Bacharach Carbon Monoxide Alarm 24 9351 User Manual

Portable Combustion  
Analyzer (PCA)  
INSTRUCTION 24-9351  
Operation & Maintenance  
Rev. 11 – May 2004  
O
2
4
.
0
C
O
1
2
H
L
N
D
C
2
9
.
5
C
U
1
5
G
T
q
L
A
2
0
8
.
.
0
1
T
L
A
A
1
.
9
2
0
4
P
S
1
M
E
N
U
P
C
A
Product Leadership • Training • Service • Reliability  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Contents  
Contents  
1.0 INTR ODUCTION ............................................................................ 1-1  
1.1 The Portable Combustion Analyzer ........................................... 1-1  
1.2 Displayed Data ........................................................................... 1-2  
1.3 Sensor Configurations ................................................................ 1-3  
2.0 TE CH NICAL CH AR ACTE R ISTICS ............................................ 2-1  
3.0 SE TTING UP TH E P CA................................................................. 3-1  
3.1 Scope ........................................................................................... 3-1  
3.2 Power .......................................................................................... 3-1  
3.2.1 Checking and Replacing the Batteries ........................... 3-1  
3.2.2 Using the Optional Power Supply .................................. 3-2  
3.3 Connecting the Probe ................................................................ 3-2  
3.4 Configuring the PCA ................................................................. 3-4  
4.0 OP E R ATION .................................................................................... 4-1  
4.1 Key Pad Functions ..................................................................... 4-1  
4.2 Sampling Hole Location ............................................................ 4-2  
4.3 Combustion Test ........................................................................ 4-3  
4.3.1 Analyzer Turn On and Warm Up .................................. 4-3  
4.3.2 Installing Probe in the Stack ......................................... 4-4  
4.3.3 Performing a Combustion Test ...................................... 4-5  
4.3.4 Ending a Combustion Test ............................................. 4-5  
4.3.5 Turning Off the Analyzer and Purging the CO Sensor .... 4-6  
4.4 Differential Pressure Measurement ......................................... 4-7  
4.5 Warm-up Screen ........................................................................ 4-8  
4.6 Sensor Status Screen ................................................................. 4-9  
4.7 Combustion Test Screen .......................................................... 4-10  
4.8 Fuel Selection Screen .............................................................. 4-11  
4.9 Draft Screens ........................................................................... 4-12  
4.10 Memory Directory Screen ..................................................... 4-13  
4.11 Memory to PC Screen ............................................................ 4-14  
4.12 ID Setup Screens ................................................................... 4-17  
4.13 Temperature Setup Screen ................................................... 4-19  
4.14 Draft Unit Setup Screen ....................................................... 4-20  
4.15 Language Setup Screen ......................................................... 4-21  
4.16 Display Mode Setup Screen .................................................. 4-22  
4.17 Time/Date Setup Screen ........................................................ 4-23  
4.18 Printer Setup Screen ............................................................. 4-24  
4.19 Maintenance Password Screen ............................................. 4-25  
4.20 Maintenance Screen .............................................................. 4-26  
4.21 User Name Screens ............................................................... 4-27  
Instruction 24-9351  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
PCA  
4.22 Saving Test Data ................................................................... 4-28  
4.23 Printing Test Data ................................................................. 4-29  
4.24 Clear Memory Screen ............................................................ 4-31  
4.25 Resetting the Microprocessor ................................................ 4-31  
5.0 CALIBR ATION ................................................................................ 5-1  
5.1 Sensor Check ............................................................................. 5-1  
5.2 Calibration Fixtures .................................................................. 5-2  
5.3 Calibrate Menu Screen .............................................................. 5-3  
5.4 Calibrate TA-Zero ...................................................................... 5-4  
5.5 Calibrate TA-Span ..................................................................... 5-5  
5.6 Calibrate TL-Zero ...................................................................... 5-6  
5.7 Calibrate TL-Span ..................................................................... 5-7  
5.8 Calibrate NX .............................................................................. 5-8  
5.9 Calibrate CO .............................................................................. 5-9  
5.10 Calibrate Draft ....................................................................... 5-11  
6.0 MAINTE NANCE .............................................................................. 6-1  
6.1 Routine Maintenance ................................................................ 6-1  
6.2 Disassembly ............................................................................... 6-2  
6.3 Cleaning the Probe .................................................................... 6-4  
6.4 Water Trap/Filter Assembly Maintenance ............................... 6-5  
6.5 Replacing the Particulate Filter ............................................... 6-6  
6.6 Replacing the Oxygen Sensor ................................................... 6-7  
6.7 Replacing the Nitric Oxide Sensor ........................................... 6-8  
6.7.1 Replacing the Nitric Oxide Sensor Filter ...................... 6-8  
6.7.2 Replacing the Nitric Oxide Sensor Bias Battery .......... 6-9  
6.8 Replacing the Carbon Monoxide Sensor ................................. 6-10  
6.8.1 Replacing the Carbon Monoxide Sensor Filter ........... 6-11  
6.9 Replacing the Pump Assembly ............................................... 6-12  
7.0 TR OUBLE SH OOTING .................................................................... 7-1  
7.1 Analyzer Repair ......................................................................... 7-1  
7.2 Error Codes ................................................................................ 7-2  
7.3 Troubleshooting Guide .............................................................. 7-3  
8.0 P AR TS & SE R VICE ........................................................................ 8-1  
8.1 Replacement Parts ..................................................................... 8-1  
8.2 Accessories ................................................................................. 8-2  
8.3 Sales/Service Centers ................................................................ 8-6  
AP P E NDIX A – Disp la y Scr een Tr a n sla t ion s ................................... A-1  
AP P E NDIX B – P r in t ou t Tr a n sla t ion s ..............................................B-1  
ii  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Introduction  
1.0 INTRODUCTION  
1.1 The Portable Combustion Analyzer  
The Portable Combustion Analyzer (PCA) (Figure 1-1) is a commercial  
grade, hand held, combustion efficiency analyzer that is designed for  
continuous (on demand) sampling of light industrial and residential  
furnaces, appliances, and boilers. The basic instrument is supplied with a  
probe, instruction manual, batteries, and carrying case.  
O
2
4
.
0
C
O
1
2
H
L
N
D
C
2
9
.
5
C
U
1
5
G
T
q
L
A
2
0
8
.
.
0
1
T
L
A
A
1
.
9
2
0
4
P
S
1
M
E
N
U
P
C
A
Figure 1-1. PCA  
Instruction 24-9351  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
PCA  
1.2 Displayed Data  
The PCA directly measures, displays, and stores the following data:  
Room Temperature in °C or °F (Primary Air/Ambient Temperature)  
Flue Gas Oxygen Content in %  
Flue Gas Temperature in °C or °F  
Flue Gas Carbon Monoxide Content (H2 Compensated) in ppm  
(For analyzers having a Carbon Monoxide sensor)  
Flue Gas Nitric Oxide content in ppm  
(For analyzers having a Nitric Oxide sensor)  
Pressure/Draft in Millibars, Pascals, or Inches of Water Column  
(For analyzers having a draft sensor)  
Differential Pressure in Millibars, Pascals, or Inches of Water Column  
(For analyzers having a draft sensor)  
The PCA will compute, display, and store the following data for any of the  
standard fuels:  
Stack Loss in %  
Lambda  
Flue Gas Carbon Dioxide Content in %  
Flue Gas Carbon Monoxide Content referenced to 0% Oxygen in ppm  
(For analyzers having a Carbon Monoxide sensor)  
Flue Gas Nitric Oxide Content referenced to 0% Oxygen in ppm  
(For analyzers having a Nitric Oxide sensor)  
The standard fuels are:  
Natural Gas  
Oil #2  
Oil #6  
LPG  
Koks  
Low Energy Gas  
P-Coal (available in English, German, Dutch, French, Italian,  
Polish, and Spanish languages)  
Biofuel (available in Danish, Finnish, and Swedish languages)  
The PCA continuously monitors flue-gas-exhaust conditions and updates  
the above displayed values during a combustion test. If the analyzer is  
equipped with an optional pressure sensor, draft measurements can be  
made simultaneously with the combustion test, or made separately.  
The analyzer has the ability to store data that was collected during a  
combustion test or draft measurement. The stored data can then either be  
viewed on the PCAs display, printed using an optional printer, or down-  
loaded to a computer.  
1-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Introduction  
1.3 Sensor Configurations  
TABLE 1-1. P CA SE NSOR CONF IGUR ATIONS  
PCA Models  
Sensors Installed  
Standard  
Advanced  
PCA Part No.* Part No.** PCA Part No.* Part No.** Stack Temp.,  
CO NX Draft  
Model  
10  
15  
20  
25  
24-  
24-  
Model  
40  
45  
50  
55  
24-  
24-  
Air Temp. & O2  
(P)  
7181  
7182  
7183  
7184  
7185  
7186  
7281  
7282  
7283  
7284  
7285  
7286  
7241  
7242  
7243  
7244  
7245  
7246  
7251  
7252  
7253  
7254  
7255  
7256  
X
X
X
X
X
X
X
X
X
X
X
X
X
30  
35  
60  
65  
X
X
* English, Danish, Dutch, German, Finnish & Swedish languages  
** English, French, German, Italian, Polish & Spanish languages  
PCA 10 & 40  
These basic instruments have the capability of measuring, displaying,  
and storing combustion tests. They will also display flue gas Oxygen  
content, flue gas Carbon Dioxide (CO2) content, air temperature, flue gas  
temperature, stack loss, Lambda, and the current fuel selected. The  
standard’ PCA 10 stores up to 10 combustion tests, while the advanced’  
PCA 40 can store up to 100 tests.  
PCA 15 & 45 with Draft  
In addition to the features of the basic PCAs described above, these instru-  
ments have the added capability of measuring, displaying, and saving draft  
or differential pressure in either Millibars, Pascals, or Inches-of-Water  
Column.  
PCA 20 & 50 with Carbon Monoxide Measurement  
In addition to the features of the basic PCAs described above, these  
instruments have the added capability of measuring, displaying, and  
saving Carbon Monoxide (CO) content, as well as calculating CO refer-  
enced to 0% Oxygen.  
PCA 25 & 55 with Draft and CO Measurement  
These instruments combine the features of all the PCAs listed above.  
PCA 30 & 60 with CO and Nitric Oxide Measurement  
In addition to the features of the basic PCAs with CO measurement, these  
instruments have the added capability of measuring, displaying, and  
saving Nitric Oxide (NX) content, as well as calculating NX referenced to  
0% Oxygen.  
Instruction 24-9351  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
PCA  
PCA 35 & 65 with Draft, CO & NX  
These instruments are capable of measuring, displaying, and saving all  
measurements as previously described.  
Printout Capability  
All PCAs have the ability to print the latest test data, or any of the saved  
tests, to an optional printer using HP, IrDA or RS-232 protocol.  
Advanced PCA Model Features  
Advanced’ models of the PCA contain the following features that are in  
addition to the features of their corresponding standard’ PCAs:  
100 memory locations  
RS232 output for transferring saved data to a Personal Computer  
Ability to enter three lines of user-identification information that is  
printed at the top of each printout  
Ability to enter three lines of customer-identification information that  
is printed with each test record.  
Automatic CO sensor purge on analyzers equipped with a CO sensor  
1-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Technical Characteristics  
2.0 TECHNICAL CHARACTERISTICS  
Th e P CA Dir ectly Measu r es an d Displays:  
Oxygen content in flue gas in the range of 0.1 to 20.9 % O2  
Flue gas temperature in the range of –18 to 1200 ºC (0 to 2192 ºF)  
Primary-air / ambient temperature is in the range of –18 to 999 ºC  
(0 to 999 ºF)  
Op t ion a l . . .  
Differential pressure/draft in the range of ±70.0 mb (±28" H2O)  
Carbon Monoxide* content in flue gas in the range of 0 to 4000 ppm  
Nitric Oxide* content in flue gas in the range of 0 to 1000 ppm  
Th e P CA Com pu tes an d Displays:  
(When the measured oxygen level is not above 18.8%, and the Stack (Flue  
Gas) temperature is not above 1200 ºC (2192 °F)  
Stack loss in the range of 0.1 to 99.9%  
Carbon Dioxide content in flue gas from 0.1 to a fuel dependent  
maximum value in percent  
Lambda in the range of 1 to 9.95  
Carbon Monoxide* content referenced to 0% Oxygen in the range of  
0 to 9,999 ppm on analyzers equipped with a CO sensor.  
Nitrix Oxide* content referenced to 0% Oxygen in the range of 0 to  
9,999 ppm on analyzers equipped with a NX sensor.  
Stan dar d Fu els** Available for  
Com bu stion Calcu lation s:  
Natural Gas  
Koks  
LEG  
LPG  
Oil #2  
Oil #6  
P-Coal (available in English, German, Dutch, French, Italian,  
Polish, and Spanish languages)  
Biofuel (available in Danish, Finnish, and Swedish languages)  
* For the PCA 30, 35, 60, & 65, the display can be set up to show either  
measured values of Carbon Monoxide and Nitric Oxide (CO & NX), or  
show the calculated values of these gases (CU & NU) referenced to  
0% Oxygen. In either case, all values are listed on the printout of  
analzyers equipped with a printer.  
** Custom fuels available upon request. Contact factory for details.  
Instruction 24-9351  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Characteristics  
PCA  
Nor m al Oper atin g Con dition s:  
Temperature:  
Analyzer ................... 0 to 40 ºC (32 to 104 ºF)  
Probe ......................... 800 ºC (1472 ºF) Max.  
Humidity:  
Analyzer ................... 15 to 90% Relative Humidity, Non-Condensing  
Air Pressure:  
Analyzer ................... Atmospheric  
Probe ......................... 25 mb (10" H2O) draft max at probe tip  
P er for m a n ce:  
Accuracy:  
Oxygen* .................... ±0.3% O2  
Carbon Monoxide ..... ±5% of reading or ±10 ppm, whichever is greater  
between 0 – 2000 ppm, and ±10% of reading  
between 2001 – 4000 ppm.  
Nitric Oxide .............. ±5% of reading or ±5 ppm, whichever is greater  
Flue Gas Temp. ........ ±2 ºC between 0 and 124 ºC  
(±4 ºF between 32 and 255 ºF)  
±3 ºC between 125 and 249º C  
(±6 ºF between 256 and 480 ºF)  
±4 °C between 250 and 400 °C  
(±8 ºF between 481 and 752 ºF)  
Ambient Temp. ........ ±1 °C between 0 and 100° C  
(±2 ºF between 32 and 212 ºF)  
Pressure ................... ±2% of reading or ±0.05 mb (±0.02 inches of  
Water Column), whichever is greater  
System Flow Rate:  
With probe ................ 200 cc/min minimum  
Fr on t P an el Con tr ols:  
Seven embossed push-button switches with tactile feedback (refer to  
Section 4.1)  
* Accuracy referenced in practical flue gas concentrations  
(mixtures of O2, CO2 and N2)  
2-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Technical Characteristics  
Display:  
20 character by 4 line alphanumeric LCD panel with a green backlight.  
P ower Requ ir em en ts:  
Four disposable AA alkaline batteries. Battery backup for the real-time  
clock, RAM, and bias voltage for the Nitric Oxide sensor are provided by  
internal lithium batteries. Optional AC Power Supplies (110 VAC &  
230 VAC) are also available.  
Oper atin g Tim e:  
A fresh set of four disposable AA alkaline batteries provides at least  
8 hours of continuous operation with the pump running and the backlight  
turned on.  
Wa r m Up Tim e:  
60 seconds.  
P r in ter In ter face:  
Infrared & RS-232 Communications (refer to Section 4.23).  
Mater ials:  
High impact ABS plastic case  
Polycarbonate window over the display  
Nickel plated, brass quick-connect hose fitting  
Stainless steel probe  
Dim en sion s:  
Height: .................. 215 mm (8.5 in.)  
Width: ................... 96 mm (3.8 in.) at display (75 mm [3.0 in.] at controls)  
Depth: ................... 50 mm (2 in.)  
Weigh t:  
With Batteries:..... Approximately 0.7 Kg (1.5 lbs)  
Instruction 24-9351  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Characteristics  
PCA  
Agen cy Appr ovals:  
TÜV Agency Approved (1.BImSchV - First Ordinance of the German  
Federal Emissions Law)  
Approval Number: TÜV By RgG 168  
CE declaration of conformity  
Manufacturer's name:  
Manufacturer's address:  
Bacharach, Inc.  
