen. The 8842A responds to the following  
addressed commands:  
GET Group Execute Trigger -- (Not to be confused with the device-dependent Get  
commands.) GET loads a trigger command into the input buffer and also  
terminates the string at that point. Only a single character is loaded into the input  
buffer. The trigger command is executed in its proper turn in the input string,  
rather than immediately. When executed, GET initiates a measurement.  
GTL Go To Local -- Causes the 8842A to switch to local. This command does not  
enter the input buffer. If the display has been blanked (with a D1 command),  
issue a D0 command before sending GTL.  
SDC Selected Device Clear -- Identical to the universal command DCL, but is  
accepted and interpreted by current listeners only. Therefore, it clears the 8842A  
only if it is addressed to listen.  
3-56.TALK-ONLY MODE  
The talk-only mode lets you take advantage of the remote capability of the 8842A  
without having to use an instrument controller. To put the 8842A in the talk-only mode:  
1. Turn the 8842A POWER switch OFF.  
2. Set the rear panel TALK ONLY bit switch to 1 (the up position).  
3. Connect the 8842A via the IEEE-488 bus to your printer, data logger, or other device.  
4. Configure the other device as a listener only.  
5. Turn the 8842A POWER switch ON.  
3-30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
REMOTE CALIBRATION  
3
6. Configure the 8842A with the front panel controls.  
The 8842A reads the TALK ONLY bit switch on power-up and when it receives the  
interface command IFC. You can therefore set the TALK ONLY switch to 1 after power-  
up as long as you then send the 8842A the IFC command.  
3-57.REMOTE CALIBRATION  
The 8842A can be calibrated over the IEEE-488 bus using remote commands. Refer to  
the Maintenance section for details.  
3-58.TIMING CONSIDERATIONS  
To help you take the best advantage of the speed of the 8842A, external trigger timing  
for the IEEE-488 Interface is described in the specifications in Section 1.  
3-59.IMMEDIATE MODE COMMANDS  
Many controllers have some amount of "immediate mode" capability. That is, commands  
may be given interactively to the 8842A via the run-time-system without the need for  
actually running a program. The controller accepts and executes these commands one at a  
time. Some commonly used commands are listed in Table 3-3 for several controllers.  
These are provided for the convenience of instrument demonstrations, set-up and check-  
out, and for those with limited experience with IEEE-488 bus operations.  
As a general note: The entire 8842A command set should work well provided the "port  
clear" and "device clear" commands are given first. You should then be able to send any  
other commands in the 8842A command set without repeating the clearing commands.  
3-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 3-3. Immediate-Mode Commands for Various Controllers  
FUNTION  
PERFORMED  
FLUKE-BASIC on  
1720A or 1722A  
HP-HPL on  
HP9825  
Calculator  
HP-BASIC on  
HP9816-PC and HP-  
85 Calculator  
TEK-BASIC on  
4051 Graphics  
System  
INITIALIZE Port  
INIT PORT 0  
cli 7  
CLEAR 7  
INIT  
CLEAR Instrument  
REMOTE Commands  
LOCAL Control  
CLEAR @4  
REMOTE @4  
LOCAL @4  
cir 704  
CLEAR 704  
PRINT @4: “*”  
WBYTE @36, 17:  
WBYTE @36, 1:  
PRINT @4:”T1”  
PRINT @4:”?”  
INPUT @4:A  
rem 704  
Icl 704  
REMOTE 704  
LOCAL 704  
EXTERNAL TRIGGER  
TRIGGER Instrument  
GET Output Data  
PRINT @4,”T1”  
TRIG @4  
wrt 704,”T1”  
trg 704  
OUTPUT 704;”T1”  
TRIGGER 704  
INPUT @4,A  
PRINT A  
red 704, A  
prt A  
ENTER 704;A (Note 2)  
PRINT A (Note 2)  
OUTPUT 704;”F2”  
OUTPUT 704;”R4”  
PRINT Data to Screen  
CONFIGURE for VAC  
PRINT A  
PRINT @4,”F2”  
PRINT @4,”R4”  
wrt 704,”F2”  
wrt 704;”R4”  
PRINT @4:”F2”  
PRINT @4:”R4”  
CONFIGURE for  
200Vac  
TRIGGER Continuously PRINT @4,”TO”  
wrt 704,”TO”  
wrt 704,”Y1”  
OUTPUT 704; “TO”  
OUTPUT 704;”Y1”  
PRINT @4:”TO”  
PRINT @4:”Y1”  
SUFFIXES Enable  
GET Data & Suffix  
PRINT @4,”Y1”  
INPUT @4, A,A$  
Red 704,A,A$  
(Note 1)  
ENTER 704;A,A$ (Note INPUT %4:A,A$  
2)  
PRINT Data & Suffix  
Notes:  
PRINT A,A$  
Prt A$,A  
PRINT A,A$ (Note 2)  
PRINT A,A$  
1. Before using A$ on the 9825 is necessary to enter ”dimA$[6]” to allocate a string variable. This statement allows  
six characters.  
2. In the HP9816 system, variables cannot be created from the keyboard; they must be created by running a  
program. (See error 910 for that system.) To get around this, type in a very short program as follows:  
SCRATCH  
10 A = 0  
20 A$ = ‘’’’  
30 END  
(Hit “EXEC” key)  
(Hit “ENTER” key)  
(Hit “ENTER” key)  
(Hit “ENTER” key)  
(Hit “RUN” Key)  
This program creates the variables ‘A’ and ‘A$’ so that they may be accessed in immediate mode and changed at  
will. This program is not necessary for the HP-85 Calculator.  
3-60.EXAMPLE PROGRAMS  
Several example programs for the 8842A using various controllers are presented in the  
remaining figures in this section. In all of these examples, the 8842A is set to IEEE-488  
address 4 (rear panel switch setting 000100). Of course, any other address (00 to 30)  
could be substituted.  
In each of these examples, the instrument is cleared prior to configuration set-ups. This  
ensures that the 8842A configuration has been completely defined.  
To run these programs, it is not necessary to type in all the comments (which appear to  
the right of the exclamation marks). Also, spaces are placed between commands for ease  
of reading; they are not required.  
3-32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-10.wmf  
Figure 3-9. Example Program  
3-33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f3-11.wmf  
Figure 3-10. Example Program: Taking Readings with Local Control  
3-34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-12.wmf  
Figure 3-11. Example Program: Using the Serial Poll Register  
3-35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f3-13.wmf  
Figure 3-12. Example Program: Record Errors During Selftest  
3-36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_01.wmf  
Figure 3-13. Example Programs: Using the IBM PC  
3-37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f3-14_01.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_02.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f3-14_03.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_04.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f3-14_05.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_06.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f3-14_07.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_08.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f3-14_09.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
f3-14_10.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f3-14_11.wmf  
Figure 3-14. Example Programs: Using the IBM PC (cont)  
3-48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Programming  
EXAMPLE PROGRAMS  
3
Table 3-4. ASCII/IEEE Std 488-1978 Bus Codes  
3-49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
NOTE  
For the examples using the Fluke 1720A or 1722A, the 8842A is plugged  
into port 0. The port is initialized by the INIT statement, which sends IFC  
(interface clear).  
3-50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4  
Measurement Tutorial  
Title  
Page  
4-1.  
4-2.  
4-3.  
4-4.  
4-5.  
4-6.  
4-7.  
4-8.  
4-9.  
4-10.  
4-11.  
4-12.  
INTRODUCTION................................................................................4-2  
Circuit Loading Error.......................................................................4-2  
Input Bias Current Error...................................................................4-3  
2-Wire Ohms....................................................................................4-4  
Correcting for Test Lead Resistance in 2-Wire Ohms .....................4-5  
4-Wire Ohms....................................................................................4-5  
Applications of the Ohms Functions................................................4-7  
TESTING ELECTROLYTIC CAPACITORS ............................4-7  
A PRECISION CURRENT SOURCE.........................................4-8  
4-15. AC VOLTAGE AND CURRENT MEASUREMENT........................4-10  
4-16.  
4-17.  
4-18.  
4-19.  
4-20.  
4-21.  
4-22.  
True RMS Measurement..................................................................4-10  
Waveform Comparison....................................................................4-10  
Crest Factor......................................................................................4-12  
AC-Coupled AC Measurements ......................................................4-12  
Combined AC and DC Measurements.............................................4-13  
Bandwidth ........................................................................................4-13  
Zero-Input VAC Error......................................................................4-13  
4-23. MAKING ACCURATE MEASUREMENTS ON THE 20 mV  
4-24. MAKING ACCURATE HIGH-RESISTANCE MEASUREMENTS .4-15  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
4-1. INTRODUCTION  
This section discusses considerations and techniques to help you use the 8842A  
effectively. Among other things, this section discusses sources of error which are an  
inherent part of the measurement process and which occur for all multimeters. By  
understanding why and when these errors occur, and by knowing how and when to  
correct for them, you can make accurate measurements with confidence.  
This section also discusses the relative benefits of 2-wire and 4-wire ohms, describes  
special considerations for making ac measurements, and presents some unusual  
applications--for example, using the test current in the 2-wire ohms function as a  
troubleshooting tool in itself.  
4-2. DC VOLTAGE MEASUREMENT  
When measuring dc voltages in high-impedance circuits, there are two possible sources  
of error to consider: circuit loading and input bias current.  
4-3.  
Circuit Loading Error  
Whenever a voltmeter is connected to a circuit, the voltmeters internal resistance changes  
the voltage of the circuit under test. The resulting error is called circuit loading error. The  
error is negligible as long as the resistance of the circuit under test (the source  
impedance) is small compared to the input impedance of the meter. As the source  
impedance approaches the input impedance of the voltmeter, the error can be  
considerable. The percentage of error can be calculated using the formula in Figure 4-1.  
f4-01.wmf  
Figure 4-1. Circuit Loading Error Calculation  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
DC VOLTAGE MEASUREMENT  
4
The input impedance of the 8842A is 10 Min the 200V and 1000V dc ranges, and is  
greater than 10,000 Min the 20 mV, 200 mV, 2V, and 20V ranges. Therefore, for the  
8842A, circuit loading error is less than 0.01% as long as the source impedance is less  
than 1 Min the 20 mV, 200 mV, 2V, and 20V ranges, and less than 1 kin the 200V  
and 1000V ranges. The exceptionally high input impedance on the 20V dc range allows  
high-accuracy readings in CMOS and high-impedance analog circuitry.  
NOTE  
Input protection circuitry can reduce the input impedance to as low as 100  
kwhen the input is overrange. This may also occur momentarily when the  
instrument autoranges to a higher range.  
4-4.  
Input Bias Current Error  
Input bias current error occurs because a voltmeters input bias current always changes  
the voltage of the circuit under test. However, the error is significant only when  
measuring voltages in circuits with very high source impedance. The error can be  
measured as shown in Figure 4-2.  
f4-02.wmf  
Figure 4-2. Measuring Input Bias Current Error  
With the 8842A, it is easy to correct for this error using the OFFSET button:  
1. Select the VDC function and the desired range.  
2. Connect the 8842A INPUT terminals to a resistor which matches the source  
impedance of the circuit to be tested.  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3. Allow the displayed reading to settle.  
4. Press the OFFSET button.  
5. Remove the resistor.  
6. Proceed with the desired measurement.  
Example:  
Measure a 1.5V source with 1 Msource impedance, correcting for input bias current.  
1. Connect a 1 Mresistor between the INPUT HI and INPUT LO terminals.  
2. Select the VDC function and the 2V range.  
3. Allow the display to settle.  
4. Press OFFSET. (This zeroes the input bias current error.)  
5. Remove the 1 Mresistor.  
6. Measure the voltage of the circuit under test.  
Note that this procedure does not correct for circuit loading error. Also note that if input  
bias current error is not corrected for, it may be added to the circuit loading error.  
4-5. RESISTANCE MEASUREMENT  
The 8842A allows you to measure resistance in both 2-wire and 4-wire configurations.  
Each has its benefits.  
4-6.  
2-Wire Ohms  
Two-Wire ohms measurements are simple to set up and yield good results for most  
measurement conditions. Measurements are made as shown in Figure 4-3. An internal  
current source (the "ohms current source") passes a known test current (Itest) through the  
resistance being tested (Runknown). The 8842A measures the voltage drop across  
Runknown, calculates Runknown using Ohms law (Runknown = Vtest/Itest), and  
displays the result.  
f4-03.wmf  
Figure 4-3. Wire Ohms Measurement  
4-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
RESISTANCE MEASUREMENT  
4
The test current and full-scale voltage for each resistance range are shown in Table 4-1.  
Since the HI INPUT test lead is positive with respect to the LO INPUT lead, these test  
leads are not interchangeable when a semiconductor device is being measured.  
4-7.  
Correcting for Test Lead Resistance in 2-Wire Ohms  
In 2-wire ohms, the resistance of the test leads can introduce error when measuring low  
resistances. Typical test leads may add as much as 0.5Q to 2-wire ohms readings.  
With the 8842A, it is easy to correct for this error using the OFFSET button:  
1. Select the 2-wire ohms function.  
2. Touch the test leads together. The 8842A should indicate the resistance of the test  
leads.  
3. With the test leads still touching, press the OFFSET button.The 8842A should read  
0.  
4-8.  
4-Wire Ohms  
Four-Wire ohms measurements provide the highest accuracy for low resistance  
measurements. The 4-wire configuration automatically corrects for both test lead  
resistance and contact resistance. Contact resistance (the resistance between the test probe  
tips and the circuit being tested) is unpredictable, and therefore cannot be reliably  
corrected with a fixed offset.  
Four-Wire ohms measurements are especially important when using long test leads. In a  
typical automated test system, for example, the test leads could be connected through  
four or five switching relays, each with 2of resistance!  
NOTE  
Instability of the test leads resistance can cause significant error on low  
ohms ranges, particularly on the 20and 200ranges. Therefore, only 4-  
wire ohms measurement is permitted in the 20range.  
The 8842A makes 4-wire ohms measurements as shown in Figure 4-4. The HI and LO  
INPUT leads apply a known, internal current source to the unknown resistance, just as in  
2-wire ohms. (See Table 4-1.) However, the voltage drop across the unknown resistance  
is measured with the SENSE leads rather than the INPUT leads. Since the current flow in  
the SENSE leads is negligible, the error caused by the voltage drop across the leads is  
also negligible.  
4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f4-04.wmf  
Figure 4-4. Wire Ohms Measurement  
Table 4-1. Ohms Test Current  
RANGE  
TEST CURRENT  
FULL SCALE VOLTAGE  
20  
200Ω  
1 mA  
1 mA  
0.02V  
0.2V  
2 kΩ  
1 mA  
2.0V  
20 kΩ  
200 kΩ  
2000 kΩ  
20 MΩ  
100 µA  
10 µA  
5 µA  
2.0V  
2.0V  
10.0V  
10.0V  
500 nA  
4-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
RESISTANCE MEASUREMENT  
4
NOTE  
In the 2 Mand 20 Mranges of 4-wire ohms, the voltage across the  
unknown resistance is sensed between the HI SENSE and LO INPUT  
terminals. Accuracy is not affected as long as the resistance of the LO  
INPUT lead is less than 10in the 2 Mrange, and less than 100in the  
20 Mrange.  
4-9.  
Applications of the Ohms Functions  
The 2-wire and 4-wire ohms functions can be used for a variety of purposes in addition to  
measuring resistance, as the following applications show.  
4-10. TESTING DIODES  
The 2-wire ohms function can also be used to test diodes.  
1. Select the 2-wire ohms function and the 2 krange.  
2. Measure the resistance of the diode. If the diode is good, when forward-biased it will  
measure about 0.6 kto 0.7 kfor silicon (0.25 kto 0.3 kfor germanium), and  
when reverse-biased it will cause the 8842A to indicate overrange. (The forward-  
biased reading depends upon the range used.)  
The 2 krange is used because its 1 mA test current provides a typical operating point,  
and its 2V full-scale voltage is sufficient to turn on most diodes (even two diodes in  
series).  
4-11. TESTING ELECTROLYTIC CAPACITORS  
The 2-wire ohms function can also give a rough test of an electrolytic capacitors leakage  
and dielectric absorption. This test works well for capacitors 0.5 µF and larger.  
1. Select the 2-wire ohms function, the 2 krange, and the medium reading rate.  
2. Connect the test leads to the capacitor (with the INPUT HI lead to the + lead and the  
INPUT LO lead to the - lead). The 8842A attempts to charge it to the open-circuit  
voltage of the 2 krange (about 6V).  
3. Disconnect the + test lead.  
4. To test for leakage, select the VDC function and the 20V range (leave the 8842A in  
the medium reading rate), and measure the voltage that was stored on the capacitor  
during step 2.  
a. If the capacitor is good, the voltage across the capacitor will be about 6V, and  
will be relatively stable.  
b. If the capacitor is leaky, the voltage across the capacitor will be much less than  
6V, and the voltage will be decreasing. The rate of change depends on how leaky  
the capacitor is.  
c. With some electrolytic capacitors, the reading will increase. This usually  
indicates the capacitor is defective.  
5. To test the capacitors dielectric absorption, briefly short the capacitors leads together  
and then measure the voltage across the capacitor.  
a. If the dielectric is good (i.e., has low dielectric absorption), the voltage across the  
capacitor will be nearly zero volts.  
b. If the dielectric is poor (i.e., has high dielectric absorption), the voltage across the  
capacitor will be significantly above zero.  
4-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
4-12. A PRECISION CURRENT SOURCE  
The ohms current source (the internal current source used in the ohms functions) makes a  
useful troubleshooting tool in itself. It has excellent linearity and temperature stability. Its  
compliance voltage is typically 5V in the lower five ohms ranges, and 12V in the upper  
two ohms ranges. The inputs are protected against accidental applications of voltage up to  
300V rms.  
To use the ohms current source, connect the test leads to the HI and LO INPUTS, and  
select either the 2-wire or 4-wire ohms function. Press the range buttons to select any of  
the current levels shown in Table 4-1.  
The ohms current source can be used to troubleshoot circuits by injecting current into  
selected nodes, forcing the circuits to be in a specific test state. For example, the ohms  
current source can be used to set or modify the bias of amplifier circuits. The current  
level can be changed simply by changing range.  
The ohms current source can also be used to test mA or µA panel meters. The accuracy  
of the current source is more than enough to verify panel meters, whose accuracy is  
typically 1% to 5%. To test an analog panel meter, simply connect the current source  
across the meter movement (as though measuring its resistance). A 1 mA meter should  
show full scale when the ohms function is set on the 2 krange. The same technique also  
works with digital panel meters.  
4-13.DC CURRENT MEASUREMENT  
To get the best accuracy using the mA DC function, it is important to understand the  
concept of burden voltage error.  
When a meter is placed in series with a circuit to measure current, error can be caused by  
the small voltage drop across the meter (in this case, across the protective fuses and  
current shunt). This voltage drop is called the burden voltage, and it is highest for full-  
scale measurements. The full-scale burden voltage for the 8842A is typically less than  
1V.  
The burden voltage can present a significant error if the current source being measured is  
unregulated (i.e., not a true current source) and if the resistance of the fuse and shunt is a  
significant part of the source resistance. If burden voltage does present a significant error,  
the percentage of error can be calculated and corrected for using the formulas in Figure 4-  
5.  
4-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
REDUCING THERMAL VOLTAGES  
4
f4-05.wmf  
Figure 4-5. Burden Voltage Error Calculation  
4-14.REDUCING THERMAL VOLTAGES  
When making very low-level dc measurements, thermal voltages can present an  
additional source of error. Thermal voltages are the thermovoltaic potentials generated at  
the junction between dissimilar metals. Thermal voltages typically occur at binding posts  
and can be greater than 10 µV.  
Thermal voltages can also cause problems in the low dc and ohms ranges, particularly in  
the 20 mV and 20ranges. Some low-value resistors are constructed with dissimilar  
metals. Just handling such resistors can cause thermal voltages large enough to introduce  
measurement errors.  
The effect of thermal voltages can be reduced by using the following techniques:  
1. Use tight connections.  
2. Use clean connections (especially free of grease and dirt).  
3. Use similar metals for connections wherever possible (e.g., copper-to-copper, gold-  
to-gold, etc.).  
4-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
4. Use caution when handling the circuit under test.  
5. Wait for the circuit to reach thermal equilibrium. (Thermal voltages are generated  
only where there is a temperature gradient.)  
4-15.AC VOLTAGE AND CURRENT MEASUREMENT  
When making precise measurements of ac voltage and current, there are several  
considerations in addition to those discussed under dc voltage and current measurement.  
These include the concepts of rms conversion, crest factor, bandwidth, and zero-input  
error.  
4-16. True RMS Measurement  
The True RMS AC Option measures the true rms value of ac voltages and currents. In  
physical terms, the rms (root-mean-square) value of a waveform is the equivalent dc  
value that causes the same amount of heat to be dissipated in a resistor. True rms  
measurement greatly simplifies the analysis of complex ac signals. Since the rms value is  
the dc equivalent of the original waveform, it provides a reliable basis for comparing  
dissimilar waveforms.  
By contrast, many meters in use today use average-responding ac converters rather than  
true rms converters. The scale factor in these meters is adjusted so that they display the  
rms value for harmonic-free sinusoids. However, if a signal is not sinusoidal, average-  
responding meters do not display correct rms readings.  
The 8842A actually derives the rms value using analog computation. This means that the  
8842A readings represent true rms values not only for harmonic-free sinusoids, but also  
for mixed frequencies, modulated signals, square waves, sawtooths, random noise,  
rectangular pulses with 10% duty cycle, etc.  
4-17. Waveform Comparison  
Figure 4-6 illustrates the relationship between ac and dc components for common  
waveforms, and compares readings for true rms meters and average-responding meters.  
For example, consider the first waveform, a 1.41421V (zero-to-peak) sine wave. Both the  
8842A and rms-calibrated average-responding meters display the correct rms reading of  
1.00000V (the dc component equals 0). However, consider the 2V (peak-to-peak) square  
wave. Both types of meter correctly measure the dc component (0V), but only the 8842A  
correctly measures the ac component (1.00000V). The average-responding meter  
measures 1.110V, which amounts to an 11% error.  
4-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
AC VOLTAGE AND CURRENT MEASUREMENT  
4
f4-06.wmf  
Figure 4-6. Waveform Comparison Chart  
Since average-responding meters have been in use for so long, you may have  
accumulated test or reference data based on them. The conversion factors in Figure 4-6  
should help you convert between the two measurement methods.  
4-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
4-18. Crest Factor  
Crest factors are useful for expressing the ability of an instrument to measure a variety of  
waveforms accurately. The crest factor of a waveform is the ratio of its peak voltage to its  
rms voltage. (For waveforms where the positive and negative half-cycles have different  
peak voltages, the more extreme peak is used in computing the crest factor.) Crest factors  
start at 1.0 for square waves (for which the peak voltage equals the rms voltage) and  
increase for more "pointed" waveforms as shown in Figure 4-7.  
f4-07.wmf  
Figure 4-7. Typical Crest Factors for Various Waveforms  
The 8842A has a full-scale crest factor limit of 3.0 for the 20V and 700V ranges, and 6.0  
for the other ranges. For full-scale input signals with a crest factor above these limits,  
dynamic range limitations can begin to cause large errors. However, as Figure 4-7 shows,  
signals with a crest factor above 3.0 are unusual.  
If you dont know the crest factor of a particular waveform but wish to know if it falls  
within the crest factor limit of the 8842A, measure the signal with both the 8842A and an  
ac-coupled oscilloscope. If the rms reading on the 8842A is 1/3 or more of the  
waveforms zero-to-peak voltage, the crest factor is 3.0 or less.  
4-19. AC-Coupled AC Measurements  
Input signals are ac-coupled in the ac functions. One of the advantages of ac coupling is  
that ac measurements can be made on power supply outputs, phone lines, etc. Ripple  
4-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
AC VOLTAGE AND CURRENT MEASUREMENT  
4
measurements, for instance, cannot be made with dc coupling. Remember, however, that  
when the 8842A measures signals with the ac functions, the reading on the display does  
not include the dc component (if one exists). For example, consider Figure 4-8, which  
shows a simple ac signal riding on a dc level. The VAC function would measure the ac  
component only.  
f4-08.wmf  
Figure 4-8. Combined AC and DC Measurement  
4-20. Combined AC and DC Measurements  
The 8842A can be used to evaluate the true rms value of waveforms such as the one  
shown in Figure 4-8, which includes both ac and dc components. First, measure the rms  
value of the ac component using the VAC function. Next, measure the dc component  
using the VDC function. Finally, calculate the total rms value as follows:  
2
V
RMS  
= VAC 2 +VDC  
4-21. Bandwidth  
Bandwidth defines the range of frequencies to which an instrument can respond  
accurately. The accuracy of the 8842A is specified for sinusoidal waveforms up to 100  
kHz, or for nonsinusoidal waveforms with frequency components up to 100 kHz. The  
small-signal bandwidth (the frequency at which the response is 3 dB down) is typically  
around 300 kHz.  
For signals with components greater than 100 kHz, the measurement accuracy is reduced  
because of frequency bandwidth and slew-rate limitations. Because of this, accuracy may  
be reduced when measuring signals with fast rise times, such as high-frequency square  
waves or switching supply waveforms. As a rule of thumb, an ac voltage input signal is  
within the bandwidth limitations if the rise time is longer than 2 us, and within the slew-  
rate limitations if the input slew rate is slower than (1V/µs)x(full scale of range).  
4-22. Zero-Input VAC Error  
If the 8842A input terminals are shorted while the VAC function is selected, the 8842A  
displays a non-zero reading (typically less than 80 digits in the highest four ranges, and  
less than 300 digits in the 200 mV range). Such readings are due to random noise  
combined with the inherent nonlinear response of computing-type rms converters to very  
small input signals.  
4-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
The zero-input error is quickly reduced when the input is increased. The rms converter  
error (a dc error) and the internally generated noise (a random ac error) are both  
uncorrelated with the input signal. Therefore, when a signal is applied, the resulting  
reading is not the simple addition of the signal and the zero-input error, but the square  
root of the sum of their squares. This reduces the effect of the error, as shown in the  
example in Figure 4-9.  
f4-09.wmf  
Figure 4-9. Reduction of Zero-Input Error  
As long as the 8842A reading is 1,000 counts or more, readings will still be within  
specified accuracy.  
4-23.MAKING ACCURATE MEASUREMENTS ON THE 20 mV  
AND 20RANGES  
NOTE  
When making low-level (uV) measurements after a large step signal has  
been applied to the inputs, allow sufficient time for thermal emfs and other  
system-related sources of error to settle before taking readings.  
The 20 mV dc and 20ranges are the 8842As most sensitive ranges. For that reason,  
they are also the most susceptible to error from electrical noise, thermal voltages, and (for  
resistance measurements) test lead resistance. You can minimize these sources of error by  
using good measurement practices.  
The most common source of error is electrical noise. Typical sources of noise include  
electrostatic noise, inductive pickup noise, radio frequency noise, power line noise and  
noise generated by ground loop currents. Noise pickup can be minimized by properly  
shielding the test leads between the 8842A and the signal source.  
For voltage measurements in most system applications, where common-mode voltages  
are typically present, connect the test lead shielding to the 8842A INPUT LO terminal as  
shown in Figure 4-10. This configuration minimizes the error caused by current that  
would flow in the leads due to common-mode voltages between the measurement and  
stimulus points and instrument ground. The 8842As INPUT LO terminal is internally  
connected to the instruments internal guard, which provides a shield between the  
instruments ground and its sensitive analog circuits. The 8842As analog circuits are  
isolated from its digital circuits by an electrostatically shielded transformer, whose shield  
is also connected to the guard.  
4-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Measurement Tutorial  
MAKING ACCURATE HIGH-RESISTANCE MEASUREMENTS  
4
f4-10.wmf  
Figure 4-10. Shielding for Low Voltage Measurements  
For low-level resistance measurements, connect the test lead shielding as shown in Figure  
4-11. Use the 4-wire ohms function to minimize the error caused by the resistance of the  
test leads.  
f4-11.wmf  
Figure 4-11. Shielding for Low Resistance Measurements  
Errors due to thermal voltages should also be considered when making low-level voltage  
or resistance measurements. Techniques for reducing thermal voltages are presented  
earlier in this section.  
4-24.MAKING ACCURATE HIGH-RESISTANCE  
MEASUREMENTS  
When high resistances are measured (typically 1 Mor greater), leakage resistance at the  
test circuit can provide enough shunt resistance to degrade the accuracy of the  
measurement (see Figure 4-12). To minimize leakage resistance, watch out for  
contamination, high humidity, poor-quality interconnections, and poor-quality insulation  
of the stand-offs on which test resistors are mounted.  
High-resistance measurements are also susceptible to error from electrical noise pickup.  
For accurate measurements, use short test leads and enclose the test leads and test circuit  
in a proper shield that is connected to the 8842As INPUT LO terminal. Pickup of slowly  
fluctuating noise can also be reduced by using the OFFSET feature as described in  
paragraph 4-7.  
4-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f4-12.wmf  
Figure 4-12. Leakage Resistance in High Resistance Measurement  
4-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 5  
Theory of Operation  
Title  
Page  
5-1.  
5-2.  
5-3.  
5-4.  
5-5.  
5-6.  
5-7.  
5-8.  
