ADDENDUM
to the
Agilent 6811B, 6812B, and 6813B
User’s Guide and Programming Guide
for the
Dual Power Analyzer Option 020/022
Agilent P/N 5964-8134
Microfiche No 5964-8135
Printed in U.S.A.
May, 2000
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent Dual Power Analyzer Description
The Agilent Technologies Dual Power Analyzer Option 020 is available on the following ac sources:
Agilent 6811B AC Power Source (375VA)
Agilent 6812B AC Power Source (750VA)
Agilent 6813B AC Power Source (1750VA)
Agilent 6813B Option 019 AC Power Source (2000VA)
Option 020 consists of an additional Power Analyzer input on the back of the ac source, an external
current shunt, and a cable to connect the current shunt to the ac source. Option 020 adds the following
capabilities to the above listed ac source models:
•
•
•
•
•
Voltage and current measurement inputs on the rear panel, with capabilities similar to the output
voltage and current measurements available in standard Agilent ac sources.
The ability to use any shunt for the external current measurement, allowing current measurements
over a very wide range of magnitudes.
Precisely synchronized measurements between the rear panel Power Analyzer input and the ac source
output. This is possible because the Power Analyzer input has its own dedicated measurement buffer.
New measurement commands that may be used in evaluating uninterruptible power supplies (To
measure transfer time, peak voltage, and phase locked loop performance for example).
Frequency modulation of the ac source output.
As shown in the following figure, the Agilent Dual Power Analyzer also includes an Agilent supplied
current shunt, which must be connected to the output of the equipment under test to make accurate
voltage and current measurements. The specifications and supplemental characteristics of the Agilent
Dual Power Analyzer are documented toward the end of this document.
Power
Analyzer Input
(V & I sense)
Agilent supplied cable
Agilent supplied shunt
Agilent AC SOURCE
(side view)
Line
Load
ac in
output
Equipment
Under Test
Shunt
AC Source output
ac line cables
(customer supplied)
Option 020 Typical Connections
Agilent Option 022 Description
Option 022 is simply an Agilent 6813B Option 019 AC Power Source combined with an Agilent Dual
Power Analyzer Option 020. All of the Option 020 information in this document also applies to units that
are configured as Option 022 units.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Menus - Additions
The Power Analyzer input uses the same front panel menu commands that are used to measure the Main
Output of the ac source. These are documented in the ac source User’s Guide. The following menu
commands have been added to let you access the Power Analyzer input, control the output frequency
modulation, and calibrate the Power Analyzer input. In most cases you will need to press the
Function key multiple times to access the menu item. Press the
Entry key to access the parameter.
φ1 115V 60 Hz
φ2 120V 60 Hz
FM OFF
Displays the Main Output measurement when φ1 is selected
Displays the Power Analyzer measurement when φ2 is selected
Enable or disable output frequency modulation (OFF or ON)
Specifies the modulating frequency of the output
Specifies the peak deviation of the output frequency modulation
Specifies the frequency input source (VOLT or CURR)
Begin external current calibration
Phase
Select
Freq
FM:FREQ 50
FM:DEV 10
Input
Shift
FREQ:SRC VOLT
CAL:CURR: EXT
CAL:VOLT: EXT
SHUNT
Cal
Begin external voltage calibration
Enter a value for the external shunt in ohms
NOTE:
With no signal connected to the Power Analyzer input, the φ2 front panel display may
return a very high frequency value. This is normal, since there is no external frequency
for the measurement circuit to detect.
SCPI Programming Commands - Additions
The Power Analyzer input uses the same measurement commands that are used to measure the Main
Output of the ac source. These are documented in the ac source User’s Guide. The following SCPI
commands have been added to let you access the Power Analyzer input, control the output frequency
modulation, and calibrate the Power Analyzer input. Additional measurement functions let you measure
transfer time, peak voltage, and phase locked loop performance when evaluating devices such as
uninterruptible power supplies.
CALibrate
:CURRent
:EXTernal
:VOLTage
:EXTernal
MEASure | FETCh
:VOLTage
:RANKed? <percentile>
:TVOLt? <level><occurrence>
:ABSolute? <level><start_time><min_pulse_width>
:SHUNt
FM
SENSe
:STATe <bool>
:NSELect 1 | 2
:DEViation <value>
:FREQuency <value>
:FREQuency:SOURce VOLT | CURR
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Rear Panel Connections
Power Analyzer input.
Connector plug is
removable.
HP-IB
LO HI
LO HI
RS 232
I
V
SENSE
. . . . .
SENSE
FLT INH
_
+
-
+
OPTIONS
LINE RATING
CAUTION
. . . .
TRIGGER
OUT
1
WARNING
WARNING
SENSE
O
1
COM COM
L
1
L
2(
N
)
_
300 VAC MAX TO
Rear Panel, Overall View
Installation
Turn the unit off before connecting any wires.
The Power Analyzer input to the ac source has both voltage and current sense inputs. Each sense input
has a HI and a LO connection. The center pin of the connector is not used. The maximum isolation
voltage to ground is 300 Vac (±425 Vdc).
Use the 1-meter, Agilent supplied cable and connect the Power Analyzer input to the external shunt. Note
that the cable connectors are keyed so that the cable will only fit one way. Disconnect the cable from the
unit by pulling it straight back. Cable connections are shown in the following figure.
IMPORTANT: You should always connect the V SENSE inputs for the Power Analyzer input to operate
properly, even if you are not making voltage measurements. This is because frequency
measurements are made by sensing the voltage signal, and many of the power analyzer
measurements depend on an accurate frequency measurement. If a voltage signal is not
available, use SENSe:FREQuency:SOURce (or its front panel equivalent) to select and
measure the frequency from the current signal instead.
