USER’S GUIDE
Agilent Technologies
DC Electronic Loads
Models N3300A, N3301A, N3302A,
N3303A, N3304A, N3305A and N3306A
Part No. 5964-8196
Microfiche No. 5964-8197
Printed in Malaysia
October, 2001
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety Summary
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to
comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to
comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this
product may be impaired if it is used in a manner not specified in the operation instructions.
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to
operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for
the ac mains voltage requirements and ambient operating temperature range.
BEFORE APPLYING POWER
Verify that all safety precautions are taken. Note the instrument's external markings described under "Safety Symbols".
GROUND THE INSTRUMENT
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the
instrument chassis and cover must be connected to an electrical ground. The instrument must be connected to the ac power
mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at
the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth
terminal will cause a potential shock hazard that could result in personal injury.
ATTENTION: Un circuit de terre continu est essentiel en vue du fonctionnement sécuritaire de l'appareil. Ne
jamais mettre l'appareil en marche lorsque le conducteur de mise … la terre est d‚branch‚.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS
Operating personnel must not remove instrument covers except as instructed in this Guide for installing or removing
electronic load modules. Component replacement and internal adjustments must be made only by qualified service
personnel. Do not replace components with power cable connected. Under certain conditions dangerous voltages may exist
even with the power cable removed. To avoid injuries always disconnect power, discharge circuits, and remove external
voltage sources before touching components.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt internal service or adjustment unless another person capable of rendering first aid resuscitation is present.
DO NOT EXCEED INPUT RATINGS
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a
properly grounded receptacle to minimize shock hazard. Operation at line voltages or frequencies in excess of those stated
on the data plate may cause leakages in excess of 5.0mA peak.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until
they can be repaired by qualified service personnel.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
SAFETY SYMBOLS
Direct current
Alternating current
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective earth (ground) terminal
Frame or chassis terminal
Terminal is at earth potential. Used for measurement and control circuits designed to
be operated with one terminal at earth potential.
Terminal for Neutral conductor on permanently installed equipment
Terminal for Line conductor on permanently installed equipment
On (supply)
Off (supply)
Standby (supply). Units with this symbol are not completely disconnected from ac
mains when this switch is off. To completely disconnect the unit from ac mains, either
disconnect the power cord or have a qualified electrician install an external switch.
In position of a bi-stable push control
Out position of a bi-stable push control
Caution, risk of electric shock
Caution, hot surface
Caution (refer to accompanying documents)
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or
the like, which, if not correctly performed or adhered to, could result in personal
injury. Do not proceed beyond a WARNING sign until the indicated conditions are
fully understood and met.
WARNING
Caution
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or
the like, which, if not correctly performed or adhered to, could result in damage to or
destruction of part or all of the product. Do not proceed beyond a CAUTION sign
until the indicated conditions are fully understood and met.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Declaration Page
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Agilent Technologies, Inc.
140 Green Pond Road
Rockaway, New Jersey 07866
U.S.A.
Manufacturer's Name
Manufacturer's Address
declares that the Products
a) dc Electronic Load Mainframes
Product Names
b) dc Electronic Load Modules for Mainframe
a) N3300A, N3301A
b) N3302A, N3303A, N3304A, N3305A and N3306A
Model Numbers
Product Options
This declaration covers all options and customized products based on the above
as stated in TCF Certificate of Conformance CC/TCF/00/084
EMC information:
Conforms with the protection requirements of European Council EMC Directive 89/336/EEC on the
approximation of the laws of the member states relating to electromagnetic compatibility:
As detailed in:
Assessed by:
Electromagnetic Compatibility (EMC), Certificate of Conformance Number
CC/TCF/00/084
based on Technical Construction File (TCF) ANJ10, dated Sept. 26, 2000
Celestica Ltd, Appointed Competent Body
Westfields House, West Avenue
Kidsgrove, Stoke-on-Trent
Straffordshire, ST7 1TL
United Kingdom
Safety information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC
and carries the CE-marking accordingly.
Supplementary Information:
The product conforms to the following safety standards:
IEC 1010-1:1990+A1+A2 / EN 61010-1:1993 +A2
UL 3111-1:1994
CSA C22.2 No. 1010.1:1993
January, 2001
Date
Hank Kowalla / Quality Manager
European contact for regulatory topics only
Your local Agilent Technologies Sales and Service Office or
Agilent Technologies GmbH, HQ-TRE, Herrenberger Straße 130,
D-71034 Böblingen (FAX: +49-7031-14-3143)
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Acoustic Noise Information
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenlä
minformationsverordnung vom 18 Januar 1991.
* Schalldruckpegel Lp <70 dB(A)
* Am Arbeitsplatz
* Normaler Betrieb
* Nach EN 27779 (Typprüfung).
Manufacturer's Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive,
from 18 January 1991.
* Sound Pressure Lp <70 dB(A)
* At Operator Position
* Normal Operation
* According to EN 27779 (Type Test).
Printing History
The edition and current revision of this manual are indicated below. Reprints of this manual containing minor
corrections and updates may have the same printing date. Revised editions are identified by a new printing date.
A revised edition incorporates all new or corrected material since the previous printing date.
Changes to the manual occurring between revisions are covered by change sheets shipped with the manual. In
some cases, the manual change applies only to specific instruments. Instructions provided on the change sheet
will indicate if a particular change applies only to certain instruments.
This document contains proprietary information protected by copyright. All rights are reserved. No part of this
document may be photocopied, reproduced, or translated into another language without the prior consent of
Agilent Technologies. The information contained in this document is subject to change without notice.
Copyright 2000, 2001 Agilent Technologies, Inc.
Edition 1 ________ August, 2000
Update 1 ________ November 2000
Update 2 ________ June 2001
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
Warranty Information ...........................................................................................................................................................2
Safety Summary......................................................................................................................................................................3
Declaration Page.....................................................................................................................................................................5
Acoustic Noise Information....................................................................................................................................................6
Printing History ......................................................................................................................................................................6
Table of Contents....................................................................................................................................................................7
QUICK REFERENCE ............................................................................................................ 11
The Front Panel -At a Glance..............................................................................................................................................11
The Rear Panel At a Glance.................................................................................................................................................12
Instrument Configuration....................................................................................................................................................12
Front Panel Number Entry..................................................................................................................................................13
Front Panel Annunciators....................................................................................................................................................14
Immediate Action Keys ........................................................................................................................................................14
Front Panel Menus - At a Glance........................................................................................................................................15
SCPI Programming Commands - At a Glance ..................................................................................................................17
GENERAL INFORMATION................................................................................................... 19
Document Orientation..........................................................................................................................................................19
Safety Considerations...........................................................................................................................................................20
Options and Accessories.......................................................................................................................................................20
Description ............................................................................................................................................................................20
Features and Capabilities....................................................................................................................................................21
Front Panel Controls...........................................................................................................................................................21
Remote Programming.........................................................................................................................................................21
Operating Modes ..................................................................................................................................................................22
Constant Current CC (Mode) .............................................................................................................................................22
Constant Resistance (CR) Mode ........................................................................................................................................23
Constant Voltage (CV) Mode.............................................................................................................................................24
Transient Operation.............................................................................................................................................................26
List Operation.......................................................................................................................................................................27
Triggered Operation.............................................................................................................................................................27
Input Control ........................................................................................................................................................................30
Protection Features...............................................................................................................................................................30
Saving and Recalling Settings..............................................................................................................................................33
External Control Signals......................................................................................................................................................33
Remote Sensing..................................................................................................................................................................33
Monitor Outputs .................................................................................................................................................................33
External Programming Input..............................................................................................................................................33
Fault....................................................................................................................................................................................34
Port On/Off.........................................................................................................................................................................34
Input Measurements.............................................................................................................................................................34
DC Measurements..............................................................................................................................................................35
RMS Measurements ...........................................................................................................................................................35
Minimum and Maximum Measurements ...........................................................................................................................35
Power Measurements .........................................................................................................................................................35
Measurement Ranges .........................................................................................................................................................35
INSTALLATION..................................................................................................................... 37
Inspection ..............................................................................................................................................................................37
Damage ..............................................................................................................................................................................37
Packaging Material.............................................................................................................................................................37
Items Supplied....................................................................................................................................................................37
7
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning .............................................................................................................................................................................37
Installing the Modules ..........................................................................................................................................................38
Procedure............................................................................................................................................................................38
Channel Number ................................................................................................................................................................39
Location.................................................................................................................................................................................40
Bench Operation.................................................................................................................................................................41
Rack Mounting...................................................................................................................................................................41
Input Connections.................................................................................................................................................................42
Power Cord.........................................................................................................................................................................42
Manually-Tightened Connectors........................................................................................................................................42
8mm Screw Terminal Connector (option UJ1) ..................................................................................................................43
Wire Considerations...........................................................................................................................................................44
Control Connector..............................................................................................................................................................46
Sense Switch ......................................................................................................................................................................47
Trigger and Digital Connections ........................................................................................................................................47
Computer Connections.........................................................................................................................................................48
GPIB Interface....................................................................................................................................................................48
RS-232 Interface ................................................................................................................................................................48
Application Connections ......................................................................................................................................................49
Local Sense Connections....................................................................................................................................................49
Remote Sense Connections ................................................................................................................................................49
Parallel Connections...........................................................................................................................................................49
Low Voltage Operation......................................................................................................................................................51
TURN-ON CHECKOUT......................................................................................................... 53
Introduction ..........................................................................................................................................................................53
Checkout Procedure .............................................................................................................................................................53
In Case of Trouble ................................................................................................................................................................54
Error Messages...................................................................................................................................................................54
Selftest Errors.....................................................................................................................................................................54
FRONT PANEL OPERATION............................................................................................... 55
Introduction ..........................................................................................................................................................................55
Front Panel Description .......................................................................................................................................................55
System Keys...........................................................................................................................................................................57
Function keys ........................................................................................................................................................................58
Immediate Action Keys......................................................................................................................................................58
Scrolling Keys....................................................................................................................................................................59
Metering Keys....................................................................................................................................................................59
Input Control Keys.............................................................................................................................................................60
Transient Control Keys ......................................................................................................................................................61
Trigger Control Keys..........................................................................................................................................................61
List Control Keys ...............................................................................................................................................................61
Entry Keys.............................................................................................................................................................................62
Examples of Front Panel Programming .............................................................................................................................63
1 - Using the Front Panel Display ......................................................................................................................................63
2 - Programming Constant Current, Voltage and Resistance Modes .................................................................................63
3 - Programming Transient Operation................................................................................................................................65
4 - Programming Lists........................................................................................................................................................67
5 - Querying and Clearing Output Protection and Errors...................................................................................................68
6 - Making Basic Front Panel Measurements.....................................................................................................................69
7 - Setting the GPIB Address .............................................................................................................................................70
8 - Storing and Recalling Instrument States .......................................................................................................................70
SPECIFICATIONS.................................................................................................................71
8
Download from Www.Somanuals.com. All Manuals Search And Download.
PERFORMANCE TEST AND CALIBRATION PROCEDURES............................................ 75
Introduction ..........................................................................................................................................................................75
Equipment Required ............................................................................................................................................................75
Performance Tests ................................................................................................................................................................76
IMON Zero Verification ....................................................................................................................................................76
CC Mode Tests...................................................................................................................................................................76
CV Mode Tests...................................................................................................................................................................77
CR Mode Tests...................................................................................................................................................................79
Transient Generator Mode Tests ........................................................................................................................................81
Agilent N3302A Verification Test Record.........................................................................................................................83
Agilent N3303A Verification Test Record.........................................................................................................................84
Agilent N3304A Verification Test Record.........................................................................................................................85
Agilent N3305A Verification Test Record.........................................................................................................................86
Agilent N3306A Verification Test Record.........................................................................................................................87
Calibration ............................................................................................................................................................................90
Parameters Calibrated.........................................................................................................................................................90
IMON, IPROG and CURRENT Calibration Program .......................................................................................................91
VOLTAGE Calibration Program .......................................................................................................................................95
RESISTANCE Calibration Program..................................................................................................................................97
INDEX.................................................................................................................................. 101
Manual Updates................................................................................................................................................................105
9
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Quick Reference
The Front Panel -At a Glance
j
15-character display shows
channel, voltage and current
measurements.
k
Annunciators indicate
operating modes and status
conditions.
l
System keys:
♦ Return to Local mode.
♦ Set the GPIB address.
♦ Set the RS-232 interface.
♦ Display SCPI error codes.
♦ Save and recall instrument
states.
1
4
2
3
5
N3300A
SYSTEM DC ELECTRONIC LOAD
SYSTEM
Ident
Local
FUNCTION
Func
ENTRY
CHANNEL
VOLTS
AMPS
Sense
Meter
Step
7
4
8
5
2
9
Input
Error
Address
Channel
Channel
Step
CV CC CR
Unr Dis Tran
Prot Cal Shift Rmt Addr Err SQR
6
Current
Res
Input
Enter
Save
Recall
Prot Clear
Protect
List
Tran
1
3
Clear Entry
E
-
.
Trigger
Trigger
Control
Voltage
Input
0
on/off
LINE
ON
OFF
6
m
Function keys:
n
Entry keys:
oTurns the electronic load on
and off.
♦ Select metering functions.
♦ Enable/disable input.
♦ Program current, resistance and
voltage modes.
♦ Enter values.
♦ Increment or decrement values.
♦ Set and clear protection
functions.
♦ Scroll through front panel
menu commands.
11
Download from Www.Somanuals.com. All Manuals Search And Download.
1 - Quick Reference
The Rear Panel At a Glance
Q
Control connector
R
Input binding post
S
GPIB connector
T
RS-232 interface
connector
1
2
3
4
5
7
6
U
Power cord
connector
V
Trigger/Digital
connector
W
Sense switch
Instrument Configuration
Use the front panel Address menu to
♦ Select GPIB or RS-232 interface (see Chapter 5 in User's Guide).
♦ Select the GPIB bus address (see Chapter 5 in User's Guide).
♦ Configure the RS-232 interface (see Chapter 5 in User's Guide).
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Quick Reference - 1
Front Panel Number Entry
Enter numbers from the front panel as follows:
Use the Entry Scroll keys to adjust the input setting in Meter mode.
Meter
AND
Input
Input
If CC is lit, the input current changes.
If CV is lit, the input voltage changes.
If CR is lit, the input resistance changes.
NOTE
The input must be on for input values to change.
Use the Function keys and Entry keys to enter a new value
NOTE
If you make a mistake use the Backspace key to delete the number, or press the Meter key to
return to the Meter mode.
Current
Res
7
4
8
5
2
9
6
Input
Input
Enter
Enter
AND
AND
1
3
Clear Entry
E
-
.
0
Voltage
13
Download from Www.Somanuals.com. All Manuals Search And Download.
1 - Quick Reference
Front Panel Annunciators
A list is initiated or running.
Prot
Indicates that a channel protection feature is active
on any channel. Press the Prot Clear key to clear
the protection condition.
φ1
CV
CC
CR
The selected input channel is in the
constant voltage (CV) mode.
Cal
Calibration mode is ON. Calibration can only be
done through the computer interface.
The selected input channel is in the
constant current (CC) mode.
Shift
Rmt
Indicates that the shift key has been pressed.
The selected input channel is in the
constant resistance (CR) mode.
Indicates that the electronic load is in remote state
(either GPIB or RS-232). In the remote state, only
the active key is the Local key.
Unr
Dis
The selected input channel is unregulated.
Addr
Err
The electronic load is addressed to talk or listen.
A remote programming error(s) have occurred.
The input is OFF. Press the Input on/off
key to turn the input on.
Tran
The selected input channel is enabled for
transient operation.
SQR
The electronic load is requesting a service.
Immediate Action Keys
A toggle switch that turns the input of the electronic load on or off.
Input
On/Off
Activates front panel control when the unit is in remote mode
(unless a Lockout command is in effect).
Local
Increases the input current (CC), voltage (CV), or resistance (CR) in Meter mode.
Q
Input
Decreases the input current (CC), voltage (CV), or resistance (CR) in Meter mode.
Selects another channel.
Input
Q
Channel
Causes an initiate and trigger to occur. Used with transient subsystem or list.
Displays any protection functions that are tripped.
+
Shift
Protect
Shift
Trigger
Resets the protection circuit and allows the unit to return to its last programmed state.
Identifies the module installed in the selected channel location. (not available)
Returns the front panel to metering mode from any other mode.
+
+
Prot
Clear
Shift
Ident
Meter
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Quick Reference - 1
Front Panel Menus - At a Glance
Address
Sets the GPIB Address
Selects an interface (GPIB or RS232)
Selects baud rate (300, 600, 1200, 2400, 4800, 9600) *
Selects message parity (NONE, EVEN, ODD, MARK, SPACE) *
Selects flow control (XON-XOFF, RTS-CTS, DTR-DSR, NONE) *
Recalls the instrument state
Resets the instrument to its power-on state
Saves the present instrument state
Displays the number of errors in the SCPI error queue
Allows selection of channel to be controlled by the front panel
Defines the number of data points in the measurement
Sets the digitizer sample spacing
ADDRESS 5
INTF GPIB
BAUDRATE 300
PARITY NONE
FLOW NONE
*RCL 0
*RST
*SAV 0
ERROR 0
CHANNEL 1
S:PNT
R
R
R
R
Recall
Shift
R
Save
Error
Channel
Sense
R
Shift
Shift
Shift
S:TIN
Sets the measurement window function (RECT, HANN)
Defines the data offset in the measurement
Selects the current measurement range
R
S:WIN
S:OFF
S:C:RNG
S:V:RNG
FUNC
FNC:MODE
INP:SHOR
R
R
Selects the voltage measurement range
R
Sets the regulation mode (CURR, RES, VOLT)
Selects what controls regulating mode (FIX, LIST)
Enable/disables the input short (OFF/ON)
Func
R
R
Protect
General protection status (overcurrent fault shown)
Voltage protection status (remote reverse voltage fault shown)
OC -- -- --
RRV --- --
R
Meter
Displays the input voltage and current
Displays the maximum voltage
Displays the minimum voltage
Displays the rms voltage
Displays the maximum current
Displays the minimum current
Displays the rms current
XXXX XXXX
XXXX V MAX
XXXX V MIN
XXXX V RMS
XXXX A MAX
XXXX A MIN
XXXX A RMS
XXXX WATTS
XXXX W MAX
XXXX W MIN
CURR
R
R
R
R
R
R
R
R
R
Displays the wattage
Displays the maximum wattage
Displays the minimum wattage
Sets the input current
Current
Sets the current mode (FIXED, LIST)
Sets the input current range
Sets current slew rate
Sets current slew rate for negative transitions
Sets current slew rate for positive transitions
Sets the transient input current
Sets the triggered input current
Sets the input resistance
R
R
R
R
R
R
R
C:MODE
C:RANG
C:SLEW
C:SLW:N
C:SLW:P
C:TLEV
C:TRIG
RES
Res
Sets the resistance mode (FIXED, LIST)
Sets the input resistance range
Sets resistance slew rate
Sets resistance slew rate for negative transitions
Sets resistance slew rate for positive transitions
Sets the transient input resistance
Sets the triggered input resistance
R
R
R
R
R
R
R
R:MODE
R:RANG
R:SLEW
R:SLW:N
R:SLW:P
R:TLEV
R:TRIG
*Only applicable for use with RS-232
15
Download from Www.Somanuals.com. All Manuals Search And Download.
R:RANG:0 EOL
R:SLEW:0 EOL
R:SLW:N:0 EOL
R:SLW:P:0 EOL
R:TLEV:0 EOL
TRAN:0 EOL
T:DCYC:0 EOL
T:FREQ:0 EOL
T:MODE:0 EOL
T:TWID:0 EOL
VOLT:0 EOL
V:RANG:0 EOL
V:SLEW:0 EOL
V:SLW:N:0 EOL
V:SLW:P:0 EOL
V:TLEV:0 EOL
1 - Quick Reference
Front Panel Menus - continued
Sets the input voltage
VOLT
Voltage
Sets the voltage mode (FIXED, LIST)
Sets the input voltage range
Sets voltage slew rate
Sets voltage slew rate for negative transitions
Sets voltage slew rate for positive transitions
Sets the transient input voltage
R
R
R
R
R
R
R
V:MODE
V:RANG
V:SLEW
V:SLW:N
V:SLW:P
V:TLEV
V:TRIG
TRAN
T:DCYC
T:FREQ
T:MODE
T:TWID
Sets the triggered input voltage
Tran
Enables/disables the transient generator (OFF/ON)
Sets the transient duty cycle in continuous mode
Sets the transient frequency in continuous mode
Sets the transient mode (CONT, PULSE, TOGGLE)
Sets the transient pulse width in pulse mode
R
R
R
R
Trigger
List
Initiates the trigger system
Aborts the initiated trigger
Sets the method of incrementing steps (ONCE, AUTO)
Specifies the number of times the list is cycled
Specifies the time period of each step
Specifies the current setting for each step
Specifies the current range for each step
Sets the current slew rate for each step
Sets the negative current slew rate for each step
Sets the positive current slew rate for each step
Sets the transient input current for each step
Sets the list regulation mode (CURR, RES, VOLT)
Specifies the resistance setting for each step
Specifies the resistance range for each step
Sets the resistance slew rate for each step
Sets the negative resistance slew rate for each step
Sets the positive resistance slew rate for each step
Sets the transient input resistance for each step
Enables/disables the transient level for each step
Sets the transient duty cycle for each step
Sets the transient frequency for each step
Sets the mode of the transient generator (CONT, PULSE)
Sets the transient pulse width for each step
Specifies the voltage setting for each step
Specifies the voltage range for each step
INIT:IMMED
ABORT
LST:STEP
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
LST:CNT
DWEL:0 EOL
CURR:0 EOL
C:RANG:0 EOL
C:SLEW:0 EOL
C:SLW:N:0 EOL
C:SLW:P:0 EOL
C:TLEV:0 EOL
FUNC:0 EOL
RES:0 EOL
Sets the voltage slew rate for each step
Sets the negative voltage slew rate for each step
Sets the positive voltage slew rate for each step
Sets the transient input voltage for each step
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Quick Reference - 1
SCPI Programming Commands - At a Glance
NOTE
Most [optional] commands have been omitted for clarity. Refer to the Programming Guide for
a complete description of all programming commands.
ABORt
CALibrate
[SOURce:]CURRent
[:LEVel] <n>
:DATA <n> [,<n>]
:TRIG <n>
:MODE <mode>
:PROTection
[:LEVel] <n>
:IMON:LEVel <points>
:IPRog:LEVel <points>
:LEVel <points>
:PASSword <n>
:SAVE
:DELay <n>
:STATe <bool>
:STATE <bool> [,<n>]
CHANnel | INSTrument
[:LOAD] <n>
:RANGe <n>
:SLEW
[:BOTH] <n>
:NEGative <n>
:POSitive <n>
INITiate
[:IMMediate]
:SEQuence[1] | :SEQuence2
:NAMe LIST | ACQuire
CONTinuous
:TLEVel <n>
FUNCtion | MODE
:MODE <mode>
:SEQuence1 <bool>
:NAMe LIST <bool>
INPut | OUTput
[:STATe] <bool>
:PROTection
:CLEar
LIST
:COUNt <n>
:CURRent
[:LEVel] <n> {,<n>}
:POINts?
:RANGe <n> {,<n>}
:POINts?
:SHORt
[:STATe] <bool>
:SLEW
MEASure | FETCh
:ARRay
[:BOTH] <n> {,<n>}
:POINts?
:NEGative <n> {,<n>}
: POSitive <n> {,<n>}
:CURRent?
:POWer?
:VOLTage?
:TLEVel <n> {,<n>}
:POINts?
[:SCALar]
:CURRent?
:DWELI <n> {,<n>}
:POINts?
:ACDC?
:MAX?
:MIN?
:FUNCtion | MODE <mode>
:RESistance
:POWer?
[:LEVel] <n> {,<n>}
:MAX?
:MIN?
:POINts?
:RANGe <n> {,<n>}
:POINts?
:VOLTage?
:ACDC?
:SLEW
:MAX?
:MIN?
[:BOTH] <n> {,<n>}
:POINts?
:NEGative <n> {,<n>}
: POSitive <n> {,<n>}
PORT0[:STATe] <bool>
PORT1[:LEVel] <n>
SENSe
:TLEVel <n> {,<n>}
:CURRent
:POINts?
:RANGe <n>
:STEP <step>
:TRANsient
:SWEep
:OFFSet
:POINts <n>
[:STATe] <bool> {,<bool>}
:POINts?
:TINTerval <n>
:WINDow <type>
:VOLTage
:DCYCle <n> {,<n>}
:POINts?
:FREQuency <n> {,<n>}
:POINts?
:RANGe <n>
:MODE <mode> {<mode>}
:POINts?
:TWIDth <n> {,<n>}
:POINts?
17
Download from Www.Somanuals.com. All Manuals Search And Download.
1 - Quick Reference
SCPI Commands - continued
STATus
[SOURce:]LIST (continued)
:VOLTage
:CHANnel
[:EVENt]?
:CONDition?
:ENABle <n>
[:LEVel] <n> {,<n>}
:POINts?
:RANGe <n> {,<n>}
:POINts?
