Campbell Manufacturing Oxygen Equipment CS512 User Manual

Campbell Scientific CS512  
Oxyguard Type III  
Dissolved Oxygen Probe  
Revision: 12/06  
C o p y r i g h t  
©
2 0 0 1 - 2 0 0 6  
C a m p b e l l S c i e n t i f i c , I n c .  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CS512 Table of Contents  
PDF viewers note: These page numbers refer to the printed version of this document. Use  
the Adobe Acrobat® bookmarks tab for links to specific sections.  
1. General Information.....................................................1  
2. Specifications ..............................................................2  
3. Optional Probe Accessories.......................................3  
4. Optional Agitator .........................................................3  
5. Application Information ..............................................4  
6. Wiring............................................................................4  
7. Programming ...............................................................5  
8. Calibration....................................................................6  
9. Maintenance.................................................................7  
10. Agitator Control .........................................................7  
Appendix  
A. Dissolved Oxygen in Fresh Water Table............... A-1  
Figures  
1-1. CS512......................................................................................................1  
4-1. Preventing Bio-fouling of the CS512......................................................3  
4-2. CS512 with PT4 Agitator........................................................................4  
10-1. Agitator Wiring .....................................................................................7  
Table  
6-1. Sensor Wiring..........................................................................................4  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
CS512 Table of Contents  
Examples  
1. Sample Program using P1.......................................................................... 5  
2. Sample Program using P2.......................................................................... 5  
3. Sample Program using VoltSE .................................................................. 5  
4. Sample Program using VoltDiff ................................................................ 5  
5. Sample Program using VoltSE .................................................................. 5  
6. Sample Program using P36........................................................................ 6  
7. Sample Program using Expression ............................................................ 6  
8. Sample Program using P86 and P22 Instructions ...................................... 8  
9. Sample Program using Portset ................................................................... 8  
10. Sample Program using Portset ................................................................. 8  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard  
Type III Dissolved Oxygen Probe  
An accurate and reliable sensor is a critical element in any measurement system. The  
Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen.  
1. General Information  
The Oxyguard Stationary Probe is a galvanic probe which produces a millivolt  
signal proportional to the amount of oxygen present in the measured medium.  
Oxygen diffuses through the membrane onto the cathode, reacts chemically, and  
combines with the anode. An electrical current is produced by this chemical  
reaction which is converted from microamps to millivolts by an in-line resistor.  
An in-line thermistor also conditions the signal providing automatic temperature  
compensation. With these features, the probe produces a millivolt output  
proportional to the oxygen present in the medium in which it is placed. The probe  
consists of two parts, an upper part with cathode, anode and cable, and a lower part  
comprising a screw on membrane cap with fitted membrane. The cap is filled with  
electrolyte and simply screwed onto the top component.  
The probe is self-polarizing and requires no external power source. There are two  
wires to connect.  
Because the probe's output is linear, it is possible to connect it directly to a data  
acquisition system capable of handling the small millivolt signal.  
The probe's robust construction and simple design make maintenance and  
servicing it straightforward. There is no need to send the probe back to the factory  
for servicing. It utilizes a strong, easy-to-clean and easy-to-change membrane in a  
screw-on membrane cap. Regular servicing is not required. When necessary the  
probe can be fully overhauled in five minutes.  
FIGURE 1-1. CS512  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
2. Specifications  
Principle of Measurement:  
Membrane covered galvanic oxygen probe  
2.5 to 5 mV per mg 1-1  
Output Signal:  
Repeatability:  
Better than ±0.2 mg/l calibration temperature  
equals measuring temperature ±5°C  
Output Impedance:  
Response Time:  
1 kohm  
After equilibration, 1 minute for 95% of final  
value  
Materials of Construction:  
Probe body:  
Delrin  
O-rings:  
Membrane O-ring = Buna N  
Body Seal O-ring = Viton  
0.05 mm (2 mil) Teflon  
Membrane:  
Dimensions and Weight:  
2.28” diameter x 2.3” height  
15.9 oz.  
Cable:  
standard length 7 m (23 ft); other lengths  
available upon request  
2-wire 22 awg shielded, PVC jacketed  
Operating Conditions:  
Temperature  
0° to 40°C (32° to 122°F)  
0 to 29 psig  
Pressure:  
Minimum Submersion Depth:  
Minimum Water Flow:  
60 mm (2 ½ in)  
1 cm/s (.39 in/sec) across membrane  
Calibration:  
In air or in air saturated water  
Automatic from 4° - 40°C (40° - 104°F)  
0-20 mg/l, 0-200% Sat  
Temperature Compensation:  
Range of Dissolved Oxygen:  
Ships with:  
5 spare membranes with O-rings  
50 ml electrolyte  
cathode cleaning pad  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
3. Optional Probe Accessories  
PT4-L — Agitator for stagnant conditions  
18026 — DO electrolyte, 1 liter  
18027 — Membrane kit, 10 membranes  
4. Optional Agitator  
The PT4 Agitator is a reliable and robust agitator for use in conjunction with  
probes subjected to bio-fouling in ponds and stagnant water conditions.  