If the Agilent supplied cable is too short, you may open the cable ends and replace the wiring in the cable
with longer wires. Install the new cable in the same way as the original cable. The cable connectors
accepts wire sizes from AWG 22 to AWG 12. Firmly tighten the screws when making wire connections.
WARNING: SAFETY HAZARD You must replace the cable hoods after completing all connections.
This is because the connector screw terminals will be at line potential during operation.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
LO HI
LO HI
V SENSE
HI LO
I SENSE
HI LO
I
V
SENSE
SENSE
CONNECTOR
(AC SOURCE END)
CONNECTOR
(SHUNT END)
HOODS PLACED
OVER WIRES AND
CONNECTORS
WARNING, SAFETY HAZARD:
Connector hoods must be installed
on connectors. Connector screw
terminals are at line potential
during operation.
RELEASE TAB
CONNECTOR
HALVES SNAP
TOGETHER
WIRES MUST BE INSTALLED IN
STRAIN RELIEF INSIDE CONNECTOR
RELEASE TAB
Measurement Cable Connections
Voltage Sensing at a Location Other than the External Current Shunt
You may also need to open up the cable connector if you need to voltage sense at a point other than at the
external current shunt. You must use your own sense cable to make the voltage sense connections. Please
completely remove the voltage sense wires from the Agilent supplied measurement cable if you are using
your own remote voltage sense cable. Disconnect the voltage sense wires from both cable connectors.
To use your own cable to voltage sense at a point other than the external current shunt, make the
connections at the point where you will be remote sensing. Insert the other end of your cable into the HI
and LO voltage sense terminals of the ac source connector. The cable connectors accepts wire sizes from
AWG 22 to AWG 12. Firmly tighten the screws when making wire connections.
WARNING: SAFETY HAZARD You must replace the cable hoods after completing all connections.
This is because the connector screw terminals will be at line potential during operation.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
External Shunt Line and Load Connections
The Agilent supplied current shunt has both input (line) and output (load) connectors. The customer must
supply the cables to connect the ac lines to and from the external current shunt. Agilent supplies one
hooded cable cover for the input of the ac source and two hooded cable covers for the shunt. All three of
these covers are assembled and installed in the same way. It is the customer’s responsibility to use these
covers when making line cable connections. Note that a different cable cover is provided for the ac
output of the ac source.
The following figure shows the Load and Line connectors on the external current shunt. Note the
location of the line, neutral and ground connectors and connect your cables accordingly. Note that the
connections on the output of the ac source are not in the same order on the barrier block as on the shunt
barrier block.
WARNING: SAFETY HAZARD You must replace the cable hoods after completing all connections.
This is because the screw terminals will be at line potential during operation.
N
L
1. NEUTRAL CONNECTION (BLUE OR WHITE)
2. LINE CONNECTION (BROWN OR BLACK)
3. GROUND CONNECTION (GRN/YEL OR GRN)
4. POWER CORD
1
2
3
5. CONNECTOR NUT
6. RUBBER BOOT
7. POWER SAFETY COVER
6
7
8
8. STRAIN RELIEF CONNECTOR
4
5
Line Connections
7
Download from Www.Somanuals.com. All Manuals Search And Download.
Check the External Current Shunt Setting
After you have completed your cable connections, you can turn on and check out the ac source as
described in the Turn-On Checkout chapter in the ac source User’s Guide. One additional step in
checking out a unit with Option 020/022 installed, is to make sure that the correct shunt value has been
entered in the unit. To check the shunt value setting, proceed as follows:
Action
Display
1.
2.
3.
Turn on the ac source and wait for the unit to finish its internal selftest.
Press Shift Calibration and scroll to the SHUNT command.
SHUNT 0.01
Check to see if the resistance value on the display matches the value written on the
outside of the external current shunt.
If it matches, press Meter to return to meter mode. You are done.
If it does not match, you must enter the correct value. Proceed as follows:
4.
5.
CAL ON 0.0
SHUNT 0.01
Press Shift Calibration, scroll to CAL ON and press Enter. If necessary, enter
the calibration password from Entry keypad and press Enter. If the password is
correct the Cal annunciator will come on.
Press Shift Calibration and scroll to the SHUNT command. Enter the shunt
value written on the outside of the external current shunt exactly as it is written on
the label. Then press Enter.
6.
7.
CAL:SAVE
CAL OFF
Press Shift Calibration, scroll to CAL SAVE, and press Enter.
Press Shift Calibration, select CAL OFF, and press Enter to exit Calibration.
NOTE:
The shunt value that you enter applies only to the Power Analyzer input and has no
effect on ac source output measurements. Therefore, ignore the φ1 and φ2 annunciators
while entering the shunt value.
Front Panel Programming
Setting the Output Frequency Modulation
To set the output frequency modulation:
Action
Display
1.
FM:FREQ 1
On the Function keypad, press Frequency. Then scroll to the FM:FREQ command
to enter a modulating frequency. Use the numeric keypad to enter a modulating value
and press Enter.
2.
To set the peak frequency deviation of the output frequency modulation, press
Frequency. Then scroll to the FM:DEV command to enter a frequency deviation.
Use the numeric keypad to enter a value and press Enter.
FM:DEV 10
The deviation frequency must be less than the programmed output frequency by at
least 0.001 Hz. Otherwise, it will result in Error 613.
3.
