Fisher Weather Radio 444 Alphaline User Manual

00809-0100-4263  
English  
Rev. AA  
Model 444 Alphaline®  
Temperature  
Transmitters  
Product Discontinued  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Product Manual  
1
Model 444 Alphaline®  
Temperature Transmitters  
NOTICE  
Read this manual before working with the product. For personal and system  
safety, and for optimum product performance, make sure you thoroughly  
understand the contents before installing, using, or maintaining this product.  
Within the United States, Rosemount Inc. has two toll-free assistance numbers.  
Customer Central: 1-800-999-9307 (7:00 a.m. to 7:00 p.m. CST)  
Technical support, quoting, and order-related questions.  
North American  
1-800-654-7768 (24 hours a day – Includes Canada)  
Response Center: Equipment service needs.  
For equipment service or support needs outside the United States, contact your  
local Rosemount representative.  
The products described in this document are NOT designed for nuclear-  
qualified applications.  
Using non-nuclear qualified products in applications that require nuclear-  
qualified hardware or products may cause inaccurate readings.  
For information on Rosemount nuclear-qualified products, contact your local  
Rosemount Sales Representative.  
Rosemount, the Rosemount logotype, and Alphaline are registered trademarks of Rosemount Inc.  
Chromel and Alumel are trademarks of Hoskins Mfg. Co.  
Cover Photo: 444-005AC  
Fisher-Rosemount satisfies all obligations coming from legislation  
to harmonize product requirements in the European Union.  
Rosemount Inc.  
8200 Market Boulevard  
Chanhassen, MN 55317 USA  
Tel 1-800-999-9307  
Telex 4310012  
N
TE  
R IN  
S. A.  
I
D
P
U.  
Fax (612) 949-7001  
00809-0100-4263  
© Rosemount Inc. 1998.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1  
SECTION 1  
Introduction  
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1  
General Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1  
Mechanical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Mounting Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Access Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Housing Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Terminal Side of Electronics Housing . . . . . . . . . . . . . . . . . 2-2  
Circuit Side of Electronics Housing . . . . . . . . . . . . . . . . . . 2-2  
Electrical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2  
Field Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3  
Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4  
RTD Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4  
Thermocouple or Millivolt Inputs . . . . . . . . . . . . . . . . . . . . 2-6  
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6  
Multi-Channel Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6  
Surges/Transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6  
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7  
Temperature Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7  
Moist or Corrosive Environments . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Hazardous Location Installation . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Intrinsically Safe Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10  
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10  
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11  
SECTION 2  
Installation  
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1  
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1  
Calibrating a RTD Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2  
Calibrating a Thermocouple Transmitter  
SECTION 3  
Calibration  
Using a Compensated Thermocouple Simulator . . . . . . . . . . . 3-5  
Using an Ice Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7  
Calibrating a Low-Power Transmitter . . . . . . . . . . . . . . . . . . . . . . 3-8  
Calibrating a Millivolt Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . 3-10  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1  
Hardware Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1  
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1  
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2  
Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3  
Reassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5  
Interchangeability of Parts  
SECTION 4  
Maintenance and  
Troubleshooting  
Mechanical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6  
Electrical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6  
Burnout Protection Adjustments . . . . . . . . . . . . . . . . . . . . . . . . 4-6  
Repair and Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7  
Return of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7  
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1  
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3  
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3  
LCD Meter Specifications  
SECTION 5  
Specifications  
and Reference Data  
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7  
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7  
Physical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7  
Analog Meter Specifications  
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8  
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8  
Physical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8  
Mounting Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1  
LCD / Analog Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2  
SECTION 6  
Options  
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1  
APPENDIX A  
Approval Drawings  
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1  
APPENDIX B  
Temperature Sensor  
Reference Information  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
1
Introduction  
This manual is designed to assist in installing, operating, and  
maintaining Rosemount® Model 444 Alphaline® Temperature  
Transmitters.  
OVERVIEW  
Section 2 Installation  
provides mechanical, electrical, and environmental considerations to  
guide you through a safe and effective transmitter installation.  
Section 3 Calibration  
provides different Model 444 calibration procedures.  
Section 4 Maintenance and Troubleshooting  
provides hardware diagnostics, maintenance tasks, basic hardware  
troubleshooting techniques, and considerations for returning materials.  
Section 5 Specifications and Reference Data  
provides functional, performance, and physical transmitter  
specifications; also includes transmitter dimensional drawings,  
ordering information, and spare parts.  
Section 6 Options  
provides a listing of transmitter options and a description of each.  
Appendix A Approval Drawings  
contains approval drawings for Canadian Standards Association (CSA)  
and Factory Mutual (FM) instrinsic safety installation.  
Appendix B Temperature Sensor Reference Information  
provides reference information regarding the application of various  
Rosemount temperature sensors.  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
2
Installation  
This section includes the following transmitter installation  
information:  
OVERVIEW  
• General Considerations  
• Mechanical Considerations  
Mounting Stability  
Access Requirements  
• Electrical Considerations  
Power Supply  
Field Wiring  
Sensor Connections  
Grounding  
Multi-Channel Installations  
Surges/Transients  
• Environmental Considerations  
Temperature Environment  
Moist or Corrosive Environments  
Hazardous Location Installation  
Intrinsically Safe Installation  
• Installation Procedure  
Mechanical  
Electrical  
GENERAL  
CONSIDERATIONS  
Failure to follow these installation guidelines may result in  
death or serious injury. Make sure only qualified personnel  
perform the installation.  
Explosions can cause death or serious injury. Verify that the  
operating atmosphere of the transmitter is consistent with  
the appropriate hazardous locations certifications.  
Use the Rosemount Model 444 Alphaline Temperature Transmitter  
when the temperature measurement point is remote from the control,  
readout, or recording point, or where the measurement point is exposed  
to environmental conditions that would be harmful to unprotected  
signal conditioning equipment.  
Electrical temperature sensors such as RTDs and thermocouples  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
produce low-level signals proportional to their sensed temperature. Model 444  
temperature transmitters convert the low-level sensor signal to a standard 4–20  
mA dc signal that is relatively insensitive to lead length and electrical noise.  
This current signal is then transmitted to the control room via two wires.  
Figures 2-1, and 2-2 show recommended mounting configurations for  
transmitter and sensor assemblies. See Section 6 Options for information  
regarding Model 444 transmitter accessories.  
You can attach the transmitter directly to the sensor assembly as shown in  
Figures 2-1 and 2-2. An optional mounting bracket permits the transmitter to be  
mounted remotely from the sensor(s), either on a flat surface or attached to a  
two-inch pipe (See Figure 2-11 on page 2-13). The choice of mounting method  
must take into account a number of factors:  
MECHANICAL  
CONSIDERATIONS  
Mounting stability is an important consideration. The transmitter, though  
rugged, may require supplementary support under high-vibration conditions,  
particularly if extensive thermowell lagging or long extension fittings are used.  
In such instances, the pipestand mounting technique shown in Figure 2-11 on  
page 2-13 is preferable.  
Mounting Stability  
When choosing an installation location and position, take into account the need  
for access to the transmitter.  
Access  
Requirements  
Housing Rotation  
You may rotate the transmitter in 90-degree increments to improve field access  
to both compartments.  
Terminal Side of  
Electronics Housing  
Make wiring connections through the conduit openings on the terminal side of  
the electronics housing. Mount the transmitter so the terminal side is accessible,  
and be sure to provide adequate clearance for cover removal.  
Circuit Side of  
Electronics Housing  
The transmitter electronics are installed in the circuit side of the transmitter  
housing. In case of electronic malfunction, provide adequate clearance for  
circuit-side cover removal. Also, be sure to account for additional clearance if a  
meter is to be installed. For more information regarding the meter option, refer  
to Section 6 Options.  
2-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
FIGURE 2-1. Recommended  
Process Mounting.  
Union or Coupling  
Sensor Hex  
Thermowell  
Hex  
Extension  
Nipple  
Transmitter  
Conduit for  
Field Wiring  
(dc Power)  
Thermowell  
Extension  
Length  
3.2  
(81.3)  
NOTE  
Dimensions are in inches (millimeters).  
FIGURE 2-2. Recommended  
Process Mounting with  
Drain Seal.  
Union or Coupling  
Sensor Hex  
Coupling  
Extension  
Nipple  
Close Nipple  
Street Ell  
Transmitter  
Terminal  
Side  
Drain Seal  
Conduit for  
Field Wiring  
(dc Power)  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
This section contains information that you should consider when preparing to  
ELECTRICAL  
CONSIDERATIONS  
install Model 444 transmitters. Read this section carefully before going on to the  
installation procedures. Metal conduit should be used to enclose cabling for best  
results in electrically noisy environments.  
The dc power supply should provide power with less than 2% ripple. The input  
voltage versus load limitation relationship for 4–20 mA transmitters is shown in  
Figure 2-3. Figure 2-4 shows field wiring for a standard Model 444 transmitter.  
Figure 2-5 shows field wiring for Models 444LL and 444LM low-power voltage  
output units, which require 100K ohms minimum load. The total R-load is the  
sum of the resistance of the signal leads and the load resistance of the controller,  
indicators, and related devices. Note that the resistance of intrinsic safety  
barriers, if used, must be included in the total load.  
Power Supply  
FIGURE 2-3. Model 444  
Load Limits.  
Power Supply Load Limitations  
R
= 50 ϫ (V  
– 12)  
LOAD (MAX.)  
(MIN.)  
1650  
1500  
Voltage  
Too Low  
1000  
500  
0
Operating  
Region  
12  
20  
30  
40  
Power Supply (V dc)  
Field Wiring  
Explosions may result in death or serious injury.  
Do not remove the instrument cover in explosive  
atmospheres when the circuit is alive.  
High voltage that may be present on leads can cause  
electrical shock. Avoid contact with leads and terminals.  
Do not apply high voltage (e.g. ac line voltage) to the  
transmitter terminals. Abnormally high voltage can damage  
the unit.  
All power to the transmitter is supplied over the signal wiring. Signal wiring  
need not be shielded, but use twisted pairs for best results. Do not run  
unshielded signal wiring in conduit or open trays with power wiring, or near  
heavy electrical equipment. To power the transmitter, connect the positive power  
lead to the terminal marked “+” and the negative power lead to the terminal  
marked “–” (see Figures 2-4 and 2-5). Tighten the terminal screws to ensure  
adequate contact. No additional power wiring is required.  
2-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
FIGURE 2-4. Field Wiring for  
the Standard Model 444  
Transmitters.  
Meter Connections  
and Signal  
(+)  
(+)  
(–)  
{
dc Power  
{
Test Points  
(–)  
+
+
+
RTD Input  
(typical)  
+ +  
Optional  
Ground  
Span Adjust  
Zero Adjust  
FIGURE 2-5. Field Wiring for  
Low-Power Model  
444 Transmitters  
(+) dc Power  
Output Load Limitation  
Minimum Load = 100K  
(444LL and LM).  
(–) Common  
Shield  
Output Voltage (+)  
Power  
Supply  
A to D  
Converter  
RTD Input  
Optional  
Ground  
Span Adjust  
Zero Adjust  
Sensor  
Connections  
Explosion may result in death or serious injury. Do not  
remove the instrument cover in explosive atmospheres  
when the circuit is alive.  
High voltage that may be present on leads can cause electrical  
shock. Avoid contact with the leads and the terminals.  
RTD Inputs  
Various RTD configurations are used in industry; each configuration offers a  
specific solution for compensating the effects of lead wire resistance. They  
include 3- and 4-wire designs. The correct installation for each of these RTDs is  
shown in Figures 2-6a and b on page 2-6.  
If the transmitter is mounted remotely from the RTD, operation will be  
satisfactory, without recalibration, for lead wire resistances of up to 2 ohms per  
lead (equivalent to 200 feet of 20 AWG wire). In this case, the leads between the  
RTD and transmitter should be shielded.  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
The correct connections for a compensation loop RTD and a 2-wire RTD are  
shown in Figure 2-6c and Figure 2-6d, respectively. In a 2-wire RTD, however,  
both leads are in series with the sensor element, so significant errors (0.1 °C)  
could occur if the lead lengths are greater than one foot. For longer runs when  
using a 2-wire RTD, attach a third lead and connect as shown in Figure 2-6a.  
FIGURE 2-6. Sensor Wiring Diagrams.  
Signal  
Signal  
Signal  
Signal  
+
+
+
+
Jumper  
Model 444 with 3-Wire RTD  
Model 444 with 4-Wire RTD  
Model 444 with  
Comp. Loop RTD  
Figure 2-6c  
Model 444 with 2-Wire RTD  
Figure 2-6a  
Figure 2-6b  
Figure 2-6d  
Signal  
+ Output  
Signal  
Signal  
Common –  
+
+
Low T/C  
High T/C  
+
+
+
Model 444LL or 444LM  
with 3-Wire RTD  
Figure 2-6e  
+
Model 444MV used as  
Differential Millivolt  
Transmitter (T/C Junctions  
must be ungrounded)  
Figure 2-6g  
Model 444MV with Millivolt  
Input or Model 444T Series  
with Grounded or  
Ungrounded Thermocouple  
Figure 2-6f  
FIGURE 2-7. Characteristics of Thermocouple and RTD Input Wires.  
Thermocouple Type  
Positive Lead  
Negative Lead  
J
K
T
E
R
S
Iron (Magnetic)  
Constantan (Non-magnetic)  
Chromel (Non-magnetic)  
Copper (Yellow color)  
Chromel (Shiny metal)  
Platinum 13% Rhodium  
Platinum 10% Rhodium  
Alumel (Magnetic)  
Constantan (Silver color)  
Constantan (Dull metal)  
Platinum  
Platinum  
Single Element RTD  
Red  
Compensation Loop RTD  
Red  
Dual Element RTD  
Red  
Red  
White  
White  
Black  
White  
Black  
White  
White  
Green  
Green  
Black  
2-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
In the case of thermocouples, make connections between the sensor and the  
transmitter with thermocouple wire. For process mounting applications, connect  
the thermocouple directly to the transmitter. For installations where the  
transmitter is mounted remotely from the sensor, use appropriate thermocouple  
extension wire. As with all low-level signal wiring, shielding is recommended for  
long runs. Make input connections for the Model 444MV Millivolt Transmitter  
using copper wires. The correct connections for thermocouple and millivolt  
inputs are shown in Figures 2-6f and g.  
Thermocouple or  
Millivolt Inputs  
The transmitter will operate with the current signal loop either floating or  
grounded. However, many types of readout devices are affected by the extra  
noise in floating systems. If operation appears noisy or erratic, grounding the  
current signal loop at a single point may solve the problem. The negative  
terminal of the power supply is the best place to ground the loop. Alternately,  
either side of the readout device could be grounded. Do not ground the current  
signal loop at more than one point.  
