Manual Supplement
Manual Title:
Print Date:
Revision/Date: 1, 1/08
Impulse 6000D/7000DP Users Supplement Issue:
3
3/09
10
August 2007
Issue Date:
Page Count:
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title:
CD Rev. & Date:
CD PN:
6000D/7000DP
1, 2/2008
3028681
© 2008-2009 Fluke Corporation. All rights reserved. Printed in the U.S.A.
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Supplement
Impulse 6000D/7000DP Users
Accuracy............................................................. ±0.1 ms
Voltage
Range ................................................................. 20 to 5000 V
Accuracy............................................................. ±(1 % of reading + 2 V)
Current
Range ................................................................. 0.4 to 100.0 A
Accuracy............................................................. ±(1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Interphase delay (biphasic and pulsed biphasic)
Range................................................................. 0.1 ms to 9.9 ms
Accuracy............................................................. ±0.1 ms
Frequency (pulsed biphasic only)
Range................................................................. 2000 Hz to 8000 Hz
Accuracy............................................................. ±1 % of reading
Duty cycle (pulse biphasic only)
Range................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Sample rate ........................................................... 250 kHz (4 μs sample)
Maximum Average Power.................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
2
8/08
Download from Www.Somanuals.com. All Manuals Search And Download.
Impulse 6000D/7000DP Users
Manual Supplement
Scope Output
Autorange ...........................................................2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output .............................................................BNC
Output impedance...........................................50 Ω
Amplitude Accuracy ........................................±5 %
Charge Time Measurement
Range .................................................................0.1 to 100.0 s
Accuracy .............................................................±0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement
Timing window ................................................ECG R-wave peak to the defib pulse peak
Range .............................................................-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380
ms following the R-wave peak.
Resolution.......................................................1 ms
Accuracy .........................................................±1 ms
ECG waves
Normal Sinus Rhythm (NSR)..........................10 to 180 (by 1) BPM
Atrial fibrillation................................................Coarse and fine
Monomorphic Ventricular Tachycardia ...........120 to 240 (by 5) BPM
Asystole ..........................................................Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus ......................................................10 to 300 (by 1) BPM
4/08
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Supplement
Impulse 6000D/7000DP Users
Ventricular Fibrillation......................................... Coarse and fine
Monomorphic Ventricular Tachycardia............... 120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia................. 5 types
Asystole .............................................................. Flat line
ECG Waves
ECG General
Lead configuration.............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ..................................... 1000 Ω (nominal)
Rate accuracy..................................................... ±1 % of nominal
ECG Amplitudes
Reference Lead...................................................Selectable, Lead II (default) or Lead I
Settings................................................................0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Accuracy (All Performance waves and Normal Sinus R waves)
Lead II..................................................................±2 %
All other leads......................................................±5 %
Defib paddles.......................................................±5 %
Amplitude of ECG signals relative to amplitude setting (in percent)
Lead II reference
Performance waves and R wave detection
Lead #
I
II
III
V1
V2
V3
V4
V5
V6
Ref. Amp.
70 %
100 %
30 %
100 %
100 %
100 %
100 %
100 %
100 %
4
3/09
Download from Www.Somanuals.com. All Manuals Search And Download.
Impulse 6000D/7000DP Users
Manual Supplement
Normal Sinus waves
Lead #
I
II
III
V1
V2
V3
V4
V5
V6
Ref. Amp.
70 %
100 %
30 %
24 %
48 %
100 %
120 %
112 %
80 %
Lead I reference
Performance waves and R wave detection
Lead #
Ref. Amp.
I
II
III
V1
V2
V3
V4
V5
V6
100 %
150 %
50 %
100 %
100 %
100 %
100 %
100 %
100 %
Normal Sinus waves
Lead #
I
II
III
V1
V2
V3
V4
V5
V6
Ref. Amp.