5-9.  
5-10.  
5-11.  
5-12.  
5-13.  
INTRODUCTION................................................................................ 5-3  
OVERALL FUNCTIONAL DESCRIPTION...................................... 5-3  
DETAILED CIRCUIT DESCRIPTION .............................................. 5-4  
DC SCALING ...................................................................................... 5-4  
VDC Scaling .................................................................................... 5-6  
VDC Protection................................................................................ 5-7  
mA DC Scaling ................................................................................ 5-7  
Analog Filter .................................................................................... 5-7  
TRACK/HOLD CIRCUIT ................................................................... 5-8  
Track Configuration......................................................................... 5-10  
Settling Configuration...................................................................... 5-11  
Hold Configuration .......................................................................... 5-11  
Pre-Charge Configuration................................................................ 5-11  
5-14. PRECISION VOLTAGE REFERENCE.............................................. 5-11  
5-15. OHMS CURRENT SOURCE .............................................................. 5-12  
5-16. OHMS PROTECTION......................................................................... 5-13  
5-17. OHMS FUNCTIONS ........................................................................... 5-13  
5-18.  
5-19.  
2-Wire Ohms.................................................................................... 5-13  
4-Wire Ohms.................................................................................... 5-15  
5-20. A/D CONVERTER .............................................................................. 5-15  
5-21.  
5-22.  
5-23.  
5-24.  
Timing/Data Control........................................................................ 5-17  
Precision DAC ................................................................................. 5-17  
A/D Amplifier.................................................................................. 5-18  
Bootstrap Supplies ........................................................................... 5-18  
5-25. DISPLAY ............................................................................................. 5-18  
5-26. KEYBOARD........................................................................................ 5-19  
5-27. DIGITAL CONTROLLER................................................................... 5-19  
5-28.  
5-29.  
5-30.  
5-31.  
5-32.  
5-33.  
5-34.  
In-Guard Microcomputer ................................................................. 5-21  
Function and Range Control ............................................................ 5-23  
A/D Control and Computation......................................................... 5-23  
Calibration Correction...................................................................... 5-23  
Keyboard/Display Control ............................................................... 5-24  
Troubleshooting Modes ................................................................... 5-24  
Guard-Crossing Communication ..................................................... 5-24  
5-35. GUARD CROSSING ........................................................................... 5-24  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-36. POWER SUPPLY ................................................................................ 5-25  
5-37. IEEE-488 INTERFACE (OPTION -05)............................................... 5-26  
5-38.  
5-39.  
5-40.  
5-41.  
5-42.  
Out-Guard Microcomputer............................................................... 5-26  
Guard Crossing................................................................................. 5-26  
Bus Interface Circuitry..................................................................... 5-26  
Signal Conditioning ......................................................................... 5-27  
IEEE-488 Interface Power Supply................................................... 5-27  
5-43. TRUE RMS AC (OPTION -09)........................................................... 5-27  
5-44.  
5-45.  
5-46.  
5-47.  
VAC Scaling .................................................................................... 5-27  
mA AC Scaling ................................................................................ 5-28  
Frequency Response Trimming ....................................................... 5-28  
True RMS AC-to-DC Conversion ................................................... 5-28  
5-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
INTRODUCTION  
5
5-1. INTRODUCTION  
This section presents an overall functional description of the 8842A, followed by a  
detailed circuit description. The descriptions are supported by simplified schematics in  
text and by the complete schematics in Section 10.  
5-2. OVERALL FUNCTIONAL DESCRIPTION  
A functional block diagram of the 8842A is shown in Figure 5-1. The basic signal path  
flows from left to right across the center of the page. The input is sensed at the input  
terminals, scaled, directed through the Track/Hold circuit, converted into digital  
representation by the Analog-to-Digital (A/D) Converter, processed by the Digital  
Controller, and sent to the display.  
The DC Scaling circuit, which constitutes the "front end" of the instrument, has two  
major functions. First, it senses the input and produces an equivalent dc voltage for all  
functions except VAC and mA AC. (AC inputs are converted to a dc voltage by the True  
RMS AC Option.) Resistances are sensed as a dc voltage using a known test current from  
the Ohms Current Source. A dc current input is converted to a dc voltage by a precision  
current shunt.  
Second, the DC Scaling circuit scales the equivalent dc voltages (for in-range inputs) to  
within the input range of the A/D Converter (+/-2V). In addition, the DC Scaling circuit  
provides input protection and provides analog filtering for certain ranges and reading  
rates. (AC inputs are scaled by the True RMS AC Option.)  
The Track/Hold (T/H) circuit samples the scaled dc voltage and presents the A/D  
Converter with a voltage that is constant for the input portion of each A/D conversion  
cycle. The T/H circuit also provides additional scaling for certain ranges.  
The Digital Controller controls the operation of virtually every part of the 8842A. It reads  
the front panel keyboard, configures the instrument for each function and range, triggers  
the A/D Converter, calculates the result of each A/D conversion cycle, averages A/D  
samples, controls the display, and communicates with the IEEE-488 Interface Option via  
the Guard Crossing circuit. The heart of the Digital Controller is the In-Guard  
Microcomputer (µC).  
The Guard Crossing circuit permits serial asynchronous communication between the  
Digital Controller and the IEEE-488 Interface Option, while isolating the two circuits  
electrically. Whereas the in-guard power supply floats with the voltage at the INPUT LO  
terminal, the IEEE-488 Interface Option operates with reference to earth ground. The  
"guard" is the isolation between the in-guard and out-guard circuits.  
The Power Supply provides supply voltages to all parts of the instrument. The Precision  
Voltage Reference provides precise reference voltages for the A/D Converter and the  
Ohms Current Source.  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-3. DETAILED CIRCUIT DESCRIPTION  
The following paragraphs give a detailed circuit description of each of the functional  
blocks in Figure 5-1. For clarity, measurement ranges are referred to as r1, r2, r3, etc.,  
where r1 is the lowest possible range, r2 the next higher range, and so on. Pins are  
designated by the respective integrated circuit (e.g., U101-7 for U101 pin 7).  
5-4. DC SCALING  
The DC Scaling circuit scales all in-range dc inputs so that the output of the Track/Hold  
(T/H) amplifier (U307) is within +/-2V. In addition, the DC Scaling circuit provides input  
protection and analog filtering. Additional scaling is provided by the the T/H Amplifier.  
The following paragraphs describe the configuration of the DC Scaling circuit in the  
DCV and mA DC functions and also describe the analog filter. The ohms functions are  
described under a later heading because the T/H Amplifier provides additional input  
switching for these functions.  
5-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
DC SCALING  
5
f5-01.wmf  
Figure 5-1. Overall Functional Block Diagram  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-5.  
VDC Scaling  
Scaling is performed in the VDC function by two precision resistors networks (Z301 and  
Z302). These components are configured by relay K301, switching transistor Q311, and  
quad analog switches U302A and U301B to provide the correct scaling for each range.  
Voltage follower U306 provides high input impedance for the 20V dc range. A simplified  
schematic and a switch state table for the VDC function are shown in Figure 5-2.  
f5-02.wmf  
Figure 5-2. DC Scaling (VDC and mA DC)  
5-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
DC SCALING  
5
In the 20 mV, 200 mV and and 2V ranges, the input voltage is applied directly to the T/H  
Amplifier via Q310, Q311, and U301B. In the 20 mV range, the T/H Amplifier has a gain  
of 100; in the 200 mV range, the T/H Amplifier has a gain of 10; in all other dc voltage  
ranges, the T/H Amplifier has a gain of 1.  
In the 20V range, the input voltage is buffered by unity-gain amplifier U306, and divided  
by 10 by Z301. To allow U306 to handle +/-20V inputs, its power supplies are  
"bootstrapped" by Q305 and Q306, so that the output voltage of U306 determines the  
midpoint of its supply voltages. The positive supply is approximately 6.2V above the  
input and the negative supply is approximately 6.2V below.  
In the 200V and 1000V ranges, K301 is de-energized and the input voltage is divided by  
100 by Z302. In the 200V range, the reduced input voltage is then applied directly to the  
T/H Amplifier as in the 2V range. In the 1000V range, the reduced input voltage is  
buffered by U306 and divided by 10 as in the 20V range.  
5-6.  
VDC Protection  
Input protection for the VDC function is provided by a 1K, fusible resistor (R309), four  
metal-oxide varistors (MOVs) (RV301, RV402, RV403, and RV404), and additional  
protection resistors and clamp circuits.  
WARNING  
TO AVOID INJURY OR EQUIPMENT DAMAGE, USE EXACT  
REPLACEMENT PARTS FOR ALL PROTECTION COMPONENTS.  
In all dc voltage ranges, voltage transients greater than 1560V are clamped by the MOVs.  
Extreme overvoltage conditions cause R309 to fail open-circuit.  
R309 is followed either by a 99 k, 10W resistor network (Z304) in the 20 mV, 200 mV,  
2V, and 20V ranges, or by 10 M(Z302) to ground in the 200V and 1000V ranges. Z304  
provides current limiting in extreme overvoltage conditions in the 20 mV, 200 mV, 2V,  
and 20V ranges. The non-inverting input of U306 is clamped to +/-25V by Q307 and  
Q308.  
5-7.  
5-8.  
mA DC Scaling  
In the mA DC function, the unknown current causes a voltage drop across current shunt  
R319. This voltage drop is then measured as in the VDC function. The DC Scaling circuit  
is configured as shown by the simplified switch table in Figure 5-2.  
Analog Filter  
The three-pole, low-pass analog filter (U304) has a Bessel response with corner  
frequency at 7 Hz, giving approximately 50 dB of rejection at 50 Hz. The filter is used  
for the slow reading rate and is used in the VDC ranges and lowest three ohms ranges.  
The filter is also used in the 20mV DC, 20, and 200 mA DC ranges when in the  
medium reading rate. The filter is switched into the input signal path by Q304 (Figure 5-  
2). In some ranges and functions, additional filtering is provided by Q317 and C314.  
5-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-9. TRACK/HOLD CIRCUIT  
The Track/Hold (T/H) circuit presents a stable voltage to the A/D Converter during the  
input period of the A/D conversion cycle. The circuit also provides a gain of 100 in the  
20 mV, 20and 200 mA ranges, and a gain of 10 in the 200 mV dc, 200, and 2000 mA  
dc ranges.  
The T/H circuit consists of the T/H Amplifier (Figure 5-3), T/H capacitor C308, quad  
analog switches U301, U302, and U303, and associated components. As shown in Figure  
5-3, the T/H Amplifier functions as an op amp, with Q314 supplying additional gain. In  
subsequent figures, the T/H Amplifier is represented as a single op amp.  
f5-03.wmf  
Figure 5-3. Track/Hold Amplifier  
The circuit operates by cycling between the track, settling, hold, and precharge  
configurations shown in Figure 5-4. The In-Guard µC selects a particular settling and  
hold configuration for each function and range, and suppresses the precharge  
configuration for certain ranges. This control is achieved by latching function and range  
information in U301, U302, and U303.  
Basic timing for the T/H circuit is provided by the A/D Converter over clock lines PC,  
HD1, TR1, and TR2. (See the timing diagram in Figure 5-5, top.) The T/H cycle is  
initiated when the In-Guard µC pulls line TR low.  
5-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
TRACK/HOLD CIRCUIT  
5
f5-04.wmf  
Figure 5-4. Track/Hold Circuit Configurations  
5-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f5-05.wmf  
Figure 5-5. Timing Diagram for One A/D Cycle  
5-10. Track Configuration  
In the track configuration (Figure 5-4A), the T/H circuit functions as a non-inverting  
buffer. The voltage on C308 tracks the scaled dc input voltage.  
5-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
PRECISION VOLTAGE REFERENCE  
5
5-11. Settling Configuration  
The circuit assumes a settling configuration between the track and hold configurations.  
The circuit assumes the configuration in Figure 5-4B for unity gain and the configuration  
in Figure 5-4C for gain of 10.  
During this time the DC Scaling circuit is still connected to the T/H amp. However,  
changes in the input do not affect the value to be measured, which is stored on C308.  
5-12. Hold Configuration  
The X1 hold configuration (Figure 5-4D) is used for all VDC ranges except r1 and for all  
ohms ranges except r1. The output of U307 is the negative of the input voltage.  
The X10 hold configuration (Figure 5-4E) is used for the mA DC function, the 200 mV  
dc range, and the 200range, and provides a gain of 10.  
5-13. Pre-Charge Configuration  
The pre-charge configuration (Figure 5-4F) occurs after the hold configuration in VDC  
ranges r1, r2, and r4, and ohms ranges r1, r2, r3, and r4. U306 is connected as a buffer to  
charge stray capacitance at the non-inverting input of the T/H Amplifier. The pre-charge  
configuration is not used in any other ranges.  
5-14. PRECISION VOLTAGE REFERENCE  
The Precision Voltage Reference (Figure 5-6) provides precise reference voltages of -  
7.00000 and +7.00000. The reference element is a reference amplifier (ref amp). The  
nominal ref amp voltage is 6.5V.  
f5-06.wmf  
Figure 5-6. Precision Voltage Reference  
5-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Resistor R701, precision resistor network Z701, and transistor/zener diode combination  
U701 are produced as a matched set so that the output of U702A is precisely -7.00000V.  
This output is remotely sensed at the pins of the custom A/D IC (U101). Diode CR701  
prevents the output from going positive at power-up.  
U702B functions as an inverter to provide the +7.00000V output and to supply the  
reference amplifier. The gain of U702B is set by the two 20 kresistors in the resistor  
network Z702.  
5-15. OHMS CURRENT SOURCE  
The Ohms Current Source (Figure 5-7) provides a precise test current for the ohms  
functions. The first stage (U401, R401, and Q401) produces a precise reference current,  
using precision resistor R401 and a -7.0000V reference voltage from the Precision  
Voltage Reference.  
f5-07.wmf  
Figure 5-7. Ohms Current Source  
5-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
OHMS PROTECTION  
5
The second stage (U404, precision resistor network Z401, and analog switches U402 and  
U403) is a current amplifier whose gain is controlled by the In-Guard µC. The In-Guard  
µC sets the output current for each range by controlling U402 and U403. (See switch  
state table in Figure 5-7.)  
5-16. OHMS PROTECTION  
The Ohms Protection circuit (Q402, Q403, Q404, Q405, Q406, and Q407) clamps the  
open circuit voltage of the Ohms Current Source and provides protection for the Ohms  
Current Source.  
The circuit protects the Ohms Current Source from up to +/-300V across the INPUT  
terminals. The circuit also clamps voltage transients larger than 1560V with four MOVs  
(RV401, RV402, RV403, and RV404). In addition, a 1 k, 2W fusible wire-wound  
resistor (R410) in series with the output current path fails open-circuit under extreme  
overvoltage conditions.  
Large positive input voltages are blocked by CR402. Large negative input voltages are  
dropped equally across three high-voltage transistors (Q402, Q403, and Q404). If -300V  
is present at the collector of Q404, the voltage drops equally across Z402 so that large  
negative voltages never reach the current source.  
The circuitry associated with Q408 (R406, R407, R408, R409, Q406, Q408, and CR403)  
clamps the open-circuit voltage of the Ohms Current Source below +6.5V in the lower  
four ranges and below +13V dc in the higher two ranges. The in-guard uC turns Q408 on  
or off depending on range. In the lower four ohms ranges, Q408 is on, effectively  
shorting R409; R406 and R409 then form a voltage divider which clamps the output of  
the ohms current source below +6.5V. In the higher two ohms ranges, Q408 is off,  
including R409 in the voltage divider and clamping the output below +13V.  
5-17. OHMS FUNCTIONS  
5-18. 2-Wire Ohms  
In the 2-wire ohms function, the Ohms Current Source is connected to the INPUT HI  
terminal by ohms relay K401 (Figure 5-8). The Ohms Current Source applies a known  
current to the resistance under test, and the resulting voltage drop across the resistor is  
measured ("sensed") as in the VDC function.  
5-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f5-08.wmf  
Figure 5-8. Ohms Scaling  
5-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
A/D CONVERTER  
5
The voltage sensed at the INPUT terminals is scaled as shown by the simplified switch  
table in Figure 5-8. (Refer to the track period of the track/hold cycle, during which the  
scaled input voltage is sampled.)  
In the lower five ranges, the full scale input voltage to the A/D Converter is 2V.  
However, in the 2000 kand 20 Mranges, the full-scale input voltage to the A/D  
Converter is +1V; the in-guard uC completes the scaling by multiplying the A/D result by  
2.  
5-19. 4-Wire Ohms  
In the 4-wire ohms function, the Ohms Current Source is connected to the INPUT HI  
terminal by ohms relay K401 as in 2-wire ohms (Figure 5-8). The Ohms Current Source  
applies a known current to the resistance under test through the INPUT HI and INPUT  
LO leads. The resulting voltage drop across the resistor is measured by the SENSE HI  
and SENSE LO leads.  
The voltage at the SENSE HI terminal is connected to the DC Scaling circuit by Q303  
(Figure 5-8). The voltage is then scaled exactly as in the 2-wire ohms function. (Refer to  
the track period in the switch table in Figure 5-8.) Q310 is turned off to isolate the  
SENSE HI terminal from the INPUT HI terminal.  
Additional input switching occurs during the hold period of the track/hold cycle. (Refer  
to the hold period in the switch table in Figure 5-8.) In ranges r1 through r4, and r8, the  
SENSE LO terminal is switched into the dc input path by U301D, and the INPUT LO  
terminal is switched out of the dc input path by U301C. This has the effect of measuring  
the SENSE HI terminal with respect to the SENSE LO terminal.  
In ranges r5 and r6, the SENSE LO and INPUT LO terminals are both switched into the  
dc input path by U301C and U301D during the hold period. This has the effect of  
measuring the SENSE HI terminal with respect to INPUT LO terminal rather than  
SENSE LO. Although the resistance of the INPUT LO lead is in series with the unknown  
resistance, accuracy is not affected as long as the resistance of the lead is less than 10in  
the 2000 krange and less than 100in the 20 Mrange.  
5-20. A/D CONVERTER  
The Analog-to-Digital (A/D) Converter (Figure 5-9) uses Flukes patented recirculating  
remainder technique. An input voltage (Vin) is compared to the output of the precision  
Digital-to-Analog Converter (DAC). The output of the A/D Amplifier, connected as a  
comparator, is monitored to indicate when the DAC output is larger than the input  
voltage.  
5-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f5-09.wmf  
Figure 5-9. Analog-to-Digital Converter  
The conversion process is broken up into an autozero period followed by five  
measurement intervals. (A timing diagram is shown in Figure 5-5.) Six bits of the final  
A/D sample are obtained during each interval.  
During the first compare period (shown in Figure 5-9), the A/D Converter determines the  
value of the scaled input voltage (Vin) by comparing Vin to the output of the DAC. Each  
of the DAC bit-switches is tried in sequence and kept or rejected (left closed or reopened)  
depending on the output polarity of the A/D Amplifier, which is configured as a  
comparator. This process produces a string of six bits which is stored in the Timing/Data  
Control circuit (the digital portion of U101).  
During the following remainder-store period (Figure 5-10), the difference between the  
Vin and the DAC output is multiplied by 16 by the A/D Amplifier and stored on  
capacitor C102. During subsequent compare and remainder-store periods, the remainder  
voltage is connected to the input of U103 and is resolved to six bits; the remainder  
voltage (multiplied by 16) is stored alternately on capacitor C102 and C103. Each of the  
five compare periods thus produces a six-bit nibble which is stored in the Timing/Data  
Control circuit.  
5-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
A/D CONVERTER  
5
f5-10.wmf  
Figure 5-10. First Remainder-Store Period  
This five-interval process thus generates five nibbles which are processed by the In-  
Guard µC to produce one A/D sample. After the fifth nibble is generated, U101 interrupts  
the In-Guard µC over line INT. The In-Guard µC then pulls line CS7 low five times,  
causing U101 to send the µC the five (six-bit) nibbles one-at-a-time over lines AD0-  
AD5. The In-Guard µC then weights each nibble 1/16 of the value of the previous  
number and calculates the input voltage.  
The hardware for the A/D Converter has four major sections: Timing/Data Control,  
Precision DAC, A/D Amplifier, and bootstrap supplies.  
5-21. Timing/Data Control  
The Timing/Data Control circuit (the digital portion of U101) times and controls the A/D  
Converter by manipulating the switches in the A/D Amplifier and the bit-switches in the  
Precision DAC. An A/D conversion cycle is triggered by the falling edge of line TR from  
the In-Guard µC. Once triggered, the A/D Converter (under control by U101) generates  
the five 6-bit nibbles without further interaction with the In-Guard µC.  
The Timing/Data Control circuit also provides a watch-dog timer (line RES not) which  
resets the In-Guard µC in case normal program execution is interrupted. If the timer  
senses inactivity on line CS7 for longer than 1.5 seconds, it resets the In-Guard µC by  
pulling RES not low.  
The Timing/Data Control circuit is supplied with a fixed-rate 8 MHz clock and provides a  
1 MHz output clock for the Keyboard/Display Interface (U212). In addition, four output  
lines (PC, HD1 not) TR1, and TR2) provide control signals for the Track/Hold circuit.  
5-22. Precision DAC  
The Precision Digital-to-Analog Converter (DAC) is composed of DAC Amplifier  
U102B and a binary ladder network, which consists of resistors in Z101 and digitally  
controlled analog bit-switches contained in U101.  
5-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
The bit-switches determine the output voltage of U102B by controlling the binary ladder  
network. The gain of U102B is set by the ratio of a precision feedback resistor (Z101-7, -  
8) and the equivalent output resistance of the ladder network.  
5-23. A/D Amplifier  
The A/D Amplifier is composed of a comparator/amplifier (U103), two remainder-  
storage capacitors (C103 and C102), an autozero storage capacitor (C101), and several  
digitally controlled analog switches contained in U101.  
The A/D Amplifier has three modes of operation: autozero mode, where any offsets in  
the A/D input are stored on C101 so as to be cancelled later; compare mode, where the  
A/D input is compared to the DAC output; and remainder-store mode, where U103  
amplifies and stores the difference between the A/D input and the DAC output on one of  
the two remainder-storage capacitors (C102 or C103). The autozero mode is shown in  
Figure 5-11. The other modes are shown in Figures 5-9 and 5-10.  
f5-11.wmf  
Figure 5-11. Autozero Period  
5-24. Bootstrap Supplies  
The bootstrap supplies are composed of U102A, Q101, Q102, CR103, CR104, and  
associated components. The bootstrap supplies enhance the gain accuracy of U103.  
During compare periods, the bootstrap supplies limit the output of U103 to minimize the  
time it takes to recover from being driven to a supply rail. Both functions are achieved by  
manipulating the supplies of U103 (BS1 and BS2).  
5-25. DISPLAY  
The vacuum fluorescent display is similar to a vacuum tube, containing eight control  
grids and 69 phosphor-coated plates which form the display segments and annunciators.  
(See Figure 5-12.) The filament voltage is 4.5V ac, with a +5V dc bias. Each plate is  
controlled by a G line and a P line. The G lines go to the control grids, and the P lines go  
to the plates.  
5-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
KEYBOARD  
5
f5-12.wmf  
Figure 5-12. Vacuum Fluorescent Display  
The Digital Controller sequentially enables the G lines by applying +30V dc (nominal).  
When a G line is enabled, electrons flow from the filament to the enabled grid. If a P line  
is enabled (i.e., raised to a nominal +30V dc by the Digital Controller), the electrons  
continue past the grid and strike the respective plate, causing it to glow.  
5-26. KEYBOARD  
The keyboard consists of a silicone-rubber switch matrix located over metalized epoxy  
contacts on the printed wire board. Each button contains a conductive pad that shorts two  
contacts when pressed.  
5-27. DIGITAL CONTROLLER  
The Digital Controller (Figure 5-13) consists of the In-Guard µC (U202), External  
Program Memory (U222), Calibration Memory (U220), Keyboard/Display Interface, and  
associated components.  
5-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f5-13.wmf  
Figure 5-13. Digital Controller Block Diagram  
5-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
DIGITAL CONTROLLER  
5
5-28. In-Guard Microcomputer  
The In-Guard Microcomputer (µC) is a single-chip Z8 microcomputer containing 4K  
bytes of ROM, 144 bytes of RAM, a UART, and four 8-bit I/O ports. It communicates  
with the rest of the instrument via the internal bus and dedicated I/O lines. The In-Guard  
µC is reset when pin 6 is pulled low either by C204 at power-up or by the watch-dog  
timer in the custom A/D IC (U101). Pin 6 is tied to +5V through a 100 kresistor inside  
the µC.  
All internal bus communication is memory-mapped. Each component that sends or  
receives data on the bus has a unique address or range of addresses. The internal bus  
consists of lines AD0-AD7 and A8-A11. Lines AD0-AD7 are time-multiplexed to carry  
both the least-significant address byte and the data. Lines A8-A11 carry the most-  
significant bits of the address. The µC writes to and reads from the internal bus according  
to the read and write cycles shown in Figure 5-14. During either cycle, the address strobe  
(AS) changes from low to high when an address is valid, and the data strobe (DS)  
changes from low to high when the data is valid.  
The address strobe latches the address on AD0-AD7 into U219 which then provides static  
address inputs for those devices that need it while data is on the bus. The data memory  
line (DM) divides the address space between program memory (U222) and data memory  
(all other devices on the bus). The data memory address space is further divided between  
the calibration memory (U220) and the remaining devices by A11. The addresses of the  
remaining devices are decoded from A8-A10 by U208, which combines the address with  
the data strobe (DS) to provide a chip select (CS0, CS2, CS3, CS4, or CS7) for each  
device.  
The In-Guard µC performs the following functions: range and function control; A/D  
control and computation; calibration corrections; keyboard/display control; serial  
communication with the IEEE-488 Interface; and diagnostic self-testing and  
troubleshooting.  
5-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f5-14.wmf  
Figure 5-14. Read/Write Timing Diagrams for Internal Bus  
5-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
DIGITAL CONTROLLER  
5
5-29. Function and Range Control  
The In-Guard µC configures the DC Scaling circuit, the Track/Hold circuit, and the  
Ohms Current Source to provide the proper input switching, scaling, and filtering for  
each function, range, and reading rate. It does this by controlling dedicated output lines  
which control relays and FET switches, and by sending configuration codes out on the  
bus. The quad analog switches (U301, U302, U303, U402, and U403) latch the  
configuration codes and perform any level-shifting needed to control their internal  
MOSFET switches. Some of the switches require dynamic timing signals from the  
custom A/D IC (U101); these signals are combined appropriately in the quad analog  
switches with the configuration codes.  
5-30. A/D Control and Computation  
The In-Guard µC initiates each A/D sample by pulling line TR low. When the µC is  
reset, it senses the power line frequency on line FREQ REF. The µC then sets its internal  
timer so that the A/D sample rate is as shown in Table 5-1.  
The number of readings per second for the slow and medium rates are chosen to provide  
rejection of input signals that are at the line frequencies.  
Table 5-1. Sample Rates and Reading Rates  
SLOW  
Samples  
MEDIUM  
FAST  
POWER LINE  
FREQUENCY  
Samples  
per Sec  
Samples  
per Sec  
Samples  
per  
Samples  
per Sec  
Samples  
per  
per  
Reading  
Reading  
Reading  
50 Hz  
60 Hz  
400 Hz  
66.67  
80  
32  
32  
32  
66.67  
4
4
4
100  
1
1
1
80  
100  
100  
76.19  
76.19  
The custom A/D IC (U101) generates five 6-bit numbers after each trigger from the µC  
and then pulls INT low, telling the µC that data is ready. The µC reads the five 6-bit  
numbers over the bus (CS7 pulses low five times for five read cycles) and computes the  
value of the A/D sample using calibration constants. The µC averages the appropriate  
number of samples for one reading, which is then sent to the keyboard/display interface  
for display.  
For example, with a 60-Hz power-line frequency, an externally triggered reading in the  
slow reading rate would cause the µC to send 32 pulses on TR at an 80 Hz rate. The 32  
A/D samples would be calibrated and averaged by the µC and sent for display. With  
internal triggering, the A/D runs continuously at 80 samples per second with a reading  
being sent to the display every 32 samples.  
5-31. Calibration Correction  
The calibration constants used by the In-Guard µC in computing each reading are stored  
in the EEROM (electronically erasable read-only memory) Calibration Memory (U220).  
The front panel CAL ENABLE switch protects the EEROM from accidental writes.  
5-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-32. Keyboard/Display Control  
Keyboard/Display Controller U212 communicates with the In-Guard µC over the internal  
bus. During a µC write cycle, address line A0 tells U212 whether to consider data being  
sent by the µC as configuration commands or as display data. Display data is stored in the  
Keyboard/Display Controller, which automatically scans the display. The  
Keyboard/Display Controller selects one of eight grids using decoder U213 and buffer  
U215. The numeric display data is decoded from BCD to 7-segment by decoder U216  
and buffered by U217. Additional annunciator data is buffered by U218.  
The Keyboard/Display Controller is reset by the µC whenever the µC is reset. It receives  
a 1-MHz clock signal from the custom A/D IC (U101), which uses the µC 8-MHz crystal  
for its clock input.  