Grounding  
Thermocouple and millivolt transmitters are isolated, so the input circuit also  
may be grounded at any single point (when a grounded thermocouple is used,  
this point is the grounded junction), and the signal loop may be grounded at any  
point.  
The 444RL transmitter is not isolated, so there can be no grounds in the RTD  
circuit. Since RTDs must be well-insulated from ground in order to give correct  
temperature readings, this is not normally an installation limitation. The  
positive side of the power supply should not be grounded for use with RTD input  
transmitters. The 444RI9 should be used with grounded RTDs.  
If using shielded wire, connect the shield of the sensor-to-transmitter cable to  
the shield of the transmitters-to-receiver cable. Ground the shielding only at the  
signal loop ground.  
Figure 2-8 illustrates how several transmitters can be connected to a single  
master power supply. In this instance, the system can be grounded only at the  
negative power supply terminal. Since several channels are dependent on one  
supply, an uninterruptible power supply or backup battery should be considered  
if loss of all channels would pose operational problems. The diodes shown in  
Figure 2-8 prevent unwanted charging or discharging of the battery.  
Multi-Channel  
Installations  
FIGURE 2-8. Multi-Channel  
Installation.  
+
Transmitter  
+
No. 1  
dc  
Power  
Supply  
Backup  
Battery  
Readout or  
Controller No. 1  
Transmitter  
No. 2  
Readout or  
Controller No. 2  
To Additional  
Transmitters  
2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
The transmitter will withstand electrical transients of the energy level usually  
Surges/Transients  
encountered in static discharges or induced switching transients. However, high-  
energy transients, such as those induced in wiring from nearby lightning strikes,  
can damage both the transmitter and the sensor.  
To protect against high-energy transients, install Model 444 transmitters in  
conjunction with the Rosemount Model 470 Transient Protector. The Model 470  
prevents damage from transients induced by lightning, welding, heavy electrical  
equipment, or switch gears. Refer to the Model 470 Transient Protector product  
data sheet, pub. no. 00813-0100-4191 for more information.  
ENVIRONMENTAL  
CONSIDERATIONS  
The transmitter will operate within specifications for ambient temperatures  
between –25 and 85 °C. It will function, but not necessarily within specifications,  
in ambient temperatures between  
Temperature  
Environment  
–40 and 100 °C.  
Aside from ambient temperature variations, heat from the process is transferred  
from the thermowell to the transmitter housing. If the process temperature is  
near or beyond specification limits, use excess thermowell lagging or an  
extension nipple to protect the transmitter from the high temperature condition.  
See Figure 2-10.  
EXAMPLE:  
Suppose the maximum ambient temperature is 40 °C and the temperature to  
be measured is 540 °C. The maximum allowable housing temperature rise is  
the rated temperature specification limit minus the existing ambient  
temperature (85 – 40), or 45 °C. As shown in Figure 2-9, an “E” dimension of  
3.6 inches will result in a housing temperature rise of 22 °C. An “E”  
dimension of 4 inches would therefore be the minimum recommended length,  
and would provide a safety factor of about about 25 °C. A longer “E”  
dimension, such as 6 inches, would be desirable in order to reduce errors  
caused by transmitter temperature effect, although in that case the  
transmitter would probably require extra support. If a thermowell with  
lagging is used, the “E” dimension may be reduced by the length of the  
lagging.  
2-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
FIGURE 2-9. Model 444  
Transmitter Housing  
Temperature Rise.  
60  
(108)  
Transmitter Housing Temperature Rise  
E
vs.  
Length for a Test Installation  
50  
(90)  
40  
(72)  
30  
(54)  
HOUSING  
RISE  
ABOVE  
AMBIENT  
°C (°F)  
22  
20  
(36)  
10  
(18)  
0
3
4
5
6
7
8
9
3.6  
E
Length in Inches  
FIGURE 2-10. Sensor  
Assembly Dimensional  
Drawings.  
Connection Head with Extended Cover  
Bayonet Spring Loaded Sensor  
½–14 NPT Thread  
Sensor Installed in  
Connection Head  
(Extended Cover)  
with Coupling and  
Nipple Extension  
0.25  
(6)  
L
Sensor X Length (Ref.)  
1.0  
5.5  
(140)  
(25)  
Nominal Fitting  
2.2  
(56)  
3.5  
(89)  
0.53 (13) Max.  
Thread  
E
Length⅜  
Engagement  
Spring Loaded Sensor  
¾–14 NPT  
on Thermowell  
(1)  
Connection Head with  
Flat Cover  
+ 1.75 (44)  
T
Sensor Installed in  
Connection Head (Flat  
Cover) with Union and  
Nipple Extension and  
Thermowell  
U
¾–14 NPT  
Chain  
NOTE  
(1)  
Dimensions are in inches (millimeters).  
T
= 0.0 on Standard Assembly Thermowells  
2-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
The transmitter is designed to resist attack by moisture and other corrosives.  
Moist or Corrosive  
Environments  
The coated circuit boards are mounted in a compartment completely sealed from  
the conduit entrances. O-ring seals protect the interior when the covers are  
installed. In humid environments, however, it is still possible for moisture  
“breathing” to occur in conduit lines. If the transmitter is mounted at a low point  
in the conduit run, the terminal compartment could fill with water, causing  
electrical shorting. The transmitter should be mounted so moisture from the  
conduit will not drain into the housing. In some instances a drain seal, installed  
as shown in Figure 2-2 on page 2-3, is advisable.  
Hazardous Location  
Installation  
Explosions may result in death or serious injury. Verify that  
the operating atmosphere of the transmitter is consistent  
with the appropriate hazardous locations certifications.  
Explosions may result in death or serious injury. Both  
transmitter covers must be fully engaged to meet explosion-  
proof requirements.  
The Model 444 is designed with an explosion-proof housing and circuitry  
suitable for intrinsically safe and non-incendive operation. Individual  
transmitters are clearly marked with a tag indicating the approvals they carry.  
The various approvals are available as options. Refer to Section 5  
Specifications and Reference Data for a complete listing of available  
approvals.  
To maintain certified ratings for installed transmitters, install in accordance  
with applicable installation codes and approval drawings. Refer to Appendix A  
Approval Drawings for Model 444 installation drawings. For future orders,  
refer to the current product price list for the most up-to-date information on  
these approvals.  
Intrinsically Safe  
Installation  
For explosion-proof installations, installation location must  
be made in accordance with Rosemount drawing 00444-  
0261, Rev. E.  
For intrinsically safe installations, installation location must  
be made in accordance with Rosemount drawing 00444-  
0034, Rev. C (CSA) or 00444-0264, Rev. B (SAA).  
You can use Intrinsically safe installations instead of explosion-proof  
installations in hazardous areas. In such configurations, the transmitter and  
sensor are located in a hazardous area, and the current signal leads are  
connected to equipment in a non-hazardous area through intrinsic safety  
barriers that limit the voltage and current fed into the hazardous area. Install in  
accordance with the barrier manufacturer’s instructions for the specific barrier  
used. For approval information, refer to Table 5-1 on page 5-4, and Table 5-2 on  
page 5-5. For installation information, refer to the intrinsically safe barrier  
systems reference drawings in Appendix A Approval Drawings.  
2-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Installation consists of mounting the transmitter and sensor assembly and  
making electrical connections. If mounting the transmitter directly to the sensor  
assembly, use the process mounting technique shown in Figure 2-1 or Figure 2-2,  
on page 2-3. For transmitter locations remote from the sensor, use conduit  
between the sensor and transmitter. Transmitter hubs will accept male conduit  
fittings with ½–14 NPT; ½–14 NPSM; or ½–14 taper thread per ANSIC 80.4.  
INSTALLATION  
PROCEDURE  
Explosion may result in death or serious injury. Do not  
remove the instrument cover in explosive atmospheres  
when the circuit is alive.  
High voltage that may be present on leads may cause  
electrical shock. Avoid contact with the leads and the  
terminals.  
Process leaks may result in death or serious injury. Install  
and tighten thermowells or sensors before applying  
pressure, or process leakage may result. Removing the  
thermowell or sensor while in operation may cause process  
fluid leaks.  
1. Mount the thermowell to the pipe or process container wall.  
Mechanical  
2. Attach any necessary extension nipples and adapters. Seal the nipple and  
adapter threads with silicone or tape.  
3. Screw the sensor into the thermowell.  
4. Install drain seals if required for severe environments or to satisfy code  
requirements (See Figure 2-2 on page 2-3).  
5. Attach the transmitter to the thermowell assembly. Seal the adapter  
threads with silicone or tape.  
6. Install conduit for field wiring to the remaining conduit entry of the  
transmitter. Seal conduit threads with silicone or tape.  
7. Pull field wiring leads through the conduit into the terminal side of the  
transmitter housing.  
Electrical  
For explosion-proof installations, wiring connections must  
be made in accordance with Rosemount drawing 00444-  
0261, Rev. E.  
For intrinsically safe installations, wiring connections must  
be made in accordance with ANSI/ISA-RP12.6, and  
Rosemount drawing 00444-0034, Rev. C (CSA) or 00444-  
0264, Rev. B (SAA).  
For all installations, wiring connections must follow the  
National Electric Code.  
2-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Preliminary Checkout  
1. For any Model 444 unit, first verify that the transmitter is calibrated to  
the required range. Calibration is usually performed by substituting an  
input in place of the sensor, and this is most conveniently accomplished  
prior to sensor connection. Refer to the calibration procedures in Section  
4 Maintenance and Troubleshooting.  
Input Connections  
2. Model 444RL: Connect the RTD leads as shown in Figure 2-6a, b, c, d, or  
e depending upon the lead compensation method used.  
Model 444T series: Connect the thermocouple leads as shown in Figure  
2-6f. Polarity is important; be sure to identify the leads accurately. The  
negative lead is usually red; if there is no color coding, the characteristics  
provided in Figure 2-7 may be helpful.  
Model 444MV: If using the transmitter as a millivolt-to-milliampere  
converter, use ordinary copper leads for input connections as shown in  
Figure 2-6f. If using the transmitter with two thermocouples to measure  
differential millivolt, connect the thermocouples as shown in Figure 2-6g.  
The “high” thermocouple causes the transmitter output to increase when  
its temperature increases relative to the “low” thermocouple. Grounded  
thermocouples cannot be used for differential measurements.  
Models 444LL and LM: In these low-power option packages, the RTD  
leads are connected the same as in the conventional RTD arrangements  
shown in Figure 2-6a, b, c, and d.  
Output Connections  
3. For all 4–20 mA models, use ordinary copper wire of sufficient size to  
assure that the voltage across the transmitter power terminals does not go  
below 12 V dc (See Figure 2-3). For multi-channel or intrinsically safe  
installations, see applicable paragraphs in this section.  
Model 444RL: Connect current signal leads as shown in Figure 2-6a, b, c,  
or d.  
Model 444T series: Connect current signal leads as shown in Figure 2-6f.  
Model 444MV: Connect current signal leads as shown in Figure 2-6f or g.  
Models 444LL and LM: Connect current signal leads as shown in Figure  
2-6e.  
Final Checkout  
4. For all models, recheck the polarity and correctness of connections; then  
turn the power on.  
2-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
FIGURE 2-11. Model 444  
with  
Optional Mounting Bracket.  
PANEL OR SURFACE MOUNTING  
PIPESTAND MOUNTING  
Transmitter can be Rotated 90°  
5
/16 -inch Bolts  
(four required,  
not furnished)  
Hole for  
/16 -inch  
Bolts  
5
2.81 (81)  
(four  
Mounting Bracket  
places)  
¼–20 ϫ½-inch  
Bolt (4)  
Clearance Hole  
for ¼-inch Bolt  
(eight places)  
5
/16 –18 U-bolt for  
2-inch Pipe (2)  
5.00 (127)  
NOTE  
Dimensions are in inches (millimeters).  
FIGURE 2-12. Model 444  
Dimensional Drawings  
0.75 (19)  
Clearance for  
Cover Removal  
(Typical)  
7.5 (191) Max. with Optional  
Meter  
4.5 Max.  
(114)  
4.5 Max.  
(114)  
Permanent Tag  
(Optional)  
Meter Housing  
½–14 NPT per  
ANSI C80.4 for  
Conduit or Sensor  
Connection  
4.5 Max.  
(114)  
Terminal  
Circuitry  
this Side  
(two places)  
4.2  
(117)  
Terminal  
Connections  
this Side  
Nameplate  
Explosion Proof or  
Intrinsic Safety Label  
(Optional)  
0.36 (9)  
0.72 (18)  
Mounting Holes  
¼–20 UNC–2B  
0.375 (10) Min. Dp.  
(four places)  
0.87 (22)  
NOTE  
Dimensions are in inches (millimeters).  
1.7  
(44)  
2-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
2-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
3
Calibration  
OVERVIEW  
Each transmitter is factory calibrated to the temperature range shown  
on the nameplate. If calibration to a specific range is not specified on  
the purchase order, the transmitter is calibrated to maximum span  
with a base temperature of 0 °C, and the “Calibration” entry on the  
transmitter nameplate is left blank. For more specific calibration  
information and a complete breakdown of transmitter parts, refer to  
Section 5 Specifications and Reference Data.  
Only a few calibration laboratories have the kind of precision  
temperature baths necessary for accurate direct calibration of a  
temperature sensor or sensor/transmitter system. As a result, the  
transmitter is normally calibrated by substituting a resistance decade  
box for an RTD or a compensated thermocouple simulator for a  
thermocouple.  
This section contains the following transmitter calibration information:  
• Calibrating a RTD Transmitter  
• Calibrating a Thermocouple Transmitter  
• Calibrating a Low-Power Transmitter  
• Calibrating a Millivolt Transmitter  
This section contains procedures that require removing the transmitter  
covers and making electrical connections. The following safety  
messages apply to all such procedures.  
SAFETY MESSAGES  
Explosion may result in death or serious injury. Do not  
remove the instrument cover in explosive atmospheres  
when the circuit is alive.  
High voltage that may be present on leads can cause  
electrical shock. Avoid contact with the leads and the  
terminals.  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Calibration Equipment Required:  
CALIBRATING A  
Readout Resistor. The transmitter test terminals give a 40–200 mV signal. The  
RTD TRANSMITTER  
Models 444RL and444RL ___B0912 have a jumper-selectable 4–20 mA test  
output option (2-board sets). If this is not suitable for the test equipment  
available, a 0.1% tolerance, 0.5 W precision wirewound resistor is needed.  
Suggested values include a 100-ohm resistor to give a 0.4 to 2 volt output; or 500  
ohms for 2 to 10 volts.  
