Specifications Guide
Agilent Technologies
ESA-L Series Spectrum Analyzers
This manual provides documentation for the following instruments:
ESA-L Series
E4403B (9 kHz – 3.0 GHz)
E4408B (9 kHz – 26.5 GHz)
E4411B (9 kHz – 1.5 GHz)
Manufacturing Part Number: E4403-90036
Supersedes: E4403-90032
Printed in USA
September 2003
© Copyright 2000 - 2003 Agilent Technologies
Download from Www.Somanuals.com. All Manuals Search And Download.
Warranty
This Agilent Technologies instrument product is warranted against defects in
material and workmanship for a period of one year from date of shipment. During
the warranty period, Agilent Technologies Company will, at its option, either
repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to
Agilent Technologies and Agilent Technologies shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by
Agilent Technologies for use with an instrument will execute its programming
instructions when properly installed on that instrument. Agilent Technologies does
not warrant that the operation of the instrument, or software, or firmware will be
uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about Agilent
Spectrum Analyzers, including firmware upgrades and application information,
please visit the following Internet URL:
http://www.agilent.com/find/esa
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Con t en t s
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Tracking Generator (Option 1DN or 1DQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Con t en t s
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
About This Chapter
About This Chapter
This chapter contains specifications and characteristics for the Agilent
E4403B spectrum analyzer. The distinction between specifications and
characteristics is described as follows.
•
•
•
Specifications describe the performance of parameters covered by
the product warranty. (The temperature range is 0 °C to 55 °C,
unless otherwise noted.)
Characteristics describe product performance that is useful in the
application of the product, but is not covered by the product
warranty.
Typical performance describes additional product performance
information that is not covered by the product warranty. It is
performance beyond specification that 80% of the units exhibit with
a 95% confidence level over the temperature range 20 to 30 °C.
Typical performance does not include measurement uncertainty.
•
Nominal values indicate the expected performance, or describe
product performance that is useful in the application of the product,
but is not covered by the product warranty.
The following conditions must be met for the analyzer to meet its
specifications.
o The analyzer is within the one year calibration cycle.
o If Auto Align All is selected:
— After 2 hours of storage within the operating temperature range.
— 5 minutes after the analyzer is turned on with sweep times less
than 4 seconds.
— After the front-panel amplitude reference is connected to the
INPUT, and Align Now RF has been run, after the analyzer is
turned on. And, once every 24 hours, or if ambient temperature
changes more than 30 °C.
8
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
About This Chapter
o If Auto Align Off is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and Align Now All has been run.
— When Align Now All is run:
•
•
•
Every hour
If the ambient temperature changes more than 3 °C
If the 10 MHz reference changes
— When Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
•
•
Every 24 hours
If the ambient temperature changes more than 30 °C
o If Auto Align All but RF is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and Align Now RF has been run.
— When Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
•
•
Every hour
If the ambient temperature changes more than 3 °C
Chapter 1
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Frequency
Frequency
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
F r eq u en cy R a n ge
9 kHz to 3.0 GHz
Sp ecifica t ion s
Su p p lem en t a l In for m a tion
F r eq u en cy R efer en ce
–6
–7
Aging Rate
±2 × 10 /year
±1.0 × 10 /day, characteristic
–7
Settability
±5 × 10
–6
Temperature Stability
±5 × 10
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
F r eq u en cy R ea d ou t
Accu r a cy
(Start, Stop, Center, Marker)
±((frequency indication ×
a
frequency reference error )
+ 0.75% of span
+ 15% of RBW
+ 10 Hz)
a. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma r k er F r eq u en cy
Cou n ter
Resolution
Selectable from 1 Hz to
100 kHz
a
±(marker frequency ×
Accuracy
b
frequency reference error
counter resolution)
+
a. Marker level to displayed noise level > 25 dB, RBW/ Span ≥ 0.002, frequency offset = 0
Hz.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
10
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
F r eq u en cy Sp a n
Range
0 Hz (zero span), 100 Hz to
3 GHz
Resolution
Accuracy
2 Hz
±1.0% of span
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Sw eep Tim e
a
Range
4 ms to 4000 s
Tracking Generator On
(Option 1DN)
50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
±1%
4 ms to 4000 s
b
c
Sweep Trigger
Free Run, Single, Line, Video ,
d
External, Delayed, Offset
e
Delayed Trigger,
Range
1 µs to 400 s
Resolution
delay in seconds
---------------------------------------
65000
rounded up to nearest µs
Accuracy
±(500 νσ + (0.01% of delay))
Offset Trigger
Resolution
sweep time
--------------------------
400
Range
±327 ms to ±323 ks
Where ST = sweep time
–32766 × ST 32365 × ST
------------------------------- ---------------------------
to
400
400
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is suspended in video, external, and delayed trigger modes while waiting for a
trigger event to occur.
c. Unavailable when RBW ≤ 300 Hz (Option 1DR).
d. For firmware revision A.04.00 or later.
e. Delayed trigger is available with line and external trigger.
Chapter 1
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Frequency
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Sw eep (t r a ce) P oin t s
401
Sp ecifica tion s
R esolu t ion Ba n d w id t h
(R BW)
Range
–3 dB bandwidth
1 kHz to 3 MHz, in 1-3-10
sequence, 5 MHz
a
(Option 1DR)
Adds 100, 300 Hz
–6 dB bandwidth (EMI)
9 kHz and 120 kHz
(Option 1DR)
Add 200 Hz
Accuracy
100 Hz to 300 Hz (–3 dB)
RBW
±10%
±15%
(Option 1DR)
1 kHz to 3 MHz (–3 dB)
RBW
5 MHz (–3 dB) RBW
±30%
±20%
9 kHz, 120 kHz (–6 dB)
RBW
(EMI)
200 Hz (–6 dB) RBW (EMI)
(Option 1DR)
±10%
Shape
100 Hz to 300 Hz RBW
(Option 1DR)
Digital, approximately
Gaussian shape
1 kHz to 5 MHz RBW
Synchronously tuned four
poles, approximately
Gaussian shape
Selectivity (60 dB/3 dB
bandwidth ratio)
100 Hz to 300 Hz RBW
(Option 1DR)
<5:1, nominal
<15:1, nominal
1 kHz to 5 MHz RBW
a. Only available in spans ≤5 MHz, sweep times ≥ 4 ms, and not usable with tracking gener-
ator on (Option 1DN).
12
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Vid eo Ba n d w id t h (VBW)
(–3 dB)
Range
30 Hz to 1 MHz in 1-3-10
sequence
3 MHz, characteristic
(Option 1DR)
Adds 1, 3, 10 Hz for RBW’s
<1 kHz
Accuracy
Shape
±30%, characteristic
Post detection, single pole low-
pass filter used to average
displayed noise
Video bandwidths below 30 Hz
are digital bandwidths with
anti-aliasing filtering.
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
St a b ility
Noise Sidebands
(Offset from CW signal with
1 kHz RBW, 30 Hz VBW and
sample detector)
≥1 kHz (Option 1DR)
≥10 kHz
≤ −78 dBc/Hz, typical
≤ −94 dBc/Hz, typical
≤ −105 dBc/Hz, typical
≤ −112 dBc/Hz, typical
≤ −122 dBc/Hz, typical
≤ −90 dBc/Hz
≤ −100 dBc/Hz
≤ −106 dBc/Hz
≤ −118 dBc/Hz
≥20 kHz
≥30 kHz
≥100 kHz
Residual FM
1 kHz RBW, 1 kHz VBW
≤150 Hz p–p in 100 ms
100 Hz RBW, 100 Hz VBW
(Option 1DR)
≤30 Hz p–p in 20 ms,
characteristic
System-Related Sidebands,
offset from CW signal
≥30 kHz
≤ −65 dBc
Chapter 1
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Frequency
Noise Sid eb a n d s Nor m a lized t o 1 H z Ver su s Offset fr om Ca r r ier
14
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Amplitude
Amplitude specifications do not apply for the negative peak detector
mode.
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Mea su r em en t R a n ge
Displayed Average Noise
Level to Maximum Safe Input
Level
Input Attenuator Range
0 to 65 dB, in 5 dB steps
0 to 75 dB, in 5 dB steps,
characteristic
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Ma xim u m Sa fe In p u t Level
Average Continuous Power
+30 dBm (1 W)
(Input attenuator setting
≥5 dB)
Peak Pulse Power
+50 dBm (100 W)
(for <10 µsec pulse width,
<1% duty cycle, and input
attenuation ≥30 dB)
dc
100 Vdc
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
1 d B Ga in Com p r ession
a b
Total power at input mixer ,
50 MHz to 3.0 GHz
0 dBm
a. Mixer power level (dBm) = input power (dBm) – input attenuation (dB).
b. For resolution bandwidths 1 kHz to 30 kHz, the maximum input signal amplitude must
be ≤ reference level +10 dB.
Chapter 1
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Disp la yed Aver a ge Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = –70 dBm)
1 kHz RBW
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
1 MHz to 10 MHz
10 MHz to 1.0 GHz
1.0 GHz to 2.0 GHz
2.0 GHz to 3.0 GHz
≤ −117 dBm
≤ −120 dBm
≤ −120 dBm
≤ −120 dBm
≤ −126 dBm
≤ −130 dBm
≤ −130 dBm
≤ −130 dBm
≤ −117 dBm
≤ −116 dBm
≤ −114 dBm
≤ −125 dBm
≤ −124 dBm
≤ −122 dBm
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y R a n ge
Log Scale
Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW ≥ 1 kHz
Calibrated 0 to −85 dB from
Reference Level
a
RBW ≤ 300 Hz (Option 1DR)
Calibrated 0 to −120 dB from
Reference Level
Linear Scale
Scale Units
Ten divisions
dBm, dBmV, dBµV, dBµA, A,
V, and W
a. 0 to –70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).
16
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Ma r k er R ea d ou t
R esolu tion
Log scale
RBW ≥ 1 kHz
0 to −85 dB from ref level
RBW ≤ 300 Hz (Option 1DR)
0.04 dB
0 to −120 dB from ref level 0.04 dB
Linear scale
0.01% of Reference Level
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
a
F r eq u en cy R esp on se
10 dB attenuation
9 kHz to 3.0 GHz
20 to 30 °C
±0.5 dB
±1.0 dB
0 to 55 °C
a. Frequency response values are referenced to the amplitude at 50 MHz.
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
In p u t At ten u a t ion
Sw it ch in g Un cer t a in t y a t
50 MH z
Attenuator Setting
0 dB to 5 dB
10 dB
±0.3 dB
Reference
±0.3 dB
15 dB
20 to 65 dB attenuation
±(0.1 dB + 0.01 × Attenuator
Setting)
Chapter 1
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
At t en u a t ion Accu r a cy R ela t ive t o th e 10 d B At t en u a t or Sett in g, Ch a r a cter ist ic
F r eq u en cy Ra n ge
At t en u a t ion
d c–3.0 GH z
0 dB
±0.3 dB
±0.3 dB
Reference
±0.4 dB
±0.4 dB
±0.5 dB
±0.5 dB
±0.6 dB
±0.6 dB
±0.7 dB
±0.7 dB
±0.9 dB
±0.9 dB
±1.0 dB
5 dB
10 dB
15 dB
20 dB
25 dB
30 dB
35 dB
40 dB
45 dB
50 dB
55 dB
60 dB
65 dB
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ab solu t e Am p lit u d e
Accu r a cy
a
±0.4 dB
At reference settings
Over a ll Am p litu d e
b
Accu r a cy
20 to 30 °C
± (0.6 dB + Absolute
Frequency Response)
a. Settings are: reference level –20 dBm; input attenuation 10 dB; center frequency 50
MHz; RBW 1 kHz; VBW 1 kHz; amplitude scale linear or log; span 2 kHz; sweep time
coupled, signal at reference level.
b. For reference level 0 to –50 dBm; input attenuation 10 dB; RBW 1 kHz; VBW 1 kHz;
amplitude scale log, log range 0 to –50 dB from reference level; sweep time coupled; sig-
nal input 0 to –50 dBm; span ≤20 κΗζ.