FM ON
To enable frequency modulation, press Frequency, scroll to FM OFF, select the
ON parameter and press Enter.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Using the Front Panel Display with the Power Analyzer Input
To select the measurement source:
Action
Display
φ
2 115V 60Hz
Press Meter to return the display to Meter mode. Press Phase Select at any time to
toggle between the Main Output and the Power Analyzer input. The left-most digit of
the front panel display will indicate either "φ1" for the Main Output, or "φ2" for the
Power Analyzer input. If the φ2 front panel display returns a very high frequency
value, it usually means that there is no signal connected to the Power Analyzer input.
Note that the Phase Select key also selects the measurement source to which the
current range setting applies. The current range command is located in the Input menu.
Making Power Analyzer Input Measurements
All measurements are based on acquiring and subsequently processing waveform information. Waveform
information is acquired either from the ac source output or from the Power Analyzer input. When the ac
source is on, it continually takes measurements and updates the front panel meter from whichever
measurement source is active. The Meter key accesses the measurement functions from the front panel.
This menu applies to both the ac source output and the Power Analyzer input.
NOTE:
The Shift Harmonic functions also apply to the Power Analyzer input.
If the display indicates OVLD, the measurement capability of the unit has been
exceeded.
Use the Meter menu for making Power Analyzer input measurements:
Action
Display
φ
2 115V 3.04A
1.
Press Phase Select to toggle to the Power Analyzer input. A φ2 will appear
on the display when the Power Analyzer input is selected.
2.
Press Meter and q repeatedly to access the following measurement functions:
φ
2
♦
♦
♦
♦
♦
♦
♦
♦
♦
♦
rms voltage and frequency (the default)
rms voltage and rms current
rms current and frequency (the default)
rms voltage and power
current crest factor
peak current, repetitive
peak current, non-repetitive
apparent power
<reading>V <reading>Hz
<reading>V <reading>A
<reading>A <reading>Hz
<reading>V <reading>W
<reading> CREST F
<reading>A PK REP
<reading>A PK NR
<reading> VA
φ
2
φ
2
φ
2
φ
2
φ
2
φ
2
φ
2
φ
2
reactive power
power factor
<reading> VAR
<reading> PFACTOR
φ
2
The ac source uses the output frequency value when calculating many of the output measurements. For ac
source output measurements, the ac source uses the programmed output frequency value. For the Power
Analyzer input , the ac source uses the frequency measured at the V SENSE input channel. This is the
default setting. You can also select the I SENSE input channel as the frequency measurement source.
Use the Input menu for selecting the frequency source:
Action
Display
If a voltage signal is not available to measure the output frequency, you must
select the current measurement as the frequency source. Press Input, scroll to
FREQ:SRC VOLT, select the CURR parameter and press Enter.
FREQ:SRC CURR
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Additional Measurement Commands
The following additional measurement commands are provided with the Agilent Dual Power Analyzer
Option 020/022. These measurement commands are optimized for use in specific applications such as
evaluating uninterruptible power supplies. In this application the new commands can be used to measure
transfer time, peak voltage, and phase locked loop performance of the UPS.
NOTE:
All MEASure and FETCh commands apply to both the Main Output and the Power
Analyzer input. This includes commands documented in the ac source User’s Guide and
those described in this section. See SENS:NSEL to select a power analyzer input.
MEASure:TVOLt?
FETCh:TVOLt?
Returns the time at which the measured voltage crosses a user-specified voltage level. The sign and
magnitude of occurrence define the event to be reported. For example, if occurrence is -3, the return
value is the third time at which the voltage crosses the level in the negative-going direction. If occurrence
is 4, the return value is the fourth time at which the voltage crosses the level in the positive-going
direction.
The return value is with respect to the start of the measurement buffer. If no point on the waveform
satisfies the specified conditions, the return value is 9.91000E+37.
MEASure:[SCALar]:TVOLt? <level>, <occurrence>
FETCh:[SCALar]:TVOLt? <level>, <occurrence>
−1E6 to 1E6 (for level)
Query Syntax
Parameters
−4096 to +4096, but not 0 (for occurrence)
MEAS:TVOL? 50, 3 FETC:TVOL? 10, -1
Examples
Returned Parameters
Related Commands
<NR3>
MEAS:TVOL:ABS?
10
Download from Www.Somanuals.com. All Manuals Search And Download.
MEASure:TVOLt:ABSolute?
FETCh:TVOLt:ABSolute?
Returns the time at which the absolute value of the measured voltage first exceeds level, with the
following qualifications: The behavior of the signal before start_ time is ignored, and the absolute value
of the voltage must remain above level for at least the specified min_ pulse_ width.
Both start_ time and the return value are with respect to the start of the measurement buffer. If no point
on the waveform satisfies the specified conditions, the return value is 9.91000E+37.
MEASure:[SCALar]:TVOLt:ABSolute? <level>, <start_time>,
<min_pulse_width>
Query Syntax
FETCh:[SCALar]:TVOLt:ABSolute? <level>, <start_time>,
<min_pulse_width>
0 to 1E6 (for level)
0 to 1E6 (for start time in seconds)
0 to 1E6 (for minimum pulse width in seconds)
MEAS:TVOL:ABS? 50, 0, 0.001
FETC:TVOL:ABS? 100, 0.005, 0
Parameters
Examples
<NR3>
MEAS:TVOL?
Returned Parameters
Related Commands
MEASure:VOLTage:RANKed?
FETCh:VOLTage:RANKed?
Scans through the buffer of 4096 instantaneous voltage measurements, and returns the voltage value that
corresponds to the percentile rank given. Specifying a percentile of 0 returns the lowest value in the
buffer. A percentile of 100 returns the highest value in the buffer. A percentile of 50 returns the median
value of the buffer. Stated another way, if you specify a percentile of 60, 60 percent of the voltage
readings in the buffer are less that the returned value, and 40 percent of the readings in the buffer are
greater than the returned value.