621 Hunt Valley Circle  
New Kensignton, PA 15068  
European operations:  
Bacharach Instruments  
Sovereign House, Queensway  
Royal Leamington Spa  
Warwickshire CV31 3J R  
United Kingdom  
Product name:  
Portable Combustion Analyzer (PCA)  
conforms to the following CE  
requirements:  
EN 50081-1, J anuary 1992 (Emissions)  
EN 50082-1, J anuary 1992 (Immunity)  
Equ ation s  
Where:  
21O2  
2
max  
×
C2 = CO  
O2 = Measured Oxygen in percent  
TA = Measured stack temperature in °C  
21  
TL = Measured primary air temperature in °C  
CO = Measured Carbon Monoxide in ppm  
NO = Measured Nitric Oxide in ppm  
C2 = Calculated Carbon Dioxide in percent  
qA = Calculated stack loss in percent  
LA = Calculated Lambda  
A2  
qA = (TA TL)×   
+ B  
(21O2  
)
21  
LA =  
CU = Calculated Carbon Monoxide referenced  
to 0% Oxygen  
21O2  
NU = Calculated Nitric Oxide referenced to 0%  
Oxygen  
CO2max, A2, and B are constants (see below)*:  
21  
=
×
×
CU CO  
21O2  
Fuel  
CO2max  
A2  
B
NG  
KOKS  
LEG  
11.8  
10.2  
13.1  
13.8  
15.4  
15.9  
18.7  
20.4  
0.66  
0.60  
0.63  
0.63  
0.68  
0.68  
0.60  
0.70  
0.009  
0.011  
0.011  
0.008  
0.007  
0.007  
0.007  
0.012  
21  
=
NU NO  
21O2  
LPG  
Oil#2  
Oil#6  
P-Coal  
Biofuel  
* Some constants vary for Danish, Italian, and Polish fuels  
2-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Setup  
3.0 SETTING UP THE PCA  
3.1 Scope  
Before using the PCA, you MUST:  
Check the batteries or plug in an optional Power Supply (Section 3.2)  
Connect the probe to the analyzer (Section 3.3)  
Check the analyzers configuration (Section 3.4)  
3.2 Power  
3.2.1 Checking and Replacing the Batteries  
A fresh set of batteries is supplied with the PCA. Install the batteries as  
described below. Check for a sufficient charge prior to each use. If a  
LOW BATTERYmessage is displayed, replace the batteries.  
1. Remove the battery cover from the back of the PCA (Figure 3-1).  
2. Remove (and properly dispose of) the old batteries.  
3. Install a new set of four AA alkaline batteries, making sure to prop-  
erly orient them as indicated by the +” and ” terminals in the  
battery compartment.  
4. Replace the battery cover.  
Figure 3-1. Battery Replacement  
Instruction 24-9351  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setup  
PCA  
3.2.2 Using the Optional Power Supply  
If an Optional Power Supply is to be used:  
1. Connect the output plug of the Optional Power Supply to the  
analyzers power supply jack (Figure 3-2).  
2. Plug the Optional Power Supply into an appropriate AC wall outlet.  
The analyzer will now operate and function normally.  
3.3 Connecting the Probe  
To attach the probe to the analyzer (Figure 3-2):  
1. Push the yellow-banded, quick-connect Flue Gas Hose (giving a slight  
twist) onto the analyzers GAS sample-inlet fitting.  
2. Push the blue-banded, quick-connect Draft Hose (giving a slight twist)  
onto the analyzers DRAFT sample-inlet fitting.  
3. Push the Flue Gas Thermocouple into the T-STACK jack (connector  
fits in only one way).  
NOTE : The PCA has a built in room-air thermocouple.  
Perform Step 4 only if the Optional Room Air/ Primary Air  
Thermocouple is used.  
4. Push the Optional Room Air/Primary Air Thermocouple into the  
T-AIR jack (connector fits in only one way).  
NOTE : In order for the PCA to correctly calculate combus-  
tion efficiency when the burners primary-air temperature is  
not the same as room temperature, the primary-air tempera-  
ture should be measured using the optional Primary Air  
Thermocouple.  
Inspect all the hoses for cracks. If any hose is defective, replace the entire  
probe assembly. Check that the water trap is dry and the filter is not  
dirty or saturated with water.  
3-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Setup  
Reset Button  
Room Air /  
Primary Air  
Thermocouple  
(Optional)  
Power Supply  
110V/60Hz  
230V/50Hz  
(Optional)  
Pressure Reference  
Port (Used in the  
Measurement of  
Differential Pressure)  
Draft Hose  
Flue Gas  
Thermocouple  
Flue Gas  
Hose  
Probe Tube  
Adjustable  
Probe Stop  
Water  
Trap / Filter  
Assembly  
Probe  
Handle  
Flue  
Gas  
Hose  
Figure 3-2. Connecting the Probe to the Analyzer  
Instruction 24-9351  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setup  
PCA  
3.4 Configuring the PCA  
The PCA is configured at the factory for the parameters shown below, but  
can be changed by following the instructions in the associated sections.  
F u n ct ion  
P a r a m et er s  
To Ch a n ge,  
R efer t o . . .  
Fuel  
Natural Gas  
ºC  
MB  
English  
CO & NX  
HR:MIN:SEC  
DD.MM.YY  
IrDA  
Section 4.8  
Section 4.13  
Section 4.14  
Section 4.15  
Section 4.16  
Section 4.17  
Section 4.17  
Section 4.18  
Temperature  
Optional Draft  
Language  
Display Mode*  
Time  
Date**  
Printer  
*
Available only on the PCA 30, 35, 60, & 65  
** The year displays as two digits on the instrument, and  
four digits on the printout.  
3-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.0 OPERATION  
4.1 Key Pad Functions  
Descriptions of the key pad functions are given below. Note that most of  
the front panel key pad buttons perform multiple functions as determined  
by what screen is being displayed at the time.  
Turns the analyzer ON and OFF. Note that there is a 5 second  
delay before the instrument actually turns OFF, thus allowing an  
operator to turn the instrument back ON by pressing the  
to prevent the accidental loss of test data.  
key  
Moves the cursor [z] in front of a menu item up through the  
displayed items. This key also increases alphanumeric values in  
screens requiring a value change.  
Moves the cursor [z] in front of a menu item down through the  
displayed items. This key also decreases alphanumeric values in  
screens requiring a value change.  
Chooses the highlighted item (the item with the cursor [z] in front  
of it) in all menus and screens. This key also causes the cursor to  
enter the number field in the Maintenance Password Screen, and  
causes the cursor to advance to the next field position in screens  
requiring multiple alphanumeric entries.  
Starts and stops a combustion test when the Combustion Test  
Screen is displayed. Pressing this key in any other screen almost  
always returns the instrument to the Combustion Test Screen.  
However, there are four situations where this key behaves as an  
enter key: 1) After entering a correct password in the Password  
Screen, press the  
2) After entering an offset or span value in any of the Calibration  
Edit Screens, press the key to store the new values. 3) After  
entering a time or date value in the Time/Date Setup Screen,  
press the key to store the new values and return the cursor to  
the left side of the display. 4) After entering text in either the ID  
Setup or User Name Screens, press the key to store the text.  
key to display the first calibration screen.  
Advances the display to the next menu screen.  
Turns the backlight ON and OFF.  
4-1  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.2 Sampling Hole Location  
The analyzer requires that a 13 mm (½ in.) diameter sampling hole be  
made in the furnace stack to accommodate the probe stop on the Probe  
and Hose Assembly.  
Locate the sampling hole downstream from the last heat exchanger,  
and upstream from any source of dilution, such as a draft diverter  
(Figure 4-1).  
IMP OR TANT! As the distance between the last heat ex-  
changer and sampling point increases, stack loss will falsely  
decrease due to heat loss by convection from the flue or stack.  
For residential and light-commercial combustion-equipment applications,  
the following recommendations are applicable:  
• Oil Gun Burners – Locate sampling  
hole at least 30 cm (12 in.) down-  
stream from the furnace breaching,  
and at least 15 cm (6 in.) upstream  
from the furnace side of the draft  
regulator.  
FLUE GAS FLOW  
(Downstream)  
DRAFT  
DIVERTER/  
REGULATOR  
• Gas Burners – Locate sampling hole  
at least 15 cm (6 in.) upstream from  
the furnace side of the draft diverter  
on gas-converted units. For gas-  
designed equipment, the probe may  
be inserted down into the flue  
SAMPLE  
POINT  
through the draft diverter or hood.  
(Upstream)  
FURNACE  
BREECHING  
Figure 4-1. Sampling Hole Location  
4-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.3 Combustion Test  
IMP OR TANT! Large rapid changes in the temperature of  
the analyzer can affect its accuracy. This is important to  
know if the analyzer is stored in a cold place (such as an  
unheated vehicle in the winter) and then taken into a warm  
furnace area. For the most accurate test results, allow the  
analyzer to warm up to room temperature before use (about  
10 minutes).  
4.3.1 Analyzer Turn On and Warm Up  
IMP OR TANT! Be sure the probe is at room temperature  
before performing the following steps.  
1. Make sure that the analyzer is properly set up per Section 3.0.  
2. Place probe in an area of fresh, ambient air; then press the analyzers  
key.  
3. Wait for the analyzer to countdown through its 60 second warmup  
period; then perform one of the following:  
If n o er r or s w er e d etected d u r in g w a r m u p , the Combustion  
Test Screen will be displayed. Skip Step 4, and go to Section 4.3.2.  
If a n er r or w a s d etected d u r in g w a r m u p , proceed with Step 4.  
4. If one or more errors were detected by the microprocessor during  
warmup, these errors will be displayed at the bottom of the Sensor  
Status Screen. Address any problems now per Section 7.2; then repeat  
this procedure starting with Step 1.  
NOTE : If the error detected is not critical to your test, the  
instrument can still perform any test not using the function  
disabled by the error.  
4-3  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.3.2 Installing Probe in the Stack  
1. After making a sampling hole in the stack (Section 4.2), and turning  
on the analyzer (Section 4.3.1), screw the probe stop supplied with the  
Probe and Hose Assembly into the sampling hole (Figure 4-2).  
2. Insert the probe through the hole in the probe stop, then position the  
probe tip inside the stack, near its center. Tighten the thumbscrew on  
the probe stop to secure the probe.  
PROBE STOP  
THUMBSCREW  
Figure 4-2. Installing the Probe  
4-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.3.3 Performing a Combustion Test  
IMP OR TANT: If the burners primary-air temperature is  
not the same as the room temperature, then be sure the  
Optional Room Air / Primary Air Thermocouple is installed  
per Section 3.3.  
1. With the Combustion Test Screen displayed and the probe installed in  
the stack, press the  
tion 4.7).  
key to start a combustion test (refer to Sec-  
2. Once all sensor readings are indicated on the screen: A) Loosen the  
thumbscrew on the probe stop. B) Move the probe in and out of the  
stack until the stacks core temperature (hot spot) is located.  
C) Tighten the thumbscrew to prevent further movement of the probe.  
Locating the highest stack temperature is very important for accurate  
combustion calculations.  
3. You can now begin burner-service procedures. The readings on the  
analyzer change quickly to show changes in burner performance.  
CAUTION  
With the Water Trap / Filter Assembly stood up on its Outlet  
End, do not let water condensate build up beyond the tip of  
the riser tube. The sensors could be damaged if water would  
enter the analyzer. Drain the water condensate after every  
combustion test (refer to Section 6.4).  
4. Pressing the  
key will save the Combustion Test Screen readings  
while a test is in progress. Moving the cursor (z) in front of the print  
( P) function using the s key, and then pressing  
will print the test  
information to an optional printer.  
4.3.4 Ending a Combustion Test  
1. Press the  
key to end a combustion test.  
WAR NING!  
Burn hazard! Allow a hot probe to cool for about 5 minutes  
before handling.  
4-5  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
CAUTION:  
Do not place a hot probe inside the instruments carrying case.  
Allow the probe to cool before storage.  
2. Loosen the thumbscrew on the probe stop; then remove the probe and  
probe stop from the stack.  
3. If data was saved during the combustion test, you can turn off the  
analyzer and review or print the stored data at a later time as de-  
scribed in Sections 4.10 and 4.23.  
4.3.5 Turning Off the Analyzer and  
Purging the CO Sensor  
Turn off the analyzer by pressing the key.  
If the key is pressed while the CO reading is 100 ppm or higher, the  
pump will automatically turn on (if not already running) to purge the  
analyzer of CO.  
IMP ORTANT! The analyzers probe must be removed from  
the stack during the purging process to allow fresh air to be  
drawn through the analyzer.  
The following message is displayed while the analyzer is being purged.  
PURGING CO SENSOR  
As soon as the CO level falls below 100 ppm, the pump turns off and the  
analyzer starts its normal 5 second turn-off sequence.  
To abort the purging process and immediately start the analyzers turn-off  
sequence, press the key.  
NOTE : Turning the analyzer off initiates a 5-second delay,  
during which time the unit can be turned on again without  
any warmup time. You can turn the analyzer back on during  
this 5-second delay by pressing the  
key.  
4-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.4 Differential Pressure Measurement  
The difference in pressure (P) between two areas can be measured by  
using the PCAs two pressure ports and DRAFT Screen. By using Pres-  
sure Port 2 (–) as the reference, the pressure applied to Port 1 (+) will be  
displayed on the DRAFT Screen as the differential pressure between the  
two ports.  
1. Turn on the analyzer by pressing the key; wait for the warmup cycle  
to complete; then press the MENU key until the first DRAFT Screen is  
displayed (refer to Section 4.9). If a TA-SENSOR ERROR is displayed  
because the probes thermocouple is not plugged into the analyzer, then  
press the  
key to acknowledge the error before pressing the MENU  
key.  
2. While the first DRAFT Screen is displayed, remove any hoses con-  
nected to Pressure Ports 1 and 2; then press the  
ports at atmospheric pressure.  
key to zero these  
3. Connect two sampling hoses to  
DRAFT  
2.25 MB  
HOT SPOT ---- °C P  
«S  
Pressure Ports 1 and 2  
(Figure 4-3). Then place the open  
end of each hose into the areas  
being measured.  
DRAFT  
4. The differential pressure between the two areas is now displayed on  
the third DRAFT Screen. If the pressure at Port 1 is higher than  
Port 2, then the pressure difference will be positive. But if the pres-  
sure at Port #1 is lower, then the pressure difference will be negative.  
The reading shown in this example indicates that the pressure at  
Port 1 is 2.25 mb lower than the pressure at Port 2.  
Sampling Hoses with quick-  
connect fittings are available  
as an optional accessory.  
Refer to Section 8.2.  
Pressure Port 2(–)  
Pressure Port 1(+)  
Reference  
Figure 4-3. Differential Pressure Hose Connections  
4-7  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.5 Warmup Screen  
BACHARACH, INC.  
PCA xx  
WARMUP yy  
Where: xx = Instrument Model Number  
yy = Counts down from 60 seconds  
As soon as the key is pressed, the instruments serial number and  
software version number are displayed for approximately 3 seconds. To  
continuously display these items, hold down the key at start-up. The  
warmup cycle continues after the key is released.  
The Warmup Screen is displayed during the analyzers 60 second warmup  
cycle, during which time the Warmup” value (yy) counts down to zero.  
After the warmup cycle is complete (and if the unit is working correctly)  
the instrument will flash NO ERRORS DETECTEDand go directly to  
the Combustion Test Screen (Section 4.7). If there is a problem, however,  
with one or more of the sensors, the Sensor Status Screen (Section 4.6) is  
displayed with error message(s) appearing at the bottom of the screen.  
Front Panel Key Functions:  
No Action  
No Action  
No Action  
No Action  
No Action  
– Toggle Backlight ON/OFF*  
– Turn analyzer OFF*  
* The  
button will always turn the backlight on and off, and the key will  
always turn the analyzer on and off. These two keys will not be mentioned  
in the remainder of this section.  
4-8  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.6 Sensor Status Screen  
BACHARACH, INC.  
PCA xx  
WARMUP 0  
z
Where: xx = Instrument Model Number  
z = Sensor(s) in error  
If there is problem with one or more of the sensors, the Sensor Status  
Screen will be displayed after the analyzer has gone through its warmup  
cycle (refer to Section 7.2 for a listing of the error codes).  
