Portable Combustion
Analyzer (PCA)
INSTRUCTION 24-9351
Operation & Maintenance
Rev. 11 – May 2004
O
2
4
.
0
C
O
1
2
H
L
N
D
C
2
9
.
5
C
U
1
5
G
T
q
L
A
2
0
8
.
.
0
1
T
L
A
A
1
.
9
2
0
4
P
S
1
M
E
N
U
P
C
A
Product Leadership • Training • Service • Reliability
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Contents
Contents
1.0 INTR ODUCTION ............................................................................ 1-1
1.1 The Portable Combustion Analyzer ........................................... 1-1
1.2 Displayed Data ........................................................................... 1-2
1.3 Sensor Configurations ................................................................ 1-3
2.0 TE CH NICAL CH AR ACTE R ISTICS ............................................ 2-1
3.0 SE TTING UP TH E P CA................................................................. 3-1
3.1 Scope ........................................................................................... 3-1
3.2 Power .......................................................................................... 3-1
3.2.1 Checking and Replacing the Batteries ........................... 3-1
3.2.2 Using the Optional Power Supply .................................. 3-2
3.3 Connecting the Probe ................................................................ 3-2
3.4 Configuring the PCA ................................................................. 3-4
4.0 OP E R ATION .................................................................................... 4-1
4.1 Key Pad Functions ..................................................................... 4-1
4.2 Sampling Hole Location ............................................................ 4-2
4.3 Combustion Test ........................................................................ 4-3
4.3.1 Analyzer Turn On and Warm Up .................................. 4-3
4.3.2 Installing Probe in the Stack ......................................... 4-4
4.3.3 Performing a Combustion Test ...................................... 4-5
4.3.4 Ending a Combustion Test ............................................. 4-5
4.3.5 Turning Off the Analyzer and Purging the CO Sensor .... 4-6
4.4 Differential Pressure Measurement ......................................... 4-7
4.5 Warm-up Screen ........................................................................ 4-8
4.6 Sensor Status Screen ................................................................. 4-9
4.7 Combustion Test Screen .......................................................... 4-10
4.8 Fuel Selection Screen .............................................................. 4-11
4.9 Draft Screens ........................................................................... 4-12
4.10 Memory Directory Screen ..................................................... 4-13
4.11 Memory to PC Screen ............................................................ 4-14
4.12 ID Setup Screens ................................................................... 4-17
4.13 Temperature Setup Screen ................................................... 4-19
4.14 Draft Unit Setup Screen ....................................................... 4-20
4.15 Language Setup Screen ......................................................... 4-21
4.16 Display Mode Setup Screen .................................................. 4-22
4.17 Time/Date Setup Screen ........................................................ 4-23
4.18 Printer Setup Screen ............................................................. 4-24
4.19 Maintenance Password Screen ............................................. 4-25
4.20 Maintenance Screen .............................................................. 4-26
4.21 User Name Screens ............................................................... 4-27
Instruction 24-9351
i
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
PCA
4.22 Saving Test Data ................................................................... 4-28
4.23 Printing Test Data ................................................................. 4-29
4.24 Clear Memory Screen ............................................................ 4-31
4.25 Resetting the Microprocessor ................................................ 4-31
5.0 CALIBR ATION ................................................................................ 5-1
5.1 Sensor Check ............................................................................. 5-1
5.2 Calibration Fixtures .................................................................. 5-2
5.3 Calibrate Menu Screen .............................................................. 5-3
5.4 Calibrate TA-Zero ...................................................................... 5-4
5.5 Calibrate TA-Span ..................................................................... 5-5
5.6 Calibrate TL-Zero ...................................................................... 5-6
5.7 Calibrate TL-Span ..................................................................... 5-7
5.8 Calibrate NX .............................................................................. 5-8
5.9 Calibrate CO .............................................................................. 5-9
5.10 Calibrate Draft ....................................................................... 5-11
6.0 MAINTE NANCE .............................................................................. 6-1
6.1 Routine Maintenance ................................................................ 6-1
6.2 Disassembly ............................................................................... 6-2
6.3 Cleaning the Probe .................................................................... 6-4
6.4 Water Trap/Filter Assembly Maintenance ............................... 6-5
6.5 Replacing the Particulate Filter ............................................... 6-6
6.6 Replacing the Oxygen Sensor ................................................... 6-7
6.7 Replacing the Nitric Oxide Sensor ........................................... 6-8
6.7.1 Replacing the Nitric Oxide Sensor Filter ...................... 6-8
6.7.2 Replacing the Nitric Oxide Sensor Bias Battery .......... 6-9
6.8 Replacing the Carbon Monoxide Sensor ................................. 6-10
6.8.1 Replacing the Carbon Monoxide Sensor Filter ........... 6-11
6.9 Replacing the Pump Assembly ............................................... 6-12
7.0 TR OUBLE SH OOTING .................................................................... 7-1
7.1 Analyzer Repair ......................................................................... 7-1
7.2 Error Codes ................................................................................ 7-2
7.3 Troubleshooting Guide .............................................................. 7-3
8.0 P AR TS & SE R VICE ........................................................................ 8-1
8.1 Replacement Parts ..................................................................... 8-1
8.2 Accessories ................................................................................. 8-2
8.3 Sales/Service Centers ................................................................ 8-6
AP P E NDIX A – Disp la y Scr een Tr a n sla t ion s ................................... A-1
AP P E NDIX B – P r in t ou t Tr a n sla t ion s ..............................................B-1
ii
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Introduction
1.0 INTRODUCTION
1.1 The Portable Combustion Analyzer
The Portable Combustion Analyzer (PCA) (Figure 1-1) is a commercial
grade, hand held, combustion efficiency analyzer that is designed for
continuous (on demand) sampling of light industrial and residential
furnaces, appliances, and boilers. The basic instrument is supplied with a
probe, instruction manual, batteries, and carrying case.
O
2
4
.
0
C
O
1
2
H
L
N
D
C
2
9
.
5
C
U
1
5
G
T
q
L
A
2
0
8
.
.
0
1
T
L
A
A
1
.
9
2
0
4
P
S
1
M
E
N
U
P
C
A
Figure 1-1. PCA
Instruction 24-9351
1-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Introduction
PCA
1.2 Displayed Data
The PCA directly measures, displays, and stores the following data:
• Room Temperature in °C or °F (Primary Air/Ambient Temperature)
• Flue Gas Oxygen Content in %
• Flue Gas Temperature in °C or °F
• Flue Gas Carbon Monoxide Content (H2 Compensated) in ppm
(For analyzers having a Carbon Monoxide sensor)
• Flue Gas Nitric Oxide content in ppm
(For analyzers having a Nitric Oxide sensor)
• Pressure/Draft in Millibars, Pascals, or Inches of Water Column
(For analyzers having a draft sensor)
• Differential Pressure in Millibars, Pascals, or Inches of Water Column
(For analyzers having a draft sensor)
The PCA will compute, display, and store the following data for any of the
standard fuels:
• Stack Loss in %
• Lambda
• Flue Gas Carbon Dioxide Content in %
• Flue Gas Carbon Monoxide Content referenced to 0% Oxygen in ppm
(For analyzers having a Carbon Monoxide sensor)
• Flue Gas Nitric Oxide Content referenced to 0% Oxygen in ppm
(For analyzers having a Nitric Oxide sensor)
The standard fuels are:
• Natural Gas
• Oil #2
• Oil #6
• LPG
• Koks
• Low Energy Gas
• P-Coal (available in English, German, Dutch, French, Italian,
Polish, and Spanish languages)
• Biofuel (available in Danish, Finnish, and Swedish languages)
The PCA continuously monitors flue-gas-exhaust conditions and updates
the above displayed values during a combustion test. If the analyzer is
equipped with an optional pressure sensor, draft measurements can be
made simultaneously with the combustion test, or made separately.
The analyzer has the ability to store data that was collected during a
combustion test or draft measurement. The stored data can then either be
viewed on the PCA’s display, printed using an optional printer, or down-
loaded to a computer.
1-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Introduction
1.3 Sensor Configurations
TABLE 1-1. P CA SE NSOR CONF IGUR ATIONS
PCA Models
Sensors Installed
Standard
Advanced
PCA Part No.* Part No.** PCA Part No.* Part No.** Stack Temp.,
CO NX Draft
Model
10
15
20
25
24-
24-
Model
40
45
50
55
24-
24-
Air Temp. & O2
(∆P)
7181
7182
7183
7184
7185
7186
7281
7282
7283
7284
7285
7286
7241
7242
7243
7244
7245
7246
7251
7252
7253
7254
7255
7256
X
X
X
X
X
X
X
X
X
X
X
X
X
30
35
60
65
X
X
* English, Danish, Dutch, German, Finnish & Swedish languages
** English, French, German, Italian, Polish & Spanish languages
PCA 10 & 40
These basic instruments have the capability of measuring, displaying,
and storing combustion tests. They will also display flue gas Oxygen
content, flue gas Carbon Dioxide (CO2) content, air temperature, flue gas
temperature, stack loss, Lambda, and the current fuel selected. The
‘standard’ PCA 10 stores up to 10 combustion tests, while the ‘advanced’
PCA 40 can store up to 100 tests.
PCA 15 & 45 with Draft
In addition to the features of the basic PCAs described above, these instru-
ments have the added capability of measuring, displaying, and saving draft
or differential pressure in either Millibars, Pascals, or Inches-of-Water
Column.
PCA 20 & 50 with Carbon Monoxide Measurement
In addition to the features of the basic PCAs described above, these
instruments have the added capability of measuring, displaying, and
saving Carbon Monoxide (CO) content, as well as calculating CO refer-
enced to 0% Oxygen.
PCA 25 & 55 with Draft and CO Measurement
These instruments combine the features of all the PCAs listed above.
PCA 30 & 60 with CO and Nitric Oxide Measurement
In addition to the features of the basic PCAs with CO measurement, these
instruments have the added capability of measuring, displaying, and
saving Nitric Oxide (NX) content, as well as calculating NX referenced to
0% Oxygen.
Instruction 24-9351
1-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Introduction
PCA
PCA 35 & 65 with Draft, CO & NX
These instruments are capable of measuring, displaying, and saving all
measurements as previously described.
Printout Capability
All PCAs have the ability to print the latest test data, or any of the saved
tests, to an optional printer using HP, IrDA or RS-232 protocol.
Advanced PCA Model Features
‘Advanced’ models of the PCA contain the following features that are in
addition to the features of their corresponding ‘standard’ PCAs:
• 100 memory locations
• RS232 output for transferring saved data to a Personal Computer
• Ability to enter three lines of user-identification information that is
printed at the top of each printout
• Ability to enter three lines of customer-identification information that
is printed with each test record.
• Automatic CO sensor purge on analyzers equipped with a CO sensor
1-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Technical Characteristics
2.0 TECHNICAL CHARACTERISTICS
Th e P CA Dir ectly Measu r es an d Displays:
• Oxygen content in flue gas in the range of 0.1 to 20.9 % O2
• Flue gas temperature in the range of –18 to 1200 ºC (0 to 2192 ºF)
• Primary-air / ambient temperature is in the range of –18 to 999 ºC
(0 to 999 ºF)
Op t ion a l . . .
• Differential pressure/draft in the range of ±70.0 mb (±28" H2O)
• Carbon Monoxide* content in flue gas in the range of 0 to 4000 ppm
• Nitric Oxide* content in flue gas in the range of 0 to 1000 ppm
Th e P CA Com pu tes an d Displays:
(When the measured oxygen level is not above 18.8%, and the Stack (Flue
Gas) temperature is not above 1200 ºC (2192 °F)
• Stack loss in the range of 0.1 to 99.9%
• Carbon Dioxide content in flue gas from 0.1 to a fuel dependent
maximum value in percent
• Lambda in the range of 1 to 9.95
• Carbon Monoxide* content referenced to 0% Oxygen in the range of
0 to 9,999 ppm on analyzers equipped with a CO sensor.
• Nitrix Oxide* content referenced to 0% Oxygen in the range of 0 to
9,999 ppm on analyzers equipped with a NX sensor.
Stan dar d Fu els** Available for
Com bu stion Calcu lation s:
• Natural Gas
• Koks
• LEG
• LPG
• Oil #2
• Oil #6
• P-Coal (available in English, German, Dutch, French, Italian,
Polish, and Spanish languages)
• Biofuel (available in Danish, Finnish, and Swedish languages)
* For the PCA 30, 35, 60, & 65, the display can be set up to show either
measured values of Carbon Monoxide and Nitric Oxide (CO & NX), or
show the calculated values of these gases (CU & NU) referenced to
0% Oxygen. In either case, all values are listed on the printout of
analzyers equipped with a printer.
** Custom fuels available upon request. Contact factory for details.
Instruction 24-9351
2-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Technical Characteristics
PCA
Nor m al Oper atin g Con dition s:
Temperature:
Analyzer ................... 0 to 40 ºC (32 to 104 ºF)
Probe ......................... 800 ºC (1472 ºF) Max.
Humidity:
Analyzer ................... 15 to 90% Relative Humidity, Non-Condensing
Air Pressure:
Analyzer ................... Atmospheric
Probe ......................... 25 mb (10" H2O) draft max at probe tip
P er for m a n ce:
Accuracy:
Oxygen* .................... ±0.3% O2
Carbon Monoxide ..... ±5% of reading or ±10 ppm, whichever is greater
between 0 – 2000 ppm, and ±10% of reading
between 2001 – 4000 ppm.
Nitric Oxide .............. ±5% of reading or ±5 ppm, whichever is greater
Flue Gas Temp. ........ ±2 ºC between 0 and 124 ºC
(±4 ºF between 32 and 255 ºF)
±3 ºC between 125 and 249º C
(±6 ºF between 256 and 480 ºF)
±4 °C between 250 and 400 °C
(±8 ºF between 481 and 752 ºF)
Ambient Temp. ........ ±1 °C between 0 and 100° C
(±2 ºF between 32 and 212 ºF)
Pressure ................... ±2% of reading or ±0.05 mb (±0.02 inches of
Water Column), whichever is greater
System Flow Rate:
With probe ................ 200 cc/min minimum
Fr on t P an el Con tr ols:
Seven embossed push-button switches with tactile feedback (refer to
Section 4.1)
* Accuracy referenced in practical flue gas concentrations
(mixtures of O2, CO2 and N2)
2-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Technical Characteristics
Display:
20 character by 4 line alphanumeric LCD panel with a green backlight.
P ower Requ ir em en ts:
Four disposable AA alkaline batteries. Battery backup for the real-time
clock, RAM, and bias voltage for the Nitric Oxide sensor are provided by
internal lithium batteries. Optional AC Power Supplies (110 VAC &
230 VAC) are also available.
Oper atin g Tim e:
A fresh set of four disposable AA alkaline batteries provides at least
8 hours of continuous operation with the pump running and the backlight
turned on.
Wa r m Up Tim e:
60 seconds.
P r in ter In ter face:
Infrared & RS-232 Communications (refer to Section 4.23).
Mater ials:
• High impact ABS plastic case
• Polycarbonate window over the display
• Nickel plated, brass quick-connect hose fitting
• Stainless steel probe
Dim en sion s:
Height: .................. 215 mm (8.5 in.)