The Keyboard/Display Controller scans the keyboard, sensing pressed buttons on lines  
RL0-RL7. It sends an interrupt to the µC via line KEYINT whenever a front panel button  
is pressed. The µC then reads the keycode from the Keyboard/Display Controller. (The  
status of the FRONT/REAR switch is sensed separately by line F/R SENSE.)  
5-33. Troubleshooting Modes  
In addition to running the diagnostic self-tests, the In-Guard µC has a troubleshooting  
mode which aids in finding digital hardware problems. After the uC is reset, it senses the  
relay control lines (U202-35 through U202-38) as inputs. If line U202-38 (TP205) is  
shorted to ground, the µC goes into the troubleshooting mode. (U201 provides internal  
pull-up.) The troubleshooting mode is described in detail in the Maintenance section.  
5-34. Guard-Crossing Communication  
The In-Guard µC contains a UART (universal asynchronous receiver transmitter) which  
it uses to communicate across the guard to the IEEE-488 Interface. The transmission  
speed is 62,500 bits per second.  
5-35. GUARD CROSSING  
The Guard Crossing consists of two identical circuits, each of which transmits data in one  
direction across the guard isolation between the Main Printed Circuit Assembly and the  
IEEE-488 Interface. One circuit is shown in Figure 5-15; the other circuit works  
identically. A portion of each circuit is contained in the IEEE-488 Interface.  
f5-15.wmf  
Figure 5-15. Guard Crossing Circuit  
5-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
POWER SUPPLY  
5
The circuit in Figure 5-15 has two stable states, corresponding to output high (+5V) and  
output low (0V). If the output is high, the voltage present at the non-inverting input of op  
amp A is approximately +140 mV.  
Since the inputs to op amps A and B are inverted, their outputs are always in opposite  
states. If the output of A is high, the output of B is low, forcing the inverting input of A  
(and the non-inverting input of B) to ground, hence reinforcing the existing state. The  
situation is analogous if the output of A is low.  
A positive-going transition at the input causes a positive pulse at the non-inverting input  
of A, and a corresponding negative pulse at the inverting input of A. If the output is high  
to start with (with the non-inverting input of A raised 140 mV above its inverting input),  
these pulses reinforce the existing state (raising the non-inverting input and lowering the  
inverting input). If, however, the output is low to start with, the positive pulse (which is  
greater than 140 mV) raises the non-inverting input of A above its inverting input,  
switching the output to the high state. The situation is analogous for a negative-going  
input transition.  
5-36. POWER SUPPLY  
The Power Supply provides the following in-guard outputs: +/-30V, +/-15V, -6.2V,  
+7.5V, +5V, -5V, and -8.2V dc; and 4.5V ac. The Power Supply also provides a 16V ac  
center-tapped out-guard output.  
Input line voltage is directed to the primary transformer winding through fuse F601, the  
front panel POWER switch, and the rear panel LINE SET switches. Metal oxide varistor  
RV601 clamps line transients at about 390V. The LINE SET switches configure the  
Power Supply to accept line power of 100, 120, 220, or 240V ac (+/-10% with a  
maximum of 250V) at 50, 60, or 400 Hz.  
AC voltage for the +5V supply is rectified by CR601 and CR602 and regulated by  
VR601. The +5V output supplies mostly logic circuits. The ac input to the +5V supply is  
sensed by the In-Guard µC (via R604, CR615, and U221-12, 13) to measure the line  
frequency.  
AC voltage for the +30V and -30V supplies is rectified by bridge network CR603,  
CR604, CR605, and CR606 and regulated by VR602 and VR605. The +30V and -30V  
outputs supply front-end buffer amp U306. In addition, the +30V output supplies the  
anodes of the vacuum fluorescent display. Zener diode CR612 supplies -6.2V to the A/D  
Converter clamps.  
AC voltage for the +15V and -15V supplies is rectified by bridge network CR608,  
CR609, CR610, and CR611 and regulated by VR603 and VR604. The +15V and -15V  
supply analog circuitry throughout the 8842A. Zener diodes CR613 and CR614 supply  
+7.5V and -8.2V to the A/D Converter, analog filter, and DC Scaling circuit.  
Secondary T601-14, 15, 16 supplies the vacuum fluorescent display filament with 4.5V  
ac. The center tap is connected to the in-guard +5V supply in order to correctly bias the  
display. An isolated secondary supplies 16V ac to the power supply on the IEEE-488  
Interface.  
Zener diode CR615 and SCR Q601 comprise a protective crow-bar circuit. If the line  
voltage exceeds the nominal value by approximately 30 percent or more, CR615  
conducts, turning on Q601, shorting out the power transformer secondary and blowing  
the line fuse. In normal operation, these components have no effect.  
5-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5-37. IEEE-488 INTERFACE (OPTION -05)  
The IEEE-488 Interface has five major parts, as shown in the block diagram in Figure 5-  
16. All components are contained in a single printed circuit assembly (PCA). Reference  
designations are numbered in the 900 series.  
f5-16.wmf  
Figure 5-16. IEEE-488 Interface Block Diagram  
5-38. Out-Guard Microcomputer  
The Out-Guard Microcomputer (µC) (U901) communicates with the IEEE-488  
talker/listener IC (U911) and the In-Guard µC (U202).  
The Out-Guard µC is similar to the In-Guard Z8 µC except that is contains 8K bytes of  
ROM and 236 bytes of RAM. For futher description of the Z8 µC, refer to the heading  
"In-Guard Microcimputer, " above.  
5-39. Guard Crossing  
The guard crossing circuit permits serial asynchronous communication between U901  
and U202 while isolating the two electrically. One-half of the guard crossing circuit is  
contained on the Main PCA; the other half is on the IEEE-488 Interface PCA. Operation  
of the guard crossing circuit is described in an earlier heading.  
5-40. Bus Interface Circuitry  
The IEEE-488 bus protocol is handled by the uPD7210 IEEE-488 talker/listener IC  
(U911). It is controlled by U901 as a memory mapped peripheral through an 8-bit data  
bus.  
Bus transceivers U912 and U913 buffer U911 from the IEEE-488 bus. They provide the  
bus with the required output drive capability and receiver impedance.  
5-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
TRUE RMS AC (OPTION -09)  
5
5-41. Signal Conditioning  
The SAMPLE COMPLETE and EXT TRIG signals (J903 and J904) are conditioned by  
U909. Diodes CR903, CR904, CR905, and CR906 and resistors R917 and R918 provide  
protection from excessive voltages. Jumpers E902 and E903 allow selection of the  
polarity of the EXT TRIG signal. (A polarity selection procedure is given in the  
Maintenance section.) The 8842A is configured in the factory so that it is triggered on the  
falling edge of the EXT TRIG signal.  
5-42. IEEE-488 Interface Power Supply  
The IEEE-488 Interface power supply circuit provides the IEEE-488 Interface PCA with  
+5V. The circuit consists of rectifying diodes CR908 and CR909, filter capacitor C910,  
and voltage regulator VR901. Power comes from transformer T605 on the Main PCA.  
U908 and associated circuitry resets the Out-Guard µC at power-up and following power-  
line voltage dropouts.  
5-43. TRUE RMS AC (OPTION -09)  
The True RMS AC circuit (Figure 5-17) performs two primary functions. First, it scales  
ac input voltages and ac current sense voltages to a range of 0V to 2V ac rms. Second, it  
converts the scaled ac voltages to an equivalent dc voltage which is then directed to the  
A/D Converter via the Track/Hold Amplifier. The True RMS AC circuit is trimmed for  
flat high-frequency response using a variable filter which is set by the High-Frequency  
AC Calibration procedure.  
f5-17.wmf  
Figure 5-17. True RMS AC Option Block Diagram  
The following paragraphs describe how these functions are performed. Components are  
laid out on a single printed circuit assembly (PCA). Component reference designators are  
numbered in the 800 series.  
5-44. VAC Scaling  
AC voltage inputs are directed from the HI INPUT terminal to the True RMS AC PCA  
through protection resistor R309 on the Main PCA. In this way, voltage transients greater  
than 1560V are clamped by MOVs (RV301, RV402, RV403, and RV404) as in the VDC  
function. With the VAC function selected, K801 is closed. The input voltage is thus  
applied to C801, which blocks dc inputs.  
U807 and resistor network Z801 provide selectable attenuation and 1 Minput  
impedance. In the upper two ranges, K802 is closed and Q806 is off, providing a gain of -  
1/500. In the lower three ranges, K802 is open and Q806 is on, shorting Z801-4 to  
ground; this configuration provides a gain of -1/5. CR801 and CR802 provide protection  
by clamping the inverting input of U807 to approximately +/-0.6V. Q805 shifts logic  
levels to control Q806.  
5-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
U806A, U806B, and a voltage divider (R804 and R805) provide gain which is selected  
for each range by the analog switches in U804. The configuration for each range is shown  
in Figure 5-17. (In this figure, the CMOS analog switches are represented by mechanical  
switches.) When U806A is not used, its non-inverting input is grounded by Q804. When  
U806B is not used, its non-inverting input is connected to the CURRENT SENSE line.  
5-45. mA AC Scaling  
The mA AC function uses the same current shunt and protection network which is used  
for dc current. In the mA AC function, Q802 switches the CURRENT SENSE line to the  
non-inverting input of U806B, which provides a gain of 10.  
5-46. Frequency Response Trimming  
The frequency response is trimmed by software calibration using a digitally controlled  
one-pole low-pass filter (R832 and a combination of C826, C827, C828, and C829). The  
analog switches in U808 configure the four capacitors to select one of 16 possible RC  
constants. The input of the digitally controlled filter is buffered by voltage follower  
U801A. The individual gain stages are also provided with fixed frequency compensation.  
5-47. True RMS AC-to-DC Conversion  
U801B buffers the input to rms converter U802. U802 computes the rms value of the  
scaled input voltage as shown in Figure 5-18. Rather than explicitly squaring and  
averaging the input, U802 uses an implicit method in which feedback is used to perform  
an equivalent analog computation.  
f5-18.wmf  
Figure 5-18. True RMS AC-to-DC Converter  
The filter averages the divider output signal. This filter consists of U809A, C813, R815,  
and the internal 25 kresistor and op amp between pins 8 and 9 of U802. The output is  
further filtered by a three-pole post-filter comprised of U809B and associated resistors  
and capacitors. This output is then switched into the Track/Hold Amplifier of the dc front  
end via U302 pins 15 and 14. The Track/Hold Amplifier is set up for unity gain on all ac  
ranges.  
5-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
static awareness  
A Message From  
Fluke Corporation  
Some semiconductors and custom IC's can be  
damaged by electrostatic discharge during  
handling. This notice explains how you can  
minimize the chances of destroying such devices  
by:  
1. Knowing that there is a problem.  
2. Leaning the guidelines for handling them.  
3. Using the procedures, packaging, and  
bench techniques that are recommended.  
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.  
3. DISCHARGE PERSONAL STATIC BEFORE  
HANDLING DEVICES. USE A HIGH RESIS-  
TANCE GROUNDING WRIST STRAP.  
1. MINIMIZE HANDLING  
2. KEEP PARTS IN ORIGINAL CONTAINERS  
UNTIL READY FOR USE.  
4. HANDLE S.S. DEVICES BY THE BODY.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. WHEN REMOVING PLUG-IN ASSEMBLIES  
HANDLE ONLY BY NON-CONDUCTIVE  
EDGES AND NEVER TOUCH OPEN EDGE  
CONNECTOR EXCEPT AT STATIC-FREE  
WORK STATION. PLACING SHORTING  
STRIPS ON EDGE CONNECTOR HELPS  
PROTECT INSTALLED S.S. DEVICES.  
5. USE STATIC SHIELDING CONTAINERS FOR  
HANDLING AND TRANSPORT.  
6. DO NOT SLIDE S.S. DEVICES OVER  
ANY SURFACE.  
9. HANDLE S.S. DEVICES ONLY AT A  
STATIC-FREE WORK STATION.  
10. ONLY ANTI-STATIC TYPE SOLDER-  
SUCKERS SHOULD BE USED.  
11. ONLY GROUNDED-TIP SOLDERING  
IRONS SHOULD BE USED.  
7. AVOID PLASTIC,VINYL AND STYROFOAM  
IN WORK AREA.  
PORTIONS REPRINTED  
WITH PERMISSION FROM TEKTRONIX INC.  
AND GERNER DYNAMICS, POMONA DIV.  
Dow Chemical  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6  
Maintenance  
Title  
Page  
6-1.  
6-2.  
6-3.  
6-4.  
6-5.  
6-6.  
6-7.  
6-8.  
INTRODUCTION.................................................................................. 6-3  
PERFORMANCE TEST........................................................................ 6-4  
Diagnostic Self-Tests ......................................................................... 6-4  
DC Voltage Test................................................................................. 6-5  
AC Voltage Test (Option -09 Only)................................................... 6-7  
Resistance Test................................................................................... 6-8  
DC Current Test................................................................................. 6-9  
AC Current Test (Option -09 Only)................................................... 6-10  
CALIBRATION..................................................................................... 6-10  
Basic Calibration Procedure............................................................... 6-11  
INITIAL PROCEDURE................................................................ 6-11  
A/D CALIBRATION .................................................................... 6-12  
OFFSET AND GAIN CALIBRATION ........................................ 6-14  
HIGH-FREQUENCY AC CALIBRATION.................................. 6-15  
Advanced Features and Special Considerations................................. 6-16  
STORING VARIABLE INPUTS.................................................. 6-16  
CALIBRATING INDIVIDUAL RANGES................................... 6-17  
VERIFYING CALIBRATION...................................................... 6-17  
ERASING CALIBRATION MEMORY....................................... 6-18  
TOLERANCE CHECK................................................................. 6-18  
AC CALIBRATION AT OTHER FREQUENCIES ..................... 6-19  
OPTIMIZING USE OF THE 5450A............................................. 6-19  
Remote Calibration ............................................................................ 6-20  
TIMING CONSIDERATIONS ..................................................... 6-23  
REMOTE ERASURE.................................................................... 6-24  
EXAMPLE CALIBRATION PROGRAM.................................... 6-24  
6-9.  
6-10.  
6-11.  
6-12.  
6-13.  
6-14.  
6-15.  
6-16.  
6-17.  
6-18.  
6-19.  
6-20.  
6-21.  
6-22.  
6-23.  
6-24.  
6-25.  
6-26.  
6-27. DISASSEMBLY PROCEDURE............................................................ 6-25  
6-28.  
6-29.  
6-30.  
6-31.  
6-32.  
Case Removal..................................................................................... 6-26  
True RMS AC PCA Removal (Option -09 Only).............................. 6-30  
IEEE-488 Interface PCA Removal (Option -05 Only) ...................... 6-30  
Main PCA Removal ........................................................................... 6-30  
Front Panel Disassembly.................................................................... 6-32  
6-33. REASSEMBLY PROCEDURE............................................................. 6-32  
6-34. INTERNAL FUSE REPLACEMENT ................................................... 6-36  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-35. EXTERNAL TRIGGER POLARITY SELECTION (Option -05 Only) 6-36  
6-36. TROUBLESHOOTING ......................................................................... 6-36  
6-37.  
6-38.  
6-39.  
6-40.  
6-41.  
6-48.  
6-59.  
6-62.  
6-63.  
6-64.  
6-65.  
6-66.  
6-67.  
6-68.  
6-69.  
6-70.  
6-71.  
6-72.  
6-73.  
6-74.  
6-75.  
Initial Troubleshooting Procedure...................................................... 6-36  
Diagnostic Self-Tests ......................................................................... 6-41  
Self-Test Descriptions........................................................................ 6-42  
Digital Controller Troubleshooting.................................................... 6-45  
IN-GUARD MICROCOMPUTER SYSTEM............................... 6-45  
DISPLAY SYSTEM...................................................................... 6-48  
ANALOG CONTROL SIGNALS................................................. 6-51  
DC Scaling Troubleshooting.............................................................. 6-54  
Track/Hold Troubleshooting.............................................................. 6-56  
Ohms Current Source Troubleshooting.............................................. 6-56  
Precision Voltage Reference Troubleshooting................................... 6-58  
A/D Converter Troubleshooting......................................................... 6-59  
Power Supply Troubleshooting.......................................................... 6-61  
IEEE-488 Interface Troubleshooting (Option -05) ............................ 6-64  
SERVICE POSITION.................................................................... 6-64  
DIAGNOSTIC PROGRAM .......................................................... 6-64  
True RMS AC Troubleshooting (Option -09).................................... 6-66  
SERVICE POSITION.................................................................... 6-66  
MAJOR PROBLEMS.................................................................... 6-67  
MORE OBSCURE PROBLEMS .................................................. 6-69  
Guard Crossing Troubleshooting ....................................................... 6-70  
6-76. INTERNAL CLEANING....................................................................... 6-70  
6-77.  
6-78.  
Cleaning Printed Circuit Assemblies ................................................. 6-70  
Cleaning After Soldering ................................................................... 6-71  
6-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
INTRODUCTION  
6
WARNING  
THESE SERVICE INSTRUCTIONS ARE FOR USE BY QUALIFIED  
PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT  
PERFORM ANY PROCEDURES IN THIS SECTION UNLESS YOU  
ARE QUALIFIED TO DO SO.  
6-1. INTRODUCTION  
This section presents maintenance information for the 8842A. The section includes a  
performance test, a calibration procedure, troubleshooting information, and other general  
service information.  
Test equipment recommended for the performance test and calibration procedure is listed  
in Table 6-1. If the recommended equipment is not available, equipment that meets the  
indicated minimum specifications may be substituted.  
Table 6-1. Recommended Test Equipment  
INSTRUMENT  
TYPE  
MINIMUM SPECIFICATIONS  
RECOMMENDED MODEL  
DC Calibrator  
PREFERRED:  
Fluke 5700A or Fluke 5440A  
Voltage Range: 0-1000V dc  
Voltage Accuracy: 10 ppm  
Absolute Linearity: ±1.0 ppm  
ALTERNATIVE:  
Fluke 720A  
Fluke 720A  
(Must be used with Kelvin-  
Varley Voltage Divider)  
Voltage Range: 0-1000V dc  
Voltage Accuracy: 20 ppm + 20 ppm  
of range  
Kelvin-Varley Voltage Divider:  
Ratio Range: 0-1.0  
Absolute Linearity: ± 1 ppm of input  
at dial setting  
Resistor Calibrator  
Resistance Accuracy: 0.0005%  
Fluke 5700A or Fluke 5450A, ESI  
DB62  
DC Current Source Accuracy: ±0.025%  
Fluke 5700A or Fluke 5100B  
Philips PM3055 or PM3355  
Oscilloscope  
General porpose, 60 MHz, with 10  
Mprobe  
Digital Multimeter  
Voltage Accuracy  
Fluke 8842A (with AC Option –09)  
:0.01% in V dc  
1.0% for 1V in V ac @ 100 kHz  
Input Impedance:  
10 Mor greater in V dc;  
1 Mor greater in parallel with <100  
pF in V ac  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
AC Calibrator  
Fluke 5700A and Fluke 5725A  
Minimum Required Accuracy (By Range)  
Frequency Range  
1, 10, 100 mV1  
1, 10, 100V2  
1000V2  
20 Hz – 30 Hz  
30 Hz – 20 kHz  
20 kHz – 50 kHz  
50 kHz – 100 kHz  
.1 + .005  
.02 + 10  
.05 + 20  
.05 + 20  
.1 + .005  
0.2 + .002  
.05 + .005  
.05 + .005  
.12 + .005  
.04 + .004  
.08 + .005  
.1 + .01  
1. ±(% of setting + µV) 2. +/-(% of setting + % of range)  
AC Current Source  
Fluke 5700A  
Frequency Range  
Minimum Required Accuracy (All Ranges)  
30 Hz – 1 kHz  
1kHz – 5 kHz  
±.07% + 1 mA)  
±(.07% + 1 mA) X frequency in kHz  
Shorting Bar  
Resistance <1.5 mΩ  
Construction Soldered (not  
rivetted)  
Pomona MDP-S-0  
6-Inch Jumper  
----  
E-Z-Hook 204-6W-S or equivalent  
Optional Test  
Equipment  
9010A, 9005A or Micro-System Troubleshooter; 9000A-8048 Interface Pod.  
6-2. PERFORMANCE TEST  
This test compares the performance of the 8842A with the specifications given in Section  
1. The test is recommended as an acceptance test when the instrument is first received,  
and as a verification test after performing the calibration procedure. If the instrument  
does not meet the performance test, calibration or repair is needed.  
To ensure optimum performance, the test must be performed at an ambient temperature  
of 18°C to 28°C, with a relative humidity of less than 75%. Also, the 8842A should be  
allowed to warm up for one hour prior to beginning any test other than the self-test.  
6-3.  
Diagnostic Self-Tests  
The diagnostic self-tests check the analog and digital circuitry in the 8842A. There are 21  
analog tests followed by in-guard program memory, calibration memory, and display  
tests. Out-guard program memory is tested when self-test is initiated by a remote  
command. Microcomputer RAM tests are done only at powerup. Each test is described  
in detail under the heading Troubleshooting. All five digital tests are performed at  
powerup.  
NOTE  
The inputs must be left open-circuited while the self-tests are performed.  
Otherwise, the 8842A may indicate errors are present. Errors may also be  
caused by inductive or capacitive pick-up from long test leads.  
If the FRONT/REAR switch is in the REAR position, the 8842A skips tests 3  
and 4. Also, if Option -09 is not installed, the 8842A skips tests 1, 2, and 3.  
6-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
PERFORMANCE TEST  
6
To initiate the self-tests, press the SRQ button for 3 seconds. The TEST annunciator will  
then light up, and the 8842A will run through the analog tests in sequence. Each test  
number is displayed for about 1 second. The instrument can be stopped in any of the test  
configurations by pressing the SRQ button while the test number is displayed. Pressing  
any button continues the tests.  
After the last analog test is performed, all display segments light up while the instrument  
performs the in-guard program memory, calibration memory, and display tests. The  
instrument then assumes the power-up configuration: VDC, autorange, slow reading rate,  
offset off, local control.  
If the 8842A detects an error during one of the tests, it displays the ERROR annunciator  
and the test number for about 2-1/2 seconds, and then proceeds to the next test. The test  
number thus becomes an error code. (Error codes are listed in Table 2-1, Section 2.)  
Passing all diagnostic self-tests does not necessarily mean the 8842A is 100% functional.  
The test, for example, cannot check the accuracy of the analog circuitry. If one or more  
errors are displayed, the 8842A probably requires service.  
6-4.  
DC Voltage Test  
The following procedure may be used to verify the accuracy of the VDC function:  
1. Ensure the 8842A is on and has warmed up for at least 1 hour.  
2. Select the VDC function.  
3. Connect the DC Calibrator (see Table 6-1) to provide a voltage input to the HI and  
LO INPUT terminals. Connections for the Kelvin-Varley Voltage Divider and the  
Fluke 5440A are shown in Figure 6-1.  
4. For each step in Table 6-2, select the indicated range, set the DC Calibrator for the  
specified input, and verify that the displayed reading is within the limits shown for  
each reading rate. (For step A, connect a short across the HI and LO INPUT  
terminals and press OFFSET. The measurement in step C should be relative to this  
offset.)  
5. Set the DC Calibrator to input negative voltage, and repeat steps C thrugh G of Table  
6-2.  
6. With the unit in the 2V range, check the A/D linearity by setting the DC Calibrator  
for each step in Table 6-9, while verifying the display reading is within the limit  
shown. Set the DC Calibrator for zero volts and disconnect if from the 8842A.  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f6-01.wmf  
Figure 6-1. DC Calibration Connections  
Table 6-2. DC Voltage Test  
Displayed Reading  
MEDIUM  
MIN  
STEP  
RANGE  
INPUT  
(V dc)  
SLOW  
FAST  
MIN  
MAX  
MAX  
MIN  
MAX  
A1  
B1  
C1  
20 mV  
0V  
(short)  
-0.0030  
-00.003  
+0.0030  
-0.0050  
-00.005  
+0.0050  
-0.030  
-00.03  
+0.030  
+00.03  
200 mV  
0V  
(short)  
+00.003  
+00.005  
2V, 20V,  
200V,  
0V  
(short)  
-2 counts +2 counts  
-4 counts + 4 counts -3 counts  
+3  
counts  
1000V  
D
E
F
G
H
I
20 mV  
200 mV  
2V  
10 mV  
+9.9963  
+10.0037  
+100.007  
+1.00004  
+10.0005  
+100.005  
+1000.05  
+9.9943  
+99.991  
+.99994  
+9.9993  
+99.993  
+999.93  
+10.0057  
+100.009  
+1.00006  
+10.0007  
+100.007  
+1000.07  
+9.970  
+99.97  
+.9997  
+9.997  
+99.97  
+999.7  
+10.030  
+100.03  
+1.0003  
+10.003  
+100.03  
+1000.3  
100 mV +99.993  
1V  
+.99996  
+9.9995  
+99.995  
+999.95  
20V  
10V  
200V  
1000V  
100V  
1000V  
1. Relative to high-quality short stored using OFFSET feature.  
6-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
PERFORMANCE TEST  
6
6-5.  
AC Voltage Test (Option -09 Only)  
The following procedure may be used to verify the accuracy of the VAC function:  
1. Ensure the 8842A is on and warmed up for at least 1 hour.  
2. Select the VAC function and the slow (S) reading rate.  
3. Connect the AC Calibrator to provide a voltage input to the HI and LO INPUT  
terminals.  
4. (Low- and Mid-Frequency Test.) For each step in Table 6-3, select the indicated  
range, set the AC Calibrator for the specified input, and verify that the displayed  
reading is within the limits shown for each reading rate.  
NOTE  
This procedure tests the extremes of each range. You may shorten the  
procedure by testing only the "quick test points" indicated in Table 6-3 with  
asterisks.  
5. (High-Frequency Test.) For each step in Table 6-4, select the indicated range, set the  
AC Calibrator for the specified input, and verify that the displayed reading is within  
the limits shown for each reading rate.  
NOTE  
This procedure tests the extremes of each range. You may shorten the  
procedure by testing only the "quick test points" indicated in Table 6-4 with  
asterisks.  
6. Set the AC calibrator to standby and disconnect it from the 8842A.  
Table 6-3. Low- and Mid-Frequency AC Voltage Test  
Step  
Number  
Range  
Input  
Frequency  
Error in  
Counts  
Display Readings  
Voltage  
Minimum  
Maximum  
1
2
2V  
2V  
0.01000V  
0.10000V  
0.30000V  
1.00000V  
1.90000V  
0.10000V  
1.90000V  
.10000V  
200 Hz  
200 Hz  
200 Hz  
200 Hz  
200 Hz  
20 Hz  
181  
87  
0.00819 VAC  
.09913 VAC  
0.29899 VAC  
0.99850 VAC  
1.89787 VAC  
0.09780 VAC  
1.87620 VAC  
0.09865 VAC  
1.89235 VAC  
0.01181 VAC  
.10087 VAC  
0.30101 VAC  
1.00150 VAC  
1.90213 VAC  
0.10220 VAC  
1.92380 VAC  
0.10135 VAC  
1.90765 VAC  
3*  
4
2V  
101  
150  
213  
220  
2380  
135  
765  
204  
252  
2V  
5*  
6
2V  
2V  
7
2V  
20 Hz  
8
2V  
45 Hz  
9
2V  
1.90000V  
0.001000V  
0.190000V  
45 Hz  
10*  
11  
200 mV  
200 mV  
100 Hz  
20 kHz  
00.796 mVAC 190.252 mVAC  
189.748 mVAC 190.252 MvAC  
* Quick Test points  
6-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 6-4. High-Frequency AC Voltage Test  
STEP  
NUMBER  
RANGE  
INPUT  
FREQUENCY  
ERROR  
IN  
COUNTS  
TEST LIMITS (IN VOLTS)  
VOLTAGE  
MIN  
MAX  
1
200 mV  
200 mV  
2V  
0.010000V  
0.010000V  
0.10000V  
1.0000V  
50 kHz  
100 kHz  
100 kHz  
100 kHz  
100 kHz  
100 kHz  
100 kHz  
100 kHz  
169  
350  
09.831 mVAC  
09.650 mVAC  
0.09650 VAC  
0.9650 VAC  
09.650 VAC  
96.50 VAC  
10.169 mVAC  
10.350 mVAC  
0.10350 VAC  
1.0350 VAC  
10.350 VAC  
103.50 VAC  
19.1250 VAC  
2 *  
3 *  
4 *  
5 *  
6 *  
7
350  
20V  
350  
200V  
700V  
20V  
10.000V  
350  
100.00V  
350  
19.0000V  
0.190000V  
1250  
1250  
18.8750 VAC  
8
200 mV  
188.750 mVAC 191.250 mVAC  
* Quick test points.  
6-6.  
Resistance Test  
The following procedure may be used to verify the accuracy of the 2-wire and 4-wire  
ohms functions.  
1. Ensure the 8842A is on and has warmed up for at least 1 hour.  
2. Connect the Resistance Calibrator to the 8842A for 4-wire ohms.  
3. For each step in Table 6-5, select the indicated range, set the Resistance Calibrator  
for the specified nominal input. When the input is 0(short), press the OFFSET  
switch and verify that the OFFSET anunciator is illuminated.  
NOTE  
A new OFFSET must be stored for each new range selected.  
a. Test the 4-wire ohms function:  
1.  
2.  
Select the 4-wire ohms function.  