Front Panel Key Functions:  
No Action  
No Action  
No Action  
– Go to Combustion Test Screen  
No Action  
4-9  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.7 Combustion Test Screen  
PCA models 10–25, 40–55  
PCA models 30, 35, 60, & 65  
O2 4.0 CO 12 HLD  
C2 9.5 CU 15 NG  
O2 4.0 CO 12 HLD  
C2 9.5 NX  
TL 20.0 TA 190  
TL 20.0 TA 190  
P
P
qA 8.1 LA 1.24 «S  
qA 8.1 LA 1.24 «S  
- OR -  
This screen shows:  
O2 4.0 CU 15 HLD  
C2 9.5 NU 12 NG  
TL 20.0 TA 190  
O2 ............... Oxygen content in flue gas (%)  
C2 ............... Carbon Dioxide content  
present in flue gas (%)  
TL ............... Primary/Ambient air temp. (°F)  
qA ............... Stack Loss  
P
qA 8.1 LA 1.24 «S  
CO* ............. Carbon Monoxide content in flue gas (ppm)  
CU* ............. Carbon Monoxide content referenced to 0% Oxygen (ppm)  
TA ............... Stack (Flue gas) temperature (°F)  
LA ............... LAMBDA  
NX* ............. Nitric Oxide content in flue gas (ppm)  
NU* ............ Nitric Oxide content referenced to 0% Oxygen (ppm)  
HLD/RUN .. PCA on hold / PCA running test  
NG .............. Fuel code for natural gas (see Section 4.8 for other codes)  
P .................. Print Data  
S .................. Save Data  
* For PCA models 30, 35, 60 and 65, you have the option of displaying either  
CO & NX, or CU & NU. Refer to Section 4.16 for setup instructions.  
NOTE: Refer to Section 7.3 if stars (****) , dashes (----) , or  
Xs (XXXX) appear in the display.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Save or Print screen data  
– Run test / Stop test  
– Go to Fuel Selection Screen  
4-10  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.8 Fuel Selection Screen  
«NATGAS  
KOKS  
LEG  
FUEL  
OIL NO.2  
OIL NO.6  
P-COAL  
LPG  
This screen is displayed by pressing the MENU key from the Combustion  
Test Screen. This screen is used to select the fuel being burned.  
To select a fuel, first use the stkeys to move the cursor (z) in front of the  
desired fuel, and then press the  
key.  
NOTE : The fuel selected is saved as the default, and re-  
mains in memory after the PCA is turned off.  
The fuel codes as displayed in the Combustion Test Screen:  
NG = Natural Gas  
KOK = Coal Gas  
LPG = Propane  
O#2 = Oil No. 2  
O#6 = Oil No. 6  
PC = P-Coal (English, German, Dutch,  
French, Italian, Polish, and Spanish) or Biofuel  
(Danish, Finnish, and Swedish)  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select Fuel  
– Go to Combustion Test Screen  
– Go to Draft Screens  
4-11  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.9 Draft Screens  
The first Draft Screen is displayed by  
repeatedly pressing the MENU key  
from the Combustion Test Screen.  
DRAFT  
DISCONNECT DRAFT  
HOSE  
PRESS ±  
To measure draft, first zero the  
analyzers pressure sensor to atmo-  
spheric pressure by disconnecting the  
draft hose from the bottom of the  
instrument, and then pressing the  
key. Reconnect the draft hose after  
the second Draft Screen appears  
(shown for 3 seconds). The third  
screen shows the current values of  
draft and stack temperature as  
measured by the analyzer.  
DRAFT  
RECONNECT DRAFT  
HOSE  
DRAFT  
0.25 XX  
DRAFT  
HOT SPOT  
190 °C P  
«S  
Where: xx = Unit of measure.  
Default is millibars (MB). See  
Optional Draft SETUP Screen  
(Section 4.14) for other choices.  
When using the analyzer to make a  
differential pressure measurement  
(Section 4.4), the differential pressure  
value will be displayed on the third  
Draft Screen.  
To save (S) or print (P) the screen  
data, first use the st keys to move  
the cursor (z) in front of the desired  
function, and then press the  
key.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Save or Print screen data  
– Go to Combustion Test Screen  
– Go to Memory Directory Screen  
4-12  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.10 Memory Directory Screen  
‘Standard’ PCA Screen  
‘Advanced’ PCA Screen  
MEMORY DIRECTORY  
MEMORY DIRECTORY  
«M8 28.7.97 15:45  
M9 MEMORY EMPTY  
CLEAR MEMORY  
«98 28.7.97 15:45  
99 MEMORY EMPTY  
CLEAR MEMORY  
The Memory Directory Screen is displayed by repeatedly pressing the  
MENU key from the Combustion Test Screen. This screen is used to select  
a memory location that contains saved data which an operator can re-  
view.  
NOTE: A standard’ PCA has 10 memory locations num-  
bered M0 thru M9, while an advanced PCA has 100 memory  
locations numbered 0 thru 99.  
To select a data-memory location, first use the st keys to move the  
cursor (z) in front of the desired memory location; then press the  
key.  
The saved data is now displayed in either the Combustion Test Screen or  
Draft Screen, depending on whether the chosen memory location contains  
combustion or draft information. To print the saved data, refer to Sec-  
tion 4.23.  
After viewing or printing the saved data, use the st keys to move the  
cursor (z) to the exit (E) function; then press  
. This will redisplay the  
memory directory.  
Selecting the CLEAR MEMORYfunction displays the Clear Memory  
Screen from where all saved data can be erased (refer to Section 4.24).  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Display the data saved at the chosen memory location  
– Go to Combustion Test Screen  
– Go to Temperature Setup Screen, or the  
Memory To PC Screen for Advanced units  
4-13  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.11 Memory to PC Screen  
(For ‘Advanced’ PCA Models 40, 45, 50, 55, 60 & 65)  
MEMORY TO PC  
«TRANSMIT DATA  
CLEAR MEMORY  
The Memory To PC Screen is displayed by repeatedly pressing the MENU key  
from the Combustion Test Screen. Use this screen to either transmit all  
stored memory locations to a computer, or clear all memory locations.  
TRANSMIT DATA  
Before data can be transmitted to a personal computer, the PCAs RS-232  
output must first be connected to an unused COM port on the computer  
using serial data cable Part No. 24-1073 (see Figure 4-4). Also, a commu-  
nications program (i.e., ProcommPlus®, Windows 3.x Terminal, or Win-  
dows 9x Hyper Terminal) must be installed, and its communications  
parameters configured for: 9600 ba u d , 8 d a ta bits, 1 stop bit, n o p a r ity,  
a n d n o h a n d sh a k in g.  
A
C
u
t
t
a
c
h
D
B
9
C
o
n
n
e
c
t
o
r
t
o
c
o
m
p
u
t
e
r
'
s
O
M
1
o
r
C
O
M
2
p
o
r
t
.
I
f
n
e
c
e
s
s
a
r
y
,
s
e
a
2
5
-
p
i
n
t
o
9
-
p
i
n
a
d
a
p
t
e
r
.
S
P
e
a
r
r
i
t
a
l
D
o
a
t
a
C
a
b
l
3
e
N
.
2
4
-
1
0
7
R
S
-
2
3
2
C
o
n
n
e
c
t
o
r
Figure 4-4. Connecting the Serial Data Cable  
4-14  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
Data is transmitted to a computer in ASCII comma-delimited format,  
which can be captured as a text file and then opened in most commer-  
cially available spreadsheet programs. Note that each data record con-  
sists of 20 fields, some of which may be blank for different tests and PCA  
models as listed in Tables 4-1 & 4-2.  
Use the communication software to capture and save the received data as  
an ASCII text file. Consult the softwares documentation for detailed  
instructions on how to perform this procedure.  
To start transmitting data, first use the st keys to position the cursor (z)  
in front of TRANSMIT DATAand then press the  
key. Observe that  
as PCA downloads its data, the word TRANSMITTINGappears on the  
display.  
CLEAR MEMORY  
To clear all memory locations, first use the st keys to position the cursor  
(z) in front of CLEAR MEMORYand then press the  
key. The Clear  
Memory Screen will then appear from where all saved data can be erased  
(refer to Section 4.24).  
Front Panel Key Functions:  
– Toggle cursor (z) position  
– Toggle cursor (z) position  
– Select function next to cursor  
– Go to Combustion Test Screen  
– Go to ID Setup Screen  
4-15  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
TABLE 4-1. COMMA-DE LIMITE D F IE LDS  
Field  
Data Name or Value  
Label in Column  
Headings  
1
2
3
4
Instrument serial number  
ID line 1 (up to 16 characters)  
ID line 2 (up to 16 characters)  
ID line 3 (up to 16 characters)  
Time of test (hh:mm:ss)  
Date of test (dd.mm.yyyy)  
Name of fuel (up to 16 characters)  
Flue gas temperature  
SN  
ID1  
ID2  
ID3  
5
TIME  
DATE  
FUEL  
TA  
6
71  
81  
91  
Air temperature  
TL  
101  
111  
121  
131,2  
141,2  
151,3  
161,3  
171  
181  
19  
20  
Temperature unit of measure (F or C)  
O2 concentration in %  
C/F  
O2  
C2  
CO  
CU  
NX  
NU  
qA  
CO2 concentration in %  
CO concentration in ppm  
CO referenced to 0% O2 in ppm  
NO concentration in ppm  
NO referenced to 0% O2 in ppm  
qA in %  
Lambda  
Draft measurement  
Draft unit of measure  
LA  
DR  
MB/PA/WC  
1
Empty data field for draft tests  
Empty data field for PCA Models 40 and 45  
Empty data field for PCA Models 40, 45, 50, and 55  
2
3
TABLE 4-2. TYP ICAL SP R E ADSH E E T F OR A P CA 65  
SN  
ID1  
ID2  
ID3  
TIME  
DATE  
FUEL  
TA TL C/F O2  
AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:03:27 19.01.1999 NATGAS 190 20  
AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:10:35 19.01.1999  
C
4
C2  
CO  
CU  
NX  
NU  
qA  
LA  
DR  
MB/PA/WC  
9.5  
12  
15  
10  
12  
8.1  
1.24  
-0.25 MB  
-0.25 MB  
Line 1: Column Headings  
Line 2: Typical Combustion Readings  
Line 3: Typical Draft Reading  
4-16  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.12 ID Setup Screens  
(For ‘Advanced’ PCA Models 40, 45, 50, 55, 60 & 65)  
SETUP  
«ID #1  
ID #2  
ID #3  
This initial ID Setup Screen is displayed by repeatedly pressing the MENU  
key from the Combustion Test Screen. Use this screen to edit three lines  
of customer information (i.e., the customers name, location, and burner  
reference number).  
Each ID line can be up to 16 alphanumerical characters in length. All  
three lines will appear at the top of each test record for the purpose of  
identifying individual tests.  
Front Panel Key Functions for the Initial ID SETUP Screen:  
– Move cursor (z) upward  
– Move cursor (z) downward  
– Select ID Number that is next to the cursor for editing  
– Go to Combustion Test Screen  
– Go to Temperature Setup Screen  
To enter a line of text, first use the st keys to position the cursor (z) in  
front of the desired ID line; then press  
Screen will then appear.  
. The selected ID Line Number  
SETUP  
ID #1  
«
Now press the st keys until the desired letter or number is displayed.  
Available characters include:  
“(space)ABCDEFGHIJKLMNOPQRSTUVWXYZüäöß0123456789”  
4-17  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
Press  
to save the selected character and advance to the next position.  
until the cursor is over the incorrect  
character and make your correction by again using the st keys. After all  
If you make a mistake, press  
the desired characters have been selected, press the  
text line and return to the initial ID SETUP Screen.  
key to save the  
NOTE: The entered ID information will be saved with all  
future memory records until it is modified or deleted.  
Front Panel Key Functions for the Individual ID SETUP Screens:  
– Increment character  
– Decrement character  
– Select the displayed character and advance to the next  
character position  
– Save the text line and return to the initial ID SETUP Screen  
– Abort any changes to the text line and return to the initial  
ID SETUP Screen  
4-18  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.13 Temperature Setup Screen  
SETUP  
TEMPERATURE UNIT «°C  
°F  
The Temperature Setup Screen is displayed by repeatedly pressing the  
MENU key from the Combustion Test Screen. Use this screen to setup the  
analyzer to display temperature in either °C or °F.  
To select the instruments temperature unit-of-measure, first use the st  
keys to move the cursor (z) in front of °Cor °F, and then press the  
key.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select temperature unit of measure next to cursor  
– Go to Combustion Test Screen  
– Go to Draft Unit Setup Screen  
4-19  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.14 Draft Unit Setup Screen  
SETUP  
DRAFT UNIT  
«MB  
PA  
WC  
The Draft Unit Setup Screen is displayed by repeatedly pressing the  
MENU key from the Combustion Test Screen. Use this screen to setup the  
analyzer to display draft in either millibars (MB), Pascals (PA), or inches-  
of-water column (WC).  
To select the draft unit-of-measure, first use the st keys to move the  
cursor (z) in front of MB, PAor WC, and then press the  
key.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select draft unit of measure  
– Go to Combustion Test Screen  
– Go to Language Setup Screen  
4-20  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.15 Language Setup Screen  
SETUP  
SETUP  
LANGUAGE DAN NED  
DEU SVE  
LANGUAGE DEU FRA  
«ENG ITA  
- OR -  
«ENG FIN  
ESP POL  
The Language Setup Screen is displayed by repeatedly pressing the MENU  
key from the Combustion Test Screen. Use this screen to select the  
language displayed on the analyzer. The languages available for selection  
include: Danish, German, English, Dutch, Swedish and Finnish; or  
German, English, Spanish, French, Italian and Polish. The languages  
displayed depend on the model of the analyzer (refer to Table 1-1).  
To select a language, first use the st keys to move the cursor (z) in front  
of the desired language, and then press the  
key.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select Language  
– Go to Combustion Test Screen  
– Go to Display Mode Setup Screen  
4-21  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.16 Display Mode Setup Screen  
(For PCA Models 30, 35, 60 & 65)  
SETUP  
DISPLAY  
«CO NX  
CU NU  
The Display Setup Screen is displayed by repeatedly pressing the MENU  
key from the Combustion Test Screen. Use this screen to select whether  
the Combustion Test Screen will display the measured values of Carbon  
Monoxide and Nitric Oxide (CO and NX), or the calculated values of these  
gases (CU and NU) referenced to 0% Oxygen.  
To setup the display, first use the st keys to move the cursor (z) in front  
of either CO NX(measured values), or CU NU(calculated values), and  
then press the  
key.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select Display Mode  
– Go to Combustion Test Screen  
– Go to Time/Date Setup Screen  
4-22  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.17 Time/Date Setup Screen  
SETUP  
«TIME 15:45:06  
DATE 31.10.01  
The Time/Date Setup Screen is displayed by repeatedly pressing the  
MENU key from the Combustion Test Screen. Use this screen to enter the  
current time and date.  
To enter the correct time or date, first use the st keys to move the  
cursor (z) in front of the function you wish to change. Each position in the  
TIMEor DATEnumber fields can then be changed by first pressing the  
key to move the cursor into the desired position, then pressing the st  
keys to increase or decrease the value. Pressing  
moves the cursor to  
the next position in the number field. Each individual position can be  
edited in the same manner. Once the time or date values have been  
entered, press  
of the screen.  
to save the values and return the cursor to the left side  
Front Panel Key Functions:  
– Move cursor (z) up, or Increase value in number fields  
– Move cursor (z) down, or Decrease value in number fields  
– Select Time or Date to be changed, or move cursor (z) to next  
position in the number field  
– Go to the Combustion Test Screen, or save the time and date  
values and return the cursor to the left side of the display  
– Go to Printer Setup Screen  
4-23  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.18 Printer Setup Screen  
SETUP  
«IR - HP  
IR - IRDA  
RS232  
PRINTER  
The Printer Setup Screen is displayed by repeatedly pressing the MENU  
key from the Combustion Test Screen. Use this screen to choose the type  
of connection and printer being used.  