18
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
R F In p u t VSWR (at tuned
frequency)
Attenuator setting 0 dB
100 kHz to 3 GHz
≤3.0:1, characteristic
≤1.6:1, characteristic
Attenuator setting 5 dB
100 kHz to 3 GHz
Attenuator setting 10 to 65 dB
9 kHz to 100 kHz
≤2.0:1, characteristic
≤1.4:1, characteristic
100 kHz to 3 GHz
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
a
Au t o Align m en t
Sweep-to-sweep variation
±0.1 dB, characteristic
a. Set Auto Align to Off and use Align Now, All to eliminate this variation.
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
R esolu tion Ba n d w id t h
Sw it ch in g Un cer t a in t y
(at Reference Level)
1 kHz RBW
Reference
±0.3 dB
±0.6 dB
±0.3 dB
3 kHz to 3 MHz RBW
5 MHz RBW
100 Hz to 300 Hz RBW
(Option 1DR)
Chapter 1
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R efer en ce Level
Range
–149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale
±0.1 dB
Linear Scale
±0.12% of Reference Level
Accuracy (at a fixed frequency,
a fixed attenuator, and
referenced to −30 dBm)
Reference Level (dBm) – input
attenuator setting (dB)
–10 dBm to > –60 dBm
–60 dBm to > –85 dBm
–85 dBm to –90 dBm
±0.3 dB
±0.5 dB
±0.7 dB
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y Sca le Sw it ch in g
Un cer ta in t y
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching
No error
20
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Disp la y Sca le F id elit y
Log Maximum Cumulative
RBW ≥ 1 kHz
0 to 85 dB Below
Reference
±(0.3 dB + 0.01 × dB from
reference level)
Level
RBW ≤ 300 Hz (Option 1DR)
Span > 0 Hz
Auto range On
a
±(0.3 dB + 0.01 × dB from
reference level)
0 to 98 dB below
reference level
> 98 to 120 dB below
reference level
±2.0 dB, characteristic
b
Auto range Off
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Span = 0 Hz
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Log Incremental Accuracy
a,c
±0.4 dB/4 dB
0 to 80 dB below reference
level
Linear Accuracy
±2% of Reference Level
a. 0 to 30 dB for RBW = 200 Hz
b. The SCPI command for auto range off is:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF)
c. 0 to 50 dB for RBWs ≤ 300 Hz and span = 0 Hz, or when auto ranging is off.
Chapter 1
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Sp u r iou s R esp on ses
Second Harmonic Distortion
Input Signal
10 MHz to 500 MHz
< −60 dBc for −30 dBm signal
at input mixer
+30 dBm SHI (second
harmonic intercept)
a
500 MHz to 1.5 GHz
< −70 dBc for −30 dBm signal
at input mixer
+40 dBm SHI
Third Order Intermodulation
Distortion
10 MHz to 100 MHz
+5 dBm TOI (third order
intercept), characteristic
100 MHz to 3 GHz
< −75 dBc for two −30 dBm
signals at input mixer and
+7.5 dBm TOI
>50 kHz separation
Other Input Related Spurious
>30 kHz offset
< −65 dBc for −20 dBm signal
at input mixer
a. Mixer power level (dBm) = input power (dBm – input attentuation (dB)
22
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Amplitude
Sp ecified Dyn a m ic R a n ge
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
R esid u a l R esp on ses (Input
terminated and 0 dB
attenuation)
150 kHz to 3 GHz
< −90 dBm
Chapter 1
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Options
Options
Tracking Generator (Option 1DN)
The spectrum analyzer/tracking generator combination will meet its
specification after a cable (8120-5148) and adapter are connected
between RF OUT and INPUT and Align Now, TG has been run.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Wa r m -u p
5 minutes
Sp ecifica tion s
Ou t p u t F r eq u en cy R a n ge
Minimum Resolution BW
9 kHz to 3.0 GHz
Sp ecifica tion s
1 kHz
Not usable with resolution
bandwidths ≤300 Hz
(Option 1DR)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t P ow er Level
Range
–2 to −66 dBm
0.1 dB
Resolution
Absolute Accuracy (at 50 MHz
with coupled source
attenuator, referenced to
–20 dBm)
± 0.75 dB
Vernier
Range
8 dB
Accuracy (with coupled
source attenuator, 50 MHz,
–20 dBm)
Incremental
Cumulative
±0.2 dB/dB
±0.5 dB, total
Output Attenuator Range
0 to 56 dB in 8 dB steps
24
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Options
Sp ecifica t ion s
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Ma xim u m Sa fe R ever se
Level
+30 dBm (1 W), 50 Vdc,
characteristic
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Ou t p u t P ow er Sw eep
Range
(−10 to −2 dBm) − (Source
Attenuator Setting)
Resolution
0.1 dB
Accuracy (zero span)
<1 dB peak-to-peak
Sp ecifica t ion s
Ou t p u t F la t n ess
Referenced to 50 MHz,
–20 dBm
9 kHz to 10 MHz
10 MHz to 3 GHz
±3 dB
±2 dB
Sp ecifica t ion s
Sp u r iou s Ou t p u t s
(–2 dBm output)
Harmonic Spurs
TG Output 9 kHz to 20 kHz
≤ −15 dBc
TG Output 20 kHz to 3 GHz ≤ −25 dBc
Non-harmonic Spurs
TG Output 9 kHz to 2 GHz
TG Output 2 GHz to 3 GHz
LO Feedthrough
≤ −27 dBc
≤ −23 dBc
LO Frequency
3.921409 GHz to
6.9214 GHz
≤ −16 dBm
Chapter 1
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Options
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Dyn a m ic Ra n ge
Maximum Output Power Level
– Displayed Average Noise
Level
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t Tr a ck in g
Drift
1.5 kHz/5 minute,
characteristic
Swept Tracking Error
Usable in 1 kHz RBW after 5
minutes of warm-up
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R F P ow er -Off R esid u a ls
9 kHz to 3 GHz
< −120 dBm, characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t At t en u a t or
R ep ea ta b ilit y
9 kHz to 300 MHz
300 MHz to 2 GHz
2 GHz to 3 GHz
±0.1 dB, characteristic
±0.2 dB, characteristic
±0.3 dB, characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t VSWR
0 dB attenuation
≥ 8 dB attenuation
<2.0:1, characteristic
<1.5:1, characteristic
26
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Options
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Ou t p u t At ten u a tor
Accu r a cy
0 dB
±0.5 dB, characteristic
±0.5 dB, characteristic
8 dB
16 dB
24 dB
32 dB
40 dB
48 dB
56 dB
Reference
±0.5 dB, characteristic
±0.6 dB, characteristic
±0.8 dB, characteristic
±1.0 dB, characteristic
±1.1 dB, characteristic
Tr a ck in g Gen er a t or Ou t p u t Accu r a cy
R ela t ive Accu r a cy (R efer r ed to –20 d Bm ) =
Ou t p u t At t en u a t or Accu r a cy + Ver n ier Accu r a cy + Ou t p u t F la t n ess
Absolute Accuracy =
Relative Accuracy (Referred to –20 dBm) + Absolute Accuracy at 50 MHz
Chapter 1
27
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
General
General
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Tem p er a tu r e R a n ge
Operating
Storage
0 to 55 °C
Floppy disk 10 to 40 °C
∠40 to 75 °C
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Au d ib le Noise
(ISO 7779)
Sound Pressure at 25 °C
<40 dBa, (<4.6 Bels power)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Milit a r y Sp ecifica tion
Has been type tested to the
environmental specifications
of MIL-PRF-28800F class 3.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
EMI Com p a t ib ilit y
Conducted and radiated
emission is in compliance with
CISPR Pub. 11/1990 Group 1
Class A.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Im m u n it y Test in g
Radiated Immunity
Testing was done at 3 V/m
according to IEC 801-3/1984.
When the analyzer tuned
frequency is identical to the
immunity test signal
frequency, there may be
signals of up to −60 dBm
displayed on the screen.
Electrostatic Discharge
Air discharges of up to 8 kV
were applied according to IEC
801-2/1991. Discharges to
center pins of any of the
connectors may cause damage
to the associated circuitry.
28
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
General
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
P ow er Req u ir em en t s
ac Operation
Voltage, frequency
90 to 132 Vrms, 47 to 440 Hz
195 to 250 Vrms, 47 to 66 Hz
Power Consumption, On
<300 W
<5 W
Power Consumption,
Standby
dc Operation
Voltage
12 to 20 Vdc
<200 W
Power Consumption
Power Consumption,
Standby
<100 mW
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Mea su r em en t Sp eed
Local Measurement and
a
≥ 30/s, characteristic
≥30/s, characteristic
Display Update rate
Remote Measurement and
b c
GPIB Transfer Rate , (Option
A4H)
RF Center Frequency Tune,
Measure, and GPIB Transfer
≤ 90 ms, characteristic
b d
Time , (Option A4H)
a. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, and spans >10 MHz
and ≤600 MHz.
b. Display Off (:DISPlay:ENABle OFF), and 32-bit integer data format (:FORMat:DATA
INT,32), if Option A4J is installed, disable sweep ramp, (:SYS-
Tem:PORTs:IFVSweep:ENABle OFF), markers Off, single sweep, measured with IBM
compatible PC with 550 MHz Pentium III running
Windows NT 4.0, one meter GPIB cable, National Instruments PCI-GPIB card and
NI-488.2 DLL.
c. Factory preset, auto align Off, RBW = 1 MHz, span= 20 MHz, fixed center frequency,
average of 100 measurements.
d. Factory preset, auto align Off, RBW = 1 MHz, span= 20 MHz, and center frequency tune
step size = 50 MHz.
Chapter 1
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
General
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Da t a St or a ge
a
8.0 MB, nominal
Internal
External
(10 to 40 °C)
3.5" 1.44 MB, MS-DOS
compatible floppy disk
a. For serial numbers prior to US41440000 or MY41440000, 1 MB without Option B72, 8
MB with Option B72.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Memory Usage
State
a
20 kB , nominal
State plus 401-point trace
Applications memory usage
21 kB , nominal
8590 Compatibility
0.7 MB, nominal
(Option 290)
a. The size of state will increase depending on installed applications.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Dem od Tu n e a n d List en
Internal speaker, front-panel
earphone jack and front-panel
volume control.
Demod
AM
(Option A4J )
An uncalibrated demodulated
signal is available on the AUX
VIDEO OUT connector at the
rear panel.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Weigh t (without options)
Net
15.5 kg (34.2 lb),
characteristic
Shipping
27.4 kg (60.4 lb),
characteristic
30
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
General
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
a
Disp la y
Resolution
640 × 480
a. The LCD display is manufactured using high precision technology. However, there may
be up to six bright points (white, blue, red or green in color) that constantly appear on the
LCD screen. These points are normal in the manufacturing process and do not affect the
measurement integrity of the product in any way.
