Compur Portable Generator Gas Generator User Manual

Download from Www.Somanuals.com. All Manuals Search And Download.  
Inhaltsverzeichnis / Table of Contents  
Seite / Page  
Deutsch  
1. Bedienungsanleitung  
1.1 Funktionsbeschreibung  
1.2 Testablauf  
3
4
4
2. Wartung  
5
5
6
6
7
8
2.1 Batteriewechsel  
2.2 Austausch der Generatorzelle  
2.2.1 H2S und HCN Generator  
2.2.2 CO, Cl2 und NO2 Generator  
2.2.3 COCl2 und SO2 Generator  
English  
3. Operating Instructions  
3.1 Principle of Operation  
3.2 How To Run a Gas Test  
9
9
9
4. Maintenance  
10  
10  
11  
11  
12  
13  
4.1 Battery Replacement  
4.2 Generator Cell Replacement  
4.2.1 H2S and HCN Generator  
4.2.2 CO, Cl2 and NO2 Generator  
4.2.3 COCl2 and SO2 Generator  
5. Ersatzteile /Spare Parts  
14  
14  
6. Technische Daten / Technical Data  
7. Abbildungen / Drawings  
7.1 H2S, HCN Generator  
15  
15  
16  
17  
7.2 CO, Cl2, NO2 Generator  
7.3 COCl2 and SO2 Generator  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.  
Bedienungsanleitung  
1.1  
Funktionsbeschreibung  
Der Compur Gasgenerator ermöglicht die Überprüfung von Gasdetektoren. Das Gerät ist mit  
einer gasspezifischen Generatorzelle bestückt. Je nach Gasart wird entweder elektrochemisch  
oder thermisch eine über dem MAK1-Wert liegende Gaskonzentration erzeugt. Mit dem  
erzeugten Gas wird der Gasdetektor auf sicheres Ansprechen getestet. Der Gasgenerator ist  
nicht geeignet, ein Prüfgas definierter Konzentration zu erzeugen. Er dient also nicht zur  
Kalibration eines Detektors.  
1.2  
Testablauf  
Zum Testen wird der Gasdetektor mit dem Sensor auf den Adapter des Gasgenerators  
aufgesetzt.  
Beim Herunterdrücken wird automatisch der Gasgenerator angeschaltet und erzeugt nun  
das Testgas. Zum Testen der Compur Gasdetektoren Monitox plus bzw. Dositox ist bei  
jedem Gasgenerator ein Verlängerungsstück für den Schaltstift aufgesteckt.  
Monitox S/SD und Minitox Detektoren werden ohne Verlängerungsstück getestet. Bei  
Fremdfabrikaten ist jeweils zu prüfen, welcher Stift am besten passt.  
Der Gasgenerator erzeugt 10 Sekunden lang Gas. Dabei leuchtet die grüne LED auf.  
Leuchtet die rote LED, muß die Batterie des Gasgenerators gewechselt werden.  
Hinweis:  
) Die Generatorzelle kann bei geringer Luftfeuchtigkeit austrocknen. Verschließen Sie  
deshalb den Generator mit der mitgelieferten Feuchthaltekappe. Achten Sie darauf, dass  
der Schwamm in der Kappe stets feucht, aber nicht naß gehalten wird.  
Der Compur Gasgenerator darf nicht im explosionsgefährdeten  
Bereich benutzt oder abgestellt werden.  
1 MAK = Maximale Arbeitsplatzkonzentration  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2.  
Wartung  
Sollte eine Störung an Ihrem Gasgenerator auftreten, so wenden Sie sich bitte an den Compur  
Monitors Kundendienst in München oder an Ihren lokalen Repräsentanten. Kleinere  
Wartungsarbeiten wie den Batteriewechsel oder Austausch der Generatorzelle können Sie  
selbst vornehmen.  
Beachten Sie bitte dazu folgende Hinweise:  
) Führen Sie die Arbeiten möglichst auf einer leitfähigen, geerdeten Arbeitsmatte aus.  
) Tragen sie möglichst ein leitfähiges Masseband (Armband, verbunden mit der  
Arbeitsmatte).  
2.1  
Batteriewechsel  
Leuchtet die rote LED am Gasgenerator auf, muß die Batterie gewechselt werden.  