IR-HP:  
Infrared connection to a printer manufactured by Hewlett  
Packard, which uses their proprietary infrared communica-  
tions protocol  
IR-IRDA: Infrared connection to a printer that uses a standard IrDA  
protocol  
RS232:  
Cable connection between the PCA and any serial printer  
capable of 9600 baud operation  
Use the st keys to move the cursor (z) in front of the desired connection  
and printer, and then press the  
to the Combustion Test Screen.  
key to make the selection and return  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select connection and printer next to cursor  
– Go to Combustion Test Screen  
– Go to Maintenance Password Screen  
4-24  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.19 Maintenance Password Screen  
MAINTENANCE  
«PASSWORD  
XXX  
Where: xxx = Password number  
The Maintenance Password Screen is displayed by repeatedly pressing  
the MENU key from the Combustion Test Screen. From here a three-digit  
password must be entered to access the instruments Maintenance  
Screens. The password number is provided on the Portable Combustion  
Analyzer Calibration Password card that was supplied with the analyzer.  
To enter the password, first press the  
the first number field, and then press the st keys until the first digit of  
the password is displayed. Press to advance to the next number field  
and enter the second digit. Perform the same sequence a third time to  
complete the password. Press the key after the correct password is  
key to move the cursor (z) into  
entered to display the Maintenance Screen.  
Front Panel Key Functions:  
No action, or increase value in password number field  
No action, or decrease value in password number field  
– Move cursor (z) to next position in password number field  
– Go to Combustion Test Screen (if cursor is on left side of  
screen), or go to Maintenance Screen (if the proper password  
was entered), or return cursor to left side of screen (if the  
wrong password was entered)  
– Go to Combustion Test Screen (if cursor is on left side of  
screen), or return cursor to left side of screen (if cursor is in  
the password number field)  
4-25  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
4.20 Maintenance Screen  
MAINTENANCE  
«CALIBRATION  
USER NAME  
The Maintenance Screen is displayed after entering the correct password  
in the Maintenance Password Screen (Section 4.19). Use this screen to  
enter either the analyzers Calibration Screen or User Name Screen.  
To enter the Calibration Screen, first use the st keys to position the  
cursor (z) in front of CALIBRATION, and then press the  
key.  
NOTE : Section 5.0 contains detailed calibration procedures.  
To enter the User Name Screen, first use the st keys to position the  
cursor (z) in front of USER NAME, and then press the  
key.  
Front Panel Key Functions:  
– Toggle cursor (z) position  
– Toggle cursor (z) position  
– Select function next to cursor  
– Go to Combustion Test Screen  
No action  
4-26  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.21 User Name Screens  
USER NAME  
«LINE 1  
LINE 2  
LINE 3  
This initial User Name Screen is displayed after selecting USER NAME  
from the Maintenance Screen (Section 4.20). Use this screen to either  
enter or edit three lines of user-name information.  
Each user-name line can be up to 20 alphanumerical characters in length.  
All three lines will appear at the top of each printout for the purpose of  
identifying the user or owner of the instrument (i.e., your companys  
name and address).  
Front Panel Key Functions for Initial User Name Screen:  
– Move cursor (z) upward  
– Move cursor (z) downward  
– Select Line Number that is next to the cursor for editing  
– Go to Combustion Test Screen  
– Return to Maintenance Screen  
To enter text, first use the st keys to position the cursor (z) in front of  
the desired line number; then press  
Number Screen will then appear.  
. The selected User Name Line  
USER NAME  
LINE 1  
«
Now press the st keys until the desired letter or number is displayed.  
Available characters include:  
“(space)ABCDEFGHIJKLMNOPQRSTUVWXYZüäöß0123456789”  
4-27  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
Press  
to save the selected character and advance to the next position.  
If you make a mistake, press  
until the cursor is over the wrong charac-  
ter and make your correction by again using the st keys.  
After all the desired characters have been selected, press  
text line and return to the initial User Name Screen.  
to save the  
Front Panel Key Functions for Individual User Name Screens:  
– Increment character  
– Decrement character  
– Select the displayed character and advance to the next  
character position  
– Save the text line and return to the initial User Name Screen  
– Abort any changes to the text line and return to the initial  
User Name Screen  
4.22 Saving Test Data  
O2 4.0 CO 12 HLD  
C2 9.5 CU 15 NG DRAFT  
DRAFT  
0.25 MB  
TL 20.0 TA 190 P HOT SPOT 190 °C P  
qA 8.1 LA 1.24 «S  
«S  
To save the data displayed in either the Combustion Test or Draft  
Screens, first use the st keys to move the cursor (z) in front of the save  
(S) function and then press the key. The data will be saved in  
memory, and can be recalled at any time from the Memory Directory  
Screen (Section 4.10).  
NOTE: Data will be automatically stored in the next free  
memory location. After all memory locations are filled, any  
additional data that is saved will start overwriting previ-  
ously saved data starting at the first memory location.  
4-28  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.23 Printing Test Data  
O2 4.0 CO 12HLD  
C2 9.5 CU 15 NG  
TL 20.0 TA 190 «P  
qA 8.1 LA 1.24 S  
DRAFT  
0.25 MB  
HOT SPOT 190 °C «P  
S
DRAFT  
Before printing, ensure that the correct  
connection and printer has been selected per  
Section 4.18.  
The Print function is available in either the  
Combustion Test Screen or the Draft Screen*.  
NOTE : The data which is stored in  
memory can also be printed. First go to  
the Memory Directory Screen (Section  
4.10) and display the data to be printed;  
then print the data as described below.  
0
I
0
When using an infrared printer:  
4
5
c
m
(
1
8
i
n
.
)
6
0
°
m
a
x
.
1. Place analyzer in-line with the printers  
IR input (see Figure 4-5).  
m
a
x
.
2. Use the st keys to move the cursor (z) in  
front of the print (P) function.  
3. Press the  
key to start printing.  
M
E
N
U
P
C
A
* The HOT SPOT line shown in the Draft  
Screen does not appear on the printout.  
Figure 4-5. Aligning the Printer  
4-29  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
When using a serial printer:  
1. First connect the analyzer to the printer using the optional RS-232  
cable (see Figure 4-6).  
2. Set the printers communication parameters to 9600 baud, 8 data bits,  
1 stop bit, no parity, and no handshaking.  
3. Use the st keys to move the cursor (z) in front of the print (P)  
function.  
4. Press the  
key to start printing.  
A
s
u
t
t
a
c
h
D
B
9
C
o
n
n
e
c
t
o
r
t
o
p
r
i
n
t
e
r
'
s
e
r
e
i
a
l
i
n
p
u
t
p
o
r
t
.
I
f
n
e
c
e
s
s
a
r
y
,
s
a
2
5
-
p
i
n
t
o
9
-
p
i
n
a
d
a
p
t
e
r
.
S
P
e
a
r
r
i
t
a
l
D
o
a
t
a
C
a
b
l
3
e
N
.
2
4
-
1
0
7
R
S
-
2
3
2
C
o
n
n
e
c
t
o
r
Figure 4-6. Connecting a Serial Printer to the Analyzer  
4-30  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Operation  
4.24 Clear Memory Screen  
CLEAR MEMORY  
C
«E  
The Clear Memory Screen is accessed from either the Memory Directory  
Screen (Section 4.10) or the Memory to PC Screen (Section 4.11).  
To clear a ll memory locations, use the st keys to place the cursor (z) in  
front of the clear (C) function, and then press the  
key.  
To return to the previous screen without clearing any memory locations, use  
the st keys to place the cursor (z) in front of the exit (E) function; then  
press  
.
4.25 Resetting the Microprocessor  
If the analyzer locks-up” and cannot be turned OFF, reset the micropro-  
cessor by pressing the RESET button (Figure 4-7). The button can be  
activated using the end of a paper clip.  
RESET  
BUTTON  
Figure 4-7. Reset Button  
4-31  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
PCA  
NOTES:  
4-32  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.0 CALIBRATION  
NOTE : Bacharach recommends that the PCA be calibrated  
by your nearest Bacharach Service Center. Calibration,  
however, can be performed in the field if your facility has the  
necessary equipment and qualified personnel to perform the  
procedures described in the sections that follow.  
IMP OR TANT! To prevent the loss of data during calibra-  
tion, perform the following procedures with fresh batteries, or  
using an optional Power Supply (see Section 3.2).  
5.1 Sensor Check  
IMP OR TANT! Before turning on the analyzer or perform-  
ing any of the calibration procedures, ensure that the ana-  
lyzer will be sampling fresh air, and that the probe is at  
room temperature.  
When the analyzer is first turned on and allowed to cycle through its  
60 second warmup period, and while sampling fresh air, the sensors are  
checked (read) and calibrated (set) to the following ambient conditions:  
Oxygen sensor is spanned to 20.9%  
Carbon Monoxide sensor (if installed) is zeroed  
Nitric Oxide sensor (if installed) is zeroed  
Pressure sensor (if installed) is zeroed  
Instruction 24-9351  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
5.2 Calibration Fixtures  
A gas and a draft fixture will be required to perform the various calibra-  
tion procedures described in this manual.  
Material Required:  
Calibration Kit (Refer to Section 8.2)  
Calibration Gas Cylinder (Refer to Section 8.2)  
Bellows  
Micromanometer  
Procedure:  
Assemble the appropriate fixture, shown in Figure 5-1, as required by the  
calibration procedure being performed.  
Leave Top  
Port Open  
GAS  
FIXTURE  
2
1
3
4
3
5
M
E
N
U
3
P
C
A
Parts Shown:  
6
1. Gas Cylinder  
2. Regulator*  
3. Tubing*  
DRAFT  
FIXTURE  
4. Tee*  
5. Flowmeter*  
6. Fitting, Gas*  
7. Fitting, Draft*  
* Contained in Calibration Kit  
M
E
N
U
P
C
A
3
4
3
Bellows  
7
Micromanometer  
Figure 5-1. Calibration Fixtures  
5-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.3 Calibrate Menu Screen  
«TA-ZERO  
TA-SPAN  
TL-ZERO  
TL-SPAN  
CALIBRATE  
NX  
CO  
DRAFT  
The Calibrate Menu Screen is displayed after entering the correct pass-  
word in the Maintenance Password Screen (Section 4.19) and selecting  
CALIBRATION from the Maintenance Screen (Section 4.20). Use this  
screen to select the sensor to be calibrated.  
Press the st keys until the cursor (z) is in front of the desired function,  
and then press the  
key.  
NOTE : TA is the stack temperature sensor, while TL is the  
optional Room Air/ Primary Air temperature sensor.  
NOTE : If a sensor is not installed (i.e., the instrument does  
not have a Nitric Oxide sensor or thermocouple simulator  
installed), the corresponding calibration screen will not be  
displayed.  
Front Panel Key Functions:  
– Move cursor (z) up  
– Move cursor (z) down  
– Select sensor to be calibrated  
– Go to Combustion Test Screen  
– Return to Maintenance Screen  
Instruction 24-9351  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
5.4 Calibrate TA-Zero  
Material Required:  
Thermocouple Simulator (K-type) Range: 0 to 300°C  
Accuracy: ±0.3°C  
Procedure:  
1. With the analyzer turned off, first plug the simulator's K-type connec-  
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and  
wait for its warmup cycle to complete.  
2. Enter the Calibration Menu Screen per Section 5.3; then choose  
TA-ZERO to zero the analyzers stack temperature channel.  
3. Adjust the simulator to 0° C (32° F).*  
4. Wait until the MEASURED reading on the screen stabilizes. Then use  
the st and  
keys to enter an APPLIED value that equals 0 °C  
(32 °F).  
Typical Calibrate TA-Zero Screen During Calibration Procedure:  
CALIBRATE TA-ZERO  
MEASURED  
APPLIED  
3.0°C  
0000.0°C  
4. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 0 – 5 °C (32 – 41°F). Any  
attempt to calibrate outside this range will cause the analyzer to  
display an error message.  
5-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.5 Calibrate TA-Span  
Material Required:  
Thermocouple Simulator (K-type) Range: 0 to 300°C  
Accuracy: ±0.3°C  
Procedure:  
1. With the analyzer turned off, first plug the simulator's K-type connec-  
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and  
wait for its warmup cycle to complete.  
2. Enter the Calibration Menu Screen per Section 5.3; then choose  
TA-SPAN to span the analyzers stack temperature channel.  
3. Set the simulator to 300° C (572° F).*  
4. Wait until the MEASURED reading on the screen stabilizes. Then use  
the st and  
keys to enter an APPLIED value that equals 300 °C  
(572 °F).  
Typical Calibrate TA-Span Screen During Calibration Procedure:  
CALIBRATE TA-SPAN  
MEASURED  
APPLIED  
295.0°C  
0300.0°C  
4. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 270 – 330 °C (518 – 626 °F).  
Any attempt to calibrate outside this range will cause the analyzer to  
display an error message.  
Instruction 24-9351  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
5.6 Calibrate TL-Zero  
Material Required:  
Thermocouple Simulator (K-type) Range: 0 to 300°C  
Accuracy: ±0.3°C  
Procedure:  
1. With the analyzer turned off, first plug the simulators K-type connec-  
tor into the T-AIR jack (Figure 3-2); then turn on the analyzer and  
wait for its warmup cycle to complete.  
2. Enter the Calibration Menu Screen per Section 5.3; then choose  
TL-ZERO to zero the analyzers room-air/primary-air temperature  
channel.  
3. Set the simulator to 0 °C (32 °F).*  
4. Wait until the MEASURED reading on the screen stabilizes. Then use  
the st and  
keys to enter an APPLIED value that equals 0 °C  
(32 °F).  
Typical Calibrate TL-Zero Screen During Calibration Procedure:  
CALIBRATE TL-ZERO  
MEASURED  
APPLIED  
3.0°C  
0000.0°C  
4. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 0 – 5 °C (32 – 41 °F). Any  
attempt to calibrate outside this range will cause the analyzer to  
display an error message.  
5-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.7 Calibrate TL-Span  
Material Required:  
Thermocouple Simulator (K-type) Range: 0 to 300°C  
Accuracy: ±0.3°C  
Procedure:  
1. With the analyzer turned off, first plug the simulators K-type connec-  
tor into the T-AIR jack (Figure 3-2); then turn on the analyzer and  
wait for its warmup cycle to complete.  
2. Enter the Calibration Menu Screen per Section 5.3; then choose  
TL-SPAN to span the analyzers room-air/primary-air temperature  
channel.  
3. Set the simulator to 100 °C (212 °F).*  
4. Wait until the MEASURED reading on the screen stabilizes. Then use  
the st and  
keys to enter an APPLIED value that equals 100 °C  
(212 °F).  
Typical Calibrate TL-Span Screen During Calibration Procedure:  
CALIBRATE TL-SPAN  
MEASURED  
APPLIED  
102.0°C  
0100.0°C  
4. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 90 – 110 °C (194 – 230 °F).  
Any attempt to calibrate outside this range will cause the analyzer to  
display an error message.  
Instruction 24-9351  
5-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
5.8 Calibrate NX  
(For PCA Models 30, 35, 60 & 65)  
The Nitric Oxide sensor needs to be spanned at regular intervals to  
determine that it still meets its accuracy specification. Because of the  
toxicity of Nitric Oxide gas, however, unless your facility has the neces-  
sary gas cylinders and personnel trained in the handling of toxic gases,  
we recommend that the Nitric Oxide sensor be spanned by an authorized  
Bacharach Service Center.  
Material Required:  
Calibration Gas Fixture (Section 5.2)  
Gas Cylinder, 50 to 150 ppm Nitric Oxide with an analytical accuracy of  
±1% (customer supplied)  
Procedure:  
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to  
calibrate the NX sensor.  
2. At the conclusion of Step 1 the pump should start running.  
3. Using the Gas Fixture shown in Figure 5-1, attach the Nitric Oxide  
calibration-gas cylinder to the analyzers GAS inlet.  
4. Adjust the regulator of the calibration fixture for a flowmeter indica-  
tion of approximately 2 SCFH.  
5. Wait until the MEASURED reading on the screen stabilizes (approxi-  
mately 3 minutes). Then use the st and  
keys to enter an  
APPLIED value* that equals the concentration that is stamped on the  
NX calibration-gas cylinder.  
Typical Calibrate NX Screen During Calibration,  
Using 100 ppm Nitric Oxide Calibration Gas:  
CALIBRATE NX  
MEASURED  
APPLIED  
092 PPM  
0100 PPM  
6. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 50 – 150 ppm. Any attempt to  
calibrate outside range will cause the analyzer to display an error  
message.  