Dim en sion s
Chapter 1
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Inputs and Outputs
Inputs and Outputs
Front Panel
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
INP UT 50 Ω
Connector
Type-N female
Impedance
50 Ω, nominal
LO Emissions
< –70 dBm, characteristic.
Average level of 1st LO,
3.9214 to 6.9214 GHz, present
at INPUT 50 Ω connector.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R F OUT 50 Ω, (Option 1DN)
Connector
Type-N female
Impedance
50 Ω, nominal
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
a
Amplitude Reference
AMP TD RE F OUT
Connector
BNC female
Impedance
50 Ω, nominal
Frequency
50 MHz
b
Frequency Accuracy
Frequency reference error
c
−20 dBm, nominal
50 Ω Amplitude
a. Turn the amplitude reference on/off by pressing the keys: Input/Output, Amptd Ref Out.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
c. The internal amplitude reference actual power is stored internally.
32
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
P R OBE P OWER
Voltage/Current
+15 Vdc, ±7% at 150 mA max.,
characteristic
−12.6 Vdc ±10% at 150 mA
max., characteristic
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
a
Used for entering screen titles
and filenames only. Interface
compatible with most
IBM-compatible PC
E XT KE YBOAR D
keyboards.
Connector
6-pin mini-DIN
a. The feature is not implemented in firmware revisions prior to A.04.00.
Sp ecifica t ion s
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Sp ea k er
Front panel knob controls
volume
Su p p lem en t a l In for m a t ion
H ea d p h on e
Connector
Front panel knob controls
volume
3.5 mm (1/8 inch) miniature
audio jack
Power Output
0.2 W into 4 Ω, characteristic
Rear Panel
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
10 MH z R E F OUT
Connector
BNC female
Impedance
50 Ω, nominal
Output Amplitude
>0 dBm, characteristic
Chapter 1
33
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
10 MH z R EF IN
Connector
BNC female
Note: Analyzer noise
sidebands and spurious
response performance may be
affected by the quality of the
external reference used.
Impedance
50 Ω, nominal
Input Amplitude Range
Frequency
−15 to +10 dBm, characteristic
10 MHz, nominal
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GATE TR IG/E XT TR IG IN
Connector
BNC female
External Trigger Input
Trigger Level
Selectable positive or negative
edge initiates sweep in EXT
TRIG mode (5 V TTL)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GATE/H I SWP OUT
Connector
BNC female
High Sweep Output
Level
a
High = sweep ; Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
VGA OUTP UT
Connector
VGA compatible, 15-pin mini
D-SUB
Format
VGA (31.5 kHz horizontal,
60 Hz vertical sync rates,
non-interlaced) Analog RGB
Resolution
640 × 480
34
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
AUX IF OUT
(Option A4J )
ΡΒΩ ≥ 1 κΗζ
Connector
Frequency
BNC female
21.4 MHz, nominal
Amplitude (for signal at
reference level and for
reference levels – input
attenuation of –10 to
–70 dBm)
−10 dBm (uncorrected),
characteristic
Impedance
50 Ω, nominal
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
AUX VIDEO OUT
(Option A4J )
ΡΒΩ ≥ 1 κΗζ
Connector
BNC female
Amplitude Range (into
>10 kΩ)
0 to 1 V (uncorrected),
characteristic
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
H I SWP IN
(Option A4J )
Connector
Input
BNC female
Open collector, low resets and
holds the sweep (5 V TTL)
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
H I SWP OUT
(Option A4J )
Connector
Output
BNC female
a
High = sweep , Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Chapter 1
35
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
SWP OUT
(Option A4J )
Connector
Amplitude
BNC female
0 to +10 V ramp,
characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GP IB In t er fa ce
(Option A4H)
Connector
IEEE-488 bus connector
GPIB Codes
SH1, AH1, T6, SR1, RL1, PP0,
DC1, C1, C2, C3 and C28
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ser ia l In ter fa ce
(Option 1AX)
Connector
9-pin D-SUB male
RS-232
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
P a r a llel In t er fa ce
Printer port only
(Option A4H or 1AX)
Connector
25-pin D-SUB female
36
Chapter1
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4403B Specifications and Characteristics
Regulatory Information
Regulatory Information
CAUTION
NOTE
This product is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 664 respectively.
This product has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronic Measuring
Apparatus, and has been supplied in a safe condition. The instruction
documentation contains information and warnings which must be
followed by the user to ensure safe operation and to maintain the
product in a safe condition.
The CE mark is a registered trademark of the European Community (if
accompanied by a year, it is the year when the design was proven).
The CSA mark is the Canadian Standards Association safety mark.
ISM 1-A
This is a symbol of an Industrial Scientific and Medical Group 1 Class A
product. (CISPR 11, Clause 4)
Chapter 1
37
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
About This Chapter
Ab ou t Th is Ch a p ter
This chapter contains specifications and characteristics for the Agilent
E4408B spectrum analyzer. The distinction between specifications and
characteristics is described as follows.
•
•
•
Specifications describe the performance of parameters covered by
the product warranty. (The temperature range is 0 °C to 55 °C,
unless otherwise noted.)
Characteristics describe product performance that is useful in the
application of the product, but is not covered by the product
warranty.
Typical performance describes additional product performance
information that is not covered by the product warranty. It is
performance beyond specification that 80% of the units exhibit with
a 95% confidence level over the temperature range 20 to 30 °C.
Typical performance does not include measurement uncertainty.
•
Nominal values indicate the expected performance, or describe
product performance that is useful in the application of the product,
but is not covered by the product warranty.
The following conditions must be met for the analyzer to meet its
specifications.
o The analyzer is within the one year calibration cycle.
o If Auto Align All is selected:
— After 2 hours of storage within the operating temperature range.
— 5 minutes after the analyzer is turned on with sweep times less
than 4 seconds.
— After the front-panel amplitude reference is connected to the
INPUT, and Align Now RF has been run, after the analyzer is
turned on. And, once every 24 hours, or if ambient temperature
changes more than 30 °C.
40
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
About This Chapter
o If Auto Align Off is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and Align Now All has been run.
— When Align Now All is run:
•
•
•
Every hour
If the ambient temperature changes more than 3 °C
If the 10 MHz reference changes
— When Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
•
•
Every 24 hours
If the ambient temperature changes more than 30 °C
o If Auto Align All but RF is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and Align Now RF has been run.
— When Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
•
•
Every hour
If the ambient temperature changes more than 3 °C
Chapter 2
41
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
F r eq u en cy
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
F r eq u en cy R a n ge
9 kHz to 26.5 GHz
a
Band
Harmonic Mixing Mode (N )
0 (0 Hz to 3.0 GHz)
1–
1–
2–
4–
4–
1 (2.85 GHz to 6.7 GHz)
2 (6.2 GHz to 13.2 GHz)
3 (12.8 GHz to 19.2 GHz)
4 (18.7 GHz to 26.5 GHz)
a. N is the harmonic mixing mode. For negative mixing modes (as indicated by the “–”), the
desired 1st LO harmonic is higher than the tuned frequency by the 1st IF (3.9214 for the
9 kHz to 3 GHz band, 321.4 MHz for all other bands).
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
F r eq u en cy R efer en ce
–6
–7
Aging Rate
±2 × 10 /year
±1.0 × 10 /day, characteristic
–7
Settability
±5 × 10
±5 × 10
–6
Temperature Stability
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
F r eq u en cy R ea d ou t
Accu r a cy
(Start, Stop, Center, Marker)
±((frequency indication ×
frequency reference error )
+ 0.75% of span
a
+ 15% of RBW
+ 10 Hz + 1 Hz × N )
b
a. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
b. N is the harmonic mixing mode.
42
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma r k er F r eq u en cy
Cou n t er
Resolution
Selectable from 1 Hz to
100 kHz
a
±(marker frequency ×
Accuracy
b
frequency reference error +
c
counter resolution)
a. Marker level to displayed noise level > 25 dB, RBW/ Span ≥ 0.002, frequency offset = 0
Hz.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
c. For firmware revisions prior to A.03.00, add 1 Hz x N, where N is the harmonic mixing
mode.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
F r eq u en cy Sp a n
Range
0 Hz (zero span), 100 Hz to
26.5 GHz
a
Resolution
Accuracy
2 Hz x N
±1.0% of span
a. N is the harmonic mixing mode.
Chapter 2
43
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sw eep Tim e
a
Range
4 ms to 4000 s
Tracking Generator On
(Option 1DN)
50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
±1%
4 ms to 4000 s
b
c
Sweep Trigger
Free Run, Single, Line, Video ,
d
External, Delayed, Offset
,e
Delayed Trigger
Range
1 µs to 400 s
Resolution
delay in seconds
---------------------------------------
65000
rounded up to nearest µs
Accuracy
±(500 ns + (0.01% of delay))
Offset Trigger
Resolution
sweep time
--------------------------
400
Range
±327 ms to ±323 ks
Where ST = sweep time
–32766 × ST 32365 × ST
------------------------------- ---------------------------
to
400
400
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is suspended in video, external, and delayed trigger modes while waiting for a
trigger event to occur.
c. Unavailable when RBW ≤ 300 Hz (Option 1DR).
d. For firmware revision A.04.00 or later.
e. Delayed trigger is available with line and external trigger.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sw eep (t r a ce) P oin t s
401
44
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en t a l In for m a tion
R esolu tion Ba n d w id t h
(R BW)
Range
–3 dB bandwidth
1 kHz to 3 MHz, in 1-3-10
sequence, 5 MHz
a
(Option 1DR)
Adds 100, 300 Hz
9 kHz and 120 kHz
–6 dB bandwidth (EMI)
(Option 1DR)
Add 200 Hz
Accuracy
100 Hz to 300 Hz (–3 dB)
RBW
±10%
(Option 1DR)
1 kHz to 3 MHz (–3 dB)
RBW
±15%
±30%
5 MHz (–3 dB) RBW
9 kHz, 120 kHz (–6 dB) RBW ±20%
(EMI)
200 Hz (–6 dB) RBW (EMI)
(Option 1DR)
±10%
Shape
100 Hz to 300 Hz RBW
(Option 1DR)
Digital, approximately
Gaussian shape
1 kHz to 5 MHz RBW
Synchronously tuned four
poles, approximately
Gaussian shape
Selectivity (60 dB/3 dB
bandwidth ratio)
100 Hz to 300 Hz RBW
(Option 1DR)
<5:1, nominal
<15:1, nominal
1 kHz to 5 MHz RBW
a. Only available in spans ≤5 MHz, sweep times ≥ 4 ms, and not usable with tracking gener-
ator on (Option 1DN).