Verify that the displayed reading is within the limits shown for  
each reading rate.  
b. Test the 2-wire ohms function:  
1. Select the 2-wire ohms function. (The SENSE test leads need not  
be disconnected.)  
2. Verify that the displayed reading is within the limits shown for  
each reading rate.  
6-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
PERFORMANCE TEST  
6
Table 6-5. Resistance Test  
Step  
Range  
Input  
Error From Input1 (in counts)  
(Nominal)  
Slow  
Medium  
Fast2  
3
1
2
20Ω  
0(short)  
10Ω  
±40  
±49  
±4  
±60  
±69  
±6  
±20  
±21  
±3  
±4  
±2  
±3  
±2  
±3  
±2  
±3  
±2  
±5  
±2  
±6  
3
20Ω  
3
3
200Ω  
0(short)  
100Ω  
3
4
200Ω  
±11  
±3  
±13  
±5  
5
2 kΩ  
2 kΩ  
0(short)  
1 kΩ  
6
±8  
±10  
±5  
7
20 kΩ  
0(short)  
10 kΩ  
±3  
8
20 kΩ  
±8  
±10  
±5  
9
200 kΩ  
200 kΩ  
2000 kΩ  
2000 kΩ  
20 MΩ  
20 MΩ  
0(short)  
100 kΩ  
0(short)  
1 MΩ  
±3  
10  
11  
12  
13  
14  
±9  
±11  
±6  
±3  
±28  
±4  
±31  
±7  
0(short)  
10 MΩ  
±44  
±47  
1 – Using offset control  
2 = 4 ½ digit counts  
3 = Applies to 4-wire ohms only  
6-7.  
DC Current Test  
The following procedure may be used to test the mA DC function:  
1. Ensure the 8842A is on and has warmed up for at least 1 hour.  
2. Select the mA DC function.  
3. Connect the Current Source to the 2A and LO INPUT terminals.  
4. For each step in Table 6-6, set the Current Source for the indicated input and verify  
that the displayed reading is within the limits shown for each reading rate.  
5. Set the Current Source for zero mA and disconnect it from the 8842A.  
Table 6-6. DC Current Test  
DISPLAYED READING  
STEP  
RANGE  
INPUT  
SLOW  
MEDIUM  
FAST  
MIN  
MAX  
MIN  
MAX  
MIN  
MAX  
1
2
3
4
200 mA  
200 mA  
2000 mA  
0 mA  
100 mA  
0 mA  
-00.040  
99.920  
-000.04  
999.56  
+00.040  
100.080  
+000.04  
1000.44  
-00.060  
99.900  
-000.06  
999.54  
+00.060  
100.100  
+000.06  
1000.46  
-00.20  
99.76  
-000.2  
999.4  
+00.20  
100.24  
+000.2  
1000.6  
2000 mA 1000 mA  
6-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-8.  
AC Current Test (Option -09 Only)  
The following procedure may be used to test the mA AC function:  
1. Ensure the 8842A is on and warmed up for at least 1 hour.  
2. Select the mA AC function.  
3. Connect the AC Current Source to provide a current input to the 2A and LO INPUT  
terminals. If an ac current source is not available, the functionality of the 8842A can  
be checked at 10 mA by using a Fluke 5200A set at 100V and connected to the  
8842A 2A terminal through a 10 k, 2W, 1% resistor.  
4. For each step in Table 6-7, set the AC Current Source for the indicated input and  
verify that the displayed reading is within the limits shown for each reading rate.  
5. Set the AC Current Source to Standby and disconnect it from the 8842A.  
Table 6-7. AC Current Test  
STEP  
NUMBER  
RANGE  
INPUT  
FREQUENCY  
TEST LIMITS  
CURRENT  
MINIMUM  
MAXIMUM  
1
2
2000 mA  
2000 mA  
1900.00 mA  
100.00 mA  
1 kHz  
1 kHz  
1890.40  
97.60  
1909.60  
102.40  
6-9. CALIBRATION  
CAUTION  
To avoid uncalibrating the 8842A, never cycle power on or off  
while the CAL ENABLE switch is on.  
NOTE  
If U220 is replaced, perform the Erase Cal Memory procedure (located  
later in this section) before attempting calibration. Failure to do so may  
result in an "ERROR 29" on the 8842A front panel display.  
The 8842A features closed-case calibration using known reference sources. The 8842A  
automatically prompts you for the required reference sources, measures them, calculates  
correction factors, and stores the correction factors in the nonvolatile calibration memory.  
Closed-case calibration has many advantages. There are no parts to disassemble, no  
mechanical adjustments to make, and if the IEEE-488 Interface is installed, the 8842A  
can be calibrated by an automated instrumentation system.  
The 8842A should normally be calibrated on a regular cycle, typically every 90 days or 1  
year. The frequency of the calibration cycle depends on the accuracy specification you  
wish to maintain. The 8842A should also be calibrated if it fails the performance test or  
has undergone repair. To meet the specifications in Section 1, the 8842A should be  
calibrated with equipment meeting the minimum specifications given in Table 6-1.  
The following paragraphs first present a basic calibration procedure. This is followed by  
a description of advanced features and special considerations, and by a description of  
remote calibration using the IEEE-488 Interface.  
6-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
6-10. Basic Calibration Procedure  
The basic calibration procedure consists of the following four parts. These parts must be  
performed in the order shown.  
1. Initial Procedure.  
2. A/D Calibration.  
3. Offset and Gain Calibration for each function and range.  
4. High-Frequency AC Calibration (True RMS AC option only).  
Normally, it is recommended that the entire calibration procedure be performed.  
However, under some circumstances the earlier parts may be omitted. For example, if  
installing the True RMS AC option, it may be necessary only to perform Offset and Gain  
Calibration for the ac functions, followed by High-Frequency AC Calibration. But if the  
A/D Calibration is performed, it must be followed by a complete Offset and Gain  
Calibration for all functions and then by High-Frequency AC Calibration.  
Some of the calibration calculations are complex and take the 8842A some time to  
execute. For example, when you store the zero input during the Offset and Gain  
Calibration for the VDC function, it takes around 22 seconds before the next prompt  
appears. (The 8842A automatically uses this input to calibrate the offset for all ranges.)  
While the 8842A is executing a calibration step, it ignores all of the front panel buttons  
and postpones execution of all remote commands.  
6-11. INITIAL PROCEDURE  
Always begin the calibration procedure as follows:  
1. Allow the 8842A to stabilize in an environment with ambient temperature of 18°C to  
28°C and relative humidity less than 75%.  
2. Turn the 8842A on and allow it to warm up for at least 1 hour.  
3. Enable the calibration mode by pressing the CAL ENABLE switch with a small  
screw-driver or other suitable instrument. (The CAL ENABLE switch is located on  
the right side of the display and is normally covered by a calibration seal.)  
When the calibration mode is enabled, the CAL annunciator lights up, and the 8842A  
displays the first prompt for the A/D Calibration procedure (Figure 6-2). To exit the  
calibration mode, press the CAL ENABLE switch again.  
f6-02.wmf  
Figure 6-2. First A/D Calibration Prompt  
In the calibration mode, the front panel controls assume the functions described in Figure  
6-3. Some of these functions are advanced features and are not required for the basic  
calibration procedure. The display blanks briefly when a button is pressed.  
6-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f6-03.wmf  
Figure 6-3. Calibration Functions  
The following functions are inappropriate during calibration, and are therefore  
unavailable:  
Offset  
Autoranging  
External Trigger  
Front Panel Trigger  
Front panel SRQ (Under local control)  
Diagnostic self-tests  
6-12. A/D CALIBRATION  
The A/D Calibration procedure calibrates the analog-to-digital converter for offset, gain  
and linearity. The 8842A automatically selects the A/D calibration procedure when the  
CAL ENABLE switch is first pressed. The procedure must be performed in its entirety,  
and may not be performed in part. If the A/D calibration is discontinued prior to  
completion, the last complete set of A/D calibration constants will be retained unchanged.  
To perform A/D Calibration, proceed as follows:  
1. Ensure the Initial Procedure has been completed. The 8842A then displays the  
prompt for the first reference source, zero volts (i.e., a short).  
6-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
2. Each time the 8842A prompts you for a reference source, apply the requested source  
to the HI and LO INPUT terminals, and press the STORE button. When STORE is  
pressed, the numeric display field blanks while the 8842A performs the necessary  
calculations. (Do not change the reference source while the display is blank.) The  
8842A then displays the next prompt. For reference, all prompts are shown in Table  
6-8.  
NOTE  
The 8842A automatically checks that the reference input is near the value  
prompted, and displays ERROR 41 if it exceeds a specific tolerance. (See  
Advanced Features and Special Considerations, later in this section.)  
3. After the last input is stored, the 8842A begins taking readings in the 2V range of the  
VDC function. (The CAL annunciator remains on.) Verify the A/D calibration using  
the test points in Table 6-9. If you wish to repeat the A/D Calibration procedure,  
press the A/D button.  
Table 6-8. A/D Calibration Steps  
STEP  
DISPLAYED PROMPT  
A
B
C
D
E
F
G
H
I
.0 V DC (short)  
- .03 V DC  
- 1.01 V DC  
+ .99 V DC  
+ .51 V DC  
- .51 V DC  
- .26 V DC  
+ .26 V DC  
+ .135 V DC  
- .135 V DC  
- .0725 V DC  
+ .0725 V DC  
J
K
L
Table 6-9. A/D Calibration Verification Test  
STEP  
INPUT  
A L L O W A B L E  
ERROR  
A
B
C
D
E
F
0V (short)  
-0.03V  
±2 counts  
±2 counts  
±2 counts  
±3 counts  
±3 counts  
±4 counts  
±4 counts  
+0.03V-0.660V  
-0.660V  
+0.660V  
-1.970V  
G
+1.970V  
6-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
The A/D Calibration procedure is an iterative process. Each pass through the procedure  
uses the constants stored previously and improves them. Normally, one pass is adequate.  
However, if the calibration memory has been erased or replaced, or the A/D Converter  
has undergone repair, the A/D Calibration procedure must be performed twice.  
Because the A/D Calibration procedure is iterative, the instruments performance can be  
slightly enhanced by going through the procedure more than twice. However, this is not  
necessary to meet the published specifications.  
6-13. OFFSET AND GAIN CALIBRATION  
This procedure calibrates the instruments offsets and gains by applying a high and low  
input for every range of each function. To save time, the 8842A uses each input for as  
many ranges as possible.  
A function is calibrated by pressing the corresponding function button. Once a function is  
selected, the 8842A automatically steps through each range of that function, prompting  
you for the necessary reference sources. (The prompts are shown in Table 6-10.) The  
8842A does not automatically select another function after one function has been  
completely calibrated; therefore, the functions may be calibrated in any order.  
To perform Offset and Gain Calibration, proceed as follows:  
Table 6-10. Offset and Gain Calibration Steps  
STEP  
DISPLAYED PROMPT  
VDC  
VAC1  
2 WIRE kΩ  
4 WIRE kΩ  
mA DC  
mA AC1  
A
B
C
+00.0 mV DC (short) 10.0 mV AC  
0.00(short) 00.0 mA DC (open)  
100 mA AC  
--  
+19.0 mV DC  
+190.0 mV DC  
--  
10.0(4 wire)  
100.0Ω  
100 mA DC  
100.0 mV AC  
1000. mA DC  
1000.0 mA AC  
D
E
F
+1.900 V DC  
+19.00 V DC  
+190.0 V DC  
+1000. V DC  
1.000 V AC  
10.00 V AC  
100.0 V AC  
500. V AC  
1.000 kΩ  
10.00 kΩ  
100.0 kΩ  
1000. kΩ  
10.00 MΩ  
Steps D-H not applicable for these  
functions.  
G
H
Step H not applicable for these  
functions.  
1 Inputs should be at 1 kHz +/-10%. Performance may be enhanced for specific frequencies (see text).  
1. Ensure the A/D Calibration procedure has been completed.  
2. Select the desired function by pressing the corresponding function button. The 8842A  
will display the first prompt for that function.  
3. Each time the 8842A prompts you for a reference source, apply this source to the  
appropriate terminals, and press the STORE button. When STORE is pressed, the  
numeric display field blanks while the 8842A performs the necessary calculations.  
(Do not change the reference source while the display is blank.) The 8842A then  
displays the next prompt. For reference, all prompts are shown in Table 6-10.  
6-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
NOTE  
To use reference sources that differ from the prompted values, see Storing  
Variable Inputs later in this section.  
4. After the last range is calibrated, the 8842A begins taking readings in the highest  
range so that you may verify its calibration. (The CAL annunciator remains on.) To  
verify the calibration for the other ranges, press the corresponding range button.  
(Pressing a function button begins the Offset and Gain Calibration procedure for that  
function.)  
5. Repeat steps 2, 3, and 4 for the remaining functions. Note that both 2-wire and 4-wire  
ohms functions must be calibrated. (The VAC and mA AC functions require  
calibration only if the True RMS AC option is installed.)  
NOTE  
(True RMS AC option only.) The VAC and mA AC functions should  
normally be calibrated using reference sources at 1 kHz (+/-10%). For  
special applications, performance may be optimized at other frequencies.  
See Optimizing AC Calibration at Other Frequencies, later in this section.  
6. When all functions have been calibrated, exit the calibration mode by pressing the  
CAL ENABLE switch and attach a calibration certification sticker over the CAL  
ENABLE switch. (If the True RMS AC option is installed, instead proceed to the  
High-Frequency AC Calibration procedure which follows.)  
6-14. HIGH-FREQUENCY AC CALIBRATION  
The High-Frequency AC Calibration procedure calibrates the response of the VAC  
function from 20 kHz to 100 kHz. If the True RMS AC option is not installed, selecting  
this procedure results in an error message.  
The reference sources used in this procedure should normally be between 90 kHz and 100  
kHz. 100 kHz (nominal) is recommended. For special applications, performance may be  
optimized at other frequencies. See Optimizing AC Calibration at Other Frequencies,  
later in this section.  
To perform High-Frequency AC Calibration, proceed as follows:  
1. Ensure Offset and Gain Calibration has been completed for the VAC function.  
2. Select the High-Frequency AC Calibration procedure by pressing the HF AC button.  
The 8842A will display the first prompt (100 mV AC). The "U" in the display  
indicates the High-Frequency AC Calibration procedure has been selected.  
3. Each time the 8842A prompts you for a reference amplitude, apply this amplitude to  
the HI and LO INPUT terminals, and press the STORE button. When STORE is  
pressed, the numeric display field blanks while the 8842A performs the necessary  
calculations. (Do not change the reference source while the display is blank.) The  
8842A then displays the next prompt. For reference, all prompts are shown in Table  
6-11.  
NOTE  
To use reference amplitudes that differ from the prompted values, see  
Storing Variable Inputs later in this section.  
4. After the last range is calibrated, the 8842A begins taking readings in the highest  
range so that you may verify its calibration. To verify the calibration for the other  
ranges, push the corresponding range button. The CAL annunciator remains on.  
6-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
5. The calibration procedure is now completed. Exit the calibration mode by pressing  
the CAL ENABLE switch, and attach a calibration certification sticker over the CAL  
ENABLE switch.  
6-15. Advanced Features and Special Considerations  
Table 6-11. High-Frequency AC Calibration Steps  
STEP  
DISPLAYED PROMPT1,2  
A
B
C
D
E
100.0 mV AC  
1.000 V AC  
10.00 V AC  
100.0 V AC  
200.0 V AC  
1. The display also indicates “U” to show that HF AC cal is selected.  
2. Inputs should be between 90 kHz and 100 kHz. (nominal) is recommended.  
The 8842A has several advanced calibration features which are not necessary for the  
basic calibration procedure, but which can make calibration easier. The following  
paragraphs describe these features and also discuss special considerations for optimizing  
the performance of the 8842A in special situations.  
6-16. STORING VARIABLE INPUTS  
As a convenience, the VAR IN (variable input) feature lets you calibrate the 8842A using  
reference source values which differ from the values prompted by the 8842A. For  
example, you may want to calibrate the 200range using a reference resistor with a  
precisely known value of 99.875, rather than 100as prompted. This feature is not  
available during A/D Calibration.  
To use the variable input feature, proceed as follows:  
1. When the 8842A prompts you for an input, press the VAR IN button. The blank  
digits will be replaced with zeroes. You can then increment each digit of the display  
by pressing the range buttons. The 20/mV button toggles the displayed sign.  
2. Change the displayed prompt to correspond to the desired reference source by  
pressing the appropriate range buttons.  
3. Connect the desired reference source to the appropriate input terminals of the 8842A.  
4. Press the STORE button.  
To meet the specifications over all ranges, the reference source for the high prompts must  
be between half and full scale. (The high prompts are those prompts that are between  
50% and 100% of full scale.) The reference source for the low prompts must be equal to  
or greater than the prompted value, but not more than that value plus 4000 counts. (The  
low prompts are those prompts that are zero or 5% of full scale.)  
For special applications, the 8842A can be calibrated at values outside the recommended  
range. This can enhance the performance at the calibration value. However, performance  
at other values may be degraded.  
6-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
6-17. CALIBRATING INDIVIDUAL RANGES  
During Offset and Gain and High-Frequency AC Calibration, it is possible to calibrate  
individually selected ranges. This feature does not apply to the mA DC and mA AC  
functions and is not permitted during A/D Calibration.  
To calibrate an individual range, proceed as follows:  
1. Select the desired calibration procedure by pressing the appropriate function button  
(or press the HF AC button if High-Frequency AC Calibration is desired).  
2. Press the range button for the range to be calibrated. The 8842A then prompts for a  
low reference source for that range. (See Table 6-12.) (During High-Frequency AC  
Calibration, the 8842A prompts only for a high reference source. In this case,  
proceed to step 4.)  
3. Apply the requested reference source and press STORE. The display will blank  
briefly and then prompt for a high reference source. (See Table 6-12.)  
4. Apply the requested reference source and press STORE. The display will blank  
briefly, and the 8842A will then begin taking readings in the selected range so that  
you may verify the calibration. The CAL annunciator remains on.  
5. To continue, select another range. You may restart any of the calibration procedures  
by pressing the appropriate function button, the A/D button, or the HF AC button.  
6-18. VERIFYING CALIBRATION  
Table 6-12. Prompts When Calibrating Individual Ranges  
PROCEDURE  
FUNCTION  
LOW PROMPT  
Zero  
HIGH PROMPT  
95% of full scale 1  
50% of full scale  
50% of full scale 2  
Offset and Gain  
Calibration  
VDC  
kΩ  
Zero  
VAC  
5% of full scale 2  
(No low prompt)  
3
High-Frequency AC  
Calibration  
(Not applicable)  
50% of full scale  
1. Exception: The 1000V dc range has a high prompt of 1000V dc.  
2. Exception: The 700V ac range has a low prompt of 100V ac and a high prompt of 500V ac.  
3. Exception: The 700V ac range has a prompt of 200V ac.  
When you complete certain parts of the calibration procedure, the 8842A automatically  
begins taking readings so that you can verify the calibration is correct. (It is  
recommended that you do so.) The CAL annunciator remains lit. The 8842A continues to  
take readings until you select another calibration procedure or exit the calibration mode.  
Specifically, the 8842A begins taking verification readings after:  
1. Completing A/D Calibration.  
2. Completing Offset and Gain or High-Frequency AC Calibration.  
3. Completing the calibration of an individually selected range.  
While the 8842A is taking verification readings, certain buttons are active or function  
differently:  
6-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
1. If you just completed the Offset and Gain or High-Frequency AC Calibration for an  
entire function (not just one range), the range buttons can be used to change ranges in  
order to verify all ranges were calibrated correctly.  
2. If you just calibrated an individually selected range, pressing another range button  
begins the Offset and Gain Calibration procedure for the new range.  
3. You can use the RATE button to verify the calibration at other reading rates.  
CAUTION  
It is still possible to erase the calibration memory while the  
8842A is taking verification readings.  
6-19. ERASING CALIBRATION MEMORY  
The 8842A allows you to erase some or all of the correction constants stored in  
calibration memory (U220). It is recommended that you erase the entire calibration  
memory before beginning calibration if the calibration memory is replaced or  
accidentally altered. The capability of erasing particular parts of the memory is mainly  
intended as a troubleshooting aid to the technician.  
CAUTION  
Once the calibration memory is erased, the 8842A must be  
recalibrated.  
To erase all or part of the calibration memory, proceed as follows:  
1. Press the front panel ERASE button. The display should show the erase prompt "cl"  
(for "clear"). If you do not press another button within 1-1/2 seconds, the 8842A  
returns to its previous state.  
2. To complete an erasure, press one of the following buttons within 1-1/2 seconds of  
pressing the ERASE button:  
a. STORE -- Erases the entire memory.  
b. A/D -- Erases the A/D Calibration constants  
c. Any function button -- Erases the Offset and Gain Calibration constants for all  
ranges of that function.  
d. HF AC -- Erases the High-Frequency AC Calibration constants.  
3. After an erasure is finished (a complete erasure takes about 3 seconds), the 8842A  
returns to one of the following states:  
a. After complete erasure: Begins A/D Calibration.  
b. After A/D erasure: Begins A/D Calibration.  
c. After Offset and Gain erasure: Begins Offset and Gain Calibration for erased  
function.  
d. After High-Frequency AC erasure: Begins High-Frequency AC Calibration.  
6-20. TOLERANCE CHECK  
The 8842A automatically checks that the reference input is near the value prompted. This  
minimizes common errors such as applying a reference source with the wrong sign. If the  
reference input exceeds the tolerances shown in Table 6-13, the 8842A displays ERROR  
41.  
6-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
If ERROR 41 occurs, the most likely cause is that the reference input is incorrect (e.g.,  
has the wrong polarity). If the input is in fact correct, refer to the Troubleshooting  
heading in this section.  
Table 6-13. Tolerance Limits  
CALIBRATION  
TOLERANCE  
1.  
A/D Calibration  
±244 counts from prompt  
2.  
Offset and Gain Calibration  
VDC, mA DC  
±488 counts from prompt  
±3002 counts from prompt  
±3002 counts from prompt  
±9999 counts from prompt  
Ohms  
VAC, mA AC  
3.  
HF AC Calibration  
6-21. AC CALIBRATION AT OTHER FREQUENCIES  
For special applications where the 8842A is to be used to measure ac voltages or currents  
exclusively at a single frequency or narrow range of frequencies, accuracy may be  
enhanced at that frequency (or range of frequencies) by performing calibration according  
to the following procedure. Note that this may degrade the accuracy at frequencies  
significantly removed from the frequency of optimization.  
To optimize performance at a frequency less than 1 kHz, perform the offset and gain  
calibration procedure, above, using the frequency at which measurements will be made  
rather than 1 kHz. This technique may be used for both the VAC and mA AC functions.  
At the calibration frequency, the 8842A will yield accuracy closely approaching the  
specified mid-band performance.  
To optimize performance at a frequency greater than 1 kHz, perform calibration as  
follows:  
1. Perform the Offset and Gain Calibration procedure using inputs at 1 kHz.  
2. Perform the High-Frequency AC Calibration procedure using inputs at the desired  
frequency of optimization rather than at 100 kHz. Skip step 5 in that procedure  
3. Again perform the Offset and Gain Calibration procedure, this time using inputs at  
the desired frequency of optimization rather than at 1 kHz.  
6-22. OPTIMIZING USE OF THE 5450A  
If the Fluke 5450A Resistance Calibrator is used to calibrate the 2-wire ohms function,  
the following procedure is recommended to optimize the calibration of the lowest two  
ranges. (The 5450A has a 25 m"floor" which would otherwise result in 25 digits of  
error in the 200range of the 8842A.) In this procedure, the 8842A is referred to as the  
unit under test (UUT).  
1. Complete Offset and Gain Calibration for the UUTs 4-wire ohms function. The  
UUT will then be taking verification readings.  
2. Connect the UUT to the 5450A as shown in Figure 6-4.  
3. Select the "SHORT" from the 5450A, and measure this value at the 5450A OUTPUT  
terminals using the UUT in 4-wire ohms. If in remote, take the average of four  
6-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
readings. (In remote calibration, the averaged value can be stored in the controller.)  
Record the value.  
4. Select the "100" output from the 5450A, and measure this value as in step 3.  
5. Find and record the numerical difference between the values calculated in steps 3 and  
4. This value will be used as the variable input for calibrating the 200range in 2-  
wire ohms.  
6. Repeat steps 4 and 5 using the "1 k" output from the 5450A; find and record the  
numerical difference between this and the "SHORT" measured in step 3. This value  
will be used as the variable input for calibrating the 2 krange in 2-wire ohms.  
7. Press the UUTs 2 WIRE kbutton. This selects the Offset and Gain calibration  
procedure for the 2-wire ohms function and prompts for zero input. Select the  
"SHORT" from the 5450A, and calibrate all the zeros by pressing STORE.  
8. Select the "100" output from the 5450A and calibrate the high point for the 200Ω  
range, entering the value computed in step 5 as a variable input.  
9. Select the "1 k" output from the 5450A and calibrate the high point for the 2 kΩ  
range, entering the value computed in step 6 as a variable input.  
10. Calibrate the remaining ranges (steps D-G of Table 6-10) using the 5450A outputs.  
11. Recalibrate the low point for each 2-wire ohms range using a shorting link (Pomona  
MDP-S-0 or equivalent) across the UUTs HI and LO INPUT terminals.  
12. Exit the calibration mode by pressing the CAL ENABLE switch.  
13. Using the same configuration shown in Figure 6-4, verify that the UUT measures the  
same value (within 1 digit) in 2-wire ohms (using the offset feature to correct for  
5450A floor error) as in 4-wire ohms. If the readings differ by more than 1 digit,  
reenable the calibration mode and repeat steps 2 through 8.  
NOTE  
Only 4-wire ohm calibration is allowed in the 20range.  
14. Cover the CAL ENABLE switch with a calibration certification sticker.  
f6-04.wmf  
Figure 6-4. Optimizing Use of the 5450A  
6-23. Remote Calibration  
If the IEEE-488 Interface is installed, the 8842A can be calibrated under remote control.  
Remote calibration is very similar to local (front-panel controlled) calibration. Table 6-14  
shows the remote commands which correspond to the front panel features.To facilitate  
remote calibration, there are some differences from local calibration:  
6-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
CALIBRATION  
6
Table 6-14. Commands Used During remote Calibration  
FRONT PANEL  
FEATURE  
CORRESPONDING  
COMMAND  
COMMENTS  
Display  
Function Buttons  
Range  
G2  
Loads the calibration prompt into the output buffer. Not valid  
when the 8842A is taking verification readings.  
F1 through F6  
In the calibration mode, these select the Offset and Gain  
Calibration procedure for the corresponding funtion.  
R1 through R6 and R8 In the calibration mode, these select the Offset and Gain  
Calibration Procedure for the corresponding range in the  
presently selected function. (For entering variable inputs, see  
VAR IN below.)  
STORE  
C0  
Tells the 8842A that the requested calibration input is valid.  
This command causes the 8842A to take readings, and  
compute and store calibration constants.  
NOTE  
The C0 command can take up to 22 seconds to execute.  
You must determine when this command is complete  
before sending more commands. See Timing  
Considerations in text.  
A/D  
C1  
C2  
C3  
Selects the A/D Calibration procedure.  
HF AC  
ERASE  
Selects the High-Frequency AC Calibration procedure.  
After receiving this command the display shows the erase  
mode prompt (‘cl’). (The prompt is not loaded into the output  
buffer.) To complete the erasure you must then send C0.  
Sending any other command after the C3 command causes  
the 8842A to return to its previous state. There is no timeout  
as with the front panel ERASE button.  
CAUTION  
The command string “C3 C0” erases the entire  
calibration memory. A complete calibration must then  
be performed.  
RATE  
S0 through S2  
Changes the reading rate while the 8842A is taking  
verification readings. Causes an error at any other time  
during calibration.  
VAR IN  
N<value>P2  
Enters <value> as a variable input. (See Entering Variable  
Inputs, in text.) Causes an error if sent during A/D Calibration  
or when the 8842A is taking verification readings. You can  
check that the command was successful by checking the  
error status or by sending the Get Calibration Input command  
(G2).  
--  
P3<string>  
Puts the <string> into calibration memory. The string must  
contain up to 16 ASCII characters, and be recalled with the  
G3 command.  
NOTE: Other commands that may be used during calibration are: P1 (Put SRQ mask); the remaining Get  
commands; and X0 (clear Error Register).  
6-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
1. In remote calibration, you can store a 16-character message in the calibration  
memory which can be read by the system controller. Possible uses include  
storing the calibration date, instrument ID, etc.  
2. Although some buttons are ignored in local calibration (e.g., the AUTO button),  
the corresponding remote commands (e.g., R0) load the output buffer with an  
error message.  
3. The calibration memory is erased differently. (This is explained later.)  
4. The rear panel SAMPLE COMPLETE signal acts slightly differently. During  
calibration, the SAMPLE COMPLETE signal is inactive. When the 8842A is  
taking verification readings, the SAMPLE COMPLETE signal acts the same as  
in normal operation.  
Note that a command may be valid in some parts of the calibration procedure but not in  
others. The Get Input Prompt (G2) command, for instance, is not valid when the 8842A is  
taking verification readings. The Rate (Sn) commands, for instance, are valid when the  
8842A is taking verification readings, but they are not valid at any other time during  
calibration. Table 6-15 shows when commands are invalid.  