If the voltage waveform consists of rectangular pulses that have narrow overshoots, using percentiles of
approximately 5 and 95 (depending on the width of the pulses and the width of the overshoots) is a good
way to determine the voltages at the flat portions of the pulses.
MEASure:[SCALar]:VOLTage:RANKed? <percentile>
FETCh:[SCALar]:VOLTage:RANKed? <percentile>
0 to 100
Query Syntax
Parameters
Examples
MEAS:VOLT:RANK? 50
<NR3>
Returned Parameters
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Additional Sense Commands
The following additional sense commands are provided with the Agilent Dual Power Analyzer Option
020/022. These commands let you select a measurement source and the frequency measurement channel
for the Power Analyzer input.
SENSe:NSELect
Selects the measurement source that will return data when a query is sent. A parameter value of 1 selects
the Main Output, which measures the actual output voltage and current of the ac source. A value of 2
selects the Power Analyzer input, which measures the voltage and current at the V SENSE and I SENSE
input channels. Note that this command is similar to, but not the same as, the command
INSTrument:NSELect that is used in 3-phase sources.
Only the following commands are affected by the measurement source selection:
•
•
Queries beginning with MEASure or FETCh
SENSe:CURRent:ACDC:RANGe <range>
SENSe:NSELect <NR1>
1 | 2
Command Syntax
Parameters
1
*RST Value
SENS:NSEL 2
Examples
SENSe:NSELect?
<NR1>
Query Syntax
Returned Parameters
SENSe:FREQuency:SOURce
This command selects the frequency measurement source for the Power Analyzer input. It only applies
when the Power Analyzer input is active.
Accurate frequency measurements are critical when making all Harmonic measurement calculations.
Other measurement calculations also use frequency measurement values, but they only slightly affect the
accuracy of the measurement. The ac source normally measures the frequency at the V SENSE input
channel. With this comand you can select the I SENSE input channel as the frequency measurement
source for the Power Analyzer input.
NOTE:
For Main Output measurements, the ac source uses the programmed output frequency
value when making measurement calculations.
SENSe:FREQuency:SOURce <source>
VOLTage | CURRent
Command Syntax
Parameters
VOLTage
*RST Value
SENS:FREQ:SOUR CURR
Examples
SENSe:FREQuency:SOURce?
<CRD>
Query Syntax
Returned Parameters
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Additional Frequency Modulation Commands
The following SCPI commands are used to program the output frequency of the ac source. These
commands apply only to the Main Output of the ac source and do not affect the measurement capability
of the unit. These commands are only provided with the Agilent Dual Power Analyzer Option 020/022.
FM
Turns off or on frequency modulation of the Main Output. Frequency modulation operates only if
FREQuency:MODE is set to FIXed.
[SOURce:]FM[:STATe] <bool>
Command Syntax
Parameters
0 | 1 | OFF | ON
OFF
*RST Value
FM 1
FM:STATE ON
Examples
[SOURce:]FM[:STATe]?
0 | 1
FM:DEV FM:FREQ
Query Syntax
Returned Parameters
Related Commands
FM:DEViation
Sets the peak frequency deviation of the output frequency modulation in Hertz. The frequency deviation
must be less than the programmed output frequency by at least 0.001 Hz. For example, if the output
frequency is set to 60 Hz, and you program a deviation of 10 Hz, the output frequency will modulate
between 50 and 70 Hz at a rate determined by the FM:FREQuency command.
[SOURce:]FM:DEViation <NRf>
0 to 1000 Hz
0
Command Syntax
Parameters
*RST Value
Hz (Hertz)
FM:DEV 10
Unit
Examples
[SOURce:]FM:DEViation?
<NR3>
FM FM:FREQ
Query Syntax
Returned Parameters
Related Commands
FM:FREQuency
Sets the modulating frequency of the output frequency modulation in Hertz.
[SOURce:]FM:FREQuency <NRf>
0.001 to 1000 Hz
0.1
Command Syntax
Parameters
*RST Value
Hz (Hertz)
FM:FREQ 1
Unit
Examples
[SOURce:]FM:FREQuency?
<NR3>
FM FM:DEV
Query Syntax
Returned Parameters
Related Commands
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Additional Calibration Commands
The following SCPI commands are used to calibrate the Power Analyzer input and external current
shunt:
CALibrate:VOLTage:EXTernal
Initiates calibration of the external voltage measurement.
CALibrate:VOLTage:EXTernal
None
CAL:VOLT:EXT
Command Syntax
Parameters
Examples
CALibrate:CURRent:EXTernal
Initiates calibration of the external current measurement.
CALibrate:CURRent:EXTernal
None
CAL:CURR:EXT
Command Syntax
Parameters
Examples
CALibrate:SHUNt
Lets you enter the external current shunt value (in ohms). The external current shunt is used for external
current measurement. The programming range is 1E-6 to 1E6. This parameter is nonvolatile, so there is
no *RST value. As with other CAL commands, calibration must first be enabled with CAL:STATe, and
the value must be made permanent with CAL:SAVE.
CALibrate:SHUNt <value>
Current shunt value in ohms
CAL:SHUN .01
Command Syntax
Parameters
Examples
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Behavior Differences of Existing SCPI Commands
This section documents the behavior differences of existing SCPI commands when the Power Analyzer
input is selected or is active.
MEASure | FETCh
All standard ac source MEASure or FETCh functions are available for the Main Output of the ac source
as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that
will return data.
MEASure:FREQuency?
When the Power Analyzer input is selected as the measurement source, this command returns the actual
frequency measured at the V SENSE input channel. When the Main Output of the ac source is selected as
the measurement source, this command returns the programmed output frequency.