Chapter 2
45
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Vid eo Ba n d w id t h (VBW)
(–3 d B)
Range
30 Hz to 1 MHz in 1-3-10
sequence
3 MHz, characteristic
(Option 1DR)
Adds 1, 3, 10 Hz for RBW’s
<1 kHz
Accuracy
Shape
±30%, characteristic
Post detection, single pole low-
pass filter used to average
displayed noise
Video bandwidths below 30 Hz
are digital bandwidths with
anti-aliasing filtering.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sta b ilit y
Noise Sidebands
(Offset from CW signal with
1 kHz RBW, 30 Hz VBW and
sample detector)
≥1 kHz (Option 1DR)
≤ −78 dBc/Hz , typical
a
≥10 kHz
≥20 kHz
≥30 kHz
≤ −90 dBc/Hz
≤ −94 dBc/Hz , typical
≤ −100 dBc/Hz
≤ −105 dBc/Hz , typical
≤ −106 dBc/Hz
≤ −112 dBc/Hz , typical
≥100 kHz
≤ −118 dBc/Hz
≤ −122 dBc/Hz , typical
Residual FM
1 kHz RBW, 1 kHz VBW
≤150 Hz × N p–p in 100 ms
100 Hz RBW, 100 Hz VBW
(Option 1DR)
≤30 Hz × N p–p in 20 ms,
characteristic
System-Related Sidebands,
offset from CW signal
≥30 kHz
≤ −65 dBc
a. Add 20 Log(N) for frequencies > 6.7 GHz.
46
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Frequency
Noise Sid eb a n d s Nor m a lized t o 1 H z Ver su s Offset fr om Ca r r ier
Chapter 2
47
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Am p lit u d e
Amplitude specifications do not apply for the negative peak detector
mode.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Mea su r em en t R a n ge
Displayed Average Noise
Level to Maximum Safe Input
Level
Input Attenuator Range
0 to 65 dB, in 5 dB steps
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ma xim u m Sa fe In p u t Level
Average Continuous Power
+30 dBm (1 W)
(Input attenuator setting
≥5 dB)
Peak Pulse Power
+50 dBm (100 W)
(for <10 µsec pulse width,
<1% duty cycle, and input
attenuation ≥30 dB)
dc
0 Vdc
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
1 d B Ga in Com p r ession
a,b
Total power at input mixer
50 MHz to 3.0 GHz
3.0 GHz to 6.7 GHz
6.7 GHz to 13.2 GHz
13.2 GHz to 26.5 GHz
0 dBm
0 dBm
–3 dBm
–5 dBm
a. Mixer power level (dBm) = input power (dBm) – input attenuation (dB).
b. For resolution bandwidths 1 kHz to 30 kHz, the maximum input signal amplitude must
be ≤ reference level +10 dB.
48
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Disp la yed Aver a ge Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = –70 dBm)
1 kHz RBW,
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
1 MHz to 10 MHz
10 MHz to 1.0 GHz
1.0 GHz to 2.0 GHz
2.0 GHz to 3.0 GHz
3.0 GHz to 6.0 GHz
6.0 GHz to 12 GHz
12 GHz to 22 GHz
22 GHz to 26.5 GHz
≤ −117 dBm
≤ −119 dBm
≤ −120 dBm
≤ −118 dBm
≤ −118 dBm
≤ −117 dBm
≤ −114 dBm
≤ −112 dBm
≤ −127 dBm
≤ −129 dBm
≤ −130 dBm
≤ −128 dBm
≤ −128 dBm
≤ −127 dBm
≤ −124 dBm
≤ −122 dBm
≤ −116 dBm
≤ −115 dBm
≤ −112 dBm
≤ −112 dBm
≤ −110 dBm
≤ −107 dBm
≤ −101 dBm
≤ −124 dBm
≤ −123 dBm
≤ −120 dBm
≤ −120 dBm
≤ −118 dBm
≤ −115 dBm
≤ −109 dBm
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y R a n ge
Log Scale
Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW ≥ 1 kHz
Calibrated 0 to −85 dB from
Reference Level
a
RBW ≤ 300 Hz (Option 1DR)
Calibrated 0 to −120 dB from
Reference Level
Linear Scale
Scale Units
Ten divisions
dBm, dBmV, dBµV, dBµA, A,
V, and W
a. 0 to –70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).
Chapter 2
49
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ma r k er R ea d ou t
R esolu t ion
Log scale
RBW ≥ 1 kHz
0 to −85 dB from ref level
0.04 dB
RBW ≤ 300 Hz (Option 1DR)
0 to −120 dB from ref level 0.04 dB
Linear scale
0.01% of Reference Level
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
F r eq u en cy R esp on se
a
Absolute /Relative
10 dB attenuation
9 kHz to 3.0 GHz
20 to 30 °C
±0.5 dB
±1.0 dB
0 to 55 °C
Preselector centered for
frequency >3.0 GHz
10 dB attenuation
3.0 GHz to 6.7 GHz
Absolute
20 to 30 °C
0 to 55 °C
±1.5 dB
±2.5 dB
Relative
20 to 30 °C
±1.3 dB
±1.5 dB
0 to 55 °C
6.7 GHz to 13.2 GHz
Absolute
20 to 30 °C
0 to 55 °C
±2.0 dB
±3.0 dB
50
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Relative
20 to 30 °C
±1.8 dB
±2.0 dB
0 to 55 °C
13.2 GHz to 26.5 GHz
Absolute
20 to 30 °C
0 to 55 °C
Relative
±2.0 dB
±3.0 dB
20 to 30 °C
0 to 55 °C
±1.8 dB
±2.0 dB
a. Absolute frequency response values are referenced to the amplitude at 50 MHz.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
In p u t At ten u a t ion
Sw it ch in g Un cer t a in t y a t
50 MH z
Attenuator Setting
0 dB to 5 dB
10 dB
±0.3 dB
Reference
15 dB
±0.3 dB
20 to 65 dB attenuation
±(0.1 dB + 0.01 × Attenuator
Setting)
Chapter 2
51
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Att en u a t ion Accu r a cy R ela tive t o t h e 10 d B At ten u a tor Set t in g, Ch a r a ct er istic
F r eq u en cy R a n ge
At ten u a tion
0 dB
d c–3 GH z
3.0–13.2 GH z 13.2–19 GH z 19–22 GH z 22–26.5 GH z
±0.3 dB
±0.3 dB
Reference
±0.4 dB
±0.4 dB
±0.5 dB
±0.5 dB
±0.6 dB
±0.6 dB
±0.7 dB
±0.7 dB
±0.9 dB
±0.9 dB
±1.0 dB
±0.5 dB
±0.5 dB
Reference
±0.5 dB
±0.5 dB
±0.6 dB
±0.6 dB
±0.7 dB
±0.7 dB
±1.0 dB
±1.0 dB
±1.1 dB
±1.1 dB
±1.6 dB
±0.8 dB
±0.8 dB
Reference
±0.8 dB
±0.8 dB
±0.8 dB
±0.8 dB
±1.0 dB
±1.0 dB
±1.3 dB
±1.3 dB
±1.6 dB
±1.6 dB
±2.0 dB
±0.9 dB
±0.9 dB
Reference
±1.0 dB
±1.0 dB
±1.2 dB
±1.2 dB
±1.8 dB
±1.8 dB
±2.2 dB
±2.2 dB
±2.7 dB
±2.7 dB
±3.2 dB
±1.0 dB
±1.0 dB
Reference
±1.5 dB
±1.5 dB
±2.0 dB
±2.0 dB
±3.0 dB
±3.0 dB
±3.4 dB
±3.4 dB
±3.5 dB
±3.5 dB
±3.8 dB
5 dB
10 dB
15 dB
20 dB
25 dB
30 dB
35 dB
40 dB
45 dB
50 dB
55 dB
60 dB
65 dB
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ab solu t e Am p lit u d e
Accu r a cy
a
±0.4 dB
At reference settings
b
Overall Amplitude Accuracy
20 to 30 °C
± (0.6 dB + Absolute
Frequency Response)
a. Settings are: reference level –20 dBm; input attenuation 10 dB; center frequency 50
MHz; RBW 1 kHz; VBW 1 kHz; amplitude scale linear or log; span 2 kHz; sweep time
coupled, signal at reference level.
b. For reference level 0 to –50 dBm; input attenuation 10 dB; RBW 1 kHz; VBW 1 kHz;
amplitude scale log, log range 0 to –50 dB from reference level; sweep time coupled; sig-
nal input 0 to –50 dBm; span ≤20 kHz.
52
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a tion
R F In p u t VSWR (at tuned
frequency)
Attenuator setting 0 dB
9 kHz to 26.5 GHz
≤3.0:1,
characteristic
Attenuator setting 5 dB
9 kHz to 100 kHz
≤2.0:1,
characteristic
100 kHz to 6.7 GHz
6.7 GHz to 13.2 GHz
13.2 GHz to 22.0 GHz
22.0 GHz to 26.5 GHz
≤1.4:1,
characteristic
≤1.7:1,
characteristic
≤2.3:1,
characteristic
≤2.6:1,
characteristic
Attenuator setting 10 to 65 dB
9 kHz to 6.7 GHz
≤1.3:1,
characteristic
6.7 GHz to 13.2 GHz
13.2 GHz to 22.0 GHz
22.0 GHz to 26.5 GHz
≤1.5:1,
characteristic
≤2.0:1,
characteristic
≤2.2:1,
characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
a
Au t o Align m en t
Sweep-to-sweep variation
±0.1 dB, characteristic
a. Set Auto Align to Off and use Align Now, All to eliminate this variation.
Chapter 2
53
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
R esolu t ion Ba n d w id t h
Sw it ch in g Un cer ta in ty
(at Reference Level)
1 kHz RBW
Reference
±0.3 dB
±0.6 dB
±0.3 dB
3 kHz to 3 MHz RBW
5 MHz RBW
100 Hz to 300 Hz RBW
(Option 1DR)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
R efer en ce Level
Range
–149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale
±0.1 dB
Linear Scale
±0.12% of Reference Level
Accuracy (at a fixed frequency,
a fixed attenuator, and
referenced to −30 dBm)
Reference Level (dBm) – input
attenuator setting (dB)
–10 dBm to > –60 dBm
–60 dBm to > –85 dBm
–85 dBm to –90 dBm
±0.3 dB
±0.5 dB
±0.7 dB
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Disp la y Sca le Sw it ch in g
Un cer ta in t y
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching
No error
54
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y Sca le F id elit y
Log Maximum Cumulative
RBW ≥ 1 kHz
0 to 85 dB Below
Reference
±(0.3 dB + 0.01 × dB from
reference level)
Level
RBW ≤ 300 Hz (Option 1DR)
Span > 0 Hz
Auto range On
a
±(0.3 dB + 0.01 × dB from
reference level)
0 to 98 dB below
reference level
> 98 to 120 dB below
reference level
±2.0 dB, characteristic
b
Auto range Off
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Span = 0 Hz
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Log Incremental Accuracy
a,c
±0.4 dB/4 dB
0 to 80 dB below reference
level
Linear Accuracy
±2% of Reference Level
a. 0 to 30 dB for RBW = 200 Hz
b. The SCPI command for auto range off is:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF)
c. 0 to 50 dB for RBWs ≤ 300 Hz and span = 0 Hz, or when auto ranging is off.