Interne Stromversorgung (9V Block-Batterie)  
1 Lösen Sie die vier Schrauben an der Rückseite des Gerätes.  
2 Wenden Sie das Gerät und heben Sie das Gerätevorderteil vorsichtig ab.  
3 Entfernen sie die alte Batterie aus der Halterung.  
4 Setzen Sie die neue Batterie ein.  
5 Setzen Sie das Gerätevorderteil wieder auf und schrauben Sie es fest.  
Externe Stromversorgung (Batteriebehälter) – nur bei COCl2  
1 Entfernen Sie die Steckverbindung zum Gasgenerator und lösen Sie die vier Schrauben an  
der Rückseite des Batteriebehälters.  
2 Öffnen Sie den Behälter und entnehmen Sie die alten Batterien.  
3 Setzen Sie einen neuen Batteriesatz (4 x 1,5 V Mono-Zellen, wir empfehlen Varta High  
Energy Alkaline oder Duracell Ultra M3) ein und schrauben Sie den Behälter wieder zu.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.2  
Austausch der Generatorzelle  
Die Lebensdauer der Generatorzelle beträgt ca. ein Jahr bzw. 2.500 Tests. Danach sollte sie  
ausgetauscht werden. Die Generatorzelle sollte auch in jedem Fall getauscht werden, wenn  
ein nachweislich intakter Gasdetektor beim Funktionstest nicht mehr anspricht.  
2.2.1  
H2S und HCN Generator (siehe Abbildung 1)  
1 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das  
Gerätevorderteil ab.  
2 Entnehmen Sie die Batterie.  
3 Löten Sie die zwei Anschlüsse der alten Generatorzelle ab und ziehen Sie den Schlauch  
von der Pumpe ab.  
4 Lösen Sie die vier Befestigungsschrauben der Leiterplatte und entnehmen Sie die  
Leiterplatte.  
5 Lösen Sie die drei Befestigungsschrauben der Generatorzelle und ziehen Sie diese nach  
vorne heraus. Nehmen Sie die neue Generatorzelle aus der Verpackung und befestigen Sie  
die neue Generatorzelle mit den Befestigungsschrauben.  
6 Bauen Sie die Leiterplatte wieder ein und schrauben Sie sie fest. Bitte achten Sie darauf,  
dass der Schaltstift den Schalter auf der Leiterplatte sicher betätigt.  
7 Schließen Sie die Batterie wieder an und legen Sie sie in das dafür vorgesehene Fach ein.  
8 Führen Sie gegebenenfalls einen Abgleich des Generatorstroms durch.  
-
-
Amperometer an Pin 4 (mittlerer Pin) und Pin 7 (linker Pin) anschließen.  
Schalter S101 drücken und gleichzeitig mit Poti R114 den Generatorstrom einstellen:  
H2S = 3,5 mA; HCN 100 ppm = 0,7 mA; HCN 20 ppm = 0,35 mA  
9 Löten Sie die Kabel der Generatorzelle fest:  
-
-
schwarzes Kabel an Pin 4 (mittlerer Pin)  
gelbes Kabel an Pin 7 (linker Pin)  
und stecken Sie den Schlauch auf den Stutzen der Pumpe, der mit „out“ gekennzeichnet  
ist.  
10 Setzen Sie das Gerätevorderteil wieder auf und schrauben Sie es fest.  
Hinweis:  
) Führen Sie mit einem funktionsfähigen Gasdetektor den Funktionstest durch.  
Nach spätestens 10 Sekunden muß der Detektor in den Alarmzustand gehen.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.2.2  
CO, Cl2 und NO2 Generator (siehe Abbildung 2)  
Diese Generatorzellen enthalten einen Lüfter, der fest mit der Generatorzelle verbunden ist.  
Beim Austausch verfahren Sie wie folgt:  
1 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das  
Gerätevorderteil ab.  
2 Entnehmen Sie die Batterie.  
3 Löten Sie die drei Anschlüsse der alten Generatorzelle ab.  
4 Lösen Sie die vier Befestigungsschrauben der Leiterplatte und entnehmen Sie die  
Leiterplatte.  
5 Lösen Sie die drei Befestigungsschrauben der Generatorzelle und ziehen Sie diese nach  
vorne heraus. Nehmen Sie die neue Generatorzelle aus der Verpackung und befestigen  
Sie die neue Generatorzelle mit den Befestigungsschrauben.  