6-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
CALIBRATION  
6
6-24. TIMING CONSIDERATIONS  
Table 6-15. Error Numbers Which Are Displayed When Commands Are Not Valid  
COMMAND  
NORMAL  
MODE  
SELFTEST  
CALIBRATION MODE  
A/D CAL  
OFFSET &  
GAIN CAL  
HF AC CAL VERIFICAT  
ION MODE  
Bn  
60  
52  
52  
52  
52  
52  
C0  
C1  
C2  
C3  
Dn  
51  
60  
60  
60  
60  
60  
54  
51  
51  
51  
52  
52  
52  
Fn  
G0  
G1  
G2  
G3  
G4  
G5  
G6  
G7  
N
60  
51  
60  
60  
54  
P0  
P1  
P2  
P3  
R0  
R1-6  
R7  
R8  
Sn  
Tn  
Wn  
X0  
Yn  
Z0  
*
60  
52  
56  
52  
52  
52  
54  
51  
51  
60  
60  
60  
60  
60  
60  
60  
60  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
52  
60  
60  
60  
52  
52  
52  
52  
52  
52  
52  
52  
?
6-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
The C0 command can take up to 22 seconds. If during this time the controller continues  
to send the 8842A more commands, the commands may fill up the 8842As input buffer.  
If this happens, errors will probably occur.  
You can avoid this problem by knowing when these commands are completed. There are  
three ways to determine this:  
1. Monitoring the Cal Step Complete bit in the serial poll status register. This status bit  
is set false every time the remote processor sends a command to the in guard  
processor. It is then set true when the in guard processor completes the command and  
is ready to accept more. So you can send a command and loop on a check of the  
status, until the command is complete.  
2. Setting the SRQ mask to generate an SRQ on Cal Step Complete. An SRQ is  
generated and the Cal Step Complete bit in the serial poll status response is set when  
a cal command is complete. This approach depends on capabilities of the controller  
being used.  
3. Executing a delay in controller software after sending each command. (Not  
recommended.)  
Although not usually necessary, these methods can be used for other commands as well.  
6-25. REMOTE ERASURE  
The C3 command allows you to erase the entire calibration memory. The erasure is  
executed by sending the string "C3 C0" (equivalent to pressing ERASE and then  
STORE). Any command other than C0 after C3 will abort the erasure. To facilitate  
remote calibration, the C0 command does not timeout as does the front panel ERASE  
button. The selective erasure that is possible from the front panel is intended as a  
troubleshooting aid, and is not available over the IEEE-488 Interface.  
Note that the erase command can take up to 3 seconds to execute. To prevent timeout  
problems with the controller, you must determine when the command is completed  
before continuing. Several methods are presented in Timing Considerations, above.  
NOTE  
When erasing calibration memory, it is good practice to send the  
commands C3 and C0 in the same command string. Sending C3 by itself  
could lead to accidentally erasing calibration memory, since the C3  
command does not time out as does the ERASE button.  
6-26. EXAMPLE CALIBRATION PROGRAM  
An example A/D calibration program is shown in Figure 6-5. The program is written in  
Fluke BASIC for the Fluke 1722A Instrument Controller. It uses the Fluke 5440A Direct  
Voltage Calibrator to perform and then verify the A/D Calibration procedure. In this  
program, the 8842A is at bus address 1, and the 5440A is at bus address 7.  
6-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
DISASSEMBLY PROCEDURE  
6
f6-05.wmf  
Figure 6-5. Example A/D Calibration Program  
6-27.DISASSEMBLY PROCEDURE  
WARNING  
TO AVOID ELECTRIC SHOCK, REMOVE THE POWER CORD  
AND TEST LEADS BEFORE DISASSEMBLING THE  
INSTRUMENT. OPENING COVERS MAY EXPOSE LIVE PARTS.  
CAUTION  
To avoid contaminating the printed circuit assemblies (PCAs),  
handle the PCAs by their edges. Do not handle the areas of the  
PCAs that are not solder masked unless absolutely necessary.  
These areas must be cleaned if contaminated.  
The following paragraphs present a disassembly procedure for the 8842A. The procedure  
should be performed in the order presented. Remove the case first, and then remove  
Option -09, the True RMS AC PCA, Option -05, the IEEE-488 Interface PCA, the Main  
PCA, and the front panel. For reference, see the final assembly drawing in Section 7.  
6-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-28. Case Removal  
1. Remove the grounding screw from the bottom of the case. Remove two rear bezel  
mounting screws. (See Figure 6-6A.)  
2. While holding the front panel, slide the case and rear bezel off the chassis (See  
Figure 6-6B). (At this point, the rear bezel is not secured to the case.)  
f6-06_1.wmf  
Figure 6-6. 8842A Disassembly  
6-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
DISASSEMBLY PROCEDURE  
6
f6-06_2.wmf  
Figure 6-6. 8842A Disassembly (cont)  
6-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f6-06_3.wmf  
Figure 6-6. 8842A Disassembly (cont)  
6-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
DISASSEMBLY PROCEDURE  
6
f6-06_4.wmf  
Figure 6-6. 8842A Disassembly (cont)  
6-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
6-29. True RMS AC PCA Removal (Option -09 Only)  
The True RMS AC PCA should be removed by reversing the last three steps in Figure  
809-1 (see Section 8).  
1. Release the True RMS AC PCA from the chassis by pulling the four plastic latches  
upward (Figure 809-1E).  
2. Raise the True RMS AC PCA slightly, and disconnect the red lead from the  
connector (J301) located on the Main PCA (Figure 809-1C).  
3. Disconnect the ribbon cable from the Main PCA by releasing the ribbon connector  
latches (push outward; see Figure 809-1D) and pulling the ribbon cable directly  
outward from the connectors (Figure 809-1C).  
4. Lift the True RMS AC PCA out of the chassis.  
6-30. IEEE-488 Interface PCA Removal (Option -05 Only)  
The IEEE-488 Interface PCA should be removed by reversing the last four steps in  
Figure 805-1 (see Section 8):  
1. Remove the two jack screws and washers from the rear panel IEEE-488 connector  
(Figure 805-1H).  
2. Release the IEEE-488 Interface PCA from the chassis by pulling the two plastic  
latches upward (Figure 805-1H).  
3. Raise the forward edge of the IEEE-488 Interface PCA slightly, pull the PCA  
towards the front panel (guiding the IEEE-488 jack and BNC connectors out of the  
rear panel), and lift the PCA out of the chassis (Figure 805-1G).  
4. Disconnect the ribbon cable from the IEEE-488 Interface PCA by releasing the  
connector latches (push outward; see Figure 805-1F) and pulling the ribbon cable out  
from the connector (Figure 805-1E).  
6-31. Main PCA Removal  
1. Disconnect the leads from the four front panel input terminals and the four rear panel  
input terminals by unplugging them. (Refer to Figure 6-6C.)  
2. Remove the cable harness from the two cable clamps on the side of the instrument  
chassis. (Figure 6-6D.) Lift the cable harness clear of the sidewall cable guide.  
3. Remove the front panel fuse by pressing in the lip of the 2A input terminal slightly  
and rotating it 1/4-turn counterclockwise (Figure 6-6E).  
4. (Disregard this step if the IEEE-488 Interface was installed.) Disconnect the ribbon  
cable from the rear panel insert by pushing outward on the snap tab on either side of  
the ribbon cable connector.  
5. Disconnect the two ribbon cables from the Display PCA by pulling the two plastic  
pull tabs directly outward from the Display PCA.  
6. Remove the two mounting screws on either side of the rear panel power receptacle.  
7. Disconnect the green power supply ground lead from the rear panel mounting stud.  
(The stud is located near the rear panel power receptacle. See Figure 6-6F.)  
8. Remove the Line Voltage Selection Switch (LINE SET) PCA as follows (Figure 6-  
6G):  
a. Remove the upper screw that holds the LINE SET PCA to the upper rear panel  
standoff.  
6-30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
DISASSEMBLY PROCEDURE  
6
b. Unplug the ribbon cable from the Main PCA and lift out the LINE SET PCA.  
9. Remove the push rod for the CAL ENABLE switch as follows (Figure 6-6H):  
a. While supporting the white plunger of the CAL ENABLE switch with a finger,  
pop the push rod off the switch plunger by pulling the push rod directly upward.  
b. Rotate the push rod 90 degrees toward the center of the instrument.  
c. Pull the push rod toward the rear panel and remove it.  
10. Remove the FRONT/REAR switch push rod as follows (Figure 6-6I):  
a. Insert a blade-type screw driver in the slot visible on the top of the  
FRONT/REAR switch push rod at the junction of the push rod and the switch.  
b. Twist the screwdriver slightly to release the push rod from the switch shaft, then  
pull the FRONT/REAR switch push rod out through the front panel.  
11. Place the chassis on its side.  
12. Remove the POWER switch push rod as follows (Figure 6-6J):  
a. Insert a blade-type screwdriver in the slot visible on the top of the POWER  
switch push rod at the junction of the push rod and the switch.  
b. Twist the screwdriver slightly to release the push rod from the switch shaft, then  
pull the rear of the POWER switch push rod out through the bottom of the  
chassis.  
c. Lift the push rod out and toward the rear panel, and remove it.  
13. Remove the two screws fastened to the transformer bracket, then remove the bracket.  
14. Remove the Main Shield as follows (Figure 6-6K):  
a. Remove the screw that fastens the Main Shield to the Main PCA.  
b. Grasp the Main Shield supports on one side of the instrument and pull the  
supports toward the center of the chassis, bowing the Main Shield. Remove the  
main shield.  
15. Release the six plastic latches that hold the Main PCA to the chassis by pulling the  
latches upward (Figure 6-6L).  
16. Lift the front end of the Main PCA upward about 3 inches.  
17. Free the white lead from the 2A INPUT tower as follows (Figure 6-6M):  
a. Guide the wire, spring and fuse contact toward the front panel.  
b. Thread the spring and fuse contact through the hole in the front end of the tower.  
18. Slide the Main PCA forwards until it is free of the chassis.  
6-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
6-32. Front Panel Disassembly  
1. Holding the chassis vertically (with the front panel downward), remove the mounting  
screws from the four corners of the Display PCA (Figure 6-7A).  
2. Holding the chassis vertically (now standing the instrument on the rear panel), pull  
the front panel off the chassis and set it aside (Figure 6-7B).  
3. Remove the the display assembly (Display PCA, spacer matrix, and keypad) from the  
chassis as follows (see Figure 6-7C):  
a. Release the two plastic tabs on the front of the chassis.  
b. Let the bottom edge of the display assembly swing toward the rear of the  
instrument.  
c. Pull the Display PCA toward the bottom of the chassis.  
4. Separate the spacer matrix from the Display PCA by releasing the two pairs of plastic  
snap tabs on the back of the Display PCA (Figure 6-7D).  
5. Remove the keypad from the spacer matrix.  
CAUTION  
The vacuum fluorescent display should not be removed from  
the Display PCA; these are supplied as one part.  
6. Remove the display window from the front panel as follows:  
a. Slide the window upward (away from the buttons) about 1/32 inch (Figure 6-8A).  
b. Push the window directly outward from the front panel (Figure 6-8B).  
6-33.REASSEMBLY PROCEDURE  
To reassemble the instrument, proceed as follows:  
1. Assemble the front panel assembly by reversing the front panel disassembly  
procedure (Figure 6-7). (It is easiest to lay the keypad on the Display PCA before  
installing the spacer matrix on the Display PCA.)  
CAUTION  
The four Display PCA mounting screws are self-tapping. To  
avoid damaging the threads, ensure the screws are threaded  
properly before tightening. Do not overtighten them.  
2. Turn the chassis upside down.  
3. Install the Main PCA through the bottom of the chassis as follows:  
6-32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
REASSEMBLY PROCEDURE  
6
NOTE  
When installing the Main PCA, guide the rear ribbon cable around the  
shield connected to the rear panel so that the cable is next to the side of the  
chassis. Make certain that the cable is not pinched between the shield and  
the Main PCA.  
a. Slide the Main PCA toward the rear panel, and position the power connector and  
fuse to fit through their respective openings in the rear panel.  
b. Reinstall the white lead in the 2A INPUT tower and reinstall the 2A fuse in the  
front panel.  
c. Make sure the six plastic latch heads are extended. Lower the Main PCA into  
position on the chassis and guide the six plastic latches into the circuit board  
supports on the chassis. Press the latch heads to lock the board in the chassis.  
Refer to Figure 6-6L. 4. Install one side of the Main Shield; bow it to install the  
the other side, and secure it to the Main PCA with the retaining screw. 5. Place the  
transformer bracket back into position and fasten down with the two screws.  
6. Replace the push rod for the POWER switch as follows:  
a. Insert the button end of the push rod into the rear of the front panel.  
b. While supporting the opposite side of the plunger with your finger, snap the other  
end of the push rod onto the POWER switch plunger. Refer to Figure 6-6J.  
7. Set the chassis right side up.  
8. Insert the FRONT/REAR switch push rod through the front panel and snap it into  
place.  
9. Reinstall the CAL ENABLE switch push rod by inserting the cylindrical end of the  
push rod into the rear of the front panel, then snapping it onto the CAL ENABLE  
switch plunger.  
6-33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f6-07.wmf  
Figure 6-7. Front Panel Disassembly  
6-34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
REASSEMBLY PROCEDURE  
6
f6-08.wmf  
Figure 6-8. Removing the Display Window  
CAUTION  
Make certain that the CAL ENABLE switch shaft is in the out  
(disabled) position after the CAL ENABLE push rod is installed.  
If the 8842A is switched on with the CAL ENABLE switch in the  
enabled position, the 8842A may require recalibration.  
10. Position the slot in the lower edge of the Line Voltage Selection Switch PCA in the  
slot on the lower rear panel standoff. Secure the top of the Line Voltage Selection  
Switch PCA to the upper standoff using the single mounting screw, and plug the  
ribbon cable into the Main PCA.  
11. Connect the power supply ground lead to the rear panel mounting stud. (The stud is  
located near the rear panel power receptacle as shown in Figure 6-6F.)  
WARNING  
TO AVOID ELECTRIC SHOCK, ENSURE THAT THE POWER  
SUPPLY GROUND LEAD IS FIRMLY ATTACHED TO THE  
REAR PANEL MOUNTING STUD.  
12. Attach the two mounting screws on either side of the rear panel power receptacle.  
13. Connect the two ribbon cables to the Display PCA to the connectors. Push the cables  
straight in to avoid damage.  
14. Reinstall the harness in the sidewall cable guide, and secure the harness to the chassis  
with the cable clamps.  
15. Connect the leads to the four front panel input terminals according to the color codes  
marked on the rear side of the Display PCA.  
16. Connect the leads to the four rear panel input terminals following the color codes as  
shown in Figure 6-6B.  
17. (Option -05 only) Install the IEEE-488 Interface PCA according to the instructions in  
Section 8.  
18. (Option -09 only) Install the True RMS AC PCA according to the instructions in  
Section 8.  
19. Slide the case and rear bezel onto the chassis.  
20. Install the two rear panel mounting screws.  
21. Install the case grounding screw in the bottom of the case.  
6-35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
WARNING  
TO AVOID ELECTRIC SHOCK, ENSURE THAT  
THE GROUNDING SCREW IS FIRMLY  
ATTACHED TO THE CASE BOTTOM.  
6-34.INTERNAL FUSE REPLACEMENT  
CAUTION  
For fire protection, use exact fuse replacements only.  
The 8842A has an internal 3A 600V slow-blow fuse (F301) in series with the 2A input  
terminal. To replace this fuse, remove the case according to the disassembly instructions.  
The fuse is held in fuse clips on the Main PCA. Do not use makeshift fuses or short-  
circuit the fuse holder.  
6-35.EXTERNAL TRIGGER POLARITY SELECTION (Option -  
05 Only)  
The EXT TRIG input is factory-configured with negative polarity (trigger on falling-  
edge). This polarity is set by jumper E902 on the IEEE-488 Interface PCA. To select  
positive polarity (trigger on rising-edge), remove jumper E902 and add jumper E903.  
6-36.TROUBLESHOOTING  
The 8842A is designed to be easily maintained and repaired. Both the analog and digital  
circuits have built-in diagnostic self-tests and troubleshooting modes to facilitate  
troubleshooting and repair. The instruments circuits allow troubleshooting and repair  
with basic electronic troubleshooting equipment such as a multimeter and oscilloscope.  
The troubleshooting mode in the digital controller circuitry generates special test signals  
to allow troubleshooting and repair without a special test signal generator or complex  
logic analyzer. Using the information in this section, a technician should be able to  
troubleshoot and repair the 8842A very efficiently. There is also a troubleshooting  
package available which utilizes the Fluke 9010A System Troubleshooter. The 8842A-  
9000 Troubleshooting Kit is described in detail in Section 8.  
6-37. Initial Troubleshooting Procedure  
WARNING  
TO AVOID INJURY OR EQUIPMENT DAMAGE, USE EXACT  
REPLACEMENT PARTS FOR ALL PROTECTION COMPONENTS.  
When a problem occurs in the 8842A, first verify the problem is actually in the  
instrument. If the problem occurs when the instrument is in a system, check to see if the  
same problem exists when under local control. Watch the display as the instrument is  
turned on to see if any of the digital self-test error codes appear indicating a digital  
failure. If the malfunction does not involve the True RMS AC or IEEE-488 options,  
remove the option(s) from the instrument before proceeding.  
6-36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
If the display lights up, perform the self-test by pressing the SRQ button for 3 seconds.  
(Remember, the input terminals must be disconnected from the test leads during the self-  
tests. Otherwise, the 8842A may indicate errors are present.) The test numbers will  
appear consecutively. "ERROR" will appear on the display if a test should fail. The  
8842A can be held in each of the test configurations by momentarily pressing the SRQ  
button. (Press any button to continue the tests.) With the description of the self-tests  
given below, it may be possible to isolate the failure. For reference, the states of various  
switches and logic lines are shown in Table 6-16 for each function, range, and reading  
rate.  
If only one or a few failures appear in the self-tests, the problem is usually in the DC  
Scaling circuit. By carefully analyzing which failure(s) occurred, the fault can be located  
to within a few components. (Table 6-17 shows which components are exercised by each  
of the analog tests.) The heading DC Scaling Troubleshooting provides detailed  
instructions on locating and repairing DC Scaling circuit problems. However, before  
troubleshooting the DC Scaling circuit, all of the power supply levels should be measured  
to verify they are within the limits specified in Table 6-23 under Power Supply  
Troubleshooting, later in this section.  
6-37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
Table 6-16. Overall State Table  
t6-16_1.wmf  
6-38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
Table 6-16. Overall State Table (cont)  
t6-16_2.wmf  
6-39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
Table 6-17. Circuitry Tested by the Analog Self-Tests  
t6-17.wmf  
Some failures will cause many self-tests to fail. If this occurs, the fault is usually in the  
Track/Hold circuit, the A/D Converter, the Digital Controller circuit, or the Power  
Supply. Again, measure all of the power supply levels according to the limits specified in  
Table 6-23. The next step is to isolate the problem to a specific section.  
If the self-tests display a large number of errors or if readings are noisy and/or in error,  
the problem is usually in the A/D Converter or Track/Hold circuit. (A large number of  
errors can also be caused by a problem in the Ohms Current Source.) To isolate the  
problem, connect a jumper between TP103 and Reference Low (TP306, or the L-shaped  
shield around U202). The display should typically read less than approximately 35 counts  
(i.e., +/-.000XX where XX is less than 35) on the 2V dc range. If a good reading can be  
obtained (less than approximately 35 counts), the A/D Converter and Precision Voltage  
Reference circuits are most likely good. A more conclusive test can be made by  
connecting a low-impedance dc source between Reference Low and TP103 with an  
output voltage between -2.0V and +2.0V. The reading on the display will be of opposite  
polarity to the voltage applied to TP103. (Disconnecting one end of R318 will usually  
make it possible to display readings within 0.1% to 0.5% of the actual input.) After it has  
been determined that the A/D Converter or the Track/Hold circuit is not functioning  
properly, proceed to the corresponding heading for detailed troubleshooting instructions  
and guidelines.  
6-40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
A failure in the instrument may cause the 8842A to display random patterns or nothing at  
all. Usually, analog circuit failures do not cause the display to go blank or display random  
patterns. The best place to start troubleshooting a "dead" instrument or an instrument with  
a non-functional display is to check the power supply with a voltmeter for proper levels  
and to use an oscilloscope to check the supplies for oscillations. If all of the supplies are  
working correctly, check the clock for the In-Guard µC at U202-2. The signal should be  
an 8 MHz sine wave approximately 3.5V peak-to-peak. Then check the 1 MHz output of  
the A/D IC (U101) at U212-3. (If not present, check at the A/D IC at U101-14.) The  
signal should be a 1 MHz square wave approximately 5V peak-to-peak. The 8 MHz sine  
wave is generated by the clock circuit of the In-Guard µC, and the 1 MHz signal is the 8  
MHz signal divided by a counter in the A/D IC. If the clock signals are correct, proceed  
to the heading Digital Controller Troubleshooting, below, for detailed troubleshooting  
instructions.  
If a problem occurs in the keyboard/display area, the instrument may appear to be totally  
inoperative even when the measurement circuitry is still functional. The heading Digital  
Controller Troubleshooting provides detailed instructions on locating problems in the  
display/keyboard system.  
Finally, as in most processor-based systems, there are communication links between the  
various parts of the system. Specifically, in the 8842A, there is a bus interface between  
analog and digital control circuits and a guard-crossing interface between logic circuits  
which may be separated by large potentials. Failures in these links can generate problems  
that may be difficult to locate and repair. However, such failures will in turn cause  
failures in some analog and or digital section. Thus, indirectly, troubleshooting the  
affected section will lead to correction of problems in the internal bus or guard-crossing  
circuit.  
6-38. Diagnostic Self-Tests  
To run the diagnostic self-tests, disconnect the test leads and press the SRQ button for 3  
seconds. If the test leads are left attached to the input terminals the 8842A may indicate  
errors are present (most likely, errors 5, 7, 8, 9, and 10). Also, if the FRONT/REAR  
switch is in the REAR position, the 8842A skips tests 3 and 4, and if Option -09 is not  
installed, the 8842A skips tests 1, 2, and 3. For all tests, there is a 0.5 second delay period  
before any readings are taken. The tests are all contingent on the A/D Converter being  
properly calibrated, but do not depend on the Offset and Gain Calibration constants.  
Failing the tests indicates that key portions of the 8842A are not performing properly.  
Passing the tests gives approximately a 90% probability that all VDC ranges and range r6  
of 2-wire ohms can be calibrated. Passing the tests also gives a reasonable probability  
that it will give accurate measurements in VDC and range r6 of 2-wire ohms. However,  
passing the tests does not guarantee that the instrument can be calibrated in VAC, mA  
DC, mA AC, 4-wire ohms, or ranges r1 to r5 of 2-wire ohms.  
NOTE  
If the A/D Converter or Precision Voltage Reference is not working, all  
analog tests would show an error. If the A/D Converter is not calibrated,  
tests 7, 15, 19 could show an error.  
If the analog self-tests indicate an error, it may be possible to isolate the problem as  
follows:  
1. While the error code is being displayed, press the SRQ button. This latches the  
8842A into the particular test configuration.  
2. Referring to Table 6-18, check that the test point voltages are as shown using another  
DMM.  
6-41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-39. Self-Test Descriptions  
Table 6-18. Self-Test Voltages  
TEST POINT  
TEST NUMBER  
VOLTAGE  
1
2
3
4
TP803  
TP803  
TP803  
TP103  
±5 mV dc  
±5 mV dc  
±5 mV dc  
T/H output waveform for zero input (Figure 6-14)  
5
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
TP302  
±5 mV dc  
6
±5 mV dc  
7
+50 mV dc typical  
8
+11.5V dc typical  
9
+11.5V dc typical  
10  
11  
12  
13  
14  
+4.5V dc with possibly 1V ac (p-p) at 10 Hz  
+4.5V dc with possibly 1V ac (p-p) at 10 Hz  
+4.5V dc with possibly 1V ac (p-p) at 10 Hz  
+4.5V dc with possibly 1V ac (p-p) at 10 Hz  
+4.5V dc with possibly 1V ac (p-p) at 10 Hz  
+50 mV dc typical  
15,22  
16  
+49 mV dc typical  
17  
+53 mV dc typical  
18  
+59 mV dc typical  
19  
<±5 mV dc  
20  
+59 mV dc  
21  
<±5 mV dc  
Note: To measure these correctly, each test must be stopped using the SRQ button.  
Also use TP306 (or L-shaped shield around U202) as Reference low.  
TEST 1: 200 VAC, Zero  
Configures the 8842A in the 200V ac range (except that K801 is opened) and measures  
the open-circuit floor reading. In this range, the first and second stage buffers effectively  
divide any noise at the input terminals by 100. This test should be fairly immune from  
noise because the input terminals are always open-circuited except for capacitive  
feedthrough across K801.  
TEST 2: 700 VAC, Zero  
Configures the 8842A in the 700V ac range and measures the open-circuit floor reading.  
In this range, the open-circuit reading is divided by 1000. Again, K801 is opened to  
reduce sensitivity to external noise.  
TEST 3: mA AC, Zero  
Configures the 8842A exactly as in the mA AC function and takes a reading of the  
voltage across the 0.1current shunt at the slow reading rate.  
TEST 4: mA DC, Zero  
6-42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
Configures the 8842A in the mA DC function and the slow reading rate, and measures  
the reading across the 0.1current shunt. This test should be fairly immune to outside  
noise because the total driving impedance is typically less than 1 k. The reading is not a  
perfect zero because of the offsets generated by charge injection of U302 and the T/H  
Amplifier (X10 configuration).  
TEST 5: 200 VDC, Zero  
Configures the 8842A in the 200V dc range and slow reading rate. Input noise is divided  
by 100. Assuming no input of any kind, the T/H Amplifier is essentially shorted to  
ground by 100 kand filtered by the 3-pole analog filter. Any non-zero reading under  
quiet input conditions is due to the offset of the T/H Amplifier (X1 configuration).  
TEST 6: 1000 VDC, Zero  
Configures the 8842A exactly as in the 1000V dc range and slow reading rate, with input  
noise being divided by 1000. The reading is very close to zero because of the inherent 2  
kdriving impedance to the T/H Amplifier (X1 configuration).  
TEST 7: 1000 VDC + 20 MΩ  
Couples the 1000V dc range and 20 Mcurrent source together. The result is nominally  
500 nA through the 10 Minput divider. Since the 1000V dc range senses this voltage at  
the divide-by-1000 point of the scaling circuit, the reading should be 5 mV, or 500 counts  
at the A/D Converter. This test could indicate an error if input capacitance is greater than  
1000 pF.  
TEST 8: 20 VDC + 20 MΩ  
Puts the DC Scaling circuit into the 20V dc range and the Ohms Current Source into the  
20 Mrange. The infinite input impedance of the 20V dc range causes the 20 MΩ  
current source to be clamped at its maximum open circuit voltage, typically 12V. The  
20V dc range scales this voltage and presents the A/D Converter with 1.15V, or 115,000  
counts.This is a good test to ensure that the maximum open-circuit voltage of the Ohms  
Current Source is less than 13V. This test is susceptible to capacitance greater than 0.01  
µF at the input terminals.  
TEST 9: 20 VDC + 2000 kΩ  
Puts the DC Scaling circuit in the 20V dc range and the Ohms Current Source in the 2000  
krange. The infinite input impedance of the 20V dc range causes the 2000 kcurrent  
source to be clamped at typically 11.5V. The reading at the A/D Converter should be  
1.15V. Again tests that the maximum open-circuit voltage of the Ohms Current Source is  
less than 13V. Capacitances greater than 0.1 µF at the input terminals can cause an error.  
TEST 10: 2 VDC + 2000 kΩ  
Puts the DC Scaling circuit in the 2V dc range and the Ohms Current Source in the 2000  
krange, except that its maximum open-circuit voltage is limited to less than 6.5V in  
this configuration. This test, as well as tests 11-13, checks clamps Q312 and Q313 and  
the analog filter. The reading at the A/D Converter should be an overload. Capacitances  
greater than 0.5 µF at the input terminals can cause an error.  
TEST 11: 200, Overrange  
TEST 12: 2 k, Overrange  
TEST 13: 20 k, Overrange  
TEST 14: 200 k, Overrange  
6-43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
These tests put the 8842A in the respective range of the 2-wire ohms function. They  
check that each range of the Ohms Current Source has enough compliance voltage to  
overload the dc front end.  
TEST 15: 1000 VDC + X10 T/H + 20 MΩ  
Puts the Ohms Current Source in the 500 nA range. The resulting current through Z302  
(the 10 Minput divider) causes a nominal divider output voltage of 50 mV. The T/H  
Amplifier is in X10; thus the A/D Converter sees 50 mV, or 5000 counts. This test can be  
susceptible to input capacitances above 1000 pF.  