Chapter 2
55
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sp u r iou s R esp on ses
Second Harmonic Distortion
Input Signal
10 MHz to 500 MHz
< −60 dBc for −30 dBm signal
at input mixer
+30 dBm SHI (second
harmonic intercept)
a
500 MHz to 1.5 GHz
1.5 GHz to 2.0 GHz
2.0 GHz to 3.35 GHz
< −70 dBc for −30 dBm signal
at input mixer
+40 dBm SHI
+70 dBm SHI
+85 dBm SHI
< −80 dBc for −10 dBm signal
at input mixer
b
< −95 dBc for −10 dBm signal
at input mixer
3.35 GHz to 6.6 GHz
6.6 GHz to 13.25 GHz
+85 dBm SHI
+85 dBm SHI
< −95 dBc for −10 dBm signal
at input mixer
< −95 dBc for −10 dBm signal
at input mixer
Third Order Intermodulation
Distortion
10 MHz to 100 MHz
+5 dBm TOI (third order
intercept), characteristic
100 MHz to 3 GHz
< −75 dBc for two −30 dBm
signals at input mixer and
>50 kHz separation
+7.5 dBm TOI
+7.5 dBm TOI
+5.0 dBm TOI
+5.0 dBm TOI
3.0 GHz to 6.7 GHz
6.7 GHz to 13.2 GHz
13.2 GHz to 26.5 GHz
< −75 dBc for two −30 dBm
signals at input mixer and
>50 kHz separation
< −70 dBc for two −30 dBm
signals at input mixer and
>50 kHz separation
< −70 dBc for two −30 dBm
signals at input mixer and
>50 kHz separation
Other Input Related Spurious
Inband Responses
>30 kHz offset
< −65 dBc for −20 dBm signal
at input mixer
56
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Out-of-band Responses
< −80 dBc for –10 dBm signal
at input mixer
a. Mixer power level (dBm) = input power (dBm – input attentuation (dB)
b. or signal below displayed average noise level.
Sp ecified Dyn a m ic R a n ge
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R esid u a l R esp on ses (Input
terminated and 0 dB
attenuation)
150 kHz to 6.7 GHz
< −90 dBm
Chapter 2
57
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Options
Op tion s
Tr a ck in g Gen er a tor (Op tion 1DN)
The spectrum analyzer/tracking generator combination will meet its
specification after a cable (8120-5148) and adapter are connected
between RF OUT and INPUT and Align Now, TG has been run.
Sp ecifica t ion s
5 minutes
Su p p lem en ta l In for m a t ion
Su p p lem en ta l In for m a t ion
Su p p lem en ta l In for m a t ion
Wa r m -u p
Sp ecifica t ion s
9 kHz to 3.0 GHz
Ou t p u t F r eq u en cy R a n ge
Min im u m R esolu t ion BW
Sp ecifica t ion s
1 kHz
Not usable with resolution
bandwidths ≤300 Hz
(Option 1DR)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t P ow er Level
Range
–2 to −66 dBm
0.1 dB
Resolution
Absolute Accuracy (at 50 MHz
with coupled source
attenuator, referenced to
–20 dBm)
± 0.75 dB
Vernier
Range
8 dB
Accuracy (with coupled
source attenuator, 50 MHz,
–20 dBm)
Incremental
Cumulative
±0.2 dB/dB
±0.5 dB, total
Output Attenuator Range
0 to 56 dB in 8 dB steps
58
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Options
Sp ecifica tion s
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma xim u m Sa fe R ever se
Level
+30 dBm (1 W), 50 Vdc,
characteristic
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Su p p lem en t a l In for m a t ion
Ou t p u t P ow er Sw eep
Range
(−10 to −2 dBm) − (Source
Attenuator Setting)
Resolution
0.1 dB
Accuracy (zero span)
<1 dB peak-to-peak
Sp ecifica tion s
Ou t p u t F la t n ess
Referenced to 50 MHz,
–20 dBm
9 kHz to 10 MHz
10 MHz to 3 GHz
±3 dB
±2 dB
Sp ecifica tion s
Sp u r iou s Ou t p u t s
(–2 dBm output)
Harmonic Spurs
TG Output 9 kHz to 20 kHz
≤ −15 dBc
TG Output 20 kHz to 3 GHz ≤ −25 dBc
Non-harmonic Spurs
TG Output 9 kHz to 2 GHz
TG Output 2 GHz to 3 GHz
LO Feedthrough
≤ −27 dBc
≤ −23 dBc
LO Frequency
3.921409 GHz to
6.9214 GHz
≤ −16 dBm
Chapter 2
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Options
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Dyn a m ic Ra n ge
Maximum Output Power Level
– Displayed Average Noise
Level
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t Tr a ck in g
Drift
1.5 kHz/5 minute,
characteristic
Swept Tracking Error
Usable in 1 kHz RBW after 5
minutes of warm-up
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
R F P ow er -Off R esid u a ls
9 kHz to 3 GHz
< −120 dBm, characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t At t en u a t or
R ep ea ta b ilit y
9 kHz to 300 MHz
300 MHz to 2 GHz
2 GHz to 3 GHz
±0.1 dB, characteristic
±0.2 dB, characteristic
±0.3 dB, characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t VSWR
0 dB attenuation
≥ 8 dB attenuation
<2.0:1, characteristic
<1.5:1, characteristic
60
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Options
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t At ten u a tor
Accu r a cy
0 dB
±0.5 dB, characteristic
±0.5 dB, characteristic
8 dB
16 dB
24 dB
32 dB
40 dB
48 dB
56 dB
Reference
±0.5 dB, characteristic
±0.6 dB, characteristic
±0.8 dB, characteristic
±1.0 dB, characteristic
±1.1 dB, characteristic
Tr a ck in g Gen er a t or Ou t p u t Accu r a cy
Relative Accuracy (Referred to –20 dBm) =
Output Attenuator Accuracy + Vernier Accuracy + Output Flatness
Absolute Accuracy =
Relative Accuracy (Referred to –20 dBm) + Absolute Accuracy at 50 MHz
Chapter 2
61
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
General
Gen er a l
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Tem p er a tu r e R a n ge
Operating
Storage
0 to 55 °C
Floppy disk 10 to 40 °C
–40 to 75 °C
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Au d ib le Noise
(ISO 7779)
Sound Pressure at 25 °C
<40 dBa, (<4.6 Bels power)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Milit a r y Sp ecifica tion
Has been type tested to the
environmental specifications
of MIL-PRF-28800F class 3.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
EMI Com p a t ib ilit y
Conducted and radiated
emission is in compliance with
CISPR Pub. 11/1990 Group 1
Class A.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Im m u n it y Test in g
Radiated Immunity
Testing was done at 3 V/m
according to IEC 801-3/1984.
When the analyzer tuned
frequency is identical to the
immunity test signal
frequency, there may be
signals of up to −60 dBm
displayed on the screen.
Electrostatic Discharge
Air discharges of up to 8 kV
were applied according to IEC
801-2/1991. Discharges to
center pins of any of the
connectors may cause damage
to the associated circuitry.
62
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
General
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
P ow er Req u ir em en t s
ac Operation
Voltage, frequency
90 to 132 Vrms, 47 to 440 Hz
195 to 250 Vrms, 47 to 66 Hz
Power Consumption, On
<300 W
<5 W
Power Consumption,
Standby
dc Operation
Voltage
12 to 20 Vdc
<200 W
Power Consumption
Power Consumption,
Standby
<100 mW
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Mea su r em en t Sp eed
Local Measurement and
a
≥ 28/s, characteristic
≥ 28/s, characteristic
Display Update rate
Remote Measurement and
b,c
GPIB Transfer Rate
(Option A4H)
RF Center Frequency Tune,
Measure, and GPIB Transfer
≤ 90 ms, characteristic
b,d
Time (Option A4H)
a. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, spans >10 MHz and
≤600 MHz, and stop frequency ≤3 GHz.
b. Display Off (:DISPlay:ENABle OFF), and 32-bit integer data format (:FORMat:DATA
INT,32), if Option A4J is installed, disable sweep ramp, (:SYS-
Tem:PORTs:IFVSweep:ENABle OFF), markers Off, single sweep, measured with IBM
compatible PC with 550 MHz Pentium III running Windows NT 4.0, one meter GPIB
cable, National Instruments PCI-GPIB card and NI-488.2 DLL
c. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, and span = 20 MHz,
stop frequency ≤3 GHz, average of 100 measurements.
d. Factory preset, auto align Off, RBW = 1 MHz, span = 20 MHz, stop frequency ≤3 GHz,
center frequency tune step size = 50 MHz.
Chapter 2
63
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
General
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Da t a St or a ge
a
8.0 MB, nominal
Internal
External
(10 to 40 °C)
3.5" 1.44 MB, MS-DOS
compatible floppy disk
a. For serial numbers prior to US41440000 or MY41440000, 1 MB without Option B72, 8
MB with Option B72.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Mem or y Usa ge
a
State
20 kB , nominal
State plus 401-point trace
Applications memory usage
21 kB , nominal
8590 Compatibility
0.7 MB, nominal
(Option 290)
a. The size of state will increase depending on installed applications.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Dem od Tu n e a n d List en
Internal speaker, front-panel
earphone jack and front-panel
volume control.
Demod
AM
An uncalibrated demodulated
signal is available on the AUX
VIDEO OUT connector at the
rear panel.
(Option A4J )
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Weigh t (without options)
Net
17.1 kg (37.7 lb),
characteristic
Shipping
31.9 kg (70.3 lb),
characteristic
64
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
General
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
a
Disp la y
Resolution
640 × 480
a. The LCD display is manufactured using high precision technology. However, there may
be up to six bright points (white, blue, red or green in color) that constantly appear on the
LCD screen. These points are normal in the manufacturing process and do not affect the
measurement integrity of the product in any way.
Dim en sion s
Chapter 2
65
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Inputs and Outputs
In p u ts a n d Ou t p u t s
Fr on t Pa n el
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
INP UT 50 Ω
Connector
Type-N female
APC 3.5 male
(Option BAB)
Impedance
50 Ω, nominal
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
R F OUT 50 Ω, (Option 1DN)
Connector
Type-N female
Impedance
50 Ω, nominal
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
a
Amplitude Reference
AMP TD RE F OUT
Connector
BNC female
Impedance
50 Ω, nominal
Frequency
50 MHz
b
Frequency Accuracy
Frequency reference error
c
−20 dBm, nominal
50 Ω Amplitude
a. Turn the amplitude reference on/off by pressing the keys: Input/Output, Amptd Ref Out.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
c. The internal amplitude reference actual power is stored internally.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
P R OBE P OWE R
Voltage/Current
+15 Vdc, ±7% at 150 mA max.,
characteristic
−12.6 Vdc ±10% at 150 mA
max., characteristic
66
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
a
Used for entering screen titles
and filenames only. Interface
compatible with most
IBM-compatible PC
E XT KE YBOAR D
keyboards.
Connector
6-pin mini-DIN
a. The feature is not implemented in firmware revisions prior to A.04.00.
Sp ecifica tion s
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Sp ea k er
Front panel knob controls
volume
Su p p lem en t a l In for m a t ion
H ea d p h on e
Connector
Front panel knob controls
volume
3.5 mm (1/8 inch) miniature
audio jack
Power Output
0.2 W into 4 Ω, characteristic
Rea r Pa n el
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
10 MH z R E F OUT
Connector
BNC female
Impedance
50 Ω, nominal
Output Amplitude
>0 dBm, characteristic
Chapter 2
67
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
10 MH z R EF IN
Connector
BNC female
Note: Analyzer noise
sidebands and spurious
response performance may be
affected by the quality of the
external reference used.