NOTE:
With no signal connected to the Power Analyzer input, this command may return a very
high frequency value. This is normal, since there is no external frequency for the
measurement circuit to detect.
SENSe:CURRent:ACDC:RANGe <value>
Sets the current measurement range of the selected measurement source (see SENSe:NSELect). Normally
when using the Main Output as the measurement source, you would put in a value and the ac source will
automatically select the current range. If you are using the Power Analyzer input with an external current
shunt as the measurement source, the crossover value of the current ranges can be determined with the
following formula: Crossover_value = 0.05713/shunt_resistance. Values greater than the crossover
value select the high current range. Alternatively, you can simply program MIN or MAX to select the
low or high current range.
TRIGger:ACQuire and other measurement trigger commands
SENSe:SWEep:TINTerval and other measurement timebase commands
SENSe:WINDow
These commands always apply to both measurement sources, regardless of which source is selected.
*IDN?
Returns the same response as a standard ac source, except for the firmware revision number.
*OPT?
Returns 20, except that the Agilent 6813B 2000 VA model returns 22.
SYSTem:CONFigure <NORMal | IEC >
All Option 020 capabilities are only available in NORMal mode.
SYSTem:LANGuage <SCPI | E9012>
All Option 020 capabilities are only available in SCPI mode.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming Example
UPS Transfer Time Measurement
The following programming example illustrates how to use the ac source with the Dual Power Analyzer
Option 020/022 to measure transfer time on a UPS. Transfer time is defined as the time it takes a UPS to
go from online-operation to battery-backup operation when the ac line fails.
This example starts with the ac source output voltage (UPS input voltage) at 120 Vac. The UPS output
voltage is connected to a load and to the Power Analyzer input. The ac source output turns on at 120 Vac,
then does a 50 ms dropout to 0 volts. Both the dropout and measurements are triggered at the same time.
The FETC:TVOLT:ABS? command is then used to return the transfer time, here defined as the time
between the start of the dropout and the time when the absolute value of the UPS output voltage exceeds
50 volts for at least 0.001 seconds.
100 ASSIGN @Ps TO 705
110 !
! AC SOURCE IS AT ADDRESS 5, INTERFACE 700
120 !
130 OUTPUT @Ps;"*RST"
140 OUTPUT @Ps;"VOLT 120" ! SET INITIAL INPUT VOLTAGE FOR UPS TO 120 VAC
150 OUTPUT @Ps;"OUTP ON" ! TURNS AC SOURCE OUTPUT ON AT 120 VAC
160 WAIT 2
170 !
180 !
190 OUTPUT @Ps;"VOLT:MODE PULS" ! SET UP PULSE MODE FOR AC SOURCE OUTPUT
200 OUTPUT @Ps;"PULS:WIDT 0.05" ! SET PULSE WIDTH TO 50 MILLISECONDS
210 OUTPUT @Ps;"VOLT:TRIG 0"
220 !
230 !
240 OUTPUT @Ps;"TRIG:SYNC:SOUR PHAS" ! SETS DROPOUT SYNCHRONIZATION TO PHASE
250 OUTPUT @Ps;"TRIG:SYNC:PHAS 90" ! SETS DROPOUT START PHASE TO 90 DEGREES
260 OUTPUT @Ps;"TRIG:SOUR BUS"
! BEGIN FROM A KNOWN STARTING POINT WITH RESET
! MAKE SURE UPS OUTPUT HAS TIME TO SETTLE
! SET PULSED (TRIGGERED) VALUE TO 0 VOLTS
! SETS DROPOUT TRIGGER AND MEASUREMENT
! TRIGGER SOURCE TO BUS TRIGGER
270
280 !
290 !
300 OUTPUT @Ps;"OUTP:TTLT:SOUR BOT" ! SETS SIGNAL TRIGGER SOURCE TO BEGINNING
310
! OF TRANSIENT (BOT)
320 OUTPUT @Ps;"OUTP:TTLT:STAT ON" ! ENABLES TRIGGER OUTPUT SIGNAL
330 OUTPUT @Ps;"TRIG:ACQ:SOUR TTLT" ! SETS MEASUREMENT TRIGGER SOURCE TO
340
350 !
360 !
! TRIGGER OUTPUT SIGNAL
370 OUTPUT @Ps;"INIT:IMM:SEQ1" ! ARMS THE PULSED DROPOUT TO WAIT FOR TRIGGER
380 OUTPUT @Ps;"INIT:IMM:SEQ3" ! ARMS MEASUREMENT PROCESS TO WAIT FOR TRIGGER
390 OUTPUT @Ps;"TRIG:IMM"
400
! TRIGGERS PULSED VOLTAGE DROPOUT TO 0 VOLTS
! AND STARTS THE MEASUREMENT PROCESS
410 OUTPUT @Ps;"SENS:NSEL 2" ! SELECTS POWER ANALYZER INPUT FOR MEASUREMENTS
420 OUTPUT @Ps;"FETC:TVOLT:ABS? 50, 0, .001"
! FETCHES TRANSFER TIME
! RETURNS THE TRANSFER TIME MEASUREMENT
! PRINTS THE TRANSFER TIME MEASUREMENT
430 ENTER @Ps;A
440 PRINT A
450 END
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Using the Power Analyzer Input with the GUI
The Agilent AC Source Graphical User Interface (GUI) is an easy to use soft front panel for the Agilent
6800-series AC Power Source/Analyzers. It is supplied with your ac source. With it, you can also control
the Power Analyzer input on the back of the unit.
