BRONCHOSCOPE DISINFECTION GUIDE
®
for OES BF-40 Series Bronchofiberscopes
Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.
HIGH-LEVEL DISINFECTION
1
2
3
4
Suction Cleaning Adapter
Fill basin with disinfectant. Attach a
suction cleaning adapter (MAJ-222) to
the previously cleaned bronchoscope.
Immerse completely in disinfectant
solution.
Attach a syringe to the adapter.
Pull the plunger of the syringe to
fill the instrument/suction channels
and channel cleaning adapter
with disinfectant solution.
While immersed, detach the adapter
Soak all equipment according
to the time and temperature
recommended by the disinfectant
manufacturer. Use a timer to mea-
sure the exact contact time.
from the bronchoscope. Use a clean,
lint-free cloth to wipe away any bub-
bles clinging to external surfaces.
RINSING
To Suction Pump
5
6
7
8
To Suction Pump
Alcohol
Connect suction cleaning adapter
to bronchoscope and to a tube
from a suction pump. Aspirate
water through channels for 30
seconds. Remove from water and
aspirate air through channels for
60 seconds.
Disconnect the suction cleaning
adapter and immerse the broncho-
scope and adapter in sterile water.
Using a sterile, lint-free cloth, thor-
oughly rinse and wipe all surfaces.
Connect the suction cleaning adapter
and syringe. Inject air to remove
Immerse the bronchoscope's distal
end in 70% alcohol and aspirate
alcohol for 5 seconds. Remove the
distal end from the alcohol and
aspirate air for 20 seconds.
disinfectant from the instrument suction
/
channels, then remove the broncho-
scope from the disinfectant.
TRANSPORTATION
DRYING
9
10
Dry the bronchoscope and adapter with an
alcohol-moistened cloth. Make sure that the
equipment is completely dry prior to storage.
Turn the suction pump off. Hold the control
section with the openings pointing down and
disconnect the suction cleaning adapter.
Hold all parts of the endoscope when carrying. Do
not let the ends of the bronchoscope swing freely.
NOMENCLATURE
IMPORTANT
• Wear appropriate personal protective
equipment during reprocessing.
Single-Use Suction Valve (MAJ-209)
Suction Cylinder
Universal Cord
Light Guide
• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically
with all valves and caps removed.
• Use a compatible disinfectant and test it
for proper concentration as recommended
by the manufacturer.
Single-Use Biopsy Valve (MAJ-210)
Instrument/Suction Channel Port
Venting Connector
• Do not use the carrying case to store
"patient-ready" endoscopes. Any
bronchoscope placed in the carrying case
must be reprocessed prior to patient use.
Insertion Tube
Control Section
Boot
Bending Section
Distal Tip
UP/DOWN Angulation Control Lever
Clean the bronchoscope
meticulously prior to disinfection.