Voltmeter (such as a 5-digit DVM). Voltage rating is dependent upon the test  
signal. Accuracy is 0.01%; resolution is 1 mV.  
dc Power Supply. Power capability is 24 V dc at 35 mA.  
Resistance Decade Box. Precision type, 5-dial, with largest dial providing 100-  
ohm steps. Accuracy is 0.02 ohm. The decade box should be periodically  
calibrated against a 5-dial Wheatstone bridge.  
Lead Simulation Resistors. If the transmitter is to be mounted remote from the  
RTD, and the lead resistance between the transmitter and the RTD is greater  
than 2 ohms per lead (equivalent to 200 ft of 20 AWG wire), the transmitter  
should be trimmed with simulated lead resistances for best accuracy. This  
requires wirewound resistors with resistance values equal to the nominal lead  
resistance of the RTD.  
Calibration Procedure  
To calibrate a model 444RL or 444RL___B0912, perform the following procedure:  
1. The Models 444RL and 444RL ___B0912 have a jumper-selectable 4–20  
mA test output option. If a 4–20 mA test output is required, reposition the  
test terminal output jumper on the range board (the default setting is 40–  
200 mV). Refer to steps 2 through 4 of the disassembly procedure, on page  
4-4, for information on removing the circuit board assembly.  
Place the jumper in the position labelled “A” for a 4–20 mA test output. See  
Figure 3-1. Refer to steps 5 through 12 of the reassembly procedure, on  
page 4-5, for information on reinstalling the circuit board assembly.  
2. Remove the cover from the terminal side of the transmitter housing.  
3. If an RTD is already connected, remove all RTD lead connections.  
4. Attach the calibration test equipment as shown in Figure 3-2. Use  
miniature banana plugs to make terminal connections. Use simulated lead  
resistors only if necessitated by long lead wire lengths, as discussed above.  
NOTE  
If using RTD configurations other than the 3-wire design shown in Figure 3-2,  
refer to Figure 2-6 on page 2-6 for the correct wiring.  
5. If trimming the transmitter to a new range, you may have to reposition the  
Coarse Zero Jumper on the Range Board. If so, see the disassembly  
procedure on page 4-4. Position the jumper in the location shown in Table  
3-1. (A transmitter with a Base Temperature outside the regions shown in  
Table 3-1 is a special design, and does not contain a Coarse Zero Jumper.)  
Reassemble the circuit boards.  
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
6. Determine the RTD resistance at the desired base and full scale  
temperatures. For Calibration Code 1 (see Table 5-3, on page 5-9), these  
resistances are listed in Table B-1.  
7. Turn the power on.  
8. Set the decade box to the resistance corresponding to the desired base  
temperature. Adjust the zero potentiometer until the output is 4 mA.  
Remember that recovery time of the unit from an underscale condition is  
longer than from an over-scale condition. Therefore, set the box to a higher  
resistance than that desired, then bring it down to the correct value.  
9. Set the decade box to the resistance corresponding to the desired full scale  
temperature. Adjust the span potentiometer until the output is 20 mA.  
10. Repeat steps 8 and 9 until you obtain the 4 and 20 mA readings without  
readjusting the span and zero potentiometers. Complete this process more  
quickly by noting the full scale reading before readjusting the span pot,  
using the span pot to overshoot the desired reading by 20%, and then  
using the zero pot to readjust the full scale reading to 20 mA.  
EXAMPLE:  
To calibrate the Model 444RL1U1 for a range of 100 to 150 °F (38 to 66 °C),  
first consult Table 3-1, and plug the jumper into pins Z2. From Table B-1,  
trim points are 114.68 and 125.37 ohms corresponding to 100 °F and 150 °F  
respectively. After adjusting the base to 4 mA, and setting the decade to full  
scale resistance, output equals 22.5 mA, or 2.5 mA greater than desired. Set  
the span pot to an output lower than 20 mA by the amount equal to 20% of  
2.5 equals 0.5 mA, or 19.5 mA. Reset the zero pot so the output equals 20 mA.  
Repeat steps 8 and 9 and this procedure until readjustments are no longer  
necessary.  
11. Disconnect the decade box and the readout resistor. Reconnect the RTD  
and power leads. Replace the terminal cover.  
12. Mark the correct range in the “Calibration” space on the nameplate  
.
TABLE 3-1. Coarse Jumper  
Location, Model 444R.  
Base Temperature  
Jumper Location  
Region  
° C  
° F  
444RL1  
444RL2  
444RL3  
–50 to 0  
0 to 50  
50 to 100  
100 to 150  
–58 to 32  
32 to 122  
122 to 212  
212 to 302  
Z1  
Z2  
Z3  
Z4  
Z1  
Z1  
Z2  
Z2  
Continuously  
adjustable  
(no jumper)  
NOTE  
If the base temperature is at a dividing point between regions, use the  
lower jumper position optimum performance; i.e., use location Z1 rather  
than Z2 for Model 444RL1 with a base temperature of 0 °C.  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
FIGURE 3-1. Location of  
Test Input and Burnout  
Protection Jumper on Model  
444RL Range Board.  
Test Output  
Jumper Position  
Burnout Protection  
Jumper Position  
FIGURE 3-2. RTD  
Transmitter Calibration  
Diagram.  
Transmitter  
Power  
Supply  
DVM  
Readout  
Resistor  
Decade Box  
DVM  
Alternate Readout  
Lead Simulator Resistors  
(If required)  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
CALIBRATING A  
THERMOCOUPLE  
TRANSMITTER  
Calibration Equipment Required  
Using a  
Compensated Thermocouple Simulator. Precision voltage source providing  
conformity to NIST Monograph 125 thermocouple curves. Reflect accuracy of  
simulator to desired calibration span. A simulator accuracy four times better  
than the transmitter is recommended (0.05% of calibrated span or 0.005mV  
whichever is greater). Simulator inaccuracies greater than this will degrade  
system accuracy and factory calibration is recommended.  
Compensated  
Thermocouple  
Simulator  
Voltmeter. Such as a 5-digit DVM. Accuracy is 0.01%; resolution is 1 mV.  
dc Power Supply. Power capability is 24 Vdc at 35 mA.  
Thermocouple Wire.Use the same type as that used in the construction of the  
thermocouple.  
Readout Resistor. The transmitter test terminals give a 40–200 mV signal. If  
this is not suitable for the test equipment available, a 0.1% tolerance, 0.5 W  
precision wirewound resistor is needed. Suggested values include a 100-ohm  
resistor to give a 0.4 to 2 volt output; 250 ohms for 1 to 5 volts; or 500 ohms for  
2 to 10 volts.  
Calibration Procedure  
1. Remove the cover from the terminal side of the transmitter housing.  
2. If a thermocouple is already connected, remove all thermocouple lead  
connections.  
3. Connect the equipment as shown in Figure 3-4. Be sure to maintain  
polarity from the transmitter to the thermocouple simulator. Make  
terminal connections using miniature banana plugs.  
4. If trimming the transmitter to a new range, you may have to reposition the  
Coarse Zero Jumper on the Range Board. If so, see the Disassembly  
Procedure on page 4-4. Position the jumper in the location shown in Table .  
(A transmitter with a base temperature outside the regions shown in Table  
is a special design and does not contain a Coarse Zero Jumper. Also, Model  
444 TR and TS transmitters do not have Coarse Zero Jumpers.)  
Reassemble the circuit boards.  
5. Determine the base and full scale temperatures.  
6. Turn the power on.  
7. Refer to the thermocouple simulator instructions for setting the  
thermocouple type and engineering units. Set the simulator to the base  
(zero) temperature and adjust the zero pot until the output is 4 mA (or 40  
mV at the test terminals).  
8. Set the simulator to the full scale temperature and adjust the span pot  
until the output is 20 mA (or 200 mV at the test terminals).  
9. Repeat steps 7 and 8 until you obtain the 4 and 20 mA readings without  
readjusting the pots. Use the “overshoot” technique described in step 9 of  
the RTD calibration procedure, if desired.  
10. Disconnect the simulator leads. Reconnect the thermocouple and power  
leads, if required. Replace the terminal side housing cover.  
11. Mark the new range in the “Calibration” space on the nameplate.  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 3-2. Coarse Zero  
Jumper Location, Model  
444TJ, TK, TF, TT, and MV  
Range Code 1.  
Transmitter Base Region  
Coarse Zero  
Jumper Location  
°C  
°F  
mV  
–50 to 50  
50 to 100  
–58 to 122  
122 to 302  
-2 to 3  
3 to 8  
Z1  
Z2  
NOTE  
Range Codes 2 and 3 are continuously adjustable over the range  
shown in Table 1. (No Coarse Zero Jumper)  
FIGURE 3-3. Location of  
Burnout Protection Jumper  
on Model 444T Range  
Board.  
Burnout R2, R3  
Protection Resistors  
.
FIGURE 3-4. Compensated  
Thermocouple Simulator  
Calibration Diagram.  
DVM  
+
Thermocouple  
Wire  
+
Power  
Supply  
Readout Resistor  
Thermocouple  
Simulator  
+
+
+
Transmitter  
DVM  
(Alternate Readout)  
3-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
Calibration Equipment Required  
Using an Ice Bath  
Millivolt Source. Precision voltage source providing outputs from –10 to 100  
mV. Reflect accuracy of four times better that the 444 transmitter is  
recommended (0.05% of calibrated span or 0.005 mV which ever is greater).  
Voltmeter. Such as a 5-digit DVM. Accuracy is 0.01%; resolution is 1 mV.  
dc Power Supply. Power capability is 24 V dc at  
35 mA.  
Thermocouple Wire.Use the same type as that used in the construction of the  
thermocouple.  
Readout Resistor. The transmitter test terminals give a 40–200 mV signal. If  
this is not suitable for the test equipment available, a 0.1% tolerance, 0.5 W  
precision wirewound resistor is needed. Suggested values include a 100-ohm  
resistor to give a 0.4 to 2 volt output; 250 ohms for 1 to 5 volts; or 500 ohms for  
2 to 10 volts.  
Ice Bath. For highest accuracy, a stirred ice bath (such as a Rosemount 911A)  
should be used, as well as ice made from deionized or distilled water.  
Input Monitor Voltmeter. Use to monitor source when required. Resolution of  
0.001 mV for ranges up to 100 mV. This can be the same as voltmeter used to  
measure transmitter output if rangeability and resolution are sufficient for  
both levels.  
Calibration Procedure  
1. Remove the cover from the terminal side of the transmitter housing.  
2. Install the equipment as shown in Figure 3-5 and allow the thermocouple  
junctions to stabilize at the ice point. Make terminal connections with  
miniature banana plugs.  
3. If trimming the transmitter to a new range, you may have to reposition the  
Coarse Zero Jumper on the Range Board. If so, see the disassembly  
procedure on page 4-4. Position the jumper in the location shown in Table .  
(A transmitter with a base temperature outside the regions shown in Table  
is a special design and does not contain a Coarse Zero Jumper. Also, Model  
444 TR and TS transmitters do not have Coarse Zero Jumpers.)  
4. Determine the thermocouple millivolt levels at the desired base and full  
scale temperatures. See NIST Monograph 125 or  
Table B-1.  
5. Turn the power on.  
6. Set the millivolt source until the monitoring voltmeter reads the emf  
corresponding to the desired base temperature. Adjust the zero pot until  
the output is 4 mA.  
7. Set the millivolt source until the monitoring voltmeter reads the emf  
corresponding to the desired full scale temperature. Adjust the span pot  
until the output is 20 mA.  
8. Repeat steps 6 and 7 until you obtain the 4 and 20 mA readings without  
readjusting the pots. Use the “overshoot” technique described in step 9 of  
the RTD calibration procedure, if desired.  
9. Mark the correct range in the “calibration” space on the nameplate.  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
FIGURE 3-5. Ice Bath  
Calibration Diagram.  
+
Input  
DVM  
Monitor  
DVM  
+
+
Thermocouple  
Wire  
Power  
Supply  
+
+
Millivolt  
Source  
+
DVM  
(Alternate Readout)  
Ice Bath  
Calibration Equipment Required  
CALIBRATING A  
Voltmeter. Such as a 5-digit DVM. Accuracy is 0.01%; resolution is 1 mV.  
LOW-POWER  
TRANSMITTER  
dc Power Supply. Power capability is 5 V dc at 1.5 mA for Model 444LL and 8 V  
dc at 2 mA for Model 444LM.  
Resistance Decade Box. Precision type, 5-dial, with largest dial providing 100-  
ohm steps. Accuracy is 0.02 ohm. The decade box should be periodically  
calibrated against a 5-dial Wheatstone bridge.  
Lead Simulation Resistors. If the transmitter is to be mounted remote from the  
RTD, and the lead resistance between the transmitter and the RTD is greater  
than 2 ohms per lead (equivalent to 200 ft of 20 AWG wire), the transmitter  
should be trimmed with simulated lead resistances for best accuracy. This  
requires wirewound resistors with resistance values equal to the nominal lead  
resistance of the RTD.  
Load Resistor. If the transmitter is to be operated with a load that is  
significantly different from the DVM used for calibration, a load resistor can be  
used for best calibration accuracy. A metal film, carbon comp, or wirewound  
resistor, as well as a decade box can be used to simulate the actual load.  
3-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
The following steps describe the procedure for calibrating a low-power  
Calibration Procedure  
transmitter, as shown in Figure 3-6:  
1. Remove the cover from the terminal side of the transmitter housing.  
2. If an RTD is already connected, remove all RTD lead connections.  
3. Attach the calibration test equipment as shown in Figure 3-6. Make  
terminal connections using miniature banana plugs. Use simulated lead  
resistors only if necessitated by long lead wire lengths, as discussed above.  
NOTE  
If using RTD configurations other than the 3-wire design shown in Figure 3-2,  
refer to Figure 2-6 on page 2-6 for the correct wiring.  
4. Determine the RTD resistance at the desired base- and full-scale  
temperatures. For Calibration Code 1 (see Table 5-5), obtain these  
resistances from Table B-1.  
5. Turn the power on.  
6. Set the decade box to the resistance corresponding to the desired base  
temperature. Adjust the zero potentiometer until the output is 0.8 V for  
Model 444LL or 1.0 V for Model 444LM.  
7. Set the decade box to the resistance corresponding to the desired full-scale  
temperature. Adjust the span potentiometer until the output is 3.2 V for  
Model 444LL or 5.0 V for Model 444LM.  
8. Repeat steps 6 and 7 until you obtain both the zero- and full-scale readings  
without adjusting the span and zero potentiometers. Complete this process  
more quickly by noting the full-scale reading before readjusting the span  
pot, using the span pot to overshoot the desired reading by 20%, and then  
using the zero pot to readjust the full scale reading.  
