Epson 9840 User Manual

Advanced Force Measurement  
Model 9840  
Digital Process Monitor  
Interface, Inc. 7401 E. Butherus Dr. Scottsdale, Arizona 85260 USA  
Tel: (480) 948-5555, Fax: (480) 948-1924  
INTERFACE MODEL 9840 VER 6.0.2  
MODEL 9840  
PG 2 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
Under the copyright laws, this publication may not be reproduced or transmitted in any  
form, electronic or mechanical, including photocopying, recording, storing in an  
information retrieval system, or translating, in whole or part, without the prior consent of  
Interface, Inc. Copyright January 8, 2008 all rights reserved.  
MODEL 9840  
PG 3 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
INTRODUCTION  
The Model 9840 is a CE compliant and versatile precision instrument intended for the  
digital readout of strain gage sensors such as load cells and torque cells. Optical  
encoders are also supported.  
Here is a quick listing of its features:  
Full bi-polar 6 digit display (±999,999) using two line, 20 character, vacuum  
fluorescent display for clear, precise indication of measured quantities and limit status.  
No knobs or dials. No internal jumpers. All adjustment and calibration is done using  
user-friendly front panel menus. Keylock option to prevent accidental or unauthorized  
changes to the unit’s setup.  
Measure load and gross weight in pounds, kilograms, Newtons, PSI, MegaPascals,  
kilo-pounds, kilo-Newtons, metric tons, mV/V, or grams. User defined base area for  
PSI and MPa. Tare at any time. The sum of the loads on both channels is also  
availabe.  
Measure torque and gross torque in Pound-Inches, Newton-Meters, Ounce-Inches,  
or mV/V. Tare at any time. The sum of the torques on both channels is also availabe.  
Capture the load peak and valley, display in pounds, kilograms, Newtons, PSI,  
MegaPascals, kilo-pounds, kilo-Newtons, metric tons, mV/V, or grams. Reset at any  
time.  
Capture the torque peak and valley, display in in Pound-Inches, Newton-Meters,  
Ounce-Inches, or mV/V. Reset at any time.  
Measure position and velocity with optional encoder input. Display in English units,  
metric units, or in percent of a user defined base length. Reset at any time.  
Four contact closure Limit switches. You can set which signal drives the limit (load,  
torque, peak, valley, position, or velocity), normally open / normally closed, set point,  
and manual reset (latched) or automatic reset (not latched).  
Print current readings of load, torque, peak, valley, position, and velocity using the  
currently selected units at the touch of one button (optional).  
Automatic identification of calibrated cells with auto-id option.  
Calibration for cells using manufacturer’s milli-volt per volt calibration factor, 6 point  
mass and milli-volt per volt, 2 or 5 point known mass, 6 point torque and milli-volt per  
volt, 2 or 5 point known torque, or internal precision shunt calibration. A back panel  
switch allows you to choose between two values used for the shunt calibration so  
there is no need to open the unit.  
Analog output of any signal (load, torque, peak, valley, position, velocity) with user  
defined scale factor and offset. Large, buffered ± 10.00 volt output.  
MODEL 9840  
PG 4 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Remote operation using standard RS232 ASCII command set.  
Each unit is set up to interact with the InterView software that can save tests, export  
data to a spreadsheet, and much more.  
Back panel digital inputs for tare, peak and valley reset, position reset, and print.  
Other functions available as custom programmed options.  
MODEL 9840  
PG 5 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
QUICK START  
This section will help you get your Model 9840 set up and operating in just a couple of  
minutes.  
The back panel is shown below. You will need to make at least two connections here.  
First be sure that the power switch is off (the lower part of the rocker should be pressed  
in). Connect the AC power line and plug it in.  
MODEL 9840 BACK PANEL  
Next connect a load or torque cell to the 9 pin female connector labeled “Load A”. You  
should use the cabling supplied with your Model 9840 and Interface load or torque cell. If  
you do not have a factory supplied cable see Appendix B for the recommended cell  
connection.  
You are now ready to turn on the power. Each time power is applied to the Model 9840  
you will see a sequence of 6 messages appear on the front display. For example,  
MODEL 9840  
PG 6 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
#
1
2
Message  
Meaning  
Interface X.Y.Z  
Serial # 12345  
This is the version number.  
This is the serial number.  
This is the option number.  