:CSUMmary
[:EVENt]?
:ENABle <n>
:OPERation
:SLEW
[:BOTH] <n> {,<n>}
:POINts?
:NEGative <n> {,<n>}
: POSitive <n> {,<n>}
[:EVENt]?
:CONDition?
:ENABle <n>
:NTRansition <n>
:PTRansition <n>
:TLEVel <n> {,<n>}
:POINts?
RESistance
[:LEVel] <n>
:TRIG <n>
:QUEStionable
[:EVENt]?
:CONDition?
:ENABle <n>
:MODE <mode>
:RANGe <n>
:SLEW
SYSTem
TRIGger
:ERRor?
:VERSion?
:LOCal
[:BOTH] <n>
:NEGative <n>
:POSitive <n>
:REMote
:RWLock
:TLEVel <n>
TRANsient
VOLTage
[:STATe] <bool>
:DCYCle <n>
:FREQuency <n>
:MODE <mode>
:TWIDth <n>
[:IMMediate]
:DELay
:SOURce <source>
:TIMer
:SEQuence2 | ACQuire
:COUNt
[:LEVel] <n>
:TRIG <n>
:MODE <mode>
:RANGe <n>
:SLEW
[:BOTH] <n>
:NEGative <n>
: POSitive <n>
:TLEVel <n>
18
Download from Www.Somanuals.com. All Manuals Search And Download.
2
General Information
Document Orientation
This manual describes the operation of the Agilent Model N3300A, N3301A, N3302A, N3303A N3304A,
N3305A and N3306A DC Electronic Loads. Unless otherwise noted, all units will be referred to by the
description "electronic load" throughout this manual. The following documents and software are shipped with
your electronic load:
♦ A User's Guide (this document), contains installation, checkout and front panel information.
♦ A Programming Guide, contains detailed GPIB programming information.
The following Getting Started Map will help you find the information you need to complete the specific task
that you want to accomplish. Refer to the table of contents or index of each guide for a complete list of the
information contained within.
Getting Started Map
Task
Where to find information
User's Guide
Installing the unit
Line voltage connections
Installing modules
Load connections
User's Guide
User's Guide
Checking out the unit
Verifying proper operation
Using the front panel
Calibrating the unit
Using the front panel
Front panel keys
Front panel examples
User's Guide
Programming Guide
Using the programming interface
GPIB interface
RS-232 interface
Programming Guide
Programming the unit using SCPI commands
SCPI commands
SCPI programming examples
Programming the unit using VXIplug&play
instrument driver
VXIplug&play on-line help
NOTE:
Installing the instrument driver
Instrument driver functions
C/C++ example programs
Visual BASIC example programs
Lab VIEW example programs
Agilent VEE example programs
The driver must be installed on your computer to
access the on-line information.
Drivers for Agilent instruments are available on the
web at www.agilent.com/find/drivers
19
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
Safety Considerations
This electronic load is a Safety Class 1 instrument, which means it has a protective earth terminal. That terminal
must be connected to earth ground through power source equipped with a ground receptacle. Refer to the Safety
Summary page at the beginning of this guide for general safety information. Before installation or operation,
check the electronic load and review this guide for safety warnings and instructions. Safety warnings for
specific procedures are located at appropriate places in the Guide.
Options and Accessories
Table 2-1 Options
Option
Description
800
One rack mount kit for two N3301A half-rack units side by side*.
Consists of: Lock-link kit (p/n 5061-9694) and Flange kit (p/n 5062-9215)
908
One rack mount kit*.
Consists of: Flange kit for N3300A (order 2 p/n 5063-9212)
Flange kit for N3301A (p/n 5062-3960)
909
UJ1
One rack mount kit with handles for N3300A* (order 2 p/n 5063-9219).
8mm input screw terminal connectors (see chapters 2 and 3)
*Support rails (p/n E3663AC) are required.
Table 2-2 Accessories
GPIB cables
Agilent Part Number
10833A
1.0 meter (3.3 ft)
2.0 meters (6.6 ft)
4.0 meters (13.2 ft)
0.5 meters (1.6 ft)
RS-232 cable
10833B
10833C
10833D
34398A
(9-pin F to 9-pin F, 2.5 meter, null modem/printer cable with one
9-pin M to 25 pin F adapter)
RS-232 adapter kit (contains 4 adapters)
9-pin M to 25-pin M for pc or printer
9-pin M to 25-pin M for pc or printer
9-pin M to 25-pin M for modem
34399A
9-pin M to 9-pin M for modem
Description
The N3300A is a DC Electronic Load Mainframe used for design, manufacturing, and evaluation of dc power
supplies, batteries, and power components. Other applications include use as a power circuit breaker or crowbar,
high current function or pulse generator, fuel-cell and photovoltaic cell test, and de-energizing superconducting
magnets.
The mainframe contains six slots for load modules. Load modules occupy either 1 or 2 slots depending on the
power rating of the module. The mainframe can dissipate up to 300 watts per slot, to a total of 1800 watts for a
fully loaded mainframe. Each individual module has its own channel number and contains its own input
connector. The mainframe contains a processor, GPIB connector, RS-232 connector and interface circuits,
trigger circuits, front-panel keypad and display, and other circuits common to all the load modules.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
The N3301A is a DC Electronic Load Mainframe that is functionally identical to the N3300A, but is a half-rack
width with only two slots for load modules. The mainframe can dissipate up to 300 watts per slot, to a total of
600 watts for a fully loaded mainframe.
The N3302A, N3303A N3304A, N3305A and N3306A are electronic load modules that can be installed in the
N3300A and N3301A mainframes. The module specific pages in Appendix A include specifications and other
information pertinent to a particular model. Each module can operate independently in constant current (CC)
mode, constant voltage (CV) mode, or constant resistance (CR) mode. In addition, each input can be turned on
or off (open-circuited), or short-circuited.
Features and Capabilities
♦ Constant current (CC), constant voltage (CV), or constant resistance (CR) mode operation.
♦ Built-in GPIB and RS-232 interface programming with SCPI command language.
♦ Triggered input and measurement functions.
♦ Front panel control with keypad.
♦ Independent channel operation.
♦ Built-in pulse generator for continuous, pulsed, and toggled transient operation.
♦ Overvoltage, overcurrent, overpower, and overtemperature protection.
♦ Extensive selftest, status reporting and software calibration.
♦ Fan speed control for reduced acoustic noise under light load conditions.
Front Panel Controls
The front panel has keyboard controls for setting the input voltage, current and resistance. The panel display
provides digital readouts of a number of functions including the inputs. Annunciators display the operating
status of the electronic load. System keys let you perform system functions such as setting the GPIB address
and recalling operating states. Front panel function keys access the electronic load function menus. Front panel
Entry keys let you select and enter parameter values. Refer to chapter 5 for a complete description of the front
panel controls.
Remote Programming
The electronic load may be remotely programmed via the GPIB bus and/or an RS-232 serial port. GPIB
programming is done with SCPI (Standard Commands for Programmable Instruments) commands, which make
the electronic load programs compatible with those of other GPIB instruments that are also SCPI compatible.
Local (front panel) control is in effect immediately after power is applied. The front panel keypad and display
allow manual control of each individual module when the electronic load is used in bench test applications.
Remote (computer) control goes into effect (front panel Rmt annunciator is on) as soon as the mainframe
receives a command via the GPIB. A built-in GPIB interface and SCPI commands allow control and readback
of all functions when the electronic load is used in computer controlled applications.
With remote control in effect, only the computer can control the electronic load; the front panel keypad has no
effect. You can still use the front panel display to view the input voltage and current readings. To return the
electronic load to local control, press the Local key. This will return the electronic load to local control, unless
the local-lockout command has been received from the GPIB controller.
Most functions that can be performed remotely over the GPIB or RS-232 can also be performed from the front
panel. Whenever possible the function menu commands reflect their corresponding SCPI commands. Thus,
learning to operate the electronic load from the front panel will aid you when writing computer programs.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
Operating Modes
The three modes of operation are:
♦ Constant current (CC).
♦ Constant voltage (CV).
♦ Constant resistance (CR).
When programmed to a mode, a module remains in that mode until the mode is changed or until a fault
condition, such as an overpower or overtemperature, occurs.
The current, resistance, and voltage mode parameters described in subsequent paragraphs can be programmed
whether or not the mode is presently selected. When a mode is selected via the front panel or via the GPIB or
RS-232, most of the associated parameters will take effect at the input (exceptions are noted in the mode
descriptions).
Constant Current CC (Mode)
In this mode, the module will sink a current in accordance with the programmed value regardless of the input
voltage (see Figure 2-1). The CC mode can be set with front panel keys or via the GPIB or RS-232. The CC
mode parameters are discussed in the following paragraphs.
Figure 2-1. Constant Current Mode
Ranges
Current may be programmed in either of two overlapping ranges, a Low range and a High range. The low range
provides better resolution at low current settings. The range can be set at the front panel or via the GPIB
(CURR:RANG command). When you program a current value, the electronic load automatically selects the
range that corresponds to the value that you program. If the value falls in a region where ranges overlap, the
electronic load selects the Low range. If the present input setting is outside the Low range, the electronic load
will automatically adjust the input setting to the highest value available in the Low range. If you subsequently
program an input value that is outside the Low range, an OUT OF RANGE message will appear on the front
panel display.
Immediate Current Level
The current level can be set at the front panel or via the GPIB (CURR command). If the CC mode is the active
mode, the new setting immediately changes the input at a rate determined by the slew setting (described below).
If the module is not in the CC mode, the new setting is saved for use when the mode is changed to CC.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
Triggered Current Level
A current level can be preset (stored in the electronic load) allowing the input to be updated when a trigger is
received instead of immediately as described above.
If the CC mode is the active mode, the preset current level will become the actual value and the input will be
updated when a trigger occurs. If the CC mode is not the active mode, the preset current level will become the
actual value when a trigger occurs but there will be no effect on the input until the CC mode becomes active.
Once a level is triggered, subsequent triggers will have no effect on the input unless another CURR:TRIG
command is sent. The trigger sources available to the electronic load are described later in this chapter. The
electronic load has a status reporting capability to keep track of pending triggers and other operating conditions.
The status reporting capability is described in detail in the Programming Guide.
Transient Current Level
The transient current level can be set at the front panel or via the GPIB. The transient current level can be
higher or lower than the main current level. The module input will switch between the main level and the
transient level when transient operation is turned on.
Software Current Limit
The electronic load allows the user to set a current limit (0 to 102% of full scale) for each module via the GPIB
(CURR:PROT command) which will shut down the input if the current limit is exceeded beyond a
programmable time delay. Note that the software current limit is in effect for any mode of operation (not just
the CC mode). The software current limit feature is described later in this chapter under Protection Features.
Slew Rate
The current slew rate determines the rate at which the input current to a module changes to a new programmed
value. Current slew rates are programmed in amperes per second. Slew rates can be set at the front panel or via
the GPIB (CURR:SLEW command). The programmed slew rate remains in effect for the immediate, triggered,
and transient level changes previously described.
Any slew rate can be programmed provided that it falls between the fastest and the slowest slew rates shown in
Figure 2-8A. If a value is programmed that is outside these limits, the module will automatically adjust the
programmed value to either the fastest or the slowest slew rates shown in the figure.
Constant Resistance (CR) Mode
In this mode, the module will sink a current linearly proportional to the input voltage in accordance with the
programmed resistance (see Figure 2-2). The CR mode can be set at the front panel or via the GPIB
(MODE:RES command). The CR mode parameters are described in the following paragraphs.
Ranges
Resistance may be programmed in any of four overlapping ranges. The range can be set at the front panel or via
the GPIB (RES:RANG command). When you program a resistance value, the electronic load automatically
selects the range that corresponds to the value that you program. If the value falls in a region where ranges
overlap, the electronic load selects the range with the highest resolution. If the present input setting is outside the
range that you select, the electronic load will automatically adjust the input setting to the closest available value
within the newly selected range. If you subsequently program an input value that is outside the newly selected
range, an OUT OF RANGE message will appear on the front panel display.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
Figure 2-2. Constant Resistance Mode
Immediate Resistance Level
The resistance level can be set at the front panel or via the GPIB (RES command). If the CR mode is active, the
new setting immediately changes the input at a rate determined by the voltage or current slew setting (see
description below). If the module is not in the CR mode, the new setting is saved for use when the mode is
changed to CR.
Triggered Resistance Level
A resistance level can be preset (stored in the electronic load) allowing the input to be updated when a trigger is
received instead of immediately as described above.
If the CR mode is active, the preset resistance level will become the actual value and the input will be updated
when a trigger occurs. If the CR mode is not the active mode, the preset resistance level will become the actual
value when a trigger occurs but there will be no effect on the input until the CR mode becomes active. Once a
level is triggered, subsequent triggers will have no effect on the input unless another RES:TRIG command is
sent.
Transient Resistance Level
The transient resistance level can be set at the front panel or via the GPIB (RES:TLEV command). The
transient level and the main level are used in transient operation, which is described later in this chapter.
Slew Rate
The resistance slew rate determines the rate at which the input resistance to a module changes to a new
programmed value. Resistance slew rates are programmed in ohms per second. Slew rates can be set at the front
panel or via the GPIB (RES:SLEW command). The programmed slew rate remains in effect for the immediate,
triggered, and transient level changes previously described.
Constant Voltage (CV) Mode
In this mode, the module will attempt to sink enough current to control the source voltage to the programmed
value (see Figure 2-3). The module acts as a shunt voltage regulator when operating in the CV mode. The CV
mode can be set at the front panel or via the GPIB (MODE:VOLT command). The CV mode parameters are
described in the following paragraphs.
Ranges
Voltage may be programmed in either of two overlapping ranges, a low range and a high range. The low range
provides better resolution at low voltage settings. The range can be set at the front panel or via the GPIB
24
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
(VOLT:RANG command). When you program a voltage value, the electronic load automatically selects the
range that corresponds to the value that you program. If the value falls in a region where ranges overlap, the
electronic load selects the Low range. If the present input setting is outside the Low range, the electronic load
will automatically adjust the input setting to the highest value available in the Low range. If you subsequently
program an input value that is outside the Low range, an OUT OF RANGE message will appear on the front
panel display.
Figure 2-3. Constant Voltage Mode
Immediate Voltage Level
The voltage level can be set at the front panel or via the GPIB (VOLT command). If the CV mode is active, the
new setting immediately changes the input at a rate determined by the voltage slew setting. If the module is not
in the CV mode, the new setting is saved for use when the mode is changed to CV.
Triggered Voltage Level
The voltage level can be preset (stored in the electronic load) allowing the input to be updated when a trigger is
received instead of immediately as described above.
If the CV mode is the active mode, the preset current level will become the actual value and the input will be
updated when a trigger occurs. If the CV mode is not the active mode, the preset current level will become the
actual value when a trigger occurs, but there will be no effect on the input until the CV mode becomes active.
Once a level is triggered, subsequent triggers will have no effect on the input unless another VOLT:TRIG
command is sent.
Transient Voltage Level
The transient voltage level can be set at the front panel or via the GPIB (VOLT:TLEV command). The module
input will switch between the main level and the transient level when transient operation is turned on. The
transient voltage level determines the higher voltage level.
Slew Rate
The voltage slew rate determines the rate at which the input voltage to a module changes to a new programmed
value. Voltage slew rates are programmed in volts per second. Slew rates can be set at the front panel or via the
GPIB (VOLT:SLEW command). The programmed slew rate remains in effect for the immediate, triggered, and
transient level changes previously described.
Any slew rate can be programmed provided that it falls between the fastest and the slowest slew rates shown in
Figure 2-8B. If a value is programmed that is outside these limits, the module will automatically adjust the
programmed value to either the fastest or the slowest slew rates shown in the figure.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
Transient Operation
Transient operation enables the module to periodically switch between two load levels, as might be required for
testing power supplies. A power supply's regulation and transient characteristics can be evaluated by monitoring
the supply's output voltage under varying combinations of load levels, frequency, duty cycle, and slew rate.
Transient operation can be turned on and off at the front panel or via the GPIB (TRAN ON and TRAN OFF
commands). Before you turn on transient operation, you should set the desired mode of operation as well as all
of the parameters associated with transient operation. Transient operation may be used in the CC, CR, or CV
modes and can be continuous, pulsed, or toggled.
Generates a repetitive pulse stream the toggles between two load levels.
Continuous
Pulse
Generates a load change that returns to its original state after some time period.
Generates a repetitive pulse stream that toggles between two load levels. Similar to
Continuous mode except that the transient points are controlled by explicit triggers instead of
an internal transient generator.
Toggled
Continuous Transient Operation
In continuous operation, a repetitive pulse train switches between two load levels. In the front panel, the
transient commands are located under the TRAN key. Continuous transient operation is selected via the GPIB
using the TRAN:MODE CONT command.
The two load levels in the transient operation are the previously described main level (immediate or triggered)
and transient level for current, resistance, or voltage. The rate at which the level changes is determined by the
slew rate (see slew rate descriptions for CV, CR, or CV mode as applicable). In addition, the frequency and
duty cycle of the continuous pulse train are programmable. The frequency can be set from 0.25 to 10000 Hz at
the front panel or via the GPIB (TRAN:FREQ command). The duty cycle can be set from 3% to 97% (0.25 Hz
to 1 kHz) or from 6% to 94% (above 1 kHz) at the front panel or via the GPIB (TRAN:DCYC command).
Pulsed Transient Operation
Pulsed transient operation is similar to continuous operation with the following exceptions:
a. In order to get a pulse, an explicit trigger is required. The trigger can be an external trigger signal
received via the TRIGGER input on the rear panel, the TRIG:SOUR function, the *TRG or TRIG
commands, the ac line, the internal timer signal, or the front panel Trigger key.
b. One pulse results from each trigger. Therefore, frequency cannot be programmed. The main level,
transient level, and slew rate are programmed as described for continuous operation. The pulse width is
programmable from 0.00005 to 4 seconds via the GPIB (TRAN:TWID command).
Toggled Transient Operation
Toggled transient operation causes the module input to alternate between two pre-defined levels as in
continuous operation except that the transient points are controlled by explicit triggers instead of the internal
transient generator. As in pulsed transient operation, the trigger signal can be an external trigger signal, the
GPIB GET function, the *TRG command, the TRIG command, or the ac line or internal timer signals.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
List Operation
List mode lets you generate complex sequences of input changes with rapid, precise timing, which may be
synchronized with internal or external signals. This is useful when running test sequences with a minimum
amount of programming overhead.
You can program up to 50 settings (or steps) in the list, the time interval (dwell) that each setting is maintained,
the number of times that the list will be executed, and how the settings change in response to triggers. All listed
data is stored in a non-volatile memory when the *SAV command is executed. This means that the programmed
data for any list will be retained when the electronic load is turned off. Note that lists data can only be saved in
nonvolatile memory locations 0, 7, 8, or 9. List data will not be saved in other memory locations. Use the *RCL
command to recall the saved state.
List steps can be either individually triggered, or paced by a separate list of dwell times which define the
duration of each step. Therefore, each of the up to 50 steps has an associated dwell time, which specifies the
time (in seconds) that the input remains at that step before moving on to the next step. See chapter 5 for detailed
information about programming lists from the front panel.
Triggered Operation
The electronic load has various triggering modes to allow synchronization with other test equipment or events.
The triggering circuits are located in the mainframe, and all modules receive the trigger simultaneously
(although each module is programmed individually as to what operation, if any, will be triggered. As described
previously, triggering can be used for the following applications:
Transfers all pending preset levels to the actual level. For the presently active
mode, the new level appears at the input. For the modes which are not presently
active, the preset levels will not take effect at the input until the applicable
mode becomes active.
Triggering a preset level
Generates a transient pulse of programmable width when pulsed transient
operation is in effect.
Triggering a transient pulse
Toggling
Changes the input between the main level and the transient level when toggled
transient operation is in effect.
Triggers can be sent from the front panel by pressing the Trigger key. However you must first initiate the
trigger function by executing the TRIG:IMMED command located in the Trigger Control menu.
Three triggering methods are available over the GPIB: the GET function, the *TRG common SCPI command,
and the TRIG subsystem SCPI command (refer to Programming Guide). The SCPI TRIG subsystem allows
you to select either the ac line frequency, internal timer, or TRIG command as the trigger source. There is also
a TRIGGER connector on the rear panel for external trigger inputs.
*TRG and the TRIG command are both synchronous with other commands; that is, the modules are not
triggered until pending operations are completed. GET, external triggers, ac-line triggers, and internal-timer
triggers are all asynchronous; that is, the modules are triggered as soon as the trigger signal is received.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
If the ac line is selected via the GPIB as the trigger source, triggers will be generated once for each cycle of ac
input power. An ac line frequency of 60 Hz produces a trigger period of 16.67 ms; 50 Hz line frequency
produces a trigger period of 20 ms.
The rear-panel TRIGGER connector also provides a trigger output signal. This signal is generated
synchronously with the trigger signal sent by the mainframe to the modules. The trigger output signal can be
used to trigger an external device such as an oscilloscope, DVM, or another electronic load mainframe.
The electronic load has a status reporting capability to keep track of trigger operations. Refer to 'Status
Reporting' in the Programming Guide.
Slew Rate and Minimum Transition Time
Slew rate is defined as the change in current, resistance, or voltage over time. A programmable slew rate allows
a controlled transition from one load setting to another to minimize induced voltage drops on inductive power
wiring, or to control induced transients on a test device (such as would occur during power supply transient
response testing).
In cases where the transition from one setting to another is large, the actual transition time can be calculated by
dividing the voltage or current transition by the slew rate. The actual transition time is defined as the time
required for the input to change from 10% to 90% or from 90% to 10% of the programmed excursion. In cases
where the transition from one setting to another is small, the small signal bandwidth of the load limits the
minimum transition time for all programmable slew rates. Because of this limitation, the actual transition time is
longer than the expected time based on the slew rate, as shown in Figure 2-7.
Voltage,
Current, or
Resistance
Change
Slew Rate
100%
90%
10%
0%
Time
Expected Time
Actual Time
Figure 2-7. Risetime Transition Limitation
28
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
Therefore, both minimum transition time and slew rate must be considered when determining the actual
transition time. This is shown in Figure 2-8, which shows the minimum transition time for a given slew rate as a
horizontal line, and at about a 13.3% or greater load change, the slew rate increases from the minimum
transition time to the Maximum transition time at a 100% load change. The actual transition time will be either
the minimum transition time, or the total slew time (transition divided by slew rate), whichever is longer.
Use the following formula to calculate the minimum transition time (MinTT) for a given slew rate:
__________8__________
MinTT (in seconds) =
slew rate (in amps/second)
Use the following formula to calculate the maximum transition time (MaxTT) for a given slew rate:
__________60__________
MaxTT (in seconds) =
slew rate (in amps/second)
NOTE:
In voltage mode, all minimum transition times are based on a low-capacitance current source.
These transition times are affected by capacitive loading of the inputs. For example, a
capacitance of 2.2 microfarads increases the 85 microsecond minimum transition time (shown
in the table) to 110 microseconds.
Maximum
Maximum
Minimum
Transition
Time
Minimum
Transition
Time
60000
µ
s
60000
µ
s
8000
µ
s
8000
µ
s
6000
µ
s
6000
µ
s
800µ
s
800µ
s
∆
Time
∆
Time
600µ
s
600
µ
s
s
120µ
85µ
s
80µ s
60µ
s
16µ
s
12µs
13.3%
16.7%
50%
100%
13.3%
16.7%
50%
100%
5%
5%
∆
Current (% of full scale)
∆
Voltage (% of full scale)
A.
B.
Figure 2-8. Transition Time Slew Rate Examples
29
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
Input Control
Short On/Off
A module can simulate a short circuit at its input by turning the load on with full-scale current. The short circuit
can be toggled on/off at the front panel using the SHORT command in the Func menu, or via the GPIB
(INPUT:SHORT ON|OFF command). The short on/off change uses the slew rate setting of the active mode
and range.
The actual value of the electronic short is dependent on the mode and range that are active when the short is
turned on. In CV mode it is equivalent to programming zero volts. In CC mode it is equivalent to programming
full-scale current for the present current range. In CR mode it is equivalent to programming the minimum
resistance for the present resistance range.
Note that turning the short on in CV mode may cause the load to draw so much current that the software current
limit operates, which may turn the input off.
Turning the short circuit on does not affect the programmed settings, and the load input will return to the
previously programmed values when the short is turned off.
Input On/Off
A module's input can be toggled on/off at the front panel, or via the GPIB (INPUT ON|OFF command). The
input on/off change does not use the slew rate setting so the input will change at the maximum slew rate.
Turning the input off (zero current) does not affect the programmed settings. The input will return to the
previously programmed values when the input is turned on again. Note that the Input On/Off command
supersedes the mode commands and Short On/Off command.
Protection Features
Each load module includes the following protection features:
•
•
•
•
•
Overvoltage.
Overcurrent (hardware and software).
Overpower.
Overtemperature.
Reverse Voltage.
The appropriate bit(s) in the mainframe's status registers are set when any of the above protection features are
active. Also, the Prot annunciator comes on and the front-panel alphanumeric display indicates which
condition(s) have been detected. For example, if an overtemperature (OT) condition has been detected causing a
module's input to be turned off (protection shutdown, PS), the display will indicate "PS OT".