9. Disconnect the decade box and the readout. Reconnect the RTD and power  
leads. Replace the terminal cover.  
10. Mark the correct range in the “Calibration” space on the nameplate.  
FIGURE 3-6. Low-Power  
Transmitter Calibration  
Diagram.  
Load Resistors ➀  
(If required)  
444LL and LM Transmitter  
+
+
dc Power  
Source  
DVM  
Decade Box  
Lead Simulator Resistors  
(If required)  
Transmitters are calibrated  
at the factory with a 220 K load.  
3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Calibration is identical to the thermocouple type (see Figure 3-5 on page 3-8),  
CALIBRATING A  
MILLIVOLT  
TRANSMITTER  
except that a reference junction and ice bath are not used. The millivolt source is  
connected directly to the transmitter input terminals with copper wire, and the  
desired millivolt levels are entered directly. See Table 3-2 for Coarse Zero  
Jumper locations.  
3-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
4
Maintenance and  
Troubleshooting  
This section contains the following transmitter maintenance and  
troubleshooting information:  
OVERVIEW  
Hardware Diagnostics  
• Troubleshooting  
• Repair  
Disassembly Procedure  
Reassembly Procedure  
Interchangeability of Parts  
Burnout Protection Adjustments  
• Repair and Warranty Service  
• Return of Materials  
Use only the procedures and new parts specifically  
referenced in this manual. Unauthorized procedures or  
parts can affect product performance and the output signal  
used to control a process, and may render the instrument  
dangerous. Direct any questions concerning these  
procedures or parts to Rosemount Inc.  
If you suspect a malfunction, refer to Table 4-1 to verify that  
transmitter hardware and process connections are in good working  
order. Under each of the seven major symptoms, you will find specific  
suggestions for solving the problem. Always deal with the most likely  
and easiest-to-check conditions first.  
HARDWARE  
DIAGNOSTICS  
This section offers tips for troubleshooting several kinds of potential  
malfunctions. To determine a malfunction, use pin-like probes to break  
through the protective coating to make measurements on a circuit  
board.  
TROUBLESHOOTING  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 4-1. Transmitter  
Troubleshooting Symptoms  
and Corrective Actions.  
Symptom  
Potential Source  
Corrective Action  
High Output  
Sensor  
Check for a sensor or thermocouple open circuit. (RL, MV, T-Series with upscale burnout protection only)  
Check for dirty or defective terminals, interconnecting pins, or receptacles.  
Check for dirty or defective interconnecting pins.  
Loop Wiring  
Electronics Assembly  
Loop Wiring  
Erratic Output  
Check for adequate voltage to the transmitter.  
Check for intermittent shorts, open circuits, and multiple grounds.  
Check for dirty or defective terminals or interconnection pins.  
Electronics Assembly  
Sensor  
Check for dirty or defective interconnecting pins.  
Low Output or  
No Output  
Check RTD leads to ensure that they are not shorting together or to ground. (RL only)  
Check for correct RTD lead connection. (RL only)  
Check for open RTD lead on double-lead side. (RL only)  
Loop Wiring  
Loop Wiring  
Check for adequate voltage to the transmitter. (RL only)  
Check for intermittent shorts, open circuits, and multiple grounds. (MV, T series only)  
Check for proper polarity at the signal terminal. (MV, T series only)  
Check for dirty or defective terminals or interconnection pins.  
Excessive  
Current  
(over 30 mA)  
Check for short between current signal leads.  
Check to ensure that current signal leads ARE NOT connected to sensor terminals.  
Check that sensor leads ARE NOT grounded when positive side of power supply is grounded (RL,  
RL___B0912, LL, and LM)  
Electronics Assembly  
Sensor  
Check for defective components in amplifier or current control section.  
Excessive  
Check for incorrect thermocouple type or Incorrect thermocouple polarity connection (T series only)  
Output Shift  
with Ambient  
Temperature  
Electronics Assembly  
Check to ensure that the burnout -protection jumpers positioned correctly (MV, T series only).  
Check for defective components in voltage regulator or dc-to-ac converter section (MV, T series only).  
Check for defective components in amplifier or current control section (all models).  
May require replacement electronics assembly.  
Unit Cannot be  
Trimmed to  
Desired Base  
Temperature  
Transmitter  
Check to ensure that unit is capable of desired range.  
Electronics Assembly  
Check to ensure that the range board jumper is in the correct position.  
Check to ensure that the burnout -protection jumpers positioned correctly. (MV, T series only).  
Check for defective zero pot.  
Unit Cannot be  
Trimmed to  
Desired Span.  
Transmitter  
Check to ensure that unit is capable of desired range.  
Check for adequate voltage to the transmitter.  
Loop Wiring  
Electronics Assembly  
Check for defective components in amplifier or current control section.  
Check for defective components in voltage regulator section.  
Check for defective span pot.  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance and Troubleshooting  
REPAIR  
Exposure to hazardous substances can cause death or  
serious injury. If a hazardous substance is identified, a  
Material Safety Data Sheet (MSDS), required by law to be  
available to people exposed to specific hazardous  
substances, must be included with the returned materials.  
In case of a failure, particularly one in which the transmitter’s output goes to  
one extreme and stays there, the first step is to determine whether the fault lies  
with the sensor(s) or the transmitter. Although only a careful calibration can  
determine sensor shifts, catastrophic failures (such as an open or shorted sensor  
element) can be checked with an ohmmeter at the time the transmitter is  
disconnected from the sensor(s).  
NOTE  
The Resistance vs. Temperature and Millivolt vs. Temperature tables for the  
standard Model 444 sensor input types are presented in Table B-1 in Appendix  
B Temperature Sensor Reference Information.  
RTD Test: A platinum RTD with an ice-point (°C) resistance of 100 ohms should  
read approximately as shown in Table B-1 at other temperatures. The resistance  
between the two leads on the same side of the sensing element should be low, a  
few ohms at most. Resistance between any of the RTD leads and the sensor  
sheath should be high  
(1 megohm or greater).  
Thermocouple Test: Thermocouple resistance should be low (10 ohms or less  
for short runs of heavy wire). For longer runs of extension wire, resistance will  
be roughly ten times the resistance of copper wire of the same diameter. If the  
sensor and receiving equipment are functioning properly, the transmitter will  
probably require repair.  
The transmitter is designed for easy replacement of its plug-in, modular circuit  
boards. A malfunction can be most easily isolated by substituting boards one at a  
time until the unit functions properly.  
It is recommended that customers return defective circuit boards to Rosemount  
Inc. for repair (see “RETURN OF MATERIALS” on page 4-6). This ensures that  
replacement parts meet the design criteria for the board and that the  
malfunctioning board is completely checked and repaired.  
Rosemount Inc. offers a circuit board repair/replacement program through its  
many service centers. Please contact your Rosemount field sales office for price  
and delivery information.  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Disassembly  
Procedure  
Explosion can cause death or serious injury. Do not remove  
the instrument cover in explosive atmospheres when the  
circuit is alive.  
High voltage that may be present on leads can cause  
electrical shock. Make sure all power to the transmitter is off  
before wiring.  
NOTE  
The numbers in parentheses refer to parts shown in the Illustrated Parts List,  
Table 5-7, in Section 5 Specifications and Reference Data.  
1. Terminal blocks for making all field wiring electrical connections are  
located in a compartment identified as “Terminal Side” on the nameplate.  
The sensor terminals, power supply and signal-test terminals, as well as  
the zero and span adjustments are accessible by removing the Electronics  
Housing Cover (2) from the terminal side. The terminals are permanently  
attached to the housing and must not be removed.  
2. The transmitter electronics Circuit Board Assembly (5, 6, 7) is located in a  
separate compartment, identified as “Circuit Side” on the nameplate.  
Make sure power is off. Then remove Circuit Side Cover (2).  
NOTE  
On the standard RTD input (Model 444RL), fast turn-on (Model 444RL  
___B0912), and low-power (Models 444LL and LM) units, boards 5 and 6 are  
integrated into one board.  
3. Remove the three Circuit Board Assembly Screws (4).  
4. Push equally on the zero and span adjustment pot shafts from the  
terminal side. This will allow you to grasp and remove the Circuit Board  
Assembly.  
5. If troubleshooting is required, it is best to keep the Circuit Board  
Assembly together for initial evaluation. Otherwise, the board assembly  
may be disassembled by grasping the Output Board (5) around its  
circumference and pulling it gently and evenly away from the other two  
boards. Remove the Amplifier Board (6) in the same manner. Take care not  
to bend the interconnection pins.  
6. The adjustment pot shafts are sealed by two small O-rings (1D). Remove, if  
necessary, by taking off the O-ring Retainer Plate (1B), which is held in  
place by two small screws (1C).  
7. The Nameplate (1E), Hazardous Service Certification Label (9), and  
Instrument Tag (8) are held in place with Drive Screws (1F). Remove any  
of these by gripping the head of the Drive Screw with pliers and carefully  
turning counter-clockwise.  
4-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance and Troubleshooting  
Reassembly  
Procedure  
Explosions can cause death or serious injury. Both  
transmitter covers must be fully engaged to meet explosion-  
proof requirements.  
1. Inspect all O-rings (1D,3) and replace if necessary. Lightly grease new O-  
rings with silicone grease to guarantee an adequate seal.  
2. If the O-ring Retainer Plate (1B) has been removed, be sure the correct  
side is facing outward. The resistor symbol should be visible on RTD  
Transmitters, while a thermocouple symbol should be visible on  
Thermocouple or Millivolt transmitters.  
3. Inspect threaded connections on the housing and covers to make sure a  
minimum of five undamaged threads will be fully engaged. If the threads  
are shiny, apply a thin layer of molybdenum disulphide thread coating  
(such as Moly-Kote) to prevent galling of the aluminum threads.  
4. If the Range Board (7A) requires a Coarse Zero Jumper (7B) check to make  
sure it is in the correct location for the desired temperature range. See  
Table 3-1, on page 3-3 or Table 3-2, on page 3-6.  
5. Orient the Range Board (7A) and Amplifier Board (6) as shown in the  
Illustrated Parts List. Taking care not to bend the pins, plug the Amplifier  
Board into the Range Board. Press together until all three standoffs on the  
Amplifier Board rest against the Range Board.  
6. Orient the Output Board (5) so its standoffs line up with the standoffs of  
the Amplifier/Range Board combination. Carefully and evenly, plug the  
output board into the Amplifier/Range Board. Take care not to bend the  
pins. Press together until all three standoffs on the Output Board rest  
against the Amplifier Board.  
7. The circuit board assembly may be bench-tested, or calibrated outside the  
housing through the use of Test Terminal Strips (11 and 12). See the  
discussion of Calibration and Troubleshooting in this section.  
8. If the circuit board assembly has been calibrated outside the housing, be  
very careful to ensure the zero and span adjustment pots are not moved  
while inserting the circuit board assembly into the housing.  
9. Orient the circuit board assembly so the pot shafts line up with the pot  
holes in the housing.  
10. Insert the circuit board assembly firmly into the housing.  
11. Replace the three Circuit Board Assembly Screws (4).  
12. Replace the transmitter covers (2). Tighten the covers hand-tight.  
Interchangeability  
of Parts  
Use only the procedures and new parts specifically  
referenced in this manual. Unauthorized procedures or  
parts can affect product performance and the output signal  
used to control a process, and may render the instrument  
dangerous. Direct any questions concerning these  
procedures or parts to Rosemount Inc.  
4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Mechanical Parts  
• All mechanical hardware is interchangeable among units without regard  
to model numbers.  
• Nameplates are interchangeable only among units that share the same  
input types (i.e. RTD, Thermocouple, or Millivolt).  
Electrical Parts  
• Amplifier Board: Interchangeable among Models 444T and 444M.  
• Output Board: 444T series (TE, TJ, TK, TT, TR, TS) and 444MV share a  
common output board.  
• Range Board: Interchangeable among units of the same input code (e.g.,  
444RL1).  
The Model 444T series (TE, TJ, TK, TT, TR, TS) and Model 444MV have a  
resistor network that drives the output either upscale or downscale if an open  
occurs in the input circuit. This option is identified in the model number. To  
convert from upscale to downscale, disassemble the circuit board assembly and  
remove R3 (22 meg, ¼WCC resistor) from the range board. To convert from  
downscale to upscale, replace R3. If no burnout protection is desired (as in some  
instances where the source has a high input impedance), remove both R2 and  
R3. See Figure 3-3, on page 3-6.  
Burnout Protection  
Adjustments  
Model 444RL has a jumper on the range board to select burnout protection.  
Placing the jumper in the “U” position will cause the output to be driven upscale  
if the RTD opens. If the jumper is in the “D” position the output will be driven  
downscale. See Figure 3-1, on page 3-3. Models 444LL and 444LM have inherent  
upscale burnout protection that cannot be changed. Model 444RL also has a  
jumper on the range board to specify mA output at the transmitter test  
terminals. Setting the jumper at the “V” position produces a 40–200 mV output  
at the test terminals. Setting the jumper at the “A” position produces a  
4–20 mA output at the test terminals. See Figure 3-1, on page 3-4.  
Repair and warranty service is available through the Rosemount Regional  
Service Centers. Submit damage claims directly to the carrier.  
REPAIR AND  
WARRANTY  
SERVICE  
RETURN OF  
MATERIALS  
Exposure to hazardous substances can cause death or  
serious injury. If a hazardous substance is identified, a  
Material Safety Data Sheet (MSDS), required by law to be  
available to people exposed to specific hazardous  
substances, must be included with the returned materials.  
To expedite the return process, call the Rosemount North American Response  
Center toll-free at 800-654-RSMT (7768). This center, available 24 hours a day,  
will assist you with any needed information or materials.  
The center will ask for product model and serial numbers, and will provide a  
Return Material Authorization (RMA) number. The center will also ask for the  
name of the process material the product was last exposed to.  
The Rosemount North American Response Center will detail the additional  
information and procedures necessary to return goods exposed to hazardous  
substances.  
4-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
5
Specifications  
and Reference Data  
Inputs  
FUNCTIONAL  
Models 444RL, LL, and LM  
100 R0 platinum RTD per IEC 751.  
SPECIFICATIONS  
Model 444T  
Thermocouple types E,J,K,T,R, and S per NIST (grounded or  
ungrounded).  
Model 444MV  
Millivolt input (grounded or ungrounded) source impedance less than  
100 .  
R-numbers, specials  
Special inputs other than standards, consult factory.  
Spans  
RTD  
Platinum  
45 to 135 °F (25 to75 °C).  
125 to 380 °F (70 to 210 °C).  
360 to 1080 °F (200 to 600 °C).  