Impedance
50 Ω, nominal
Input Amplitude Range
Frequency
−15 to +10 dBm, characteristic
10 MHz, nominal
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
GATE TR IG/E XT TR IG IN
Connector
BNC female
External Trigger Input
Trigger Level
Selectable positive or negative
edge initiates sweep in EXT
TRIG mode (5 V TTL)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
GATE/H I SWP OUT
Connector
BNC female
High Sweep Output
Level
a
High = sweep ; Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
VGA OUTP UT
Connector
VGA compatible, 15-pin mini
D-SUB
Format
VGA (31.5 kHz horizontal,
60 Hz vertical sync rates,
non-interlaced) Analog RGB
Resolution
640 × 480
68
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
AUX IF OUT
(Option A4J )
RBW ≥ 1 kHz
Connector
Frequency
BNC female
21.4 MHz, nominal
Amplitude (for signal at
reference level and for
reference levels – input
attenuation of –10 to
–70 dBm)
−10 dBm (uncorrected),
characteristic
Impedance
50 Ω, nominal
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
AUX VIDEO OUT
(Option A4J )
RBW ≥ 1 kHz
Connector
BNC female
Amplitude Range (into
>10 kΩ)
0 to 1 V (uncorrected),
characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
H I SWP IN
(Option A4J )
Connector
Input
BNC female
Open collector, low resets and
holds the sweep (5 V TTL)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
H I SWP OUT
(Option A4J )
Connector
Output
BNC female
a
High = sweep , Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Chapter 2
69
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
SWP OUT
(Option A4J )
Connector
Amplitude
BNC female
0 to +10 V ramp,
characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
P R E SEL TUNE OUTP UT
Connector
BNC female
Load Impedance (dc coupled)
Range
> 10 kΩ, nominal
0 to +10 V, characteristic
Sensitivity
0.33 V/GHz of tuned frequency
> 3 GHz, characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
GP IB In t er fa ce
(Option A4H)
Connector
IEEE-488 bus connector
GPIB Codes
SH1, AH1, T6, SR1, RL1, PP0,
DC1, C1, C2, C3 and C28
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ser ia l In ter fa ce
(Option 1AX)
Connector
9-pin D-SUB male
RS-232
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
P a r a llel In t er fa ce
Printer port only
(Option A4H or 1AX)
Connector
25-pin D-SUB female
70
Chapter2
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4408B Specifications and Characteristics
Regulatory Information
R egu la t or y In for m a tion
CAUTION
NOTE
This product is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 664 respectively.
This product has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronic Measuring
Apparatus, and has been supplied in a safe condition. The instruction
documentation contains information and warnings which must be
followed by the user to ensure safe operation and to maintain the
product in a safe condition.
The CE mark is a registered trademark of the European Community (if
accompanied by a year, it is the year when the design was proven).
The CSA mark is the Canadian Standards Association safety mark.
ISM 1-A
This is a symbol of an Industrial Scientific and Medical Group 1 Class A
product. (CISPR 11, Clause 4)
Chapter 2
71
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
About This Chapter
Ab ou t Th is Ch a p ter
This chapter contains specifications and characteristics for the Agilent
E4411B spectrum analyzer. The distinction between specifications and
characteristics is described as follows.
•
•
•
Specifications describe the performance of parameters covered by
the product warranty. (The temperature range is 0 °C to 55 °C,
unless otherwise noted.)
Characteristics describe product performance that is useful in the
application of the product, but is not covered by the product
warranty.
Typical performance describes additional product performance
information that is not covered by the product warranty. It is
performance beyond specification that 80% of the units exhibit with
a 95% confidence level over the temperature range 20 to 30 °C.
Typical performance does not include measurement uncertainty.
•
Nominal values indicate the expected performance, or describe
product performance that is useful in the application of the product,
but is not covered by the product warranty.
The following conditions must be met for the analyzer to meet its
specifications.
o The analyzer is within the one year calibration cycle.
o If Auto Align All is selected:
— After 2 hours of storage within the operating temperature range.
— 5 minutes after the analyzer is turned on with sweep times less
than 4 seconds.
o If Auto Align Off is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes and
Align Now All has been run.
— When Align Now All is run:
•
•
•
Every hour
If the ambient temperature changes more than 3 °C
If the 10 MHz reference changes
74
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
About This Chapter
o If Auto Align All but RF is selected:
— When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
— After the analyzer is turned on for a minimum of 90 minutes and
Align Now RF has been run.
— When Align Now RF is run:
•
•
Every hour
If the ambient temperature changes more than 3 °C
Chapter 3
75
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Frequency
F r eq u en cy
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
F r eq u en cy R a n ge
50 Ω
9 kHz to 1.5 GHz
1 MHz to 1.5 GHz
75 Ω (Option 1DP)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
F r eq u en cy R efer en ce
–6
–7
Aging Rate
±2 × 10 /year
±1.0 × 10 /day, characteristic
–7
Settability
±5 × 10
–6
Temperature Stability
±5 × 10
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
F r eq u en cy R ea d ou t
Accu r a cy
(Start, Stop, Center, Marker)
±((frequency indication ×
a
frequency reference error )
+ 0.75% of span
+ 15% of RBW
+ 10 Hz)
a. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ma r k er F r eq u en cy
Cou n ter
Resolution
Selectable from 1 Hz to
100 kHz
a
±(marker frequency ×
Accuracy
b
frequency reference error
counter resolution)
+
a. Marker level to displayed noise level > 25 dB, RBW/ Span ≥ 0.002, frequency offset = 0
Hz.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
76
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Frequency
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
F r eq u en cy Sp a n
Range
0 Hz (zero span), 100 Hz to
1.5 GHz
Resolution
Accuracy
2 Hz
±1.0% of span
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Sw eep Tim e
a
Range
4 ms to 4000 s
Tracking Generator On
(Option 1DN or 1DQ)
50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
±1%
4 ms to 4000 s
b
c
Sweep Trigger
Free Run, Single, Line, Video ,
d
External, Delayed, Offset
,e
Delayed Trigger
Range
1 µs to 400 s
Resolution
delay in seconds
---------------------------------------
65000
rounded up to nearest µs
Accuracy
±(500 ns + (0.01% of delay))
Offset Trigger
Resolution
sweep time
--------------------------
400
Range
±327 ms to ±323 ks
Where ST = sweep time
–32766 × ST 32365 × ST
------------------------------- ---------------------------
to
400
400
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is suspended in video, external, and delayed trigger modes while waiting for a
trigger event to occur.
c. Unavailable when RBW ≤ 300 Hz (Option 1DR).
d. For firmware revision A.04.00 or later.
e. Delayed trigger is available with line and external trigger.
Chapter 3
77
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Frequency
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Su p p lem en ta l In for m a t ion
Sw eep (t r a ce) P oin t s
401
Sp ecifica t ion s
R esolu t ion Ba n d w id t h
(R BW)
Range
–3 dB bandwidth
1 kHz to 3 MHz, in 1-3-10
sequence, 5 MHz
a
(Option 1DR)
Adds 100, 300 Hz
9 kHz and 120 kHz
–6 dB bandwidth (EMI)
(Option 1DR)
Add 200 Hz
Accuracy
100 Hz to 300 Hz (–3 dB)
RBW
±10%
±15%
(Option 1DR)
1 kHz to 3 MHz (–3 dB)
RBW
5 MHz (–3 dB) RBW
±30%
±20%
9 kHz, 120 kHz (–6 dB)
RBW
(EMI)
200 Hz (–6 dB) RBW (EMI)
(Option 1DR)
±10%
Shape
100 Hz to 300 Hz RBW
(Option 1DR)
Digital, approximately
Gaussian shape
1 kHz to 5 MHz RBW
Synchronously tuned four
poles, approximately
Gaussian shape
Selectivity (60 dB/3 dB
bandwidth ratio)
100 Hz to 300 Hz RBW
(Option 1DR)
<5:1, nominal
<15:1, nominal
1 kHz to 5 MHz RBW
a. Only available in spans ≤5 MHz, sweep times ≥ 4 ms, and not usable with tracking gener-
ator on (Option 1DN, or Option 1DQ).
78
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Frequency
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Vid eo Ba n d w id t h (VBW)
(–3 d B)
Range
30 Hz to 1 MHz in 1-3-10
sequence
3 MHz, characteristic
(Option 1DR)
Adds 1, 3, 10 Hz for RBW’s
<1 kHz
Accuracy
Shape
±30%, characteristic
Post detection, single pole low-
pass filter used to average
displayed noise
Video bandwidths below 30 Hz
are digital bandwidths with
anti-aliasing filtering.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
St a b ility
Noise Sidebands
(Offset from CW signal with
1 kHz RBW, 30 Hz VBW and
sample detector)
≥1 kHz (Option 1DR)
≥10 kHz
≤ −79 dBc/Hz, typical
≤ −95 dBc/Hz, typical
≤ −102 dBc/Hz, typical
≤ −106 dBc/Hz, typical
≤ −116 dBc/Hz, typical
≤ −93 dBc/Hz
≤ −100 dBc/Hz
≤ −104 dBc/Hz
≤ −113 dBc/Hz
≥20 kHz
≥30 kHz
≥100 kHz
Residual FM
1 kHz RBW, 1 kHz VBW
≤150 Hz p–p in 100 ms
100 Hz RBW, 100 Hz VBW
(Option 1DR)
≤30 Hz p–p in 20 ms,
characteristic
System-Related Sidebands,
offset from CW signal
≥30 kHz
≤ −65 dBc
Chapter 3
79
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Frequency
Noise Sid eb a n d s Nor m a lized t o 1 H z Ver su s Offset fr om Ca r r ier
80
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Am p litu d e
Amplitude specifications do not apply for the negative peak detector
mode.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Mea su r em en t R a n ge
Displayed Average Noise
Level to Maximum Safe Input
Level
Input Attenuator Range
0 to 60 dB, in 5 dB steps
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma xim u m Sa fe In p u t Level
Input attenuator setting
≥15 dB
Signals > +33 dBm (2 W) or
+79 dBmV (1 W) (Option 1DP)
nominal may trigger input
protection, which disconnects
the input path.
Average Continuous Power
or Peak Pulse Power
50 Ω
+30 dBm (1 W)
+75 dBmV (0.4 W)
100 Vdc
75 Ω (Option 1DP)
dc
dc transients may
momentarily trigger input
protection.
Input attenuator setting
<15 dB
Signals > +6 dBm (4 mW) or
+61 dBmV (15 mW)
(Option 1DP) nominal may
trigger input protection,
which automatically increases
input attenuation to 15 dB.
Average Continuous Power
or Peak Pulse Power
50 Ω
+3 dBm (2 mW)
+59 dBmV (10 mW)
100 Vdc
75 Ω (Option 1DP)
dc
dc transients may trigger
input protection.
Chapter 3
81
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
1 d B Ga in Com p r ession
a,b
Total power at input mixer
50 MHz to 1.5 GHz
50 Ω
0 dBm
75 Ω (Option 1DP)
+46.75 dBmV
a. Mixer power level (dBm) = input power (dBm) – input attenuation (dB).
b. For resolution bandwidths 1 kHz to 30 kHz, the maximum input signal amplitude must
be ≤ reference level +10 dB. (Option 1DP: For resolution bandwidths 1 kHz to 30 kHz, the
maximum input signal amplitude must be ≤ reference level +5 dB.)