Resetting Latched Protection
All of the protection features latch (remain set) when they are tripped, except for the hardware overcurrent and
reverse voltage. The latched protection features can be reset via the GPIB (*RST or INP:PROT:CLE
commands) or at the front panel. Of course, the condition that caused the protection feature to trip must be
removed or it will trip again as soon as it is reset.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
Caution
To protect the electronic load from possible damage, the input voltage must not exceed the
maximum input voltage rating specified in the module-specific pages supplied with each
module. Never apply the ac line voltage to a module's input connectors.
Overvoltage
The overvoltage protection circuit is set at a predetermined voltage level, which cannot be changed. If the
overvoltage circuit has tripped, the module will attempt to limit the voltage level by drawing current from the dc
source. The module limits the value of current drawn such that the resulting power is within the power rating.
The overvoltage (OV) and voltage fault (VF) status register bits are set when the OV condition occurs, and will
remain set until they are reset as previously described.
An overvoltage condition does not cause the module's input to be turned off. However, a Fault signal (pin A6)
output at the module's rear-panel control connector will indicate when either an overvoltage condition or a
reverse voltage condition has occurred. The Fault signal is latched true (high TTL level) when the VF bit in the
status register goes true. The Fault output signal can be used to trip an external circuit breaker or control a relay
in order to disconnect the electronic load input from the source it is testing when an overvoltage or a reverse
voltage condition occurs.
Overcurrent
The electronic load includes both hardware and software overcurrent protection features.
Hardware. When operating in the CR or CV mode, it is possible for a module to attempt to sink more current
than it is rated for. Under this condition, the load current will be limited by a current limit circuit, which is set at
a value slightly above the current rating of the module. It protects both the electronic load and the device under
test from operating too far beyond specified limits. The hardware current limit circuit does not turn the module's
input off. The overcurrent (OC) bit in the status register is set when an OC condition occurs, and is reset when
the OC condition is removed.
Software. In addition to the hardware overcurrent protection circuit, the electronic load allows the user to
define a current protection limit in software which will shut down a module's input if the limit is exceeded. This
feature can only be programmed via the GPIB. It is turned on/off using the CURR:PROT:STATE ON|OFF
command. The software current limit level (in amps) is set using the CURR:PROT command. A
programmable delay (in seconds) before trip is also provided with the CURR:PROT:DEL command. If the
software overcurrent limit is exceeded and persists beyond the specified delay time, the module is turned off.
Also, for these conditions, the OC and PS (protection shutdown) status register bits are set and will remain set
until the OC condition is removed and the bits are reset as previously described.
Overpower
The power-limit boundary is set by software that monitors the input current and voltage. If the input power
exceeds the power limit, the load module sets the overpower status bit, which will reset if the overpower
condition ceases. If the overpower condition persists for 3 seconds, the load module's input circuit turns off, and
the OP and PS status bits are both latched on. The input circuit remains off, and the OP and PS status bits
remain set, until protection clear occurs. Of course, if the overpower condition is not corrected, the load will
turn off again.
Overtemperature
Each module has an overtemperature (OT) protection circuit, which will turn off the input if the internal
temperature exceeds safe limits. If the OT circuit activates, the OT and PS status register bits are set and will
31
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
remain set until they are reset. If the OT condition still exists when the reset is executed, the module's input will
remain off. You must wait until the module cools down before you can reset the OT circuit. The fan(s) will
continue to operate to cool the unit as quickly as possible.
Reverse Voltage
Caution
This feature protects the load module in case the input dc voltage lines are connected with the
wrong polarity. If a reverse voltage (LRV or RRV) condition is detected, turn off power to the
dc source and the electronic load and make the correct connections.
The electronic load conducts reverse current when the polarity of the dc source connection is incorrect. The
reverse voltage (LRV for local reverse voltage; RRV for remote reverse voltage) and voltage fault (VF) bits in
the status register are set when reverse voltage is applied. When the reverse voltage is removed the LRV and
RRV bits are cleared. However, the VF bit remains set until it is reset. As previously described, the Fault
output signal at the control connector tracks the state of the VF bit. The Fault signal can be used to control an
external relay in order to disconnect the module from the dc source if an RV condition occurs. This feature also
detects reverse voltage at the Sense terminals.
Reading Remote Programming Errors
Remote programming errors can be read via the GPIB (SYST:ERR? query) or at the front panel. The Err
annunciator indicates when remote programming errors have occurred. The errors are negative numbers
grouped into blocks of 100 as follows:
-lxx
Command errors
-2xx Execution errors
-3xx Device-specific errors
-4xx Query errors
The SYST:ERR? query reads back the errors in the order in which they occurred (the error queue can hold up
to 30 entries). Once the error is read back it is removed from the list. A value 0 indicates there is no error; and
0 will be returned when all errors in the list have been read. Pressing the Error key displays just the error
number. The SYST:ERR? query returns the error number and a short description of the error to the computer.
Refer to Appendix B in the Programming Guide.
Local programming errors generated by front panel operations are not put into the error list, but are immediately
put on the electronic load's front panel display; e.g., 'OUT OF RANGE'.
Status Reporting
The electronic load incorporates a status reporting capability. Various status conditions within the electronic
load can be reported using this capability. The user determines which condition(s) will be reported. Chapter 5
of the Programming Guide describes each of the status registers in the electronic load. (These registers,
including the channel status registers, are all maintained in the mainframe.) Notice that the same information is
available in both the channel status and questionable status registers, but the channel registers are organized by
channel, and the questionable registers are organized by fault. Therefore, depending on which channels and/or
faults are most critical in your application, you can use one branch to localize selected faults quickly, and use
the other branch for broader fault reporting. By knowing that only a particular fault (questionable branch) or a
particular channel (channel branch) is enabled to initiate a service request, you can eliminate the need to read
one or more registers to locate a fault.
32
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
Saving and Recalling Settings
The electronic load has internal registers in which settings (mode, current, voltage, resistance, slew, transient
level, etc.) for each module can be saved. By saving settings and recalling them later you can save
programming time.
The present settings for all channels are saved in the specified register (0 to 9) at the front panel or via the GPIB
(*SAV command). All of the settings are saved in the specified location in the mainframe's memory. Settings
saved in locations 1 through 6 will be lost when ac line power is cycled. However, the *SAV 0 command will
cause the settings to be stored in a non-volatile memory; and, the next time the electronic load is turned on, these
settings will become the power-on settings. In addition, locations 7 through 9 are also saved in non-volatile
memory. These locations are used to store lists.
You can recall the saved settings from the specified register (0 to 9) at the front panel or via the GPIB (*RCL
command). All of the parameters for each module which were saved by the *SAV command are set to the
saved values. At power-on, the electronic load automatically executes a *RCL 0, which recalls the values saved
in location 0 of non-volatile memory.
You can recall the factory default settings at the front panel or via the GPIB (*RST command).
Remember that Save and Recall operate on all channels, not just the presently addressed or selected channel.
External Control Signals
Each module has a 14-pin connector mounted on its rear panel. These signals are described in the following
paragraphs. See Chapter 3 for connection details.
Remote Sensing
The remote sensing inputs, + S and - S, can be used in CV or CR modes. By eliminating the effect of the
inevitable voltage drop in the load leads, remote sensing provides greater accuracy by allowing the load to
regulate directly at the source's output terminals, as well as measure the voltage there.
Monitor Outputs
The current monitor (pin A1) and voltage monitor (pin A2) output signals indicate the input current and voltage.
A 0-to-+10V signal at the appropriate output indicates the zero-to-full scale input current or voltage. An
external DVM or oscilloscope can be connected to monitor the input voltage and current.
External Programming Input
CC and CV modes can be programmed with a signal (ac or dc) connected to the external programming (pin A4)
input. A 0-to-10V external signal corresponds to the 0-to-full scale input range in CV mode or in CC mode.
The external programming signal is combined with the value programmed via the GPIB or the front panel, so
that, for example, a programmed value of one-half full scale and a 5-volt external programming input would
produce a full-scale value at the input.
Figure 2-9 shows the input waveform that would result from the following setup:
CC Mode
60A Range
33
Download from Www.Somanuals.com. All Manuals Search And Download.
2 - General Information
20A Input (programmed via GPIB or front panel)
± 1V (2 V peak-peak) 1 kHz external programming signal
The external programming signal (+ 1 and - 1 volt) corresponds to + 6 and - 6 amps at the input (1 volt external
programming input = 1/10 full scale). Therefore, the input varies 6A at the 20A level.
Fault
The Fault signal becomes active if an overvoltage or reverse voltage occurs at the input, as described in the
Protection Features paragraphs.
Figure 2-9. External Programming Example
Port On/Off
Port is a general purpose output port that can be used to control an external device such as a relay for power
supply test purposes. The output is toggled on and off via the GPIB (PORT0 ON | OFF command). It cannot
be controlled from the front panel.
The Port output signal is a TTL compatible signal that becomes active (high level) when the PORT command is
programmed ON and becomes inactive (low level) when the PORT command is programmed OFF.
Input Measurements
Each module's input current, voltage, and power are continuously measured at the front panel.
With remote control in effect, a module may be instructed to measure its dc input voltage, current, or power by
sending the appropriate query command (e.g. MEAS:CURR). The results will be read back when the
electronic load is addressed to talk. Voltage and current measurements are performed with approximately 16-bit
resolution of full-scale ratings. Power is computed from this information. Voltage and current are measured
simultaneously, so that power data is correct.
All measurements are performed by digitizing the instantaneous input voltage or current for a defined number of
samples and sample interval, storing the results in a buffer, and then calculating the measured result. Many
parameters of the measurement are programmable. These include the number of samples, the time interval
between samples, and the method of triggering. Note that there is a tradeoff between these parameters and the
speed, accuracy, and stability of the measurement in the presence of noise. This is described in the
Programming Guide.
34
Download from Www.Somanuals.com. All Manuals Search And Download.
General Information - 2
As shipped from the factory, front panel measurements for input voltage and current are calculated from a total
of 1000 readings taken at a 10 microsecond sampling rate. There are no trigger controls for front panel
measurements. However, you can program both the sampling rate and the number of data points in each front
panel measurement using commands in the Sense menu. With this flexibility, measurement accuracy can be
improved for waveforms with frequencies as low as several Hertz. The sample buffer size may be varied from 1
to 4096 data points. The sampling rate may be varied from 0.000001 seconds to 0.032 seconds. Values are
rounded to the nearest 10-microsecond interval.
DC Measurements
DC voltage and current is measured by acquiring a number of readings at the selected time interval, applying a
windowing function to the readings, and averaging the readings. Windowing is a signal conditioning process
that reduces the error in dc measurements made in the presence of periodic signals such as line ripple. At
power-on and after a *RST command, the interval and number of sample points used for the measurement
acquisition is set. Ripple rejection is a function of the number of cycles of the ripple frequency contained in the
acquisition window. More cycles in the acquisition window results in better ripple rejection. In addition, the
speed of the measurement can be increased by reducing the number of sample points.
To measure the dc input voltage or current using the front panel, press the Meter key. The display indicates the
present voltage and current reading. Scroll to the appropriate measurement. To measure the voltage or current
over the GPIB, use the MEAS:VOLT? or MEAS:CURR? command.
RMS Measurements
To measure the rms of the input voltage or current, press the Meter key and scroll to the V RMS or A RMS
menu commands. To measure the rms voltage or current over the GPIB, use MEAS:VOLT:ACDC? or
MEAS:CURR:ACDC?
This returns the total rms measurement, including the dc portion.
Minimum and Maximum Measurements
To measure the maximum or minimum voltage or current, press the Meter key and scroll to the
V MAX, V MIN, A MAX, or A MIN menu commands. Minimum returns the lowest value in the array;
maximum returns the highest value in the array. To measure the maximum/minimum voltage or current over the
GPIB, use MEAS:VOLT:MAX?/MEAS:VOLT:MIN? or MEAS:CURR:MAX?/MEAS:CURR:MIN?
Power Measurements
To measure the average, maximum, or minimum power, press the Meter key and scroll to the
WATTS, W MAX, or W MIN menu commands. Power is calculated from the latest voltage and current
measurements. Use MEASure:POWer? to measure power over the GPIB.
Measurement Ranges
The electronic load has two measurement ranges. The commands that control the measurement ranges are
located in the front panel Sense menu. To change measurement ranges, scroll to the S:C:RNG or the S:V:RNG
commands and enter a value that falls within the range that you wish to set. The electronic load will pick the
range with the best resolution for that value. The corresponding SCPI range commands are:
SENS:CURR:RANG and SENS:VOLT:RANG.
35
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Installation
Inspection
Damage
When you receive your electronic load, inspect it for any obvious damage that may have occurred during
shipment. If there is damage, notify the shipping carrier and nearest Agilent Sales and Support Office
immediately. The list of Agilent Sales and Support Offices is at the back of this guide. Warranty information is
printed in the front of this guide.
Packaging Material
Until you have checked out the electronic load, save the shipping carton and packing materials in case the unit
has to be returned. If you return the electronic load for service, attach a tag identifying the model number and
the owner. Also include a brief description of the problem.
Items Supplied
The following user replaceable items are included with your electronic load. Some of these items are installed
in the unit.
Table 3-1. Items Supplied
Item
Part Number
Description
Power Cord
Contact nearest Agilent
Sales and Support Office
A power cord appropriate for your location.
Trigger/Digital Connector
Output Connector
0360-2693
A 6-pin connector is provided for input and
output trigger and digital signals.
0360-2870
A 14-pin control connector is provided for
connecting remote sense leads.
Feet
5041-8801
5964-8196
Feet for bench mounting.
User's Guide
Contains installation, checkout, and front panel
information.
Programming Guide
5964-8198
Contains detailed GPIB programming
information.
Cleaning
Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean
internally.
WARNING: To prevent electric shock, unplug unit before cleaning.
37
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
Installing the Modules
Procedure
1. With the mainframe off, disconnect the power cord and remove the top cover by loosening the screws with a
flat-bladed screwdriver.
2. Remove any packing material from inside the mainframe.
3. Grasp the module using the quarter-turn locking fastener and the input connectors. This reduces the
possibility of damage to static sensitive components on the pc board.
4. Start installing the modules in the slot next to the GPIB board (see figure 3-1).
Figure 3-1. Module Installation Diagram
5. Lock the module in place using the quarter-turn locking fastener and the rear panel thumbscrew. Hand-
tighten only.
38
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
6. Connect the ribbon cable to the adjacent connector pins in the GPIB board (or adjacent module). Make sure
the connectors are properly seated.
7. If applicable, install each module in the slot next to the previous module in the same manner (step 3 through
6).
8. Replace the top cover after all modules are installed.
9. Reconnect the power cord.
Channel Number
The channel number of a specific module is determined by the location of that module in relation to the GPIB
board. For example, the module next to the GPIB board is always channel number one. Numbering continues
sequentially so that the module furthest from the GPIB board is the highest numbered channel in your system.
Figure 3-2 shows the channel assignments for an Agilent N3300A electronic load mainframe containing a
N3304A single-width module and a N3305A double-width module. One channel number is automatically
assigned to each module according to the order in which it was installed in the mainframe. The maximum
number of channels is six for N3300A mainframes, and two for N3301A mainframes.
CHAN 2
CHAN 1
SINGLE
WIDTH
DOUBLE
WIDTH
GP-IB
BOARD
Figure 3-2. Channel Number Example
Location
The outline diagram in figure 3-3A and figure 3-3B gives the dimensions of your electronic load. The electronic
load must be installed in a location that allows sufficient space at the sides and back of the unit for adequate air
circulation (see Bench Operation).
39
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
Figure 3-3A. N3300A Outline Diagram
Figure 3-3B. N3301A Outline Diagram
mm
mm
mm
mm
mm
Figure 3-3C. N3300A Connector Diagram
40
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
Bench Operation
A fan cools the electronic load by drawing air through the top and sides and exhausting it out the back.
Minimum clearances for bench operation are 1 inch (25 mm) along the sides.
Do not block the fan exhaust at the rear of the unit.
Rack Mounting
The N3300A electronic load can be mounted in a standard 19-inch rack panel or cabinet. Rack mount kits are
available as Option 908 and 909 (with handles). Support rails are also required for rack mounting. These are
normally supplied with the cabinet and are not included with the rack mount options.
The N3301A electronic load can also be mounted in a standard 19-inch rack panel or enclosure using an Option
908 rack mount kit (see Figure 3-4). A rack mount kit for joining two half-rack units is available as Option 800.
Option 800 must also be used if you are mounting other instruments next to a N3301A electronic load. Support
rails are also required for rack mounting. These are normally supplied with the cabinet and are not included
with the rack mounting options.
If you are installing equipment on top of your electronic load in the cabinet, use a filler panel above the unit to
ensure adequate space for air circulation. A 1U panel (EIA Standard RS-310-C) as shown in Figure 3-4 is
sufficient. If your cabinet has a circulation fan, avoid installing the Electronic Load too close to the cabinet fan.
The cabinet fan may restrict the airflow required through the Electronic Load.
Figure 3-4. Rack Installation
41
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
Input Connections
Power Cord
1. Connect the power cord to the IEC 320 connector on the rear of the unit. If the wrong power cord was
shipped with your unit, contact your nearest Agilent Sales and Support Office (refer to the list at the back of
this guide) to obtain the correct cord. See Figure 3-5 for the part number and ordering options.
Figure 3-5. Power Cord Configurations
Manually-Tightened Connectors
The standard manually-tightened input connectors are located on the rear panel and are used for connecting
unterminated wires directly to the back of the modules.
Two screw-down connectors (+ and -) are provided on each module for connecting the input wires to the
electronic load (see figure 3-6a). Connections are made as follows:
42
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
Figure 3-6a. Manual Connector
1. Strip the back wire insulation as indicated
Wire Size
Strip back
AWG 4
6 mm (0.65 in)
13 mm (0.5 in)
10 mm (0.4 in)
AWG 6 or 8
AWG 10 or smaller
2. AWG 4 is the maximum wire size. Stranded copper wire size, AWG 6 or 8 is the recommended wire. If
you are connecting more than one wire on each connector, twist the wires to ensure a good contact when the
adjustment knob is tightened.
3. Insert the wire into the connector.
WARNING
To prevent accidental contact with hazardous voltages, do not extend the wire beyond the
contact area inside the input connector.
4. Hand-tighten the adjustment knob to secure the wire in the connector. If you are using a slotted
screwdriver, tighten the knob to 8 in-lb (90 N-cm) for a secure connection.
Caution
Do not use lubricants or contact cleaners on the connectors. Certain chemical agents can
damage the LEXAN material of the connector, causing the part to fail.
8mm Screw Terminal Connector (option UJ1)
The 8mm screw terminal connectors (option UJ1) are located on the rear panel and are used for connecting
wires that are terminated by wire lugs to the load modules.
Two 8mm-diameter bolts (+ and −) are provided on each module for connecting the input wires (see figure 3-
6b). Connections are made as follows:
43
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
conical
washer
spacer
(0380-4835)
(3050-1924)
insert
tabs
4 AWG
wire max.
cover
(5040-1736)
insulated
wire lug
spacer
(AMP 52266-3)
connector
assembly
with bolt
breakaway
tab
(5040-1739)
tighten
cover
screw
spacer
safety cover secured
wires exiting at bottom
wires exiting at top
Figure 3-6b. 8mm Screw Terminal Connectors
1. Attach a connecting lug with an insulated ring terminal to the input wires. Connecting lugs must have an
opening of at least 8mm in diameter. The recommended wire lug for connecting AWG 4 wire to an 8mm
2. Wire sizes smaller than the maximum of AWG 4 may be used, provided the wire lugs have the required
8mm diameter opening.
Caution
To prevent damaging the wiring insulation, which may result in shorting the input, use the
supplied spacer to offset the input wires from each other.
3. Attach the wire lugs to the connector. Place the conical washer between the bolt and the wire lug (refer to
the exploded view in the upper left corner of figure 3-6b). The wires can exit either out of the top or out of
the bottom. Place the supplied spacer under the appropriate wire lug to prevent interference between the
wires. If the wires exit out of the top, place the spacer on the bottom terminal. If the wires exit out of the
bottom, place the spacer on the top terminal.
4. Tighten the bolts to secure the wires to the connectors. Torque the bolts to between 20 and 25 in-lbs
(between 225 and 280 N-cm) for a secure connection.
5. Remove the appropriate breakaway tab and install the safety cover over the connectors. Hand-tighten the
Torx T10 cover screw.
Wire Considerations
WARNING
To satisfy safety requirements, load wires must be heavy enough not to overheat while carrying
the short-circuit output current of the device connected to the electronic load. Refer to Table 3-
2 for the ampere capacity of various stranded wire sizes.
Input connections are made to the + and − connectors on the back of each module. A major consideration in
making input connections is the wire size. The minimum wire size required to prevent overheating may not be
large enough to maintain good regulation. It is recommended that stranded, copper wires be used. The wires
should be large enough to limit the voltage drop to no more than 0.5 V per lead. Table 3-3 gives the maximum
load lead length to limit the voltage drop to the specified limit.
44
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
Table 3-2. Stranded Copper Wire Ampere Capacity
Wire Size
Ampacity
Notes:
1. Ratings for AWG-sized wires derived from MIL-W-
5088B. Ratings for metric-sized wires derived from
IEC Publication 335-1.
AWG
Cross Section
Area in mm2
22
20
5.0
8.33
10
15.4
13.5
19.4
16
31.2
25
40
32
55
40
75
63
100
135
2. Ampacity of aluminum wire is approximately84% of
that listed for copper wire.
0.75
1
18
16
14
12
10
8
3. When two or more wires are bundled together, ampacity
for each wire must be reduced to the following
percentages:
1.5
2.5
4
2 conductors 94%
3 conductors 89%
4 conductors 83%
5 conductors 76%
6
4. Maximum temperatures:
Ambient = 50° C
10
6
4
Conductor = 105° C
Table 3-3. Maximum Wire Lengths to Limit Voltage Drops
Resistivity Maximum Length in Meters (Feet) to Limit
Voltage Drop to 0.5 V or Less
Wire Size
Cross
Section
Area in
mm2
AWG
5 A 10 A
20 A
30 A
40 A
50 A
60 A
Ω/kft
Ω/km
22
20
18
16
14
12
10
8
16.15
10.16
(6)
2.5
(9.5) (4.5)
3.7
(15.5)
5.0
(24.5)
7.3
(3)
1.2
(1.5)
0.6
(2)
0.9
(3.5)
1.3
(1)
0.4
(1.5)
0.6
(2.5)
0.8
(4)
1.2
(6.5)
2.0
(0.77)
0.31
(1.23)
0.47
(2.0)
0.63
(3.1)
0.91
(4.9)
1.52
(7.9)
2.46
(0.62)
0.25
(0.98)
0.37
(1.57)
0.50
(2.49)
0.73
(3.46)
1.22
(6.29)
1.96
(0.52)
0.21
(0.82)
0.31
(1.30)
0.42
(2.07)
0.61
(3.30)
1.01
(5.24)
1.64
(8.34)
2.96
0.5
0.75
1
40.1
26.7
20.0
13.7
8.21
5.09
3.39
1.95
1.24
1.9
(7.5)
2.5
(12)
3.6
6.388
4.018
2.526
(6)
1.8
1.5
2.5
4
(39.5) (19.5)
(9.5)
3.0
(15.5) (10.5)
4.9
(25)
7.4
(39.5)
12.8
(63)
20
12.2
(62.5)
19.6
(100)
29
(159)
51
(252)
80
6.1
(31)
9.8
(50)
14.7
(79)
25
(126)
40
(201)
1.589
3.3
(17)
4.9
(27)
8.5
(40)
13.4
(68)
0.9994
0.6285
0.3953
0.2486
(12.5) (10.00)
3.69 2.95
(19.9) (15.91) (13.25)
6.41 5.13 4.27
(31.6) (25.30) (21.07)
10.08 8.06 6.72
6
10
16
6
4
(402)
(100)
(50.37) (40.23) (33.51)
45
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
Control Connector
A 14-pin connector and a quick-disconnect mating plug are provided on each module for connecting remote
sense leads, external V/I monitors, an external programming input, and external control lines (see figure 3-7).
The mating plug is packaged in an envelope that is included with the module.
Consistent with good engineering practice, all leads connected to the control connector should be twisted and
shielded to maintain the instrument's specified performance. Make all wire connections to the mating plug as
required before installing the connector in the module.
Not available
A1-A3
A4
Provides the common connection for the A5 and A6 pins.
A TTL-compatible output signal that becomes active (high level) when the PORT0 command is
programmed ON. This signal can be used to control an external device such as a relay for
shorting or disconnecting the module's input terminals or as a general purpose digital output
port. This signal powers up in the inactive (low-level) state.
A5
A TTL-compatible output (fault) signal that becomes active (high level) when an overvoltage,
reverse voltage condition or fault occurs. This signal powers up in the inactive (low-level) state.
A6
Provides the common connection for the external programming input (pin A8).