Copper  
180 to 540 °F (100 to 300 °C).  
45 to 360 °F (25 to 200 °C).  
Nickel  
Thermocouples  
Type J, K, E, T 180 to 540 °F (100 to 300 °C).  
Type J  
Type K, E  
Type K  
504 to 1458 °F (280 to 810 °C).  
504 to 1510 °F (280 to 840 °C).  
845 to 2540 °F (470 to 1410 °C).  
1467 to 3000 °F (815 to 1670 °C).  
Type R, S  
Millivolt  
5 to 15 mV.  
15 to 45 mV.  
Outputs  
Linear with temperature for RTD inputs.  
Linear with millivolt input signal for thermocouple or millivolt inputs;  
thermocouple and millivolt models input/output isolated to 500 V dc.  
Models 444RL, T, MV  
4–20 mA.  
Model 444LL  
0.8–3.2 V dc.  
Model 444LM  
1.0–5.0 V dc.  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Output Limits (approximate)  
Models 444RL, T, MV  
Low: 3.9 mA dc.  
High: 30.0 mA dc.  
Model 444LL  
Low: 0.1 V dc.  
High: 4.2 V dc.  
Model 444LM  
Low: 0.125 V dc.  
High: 6.2 V dc.  
Power Supply  
Models 444RL, T, and MV  
12 to 45 V dc at terminals of transmitter.  
Model 444LL  
5 to 12 V dc (overvoltage protected to 24 V dc)  
max current = 1.5 mA.  
Model 444LM  
8 to 12 V dc (overvoltage protected to 24 V dc)  
max current = 2.0 mA.  
Load Limits  
Models 444RL, T, and MV  
4–20 mA.  
4–20 mA dc  
1650  
1500  
Voltage  
Too Low  
1000  
500  
0
Operating  
Region  
12  
20  
30  
40  
Power Supply (V dc)  
Maximum Load = 50 ϫ (Supply Voltage – 12)  
Span and Zero  
Continuously adjustable, as defined in the ordering table. Adjustments are  
accessible from the terminal side of the transmitter housing.  
Transmitter Temperature Limits  
–13 to 185 °F (–25 °C to 85 °C), transmitter operates within specifications.  
–40 to 212 °F (–40 °C to 100 °C), transmitter operates without damage.  
–58 to 248 °F (–50 °C to 120 °C), storage.  
–13 to 149 °F (–25 °C to 65 °C), transmitter operates within specifications for  
meter option.  
Loss of Input  
Upscale burnout indication standard for RTD inputs, downscale burnout  
indication optional. Upscale burnout indication standard for thermocouple and  
millivolt inputs; downscale burnout indication or no indication optional.  
Turn-on Time  
2 seconds. No warm-up required.  
5-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
Accuracy  
PERFORMANCE  
SPECIFICATIONS  
0.2% of calibrated span (or, for thermocouple and millivolt inputs, 0.02  
millivolts, whichever is greater). 0.5% for copper, nickel, and isolated RTD  
inputs, 0.1% for differential RTD inputs. Includes combined effects of  
transmitter repeatability, hysteresis, linearity (conformity instead of linearity  
for thermocouple input), and adjustment resolution. Does not include sensor  
error.  
Stability  
0.2% of calibrated span for six months.  
Ambient Temperature Effect  
Errors for 50 °F (28 °C) change in ambient temperature.  
RTD Inputs  
Zero: 0.17 °C,  
plus  
Span: 0.22%,  
plus  
Elevation/Suppression: 0.083% of base temperature in °C.  
T/C Inputs (Includes Effect of Cold Junction)  
Zero: 1.38 °C,  
plus  
Span: 0.28% of span,  
plus  
Elevation/Suppression: 0.11% of base  
temperature in °C.  
Millivolt Inputs  
Zero: 0.038 mV,  
plus  
Span: 0.28% of span,  
plus  
Elevation/Suppression: 0.11% of base input in mV.  
Input Impedance (Thermocouple and mV Inputs)  
More than 1 megohm—burnout resistors disconnected.  
Power Supply Effect  
0.005% per volt.  
Load Effect  
No load effect other than the change in voltage supplied to the transmitter.  
Vibration Effect  
0.05% of span per g to 200 Hz in any axis for 3 g’s up to 33 Hz, 2 g’s from 33 to  
70 Hz and 1 g from 70 to 200 Hz.  
Mounting Position Effect  
None.  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Materials of Construction  
PHYSICAL  
SPECIFICATIONS  
Electronics Housing  
Low-copper aluminum. (NEMA 4X). IP 54, IP 65, IP 66.  
Housing Paint  
Polyurethane.  
Housing O-rings  
Buna-N.  
Sensor and Conduit Connections  
1/2-inch conduit on electronics housing. Screw terminals and integral test jacks  
compatible with miniature banana plugs (Pomona 2944, 3690 or equal).  
Weight  
Transmitter: 3 lb (1.4 kg).  
Transmitter with mounting bracket: 4 lb (1.8 kg).  
Hazardous Location Certifications  
Factory Mutual (FM) Approvals  
E5 Explosion Proof: Class I, Division 1, Groups B, C, and D. Dust Ignition  
Proof: Class II, Division 1, Groups E, F, and G; Class III, Division 1  
hazardous locations. Indoor and outdoor use. NEMA Enclosure Type 4X.  
Refer to Factory Mutual Explosion Proof Drawing 00444-0261.  
I5 Intrinsic Safety: Class I, Division 1, Groups A, B, C, and D; Class II,  
Division 1, Groups E, F, and G; Class III, Division 1 hazardous locations;  
Intrinsically safe system only when applying Table 5-1 entity parameters.  
Nonincendive: Class I, Division 2, Groups A, B, C, and D; Indoor and  
outdoor use. NEMA Enclosure Type 4X.  
Refer to Factory Mutual Intrinsic Safety Drawing 01151-0214.  
K5 Combination of E5 and I5.  
TABLE 5-1. FM Entity  
Parameters.  
FM Approved for  
Associated  
Equipment  
Parameters  
Class I, II, III,  
Division 1, Groups  
Model 444  
Parameters  
A
B
C thru G  
V
= 40 V  
= 165 mA  
V
I
C
or V 40 V  
MAX  
OC  
T
I
or I 165 mA  
MAX  
SC  
T
C = 0.044µF  
> 0.044µF  
> 0  
i
A
L = 0  
L
A
i
V
I
MAX = 40 V  
MAX = 225 mA  
C = 0.044µF  
V
I
C
or V 40 V  
NA  
NA  
OC  
T
or I 225 mA  
SC  
T
A > 0.044µF  
i
L = 0  
LA > 0  
i
Canadian Standards Association (CSA) Approvals  
E6 Explosion Proof: Class I, Division 1, Groups C and D; Dust-ignitionproof  
Class II, Division 1, Groups E, F, and G; Class III, Division 1 hazardous  
locations; Suitable for Class I, Division 2, Groups A, B, C, and D; CSA  
Enclosure Type 4X.  
I6 Intrinsic Safety: Class I, Division 1, Groups A, B, C, and D. Intrinsically  
safe system only when applying Table 5-2 parametric parameters.  
Temperature code T2D. CSA Enclosure Type 4X.  
Refer to CSA Intrinsic Safety Drawing 00444-0034.  
C6 Combination of E6 and I6.  
5-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
TABLE 5-2. CSA Parametric  
Parameters.  
CSA Approved for  
Class I,  
Barrier Manufacturer/Model  
Division 1 Groups  
A
B
C
D
Any CSA approved zener barrier  
30 V, 330 or  
28 V, 300 or  
22 V, 180 ⍀  
Foxboro Converters  
2AI-I2V-CGB  
NA  
2AI-I3V-CGB  
2AS-I2I-CGB  
2AS-I3I-CGB  
3AD-I3IA CS-E/CGB-A  
3A2-I2D CS-E/CGB-A,  
3A2-I3D CS-E/CGB-A  
3A4-I2DA CS-E/CGB-A  
3F4-I2DA1 CS-E/CGB-A  
Any CSA approved zener barrier  
NA  
NA  
30 V, 150⍀  
Standards Association of Australia (SAA) Certifications  
E7 Flameproof:  
Ex d IIB+H2 T6  
Class I, Zone 1.  
DIP T6  
Class II.  
Special Conditions for Safe Use (“X”):  
For transmitters having NPT or PG cable entry thread, an appropriate  
flameproof thread adaptor shall be used to facilitate application of certified  
flameproof cable glands. Only SAA-certified flameproof temperature  
sensors shall be used with the Model 444 Temperature Transmitter if  
fitted directly into the tapped entry of the enclosure.  
I7 Intrinsic Safety:  
Ex ia IIC T6 (Tamb = 40 °C)  
Ex ia IIC T5 (Tamb = 70 °C)  
Class I, Zone 0  
Entity Parameters:  
Ui = 30 V  
Ii = 200 mA  
Pi = 1.0 W  
Ci = 0.024 F  
Li = 0  
Refer to SAA Intrinsic Safety Drawing 00444-0264.  
Special Conditions for Safe Use (“X”):  
The equipment has been assessed to the “Entity” concept and upon  
installation the barrier/entity parameters must be taken into account.  
N7 Type “n”:  
Ex n IIC T6 (Tamb = 40 °C)  
Ex n IIC T5 (Tamb = 70 °C)  
Class I, Zone 2  
Special Conditions for Safe Use (“X”):  
The equipment must be connected to a supply voltage which does not  
exceed the rated voltage. The enclosure end caps must be correctly fitted  
while the equipment is powered.  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Centro Elettrotecnico Sperimentale Italiano (CESI/CENELEC) Certifications  
E8 Flameproof:  
EEx d IIC T6  
I8 Intrinsic Safety:  
Model 444RL  
EEx ia IIC T6 (Tamb = 40 °C) [Pi = 0.75 W]  
EEx ia IIC T5 (Tamb = 55 °C) [Pi = 1.0 W]  
EEx ia IIC T4 (Tamb = 80 °C) [Pi = 1.0 W]  
Model 444T_ & MV  
EEx ib IIB T6 (Tamb = 40 °C) [Pi = 0.75 W]  
EEx ib IIB T5 (Tamb = 55 °C) [Pi = 1.0 W]  
EEx ib IIB T4 (Tamb = 80 °C) [Pi = 1.0 W]  
Entity Parameters:  
Vi = 30 V dc  
Ii = 200 mA  
Pi = 0.75 W (T6)  
Pi = 1.0 W (T5)  
Pi = 1.0 W (T4)  
Ci = 0.024 F (Model 444RL)  
Ci = 0.006 F (Model 444T_ & MV)  
Li = 0.  
Special Conditions for Safe Use (“X”):  
If the temperature sensor connected to the input circuit does not tolerate  
an alternating tension of 500V for 60 seconds, the certified transducers  
must be powered by galvanically-insulated equipment. Model 444RL  
temperature transducers must be connected to associated electronic  
equipment certified to EN 50.014/EN 50.020 standards.  
British Approvals Service for Electrical Equipment Flammable Atmospheres  
(BASEEFA) Type N Certification  
N1 Ex N II T5  
Maximum Voltage: 35 V dc  
Maximum Current: 30 mA dc  
An appropriate stainless steel tag will be supplied according to the certification  
option selected.  
5-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
LCD METER  
SPECIFICATIONS  
(OPTIONS D AND E)  
Configuration  
Functional  
4 mA point limits: –999 to 1000.  
Span limits: 0200 to 9999.  
Specifications  
The sum of the 4 mA point and span must not exceed 9999. Adjustments are  
made using non-interactive zero and span buttons.  
Temperature Limits  
Storage: –40 to 85 °C (–40 to 185 °F).  
Operating: –20 to 70 °C (–4 to 158 °F).  
Between –40 to –20 °C (–40 to –4 °F) loop is intact and the meter is not damaged.  
Humidity Limitation  
0 to 95% non-condensing relative humidity.  
Update Period  
750 ms.  
Response Time  
Responds to changes in input within a maximum of two update periods. If the  
filter is activated, then the display responds to the change within nine update  
periods.  
Digital Display Resolution  
0.05% of calibrated range 1 digit.  
Analog Bar Graph Resolution  
0.05% of calibrated range.  
Performance  
Specifications  
Indication Accuracy  
0.25% of calibrated range 1 digit.  
Stability  
Over Time: 0.1% of calibrated range 1 digit per six months.  
Temperature Effect  
0.01% of calibrated range per °C on zero.  
0.02% of calibrated range per °C on span over the operating temperature range.  
Power Interrupt  
All calibration constants are stored in EEPROM memory and are not affected by  
power loss.  
Failure Mode  
LCD meter failure will not affect transmitter operation.  
Under/Over Range Indication  
Input current < 3.5 mA: Display blank.  
Input current > 22.0 mA: Display flashes 112.5% of full scale value or 9999,  
whichever is less.  
Meter Size  
Physical  
2¼-inch diameter face with four, ½-inch high characters.  
Specification  
5-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
ANALOG METER  
SPECIFICATIONS  
(OPTIONS B AND C)  
Meter Indication  
Functional  
0 to 100% linear scale.  
Special optional ranges.  
Temperature Limits  
Specifications  
–40 to 65 °C (–40 to 150 °F).  
Humidity Limits  
0 to 100% relative humidity.  
Zero Adjustment  
Adjustment screw on face of meter.  
Indication Accuracy  
2% of calibrated span.  
Performance  
Specifications  
Temperature Effect  
Less than 2% of full scale at any point within the temperature limits.  
Meter Size  
Physical  
2¼-inch diameter face with 2-inch long scale.  
Specification  
5-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
TABLE 5-3. 444 Model Structure.  