O2 probes require a minimum water velocity across their membranes to function  
properly. Therefore, to measure DO in stagnant water conditions, it is necessary to  
move the water past the membrane to get accurate and reliable DO measurements.  
In many instances the water also has a high bio-loading and the probes become  
fouled resulting in inaccurate DO measurements.  
FIGURE 4-1. Preventing Bio-fouling of the CS512  
The PT4 Agitator overcomes these problems. The device is designed so that a soft  
bristle brush sweeps across the probe membrane or sensor tip. This sweeping  
action of the brush provides the required water velocity as well as prevents the  
membrane from becoming bio-fouled.  
The optimum sweeping frequency depends upon the design of probe and type of  
membrane used and water conditions. An ON-time of 0.25 seconds and OFF-time  
of 5 seconds is suitable in most circumstances.  
Agitator’s overall size: 7-1/8" x 3-1/4" diameter, 1-1/4 lb. (180mm x 83mm, 0.6 kg)  
Supplied with 10 ft. (3 meters) cable.  
Power required: 10.5 to 18 VDC at the agitator, 1.1 amps. Maximum ON-time is  
3 seconds.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
FIGURE 4-2. CS512 with PT4 Agitator  
Optional Repeat Cycle Timer for Agitators: Reciprocating action may be  
controlled by the optional solid state Repeat Cycle Timer. It sends 12 VDC pulses  
to the agitator coil. The ON-time is 0.25 sec. The OFF-time is adjustable from 3  
to 12 sec.; requires supply voltage 10 to 17 VDC; housed in watertight cylinder  
6.6" x 2" diameter (170 mm X 50 mm).  
5. Application Information  
SCWin users: This manual was written primarily for those  
whose needs are not met by SCWin. Your procedure is much  
simpler: just add the CS512 probe (it’s in the water folder), save  
your program, and follow the wiring shown in Step 2 of SCWIN.  
NOTE  
6. Wiring  
The CS512 probe can use one differential channel or one single-ended channel.  
Differential wiring is better at rejecting electrical noise and ground loop error.  
TABLE 6-1. Sensor Wiring  
CR510, CR10X, CR800, CR850,  
Colo  
r
Functio  
n
CR23X, CR1000, CR3000  
CR200 Series  
Whit  
e
Signal +  
Differential High,  
or Single-Ended Channel  
Single-Ended  
Channel  
Blue  
Signal -  
Differential Low or AG  
Ground  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
7. Programming  
In the CR510, CR10X, and CR23X datalogger, Instruction P1 or P2 can be used.  
Example 1 uses a single-ended measurement for the CS512 probe; example 2 uses  
a differential measurement. The example measurement instructions that follow do  
not store data to final storage. Additional instructions (typically P92, P77, and  
output processing instructions such as P70) are required to store data permanently.  
EXAMPLE 1. Sample Program using P1  
1: Volt (SE) (P1)  
1:  
1
Reps  
2: 24  
250 mV 60 Hz Rejection Range ; code 23 used for CR23X  
3:  
4:  
5: 1.0  
6: 0.0  
1
1
SE Channel  
Loc [ DOmV  
Multiplier  
Offset  
]
*See Calibration*  
EXAMPLE 2. Sample Program using P2  
1: Volt (Diff) (P2)  
1:  
1
Reps  
2: 24  
250 mV 60 Hz Rejection Range ; code 23 used for CR23X  
3:  
4:  
5: 1.0  
6: 0.0  
1
1
DIFF Channel  
Loc [ DOmV  
Multiplier  
Offset  
]
*See Calibration*  
In the CR1000, CR800, CR850, and CR3000 dataloggers, instruction VoltDiff  
or VoltSE is used to measure the CS512 probe.  
EXAMPLE 3. Sample Program using VoltSE  
VoltSE (DOmV,1,mV250,1,1,0,_60Hz,1.0,0)  
EXAMPLE 4. Sample Program using VoltDiff  
VoltDiff (DOmV,1,mV250,1,True ,0,_60Hz,1.0,0)  
In the CR200 series dataloggers, only the instruction VoltSE is used to  
measure the CS512 probe.  
EXAMPLE 5. Sample Program using VoltSE.  
VoltSE (DOmV,1,1,1.0,0)  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
8. Calibration  
The multiplier is used to calibrate the CS512 probe. To calculate the multiplier:  
1) Program the datalogger using a multiplier of one.  
2) Place the OxyGuard probe in the air, shaded from the sun. Wait for readings  
to stabilize. This may take 15 minutes or more.  
3) Determine the air temperature and barometric pressure.  
4) Using a calibration chart such as that provided in the probe’s manual,  
determine the oxygen concentration of the air.  
5) Use the following equation to calculate the multiplier:  
M = P/R  
M = Multiplier  
P = Concentration in PPM of the air (from the calibration chart in Appendix A)  
R = The signal output of the OxyGuard probe when using a multiplier of one  
6) Change the multiplier in the datalogger program from one to the calculated  
number.  
A more common way to enter the multiplier is to insert a separate instruction in  
the program. This will allow a new multiplier to be added to the program  
without rewriting, compiling, and downloading the program to the data logger.  