5-8  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.9 Calibrate CO  
(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)  
Material Required:  
Calibration Gas Fixture (Section 5.2)  
Gas Cylinder, 500 ppm CO in air (Refer to Section 8.2)  
Gas Cylinder, CO (1000 ppm) and H2 (1000 ppm) in Nitrogen  
(Refer to Section 8.2)  
Procedure:  
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to  
calibrate the CO sensor.  
2. At the conclusion of Step 1 the pump should start running.  
3. Using the Gas Fixture shown in Figure 5-1, attach the CO calibration-  
gas cylinder to the analyzers GAS inlet.  
4. Adjust the regulator of the calibration fixture for a flowmeter indica-  
tion of approximately 2 SCFH.  
5. Wait until the MEASURED reading on the screen stabilizes (approxi-  
mately 3 minutes). Then use the st and  
keys to enter an  
APPLIED value* that equals the concentration which is stamped on  
the CO calibration-gas cylinder.  
Typical CALIBRATE CO Screen During Calibration Procedure,  
Using 500 ppm CO Calibration Gas:  
CALIBRATE CO  
MEASURED  
APPLIED  
492 PPM  
0500 PPM  
6. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Test Gas CO/H2 Screen  
is displayed.  
* The calibration range for this screen is 250 – 1500 ppm. Any attempt  
to calibrate outside this range will cause the analyzer to display an  
error message.  
Instruction 24-9351  
5-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
7. Turn off the flow of CO calibration-gas; then remove the calibration-  
gas cylinder from the calibration fixture.  
8. Attach a CO/H2 calibration-gas cylinder to the calibration fixture;  
then adjust the regulator of the calibration fixture for a flowmeter  
reading of approximately 2 SCFH.  
9. Use the stand  
keys to enter a CO-VALUE that is the same as  
the CO concentration which is stamped on the CO/H calibration-gas  
2
cylinder.  
Typical Test Gas CO/H2 Screen During Calibration Procedure:  
TEST GAS CO/H2  
-------------------  
CO-VALUE  
XXXX  
ENTER CO-VALUE  
10. Press the  
Screen.  
key to save the CO-VALUE and display the Calibrate H2  
11. After calibration gas has been applied for approximately 3 minutes (to  
allow for stabilization), use the stand keys to enter an APPLIED  
value* that equals the H concentration stamped on the CO/H  
2
2
calibration-gas cylinder.  
Typical CALIBRATE H2 Screen During Calibration Procedure:  
CALIBRATE H2  
MEASURED  
APPLIED  
1050 PPM  
1000 PPM  
12. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
* The calibration range for this screen is 500–1500 ppm, any attempt to  
calibrate outside this range will cause the unit to display an error  
message.  
5-10  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Calibration  
5.10 Calibrate Draft  
(For PCA Models 15, 25, 35, 45, 55 & 65)  
Material Required:  
Calibration Fixture (Section 5-2)  
Bellows (adjustable)  
Micromanometer  
Range:  
±20 mb (±8 in. H2O column)  
Accuracy: ±0.025 mb (±0.01 in. H2O column)  
Procedure:  
IMP OR TANT! In Step 1, d o n ot connect the draft calibra-  
tion fixture to the analyzer until the Calibrate Draft Screen  
has been selected and displayed.  
1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to  
calibrate the draft sensor.  
2. With the Calibrate Draft Screen displayed, connect the hose from the  
calibration fixture to the analyzers DRAFT port; then adjust the  
bellows for a micromanometer reading of –10 mb (–4" H O column).  
2
3. Wait until the MEASURED reading on the screen stabilizes. Then use  
the stand  
keys to enter an APPLIED reading* which equals the  
Micromanometer reading.  
Typical CALIBRATE DRAFT Screen During Calibration Procedure:  
CALIBRATE DRAFT  
MEASURED  
APPLIED  
- 9.00 MB  
-10.00 MB  
4. Press the  
key to calibrate the analyzers MEASURED value to  
that of the APPLIED value. At this time the Calibrate Menu Screen is  
redisplayed.  
5. When all desired calibrations have been completed, press  
the Combustion Test Screen.  
to exit to  
* The calibration range for this screen is –5 to –15 mb, any attempt to  
calibrate outside this range will cause the unit to display an error  
message.  
Instruction 24-9351  
5-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
PCA  
NOTES:  
5-12  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.0 MAINTENANCE  
6.1 Routine Maintenance  
Routine maintenance of the analyzer consists of: replacing the batteries,  
cleaning the probe, draining the water trap, replacing the water trap  
filter, and performing periodic calibration checks to ensure that the  
analyzer is providing accurate readings.  
Replace the Batteries per Section 3.2.  
Clean the Probe per Section 6.3.  
Maintain the Water Trap/Filter Assembly per Section 6.4.  
Replace the Particulate Filter per Section 6.5  
Calibrate the analyzer per Section 5.0.  
Instruction 24-9351  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.2 Disassembly  
Perform the following when a maintenance procedure calls for removing  
the case, printed circuit board, pump, or sensors:  
1. Remove the batteries (Section 3.2.1)  
2. Place the analyzer face down on a work surface, then remove the  
units four rear-case screws.  
3. Carefully lift the rear case from the analyzer, unplug the battery  
compartment wires, then place the rear housing on a work surface  
(see Figures 6-1 & 6-2).  
B
W
A
T
T
E
E
R
Y
I
R
S
Figure 6-1. Disassembling the Analyzer  
6-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Maintenance  
TO  
BATTERY  
COMPARTMENT  
BIAS  
BATTERY  
FUSE  
CO PCB  
NX PCB  
OPTIONAL  
OPTIONAL  
NITRIC  
OXIDE  
CARBON  
MONOXIDE  
SENSOR  
SENSOR  
PUMP  
MTG.  
SCREWS  
OXYGEN  
SENSOR  
PUMP  
+
RED  
BLACK  
4
PARTICULATE  
FILTER  
DRAFT  
SENSOR  
Figure 6-2. Rear View of the PCB and Sensors  
Instruction 24-9351  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.3 Cleaning the Probe  
The Probe Tube and the Probe Body will become dirty under normal use  
(the water traps filter element should prevent soot from reaching the  
analyzers internal components). If the probe assembly is not kept clean,  
it could become clogged and restrict the flow of gas to the analyzer,  
resulting in incorrect readings and calculations.  
NOTE : An analyzer that is used to sample natural-gas  
furnaces normally requires less frequent cleaning than an  
analyzer used to sample oil or coal fired furnaces.  
Equipment Required:  
Alcohol  
Aerosol can of Automotive Carburetor Cleaner  
Clean Rag  
Source of Compressed Air (optional)  
Procedure:  
1. Remove the rubber tubing from the barbed fitting(s) on the probe  
handle (Figure 3-2).  
CAUTION:  
Carburetor cleaner attacks plastic components. Take precau-  
tions not to spray cleaner onto the probe handle or analyzer.  
2. Insert the plastic-spray tube of the carburetor cleaner into the barbed  
fitting(s) of the probe handle; then liberally spray carburetor cleaner  
through the probe.  
3. After spraying, remove all the residual cleaner by repeatedly flushing  
the probe with alcohol.  
4. Wipe off the surfaces of the probe and tubing with a clean rag.  
5. Allow the parts to dry completely. If available, blow compressed air  
through the probe to expedite the drying process.  
6. Reassemble the parts of the probe assembly.  
6-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Maintenance  
6.4 Water Trap/Filter Assembly Maintenance  
The Water Trap / Filter Assembly removes water condensate from the gas  
sample, and also prevents soot from contaminating the internal compo-  
nents of the analyzer.  
Drain the water condensate after every use.  
Procedure:  
1. Pull off the end-cap from the Inlet End of the Water Trap / Filter  
Assembly (Figure 6-3).  
2. Pour out all of the water condensate, and replace the end-cap.  
Replace the filter element when it becomes  
excessively dirty.  
Equipment Required:  
Filter Element (Refer to Section 8.0)  
Procedure:  
1. Pull off the end-cap from the Outlet End of the Water Trap / Filter  
Assembly (Figure 6-3).  
2. Remove and discard the old filter element.  
3. Install a new filter element and replace the end-cap.  
F
i
l
t
e
r
E
l
e
m
e
n
t
R
i
s
e
r
T
u
b
e
I
E
n
l
e
t
O
E
u
t
l
e
t
n
d
n
d
Figure 6-3. Water Trap/ Filter Assembly  
Instruction 24-9351  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.5 Replacing the Particulate Filter  
The internal particulate filter (Figure 6-2) prevents small dust and dirt  
particles from entering and damaging the pump. Depending on your  
environmental conditions, it is recommended to change the particulate  
filter and fitting approximately every six months, or sooner if it becomes  
blocked.  
Equipment Required:  
Particulate Filter and Fitting (Refer to Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Carefully pull off the tubing from the filter and fitting. Note the  
orientation of tubing and fitting before removing.  
3. Remove and discard the old filter and fitting and replace with new  
ones. Be careful not to not pinch the tubing during reassembly.  
4. Reassemble the analyzer in the reverse order of disassembly.  
6-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.6 Replacing the Oxygen Sensor  
Replace the Oxygen Sensor when it has expired (when the analyzers  
automatic O2 calibration fails and the unit displays the message O2-  
Sensor Error).  
NOTE : A "O2-SENSOR ERROR" displayed in the Sensor  
Status Screen does not necessarily mean that the sensor has  
expired. Before replacing the sensor, refer to Section 7.2 for  
other possible causes of the error.  
Equipment Required:  
Small Flat Blade Screw Driver  
Oxygen Sensor (Refer to Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Remove the Oxygen Sensor connector from the printed circuit board;  
then carefully pull off the two wires connected to the pins of the  
sensor (see Figure 6-2).  
3. Push in and turn the sensor counterclockwise (watching the notch on  
top) until it stops; then pull the sensor out of its housing.  
4. To install a new sensor, push the sensor into the housing and turn it  
clockwise until it locks in place.  
5. Connect the black (–) and red (+) wires, which were removed from the  
old sensor in Step 2, to the pins of the new sensor. Be sure to observe  
polarity as marked on the sensor.  
6. Reinstall the sensor connector onto the printed circuit board.  
7. Reassemble the analyzer. Then allow the sensor to be connected in  
the circuit for at least 1 hour before continuing.  
8. Place the analyzer in an area of fresh air and turn it ON. After the  
warmup cycle, observe that the Sensor Status Screen should not show  
an O2 sensor error.  
Instruction 24-9351  
6-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.7 Replacing the Nitric Oxide Sensor  
(For PCA Models 30, 35, 60 & 65)  
Replace the Nitric Oxide sensor when it has expired (can no longer be  
calibrated).  
NOTE : A "NX-SENSOR ERROR" displayed in the Sensor  
Status Screen does not necessarily mean that the sensor has  
expired. Before replacing the sensor, refer to Section 7.2 for  
other possible causes of the error.  
Equipment Required:  
Small Flat Blade Screw Driver  
Nitric Oxide Sensor (Refer to Section 8.0)  
Nitric Oxide Sensor Filter (Refer to Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Carefully pull the printed circuit board off the rear of the Nitric Oxide  
sensor (see Figure 6-2).  
3. Push in and turn the sensor counterclockwise (watching the notch on  
top) until it stops; then pull the sensor out of its housing.  
4. Install a new sensor by first pushing it into its housing, and then  
turning it clockwise until it locks in place.  
5. Install the circuit board, which was removed in Step 2, onto the rear  
of the sensor.  
6. Reassemble the analyzer and allow the sensor to be connected in the  
circuit for at least 4 hours before continuing.  
7. Place the analyzer in an area of fresh air and turn it ON.  
8. Calibrate the analyzer per Section 5.0.  
6.7.1 Replacing the Nitric Oxide Sensor Filter  
Replacing the orange filter on the Nitric Oxide sensor once a year could  
increase the life of the sensor.  
Procedure:  
1. Remove the Nitric Oxide sensor per Section 6.7.  
2. Pry the orange filter from the Nitric Oxide sensor and replace it with  
a new one.  
3. Reinstall the sensor.  
6-8  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.7.2 Replacing the Nitric Oxide Sensor  
Bias Battery  
A single lithium battery, located on the Nitric Oxide printed circuit board  
(see Figure 6-2), applies a constant bias voltage to the Nitric Oxide sensor  
even while the instrument is turned off. This battery has a life expectancy  
of at least 2 years.  
Replace the Nitric Oxide bias battery toward the end of its expected life.  
Equipment Required:  
Bias Battery (See Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Remove old battery from its holder (see Figure 6-2).  
3. Insert the new battery (positive side facing upwards) into its holder.  
4. Reassemble the analyzer.  
5. Before powering up and using the instrument, allow the Nitric Oxide  
sensor to stabilize as described below. Recalibration of the Nitric  
Oxide sensor is usually not required.  
Depending on how long the Nitric Oxide sensor was without bias  
voltage, the time required for the sensor to completely stabilize varies  
from less than a minute to several days. Typical stabilization times  
are shown below. Generally, however, the sensor is sufficiently stable  
after 4 hours for measurement purposes.  
Bia s r em oved for . . Sta biliza tion tim e  
Less than 15 min.  
Less than 1 hr.  
Less than 2 days  
Greater than 2 days  
Less than 1 min.  
Less than 5 min.  
Less than 4 hr.  
Up to 2 days  
Instruction 24-9351  
6-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Maintenance  
6.8 Replacing the Carbon Monoxide Sensor  
(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)  
Replace the Carbon Monoxide sensor when it has expired (can no longer  
be calibrated).  
NOTE : A "CO-SENSOR ERROR" displayed in the Sensor  
Status Screen does not necessarily mean that the sensor has  
expired. Before replacing the sensor, refer to Section 7.2 for  
other possible causes of the error.  
Equipment Required:  
Small Flat Blade Screw Driver  
Carbon Monoxide Sensor (Refer to Section 8.0  
Carbon Monoxide Sensor Filter (Refer to Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Carefully pull the printed circuit board off the rear of the Carbon  
Monoxide sensor (see Figure 6-2).  
3. Push in and turn the sensor counterclockwise (watching the notch on  
top) until it stops; then pull the sensor out of its housing.  
4. Remove the wire jumper from the pins of the new sensor.  
5. Install the new sensor by first pushing the sensor into its housing,  
and then turning clockwise until it locks in place.  
6. Install the circuit board, which was removed in Step 2, onto the rear  
of the sensor.  
7. Reassemble the analyzer; then allow the sensor to be connected in the  
circuit for at least 1 hour before continuing.  
8. Place the analyzer in an area of fresh air and turn it ON.  
9. Calibrate the analyzer per Section 5.0.  
6-10  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Maintenance  
6.8.1 Replacing the Carbon Monoxide Sensor Filter  
Replacing the red filter on the Carbon Monoxide sensor once a year could  
increase the life of the sensor.  
Procedure:  
1. Remove the Carbon Monoxide sensor per Section 6.8.  
2. Pry the red filter from the Carbon Monoxide sensor and replace it with a  
new one.  
3. Reinstall the sensor.  
Instruction 24-9351  
6-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
PCA  
6.9 Replacing the Pump Assembly  
Replace the Pump Assembly if it is found to be defective.  
Equipment Required:  
Small Flat Blade Screw Driver  
No. 1 Phillips Screw Driver  
Pump Assembly (Refer to Section 8.0)  
Procedure:  
1. Remove the analyzers rear case and lay it aside (refer to Section 6.2).  
2. Remove the two self tapping Phillips screws holding the Pump Assem-  
bly (See Figure 6-1).  
3. Unplug the pump connector from the printed circuit board; slide off  
the two hoses from the pump noting their orientation; then remove  
the entire assembly.  
4. Install the new assembly and reassemble the analyzer in the reverse  
order of disassembly.  
6-12  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
PCA  
7.0 TROUBLESHOOTING  
7.1 Analyzer Repair  
It is recommended that field repair of the PCA be limited to:  
Simple checks of the printed circuit boards  
Replacing the Probe Assembly  
Replacing the filter element in the Water Trap / Filter Assembly  
Replacing the Particulate Filter  
Replacing the Pump Assembly  
Replacing Sensors and Sensor Filters  
Replacing Batteries  
All other repairs should be performed by an authorized Bacharach Ser-  
vice Center. Any repairs performed by an unauthorized service organiza-  
tion will void the analyzers warranty and release Bacharach, Inc. of any  
implied or written product liability.  
