EXFO Photonic Solutions Div Power Supply IQS 1500 User Manual

C A LIB R ATIO N P O WE R M E TE R  
IQ S -1 5 0 0  
15 0 0  
R &D AN D MAN U FAC TU R IN G  
Operates in the WDM window  
Enables on-site power-meter calibration  
Linearity of ± 0.01 dB  
5 mm cooled detector  
Telecommunications Test and Measurement  
IQ S -1 5 0 0  
C a lib ra tio n Mo d e  
The IQS-1500s Windows-compatible, highly intuitive GUI allows easy control via mouse, touchscreen or keyboard. Also, it lets you save multiple  
user configurations and access the instruments online user guide.  
A Start  
Allows the user to select the calibration mode, start a new calibration  
sequence with template file or view a report file.  
A
B Information  
B
C
D
Allows the user to enter general and specific information about current  
E
F
measurement conditions. This information will be included in the  
calibration summary.  
G
H
C General  
I
J
Contains user and module information currently used to perform  
calibration.  
K
L
D Conditions  
Contains all the information pertaining to the conditions at which the  
calibration is performed: source wavelength, power and bandwidth,  
temperature and humidity.  
E Device under Test (DUT)  
Contains all the information pertaining to the DUT under calibration.  
J DUT Measurements  
F Uncertainties  
Allows the user to enter DUT power measurements.  
Allows the user to enter specific information pertaining to the  
calculation of the calibration uncertainty.  
K Summary  
G Null  
Displays complete calibration results, including mean power, calibration  
factor and total uncertainty.  
Allows the user to eliminate electronic offsets and dark currents to  
ensure accuracy.  
L Detailed calibration reports  
H Calibration  
Generates detailed calibration reports that include instrumentation  
identification, environmental information, reference conditions and  
a calibration summary for each wavelength.  
Allows the user to perform calibration measurements and calculation.  
I Ref. Meter Measurement  
This auto-save function allows the user to select the number of  
measurements and the intervals between readings for automated  
measurements on the reference power meter (IQS-1500).  
The IQS-1500s Summary Window  
Calibration factor for DUT  
readings (in W)  
1
Source wavelength  
5
Wavelength setting of  
power meter under test  
2
Standard deviation of the  
calibration factor  
6
7
1
2
3
4
5
6
7
Mean power measured  
by the IQS-1500  
3
4
Expanded uncertainty of  
corrected DUT readings  
Number of calibration  
measurements  
Download from Www.Somanuals.com. All Manuals Search And Download.  
IQ S -1 5 0 0  
S P E C IF IC ATIO N S 1  
Model  
Detector type  
Detector size (mm)  
Power range1 (dBm)  
Wavelength range (nm)  
Linearity1,2 (dB)  
IQS-1502-Q0  
Ge  
5
+5 to 60  
750 to 1800  
± 0.01  
(+5 dBm to 50 dBm)  
± 2 %3  
IQS-1502-Q1  
Ge  
5
+5 to 60  
750 to 1800  
± 0.01  
(+5 dBm to 50 dBm)  
± 0.9 %4 at 3 λ  
(10 dBm CW)  
STANDARD ACCES S ORIES  
User guide, fiber-optic adapter  
(FOA), one reference test jumper  
and Certificate of Calibration  
Notes  
1. At 1310 nm and 1550 nm.  
2. At a ± 1 °C constant temperature.  
3. Q0 option, uncertainty at EXFO  
reference conditions:  
Uncertainty6  
(at 1310.0 ± 0.1 nm and 1550.0 ± 0.1 nm)  
(0 dBm to 10 dBm CW)  
• ± 2 % uncertainty at (1310.0 ± 0.1) nm  
and (1550.0 ± 0.1) nm, with 9/125 µm  
(B) fiber, and source spectral width  
(FWHM) < 10 nm.  
• ± 3 % uncertainty from 750 nm to  
1700 nm, with 62.5/125 µm (D) fiber,  
and source spectral width (FWHM)  
12 nm.  
± 3 %3  
(750 nm to 1700 nm)  
(30 dBm to 40 dBm CW)  
0.001  
± 3 %4  
(750 nm to 1700 nm)  
(30 dBm to 40 dBm CW)  
0.001  
Power resolution (dB)  
Wavelength resolution (nm)  
Applicable fiber type (µm)  
Connector type5  
0.1  
0.1  
9/125 (B); 50/125 (C); 62.5/125 (D)  
B: 89  
9/125 (B); 50/125 (C); 62.5/125 (D)  
B: 58; 88; 89; 90; 91  
C/D: 50; 54; 74  
• FC/UPC connector (ceramic ferrule)  
with FOA-322.  