Model  
Product Description  
444  
Alphaline Temperature Transmitter  
Temperature Span  
Minimum Maximum  
45 °F (25 °C)  
125 °F (70 °C)  
360 °F (200 °C)  
Base Temperature  
(1)  
Code  
Input Type  
Upper Range Limit  
Minimum  
Maximum  
RL1  
RL2  
RL3  
Platinum RTD  
100 R0  
135 °F (75 °C)  
380 °F (210 °C)  
1080 °F (600 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
300 °F (150 °C)  
300 °F (150 °C)  
300 °F (150 °C)  
435 °F (225 °C)  
680 °F (360 °C)  
1380 °F (750 °C)  
linearized output  
RL9  
Special Input, Range or Accuracy (minimum span 3 )  
Thermocouple  
TJ1  
TJ2  
TK1  
TK2  
TK3  
TE1  
TE2  
TT1  
TR1  
TS1  
Type J  
Type J  
Type K  
Type K  
Type K  
Type E  
Type E  
Type T  
Type R  
Type S  
180 °F (100 °C)  
504 °F (280 °C)  
180 °F (100 °C)  
504 °F (280 °C)  
845 °F (470 °C)  
180 °F (100 °C)  
504 °F (280 °C)  
180 °F (100 °C)  
1467 °F (815 °C)  
1467 °F (815 °C)  
540 °F (300 °C)  
1458 °F (810 °C)  
540 °F (300 °C)  
1510 °F (840 °C)  
2540 °F (1410 °C)  
540 °F (300 °C)  
1510 °F (840 °C)  
540 °F (300 °C)  
3000 °F (1670 °C)  
3000 °F (1670 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
–58 °F (–50 °C)  
0 °F (–18 °C)  
300 °F (150 °C)  
930 °F (500 °C)  
300 °F (150 °C)  
930 °F (500 °C)  
930 °F (500 °C)  
300 °F (150 °C)  
930 °F (500 °C)  
300 °F (150 °C)  
1500 °F (815 °C)  
1500 °F (815 °C)  
840 °F (450 °C)  
1400 °F (760 °C)  
840 °F (450 °C)  
2440 °F (1340 °C)  
2500 °F (1370 °C)  
840 °F (450 °C)  
1830 °F (1000 °C)  
750 °F (400 °C)  
3200 °F (1760 °C)  
3200 °F (1760 °C)  
0 °F (–18 °C)  
T_9  
Special Range or Accuracy (minimum span 3 mV, maximum span 100 mV)  
MV1  
MV2  
Millivolt  
Millivolt  
5 mV  
15 mV  
15 mV  
45 mV  
–2 mV  
–2 mV  
8 mV  
20mV  
23 mV  
65 mV  
MV9  
Special Range or Accuracy (minimum span 3 mV, maximum span 100 mV)  
Code  
Loss of Input Indication  
U
D
N
Upscale (standard for all input types)  
Downscale  
None (not available for platinum RTD inputs)  
Code  
Calibration  
1
2
3
Trim to IEC 751 Class B (RTD) or NIST Curve (thermocouple)  
Trim to Specific Model 68/78/88 Calibration Schedule  
Trim to Other Nominal Curve (customer must specify separately) (Note: Millivolt input must use Code 3.)  
Code  
Meter Options  
A
B
C
D
E
None  
Integral Analog Meter, Special Scale (same as calibrated range)  
Integral Analog Meter, 0–100% Scale  
Integral LCD Meter, 0–100% Scale(2)  
Integral LCD Meter, Special Scale (specify range, mode, and engineering units)(1)  
Code  
Mounting Bracket  
1
2
None  
Mounting Bracket for 2-inch Pipe or Surface Mounting  
Code  
Hazardous Area Certifications  
NA  
E5  
I5  
E6  
I6  
E7  
I7  
N7  
E8  
I8  
No Certification Required  
FM Explosion-Proof Approval  
FM Intrinsic Safety and Non-incendive Approval  
CSA Explosion-Proof Approval  
CSA Intrinsic Safety Approval (444RL and 444T only)  
SAA Explosion-Proof Certification  
SAA Intrinsic Safety Certification  
SAA Non-incendive Certification  
CESI Explosion-Proof Certification (When ordering a transmitter with this option, place a W before the model number: W444.)  
CESI Intrinsic Safety Certification (When ordering a transmitter with this option, place a W before the model number: W444.)  
BASEEFA Non-incendive Certification (When ordering a transmitter with this option, place a T before the model number: T444.)  
N1  
Code  
Options  
Q4  
A1  
A2  
2-Point Calibration Certificate  
One (1) ½ NPT to M20 (CM 20) SST Thread Adapter  
One (1) ½ NPT to M20 (CM 20) SST Thread Adapter  
Code  
Special  
RXXXX  
Unique Range (use with RL9, T_9, and MV9 inputs)  
Typical Model Number:  
0444 RL3  
U
1
A
2
E5 Q4  
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information  
about Rosemount sensor assemblies.  
(2) LCD Meters are only available with RL1, RL2, or RL3. (may be reconfigured in the field.)  
5-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 5-4. 444RL Fast Turn-on Model Structure.  
Model  
Product Description  
444  
Alphaline Fast Turn-On Temperature Transmitter  
Temperature Span  
Base Temperature (4mA Point)  
Minimum Maximum  
Upper Range  
Limit  
(1)  
Code  
Input Type  
Minimum  
Maximum  
Platinum RTD  
100 R  
linearized output  
linearized output  
RL1  
RL2  
RL3  
45 °F (25 °C)  
125 °F (70 °C)  
360 °F (200 °C)  
135 °F (75 °C)  
380 °F (210 °C)  
1080 °F (600 °C)  
–58 °F (–50 °C) 300 °F (150 °C)  
–58 °F (–50 °C) 300 °F (150 °C)  
–58 °F (–50 °C) 300 °F (150 °C)  
435 °F (225 °C)  
680 °F (360 °C)  
1380 °F (750 °C)  
0
Code  
U
Loss of Input Indication  
Upscale (standard for all input types)  
Calibration  
Code  
1
2
3
Trim to IEC 751 Curve (RTD)  
Trim to Specific Model 68/78/88 Calibration Schedule  
Trim to Other Nominal Curve (customer must specify separately)  
Code  
Meter Options  
A
B
C
None  
Integral Analog Meter, Special Scale (must specify range, mode, and engineering units)  
Integral Analog Meter, 0–100% Scale  
Code  
Mounting Bracket  
1
2
None  
Mounting Bracket for 2-inch Pipe or Surface Mounting  
Code  
Hazardous Area Certifications  
NA  
E5  
E6  
I6  
No Certification Required  
FM Explosion-Proof Approval  
CSA Explosion-Proof Approval  
CSA Intrinsic Safety Approval  
Code  
Options  
Q4  
A1  
A2  
2-Point Calibration Certificate  
One (1) ½ NPT to M20 (CM 20) SST Thread Adapter  
One (1) ½ NPT to M20 (CM 20) SST Thread Adapter  
Code  
Special  
B0912  
Fast Turn-on Electronics  
Typical Model Number:  
0444 RL3  
U
1
A
2
NA B0912  
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information  
about Rosemount sensor assemblies.  
5-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
TABLE 5-5. 444LL, LM Model Structure.  
Model  
Product Description  
444  
Alphaline Low-Power Temperature Transmitter  
Temperature Span  
Base Temperature (4mA Point)  
(1)  
Code  
Input Type  
Minimum  
Maximum  
Minimum  
Maximum  
Platinum RTD  
LL1  
LM1  
100 R  
75 °F (42 °C)  
75 °F (42 °C)  
150 °F (83 °C)  
150 °F (83 °C)  
–25 °F (–32 °C) 50 °F (10 °C)  
–25 °F (–32 °C) 50 °F (10 °C)  
0
Linearized Output  
Loss of Input Indication  
Upscale  
Code  
U
Code  
Calibration  
1
2
Trim to IEC 751 Curve (RTD)  
Trim to Specific Model 68/78/88 Calibration Schedule  
Code  
A
Meter Options  
None  
Code  
Mounting Bracket  
1
2
None  
Mounting Bracket for 2-inch Pipe or Surface Mounting  
Code  
Options  
NA  
E5  
E6  
I6  
No Certification Required  
FM Explosion-Proof Approval  
CSA Explosion-Proof Approval  
CSA Intrinsic Safety Approval (444RL and 444T only)  
Typical Model Number:  
0444 LM1  
U
1
A
2
E5  
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information  
about Rosemount sensor assemblies.  
TABLE 5-6. Meter Kits.  
(1)  
Part Description  
Part Number  
Spares Category  
(2)  
40–200 mV Analog Meter Kit  
00444-0194-0001  
00444-0194-0005  
01151-0687-0003  
01151-0744-0003  
00444-0194-0004  
01151-1300-1000  
00444-0194-0002  
00444-0194-0003  
01151-2610-0001  
scale: XX to XX  
C
C
C
C
C
C
C
C
C
C
(2)  
40–200 mV Analog Meter Kit (CENELEC and I.S. approved)  
40–200 mV Analog Meter  
40–200 mV Analog Meter (CENELEC approved)  
(2) (3)  
4–20 mA LCD Meter Kit  
4–20 mA LCD Meter  
Mounting Hardware and Cover Kit  
Mounting Hardware Kit  
Cover Assembly  
(4)  
Meter with Special Scale  
(1) Normally, no spare parts required for C classification.  
(2) Meter kit includes meter, mounting hardware, and cover assembly.  
(3) LCD Meters are only available with RL1, RL2, and RL3 (may be reconfigured in the field).  
(4) For Special Scale Meters, order appropriate meter and indicate special scale desired. Mounting hardware and cover  
assembly must be ordered separately  
5-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 5-7. Standard Parts List.  
Item  
Number  
Spares  
Category  
Model  
444R  
444T  
444M  
Part Description  
Part Number  
(1)  
1
Housing Kit—RTD Input  
Housing Kit—Thermocouple Input  
Housing kit—Millivolt Input  
00444-0028-0001  
00444-0028-0002  
00444-0028-0003  
C
C
C
Electronics housing kit consists of:  
1 each electronics housing (item 1A )  
1 each O-ring retainer plate  
3
1B  
(item )  
2 each 6–32 ϫ /16 retainer plate screws (item 1C )  
2 each pot O-rings (item 1D )  
1 each nameplate (item 1E )  
3
1F  
2 each #4–40 ϫ /16 screws (item )  
2
3
1D  
4
3
5
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
Housing Cover—Quantity One  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
00444-0030-0001  
C
C
B
C
B
Housing Cover O-ring  
Adjustment Potentiometer O-ring (pkg. of 12)  
7
Electronics Assembly Screws, 6–32 ϫ1 /8 (pkg. of 12)  
Cover O-rings. (pkg. of 12)  
A/O Board—RTD Input (2 board set, Figure 5-1)  
Output Board—T/C, mV Input (3 board set, Figure 5-2)  
00444-0228-0001  
00444-0015-0001  
A
A
6
ALL  
Amplifier Board—All Inputs (3 board set, Figure 5-2)  
00444-0007-0003  
A
7A  
444R  
Range Board—RTD Inputs  
25 to 75 °C Span (Input Code RL1)  
00444-0221-0001  
00444-0221-0002  
00444-0221-0003  
00444-0003-XXXX  
A
A
A
A
70 to 210 °C Span (Input Code RL2)  
200 to 600 °C Span (Input Code RL3)  
Special Range RL9 (requires 3 board set – consult factory)  
7A  
444T  
Range Board—Thermocouple Inputs  
Type J, 100 to 300 °C Span (Input Code TJ1)  
Type J, 280 to 840 °C Span (Input Code TJ2)  
Type K, 100 to 300 °C Span (Input Code TK1)  
Type K, 280 to 840 °C Span (Input Code TK2)  
Type K, 470 to 1410 °C Span (Input Code TK3)  
00444-0262-0002  
00444-0262-0006  
00444-0262-0003  
00444-0262-0007  
00444-0262-0008  
00444-0262-0001  
00444-0262-0005  
00444-0262-0004  
00444-0262-0009  
00444-0262-0010  
00444-0013-XXXX  
A
A
A
A
A
A
A
A
A
A
A
(2)  
Type E, 100 to 300 °C Span (Input Code TE1)  
Type E, 280 to 840 °C Span (Input Code TE2)  
Type T, 100 to 300 °C Span (Input Code TT1)  
Type R, 815 to 1670 °C Span (Input Code TR1)  
Type S, 815 to 1670 °C Span (Input Code TS1)  
Special Range T-9 Kit (consult factory)  
7A  
444M  
Range Board—Millivolt Inputs  
5 to 15 mV (Input Code MV1)  
15 to 45 °C Span (Input Code MV2)  
Special Range MV9 (consult factory)  
00444-0023-0001  
00444-0023-0002  
00444-0023-XXXX  
A
A
A
7B  
8
9
444R  
ALL  
ALL  
ALL  
Range Board Zero Jumpers (pkg. of 12)  
Instrument Tag (optional)  
Certification Label  
00444-0036-0001  
01151-0148  
A
C
C
C
10  
Mounting Bracket Kit  
00444-0022-0001  
10A  
1 each mounting bracket (item  
)
10B  
4 each ¼–20 ϫ ½ bolts(item  
)
10C  
2 each U-bolts (item  
)
10D  
4 each ¼ washers (item  
)
5
10E  
4 each /16 –18 nuts (item  
)
2 each washer plates (item 10F )  
11  
ALL  
Terminal Block, Three-position  
Terminal Block, Four-position  
C10448-0106  
C10448-0109  
D
D
12  
(1) Spares Categories (2) Thermocouple Range Board Kit Consists of:  
Code A – Recommended 1 spare part per 25 transmitters. 1 each T/C Range Board.  
Code B – Recommended 1 spare part per 50 transmitters. 1 each External CJC Block.  
Code C – None normally required.  
Code D – Not used on transmitters, but recommended for repair actiivities.  
5-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
TABLE 5-8. Parts List for Model W444 Transmitter (CESI/CENELEC Approval).  
Item  
(1)  
Part Description  
Number  
Part No.  
Spares Category  
(2)  
1
Housing Kit — RTD Input  
00444-0028-0001  
00444-0028-0002  
00444-0028-0003  
C
C
C
(2)  
Housing Kit — Thermocouple Input  
(2)  
Housing kit — Millivolt Input  
2
3
1D  
4
Housing Cover—Quantity One  
Housing Cover O-rings  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
C
B
B
C
Adjustment Potentiometer O-rings  
Electronics Assembly Screws  
5
A/O Board—RTD Input  
Output Board—T/C, mV Input  
Output Board—RD Input  
00444-0228-0001  
00444-0015-0002  
00444-0111-0001  
A
A
A
6
Amplifier Board  
00444-0007-0003  
A
7A  
Range Board  
W444 RL1  
00444-0221-0001  
00444-0221-0002  
00444-0221-0003  
00444-0003-XXXX  
00444-0262-0002  
00444-0262-0006  
00444-0262-0003  
00444-0262-0007  
00444-0262-0008  
00444-0262-0001  
00444-0262-0005  
00444-0262-0004  
00444-0262-0009  
00444-0262-0010  
00444-0013-XXXX  
00444-0023-0001  
00444-0023-0002  
00444-0023-XXXX  
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
W444 RL2  
W444 RL3  
W444 RL9 (consult factory)  
W444 TJ1  
W444 TJ2  
W444 TK1  
W444 TK2  
W444 TK3  
W444 TE1  
W444 TE2  
W444 TT1  
W444 TR1  
W444 TS1  
W444 T_9 (consult factory)  
W444 MV1  
W444 MV2  
W444 MV9 (consult factory)  
7B  
Range Board Zero Jumpers  
Mounting Bracket Kit  
00444-0036-0001  
00444-0022-0001  
A
C
10  
11  
Terminal Block, Three-position  
Terminal Block, Four-position  
C10448-0106  
C10448-0109  
D
D
12  
(1) Spares Categories  
Code A – Recommended one spare part per 25 transmitters.  