Width: ................... 96 mm (3.8 in.) at display (75 mm [3.0 in.] at controls)
Depth: ................... 50 mm (2 in.)
Weigh t:
With Batteries:..... Approximately 0.7 Kg (1.5 lbs)
Instruction 24-9351
2-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Technical Characteristics
PCA
Agen cy Appr ovals:
TÜV Agency Approved (1.BImSchV - First Ordinance of the German
Federal Emissions Law)
Approval Number: TÜV By RgG 168
CE declaration of conformity
Manufacturer's name:
Manufacturer's address:
Bacharach, Inc.
621 Hunt Valley Circle
New Kensignton, PA 15068
European operations:
Bacharach Instruments
Sovereign House, Queensway
Royal Leamington Spa
Warwickshire CV31 3J R
United Kingdom
Product name:
Portable Combustion Analyzer (PCA)
conforms to the following CE
requirements:
EN 50081-1, J anuary 1992 (Emissions)
EN 50082-1, J anuary 1992 (Immunity)
Equ ation s
Where:
21− O2
2
max
×
C2 = CO
O2 = Measured Oxygen in percent
TA = Measured stack temperature in °C
21
TL = Measured primary air temperature in °C
CO = Measured Carbon Monoxide in ppm
NO = Measured Nitric Oxide in ppm
C2 = Calculated Carbon Dioxide in percent
qA = Calculated stack loss in percent
LA = Calculated Lambda
A2
qA = (TA − TL)×
+ B
(21− O2
)
21
LA =
CU = Calculated Carbon Monoxide referenced
to 0% Oxygen
21− O2
NU = Calculated Nitric Oxide referenced to 0%
Oxygen
CO2max, A2, and B are constants (see below)*:
21
=
×
×
CU CO
21− O2
Fuel
CO2max
A2
B
NG
KOKS
LEG
11.8
10.2
13.1
13.8
15.4
15.9
18.7
20.4
0.66
0.60
0.63
0.63
0.68
0.68
0.60
0.70
0.009
0.011
0.011
0.008
0.007
0.007
0.007
0.012
21
=
NU NO
21− O2
LPG
Oil#2
Oil#6
P-Coal
Biofuel
* Some constants vary for Danish, Italian, and Polish fuels
2-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Setup
3.0 SETTING UP THE PCA
3.1 Scope
Before using the PCA, you MUST:
• Check the batteries or plug in an optional Power Supply (Section 3.2)
• Connect the probe to the analyzer (Section 3.3)
• Check the analyzer’s configuration (Section 3.4)
3.2 Power
3.2.1 Checking and Replacing the Batteries
A fresh set of batteries is supplied with the PCA. Install the batteries as
described below. Check for a sufficient charge prior to each use. If a
LOW BATTERYmessage is displayed, replace the batteries.
1. Remove the battery cover from the back of the PCA (Figure 3-1).
2. Remove (and properly dispose of) the old batteries.
3. Install a new set of four AA alkaline batteries, making sure to prop-
erly orient them as indicated by the “+” and “–” terminals in the
battery compartment.
4. Replace the battery cover.
Figure 3-1. Battery Replacement
Instruction 24-9351
3-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Setup
PCA
3.2.2 Using the Optional Power Supply
If an Optional Power Supply is to be used:
1. Connect the output plug of the Optional Power Supply to the
analyzer’s power supply jack (Figure 3-2).
2. Plug the Optional Power Supply into an appropriate AC wall outlet.
The analyzer will now operate and function normally.
3.3 Connecting the Probe
To attach the probe to the analyzer (Figure 3-2):
1. Push the yellow-banded, quick-connect Flue Gas Hose (giving a slight
twist) onto the analyzer’s GAS sample-inlet fitting.
2. Push the blue-banded, quick-connect Draft Hose (giving a slight twist)
onto the analyzer’s DRAFT sample-inlet fitting.
3. Push the Flue Gas Thermocouple into the T-STACK jack (connector
fits in only one way).
NOTE : The PCA has a built in room-air thermocouple.
Perform Step 4 only if the Optional Room Air/ Primary Air
Thermocouple is used.
4. Push the Optional Room Air/Primary Air Thermocouple into the
T-AIR jack (connector fits in only one way).
NOTE : In order for the PCA to correctly calculate combus-
tion efficiency when the burner’s primary-air temperature is
not the same as room temperature, the primary-air tempera-
ture should be measured using the optional Primary Air
Thermocouple.
Inspect all the hoses for cracks. If any hose is defective, replace the entire
probe assembly. Check that the water trap is dry and the filter is not
dirty or saturated with water.
3-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Setup
Reset Button
Room Air /
Primary Air
Thermocouple
(Optional)
Power Supply
110V/60Hz
230V/50Hz
(Optional)
Pressure Reference
Port (Used in the
Measurement of
Differential Pressure)
Draft Hose
Flue Gas
Thermocouple
Flue Gas
Hose
Probe Tube
Adjustable
Probe Stop
Water
Trap / Filter
Assembly
Probe
Handle
Flue
Gas
Hose
Figure 3-2. Connecting the Probe to the Analyzer
Instruction 24-9351
3-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Setup
PCA
3.4 Configuring the PCA
The PCA is configured at the factory for the parameters shown below, but
can be changed by following the instructions in the associated sections.
F u n ct ion
P a r a m et er s
To Ch a n ge,
R efer t o . . .
Fuel
Natural Gas
ºC
MB
English
CO & NX
HR:MIN:SEC
DD.MM.YY
IrDA
Section 4.8
Section 4.13
Section 4.14
Section 4.15
Section 4.16
Section 4.17
Section 4.17
Section 4.18
Temperature
Optional Draft
Language
Display Mode*
Time
Date**
Printer
*
Available only on the PCA 30, 35, 60, & 65
** The year displays as two digits on the instrument, and
four digits on the printout.
3-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.0 OPERATION
4.1 Key Pad Functions
Descriptions of the key pad functions are given below. Note that most of
the front panel key pad buttons perform multiple functions as determined
by what screen is being displayed at the time.
Turns the analyzer ON and OFF. Note that there is a 5 second
delay before the instrument actually turns OFF, thus allowing an
operator to turn the instrument back ON by pressing the
to prevent the accidental loss of test data.
key
Moves the cursor [z] in front of a menu item up through the
displayed items. This key also increases alphanumeric values in
screens requiring a value change.
Moves the cursor [z] in front of a menu item down through the
displayed items. This key also decreases alphanumeric values in
screens requiring a value change.
Chooses the highlighted item (the item with the cursor [z] in front
of it) in all menus and screens. This key also causes the cursor to
enter the number field in the Maintenance Password Screen, and
causes the cursor to advance to the next field position in screens
requiring multiple alphanumeric entries.
Starts and stops a combustion test when the Combustion Test
Screen is displayed. Pressing this key in any other screen almost
always returns the instrument to the Combustion Test Screen.
However, there are four situations where this key behaves as an
enter key: 1) After entering a correct password in the Password
Screen, press the
2) After entering an offset or span value in any of the Calibration
Edit Screens, press the key to store the new values. 3) After
entering a time or date value in the Time/Date Setup Screen,
press the key to store the new values and return the cursor to
the left side of the display. 4) After entering text in either the ID
Setup or User Name Screens, press the key to store the text.
key to display the first calibration screen.
Advances the display to the next menu screen.
Turns the backlight ON and OFF.
4-1
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.2 Sampling Hole Location
The analyzer requires that a 13 mm (½ in.) diameter sampling hole be
made in the furnace stack to accommodate the probe stop on the Probe
and Hose Assembly.
Locate the sampling hole downstream from the last heat exchanger,
and upstream from any source of dilution, such as a draft diverter
(Figure 4-1).
IMP OR TANT! As the distance between the last heat ex-
changer and sampling point increases, stack loss will falsely
decrease due to heat loss by convection from the flue or stack.
For residential and light-commercial combustion-equipment applications,
the following recommendations are applicable:
• Oil Gun Burners – Locate sampling
hole at least 30 cm (12 in.) down-
stream from the furnace breaching,
and at least 15 cm (6 in.) upstream
from the furnace side of the draft
regulator.
FLUE GAS FLOW
(Downstream)
DRAFT
DIVERTER/
REGULATOR
• Gas Burners – Locate sampling hole
at least 15 cm (6 in.) upstream from
the furnace side of the draft diverter
on gas-converted units. For gas-
designed equipment, the probe may
be inserted down into the flue
SAMPLE
POINT
through the draft diverter or hood.
(Upstream)
FURNACE
BREECHING
Figure 4-1. Sampling Hole Location
4-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.3 Combustion Test
IMP OR TANT! Large rapid changes in the temperature of
the analyzer can affect its accuracy. This is important to
know if the analyzer is stored in a cold place (such as an
unheated vehicle in the winter) and then taken into a warm
furnace area. For the most accurate test results, allow the
analyzer to warm up to room temperature before use (about
10 minutes).
4.3.1 Analyzer Turn On and Warm Up
IMP OR TANT! Be sure the probe is at room temperature
before performing the following steps.
1. Make sure that the analyzer is properly set up per Section 3.0.
2. Place probe in an area of fresh, ambient air; then press the analyzer’s
key.
3. Wait for the analyzer to countdown through its 60 second warmup
period; then perform one of the following:
•
•
If n o er r or s w er e d etected d u r in g w a r m u p , the Combustion
Test Screen will be displayed. Skip Step 4, and go to Section 4.3.2.
If a n er r or w a s d etected d u r in g w a r m u p , proceed with Step 4.
4. If one or more errors were detected by the microprocessor during
warmup, these errors will be displayed at the bottom of the Sensor
Status Screen. Address any problems now per Section 7.2; then repeat
this procedure starting with Step 1.
NOTE : If the error detected is not critical to your test, the
instrument can still perform any test not using the function
disabled by the error.
4-3
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.3.2 Installing Probe in the Stack
1. After making a sampling hole in the stack (Section 4.2), and turning
on the analyzer (Section 4.3.1), screw the probe stop supplied with the
Probe and Hose Assembly into the sampling hole (Figure 4-2).
2. Insert the probe through the hole in the probe stop, then position the
probe tip inside the stack, near its center. Tighten the thumbscrew on
the probe stop to secure the probe.
PROBE STOP
THUMBSCREW
Figure 4-2. Installing the Probe
4-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.3.3 Performing a Combustion Test
IMP OR TANT: If the burner’s primary-air temperature is
not the same as the room temperature, then be sure the
Optional Room Air / Primary Air Thermocouple is installed
per Section 3.3.
1. With the Combustion Test Screen displayed and the probe installed in
the stack, press the
tion 4.7).
key to start a combustion test (refer to Sec-
2. Once all sensor readings are indicated on the screen: A) Loosen the
thumbscrew on the probe stop. B) Move the probe in and out of the
stack until the stack’s core temperature (hot spot) is located.
C) Tighten the thumbscrew to prevent further movement of the probe.
Locating the highest stack temperature is very important for accurate
combustion calculations.
3. You can now begin burner-service procedures. The readings on the
analyzer change quickly to show changes in burner performance.
CAUTION
With the Water Trap / Filter Assembly stood up on its Outlet
End, do not let water condensate build up beyond the tip of
the riser tube. The sensors could be damaged if water would
enter the analyzer. Drain the water condensate after every
combustion test (refer to Section 6.4).
4. Pressing the
key will save the Combustion Test Screen readings
while a test is in progress. Moving the cursor (z) in front of the print
( P) function using the s key, and then pressing
will print the test
information to an optional printer.
4.3.4 Ending a Combustion Test
1. Press the
key to end a combustion test.
WAR NING!
Burn hazard! Allow a hot probe to cool for about 5 minutes
before handling.
4-5
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
CAUTION:
Do not place a hot probe inside the instrument’s carrying case.
Allow the probe to cool before storage.
2. Loosen the thumbscrew on the probe stop; then remove the probe and
probe stop from the stack.
3. If data was saved during the combustion test, you can turn off the
analyzer and review or print the stored data at a later time as de-
scribed in Sections 4.10 and 4.23.
4.3.5 Turning Off the Analyzer and
Purging the CO Sensor
Turn off the analyzer by pressing the key.
If the key is pressed while the CO reading is 100 ppm or higher, the
pump will automatically turn on (if not already running) to purge the
analyzer of CO.
IMP ORTANT! The analyzer’s probe must be removed from
the stack during the purging process to allow fresh air to be
drawn through the analyzer.
The following message is displayed while the analyzer is being purged.
PURGING CO SENSOR
As soon as the CO level falls below 100 ppm, the pump turns off and the
analyzer starts its normal 5 second turn-off sequence.
To abort the purging process and immediately start the analyzer’s turn-off
sequence, press the key.
NOTE : Turning the analyzer off initiates a 5-second delay,
during which time the unit can be turned on again without
any warmup time. You can turn the analyzer back on during
this 5-second delay by pressing the
key.
4-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.4 Differential Pressure Measurement
The difference in pressure (∆P) between two areas can be measured by
using the PCA’s two pressure ports and DRAFT Screen. By using Pres-
sure Port 2 (–) as the reference, the pressure applied to Port 1 (+) will be
displayed on the DRAFT Screen as the differential pressure between the
two ports.
1. Turn on the analyzer by pressing the key; wait for the warmup cycle
to complete; then press the MENU key until the first DRAFT Screen is
displayed (refer to Section 4.9). If a TA-SENSOR ERROR is displayed
because the probe’s thermocouple is not plugged into the analyzer, then
press the
key to acknowledge the error before pressing the MENU
key.
2. While the first DRAFT Screen is displayed, remove any hoses con-
nected to Pressure Ports 1 and 2; then press the
ports at atmospheric pressure.
key to zero these
3. Connect two sampling hoses to
DRAFT
– 2.25 MB
HOT SPOT ---- °C P
«S
Pressure Ports 1 and 2
(Figure 4-3). Then place the open
end of each hose into the areas
being measured.
DRAFT
4. The differential pressure between the two areas is now displayed on
the third DRAFT Screen. If the pressure at Port 1 is higher than
Port 2, then the pressure difference will be positive. But if the pres-
sure at Port #1 is lower, then the pressure difference will be negative.
The reading shown in this example indicates that the pressure at
Port 1 is 2.25 mb lower than the pressure at Port 2.
Sampling Hoses with quick-
connect fittings are available
as an optional accessory.
Refer to Section 8.2.
Pressure Port 2(–)
Pressure Port 1(+)
Reference
Figure 4-3. Differential Pressure Hose Connections
4-7
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.5 Warmup Screen
BACHARACH, INC.
PCA xx
WARMUP yy
Where: xx = Instrument Model Number
yy = Counts down from 60 seconds
As soon as the key is pressed, the instrument’s serial number and
software version number are displayed for approximately 3 seconds. To
continuously display these items, hold down the key at start-up. The
warmup cycle continues after the key is released.
The Warmup Screen is displayed during the analyzer’s 60 second warmup
cycle, during which time the “Warmup” value (yy) counts down to zero.
