HP Hewlett Packard IAP Monitor User Manual

Spiegelberg:  
Manufacturer  
Spiegelberg  
IAP-Monitor  
IAP 32.1  
(GmbH & Co.) KG  
Tempowerkring 4  
21079 Hamburg  
Germany  
Operating-Instructions  
Phone: +49-40-790-178-0  
Fax:  
+49-40-790-178-10  
0297  
CE-marking according to directive  
93/42/EEC  
Technical alterations reserved.  
Innovative  
Measurement  
Technology for  
Medicine  
BEDIAP32.1.91 V 2.0  
© by Spiegelberg (GmbH & Co.) KG, 2003  
7
4
Faults  
Monitor operation  
The air-tightness of the air-pouch  
system is checked regularly. If a leak  
is detected, the red warning  
indicator shows alarm. When the  
leak is removed the red warning  
indicator disappears and the  
After placement of probe the tube  
connector is inserted into the  
monitor socket and fastened by a  
slight clockwise turn. The soft  
material of the connector  
guarantees airtightness.  
measurement is continued. If the  
leak has not been removed after  
one minute, in addition to the alarm  
display a chirp sound is emitted. To  
reset the alarm the monitor must be  
switched off and on again. The  
chirp sound stops after one minute,  
unless the monitor has been  
switched off before, and the digital  
display dissapears. The warning  
indicator continues to flash.  
Pressure measurements are started  
by switching on the monitor. No  
manual zero adjustment or  
calibration is required.  
If a pressure limit is exceeded during  
probe filling the filling process is  
interrupted, the digital display  
shows ´E1´ and the chirp sound is  
emitted. This can happen if the air-  
pouch resists unfolding or if the  
tube is clamped. Switching off and  
on the monitor resets the system.  
If the digital display shows ´E2´ or  
´E3´ during operation an internal  
fault has been detected. Contact  
the service in this case.  
Warning  
The unit cannot be  
repaired by the user.  
5
6
Connections  
On the digital display the mean  
pressure is shown in mmHg.  
Connection of a Patient  
Monitor  
Connection of a Strip-Chart  
Recorder  
To connect a patient monitor the  
connecting cable is inserted into the  
socket of the IAP-Monitor and  
connected to the pressure  
transducer input of the patient  
monitor. Connect probe to the IAP-  
Monitor. Switch on the Brain-  
Pressure Monitor. Wait for the  
digital display of the mean ICP. The  
pulsatile signal is displayed now on  
the monitor screen.  
To connect a strip-chart recorder  
with voltage input the appropriate  
cable is connected with the socket  
at the rear panel of the monitor and  
the input socket of the recorder.  
During the first ten minutes after  
switching on the pulse amplitude  
(i.e. systolic pressure minus diastolic  
pressure) is additionally displayed  
marked by an ´A´.  
Connection of a Computer  
The RS 232 C interface permits  
communication with any computer.  
During this time the mean pressure  
is displayed for three seconds, then  
the amplitude is displayed for one  
second.  
Zero Adjustment of the  
Patient Monitor  
Disconnect the catheter from the  
IAP-Monitor. Leave the IAP-Monitor  
and the patient monitor switched  
on. Zero patient monitor using the  
zero button. Reconnect the probe.  
Switch off the IAP-Monitor. Switch  
it on again. Observe the display of  
the mean ICP on the IAP-Monitor  
and the pulsatile signal on the  
patient monitor.  
Cleaning  
After removal of the power cord the  
unit can be wiped off with a damp  
rag using a common disinfectant.  
No liquid whatsoever must be  
Warning  
Unplug the power cord  
prior to cleaning.  
Connection of a CPP-Monitor  
Connect the IAP-Monitor to the RS  
232 IAP socket of the CPP-Monitor.  
introduced into the catheter or into  
the connector of the monitor.  
Warning  
No fluid should be  
allowed to enter the  
probe connector.  
Warning  
In case of unintentional  
wettening of the device,  
the power cord is to be  
disconnected immediately.  
3
8
Indication and Method  
Technical specifications  
The IAP-Monitor and the catheters  
are intended for the monitoring of  
intraabdominal pressure in the  
stomach.  