6 Bauen Sie die Leiterplatte wieder ein und schrauben Sie sie fest. Bitte achten Sie darauf,  
dass der Schaltstift den Schalter auf der Leiterplatte sicher betätigt.  
7 Schließen Sie die Batterie wieder an und legen Sie sie in das dafür vorgesehene Fach ein.  
8 Führen Sie gegebenenfalls einen Abgleich des Generatorstroms durch.  
-
-
Amperometer an Pin 4 (mittlerer Pin) und Pin 7 (linker Pin) anschließen.  
Schalter S101 drücken und gleichzeitig mit Poti R114 den Generatorstrom einstellen:  
CO = 13 mA; Cl2 = 0,3 mA; NO2 = 5 mA  
9 Löten Sie die Kabel der Generatorzelle mit Lüfter fest:  
-
-
-
schwarzes Kabel an Pin 4 (mittlerer Pin)  
gelbes Kabel an Pin 7 (linker Pin)  
rotes Kabel an Pin 3 (rechter Pin)  
10 Setzen Sie das Gerätevorderteil wieder auf und schrauben Sie es fest.  
Hinweis:  
) Führen Sie mit einem funktionsfähigen Gasdetektor den Funktionstest durch.  
Nach spätestens 10 Sekunden muß der Detektor in den Alarmzustand gehen.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.2.3  
COCl2 und SO2 Generator (siehe Abbildung 3)  
1 Stecken Sie den externen Batteriebehälter aus (nur bei COCl2).  
2 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das  
Gerätevorderteil ab.  
3 Entnehmen Sie die Batterie.  
4 Schieben Sie die beiden Isolierschläuche zurück und löten Sie die Anschlussdrähte der  
Generatorzelle von den Litzen.  
5 Schrauben Sie das Klemmstück (KS) ab und schieben Sie die Generatorzelle rückwärts  
heraus.  
6 Führen Sie die neue Generatorzelle vorsichtig in die Adapterkappe ein und schrauben Sie  
das Klemmstück fest.  
7 Löten Sie die Drähte der neuen Zelle an (Polung beliebig) und schieben Sie die  
Isolierschläuche über die Lötstellen.  
8 Setzen Sie die Batterie wieder ein.  
9 Setzen Sie das Gerätevorderteil wieder auf und schrauben Sie es fest. Stecken Sie den  
Batteriebehälter wieder ein.  
Hinweis:  
) Führen Sie mit einem funktionsfähigen Gasdetektor den Funktionstest durch.  
Nach spätestens 10 Sekunden muß der Detektor in den Alarmzustand gehen.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.  
Operating Instructions  
3.1  
Principle of Operation  
The Compur gas generator is a device to check the proper function of gas detectors. The  
instrument contains a gas specific generator cell. Depending on the type of gas, the generator  
will produce, either electrochemically or thermally, a gas concentration above the TLV1-level.  
The response of a gas detector can be checked with the small gas concentration produced by  
the gas generator. However, since the gas generator does not produce a certifiable known gas  
concentration, it is not suitable for detector calibration.  
3.2  
How To Run a Gas Test  
Insert the sensor of the gas detector into the adapter of the gas generator.  
When the detector is pressed down, the generation of gas starts automatically. If you want  
to test the Compur Monitors gas detectors Monitox plus or Dositox, please use the switch  
pin extension included with the generator. Compur gas detectors Monitox S/SD and  
Minitox are tested without extension. If you want to check gas detectors of other  
manufacturers, please check whether it works with or without the switch pin extension.  
The gas generator will produce gas for a period of 10 seconds. During gas generation, the  
green LED will turn on. If the red LED illuminates, the battery needs replacement.  
Note:  
) If the generator cell is exposed to low humidity, it may dry out. Therefore, please cover the  
generator with the included humidity cap. Make sure the sponge inside the cap is always  
moist, but not dripping wet.  
Do not store or use a Compur gas generator in classified areas.  
1 TLV = Threshold Limit Value  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4.  
Maintenance  
In service is needed, please contact Compur Monitors Service in Munich or your local  
representative. Minor maintenance work such as battery replacement, or generator cell  
exchange, can be performed by observing the following instructions.  
Please observe the following notes:  
) Use a grounded conductive pad.  
) Wear a conductive wrist strap connected to your work pad.  
4.1  
Battery Replacement  
If the red LED turns on, the battery needs replacement.  