Instructions to install and run the Graphical User Interface are provided in the Quick Start Guide that
accompanies the software. Briefly:
1. Place Disk #1 in the A: drive of your computer and run SETUP.EXE.
2. Follow the directions on the screen to install the software.
3. To run the Agilent AC Source GUI, click on its desktop icon:
You can also click on the Start button and select:
Programs | Agilent AC Source | AC Source GUI.
The Agilent AC Source Graphical User Interface software recognizes if your ac source includes Option
020/022, and automatically activates the controls for these options when the software is installed.
Making a Measurement
First, you must select the type of waveform to display and then make the measurement.
ñ
ñ
Click on the Config button and then selecting the Config Waveform Display tab.
Select the type of waveform to display from the Ext Power Analyzer dropdown box. Selections
include voltage, current, voltage+current, or harmonic display functions.
ñ
For the Main Output, select from voltage, current, voltage+current, or harmonic display functions. If
you only want one measurement window to appear, select None in the drop-down box of the
measurement window that you do not want to appear.
ñ
ñ
Click OK to return to the Main window.
Click on the Measure button to make a single measurement.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
If you have both the Main Output and the Power Analyzer input displayed, the Main window should
display both waveforms, as illustrated in the following screen shot.
Making a Voltage Phase Difference Measurement
The voltage phase difference measures the phase difference between the output of the ac source and the
output of a UPS after ac power is restored following an ac line dropout. This measurement requires that
the output of the UPS be connected to the Power Analyzer input. Also, you must have already measured
and displayed both the ac source output and the UPS output waveforms in the Main window.
To access the Voltage Phase Difference measurement, go to the Tests menu and select the Voltage
Phase Difference command.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
In the dialog box that appears on the screen, enter values for the indicated parameters. The test uses these
parameters to make the measurement. For a detailed explanation of the Voltage Phase Difference
parameters refer to the online help by pressing the F1 key.
When you are ready to make the measurement, press Measure.
Performing a Transfer Time Test
The transfer time test measures the time it takes the UPS to go from online-operation to battery backup
operation when the ac line fails.
To access the Transfer Time Test, go to the Tests menu and select the Transfer Time Test command.
In the dialog box that appears on the screen, enter values for the indicated parameters. The test uses these
parameters to run the test. For a detailed explanation of the Transfer Time test parameters refer to the
online help by pressing the F1 key.
When you are ready to run the test, press Run Test.
Modulating the Output Frequency
AC sources equipped with Option 020/022 have the capability to modulate the output frequency.
To access the Frequency Modulation dialog box, go to the Source menu and select the Frequency
Modulation command.
In the dialog box that appears on the screen, enter values for the indicated parameters. For a detailed
explanation of the Output Frequency parameters, refer to the online help by pressing the F1 key.
When you are finished, press Close. If FM has been enabled, the output frequency of the ac source starts
modulating immediately.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Power Analyzer Input Specifications
Table A-1 lists the specifications of the Agilent Dual Power Analyzer Option 020/022. Performance
o
specifications are warranted over the ambient temperature range of 0 to 40 C. Unless otherwise noted,
specifications are for a sinewave with a resistive load at a frequency range of 45 Hz to 1 kHz, with the
Agilent supplied current shunt, after a 30-minute warmup.
Table A-1. Performance Specifications
Measurement Accuracy
(@25 C ±5 C), ± (% of output + offset)
0.03% + 100 mV
0.1% + 100 mV
0.2% + 100 mV
rms Voltage (45−100 Hz):
(>100−500 Hz):
o
o
(>500 Hz−1 kHz):
dc Voltage:
0.03% + 150 mV
rms Current High Range
(45−100 Hz):
0.05% + 10 mA
0.05% + 15 mA
0.05% + 30 mA
(>100−500 Hz):
(>500 Hz−1 kHz):
rms Current Low Range
(45−100 Hz):
0.05% + 1.5 mA
0.05% + 8 mA
0.05% + 25 mA
(>100−500 Hz):
(>500 Hz−1 kHz):
repetitive pk current High Range
(45 Hz−1 kHz):
0.05% + 150 mA
0.03% + 150 mA
repetitive pk current Low Range
(45 Hz−1 kHz):
Power (VA) Low Range
0.1% + 1.5 VA + 1.2 mVA/V
0.1% + 2 VA + 1.2 mVA/V
0.1% + 6 VA + 1.2 mVA/V
(45−100 Hz):
(>100−500 Hz):
(>500 Hz-1 kHz):
Power (VA) High Range
0.1% + 1.5 VA + 12 mVA/V
0.1% + 2 VA + 12 mVA/V
0.1% + 6 VA + 12 mVA/V
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
Power (Watts) Low Range
(45−100 Hz):
0.1% + 0.3 W + 1.2 mW/V
0.1% + 1.2 W + 1.2 mW/V
0.1% + 2.5 W + 1.2 mW/V
(>100−500 Hz):
(>500 Hz−1 kHz):
Power (Watts) High Range
(45−100 Hz):
0.1% + 0.3 W + 12 mW/V
0.1% + 1.2 W + 12 mW/V
0.1% + 2.5 W + 12 mW/V
(>100−500 Hz):
(>500 Hz−1 kHz):
Power Factor:
0.01
Harmonic Measurement Accuracy
(50/60 Hz, @25 C ±5 C),
± (% of output + offset)
Voltage Magnitude:
Current Magnitude (Low Range)
Fundamental:
0.03% + 100 mV + 0.2%/kHz
o
o
0.03% + 1.5 mA
0.03% + 1 mA + 0.2%/kHz
Harmonics 2−49:
Current Magnitude (High Range)
Fundamental:
0.05% + 5 mA
0.05% + 3 mA + 0.2%/kHz
Harmonics 2−49:
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Table A-2. Supplemental Characteristics
Isolation to Ground:
Measurement Ranges
300 Vrms/425 Vdc
AC Voltage:
DC Voltage:
Current:
0−300 Vrms
± 425 V
± 80 A peak (high range)
± 8 A peak (low range)
13 A rms max. continuous
Average Measurement Resolution
rms Voltage:
rms Current:
10 mV
2 mA
5% of reading + 0.1%
THD (for a fundamental amplitude ≥ 5% of full scale):
Measurement System
Measurement Source Synchronization:
Measurement Buffer Length:
< 100 ns
4096 points
≤ 50 µs
25−250 µs
12 bits
Measurement/Generation Synchronization:
Measurement Acquisition Sampling Rate Range:
Voltage/Current Digitization Accuracy:
Voltage/Current Digitization Resolution:
Harmonic Measurement Time (amplitude):
16 bts
Meas:Curr:Harm? <n>
Meas:Array:Curr:Harm?