After the warmup cycle is complete (and if the unit is working correctly)
the instrument will flash NO ERRORS DETECTEDand go directly to
the Combustion Test Screen (Section 4.7). If there is a problem, however,
with one or more of the sensors, the Sensor Status Screen (Section 4.6) is
displayed with error message(s) appearing at the bottom of the screen.
Front Panel Key Functions:
– No Action
– No Action
– No Action
– No Action
– No Action
– Toggle Backlight ON/OFF*
– Turn analyzer OFF*
* The
button will always turn the backlight on and off, and the key will
always turn the analyzer on and off. These two keys will not be mentioned
in the remainder of this section.
4-8
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.6 Sensor Status Screen
BACHARACH, INC.
PCA xx
WARMUP 0
z
Where: xx = Instrument Model Number
z = Sensor(s) in error
If there is problem with one or more of the sensors, the Sensor Status
Screen will be displayed after the analyzer has gone through its warmup
cycle (refer to Section 7.2 for a listing of the error codes).
Front Panel Key Functions:
– No Action
– No Action
– No Action
– Go to Combustion Test Screen
– No Action
4-9
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.7 Combustion Test Screen
PCA models 10–25, 40–55
PCA models 30, 35, 60, & 65
O2 4.0 CO 12 HLD
C2 9.5 CU 15 NG
O2 4.0 CO 12 HLD
C2 9.5 NX
TL 20.0 TA 190
TL 20.0 TA 190
P
P
qA 8.1 LA 1.24 «S
qA 8.1 LA 1.24 «S
- OR -
This screen shows:
O2 4.0 CU 15 HLD
C2 9.5 NU 12 NG
TL 20.0 TA 190
O2 ............... Oxygen content in flue gas (%)
C2 ............... Carbon Dioxide content
present in flue gas (%)
TL ............... Primary/Ambient air temp. (°F)
qA ............... Stack Loss
P
qA 8.1 LA 1.24 «S
CO* ............. Carbon Monoxide content in flue gas (ppm)
CU* ............. Carbon Monoxide content referenced to 0% Oxygen (ppm)
TA ............... Stack (Flue gas) temperature (°F)
LA ............... LAMBDA
NX* ............. Nitric Oxide content in flue gas (ppm)
NU* ............ Nitric Oxide content referenced to 0% Oxygen (ppm)
HLD/RUN .. PCA on hold / PCA running test
NG .............. Fuel code for natural gas (see Section 4.8 for other codes)
P .................. Print Data
S .................. Save Data
* For PCA models 30, 35, 60 and 65, you have the option of displaying either
CO & NX, or CU & NU. Refer to Section 4.16 for setup instructions.
NOTE: Refer to Section 7.3 if stars (****) , dashes (----) , or
Xs (XXXX) appear in the display.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Save or Print screen data
– Run test / Stop test
– Go to Fuel Selection Screen
4-10
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.8 Fuel Selection Screen
«NATGAS
KOKS
LEG
FUEL
OIL NO.2
OIL NO.6
P-COAL
LPG
This screen is displayed by pressing the MENU key from the Combustion
Test Screen. This screen is used to select the fuel being burned.
To select a fuel, first use the stkeys to move the cursor (z) in front of the
desired fuel, and then press the
key.
NOTE : The fuel selected is saved as the default, and re-
mains in memory after the PCA is turned off.
The fuel codes as displayed in the Combustion Test Screen:
NG = Natural Gas
KOK = Coal Gas
LPG = Propane
O#2 = Oil No. 2
O#6 = Oil No. 6
PC = P-Coal (English, German, Dutch,
French, Italian, Polish, and Spanish) or Biofuel
(Danish, Finnish, and Swedish)
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select Fuel
– Go to Combustion Test Screen
– Go to Draft Screens
4-11
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.9 Draft Screens
The first Draft Screen is displayed by
repeatedly pressing the MENU key
from the Combustion Test Screen.
DRAFT
DISCONNECT DRAFT
HOSE
PRESS ±
To measure draft, first zero the
analyzer’s pressure sensor to atmo-
spheric pressure by disconnecting the
draft hose from the bottom of the
instrument, and then pressing the
key. Reconnect the draft hose after
the second Draft Screen appears
(shown for 3 seconds). The third
screen shows the current values of
draft and stack temperature as
measured by the analyzer.
DRAFT
RECONNECT DRAFT
HOSE
DRAFT
– 0.25 XX
DRAFT
HOT SPOT
190 °C P
«S
Where: xx = Unit of measure.
Default is millibars (MB). See
Optional Draft SETUP Screen
(Section 4.14) for other choices.
When using the analyzer to make a
differential pressure measurement
(Section 4.4), the differential pressure
value will be displayed on the third
Draft Screen.
To save (S) or print (P) the screen
data, first use the st keys to move
the cursor (z) in front of the desired
function, and then press the
key.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Save or Print screen data
– Go to Combustion Test Screen
– Go to Memory Directory Screen
4-12
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.10 Memory Directory Screen
‘Standard’ PCA Screen
‘Advanced’ PCA Screen
MEMORY DIRECTORY
MEMORY DIRECTORY
«M8 28.7.97 15:45
M9 MEMORY EMPTY
CLEAR MEMORY
«98 28.7.97 15:45
99 MEMORY EMPTY
CLEAR MEMORY
The Memory Directory Screen is displayed by repeatedly pressing the
MENU key from the Combustion Test Screen. This screen is used to select
a memory location that contains saved data which an operator can re-
view.
NOTE: A ‘standard’ PCA has 10 memory locations num-
bered M0 thru M9, while an ‘advanced’ PCA has 100 memory
locations numbered 0 thru 99.
To select a data-memory location, first use the st keys to move the
cursor (z) in front of the desired memory location; then press the
key.
The saved data is now displayed in either the Combustion Test Screen or
Draft Screen, depending on whether the chosen memory location contains
combustion or draft information. To print the saved data, refer to Sec-
tion 4.23.
After viewing or printing the saved data, use the st keys to move the
cursor (z) to the exit (E) function; then press
. This will redisplay the
memory directory.
Selecting the CLEAR MEMORYfunction displays the Clear Memory
Screen from where all saved data can be erased (refer to Section 4.24).
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Display the data saved at the chosen memory location
– Go to Combustion Test Screen
– Go to Temperature Setup Screen, or the
Memory To PC Screen for Advanced units
4-13
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.11 Memory to PC Screen
(For ‘Advanced’ PCA Models 40, 45, 50, 55, 60 & 65)
MEMORY TO PC
«TRANSMIT DATA
CLEAR MEMORY
The Memory To PC Screen is displayed by repeatedly pressing the MENU key
from the Combustion Test Screen. Use this screen to either transmit all
stored memory locations to a computer, or clear all memory locations.
TRANSMIT DATA
Before data can be transmitted to a personal computer, the PCA’s RS-232
output must first be connected to an unused COM port on the computer
using serial data cable Part No. 24-1073 (see Figure 4-4). Also, a commu-
nications program (i.e., ProcommPlus®, Windows 3.x Terminal, or Win-
dows 9x Hyper Terminal) must be installed, and its communications
parameters configured for: 9600 ba u d , 8 d a ta bits, 1 stop bit, n o p a r ity,
a n d n o h a n d sh a k in g.
A
C
u
t
t
a
c
h
D
B
9
C
o
n
n
e
c
t
o
r
t
o
c
o
m
p
u
t
e
r
'
s
O
M
1
o
r
C
O
M
2
p
o
r
t
.
I
f
n
e
c
e
s
s
a
r
y
,
s
e
a
2
5
-
p
i
n
t
o
9
-
p
i
n
a
d
a
p
t
e
r
.
S
P
e
a
r
r
i
t
a
l
D
o
a
t
a
C
a
b
l
3
e
N
.
2
4
-
1
0
7
R
S
-
2
3
2
C
o
n
n
e
c
t
o
r
Figure 4-4. Connecting the Serial Data Cable
4-14
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
Data is transmitted to a computer in ASCII comma-delimited format,
which can be captured as a text file and then opened in most commer-
cially available spreadsheet programs. Note that each data record con-
sists of 20 fields, some of which may be blank for different tests and PCA
models as listed in Tables 4-1 & 4-2.
Use the communication software to capture and save the received data as
an ASCII text file. Consult the software’s documentation for detailed
instructions on how to perform this procedure.
To start transmitting data, first use the st keys to position the cursor (z)
in front of TRANSMIT DATAand then press the
key. Observe that
as PCA downloads its data, the word TRANSMITTINGꢀappears on the
display.
CLEAR MEMORY
To clear all memory locations, first use the st keys to position the cursor
(z) in front of CLEAR MEMORYand then press the
key. The Clear
Memory Screen will then appear from where all saved data can be erased
(refer to Section 4.24).
Front Panel Key Functions:
– Toggle cursor (z) position
– Toggle cursor (z) position
– Select function next to cursor
– Go to Combustion Test Screen
– Go to ID Setup Screen
4-15
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
TABLE 4-1. COMMA-DE LIMITE D F IE LDS
Field
Data Name or Value
Label in Column
Headings
1
2
3
4
Instrument serial number
ID line 1 (up to 16 characters)
ID line 2 (up to 16 characters)
ID line 3 (up to 16 characters)
Time of test (hh:mm:ss)
Date of test (dd.mm.yyyy)
Name of fuel (up to 16 characters)
Flue gas temperature
SN
ID1
ID2
ID3
5
TIME
DATE
FUEL
TA
6
71
81
91
Air temperature
TL
101
111
121
131,2
141,2
151,3
161,3
171
181
19
20
Temperature unit of measure (F or C)
O2 concentration in %
C/F
O2
C2
CO
CU
NX
NU
qA
CO2 concentration in %
CO concentration in ppm
CO referenced to 0% O2 in ppm
NO concentration in ppm
NO referenced to 0% O2 in ppm
qA in %
Lambda
Draft measurement
Draft unit of measure
LA
DR
MB/PA/WC
1
Empty data field for draft tests
Empty data field for PCA Models 40 and 45
Empty data field for PCA Models 40, 45, 50, and 55
2
3
TABLE 4-2. TYP ICAL SP R E ADSH E E T F OR A P CA 65
SN
ID1
ID2
ID3
TIME
DATE
FUEL
TA TL C/F O2
AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:03:27 19.01.1999 NATGAS 190 20
AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:10:35 19.01.1999
C
4
C2
CO
CU
NX
NU
qA
LA
DR
MB/PA/WC
9.5
12
15
10
12
8.1
1.24
-0.25 MB
-0.25 MB
Line 1: Column Headings
Line 2: Typical Combustion Readings
Line 3: Typical Draft Reading
4-16
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.12 ID Setup Screens
(For ‘Advanced’ PCA Models 40, 45, 50, 55, 60 & 65)
SETUP
«ID #1
ID #2
ID #3
This initial ID Setup Screen is displayed by repeatedly pressing the MENU
key from the Combustion Test Screen. Use this screen to edit three lines
of customer information (i.e., the customer’s name, location, and burner
reference number).
Each ID line can be up to 16 alphanumerical characters in length. All
three lines will appear at the top of each test record for the purpose of
identifying individual tests.
Front Panel Key Functions for the Initial ID SETUP Screen:
– Move cursor (z) upward
– Move cursor (z) downward
– Select ID Number that is next to the cursor for editing
– Go to Combustion Test Screen
– Go to Temperature Setup Screen
To enter a line of text, first use the st keys to position the cursor (z) in
front of the desired ID line; then press
Screen will then appear.
. The selected ID Line Number
SETUP
ID #1
«
Now press the st keys until the desired letter or number is displayed.
Available characters include:
“(space)ABCDEFGHIJKLMNOPQRSTUVWXYZüäöß0123456789”
4-17
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
Press
to save the selected character and advance to the next position.
until the cursor is over the incorrect
character and make your correction by again using the st keys. After all
If you make a mistake, press
the desired characters have been selected, press the
text line and return to the initial ID SETUP Screen.
key to save the
NOTE: The entered ID information will be saved with all
future memory records until it is modified or deleted.
Front Panel Key Functions for the Individual ID SETUP Screens:
– Increment character
– Decrement character
– Select the displayed character and advance to the next
character position
– Save the text line and return to the initial ID SETUP Screen
– Abort any changes to the text line and return to the initial
ID SETUP Screen
4-18
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.13 Temperature Setup Screen
SETUP
TEMPERATURE UNIT «°C
°F
The Temperature Setup Screen is displayed by repeatedly pressing the
MENU key from the Combustion Test Screen. Use this screen to setup the
analyzer to display temperature in either °C or °F.
To select the instrument’s temperature unit-of-measure, first use the st
keys to move the cursor (z) in front of °Cor °F, and then press the
key.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select temperature unit of measure next to cursor
– Go to Combustion Test Screen
– Go to Draft Unit Setup Screen
4-19
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.14 Draft Unit Setup Screen
SETUP
DRAFT UNIT
«MB
PA
WC
The Draft Unit Setup Screen is displayed by repeatedly pressing the
MENU key from the Combustion Test Screen. Use this screen to setup the
analyzer to display draft in either millibars (MB), Pascals (PA), or inches-
of-water column (WC).
To select the draft unit-of-measure, first use the st keys to move the
cursor (z) in front of MB, PAor WC, and then press the
key.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select draft unit of measure
– Go to Combustion Test Screen
– Go to Language Setup Screen
4-20
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.15 Language Setup Screen
SETUP
SETUP
LANGUAGE DAN NED
DEU SVE
LANGUAGE DEU FRA
«ENG ITA
- OR -
«ENG FIN
ESP POL
The Language Setup Screen is displayed by repeatedly pressing the MENU
key from the Combustion Test Screen. Use this screen to select the
language displayed on the analyzer. The languages available for selection
include: Danish, German, English, Dutch, Swedish and Finnish; or
German, English, Spanish, French, Italian and Polish. The languages
displayed depend on the model of the analyzer (refer to Table 1-1).
To select a language, first use the st keys to move the cursor (z) in front
of the desired language, and then press the
key.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select Language
– Go to Combustion Test Screen
– Go to Display Mode Setup Screen
4-21
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.16 Display Mode Setup Screen
(For PCA Models 30, 35, 60 & 65)
SETUP
DISPLAY
«CO NX
CU NU
The Display Setup Screen is displayed by repeatedly pressing the MENU
key from the Combustion Test Screen. Use this screen to select whether
the Combustion Test Screen will display the measured values of Carbon
Monoxide and Nitric Oxide (CO and NX), or the calculated values of these
gases (CU and NU) referenced to 0% Oxygen.
To setup the display, first use the st keys to move the cursor (z) in front
of either CO NX(measured values), or CU NU(calculated values), and
then press the
key.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select Display Mode
– Go to Combustion Test Screen
– Go to Time/Date Setup Screen
4-22
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.17 Time/Date Setup Screen
SETUP
«TIME 15:45:06
DATE 31.10.01
The Time/Date Setup Screen is displayed by repeatedly pressing the
MENU key from the Combustion Test Screen. Use this screen to enter the
current time and date.