A7
A8
Connects an external programming input. The CC and CV mode can be programmed with a
0V-to-+10V signal (ac or dc). This signal can act alone or can be combined with values
programmed over the GPIB or RS-232. Thus, it is possible to superimpose an ac signal upon a
dc level.
Provides the common connection for the current and voltage monitor signals (pins A10 and
A11).
A9
Used to monitor the modules input current and voltage. A 0V-to-+10V signal at the appropriate
pin indicates the zero-to-full scale current or voltage. Pin A11 monitors current and pin A10
monitors voltage.
A11 and A10
Used to connect the remote sense leads to the power source. Pin +S connects the +S signal and
pin -S connects the -S signal. Remote sensing can only be used in CV and CR modes.
+S and -S
Figure 3-7. Control Connector
46
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
Sense Switch
A local/remote sense switch is provided on each module. Unless you are using remote sensing, make sure that
the sense switch is set to LCL (depressed). Remote sensing is used in certain applications to achieve greater
accuracy (refer to Remote Sense Connections for more information).
NOTE
If the sense switch is set to remote operation without having sense leads connected to the sense
inputs, the module will continue to work in the CC mode, but the input will turn off in CV and
CR modes. Voltage readback will not work in any mode.
Trigger and Digital Connections
A connector is provided on each mainframe for a ground reference of input and output trigger signals (see
Figure 3-8). The remote voltage sense, current and voltage monitor signals, analog programming input, fault
output pin and a programmable digital output pin are referenced to the trigger input.
Figure 3-8. Control Connector
A TTL-compatible input that responds to low-level external trigger signals. A trigger applied to
this input can be used to change settings (voltage, current, resistance, etc.), toggle between
settings in transient-toggle mode, or generate a pulse in transient-pulse mode. An external trigger
affects any module that has its external trigger input enabled by the TRIG:SOUR:EXT command.
TRIG IN
A TTL-compatible output signal that becomes active (low-level) whenever the electronic load is
triggered by a GPIB command or TRIG IN signal. This signal can be used to trigger external
equipment such as oscilloscopes, digitizers, or another electronic load.
TRIG
OUT
Provides the common connection for the trigger signals. This common is directly connected to
the chassis.
TRIG
GND
A programmable digital output line.
DIG 1
A second programmable digital output line.
DIG 2
Provides the common connection for the digital signals. This common is directly connected to
the chassis.
DIG GND
Consistent with good engineering practice, all leads connected to the connector should be twisted and shielded
to maintain the instrument's specified performance.
47
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
Computer Connections
The electronic load can be controlled through a GPIB interface or through an RS-232 interface.
GPIB Interface
Each electronic load has its own GPIB bus address, which can be set using the front panel Address key as
described in Chapter 5. GPIB address data is stored in a non-volatile memory. The electronic load is shipped
with its GPIB address set to 5.
Electronic loads may be connected to the GPIB interface in series configuration, star configuration, or a
combination of the two, provided the following rules are observed:
♦
♦
The total number of devices including the GPIB interface card is no more than 15.
The total length of all cables used is no more than 2 meters times the number of devices connected together,
up to a maximum of 20 meters. (Refer to Table 2-2 for a list of GPIB cables available from Agilent
Technologies.)
♦
♦
Do not stack more than three connector blocks together on any GPIB connector.
Make sure all connectors are fully seated and the lock screws are firmly finger-tightened.
RS-232 Interface
The electronic loads have an RS-232 programming interface, which is activated by commands located in the
front panel Address menu. All applicable SCPI commands are available through RS-232 programming. When
the RS-232 interface is selected, the GPIB interface is disabled.
The RS-232 connector is a DB-9, male connector as shown in figure 3-9. Adapters are available to connect the
electronic load to any computer or terminal with a properly configured DB-25 connector (see Table 2-2).
1 2 3 4 5
6 7 8 9
Figure 3-9. RS-232 Connector
Pin
1
2
3
4
Input/Output Description
-
Input
Output
-
no connection
Receive Data (RxD)
Transmit Data (TxD)
not applicable
5
6
Common
-
Signal ground
not applicable
7
8
9
Output
Input
-
Request to Send (RTS)
Clear to Send (CTS)
No connection
48
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
Application Connections
Local Sense Connections
Figure 3-10 illustrates a typical setup with module number 1 connected for constant current or constant
resistance operation. Local sensing is used in applications where lead lengths are relatively short, or where load
regulation is not critical. The sense switch must be set to LCL. Load leads should be bundled or tie-wrapped
together to minimize inductance.
N3300A
Figure 3-10. Local Sensing
Remote Sense Connections
Figure 3-11 illustrates a typical setup with module number 1 connected for remote sense operation. The remote
sense terminals of module 1 are connected to the output of the power supply. Remote sensing compensates for
the voltage drop in applications that require long lead lengths. It is only useful when module 1 is operating in
CV or CR mode, or when using voltage readback. The sense switch must be set to RMT. Load leads should be
bundled or tie wrapped together to minimize inductance.
Parallel Connections
Figure 3-12 illustrates how modules can be paralleled for increased power dissipation. Up to six modules can be
directly paralleled in CC or in CR mode. Modules cannot be paralleled in CV mode. Each module will
dissipate the power it has been programmed for. For example, if two modules are connected in parallel, with
module number 1 programmed for 10 A and module number 2 programmed for 20A, the total current drawn
from the source is 30 A.
49
Download from Www.Somanuals.com. All Manuals Search And Download.
3 - Installation
N3300A
Figure 3-11. Remote Sensing
N3300A
Figure 3-12. Parallel Operation
50
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation - 3
In Figure 3-12, all lead connections are terminated at the source. Each module is connected to the source using
separate wires. Using the source as the current distribution point allows larger wires to be used for each module
connection and also reduces the common impedance inherent in daisy-chained configurations.
If because of lead length or other considerations, lead connections cannot be made at the source, a remote
distribution terminal may be required. Lead connections can also be daisy-chained across the module input
connectors as long as the total current draw is less than the ampere-rating of AWG 8 wire (see Table 3-2). This
is because two wires larger than AWG 8 cannot both fit inside the module input connectors.
Low Voltage Operation
When operating below 3 volts, the slew rate and input current are derated, as shown in figure A-1. If these
conditions are not acceptable for your application, then it is recommended that you use an external dc power
supply to boost the voltage across the load terminals to above 3 volts, as shown in figure 3-13.
Remote sensing is recommended in this configuration, as illustrated in Figure 3-13. The load leads connect to
both the boost supply and the DUT, but the remote sense leads connect directly to the DUT. This allows the
voltage readback to measure the voltage at the DUT alone, and also improves load regulation in CV and CR
modes.
In this configuration, power from both the boost supply and the DUT must be absorbed by the electronic load.
A higher power load module may now be needed. Also, the boost supply must have adequate ratings to allow
the DUT to produce its full rated current. Finally, and noise generated by the boost supply will affect
measurements made on the DUT. A boost supply with suitable noise specifications should be selected.
N3300A
Figure 3-13. Zero Volt Loading
51
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Turn-On Checkout
Introduction
Successful tests in this chapter provide a high degree of confidence that the electronic load is operating properly.
For verification tests, see Appendix B.
NOTE:
This chapter provides a preliminary introduction to the electronic load front panel. See Chapter
5 for more detail.
Checkout Procedure
The tests in this section checks for proper operation of the electronic load. If you have not already done so,
connect the power cord to the unit and plug it in.
The following procedure assumes that the unit turns in the factory-default state. If you need more information
about the factory default state, refer to the *RST command in Chapter 4 of the Programming Guide. Note that
the values shown in the Display column may not exactly match the values that appear on the front panel of your
unit.
Table 4-1. Checkout Programming Values
Model
All
Voltage Current
10V 10A
Note
The settings of the power supply and the values used in the procedure were
selected so that they can be used with any module. You can use different
settings, but the results of the test will differ from the results shown below.
Procedure
Display
Explanation
Turn the unit on. The electronic
load undergoes a self-test when you
first turn it on.
**************
During selftest, all display segments are
briefly lit, followed by the channel number
displayed and the CC annunciator on.
1.
1
0.00
0.00
Check that the electronic load fan is
on.
You should be able to hear the fan and feel
the air coming from the back of the unit.
2.
3.
Connect a 10V, 10A power supply
to the input connectors of the first
channel of the electronic load.
Set the power supply voltage to
10V and current limit to 10A.
1
10.00 0.00
Observe the front panel. It should display
the voltage that the power supply was set to.
4.
5.
Depress the following front panel
keys in the indicated order:
Current , 5, Enter.
1 10.00 4.99
The electronic load is drawing 5A and is
operating in the CC mode. The CC
annunciator is on. The power supply should
be operating in the CV mode.
53
Download from Www.Somanuals.com. All Manuals Search And Download.
4 - Turn-On Checkout
Depress the Meter key followed by
the scroll key.
1
50 W
The display shows the computed input
power for the selected channel.
6.
7.
8.
Repeat steps 3 through 6 for any
other modules in the mainframe.
Turn off electronic load, disconnect
power supply and continue with
rear panel connections.
NOTE:
If the Err annunciator on the display is on, press the Shift key followed by the Error key to see
the error number. See Table 4-2 for a list of errors.
In Case of Trouble
Error Messages
Electronic load failure may occur during power-on Selftest or during operation. In either case, the display may
show an error message that indicates the reason for the failure.
Selftest Errors
Pressing the Shift, Error keys will show the error number. On the front panel, selftest error messages appear as;
ERROR <n> where "n" is a number listed in the following table. Some values consist of a number followed by
the affected channel. If this occurs, turn the power off and then back on to see if the error persists. If the error
message persists, the electronic load requires service. Refer to the list of Agilent Sales and Support Offices at
the back of this manual.
Table 4-2 Selftest Error Codes
Error
Failed Test
0
No error
2
4
5
10
Non-volatile RAM CONFIG section checksum failed
Non-volatile RAM STATE section checksum failed
Non-volatile RAM RST section checksum failed
RAM selftest failed
11<channel>
12<channel>
13<channel>
14<channel>
15<channel>
16<channel>
17
CVDAC selftest 1 failed on specified channel (01 - 06)
CVDAC selftest 2 failed on specified channel (01 - 06)
CCDAC selftest 1 failed on specified channel (01 - 06)
CCDAC selftest 2 failed on specified channel (01 - 06)
CRDAC selftest 1 failed on specified channel (01 - 06)
CRDAC selftest 2 failed on specified channel (01 - 06)
ADC buffer failed
18<channel>
19<channel>
20
LIMIT section checksum failed on specified channel (01 - 06)
CAL section checksum failed on specified channel (01 - 06)
Input Down
40
Flash write failed
42
Flash erase failed
80
Digital I/O selftest error
54
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Front Panel Operation
Introduction
Here is what you will find in this chapter:
•
•
A complete description of the front panel controls
Front panel programming examples.
NOTE:
The electronic load must be set in Local Mode to use the front panel controls. Press the Local
key on the front panel to put the unit in Local Mode.
Front Panel Description
1
4
2
3
5
N3300A
SYSTEM DC ELECTRONIC LOAD
SYSTEM
Ident
Local
FUNCTION
Func
ENTRY
9
CHANNEL
VOLTS
AMPS
Sense
Meter
Step
7
4
8
5
2
Input
Error
Address
Channel
Channel
Step
CV CC CR
Unr Dis Tran
Prot Cal Shift Rmt Addr Err SQR
6
Current
Res
Input
Enter
Save
Recall
Prot Clear
Protect
List
Tran
1
3
Clear Entry
E
-
.
Trigger
Trigger
Voltage
Input
0
on/off
Control
LINE
ON
OFF
6
Figure 5-1 Front Panel, Overall View
55
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
15-character fluorescent display for showing measurements and programmed values.
Annunciators light to indicate operating modes and status conditions:
j
Display
kAnnunciators
φ1
A list is in progress
CV
CC
CR
Unr
Dis
The electronic load channel is in constant-voltage mode.
The electronic load channel is in constant-current mode.
The electronic load channel is in constant-resistance mode.
The electronic load channel is in an unregulated state.
The electronic load channel measurement is disabled (off).
Tran The electronic load channel is in transient operation.
Prot One of the electronic load's channel protection feature is activated.
Cal
The electronic load is in calibration mode.
Shift The Shift key is pressed to access an alternate key function.
Rmt The selected interface (GPIB or RS-232) is in a remote state.
Addr The interface is addressed to talk or to listen.
Err
There is a message in the SCPI error queue.
SQR The interface is requesting service from the controller.
The system keys let you:
Return to Local mode (front panel control).
Set the electronic load GPIB address.
lSystem Keys
Set the RS-232 interface communication baud rate and parity bit.
Display SCPI error codes and clear the error queue.
Save and recall up to 10 instrument operating configurations.
Function access command menus that let you:
Select CC, CR and CV modes.
m
Function Keys
Select channel.
Select Current, Resistance and Voltage levels.
Set Trigger and Transient levels.
Set up front panel measurements.
Entry keys let you:
Enter programming values.
Increment or decrement programming values.
▲ Enter and ▼ Enter select the front panel menu parameters.
n
Entry Keys
Line
This turns the electronic load on or off.
o
56
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
System Keys
Refer to the examples later in this chapter for more details on the use of these keys.
SYSTEM
Ident
Local
Error
Address
Save
Recall
Figure 5-2. System Keys
This is the blue, unlabeled key, which is also shown as shift in this guide. Pressing this
key accesses the alternate or shifted function key (such as ERROR). Release the key after
you press it. The Shift annunciator is lit, indicating that the shifted keys are active.
Press to change the electronic load's selected interface from the remote operation to local
(front panel) operation. Pressing the key will have no effect if the interface state is
already Local, Local-with-Lockout, or Remote-with-Lockout.
Local
Press to access the system address menu. This menu lets you configure the electronic
load's interface. Address Menu entries are stored in non-volatile memory.
Address
Display
Command Function
ADDRESS <value>
Sets the GPIB address
INTF <char>
Selects an interface (GPIB or RS-232)
Selects baud rate (300, 600, 1200, 2400, 4800, 9600)
Message parity (NONE, EVEN, ODD, MARK, SPACE)
Flow control ( RTS-CTS, NONE)
BAUDRATE <value>
PARITY <char>
FLOW <char>
value = numeric value
char = a character string parameter
~
Use
and
to scroll through the command list
~Input
Input
Use
and
to scroll through the parameter list
Press to place the electronic load into a previously stored state. You can recall up to 10 (0
through 9) previously stored states.
Recall
Shift
Ident
Displays full scale voltage and current for the present channel. It can be used to identify
which modules are installed in the selected channel locations.
Press to display the system error codes stored in the SCPI error queue. This action also
clears the queue. If there is no error in the queue, 0 is displayed.
Shift
Shift
Error
Save
Press to store an existing electronic load state in non-volatile memory. The parameters
saved are listed under *SAV in the electronic load Programming Guide. You can save up
to 10 states (0 through 9).
57
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Function keys
Refer to the examples later in this chapter for more details on the use of these keys.
FUNCTION
Sense
Meter
Step
Step
Func
Channel
Channel
Current
Res
Prot Clear
Protect
List
Tran
Trigger
Voltage
Input
on/off
Trigger
Control
Figure 5-3. Function Keys
Immediate Action Keys
Immediate action keys immediately execute their corresponding function when pressed. Other function keys
have commands underneath them that are accessed when the key is pressed.
Input
On/Off
This key toggles the input of the electronic load between the on and off states. It
immediately executes its function as soon as you press it. When off, the electronic
load input is disabled and the Dis annunciator is on.
Q
Selects another channel.
Channel
Causes a trigger to occur. The trigger system must first be initiated by pressing
Shift
Trigger
Trigger Control and Enter. Triggers are used to generate transients or lists.
Displays any protection functions that are tripped.
Protect
Shift
Prot
Clear
Press this key to reset the protection circuit and allow the unit to return to its last
programmed state. The condition that caused the protection circuit to become active
must be removed prior to pressing this key, or the unit will shut down again and
display the Prot annunciator again.
Takes the front panel back to the Metering mode from any other mode.
Meter
58
Download from Www.Somanuals.com. All Manuals Search And Download.
Q
These keys let you scroll through points in a list function. When the display indicates
EOL, the end of the list has been reached.
Press this key to scroll through all channels in the mainframe
Front Panel Operation - 5
Scrolling Keys
Q
R
Scrolling keys let you move through the commands in the presently selected function
menu. Press ▼ to bring up the next command in the list. Press ▲ to go back to the
previous command in the list. Function menus are circular; you can return to the starting
position by continuously pressing either key
R
Step
Step
Q
Channel
Metering Keys
Metering keys control the metering functions of the electronic load. As set from the factory, all front panel
measurements are calculated from a total of 1000 readings taken at a 10 microsecond sampling rate. Therefore,
the factory default acquisition time for a single front panel measurement is about 10 milliseconds. Refer to
“Making Front Panel Measurements” for more information about changing the front panel sampling rate and the
number of measurement points.
Meter
Takes the front panel back to the Metering mode from any other mode.
Also accesses the meter menu. This menu lets you measure various parameters of the
selected input.
Display
Measurement
<chan> <voltage> <current> Displays the voltage and current of the selected channel
<chan> <value> V MAX
<chan> <value> V MIN
<chan> <value> V RMS
<chan> <value> A MAX
<chan> <value> A MIN
<chan> <value> A RMS
<chan> <value> WATTS
<chan> <value> W MAX
<chan> <value> W MIN
Displays the maximum voltage
Displays the minimum voltage
Displays the rms voltage
Displays the maximum current
Displays the minimum current
Displays the rms voltage
Displays the input power
Displays the maximum power
Displays the minimum power
Shift
Sense
Specifies the measurement functions and ranges.
Display
S:PNT
Command Function
Defines the number of data points in the measurement
S:TIN
Sets the digitizer sample spacing
S:WIN
S:OFF
S:C:RNG
S:V:RNG
Sets the measurement window function
Defines the data offset in the measurement
Selects the current measurement range
Selects the voltage measurement range
R
Q
Use
Use
and
and
to scroll through the menu commands
to scroll through the command parameters
Notes:
R Input
Q Input
59
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Input Control Keys
Input control keys control the input functions of the electronic load.
Shift
Channel
Press this key to select another channel. Enter the channel number using the Entry Keys.
(This key performs the same function as ▼ Channel.)
Current
Press this key to access the current menu.
Display
Command Function
CURR <value> Sets the current of input
C:MODE
C:RANG
C:SLEW
C:SLW:N
C:SLW:P
C:TLEV
C:TRIG
Sets the current mode to FIX or LIST
Sets the current range
Sets the current slew rate for both positive and negative transitions
Sets the current slew rate for the negative transitions
Sets the current slew rate for the positive transitions
Sets the current transient input level
Sets the current trigger input level
Press this key to access the resistance menu.
Res
Voltage
Func
Display
Command Function
RES <value>
R:MODE
R:RANG
R:SLEW
R:SLW:N
R:SLW:P
R:TLEV
Sets the resistance of input
Sets the resistance mode to FIX or LIST
Sets the resistance range
Sets the resistance slew rate for both positive and negative transitions
Sets the resistance slew rate for the negative transitions
Sets the resistance slew rate for the positive transitions
Sets the resistance transient input level
Sets the resistance trigger input level
R:TRIG
Press this key to access the voltage menu.
Display Command Function
VOLT <value> Sets the voltage of input
V:MODE
V:RANG
V:SLEW
V:SLW:N
V:SLW:P
V:TLEV
V:TRIG
Sets the voltage mode to FIX or LIST
Sets the voltage range
Sets the voltage slew rate for both positive and negative transitions
Sets the voltage slew rate for the negative transitions
Sets the voltage slew rate for the positive transitions
Sets the voltage transient input level
Sets the voltage trigger input level
Press this key to access the function menu.
Display
Command Function
FUNC <char>
FNC:MODE
INP:SHOR
Selects the input regulation mode (CURR, RES, VOLT)
Selects what controls the regulation mode (FIX or LIST)
Enables or disables the input short (ON, OFF)
R
Q
Use
Use
and
and
to scroll through the menu commands
to scroll through the command parameters
Notes
R Input
Q Input
60
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
Transient Control Keys
Transient control keys control the transient functions of the electronic load.
Tran
Press this key to access the function menu.
Display
Command Function
TRAN <value> Sets the current of input
T:DCYC
T:FREQ
T:MODE
T:TWID
Sets the current mode to FIX or LIST
Sets the current slew rate for both positive and negative transitions
Sets the current slew rate for the negative transitions
Sets the current slew rate for the positive transitions
Trigger Control Keys
Trigger control keys control the trigger functions of the electronic load.
Trigger
Control
Press this key to access the function menu.
Display
Command Function
INIT:IMMED
Sets the current of input
ABORT
Sets the current mode to FIX or LIST
List Control Keys
List control keys control the list functions of the electronic load.
List
Press this key to access the function menu.
Display
Command Function
LIST:STEP
Sets the method of incrementing steps (ONCE, AUTO)
Specifies the number of times the list is cycled
Specifies the time period of each step
LST:CNT
DWEL:0 EOL
CURR:0 EOL
C:RANG:0 EOL
C:SLEW:0 EOL
C:SLW:N:0 EOL
C:SLW:P:0 EOL
C:TLEV:0 EOL
FUNC:0 EOL
RES:0 EOL
R:RANG:0 EOL
R:SLEW:0 EOL
R:SLW:N:0 EOL
R:SLW:P:0 EOL
R:TLEV:0 EOL
TRAN:0 EOL
T:DCYC:0 EOL
T:FREQ:0 EOL
T:MODE:0 EOL
T:TWID:0 EOL
VOLT:0 EOL
V:RANG:0 EOL
V:SLEW:0 EOL
Specifies the current setting for each step
Specifies the current range for each step
Sets the current slew rate for each step
Sets the negative current slew rate for each step
Sets the positive current slew rate for each step
Sets the transient input current for each step
Sets the list regulation mode (CURR, RES, VOLT)
Specifies the resistance setting for each step
Specifies the resistance range for each step
Sets the resistance slew rate for each step
Sets the negative resistance slew rate for each step
Sets the positive resistance slew rate for each step
Sets the transient input resistance for each step
Enables/disables the transient level for each step
Sets the transient duty cycle for each step
Sets the transient frequency for each step
Sets the mode of the transient generator (CONT, PULSE)
Sets the transient pulse width for each step
Specifies the voltage setting for each step
Specifies the voltage range for each step
Sets the voltage slew rate for each step
V:SLW:N:0 EOL Sets the negative voltage slew rate for each step
V:SLW:P:0 EOL
V:TLEV:0 EOL
Sets the positive voltage slew rate for each step
Sets the transient input voltage for each step
61
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Entry Keys
7
4
8
5
2
9
6
Input
Input
Enter
1
3
Clear Entry
E
-
.
0
Figure 5-4 Entry Keys
Q
R
These keys perform two functions. In Meter mode, these keys can be used to adjust
the present input current, voltage, or resistance - depending on which function is
presently active (indicated by the CC, CR, or CV annunciator).
Input
Input
In menu mode, these keys let you scroll through choices in a parameter list that
apply to a specific command. Parameter lists are circular; you can return to the
starting position by continuously pressing either key. If the command has a numeric
range, these keys increment or decrement the existing numeric value.
S
The backspace key deletes the last digit entered from the keypad. This key lets you
correct one or more wrong digits before they are entered.
0
9
.
0 through 9 are used for entering numeric values. . is the decimal point. For example,
to enter 33.6 press: 3, 3, . , 6, Enter.
-
,
This key executes the entered value or parameter of the presently accessed command.
Until you press this key, the parameters you enter with the other Entry keys are
displayed but not entered into the electronic load. Before pressing Enter, you can
change or abort anything previously entered into the display. After Enter is pressed,
the electronic load returns to Meter mode.
Enter
Shift
Shift
Shift
E
-
This is used to enter an exponent value in the numeric field. Pressing these keys
displays an E in the field. Enter the value of the exponent following the E, then press
Enter.
This is used to enter either a negative number or a negative exponent value in the
numeric field. Pressing these keys displays an - in the field. Enter the numeric value
then press Enter.
Clear
Entry
This is used to clear the numeric field of an entered value before the Enter key has
been pressed. After clearing the field, a new number can be entered.
62
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
Examples of Front Panel Programming
You will find these examples on the following pages:
1. Using the front panel display.
2. Programming constant current, voltage, and resistance modes.
3. Programming transient operation.
4. Programming lists.
5. Querying and Clearing Output Protection and Errors
6. Making basic front panel measurements
7. Setting GPIB address or RS-232 parameters.
8. Saving and recalling operating states.
Similar examples are given in the electronic load Programming Guide using SCPI commands.
1 - Using the Front Panel Display
Action
Display
1.
Press Meter to return the display to Meter mode. Press QChannel to scroll
through the channels. The left-most digit of the front panel display identifies the
input channel that is presently being controlled by the front panel.
1 7.003V 0.004A
You can only select an input when the unit is in metering mode. Once an input has
been selected, only the menu commands that apply to that input will appear on
the display. The CV, CC, and UNR annunciators apply to the selected channel.
2 - Programming Constant Current, Voltage and Resistance Modes
This example shows you how to set the input voltage, current and resistance modes.
Set the constant current mode input
Action
Display
1.