TEST 16: 200 VDC + 200 kΩ  
TEST 17: 200 VDC + 20 kΩ  
TEST 18: 200 VDC + 2 kΩ  
These three tests put the DC Scaling circuit in the 200V dc range and the Ohms Current  
Source in the respective current range. The 10M ohm input divider (Z302) senses the  
maximum open-circuit voltage of each range of the current source, and the T/H Amplifier  
(X1) presents the compliance voltage divided by 100 to the A/D Converter. Nominal  
readings should be 49 mV for Test 16, 53 mV for Test 17, and 59 mV for Test 18. All  
three tests have a pass limit of 65 mV, insuring that no more than 6.5 volts appears at the  
input terminals.  
TEST 19: 200 VDC, Filter On  
Test 18 (above) ties the 1 mA range of the Ohms Current Source into the 200V dc range,  
with the three-pole analog filter on, such that the A/D reads 59 mV nominal. Test 19  
decouples the DC Scaling circuit from the Ohms Current Source; the In-Guard µC waits  
28 ms and determines if the voltage at Z302-3 (the divide-by-100 point of the 10 MΩ  
input divider) has not discharged to zero volts, due to the long time constant of the filter.  
TEST 20: 200 VDC + 2 k, Filter Off  
TEST 21: 200 VDC, Filter Off  
TEST 22: x100 T/H  
The T/H Amplifier is configured in x100 mode. DC stimulus of 5 mV (similar to test 15)  
is amplified to 500 mV, which is applied to the A/D converter. This test can be  
susceptible to input capacitance above 1000 pF.  
Test 20 ties the 1 mA range of the Ohms Current Source into the 200V dc range, with the  
3-pole analog filter off, such that the A/D Converter reads 59 mV nominal. Test 21 then  
decouples the DC Scaling circuit from the ohms current source; the In-Guard µC waits 28  
ms and determines if the voltage at Z302-3 is at zero volts.  
TEST 25: In-Guard µC Internal RAM (U202)  
A GALPAT test is performed on the internal RAM of the In-Guard µC. If there are any  
errors, ERROR 25 is displayed. This test is performed only upon powerup.  
TEST 26: Display RAM (U212)  
A pattern is written to the Display RAM and read back for comparison.If there are any  
differences ERROR 26 is displayed.  
TEST 27: In-Guard µC Program Memory (U202)  
A two-byte check sum is calculated over the entire 4K Internal Program Memory and  
compared with the checksum bytes at the end of that memory. A special add and shift  
algorithm minimizes the possibility of double errors cancelling. If something is wrong  
with the Internal Program Memory, ERROR 27 is displayed.  
6-44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
TROUBLESHOOTING  
6
TEST 28: External Program Memory (U222)  
A two-byte check sum is calculated over the entire 4K External Program Memory and  
compared with the checksum bytes at the end of that memory. A special add and shift  
algorithm minimizes the possibility of double errors cancelling. If something is wrong  
with the External Program Memory, ERROR 28 is displayed.  
TEST 29: Calibration Memory (U220)  
Numerous single-byte checksums are placed in the Calibration Memory, one at the end of  
each group of calibration constants. They are calculated in such a way that the single-byte  
sum of all bytes in the Calibration Memory add to zero (all carries discarded) and the  
single-byte sums of each group also add to zero. A new checksum is calculated and  
written to Calibration Memory each time a full or partial calibration is performed. If the  
Calibration Memory is not properly configured or not working correctly, ERROR 29 is  
displayed. The accuracy of the 8842A is suspect.  
6-40. Digital Controller Troubleshooting  
The basic strategy in troubleshooting the Digital Controller circuit is to check first  
whether the In-Guard Microcomputer (µC) system is functional, starting with the In-  
Guard µC itself (U202). Most of this circuitry is tested using the specially provided In-  
Guard Troubleshooting Mode.  
If the In-Guard µC system proves to be functional, then basic instrument control is  
assured and troubleshooting efforts can proceed in one of two directions. If the display  
and keyboard appear to be malfunctioning, then they should be checked next. (See  
Display System, below.) If the display and keyboard are functioning correctly, you can  
omit Display System troubleshooting and proceed to verify that signals are arriving  
correctly at the inputs of the analog control devices. (See Analog Control Signals, below.)  
If these are also correct, the digital controller is functioning correctly, and you can  
proceed to the appropriate analog troubleshooting procedure.  
NOTE  
For the convenience of the following tests, Options -05 and -09 should be  
removed if present. They should only be removed in the power-off  
condition.  
6-41. IN-GUARD MICROCOMPUTER SYSTEM  
This procedure is performed entirely in the In-Guard Troubleshooting Mode. This mode  
is established by shorting TP205 (U202-38) to Reference Low (TP306, or the L-shaped  
shield around U202) prior to turning on the instrument. Refer to Figure 6-9. To maintain  
this mode, the short must remain in effect after the instrument is turned on. When this is  
done, the µC programs U202-38 as an input (it is normally an output) to preclude any  
possibility of damage due to the short.  
CAUTION  
To avoid damaging the µC, the short must be initiated before  
the instrument is turned on, not after.  
The In-Guard Troubleshooting Mode also programs all the normal port outputs to display  
a 1 kHz square wave except that the IEEE-488 output (U202-4) sends the word "55"  
repeatedly at a rate of 2,000 words/second and the A/D trigger (U202-40) is a square  
wave at its normal frequency of 80 Hz, and DM and P23 stay high. (The data received at  
U202-5 is meaningless.) Adjacent port outputs display opposite phases of the 1 kHz  
square wave. All µC pins that are normally only programmed as inputs are also  
6-45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
programmed as inputs to prevent contentions between them and the outputs from other  
ICs which drive them. Data coming into all µC inputs (except pin 38) is ignored.  
NOTE  
If the A/D IC (U101) is working properly, its watchdog timer briefly  
interrupts all of the In-Guard Troubleshooting Mode signals every 1.5 sec  
for a period of about 0.2 sec. (The signals are then re-established.) If this  
occurs, the main counter in U101 and its watchdog timer are operating  
correctly. (See step 6, below.)  
When the test is complete, turn off the 8842A and remove the short from TP205.  
6-42. In-Guard Microcomputer  
While the 8842A is in the In-Guard Troubleshooting Mode, check the following in the  
order shown:  
f6-09.wmf  
Figure 6-9. U202 Pin Diagram  
6-46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
1. Power supplies: +5V dc at U202-1; 0V dc U202-11.  
2. µC clock output: 8 MHz at U202-2,-3.  
3. Trigger line U202-40 (TP201): Square wave, 50% duty, low 0V, high 3.8V  
(nominal). The period of the trigger signal should be 12.500 ms for 60 Hz line.  
4. Interrupt from A/D (U202-39): Normally low, duration 48 µs occurs approximately  
5450 µs after falling edge of trigger signal on TP201 (U202-40).  
5. Guard-crossing test pattern (U202-4): Waveform C (see Figure 6-10).  
6. Interrupts from watchdog timer (U202-6): Pulsed low for 0.2 sec every 1.5 sec,  
exponential rise between pulses.  
7. Output test patterns (see Waveforms A and B in Figure 6-10): 1 kHz square wave on  
indicated pins, 50% duty cycle, low 0V, high 3.8V nominal. (The waveforms are  
interrupted every 1.5 sec for 0.2 sec due to interrupts from the watchdog timer.) To  
observe these patterns, remove U220, attach a logic clip to address latch U219 and  
sync on U219-3 for Reference Waveform A on channel 1 of a dual trace scope.  
Compare channel 1 with waveforms at U202-10,-14,-16,-18,-20,-22,-24,-26,-28.  
These should all be the same as reference Waveform A (including phase). Then  
compare channel 1 with waveforms at U202-13,-15,-17,-19,-21,-23,-25,-27,-29,-  
33.These should be the same as Waveform B, which is simply the opposite phase of  
Waveform A.  
6-43. Address Latch (U219)  
Verify that U219-2, -6, -7, -12, -13, -16, -17 are the same as Waveform A (see Figure 6-  
10) on U219-3. Verify that U219-4, -5, -8, -9, -14, -15, -18, -19 are the same as  
Waveform B. The waveforms should be interrupted every 1.5 sec for 0.2 sec due to  
interrupts from the watchdog timer.  
f6-10.wmf  
Figure 6-10. Waveforms for In-Guard Troubleshooting Mode  
6-44. External Program Memory (XU222)  
Sync on U219-3. Verify that XU222-3, -5, -9, -12, -15, -17, -19, -23, -24 are the same as  
Waveform A (see Figure 6-10) on U219-3. Verify that XU222-4, -6, -8, -10, -11, -13, -  
6-47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
16, -18, -21, -25 are the same as Waveform B. The waveforms should be interrupted  
every 1.5 sec for 0.2 sec due to interrupts from the watchdog timer. (Note: XU222 pins  
refer to a 28-pin socket.)  
6-45. Calibration Memory (U220)  
Sync on U219-3. Verify that U220-1, -3, -5, -7, -10, -13, -15, -17, -22 are the same as  
Waveform A (U219-3). Verify that U220-2, -4, -6, -8, -9, -11, -14, -16, -19, -23 are the  
same as Waveform B. The waveforms should be interrupted every 1.5 sec for 0.2 sec due  
to interrupts from the watchdog timer.  
6-46. Relay Buffer (U201)  
At this point it is necessary to return the 8842A to the normal operating mode by turning  
the power switch off, removing the short from TP205 (U202-38), and installing the True  
RMS AC option, if present, so that U201-14, -15 may be checked. Power up the  
instrument. Unlike the previous checks, outputs are steady state and therefore do not  
require a sync signal. Logic "1" is approximately 4.3V dc.  
Check that U201-14 is high (4.3V) for mA AC and all ranges of VAC, and low for all  
other functions.  
Check that U201-15 is high (4.3V) for mA AC and 200V ac and 700V ac ranges, and low  
for all other functions.  
Check that U201-16 is high (4.3V) for all 2-wire and 4-wire ohms ranges, and low for all  
other functions.  
Check that U201-17 is high (4.3V) for the lowest three VDC ranges and all 2-wire and 4-  
wire ohms ranges, and low for all other functions.  
6-47. 3-to-8 Chip Select Decoder (U208)  
Make the following checks in the normal operating mode using the fast reading rate and  
any function and range. These sequences begin 5.5 ms after the A/D trigger, which is the  
falling edge at U202-40.  
Check U208-13 for 0.2 µs pulses, normally high, groups of 1, pulse spacing: 10 ms.  
Check U208-12 for 0.2 µs pulses, normally high, groups of 1, pulse spacing: 10 ms (10  
µs after pin 13).  
Check U208-11 for 0.2 µs pulses, normally high, groups of 1, pulse spacing: 10 ms (46  
us after pin 13).  
Check U208-15 for 0.2 µs pulses, normally high, groups of 13, group width: 100 µs,  
group spacing: 10 ms (230 µs after pin 13).  
Check U208-7 for 0.6 µs pulses, normally high, groups of 5, group width: 50 µs, group  
spacing: 10 ms (380 µs after pin 13).  
This concludes testing of the basic µC system. If the keyboard or display is still suspect at  
this point, proceed to Display System, below. Otherwise proceed to Analog Control  
Signals, below.  
6-48. DISPLAY SYSTEM  
The display/keyboard system is operated by a complex LSI IC (U212). Generally, this IC  
is checked indirectly by observing behavior of the simpler logic devices which it drives.  
If the keyboard is working at all, the 8842A display should be "frozen" to make the  
following tests. This places the 8842A in a special display test configuration. If it is not  
6-48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
possible to freeze the display, it should still be possible to observe the waveforms at  
U215, U213, U221, and U211 as described in the following paragraphs.  
To freeze the display, turn off the instrument, press the POWER switch and within 1  
second press the SRQ button. If all is well, all display segments will light and remain lit.  
Do not press any other buttons as that will release the display, allowing the instrument to  
resume its normal power-up sequence. This state should remain in effect for all of the  
following tests.  
f6-11.wmf  
Figure 6-11. Waveforms for Display Logic  
6-49. Display Control (U212)  
Check for the 1 MHz clock from the A/D IC at U212-3.  
NOTE  
The following waveforms are illustrated in Figure 6-11.  
6-50. 8-Bit Digit Driver (U215)  
Check for strobe waveforms 0-7 on U215-8, -1. Reference U215-8 for waveform  
STROBE ZERO. U215-7 is STROBE ONE, U215-6 is STROBE TWO etc. High level is  
3.8V to 4.3V and low is near 0V.  
6-49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Check for the same waveforms at outputs U215-11 through U215-18. (However, the high  
level should be approximately 30V.) If these waveforms are OK, then strobe decoder  
U213 and display control U212 are OK in this regard.  
6-51. 3-to-8 Strobe Decoder (U213)  
Check for strobe waveforms 0-7 on U213-4, -5, -6, -7, -9, -10, -11, -12. Reference U213-  
4 for STROBE ZERO. Check for strobe decoder inputs SL0, SL1, SL2 on U213-1, -2, -3  
respectively.  
6-52. 8-Bit Segment Driver (U217)  
Check that U217-1 through U217-7 all look like the waveform DIGIT DATA. High level  
is 3.8V to 4.3V.  
Check that U217-12 through U217-18 all look like the waveform DIGIT DATA except  
high level is approximately 30V.  
6-53. 4-to-7 Segment Decoder (U216)  
Check that U216-1, -2, -7 are low and U216-4, -6 and U216-9 through U216-15 look like  
the DIGIT DATA waveform.  
6-54. 8-Bit Digit Driver (U218)  
Check that U218-1 through U218-4 all look like the waveform DIGIT DATA. High level  
is 3.8V to 4.3V.  
Check that U218-15 through U218-18 all look like the waveform DIGIT DATA, except  
that the high level should be approximately 30V.  
6-55. Hex Inverter (U203)  
At this point the display should be "unfrozen" by pressing any button. The instrument  
should then complete the power-up self-test and begin normal operation. Then do the  
following:  
1. Check that U203-9 is the same as STROBE ONE and that U203-10 is STROBE ONE  
inverted.  
2. Check that U203-5 is the same as STROBE TWO and that U203-6 is STROBE TWO  
inverted.  
3. Check that U203-11 shows positive pulses 50 us to 300 us while repeatedly pushing  
front panel buttons in normal mode and that U203-10 shows the inverse.  
4. Check that the waveform seen at U208-15 is the same at U203-13, -4 and inverted at  
U203-3, -12.  
6-50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
TROUBLESHOOTING  
6
6-56. Hex Inverter (U221)  
Check that U221-5 is the same as STROBE ZERO and that U221-6 is STROBE ZERO  
inverted.  
6-57. Quad OR Gate (U211)  
Check U211-6 for 0.2 us pulses, normally high, in two groups of 3 and 15, group widths:  
50 and 100 us, group spacing: 10 ms (in fast reading rate).  
Check U211-8 for 0.4 us pulses, normally high, groups of hundreds, group widths: 3.5-4  
ms, group spacing: about 8 ms (variable).  
6-58. Keyboard Wiring  
Table 6-19 indicates which waveforms in Figure 6-11 are seen on keyboard inputs to  
U212 when each front panel button is pressed and held. For example, if the SRQ button is  
pressed and held, an inverted version of STROBE ZERO waveform is applied to U212-  
38. If range button "20" is pressed then inverted STROBE ONE is applied to U212-1, and  
so forth. Note that these waveforms are very noisy with many spikes. That is normal.  
Compare these waveforms with normal STROBE ZERO at U215-8.  
6-59. ANALOG CONTROL SIGNALS  
Table 6-19. Keyboard Wiring  
SIGNAL NAME  
PIN  
BUTTONS THAT  
PRODUCE  
BUTTONS THAT  
PRODUCE  
BUTTONS THAT  
PRODUCE  
INVERTED  
INVERTED  
INVERTED  
(STROBE ZERO)  
(STROBE ONE)  
(STROBE TWO)  
RL0  
RL1  
RL2  
RL3  
RL4  
RL5  
RL6  
U212-38  
U212-39  
U212-1  
U212-2  
U212-5  
U212-6  
U212-7  
SRQ  
LOCAL  
RATE  
20/mV  
200/mV  
2
VDC  
VAC  
2 WIRE kΩ  
4 WIRE KΩ  
mA DC  
OFFSET  
20MΩ  
20  
200  
TRIG  
2000  
mA AC  
EX TRIG  
Not used  
AUTO  
Table 6-20 is useful for determining whether the correct digital signals are being applied  
to the analog control devices indicated. Since most of these devices (the quad analog  
switches in particular) have no digital outputs, it cannot be determined directly whether  
the correct pattern is being latched. That determination must be made indirectly by analog  
means. Nevertheless, it is valuable to know whether the correct digital signals are  
reaching those devices.  
6-51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 6-20. Analog Control Devices  
DEVICE  
REF. DES.  
Relay Buffer  
U201  
U305  
U301  
U302  
U303  
U402  
U403  
U803 *  
U804 *  
U808 *  
Quad Comparator  
Quad Analog Switch  
Quad Analog Switch  
Quad Analog Switch  
Quad Analog Switch  
Quad Analog Switch  
8-Bit Latch  
Quad Analog Switch  
Quad Analog Switch  
* Option -09 only  
Correct operation of 8-bit latch U803, situated on the True RMS AC PCA (Option -09  
only), can be determined directly since all of its inputs and outputs are available. Again,  
outputs of quad analog switches U804 and U808 are not available and must be  
determined by analog means.  
6-60. Evaluating Static Signals  
Table 6-21 may be used to determine whether or not proper signals are reaching any  
particular analog control device. It may also be used to quickly exercise all of the devices  
before beginning analog troubleshooting if it is still unclear as to which devices are  
suspect. A number of the inputs to these devices are static which makes them particularly  
easy to check.  
For example, suppose quad comparator U305 appears not to be working. Connect a scope  
to U305-11 and step the 8842A through all functions and ranges in the following  
sequence:  
1. VDC: 20 mV, 200 mV, 2V, 20V, 200V, 1000V  
2. VAC: 200 mV, 2V, 20V, 200V, 700V  
3. 2 WIRE k: 200, 2k, 20k, 200k, 2M, 20M  
4. 4 WIRE k: 20, 200, 2k, 20k, 200k, 2M, 20M  
5. mA DC: 200 mA, 2000 mA  
6. mA AC (one range only)  
While doing this, observe the state of U305-11. As shown in Table 6-21, this 24-range  
sequence will produce the following pattern at U305-11:  
111010 00000 111100 1111100 000  
6-52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
Table 6-21. Analog Control Logic States  
t6-21.wmf  
Next move to U305-5 and repeat. The slow reading rate gives the following pattern at  
U305-5:  
111111 00000 111000 1111000 000  
6-53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
If the instrument is not in the slow reading rate, it gives the following pattern at U305-5:  
000000 00000 000000 0000000 000  
Next move to U305-7 and repeat. The pattern at U305-7 will be:  
000000 00000 000000 1111111 000  
6-61. Evaluating Dynamic Signals  
The procedure for evaluating the dynamic signals is only slightly more involved. For  
example, consider U302-7 in Table 6-21. At the right end of that row the table says to  
sync on U302-3. The sync pulse is negative-going. Apply it to channel 1 of a dual-trace  
scope syncing on the leading (negative-going) edge. Observe the target pin (U302-7) on  
channel 2 of the scope. While stepping through the 24 ranges observe the state of the  
target pin exactly when the sync pulse goes from low to high. (See Figure 6-12.) (This  
procedure works best in the fast reading rate since the repetition rate of the sync pulse on  
U302-3 is greater.) Using this procedure, the following pattern should be seen:  
00000 11111 000000 000000 0 1  
Note that the last eight rows in Table 6-21 are actually outputs of U803. Therefore,  
observing those pins proves not only that the control signals are correct but also that  
U803 itself is functioning correctly.  
6-62. DC Scaling Troubleshooting  
Whenever there is a failure in the DC Scaling circuit, first check the power supply  
voltages for all active components. (Supply voltages and pin numbers are listed in Table  
6-22.) A test of the bootstrap supplies for U306 is described later under this heading.  
After checking the power supplies, use an oscilloscope to check the digital logic input  
pins of quad analog switches U301, U302 and U303. These should show digital signals  
with high =>+3V and low =<+0.5V.  
f6-12.wmf  
Figure 6-12. Typical Dynamic Control Signals  
6-54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
Table 6-22. DC Scaling and Track/Hold Supply Voltages  
SUPPLY VOLTAGE PIN OR DEVICE  
+5V U303-20  
PIN OR DEVICE  
U301-6  
SUPPLY VOLTAGE  
+7.5V  
U301-10  
U301-20  
U301-11  
U302-6  
0V  
U303-11  
U304-4  
U304-7  
U305-3  
U305-12  
U307-4  
U307-7  
Q305,c  
Q306,c  
-8.2V  
-8.2V  
+7.5V  
+5V  
+7.5V  
-5V  
+5V  
0V  
U302-10  
U302-20  
U302-11  
U303-6  
-5.5V(Nom)*  
-15V  
+7.5V  
-5V  
+15V  
+5V  
0V  
+30V  
U303-10  
*With 0V input.  
-30V  
In the 20V range, any voltage applied to the HI INPUT terminal (relative to Reference  
Low) should be present at U306-3. If it isnt, trace the voltage from the HI INPUT  
terminal to U306-3 to isolate the problem.  
To check U306, select the VDC function and the 2V range. Measure the voltage at TP302  
while applying first 1V and then 0V (a short) across the HI and LO INPUT terminals. If  
1V and then 0V appear at TP302, U306 is probably OK. If not, the problem is in U306 or  
its bootstrap supplies (TP301 and TP303).  
To check the bootstrap supplies, put the 8842A in the 20V range and measure the voltage  
at TP301, TP302, and TP303. TP301 should be 6.3V (nominal) above TP302, and TP303  
should be 6.2V (nominal) below TP302. If the bootstrap supplies are operating correctly,  
measure the voltage at U306-3 and U306-6 for input voltages of +20V and -20V; if the  
voltage at U306-3 differs from U306-6, then U306 is bad.  
To check the dc input path after U306, short the HI and LO INPUT terminals and read  
the display. If zero is displayed for ranges r3 and r5 but not for r1, r2, and r4, then the  
signal path including Q311 and U301B is suspect. To check Q311, apply a 1V dc input in  
the 2V range and check that the voltage at the drain and source of Q311 is 1V. If not,  
Q311 or its driver is bad. If 1V appears at U301-16, but not at the display, then U301  
may be bad.  
If zero is not displayed for r3 and r5 with the HI and LO INPUT terminals shorted, then  
Z301 or U302D is probably bad.  
6-55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-63. Track/Hold Troubleshooting  
If a problem is suspected in the Track/Hold (T/H) circuit, first check the power supply  
voltages of all active components. (See Table 6-22.)  
Next, check the T/H output waveform at TP103 with an oscilloscope. Set the 8842A to  
the VDC function and 2V dc range, apply +1V dc across the HI and LO INPUT  
terminals, and trigger the scope from the falling edge of line not-TR (TP201). The  
waveform should look like that in Figure 6-13. The circuit may be checked as follows:  
1. Short the HI and LO INPUT terminals, and select the 2V dc range.  
2. Short U301-14 to ground (Reference Low). The 8842A should read within 10 counts  
of zero. (The actual value is not as important as its stability.)  
3. Connect U307-6 to U303-18, and monitor the voltage at TP103 using another  
multimeter. The 8842A should read about the same as the external multimeter, but  
with opposite sign.  
If the 8842A fails step 2 but not step 3, then U303 is bad. If the 8842A fails both steps 2  
and 3, then the fault is in the A/D Converter or the T/H Amplifier. To tell which, lift the  
end of R318 closest to the front panel (connected to U307-6) and apply an input of less  
than +2V to TP103. If the A/D Converter is OK, the 8842A will display the applied  
voltage with the opposite polarity. (For example, if you apply +1V, it should display -  
1V.) The readings may differ by a slight offset.  
6-64. Ohms Current Source Troubleshooting  
Malfunctions in the ohms functions can be caused by a fault in the Precision Voltage  
Reference, Ohms Current Source, or Ohms Protection. Malfunctions can also be caused  
by a fault in the DC Scaling circuit which loads the Ohms Current Source.  
First check the power-supply levels (see schematic). Then check all digital logic input  
pins of the quad analog switches (U402 and U403). These should show digital signals  
with high =>+3V and low =<+0.5V.  
To determine whether the Ohms Current Source is being loaded down by the DC Scaling  
circuit, select the VDC function and connect a 10 kresistor between the collector of  
Q404 and ground (Reference Low). (Selecting the VDC function opens K401, and  
configures the Ohms Current Source in the 20 krange.) If the voltage across the 10 kΩ  
resistor is 1V, then the Ohms Current Source is working (at least in the 20 krange), and  
the problem is probably due to a defect in the DC Scaling circuit.  
6-56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
f6-13.wmf  
Figure 6-13. Typical Output Waveforms for Track/Hold Circuit (TP103)  
To test whether the Ohms Current Source is actually being sourced out the HI and LO  
OUTPUT terminals, select the 20 krange and the 2-wire ohms function, connect a 10  
kresistor across the HI and LO INPUT terminals, and measure the voltage across this  
resistor with another voltmeter. There should be a 1V drop across the resistor.  
6-57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
If the ohms functions do not work in any range, check the supplies at U401 (+/-15V),  
U404 (+30V and -5V), U402 (+15V, +5V, and 0V), and U403 (+15V, +5V, and 0V), and  
check the -7V reference at R416. Also, test the Ohms Protection circuitry as follows:  
Select the 20 krange and 2-wire ohms function, connect a 10 kresistor to the HI and  
LO INPUT terminals, and bypass the protection circuitry by connecting the emitter of  
Q402 to the junction of R410 and R309. If a reading of 10 kis displayed, the protection  
circuitry is defective. To isolate the problem, successively short each part of the  
protection circuitry that is in series with the Ohms Current Source, until the display reads  
10 k.  
If the ohms functions work in only certain ranges, suspect resistor network Z401 or  
analog switches U402 or U403. To test the analog switches, select a defective range and  
connect a short across the switches that are supposed to be closed in that range. If the  
Ohms Current Source then works, one of the analog switches is probably bad. If the  
range still doesnt work, then Z401 is probably bad.  
To test the first stage of the Ohms Current Source, short U402-19 to Reference Low  
through a 2 kresistor and check that the voltage across R401 is 7.0V (nominal) and that  
the voltage at U401-6 is -4V (nominal). If the voltages are correct, the first stage of the  
Ohms Current Source (U401 and Q401) is working. If not, suspect U401 or Q401.  
(Under no circumstances should U401-6 ever be positive.)  
If the first stage of the Ohms Current Source is working, test the second stage as follows:  
1. Select the 20 krange and apply a 10 kinput.  
2. Check that the voltage between TP403 and U402-16 is +5V.  
3. Measure the voltage at U404-6 with respect to ground. If the voltage at U404-6 is  
negative, U404 is bad. (Under no condition should U404-6 ever be negative.)  
6-65. Precision Voltage Reference Troubleshooting  
If there is a failure of the Precision Voltage Reference, check the power supply levels at  
U702. U702 requires two supplies, +15V and -15V, which must be within approximately  
+/-5% of their nominal value. Using an oscilloscope, check that the power supplies and  
op amp outputs (U702-1 and U702-7) are free from ripple and oscillations.  
If the supplies are correct, check the output voltage levels at TP701 and TP702. The  
voltages should be +7.00000V +/-1000 ppm and -7.00000V +/-250 ppm. Also check that  
the reference amplifier output voltage (U702-1) is nominally +6.5V.  
If the outputs are grossly out of tolerance (e.g., stuck at +15V or -15V), the most likely  
cause is a bad op amp (U702) or open resistor network (Z701 or Z702). If the outputs are  
slightly out of tolerance, the most likely cause is a defective or out-of-tolerance resistor in  
Z701 or Z702. Because Z701 is precisely matched with U701, Z701 and U701 must be  
replaced as a matched set.  
Shorts between lands or runs can also cause small errors (10 ppm to several hundred  
ppm). Shorts between sense and output lands can cause small errors that are not related to  
resistor networks. Load regulation problems can also be caused by shorts between sense  
and load lines.  
In some rare cases, the op amps (U702A and U702B) could be out of spec, causing a  
small error. The maximum input offset voltage of the op amps used in the circuit is 3 mV.  
6-58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
6-66. A/D Converter Troubleshooting  
If there is a failure of the A/D Converter, all power supply levels should be checked at the  
op amps (U102 and U103) and the A/D IC (U101). The A/D Converter has a total of  
seven supplies: +15V, -15V, +5V, +7.5V, -8.2V, +7.00000V, and -7.00000V. All  
supplies should be within 5% of their nominal values except for the +7.00000V and -  
7.00000V reference supplies, which should be within +/-1000 ppm and +/-250 ppm  
respectively. The bootstrap supplies (lines BS1 and BS2) should be +7V and -7V (+/-  
10%) referenced to the + input of the A/D amplifier (U103-3).  
Troubleshooting the bootstrap supplies can often be made easier by putting the 8842A in  
EX TRIG (to stop the A/D Converter) and connecting the input of the A/D Converter  
(TP103) to INPUT LO (Reference Low on the schematic). The bootstrap supplies are  
then referenced to instrument common (Reference Low).  
NOTE  
For the following tests, set the 8842A to the VDC function and the 2V  
range, and trigger the oscilloscope from the falling edge of line not-TR  
(TP201).  