To enter the correct time or date, first use the st keys to move the
cursor (z) in front of the function you wish to change. Each position in the
TIMEor DATEnumber fields can then be changed by first pressing the
key to move the cursor into the desired position, then pressing the st
keys to increase or decrease the value. Pressing
moves the cursor to
the next position in the number field. Each individual position can be
edited in the same manner. Once the time or date values have been
entered, press
of the screen.
to save the values and return the cursor to the left side
Front Panel Key Functions:
– Move cursor (z) up, or Increase value in number fields
– Move cursor (z) down, or Decrease value in number fields
– Select Time or Date to be changed, or move cursor (z) to next
position in the number field
– Go to the Combustion Test Screen, or save the time and date
values and return the cursor to the left side of the display
– Go to Printer Setup Screen
4-23
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.18 Printer Setup Screen
SETUP
«IR - HP
IR - IRDA
RS232
PRINTER
The Printer Setup Screen is displayed by repeatedly pressing the MENU
key from the Combustion Test Screen. Use this screen to choose the type
of connection and printer being used.
IR-HP:
Infrared connection to a printer manufactured by Hewlett
Packard, which uses their proprietary infrared communica-
tions protocol
IR-IRDA: Infrared connection to a printer that uses a standard IrDA
protocol
RS232:
Cable connection between the PCA and any serial printer
capable of 9600 baud operation
Use the st keys to move the cursor (z) in front of the desired connection
and printer, and then press the
to the Combustion Test Screen.
key to make the selection and return
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select connection and printer next to cursor
– Go to Combustion Test Screen
– Go to Maintenance Password Screen
4-24
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.19 Maintenance Password Screen
MAINTENANCE
«PASSWORD
XXX
Where: xxx = Password number
The Maintenance Password Screen is displayed by repeatedly pressing
the MENU key from the Combustion Test Screen. From here a three-digit
password must be entered to access the instrument’s Maintenance
Screens. The password number is provided on the Portable Combustion
Analyzer Calibration Password card that was supplied with the analyzer.
To enter the password, first press the
the first number field, and then press the st keys until the first digit of
the password is displayed. Press to advance to the next number field
and enter the second digit. Perform the same sequence a third time to
complete the password. Press the key after the correct password is
key to move the cursor (z) into
entered to display the Maintenance Screen.
Front Panel Key Functions:
– No action, or increase value in password number field
– No action, or decrease value in password number field
– Move cursor (z) to next position in password number field
– Go to Combustion Test Screen (if cursor is on left side of
screen), or go to Maintenance Screen (if the proper password
was entered), or return cursor to left side of screen (if the
wrong password was entered)
– Go to Combustion Test Screen (if cursor is on left side of
screen), or return cursor to left side of screen (if cursor is in
the password number field)
4-25
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
4.20 Maintenance Screen
MAINTENANCE
«CALIBRATION
USER NAME
The Maintenance Screen is displayed after entering the correct password
in the Maintenance Password Screen (Section 4.19). Use this screen to
enter either the analyzer’s Calibration Screen or User Name Screen.
To enter the Calibration Screen, first use the st keys to position the
cursor (z) in front of CALIBRATION, and then press the
key.
NOTE : Section 5.0 contains detailed calibration procedures.
To enter the User Name Screen, first use the st keys to position the
cursor (z) in front of USER NAME, and then press the
key.
Front Panel Key Functions:
– Toggle cursor (z) position
– Toggle cursor (z) position
– Select function next to cursor
– Go to Combustion Test Screen
– No action
4-26
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.21 User Name Screens
USER NAME
«LINE 1
LINE 2
LINE 3
This initial User Name Screen is displayed after selecting USER NAME
from the Maintenance Screen (Section 4.20). Use this screen to either
enter or edit three lines of user-name information.
Each user-name line can be up to 20 alphanumerical characters in length.
All three lines will appear at the top of each printout for the purpose of
identifying the user or owner of the instrument (i.e., your company’s
name and address).
Front Panel Key Functions for Initial User Name Screen:
– Move cursor (z) upward
– Move cursor (z) downward
– Select Line Number that is next to the cursor for editing
– Go to Combustion Test Screen
– Return to Maintenance Screen
To enter text, first use the st keys to position the cursor (z) in front of
the desired line number; then press
Number Screen will then appear.
. The selected User Name Line
USER NAME
LINE 1
«
Now press the st keys until the desired letter or number is displayed.
Available characters include:
“(space)ABCDEFGHIJKLMNOPQRSTUVWXYZüäöß0123456789”
4-27
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
Press
to save the selected character and advance to the next position.
If you make a mistake, press
until the cursor is over the wrong charac-
ter and make your correction by again using the st keys.
After all the desired characters have been selected, press
text line and return to the initial User Name Screen.
to save the
Front Panel Key Functions for Individual User Name Screens:
– Increment character
– Decrement character
– Select the displayed character and advance to the next
character position
– Save the text line and return to the initial User Name Screen
– Abort any changes to the text line and return to the initial
User Name Screen
4.22 Saving Test Data
O2 4.0 CO 12 HLD
C2 9.5 CU 15 NG DRAFT
DRAFT
– 0.25 MB
TL 20.0 TA 190 P HOT SPOT 190 °C P
qA 8.1 LA 1.24 «S
«S
To save the data displayed in either the Combustion Test or Draft
Screens, first use the st keys to move the cursor (z) in front of the save
(S) function and then press the key. The data will be saved in
memory, and can be recalled at any time from the Memory Directory
Screen (Section 4.10).
NOTE: Data will be automatically stored in the next free
memory location. After all memory locations are filled, any
additional data that is saved will start overwriting previ-
ously saved data starting at the first memory location.
4-28
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.23 Printing Test Data
O2 4.0 CO 12HLD
C2 9.5 CU 15 NG
TL 20.0 TA 190 «P
qA 8.1 LA 1.24 S
DRAFT
– 0.25 MB
HOT SPOT 190 °C «P
S
DRAFT
Before printing, ensure that the correct
connection and printer has been selected per
Section 4.18.
The Print function is available in either the
Combustion Test Screen or the Draft Screen*.
NOTE : The data which is stored in
memory can also be printed. First go to
the Memory Directory Screen (Section
4.10) and display the data to be printed;
then print the data as described below.
0
I
0
When using an infrared printer:
4
5
c
m
(
1
8
i
n
.
)
6
0
°
m
a
x
.
1. Place analyzer in-line with the printer’s
IR input (see Figure 4-5).
m
a
x
.
2. Use the st keys to move the cursor (z) in
front of the print (P) function.
3. Press the
key to start printing.
M
E
N
U
P
C
A
* The HOT SPOT line shown in the Draft
Screen does not appear on the printout.
Figure 4-5. Aligning the Printer
4-29
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
When using a serial printer:
1. First connect the analyzer to the printer using the optional RS-232
cable (see Figure 4-6).
2. Set the printer’s communication parameters to 9600 baud, 8 data bits,
1 stop bit, no parity, and no handshaking.
3. Use the st keys to move the cursor (z) in front of the print (P)
function.
4. Press the
key to start printing.
A
s
u
t
t
a
c
h
D
B
9
C
o
n
n
e
c
t
o
r
t
o
p
r
i
n
t
e
r
'
s
e
r
e
i
a
l
i
n
p
u
t
p
o
r
t
.
I
f
n
e
c
e
s
s
a
r
y
,
s
a
2
5
-
p
i
n
t
o
9
-
p
i
n
a
d
a
p
t
e
r
.
S
P
e
a
r
r
i
t
a
l
D
o
a
t
a
C
a
b
l
3
e
N
.
2
4
-
1
0
7
R
S
-
2
3
2
C
o
n
n
e
c
t
o
r
Figure 4-6. Connecting a Serial Printer to the Analyzer
4-30
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Operation
4.24 Clear Memory Screen
CLEAR MEMORY
C
«E
The Clear Memory Screen is accessed from either the Memory Directory
Screen (Section 4.10) or the Memory to PC Screen (Section 4.11).
To clear a ll memory locations, use the st keys to place the cursor (z) in
front of the clear (C) function, and then press the
key.
To return to the previous screen without clearing any memory locations, use
the st keys to place the cursor (z) in front of the exit (E) function; then
press
.
4.25 Resetting the Microprocessor
If the analyzer “locks-up” and cannot be turned OFF, reset the micropro-
cessor by pressing the RESET button (Figure 4-7). The button can be
activated using the end of a paper clip.
RESET
BUTTON
Figure 4-7. Reset Button
4-31
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation
PCA
NOTES:
4-32
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.0 CALIBRATION
NOTE : Bacharach recommends that the PCA be calibrated
by your nearest Bacharach Service Center. Calibration,
however, can be performed in the field if your facility has the
necessary equipment and qualified personnel to perform the
procedures described in the sections that follow.
IMP OR TANT! To prevent the loss of data during calibra-
tion, perform the following procedures with fresh batteries, or
using an optional Power Supply (see Section 3.2).
5.1 Sensor Check
IMP OR TANT! Before turning on the analyzer or perform-
ing any of the calibration procedures, ensure that the ana-
lyzer will be sampling fresh air, and that the probe is at
room temperature.
When the analyzer is first turned on and allowed to cycle through its
60 second warmup period, and while sampling fresh air, the sensors are
checked (read) and calibrated (set) to the following ambient conditions:
• Oxygen sensor is spanned to 20.9%
• Carbon Monoxide sensor (if installed) is zeroed
• Nitric Oxide sensor (if installed) is zeroed
• Pressure sensor (if installed) is zeroed
Instruction 24-9351
5-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
5.2 Calibration Fixtures
A gas and a draft fixture will be required to perform the various calibra-
tion procedures described in this manual.
Material Required:
• Calibration Kit (Refer to Section 8.2)
• Calibration Gas Cylinder (Refer to Section 8.2)
• Bellows
• Micromanometer
Procedure:
Assemble the appropriate fixture, shown in Figure 5-1, as required by the
calibration procedure being performed.
Leave Top
Port Open
GAS
FIXTURE
2
1
3
4
3
5
M
E
N
U
3
P
C
A
Parts Shown:
6
1. Gas Cylinder
2. Regulator*
3. Tubing*
DRAFT
FIXTURE
4. Tee*
5. Flowmeter*
6. Fitting, Gas*
7. Fitting, Draft*
* Contained in Calibration Kit
M
E
N
U
P
C
A
3
4
3
Bellows
7
Micromanometer
Figure 5-1. Calibration Fixtures
5-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.3 Calibrate Menu Screen
«TA-ZERO
TA-SPAN
TL-ZERO
TL-SPAN
CALIBRATE
NX
CO
DRAFT
The Calibrate Menu Screen is displayed after entering the correct pass-
word in the Maintenance Password Screen (Section 4.19) and selecting
CALIBRATION from the Maintenance Screen (Section 4.20). Use this
screen to select the sensor to be calibrated.
Press the st keys until the cursor (z) is in front of the desired function,
and then press the
key.
NOTE : TA is the stack temperature sensor, while TL is the
optional Room Air/ Primary Air temperature sensor.
NOTE : If a sensor is not installed (i.e., the instrument does
not have a Nitric Oxide sensor or thermocouple simulator
installed), the corresponding calibration screen will not be
displayed.
Front Panel Key Functions:
– Move cursor (z) up
– Move cursor (z) down
– Select sensor to be calibrated
– Go to Combustion Test Screen
– Return to Maintenance Screen
Instruction 24-9351
5-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
5.4 Calibrate TA-Zero
Material Required:
• Thermocouple Simulator (K-type) Range: 0 to 300°C
Accuracy: ±0.3°C
Procedure:
1. With the analyzer turned off, first plug the simulator's K-type connec-
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TA-ZERO to zero the analyzer’s stack temperature channel.
3. Adjust the simulator to 0° C (32° F).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the st and
keys to enter an APPLIED value that equals 0 °C
(32 °F).
Typical Calibrate TA-Zero Screen During Calibration Procedure:
CALIBRATE TA-ZERO
MEASURED
APPLIED
3.0°C
0000.0°C
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 0 – 5 °C (32 – 41°F). Any
attempt to calibrate outside this range will cause the analyzer to
display an error message.
5-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.5 Calibrate TA-Span
Material Required:
• Thermocouple Simulator (K-type) Range: 0 to 300°C
Accuracy: ±0.3°C
Procedure:
1. With the analyzer turned off, first plug the simulator's K-type connec-
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TA-SPAN to span the analyzer’s stack temperature channel.
3. Set the simulator to 300° C (572° F).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the st and
keys to enter an APPLIED value that equals 300 °C
(572 °F).
Typical Calibrate TA-Span Screen During Calibration Procedure:
CALIBRATE TA-SPAN
MEASURED
APPLIED
295.0°C
0300.0°C
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 270 – 330 °C (518 – 626 °F).
Any attempt to calibrate outside this range will cause the analyzer to
display an error message.
Instruction 24-9351
5-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
5.6 Calibrate TL-Zero
Material Required:
• Thermocouple Simulator (K-type) Range: 0 to 300°C
Accuracy: ±0.3°C
Procedure:
1. With the analyzer turned off, first plug the simulator’s K-type connec-
tor into the T-AIR jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TL-ZERO to zero the analyzer’s room-air/primary-air temperature
channel.
3. Set the simulator to 0 °C (32 °F).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the st and
keys to enter an APPLIED value that equals 0 °C
(32 °F).
Typical Calibrate TL-Zero Screen During Calibration Procedure:
CALIBRATE TL-ZERO
MEASURED
APPLIED
3.0°C
0000.0°C
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 0 – 5 °C (32 – 41 °F). Any
attempt to calibrate outside this range will cause the analyzer to
display an error message.
5-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.7 Calibrate TL-Span
Material Required:
• Thermocouple Simulator (K-type) Range: 0 to 300°C
Accuracy: ±0.3°C
Procedure:
1. With the analyzer turned off, first plug the simulator’s K-type connec-
tor into the T-AIR jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TL-SPAN to span the analyzer’s room-air/primary-air temperature
channel.
3. Set the simulator to 100 °C (212 °F).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the st and
keys to enter an APPLIED value that equals 100 °C
(212 °F).
Typical Calibrate TL-Span Screen During Calibration Procedure:
CALIBRATE TL-SPAN
MEASURED
APPLIED
102.0°C
0100.0°C
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 90 – 110 °C (194 – 230 °F).
Any attempt to calibrate outside this range will cause the analyzer to
display an error message.
Instruction 24-9351
5-7
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
5.8 Calibrate NX
(For PCA Models 30, 35, 60 & 65)
The Nitric Oxide sensor needs to be spanned at regular intervals to
determine that it still meets its accuracy specification. Because of the
toxicity of Nitric Oxide gas, however, unless your facility has the neces-
sary gas cylinders and personnel trained in the handling of toxic gases,
we recommend that the Nitric Oxide sensor be spanned by an authorized
Bacharach Service Center.
Material Required:
• Calibration Gas Fixture (Section 5.2)
• Gas Cylinder, 50 to 150 ppm Nitric Oxide with an analytical accuracy of
±1% (customer supplied)
Procedure:
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to
calibrate the NX sensor.
2. At the conclusion of Step 1 the pump should start running.
3. Using the Gas Fixture shown in Figure 5-1, attach the Nitric Oxide
calibration-gas cylinder to the analyzer’s GAS inlet.