On the Function keypad, press Q Channel to scroll through the channels until
1
channel 1 appears on the display.
2.
3.
On the Function keypad, press Input On/Off to turn input off.
Dis annunciator on
On the Function keypad, press Func. On the Entry keypad, press Q Input or
R Input key until the current function is displayed, then press Enter.
FUNC CURR
4.
On the Function keypad, press Current. On the Function keypad, press Q key until CURR
current is displayed.
5.
6.
On the Entry keypad press 1.25 Enter.
CURR 1.25
C:SLW
On the Function keypad, press Current. On the Function keypad, press Q key until
current slew is displayed.
7.
8.
On the Entry keypad press 30000 Enter.
C:SLW 30000
C:TLEV
On the Function keypad, press Current. On the Function keypad, press Q key until
current transient is displayed.
9.
On the Entry keypad press 2 Enter.
C:TLEV 2
63
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
10. On the Function keypad, press Input On/Off to turn input on.
Dis annunciator off
11. To make minor changes to an existing value: On the Function keypad, press
Current. On the Entry keypad, press QInput or R Input to scroll from 1.25 to
2.25. Then press Enter.
CURR 2.25
Set the constant resistance mode input
Action
Display
1.
On the Function keypad, press Q Channel to scroll through the channels until
1
channel 1 appears on the display.
2.
3.
On the Function keypad, press Input On/Off to turn input off.
Dis annunciator on
On the Function keypad, press Func. On the Entry keypad, press Q Input or
R Input key until the resistance function is displayed, then press Enter.
FUNC:RES
4.
On the Function keypad, press Res. On the Function keypad, press Q key until
RES
resistance is displayed.
5.
6.
On the Entry keypad press 50 Enter.
RES 50
R:TLEV
On the Function keypad, press Res. On the Function keypad, press Q key until
resistance transient is displayed.
7.
8.
9.
On the Entry keypad press 40 Enter.
R:TLEV 40
Dis annunciator off
RES 60
On the Function keypad, press Input On/Off to turn input on.
To make minor changes to an existing value: On the Function keypad, press Res.
On the Entry keypad, press QInput or R Input to scroll from 50 to 60. Then press
Enter.
Set the constant voltage mode input
Action
Display
1.
On the Function keypad, press Q Channel to scroll through the channels until
1
channel 1 appears on the display.
2.
3.
On the Function keypad, press Input On/Off to turn input off.
Dis annunciator on
On the Function keypad, press Func. On the Entry keypad, press Q Input or
R Input key until the voltage function is displayed, then press Enter.
FUNC VOLT
4.
On the Function keypad, press Voltage. On the Function keypad, press Q key until
VOLT
voltage is displayed.
5.
6.
On the Entry keypad press 6 Enter.
VOLT 6
V:SLW
On the Function keypad, press Voltage. On the Function keypad, press Q key until
voltage slew is displayed.
7.
8.
On the Entry keypad press 30000 Enter.
V:SLW 30000
V:TLEV
On the Function keypad, press Voltage. On the Function keypad, press Q key until
voltage transient is displayed.
9.
On the Entry keypad press 4 Enter.
V:TLEV 4
10. On the Function keypad, press Input On/Off to turn input on.
Dis annunciator off
VOLT 7.000
11. To make minor changes to an existing value: On the Function keypad, press
Voltage. On the Entry keypad, press QInput or R Input to scroll from 6.000 to
7.000. Then press Enter.
64
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
3 - Programming Transient Operation
Transient operation can be used in the CC, CR or CV mode. It causes the selected channel to switch between
two load levels.
Set transient operation for Continuous mode
In the following example, assume that the CC mode is active, the slew rate is at the default setting (maximum
rate), and the applicable transient operation parameters have been set as follows: continuous mode, main current
level = 5A, transient current level = 10 A, transient frequency = 1kHz, transient duty cycle = 40.
The load module starts conduction at the main level (in this case 5 amps). When transient operation is turned on
the module input current will slew to and remain at 10 A for 40% of the period (400 µs), then slew to and
remain at 5 A for the remaining 60% (600 µs) of that cycle. This cycle is repeated on a continuous basis. Figure
5-5 shows the waveform that would result in this example.
Figure 5-5. Continuous Transient Operation
Action
Display
1.
2.
On the Function keypad, press Current. On the Entry keypad, press 5 Enter .
CURR 5
On the Function keypad, press Current. On the Function keypad, press Q key until C:TLEV
transient current is displayed.
3.
4.
On the Entry keypad press 10 Enter.
C:TLEV 10
On the Function keypad, press Tran. On the Function keypad, press Q key until
T:DCYC
transient duty cycle is displayed.
5.
6.
On the Entry keypad press 40 Enter.
T:DCYC 40
T:FREQ
On the Function keypad, press Tran. On the Function keypad, press Q key until
transient frequency is displayed.
7.
8.
On the Entry keypad press 1000 Enter.
1000
On the Function keypad, press Tran. On the Function keypad, press Q key until
T:MODE
transient mode is displayed.
9.
To select the continuous mode, press the QInput or R Input key until the
Continuous mode is displayed. Then press Enter.
T:MODE CONT
TRAN ON
10. On the Function keypad, press the Tran key and select ON to activate the transient
mode.
65
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Set transient operation for Pulse mode
In this example, assume that the CC mode is active, the slew rate is at the factory default setting (maximum
rate), and the applicable transient operation parameters have been set as follows: pulse mode, main current level
= 5 A, transient current level = 10 A, pulse width = 1 millisecond.
In this example, the electronic load starts conduction at the main current level setting (5 amps). When the
transient mode is turned on and an external trigger signal is received, module input current will slew to and
remain at 10 A for the remainder of the 10 millisecond time period, then slew to and remain at 5 A until another
trigger is received. Any triggers that occur during the time the transient level is in effect will extend the pulse
by another 10 milliseconds. Figure 5-6 shows the waveform that would result in this pulsed transient example.
Figure 5-6. Pulsed Transient Operation
Action
Display
1.
2.
On the Function keypad, press Current. On the Entry keypad, press 5 Enter .
CURR 5
On the Function keypad, press Current. On the Function keypad, press Q key until C:TLEV
transient current is displayed.
3.
4.
On the Entry keypad press 10 Enter.
C:TLEV 10
On the Function keypad, press Tran. On the Function keypad, press Q key until
T:TWID
transient pulse width is displayed.
5.
6.
On the Entry keypad press 0.01 Enter.
T:TWID 0.01
T:MODE
On the Function keypad, press Tran. On the Function keypad, press Q key until
transient mode is displayed.
7.
8.
9.
To select the pulse mode, press the QInput or R Input key until the Pulse mode
is displayed. Then press Enter.
T:MODE PULS
TRAN ON
On the Function keypad, press the Tran key and select ON to activate the transient
mode.
On the Function keypad, press Trigger Control, then press Enter. This initiates
INIT:IMMED
the trigger system.
10 Press Shift and Trigger. This starts the pulse. Repeat this step for additional pulses.
Set transient operation for Toggled mode
In this example, assume that the CC mode is active, the slew rate is at the factory default setting (maximum
rate), and the applicable transient operation parameters have been set as follows: toggle mode, main current
level = 5 A, transient current level = 10 A.
The operation of toggled mode is similar to that described for continuous and pulse operation, except that each
time a trigger is received the input alternates between the main and transient current levels. Figure 5-7 shows the
waveform that would result for this toggled transient operation example
66
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
Figure 5-7. Toggled Transient Operation
Action
On the Function keypad, press Current. On the Entry keypad, press 5 Enter .
Display
1.
2.
CURR 5
On the Function keypad, press Current. On the Function keypad, press • key until C:TLEV
transient current is displayed.
3.
6.
On the Entry keypad press 10 Enter.
C:TLEV 10
T:MODE
On the Function keypad, press Tran. On the Function keypad, press • key until
transient mode is displayed.
7.
8.
9.
To select the pulse mode, press the •Input or ‚ Input key until the Toggle mode T:MODE TOGG
is displayed. Then press Enter.
On the Function keypad, press the Tran key and select ON to activate the transient
TRAN ON
mode.
On the Function keypad, press Trigger Control, then press Enter. This initiates
INIT:IMMED
the trigger system.
10 Press Shift and Trigger. This switches to the toggled level. Each time a trigger is
sent, the input switches to the alternate level.
NOTE:
The difference between programming a toggled level (C:TLEV) and a triggered level (C:TRIG),
is that a toggled level allows you to switch back and forth between the main (or immediate)
level and the toggled level. A triggered level, on the other hand, becomes the main level once it
is triggered. Subsequent triggers have no effect on the input level.
4 - Programming Lists
Lists are the most flexible means of generating complex input sequences. The following figure shows a current
input sequence generated from a list. The input current shown represents three different current pulses (8 A for 2
seconds, 6 A for 5 seconds, and 4 A for 7 seconds) separated by 3-second, 0-ampere intervals.
The list specifies the pulses as three current points (point 0, 2, and 4), each with its corresponding dwell point.
The intervals are three zero-current points (point 1, 3, and 5) of equal intervals. The count parameter causes the
list to execute twice when started by a single trigger.
NOTE:
From the Recall Menu, execute the *RST command to reset the load module. This is necessary
because any previously programmed functions remain in effect until cleared.
Action
Press Current to access the Current Menu. Then press ‚ to access the current mode
Display
1.
2.
3.
C:MODE FIXED
C:MODE LIST
LST:CNT 2
command.
On the Entry keypad, press •Input or ‚Input to scroll through the mode parameters
to obtain LIST and press Enter.
Access the List Menu by pressing Shift List. Press ‚ to scroll to the count command.
From the Entry keypad, change the list count from the default (1) to 2. Press Enter.
67
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Trigger
1
2
3
4
5
List Count = 1
List Count = 2
0
Figure 5-8. List Sequence
Action
Display
4.
5.
Access the List menu again and press ‚ until you access the dwell time. This specifies
the time for each current point, which is effectively its width. The first dwell point (0)
appears in the display. On the Entry keypad, press 2 and Enter.
DWEL:0
2
Pressing the Enter key automatically advances to the step in the list. Enter the following
values for dwell list points 1 through 5: 3, 5, 3, 7, 3. Press Enter to enter each value.
When you finish, you will be at point 6, which is the end of the list.
DWEL:1
DWEL:2
DWEL:3
DWEL:4
DWEL:5
3
5
3
7
3
Note: Press Shift •Step or Shift ‚Step to access and edit any list point.
DWEL:6 EOL
6.
7.
Press ‚ until you access the current list. This specifies the value of each current point
during its corresponding dwell period. The first current list point (0) appears in the
display. On the Entry keypad, press 8 and Enter.
CURR:0
8
Pressing the Enter key automatically advances to the step in the list. Enter the following
values for current list points 1 through 5: 0, 6, 0, 4, 0. Press Enter to enter each value.
When you finish, you will be at point 6, which is the end of the list.
CURR 1
CURR 2
CURR 3
CURR 4
CURR 5
0
6
0
4
0
Note: Press Shift •Step or Shift ‚Step to access and edit any list point.
CURR 6 EOL
8.
Progam at least one lis value for the remaining functions in the current list system.
Press ‚ until you access the following current list functions:
C:RANG
C:SLEW
C:TLEV
C:RANG:0 60
C:SLEW:0 5E6
C:TLEV:0
0
The value programmed for C:SLEW will also be applied to C:SLW:P and S:SLW:N.
If you will not be programming different values for each list step, you only need to
program an initial value in step 0 that will be applied to all the steps for that function.
Press ‚ until you access the step command. Check that it is at the default mode (AUTO).
This lets a single trigger run your list for the specified count.
Press Trigger Control and Enter to initiate the trigger system. The small Æ1 annunciator
STEP AUTO
INIT:IMMED
9.
10.
on the left of the display comes on to indicate that the list is initialized.
If the message LIST LENGTH appears on the display, it usually means that not all
current functions have had a list value programmed, or else one of the programmed lists
is longer or shorter than the other lists.
11. Press Shift Trigger. This sends the ac source an immediate trigger to generate the
0 V 60 Hz
output list sequence. The output returns to the immediate value at the end of the list.
Note: To clear a list, press Clear Entry. This truncates or clears the list at the presently
displayed list point. Each list must be accessed and cleared separately.
68
Download from Www.Somanuals.com. All Manuals Search And Download.
Front Panel Operation - 5
5 - Querying and Clearing Output Protection and Errors
When overvoltage, overcurrent, or overtemperature condition occurs, the Prot annunciator on the front panel
will be on and the electronic load will disable its output.
Error messages can occur at any time during the operation of the unit. When the Err annunciator on the front
panel is on, it means that either an error has occurred on the GPIB bus, or a selftest error has occurred. Appendix
C lists error numbers and descriptions.
Query and clear the electronic load overcurrent protection as follows:
Action
Display
1.
2.
On the Function keypad, press Protect. In this example, an overcurrent condition
has occurred.
OC
To restore normal operation after the cause of the overcurrent condition has been
removed, press Shift, Prot Clr. The Prot annunciator then will go off.
Query and Clear Errors as follows:
1.
On the Function keypad, press Shift, Error. This displays and clears the error in
the error queue. Repeatedly press these keys to clear all errors in the queue. If errors
persist, your unit may require service.
ERROR 0
6 - Making Basic Front Panel Measurements
As shipped from the factory, front panel measurements for the input are calculated from a total of 1000 readings
taken at a 10 microsecond sampling rate. The unit alternates between voltage and current measurements.
Therefore, the data acquisition time for a single front panel voltage or current measurement is about 100
milliseconds.
There are no trigger controls for front panel measurements. However, you can program both the sampling rate
and the number of data points in each front panel measurement using commands in the Sense menu. With this
flexibility, measurement accuracy can be improved for waveforms with frequencies as low as several Hertz. The
sample buffer size may be varied from 1 to 4096 data points. The sampling rate may be varied from 10
microseconds to 32 milliseconds. Values are rounded to the nearest 10 microsecond interval.
To have the unit turn on with the reconfigured buffer size and sampling rate, save this state in location 0.
NOTE:
If the front panel display indicates OVLD, the output has exceeded the measurement capability
of the instrument. This can occur when the measurement has been set to the Low range. If the
front panel display indicates -- -- -- -- -- -- , a GPIB measurement is in progress.
Use the Meter menu for making front panel measurements:
Action
Display
1.
On the Function keypad press Meter to access the following measurement
parameters:
Displays the voltage and current of the selected channel
Displays the maximum voltage
Displays the minimum voltage
Displays the rms voltage
Displays the maximum current
<chan> <voltage> <current>
<chan> <value> V MAX
<chan> <value> V MIN
<chan> <value> V RMS
<chan> <value> A MAX
69
Download from Www.Somanuals.com. All Manuals Search And Download.
5 - Front Panel Operation
Action
Display
Displays the minimum current
Displays the rms voltage
Displays the input power
Displays the maximum power
Displays the minimum power
<chan> <value> A MIN
<chan> <value> A RMS
<chan> <value> WATTS
<chan> <value> W MAX
<chan> <value> W MIN
1.
‚
•
Use
and
to scroll through the measurement selections
7 - Setting the GPIB Address
Your electronic load is shipped with the GPIB address set to 5. This address can only be changed from the front
panel using the Address menu located under the Address key
Set the GPIB address as follows:
Action
Display
1.
2.
On the System keypad, press Address.
Enter the new address. For example, Press 7, Enter.
ADDRESS 5
ADDRESS 7
8 - Storing and Recalling Instrument States
You can save up to 10 states (from location 0 to location 9) in non-volatile memory and recall them from the
front panel. All programmable settings are saved.
NOTE:
The SAV0 state is the power up state.
Save an instrument state in location 0 as follows:
Action
Display
1.
2.
Set the instrument to the state that you want to save.
Save this state to location 0. Press Shift, Save, 0, Enter.
*SAV 0
Recall a saved state as follows:
Action
Display
1.
Recall the state saved in location 0 by pressing Recall, 0, Enter.
*RCL 0
Clear the non-volatile memory of the electronic load as follows:
Action
Display
1.
On the System keypad, press the Recall key and scroll to the reset command. Then
press Enter.
*RST
This returns the unit to the factory-default settings.
Save these settings to location 0. Press Shift, Save, 0, Enter.
Repeat step #2 for memory locations 7 through 9.
2.
3.
*SAV 0
*SAV 7
*SAV 8
*SAV 9
70
Download from Www.Somanuals.com. All Manuals Search And Download.
A
Specifications
Table A-1 lists the specifications for the different load models. Specifications indicate warranted performance in
the 25°C 5°C region of the operating temperature range. Specifications apply to normal and transient modes
unless otherwise noted.
Table A-1. Specifications
N3302A
N3303A
N3304A
N3305A
N3306A
Input Ratings
Current
0 - 30 A
0 - 60 V
150 W
0 - 10 A
0 - 240 V
250 W
0 - 60 A
0 - 60 V
300 W
0 - 60 A
0 - 150V
500 W
0 - 120 A
0 - 60V
600 W
Voltage
Maximum Power @ 40°C 1
Input Characteristic
OPERATING CONTOUR
DERATED CURRENT DETAIL
Voltage
Voltage
full
scale
Max Power Contour
all specifications apply
3
2
1
0
slew rate limitations apply
(see Table A-2)
0
Current
Current
full
full
scale
scale
Specified Current @
Low Voltage Operation
2.0 V
1.5 V
1.0 V
0.5 V
0 V
30 A
10 A
7.5 A
5 A
60 A
45 A
30 A
15 A
0 A
60 A
45 A
30 A
15 A
0 A
120 A
90 A
60 A
30 A
0 A
22.5 A
15 A
7.5 A
0 A
2.5 A
0 A
1Maximum continuous power available is derated linearly from 100% of maximum at 40°C, to 75% of maximum at 55°C.
Typical Minimum Operating Voltage @ Full Scale Current
N3302A
N3303A
N3304A
N3305A
N3306A
Table A-1 states that maximum
current is available down to 2
volts. Typically, under normal
operating conditions, the load
can sink the maximum current
down to the following voltages:
1.2 V
1.2 V
1.2 V
1.4 V
1.4 V
71
Download from Www.Somanuals.com. All Manuals Search And Download.
A - Specifications
Table A-1. Specifications (continued)
N3302A
N3303A
N3304A
N3305A
N3306A
Constant Current Mode2
Low Range/High Range
Regulation
3A / 30A
10mA
1A / 10A
8mA
6A / 60A
10mA
6A / 60A
10mA
12A / 120A
10mA
Low Range Accuracy
High Range Accuracy
Constant Voltage Mode2
Low Range/High Range
Regulation
0.1% + 5mA
0.1% + 10mA
0.1% + 4mA
0.1% + 7.5mA
0.1% + 7.5mA
0.1% + 15mA
0.1% + 7.5mA
0.1% + 15mA
0.1% + 15mA
0.1% + 37.5mA
6V / 60V
5mV
24V / 240V
10mV
6V / 60V
10mV
15V / 150V
10mV
6V / 60V
20mV
Low Range Accuracy
High Range Accuracy
0.1% + 3 mV
0.1% + 8 mV
0.1% + 10mV
0.1% + 40mV
0.1% + 3 mV
0.1% + 8mV
0.1% + 10mV
0.1% + 20mV
0.1% + 3mV
0.1% + 8mV
Constant Resistance Mode2
Range 1 (I >10% of current rating)
Accuracy up to 25% of range
0.067-4 Ω
0.2-48 Ω
0.033-2 Ω
0.033-5 Ω
0.017-1 Ω
0.4% ± 24mΩ
0.4% ± 24mΩ
0.8% ± 24mΩ
3.6-40 Ω
1% ± 100mΩ
2% ± 100mΩ
3% ± 100mΩ
44-480 Ω
0.4% ±12mΩ
0.4% ±12mΩ
0.4% ±12mΩ
1.8-20 Ω
0.4% ± 24mΩ
0.4% ± 24mΩ
0.8% ± 24mΩ
4.5-50 Ω
0.4% ± 6mΩ
0.4% ± 6mΩ
0.4% ± 6mΩ
0.9-10 Ω
Accuracy >25% to 50% of range
Accuracy >50% to 100% of range
Range 2 (I >1% of current rating)
Accuracy up to 25% of range
Accuracy >25% to 50% of range
Accuracy >50% to 100% of range
1.5% ±80mΩ
2% ±80mΩ
3% ±80mΩ
36-400 Ω
8% ±240mΩ
16% ±240mΩ
32% ±240mΩ
5% ± 500mΩ
10% ± 500mΩ
20% ± 500mΩ
440-4800 Ω
-30/+50%
1.5% ±40mΩ
2% ±40mΩ
3% ±40mΩ
18-200 Ω
6% ±120mΩ
10% ±120mΩ
20% ±120mΩ
1.5% ±80mΩ
2% ±80mΩ
3% ±80mΩ
45-500 Ω
-10/+25%
-20/+50%
-30/+150%
1.5% ±20mΩ
2% ±20mΩ
3% ±20mΩ
9-100 Ω
6% ±60mΩ
10% ±60mΩ
20% ±60mΩ
Range 3 (I >0.1% of current rating)
Accuracy up to 25% of range
Accuracy >25% to 50% of range
Accuracy >50% to 100% of range
-40/+133%
-50/+600%
Range 4 (I >0.01% of current rating)
Accuracy up to 25% of range
Accuracy >25% to 50% of range
Accuracy >50% to 80% of range
Accuracy >80% to 100% of range
360-2000 Ω
-25/+100%
-25/+100%
-40/+400%
-40/+400%
4400-12000 Ω
-66/+800%
-66/+800%
-66/+800%
-66/+800%
180-2000 Ω
450-2500 Ω
-35/+200%
-45/+500%
-50/+1200%
-50/+2000%
90-1000 Ω
-20/+66%
-20/+66%
-33/+200%
-45/+600%
-50/+2000%
-33/+200%
-45/+600%
-50/+2000%
Transient Generator
Frequency Range
Frequency Accuracy
Duty Cycle Range
0.25Hz -10kHz
0.5%
3 to 97% (0.25Hz - 1kHz)
6 to 94% (1kHz - 10kHz)
1%
Duty Cycle Accuracy3
Pulse Width
50µs ± 1% to 4 seconds ± 1%
Current Measurement2
Low Range / High Range
Low Range Accuracy4
High Range Accuracy4
Voltage Measurement2
Low Range / High Range
Low Range Accuracy
High Range Accuracy
Power Measurement2
Accuracy
3A / 30A
0.05% + 3mA
0.05% + 6mA
1A / 10A
0.05% + 2.5mA
0.05% + 5mA
6A / 60A
0.05% + 5mA
0.05% + 10mA
6A / 60A
0.05% + 5mA
0.05% + 10mA
12A / 120A
0.05% + 10mA
0.05% + 20mA
6V / 60V
0.05% + 3mV
0.05% + 8mV
24V / 240V
0.05% + 10mV
0.05% + 20mV
6V / 60V
0.05% + 3 mV
0.05% + 8 mV
15V / 150V
0.05% + 8mV
0.05% + 16mV
6V / 60V
0.05% + 3mV
0.05% + 8mV
0.1% + 0.5W
0.1% + 1.2W
0.1% + 0.5W
0.1% + 1.5W
0.1% + 1.2W
2Specification is +/-(% of reading + fixed offset). Measurement is 1000 samples. Specification may degrade when the unit is subject to an
RF field of 3V/meter, the unit is subject to line spikes of 500V, or an 8kV electrostatic discharge.
3For example, if the duty cycle setting is 20%, the actual duty cycle would be in the range of 19% to 21%.
4DC current accuracy specifications apply 30 seconds after input current is applied.
72
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications - A
Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.