In the CR510, CR10X, and CR23X, use Instruction P36. The multiplier is  
entered into an input location called DOmult using the numeric display in  
PC200W, PC208W, LoggerNet, PC400, PConnect, PConnectCE, or the  
CR10KD.  
EXAMPLE 6. Sample Program using P36  
57: Z=X*Y (P36)  
1:  
2:  
3:  
1
2
3
X Loc [ DOmV  
Y Loc [ DOmult  
Z Loc [ DOppm  
]
]
]
In all dataloggers that use CRBasic, an expression is written. The multiplier  
value is entered into the expression through the Public Table using the numeric  
display in PC200W, LoggerNet, PC400, PConnect, and PConnectCE.  
EXAMPLE 7. Sample Program using Expression  
DOppm = DOMult * DOmV  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
9. Maintenance  
The CS512 probe needs little maintenance. Regular cleaning of the membrane is  
all that is required. The membrane is very durable and can be cleaned with a cloth  
or soft paper. Do not scratch it clean with your fingernail.  
10. Agitator Control  
In low flow conditions (less than about 5 cm/sec), it may be necessary to add  
an agitator to the CS512 probe. Campbell Scientific ships the agitator with a  
repeat cycle timer. Using the repeat cycle timer requires no datalogger  
programming. However, some users choose to use a solid state relay and have  
the datalogger agitate the water on the probe face either periodically  
throughout the day or just before measurement. Agitating just before the  
measurement saves on power and causes less wear and tear on the agitator and  
probe membrane.  
The wiring for the agitator as controlled by this example program would be as  
follows:  
BLACK  
1
Ground/G  
Datalogger  
4
3
Agitator  
#7321 Relay  
Crydom D1D07  
Port C1  
WHITE  
2
+
-
12 V Battery  
FIGURE 10-1. Agitator Wiring  
The following instructions would trigger the agitator as discussed in the  
agitator manual.  
In the CR510, CR10X, and CR23X, use instruction P86 and P22.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe  
EXAMPLE 8. Sample Program using P86 and P22 Instructions  
45: Do (P86)  
1: 41  
Set Port 1 High  
46: Excitation with Delay (P22)  
1:  
1
Ex Channel  
2: 20  
Delay W/Ex (units = 0.01 sec)  
Delay After Ex (units = 0.01 sec)  
mV Excitation  
3:  
4:  
0
0
47: Do (P86)  
1: 51  
Set Port 1 Low  
48: End (P95)  
In the CR200 series dataloggers, use the Portset instruction.  
EXAMPLE 9. Sample Program using Portset  
Portset (1,1)  
Delay(500,msec)  
Portset(1,0)  
In the CR1000, CR800, CR850, and CR3000 dataloggers, use the Portset  
instruction.  
EXAMPLE 10. Sample Program using Portset  
Portset (1,1)  
Delay (1,500,msec)  
Portset (1,0)  
The above example is not as power efficient as possible and would require AC  
power to maintain a sufficient battery charge. If it is necessary to operate an  
agitator without AC power available, write the program so that the agitator is  
only operated for a short period of time just before the measurement is to be  
taken.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A. Dissolved Oxygen in  
Fresh Water Table  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A. Dissolved Oxygen in Fresh Water Table  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
This is a blank page.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Campbell Scientific Companies  
Campbell Scientific, Inc. (CSI)  
815 West 1800 North  
Logan, Utah 84321  
UNITED STATES  
Campbell Scientific Africa Pty. Ltd. (CSAf)  
PO Box 2450  
Somerset West 7129  
SOUTH AFRICA  
Campbell Scientific Australia Pty. Ltd. (CSA)  
PO Box 444  
Thuringowa Central  
QLD 4812 AUSTRALIA  
Campbell Scientific do Brazil Ltda. (CSB)  
Rua Luisa Crapsi Orsi, 15 Butantã  
CEP: 005543-000 São Paulo SP BRAZIL  
Campbell Scientific Canada Corp. (CSC)  
11564 - 149th Street NW  
Edmonton, Alberta T5M 1W7  
CANADA  
Campbell Scientific Ltd. (CSL)  
Campbell Park  
80 Hathern Road  
Shepshed, Loughborough LE12 9GX  
UNITED KINGDOM  
Campbell Scientific Ltd. (France)  
Miniparc du Verger - Bat. H  
1, rue de Terre Neuve - Les Ulis  
91967 COURTABOEUF CEDEX  
FRANCE  
Campbell Scientific Spain, S. L.  
Psg. Font 14, local 8  
08013 Barcelona  
SPAIN  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Beko Washer WMD77107 User Manual
Bernard Welder SPEC FN 11 User Manual
Bertazzoni Cooktop CB36500X User Manual
Blue Rhino Gas Grill NBC1218G C User Manual
Braun Coffeemaker KF 185 User Manual
Brinkmann Gas Grill Series 2630 User Manual
Casio Clock Radio MA0605 EA User Manual
Char Broil Charcoal Grill 10201595 User Manual
Chauvet Stereo System 5162 User Manual
Chicago Electric Power Hammer 68696 User Manual