Before returning your analyzer for repair, you may be able to determine  
and resolve a problem using the Troubleshooting Guide in Section 7.3.  
Instruction 24-9351  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
PCA  
7.2 Error Codes  
If one of the following messages appear at the bottom of the Sensor Status  
Screen, refer to Section 7.3 Troubleshooting Guide for information on how  
to correct the error.  
O2 - SENSOR ERROR  
CO - SENSOR ERROR  
NX - SENSOR ERROR  
O2 Sensor not connected, or is expired, or  
was exposed to combustion gases during  
warmup.  
Carbon Monoxide sensor is expired, or was  
exposed to Carbon Monoxide during  
warmup.  
Nitric Oxide sensor is expired, or was  
exposed to Nitric Oxide during warm up,  
or the bias battery is dead.  
TL-SENSOR ERROR  
TA-SENSOR ERROR  
Room air thermocouple is outside the  
range of –20 to 100 °C (–4 to 212 °F)  
Flue Gas thermocouple is not connected  
or is outside the range of –20 to 1200 °C  
(–4 to 2192 °F)  
DRAFT-SENSOR ERROR  
LOW BATTERY  
Outside the range of –7.5 to +7.5 mb  
(– 3 to + 3inches of water column)  
Battery voltage has dropped below 3.9 volts.  
The instrument will shut off when battery  
voltage drops below 3.5 volts.  
“– – – ”  
Not calculated (O2 above 18.8%,  
or the temperature is above 1200 °C  
(2192 °F), or sensor error  
“* * * *”  
“XXXX”  
Sensor not installed  
Overrange (numeric)  
7-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Troubleshooting  
7.3 Troubleshooting Guide  
The following table lists the causes and remedies for most of the problems  
that may arise with the analyzer. For help with any problem not dis-  
cussed here, contact your nearest Bacharach Service Center.  
TABLE 7-1. TR OUBLE SH OOTING GUIDE  
F a u lt  
P r ob a b le Ca u se & R em ed y  
Analyzer completely nonfunctional; a. Batteries dead. Replace batteries  
per Section 3.2.  
wont turn on when the  
pressed.  
key is  
b. Loose battery connector. Disas-  
semble analyzer and ensure that  
the battery connector is attached  
to the printed circuit board.  
c. Microprocessor needs to be reset.  
Press RESET button (Fig. 4-7).  
d. Optional Power Supply defective.  
Replace Power Supply.  
e. An a lyzer defect ive. Ret u r n t o  
Bacharach for repair.  
a. Microprocessor needs to be reset.  
Press RESET button (Fig. 4-7).  
Display Screen is blank when ana-  
lyzer is turned on, but pump runs  
during warmup cycle.  
b. An a lyzer defect ive. Ret u r n t o  
Bacharach for repair.  
Batteries close to being discharged.  
Analyzer will run for several min-  
utes before the instrument shuts off.  
Replace batteries per Section 3.2.1.  
LOW BATTE R Y message appears  
at bottom of display.  
a. Calibration was attempted while  
sampling combustion gases.  
O 2-S E N S O R E R R O R code a p-  
pears in the Sensor Status Screen.  
b. O2 sensor is either expired, not  
wired correctly, or not connected to  
the circuit board. Replace or check  
wiring of sensor per Section 6.6.  
Instruction 24-9351  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
PCA  
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)  
F a u lt P r ob a b le Ca u se & R em ed y  
NX-SE NSOR E R R OR code a p-  
pears in the Sensor Status Screen.  
a. Calibration was attempted while  
sampling combustion gases.  
b. Nitric Oxide sensor is expired.  
Replace sensor per Section 6.7.  
c. Bias battery on the Nitric Oxide  
circuit board is dead. Replace bat-  
tery per Section 6.7.2.  
CO-SE NSOR E R R OR code a p-  
pears in the Sensor Status Screen.  
a. Calibration was attempted while  
sampling combustion gases.  
b. Carbon Monoxide sensor is ex-  
pired. Replace sensor per Sec-  
tion 6.8.  
DRAFT-SENSOR ERROR code ap-  
pears in the Sensor Status Screen.  
a. Sensor is exposed to pressure out-  
side of its detectable range.  
b. Sensor defective. Return analyzer  
to Bacharach for repair.  
TA-SE NSOR OR TL-SE NSOR  
ERROR code appears in the Sen-  
sor Status Screen.  
a. Calibration was attempted while  
sampling combustion gases.  
b. Th er m ocou ple defect ive. Re-  
place probe assembly.  
c. Thermocouple not connected.  
Connect thermocouple to ana-  
lyzer per Section 3.3.  
d. Instrument exposed to temp-  
erature outside its allowable  
operating temperature.  
****appears in one or more value  
fields.  
The fields associated sensor is not  
installed.  
7-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Troubleshooting  
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)  
F a u lt P r ob a b le Ca u se & R em ed y  
- - - -appears in one or more value a. The analyzer is not able to calcu-  
fields of the Combustion Test Screen.  
late a numerical value based on  
measured combustion data. The  
“- - - -” is replaced with numerical  
values when the analyzer begins  
to detect valid combustion data.  
b. Sensor in error during warm-up.  
XXXXappears in one or more value The fields associated sensor is de-  
fields.  
tecting a value that is outside the  
analyzers detection range. XXXX”  
is r epla ced wit h n u m er ica l da t a  
when the analyzer detects values  
that fall within its range.  
An a lyzer won t r espon d wh en a Microprocessor needs to be reset.  
panel key is pressed. Press RESET button (Fig. 4-7).  
Pump motor sounds sluggish, stalls, a. Flow restricted. Check that the  
or wont start.  
filter element in the Water Trap/  
Filter Assembly is clean and not  
saturated with water. Also, verify  
that the probe hose tubing is not  
pinched (Fig. 6-3).  
b. Flow restricted. Check the Par-  
ticulate Filter is clean and not  
blocked (Fig. 6-2).  
c. Loose pump connection. Disas-  
semble analyzer and ensure that  
the pump connector is securely  
attached to the circuit board  
d. Pump defective. Replace pump  
assembly.  
Instruction 24-9351  
7-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
PCA  
TABLE 7-1. TR OUBLE SH OOTING GUIDE (Con t .)  
F a u lt P r ob a b le Ca u se & R em ed y  
Backlight wont turn on.  
Backlight LED burned out. Return  
to Bacharach for repair.  
Batteries do not last 10 hours.  
Cold temperature is reducing bat-  
tery capacity. To obtain longer oper-  
ating time, keep analyzer warm.  
Erratic Combustion Test Screen a. Faulty sensor(s):  
values. - Check that the sensors are prop-  
erly installed per Sections 6.6  
through 6.8.  
- Ch eck sen sor ca libr a t ion per  
Section 5.0.  
- Repla ce sen sor (s) a n d r eca l-  
ibrate per Sections 5.0 and 6.0.  
b. Probe assembly leaking. Check  
tightness of all hose connections  
and integrity of tubing.  
c. Pump defective. Replace pump &  
motor assembly.  
d. An a lyzer defect ive. Ret u r n t o  
Bacharach for repair.  
Analyzer will not calibrate properly. a. Wrong calibration gas or insuffi-  
cient flow being applied to sensor.  
Ensure your calibration setup is  
correct.  
b. Faulty sensor. Replace sensor and  
recalibrate per Sections 5.0 and  
6.0.  
c. An a lyzer defect ive. Ret u r n t o  
Bacharach for repair.  
7-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Parts & Service  
8.0 PARTS & SERVICE  
8.1 Replacement Parts  
It em  
Descr ip t ion  
P a r t No.  
F igu r es  
8-1, 2 & 3  
1
2
3
Battery Cover  
Screw, Case Housing  
Screw, Pump Mounting  
24-0784  
501-3824  
501-3822  
8
9
Oxygen Sensor  
24-0788  
24-0789  
24-0881  
24-0863  
24-0862  
24-3009  
604-2605  
204-0020  
Carbon Monoxide Sensor  
Nitric Oxide Sensor  
Carbon Monoxide Sensor Filter  
Nitric Oxide Sensor Filter  
Pump Assembly  
10  
11  
12  
13  
14  
15  
Fuse  
Battery, Nitric Oxide Sensor Bias  
18  
19  
20  
21  
22  
23  
24  
25  
Filter Element (white)  
Draft Connector, Probe  
Gas Connector, Probe  
O-Ring, 7mm OD x 1mm wall  
O-Ring, 8mm OD x 1mm wall  
Particulate Filter  
Fitting, for Particulate Filter  
Filter Assembly (complete)  
07-1644  
24-0878  
24-0877  
105-5103  
105-5102  
07-1600  
103-5267  
24-1107  
Instruction 24-9351  
8-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Parts & Service  
PCA  
8.2 Accessories  
Descr ip t ion  
P a r t No.  
STANDAR D ACCE SSOR IE S:  
Battery, AA” Alkaline  
Complete Probe and Hose Assembly (Gas & Draft)  
Instruction Manual  
204-0004  
24-3004  
24-9351  
24-1078  
Plastic Carrying Case  
OP TIONAL ACCE SSOR IE S:  
Ambient Thermocouple, 10 ft. K-type  
Ambient Thermocouple, 1 in. K-type  
Bent Probe Tip  
104-1797  
104-1798  
24-8039  
24-7059  
24-1103  
24-0794  
24-0492  
Calibration Kit  
Differential Pressure Hose Assembly  
Gas Cylinder, 1000 ppm CO, 1000 ppm H2, in Nitrogen  
Gas Cylinder, 500 ppm CO  
Printer, IrDA (with battery charger):  
120 VAC  
24-1229  
24-1230  
06-8733  
24-0885  
24-1209  
24-1073  
230 VAC  
Printer Paper, Thermal (1 roll)  
Power Supply Adapter, 110 VAC  
Regulated Power Supply Adapter, 230 VAC  
Serial Cable  
8-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Parts & Service  
2
1
3
Figure 8-1. Replacement Parts  
Instruction 24-9351  
8-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Parts & Service  
PCA  
FUSE  
14  
15  
10  
12  
9
11  
8
+
13  
RED  
BLACK  
4
23  
24  
22  
21  
Figure 8-2. Replacement Parts  
8-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Parts & Service  
1
9
2
0
1
8
2
5
Figure 8-3. Replacement Parts  
Instruction 24-9351  
8-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Parts & Service  
PCA  
8.3 Sales/Service Centers  
Replacement parts and service can be obtained by contacting one of  
the following Bacharach Sales/Service Centers:  
UnitedStates  
Canada  
BacharachofCanada,Inc.  
250ShieldsCourtUnit#3  
Markham, OntarioL3R9W7Canada  
Phone:905-470-8985  
Fax:905-470-8963  
California  
7281GardenGroveBlvd.,  
SuiteH  
GardenGrove, CA92841  
Phone:714-895-0050  
Fax:714-895-7950  
México  
BacharachdeMéxico  
PlayaRegatasNo.473TercerPiso  
Col.MilitarMarte  
DelegaciónIztacalco,08830  
MéxicoD.F.México  
Phones: +52-555-634-7740  
+52-555-634-7741  
Indiana  
8618LouisianaPlace  
Merrillville,IN46410  
Phone:219-736-6178  
Fax:219-736-6269  
FAX: +52-555-634-7738  
NewJersey  
7300IndustrialPark  
Rte.130,Bldg.22  
Europe  
Pennsauken,NJ08110  
Phone:856-665-6176  
Fax:856-665-6661  
EuropeanHeadquarters  
BacharachInstruments  
SovereignHouse,Queensway  
RoyalLeamingtonSpa  
WarwickshireCV313JR  
UnitedKingdom  
Phone:+44-1926-338111  
Fax:+44-1926-338110  
Pennsylvania  
621HuntValleyCircle  
NewKensington, PA15068  
Phone:724-334-5051  
Fax:724-334-5723  
Sales/ServiceCenter-Denmark  
BacharachInstrumentsInt’l  
P.O.Box44  
Texas  
5151Mitchelldale,B-4  
Houston, TX77092  
Phone:713-683-8141  
Fax:713-683-9437  
39Lindegade  
DK6070ChristiansfeldDenmark  
Phone:+45-74-563171  
Fax:+45-74-563178  
8-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix A  
APPENDIX A  
Display Screen Translations  
This appendix shows the contents of the various screen displays that are  
seen while operating the PCA in all ten languages that the analyzer is  
capable of displaying. The language that is displayed on the analyzer is  
dependent on the analyzers model (refer to Table 1-1), and the language  
selected per Section 4.15.  
The languages are arranged in columns side-by-side for comparison and  
translation purposes.  
Instruction 24-9351  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
PCA  
English  
(ENG)  
German  
(DEU)  
Danish  
(DAN)  
French  
(FRA)  
Spanish  
(ESP)  
Warm Up Screen  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
PCA nn  
Warmup nn  
KALIBRIEREN nn  
OPVARMNING nn  
CYCLE D AUTOCALIB nn CALENTAMIENTO nn  
Sensor Status Screen (Errors)  
NO ERRORS DETECTED  
LOW BATTERY  
KALIBRIERUNG OK  
Batterie leer  
INGEN FEJL  
PAS D ERREUR DETECTE NO HAY ERRORES  
LAV BATTERIKAPACITET  
O2-SENSOR FEJL  
CO-SENSOR FEJL  
TRAEK-SENSOR FEJL  
ROEGGASTEMP FEJL  
LUFTTEMP FEJL  
BATTERIE FAIBLE  
BATTERIA BAJA  
O2-SENSOR ERROR  
CO-SENSOR ERROR  
DRAFT-SENSOR ERROR  
O2-Sensor  
02-ERREUR DE CAPTEUR ERROR SENSOR-O2  
ERREUR DE CAPTEUR-CO ERROR SENSOR-CO  
CO-Sensor  
Feinzug-Sensor  
ERR D CPTR TIRAGE  
ERROR SENSOR-TIRO  
TA-STACK-SENSOR ERROR Gas-F¸hler  
ERREUR D CPTR T-CHMN ERR SENSOR T-CHIMENE  
TA-AIR-SENSOR ERROR  
NX-SENSOR ERROR  
Luft-F¸hler  
NX-Sensor  
ERREUR D CPTR T-AIR  
ERREUR DE CAPTEUR  
ERROR SENSOR T-AIRE  
ERROR SENSOR-NX  
NX-SENSOR FEJL  
Sign Off Screen  
OFF IN  
n
SEC  
Aus in  
n
sec  
STOP OM n SEK  
TERMNR DANS  
n
SEC  
APAG EN n SEGS  
RUN/HLD Screen (Abbreviations)  
NG  
EG  
KG  
SG  
FG  
NG  
KOK  
BG  
GN  
GN  
KOK  
LEG  
LPG  
O#2  
O#6  
PC  
GM  
GC  
GV  
GCD  
LPG  
P#2  
P#6  
CAR  
FG  
LPG  
H#2  
H#6  
CHR  
Hˆl  
Sˆl  
KOL  
O#2  
O#6  
BF  
Fuel Selection Screen  
FUEL  
BRENNSTOFF  
Erdgas  
FUEL  
COMBSTBL  
GAZ NATU  
GAZ MANU  
GAZ VIL  
COMBUSTB  
GNATURAL  
GCARBON  
GCIUDAD  
LPG  
NATGAS  
KOKS  
N-GAS  
Kok.-Gas  
Stadtgas  
Fl.-Gas  
KOKS  
LEG  
B-GAS  
LPG  
F-GAS  
PROPANE  
FUEL DOM  
FUEL LOU  
CHARBON  
OIL NO.2  
OIL NO.6  
P-COAL  
Heizoel  
LET OLIE  
TUN OLIE  
BIOFUEL  
PETRO #2  
PETRO #6  
CARBON  
Schweroel  
Kohle  
Draft Screens  
DRAFT  
FEINZUG  
Saugschlauch  
trennen  
TRAEK  
TIRAGE  
TIRO  
DISCONNECT DRAFT  
HOSE  
AFMONTER  
TRAEKSLANGE  
TRYK  
DEBRANCHER LE TUYAU DESCONECTE MANGUERA  
DE TIRAGE  
DE ASPIRACION  
OPRIMA  
PRESS  
dr¸cken  
APPUYER SUR  
RECONNECT DRAFT  
HOSE  
Schlauch  
anschlieen  
Feinzug  
MONTER  
REBRANCHER LE TUYAU RECONECTE MANGUERA  
TRAEKSLANGE  
TRAEK  
DE TIRAGE  
TIRAGE  
DE ASPIRACION  
TIRO  
DRAFT  
HOT SPOT  
Gas-Temp  
KERNESTR  
PNT CHAUD  
PTO CLNTE  
A-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix A  
Polish  
(POL)  
Italian  
(ITA)  
Dutch  
(NED)  
Swedish  
(SVE)  
Finnish  
(FIN)  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
BACHARACH, INC.  