Table A-2. Supplemental Characteristics
N3302A
N3303A
N3304A
N3305A
N3306A
Programming Resolution
Constant Current Mode
Constant Voltage Mode
Constant Resistance Mode
0.05mA/0.5mA
0.1mV/1mV
0.07/0.7/7/70mW
0.02mA/0.2mA
0.4 mV/4mV
0.82/8.2/82mW
0.1 mA/1mA
0.1mV/1mV
0.035/0.35/3.5/
35mW
0.1mA/1mA
0.25mV/2.5mV
0.085/0.85/8.5/
85mW
0.2 mA / 2 mA
0.1mV / 1mV
0.0175/0.175/
1.75/17.5mW
Readback Resolution
Current
0.05mA/0.5mA
0.1mV / 1mV
0.02mA/0.2mA
0.4 mV / 4mV
0.1 mA / 1mA
0.1 mV / 1mV
0.1mA / 1 mA
0.25mV/2.5mV
0.2 mA / 2 mA
0.1mV / 1mV
Voltage
Programmable Slew Rate1
500A/s - 25kA/s
50kA/s - 2.5MA/s
50kA/s - 250kA/s
167A/s - 8330A/s
16.7kA/s - 833kA/s
16.7kA/s - 83.3kA/s 100kA/s - 500kA/s
1kA/s - 50kA/s
100kA/s - 5MA/s
1kA/s - 50kA/s
100kA/s - 5MA/s
100kA/s - 500kA/s
2kA/s - 100kA/s
200kA/s - 10MA/s
200kA/s - 1MA/s
Current Ranges
Voltage Ranges
Slow band
Fast band ³ 3V
Fast band <3V
Slow band
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
4kV/s - 200kV/s
400kV/s - 2MV/s
400kV/s - 200kV/
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
2.5kV/s - 125kV/s
250kV/s -1.25MV/s
250kV/s -125kV/s
1kV/s - 50kV/s
100kV/s - 500kV/s
100kV/s - 50kV/s
Fast band ³ 3V
Fast band <3V
44W/s - 1125W/s
2250W/s - 34kW/s
2250W/s - 3.4kW/s
540W/s - 13.5kW/s
27kW/s - 408kW/s
27kW/s - 40.8kW/s
22W/s - 560W/s
1120W/s - 17kW/s
1120W/s - 1.7kW/s
55W/s - 1400W/s
2800W/s - 42.5kW/s
2800W/s - 4.25kW/s
11W/s - 280W/s
560W/s - 8.5kW/s
560W/s - 850W/s
Resistance Range 1 Slow band
Fast band ³ 3V
Fast band <3V
440W/s - 11.25kW/s
22.5kW/s - 340kW/s
22.5kW/s - 34kW/s
5.4kW/s - 135kW/s
220W/s - 5600W/s
550W/s - 14kW/s
28kW/s - 425kW/s
28kW/s - 42.5kW/s
110W/s - 2800W/s
5600W/s - 85kW/s
5600W/s - 8.5kW/s
Resistance Range 2 Slow band
Fast band ³ 3V
270kW/s -4.08MW/s 11.2kW/s - 170kW/s
270kW/s -408kW/s
11.2kW/s - 17kW/s
Fast band <3V
4.4kW/s - 112.5kW/s
225kW/s - 3.4MW/s
225kW/s - 340kW/s
54kW/s - 1.35MW/s
2.7MW/s-40.8MW/s
2.7MW/s-4.08MW/s
2.2kW/s - 56kW/s
5.5kW/s - 140kW/s
1.1kW/s - 28kW/s
Resistance Range 3 Slow band
Fast band ³ 3V
112kW/s - 1.7M W/s 280kW/s -4.25MW/s 56kW/s - 850kW/s
112kW/s - 170k W/s
280kW/s -425kW/s
56kW/s - 85kW/s
Fast band <3V
44kW/s - 1.125MW/s
2.25MW/s - 34MW/s
2.25MW/s - 3.4MW/s
540kW/s -13.5MW/s 22kW/s - 560kW/s
55kW/s - 1.4MW/s
2.8MW/s-42.5MW/s
2.8MW/s-4.25MW/s
11kW/s - 280kW/s
560kW/s - 8.5MW/s
560kW/s - 850kW/s
Resistance Range 4 Slow band
Fast band ³ 3V
27MW/s - 408MW/s
27MW/s -40.8MW/s
1.12MW/s -17MW/s
1.12MW/s-1.7MW/s
Fast band <3V
within 35% of programmed value
Slew Rate Accuracy
Programmable Short
66mW max.
40mW typical
³ 20kW
200mW max.
100mW typical
³ 80kW
33mW max.
20mW typical
³ 20kW
33mW max.
25mW typical
³ 20kW
17mW max.
12mW typical
³ 20kW
Programmable Open
±300 Vdc between + or - input binding post and chassis ground
DC Isolation Voltage
Command Processing Time
Using discrete commands
Using List commands
List Dwell Characteristics
Range
3 ms
1 ms
0 - 10 s
1 ms
Resolution
Accuracy
5 ms
Measurement Time
1000 samples (default)
200 samples
100 samples
20 points
< 20 points
Measurement Bandwidth
GPIB Capabilities
20 ms (with specified measurement accuracy)
10 ms (with < 6% additional fixed offset)
9 ms (with < 10% additional fixed offset)
7 ms (with < 30% additional fixed offset)
7 ms (with > 30% additional fixed offset)
10kHz (rms)
SH1, AH1, T6, L4, SR1, RL1, DT1, CD1
1Slew rate bands are not programmable. When you program a slew rate value outside the indicated bands, the electronic load will
automatically adjust the slew rate to fit within the band that is closest to the programmed value. The slew rate accuracy specification is
only applicable to slew rates within the indicated bands. Additionally, when the transition from one setting to another is small, the small
signal bandwidth of the load limits the minimum transition time for all programmable slew rates (see chapter 2 for more information).
Below 3 volts, the maximum bandwidth of the electronic load is reduced by a factor of ten to one. For example, in the current range for
Model N3302A, the maximum slew rate is specified as 2.5MA/s, below 3 volts the maximum slew rate would be 250kA/s. Any slew rate
programmed between 2.5MA/s and 250kA/s would produce a slew rate of 250kA/s. Slew rates programmed slower than 250kA/s would
still correctly reflect their programmed value. If you are using transient mode to generate a high frequency pulse train, a reduced slew
rate might cause the load to never reach the upper programmed value before beginning the transition to the lower programmed value. So
even though the transient mode is still operational at lower voltages, a fast pulse train with large transitions may not be achievable.
73
Download from Www.Somanuals.com. All Manuals Search And Download.
A - Specifications
Table A-2. Supplemental Characteristics (continued)
N3302A
N3303A
N3304A
N3305A
N3306A
Ripple and Noise (20Hz-10MHz)
Current (rms / peak to peak)
Voltage (rms)
2mA / 20mA
5mV rms
1mA / 10mA
12mV rms
4mA / 40mA
6mV rms
4mA / 40mA
10mV rms
6mA / 60mA
8mV rms
Temperature Coefficients
Current Programming
Current Readback
Voltage Programming
Voltage Readback
Resistance Range 1 Programming
Resistance Range 2 Programming
Resistance Range 3 Programming
Resistance Range 4 Programming
Remote Sensing:
120ppm/°C + 0.5mA/°C
100ppm/°C + 0.5mA/°C
100ppm/°C + 0.5mV/°C
80ppm/°C + 0.33mV/°C
800ppm/°C + 0.4mΩ/°C
800ppm/°C + 4mΩ/°C
800ppm/°C + 40mΩ/°C
800ppm/°C + 4 Ω/°C
120ppm/°C + 0.1mA/°C 120ppm/°C + 1mA/°C
100ppm/°C + 0.1mA/°C 100ppm/°C + 1mA/°C
120ppm/°C + 1mA/°C
100ppm/°C + 1mA/°C
120ppm/°C + 2mA/°C
100ppm/°C + 2mA/°C
100ppm/°C + 2mV/°C
80ppm/°C + 1.33mV/°C 80ppm/°C + 0.33mV/°C 80ppm/°C + 0.8mV/°C
800ppm/°C + 1.6mΩ/°C 800ppm/°C + 0.2mΩ/°C 800ppm/°C + 0.4mΩ/°C 800ppm/°C + 0.1mΩ/°C
800ppm/°C + 16mΩ/°C 800ppm/°C + 2mΩ/°C 800ppm/°C + 4mΩ/°C 800ppm/°C + 1mΩ/°C
800ppm/°C + 160mΩ/°C 800ppm/°C + 20mΩ/°C 800ppm/°C + 40mΩ/°C 800ppm/°C + 10mΩ/°C
100ppm/°C + 0.5mV/°C 100ppm/°C + 1.5mV/°C 100ppm/°C + 0.5mV/°C
80ppm/°C + 0.33mV/°C
800ppm/°C + 16 Ω/°C
800ppm/°C + 2 Ω/°C
800ppm/°C + 4 Ω/°C
800ppm/°C + 1 Ω/°C
5Vdc between sense and load input
Reverse Current Capability
With input on
With input off
50A
20A
20A
10A
100A
40A
60A
30A
120A
60A
External Analog Programming
Programming Voltage
Voltage from 0 to full scale: 0-10V
Current from 0 to full scale: 0-10V
10 kHz (-3db frequency)
Bandwidth
Voltage Programming Accuracy2
Temperature Coefficient2
Current Programming Accuracy2
Temperature Coefficient2
External Monitor Ports
Monitor Voltage
0.5% + 12mV
0.5% + 48mV
0.5% + 12mV
0.5% + 30mV
0.5% + 12mV
100ppm/°C + 0.33mV/°C 100ppm/°C +1.33mV/°C 100ppm/°C +0.33mV/°C 100ppm/°C + 0.8mV/°C 100ppm/°C +0.33mV/°C
0.25% + 4.5mA
120ppm/°C + 0.5mA/°C
0.25% + 1.5mA
0.25% + 9mA
0.25% + 9mA
0.25% + 18mA
120ppm/°C + 2mA/°C
120ppm/°C + 0.1mA/°C 120ppm/°C + 1mA/°C
120ppm/°C + 1mA/°C
Voltage from 0 to full scale: 0-10V
Current from 0 to full scale: 0-10V
0.25% + 48mV 0.25% + 12mV 0.25% + 30mV
0.25% + 12mV
0.25% + 12mV
Voltage Monitor Accuracy
Temperature Coefficient
Current Monitor Accuracy
Temperature Coefficient
Drift
100ppm/°C + 0.33mV/°C 100ppm/°C +1.33mV/°C 100ppm/°C +0.33mV/°C 100ppm/°C + 0.8mV/°C 100ppm/°C +0.33mV/°C
0.1% + 4.5mA
120ppm/°C + 0.5mA/°C
0.1% + 1.5mA
0.1% + 9mA
0.1% + 9mA
0.1% + 18mA
120ppm/°C + 2mA/°C
120ppm/°C + 0.1mA/°C 120ppm/°C + 1mA/°C
120ppm/°C + 1mA/°C
Voltage Programming
Current Programming
Digital/Trigger Inputs
2mV
1mA
10mV
1mA
5mV
2mA
5mV
2mA
5mV
4mA
Vil=0.9V max at Iil=-1mA
Vih-3.15V min (pull-up resistor on input)
Vol=0.72V max at Iol=1mA
Voh=4.4V min at Ioh=-20uA
1 year
Digital/Trigger Outputs
Module Calibration Interval
(N3300A and N3301A Electronic Load mainframes do not require calibration)
Weight
Net
2.7kg (6lb)
4.1kg (9lb)
2.7kg (6lb)
4.1kg (9lb)
2.7kg (6lb)
4.1kg (9lb)
4.6kg (10lb)
6.8kg (15lb)
4.6kg (10lb)
6.8kg (15lb)
Shipping
2Applies to all ranges.
Table A-3. N3300A/N3301A Supplemental Characteristics
N3300A
N3301A
Operating Temperature Range
0°C to 55°C
Input Ratings
Operating range
100 - 250 Vac;
48 - 63 Hz
Input Current
Input VA
Inrush Current
Dimensions
Height
Depth
Width
4.2A @ 100-127 Vac ; 2.2 A @ 200-250 Vac
2.3A @ 100-250 Vac
230 VA
18A @ 115 Vac 36A @ 230 Vac
440 VA
38A
178 mm (7 in.), add 10 mm (0.4 in.) for removable feet
625 mm (24.6 in.), including input connectors on modules
425.5mm (16.75in.)
213 mm (8.4 in.)
Weight
Net
Shipping
13.2kg (29lb)
17.3kg (38lb)
7.3kg (16lb)
9.1kg (20lb)
74
Download from Www.Somanuals.com. All Manuals Search And Download.
B
Performance Test and Calibration Procedures
Introduction
This appendix contains test procedures for checking the operation and calibration of the Agilent N330xA Series
Electronic Load Modules. The tests are performed using the front panel keypad of the N3300A or N3301A
Mainframes. The required test equipment is listed in Table B-1 and sample performance test record cards are
included at the end of the performance test section. The performance tests confirm the Agilent N330xA Series
Electronic Load Modules meet all their published specifications.
The calibration ( adjustment ) procedures create new calibration constants to bring the current, voltage and
resistance values within specifications. Calibration cannot be performed using the front panel keypad, an IEEE-
488 ( GPIB ) controller is required. Note that the Electronic Load Mainframe does not require calibration.
Important Perform the verification procedures before calibrating your Electronic Load Module. If the
Electronic Load Module passes the verification procedures, the unit is operating within its
calibration limits and does not need to be re-calibrated.
If the electronic load requires service, refer to the list of Agilent Sales and Support Offices at the back of this
manual.
Equipment Required
The equipment listed in the following table, or the equivalent to this equipment, is required for verification and
calibration. An IEEE-488.2 controller is not required for verification.
Table B-1. Equipment Required
Equipment
Characteristics
0.1 Ohms @ 15 Amps, 0.04% @ 25 watts
0.01 Ohms @ 100 Amps, 0.04% @ 100 watts
0.001 Ohms @ 300 Amps, 0.04% @ 100 watts
DC accuracy 0.01% 5 digit
Recommended Models
Guildline 9230/15
Guildline 9230/100
Guildline 9230/300
Agilent 34401A or 3458A
Agilent 1146A
Current Shunts1
Voltmeter
Current Probe
Oscilloscope
Power Source
DC to 100Arms - DC to 100kHz
Agilent 54520A
61Vdc / 61A minimum ( N3302A, N3304A, N3306A )
20Vdc / 143A minimum ( N3305A, N3306A )
241Vdc / 10A minimum ( N3303A, N3305A )
GPIB ( IEEE 488.2 )
6032A or equivalent
6031A or equivalent
6035A
HP Series 200/300 or
PC with HP 82341 IEEE
488.2 Interface card and
Basic for Windows or equiv.
Controller
1 A 4-terminal resistor ( current shunt ) is required to eliminate output current measurement errors caused by voltage drops in the load
leads and connections.
75
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
Performance Tests
IMON Zero Verification
This test verifies that IMON Zero is within calibration. IMON zero must be in calibration to perform the tests in
this section. If the IMON value is not within calibration, go to the calibration section of this appendix and
calibrate IMON zero, IPROG, current programming and readback and resistance programming and readback.
See figure B-1 for IMON Zero test setup.
Action
Normal Result
1.
Turn off load module and connect DVM to IMON as per figure B-1.
Input terminals must be open ( no power source connected )
Turn on Load – press [ Recall } – scroll to *RST – press [Enter ]
Read IMON voltage from DVM.
2.
3.
Voltage reading less than 0.166mV for
models N3302A through N3305A
Voltage reading less than 0.320mV
for model N3306A
4.
If IMON in step 3 in not within specification then calibrate IMON
zero, IPROG, Current mode and Resistance mode.
CC Mode Tests
These tests verify that the module operates in the CC mode and that IMON, current programming, and readback
to the front panel display are within specification. Values read back over the GPIB are the same as those
displayed on the front panel. To read back the current value via the GPIB, use MEAS:CURR:ACDC?
Current Programming and Measurement Accuracy Test
This test verifies that the current programming and measurement accuracy are within specification. If the test
readings are out of tolerance the module may require calibration. If the test readings significantly disagree with
the specified values or no readings can be obtained go to the Turn-On Checkout procedure to verify module
operation.
Make a copy of the module test card for the model to be tested to record test values.
1. Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.
2. Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].
3. Turn on the power source. Set the power source voltage and current to the values listed in the following table.
N3302A
5V
33A
N3303A
20V
12A
N3304A
5V
61A
N3305A
8V
61A
N3306A
5V
130A
Power Source Voltage Setting
Power Source Current Setting
4. Checking high current range, Low input current. Press [ CURR ] [ 1 ] then [ Enter ]. Wait 10 seconds then record the
actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test card under
high current range low current.
5. Checking high current range, maximum input current. Press [ CURR ] [ select current from following table] [Enter ].
N3302A
30
N3303A
10
N3304A
60
N3305A
60
N3306A
120
Wait 30 seconds, then record the actual input current ( DVM reading / current shunt resistance ) and front panel current
reading on the test card under high current range high current.
76
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
6. Press [ CURR ] [ 1 ] ( except N3303A press [ CURR ] [ 0.1 ] ). Press [ CURR ]. Scroll until display reads C:RANG,
press [ 1 ] [ Enter ]. Press shift key, [ Sense ] scroll till display reads S:C:RNG press [ 1 ] [ Enter ].
7. Checking low current range, low input current Wait 10 seconds then record the actual input current ( DVM reading /
current shunt resistance ) and front panel current reading on the test card under low current range low current.
8. Checking low current range, maximum input current. Press [ CURR ] [ select current from following table ] [ Enter ].
N3302A
3
N3303A
1
N3304A
6
N3305A
6
N3306A
12
Wait 30 seconds then record the actual input current ( DVM reading / current shunt resistance ) and front panel
current reading on the test card under high current range high current.
CC Mode Regulation Test
This test verifies the Input Current remains within specification when the input voltage is changed from a low
voltage to rated voltage.
1. Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.
2. Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ]
3. Press [ CURR ] [ select current from following table ] [ Enter ].
N3302A
2.5
N3303A
1
N3304A
5
N3305A
3.3
N3306A
10
4. Turn on the power source. Set the power source voltage and current to the values listed in following table.
N3302A
4V
5A
N3303A
4V
5A
N3304A
4V
10A
N3305A
4V
10A
N3306A
4V
20A
Power Source Voltage Setting
Power Source Current Setting
5. Wait 10 seconds, then record input current reading ( DVM reading / current shunt ).
6. Reset the power source voltage level to value listed in following table.
N3302A
60V
N3303A
240V
N3304A
60V
N3305A
150V
N3306A
60V
Power Source Voltage Setting
7. Wait 10 seconds, then record input current reading ( DVM reading / shunt resistance ).
8. Subtract reading in step ‘7’ from step ‘5’. Absolute value of difference should be less then specification.
CV Mode Tests
These tests verify that the module operates in the CV mode and that voltage programming and readback to the
front panel display are within specification. Values read back over the GPIB are the same as those displayed on
the front panel. To read back the voltage value via the GPIB, use the command MEAS:VOLT:ACDC?
Voltage Programming and Measurement Accuracy Test
This test verifies that voltage programming and voltage measurement accuracy are within specification. Make a
copy of the module test card for the model being tested to record test values.
1. Connect the Electronic Load, Power source and DVM as shown in figure B-3. Be sure + sense and – sense are
connected to the +/- input terminals and the remote/local button is in the remote position.
2. Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].
3. Press [ Func ]. Scroll until display reads FUNC VOLT then press [ Enter ].
77
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
4. Turn on the power source. Set the power source voltage and current to the values listed in the following table.
N3302A
61V
2A
N3303A
246V
0.6A
N3304A
61V
3A
N3305A
152V
2A
N3306A
61V
3A
Power Source Voltage Setting
Power Source Current Setting
5. Checking high voltage range, high voltage point. Wait 10 seconds then record actual input voltage and front panel
voltage reading on test record card under high voltage range, high voltage.
6. Checking high voltage range, low voltage point. Press [ VOLT ] [ 3 ] [ Enter ]. Wait 10 seconds, then record actual
input voltage and front panel readings on test record card under high voltage range, low voltage.
7. Checking low voltage range, low voltage point. Press [ VOLT ]. Scroll until display reads V:RANG Press [ 1 ]
[
Enter ]. Press shift key then [ Sense ] scroll till display reads S:V:RNG, press [ 3 ]. Wait 10 seconds then record actual
input voltage and front panel readings on test record card under low voltage range, low voltage.
8. Checking low voltage range, high voltage point. Press [ VOLT ] [ select voltage from following table ] [ Enter ].
N3302A
6V
N3303A
24V
N3304A
6V
N3305A
15V
N3306A
6V
9. Wait 10 seconds then record actual input voltage and front panel readings on test record card under low voltage range,
low voltage.
CV Mode Regulation Test
This test verifies the Input Voltage remains within specification when the input current is changed from a low
current to rated current.
1. Connect Electronic Load, power source and DVM as shown in figure B-3. Be sure + sense and – sense are connected
to the +/- input terminals and the remote/local button is in the remote position.
2. Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].
3. Press [ Func ]. Scroll until display reads FUNC VOLT, then press [ Enter ].
4. Press [ Volt ] [ select voltage from following table ] [ Enter ].
N3302A
5V
N3303A
25V
N3304A
5V
N3305A
8.3V
N3306A
5V
5. Turn on power source. Set power source voltage and current to values listed in following table.
N3302A
6V
1A
N3303A
30V
0.4A
N3304A
6.5V
1A
N3305A
10V
1A
N3306A
6.6V
1A
Power Source Voltage Setting
Power Source Current Setting
6. Wait 10 seconds and record input voltage reading.
7. Reset power source current level to value listed in following table.
N3302A
30A
N3303A
10A
N3304A
60A
N3305A
60A
N3306A
120A
Power Source Current Setting
8. Wait 10 seconds and record input voltage reading.
9. Subtract reading in step ‘8’ from reading in step ‘6’. Absolute value of difference should be less then specification.
78
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
CR Mode Tests
These tests verify that the module operates in CR mode and the resistance programming is within specification.
The programmed resistance values are checked by recording the voltage across the current monitor resistor and
the input voltage (voltage at the modules + and – sense terminals ), then calculating the resistance as follows;
Load Resistance = Input Voltage / ( voltage across current monitor / monitor resistor value )
Resistance Range 1 Programming Accuracy
1. Connect Electronic Load, Power source and DVM’s as shown in figure B-4. Be sure + sense and – sense are
connected to the +/- input terminals and the remote/local button is in the remote position.
2. Turn on the Electronic Load and press [ Recall ], scroll till display reads *RST and press [ Enter ].
3. Press [ Func ]. Scroll until display reads FUNC RES then press [ Enter ].
4. Press [ RES ]. Scroll until display reads RES:RANG, press [ 1 ] then [ Enter ].
5. Turn on the power source. See power source voltage and current values as listed in following table.
N3302A
24V
7A
N3303A
48V
7A
N3304A
12V
10A
N3305A
30V
10A
N3306A
15V
20A
Power Source Voltage Setting
Power Source Current Setting
6. Wait 30 seconds. Checking resistance range 1, high resistance point. Calculate and record resistance on test card.
Input resistance = Input Voltage / ( DVM voltage reading from current shunt/ current shunt resistance).
7. Checking resistance range 1, low resistance point. Reset power source voltage to following values.
N3302A
4V
25A
N3303A
6V
10A
N3304A
10V
30A
N3305A
5V
50A
N3306A
6V
70A
Power Source Voltage Setting
Power Source Current Setting
8. Press [ RES ] [ select resistance from following table ] [ Enter ].
N3302A
0.2Ω
N3303A
1.2 Ω
N3304A
0.5 Ω
N3305A
0.125 Ω
N3306A
0.100 Ω
9. Wait 30 seconds. Calculate and record resistance range 1, low resistance point.
Resistance Range 2 Programming Accuracy
1. Press [ RES ]. Scroll until display reads RES:RANG, For models N3302A, N3304A, N3305A and N3306A press
[ 9 ] then [ Enter ]. For model N3303A Press [ 5 ] [ 0 ] then [ Enter ].
2. Checking resistance range 2, low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
3.6 Ω
44 Ω
1.8 Ω
4.5 Ω
0.9 Ω
3. Set power source voltage and current values as listed in following table.
N3302A
20V
7A
N3303A
96V
3A
N3304A
6V
7A
N3305A
30V
15A
N3306A
6V
8A
Power Source Voltage Setting
Power Source Current Setting
4. Wait 15 seconds. Calculate and record resistance range 2 low resistance point.
5. Checking resistance range 2, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
40 Ω
480 Ω
20 Ω
50 Ω
10 Ω
79
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
6. Set power source voltage and current values as listed in following table.
N3302A
24V
7A
N3303A
96V
3A
N3304A
12V
7A
N3305A
30V
15A
N3306A
12V
8A
Power Source Voltage Setting
Power Source Current Setting
7. Wait 15 seconds. Calculate and record resistance range 2 high resistance point
Resistance Range 3 Programming Accuracy
1. Press [ RES ]. Scroll until display reads RES:RANG, for models N3302A, N3304A, N3305A and N3306A press
[ 1 ] [ 0 ] [ 0 ] then [ Enter ]. For model N3302A press [ 5 ] [ 0 ] [ 0 ] then [ Enter ].
2. Checking resistance range 3, low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
36 Ω
440 Ω
18 Ω
45 Ω
9 Ω
3. Set power source voltage and current values as listed in following table.
N3302A
40V
3.5A
N3303A
96V
1A
N3304A
20V
3A
N3305A
50V
3.5A
N3306A
20V
3.5A
Power Source Voltage Setting
Power Source Current Setting
4. Wait 7 seconds. Calculate and record resistance range 3 low resistance point.
5. Checking resistance range 3, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
400 Ω
4800 Ω
200 Ω
500 Ω
100 Ω
6. Set power source voltage and current values as listed in following table.
N3302A
40V
3.5A
N3303A
96V
1A
N3304A
20V
3A
N3305A
50V
3.5A
N3306A
20V
3.5A
Power Source Voltage Setting
Power Source Current Setting
7. Wait 7 seconds. Calculate and record resistance range 3 high resistance point.
Resistance Range 4 Programming Accuracy
1. Press [ RES ]. Scroll until display reads RES:RANG, For models N3302A, N3304A, N3305A and N3306A press
[ 1 ] [ 0 ] [ 0 ] [ 0 ] then [ Enter ]. For model N3303A press [ 5 ] [ 0 ] [ 0 ] [ 0 ] then [ Enter ].