GENERAL S PECIFICATIONS  
4. Q1 option, calibration at NIST  
a) • ± 0.9 % uncertainty at three  
user-specified wavelengths within  
the following ranges at NIST  
reference conditions:  
1
Size (H × W × D)  
Weight  
12.5 cm × 3.6 cm × 28.2 cm  
0.63 kg  
23 °C ± 5 °C  
(4 15/16 in × 1 7/16 in × 11  
(1.4 lb)  
(73 °F ± 9 °F)  
/
8
in)  
Temperature  
operating  
storage  
40 °C to 70 °C  
(40 °F to 158 °F)  
Relative humidity  
0 % to 80 % non-condensing  
786 nm ± 0.5 nm  
(840 to 860) nm ± 0.25 nm  
(1280 to 1330) nm ± 0.25 nm  
(1520 to 1625) nm ± 0.25 nm  
• 9/125 to 62.5/125 µm (B to D) fiber.  
• FC, ST or SC connector.  
• Fiber-optic adapter (FOA) used  
at NIST.  
O R D E R IN G IN F O R M ATIO N  
IQ S -1 5 0 X -X X -X -X X  
Model  
Source spectral width (FWHM)  
< 10 nm.  
IQS-1502-Q0-B = 5 mm TEC-Ge detector, 2 % uncertainty at (1310 ± 0.1) nm and (1550 ± 0.1) nm, Singlemode 9/125 µm7  
IQS-1502-Q1-B = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Singlemode 9/125 µm  
IQS-1502-Q1-C = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 50/125 µm  
IQS-1502-Q1-D = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 62.5/125 µm  
b) ± 3 % uncertainty from 750 nm to  
1700 nm, 62.5/125 µm (D) fiber,  
source spectral width (FWHM)  
< 12 nm, and FC/UPC connector  
(ceramic ferrule) with FOA-322.  
5. See Ordering Information for details.  
6. All uncertainties are valid on the  
day of calibration after a warm-up  
time of 20 minutes, and specified  
with a 95 % confidence level.  
7. Test jumper FC/UPC only  
8. Singlemode only  
Test jumper connector  
50 = FC/PC9  
54 = SC/PC9  
58 = FC/APC narrow key8  
74 = ST/PC9  
Example: IQS-1502-Q1-B-89  
88 = SC/APC8  
89 = FC/UPC narrow key8  
90 = ST/UPC8  
91 = SC/UPC8  
9. Multimode only  
Corporate Headquarters > 400 Godin Avenue, Vanier (Quebec) G1M 2K2 CANADA Tel.: 1 418 683-0211 Fax: 1 418 683-2170 info@exfo.com  
Toll-free: 1 800 663-3936 (USA and Canada) www.exfo.com  
EXFO America  
EXFO Europe  
EXFO Asia-Pacific  
EXFO China  
4275 Kellway Circle, Suite 122  
Addison, TX 75001 USA  
Tel.: 1 800 663-3936  
Tel.: +33.1.40.83.85.85  
Tel.: +65 6333 8241  
Tel.: +86 (10) 6849 2738  
Fax: 1 972 836-0164  
Fax: +33.1.40.83.04.42  
Fax: +65 6333 8242  
Fax: +86 (10) 6849 2662  
Le Dynasteur, 10/12 rue Andras Beck  
92366 Meudon la Forêt Cedex FRANCE  
SINGAPORE 169876  
151 Chin Swee Road, # 03-29 Manhattan House  
Beijing New Century Hotel Office Tower, Room 1754-1755  
No. 6 Southern Capital Gym Road  
Beijing 100044 P. R. CHINA  
EXFO is certified ISO 9001 and attests to the quality of these products. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:  
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. EXFO  
has made every effort to ensure that the information contained in this specification sheet is accurate. However, we accept no responsibility for any errors or omissions, and we  
reserve the right to modify design, characteristics and products at any time without obligation. Units of measurement in this document conform to SI standards and practices.  
Contact EXFO for prices and availability or to obtain the phone number of your local EXFO distributor.  
For the most recent version of this spec sheet, please go to the EXFO website at http://www.exfo.com/specs  
In case of discrepancy, the Web version takes precedence over any printed literature.  
SPIQS1500.2AN  
©
2004 EXFO Electro-Optical Engineering Inc. All rights reserved.  
Printed in Canada 04/ 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Clarion Car Video System OHM642 User Manual
Conair Automobile Accessories HMC1 User Manual
Cradlepoint Cell Phone Accessories PS6PMCW User Manual
Craftsman Lawn Mower Accessory 944511591 User Manual
Crown Audio Stereo Amplifier 1 VCAP User Manual
Cuisinart Oven PIZ 100 User Manual
Curtis Battery Charger 1620S User Manual
Desa Outdoor Fireplace VTGF33NRA User Manual
Frigidaire Oven 318205325 User Manual
Garland Range GXHP11 User Manual