Internal battery (9V battery)  
1 Remove the four screws on the backside of the instrument.  
2 Separate the top and bottom halves of the generator housing. (Use caution as the LED  
wires are attached to the top case)  
3 Remove the old battery.  
4 Insert the new battery.  
5 Attach the top cover and tighten the screws.  
External battery pack (COCl2 only)  
1 Unplug the battery pack from the gas generator and remove the four screws on its  
backside.  
2 Open the battery pack and remove the old batteries.  
3 Insert the new batteries (4x1,5 V size C), close the battery pack and tighten the screws.  
) For the 1.5 V size C batteries, it is recommended to only use Duracell Ultra M3 or Varta  
High Efficiency Alkaline. Use of other batteries may cause flashing of the battery LED  
during a gas test.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.2  
Generator Cell Replacement  
The generator cell has an approximate life time of 1 year or 2,500 tests. After this time it  
needs replacement. It also needs replacement, if a gas detector known to be good does not  
respond to a gas test.  
4.2.1  
H2S and HCN Generator (see Picture 1)  
1 Loosen the 4 screws on the backside of the instrument and separate the two halves. (Use  
caution as the LED’s are attached to the top case)  
2 Remove the battery.  
3 Unsolder the 2 wires from the old generator cell and remove the tubing from the pump.  
4 Unsrew the 4 screws securing the PCB and remove the PCB.  
5 Unscrew the 3 screws of the generator cell, which secure it to the bottom case, and  
remove the old generator cell. Insert the new generator cell and tighten the 3 screws.  
6 Reinsert the PCB and tighten the screws. Please make sure that the switch pin has not  
moved and that it easily activates the switch on the PCB.  
7 Reconnect the battery.  
8 If required, adjust the generator current to the following specifications:  
-
-
Connect current meter to PCB pin 4 (center pin) and pin 7 (left pin).  
Press switch S101 and adjust generator current with potentiometer R114 to the  
applicable current:  
H2S = 3,5 mA; HCN 100 ppm = 0,7 mA; HCN 20 ppm = 0,35 mA  
9 Solder the wires of the generator cell to the PCB pins:  
-
-
black cable to pin 4 (center pin)  
yellow cable to pin 7 (left pin)  
Connect the tubing to the pump port marked “out”.  
10 Close the housing of the generator and tighten the 4 screws in the back.  
Note:  
) Whenever you have done maintenance work on a gas generator, test the function with a  
working gas detector. The detector should give an alarm within 10 seconds.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.2.2  
CO, Cl2 and NO2 Generator (see Picture 2)  
The CO, Cl2 and NO2 generator cells contain an internal fan. When replacing, please proceed  
as follows:  
1 Loosen the 4 screws on the backside of the instrument and seperate the two halves. (Use  
caution as the LED’s are attached to the top case)  
2 Remove the battery.  
3 Unsolder the 3 wires from the old generator cell.  
4 Unscrew the 4 screws securing the PCB and remove the PCB.  
5 Unscrew the 3 screws of the generator cell , which secure it to the bottom case, and  
remove the old generator cell. Insert the new generator cell and tighten the 3 screws.  
6 Reinsert the PCB and tighten the screws. Please make sure that the switch pin has not  
moved and that it easily activates the switch on the PCB.  
7 Reconnect the battery.  
8 If required, readjust the generator current to the following specifications:  
-
-
Connect current meter to PCB pin 4 (center pin) and pin 7 (left pin).  
Press switch S101 and adjust generator current with potentiometer R114 to the  
applicable current:  
CO = 13 mA; Cl2 = 0,3 mA; NO2 = 5 mA  
9 Solder the wires of the generator cell and the fan to the PCB pins:  
-
-
-
black cable to pin 4 (center pin)  
yellow cable to pin 7 (left pin)  
red cable to pin 3 (right pin)  
10 Close the housing of the generator and tighten the screws in the back.  
Note:  
) Whenever you have done maintenance work on a gas generator, test the function with a  
working gas detector. The detector should give an alarm within 10 seconds.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.2.3  
COCl2 and SO2 Generator (see Picture 3)  
1 Unplug the external battery (COCl2 only).  
2 Loosen the 4 screws on the backside of the instrument and separate the two halves. (Use  
caution as the LED’s are attached to the top case)  
3 Remove the battery.  
4 Push back the 2 isolation tubes and unsolder the generator cell from the wires.  
5 Unscrew the clamp (KS) and push the generator cell backwards. Remove old generator  
cell.  