Sp ecifica t ion s
Su p p lem en t a l In for m a t ion
Disp la yed Aver a ge Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = –70 dBm)
(75 Ω: Reference Level =
–21.24 dBmV)
50 Ω
1 kHz RBW
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
400 kHz to 10 MHz
10 MHz to 500 MHz
500 MHz to 1.0 GHz
1.0 GHz to 1.5 GHz
≤ −115 dBm
≤ −119 dBm
≤ −117 dBm
≤ −113 dBm
≤ −123 dBm
≤ −127 dBm
≤ −125 dBm
≤ −121 dBm
≤ −119 dBm
≤ −121 dBm
≤ −121 dBm
≤ −118 dBm
≤ −129 dBm
≤ −131 dBm
≤ −130 dBm
≤ −128 dBm
75 Ω, (Option 1DP)
1 MHz to 10 MHz
10 MHz to 500 MHz
500 MHz to 1.0 GHz
1.0 GHz to 1.5 GHz
≤ −63 dBmV
≤ −65 dBmV
≤ −60 dBmV
≤ −53 dBmV
≤ −71 dBmV
≤ −73 dBmV
≤ −68 dBmV
≤ −61 dBmV
≤ −69 dBmV
≤ −70 dBmV
≤ −66 dBmV
≤ −62 dBmV
≤ −79 dBmV
≤ −80 dBmV
≤ −75 dBmV
≤ −71 dBmV
82
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y R a n ge
Log Scale
Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW ≥ 1 kHz
Calibrated 0 to −85 dB from
Reference Level
a
RBW ≤ 300 Hz (Option 1DR)
Calibrated 0 to −120 dB from
Reference Level
Linear Scale
Scale Units
Ten divisions
dBm, dBmV, dBµV, dBµA, A,
V, and W
a. 0 to –70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma r k er R ea d ou t
R esolu tion
Log scale
RBW ≥ 1 kHz
0 to −85 dB from ref level
RBW ≤ 300 Hz (Option 1DR)
0.04 dB
0 to −120 dB from ref level 0.04 dB
Linear scale 0.01% of Reference Level
Chapter 3
83
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
a
F r eq u en cy R esp on se
50 Ω
9 kHz to 1.5 GHz
10 dB attenuation
20 to 30 °C
±0.5 dB
±1.0 dB
0 to 55 °C
0 dB, 5 dB, 15 to 60 dB
attenuation
±1.0 dB, characteristic
75 Ω (Option 1DP)
1 MHz to 1.5 GHz
10 dB attenuation
20 to 30 °C
±0.5 dB
±1.0 dB
0 to 55 °C
0, 5, 15 to 50 dB attenuation
55 to 60 dB attenuation
1 MHz to 1 GHz
±1.0 dB, characteristic
±1.0 dB, characteristic
±1.25 dB, characteristic
1 GHz to 1.5 GHz
a. Frequency response values are referenced to the amplitude at 50 MHz.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
In p u t At t en u a t ion
Sw it ch in g Un cer ta in ty a t
50 MH z
Attenuator Setting
0 dB to 5 dB
10 dB
±0.3 dB
Reference
15 dB
±0.3 dB
20 to 60 dB attenuation
±(0.1 dB + 0.01 × Attenuator
Setting)
84
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ab solu t e Am p lit u d e
Accu r a cy
a
±0.4 dB
At reference settings
b
Overall Amplitude Accuracy
20 to 30 °C
± (0.6 dB + Absolute
Frequency Response)
a. Settings are: reference level −25 dBm; (75 Ω reference level +28.75 dBmV); input attenu-
ation 10 dB; center frequency 50 MHz; RBW 1 kHz; VBW 1 kHz; amplitude scale linear
or log; span 2 kHz; sweep time coupled, signal at reference level.
b. For reference level 0 to –50 dBm; input attenuation 10 dB; RBW 1 kHz; VBW 1 kHz;
amplitude scale log, log range 0 to –50 dB from reference level; sweep time coupled; sig-
nal input 0 to –50 dBm; span ≤20 kHz.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R F In p u t VSWR (at tuned
frequency)
Attenuator setting
50 Ω
0 to 5 dB attenuation
10 to 60 dB attenuation
75 Ω
≤1.55:1, characteristic
≤1.35:1, characteristic
1 MHz to 1.1 GHz
0 to 5 dB attenuation
10 to 60 dB attenuation
1.1 GHz to 1.5 GHz
0 to 60 dB attenuation
Input protection is tripped
Amptd Ref is On
≤1.55:1, characteristic
≤1.35:1, characteristic
≤2.0:1, characteristic
Open input, characteristic
Open input, characteristic
Auto Align All is selected
Open input momentarily
during retrace, characteristic
Chapter 3
85
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
a
Au t o Align m en t
Sweep-to-sweep variation
±0.1 dB, characteristic
a. Set Auto Align to Off and use Align Now, All to eliminate this variation.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
R esolu t ion Ba n d w id t h
Sw it ch in g Un cer ta in ty
(at Reference Level)
1 kHz RBW
Reference
3 kHz to 3 MHz RBW
5 MHz RBW
±0.3 dB
±0.6 dB
±0.3 dB
100 Hz to 300 Hz RBW
(Option 1DR)
86
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en t a l In for m a tion
R efer en ce Level
Range
–149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale
±0.1 dB
Linear Scale
±0.12% of Reference Level
50 Ω, Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
−35 dBm)
Reference Level (dBm) – input
attenuator setting (dB)
–10 dBm to > –60 dBm
–60 dBm to > –85 dBm
–85 dBm to –90 dBm
±0.3 dB
±0.5 dB
±0.7 dB
75 Ω (Option 1DP), Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
18.75 dBmV)
Reference Level (dBmV) –
input attenuator setting (dB)
38.75 dBmV to
> –11.25 dBmV
±0.3 dB
±0.5 dB
±0.7 dB
–11.25 dBmV to
> –26.25 dBmV
–26.25 dBmV to
–41.25 dBmV
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Disp la y Sca le Sw itch in g
Un cer t a in t y
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching
No error
Chapter 3
87
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Disp la y Sca le F id elit y
Log Maximum Cumulative
RBW ≥ 1 kHz
0 to 85 dB Below
Reference
±(0.3 dB + 0.01 × dB from
reference level)
Level
RBW ≤ 300 Hz (Option 1DR)
Span > 0 Hz
Auto range On
a
±(0.3 dB + 0.01 × dB from
reference level)
0 to 98 dB below
reference level
> 98 to 120 dB below
reference level
±2.0 dB, characteristic
b
Auto range Off
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Span = 0 Hz
±(0.3 dB + 0.015 × dB from
reference level)
0 to 60 dB below
reference level
> 60 to 70 dB below
reference level
±1.5 dB
Log Incremental Accuracy
±0.4 dB/4 dB
0 to 80 dB below reference
level
Linear Accuracy
±2% of Reference Level
a. 0 to 30 dB for RBW = 200 Hz
b. The SCPI command for auto range off is:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF)
c. 0 to 50 dB for RBWs ≤ 300 Hz and span = 0 Hz, or when auto ranging is off.
88
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Sp u r iou s R esp on ses
50 Ω
Second Harmonic Distortion
Input Signal
2 MHz to 750 MHz
< −75 dBc for −40 dBm signal
at input mixer
+35 dBm SHI (second
harmonic intercept)
a
Third Order Intermodulation
Distortion
2 MHz to 10 MHz
+14.5 dBm TOI (third order
intercept), typical
10 MHz to 1.5 GHz
< −75 dBc for two −30 dBm
signals at input mixer and
+7.5 dBm TOI
>50 kHz separation.
Other Input Related Spurious
30 kHz ≤ offset ≤1200 MHz
< −65 dBc for −20 dBm signals
at input mixer ≤1.5 GHz.
Offset >1200 MHz
< −45 dBc for −20 dBm signal
at input mixer ≤1.5 GHz.
Noise Floor Degradation
Input frequency =
1210.7 MHz ± RBW
< −62 dBc for −45 dBm signal
at input mixer
a. Mixer Power Level (dBm) = Input Power (dBm) − Input Attenuation (dB).
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Sp u r iou s R esp on ses
75 Ω, (Option 1DP)
Second Harmonic Distortion
Input signal
2 MHz to 750 MHz
< −75 dBc for +8.75 dBmV
signal at input mixer
a
Chapter 3
89
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Third Order
Intermodulation Distortion
10 MHz to 1.5 GHz
< −75 dBc for two
+18.75 dBmV signals at input
mixer and >50 kHz
separation.
Other Input Related
Spurious
30 kHz ≤ offset
≤1200 MHz
< −65 dBc for +28.75 dBmV
signal at input mixer
≤1.5 GHz.
Offset >1200 MHz
< −45 dBc, for +28.75 dBmV
signal at input mixer
≤1.5 GHz.
Noise Floor Degradation
Input frequency =
1210.7 MHz ± RBW
< −62 dBc, for +3.75 dBmV
signal at input mixer
a. Mixer Power Level (dBmV) = Input Power (dBmV) − Input Attenuation (dB)
90
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Amplitude
Sp ecified Dyn a m ic R a n ge
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R esid u a l R esp on ses (Input
terminated and 0 dB
attenuation)
50 Ω
150 kHz to 1.5 GHz
75 Ω, (Option 1DP)
1 MHz to 1.5 GHz
< −90 dBm
< −36 dBmV
Chapter 3
91
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Options
Op tion s
Tr a ck in g Gen er a tor (Op tion 1DN or 1DQ)
Sp ecifica t ion s
5 minutes
Su p p lem en ta l In for m a t ion
Wa r m -u p
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t F r eq u en cy R a n ge
50 Ω (Option 1DN)
9 kHz to 1.5 GHz
1 MHz to 1.5 GHz
75 Ω (Option 1DQ)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Min im u m R esolu t ion BW
1 kHz
Not usable with resolution
bandwidths ≤300 Hz
(Option 1DR)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t P ow er Level
20 to 30 °C
Range
50 Ω (Option 1DN)
75 Ω (Option 1DQ)
Resolution
0 to −70 dBm
+42.75 to −27.25 dBmV
0.1 dB
Absolute Accuracy (at
50 MHz with coupled source
attenuator)
50 Ω (Option 1DN)
± 0.5 dB
± 1.5 dB
referenced to 0 dBm
75 Ω (Option 1DQ)
referenced to +42.75 dBmV
92
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Options
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Vernier
Range
10 dB
Accuracy (with coupled
source attenuator)
50 Ω (Option 1DN)
±0.75 dB, for 0 to −10 dBm
±0.9 dB, for +42.75 to
referenced to 0 dBm
75 Ω (Option 1DQ)
referenced to 42.75 dBmV +32.75 dBmV
Output Attenuator Range
0 to 60 dB in 10 dB steps
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ma xim u m Sa fe R ever se
Level
a
+20 dBm (0.1 W), 100 Vdc,
characteristic
50 Ω (Option 1DN)
a
+69 dBmV (0.1 W), 100 Vdc,
characteristic
75 Ω (Option 1DQ)
a. dc transients may trigger reverse power protection.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t P ow er Sw eep
20 to 30 °C
Range
50 Ω (Option 1DN)
75 Ω (Option 1DQ)
Resolution
(−15 to 0 dBm) − (Source
Attenuator Setting)
(27.75 to 42.75 dBmV) −
(Source Attenuator Setting)
0.1 dB
Accuracy (zero span)
50 Ω (Option 1DN)
<1.5 dB peak-to-peak
<1.8 dB peak-to-peak
75 Ω (Option 1DQ)
Chapter 3
93
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Options
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t F la t n ess
Referenced to 50 MHz, 0 dB
attenuator
50 Ω (Option 1DN)
9 kHz to 10 MHz
10 MHz to 1.5 GHz
75 Ω (Option 1DQ)
1 MHz to 10 MHz
10 MHz to 1.5 GHz
±2 dB
±1.5 dB
±2.5 dB
±2 dB
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sp u r iou s Ou t p u t s
50 Ω (Option 1DN)
(0 dBm output)
75 Ω (Option 1DQ)
(+42.75 dBmV output)
Harmonic Spurs
9 kHz to 20 MHz
20 MHz to 1.5 GHz
Non-harmonic Spurs
< −20 dBc
< −25 dBc
< −35 dBc
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Dyn a m ic Ra n ge
Maximum Output Power Level
– Displayed Average Noise
Level
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Ou t p u t Tr a ck in g
Drift
No error
Swept Tracking Error
No error for coupled sweep
times
94
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Options
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R F P ow er -Off R esid u a ls
50 Ω (Option 1DN)
100 kHz to 1.5 GHz
< −120 dBm, characteristic
75 Ω (Option 1DQ)
1 MHz to 1.5 GHz
< 65 dBmV, characteristic
Sp ecifica tion s
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t At ten u a tor
R ep ea t a b ilit y
±0.2 dB, characteristic
Su p p lem en t a l In for m a t ion
Ou t p u t VSWR
50 Ω (Option 1DN)
75 Ω (Option 1DQ)
<2.5:1, characteristic
<2.0:1, characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ou t p u t At ten u a tor
Accu r a cy
0 dB
Reference
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
±0.6 dB, characteristic
±0.9 dB, characteristic
±1.2 dB, characteristic
±1.5 dB, characteristic
±1.8 dB, characteristic
±2.1 dB, characteristic
Tr a ck in g Gen er a tor Ou tp u t Accu r a cy 50 Ω (Op t ion 1DN)
Relative Accuracy (Referred to 0 dBm) =
Output Attenuator Accuracy + Vernier Accuracy + Output Flatness
Absolute Accuracy =
Relative Accuracy (Referred to 0 dBm) + Absolute Accuracy at 50 MHz
Chapter 3
95
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Options
Tr a ck in g Gen er a t or Ou t p u t Accu r a cy 75 Ω (Op t ion 1DQ)
Relative Accuracy (Referred to +42.75 dBmV) =
Output Attenuator Accuracy + Vernier Accuracy + Output Flatness
Absolute Accuracy =
Relative Accuracy (Referred to +42.75 dBmV) + Absolute Accuracy at 50 MHz
96
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
General
Gen er a l
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Tem p er a t u r e R a n ge
Operating
Storage
0 to 55 °C
–40 to 75 °C
Floppy disk 10 to 40 °C
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Au d ib le Noise
(ISO 7779)
Sound Pressure at 25 °C
<40 dBa, (<4.6 Bels power)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Milita r y Sp ecifica t ion
Has been type tested to the
environmental specifications
of MIL-PRF-28800F class 3.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
E MI Com p a t ib ilit y
Conducted and radiated
emission is in compliance with
CISPR Pub. 11/1990 Group 1
Class A.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Im m u n it y Test in g
Radiated Immunity
Testing was done at 3 V/m
according to IEC 801-3/1984.