The serial number of the load cell on Ch A.  
Option # 114236  
Last ID A LD90437  
3
4
5
Last ID B TQ12030  
Ch A Max 4.00Lb  
Ch B Max 4.00LbI  
The serial number of the torque cell on Ch B  
Maximum rated load of the sensor on channel A.  
Maximum rated torque of the sensor on channel B.  
Ch A Cal 4.00000 mVv The mV/V constant used with the load cell on channel A.  
Ch B Cal 4.00000 mVv The mV/V constant used with the torque cell on channel  
B.  
Note: If either of these sensors was calibrated by 5 –Point  
Mass, then 4 mV/V constants will be displayed here.  
6
Ch A Exc 5.00 Volt  
Ch B Exc 10.00 Volt  
Excitation voltage used with the load cell on channel A.  
Excitation voltage used with the torque cell on channel B.  
Each of these messages will be displayed for about 3 seconds. After that the front display  
will start showing the current readings from the sensor.  
MODEL 9840 FRONT PANEL  
The example above shows that Load A is 0.0000 Lb. The second line of the display  
shows text.  
To change what item is being displayed press the leftmost button. This will cycle through  
the list of items (load A, peak A, valley A, gross A, load B, peak B, valley B, gross B,  
channel A+B, pos, velocity, lim). If a torque cell is loaded on either channel, the display  
will show torq rather than load for the appropriate channel. Channel A+B will only appear  
when both channels are attached to sensors of the same type: load or torque.  
MODEL 9840  
PG 7 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
The rightmost button will change the units that are being used. Repeated presses of this  
button will cycle through the list of units available for the currently displayed item. For  
example, with Load A showing you would see Lb, kg, N, PSI, MPa, Klb, kN, t, mV/V,  
grams, and then back to Lb. With Torq A, you will see LbI, NM, OzI, and mV/V.  
Switching from a load to a torque cell and back will automatically reset Analog, Limit, and  
Filter Window settings to zero and/or disabled for that particular channel. See information  
under each of these options for details on how they work.  
MODEL 9840  
PG 8 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
RUN MODE  
The Model 9840 has three modes of operation, Run mode, Setup mode, and System  
Calibration mode. At power-on it will be in Run mode.  
The first line of the display show an item label (Load A), a value (0.0000), and a unit label  
(Lb).  
The second line of the display can show the status of the 4 contact closure limit switches,  
show another display, show text, or be blank. The above example shows text  
(“INTERFACE MODEL 9840”).  
To change what item is being displayed press the leftmost button. This will cycle through  
the list of items (Load A, Peak A, Vall A, Grs A, Load B, Peak B, Vall B, Grs B, Ch A+B,  
Pos, Vel, and Limits). If a torque cell is attached rather than a Load Cell, Torq A or B will  
replace Load A or B. Channel A+B will only appear when both channels are attached to  
sensors of the same type: load or torque.  
The rightmost button will change the units that are being used. Repeated presses of this  
button will cycle through the list of units available for the currently displayed item. For  
example, with Load showing you would see Lb, kg, N, PSI, MPa, Klb, kN, t, mV/V, grams,  
and then back to Lb. Torq will show LbI, NM, OzI and mV/V.  
MODEL 9840  
PG 9 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
The table below summarizes the options. Whether the user sees torque or load for each  
channel depends upon the cell type that was calibrated or selected for that channel.  
Load A/B  
Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds,  
Kilo-Newtons, Metric Tons, mV/V, grams.  
Peak A/B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds,  
(Load) Kilo-Newtons, Metric Tons, mV/V, grams.  
Valley A/B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds,  
(Load) Kilo-Newtons, Metric Tons, mV/V, grams.  
Gross A/B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds,  
(Load)  
Kilo-Newtons, Metric Tons, mV/V, grams.  
Channel  
A+B (Ld)  
Torq A/B  
Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds,  
Kilo-Newtons, Metric Tons, mV/V, grams.  
Pound-Inches, Newton-Meters, Ounce-Inches, mV/V.  
Peak A/B Pound-Inches, Newton-Meters, Ounce-Inches, mV/V.  
(Torque)  
Valley A/B Pound-Inches, Newton-Meters, Ounce-Inches, mV/V.  
(Torque)  
Gross A/B Pound-Inches, Newton-Meters, Ounce-Inches, mV/V.  