If all supplies are correct, the next most useful troubleshooting tool is the A/D output  
waveform at TP101, which can be checked with an oscilloscope. The waveform should  
look like the one shown in Figure 6-14 when the input voltage is at 50% of the selected  
range. Various portions of the waveform correspond to different parts of the A/D cycle.  
By examining the waveform, problems in the A/D Converter can be isolated down to one  
or two components.  
f6-14.wmf  
Figure 6-14. Output of A/D Amplifier (TP101)  
6-59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f6-15.wmf  
Figure 6-15. Waveforms at U101-24 and U101-25  
f6-16.wmf  
Figure 6-16. Typical Bus Data Line Waveform  
The waveform at the storage capacitors can often be used to locate leakage problems. The  
leakage can be due to contamination on the Main PCA or to defective switches in U101.  
Figure 6-15 shows the waveforms across storage capacitors C102 and C103 (U101-24  
and U101-25, respectively) for a specific input.  
The A/D Converter communicates with the In-Guard µC via the internal bus, which also  
goes to several other sections of the instrument. What looks like a problem in the A/D  
Converter may actually be caused by a problem in another section of the instrument  
which is loading down the bus data lines (U101-1, -2, -3, -38, -39, -40). A typical  
waveform at one of the data lines is shown in Figure 6-16. One of the data lines can be  
loaded down so that the In-Guard µC fails to recognize data sent over that line. If so, the  
amplitude of the signal of the bad line would be less than 3V peak-to-peak.  
One technique of finding an overloaded or shorted data line is to remove the In-Guard µC  
and drive one data line at a time through a 1 kresistor. Measure the voltage drop across  
a length of the line. Normally the voltage drop across the line is zero volts (less than 5  
µV). Voltage drops larger than 5 µV indicate a short. (The voltage drop is caused by  
excessive current flowing through the line.)  
When troubleshooting the A/D Converter it may be desirable to determine what the  
reading is at the A/D Converter when the display is definitely incorrect. A digital  
6-60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
problem between the A/D Converter and the In-Guard µC can cause erroneous or noisy  
readings or offsets. Similar problems may be caused by a failure of the Calibration  
Memory (U220) or by bad A/D calibration constants. (To check for bad A/D calibration  
constants, clear the calibration memory.)  
Readings at the A/D Converter can be determined by interpreting the waveform at the  
DAC output (TP102). Waveforms at TP102 for several input levels are shown in Figure  
6-18. The A/D reading can be calculated by knowing the weight of each bit and by  
weighting each nibble correctly. (The first nibble is weighted 1, the second 1/16, the third  
1/16 2 the fourth 1/16 3 etc.) Figure 6-18 shows how to read the A/D output for an input  
of 0.66V by interpreting the waveform at TP102 using the first three nibbles.  
Troubleshooting the A/D Converter for defective components can be simplified by  
setting the circuit in a quiescent state. This can be done by selecting EX TRIG, which  
causes all A/D activity to stop. The A/D Converter is then in the autozero configuration,  
and the offset of the amplifiers and the various levels in the bootstrap circuits can be  
easily measured with a voltmeter. Oscillations at the outputs of the amplifiers and other  
abnormal signals can easily be identified with an oscilloscope.  
6-67. Power Supply Troubleshooting  
If the display does not light up, first check the following:  
1. Is the instrument plugged in to an energized outlet providing alternating current at a  
frequency of 47 Hz to 440 Hz and a voltage within +/-10% of that selected by the  
rear panel line voltage selection switches?  
2. Is the POWER switch ON (pushed in)?  
3. Is the rear panel fuse blown?  
CAUTION  
For fire protection, use exact fuse replacement only.  
If the rear panel fuse is blown, replace it with a 250V fuse of the proper rating for the line  
voltage selected. Use 1/4A slow-blow for 100V and 120V power-line voltage and 1/8A  
slow-blow for 200V and 240V power-line voltage. If the fuse keeps blowing, measure the  
resistances of the power transformer (T601) windings. They should be within 10% of the  
values shown on the schematic. If not, the transformer is probably shorted. Also inspect  
the area around the transformer POWER switch and power-cord connector to make sure  
there isnt something shorting out the traces. If the IEEE-488 Interface board is suspected  
of causing the problem, it can easily be unplugged. Check the crowbar circuit (CR615  
and Q601, and R605). If CR615 or Q601 is shorted or if there is a large amount of  
leakage or R605 is open, fuses may continue to blow.  
If everything looks OK but the fuse keeps blowing, troubleshooting may be performed by  
powering the instrument through a variac, applying only enough line voltage to find the  
problem without blowing the fuse. NEVER USE A LARGER FUSE. To do so will only  
turn a small problem into a big one.  
6-61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f6-17.wmf  
Figure 6-17. Waveforms at TP102 for Several Inputs on 2V DV Range  
Since power supply problems can produce symptoms in many different sections of the  
instrument, the first step in troubleshooting any problem should usually be a quick check  
of the power supplies. For each power supply (TP801-TP806), check the level with a  
voltmeter and check for ripple with an ac-coupled oscilloscope. The dc voltages should  
be within the limits given in Table 6-23.  
If a supply is too high, either its three-terminal regulator has failed or a fault elsewhere in  
the instrument has shorted two supplies together. After repairing such a problem, make  
certain that nothing else was damaged by the overvoltage.  
6-62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
f6-18.wmf  
Figure 6-18. Calculating the A/D Reading From TP102 Waveform  
Table 6-23. Power Supply Voltages  
TEST POINT  
LIMITS (in volts)  
MINIMUM  
MAXIMUM  
+5V  
4.75  
5.25  
+7.5V  
+15V  
+30V  
-8.2  
7.00  
7.87  
14.25  
28.45  
-8.61  
-15.75  
-31.55  
15.75  
31.55  
-7.60  
-14.25  
-28.45  
-15V  
-30V  
If a supply is too low, there are a number of possible causes. First check the input to the  
affected regulator. If it is not at least 1V above the maximum output given in Table 6-23,  
the cause may be a bad transformer winding (check the resistance), open or shorted  
rectifiers, a shorted filter capacitor, or a shorted regulator. The latter two failures will  
usually blow the line fuse.  
All regulators incorporate current-limiting which allows them to shut down in the event  
of a load failure. Therefore if the power supply output is too low, the first step should be  
to determine if it is due to a high load caused by a failure elsewhere in the instrument.  
Frequently the faulty component can be found by using a multimeter with at least 5 digits  
resolution to check the supply pins of all components powered from that supply. Connect  
one lead of the voltmeter to the appropriate test point for the power supply under test and  
use the other lead to probe the loads. Small voltage drops across the PCA traces can be  
detected in this way, and the fault isolated. If any component other than one of the  
regulators is too hot to touch, there is something wrong with it or with something  
connected to it.  
6-63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
The True RMS AC PCA, if installed, uses +5V and +/-15V. If there is a problem with  
one of those supplies, first disconnect the True RMS AC PCA. If the problem goes away,  
troubleshoot the True RMS AC PCA using the procedure given later in this section.  
With most power supply problems, the output voltage is too low or too high. More subtle  
problems that may be encountered include high ripple or oscillation. If more than 10 mV  
of line-frequency ripple exists on one of the power supply outputs, it is usually caused by  
the input being too low, causing the regulator to drop out of regulation. It is also possible  
(but not likely) that the regulator itself is defective. High-frequency oscillation  
(frequently synchronized with the 1 Mhz or 8 Mhz clock) is usually the result of a bad  
regulator or output bypass capacitor. A fair amount of high-frequency noise is generally  
present on all the supplies, particularly +5V, and should not cause any concern unless the  
instrument behaves erratically or the reading is noisy.  
6-68. IEEE-488 Interface Troubleshooting (Option -05)  
6-69. SERVICE POSITION  
To provide easy access to the IEEE-488 Interface PCA and the Main PCA, the IEEE-488  
Interface PCA can be placed in the specially provided service position as follows:  
1. Remove the case from the chassis according to the Case Disassembly procedure  
provided earlier in this section.  
2. Release the two nylon latches that hold the IEEE-488 Interface PCA in place by  
pulling the latches upward.  
3. Position the IEEE-488 Interface PCA vertically as shown in Figure 6-19 and latch it  
in place be pressing the two nylon latches into the mounting supports specially  
provided on the chassis.  
6-70. DIAGNOSTIC PROGRAM  
To facilitate troubleshooting, the IEEE-488 Interface provides a diagnostic program  
which places the instrument in known configurations. To initiate the diagnostic program,  
proceed as follows:  
CAUTION  
To avoid damage to the 8842A or other equipment, the 8842A  
must be disconnected from all other IEEE-488 Interface  
instruments while the diagnostic program is running.  
1. Ensure the 8842A POWER switch is OFF.  
2. Disconnect all cables from the rear panel IEEE-488 connector.  
3. Short TP903 to TP905.  
4. Power up the 8842A. The 8842A should display ERROR 50. To exit the  
troubleshooting mode, open the jumper and cycle the POWER switch from off to on.  
Once the diagnostic program is started, rear-panel IEEE-488 address switches A3, A2,  
and A1 can be used to select one of four diagnostic modes, as shown in Table 6-24. In  
this table, Configuration indicates which Out-Guard µC I/O port bits are programmed as  
outputs and driven with a signal, as shown in Table 6-25.  
6-64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
f6-19.wmf  
Figure 6-19. Option -05 Service Position  
Table 6-24. Diagnostic Modes  
SWITCHES  
A2  
CONFIGURATION  
Static, odd-port bits = 1, even-port bits = 0  
A3  
A1  
1
1
0
1
1
1
Static, odd-port bits = , even-port bits = 1  
1
0
X
X
0
x
Dynamic  
Read/Write  
NOTES:  
“x” means switch setting does not matter.  
“Static” means the Out-Guard µC I/O port bits programmed as outputs are driven to a constant  
logic 1 or 0 level (as determined by switch A2).  
“Dynamic” means the Out-Guard µC I/O port bits programmed as outputs are driven with a 610  
Hz, 50% duty cycle square wave. All odd port bit numbers are 180 degrees out of phase with  
even port bit numbers.  
“Read/Write” means that data is read from and written to the NEC7210 IEEE chip (U901) when DS  
(U901-8) is low. R/W (U901-7) determines whether the data is being read from or written to the  
NEC7120. The address bits are always 3 (0011) and the data bits are incremented each time.  
6-65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 6-25. I/O Port Configurations  
CONFIGURATION  
CONFIGURATION  
PORT BIT  
PORT BIT  
Static  
Read/Write  
Static  
Read/Write  
Dynamic  
Dynamic  
PO-0  
PO-1  
PO-2  
PO-3  
PO-4  
PO-5  
PO-6  
PO-7  
P2-0  
P2-1  
P2-2  
P2-3  
P2-4  
P2-5  
P2-6  
P2-7  
OUT  
OUT  
OUT  
OUT  
IN  
Address  
P1-0  
P1-1  
P1-2  
P1-3  
P1-4  
P1-5  
P1-6  
P1-7  
P3-0  
P3-1  
P3-2  
P3-3  
P3-4  
P3-5  
P3-6  
P3-7  
OUT  
OUT  
OUT  
OUT  
OUT  
OUT  
OUT  
OUT  
IN  
Data  
Data  
Data  
Data  
Data  
Data  
Data  
Data  
IN  
Address  
Address  
Address  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
IN  
Clock  
serial  
Clock  
Serial  
IN  
IN  
NOTES:  
Due to external hardware conflicts, the following bits are NEVER driven by the Out-Guard µC in  
ANY diagnostic mode: P0-4,5,6,7; P2 (all bits); P3-1,2,3,4.  
P3-6 is the 4 MHz clock for the NEC7210 IEEE chip (U901).  
P3-7 is programmed as the serial output, and constantly transmits hex 55 every 820 µs at 62,000  
baud in all four diagnostic modes. This causes the front panel error message.  
6-71. True RMS AC Troubleshooting (Option -09)  
6-72. SERVICE POSITION  
To provide easy access to the True RMS AC PCA and the Main PCA, the True RMS AC  
PCA can be placed in the specially provided service position as follows:  
1. Remove the case from the chassis using the Case Disassembly procedure provided  
earlier in this section.  
2. Release the four nylon latches that hold the True RMS AC PCA in place by pulling  
the latches upward. (See Figure 809-1E in Section 8.)  
3. Disconnect the red ac input lead from both the True RMS AC PCA and the Main  
PCA.  
6-66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
4. Position the True RMS AC PCA vertically as shown in Figure 6-20 and latch it in  
place by pressing the bottom two nylon latches into the specially provided mounting  
supports on the chassis.  
5. Connect the Main PCA ac take-off point (stud connector W301) to the True RMS AC  
PCA input (the stud connector labeled AC IN) with a 6-inch jumper (E-Z-Hook 204-  
6W-S or equivalent).  
6-73. MAJOR PROBLEMS  
The signal flow through the True RMS AC option is straightforward, with no feedback  
paths between individual stages. This simplifies troubleshooting and often makes it  
possible to isolate a single defective stage without removing the instrument cover.  
Test the mid-frequency response of the VAC function around 1 kHz. If an accurate  
reading can be obtained on at least one range, the rms converter (U802) is working  
properly. Test the high-frequency response around 100 kHz. If, after calibration, an  
accurate reading can be obtained on at least one range, the digitally controlled filter  
(U801, U808, R832, and C826-829) is OK. If some ranges are good and others are bad,  
the defective stage may be isolated using Table 6-26. If this table is used, the bad ranges  
must correspond exactly to the ranges listed in the first column and all other ranges must  
be good.  
Most ac troubleshooting can be performed with the shields removed. To remove both  
shields, unscrew the Phillips screw on the back of the True RMS AC PCA. The only time  
it should be necessary to work on the PCA with the shields in place is when there is  
subtle high-frequency (>20 kHz) or low-level (<10 mV) error. In that case, the PCA  
should be left in its operating position, and the test points probed from the foil side of the  
PCA. Test points are labeled on both sides to facilitate such troubleshooting.  
Table 6-26. Isolating a Defective AC Stage  
DEFECTIVE RANGES  
200 mV, 2000 mA  
DEFECTIVE STAGE  
U806B  
U806A  
U806A  
20V, 700V  
2V, 200V  
200V, 700V  
2V, 20V, 200V  
Input (Q806, K802, Z801)  
Input (Q806, K802, Z801)  
If no ranges work, the signal should be traced from input to output. At any point where  
the signal disappears, the preceding stage should be searched thoroughly. To trace the  
signal, lock the instrument into one range (200 mV is usually a good choice) and apply  
the appropriate voltage shown in Table 6-27 to the HI and LO INPUT terminals. The  
input voltage should appear unchanged at pin Z801-1, and should appear at TP801 and  
TP802 as shown in Table 6-27. If no ranges work, it is likely that the rest of the scaling  
circuitry (U806B) is functional.  
6-67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f6-20.wmf  
Figure 6-20. Option -09 Service Position  
If the signal at the input to U801A (pin 5) is incorrect, U804 may be defective, or the  
switch codes may be wrong. If the latter problem is suspected, refer to Table 6-28 and  
test the control lines to U804 (U804-1, 8, 9, 16). If a logic error is found, it may be due to  
excessive loading or a faulty data latch (U803), or other cabling or main-board digital  
problems. High-frequency oscillation problems are usually caused by switches being on  
when they should be off, resulting in positive feedback loops being closed around  
portions of the scaling circuitry.  
Table 6-27. AC Signal Tracing  
RANGE  
INPUT VOLTAGE (1  
kHz)  
VOLTAGE AT TP801  
VOLTAGE AT TP802  
200 mV  
100 mV  
20 mV  
1 V  
2V  
1V  
200 mV  
2V  
1 V  
20V  
200V  
700V  
10V  
10V  
100V  
1V  
20 mV  
200 mV  
100 mV  
100 mV  
If the signal at TP802 is incorrect, but U801-5 is OK, the digitally controlled filter section  
(U801A and U808) is probably defective.  
6-68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
TROUBLESHOOTING  
6
Table 6-28. Truth Table for U804 and K2  
PIN OR DEVICE  
RANGE  
U804-1  
U804-8  
U804-9  
U804-16  
K2  
0
2000 mA  
1
1
1
0
1
0
0
0
1
1
1
1
1
0
1
1
1
1
1
1
0
1
0
1
200 mV  
2V  
0
0
20V  
0
200V  
700V  
NOTE:  
1
1
For U804, logic 0 = switch on. Logic 1 is >2.4V; logic 0 is <0.8V.  
If the signal at TP802 is correct but the output signal (TP803) is incorrect, the rms  
converter is probably the source of the problem. Problems with U802 generally show up  
as an identical number of counts displayed in all ranges or as an overrange in all ranges.  
First isolate faults in the buffer amplifier (U802A) by ascertaining that the signal at  
U802-13 is the same as that at TP802 when each is observed with an ac-coupled  
oscilloscope, and that the dc offset at U802-13 is less than 4 mV. It is possible that a  
component in the loop filter (U809A and associated passive components) or the post-  
filter (U809B and associated passive components) is defective. The dc voltage at U802-6  
should be the same as that at TP803 for frequencies above 500 Hz, and should be equal to  
the rms value of the input signal.  
6-74. MORE OBSCURE PROBLEMS  
Slow settling time or excessive jitter for low-frequency inputs is caused by rms converter  
loop errors. The cause may be a fault in the rms converter or loop filter.  
If the output voltage is stuck at the supply rails, the cause is probably a fault in the rms  
converter. A less common cause is op amp oscillation; this can be checked with a scope  
at TP802.  
If one or more ranges are functional but cannot be calibrated at high frequencies, then  
either the digitally controlled filter (U801B, R832, and C826-C829) is defective, or a  
defective component elsewhere in the circuit has rendered the response out of calibration  
range. (The high-frequency calibration is designed to cover the range of error expected  
due to op amp variations, input dividers, PCA tolerances, shielding, etc.). A sweep  
generator is useful in troubleshooting difficult frequency response problems.  
The calibration control lines to U808 are set by software to store a high-frequency  
correction factor for each range. A state table cannot be given for these signals, but  
common sense will indicate if they are reasonable. For example, if all lines are at logic 0  
for all ranges, something is probably wrong. Suspect U803, U808, or digital hardware on  
the Main PCA. If the control signals do indeed change with range, U808 or some part of  
the digitally controlled filter (U801B, R832, and C826-C829) may be defective. Even  
with the worst possible error in the high-frequency calibration code, the reading should  
be within 10% of the correct value at frequencies up to 100 kHz. If the error is larger,  
there are analog problems.  
6-69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
It is safe to force one control line at a time high (+5V) or low (0V) to test the individual  
switches in U808. (The on resistance of switches in U808 should be less than 500; the  
off resistance should be greater than 10 M.) Forcing the control lines high or low  
should cause the reading to change when the voltages in Table 6-27 are applied to the  
input terminals at 100 kHz. If only certain ranges cannot be calibrated, refer to Table 6-  
26 to find the suspected stage.  
6-75. Guard Crossing Troubleshooting  
To troubleshoot the Guard Crossing circuit, place the 8842A in the In-Guard  
Troubleshooting Mode as described under Digital Controller Troubleshooting, earlier in  
this section. This causes the In-Guard µC to send a test pattern to the IEEE-488 Interface  
PCA via one-half of the Guard Crossing circuit. You should be able to observe the  
waveforms shown in the left half of Figure 6-21.  
To troubleshoot the second half of the Guard Crossing circuit, leave the 8842A in the In-  
Guard Troubleshooting Mode. This causes the IEEE-488 Interface PCA to send a test  
pattern to the In-Guard µC. You should be able to observe the waveforms shown in the  
right half of Figure 6-21. The IEEE-488 Interface PCA sends the test pattern in response  
to the test pattern sent by the In-Guard µC; therefore, the first half of the Guard Crossing  
Circuit (which was tested in the previous paragraph) must be working properly before the  
second half can be tested.  
6-76.INTERNAL CLEANING  
CAUTION  
Failures due to electrostatic discharge can be caused by  
improper handling of the PCAs and by the use of a vacuum  
cleaner with static-inducing brushes. To prevent damage by  
electrostatic discharge, observe the precautions described on  
the Static Awareness sheet in front of this section.  
If visual inspection of the instrument shows excessive dirt build-up in the instrument,  
clean the appropriate section using clean, oil-free, low-pressure air (less than 20 psi). If  
necessary, remove the option PCAs first.  
6-77. Cleaning Printed Circuit Assemblies  
If conditions warrant, individual printed circuit assemblies (PCAs) can be cleaned with  
water-based commercial cleaning systems such as dishwashers. If such systems are used,  
observe the following precautions:  
1. Remove all shield covers (applies to the True RMS AC PCA) and socketed ICs.  
2. Use Reagent Grade 2 or better water (de-ionized or distilled water) for the final rinse  
in geographic areas with exceptionally hard water. During the final rinse, spray or  
run the water so that the surface is thoroughly covered to remove all ionized material.  
3. Thoroughly dry all PCAs using one of the following methods:  
a. Preferably, the PCA should be dried in a low-temperature drying chamber or  
infrared drying rack with a temperature range of 49°C to 72°C (120°F to 160°F).  
b. If neither a drying chamber nor a drying rack is available, air dry the PCA at  
ambient room temperature for at least two days.  
A satisfactory cleaning method consists of holding the PCAs under hot running water  
until they are clean. Follow this wash with a final rinse. (See consideration 2, above.)  
6-70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
INTERNAL CLEANING  
6
6-78. Cleaning After Soldering  
CAUTION  
T.M.C. Cleaner and similar products can can attack the nylon  
latches and other plastic pieces.  
f6-21.wmf  
Figure 6-21. Guard Crossing Test Waveforms  
If a PCA has been soldered, it should first be cleaned with SPRAYON T.M.C Cleaner  
(rosin flux remover) or equivalent. The PCA should then be cleaned with water as  
described above.  
6-71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
6-72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7  
List of Replaceable Parts  
Title  
Page  
7-1.  
7-2.  
7-3.  
7-4.  
7-5.  
INTRODUCTION................................................................................7-3  
HOW TO OBTAIN PARTS.................................................................7-3  
NEWER INSTRUMENTS...................................................................7-4  
SERVICE CENTERS...........................................................................7-4  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
7-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Replaceable Parts  
INTRODUCTION  
7
7-1. INTRODUCTION  
This section contains an illustrated list of replaceable parts for the 8842A. Parts are listed  
by assembly; alphabetized by reference designator. Each assembly is accompanied by an  
illustration showing the location of each part and its reference designator. The parts lists  
give the following information:  
Reference designator  
An indication if the part is subject to damage by static discharge  
Description  
Fluke stock number  
Manufacturers supply code (code-to-name list at the end of this section)  
Manufacturers part number or generic type  
Total quantity  
Any special notes (i.e., factory-selected part)  
CAUTION  
A * symbol indicates a device that may be damaged by static  
discharge.  
7-2. HOW TO OBTAIN PARTS  
Electrical components may be ordered directly from the manufacturer by using the  
manufacturers part number, or from the Fluke Corporation and its authorized  
representatives by using the part number under the heading FLUKE STOCK NO. In the  
U.S., order directly from the Fluke Parts Dept. by calling 1-800-526-4731. Parts price  
information is available from the Fluke Corporation or its representatives. Prices are also  
available in a Fluke Replacement Parts Catalog which is available on request.  
In the event that the part ordered has been replaced by a new or improved part, the  
replacement will be accompanied by an explanatory note and installation instructions, if  
necessary.  
To ensure prompt delivery of the correct part, include the following information when  
you place an order:  
Instrument model and serial number  
Part number and revision level of the pca containing the part  
Reference designator  
Fluke stock number  
Description (as given under the DESCRIPTION heading)  
Quantity  
7-3. MANUAL STATUS INFORMATION  
The Manual Status Information table that precedes the parts list defines the assembly  
revision levels that are documented in the manual. Revision levels are printed on the  
component side of each pca.  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
7-4. NEWER INSTRUMENTS  
Changes and improvements made to the instrument are identified by incrementing the  
revision letter marked on the affected pca. These changes are documented on a  
supplemental change/errata sheet which, when applicable, is included with the manual.  
7-5. SERVICE CENTERS  
A list of service centers is located at the end of this section.  
NOTE  
This instrument may contain a Nickel-Cadmium battery. Do not mix with  
the solid waste stream. Spent batteries should be disposed of by a qualified  
recycler or hazardous materials handler. Contact your authorized Fluke  
service center for recycling information.  
WARNING  
THIS INSTRUMENT CONTAINS A FUSIBLE RESISTOR (PN  
733915).TO ENSURE SAFETY, USE EXACT REPLACEMENT  
ONLY.  
7-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
List of Replaceable Parts  
SERVICE CENTERS  
7
Manual Status Information  
REF OF OPTION NO.  
ASSEMBLY  
Main PCA  
Display PCA  
IEEE-488 Interface PCA  
True RMS AC PCA  
FLUKE PART NO.  