2. Checking resistance range 4 low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
360 Ω
4400 Ω
180 Ω
450 Ω
90 Ω
3. Set power source voltage and current values as listed in following table.
N3302A
50V
3.5A
N3303A
96V
1A
N3304A
20V
3A
N3305A
50V
3.5A
N3306A
20V
3.5A
Power Source Voltage Setting
Power Source Current Setting
4. Wait 3 seconds. Calculate and record resistance range 3 low resistance point.
5. Checking resistance range 4, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].
N3302A
N3303A
N3304A
N3305A
N3306A
2000 Ω
12,000 Ω
2000 Ω
2500 Ω
1000 Ω
80
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
6. Set power source voltage and current values as listed in following table.
N3302A
50V
3.5A
N3303A
96V
1A
N3304A
20V
3A
N3305A
50V
3.5A
N3306A
20V
3.5A
Power Source Voltage Setting
Power Source Current Setting
7. Wait 3 seconds. Calculate and record resistance range 4 high resistance point
Transient Generator Mode Tests
This test verifies that the transient generator frequency and duty cycle circuits are within specification.
Frequency and Duty Cycle Accuracy
1. Connect Electronic Load, Power Source, Current Probe and Oscilloscope as shown in figure B-5.
2. Turn on the Electronic Load and press [ Recall ], scroll till display reads *RST and press [ Enter ].
3. On the Electronic Load front panel keypad:
a. Press [ Curr ] scroll until display reads C:TLEV. Press 10 then [ Enter ].
b. Press [ Tran ] scroll until display reads T:FREQ. Press 1000 then [ Enter ].
c. Press [ Tran ], press [ Input ] key until display reads TRAN ON then press [ Enter ].
4. Turn on power source and set for 10 volts and15 amperes.
5. Turn on and set oscilloscope:
a. Adjust Vertical for 1:1 probe.
b. Adjust Vertical for 20 mV/div.
c. Adjust horizontal for 500 uS/div
d. On measurement keypads Press shift key and select frequency and duty cycle for channel 1.
6. Connect current probe to channel 1 on oscilloscope, turn on current probe and set to 10mV/div.
7. Read measurements from oscilloscope screen and enter on test card. The oscilloscope may require 2 or 3 complete
waveforms to make measurements.
81
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
Pulse Width Accuracy
This test verifies that the transient generator pulse width accuracy is within specification. Note: Use the GPIB controller to
generate a continuous pulse train.
1. Connect Electronic Load, Power Source, Current Probe and Oscilloscope as shown in figure B-5.
2. If using a controller, connect controller to Electronic Load mainframe rear panel GPIB connector.
3. Turn on the Electronic Load and press [Recall]. Scroll to until *RST is displayed and press [Enter].
4. Turn on the Power Source and program the voltage to 10 V and the current to 15 A.
5. Turn on the Oscilloscope and set to:
a. Ch 1 vertical deflection, 15 mV/div.
b. 1:1 probe, DC coupled, 50 Ohms.
c. Position or offset to 50mV.
d. Horizontal time base to 20 uS/div.
e. Trigger on positive edge at about 50mV.
6. Turn on Current Probe to the 10mV/A position.
7. Run the following program for a continuous pulse train or skip to step 8 to manually generate pulses.
10 OUTPUT 705;”*RST”
20 OUTPUT 705;”TRAN:MODE PULS”
30 OUTPUT 705;”CURR:TLEV 10”
40 OUTPUT 705;”TRAN:TWID 100E-6”
50 OUTPUT 705;”TRAN ON”
60 OUTPUT 705;”TRIG:IMM”
70 WAIT 1000 ! wait 1 second
80 GOTO 60 ! loop back to line 60 to for continuous pulse train.
90 END
8. For front panel single pulse operation: (Note: skip to step 9 if using a controller to generate a continuous pulse train)
a. Press [Current] and scroll to C:TLEV. On the Entry keypad press 10 [Enter].
b. Press [Tran] and scroll to T:TWID. On the Entry keypad enter 1E-4 then [Enter] to generate a pulse width of
100uS.
c. Press [Tran] and scroll to T:MODE. On the Entry keypad scroll to T:MODE PULS then press [Enter] .
d. Press [Tran]. Using the Entry keypad scroll to ON and press [Enter] to activate the transient mode.
e. Press [Trigger Control] and then [Enter] to initiate the trigger system.
f. Press [Shift] and then [Trigger] to generate a single current pulse from zero to 10 A..
9. Adjust the oscilloscope as needed to display a complete pulse waveform. Generate pulses as needed.
10. Measure width of the 10 A, 100 uS nominal pulse between the 50% points of waveform and record the results on
Performance Test Record. Note: Use the built-in measurement capabilities of the oscilloscope, if available, for the most
accurate results.
82
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
Agilent N3302A Verification Test Record
Test Description
Minimum
Results
Maximum
Specification
Specification
Constant Current Mode Tests
30 Ampere Range Programming and Readback
Low Current ( 1 A )
Front Panel Display
High Current ( 30 A )
Front Panel Display
0.989 A
Aout – 6.5mA
29.960 A
_______A
_______A
_______A
_______A
1.011 A
Aout + 6.5mA
30.040 A
Aout – 21mA
Aout + 21mA
3 Ampere Range Programming and Readback
Low Current ( 1 A )
0.994 A
Aout – 3.5mA
2.992 A
_______A
_______A
_______A
_______A
1.006 A
Aout + 3.5mA
3.008 A
Front Panel Display
High Current ( 3 A )
Front panel Display
Aout – 4.5mA
Aout + 4.5mA
Constant Current Regulation Test
2.5 Amp Regulation at input voltage delta 4V to 60V
______ mA
Less then 10mA
Voltage Mode Tests
60 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.989 V
Vout – 9.5mV
59.932 V
_______V
_______V
_______V
_______V
3.011 V
Vout + 9.5mV
60.068 V
Front Panel Display
High Voltage ( 60 V )
Front Panel Display
Vout – 3.8mV
Vout + 3.8mV
6 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.9905 V
Vout – 4.5mV
5.991 V
_______V
_______V
_______V
_______V
3.0095 V
Vout + 4.5mV
6.009 V
Front Panel Display
High Voltage ( 6 V )
Front Panel Display
Vout – 6.0mV
Vout + 6.0mV
Voltage Mode Regulation Test
5V regulation with input current delta 1A to 30A
______mV
Less then 5mV
Resistance Mode Tests
Resistance Range 1
High Resistance Point
Low Resistance Point 0.1 Ω
4 Ω
3.944 Ω
0.0756 Ω
______ Ω
______ Ω
4.056 Ω
0.1244 Ω
Resistance Range 2
Low Resistance Point 3.6 Ω
High Resistance Point 40 Ω
3.466 Ω
38.520 Ω
______Ω
______Ω
3.734 Ω
41.480 Ω
Resistance Range 3
Low Resistance Point 36 Ω
High Resistance Point 400 Ω
32.880 Ω
271.76 Ω
______Ω
______ Ω
39.120 Ω
528.24 Ω
Resistance Range 4
Low Resistance Point 360 Ω
High Resistance Point 2000 Ω
270 Ω
1200 Ω
______Ω
______Ω
720 Ω
10,000 Ω
Transient Generator Tests
995 Hz.
Frequency Range Tests (1000 Hz test)
_____ Hz
______ %
______s
1005 Hz
51%
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz
49%
Pulse Width Test (100µS wide; 10 A pulse)
99µs
101µs
83
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
Agilent N3303A Verification Test Record
Test Description
Minimum
Results
Maximum
Specification
Specification
Constant Current Mode Tests
10 Ampere Range Programming and Readback
Low Current ( 1 A )
Front Panel Display
High Current ( 10 A )
Front Panel Display
0.9915 A
Aout – 5.5mA
9.9825 A
_______A
_______A
_______A
_______A
1.041 A
Aout + 5.5mA
10.0175 A
Aout – 10mA
Aout + 10mA
1 Ampere Range Programming and Readback
Low Current ( 0.1 A )
Front Panel Display
0.0959 A
Aout – 2.55 mA
0.995 A
_______A
_______A
_______A
_______A
0.1041 A
Aout + 2.55mA
1.005 A
High Current ( 1 A )
Front panel Display
Aout – 3mA
Aout + 3mA
Current Regulation test
1.0 Amp Regulation at input voltage delta 4V to 240V
______mA
Less then 8mA
Voltage Mode Tests
240 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.957 V
Vout – 21mV
239.720 V
_______V
_______V
_______V
_______V
3.043 V
Vout + 21mV
240.280 V
Front Panel Display
High Voltage ( 240 V )
Front Panel Display
Vout – 140mV
Vout + 140mV
24 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.987 V
Vout – 11mV
23.966 V
_______V
_______V
_______V
_______V
3.013 V
Vout + 11mV
24.034 V
Front Panel Display
High Voltage ( 24 V )
Front Panel Display
Vout – 22mV
Vout + 22mV
Voltage Mode Regulation Test
25V regulation with input current delta 0.4A to 10A
______mV
Less then 10mV
Resistance Mode Tests
46.46 Ω
Resistance Range 1
High Resistance Point 48 Ω
Low Resistance Point 1.2 Ω
______ Ω
______ Ω
49.54 Ω
1.312 Ω
1.088 Ω
Resistance Range 2
Low Resistance Point 44 Ω
High Resistance Point 480 Ω
41.30 Ω
383.50 Ω
______Ω
______Ω
46.70 Ω
576.5 Ω
Resistance Range 3
Low Resistance Point 440 Ω
High Resistance Point 4800 Ω
325 Ω
2640 Ω
______Ω
______Ω
598.4 Ω
38,400 Ω
Resistance Range 4
Low Resistance Point 4400 Ω
High Resistance Point 12,000 Ω
1496 Ω
4080 Ω
______Ω
______Ω
39,600 Ω
108,000 Ω
Transient Generator Tests
995 Hz.
Frequency Range Tests (1000 Hz test)
_____ Hz
______ %
______s
1005 Hz
51%
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz
49%
Pulse Width Test (100µS wide; 10 A pulse)
99µs
101µs
84
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
Agilent N3304A Verification Test Record
Test Description
Minimum
Results
Maximum
Specification
Specification
Constant Current Mode Tests
60 Ampere Range Programming and Readback
Low Current ( 1 A )
Front Panel Display
High Current ( 60 A )
Front Panel Display
0984 A
Aout –10.5mA
59.925 A
_______A
_______A
_______A
_______A
1.016 A
Aout + 10.5mA
60.075 A
Aout – 40mA
Aout +40mA
6 Ampere Range Programming and Readback
Low Current ( 1 A )
0.9915
Aout – 5.5mA
5.9865 A
_______A
_______A
_______A
_______A
1.0085
Aout + 5.5mA
6.0135 A
Front Panel Display
High Current ( 6 A )
Front panel Display
Aout – 8mA
Aout + 8mA
Current Regulation Test
5 Amp Regulation at input voltage delta 4V to 60V
______mA
Less then 10 mA
Voltage Mode Tests
60 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.957 V
Vout – 21mV
59.932 V
_______V
_______V
_______V
_______V
3.043 V
Vout + 21mV
60.068 V
Front Panel Display
High Voltage ( 60 V )
Front Panel Display
Vout – 140mV
Vout + 140mV
6 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.9905 V
Vout – 4.5mV
5.991 V
_______V
_______V
_______V
_______V
3.0095 V
Vout + 4.5mV
6.009 V
Front Panel Display
High Voltage ( 6 V )
Front Panel Display
Vout – 6mV
Vout + 6mV
Voltage Mode Regulation Test
5V regulation with input current delta 1A to 60A
______mV
Less then 10mV
Resistance Mode Tests
Resistance Range 1
High Resistance Point 2 Ω
Low Resistance Point 0.50 Ω
1.980 Ω
0.486 Ω
______ Ω
______ Ω
2.020 Ω
0.514 Ω
Resistance Range 2
Low Resistance Point 1.8 Ω
High Resistance Point 20 Ω
1.733 Ω
19.360 Ω
______Ω
______ Ω
1.867 Ω
20.640 Ω
Resistance Range 3
Low Resistance Point 18 Ω
High Resistance Point 200 Ω
16.800 Ω
159.88 Ω
______Ω
______Ω
19.20 Ω
240.12 Ω
Resistance Range 4
Low Resistance Point
High Resistance Point
180 Ω
2000 Ω
144 Ω
1000 Ω
______Ω
______Ω
298.8 Ω
42,000 Ω
Transient Generator Tests
995 Hz.
Frequency Range Tests (1000 Hz test)
_____ Hz
______ %
______s
1005 Hz
51%
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz
49%
Pulse Width Test (100µS wide; 10 A pulse)
99µs
101µs
85
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
Agilent N3305A Verification Test Record
Test Description
Minimum
Results
Maximum
Specification
Specification
Constant Current Mode Tests
60 Ampere Range Programming and Readback
Low Current ( 1A )
Front Panel Display
High Current ( 60 A )
Front Panel Display
0.984 A
Aout – 10.5mA
59.925 A
_______A
_______A
_______A
_______A
1.016 A
Aout + 10.5mA
60.075 A
Aout – 40mA
Aout + 40mA
6 Ampere Range Programming and Readback
Low Current ( 1A )
0.9915 A
Aout – 5.5mA
5.9865 A
_______A
_______A
_______A
_______A
1.0085 A
Aout + 5.5mA
6.0135 A
Front Panel Display
High Current ( 6 A )
Front panel Display
Aout – 8mA
Aout + 8mA
Current Regulation Test
3.3 Amp Regulation at input voltage delta 4V to 60V
______mA
Less then 10mA
Voltage Mode Tests
150 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.977 V
Vout – 17.5mV
149.830 V
_______V
_______V
_______V
_______V
3.023 V
Vout + 17.5mV
150.170 V
Front Panel Display
High Voltage ( 150 V )
Front Panel Display
Vout – 9.1mV
Vout + 9.1mV
15 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.987 V
Vout – 9.5mV
14.975 V
_______V
_______V
_______V
_______V
3.013 V
Vout + 9.5mV
15.025 V
Front Panel Display
High Voltage ( 15 V )
Front Panel Display
Vout – 15.5mV
Vout + 15.5mV
Voltage Regulation Test
8.3V regulation with input current delta 1A to 60A
______mV
Less then 10mV
Resistance Mode Tests
Resistance Range 1
High Resistance Point 5 Ω
Low Resistance Point 0.125 Ω
4.936 Ω
0.1005 Ω
______ Ω
______ Ω
5.064 Ω
0.1495 Ω
Resistance Range 2
Low Resistance Point 4.5 Ω
High Resistance Point 50 Ω
4.352 Ω
48.42 Ω
______Ω
______Ω
4.647 Ω
51.58 Ω
Resistance Range 3
Low Resistance Point 45 Ω
High Resistance Point 500 Ω
40.50 Ω
350 Ω
______Ω
______Ω
56.25 Ω
1250 Ω
Resistance Range 4
Low Resistance Point 450 Ω
High Resistance Point 2500 Ω
292 Ω
1000 Ω
______Ω
______Ω
1575 Ω
252 k Ω
Transient Generator Tests
Frequency Range Tests (1000 Hz test)
995 Hz.
_____ Hz
______ %
______s
1005 Hz
51%
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz
Pulse Width Test (100µS wide; 10 A pulse)
49%
99µs
101µs
86
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
Agilent N3306A Verification Test Record
Test Description
Minimum
Results
Maximum
Specification
Specification
Constant Current Mode Tests
120 Ampere Range Programming and Readback
Low Current ( 1A )
Front Panel Display
High Current ( 120 A )
Front Panel Display
0.9615 A
Aout – 20.5mA
119.8425 A
_______A
_______A
_______A
_______A
1.0385 A
Aout + 20.5mA
120.1575 A
Aout – 80mA
Aout + 80mA
12 Ampere Range Programming and Readback
Low Current ( 1A )
0.9840A
Aout – 10.5mA
11.973 A
_______A
_______A
_______A
_______A
1.0160 A
Aout + 10.5mA
12.027 A
Front Panel Display
High Current ( 12 A )
Front panel Display
Aout – 16mA
Aout + 16mA
Current Regulation Test
10 Amp Regulation at input voltage delta 4V to 60V
______mA
Less then 10mA
Voltage Mode Tests
60 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.989 V
Vout – 21mV
59.932 V
_______V
_______V
_______V
_______V
3.011 V
Vout + 21mV
60.068 V
Front Panel Display
High Voltage ( 60 V )
Front Panel Display
Vout – 38mV
Vout + 38mV
6 Volt Range Programming and Readback
Low Voltage ( 3 V )
2.9905 V
Vout – 4.5m
5.991 V
_______V
_______V
_______V
_______V
3.0095 V
Vout + 4.5m
6.009 V
Front Panel Display
High Voltage ( 6 V )
Front Panel Display
Vout – 6mV
Vout + 6mV
Voltage Regulation Test
5V regulation with input current delta 1A to 120A
______mV
Less then 20mV
Resistance Mode Tests
Resistance Range 1
High Resistance Point 1 Ω
Low Resistance Point 0.100 Ω
0.990 Ω
0.0936 Ω
______ Ω
______ Ω
1.010 Ω
0.1064 Ω
Resistance Range 2
Low Resistance Point 0.9 Ω
High Resistance Point 10 Ω
0.8665 Ω
9.680 Ω
______Ω
______Ω
0.9335 Ω
10.32 Ω
Resistance Range 3
Low Resistance Point 9 Ω
High Resistance Point 100 Ω
8.400 Ω
79.940 Ω
______Ω
______Ω
9.60 Ω
120.06 Ω
Resistance Range 4
Low Resistance Point 90 Ω
High Resistance Point 1000 Ω
72 Ω
500 Ω
______Ω
______ Ω
149.4 Ω
21,000 Ω
Transient Generator Tests
Frequency Range Tests (1000 Hz test)
995 Hz.
_____ Hz
______ %
______s
1005 Hz
51%
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz
Pulse Width Test (100µS wide; 10 A pulse)
49%
99µs
101µs
87
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
N330xA Module (top view)
Location of I MON zero pins
on single-slot modules
IMON_INV
GND_AQS
A COM
I MON
Agilent 3458 or
34401 DVM
-
IMON_INV
GND_AQS
+
Ribbon
Cable
INPUT TERMINALS
MUST BE OPEN
Figure B-1. I MON ZERO CALIBRATION
N330xA Module (rear view)
+
I MON
Agilent 3458 or
34401 DVM
+
A11
_
A COM
_
A9
SHUNT
Power
Source
EXT PRG
EXT COM
A8
A7
Iprog Power Source +
& DVM
-
Agilent 3458 or
34401 DVM
If 3 DVM's are not available - move DVM
between I MON, EXT PRG and current shunt as
required by program
Figure B-2. I MON / EXT PRG and CURRENT
CALIBRATION
88
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
N330xA Module
+
+
_
_
Power
Source
Agilent 3458 or
34401 DVM
-S
+S
Be sure sense switch is in RMT position
Figure B-3. VOLTAGE CALIBRATION
N330xA Module
+
+
_
SHUNT
-S
_
Power
Source
Agilent 3458 or
34401 DVM
+S
Agilent 3458 or
34401 DVM
If 2 DVM's are not available - move DVM between
+/- sense and shunt as required by program
Be sure sense switch is in RMT position
Figure B-4. RESISTANCE CALIBRATION
+
+
_
_
Power
Source
Current
Probe
Oscilloscope
Figure B-5. FREQUENCY / DUTY CYCLE TEST
89
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
Calibration
Parameters Calibrated
The following parameters may be calibrated.
•
•
•
•
•
External Current Monitor ( Imon )
External Current Programming ( Iprog )
Input Current
Input Voltage
Input Resistance
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the current,
voltage or resistance and proceed to “Saving Calibration Constants”. However, Imon and Iprog must be
calibrated before Input Current is calibrated and both Imon and Input Current must be calibrated before Input
Resistance can be calibrated.
The following section lists calibration programs to re-calibrate ( adjust ) those specifications which are not
within specification. The programs listed are for use with either an HP Series 200/300 computer using HP Basic
( RMB ) or a PC equipped with an IEEE-488.2 interface card an HP Basic for Windows.
The variables to be used in the programs are listed in Table B-2 Variables. The connections for the test
equipment, current shunts and DVMs are shown in Figures B-1 for IMON ZERO calibration, B-2 for IMON,
IPROG and CURRENT calibration, B-3 for VOLTAGE calibration and B-4 for RESISTANCE calibration. A
listing of all programming commands and their explanations can be found in the Programming Guide, part
number 5964-8198.
Table B-2 Variables and Power Source Settings
VARIABLE
N3302A
N3303A
N3304A
N3305A
N3306A
CURRENT MODE VOLTAGE / CURRENT SETTINGS
Power source voltage setting
Power source current setting
5 V
33 A
20 V
12 A
5 V
61 A
8 V
61 A
5 V
130 A
VOLTAGE MODE VOLTAGE / CURRENTSETTINGS
Power source voltage setting
Power source current setting
61V
2 A
246 V
0.6 A
61V
3 A
152V
2 A
61V
3 A
RESISTANCE MODE VARIABLE and VOLTAGE / CURRENT SETTINGS
Resistance Range 1
Power source voltage setting for P1
Power source voltage setting for P2
Power source current setting
Resistance Range 2
24V
6V
20A
80V
15V
7A
12V
6V
40A
28V
4.5V
15A
12V
6V
80A
< resistance range 2 variable >
Power source voltage setting for P1
Power source voltage setting for P2
Power source current setting
Resistance Range 3
40 Ω
22V
6V
480 Ω
50V
30V
20 Ω
20V
12V
15A
50 Ω
45V
9V
10 Ω
20V
12V
15A
3.5A
3.5A
3.5A
< resistance range 3 variable >
Power source voltage setting for P1
Power source voltage setting for P2
Power source current setting
Resistance Range 4
400 Ω
30V
15V
2400 Ω
150V
75V
200 Ω
60V
60V
500 Ω
45V
15V
3.5
100 Ω
60V
60V
7A
3.5A
3.5A
3.5A
< resistance range 4 variable >
Power source voltage setting for P1
Power source voltage setting for P2
Power source current setting
2000 Ω
60V
30V
12000 Ω
150V
150V
2000 Ω
60V
60V
2500 Ω
150V
115V
3.5A
1000 Ω
18V
18V
3.5A
3.5A
3.5A
13.5A
90
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
IMON, IPROG and CURRENT Calibration Program
10
20
30
40
50
60
70
!
!
!
!
!
!
!
This program will calibrate Imon, Iprog and Current
for load modules N3302A, N3303A, N3304A, N3305A and N3306A.
last update 9/6/00
Edit the interface address on line 100 if different then 705
80
90
ASSIGN @Ld TO 705
!
100
110
120
130
140
150
160
170
180
190
200
OUTPUT @Ld;"*RST"
OUTPUT @Ld;"CAL:STAT ON"
OUTPUT @Ld;"FUNC CURR"
OUTPUT @Ld;"CURR:RANG MAX"
OUTPUT @Ld;"SENS:CURR:RANG MAX"
OUTPUT @Ld;"INP:STAT ON"
!
! *******
Calibrating Imon Zero *******
!
PRINT TABXY(10,8),"See figure B-1 for Imon calibration"
PRINT TABXY(10,10),"Module Input Terminals must be open - disconnect power
source"
210
220
230
240
250
260
270
280
290
300
PRINT TABXY(10,12),"Connect DVM to IMON terminals"
DISP "Press CONT when ready to calibrate Imon"
PAUSE
CLEAR SCREEN
!
OUTPUT @Ld;"CAL:IMON:LEV P1"
PRINT TABXY(20,20);"Calibrating IMON P1 - 5 SECOND WAIT"
WAIT 5
CLEAR SCREEN
INPUT "Measure IMON and enter value in volts - OBSERVE VOLTAGE
POLARITY",Vimonp1
310
320
330
340
350
360
370
380
!
OUTPUT @Ld;"CAL:DATA ";Vimonp1
!
OUTPUT @Ld;"CAL:IMON:LEV P2"
PRINT TABXY(20,20);"Calibrating IMON P2 - 5 SECOND WAIT"
WAIT 5
CLEAR SCREEN
INPUT "Measure IMON and enter value in volts - OBSERVE VOLTAGE
POLARITY",Vimonp2
390
400
410
420
430
440
450
460
470
!
OUTPUT @Ld;"CAL:DATA ";Vimonp2
!
! *******
!
PRINT TABXY(10,8),"See figure B-2 for Imon and Iprog calibration"
PRINT TABXY(10,10),"Connect module input terminals to power source"
Calibrating Imon and Iprog gain *******
PRINT TABXY(10,12),"See Table B-2 Current Mode Voltage /Current Settings"
PRINT TABXY(10,14),"Set power source voltage and current for model to be
calibrated"
480
490
500
510
520
530
PRINT TABXY(10,16),"Set IPROG power source to 1 volt"
!
DISP "Press CONT when ready to calibrate Imon and Iprog"
PAUSE
CLEAR SCREEN
!