(Torque)  
Channel  
A+B (Tq)  
Position  
Velocity  
Limits  
Pound-Inches, Newton-Meters, Ounce-Inches, mV/V.  
Inches, Centimeters, Percent of Base Length.  
Inches per Minute, Centimeters per Minute.  
Shows state of each switch, 1 for on, 0 for off, dash for  
disabled, asterisk for source error.  
All of these measurements are acquired by the Model 9840 at all times with torque  
measurements acquired when a torque cell is attached to a channel and load  
measurements acquired when a load cell is attached. The display just selects which  
particular item you wish to see and what units are used to display or print this data.  
The buttons have the functions indicated by the upper set of labels (Item, Tare, Reset,  
View, Esc, Print, Unit). Each of these is described below.  
Item Button  
The Item button cycles through the list of different items that can be displayed  
(Load A, Peak A, Vall A, Grs A, Load B, Peak B, Vall B, Grs B, Ch A+B, Pos, Vel,  
Limits). Torq A or B substitutes for Load A or B when appropriate. Use the Unit  
button to change the units that are used for the currently displayed item. Ch A+B is  
only available when the cells on both channels are of the same type (load or  
torque).  
Tare Button  
The Tare button sets the current load or torque to zero for channel A, channel B,  
or channels A and B. A submenu will allow you to select what channel you want  
to tare. It does not matter what is being displayed when you press this button.  
MODEL 9840  
PG 10 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
The load or torque is always reset to zero. A single channel unit will automatically  
set Load A or Torque A to zero.  
Reset Button  
The Reset button brings up a list of things that you might want to set to zero:  
If this is a 2-channel unit, the following menu will appear:  
> Reset Ch A + Ch B will reset the peak and valley for channels A and B.  
> Reset Ch A Only will reset the peak and valley for channel A.  
> Reset Ch B Only will reset the peak and valley for channel B.  
> Reset Position will reset the position to zero.  
> Reset Limits will reset any limits that require manual resetting.  
For a single channel unit, the following menu will appear:  
> Reset Peak+Vall will reset the peak and valley measurements.  
> Reset Peak Only will reset the peak measurement.  
> Reset Vall Only will reset the valley measurement.  
> Reset Position will reset the position to zero.  
> Reset Limits will reset any limits that require manual resetting.  
Pressing either the “+” button or the “>“ button will scroll forward through this list.  
Pressing either the “-” button or the “<“ button will scroll backward through the list.  
Pressing “Enter” will reset the currently displayed item. Pressing “Esc” will return  
to Run Mode without resetting anything. Note that all menu lists on the Model  
9840 start with the “>“ symbol and that the Esc button will always back out of a  
menu without making any changes.  
View Button  
The View button switches between two virtual displays stored in the memory.  
Each of these displays will remember the units you have selected for each item,  
and which item is currently being shown. This may save you from scrolling  
through the list of items if there are two particular results that you often use. For  
instance you might set the display to show Load in Lb, press View, and then set  
the display for Peak in PSI. Repeated presses of the View button will now switch  
back and forth between Load in Lb and Peak in PSI with a single button push. If  
you choose to have the second line show a display, then the View button toggles  
between the display currently being activated. The activated display will have a  
“>” symbol to the left of it.  
Esc Button  
In Run Mode, the Esc button will freeze and unfreeze the run mode numerical  
display. Press once to freeze the display at the current value. Press again to  
return display to actual reading from sensor. See Setup Mode information for the  
use of the Esc button during setup.  
MODEL 9840  
PG 11 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Print Button  
If the Print option is purchased, the Print Button can be used to either print data or  
to freeze the display. If the Hold option is purchased, the functionality of the  
button is set using the System Setup options. Otherwise, the button can only  
control the ability to print data.  
In Run Mode, if the System Setup Options Print/Hold button is set to Hold, the  
Print button will freeze and unfreeze the run mode numerical display. Press once  
to freeze the display at the current value. Press again to return display to actual  
reading from sensor. See Setup Mode information for the use of the Esc button  
during setup.  
If the Print button is set to Print, it will send a set of data to the printer. This will  
contain the current readings of load, peak, valley, position, velocity, and limit  
status. The units used for printing will be the units last selected on the numerical  
display.  
The printer must be a serial printer that supports ESC/P protocol. Many Epson  
printers with serial ports are known to be compatible.  