Code B – Recommended one spare part per 50 transmitters.  
Code C – None normally required.Code D – Not used on transmitters, but recommended for repair activities (see Instruction  
Manual).  
(2) Housing kit includes housing, adjustment potentionmeter O-rings, O-ring retainer plate and nameplate.  
5-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 5-9. Parts List for Model T444 Transmitter (BASEEFA Approval).  
Item  
(1)  
Part Description  
Number  
Part No.  
Spares Category  
(2)  
1
Housing Kit —RTD Input  
00444-0028-0001  
00444-0028-0002  
00444-0028-0003  
C
C
C
(2)  
Housing Kit —Thermocouple Input  
(2)  
Housing kit —Millivolt Input  
2
3
1D  
4
Housing Cover—Quantity One  
Housing Cover O-rings  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
C
B
B
C
A
Adjustment Potentiometer O-rings  
Electronics Assembly Screws  
5
A/O Board—RTD Input  
Output Board—T/C, mV Input  
00444-0228-0001  
00444-0015-0001  
6
Amplifier Board  
00444-0007-0003  
A
7A  
Range Boards  
T444 RL1  
00444-0136-0001  
00444-0136-0002  
00444-0136-0003  
00444-0136-XXXX  
00444-0273-0001  
00444-0273-0002  
00444-0273-0003  
00444-0273-0004  
00444-0273-0005  
00444-0273-0006  
00444-0273-0007  
00444-0273-0008  
00444-0273-0009  
00444-0273-0010  
00444-0138-XXXX  
00444-0139-0001  
00444-0139-0002  
00444-0139-XXXX  
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
T444 RL2  
T444 RL3  
T444 RL9 (consult factory)  
T444 TE1  
T444 TJ1  
T444 TK1  
T444 TT1  
T444 TE2  
T444 TJ2  
T444 TK2  
T444 TK3  
T444 TR1  
T444 TS1  
T444 T_9 (consult factory)  
T444 MV1  
T444 MV2  
T444 MV9 (consult factory)  
7B  
Range Board Zero Jumpers  
Mounting Bracket Kit  
00444-0036-0001  
00444-0022-0001  
A
C
10  
11  
Terminal Block, Three-position  
Terminal Block, Four-position  
C10448-0106  
C10448-0109  
D
D
12  
For meter information, see page 7.  
(1) Spares Categories  
Code A – Recommended one spare part per 25 transmitters.  
Code B – Recommended one spare part per 50 transmitters.  
Code C – None normally required.  
Code D – Not used on transmitters, but recommended for repair activities (see Instruction Manual).  
(2) Housing kit includes housing, adjustment potentiometer O-rings, O-ring retainer plate and nameplate.  
5-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
TABLE 5-10. Parts List for Model 444 RL Fast Turn On Transmitter (RTD Input).  
Item  
(1)  
Part Description  
Number  
Part No.  
Spares Category  
(2)  
1
Housing Kit  
00444-0028-0001  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
00444-0126-0001  
C
C
B
B
C
A
2
3
1D  
4
Housing Cover—Quantity One  
Housing Cover O-rings  
Adjustment Potentiometer O-rings  
Electronics Assembly Screws  
Output Board 444 RL B0912  
5
7A  
Range Boards  
RL1 B0912  
RL2 B0912  
RL3 B0912  
00444-0221-0011  
00444-0221-0012  
00444-0221-0013  
A
A
A
7B  
Range Board Zero Jumpers  
Mounting Bracket Kit  
00444-0036-0001  
00444-0022-0001  
A
C
10  
11  
Terminal Block, Three-position  
Terminal Block, Four-position  
C104480106  
C104480109  
D
D
12  
For meter information, see page 7.  
(1) Spares Categories  
Code A – Recommended one spare part per 25 transmitters.  
Code B – Recommended one spare part per 50 transmitters.  
Code C – None normally required.  
Code D – Not used on transmitters, but recommended for repair activities (see Instruction Manual).  
(2) Housing kit includes housing, adjustment potentiometer O-rings, O-ring retainer plate and nameplate.  
TABLE 5-11. Parts List for Model 444 LL, 444 LM Low Power Transmitter (RTD Input).  
Item  
Number  
(1)  
Part Description  
Part No.  
Spares Category  
(2)  
1
Housing Kit  
00444-0028-0005  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
C
C
B
B
C
2
3
1D  
4
Housing Cover—Quantity One  
Housing Cover O-rings  
Adjustment Potentiometer O-rings  
Electronics Assembly Screws  
5
Output Board 444 LL  
Output Board 444 LM  
00444-0159-0001  
00444-0159-0002  
A
A
7A  
Range Boards  
444 LL  
00444-0157-0001  
00444-0157-0002  
A
A
444 LM  
7B  
Range Board Zero Jumpers  
Mounting Bracket Kit  
00444-0036-0001  
00444-0022-0001  
C
C
10  
11  
Terminal Block, Three-position  
Terminal Block, Four-position  
C104480106  
C104480109  
D
D
12  
No meter available.  
(1) Spares Categories  
Code A – Recommended one spare part per 25 transmitters.  
Code B – Recommended one spare part per 50 transmitters.  
Code C – None normally required.  
Code D – Not used on transmitters, but recommended for repair activities (see Instruction Manual).  
(2) Housing kit includes housing, adjustment potentiometer O-rings, O-ring retainer plate and nameplate.  
5-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
TABLE 5-12. Parts List for Model 444 Increased Accuracy or Special Range Transmitters.  
Item  
Number  
(1)  
Part Description  
Part No.  
Spares Category  
(2)  
1
Housing Kit —RTD Input  
00444-0028-0001  
00444-0028-0002  
00444-0028-0003  
C
C
C
(2)  
Housing Kit —Thermocouple Input  
(2)  
Housing kit —Millivolt Input  
2
3
1D  
4
Housing Cover—Quantity One  
Housing Cover O-rings  
90032-0240-0003  
01151-0033-0003  
00444-0030-0001  
00444-0031-0001  
C
B
B
C
Adjustment Potentiometer O-rings  
Electronics Assembly Screws  
5
Output Board RL9  
Output Board T_9  
Output Board MV9  
00444-0005-0001  
00444-0015-0001  
00444-0015-0001  
A
A
A
6
Amplifier Board  
00444-0007-0003  
A
7A  
Range Boards  
RL9 (consult factory)  
T_9 (consult factory)  
MV9 (consult factory)  
00444-0003-XXXX  
00444-0013-XXXX  
00444-0023-XXXX  
A
A
A
7B  
Range Board Zero Jumpers  
Mounting Bracket Kit  
00444-0036-0001  
00444-0022-0001  
C
C
10  
11  
Terminal Block, Three-position  
Terminal Block, Four-position  
C10448-0106  
C10448-0109  
D
D
12  
For meter information, see below.  
(1) Spares Categories  
Code A – Recommended one spare part per 25 transmitters.  
Code B – Recommended one spare part per 50 transmitters.  
Code C – None normally required.  
Code D – Not used on transmitters, but recommended for repair activities.  
(2) Housing kit includes housing, adjustment potentiometer O-rings, O-ring retainer plate and nameplate.  
5-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Reference Data  
FIGURE 5-1. Transmitter Exploded View 1.  
View Corresponds to  
the Following Models:  
2
444RL1  
444RL2  
444RL3  
444LM1  
444LL1  
3
4
5
8
7A  
9
1A  
7B  
1D  
1B  
3
1C  
2
1F  
1E  
10E  
10F  
10A  
10D  
10B  
10C  
Detail of Item 1B. O-Ring Retainer Plate Wiring Diagrams  
Reversible  
Accessory Terminal Blocks For  
Bench Testing  
Thermocouple  
or Millivolt  
Side  
Models 444T, MV  
RTD Side  
Models 444RL, LL, LM  
5-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
FIGURE 5-2. Transmitter Exploded View 2.  
View Corresponds to  
the Following Models:  
2
444RL9  
444TJ1  
444TJ2  
444TK1  
444TK2  
444TK3 444TR1  
3
444TE1 444TS1  
444TE2 444MV1  
444TT1 444MV2  
444T_9 444MV9  
4
5
6
8
9
7A  
1A  
7B  
1D  
1B  
3
1C  
2
1F  
1E  
10E  
10F  
10A  
10D  
10B  
Detail of Remote Cold Junction  
Compensation Sensor. (Model 444T)  
10C  
5-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Section  
6
Options  
The mounting bracket option provides auxiliary configurations for  
mounting the Model 444 transmitter. With this option you can either  
mount the transmitter to a 2-inch pipe or a suitable flat panel. The  
bracket is constructed of carbon steel with carbon steel U-bolts. See  
Figure 6-1 for an exploded view of the Model 444 with a mounting  
bracket in both configurations.  
MOUNTING BRACKET  
FIGURE 6-1. Model 444 with  
Optional Mounting Bracket.  
PANEL OR SURFACE MOUNTING  
PIPESTAND MOUNTING  
Transmitter can be Rotated 90°  
5
/16 -inch Bolts  
(four required,  
not furnished)  
Hole for  
/16 -inch  
Bolts  
5
2.81 (81)  
(four  
Mounting Bracket  
places)  
¼–20 ϫ½-inch  
Bolt (4)  
Clearance Hole  
for ¼-inch Bolt  
(eight places)  
5
/16 –18 U-bolt for  
2-inch Pipe (2)  
5.00 (127)  
NOTE  
Dimensions are in inches (millimeters).  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
LCD / ANALOG  
METER  
The LCD and analog meters provide local indication of the transmitter output.  
Both meters attach easily to the terminal side of the transmitter.  
The Rosemount® LCD meter plugs directly into the Model 444 to provide a  
highly accurate digital display of the process variable. The meter adds no voltage  
drop in the 4–20 mA current loop when connected directly across the transmitter  
test terminals.  
LCD METER  
Configure the LCD meter to meet specific requirements by using the left and  
right calibration buttons located on the meter face as shown in Figure 6-2. The  
analog bar graph is also shown in Figure 6-2. The 20-segment bar graph is  
factory calibrated and represents  
Configuration  
4–20 mA directly.  
.
FIGURE 6-2. LCD Meter.  
Analog  
Bar Graph  
Retaining  
Ring  
Left Configuration  
Button  
Right Configuration  
Button  
No calibration equipment is required to configure the LCD meter, but between 4  
and 20 mA must exist in the loop in order for the meter to operate. The actual  
value of the current is not significant. In addition, meter configuration does not  
affect the transmitter/loop current. Use the following meter configuration  
procedure to properly configure the LCD meter.  
Remove the Cover  
1. Unscrew the retaining ring shown in Figure 6-2 and lift the transparent  
cover off of the housing.  
NOTE  
The LCD meter time-out is approximately 16 seconds. If keys are not pressed  
within this period, the indicator reverts to reading the current signal.  
6-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Options  
Position the Decimal  
Point and Select the  
Meter Function  
2. Press the left and right configuration buttons simultaneously and release  
them immediately.  
3. To move the decimal point to the desired location, press the left  
configuration button. Note that the decimal point wraps around.  
4. To scroll through the mode options, press the right configuration button  
repeatedly until the desired mode is displayed. See Table 6-1.  
TABLE 6-1. LCD Meter Modes.  
Options  
Relationship between Input Signal and Digital Display  
L in  
Linear  
L inF  
Srt  
Linear with five-second filter  
Square root  
SrtF  
Square root with five-second filter  
Square root function only relates to the digital display.  
The bar graph output remains linear with the current signal.  
Square root response  
The digital display will be proportional to the square root of the input current where 4 mA=0  
and 20 mA=1.0, scaled per the calibration procedure. The transition point from linear to  
square root is at 25% of full scale flow.  
Filter response operates upon “present input” and “input received in the previous five second  
interval” in the following manner:  
Display = (0.75 ϫ previous input) + (0.25 ϫ present input)  
This relationship is maintained provided that the previous reading minus the present reading is  
less than 25% of full scale.  
Store the Information  
5. Press both configuration buttons simultaneously for two seconds. Note  
that the meter displays “----” for approximately 7.5 seconds while the  
information is being stored.  
Set the Display  
Equivalent  
to a 4 mA Signal  
6. Press the left button for two seconds.  
7. To decrement the display numbers, press the left configuration button and  
to increment the numbers, press the right configuration button. Set the  
numbers between –999 and 1000.  
8. To store the information, press both configuration buttons simultaneously  
for two seconds.  
Set the Display  
Equivalent  
to a 20 mA Signal  
9. Press the right button for two seconds.  
10. To decrement the display numbers, press the left configuration button on  
the display and to increment the numbers, press the right configuration  
button. Set the numbers between –999 and 9999. The sum of the 4 mA  
point and the span must not exceed 9999.  
11. To store the information, press both configuration buttons simultaneously  
for two seconds. The LCD meter is now configured.  
Replace the Cover  
12. Make sure the rubber gasket is seated properly, replace the transparent  
cover, and replace the retaining ring.  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
Figure 6-3 shows the mounting hardware required to properly install the LCD  
meter on a Model 444 transmitter. This mounting hardware may also be used  
with the Rosemount universal (analog) meter.  
LCD Meter  
Assembly  
FIGURE 6-3. LCD Meter  
Exploded View  
Mounting Screw into Housing  
Strap Washer  
Retaining Straps  
Mounting Screws  
Mounting Screw into  
Mounting Plate  
Terminal Screws (Mount  
into Transmitter “Test”  
Terminal Block)  
Mounting Plate  
Meter (Meter may be rotated  
in 90 degree increments)  
Cover Bushing  
Cover Foam Spacer  
LCD Meter  
Specifications  
Functional  
Input Signal  
Specifications  
4–20 mA dc.  
Meter Indication  
4-digit LCD showing –999 to 9999. A 20-segment bar graph directly represents  
the 4–20 mA current.  
Scaling/Calibration  
4 mA Point Limits: –999 to 1000.  
Span limits: 0200 to 9999.  
The sum of the 4 mA point and span must not exceed 9999.  
Adjustments are made using non-interactive zero and span buttons.  
Hazardous Locations Certifications  
Approved for use with Rosemount Models 444, 751, 1135, 1144,  
and 1151.  
6-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Options  
Overload Limitation  
666 mA.  
Temperature Limits  
Storage: –40 to 85 °C (–40 to 185 °F).  
Operating: –20 to 70 °C (–4 to 158 °F).  