PCA nn  
NAGRZEWANIE nn  
RISCALDAMENTO nn  
INITIALISATIE nn  
UPPVARMNING nn  
MITTARIL‰MPI‰‰  
NIE WYKRYTO BLEDOW  
SLABY AKUMULATOR  
BLAD CZUJNIKA-O2  
NO RIVELAZIONE ERROR  
BATTERIA BASSA  
GEEN FOUTMELDINGEN  
VOEDINGSSPANNING  
ERR: O2-SENSOR  
Inget fel  
EI HAVAITTU VIRHEIT‰  
PARISTO LOPPUU  
Lag batterikapacitet  
O2-SENSOR FEL  
CO-SENSOR FEL  
Drag-SENSOR FEL  
ERRORE SENSORE-O2  
ERRORE SENSORE-CO  
VIKA O2-ANTURISSA  
VIKA CO-ANTURISSA  
VIKA PAINEANTURISSA  
SAVUK LT-ANTURIVIKA  
ILMAN LT-ANTURIVIKA  
VIKA NX-ANTURISSA  
BLAD CZUJNIKA-CO  
ERR: CO-SENSOR  
BLAD CZUJNIKA-CIAGU  
BLAD CZUJNIKA KOMINT  
BLAD CZUJNIKA POW-T  
BLAD CZUJNIKA-NX  
ERRORE SENSORE-TIRAG ERR: TREK-SENSOR  
ERRORE SENS-T-CAMINO ERR:T-ROOKGAS SENSOR Rokgastemp FEL  
ERRORE SENSOR-T-ARIA  
ERRORE SENSORE-NX  
ERR: T-OMG. SENSOR  
ERR: NX-SENSOR  
Lufttemp FEL  
NOX-SENSOR FEL  
WYTACZ W n SEK  
OFF IN n SEC  
UIT NA n SEC  
Stangs av om n sek  
KIINNI n SEK:SSA  
GZ  
MET  
COK  
CIT  
NG  
NG  
MKA  
KOK  
KKA  
PRO  
KˆL  
RˆL  
BP  
GWL  
GNG  
LPG  
O#2  
O#6  
PC  
KOK  
LEG  
LPG  
O#2  
O#6  
PC  
KOK  
SG  
GPL  
O#2  
O#6  
CAR  
PG  
O#2  
O#6  
BF  
PALIWO  
COMBUSTI  
METANO  
GAS COKE  
GASCITTA  
GPL  
BRANDST.  
AARDGAS  
KOKS  
BRANSLE  
N-GAS  
POLTTOAI  
MAAKAASU  
HIILIKAA  
GAZ ZIEN  
GAZWEGL  
GAZNENG  
PROP PLN  
OLEJ LEK  
OLEJ CIE  
P-COAL  
KOKS  
LEG  
Stadsgas  
P-GAS  
KAUPKAAS  
PROPAANI  
KEVYTˆLJ  
RASKˆLJY  
BIOPOAINE  
LPG  
GASOLIO  
NAFTA  
OIL NO.2  
OIL NO.6  
P-COAL  
Tunn olja  
Tjock olja  
Biofuel  
CARBON  
CIAG  
TIRAGGIO  
TREK-METING  
ONTKOPPEL LEIDING  
DRAFT HOSE  
DRAG  
VETO  
ODLACZ WAZ DO  
POMIARU CIAGU  
NACISNIJ  
DISINSERIRE  
ASPIRATORE  
PREMERE  
Tabort drag slang  
IRROITA PAINELETKUN  
LIITIN  
DRUK  
Tryck  
PAINA  
PODLACZ PONOWNIE WAZ  
DO POMIARU CIAGU  
CIAG  
REINSERIRE  
ASPIRATORE  
TIRAGGIO  
VERBIND OPNIEUW DE  
LEIDING (DRAFT HOSE)  
TREK  
Anlsut drag slang  
LIITPAINELETKUN  
LIITIN  
Drag  
PAINE  
OGNISKO  
PTO CALDO  
HOT SPOT  
Rokgas  
KUUMA KOH  
Instruction 24-9351  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
PCA  
English  
(ENG)  
German  
(DEU)  
Danish  
(DAN)  
French  
(FRA)  
Spanish  
(ESP)  
Saving Memory Screen  
SAVING MEMORY  
LOCATION nn  
Speichere Daten in  
Block nn  
LAGRING MEMORY  
BLOK nn  
SAUVEGARDE D MEMOIRE GUARDANDO UBIBACION  
ZONE nn  
DE MEMORIA nn  
Memory Directory Screen  
MEMORY DIRECTORY  
MEMORY EMPTY  
CLEAR MEMORY  
SPEICHER LISTE  
Speicher leer  
MEMORY BIBLIOTEK  
MEMORY TOM  
REPERTOIRE MEMRE  
MEMOIRE VIDE  
DIRECTORIO MEMORI  
MEMORIA VACIA  
Speicher Lˆschen  
SLET MEMORY  
EFFACER MEMOIRE  
BORRAR MEMORIA  
Draft Memory Screen  
DRAFT MEM  
DRAFT  
FEINZUG MEM  
Feinzug  
TRAEK MEMORY  
TRAEK  
TIRAGE MEM  
TIRAGE  
TIRO MEM  
TIRO  
Clear Memory Screen  
CLEAR MEMORY  
SPEICHER LˆSCHEN  
SLET MEMORY  
EFFACER MEMOIRE  
BORRAR MEMORIA  
Memory to PC Screen  
MEMORY TO PC  
TRANSMIT DATA  
CLEAR MEMORY  
TRANSMITTING...  
ABSPEICHERN AUF PC  
PC HUKOMMELSE  
OVERFOER DATA  
SLET MEMORY  
MEMOROIRE A PC  
TRANSM. DONNEES  
EFFACER MEMOIRE  
EN TRANSMISSION...  
MEMORIA AL PC  
Daten ¸bertragen  
Speicher lˆschen  
¸bertragung...  
TRANSMITIR DATOS  
BORRAR MEMORIA  
TRANSMITIENDO...  
OVERFOERELSE...  
ID Setup Screen  
SETUP  
ID #n  
KONFIGURIEREN  
ID #n  
OPSAETNING  
ID #n  
PROGRAMMER  
ID #n  
ESTABLECER  
ID #n  
Temperature Setup Screen  
SETUP  
KONFIGURIEREN  
Temp.-Einheit  
OPSAETNING  
PROGRAMMER  
UNITE DE  
ESTABLECER  
UNIDAD DE  
TEMPERATURE UNIT  
TEMPERATURENHED  
TEMPERATURE  
TEMPERATURA  
Draft Setup Screen  
SETUP  
KONFIGURIEREN  
OPSAETNING  
TRAEKENHED  
mB/PA/WC  
PROGRAMMER  
UNITE DE TIRAGE  
mB/PA/WC  
ESTABLECER  
UNIDAD DE TIRO  
mB/PA/WC  
DRAFT UNIT  
mB/PA/WC  
Feinzug-Einheit  
mB/PA/WC  
Language Setup Screen  
SETUP  
KONFIGURIEREN  
Sprache w‰hlen  
OPSAETNING  
SPROGKODE  
PROGRAMMER  
LANGUE  
ESTABLECER  
IDIOMA  
LANGUAGE  
Display Mode Setup Screen  
SETUP  
KONFIGURIEREN  
Anzeigen  
OPSAETNING  
DISPLAY  
PROGRAMMER  
AFFICHER  
ESTABLECER  
VISUALIZAR  
DISPLAY  
A-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix A  
Polish  
(POL)  
Italian  
(ITA)  
Dutch  
(NED)  
Swedish  
(SVE)  
Finnish  
(FIN)  
ZACHOWYWANIE MIEJSCE  
W PAMIECI nn  
SALVARE MEMORIA  
UBICAZIONE nn  
OPSLAAN IN GEHEUGEN  
LOCATIE nn  
Sparar minne  
i block nn  
TALLENTAA MUISTI  
PAIKKAAN nn  
KATALOG PAMIECI  
PAMIEC PUSTA  
INDIRIZZ MEMORIA  
MEMORIA VUOTA  
INHOUD GEHEUGEN  
GEHEUGEN LEEG  
GEHEUGEN WISSEN  
MINNES BIBLIOTEK  
Minnet tomt  
MUISTITIEDOSTO  
MUISTI TYHJfl  
WYCZYSC PAMIEC  
ANNULARE MEMORIA  
Rensa minnet  
TYHJENNMUISTI  
CIAG MEM  
CIAG  
TIRAGGIO MEM  
TIRAGGIO  
TREK-METING MEM  
TREK  
DRAG MINNET  
Drag  
VETO MEM  
PAINE  
WYCZYSC PAMIEC  
ANNULARE MEMORIA  
GEHEUGEN WISSEN  
Rensa minnet  
TYHJENNMUISTI  
PAMIEC DO PC  
DA MEMORIA A PC  
TRASMETTERE DATI  
ANNULARE MEMORIA  
TRANSMISSIONE...  
GEHEUGEN NAAR PC  
VERZEND GEGEVENS  
GEHEUGEN WISSEN  
VERZENDING...  
MINNE TILL PC  
Overfor data  
Rensa minnet  
Overforing...  
MUISTI PC:LLE  
SIIRRƒ TIEDOT  
TYHJENNƒ MUISTI  
SIIRT‰‰ .  
PRZESLAC DANE  
WYCZYSC PAMIEC  
TRANSMISJA...  
USTAW  
ID #n  
PREPARAZIONE  
ID #n  
INSTELLINGEN  
ID #n  
KONFIG  
ID #n  
ASETUKSET  
TUNNISTE #n  
USTAW  
PREPARAZIONE  
INSTELLINGEN  
EENHEID VAN  
TEMPERATUUR  
KONFIG  
ASETUKSET  
L‰MPˆTILAN  
YKSIKKˆ  
JEDNOSTKE  
TEMPERATURY  
UNITA DI TEMPER  
Temperaturenhet  
USTAW  
PREPARAZIONE  
UNITA DI TIRAGGIO  
mB/PA/WC  
INSTELLINGEN  
EENHEID VAN TREK  
mB/PA/WC  
KONFIG  
ASETUKSET  
JEDNOSTKE CIAGU  
mB/PA/WC  
Drag enhet  
mB/PA/WC  
PAINEEN YKSIKKˆ  
mB/PA/WC  
USTAW  
JEZYK  
PREPARAZIONE  
LINGUA  
INSTELLINGEN  
TAAL  
KONFIG  
Sprak  
ASETUKSET  
KIELI  
USTAW  
PREPARAZIONE  
VISUALIZZARE  
INSTELLINGEN  
SCHERM  
KONFIG  
DISPLAY  
ASETUKSET  
N‰YTTˆ  
WYSWIETLIC  
Instruction 24-9351  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
PCA  
English  
(ENG)  
German  
(DEU)  
Danish  
(DAN)  
French  
(FRA)  
Spanish  
(ESP)  
Time/Date Setup Screen  
SETUP  
TIME  
KONFIGURIEREN  
OPSAETNING  
TID  
PROGRAMMER  
HEURE  
ESTABLECER  
HORA  
Zeit  
DATE  
Datum  
DATO  
DATE  
FECHA  
Maintenance Password Screen  
MAINTENANCE  
PASSWORD  
INSTANDHALTUNG  
Kennwort  
VEDLIGEHOLDELSE  
KODEORD  
MAINTENANCE  
MOT DE PASSE  
MANTENIMIENTO  
CONTRASENA  
Printer Setup Screen  
SETUP PRINTER  
KONFIGURIEREN PRINTER OPSAETNING PRINTER  
PROGRAMMER PRINTER ESTABLECER PRINTER  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
Maintenance Menu Screen  
MAINTENANCE  
CALIBRATION  
USER NAME  
INDSTANDHALTUNG  
ABGLEICH  
VEDLIGEHOLDELSE  
CALIBRATION  
MAINTENANCE  
CALIBRAGE  
MANTENIMIENTO  
CALIBRACION  
BENUTZERNAME  
BRUGER NAVN  
USER NAME  
NOMBRE DEL USUARIO  
Calibrate Menu Screen  
CALIBRATE  
TA-ZERO  
TA-SPAN  
TL-ZERO  
TL-SPAN  
NX  
ABGLEICHEN  
TA-Offst  
TA-Gain  
TL-Offst  
TL-Gain  
NX  
KALIBRER  
TA-NUL  
TA-SPAN  
TL-NUL  
TL-SPAN  
NX  
CALIBRER  
TA-ZERO  
TA-FRCH  
TL-ZERO  
TL-FRCH  
NX  
CALIBRAR  
TA-CERO  
TA-TRAM  
TL-CERO  
TL-TRAM  
NX  
CO  
CO  
CO  
CO  
CO  
DRAFT  
ZUG  
TRAEK  
TIRAGE  
TIRO  
Calibrate TA-Zero Screen  
CALIBRATE  
TA-ZERO  
ABGLEICHEN  
TA-OFFSET  
Ist-Wert  
KALIBRER  
TA-NUL  
CALIBRER  
TA-ZERO  
MESURE  
APPLIQUE  
CALIBRAR  
TA-CERO  
MEDIDA  
MEASURED  
APPLIED  
MAALT  
Soll-Wert  
SAND VAERDI  
APLICADA  
Calibrate TA-Span Screen  
CALIBRATE  
TA-SPAN  
ABGLEICHEN  
TA-GAIN  
KALIBRER  
TA-SPAN  
CALIBRER  
FRCHETTE-TA  
MESURE  
CALIBRAR  
TA-TRAM  
MEDIDA  
MEASURED  
APPLIED  
Ist-Wert  
MAALT  
Soll-Wert  
SAND VAERDI  
APPLIQUE  
APLICADA  
Calibrate TL-Zero Screen  
CALIBRATE  
TL-ZERO  
ABGLEICHEN  
TL-OFFSET  
Ist-Wert  
KALIBRER  
TL-NUL  
CALIBRER  
TL-ZERO  
MESURE  
APPLIQUE  
CALIBRAR  
TL-CERO  
MEDIDA  
MEASURED  
APPLIED  
MAALT  
Soll-Wert  
SAND VAERDI  
APLICADA  
A-6  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix A  
Polish  
(POL)  
Italian  
(ITA)  
Dutch  
(NED)  
Swedish  
(SVE)  
Finnish  
(FIN)  
USTAW  
CZAS  
DATE  
PREPARAZIONE  
ORA  
INSTELLINGEN  
TIJD  
KONFIG  
Tid  
ASETUKSET  
AIKA  
DATA  
DATUM  
Datum  
PƒIVƒ  
UTRZYMANIE  
HASTO  
MANTENIMENTO  
ONDERHOUD  
PASWOORD  
UNDERHALL  
PASSWORD  
YLLƒPITO  
PAROLA DORDINE  
SALASANA  
USTAW PRINTER  
PREPARAZIONE PRINTER  
IR-HP/IR-IRDA/RS232  
INSTELLINGEN PRINTER  
IR-HP/IR-IRDA/RS232  
KONFIG PRINTER  
ASETUKSET PRINTER  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
IR-HP/IR-IRDA/RS232  
UTRZYMANIE  
MANTENIMENTO  
CALIBRAZIONE  
ONDERHOUD  
UNDERHALL  
Kalibrering  
YLLƒPITO  
WZORCOWANIE  
UZYWACZA IMIE  
IJKINGSPROCEDURE  
GEBRUIKERS ID  
KALIBROINTI  
KƒYTTƒJƒNIMI  
NOME DELLíUTENTE  
Anvandar namn  
WZORCUJ  
TA-ZERO  
TA-ROZP  
TL-ZERO  
TL-ROZP  
NX  
CALIBRARE  
TA-ZERO  
TA-DIFF  
TL-ZERO  
TL-DIFF  
NX  
CALIBREER  
T-RKG-0  
T-RKG-1  
T-OMG-0  
T-OMG-1  
NX  
CALIBRATE  
TA-ZERO  
TA-SPAN  
TL-ZERO  
TL-SPAN  
NX  
KALIBROI  
TS-NOLL  
TS-YLƒR  
TI-NOLL  
TI-YLƒR  
NX  
CO  
CO  
CO  
CO  
CO  
CIAG  
TIRAGG  
TREK  
DRAFT  
VETO  
WZORCUJ  
TA-ZERO  
CALIBRARE  
TA-ZERO  
IJKING  
CALIBRATE  
TA-ZERO  
KALIBROI  
TS-NOLLA  
MITATTU  
T-RKG-0  
MIERZONA  
STOSOWANA  
MISURATO  
APPLICATO  
GEMETEN  
AANGELEGD  
MEASURED  
APPLIED  
SY÷TETTY  
WZORCU  
CALIBRARE  
TA-DIFF  
IJKING  
CALIBRATE  
TA-SPAN  
KALIBROI  
TS-YLRAJA  
MITATTU  
TA-ROZPIETOS  
MIERZONA  
T-RKG-1  
MISURATO  
APPLICATO  
GEMETEN  
AANGELEGD  
MEASURED  
APPLIED  
STOSOWANA  
SY÷TETTY  
WZORCUJ  
TL-ZERO  
CALIBRARE  
TL-ZERO  
IJKING  
CALIBRATE  
TL-ZERO  
KALIBROI  
TI-NOLLA  
MITATTU  
T-OMG-0  
GEMETEN  
AANGELEGD  
MIERZONA  
STOSOWANA  
MISURATO  
APPLICATO  
MEASURED  
APPLIED  
SY÷TETTY  
Instruction 24-9351  
A-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
PCA  
English  
(ENG)  
German  
(DEU)  
Danish  
(DAN)  
French  
(FRA)  
Spanish  
(ESP)  
Calibrate TL-Span Screen  
CALIBRATE  
TL-SPAN  
ABGLEICHEN  
TL-GAIN  
KALIBRER  
TL-SPAN  
CALIBRER  
FRCHETTE-TL  
MESURE  
CALIBRAR  
TL-TRAM  
MEDIDA  