91
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
540
550
560
570
580
590
600
610
620
630
640
650
660
670
680
690
700
710
720
730
740
750
760
770
780
790
800
810
820
830
840
850
860
870
880
890
900
910
920
930
940
950
960
970
980
990
INPUT "Enter value of high range current shunt in ohms",Rshunt
DISP "You have entered ";Rshunt;" ohms"
INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
IF UPC$(Y$)="Y" OR Y$="" THEN 590
IF UPC$(Y$)="N" OR Y$="n" THEN 520
CLEAR SCREEN
!
OUTPUT @Ld;"CAL:IPR:LEV P1"
PRINT TABXY(20,20),"Calibrating IMON & IPROG P1 - 30 SECOND WAIT"
WAIT 30
CLEAR SCREEN
!
INPUT "Measure voltage across current shunt and enter in volts",Ip1
!
INPUT "Measure Imon voltage and enter in volts",Vimonp1
!
INPUT "Measure Iprog voltage and enter in volts",Viprogp1
!
Ip1=Ip1/Rshunt
!
OUTPUT @Ld;"CAL:DATA ";Ip1,Vimonp1,Viprogp1
!
OUTPUT @Ld;"CAL:IPR:LEV P2"
!
PRINT TABXY(20,20),"Calibrating IMON & IPROG P2 - 30 SECOND WAIT"
WAIT 30
CLEAR SCREEN
!
INPUT "Measure voltage across current shunt and enter in volts",Ip2
!
INPUT "Measure Imon voltage and enter in volts",Vimonp2
!
INPUT "Measure Iprog voltage and enter in volts",Viprogp2
!
Ip2=Ip2/Rshunt
!
OUTPUT @Ld;"CAL:DATA ";Ip2,Vimonp2,Viprogp2
!
PRINT "Set IPROG power source to 8.5 volts"
DISP "Press CONT when ready"
PAUSE
CLEAR SCREEN
!
OUTPUT @Ld;"CAL:IPR:LEV P3"
PRINT TABXY(20,20),"Calibrating IMON & IPROG P3 - 30 SECOND WAIT"
WAIT 30
1000 CLEAR SCREEN
1010
1020 INPUT "Measure voltage across current shunt and enter in volts",Ip3
1030
1040 INPUT "Measure Imon voltage and enter in volts",Vimonp3
1050
1060 INPUT "Measure Iprog voltage and enter in volts",Viprogp3
1070
1080 Ip3=Ip3/Rshunt
1090
!
!
!
!
!
1100 OUTPUT @Ld;"CAL:DATA ";Ip3,Vimonp3,Viprogp3
92
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
1110
1120 OUTPUT @Ld;"CAL:IPR:LEV P4"
1130
!
!
1140 PRINT TABXY(20,20),"Calibrating IMON & IPROG P4 - 30 SECOND WAIT"
1150 WAIT 30
1160 CLEAR SCREEN
1170
1180 INPUT "Measure voltage across current shunt and enter in volts",Ip4
1190
1200 INPUT "Measure Imon voltage and enter in volts",Vimonp4
1210
1220 INPUT "Measure Iprog voltage and enter in volts",Viprogp4
1230
1240 Ip4=Ip4/Rshunt
1250
1260 OUTPUT @Ld;"CAL:DATA ";Ip4,Vimonp4,Viprogp4
!
!
!
!
!
1270
1280
1290
!
!
!
*******
Calibrating Main Current DAC
********
1300 PRINT TABXY(10,10),"Disconnect instruments from IMON and IPROG terminals"
1310 DISP "Press CONT when ready to calibrate main current ADC"
1320 PAUSE
1330
!
1340 CLEAR SCREEN
1350 OUTPUT @Ld;"CAL:LEV P1"
1360 PRINT TABXY(10,20),"Calibrating high current range P1 - 30 SECOND WAIT"
1370 WAIT 30
1380 CLEAR SCREEN
1390
1400 INPUT "Measure voltage across current shunt and enter in volts",Ip1
1410
1420 Ip1=Ip1/Rshunt
1430
1440 OUTPUT @Ld;"CAL:DATA ";Ip1 1450
1470
!
!
!
!
1460 OUTPUT @Ld;"CAL:LEV P2"
!
1480 PRINT TABXY(10,10),"Calibrating high current range P2 - 30 SECOND WAIT"
1490 WAIT 30
1500 CLEAR SCREEN
1510
1520 INPUT "Measure voltage across current shunt and enter in volts",Ip2
1530
1540 Ip2=Ip2/Rshunt
1550
1560 OUTPUT @Ld;"CAL:DATA ";Ip2
1570
!
!
!
!
1580 OUTPUT @Ld;"INP:STAT OFF"
1590 OUTPUT @Ld;"CURR:RANG MIN"
1600 OUTPUT @Ld;"SENS:CURR:RANG MIN"
1610
1620 CLEAR SCREEN
1630
!
!
1640 INPUT "Enter value of low range current shunt in ohms",Rshuntl
1650 DISP "You have entered ";Rshuntl;" ohms"
1660 INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
1670 IF UPC$(Y$)="Y" OR Y$="" THEN 1690
1680 IF UPC$(Y$)="N" OR Y$="n" THEN 1620
93
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
1690 CLEAR SCREEN
1700
!
1710 OUTPUT @Ld;"INP:STAT ON"
1720 OUTPUT @Ld;"CAL:LEV P1"
1730
!
1740 PRINT TABXY(10,10),"Calibrating low current range P1 - 30 SECOND WAIT"
1750 WAIT 30
1760 CLEAR SCREEN
1770
1780 INPUT "Measure voltage across current shunt and enter in volts",Ip1
1790
1800 Ip1=Ip1/Rshunt
1810
1820 OUTPUT @Ld;"CAL:DATA ";Ip1
1830
1840 OUTPUT @Ld;"CAL:LEV P2"
1850
!
!
!
!
!
1860 PRINT TABXY(10,10),"calibrating low current range P2 - 30 SECOND WAIT"
1870 WAIT 30
1880 CLEAR SCREEN
1890
1900 INPUT "Measure voltage across current shunt and enter in volts",Ip2
1910
1920 Ip2=Ip2/Rshunt
1930
1940 OUTPUT @Ld;"CAL:DATA ";Ip2
1950
!
!
!
!
1960 OUTPUT @Ld;"CAL:SAVE"
1970 OUTPUT @Ld;"CAL:STAT OFF"
1980
!
1990 PRINT "IMON, IPROG and Current calibration adjustments complete"
2000 PRINT "Verify Current Adjustments"
2010
!
2020 END
94
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
VOLTAGE Calibration Program
10
20
30
40
50
60
70
!
!
!
!
!
!
!
This program calibrates the voltage mode for load modules
N3302A, N3303A, N3304A, N3305A and N3306A.
last update 9/6/00
Edit the interface address on line 100 if different then 705
80
90
ASSIGN @Ld TO 705
!
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
OUTPUT @Ld;"*RST"
OUTPUT @Ld;"CAL:STAT ON"
OUTPUT @Ld;"FUNC VOLT"
OUTPUT @Ld;"VOLT:RANG MAX"
OUTPUT @Ld;"SENS:VOLT:RANG MAX"
OUTPUT @Ld;"INP:STAT ON"
!
! *******
!
Calibrating high volt range
*******
PRINT TABXY(10,8),"See figure B-3 for voltage calibration"
PRINT TABXY(10,10),"Connect power leads to input terminals"
PRINT TABXY(10,12),"Connect + sense to + input terminal"
PRINT TABXY(10,14),"Connect - sense to - input terminal"
PRINT TABXY(10,16),"Connect DVM to +/- sense terminals"
PRINT TABXY(10,18),"See Table B-2 Voltage Mode Voltage / Current Settings"
PRINT TABXY(10,20),"Set power source to voltage and current for model to be
calibrated"
260
PRINT TABXY(10,24),"BE SURE REMOTE BUTTON ON REAR PANEL IS IN REMOTE
POSITION"
270
280
290
300
310
320
330
340
350
360
370
380
390
400
410
420
430
440
450
460
470
480
490
500
510
520
530
DISP "Press CONT when ready to calibrate voltage "
PAUSE
CLEAR SCREEN
!
PRINT TABXY(9,15);"CALIBRATING HIGH VOLTAGE RANGE"
OUTPUT @Ld;"CAL:LEV P1"
WAIT 1
INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1
!
OUTPUT @Ld;"CAL:DATA";Vp1
!
OUTPUT @Ld;"CAL:LEV P2"
WAIT 1
INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2
!
OUTPUT @Ld;"CAL:DATA";Vp2
!
! *******
!
Calibrating low voltage range *******
OUTPUT @Ld;"VOLT:RANG MIN"
OUTPUT @Ld;"SENS:VOLT:RANG MIN"
!
CLEAR SCREEN
PRINT TABXY(9,15);"CALIBRATING LOW VOLTAGE RANGE"
!
OUTPUT @Ld;"CAL:LEV P1"
WAIT 1
95
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
540
550
560
570
580
590
600
610
620
630
640
650
660
670
680
690
700
710
720
730
!
INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1
!
OUTPUT @Ld;"CAL:DATA";Vp1
!
OUTPUT @Ld;"CAL:LEV P2"
WAIT 1
!
INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2
!
OUTPUT @Ld;"CAL:DATA";Vp2
!
OUTPUT @Ld;"CAL:SAVE"
OUTPUT @Ld;"CAL:STAT OFF"
!
CLEAR SCREEN
PRINT "Voltage calibration adjustments complete"
PRINT "Verify Voltage Adjustments"
!
END
96
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
RESISTANCE Calibration Program
10
20
30
40
50
60
70
80
!
!
!
!
!
!
!
!
!
!
!
!
!
This program calibrates the resistance mode
for load modules N3302A, N3303A, N3304A, N3305A and N3306A.
last update 9/27/00
The variables for the module to be calibrated must be inserted
in the program where a < variable > is encountered. See Table B-1
for variables for the module you are calibrating.
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
270
280
290
300
310
320
330
340
350
360
370
380
390
NOTE: Delete lines 1540 to 1940 for module Model N3303
Edit the interface address on line 100 if different then 705
ASSIGN @Ld TO 705
!
OUTPUT @Ld;"*RST"
OUTPUT @Ld;"CAL:STAT ON"
OUTPUT @Ld;"FUNC RES"
OUTPUT @Ld;"INP:STAT ON"
!
PRINT TABXY(10,8);"See figure B-4 for resistance calibration set-up"
PRINT TABXY(10,10);"Connect power leads to input terminals"
PRINT TABXY(10,12);"Connect +/- sense to +/- input terminal"
PRINT TABXY(10,14);"Connect one DVM to +/- sense terminals"
PRINT TABXY(10,16);"Connect second DVM across current shunt"
DISP "Press CONT when ready to continue"
PAUSE
!
!
!
********
CALIBRATING RANGE 1 ****************
CLEAR SCREEN
INPUT "Enter value of current shunt for range 1 in OHMS",Rshunt1
PRINT TABXY(10,10);"You have entered ";Rshunt1;" ohms"
INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
IF UPC$(Y$)="Y" OR Y$="" THEN 370
IF UPC$(Y$)="N" OR Y$="n" THEN 310
CLEAR SCREEN
!
PRINT TABXY(10,12);"Turn on and set power source to voltage and current
settings from"
400
410
420
430
440
450
460
470
480
490
500
PRINT TABXY(10,13);"Table B-1 for resistance range 1 - Point P1"
DISP "Press CONT when ready to calibrate resistance range 1"
PAUSE
CLEAR SCREEN
PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 1 - P1 - 30 SECOND WAIT"
OUTPUT @Ld;"RES:RANG 1"
OUTPUT @Ld;"CAL:LEV P1"
WAIT 30
INPUT "Measure voltage across current shunt - enter value in volts",Vp1
Ip1=Vp1/Rshunt1
INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt1
510
520
530
540
Rp1=Vt1/Ip1
OUTPUT @Ld;"CAL:DATA";Rp1
!
CLEAR SCREEN
97
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
550
PRINT TABXY(10,12);"Set power source to voltage and current settings from
Table B-2"
560
570
580
590
600
610
620
630
640
650
PRINT TABXY(10,13);"for resistance range 1 - Point P2"
DISP "Press Cont to continue resistance range 1 calibration"
PAUSE
OUTPUT @Ld;"CAL:LEV P2"
CLEAR SCREEN
PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 1 - P2 - 30 SECOND WAIT”
WAIT 30
INPUT "Measure voltage across current shunt - enter value in volts",Vp2
Ip2=Vp2/Rshunt1
INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt2
660
670
680
690
700
710
720
730
740
750
760
770
780
790
800
810
820
830
840
850
860
870
880
890
900
910
920
Rp2=Vt2/Ip2
OUTPUT @Ld;"CAL:DATA";Rp2
OUTPUT @Ld;"INP:STAT OFF"
!
!
!
********** CALIBRATING RANGE 2
******************
CLEAR SCREEN
INPUT "Enter value of current shunt for range 2 in OHMS",Rshunt2
PRINT TABXY(10,10);"You have entered ";Rshunt2;" ohms"
INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
IF UPC$(Y$)="Y" OR Y$="" THEN 780
IF UPC$(Y$)="N" OR Y$="n" THEN 720
CLEAR SCREEN
!
PRINT TABXY(10,12);"Set power source to voltage and current settings from"
PRINT TABXY(10,13);"Table B-1 for resistance range 2 - Point P1"
DISP "Press CONT when ready to calibrate resistance range 2"
PAUSE
CLEAR SCREEN
PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 2 - P1 - 15 SECOND WAIT"
OUTPUT @Ld;"INP:STAT ON"
OUTPUT @Ld;"RES:RANG 20" !< resistance range 2 variable from table B-1
OUTPUT @Ld;"CAL:LEV P1"
WAIT 15
INPUT "Measure voltage across current shunt - enter value in volts",Vp1
Ip1=Vp1/Rshunt2
INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt1
930
940
950
960
970
Rp1=Vt1/Ip1
OUTPUT @Ld;"CAL:DATA";Rp1
!
CLEAR SCREEN
PRINT TABXY(10,12);"Set power source to voltage and current settings from
Table B-2"
980
990
PRINT TABXY(10,13);"for resistance range 2 - Point P2"
DISP "Press Cont to continue resistance range 2 calibration"
1000 PAUSE
1010 OUTPUT @Ld;"CAL:LEV P2"
1020 CLEAR SCREEN
1030 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 2 - P2 - 15 SECOND WAIT"
1040 WAIT 15
1050 INPUT "Measure voltage across current shunt - enter value in volts",Vp2
1060 Ip2=Vp2/Rshunt2
1070 INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt2
98
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Test and Calibration - B
1080 Rp2=Vt2/Ip2
1090 OUTPUT @Ld;"CAL:DATA";Rp2
1100 OUTPUT @Ld;"INP:STAT OFF"
1110
1120
1130
!
!
!
************ CALIBRATING RANGE 3 ******************
1140 CLEAR SCREEN
1150 INPUT "Enter value of current shunt for range 3 in OHMS",Rshunt3
1160 PRINT TABXY(10,10);"You have entered ";Rshunt3;" ohms"
1170 INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
1180 IF UPC$(Y$)="Y" OR Y$="" THEN 1200
1190 IF UPC$(Y$)="N" OR Y$="n" THEN 1140
1200 CLEAR SCREEN
1210
!
1220 PRINT TABXY(10,12);"Set power source to voltage and current settings from"
1230 PRINT TABXY(10,13);"Table B-1 for resistance range 3 - Point P1"
1240 DISP "Press CONT when ready to calibrate resistance range 3"
1250 PAUSE
1260 CLEAR SCREEN
1270 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 3 - P1 - 7 SECOND WAIT"
1280 OUTPUT @Ld;"INP:STAT ON"
1290 OUTPUT @Ld;"RES:RANG 200" !< resistance range 3 variable from table B-1
1300 OUTPUT @Ld;"CAL:LEV P1"
1310 WAIT 7
1320 INPUT "Measure voltage across current shunt - enter value in volts",Vp1
1330 Ip1=Vp1/Rshunt3
1340 INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt1
1350 Rp1=Vt1/Ip1
1360 OUTPUT @Ld;"CAL:DATA";Rp1
1370
!
1380 CLEAR SCREEN
1390 PRINT TABXY(10,12);"Set power source to voltage and current settings from
Table B-2"
1400 PRINT TABXY(10,13);"for resistance range 3 - Point P2"
1410 DISP "Press Cont to continue resistance range 3 calibration"
1420 PAUSE
1430 OUTPUT @Ld;"CAL:LEV P2"
1440 CLEAR SCREEN
1450 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 3 - P2 - 7 SECOND
1460 WAIT 7
WAIT"
1470 INPUT "Measure voltage across current shunt - enter value in volts",Vp2
1480 Ip2=Vp2/Rshunt3
1490 INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt2
1500 Rp2=Vt2/Ip2
1510 OUTPUT @Ld;"CAL:DATA";Rp2
1520 OUTPUT @Ld;"INP:STAT OFF"
1530
1540
1550
!
!
!
************
CALIBRATING RANGE 4
*********************
1560 CLEAR SCREEN F1570 INPUT "Enter value of current shunt for range 4 in
OHMS",Rshunt4
1580 PRINT TABXY(10,10);"You have entered ";Rshunt4;" ohms"
1590 INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$
1600 IF UPC$(Y$)="Y" OR Y$="" THEN 1620
1610 IF UPC$(Y$)="N" OR Y$="n" THEN 1560
99
Download from Www.Somanuals.com. All Manuals Search And Download.
B - Performance Test and Calibration
1620 CLEAR SCREEN
1630
!
1640 PRINT TABXY(10,12);"Set power source to voltage and current settings from"
1650 PRINT TABXY(10,13);"Table B-1 for resistance range 4 - Point P1"
1660 DISP "Press CONT when ready to calibrate resistance range 4"
1670 PAUSE
1680 CLEAR SCREEN
1690 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 4 - P1 - 3 SECOND WAIT"
1700 OUTPUT @Ld;"INP:STAT ON"
1710 OUTPUT @Ld;"RES:RANG 2000"!< resistance range 4 variable from table B-1
1720 OUTPUT @Ld;"CAL:LEV P1"
1730 WAIT 3
1740 INPUT "Measure voltage across current shunt - enter value in volts",Vp1
1750 Ip1=Vp1/Rshunt4
1760 INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt1
1770 Rp1=Vt1/Ip1
1780 OUTPUT @Ld;"CAL:DATA";Rp1
1790
!
1800 CLEAR SCREEN
1810 PRINT TABXY(10,12);"Set power source to voltage and current settings from
Table B-2"
1820 PRINT TABXY(10,13);"for resistance range 4 - Point P2"
1830 DISP "Press Cont to continue resistance range 4 calibration"
1840 PAUSE
1850 OUTPUT @Ld;"CAL:LEV P2"
1860 CLEAR SCREEN
1870 PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 4 - P2 - 3 SECOND
1880 WAIT 3
WAIT"
1890 INPUT "Measure voltage across current shunt - enter value in volts",Vp2
1900 Ip2=Vp2/Rshunt4
1910 INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt2
1920 Rp2=Vt2/Ip2
1930 OUTPUT @Ld;"CAL:DATA";Rp2
1940 OUTPUT @Ld;"INP:STAT OFF"
1950
!
1960 OUTPUT @Ld;"CAL:SAVE"
1970 OUTPUT @Ld;"CAL:STAT OFF"
1980
!
1990 CLEAR SCREEN
2000 PRINT TABXY(10,10);"Resistance calibration adjustments complete"
2010 PRINT TABXY(10,12);"Verify Resistance Adjustments"
2020 END
100
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
control connector, 46
controller connections, 48
CR, 23
CR mode, 63
current
-- -- -- -- --, 69
range, 22
triggered, 23
current measurement range, 69
current monitor, 33
CV, 24
0
8
0 ... 9, 62
CV mode, 63
8mm screw terminal connector, 43
D
A
damage, 37
dc measurements, 35
description, 20
accessories, 20
airflow, 41
annunciators
Addr, 56
Cal, 56
digital connector, 37, 47
dimensions, 40
CC, 56
CR, 56
CV, 56
Dis, 56
E
entry keys, 62
▲ input, 62
▼ input, 62
0 ... 9, 62
Backspace, 62
Clear Entry, 62
Enter, 62
Err, 56
Prot, 56
Rmt, 56
Shift, 56
SRQ, 56
Tran, 56
Unr, 56
error messages, 54
errors, 68
external programming, 33
B
F
binding posts, 42
fault, 34
front panel, 55
annunciators, 14
annuncuiators, 56
controls and indicators, 55
immediate action, 14
keys, 56
C
cables, 20
calibration
equipment, 75
capabilities, 21
CC, 22
CC mode, 63
channel
measurements, 69
menus, 15
using, 13
function keys, 58
current, 60
location, 39
characteristics, 73
checkout procedure, 53
cleaning, 37
clearing errors, 68
clearing protection, 68
connections
function, 60
immediate action, 58
input control, 60
Input On/Off, 58
list control, 61
metering, 58
resistance, 60
scrolling, 58
short on/off, 60
transient control, 61
trigger control, 61
voltage, 60
input, 42
local sense, 49
parallel, 49
remote sense, 49
constant current, 22
constant resistance, 23
constant voltage, 24
101
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
G
P
GPIB, 70
port, 34
address, 70
power cord, 37, 42
power measurements, 35
power receptacle, 20
print date, 6
connections, 48
ground, earth, 20
guide, user’s, 19
programming errors, 32
programming lists, 67
programming transinets, 65
protection
H
I
history, 6
OV, 68
Q
R
input
query protection, 68
current setting, 63
resistance setting, 63
voltage setting, 63
input measurements, 34
input on/off, 30, 58
inspection, 37
rack mount kit, 20
rack mounting, 41
rear panel
installing
modules, 38
at a glance, 12
recall state, 33
recalling operating states, 70
remote programming, 21
remote sense, 33
repacking, 37
reset protection, 30
resistance
L
List, 67
list mode, 27
location, 40
low voltage operation, 51
LRV, 32
range, 23
triggered, 24
reverse voltage, 32
risetime limitation, 28
rms measurements, 35
RRV, 32
RS 232
connections, 48
RS-232, 70
M
making measurements, 34, 69
manual connector, 42
manuals, 37
maximum measurements, 35
measurement ranges, 35, 69
measurements, 34
minimum measurements, 35
minimum transtion time, 28
S
safety, 3, 5
safety class, 20
safety warning, 20
save state, 33
saving operating states, 70
SCPI commands
at a glance, 17
selftest errors, 54
sense switch, 47
short on/off, 30, 60
signals
N
O
non-volatile, 33
non-volatile memory
clearing, 70
storing, 57
operating modes, 22
options, 20
output
connector, 37
overcurrent, 31
overpower, 31
overtemperature, 31
overvoltage, 31
OVLD, 69
current monitor, 33
external, 33
external programming, 33
fault, 34
port, 34
remote sense, 33
slew rate, 28
specifications, 71
status reporting, 32
support rails, 41
system keys, 57
102
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
Address, 57
Error, 57
V
Interface, 57
Local, 57
RCL, 57
Save, 57
Shift, 57
verification
equipment, 75
voltage
range, 24
triggered, 25
VXIplug&play, 19
T
W
Z
transient, 26
continuous, 26
pulse, 26
toggled, 26
warranty, 2
wire
current ratings, 44
transients
continuous mode, 65
pulse mode, 66
toggled mode, 66
trigger connector, 47
trigger mode, 27
zero-volt operation, 51
103
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent Sales and Support Office
For more information about Agilent Technologies test and measurement products, applications, services, and for
a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir
You can also contact one of the following centers and ask for a test and measurement sales representative.
United States:
Latin America:
Agilent Technologies
Test and Measurement Call Center
P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 4844
Agilent Technologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286
Canada:
Australia/New Zealand:
Agilent Technologies Australia Pty Ltd
347 Burwood Highway
Forest Hill, Victoria 3131
(tel) 1-800 629 485 (Australia)
(fax) (61 3) 9272 0749
Agilent Technologies Canada Inc.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414
(tel) 0 800 738 378 (New Zealand)
(fax) (64 4) 802 6881
Europe:
Asia Pacific:
Agilent Technologies
Test & Measurement European Marketing Organisation
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
Agilent Technologies
24/F, Cityplaza One, 1111 King's Road,
Taikoo Shing, Hong Kong
tel: (852)-3197-7777
fax: (852)-2506-9284
(tel) (31 20) 547 9999
Japan:
Agilent Technologies Japan Ltd.
Measurement Assistance Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Technical data is subject to change.
104
Download from Www.Somanuals.com. All Manuals Search And Download.
Manual Updates
The following updates have been made to this manual since its publication.
6/1/01
Table 2-1 has been updated with new option part numbers.
Information about Option UJ1 has been added to chapters 2 and 3.
Information about Low Voltage Operation has been has been added to the end of chapter 3.
Appendix A has been rewritten to include more information about:
Low voltage operation
Slew rates
List dwell characteristics
Measurement times
Temperature coefficients
N3300A and N3301A characteristics
Figure B-2 has been corrected.
A new Pulse Width Accuracy test has been added to Appendix B.
10/1/01
Figure 3-3C has been added.
The Programming Lists section in chapter 5 has been updated.
DC Isolation Voltage has been added to Table A-2.
Download from Www.Somanuals.com. All Manuals Search And Download.
|