4. Adjust the regulator of the calibration fixture for a flowmeter indica-
tion of approximately 2 SCFH.
5. Wait until the MEASURED reading on the screen stabilizes (approxi-
mately 3 minutes). Then use the st and
keys to enter an
APPLIED value* that equals the concentration that is stamped on the
NX calibration-gas cylinder.
Typical Calibrate NX Screen During Calibration,
Using 100 ppm Nitric Oxide Calibration Gas:
CALIBRATE NX
MEASURED
APPLIED
092 PPM
0100 PPM
6. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 50 – 150 ppm. Any attempt to
calibrate outside range will cause the analyzer to display an error
message.
5-8
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.9 Calibrate CO
(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)
Material Required:
• Calibration Gas Fixture (Section 5.2)
• Gas Cylinder, 500 ppm CO in air (Refer to Section 8.2)
• Gas Cylinder, CO (1000 ppm) and H2 (1000 ppm) in Nitrogen
(Refer to Section 8.2)
Procedure:
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to
calibrate the CO sensor.
2. At the conclusion of Step 1 the pump should start running.
3. Using the Gas Fixture shown in Figure 5-1, attach the CO calibration-
gas cylinder to the analyzer’s GAS inlet.
4. Adjust the regulator of the calibration fixture for a flowmeter indica-
tion of approximately 2 SCFH.
5. Wait until the MEASURED reading on the screen stabilizes (approxi-
mately 3 minutes). Then use the st and
keys to enter an
APPLIED value* that equals the concentration which is stamped on
the CO calibration-gas cylinder.
Typical CALIBRATE CO Screen During Calibration Procedure,
Using 500 ppm CO Calibration Gas:
CALIBRATE CO
MEASURED
APPLIED
492 PPM
0500 PPM
6. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Test Gas CO/H2 Screen
is displayed.
* The calibration range for this screen is 250 – 1500 ppm. Any attempt
to calibrate outside this range will cause the analyzer to display an
error message.
Instruction 24-9351
5-9
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
7. Turn off the flow of CO calibration-gas; then remove the calibration-
gas cylinder from the calibration fixture.
8. Attach a CO/H2 calibration-gas cylinder to the calibration fixture;
then adjust the regulator of the calibration fixture for a flowmeter
reading of approximately 2 SCFH.
9. Use the stand
keys to enter a CO-VALUE that is the same as
the CO concentration which is stamped on the CO/H calibration-gas
2
cylinder.
Typical Test Gas CO/H2 Screen During Calibration Procedure:
TEST GAS CO/H2
-------------------
CO-VALUE
XXXX
ENTER CO-VALUE
10. Press the
Screen.
key to save the CO-VALUE and display the Calibrate H2
11. After calibration gas has been applied for approximately 3 minutes (to
allow for stabilization), use the stand keys to enter an APPLIED
value* that equals the H concentration stamped on the CO/H
2
2
calibration-gas cylinder.
Typical CALIBRATE H2 Screen During Calibration Procedure:
CALIBRATE H2
MEASURED
APPLIED
1050 PPM
1000 PPM
12. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The calibration range for this screen is 500–1500 ppm, any attempt to
calibrate outside this range will cause the unit to display an error
message.
5-10
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Calibration
5.10 Calibrate Draft
(For PCA Models 15, 25, 35, 45, 55 & 65)
Material Required:
• Calibration Fixture (Section 5-2)
• Bellows (adjustable)
• Micromanometer
Range:
±20 mb (±8 in. H2O column)
Accuracy: ±0.025 mb (±0.01 in. H2O column)
Procedure:
IMP OR TANT! In Step 1, d o n ot connect the draft calibra-
tion fixture to the analyzer until the Calibrate Draft Screen
has been selected and displayed.
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to
calibrate the draft sensor.
2. With the Calibrate Draft Screen displayed, connect the hose from the
calibration fixture to the analyzer’s DRAFT port; then adjust the
bellows for a micromanometer reading of –10 mb (–4" H O column).
2
3. Wait until the MEASURED reading on the screen stabilizes. Then use
the stand
keys to enter an APPLIED reading* which equals the
Micromanometer reading.
Typical CALIBRATE DRAFT Screen During Calibration Procedure:
CALIBRATE DRAFT
MEASURED
APPLIED
- 9.00 MB
-10.00 MB
4. Press the
key to calibrate the analyzer’s MEASURED value to
that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
5. When all desired calibrations have been completed, press
the Combustion Test Screen.
to exit to
* The calibration range for this screen is –5 to –15 mb, any attempt to
calibrate outside this range will cause the unit to display an error
message.
Instruction 24-9351
5-11
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
PCA
NOTES:
5-12
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.0 MAINTENANCE
6.1 Routine Maintenance
Routine maintenance of the analyzer consists of: replacing the batteries,
cleaning the probe, draining the water trap, replacing the water trap
filter, and performing periodic calibration checks to ensure that the
analyzer is providing accurate readings.
• Replace the Batteries per Section 3.2.
• Clean the Probe per Section 6.3.
• Maintain the Water Trap/Filter Assembly per Section 6.4.
• Replace the Particulate Filter per Section 6.5
• Calibrate the analyzer per Section 5.0.
Instruction 24-9351
6-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.2 Disassembly
Perform the following when a maintenance procedure calls for removing
the case, printed circuit board, pump, or sensors:
1. Remove the batteries (Section 3.2.1)
2. Place the analyzer face down on a work surface, then remove the
unit’s four rear-case screws.
3. Carefully lift the rear case from the analyzer, unplug the battery
compartment wires, then place the rear housing on a work surface
(see Figures 6-1 & 6-2).
B
W
A
T
T
E
E
R
Y
I
R
S
Figure 6-1. Disassembling the Analyzer
6-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Maintenance
TO
BATTERY
COMPARTMENT
BIAS
BATTERY
FUSE
CO PCB
NX PCB
OPTIONAL
OPTIONAL
NITRIC
OXIDE
CARBON
MONOXIDE
SENSOR
SENSOR
PUMP
MTG.
SCREWS
OXYGEN
SENSOR
PUMP
–
+
RED
BLACK
4
PARTICULATE
FILTER
DRAFT
SENSOR
Figure 6-2. Rear View of the PCB and Sensors
Instruction 24-9351
6-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.3 Cleaning the Probe
The Probe Tube and the Probe Body will become dirty under normal use
(the water trap’s filter element should prevent soot from reaching the
analyzer’s internal components). If the probe assembly is not kept clean,
it could become clogged and restrict the flow of gas to the analyzer,
resulting in incorrect readings and calculations.
NOTE : An analyzer that is used to sample natural-gas
furnaces normally requires less frequent cleaning than an
analyzer used to sample oil or coal fired furnaces.
Equipment Required:
• Alcohol
• Aerosol can of Automotive Carburetor Cleaner
• Clean Rag
• Source of Compressed Air (optional)
Procedure:
1. Remove the rubber tubing from the barbed fitting(s) on the probe
handle (Figure 3-2).
CAUTION:
Carburetor cleaner attacks plastic components. Take precau-
tions not to spray cleaner onto the probe handle or analyzer.
2. Insert the plastic-spray tube of the carburetor cleaner into the barbed
fitting(s) of the probe handle; then liberally spray carburetor cleaner
through the probe.
3. After spraying, remove all the residual cleaner by repeatedly flushing
the probe with alcohol.
4. Wipe off the surfaces of the probe and tubing with a clean rag.
5. Allow the parts to dry completely. If available, blow compressed air
through the probe to expedite the drying process.
6. Reassemble the parts of the probe assembly.
6-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Maintenance
6.4 Water Trap/Filter Assembly Maintenance
The Water Trap / Filter Assembly removes water condensate from the gas
sample, and also prevents soot from contaminating the internal compo-
nents of the analyzer.
Drain the water condensate after every use.
Procedure:
1. Pull off the end-cap from the Inlet End of the Water Trap / Filter
Assembly (Figure 6-3).
2. Pour out all of the water condensate, and replace the end-cap.
Replace the filter element when it becomes
excessively dirty.
Equipment Required:
• Filter Element (Refer to Section 8.0)
Procedure:
1. Pull off the end-cap from the Outlet End of the Water Trap / Filter
Assembly (Figure 6-3).
2. Remove and discard the old filter element.
3. Install a new filter element and replace the end-cap.
F
i
l
t
e
r
E
l
e
m
e
n
t
R
i
s
e
r
T
u
b
e
I
E
n
l
e
t
O
E
u
t
l
e
t
n
d
n
d
Figure 6-3. Water Trap/ Filter Assembly
Instruction 24-9351
6-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.5 Replacing the Particulate Filter
The internal particulate filter (Figure 6-2) prevents small dust and dirt
particles from entering and damaging the pump. Depending on your
environmental conditions, it is recommended to change the particulate
filter and fitting approximately every six months, or sooner if it becomes
blocked.
Equipment Required:
• Particulate Filter and Fitting (Refer to Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Carefully pull off the tubing from the filter and fitting. Note the
orientation of tubing and fitting before removing.
3. Remove and discard the old filter and fitting and replace with new
ones. Be careful not to not pinch the tubing during reassembly.
4. Reassemble the analyzer in the reverse order of disassembly.
6-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.6 Replacing the Oxygen Sensor
Replace the Oxygen Sensor when it has expired (when the analyzer’s
automatic O2 calibration fails and the unit displays the message “O2-
Sensor Error”).
NOTE : A "O2-SENSOR ERROR" displayed in the Sensor
Status Screen does not necessarily mean that the sensor has
expired. Before replacing the sensor, refer to Section 7.2 for
other possible causes of the error.
Equipment Required:
• Small Flat Blade Screw Driver
• Oxygen Sensor (Refer to Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Remove the Oxygen Sensor connector from the printed circuit board;
then carefully pull off the two wires connected to the pins of the
sensor (see Figure 6-2).
3. Push in and turn the sensor counterclockwise (watching the notch on
top) until it stops; then pull the sensor out of its housing.
4. To install a new sensor, push the sensor into the housing and turn it
clockwise until it locks in place.
5. Connect the black (–) and red (+) wires, which were removed from the
old sensor in Step 2, to the pins of the new sensor. Be sure to observe
polarity as marked on the sensor.
6. Reinstall the sensor connector onto the printed circuit board.
7. Reassemble the analyzer. Then allow the sensor to be connected in
the circuit for at least 1 hour before continuing.
8. Place the analyzer in an area of fresh air and turn it ON. After the
warmup cycle, observe that the Sensor Status Screen should not show
an O2 sensor error.
Instruction 24-9351
6-7
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.7 Replacing the Nitric Oxide Sensor
(For PCA Models 30, 35, 60 & 65)
Replace the Nitric Oxide sensor when it has expired (can no longer be
calibrated).
NOTE : A "NX-SENSOR ERROR" displayed in the Sensor
Status Screen does not necessarily mean that the sensor has
expired. Before replacing the sensor, refer to Section 7.2 for
other possible causes of the error.
Equipment Required:
• Small Flat Blade Screw Driver
• Nitric Oxide Sensor (Refer to Section 8.0)
• Nitric Oxide Sensor Filter (Refer to Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Carefully pull the printed circuit board off the rear of the Nitric Oxide
sensor (see Figure 6-2).
3. Push in and turn the sensor counterclockwise (watching the notch on
top) until it stops; then pull the sensor out of its housing.
4. Install a new sensor by first pushing it into its housing, and then
turning it clockwise until it locks in place.
5. Install the circuit board, which was removed in Step 2, onto the rear
of the sensor.
6. Reassemble the analyzer and allow the sensor to be connected in the
circuit for at least 4 hours before continuing.
7. Place the analyzer in an area of fresh air and turn it ON.
8. Calibrate the analyzer per Section 5.0.
6.7.1 Replacing the Nitric Oxide Sensor Filter
Replacing the orange filter on the Nitric Oxide sensor once a year could
increase the life of the sensor.
Procedure:
1. Remove the Nitric Oxide sensor per Section 6.7.
2. Pry the orange filter from the Nitric Oxide sensor and replace it with
a new one.
3. Reinstall the sensor.
6-8
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.7.2 Replacing the Nitric Oxide Sensor
Bias Battery
A single lithium battery, located on the Nitric Oxide printed circuit board
(see Figure 6-2), applies a constant bias voltage to the Nitric Oxide sensor
even while the instrument is turned off. This battery has a life expectancy
of at least 2 years.
Replace the Nitric Oxide bias battery toward the end of its expected life.
Equipment Required:
• Bias Battery (See Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Remove old battery from its holder (see Figure 6-2).
3. Insert the new battery (positive side facing upwards) into its holder.
4. Reassemble the analyzer.
5. Before powering up and using the instrument, allow the Nitric Oxide
sensor to stabilize as described below. Recalibration of the Nitric
Oxide sensor is usually not required.
Depending on how long the Nitric Oxide sensor was without bias
voltage, the time required for the sensor to completely stabilize varies
from less than a minute to several days. Typical stabilization times
are shown below. Generally, however, the sensor is sufficiently stable
after 4 hours for measurement purposes.
Bia s r em oved for . . Sta biliza tion tim e
Less than 15 min.
Less than 1 hr.
Less than 2 days
Greater than 2 days
Less than 1 min.
Less than 5 min.
Less than 4 hr.
Up to 2 days
Instruction 24-9351
6-9
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Maintenance
6.8 Replacing the Carbon Monoxide Sensor
(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)
Replace the Carbon Monoxide sensor when it has expired (can no longer
be calibrated).
NOTE : A "CO-SENSOR ERROR" displayed in the Sensor
Status Screen does not necessarily mean that the sensor has
expired. Before replacing the sensor, refer to Section 7.2 for
other possible causes of the error.
Equipment Required:
• Small Flat Blade Screw Driver
• Carbon Monoxide Sensor (Refer to Section 8.0
• Carbon Monoxide Sensor Filter (Refer to Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Carefully pull the printed circuit board off the rear of the Carbon
Monoxide sensor (see Figure 6-2).
3. Push in and turn the sensor counterclockwise (watching the notch on
top) until it stops; then pull the sensor out of its housing.
4. Remove the wire jumper from the pins of the new sensor.
5. Install the new sensor by first pushing the sensor into its housing,
and then turning clockwise until it locks in place.
6. Install the circuit board, which was removed in Step 2, onto the rear
of the sensor.
7. Reassemble the analyzer; then allow the sensor to be connected in the
circuit for at least 1 hour before continuing.
8. Place the analyzer in an area of fresh air and turn it ON.
9. Calibrate the analyzer per Section 5.0.
6-10
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Maintenance
6.8.1 Replacing the Carbon Monoxide Sensor Filter
Replacing the red filter on the Carbon Monoxide sensor once a year could
increase the life of the sensor.
Procedure:
1. Remove the Carbon Monoxide sensor per Section 6.8.
2. Pry the red filter from the Carbon Monoxide sensor and replace it with a
new one.
3. Reinstall the sensor.
Instruction 24-9351
6-11
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
PCA
6.9 Replacing the Pump Assembly
Replace the Pump Assembly if it is found to be defective.