100 %
150 %
50 %
24 %
48 %
100 %
120 %
112 %
80 %
ECG Normal Sinus
Rates...................................................................10 to 360 (by 1) BPM
ECG High Level Output (BNC Jack)
Amplitude
Range .............................................................0.5 V per mV of reference lead setting
Accuracy .........................................................±5 %
Output Impedance ..............................................50 Ω
ECG on Defibrillator Input Load
Same as the LEAD II amplitude but limited to ±4 mV
ECG Performance Waves
Square wave.......................................................2.0 and 0.125 Hz
Triangular wave ..................................................2.0 and 2.5 Hz
8/08
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Supplement
Impulse 6000D/7000DP Users
Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse................................................................... 30 and 60 BPM, 60 ms pulse width
R-Wave Detection
Waveform ........................................................... Haver-triangle
Amplitude............................................................ 0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM
Widths................................................................. 8, 10, 12 ms, and 20 to 200 (by 10) ms
Accuracy............................................................. ±(1 % setting + 1 ms)
Noise Immunity
Wave................................................................... Sine
Line Frequency................................................... 50 or 60 Hz (± 0.5 Hz)
Amplitude
Range............................................................. 0.0 to 10.0 (by 0.5) mV
Accuracy......................................................... ±5 %
Transvenous Pacer Pulse Simulation
Widths
Range............................................................. 0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy......................................................... ±5 % of setting
Amplitude
Range............................................................. 0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and
±700 mV
6
4/08
Download from Www.Somanuals.com. All Manuals Search And Download.
Impulse 6000D/7000DP Users
Manual Supplement
Accuracy .........................................................±(10 % of setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)
Lead II reference
Lead #
Ref. Amp.
I
II
III
V1
V2
V3
V4
V5
V6
67 %
100 %
33 %
67 %
67 %
67 %
67 %
67 %
67 %
Lead I reference
Lead #
I
II
III
V1
V2
V3
V4
V5
V6
Ref. Amp.
100 %
150 %
50 %
100 % 100 % 100 % 100 % 100 % 100 %
Arrhythmia Selections
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)
Demand ..........................................................30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only)10 to 250 (by 10) mA
Supraventricular
Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
4/08
7
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Supplement
Impulse 6000D/7000DP Users
Change #2
On page 12, prior to Analyzing Pacemakers (7000DP only), add the following section:
SCOPE OUTPUT
Each time a defibrillator is fired into the Analyzer during an active defibrillator test, a delayed representation of the
defibrillator pulse is sent to the Scope Output jack on the rear panel of the Analyzer. The Scope Output is an
isolated signal and should not present any problems when connected to the input of an oscilloscope.
To assist with the displaying of the defibrillator pulse, synchronizing marker pulses appear on the scope output jack
along with the defibrillator pulse. Table 3A list these markers and describes their purpose.
Table 3A. Synchronizing Marker Pulses
Marker
Defib fire
Ranging
Pulse Characteristics
+2 V, 50 ms
Purpose
Indicates when the defibrillator was fired.
Indicates the scaling used to output the pulse image:
-4 V, 0.4 ms, 1 ms apart
1 pulse……….each volt out equals 80 volts pulse amplitude.
2 pulses……...each volt out equals 400 volts pulse amplitude.
3 pulses……...each volt out equals 2000 volts pulse amplitude.
8
4/08
Download from Www.Somanuals.com. All Manuals Search And Download.
Impulse 6000D/7000DP Users
Manual Supplement
Pulse
+4 V, 1 ms pulse
Start of defibrillator pulse playback [1]
playback
[1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early
transitional signals.
Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the
defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, set the oscilloscope to trigger
on the -4 V markers. To ignore the ranging information and see only the defibrillator pulse, set the oscilloscope to trigger
on levels above +2 volts.
On page 26, prior to Controlling the Analyzer Remotely, add the following section:
Setting the ECG Reference Lead
ECG signal amplitude is set by designating either lead I or lead II as the reference lead. Lead II is the default
selection. To change the reference lead, do the following:
Press Θ to reveal the main setup menu. Next press the softkey labeled More for additional setup
selections. Press the softkey labeled ECG Ref Lead to toggle between lead I and lead II. Once the
reference lead is set, press the softkey labeled Back or one of the other functions to store the selection
in non-volatile memory.
4/08
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Supplement
Impulse 6000D/7000DP Users
Change #3, 48038
On page 38, under Pacemaker Input, change:
From: Accuracy …………..±1 % non-inductive (<2 μH)
To:
Accuracy……………±2 % non-inductive (<2 μH)
10
8/08
Download from Www.Somanuals.com. All Manuals Search And Download.
|