REVISION LEVEL  
A1  
A2  
05  
09  
759365  
728873  
879267  
759266  
BE  
__  
D
N
Table 7-1. 8842A Digital Multimeter  
7-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f7-01_1.wmf  
Figure 7-1. 8842A Digital Multimeter  
7-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
List of Replaceable Parts  
SERVICE CENTERS  
7
f7-01_2.wmf  
Figure 7-1. 8842A Digital Multimeter (cont)  
7-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
f7-01_3.wmf  
Figure 7-1. 8842A Digital Multimeter (cont)  
7-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Replaceable Parts  
SERVICE CENTERS  
7
f7-01_4.wmf  
Figure 7-1. 8842A Digital Multimeter (cont)  
7-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
Table 7-2. A1 Main PCA  
Reference  
Designator  
Description  
Fluke Stock  
No  
Tot Qty  
Notes  
AR701  
REF AMP SET  
759290  
446781  
733386  
733386  
733386  
733386  
697284  
697284  
697417  
697417  
697458  
520346  
715284  
446799  
721662  
446807  
721142  
721142  
720938  
446773  
697441  
697409  
478784  
484436  
747493  
649731  
658898  
707836  
113316  
113316  
113316  
113316  
113316  
256446  
256446  
203323  
203323  
203323  
203323  
203323  
246611  
260695  
267807  
707075  
159798  
742874  
742874  
325811  
386771  
187757  
386771  
103424  
380196  
110635  
178533  
185918  
706986  
714022  
1
4
C101-103,C311  
C104,C105,C205-  
209,C304,C305,  
C315,C404,C604,  
C606,C609,C610  
C202,C203,C617,  
C618  
CAP,POLYPR,0.1UF,+-10%,160V  
CAP,CER,0.22UF,+80-20%,50V,Z5U  
15  
CAP,CER,0.01UF,+80-20%,50V,Z5V  
CAP,TA,1UF,+-20%,35V  
4
4
C204,C602,C608,  
C612  
C210  
CAP,CER,1000PF,+-20%,50V,X7R  
CAP,POLYPR,0.33UF,+-10%,160V  
CAP,POLYES,0.33UF,+-10%,50V  
CAP, POLYPR,0.22UF,+-10%,160V  
CAP, POLYPR,4700PF,+-10%,63V  
CAP,POLYPR,0.47UF,+-10%,160V  
CAP,CER,120PF,+-5%,50V,C0G  
CAP,CER,120PF,+-5%,50V,C0G  
CAP,POLYES,0.001UF,+-10%,50V  
CAP,POLYPR,0.047UF,+-10%,160V  
CAP,CER,330PF,+-5%,50V,C0G  
CAP,POLYES,0.47UF,+-10%,50V  
CAP,AL,6800UF,+/-20%,16V  
1
2
1
7
t
C301,C302  
C303  
C306  
C307  
C308  
1
1
1
1
1
1
1
1
1
2
1
1
1
10  
C309  
C310  
C312  
C314  
C402  
C403  
0601  
C603  
CAP,AL,330UF,+50-20%,100V  
C605,C607  
C611  
CAP,AL,470UF,+-20%,50V,SOLV PROOF  
CAP,AL,100UF,+50-20%,50V  
1
C701  
CAP,CER,270PF,+-5%,50V,C0G  
I-REG DIODE,0.43MA,20%,SEL,TO-0226AC  
ZENER,UNCOMP,3.9V,10%,20.0MA,0,4W  
CL301  
*
CR101,CR102,  
CR201,CR202,  
CR306,CR307,  
CR309,CR311-  
313  
CR103,CR104,  
CR613  
*
*
*
*
*
*
*
*
*
*
ZENER,UNCOMP,7.5V,5%,20.0MA,0.4W  
DIODE,SI,BV=75V,10=150MA,500MW  
3
CR105,CR106,  
CR203-206,  
CR301,CR302,  
CR401,CR404,  
CR615,CR701  
CR303  
12  
ZENER,UNCOMP,10.0V,5%,12.5MA,0.4W  
ZENER,UNCOMP,6.8V,5%,20.OMA,0.4W  
ZENER,UNCOMP,24.0V,5%,5.2MA,0.4W  
DIODE,SI,1K PIV,1.0AMP  
1
2
CR304,CR305  
CR308,CR310  
CR402,CR403  
CR405  
2
2
*
ZENER,UNCOMP,5.1V,5%,20.0MA,0  
DIODE,S1,100 PIV,1 AMP  
1
CR601-606,  
CR608-611  
CR607,CR612  
CR614  
70  
*
*
ZENER,UNCOMP,6.2V,5%,20.0MA,0.4W  
ZENER,UNCOMP,8.2V,5%,20.0MA,0.4W  
ZENER,UNCOMP,56.0V,5%,2.2MA,0.4W  
ZENER,UNCOMP,8.2V,5%,20.OMA,0.4W  
RIVET,S-TUB,OVAL,STL,.118,.156  
NUT,BROACH,STL,4-40  
2
1
1
1
2
1
1
1
1
1
1
CR616  
CR617  
H1,H51  
H3  
H5  
NUT,HEX,STL,4-40  
H6  
SCREW,PH,P,SEMS,STL,6-32,.250  
SCREW,PH,P,SEMS,STL,4-40,.250  
HEADER,2 ROW,.100CTR,20 PIN  
CABLE ASSY,FLAT,10 CONDUCT,6.0  
H12  
J201  
J202  
7-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
List of Replaceable Parts  
SERVICE CENTERS  
7
Table 7-2. A1 Main PCA (cont)  
Reference  
Designator  
Description  
Fluke Stock  
No  
Tot Qty  
Notes  
J203,J204  
CABLE, DISPLAY  
684167  
380378  
682575  
615575  
714352  
152207  
516880  
728907  
414128  
602763  
735365  
698225  
698225  
698233  
601333  
601333  
601333  
601333  
601333  
601333  
722934  
650846  
461772  
783449  
742643  
343426  
343426  
343434  
441501  
340257  
385948  
348854  
348854  
348771  
343418  
343418  
348888  
348888  
441493  
285056  
715185  
715185  
234708  
349001  
348987  
348987  
733915  
442350  
109975  
442319  
731950  
343459  
762120  
543348  
875257  
442335  
474080  
2
1
2
1
1
1
2
1
1
1
1
4
J602  
HEADER,1 ROW,.156CTR,6 PIN  
RES JUMPER,0.02,0.25W  
JPR001,JPR002  
K301  
RELAY,ARMATURE,2 FORM C,5VDC  
RELAY,REED,1 FORM A,5V,HIGH VOLTAGE  
INSUL PT,TRANSISTOR MOUNT,DAP,TO-5  
HLDR,FUSE,13/32,PWB MT  
K401  
MP2  
MP29,MP50  
MP30  
SHIELD, AID  
MP33  
HEAT DIS,VERT,1.18X1.00X0.50,T0-220  
HLDR PART,FUSE,BODY,PWB MT  
PAD, ADHESIVE  
MP35  
MP201  
Q101,Q309,Q318,  
Q408  
*
*
*
*
*
*
*
*
*
*
*
*
*
*
TRANSISTOR,SI,NPN,60V,1W,T0-92  
Q102,Q319  
Q301,  
TRANSISTOR,SI,PNP,40V,350MW,T0-92  
TRANSISTOR,SI,N-JFET,TO-92  
2
17  
Q302,Q303,Q304,  
Q307,Q308,Q310  
Q311,Q312,Q313,  
Q315,Q316,Q317,  
Q401,Q405-407  
Q305  
TRANSISTOR,SI,NPN,300V,1W,T0-92  
TRANSISTOR,SI,PNP,350V,0.6W,SEL,TO-92  
TRANSISTOR,SI,N-JFET,DUAL,SEL,TO-71  
TRANSISTOR,SI,N-MOS,350MW,T0-92  
THYRISTOR,SI,SCR,VBO=100V,0.8A  
RES,CF,1K,+-5%,0.25W  
1
4
1
1
1
4
Q306,Q402-404  
Q314  
Q320  
Q601  
R101,R602,R603,  
R605  
R102,R203,R204  
R103,R104  
R105  
RES,CF,470,+-5%,0.25W  
RES,CF,27K,+-5%,0.25W  
RES,MF,50K,+-0.1 %,0.125W,25PPM  
RES,CF,560,+-5%,0.25W  
RES,CF,15K,+-5%,0.25W  
3
2
1
1
7
R106  
R201,R215,R217,  
R305-307,R604  
R202,R315,R327  
R205,R207,R208,  
R210,R412  
R206,R209,R312,  
R313,R408  
R216,R326,R406  
R301  
RES,CF,100,+-5%,0.25W  
RES,CF,1.5K,+-5%,0.25W  
3
5
RES,CF,33K,+-5%,0.25W  
5
RES,CF,2.4K,+-5%,0.25W  
3
1
4
RES,CC,100K,+-5%,2W  
R302,R311,R324,  
R325  
RES,MF,46.4K,+-1%,0.125W,50PPM  
R303  
RES,MF,110K,+-1%,0.125W,100PPM  
FiES,CF,1.5M,+-5%,0.25W  
1
1
4
R304  
R308,R316,R317,  
R323  
RES,CF,1M,+-5%,0.25W  
R309  
RES,MF,1K,+-1%,0.5W,FLMPRF,FUSIBLE  
RES,CF,5.6K,+-5%,0.25W  
1
3
1
1
1
2
1
9
1
1
1
1
R310,R409,R415  
R314  
RES,CC,22K,+-10%,2W  
R318  
RES,CF,620,+-5%,0.25W  
R319  
W W RESISTOR  
R321,R322  
R401  
RES,MF,10K,+-0.1%,0.725W,50PPM  
RES, WW, HERM, 1/4W, 54.6K, +-.05%  
RES,CF,4.3M,+-5%,0.25W  
R402  
R403  
RES,CF,10M,+-5%,.25W  
R407  
RES,CF,910,+-5%,0.25W  
R410  
RES,MF,1K,+-1%,100PPM,FLMPRF,FUSIBLE  
1
7-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
Table 7-2. A1 Main PCA (cont)  
Reference  
Designator  
Description  
Fluke Stock  
No  
Tot Qty  
Notes  
R413  
RES,MF,4.99M,+-1%,0.125W,100PPM  
715060  
344291  
460527  
385948  
697383  
519355  
875703  
453605  
660589  
803270  
512889  
512889  
512889  
512889  
512889  
685164  
685156  
685412  
393058  
387233  
407585  
393108  
419242  
685537  
697730  
585364  
483180  
700013  
780759  
650713  
418780  
605980  
803478  
478107  
428847  
604074  
413187  
413179  
418251  
684399  
728840  
765297  
765305  
873740  
873745  
873752  
756668  
756668  
756650  
756353  
707133  
755983  
755934  
756130  
715789  
755884  
756080  
1
1
R414  
RES,MF,576K,+-1%,0.125W,l00PPM  
RES,MF,100,+-1 %,0.125W,25PPM  
RES,CF,560,+-5%,0.25W  
R416  
1
R601  
1
RV301,RV401-404  
RV601  
VARISTOR,390V,+-10%,1MA  
5
VARISTOR,430V,+-10%,1.0MA  
SWITCH,PUSHBUTTON,DPDT,PUSH-PUSH  
SWITCH,PUSHBUTTON,DPDT,PUSH-PUSH  
TRANSFORMER, PULSE  
1
S201  
1
S601  
1
T201,T202  
T601  
2
TRANSFORMER, POWER  
1
TP101-105,TP201,  
TP202,TP204-207,  
TP301-304,TP401,  
TP403,TP601-605,  
TP607-609  
U102,U702  
U103  
TERM,FASTON,TA8,.110,SOLDER  
25  
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
IC,OP AMP,DUAL,LO OFFST,VOLT,LO-DRIFT  
IC,OP AMP,LO-OFF VOLTAGE,LO-DRIFT  
IC,BIPLR,BCHNL DRIVER  
2
1
1
1
2
1
1
1
3
1
1
1
4
1
1
1
1
1
1
1
1
1
1
1
2
2
1
1
1
1
1
3
U201  
0203  
IC,LSTTL,HEX INVERTER  
U204,U305  
U208  
IC,COMPARATOR,QUAD,14 PIN DIP  
IC,LSTTL,3-8 LINE DCDR W/ENABLE  
IC,LSTTL,QUAD 2 INPUT OR GATE  
U211  
0213  
IC,LSTTL,8BIT ADDRSABLE LATCH,W/CLR  
IC,BIPLR,8CHNL FLOURSCNT DISPLY DRIVR  
IC,LSTTL,BCD TO 7 SEGMENT DCDR/DRVR  
IC,CMOS,OCTL D F/F W/3-STATE,+EDG TRG  
IC,LSTTL,HEX INVERTER W/SCHMT TRIG  
IC CMOS, QAS, PLASTIC, HIGH, B GRADE  
IC,OP AMP,FET,LO NOISE,PRECISIION,8DIP  
IC,OP AMP,SPECIAL,LOW DRIFT,TO-99  
IC,OP AMP,JFET IN,COMPENSTD,8 PIN DIP  
IC,OP AMP,LO-OFFSET VOLTAGE,LO-NOISE  
IC CMOS, QAS, PLASTIC, HIGH, A GRADE  
IC,OP AMP,GENERAL PURPOSE,8 PIN  
IC,VOLT REG,FIXED,+5 VOLTS,1.5 AMPS  
IC,VOLT REG,FIXED,+24 VOLTS,1.5 AMPS  
IC,VOLT REG,FIXED,+15 VOLTS,1.5 AMPS  
IC,VOLT REG,FIXED; 15 VOLTS,1.5 AMPS  
IC,VOLT REG,FIXED, 24 VOLTS,1.5 AMPS  
WIRE ASSEMBLY, GUARD CROSSING  
WIRE ASSY, GUARD CROSSING  
U215,U217,U218  
0216  
U219  
0221  
U301-303,U402  
U304  
0306  
U307  
0401  
U403  
U404  
VR601  
VR602  
VR603  
VR604  
VR605  
W1,W8  
W2,W9  
W5  
HARNESS, ANALOG FRONT  
W6  
HARNESS, ANALOG REAR  
W10  
CABLE, LINE VOLTAGE-GROUND  
W11  
CABLE, LINE VOLTAGE-LINE  
W12  
CABLE, LINE VOLTAGE-NEUTRAL  
XU101,XU102,  
XU212  
SOCKETJC,40 PIN,DUAL WIPE,RET  
XU220  
SOCKET,IC,24 PIN,DUAL WIPE,BEAM TYPE  
1
1
1
1
2
1
2
1
1
XU222  
Y201  
CRYSTAL,8.OOMHZ QUARTZ HC-18U  
RNET,MF,HERM,SIP,8842 A TO D CONV  
RNET,MF,HERM,SIP,8842 LO V DIVIDER  
RNET,MF,HERM,SIP,8842 HI V DIVIDER  
RNET,CERM,SIP,8840 HI V PROTECT  
RNET,MF,HERM,SIP,LO V 1 SOURCE  
RNET,MF,HERM,SIP,8842 LO V DIVIDER  
Z101  
Z301,Z303  
Z302  
*
*
*
*
*
Z304,Z402  
Z401  
Z702  
1. FUSIBLE RESISTOR. TO ENSURE SAFETY, USE EXACT REPLACEMENT ONLY.  
7-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Replaceable Parts  
SERVICE CENTERS  
7
f7-03.wmf  
Figure 7-2. A1 Main PCA  
7-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 7-3. A2 Display PCA  
7-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
List of Replaceable Parts  
SERVICE CENTERS  
7
f7-03.wmf  
Figure 7-3. A2 Display PCA  
7-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Supply Codes for Manufacturers  
supply_1.wmf  
7-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Replaceable Parts  
SERVICE CENTERS  
7
Supply Codes for Manufacturers (cont)  
supply_2.wmf  
7-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8842A  
Instruction Manual  
service_1.wmf  
Figure 7-4. Service Centers  
7-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
List of Replaceable Parts  
SERVICE CENTERS  
7
service_2.wmf  
Figure 7-4. Service Centers (cont)  
7-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
7-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 8  
Options and Accessories  
Title  
Page  
8-1.  
8-2.  
8-3.  
8-4.  
8-5.  
8-6.  
8-7.  
8-8.  
8-9.  
8-10.  
8-11.  
8-12.  
INTRODUCTION.................................................................................. 8-3  
ACCESSORIES ..................................................................................... 8-4  
Rack-Mount Kits (Y8834, Y8835 and Y8836).................................. 8-4  
Shielded IEEE-488 Interface Cables (Y8021, Y8022, and Y8023)... 8-4  
Replacement Test Leads (TL70A)..................................................... 8-4  
Deluxe Test Lead Kits (Y8134) ......................................................... 8-4  
Slim-Flex Test Leads (Y8140)........................................................... 8-4  
Temperature Probes (80T-150U, and 80TK.).................................... 8-4  
RF Probes (85RF and 83RF).............................................................. 8-4  
Current Shunt (80J-10)....................................................................... 8-5  
Current Probes (Y8100, Y8101, 80i-400 and 80i-600)...................... 8-5  
High Voltage Probes (80K-6 and 80K-40) ........................................ 8-5  
8-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
8-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Options and Accessories  
INTRODUCTION  
8
8-1. INTRODUCTION  
A number of options and accessories are available which can enhance the 8842As  
capabilities and increase operator safety. The accessories are summarized in Table 8-1  
and described in the following paragraphs. The options are summarized in Table 8-2 and  
described in the following subsections.  
Table 8-1. Accessories  
MODEL  
Y8021  
DESCRIPTION  
IEEE-488 Interface Cable (1m)  
Y8022  
Y8023  
Y8834  
Y8835  
Y8836  
TL70A  
Y8134  
Y8140  
80T-150U  
80TK  
IEEE-488 Interface Cable (2m)  
IEEE-488 Interface Cable (4m)  
Single Rack-Mount Kit  
Dual Rack-Mount Kit  
Center Rack-Mount Kit  
Replacement Test Leads  
Replacement Test Leads  
Slim-Flex Test Leads  
Temperature Probe  
K-Type Thermocouple Converter  
RF Probe (100 kHz to 500 MHz)  
RF Probe (100 kHz to 100 MHz)  
Current Shunt  
85RF  
83RF  
80J-10  
Y8100  
Y8101  
80I-400  
80I-600  
80K-6  
Current Probe (200A ac/dc)  
Current Probe (150A ac)  
Current Probe (400A ac)  
Current Probe (600A)  
High Voltage Probe (6,000V)  
High Voltage Probe (40,000V)  
80K-40  
Table 8-2. Options  
NUMBER  
-05  
OPTION  
IEEE-488 Interface  
True RMS AC  
-09  
8-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
8-2. ACCESSORIES  
Accessories include a variety of rack-mounting kits, cables, test leads, and probes. The  
accessories include installation and/or operating instructions.  
8-3.  
Rack-Mount Kits (Y8834, Y8835 and Y8836)  
The rack-mount kits allow the 8842A to be mounted in standard 19-inch rack panels. The  
Y8834 kit allows the 8842A to be mounted either on the left or the right. The Y8835 kit  
allows two 8842As to be mounted side-by-side. The Y8836 kit allows the 8842A to be  
mounted in the center of the rack. Installation instructions are given in Section 2, and are  
also included with each kit.  
8-4.  
Shielded IEEE-488 Interface Cables (Y8021, Y8022, and Y8023)  
Shielded IEEE-488 cables are available in three lengths (see Table 8-1). The cables attach  
the 8842A to any other IEEE-488 device. Each cable has double 24-pin connectors at  
both ends to allow stacking. Metric threaded mounting screws are provided with each  
connector.  
8-5.  
8-6.  
Replacement Test Leads (TL70A)  
The TL70A replacement test leads feature safety-designed input connectors.  
Deluxe Test Lead Kits (Y8134)  
Each deluxe test lead kit includes two test-tip probes, two alligator clips, two large spade  
lug tips, and one spring-loaded hook tip and probe.  
8-7.  
8-8.  
Slim-Flex Test Leads (Y8140)  
The Y8140 has adjustable, flexible, and insulated leads, and can fit into small places. The  
sharp steel needle points will pierce varnish and thin insulation.  
Temperature Probes (80T-150U, and 80TK.)  
The 80T-150U is a universal temperature probe designed to provide virtually any DMM  
with temperature measurement capability. The probe provides direct temperature  
conversion of 1 mV dc per degree. A three-position switch acts as a power switch and is  
used for selecting Celsius or Fahrenheit scaling for the output. The 80TK thermocouple  
module converts the microvolt output from a "K" type thermocouple to a 1 mV per  
degree signal. The on-off switch allows selection of degrees "C" or "F" output scaling.  
8-9.  
RF Probes (85RF and 83RF)  
The RF probes (85RF and 83RF) use the DMM dc volts function to measure radio  
frequency (RF) ac signals. The 83RF has a frequency range of 100 kHz to 100 MHz; the  
85RF has a frequency range of 100 kHz to 500 MHz. The probes are calibrated so that  
the dc output is equivalent to the rms value of a sine wave input over a range of 0.25V to  
30V rms.  
For best accuracy, the probes should be used with DMMs having 10 Minput  
impedance. This condition is satisfied by the 8842A in the 200V and 1000V ranges. The  
probes can also be used with the 8842A in the 200 mV, 2V, and 20V ranges if a 10 MΩ  
resistor is connected in parallel across the 8842A input terminals.  
8-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Options and Accessories  
ACCESSORIES  
8
8-10. Current Shunt (80J-10)  
The 80J-10 current shunt extends ac or dc current measurement up to 10A continuous, or  
to 20A for one minute without overheating.  
8-11. Current Probes (Y8100, Y8101, 80i-400 and 80i-600)  
The current probes extend the ac and dc current measurement range. The Y8100 current  
probe uses the Hall effect to measure dc or ac current up to 200A (in two ranges) without  
electrical contact. The Y8101 is a low-cost, compact current probe which uses a  
transformer to measure ac current from 2A to 150A. The 80i-600 uses a transformer to  
measure ac current from 1A to 600A, and features a large jaw opening for industrial use.  
The Y8101, 80i-400 and 80i-600 measure ac current only.  
8-12. High Voltage Probes (80K-6 and 80K-40)  
The high voltage probes extend the dc and ac voltage measurement range while  
minimizing shock hazard. The 80K-6 has a range of 0 to 6000V dc or peak ac, with  
frequency response to 1 kHz; the 80K-40 has a range of 0 to 40,000V dc or peak ac, with  
frequency response to 60 Hz. As the probes use 1000:1 dividers, the probes have a high  
input impedance and cause minimal circuit loading. The probes are impedance matched  
for both ac and dc measurements. A plastic body protects the operator from the voltage  
being measured.  
For best accuracy, the probes should be used with DMMs having 10 Minput  
impedance. This condition is satisfied by the 8842A in the 200V dc and 1000V dc ranges.  
The probes can also be used with the 8842A in the 200 mV dc, 2V dc, and 20V dc ranges  
if a 10 Mresistor is connected in parallel across the input terminals.  
8-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
8-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 805  
Option –05 IEEE-488 Interface  
Title  
Page  
805-1. INTRODUCTION................................................................................ 805-3  
805-2. CAPABILITIES ................................................................................... 805-3  
805-3. EXTERNAL CONTROLS................................................................... 805-3  
805-4. INSTALLATION................................................................................. 805-3  
805-5. PROGRAMMING INSTRUCTIONS.................................................. 805-4  
805-6. MAINTENANCE................................................................................. 805-4  
805-7. LIST OF REPLACEABLE PARTS..................................................... 805-4  
805-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
805-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Option –05 IEEE-488 Interface  
INTRODUCTION  
805-1. INTRODUCTION  
The IEEE-488 Interface turns the 8842A into a fully programmable instrument for use  
with the IEEE Standard 488-1978 interface bus (IEEE-488 bus). With the IEEE-488  
Interface, the 8842A can become part of an automated instrumentation system. The  
8842A can be under complete, interactive control from a remote bus controller, or it can  
be set to the talk-only mode, connected to a data logger or printer, and dedicated to a  
single task.  
805-2. CAPABILITIES  
The IEEE-488 Interface provides remote control of all front panel controls except for the  
POWER, CAL ENABLE, and FRONT/REAR switches. Other features include:  
A simple and predictable command set  
Fast measurement throughput  
Full talk/listen capability, including talk-only operation  
Full serial poll capability, with bit-maskable SRQ  
Full remote/local capability, including local lockout  
External Trigger and Sample Complete connectors  
Remote calibration  
Programmable trigger sources, including two bus triggers  
Informative output suffix (suppressible)  
Selectable output terminators  
The 8842A supports the following interface function subsets: SH1, AH1, T5, L4, SR1,  
RL1, DC1, DT1, E1, PP0, and C0.  
805-3. EXTERNAL CONTROLS  
When the IEEE-488 Interface is installed, the rear panel contains EXT TRIG (External  
Trigger) and SAMPLE COMPLETE connectors. These controls can be used even when  
the 8842A is disconnected from the IEEE-488 bus. Refer to Section 2 for details.  
805-4. INSTALLATION  
The IEEE-488 Interface is contained on a single, easy-to-install printed circuit assembly  
(PCA). To install the option, proceed as follows:  
WARNING  
TO AVOID ELECTRIC SHOCK, DISCONNECT THE POWER  
CORD AND ANY INPUT LEADS BEFORE REMOVING THE  
INSTRUMENT CASE.  
1. Remove the grounding screw from the bottom of the case and remove the two  
rear panel mounting screws (Figure 805-1A).  
2. Holding the front panel, slide the case and rear bezel off of the chassis (Figure  
805-1B). Note: At this point, the rear bezel is not secured to the case.  
805-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
3. Disconnect the ribbon cable from the plastic rear panel insert by pulling the tabs  
on either side of the ribbon cable connector outward. Pull the ribbon cable  
directly toward the front panel (Figure 805-1C).  
4. Remove the rear panel insert by releasing the two snap tabs inside the instrument  
(Figure 805-1D).  
NOTE  
The connection on the rear panel insert is used for factory calibration only.  
The connector is electrically isolated from all measuring circuitry.  
5. Connect the the ribbon cable from the Main PCA to the connector on the IEEE-  
488 Interface PCA (Figure 805-1E).  
6. Latch the ribbon cable in place as shown in Figure 805-1F.  
7. Make sure the heads of the plastic latches are in the extended position.  
8. With the component side down, guide the BNC and IEEE-488 connectors  
(located on the rear of the IEEE-488 Interface PCA) into the rear panel, and seat  
the IEEE-488 Interface PCA on the mounting supports on the chassis (Figure  
805-1G).  
9. Fasten the IEEE-488 Interface PCA to the chassis by pressing the two plastic  
latches into the mounting supports. (See Figure 805-1H.)  
10. Secure the IEEE-488 connector to the rear panel with the two screws and  
washers supplied.  
11. Replace the cover and rear bezel on the chassis and attach the two ear panel  
mounting screws.  
12. Attach the grounding screw to the bottom of the case.  
WARNING  
TO AVOID ELECTRIC SHOCK, ENSURE THE GROUNDING  
SCREW IS FIRMLY ATTACHED TO THE CASE BOTTOM.  
805-5. PROGRAMMING INSTRUCTIONS  
Programming instructions are presented in Section 3. That section also explains how to  
set up the 8842A on the IEEE-488 bus.  
805-6. MAINTENANCE  
All service information regarding Option -05 is contained in Section 6. The theory of  
operation is contained in Section 5.  
805-7. LIST OF REPLACEABLE PARTS  
A list of replaceable parts for the IEEE-488 Interface printed circuit assembly (PCA) is  
given in Table 805-1. Refer to Section 7 for ordering information.  
Caution  
The symbol Y indicates a device that may be damaged by static  
discharge.  
805-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Option –05 IEEE-488 Interface  
LIST OF REPLACEABLE PARTS  
f805-1_1.wmf  
Figure 805-1. Installing Option -05  
805-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f805-1_2.wmf  
Figure 805-1. Installing Option –05 (cont)  
805-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Option –05 IEEE-488 Interface  
LIST OF REPLACEABLE PARTS  
Table 805-1. Option -05A IEEE-488 Interface PCA  
805-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f805-2.wmf  
Figure 805-2. IEEE-488 Interface PCA  
805-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 809  
Option –09 True RMS AC  
Title  
Page  
809-1. INTRODUCTION.................................................................................. 809-3  
809-2. INSTALLATION................................................................................... 809-3  
809-3. OPERATING INSTRUCTIONS............................................................ 809-4  
809-4. MAINTENANCE................................................................................... 809-4  
809-5. LIST OF REPLACEABLE PARTS....................................................... 809-4  
809-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
809-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Option –09 True RMS AC  
INTRODUCTION  
809-1. INTRODUCTION  
The True RMS AC option gives the 8842A the ability to make ac voltage and current  
measurements. The ac functions are selected with the front panel VAC and mA AC  
buttons, or by remote commands if the IEEE-488 Interface option is installed.  
Specifications for Option -09 are shown in Section 1, Table 1-1.  
809-2. INSTALLATION  
The True RMS AC option is contained on a single, easy-to-install printed circuited  
assembly (PCA). To install the option, proceed as follows:  
WARNING  
TO AVOID ELECTRIC SHOCK, DISCONNECT THE POWER  
CORD AND TEST LEADS BEFORE REMOVING THE  
INSTRUMENT CASE.  
1. Remove the grounding screw from the bottom of the case and remove the two rear  
panel mounting screws (Figure 809-1A).  
2. Holding the front panel, slide the case and rear bezel off of the chassis (Figure  
809-1B). (At this point, the rear bezel is not secured to the case.)  
3. Holding the True RMS AC PCA slightly above the chassis, component side down,  
connect the the ribbon cable from the True RMS AC PCA to the Main PCA and  
latch it in place. (See Figure 809-1C and D.)  
4. Connect the red lead from the True RMS AC PCA to stud W301 on the Main  
PCA. (See Figure 809-1C.) The stud is located next to the forward end of the  
FRONT/REAR switch.  
5. Make sure the heads of the four plastic latches are in the extended position. Guide  
the PCA into the 4 circuit board supports.  
6. Fasten the True RMS AC PCA to the chassis by pressing the four nylon latches  
into the mounting supports on the chassis. (See Figure 809-1E.)  
7. Reinstall the cover and rear bezel on the chassis and attach the two rear panel  
mounting screws.  
8. Attach the grounding screw to the bottom of the case.  
WARNING  
TO AVOID ELECTRIC SHOCK, ENSURE THE GROUNDING  
SCREW IS FIRMLY ATTACHED TO THE CASE BOTTOM.  
9. Calibrate the VAC voltage and mA AC functions according to the calibration  
instructions given in the Maintenance section.  
809-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
809-3. OPERATING INSTRUCTIONS  
For operating instructions, refer to Section 2. For ac measurement considerations, refer to  
Section 4.  
809-4. MAINTENANCE  
All service information regarding Option -09 is contained in Section 6. The theory of  
operation is contained in Section 5.  
809-5. LIST OF REPLACEABLE PARTS  
A list of replaceable parts for the True RMS AC printed circuit assembly (PCA) is given  
in Table 809-1. Refer to Section 7 for ordering information.  
CAUTION  
The symbol * indicates a device that may be damaged by static  
discharge.  
809-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Option –09 True RMS AC  
LIST OF REPLACEABLE PARTS  
f809-1.wmf  
Figure 809-1. Installing Option -09  
809-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
Table 809-1. Option -09 True RMS AC PCA  
t809-1.wmf  
809-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Option –09 True RMS AC  
LIST OF REPLACEABLE PARTS  
f809-2.wmf  
Figure 809-2. True RMS AC PCA  
809-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
809-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 9  
Schematic Diagrams  
9-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
9-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-01_1.wmf  
Figure 9-1. Main PCA, DC Scaling and F/R Switch  
9-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-01_2.wmf  
Figure 9-1. Main PCA, DC Scaling and F/R Switch (cont.)  
9-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-02_1.wmf  
Figure 9-2. Main PCA, A/D Converter  
9-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-02_2.wmf  
Figure 9-2. Main PCA, A/D Converter (cont.)  
9-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-03_1.wmf  
Figure 9-3. Main PCA, Ohms Current Source  
9-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-03_2.wmf  
Figure 9-3. Main PCA, Ohms Current Source (cont.)  
9-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-04_1.wmf  
Figure 9-4. Main PCA, Digital  
9-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-04_2.wmf  
Figure 9-4. Main PCA, Digital (cont.)  
9-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-05_1.wmf  
Figure 9-5. Main PCA, Power Supply  
9-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-05_2.wmf  
Figure 9-5. Main PCA, Power Supply (cont.)  
9-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-06_1.wmf  
Figure 9-6. Display PCA  
9-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-06_2.wmf  
Figure 9-6. Display PCA (cont.)  
9-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-07_1.wmf  
Figure 9-7. IEEE-488 Interface PCA  
9-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-07_2.wmf  
Figure 9-7. IEEE-488 Interface PCA (cont.)  
9-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Diagrams  
9
f9-08_1.wmf  
Figure 9-8. IEEE-488 Interface PCA, Option -08  
9-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8842A  
Instruction Manual  
f9-08_2.wmf  
Figure 9-8. IEEE-488 Interface PCA, Option –08 (cont.)  
9-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Edelbrock Automobile Parts 9629 User Manual
Elation Professional Work Light 5R User Manual
Euro Pro Food Processor QB900W 30 User Manual
Exmark Lawn Mower TT23KAEP User Manual
Fishman Musical Instrument Pro Eq User Manual
Ford Automobile 2005 Crown Victoria User Manual
Frigidaire Air Conditioner FGHD2472PF User Manual
FujiFilm Digital Camera S600 Zoom Date User Manual
Gateway Computer Hardware E4350 User Manual
GE Monogram Beverage Dispenser ZDWI240 User Manual