Equipment Required:
• Small Flat Blade Screw Driver
• No. 1 Phillips Screw Driver
• Pump Assembly (Refer to Section 8.0)
Procedure:
1. Remove the analyzer’s rear case and lay it aside (refer to Section 6.2).
2. Remove the two self tapping Phillips screws holding the Pump Assem-
bly (See Figure 6-1).
3. Unplug the pump connector from the printed circuit board; slide off
the two hoses from the pump noting their orientation; then remove
the entire assembly.
4. Install the new assembly and reassemble the analyzer in the reverse
order of disassembly.
6-12
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Troubleshooting
PCA
7.0 TROUBLESHOOTING
7.1 Analyzer Repair
It is recommended that field repair of the PCA be limited to:
• Simple checks of the printed circuit boards
• Replacing the Probe Assembly
• Replacing the filter element in the Water Trap / Filter Assembly
• Replacing the Particulate Filter
• Replacing the Pump Assembly
• Replacing Sensors and Sensor Filters
• Replacing Batteries
All other repairs should be performed by an authorized Bacharach Ser-
vice Center. Any repairs performed by an unauthorized service organiza-
tion will void the analyzer’s warranty and release Bacharach, Inc. of any
implied or written product liability.
Before returning your analyzer for repair, you may be able to determine
and resolve a problem using the Troubleshooting Guide in Section 7.3.
Instruction 24-9351
7-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Troubleshooting
PCA
7.2 Error Codes
If one of the following messages appear at the bottom of the Sensor Status
Screen, refer to Section 7.3 Troubleshooting Guide for information on how
to correct the error.
O2 - SENSOR ERROR
CO - SENSOR ERROR
NX - SENSOR ERROR
O2 Sensor not connected, or is expired, or
was exposed to combustion gases during
warmup.
Carbon Monoxide sensor is expired, or was
exposed to Carbon Monoxide during
warmup.
Nitric Oxide sensor is expired, or was
exposed to Nitric Oxide during warm up,
or the bias battery is dead.
TL-SENSOR ERROR
TA-SENSOR ERROR
Room air thermocouple is outside the
range of –20 to 100 °C (–4 to 212 °F)
Flue Gas thermocouple is not connected
or is outside the range of –20 to 1200 °C
(–4 to 2192 °F)
DRAFT-SENSOR ERROR
LOW BATTERY
Outside the range of –7.5 to +7.5 mb
(– 3 to + 3inches of water column)
Battery voltage has dropped below 3.9 volts.
The instrument will shut off when battery
voltage drops below 3.5 volts.
“– – – –”
Not calculated (O2 above 18.8%,
or the temperature is above 1200 °C
(2192 °F), or sensor error
“* * * *”
“XXXX”
Sensor not installed
Overrange (numeric)
7-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Troubleshooting
7.3 Troubleshooting Guide
The following table lists the causes and remedies for most of the problems
that may arise with the analyzer. For help with any problem not dis-
cussed here, contact your nearest Bacharach Service Center.
TABLE 7-1. TR OUBLE SH OOTING GUIDE
F a u lt
P r ob a b le Ca u se & R em ed y
Analyzer completely nonfunctional; a. Batteries dead. Replace batteries
per Section 3.2.
won’t turn on when the
pressed.
key is
b. Loose battery connector. Disas-
semble analyzer and ensure that
the battery connector is attached
to the printed circuit board.
c. Microprocessor needs to be reset.
Press RESET button (Fig. 4-7).
d. Optional Power Supply defective.
Replace Power Supply.
e. An a lyzer defect ive. Ret u r n t o
Bacharach for repair.
a. Microprocessor needs to be reset.
Press RESET button (Fig. 4-7).
Display Screen is blank when ana-
lyzer is turned on, but pump runs
during warmup cycle.
b. An a lyzer defect ive. Ret u r n t o
Bacharach for repair.
Batteries close to being discharged.
Analyzer will run for several min-
utes before the instrument shuts off.
Replace batteries per Section 3.2.1.
LOW BATTE R Y message appears
at bottom of display.
a. Calibration was attempted while
sampling combustion gases.
O 2-S E N S O R E R R O R code a p-
pears in the Sensor Status Screen.
b. O2 sensor is either expired, not
wired correctly, or not connected to
the circuit board. Replace or check
wiring of sensor per Section 6.6.
Instruction 24-9351
7-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Troubleshooting
PCA
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)
F a u lt P r ob a b le Ca u se & R em ed y
NX-SE NSOR E R R OR code a p-
pears in the Sensor Status Screen.
a. Calibration was attempted while
sampling combustion gases.
b. Nitric Oxide sensor is expired.
Replace sensor per Section 6.7.
c. Bias battery on the Nitric Oxide
circuit board is dead. Replace bat-
tery per Section 6.7.2.
CO-SE NSOR E R R OR code a p-
pears in the Sensor Status Screen.
a. Calibration was attempted while
sampling combustion gases.
b. Carbon Monoxide sensor is ex-
pired. Replace sensor per Sec-
tion 6.8.
DRAFT-SENSOR ERROR code ap-
pears in the Sensor Status Screen.
a. Sensor is exposed to pressure out-
side of its detectable range.
b. Sensor defective. Return analyzer
to Bacharach for repair.
TA-SE NSOR OR TL-SE NSOR
ERROR code appears in the Sen-
sor Status Screen.
a. Calibration was attempted while
sampling combustion gases.
b. Th er m ocou ple defect ive. Re-
place probe assembly.
c. Thermocouple not connected.
Connect thermocouple to ana-
lyzer per Section 3.3.
d. Instrument exposed to temp-
erature outside it’s allowable
operating temperature.
“****” appears in one or more value
fields.
The field’s associated sensor is not
installed.
7-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Troubleshooting
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)
F a u lt P r ob a b le Ca u se & R em ed y
“- - - -” appears in one or more value a. The analyzer is not able to calcu-
fields of the Combustion Test Screen.
late a numerical value based on
measured combustion data. The
“- - - -” is replaced with numerical
values when the analyzer begins
to detect valid combustion data.
b. Sensor in error during warm-up.
“XXXX” appears in one or more value The field’s associated sensor is de-
fields.
tecting a value that is outside the
analyzer’s detection range. “XXXX”
is r epla ced wit h n u m er ica l da t a
when the analyzer detects values
that fall within its range.
An a lyzer won ’t r espon d wh en a Microprocessor needs to be reset.
panel key is pressed. Press RESET button (Fig. 4-7).
Pump motor sounds sluggish, stalls, a. Flow restricted. Check that the
or won’t start.
filter element in the Water Trap/
Filter Assembly is clean and not
saturated with water. Also, verify
that the probe hose tubing is not
pinched (Fig. 6-3).
b. Flow restricted. Check the Par-
ticulate Filter is clean and not
blocked (Fig. 6-2).
c. Loose pump connection. Disas-
semble analyzer and ensure that
the pump connector is securely
attached to the circuit board
d. Pump defective. Replace pump
assembly.
Instruction 24-9351
7-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Troubleshooting
PCA
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)
F a u lt P r ob a b le Ca u se & R em ed y
Backlight won’t turn on.
Backlight LED burned out. Return
to Bacharach for repair.
Batteries do not last 10 hours.
Cold temperature is reducing bat-
tery capacity. To obtain longer oper-
ating time, keep analyzer warm.
Erratic Combustion Test Screen a. Faulty sensor(s):
values. - Check that the sensors are prop-
erly installed per Sections 6.6
through 6.8.
- Ch eck sen sor ca libr a t ion per
Section 5.0.
- Repla ce sen sor (s) a n d r eca l-
ibrate per Sections 5.0 and 6.0.
b. Probe assembly leaking. Check
tightness of all hose connections
and integrity of tubing.
c. Pump defective. Replace pump &
motor assembly.
d. An a lyzer defect ive. Ret u r n t o
Bacharach for repair.
Analyzer will not calibrate properly. a. Wrong calibration gas or insuffi-
cient flow being applied to sensor.
Ensure your calibration setup is
correct.
b. Faulty sensor. Replace sensor and
recalibrate per Sections 5.0 and
6.0.
c. An a lyzer defect ive. Ret u r n t o
Bacharach for repair.
7-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Parts & Service
8.0 PARTS & SERVICE
8.1 Replacement Parts
It em
Descr ip t ion
P a r t No.
F igu r es
8-1, 2 & 3
1
2
3
Battery Cover
Screw, Case Housing
Screw, Pump Mounting
24-0784
501-3824
501-3822
8
9
Oxygen Sensor
24-0788
24-0789
24-0881
24-0863
24-0862
24-3009
604-2605
204-0020
Carbon Monoxide Sensor
Nitric Oxide Sensor
Carbon Monoxide Sensor Filter
Nitric Oxide Sensor Filter
Pump Assembly
10
11
12
13
14
15
Fuse
Battery, Nitric Oxide Sensor Bias
18
19
20
21
22
23
24
25
Filter Element (white)
Draft Connector, Probe
Gas Connector, Probe
O-Ring, 7mm OD x 1mm wall
O-Ring, 8mm OD x 1mm wall
Particulate Filter
Fitting, for Particulate Filter
Filter Assembly (complete)
07-1644
24-0878
24-0877
105-5103
105-5102
07-1600
103-5267
24-1107
Instruction 24-9351
8-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Parts & Service
PCA
8.2 Accessories
Descr ip t ion
P a r t No.
STANDAR D ACCE SSOR IE S:
Battery, “AA” Alkaline
Complete Probe and Hose Assembly (Gas & Draft)
Instruction Manual
204-0004
24-3004
24-9351
24-1078
Plastic Carrying Case
OP TIONAL ACCE SSOR IE S:
Ambient Thermocouple, 10 ft. K-type
Ambient Thermocouple, 1 in. K-type
Bent Probe Tip
104-1797
104-1798
24-8039
24-7059
24-1103
24-0794
24-0492
Calibration Kit
Differential Pressure Hose Assembly
Gas Cylinder, 1000 ppm CO, 1000 ppm H2, in Nitrogen
Gas Cylinder, 500 ppm CO
Printer, IrDA (with battery charger):
120 VAC
24-1229
24-1230
06-8733
24-0885
24-1209
24-1073
230 VAC
Printer Paper, Thermal (1 roll)
Power Supply Adapter, 110 VAC
Regulated Power Supply Adapter, 230 VAC
Serial Cable
8-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Parts & Service
2
1
3
Figure 8-1. Replacement Parts
Instruction 24-9351
8-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Parts & Service
PCA
FUSE
14
15
10
12
9
11
8
–
+
13
RED
BLACK
4
23
24
22
21
Figure 8-2. Replacement Parts
8-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Parts & Service
1
9
2
0
1
8
2
5
Figure 8-3. Replacement Parts
Instruction 24-9351
8-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Parts & Service
PCA
8.3 Sales/Service Centers
Replacement parts and service can be obtained by contacting one of
the following Bacharach Sales/Service Centers:
UnitedStates
Canada
BacharachofCanada,Inc.
250ShieldsCourtUnit#3
Markham, OntarioL3R9W7Canada
Phone:905-470-8985
Fax:905-470-8963
California
7281GardenGroveBlvd.,
SuiteH
GardenGrove, CA92841
Phone:714-895-0050
Fax:714-895-7950
México
BacharachdeMéxico
PlayaRegatasNo.473TercerPiso
Col.MilitarMarte
DelegaciónIztacalco,08830
MéxicoD.F.México
Phones: +52-555-634-7740
+52-555-634-7741
Indiana
8618LouisianaPlace
Merrillville,IN46410
Phone:219-736-6178
Fax:219-736-6269
FAX: +52-555-634-7738
NewJersey
7300IndustrialPark
Rte.130,Bldg.22
Europe
Pennsauken,NJ08110
Phone:856-665-6176
Fax:856-665-6661
EuropeanHeadquarters
BacharachInstruments
SovereignHouse,Queensway
RoyalLeamingtonSpa
WarwickshireCV313JR
UnitedKingdom
Phone:+44-1926-338111
Fax:+44-1926-338110
Pennsylvania
621HuntValleyCircle
NewKensington, PA15068
Phone:724-334-5051
Fax:724-334-5723
Sales/ServiceCenter-Denmark
BacharachInstrumentsInt’l
P.O.Box44
Texas
5151Mitchelldale,B-4
Houston, TX77092
Phone:713-683-8141
Fax:713-683-9437
39Lindegade
DK6070ChristiansfeldDenmark
Phone:+45-74-563171
Fax:+45-74-563178
8-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix A
APPENDIX A
Display Screen Translations
This appendix shows the contents of the various screen displays that are
seen while operating the PCA in all ten languages that the analyzer is
capable of displaying. The language that is displayed on the analyzer is
dependent on the analyzer’s model (refer to Table 1-1), and the language
selected per Section 4.15.
The languages are arranged in columns side-by-side for comparison and
translation purposes.
Instruction 24-9351
A-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
PCA
English
(ENG)
German
(DEU)
Danish
(DAN)
French
(FRA)
Spanish
(ESP)
Warm Up Screen
BACHARACH, INC.
PCA nn
BACHARACH, INC.
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
PCA nn
Warmup nn
KALIBRIEREN nn
OPVARMNING nn
CYCLE D AUTOCALIB nn CALENTAMIENTO nn
Sensor Status Screen (Errors)
NO ERRORS DETECTED
LOW BATTERY
KALIBRIERUNG OK
Batterie leer
INGEN FEJL
PAS D ERREUR DETECTE NO HAY ERRORES
LAV BATTERIKAPACITET
O2-SENSOR FEJL
CO-SENSOR FEJL
TRAEK-SENSOR FEJL
ROEGGASTEMP FEJL
LUFTTEMP FEJL
BATTERIE FAIBLE
BATTERIA BAJA
O2-SENSOR ERROR
CO-SENSOR ERROR
DRAFT-SENSOR ERROR
O2-Sensor
02-ERREUR DE CAPTEUR ERROR SENSOR-O2
ERREUR DE CAPTEUR-CO ERROR SENSOR-CO
CO-Sensor
Feinzug-Sensor
ERR D CPTR TIRAGE
ERROR SENSOR-TIRO
TA-STACK-SENSOR ERROR Gas-F¸hler
ERREUR D CPTR T-CHMN ERR SENSOR T-CHIMENE
TA-AIR-SENSOR ERROR
NX-SENSOR ERROR
Luft-F¸hler
NX-Sensor
ERREUR D CPTR T-AIR
ERREUR DE CAPTEUR
ERROR SENSOR T-AIRE
ERROR SENSOR-NX
NX-SENSOR FEJL
Sign Off Screen
OFF IN
n
SEC
Aus in
n
sec
STOP OM n SEK
TERMNR DANS
n
SEC
APAG EN n SEGS
RUN/HLD Screen (Abbreviations)
NG
EG
KG
SG
FG
NG
KOK
BG
GN
GN
KOK
LEG
LPG
O#2
O#6
PC
GM
GC
GV
GCD
LPG
P#2
P#6
CAR
FG
LPG
H#2
H#6
CHR
Hˆl
Sˆl
KOL
O#2
O#6
BF
Fuel Selection Screen
FUEL
BRENNSTOFF
Erdgas
FUEL
COMBSTBL
GAZ NATU
GAZ MANU
GAZ VIL
COMBUSTB
GNATURAL
GCARBON
GCIUDAD
LPG
NATGAS
KOKS
N-GAS
Kok.-Gas
Stadtgas
Fl.-Gas
KOKS
LEG
B-GAS
LPG
F-GAS
PROPANE
FUEL DOM
FUEL LOU
CHARBON
OIL NO.2
OIL NO.6
P-COAL
Heizoel
LET OLIE
TUN OLIE
BIOFUEL
PETRO #2
PETRO #6
CARBON
Schweroel
Kohle
Draft Screens
DRAFT
FEINZUG
Saugschlauch
trennen
TRAEK
TIRAGE
TIRO
DISCONNECT DRAFT
HOSE
AFMONTER
TRAEKSLANGE
TRYK
DEBRANCHER LE TUYAU DESCONECTE MANGUERA
DE TIRAGE
DE ASPIRACION
OPRIMA
PRESS
dr¸cken
APPUYER SUR
RECONNECT DRAFT
HOSE
Schlauch
anschlieflen
Feinzug
MONTER
REBRANCHER LE TUYAU RECONECTE MANGUERA
TRAEKSLANGE
TRAEK
DE TIRAGE
TIRAGE
DE ASPIRACION
TIRO
DRAFT
HOT SPOT
Gas-Temp
KERNESTR
PNT CHAUD
PTO CLNTE
A-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix A
Polish
(POL)
Italian
(ITA)
Dutch
(NED)
Swedish
(SVE)
Finnish
(FIN)
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
BACHARACH, INC.