REF  
IAP 32.1  
Measurement range  
Operating voltage  
-50 to +100 mmHg  
230 V~, 50/60 Hz  
or  
115 V~, 50/60 Hz  
The air-pouch system consists of a  
hollow body connected to a  
pressure transducer by tubing. The  
pressure transducer, the electronic  
hardware, and the device for filling  
the air-pouch are integrated in  
the IAP-Monitor.  
Current requirement  
Fuses  
0,1 A at 230 V~  
0,2 A at115 V~  
125 mA T at 230 V~  
250 mA T at 115 V~  
For intraabdominal pressure  
measurements the air- pouch is  
placed in the stomach of the  
Voltage output  
1V/50 mmHg at 1 k  
patient. The intraabdominal pressure  
is transmitted across the thin pouch  
wall to the air volume in the pouch  
and transformed into an electric  
signal by the pressure transducer.  
Pressure monitor output  
RS 232 C interface  
5 µV/mmHg/V  
9600 Bd (other baud rates  
upon request)  
Accuracy class  
2
On the digital display the mean  
pressure is shown. At the monitor  
output both the mean pressure and  
the pulsatile signal are available.  
Safety class  
I / BF  
3 kg  
Weight  
Once every hour the IAP-Monitor  
opens the pressure transducer to  
atmospheric pressure for zero  
adjustment. The air-pouch is  
then filled with the exact air volume  
required for accurate pressure  
transmission.  
In accordance with standards  
IEC 60601-1  
IEC 60601-1-2  
1
10  
Accessories  
1.9.2003 AS/az  
Description  
CPP-Monitor  
IAP-Catheter  
Order number  
CPP 21.x  
EG-Konformitätserklärung  
Declaration of Conformity  
IAP 32.1.11  
KBL 21.005.00  
Das Produkt  
Interconnect Cable for  
IAP-Monitor HDM IAP 32.1  
Spiegelberg CPP-Monitor 21.x  
entspricht den grundlegenden Anforderungen gemäß Anhang  
I der Richtlinie 93/42/EWG. Es ist als Klasse IIa klas-  
sifiziert.  
Monitor Cables  
Bruker VM800/900  
Corometrix  
Datex-Cardiocap  
Dräger VM 300  
Hellige 4. Generation up  
Hellige 3. Generation  
Hewlett-Packard  
Horizon  
KBL 13.014.00  
KBL 13.020.00  
KBL 13.007.00  
KBL 13.001.00  
KBL 13.003.00  
KBL 13.019.00  
KBL 13.004.00  
KBL 13.015.00  
KBL 13.018.00  
KBL 13.005.00  
KBL 13.005.01  
KBL 13.008.00  
KBL 13.008.01  
KBL 13.011.00  
KBL 13.009.00  
KBL 13.010.00  
KBL 13.002.00  
KBL 13.006.00  
KBL 13.013.00  
Das Qualitätsmanagementsystem entspricht den Forderungen  
der Norm ISO 13485:2001 gemäß Zertifikat Nr. 18737 der  
DQS, benannte Stelle Nr. 0297.  
The Product  
IAP-Monitor IAP 32.1  
is in accordance with the requirements of Annex I of the  
Manual Devices Directive 93/42/EEC. It is classified as  
class IIa.  
Kontron  
Marquette  
Marquette 7010  
Nihon-Kohden  
Nihon-Kohden BMS 8300  
Philips CM130  
Propaque/Mennen  
Schiller CM-8  
Siemens-Sirecust  
Space-Labs  
S&W, Danica  
The Quality Management System is in accordance with the  
requirements of the standard ISO 13485:2001 according to  
certificate no. 18737 of DQS, Notified Body No. 0297.  
Spiegelberg (GmbH & Co.) KG  
Recorder Cables for Recorders  
with Voltage Input  
BNC-Plug  
KBL 13.030.00  
KBL 13.031.00  
Dr. A. Spiegelberg  
Banana Plug 4 mm  
Computer Cable RS 232 9-pin  
KBL 13.033.00  

Hotpoint Cooktop CH 10750 GF S User Manual
Freescale Semiconductor MCF5480 User Manual
DCS C 24 User Manual
Dacor SGM466 User Manual
Cypress CY14B101Q3 User Manual
BT Diverse 2016 User Manual
Blodgett CONVEYOR OVEN MT1820E User Manual
Black Decker Perfect Broil CTO4400BUC User Manual
Asus M2N68 AM SE2 User Manual
Acer Computer Monitor P235H User Manual