6 Insert the new generator cell and fasten the clamp.  
7 Solder the wires to the new generator cell (polarity of no importance) and push the  
isolation tubes back over the connections.  
8 Insert the battery.  
9 Close the housing and tighten the screws in the back. Reconnect the external battery pack.  
Note:  
) Whenever you have done maintenance work on a gas generator, test the function with a  
working gas detector. The detector should give an alarm within 10 seconds.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Ersatzteile / Spare Parts  
Generatorzellen / Generator cells  
Bestell-Nr.:  
Article-No.  
H2S  
516722  
516847  
516730  
516850  
516912  
516938  
516755  
516763  
COCl2  
HCN 100 ppm  
HCN 20 ppm  
NO2  
CO  
Cl2  
SO2  
6.  
Technische Daten / Technical Data  
Abmessungen / Dimensions  
133 x 65 x 40 mm  
Gewicht mit Batterie / Weight incl. battery  
Temperaturbereich / Temperature range  
Interne Stromversorgung / Internal power supply  
Lebensdauer der Generatorzelle /  
250 g  
+10°C - +45°C  
9V Batterie/Battery  
typ.: 1 Jahr/Year  
bzw./aprox. 2500 Tests  
Expected life of generator cell  
Nur bei COCl2 / COCl2 only:  
Externe Stromversorgung / External power supply  
4 x 1,5V Batterie/Battery  
Wir empfehlen / We recommend:  
Varta High Energy Alkaline  
Duracell Ultra M3  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.  
Abbildungen / Drawings  
7.1  
H2S, HCN Generator  
Abbildung / Drawing 1  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.2  
CO, Cl2, NO2 Generator  
Abbildung / Drawing 2  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.3  
COCl2, SO2 Generator  
Abbildung / Drawing 3  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notizen / Remarks  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Die vorliegenden Informationen erfolgen nach bestem Wissen, gelten jedoch nur als  
unverbindlicher Hinweis auch in Bezug auf etwaige Schutzrechte Dritter.  
Die vorangegangen technischen Daten und Anwendungshinweise befreien den Anwender  
nicht von einer eigenen Prüfung unserer Produkt- und Anwendungsvorschläge im Hinblick  
auf ihre Eignung für die beabsichtigten Verfahren und Zwecke.  
Die Anwendung unserer Produkte erfolgt außerhalb unserer Kontrollmöglichkeiten und liegt  
daher ausschließlich im Verantwortungsbereich des Kunden. Der Verkauf unserer Produkte  
erfolgt nach der Maßgabe der allgemeinen Verkaufs- und Lieferbedingungen der Compur  
Monitors GmbH & Co. KG, München.  
Specifications are subject to change without notice, and are provided only for comparison of  
products. The conditions under which our products are used, are beyond our control.  
Therefore, the user must fully test our products and / or information to determine suitability  
for the intended use, application, condition or situation. All information is given without  
warranty or guarantee. Compur Monitors disclaims any liability, negligence or otherwise,  
incurred in connection with the use of products and information. Any statement or  
recommendation not contained herein is unauthorized and shall not bind Compur Monitors.  
Nothing herein shall be construed as a recommendation to use any product in conflict with  
patents covering any material or device or its use. No licence is implied or in fact granted  
under the claims of any patent. Instruments are manufactured by Compur Monitors GmbH &  
Co. KG, Munich.  
The General Conditions of Supply and Services of Compur Monitors GmbH & Co. KG are  
applicable.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Compur Monitors GmbH & Co. KG  
Weißenseestraße 101 · D-81539 München  
Tel.: ++49/89/ 6 20 38 – 268  
Fax : ++49/89/ 6 20 38 – 184  
5390 000 998 08 02 / 07.03  
550580  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Coleman Outdoor Fireplace 5059A703 User Manual
Conair Scale TH328 User Manual
Cowon Systems MP3 Player U5 User Manual
Cradlepoint Modem MC100E User Manual
Craftsman Impact Driver 87519119 User Manual
Dual Stereo System XD5125 User Manual
Electro Voice Music Mixer DM Series User Manual
ETA Systems Oxygen Equipment FC01 LQ User Manual
Extron electronic Switch VYC 100N User Manual
Frigidaire Cooktop 318200563 User Manual