PCA nn
NAGRZEWANIE nn
RISCALDAMENTO nn
INITIALISATIE nn
UPPVARMNING nn
MITTARIL‰MPI‰‰
NIE WYKRYTO BLEDOW
SLABY AKUMULATOR
BLAD CZUJNIKA-O2
NO RIVELAZIONE ERROR
BATTERIA BASSA
GEEN FOUTMELDINGEN
VOEDINGSSPANNING
ERR: O2-SENSOR
Inget fel
EI HAVAITTU VIRHEIT‰
PARISTO LOPPUU
Lag batterikapacitet
O2-SENSOR FEL
CO-SENSOR FEL
Drag-SENSOR FEL
ERRORE SENSORE-O2
ERRORE SENSORE-CO
VIKA O2-ANTURISSA
VIKA CO-ANTURISSA
VIKA PAINEANTURISSA
SAVUK LT-ANTURIVIKA
ILMAN LT-ANTURIVIKA
VIKA NX-ANTURISSA
BLAD CZUJNIKA-CO
ERR: CO-SENSOR
BLAD CZUJNIKA-CIAGU
BLAD CZUJNIKA KOMINT
BLAD CZUJNIKA POW-T
BLAD CZUJNIKA-NX
ERRORE SENSORE-TIRAG ERR: TREK-SENSOR
ERRORE SENS-T-CAMINO ERR:T-ROOKGAS SENSOR Rokgastemp FEL
ERRORE SENSOR-T-ARIA
ERRORE SENSORE-NX
ERR: T-OMG. SENSOR
ERR: NX-SENSOR
Lufttemp FEL
NOX-SENSOR FEL
WYTACZ W n SEK
OFF IN n SEC
UIT NA n SEC
Stangs av om n sek
KIINNI n SEK:SSA
GZ
MET
COK
CIT
NG
NG
MKA
KOK
KKA
PRO
KˆL
RˆL
BP
GWL
GNG
LPG
O#2
O#6
PC
KOK
LEG
LPG
O#2
O#6
PC
KOK
SG
GPL
O#2
O#6
CAR
PG
O#2
O#6
BF
PALIWO
COMBUSTI
METANO
GAS COKE
GASCITTA
GPL
BRANDST.
AARDGAS
KOKS
BRANSLE
N-GAS
POLTTOAI
MAAKAASU
HIILIKAA
GAZ ZIEN
GAZWEGL
GAZNENG
PROP PLN
OLEJ LEK
OLEJ CIE
P-COAL
KOKS
LEG
Stadsgas
P-GAS
KAUPKAAS
PROPAANI
KEVYTˆLJ
RASKˆLJY
BIOPOAINE
LPG
GASOLIO
NAFTA
OIL NO.2
OIL NO.6
P-COAL
Tunn olja
Tjock olja
Biofuel
CARBON
CIAG
TIRAGGIO
TREK-METING
ONTKOPPEL LEIDING
DRAFT HOSE
DRAG
VETO
ODLACZ WAZ DO
POMIARU CIAGU
NACISNIJ
DISINSERIRE
ASPIRATORE
PREMERE
Tabort drag slang
IRROITA PAINELETKUN
LIITIN
DRUK
Tryck
PAINA
PODLACZ PONOWNIE WAZ
DO POMIARU CIAGU
CIAG
REINSERIRE
ASPIRATORE
TIRAGGIO
VERBIND OPNIEUW DE
LEIDING (DRAFT HOSE)
TREK
Anlsut drag slang
LIIT‰PAINELETKUN
LIITIN
Drag
PAINE
OGNISKO
PTO CALDO
HOT SPOT
Rokgas
KUUMA KOH
Instruction 24-9351
A-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
PCA
English
(ENG)
German
(DEU)
Danish
(DAN)
French
(FRA)
Spanish
(ESP)
Saving Memory Screen
SAVING MEMORY
LOCATION nn
Speichere Daten in
Block nn
LAGRING MEMORY
BLOK nn
SAUVEGARDE D MEMOIRE GUARDANDO UBIBACION
ZONE nn
DE MEMORIA nn
Memory Directory Screen
MEMORY DIRECTORY
MEMORY EMPTY
CLEAR MEMORY
SPEICHER LISTE
Speicher leer
MEMORY BIBLIOTEK
MEMORY TOM
REPERTOIRE MEMRE
MEMOIRE VIDE
DIRECTORIO MEMORI
MEMORIA VACIA
Speicher Lˆschen
SLET MEMORY
EFFACER MEMOIRE
BORRAR MEMORIA
Draft Memory Screen
DRAFT MEM
DRAFT
FEINZUG MEM
Feinzug
TRAEK MEMORY
TRAEK
TIRAGE MEM
TIRAGE
TIRO MEM
TIRO
Clear Memory Screen
CLEAR MEMORY
SPEICHER LˆSCHEN
SLET MEMORY
EFFACER MEMOIRE
BORRAR MEMORIA
Memory to PC Screen
MEMORY TO PC
TRANSMIT DATA
CLEAR MEMORY
TRANSMITTING...
ABSPEICHERN AUF PC
PC HUKOMMELSE
OVERFOER DATA
SLET MEMORY
MEMOROIRE A PC
TRANSM. DONNEES
EFFACER MEMOIRE
EN TRANSMISSION...
MEMORIA AL PC
Daten ¸bertragen
Speicher lˆschen
¸bertragung...
TRANSMITIR DATOS
BORRAR MEMORIA
TRANSMITIENDO...
OVERFOERELSE...
ID Setup Screen
SETUP
ID #n
KONFIGURIEREN
ID #n
OPSAETNING
ID #n
PROGRAMMER
ID #n
ESTABLECER
ID #n
Temperature Setup Screen
SETUP
KONFIGURIEREN
Temp.-Einheit
OPSAETNING
PROGRAMMER
UNITE DE
ESTABLECER
UNIDAD DE
TEMPERATURE UNIT
TEMPERATURENHED
TEMPERATURE
TEMPERATURA
Draft Setup Screen
SETUP
KONFIGURIEREN
OPSAETNING
TRAEKENHED
mB/PA/WC
PROGRAMMER
UNITE DE TIRAGE
mB/PA/WC
ESTABLECER
UNIDAD DE TIRO
mB/PA/WC
DRAFT UNIT
mB/PA/WC
Feinzug-Einheit
mB/PA/WC
Language Setup Screen
SETUP
KONFIGURIEREN
Sprache w‰hlen
OPSAETNING
SPROGKODE
PROGRAMMER
LANGUE
ESTABLECER
IDIOMA
LANGUAGE
Display Mode Setup Screen
SETUP
KONFIGURIEREN
Anzeigen
OPSAETNING
DISPLAY
PROGRAMMER
AFFICHER
ESTABLECER
VISUALIZAR
DISPLAY
A-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix A
Polish
(POL)
Italian
(ITA)
Dutch
(NED)
Swedish
(SVE)
Finnish
(FIN)
ZACHOWYWANIE MIEJSCE
W PAMIECI nn
SALVARE MEMORIA
UBICAZIONE nn
OPSLAAN IN GEHEUGEN
LOCATIE nn
Sparar minne
i block nn
TALLENTAA MUISTI
PAIKKAAN nn
KATALOG PAMIECI
PAMIEC PUSTA
INDIRIZZ MEMORIA
MEMORIA VUOTA
INHOUD GEHEUGEN
GEHEUGEN LEEG
GEHEUGEN WISSEN
MINNES BIBLIOTEK
Minnet tomt
MUISTITIEDOSTO
MUISTI TYHJfl
WYCZYSC PAMIEC
ANNULARE MEMORIA
Rensa minnet
TYHJENNfl MUISTI
CIAG MEM
CIAG
TIRAGGIO MEM
TIRAGGIO
TREK-METING MEM
TREK
DRAG MINNET
Drag
VETO MEM
PAINE
WYCZYSC PAMIEC
ANNULARE MEMORIA
GEHEUGEN WISSEN
Rensa minnet
TYHJENNfl MUISTI
PAMIEC DO PC
DA MEMORIA A PC
TRASMETTERE DATI
ANNULARE MEMORIA
TRANSMISSIONE...
GEHEUGEN NAAR PC
VERZEND GEGEVENS
GEHEUGEN WISSEN
VERZENDING...
MINNE TILL PC
Overfor data
Rensa minnet
Overforing...
MUISTI PC:LLE
SIIRRƒ TIEDOT
TYHJENNƒ MUISTI
SIIRT‰‰ .
PRZESLAC DANE
WYCZYSC PAMIEC
TRANSMISJA...
USTAW
ID #n
PREPARAZIONE
ID #n
INSTELLINGEN
ID #n
KONFIG
ID #n
ASETUKSET
TUNNISTE #n
USTAW
PREPARAZIONE
INSTELLINGEN
EENHEID VAN
TEMPERATUUR
KONFIG
ASETUKSET
L‰MPˆTILAN
YKSIKKˆ
JEDNOSTKE
TEMPERATURY
UNITA DI TEMPER
Temperaturenhet
USTAW
PREPARAZIONE
UNITA DI TIRAGGIO
mB/PA/WC
INSTELLINGEN
EENHEID VAN TREK
mB/PA/WC
KONFIG
ASETUKSET
JEDNOSTKE CIAGU
mB/PA/WC
Drag enhet
mB/PA/WC
PAINEEN YKSIKKˆ
mB/PA/WC
USTAW
JEZYK
PREPARAZIONE
LINGUA
INSTELLINGEN
TAAL
KONFIG
Sprak
ASETUKSET
KIELI
USTAW
PREPARAZIONE
VISUALIZZARE
INSTELLINGEN
SCHERM
KONFIG
DISPLAY
ASETUKSET
N‰YTTˆ
WYSWIETLIC
Instruction 24-9351
A-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
PCA
English
(ENG)
German
(DEU)
Danish
(DAN)
French
(FRA)
Spanish
(ESP)
Time/Date Setup Screen
SETUP
TIME
KONFIGURIEREN
OPSAETNING
TID
PROGRAMMER
HEURE
ESTABLECER
HORA
Zeit
DATE
Datum
DATO
DATE
FECHA
Maintenance Password Screen
MAINTENANCE
PASSWORD
INSTANDHALTUNG
Kennwort
VEDLIGEHOLDELSE
KODEORD
MAINTENANCE
MOT DE PASSE
MANTENIMIENTO
CONTRASENA
Printer Setup Screen
SETUP PRINTER
KONFIGURIEREN PRINTER OPSAETNING PRINTER
PROGRAMMER PRINTER ESTABLECER PRINTER
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
Maintenance Menu Screen
MAINTENANCE
CALIBRATION
USER NAME
INDSTANDHALTUNG
ABGLEICH
VEDLIGEHOLDELSE
CALIBRATION
MAINTENANCE
CALIBRAGE
MANTENIMIENTO
CALIBRACION
BENUTZERNAME
BRUGER NAVN
USER NAME
NOMBRE DEL USUARIO
Calibrate Menu Screen
CALIBRATE
TA-ZERO
TA-SPAN
TL-ZERO
TL-SPAN
NX
ABGLEICHEN
TA-Offst
TA-Gain
TL-Offst
TL-Gain
NX
KALIBRER
TA-NUL
TA-SPAN
TL-NUL
TL-SPAN
NX
CALIBRER
TA-ZERO
TA-FRCH
TL-ZERO
TL-FRCH
NX
CALIBRAR
TA-CERO
TA-TRAM
TL-CERO
TL-TRAM
NX
CO
CO
CO
CO
CO
DRAFT
ZUG
TRAEK
TIRAGE
TIRO
Calibrate TA-Zero Screen
CALIBRATE
TA-ZERO
ABGLEICHEN
TA-OFFSET
Ist-Wert
KALIBRER
TA-NUL
CALIBRER
TA-ZERO
MESURE
APPLIQUE
CALIBRAR
TA-CERO
MEDIDA
MEASURED
APPLIED
MAALT
Soll-Wert
SAND VAERDI
APLICADA
Calibrate TA-Span Screen
CALIBRATE
TA-SPAN
ABGLEICHEN
TA-GAIN
KALIBRER
TA-SPAN
CALIBRER
FRCHETTE-TA
MESURE
CALIBRAR
TA-TRAM
MEDIDA
MEASURED
APPLIED
Ist-Wert
MAALT
Soll-Wert
SAND VAERDI
APPLIQUE
APLICADA
Calibrate TL-Zero Screen
CALIBRATE
TL-ZERO
ABGLEICHEN
TL-OFFSET
Ist-Wert
KALIBRER
TL-NUL
CALIBRER
TL-ZERO
MESURE
APPLIQUE
CALIBRAR
TL-CERO
MEDIDA
MEASURED
APPLIED
MAALT
Soll-Wert
SAND VAERDI
APLICADA
A-6
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix A
Polish
(POL)
Italian
(ITA)
Dutch
(NED)
Swedish
(SVE)
Finnish
(FIN)
USTAW
CZAS
DATE
PREPARAZIONE
ORA
INSTELLINGEN
TIJD
KONFIG
Tid
ASETUKSET
AIKA
DATA
DATUM
Datum
PƒIVƒ
UTRZYMANIE
HASTO
MANTENIMENTO
ONDERHOUD
PASWOORD
UNDERHALL
PASSWORD
YLLƒPITO
PAROLA DORDINE
SALASANA
USTAW PRINTER
PREPARAZIONE PRINTER
IR-HP/IR-IRDA/RS232
INSTELLINGEN PRINTER
IR-HP/IR-IRDA/RS232
KONFIG PRINTER
ASETUKSET PRINTER
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
IR-HP/IR-IRDA/RS232
UTRZYMANIE
MANTENIMENTO
CALIBRAZIONE
ONDERHOUD
UNDERHALL
Kalibrering
YLLƒPITO
WZORCOWANIE
UZYWACZA IMIE
IJKINGSPROCEDURE
GEBRUIKERS ID
KALIBROINTI
KƒYTTƒJƒNIMI
NOME DELLíUTENTE
Anvandar namn
WZORCUJ
TA-ZERO
TA-ROZP
TL-ZERO
TL-ROZP
NX
CALIBRARE
TA-ZERO
TA-DIFF
TL-ZERO
TL-DIFF
NX
CALIBREER
T-RKG-0
T-RKG-1
T-OMG-0
T-OMG-1
NX
CALIBRATE
TA-ZERO
TA-SPAN
TL-ZERO
TL-SPAN
NX
KALIBROI
TS-NOLL
TS-YLƒR
TI-NOLL
TI-YLƒR
NX
CO
CO
CO
CO
CO
CIAG
TIRAGG
TREK
DRAFT
VETO
WZORCUJ
TA-ZERO
CALIBRARE
TA-ZERO
IJKING
CALIBRATE
TA-ZERO
KALIBROI
TS-NOLLA
MITATTU
T-RKG-0
MIERZONA
STOSOWANA
MISURATO
APPLICATO
GEMETEN
AANGELEGD
MEASURED
APPLIED
SY÷TETTY
WZORCU
CALIBRARE
TA-DIFF
IJKING
CALIBRATE
TA-SPAN
KALIBROI
TS-YLflRAJA
MITATTU
TA-ROZPIETOS
MIERZONA
T-RKG-1
MISURATO
APPLICATO
GEMETEN
AANGELEGD
MEASURED
APPLIED
STOSOWANA
SY÷TETTY
WZORCUJ
TL-ZERO
CALIBRARE
TL-ZERO
IJKING
CALIBRATE
TL-ZERO
KALIBROI
TI-NOLLA
MITATTU
T-OMG-0
GEMETEN
AANGELEGD
MIERZONA
STOSOWANA
MISURATO
APPLICATO
MEASURED
APPLIED
SY÷TETTY
Instruction 24-9351
A-7
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
PCA
English
(ENG)
German
(DEU)
Danish
(DAN)
French
(FRA)
Spanish
(ESP)
Calibrate TL-Span Screen
CALIBRATE
TL-SPAN
ABGLEICHEN
TL-GAIN
KALIBRER
TL-SPAN
CALIBRER
FRCHETTE-TL
MESURE
CALIBRAR
TL-TRAM
MEDIDA
MEASURED
APPLIED
Ist-Wert