MEASURED  
APPLIED  
Ist-Wert  
MAALT  
Soll-Wert  
SAND VAERDI  
APPLIQUE  
APLICADA  
Calibrate NX Screen  
CALIBRATE NX  
MEASURED  
APPLIED  
ABGLEICHEN NX-GAIN  
KALIBRER NX  
MAALT  
CALIBRER NX  
MESURE  
CALIBRAR NX  
MEDIDO  
Ist-Wert  
Soll-Wert  
SAND VAERDI  
APPLIQUE  
APLICADO  
Calibrate CO Screen  
CALIBRATE CO  
MEASURED  
APPLIED  
ABGLEICHEN CO-GAIN  
KALIBRER CO  
MAALT  
CALIBRER CO  
MESURE  
CALIBRAR CO  
MEDIDO  
Ist-Wert  
Soll-Wert  
SAND VAERDI  
APPLIQUE  
APLICADO  
Calibrate CO/H2 Screen  
TESTGAS CO/H2  
CO VALUE  
Pr¸fgas CO/H2  
CO-Anteil  
PROEVEGAS CO/H2  
CO VAERDI  
TESTERGAZ CO/H2  
VALEUR CO  
GAS DE PRUEBA CO/H2  
VALOR DE CO  
ENTER CO VALUE  
CO-Anteil eingeben  
CO VAERDI ENTER  
ENTRER VALEUR CO  
ENTRAR VALOR DE CO  
Calibrate H2 Screen  
CALIBRATE H2  
MEASURED  
APPLIED  
ABGLEICHEN H2-GAIN  
KALIBRER H2  
MAALT  
CALIBRER H2  
MESURE  
CALIBRAR H2  
MEDIDO  
Ist-Wert  
Soll-Wert  
SAND VAERDI  
APPLIQUE  
APLICADO  
Calibrate Draft Screen  
CALIBRATE DRAFT  
MEASURED  
ABGLEICHEN FEINZUG  
KALIBRER TRAEK  
MAALT  
CALIBRER TIRAGE  
MESURE  
CALIBRAR TIRO  
MEDIDO  
Ist-Wert  
APPLIED  
Soll-Wert  
SAND VAERD  
APPLIQUE  
APLICADO  
User Name Screen  
USER NAME  
LINE n  
BENUTZERNAME  
Zeile n  
BRUGER NAVN  
LINIE n  
USER NAME  
LIGNE n  
NOMBRE DEL USUARIO  
LINEA n  
Miscellaneous  
BAD CALIBRATION ENTRY Kalibrierung nich mˆglich  
KALIBRERING IKEE MULIGT ERREUR D'ETALONNAGE  
SKYLLER CO SENSOR PURGE CELLULE CO  
CALIBRACION-INCORRECTA  
PURGANDO SENSOR CO  
PURGING CO SENSOR  
CO Sp¸lfunktion  
A-8  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix A  
Polish  
(POL)  
Italian  
(ITA)  
Dutch  
(NED)  
Swedish  
(SVE)  
Finnish  
(FIN)  
WZORCUJ  
CALIBRARE  
TL-DIFF  
IJKING  
CALIBRATE  
TL-SPAN  
KALIBROI  
TI-YLRAJA  
MITATTU  
TL-ROZPIETOS  
MIERZONA  
T-OMG-1  
GEMETEN  
AANGELEGD  
MISURATO  
APPLICATO  
MEASURED  
APPLIED  
STOSOWANA  
SY÷TETTY  
WZORCUJ NX  
MIERZONE  
CALIBRARE NX  
MISURATO  
IJKING NX-METING  
GEMETEN  
CALIBRATE NX  
MEASURED  
APPLIED  
KALIBROI NX  
MITATTU  
STOSOWANE  
APPLICATO  
AANGELEGD  
SY÷TETTY  
WZORCUJ CO  
MIERZONE  
CALIBRARE CO  
MISURATO  
IJKING CO-METING  
GEMETEN  
CALIBRATE CO  
MEASURED  
APPLIED  
KALIBROI CO  
MITATTU  
STOSOWANE  
APPLICATO  
AANGELEGD  
SY÷TETTY  
TESTUJ GAZ CO/H2  
WARTOSC CO  
PROVA GAS CO/H2  
VALORE CO  
CO/H2-METING  
CO-GEHALTE  
TESTGAS CO/H2  
CO-VALUE  
TESTIKAASU CO/H2  
CO ARVO  
WPROWADZ WARTOSC CO  
IMMETTERE VALORE CO  
GEEF DE WAARDE IN  
ENTER CO-VALUE  
SYˆ T‰ CO ARVO  
WZORCUJ H2  
MIERZONE  
CALIBRARE H2  
MISURATO  
IJKING H2-METING  
GEMETEN  
CALIBRATE H2  
MEASURED  
APPLIED  
KALIBROI H2  
MITATTU  
STOSOWANE  
APPLICATO  
AANGELEGD  
SY÷TETTY  
WZORCUJ CIAG  
MIERZONY  
CALIBRARE TIRAGGIO  
MISURATO  
IJKING TREK-METING  
GEMETEN  
CALIBRATE DRAFT  
MEASURED  
KALIBROI PAINE  
MITATTU  
STOSOWANY  
APPLICATO  
AANGELEGD  
APPLIED  
SYˆTETTY  
UZYWACZA IMIE  
LINIA n  
NOME DELLíUTENTE  
LINEA n  
GEBRUIKERS ID  
REGEL n  
ANVANDAR NAMN  
LINE n  
K‰YTT‰J‰NIMI  
RIVI n  
ZLA KAL KALIBRACJA  
ERRORE CALIBRAZIONE  
PURGA SENSORE CO  
VALSE INGAVE  
BAD CALIBRATION ENTRY  
RENSAR CO SENSOR  
VAARA KALIBROINTIARVO  
PUHDISTETAAN CO-ANTURI  
PRZEDMUCH SENSORA CO  
REINIGING CO SENSOR  
Instruction 24-9351  
A-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
PCA  
NOTES:  
A-10  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix B  
APPENDIX B – Printout Translations  
English  
(ENG)  
German  
(DEU)  
Danish  
(DAN)  
Combustion Test Data  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
SN: xxxxxx  
SN: xxxxxx  
====================  
====================  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
TIME  
DATE  
09:03:27  
28.08.1998  
Zeit  
09:03:27  
TID09:03:27  
Datum 28.08.1998  
DATO  
28.08.1998  
FUEL  
NATGAS  
BRENNSTOFF  
Erdgas  
FUEL  
N-GAS  
STACK-TEMP  
AMB.-TMP  
O2  
CO2  
CO  
0% COR CO  
NX  
0% COR NX  
qA  
190 °C  
20.0 °C  
4.0 %  
Abgas-Temp  
Raum-Temp  
O2  
CO2  
CO  
COunverd.  
NX  
NXunverd.  
qA  
190 °C  
ROEGTEMP  
LUFTTMP  
O2  
CO2  
CO  
CO UFORT  
NX  
NX UFORT  
qA  
190 °C  
20.0 °C  
4.0 %  
9.5 %  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
1.24  
20.0 °C  
4.0 %  
9.5 %  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
1.24  
9.5 %  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
LAMBDA  
1.24  
Lambda  
LAMBDA  
DRAFT  
-0.12 MB  
Feinzug  
-0.12 MB  
TRAEK  
-0.12 MB  
---------------------  
COMMENTS:  
---------------------  
Kommentar:  
---------------------  
KOMMENTAR:  
Draft Only  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
SN: xxxxxx  
SN: xxxxxx  
====================  
====================  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
TIME  
DATE  
13:03:45  
28.08.1998  
Zeit  
13:03:45  
TID13:03:45  
Datum 28.08.1998  
DATO  
28.08.1998  
DRAFT  
FEINZUGMESSUNG  
TRAEK  
DRAFT  
-0.37 MB  
Feinzug  
---------------------  
Kommentar:  
-0.37 MB  
TRAEK  
-0.37 MB  
---------------------  
COMMENTS:  
---------------------  
KOMMENTAR:  
Instruction 24-9351  
B-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B  
PCA  
French  
(FRA)  
Spanish  
(ESP)  
Polish  
(POL)  
Combustion Test Data  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
SN: xxxxxx  
SN: xxxxxx  
====================  
====================  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
HEURE 09:03:27  
DATE 28.08.1998  
HORA  
09:03:27  
CZAS  
DATE  
09:03:27  
28.08.1998  
FECHA 28.08.1998  
COMBUSTBL  
GAZ NATU  
COMBUSTB  
GNATURAL  
PALIWO  
GAZ ZIEN  
TEMP-CHMNE  
TEMP-AMBTE  
O2  
190 °C  
20.0 °C  
.0 %  
TEMP-CHIME  
TEMP-AMBTE  
O2  
190 °C  
20.0 °C  
4.0 %  
TEMP-KOMIN  
TEMP-OTOCZ  
O2  
190 °C  
20.0 °C  
4.0 %  
CO2  
9.5 %  
CO2  
9.5 %  
CO2  
9.5 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
0% CO COR  
NX  
0% NX COR  
qA  
0% CORR CO  
NX  
0% CORR NX  
qA  
0% KOR CO  
NX  
0% KOR NX  
qA  
LAMBDA  
1.24  
LAMBDA  
1.24  
LAMBDA  
1.24  
TIRAGE  
-0.12 MB  
TIRO  
-0.12 MB  
CAIG  
-0.12 MB  
---------------------  
COMMENTRES:  
---------------------  
COMENTARIO:  
---------------------  
UWAGI:  
Draft Only  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
====================  
====================  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
HEURE 13:03:45  
HORA  
13:03:45  
CZAS  
DATE  
13:03:45  
28.08.1998  
DATE  
25.08.96  
FECHA 28.08.1998  
TIRAGE  
TIRO  
DRAFT  
TIRAGE  
---------------------  
COMMENTRES:  
-0.37 MB  
TIRO  
-0.37 MB  
CIAG  
-0.37 MB  
---------------------  
COMENTARIO:  
---------------------  
UWAGI:  
B-2  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PCA  
Appendix B  
Italian  
(ITA)  
Dutch  
(NED)  
Swedish  
(SVE)  
Combustion Test Data  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
SN: xxxxxx  
SN: xxxxxx  
====================  
====================  
====================  
[ID1: optional data]  
[ID2: optional data]  
[ID3: optional data]  
[ID1: optional data]  
[ID2: optional data]  
[ID3: optional data]  
[ID1: optional data]  
[ID2: optional data]  
[ID3: optional data]  
ORA  
DATA  
09:03:27  
28.08.1998  
TIJD  
09:03:27  
Tid  
09:03:27  
DATUM 28.08.1998  
Datum 28.08.1998  
COMBUSTI  
METANO  
BRANDST.  
AARDGAS  
BRANSLE  
N_GAS  
CAMINO-TEM  
AMB-TEMP  
O2  
190 °C  
20.0 °C  
4.0 %  
ROOKGAS-T°  
OMG.-T°  
O2  
190 °C  
20.0 °C  
4.0 %  
Rokgastemp  
Luft temp  
O2  
190 °C  
20.0 °C  
4.0 %  
CO2  
9.5 %  
CO2  
9.5 %  
CO2  
9.5 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
0% COR CO  
NX  
0% COR NX  
qA  
0% CO COR  
NX  
0% COR NX  
qA  
CO outsp.  
NX  
NX outsp.  
qA  
LAMBDA  
1.24  
LAMBDA  
1.24  
LAMBDA  
1.24  
TIRAGGIO  
-0.12 MB  
TREK  
-0.12 MB  
Drag  
-0.12 MB  
---------------------  
COMMENTI:  
---------------------  
OPM.:  
---------------------  
Kommentar:  
Draft Only  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
====================  
====================  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
ORA  
13:03:45  
TIJD  
13:03:45  
Tid  
13:03:45  
DATA 28.08.1998  
DATUM 28.08.1998  
Datum 28.08.1998  
TIRAGGIO  
TREK-METING  
DRAG  
TRIAGGIO  
---------------------  
COMMENTI:  
-0.37 MB  
TREK  
-0.37 MB  
Drag  
-0.37 MB  
---------------------  
OPM.:  
---------------------  
Kommentar:  
Instruction 24-9351  
B-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B  
PCA  
Finnish  
(FIN)  
Combustion Test Data  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
====================  
[ID1: optional data]  
[ID2: optional data]  
[ID3: optional data]  
AIKA  
09:03:27  
PÄIVÄ 28.08.1998  
POLTTOAI  
MAAKAASU  
SKA LäMPöT  
YMP LäMPöT  
O2  
190 °C  
20.0 °C  
4.0 %  
CO2  
9.5 %  
CO  
12 ppm  
15 ppm  
10 ppm  
12 ppm  
8.1 %  
CO HAPETON  
NX  
NX HAPETON  
SAVUPIIPHä  
ILMäYLIMää  
1.24  
PAINE  
-0.12 MB  
---------------------  
KOMMENTIT:  
Draft Only  
[Line 1: user name]  
[Line 2: user name]  
[Line 3: user name]  
BACHARACH, INC.  
PCA 65  
SN: xxxxxx  
====================  
ID1: [optional data]  
ID2: [optional data]  
ID3: [optional data]  
AIKA  
13:03:45  
PäIVä 28.08.1998  
VETO  
PAINE  
-0.37 MB  
---------------------  
KOMMENTIT:  
B-4  
Instruction 24-9351  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
World Headquarters  
621 Hunt Valley Circle, New Kensington, PA 15068  
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 1-800-736-4666  
Website: www.bacharach-inc.com • E-mail: [email protected]  
Printed in U.S.A.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Aastra Telecom Telephone 2025 User Manual
Accusplit Fitness Electronics AE140XL User Manual
Accusplit Watch AE800AC User Manual
Acer Laptop m3400g User Manual
Acoustic Research Speaker System AW822 User Manual
Agilent Technologies Yard Vacuum 437B User Manual
Altinex Switch MT110 103 User Manual
Audiovox Stereo Amplifier AXT 500 User Manual
Axis Communications Home Security System 25733 User Manual
Axis Communications Security Camera 209FD R M12 User Manual