Between temperatures –40 to –20 °C (–40 to –4 °F), the loop is intact and the  
meter is not damaged.  
Humidity Limitation  
0 to 95% non-condensing relative humidity.  
Update Period  
750 ms.  
Response Time  
Responds to changes in input within a maximum of two update periods. If the  
filter is activated, then the display responds to the change within nine update  
periods.  
Performance  
Digital Display Resolution  
Specifications  
0.05% of calibrated range 1 digit.  
Analog Bar Graph Resolution  
0.05% of calibrated range.  
Indication Accuracy  
0.25% of calibrated range 1 digit.  
Stability  
Over Time: 0.1% of calibrated range 1 digit per six months.  
Temperature Effect  
0.01% of calibrated range per °C on zero.  
0.02% of calibrated range per °C on span over the operating temperature range.  
Power Interrupt  
All calibration constants are stored in EEPROM memory and are not affected by  
power loss.  
Failure Mode  
LCD meter failure will not affect transmitter operation.  
Under/Over Range Indication  
Input current < 3.5 mA: Display blank.  
Input current > 22.0 mA: Display flashes 112.5% of full scale value or 9999,  
whichever is less.  
Physical Specification  
Meter Size  
2¼-inch diameter face with ½-inch high characters.  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
The analog meter plugs directly into the Model 444 to provide an accurate local  
ANALOG METER  
indication of user-specified units. It requires an analog 4–20 mA dc, 10–50 mA  
dc, or 40–200 mV dc transmitter output from a two-wire transmitter, and adds  
no voltage drop in the 4–20 mA current loop when connected directly across the  
transmitter test terminals.  
The large 2¼-inch diameter meter face has a two-inch long scale for easy  
readability, as shown in Figure 6-4. A meter-zero adjustment is located on the  
meter faceplate. You can rotate the meter in 90-degree increments within the  
transmitter housing for convenient viewing.  
FIGURE 6-4. Analog Meter  
Face.  
ÿþýüþûú  
Analog Meter  
Specifications  
Functional  
Input Signal  
Specifications  
4–20 mA dc.....  
10–50 mA dc.  
.
Maximum series resistance is  
}
10 ohms for milliameters.  
40–200 mV  
Meter Indication  
0 to 100% linear scale.  
Special optional ranges.  
Overload Limit  
150% of rated end scale value for two minutes.  
Temperature Limits  
–40 to 65 °C (–40 to 150 °F).  
Humidity Limit  
0 to 100% relative humidity.  
Zero Adjustment  
Adjustment screw on face of meter.  
Performance  
Indication Accuracy  
Specifications  
2% of calibrated span.  
Temperature Effect  
Less than 2% of full scale at any point within the temperature limits.  
Physical Specification  
Meter Size  
2¼-inch diameter face with 2-inch long scale.  
6-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
A
Hazardous Locations  
Certifications Drawings  
Rosemount Drawing 00444-0261, 2 Sheets:  
Model 444 Explosion-proof Installation Drawing, Factory Mutual.  
Rosemount Drawing 01151-0214, 6 Sheets:  
Index of Intrinsically Safe Barrier Systems and Entity parameters for  
444, 1135, 1144, 1151, and 2051 Transmitters and 751 Field Indicators.  
Rosemount Drawing 00444-0034, 2 Sheets:  
CSA Intrinsic Safety Approvals for Model 444.  
Rosemount Drawing 00444-0264, 1 Sheet:  
Model 444 SAA Intrinisc Safety Configuration  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
A-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
A-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
B
Temperature Sensor  
Reference Information  
TABLE B-1. Resistance Versus Temperature Curve R =100.00 , =0.00385 Reference Standard IEC 751.  
o
°C  
°C  
0
–10  
–20  
–30  
–40  
–50  
–60  
–70  
–80  
–90  
–200  
–100  
0
18.49  
60.25  
100  
56.19  
96.09  
52.11  
92.16  
48.00  
88.22  
43.87  
84.27  
39.71  
80.31  
35.53  
76.33  
31.32  
72.33  
27.08  
68.33  
22.80  
64.30  
°C  
0
10  
20  
30  
40  
50  
60  
70  
80  
90  
0
100  
103.90  
142.29  
179.51  
215.57  
250.48  
284.22  
316.80  
348.22  
378.48  
107.79  
146.06  
183.17  
219.12  
253.90  
287.53  
319.99  
351.30  
381.45  
111.67  
149.82  
186.82  
222.65  
257.32  
290.83  
323.18  
354.37  
384.40  
115.54  
153.58  
190.45  
226.17  
260.72  
294.11  
326.35  
357.42  
387.33  
119.40  
157.31  
194.07  
229.67  
264.11  
297.39  
329.51  
360.47  
123.24  
161.04  
197.69  
233.17  
267.49  
300.65  
332.66  
363.50  
127.07  
164.76  
201.29  
236.65  
270.86  
303.91  
335.79  
366.52  
130.89  
168.46  
204.88  
240.13  
274.22  
307.15  
338.92  
369.53  
134.70  
172.16  
208.45  
243.59  
277.56  
310.38  
342.03  
372.52  
100  
200  
300  
400  
500  
600  
700  
800  
138.50  
175.84  
212.02  
247.04  
280.90  
313.59  
345.13  
375.51  
°F  
°F  
0
–10  
–20  
–30  
–40  
–50  
–60  
–70  
–80  
–90  
–300  
–200  
–100  
0
25.18  
48.46  
71.00  
93.03  
22.80  
46.17  
68.77  
90.85  
20.41  
43.87  
66.54  
88.66  
18.01  
41.56  
64.30  
86.47  
39.25  
62.05  
84.27  
36.93  
59.80  
82.07  
34.60  
57.55  
79.87  
32.26  
55.28  
77.66  
29.91  
53.02  
75.44  
27.55  
50.74  
73.22  
°F  
0
10  
20  
30  
40  
50  
60  
70  
80  
90  
0
100  
200  
300  
400  
500  
600  
700  
93.03  
114.68  
135.97  
156.90  
177.47  
197.69  
217.54  
237.04  
256.18  
274.96  
293.38  
311.45  
329.16  
346.51  
363.50  
380.13  
95.21  
116.83  
138.08  
158.97  
179.51  
199.69  
219.51  
238.97  
258.07  
276.82  
295.21  
313.24  
330.91  
348.22  
365.18  
381.77  
97.39  
118.97  
140.18  
161.04  
181.54  
201.69  
221.47  
240.90  
259.96  
278.67  
297.03  
315.02  
332.66  
349.93  
366.85  
383.41  
99.57  
121.11  
142.29  
163.11  
183.57  
203.68  
223.43  
242.82  
261.85  
280.53  
298.84  
316.80  
334.40  
351.64  
368.52  
385.05  
101.74  
123.24  
144.38  
165.17  
185.60  
205.67  
225.38  
244.74  
263.74  
282.37  
300.65  
318.58  
336.14  
353.35  
370.19  
386.68  
103.90  
125.37  
146.48  
167.23  
187.62  
207.66  
227.34  
246.65  
265.62  
284.22  
302.46  
320.35  
337.88  
355.05  
371.86  
388.31  
106.07  
127.50  
148.57  
169.29  
189.64  
209.64  
229.28  
248.57  
267.49  
286.06  
304.27  
322.12  
339.61  
356.74  
373.52  
389.94  
108.22  
129.62  
150.66  
171.34  
191.66  
211.62  
231.23  
250.48  
269.36  
287.90  
306.07  
323.88  
341.34  
358.44  
375.18  
110.38  
131.74  
152.74  
173.39  
193.67  
213.60  
233.17  
252.38  
271.23  
289.73  
307.87  
325.64  
343.07  
360.13  
376.18  
112.53  
133.86  
154.82  
175.43  
195.68  
215.57  
235.11  
254.28  
273.10  
291.56  
309.66  
327.40  
344.79  
361.81  
378.48  
800  
900  
1000  
1100  
1200  
1300  
1400  
1500  
B-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 244P Head and Rail Mount Temperature Transmitters  
B-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
A
H
M
Accuracy 5-3  
Ambient Temperature Effect 5-3  
Analog Meter Specifications 5-8  
Hazardous Location  
Certifications 5-4  
Hazardous Location Installation 2-10  
Humidity Limitation 5-7  
Maintenance 3-1, 4-1  
Burnout Protection  
Adjustments 4-6  
Disassembly Procedure 4-4  
Interchangeability of Parts 4-6  
Reassembly Procedure 4-5  
Repair 4-3  
B
I
Burnout Protection Adjustments 4-6  
Input Impedance 5-3  
Inputs 5-1  
Installation 2-1  
Warranty Service 4-7  
C
Materials of Construction 5-4  
Mechanical Installation 2-11  
Millivolt Transmitter 3-10  
Moist or Corrosive Atmospheres 2-10  
Mounting 2-2  
Mounting Bracket 2-2, 2-13, 6-1  
Mounting Position Effect 5-3  
Multi-Channel Installations 2-7  
Calibration  
Electrical Considerations  
Low-Power Transmitter 3-8  
Millivolt Transmitter 3-10  
RTD Transmitter 3-2  
Thermocouple Transmitter 3-5  
Configuration 5-7  
Intrinsically Safe  
Installations 2-10  
Multi-Channel  
Installations 2-7  
Power Supply 2-4, 2-8  
Mechanical Considerations 2-1  
D
Hazardous Location  
Installation 2-10  
Moist or Corrosive  
Atmospheres 2-10  
Mounting 2-2  
Temperature  
Environment 2-8  
Digital Display Resolution 5-7  
Dimensional Drawings  
Sensor Assembly 2-9  
Transmitter 2-13  
Disassembly Procedure 4-4  
O
Ordering Tables 5-9–5-11  
Output Limits 5-2  
Outputs 5-1  
P
Installation Procedure 2-11  
Electrical 2-12  
E
Power Supply 2-4, 2-8, 5-2  
Power Supply Effect 5-3  
Electrical Installation 2-12  
Final Checkout 2-12  
Input Connections 2-12  
Output Connections 2-12  
Preliminary Checkout 2-12  
Exploded View  
Mechanical 2-11  
Interchangeability of Parts 4-6  
Intrinsically Safe Installations 2-10  
Introduction 1-1  
R
Reassembly Procedure 4-5  
Repair 4-3  
L
Burnout Protection  
Adjustments 4-6  
Disassembly Procedure 4-4  
Interchangeability of Parts 4-6  
Reassembly Procedure 4-5  
Warranty Service 4-7  
LCD Meter 6-4  
Transmitter 5-17, 5-18  
Transmitter with Mounting  
Bracket 2-13, 6-1  
LCD Meter  
Assembly 6-4  
Specifications 6-4  
LCD Meter Specifications 5-7  
Load Effect 5-3  
Load Limitations 5-2  
Loss of Input 5-2  
G
Response Time 5-7  
RTD Inputs 2-5  
Remote RTD Mounting 2-5  
RTD Transmitter 3-2  
Grounding 2-7  
Current Signal Loop 2-7  
RTD Circuit 2-7  
Shielded Wire 2-7  
Thermocouple and Millivolt  
Transmitters 2-7  
Low-Power Transmitter 3-8  
Calibration Equipment  
Required 3-8  
Calibration Procedure 3-9  
Calibration Equipment  
Required 3-2  
Calibration Procedure 3-2  
I-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Model 444 Alphaline Temperature Transmitters  
S
T
Sensor and Conduit Connections 5-4  
Sensor Connections  
Grounding 2-7  
Temperature Environment 2-8  
Temperature Limits 5-2, 5-7  
Testing  
RTD Inputs 2-5  
RTD Test 4-3  
2-Wire RTD 2-6  
3-Wire RTD 2-6  
4-Wire RTD 2-6  
Thermocouple Test 4-3  
Thermocouple or Millivolt Inputs 2-7  
Thermocouple Transmitter 3-5  
Compensation Loop  
RTD 2-6  
Calibration Procedure for a  
Compensated Thermocouple  
Differential Millivolt 2-6  
Differential RTD 2-5  
Thermocouple or Millivolt  
Inputs 2-7  
Calibration Equipment  
Required 3-5  
Calibration Procedure 3-5  
Calibration Procedure  
for Ice Bath  
Calibration Equipment  
Required 3-7  
Calibration Procedure 3-7  
Troubleshooting 4-2  
Span and Zero 5-2  
Spans 5-1  
Spare Parts, List of 5-12–5-16  
Specifications and Reference  
Data 5-1  
Turn-on Time 5-2  
Analog Meter Specifications 5-8  
LCD Meter Specifications 5-7  
Configuration 5-7  
U
Digital Display  
Resolution 5-7  
Update Period 5-7  
Humidity Limitation 5-7  
Response Time 5-7  
Temperature Limits 5-7  
Update Period 5-7  
V
Vibration Effect 5-3  
W
Transmitter Specifications 5-1,  
5-2  
Warranty Service 4-7  
Weight 5-4  
Accuracy 5-3  
Wiring 2-4  
Ambient Temperature  
Effect 5-3  
Hazardous Location  
Certifications 5-4  
Input Impedance 5-3  
Inputs 5-1  
Load Effect 5-3  
Materials of  
Construction 5-4  
Mounting Position  
Effect 5-3  
Power Supply Effect 5-3  
Sensor and Conduit  
Connections 5-4  
Stability 5-3  
Turn-on Time 5-2  
Vibration Effect 5-3  
Weight 5-4  
Stability 5-3  
I-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
I-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rosemount Inc.  
Fisher-Rosemount Limited  
Heath Place  
Bognor Regis  
West Sussex PO22 9SH  
England  
Tel 44 (1243) 863 121  
Fax 44 (1243) 867 5541  
Fisher-Rosemount  
Singapore Pte Ltd.  
1 Pandan Crescent  
Singapore 128461  
Tel (65) 777-8211  
Fax (65) 777-0947  
Tlx RS 61117 FRSPL  
8200 Market Boulevard  
Chanhassen, MN 55317 USA  
Tel 1-800-999-9307  
Telex 4310012  
Fax (612) 949-7001  
© 1998 Rosemount Inc.  
N
IN  
S.  
¢00809-0100-4263z¤  
00809-0100-4263, Rev. AA  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Eton Home Safety Product ARCCR100RSNG User Manual
FireplaceXtrordinair Indoor Fireplace 1080 User Manual
Fisher Price Baby Furniture 79026 User Manual
Fisher Price Bouncy Seat B8786 User Manual
Flymo Trimmer Multi Trim User Manual
Fostex Printer VF 08 User Manual
FujiFilm Camcorder 16178897 User Manual
GE Clothes Dryer DCCB330GDKC User Manual
GE Microwave Oven JES1339 User Manual
Global Machinery Company Trimmer LTP25SS User Manual