MAALT
Soll-Wert
SAND VAERDI
APPLIQUE
APLICADA
Calibrate NX Screen
CALIBRATE NX
MEASURED
APPLIED
ABGLEICHEN NX-GAIN
KALIBRER NX
MAALT
CALIBRER NX
MESURE
CALIBRAR NX
MEDIDO
Ist-Wert
Soll-Wert
SAND VAERDI
APPLIQUE
APLICADO
Calibrate CO Screen
CALIBRATE CO
MEASURED
APPLIED
ABGLEICHEN CO-GAIN
KALIBRER CO
MAALT
CALIBRER CO
MESURE
CALIBRAR CO
MEDIDO
Ist-Wert
Soll-Wert
SAND VAERDI
APPLIQUE
APLICADO
Calibrate CO/H2 Screen
TESTGAS CO/H2
CO VALUE
Pr¸fgas CO/H2
CO-Anteil
PROEVEGAS CO/H2
CO VAERDI
TESTERGAZ CO/H2
VALEUR CO
GAS DE PRUEBA CO/H2
VALOR DE CO
ENTER CO VALUE
CO-Anteil eingeben
CO VAERDI ENTER
ENTRER VALEUR CO
ENTRAR VALOR DE CO
Calibrate H2 Screen
CALIBRATE H2
MEASURED
APPLIED
ABGLEICHEN H2-GAIN
KALIBRER H2
MAALT
CALIBRER H2
MESURE
CALIBRAR H2
MEDIDO
Ist-Wert
Soll-Wert
SAND VAERDI
APPLIQUE
APLICADO
Calibrate Draft Screen
CALIBRATE DRAFT
MEASURED
ABGLEICHEN FEINZUG
KALIBRER TRAEK
MAALT
CALIBRER TIRAGE
MESURE
CALIBRAR TIRO
MEDIDO
Ist-Wert
APPLIED
Soll-Wert
SAND VAERD
APPLIQUE
APLICADO
User Name Screen
USER NAME
LINE n
BENUTZERNAME
Zeile n
BRUGER NAVN
LINIE n
USER NAME
LIGNE n
NOMBRE DEL USUARIO
LINEA n
Miscellaneous
BAD CALIBRATION ENTRY Kalibrierung nich mˆglich
KALIBRERING IKEE MULIGT ERREUR D'ETALONNAGE
SKYLLER CO SENSOR PURGE CELLULE CO
CALIBRACION-INCORRECTA
PURGANDO SENSOR CO
PURGING CO SENSOR
CO Sp¸lfunktion
A-8
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix A
Polish
(POL)
Italian
(ITA)
Dutch
(NED)
Swedish
(SVE)
Finnish
(FIN)
WZORCUJ
CALIBRARE
TL-DIFF
IJKING
CALIBRATE
TL-SPAN
KALIBROI
TI-YLflRAJA
MITATTU
TL-ROZPIETOS
MIERZONA
T-OMG-1
GEMETEN
AANGELEGD
MISURATO
APPLICATO
MEASURED
APPLIED
STOSOWANA
SY÷TETTY
WZORCUJ NX
MIERZONE
CALIBRARE NX
MISURATO
IJKING NX-METING
GEMETEN
CALIBRATE NX
MEASURED
APPLIED
KALIBROI NX
MITATTU
STOSOWANE
APPLICATO
AANGELEGD
SY÷TETTY
WZORCUJ CO
MIERZONE
CALIBRARE CO
MISURATO
IJKING CO-METING
GEMETEN
CALIBRATE CO
MEASURED
APPLIED
KALIBROI CO
MITATTU
STOSOWANE
APPLICATO
AANGELEGD
SY÷TETTY
TESTUJ GAZ CO/H2
WARTOSC CO
PROVA GAS CO/H2
VALORE CO
CO/H2-METING
CO-GEHALTE
TESTGAS CO/H2
CO-VALUE
TESTIKAASU CO/H2
CO ARVO
WPROWADZ WARTOSC CO
IMMETTERE VALORE CO
GEEF DE WAARDE IN
ENTER CO-VALUE
SYˆ T‰ CO ARVO
WZORCUJ H2
MIERZONE
CALIBRARE H2
MISURATO
IJKING H2-METING
GEMETEN
CALIBRATE H2
MEASURED
APPLIED
KALIBROI H2
MITATTU
STOSOWANE
APPLICATO
AANGELEGD
SY÷TETTY
WZORCUJ CIAG
MIERZONY
CALIBRARE TIRAGGIO
MISURATO
IJKING TREK-METING
GEMETEN
CALIBRATE DRAFT
MEASURED
KALIBROI PAINE
MITATTU
STOSOWANY
APPLICATO
AANGELEGD
APPLIED
SYˆTETTY
UZYWACZA IMIE
LINIA n
NOME DELLíUTENTE
LINEA n
GEBRUIKERS ID
REGEL n
ANVANDAR NAMN
LINE n
K‰YTT‰J‰NIMI
RIVI n
ZLA KAL KALIBRACJA
ERRORE CALIBRAZIONE
PURGA SENSORE CO
VALSE INGAVE
BAD CALIBRATION ENTRY
RENSAR CO SENSOR
VAARA KALIBROINTIARVO
PUHDISTETAAN CO-ANTURI
PRZEDMUCH SENSORA CO
REINIGING CO SENSOR
Instruction 24-9351
A-9
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
PCA
NOTES:
A-10
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix B
APPENDIX B – Printout Translations
English
(ENG)
German
(DEU)
Danish
(DAN)
Combustion Test Data
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
SN: xxxxxx
SN: xxxxxx
SN: xxxxxx
====================
====================
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
TIME
DATE
09:03:27
28.08.1998
Zeit
09:03:27
TID09:03:27
Datum 28.08.1998
DATO
28.08.1998
FUEL
NATGAS
BRENNSTOFF
Erdgas
FUEL
N-GAS
STACK-TEMP
AMB.-TMP
O2
CO2
CO
0% COR CO
NX
0% COR NX
qA
190 °C
20.0 °C
4.0 %
Abgas-Temp
Raum-Temp
O2
CO2
CO
COunverd.
NX
NXunverd.
qA
190 °C
ROEGTEMP
LUFTTMP
O2
CO2
CO
CO UFORT
NX
NX UFORT
qA
190 °C
20.0 °C
4.0 %
9.5 %
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
1.24
20.0 °C
4.0 %
9.5 %
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
1.24
9.5 %
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
LAMBDA
1.24
Lambda
LAMBDA
DRAFT
-0.12 MB
Feinzug
-0.12 MB
TRAEK
-0.12 MB
---------------------
COMMENTS:
---------------------
Kommentar:
---------------------
KOMMENTAR:
Draft Only
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
SN: xxxxxx
SN: xxxxxx
SN: xxxxxx
====================
====================
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
TIME
DATE
13:03:45
28.08.1998
Zeit
13:03:45
TID13:03:45
Datum 28.08.1998
DATO
28.08.1998
DRAFT
FEINZUGMESSUNG
TRAEK
DRAFT
-0.37 MB
Feinzug
---------------------
Kommentar:
-0.37 MB
TRAEK
-0.37 MB
---------------------
COMMENTS:
---------------------
KOMMENTAR:
Instruction 24-9351
B-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
PCA
French
(FRA)
Spanish
(ESP)
Polish
(POL)
Combustion Test Data
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
SN: xxxxxx
SN: xxxxxx
SN: xxxxxx
====================
====================
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
HEURE 09:03:27
DATE 28.08.1998
HORA
09:03:27
CZAS
DATE
09:03:27
28.08.1998
FECHA 28.08.1998
COMBUSTBL
GAZ NATU
COMBUSTB
GNATURAL
PALIWO
GAZ ZIEN
TEMP-CHMNE
TEMP-AMBTE
O2
190 °C
20.0 °C
.0 %
TEMP-CHIME
TEMP-AMBTE
O2
190 °C
20.0 °C
4.0 %
TEMP-KOMIN
TEMP-OTOCZ
O2
190 °C
20.0 °C
4.0 %
CO2
9.5 %
CO2
9.5 %
CO2
9.5 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
0% CO COR
NX
0% NX COR
qA
0% CORR CO
NX
0% CORR NX
qA
0% KOR CO
NX
0% KOR NX
qA
LAMBDA
1.24
LAMBDA
1.24
LAMBDA
1.24
TIRAGE
-0.12 MB
TIRO
-0.12 MB
CAIG
-0.12 MB
---------------------
COMMENTRES:
---------------------
COMENTARIO:
---------------------
UWAGI:
Draft Only
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
SN: xxxxxx
BACHARACH, INC.
PCA 65
SN: xxxxxx
BACHARACH, INC.
PCA 65
SN: xxxxxx
====================
====================
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
HEURE 13:03:45
HORA
13:03:45
CZAS
DATE
13:03:45
28.08.1998
DATE
25.08.96
FECHA 28.08.1998
TIRAGE
TIRO
DRAFT
TIRAGE
---------------------
COMMENTRES:
-0.37 MB
TIRO
-0.37 MB
CIAG
-0.37 MB
---------------------
COMENTARIO:
---------------------
UWAGI:
B-2
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
PCA
Appendix B
Italian
(ITA)
Dutch
(NED)
Swedish
(SVE)
Combustion Test Data
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
BACHARACH, INC.
PCA 65
SN: xxxxxx
SN: xxxxxx
SN: xxxxxx
====================
====================
====================
[ID1: optional data]
[ID2: optional data]
[ID3: optional data]
[ID1: optional data]
[ID2: optional data]
[ID3: optional data]
[ID1: optional data]
[ID2: optional data]
[ID3: optional data]
ORA
DATA
09:03:27
28.08.1998
TIJD
09:03:27
Tid
09:03:27
DATUM 28.08.1998
Datum 28.08.1998
COMBUSTI
METANO
BRANDST.
AARDGAS
BRANSLE
N_GAS
CAMINO-TEM
AMB-TEMP
O2
190 °C
20.0 °C
4.0 %
ROOKGAS-T°
OMG.-T°
O2
190 °C
20.0 °C
4.0 %
Rokgastemp
Luft temp
O2
190 °C
20.0 °C
4.0 %
CO2
9.5 %
CO2
9.5 %
CO2
9.5 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
0% COR CO
NX
0% COR NX
qA
0% CO COR
NX
0% COR NX
qA
CO outsp.
NX
NX outsp.
qA
LAMBDA
1.24
LAMBDA
1.24
LAMBDA
1.24
TIRAGGIO
-0.12 MB
TREK
-0.12 MB
Drag
-0.12 MB
---------------------
COMMENTI:
---------------------
OPM.:
---------------------
Kommentar:
Draft Only
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
SN: xxxxxx
BACHARACH, INC.
PCA 65
SN: xxxxxx
BACHARACH, INC.
PCA 65
SN: xxxxxx
====================
====================
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
ORA
13:03:45
TIJD
13:03:45
Tid
13:03:45
DATA 28.08.1998
DATUM 28.08.1998
Datum 28.08.1998
TIRAGGIO
TREK-METING
DRAG
TRIAGGIO
---------------------
COMMENTI:
-0.37 MB
TREK
-0.37 MB
Drag
-0.37 MB
---------------------
OPM.:
---------------------
Kommentar:
Instruction 24-9351
B-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
PCA
Finnish
(FIN)
Combustion Test Data
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
SN: xxxxxx
====================
[ID1: optional data]
[ID2: optional data]
[ID3: optional data]
AIKA
09:03:27
PÄIVÄ 28.08.1998
POLTTOAI
MAAKAASU
SKA LäMPöT
YMP LäMPöT
O2
190 °C
20.0 °C
4.0 %
CO2
9.5 %
CO
12 ppm
15 ppm
10 ppm
12 ppm
8.1 %
CO HAPETON
NX
NX HAPETON
SAVUPIIPHä
ILMäYLIMää
1.24
PAINE
-0.12 MB
---------------------
KOMMENTIT:
Draft Only
[Line 1: user name]
[Line 2: user name]
[Line 3: user name]
BACHARACH, INC.
PCA 65
SN: xxxxxx
====================
ID1: [optional data]
ID2: [optional data]
ID3: [optional data]
AIKA
13:03:45
PäIVä 28.08.1998
VETO
PAINE
-0.37 MB
---------------------
KOMMENTIT:
B-4
Instruction 24-9351
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
World Headquarters
621 Hunt Valley Circle, New Kensington, PA 15068
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 1-800-736-4666
Website: www.bacharach-inc.com • E-mail: [email protected]
Printed in U.S.A.
Download from Www.Somanuals.com. All Manuals Search And Download.
|