400 ms
10 s
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Verification and Calibration for Option 020/022
This section includes verification and calibration procedures for the Agilent Dual Power Analyzer Option
020/022. Add these procedures to those described in Appendix B of the AC Source User’s Guide.
The verification procedures do not check all the operating parameters, but verify that the ac source is
performing properly. Performance Tests, which check all the specifications of the ac source, are given in
the applicable ac source Service Manual.
Because the output of the ac source must be enabled during verification or calibration, proceed with
caution, since voltages and currents will be active at the output terminals.
Important
Perform the verification procedures before calibrating your ac source. If the ac source
passes the verification procedures, the unit is operating within its calibration limits and
does not need to be re-calibrated.
WARNING LETHAL VOLTAGES. Ac sources can supply 424 V peak at their output. DEATH
on contact may result if the output terminals or circuits connected to the output are
touched when power is applied. These procedures must be performed by a qualified
electronics technician or engineer trained on this equipment.
Equipment Required
The equipment listed in the following table, or the equivalent to this equipment, is required for
verification and calibration.
Table B-1. Equipment Required
Equipment
Characteristics
Recommended Model
0360-2681
5-pin connector plug
This connector is supplied with the ac source
when ordered with Option 020/022.
Digital Voltmeter
Resolution: 10 nV @ 1 V
Readout: 8.5 digits
Agilent 3458A
Accuracy: >20 ppm
Current Monitor1
Load Resistor
Guildline 7320/0.01
0.01 Ω, ±200 ppm,
30 ohm, 100 Watts minimum
1.0 µF, 50 V
Capacitor
(calibration only)
0160-3490
Function Generator
50 mV, 500 mV, 60 Hz sinewave capability
Agilent 33120A
(calibration only)
GPIB Controller
Full GPIB capabilities
HP Series 200/300 or equivalent
1The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage
drops in the load leads and connections. Connect the voltmeter directly to these current-monitoring
terminals.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Verification Procedure
The following verification procedures assume you understand how to operate the ac source from the front
panel as explained in chapter 4.
Perform the following tests for operation verification in the order indicated.
1. Turn-On Checkout
2. Voltage Measurement Accuracy
3. Current Measurement Accuracy
NOTE:
The ac source must pass turn-on selftest before you can proceed with the verification.
Test Set up
Agilent 6811B/6812B/6813B
LO HI
|
LO HI
I
V
SENSE
SENSE
POWER ANALYZER
HI
MAIN OUTPUT
SENSE
Agilent 3458A
DMM
COM COM
01
01
LO
RL
Rs
RL= Load resistor for CC test (30 ohms)
Rs = Current Monitor resistor (0.01 ohms)
HI
Agilent 3458A
DMM
LO
A.
Agilent 6811B/6812B/6813B
Agilent 33120A
LO HI
|
LO HI
FUNCTION GENERATOR
I
V
SENSE
SENSE
POWER ANALYZER
HI
Agilent 3458A
DMM
C1
MAIN OUTPUT
SENSE
LO
COM COM
01
01
C1 = Capacitor (1.0 uF)
B.
Figure B-1. Verification / Calibration Test Setup
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Power Analyzer Voltage Measurement Accuracy
This test verifies the accuracy of the Power Analyzer voltage measurements. The Phase 2 annunciator on
the left side of the display, must be on for the front panel to display Power Analyzer input measurements.
Figure B-1A shows the setup. Measure the ac output voltage directly at the output terminals.
Action
Normal Result
1.
Make sure the ac source is turned off. Connect the DVM to the output
terminals the ac source. Connect the Power Analyzer voltage sense terminals
(V SENSE) to the Main Output terminals of the ac source (see figure B-1A).
2.
3.
4.
Turn on the ac source with no load. In the Output menu, execute the *RST
command to reset the unit to its factory default state.
*RST
φ2 0 V 60 Hz
If the Phase 2 annunciator does not appear on the right side of the display,
press the Phase Select key.
Program the output voltage to 150 volts and set the output current limit to its
maximum value.
CV annunciator on.
5.
6.
Output voltage near 150 V.
Enable the output by pressing Output On/Off.
Record voltage readings at the DVM and on the front panel display.
φ2 front panel reading within
145 mV of DVM reading .
7.
8.
Program the output voltage to 300 volts.
Output voltage near 300 V.
Record voltage readings at the DVM and on the front panel display.
φ2 front panel reading within
190 mV of DVM reading
Power Analyzer Current Measurement Accuracy
This test verifies the accuracy of the Power Analyzer current measurements. The Phase 2 annunciator on
the left side of the display, must be on for the front panel to display Power Analyzer input measurements.
Figure B-1A shows the setup. Use the Agilent supplied current shunt. It has the accuracy specified in
table B-1. Use wire of sufficient size to carry the maximum rated current of the ac source.