When the analyzer tuned
frequency is identical to the
immunity test signal
frequency, there may be
signals of up to −60 dBm
displayed on the screen.
Electrostatic Discharge
Air discharges of up to 8 kV
were applied according to IEC
801-2/1991. Discharges to
center pins of any of the
connectors may cause damage
to the associated circuitry.
Chapter 3
97
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
General
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
P ow er R eq u ir em en t s
ac Operation
Voltage, frequency
90 to 132 Vrms, 47 to 440 Hz
195 to 250 Vrms, 47 to 66 Hz
Power Consumption, On
<300 W
<5 W
Power Consumption,
Standby
dc Operation
Voltage
12 to 20 Vdc
<200 W
Power Consumption
Power Consumption,
Standby
<100 mW
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Mea su r em en t Sp eed
Local Measurement and
a,b
≥ 35/s, characteristic
≥ 30/s, characteristic
Display Update rate
Remote Measurement and
b,c,d
GPIB Transfer Rate
(Option A4H)
RF Center Frequency Tune,
Measure, and GPIB Transfer
≤ 90 ms, characteristic
Time
(Option A4H)
a. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, and spans >102
MHz and ≤400 MHz.
b. Sweeping through 425.6 MHz or 914.6 MHz will cause the measurement speed to
degrade.
c. Display Off (:DISPlay:ENABle OFF), and 32-bit integer data format (:FORMat:DATA
INT,32), if Option A4J is installed, disable sweep ramp, (:SYS-
Tem:PORTs:IFVSweep:ENABle OFF), markers off, single sweep, measured with IBM
compatible PC with 550 MHz Pentium III running Windows NT 4.0, one meter GPIB
cable, National Instruments PCI-GPIB card and NI-488.2 DLL.
d. Factory preset, auto align Off, RBW = 1 MHz, span= 20 MHz, fixed center frequency,
average of 100 measurements.
e. Factory preset, auto align Off, RBW = 1 MHz, span= 20 MHz, and center frequency tune
step size = 50 MHz.
98
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
General
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Da t a St or a ge
a
8.0 MB, nominal
Internal
External
(10 to 40 °C)
3.5" 1.44 MB, MS-DOS
compatible floppy disk
a. For serial numbers prior to US41440000 or MY41440000, 1 MB without Option B72, 8
MB with Option B72.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Mem or y Usa ge
a
State
20 kB , nominal
State plus 401-point trace
Applications memory usage
21 kB , nominal
8590 Compatibility
0.7 MB, nominal
(Option 290)
a. The size of state will increase depending on installed applications.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Dem od Tu n e a n d List en
Internal speaker, front-panel
earphone jack and front-panel
volume control.
Demod
AM
An uncalibrated demodulated
signal is available on the AUX
VIDEO OUT connector at the
rear panel.
(Option A4J )
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Weigh t (without options)
Net
13.2 kg (29.1 lb),
characteristic
Shipping
25.1 kg (55.4 lb),
characteristic
Chapter 3
99
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
General
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
a
Disp la y
Resolution
640 × 480
a. The LCD display is manufactured using high precision technology. However, there may
be up to six bright points (white, blue, red or green in color) that constantly appear on the
LCD screen. These points are normal in the manufacturing process and do not affect the
measurement integrity of the product in any way.
Dim en sion s
100
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
In p u t s a n d Ou tp u ts
In ter n a l
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Amplitude reference
50 MHz
a
Am p td R ef
Frequency
b
Frequency Accuracy
Frequency reference error
c
50 Ω Amplitude
−25 dBm , nominal
c
75 Ω Amplitude (Option 1DP)
+28.75 dBmV , nominal
a. Turn the amplitude reference signal on/off by pressing the keys: Input/Output,
Amptd Ref.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
c. The internal amplitude reference actual power is stored internally.
Fr on t Pa n el
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
INP UT 50 Ω
Connector
Type-N female
Impedance
50 Ω, nominal
INP UT 75 Ω (Option 1DP)
Connector
BNC female
Impedance
75 Ω, nominal
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
R F OUT 50 Ω, (Option 1DN)
Connector
Type-N female
BNC female
Impedance
50 Ω, nominal
75 Ω, nominal
R F OUT 75 Ω, (Option 1DQ)
Connector
Impedance
Chapter 3
101
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
P R OBE P OWE R
Voltage/Current
+15 Vdc, ±7% at 150 mA max.,
characteristic
−12.6 Vdc ±10% at 150 mA
max., characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
a
Used for entering screen titles
and filenames only. Interface
compatible with most
IBM-compatible PC
EXT KE YBOAR D
keyboards.
Connector
6-pin mini-DIN
a. The feature is not implemented in firmware revisions prior to A.04.00.
Sp ecifica t ion s
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
Sp ea k er
Front panel knob controls
volume
Su p p lem en ta l In for m a t ion
H ea d p h on e
Connector
Front panel knob controls
volume
3.5 mm (1/8 inch) miniature
audio jack
Power Output
0.2 W into 4 Ω, characteristic
Rea r Pa n el
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
10 MH z R EF OUT
Connector
BNC female
Impedance
50 Ω, nominal
Output Amplitude
>0 dBm, characteristic
102
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
10 MH z R E F IN
Connector
BNC female
Note: Analyzer noise
sidebands and spurious
response performance may be
affected by the quality of the
external reference used.
Impedance
50 Ω, nominal
Input Amplitude Range
Frequency
−15 to +10 dBm, characteristic
10 MHz, nominal
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GATE TR IG/EXT TR IG IN
Connector
BNC female
External Trigger Input
Trigger Level
Selectable positive or negative
edge initiates sweep in EXT
TRIG mode (5 V TTL)
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GATE /H I SWP OUT
Connector
BNC female
High Sweep Output
Level
a
High = sweep ; Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
VGA OUTP UT
Connector
VGA compatible, 15-pin mini
D-SUB
Format
VGA (31.5 kHz horizontal,
60 Hz vertical sync rates,
non-interlaced) Analog RGB
Resolution
640 × 480
Chapter 3
103
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
AUX IF OUT
(Option A4J )
RBW ≥ 1 kHz
Connector
Frequency
BNC female
21.4 MHz, nominal
Amplitude (for signal at
reference level and for
reference levels – input
attenuation of –10 to
–70 dBm)
−10 dBm (uncorrected),
characteristic
Impedance
50 Ω, nominal
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
AUX VIDE O OUT
(Option A4J )
RBW ≥ 1 kHz
Connector
BNC female
Amplitude Range (into
>10 kΩ)
0 to 1 V (uncorrected),
characteristic
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
H I SWP IN
(Option A4J )
Connector
Input
BNC female
Open collector, low resets and
holds the sweep (5 V TTL)
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
H I SWP OUT
(Option A4J )
Connector
Output
BNC female
a
High = sweep , Low = retrace
(5 V TTL)
a. High sweep may be high longer than the indicated sweep times.
Sp ecifica t ion s
Su p p lem en ta l In for m a t ion
SWP OUT
(Option A4J )
104
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
Sp ecifica tion s
BNC female
Su p p lem en t a l In for m a t ion
Connector
Amplitude
0 to +10 V ramp,
characteristic
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
GP IB In t er fa ce
(Option A4H)
Connector
IEEE-488 bus connector
GPIB Codes
SH1, AH1, T6, SR1, RL1, PP0,
DC1, C1, C2, C3 and C28
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
Ser ia l In t er fa ce
(Option 1AX)
Connector
9-pin D-SUB male
RS-232
Sp ecifica tion s
Su p p lem en t a l In for m a t ion
P a r a llel In t er fa ce
Printer port only
(Option A4H or 1AX)
Connector
25-pin D-SUB female
Chapter 3
105
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent E4411B Specifications and Characteristics
Regulatory Information
R egu la tor y In for m a t ion
CAUTION
NOTE
This product is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 664 respectively.
This product has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronic Measuring
Apparatus, and has been supplied in a safe condition. The instruction
documentation contains information and warnings which must be
followed by the user to ensure safe operation and to maintain the
product in a safe condition.
The CE mark is a registered trademark of the European Community (if
accompanied by a year, it is the year when the design was proven).
The CSA mark is the Canadian Standards Association safety mark.
ISM 1-A
This is a symbol of an Industrial Scientific and Medical Group 1 Class A
product. (CISPR 11, Clause 4)
106
Chapter3
Download from Www.Somanuals.com. All Manuals Search And Download.
|