Appendix B gives the pin assignments on the serial connector.  
Unit Button  
The Unit button cycles through the list of units that are available for the item that is  
currently being displayed.  
MODEL 9840  
PG 12 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
SETUP MODE SUMMARY  
The table below summarizes the setup mode menus. Detailed information about each  
individual item is presented in the following sections.  
Setup mode  
╠══════  
>User Data Entry  
╠════════  
>>Base Area Ch A/B  
>>Base Length  
╚════════  
╠══════  
>Analog Output  
╚════════  
Source Units Full Zero.  
╠══════  
>Sensor Select  
╚════════  
Channel Serial Number List Cal Info.  
╠══════  
(password access and cell type)  
>>Cal Check  
>>Cal by mV / volt  
>>Cal by Masses or Torque  
>>Cal by Shunt  
>>Counts / Inch  
>Calibration →  
╠════════  
╠════════  
╠════════  
╠════════  
╚════════  
╠══════  
>Limits  
╚════════  
Select NO/NC Enable →  
Source Units Set Trip →  
Latch Reset Done.  
╠══════  
>System Options  
╠════════  
╠════════  
╠════════  
╠════════  
╠════════  
╠════════  
╠════════  
╚════════  
>>Printer Baud  
>>Auto Identify  
>>Auto Zero Ch A/B  
>>Com Address  
>>Com Baudrate  
>>Com Line Feed  
>>Retain Tare  
╠══════  
>>RS232 EOT Char  
╚══════  
>Display Options  
╠════════  
╠════════  
╠════════  
╠════════  
╚════════  
>>Filter Select  
>>Decimals Ch A/B  
>>Count By Ch A/B  
>>Second Line  
>>Filter Window  
MODEL 9840  
PG 13 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
SETUP MODE  
Setup mode is used to change the setup of the Model 9840. To enter setup mode press  
the two end buttons at the same time ( “<“ and “>“ ). The row of four buttons will now  
function as indicated by the lower set of labels (Plus, Minus, Enter, and Escape).  
If the optional keylock is in the locked position you will see “>Setup is Locked” when you  
try to enter setup mode. This prevents accidental or unauthorized changes to the setup of  
the unit.  
To exit from setup mode and return to run mode press the escape button (Esc). When  
you are entering numerical data or selecting options from a sub-menu the escape button  
will back you out without making any changes.  
The main menu in setup mode contains the following 6 entries which will be described in  
detail in the following sections. Note that all main menu items begin with the “>“ symbol  
and that all submenu items begin with the “>>“ symbol.  
> User Data Entry -- allows you to set any common math options such as the base  
area used to display load in PSI or the base length used to display position in percent.  
> Analog Output -- is used to select the source and scaling for the standard analog  
output channel.  
> Sensor Select -- lets you manually set which load or torque cell calibration data is  
used. If you use auto-id cells you will never need this menu.  
> Calibration -- provides access to calibration for cells or optical encoders.  
> Limits -- allows you to set the source and options for the 4 contact closure switches.  
> System Options -- contains entries to set the printer baud rate, enable or disable  
automatic sensor identification, enable or disable automatic load or torque zeroing, set  
the serial communications address, baud rate, line feed, and RS232 options, and  
choose whether or not to retain the most recent tare value through power down.  
> Display Options -- lets you choose the filtering that is applied to the load or torque  
measurements, an optional window of data incrementation on which to turn the filter  
off, the number of decimal digits that are shown on the display and on the print, the  
count by values for channels A and B (1, 2, 5, 10, or 20), and whether the second line  
of the display shows the limit switch status, a blank line, another display, or text.  
The Plus or Right button ( “+” or “>“ ) will scroll forward through this menu. The Minus or  
Left button ( “-” or “<“ ) will scroll backward through this menu. The Enter button selects  
the currently displayed menu item. The Esc button will return you to Run Mode.  
MODEL 9840  
PG 14 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Entering Numerical Data  
At times you will need to enter numerical data into the Model 9840. The method is  
consistent in all cases. A number is presented on the left side of the display with an  
underline cursor.  
Left ( “<“ ) moves the cursor left one position. If you are on the leftmost digit a new  
zero will be added. Thus 10.000 will become 010.00 where the cursor has moved  
from the tens place to the hundreds place.  