NOTE:
Check to make sure that the internal shunt value in the ac source matches the value
written on the external current shunt. Refer to "Check the Shunt Value Setting".
Action
Normal Result
1.
Turn off the ac source. Connect the 30 ohm load resistor, current shunt,
and the DVM across the current shunt. Connect the Power Analyzer
current terminals (I SENSE) to the output of the function generator (see
figure B-1A).
2.
3.
4.
5.
Turn on the ac source. In the Output menu, execute the *RST command to
reset the unit to its factory default state.
*RST
φ2 0 V 60 Hz
If the Phase 2 annunciator does not appear on the right side of the display,
press the Phase Select key.
Program the output voltage to 100 volts and set the current limit to 3
amperes. Enable the output by pressing Output On/Off.
CC annunciator on.
Output current near 3 amperes
Record the DVM voltage reading and calculate the rms current. Divide the
DVM reading by the current monitor resistor value. Record the front panel
reading.
φ2 front panel reading within
11.5 mA of measured output
current.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Procedure
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup. The following
procedures assume you understand how to operate front panel keys (see chapter 4).
WARNING LETHAL VOLTAGES. Ac sources can supply 425 V peak at their output. DEATH
on contact may result if the output terminals or circuits connected to the output are
touched when power is applied. These procedures must be performed by a qualified
electronics technician or engineer trained on this equipment.
Summary of Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
Shift
Cal
Press this key to access the calibration menu.
Display
CAL ON <value>
Command Function
Turns calibration mode on when the correct password
value is entered.
CAL OFF
Turns calibration mode off
CAL:LEV <char>
CAL:DATA <value>
SHUNT <value>
CAL:VOLT:OFFSET
CAL:VOLT:DC
CAL:VOLT:AC
CAL:VOLT:PROT
CAL:VOLT:EXT
CAL:CURR:AC
CAL:CURR:MEAS
CAL:CURR:EXT
CAL:IMP
Advance to next step in sequence (P1, P2, P3, or P4).
Input a calibration measurement.
Enter a value for the external current shunt in ohms.
Begin voltage offset calibration
Begin dc voltage calibration sequence
Begin ac voltage calibration sequence
Begin voltage protection calibration
Begin external voltage calibration sequence
Begin ac current calibration sequence
Begin current measurement calibration sequence
Begin external current calibration sequence
Begin output impedance calibration sequence
Saves the calibration constants in non-volatile memory.
Set new calibration password.
CAL:SAVE
CAL:PASS <value>
Enable Calibration Mode
Action
Display
1.
2.
*RST
Reset the unit by selecting Shift, Output, and pressing Enter.
CAL ON 0.0
Press Shift Calibration, enter the calibration password from Entry keypad,
and press Enter. If the password is correct, the Cal annunciator will come on.
(If no password has been set, just press Enter.)
OUT OF RANGE
CAL DENIED
If the password is incorrect, an error occurs. If the active password is lost, the
calibration function can be recovered by setting an internal switch that defeats
password protection (see the Service Manual.)
If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed (see the Service Manual.)
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibrating and Entering Power Analyzer Voltage Calibration Values
Action
Display
4.
Connect an Agilent 3458A DVM (in synchronous ac volts mode) to the ac source
output terminals. Short the Power Analyzer voltage terminals (V SENSE) to
ensure no dc offset error.
5.
6.
CAL:VOLT:EXT
Press Shift Calibration, scroll to CAL VOLT EXT, and press Enter.
Leave the ac source on and remove the short from the Power Analyzer voltage
terminals.
7.
8.
Connect the Power Analyzer voltage sense terminals to the Main Output
terminals of the ac source (see figure 3-1A).
CAL:LEV P1
Press Shift Calibration, scroll to CAL LEV P1 and press Enter.
The ac source will output 300 Vac 60 Hz sine wave.
9.
CAL:DATA 0.00
CAL SAVE
Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to
enter the voltage value displayed on the Agilent 3458A DVM.
10.
Press Shift Calibration, scroll to CAL SAVE and press Enter.
Calibrating and Entering Power Analyzer Current Calibration Values
Action
Display
11.
Short the Power Analyzer current sense terminals (I SENSE) to ensure no dc
offset error.
12.
13.
CAL:CURR:EXT
Press Shift Calibration, scroll to CAL CURR EXT and press Enter.
Leave the ac source on and remove the short from the Power Analyzer current
terminals.
14
Connect the Power Analyzer current terminals to the output of the function
generator (see figure 3-1B). Set the function generator output to 50 mV, 60 Hz.
15.
Connect the Power Analyzer current terminals to the Agilent 3458A input
terminals. Note that the Agilent 3458A must be in synchronous ac volts mode and
a 1.0 uF capacitor must be installed across the Agilent 3458A input terminals.
16.
17.
CAL LEV P1
Press Shift Calibration, scroll to CAL LEV P1, and press Enter.
CAL:DATA 0.00
Press Shift Calibration and scroll to CAL DATA. Use the Entry keypad to
enter the voltage value displayed on the Agilent 3458A DVM.
18
Change the function generator output setting to 500 mV.
19.
CAL:LEV P2
CAL:DATA 0.00
CAL SAVE
Press Shift Calibration, scroll to CAL LEV P1 command, use ° to scroll to
the P2 parameter, and press Enter.
20.
21
Press Shift Calibration and scroll to the CAL DATA. Use the Entry keypad to
enter the voltage value displayed on the Agilent 3458A DVM.
Press Shift Calibration, scroll to CAL SAVE and press Enter
26
Download from Www.Somanuals.com. All Manuals Search And Download.
|