Right ( “>“ ) moves the cursor right one position. If you are at the far right a new  
decimal will be added if there is room. Thus 12.34 will become 12.340 where the  
cursor has moved from the tenths to the hundredths place.  
Plus ( “+” ) increments the digit above the cursor. Thus 12.34 becomes 13.34. Note  
that 9 will “wrap around” to 0.  
Minus ( “-” ) decrements the digit above the cursor. Thus 12.34 becomes 11.34.  
Note that 0 will “wrap around” to 9.  
Plus and Minus (both “+” and “-” at the same time). This will change the sign of the  
number. Thus 12.34 becomes -12.34. Note that cursor position does not matter when  
you change the sign.  
Enter accepts the currently displayed number.  
Esc escapes from numerical entry and discards any changes that have been made.  
Entering Alphanumeric Data  
When entering cell serial numbers, the option of using numbers and/or letters from A to Z  
is available. This works in the same way as entering numerical data with the exception  
that no decimal point exists in the display.  
>User Data Entry  
This menu is used to enter numerical constants used by the math options. These  
constants are listed in a submenu that works exactly like the main setup mode menu.  
Typical items in this submenu are:  
>>Base Area Ch A/B  
This is the area in square inches that is used to calculate load in pounds per square  
inch (PSI) or Mega Pascals (MPa).  
>>Base Length  
This is the length in inches that is used to calculate position in percent.  
>Analog Output  
This menu is used to set up the standard analog output. Instead of a submenu there is a  
series of selections. Pressing escape at any time will leave the analog output unchanged  
and return you to the setup mode main menu. Note that when you change a cell type  
(from load to torque or back again) these settings must be reset for the new cell. If the  
unit is a 2-channel unit and the Analog Output B option has been enabled, the first option  
asks the user for the Analog Output channel. Select A or B. Then begin selecting the  
following settings.  
MODEL 9840  
PG 15 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
1. Source: -- This entry allows you to scroll through a list (using the left/right or  
plus/minus buttons) to select the source of the data that will drive the analog output.  
The standard list will include load or torque, peak, valley, gross, position, and velocity.  
Channel A+B is an option which, if chosen, will appear only if the cells on both  
channels are of the same type. Press enter to select the displayed item.  
2. Units: -- This entry allows you to scroll through the list of available units for the data  
source you have chosen in the previous step.  
3. Full -- This entry specifies the value of source data (in the selected units) that will  
cause the Model 9840 to output positive full scale analog voltage (+10.0 volts).  
4. Zero -- This entry specifies the value of source data (in the selected units) that will  
cause the Model 9840 to output zero analog voltage (0.0 volts).  
Once you have entered these 4 selections the analog output is setup.  
4-20MA Output Option  
This option allows the user to output with current (milliamps) rather than voltage. It  
operates similarly to Analog Output using the same submenus, but, in this case, the Full  
and Zero data entered control the distribution of output in milliamps from the unit given  
certain inputs to the cell. Particularly, the number entered for Zero will set the  
corresponding input in Lbs to output 12 milliamps (the mid-point between four and twenty  
milliamps) and the unit will extrapolate other output values between 4 and 20 milliamps.  
If this option is enabled, the following steps will set up the unit: follow the instructions for  
Analog Output until "Source:" is displayed. Toggle through until the "Load A" is on the  
display. Press the enter button to bring up the "Lb" menu. Press the enter button again  
and the "4.0000 Lb Full" menu will be displayed. This is where the maximum Lbs for the  
cell should be entered. Press enter and the "0.0000 Lb Zero" menu will be displayed.  
This is where the offset from zero (the value in lbs that will give a 12 milliamp) should be  
entered. Press the enter button to return to the Analog Output menu and then press the  
escape button to return to the normal display.  
Example outputs with varying Full and Zero Values (assuming a 4.00 Lb calibrated cell  
with 10 V excitation):  
1. Full = 4.0000 Lb  
(Input) Mv/V  
Zero = 0.0000 Lb  
(Output) Volt  
(Output) MA  
4
3
2
1
10v  
20mA  
18mA  
16mA  
14mA  
12mA  
10mA  
8mA  
7.5v  
5.0v  
2.5v  
0v  
-2.5v  
-5.0v  
-7.5v  
-10.0v  
0
-1  
-2  
-3  
-4  
6mA  
4mA  
MODEL 9840  
PG 16 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
2. Full = 4.0000 Lb  
(Input) Mv/V  
Zero = 4.0000 Lb  
(Output)Volt  
(Output) MA  
4
3
2
1
0v  
12mA  
10mA  
8mA  
6mA  
4mA  
4mA  
4mA  
4mA  
4mA  
-2.5v  
-5.0v  
-7.5v  
-10.0v  
-10.0v  
-10.0v  
-10.0v  
-10.0v  
0
-1  
-2  
-3  
-4  
3. Full = 2.0000 Lb  
(Input) Mv/V  
Zero = 0.0000 Lb  
(Output) Volt  
(Output) MA  
4
3
2
1
10.0v  
10.0v  
10.0v  
5.00v  
0v  
-5.0v  
-10.0v  
-10.0v  
-10.0v  
20mA  
20mA  
20mA  
16mA  
12mA  
8mA  
4mA  
4mA  
4mA  
0
-1  
-2  
-3  
-4  
4. Full = 2.0000 Lb  
(Input) Mv/V  
Zero = 2.0000 Lb  
(Output) Volt  
(Output) MA  
4
3
2
1
10.0v  
5.0v  
0v  
-5.00v  
-10.00v  
-10.0v  
-10.0v  
-10.0v  
-10.0v  
20mA  
16mA  
12mA  
8mA  
4mA  
4mA  
4mA  
4mA  
4mA  
0
-1  
-2  
-3  
-4  
>Sensor Select  
This menu is used to manually select load or torque cell calibration data from a list stored  
in the Model 9840. When you select this menu on a 2 channel unit, you are asked to  
select Channel A or B. The cell serial number of the selected channel is displayed. Use  
either the left/right or plus/minus buttons to page through the list. The escape button will  
MODEL 9840  
PG 17 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
return to the setup mode main menu without changing the cell that is selected. The enter  
button will select the currently displayed cell and briefly display the rated load or torque,  
last calibration date, and excitation voltage.  
To add new cells to this list use the Calibration entry on the setup mode main menu  
described below.  
If you are using auto-id load cells this menu will allow you to see which cell is selected but  
will not allow you to change the selection. This ensures that the correct calibration  
information is always used when you use Auto-ID load cells.  
If the selected cell is of a different type (load or torque) than that previously displayed on  
the channel, the display item and units will change accordingly.  
Sensor Delete  
The Model 9840 can store calibration data for 25 different cells. To delete a cell from this  
list choose “>Sensor Select” from the setup mode main menu. Press the left/right or  
plus/minus keys until the serial number of the cell you wish to delete is displayed. (You  
may need to shut off auto-id if it is on, see the System Options menu for details). Then  
press both the plus and minus keys at the same time. The Model 9840 will ask “Delete  
Sensor?” to be sure that you intend to delete this calibration data. Pressing Enter will  
delete the cell data from the list. Pressing Escape will return to the cell select without  
changing the list.  
>Calibration  
This menu is used to access all methods of calibration that are supported by the Model  
9840. Selecting calibration from the setup mode main menu will bring up a password  
entry field. Successful entrance of the password (888) will bring up a submenu which  
asks for the cell type. Toggle through either Load or Torque and press enter when  
appropriate. Another submenu will appear listing the various types of calibration. This  
submenu is used exactly like the main menu. Press left/right or plus/minus to change the  
menu item. Press escape to return to the setup mode main menu. Press enter to select  
the displayed menu item.  
If you are re-calibrating a cell, the Model 9840 will recognize it using either the auto-id or a  
manually entered serial number. Any time you overwrite existing cell calibration data you  
will first be warned with the “Overwrite Data?” message. Pressing enter will overwrite;  
pressing escape will cancel the calibration. Another method of overwriting existing data is  
to delete the sensor first and then calibrate.  
If a cell with a new serial number (or new auto-id) is calibrated the data is added as a new  
entry in the cell list.  
Cells must be calibrated for the channel on which they will be used.  
MODEL 9840  
PG 18 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
>>Cal Check  
Selecting this item will run a calibration check on the cell attached to Channel A or B.  
The Model 9840 will measure the shunt value of the cell and compare it to the shunt  
value that was recorded when the cell was last calibrated. The rated load or torque  
and last calibration date for the cell are displayed followed by the currently measured  
shunt value and the shunt value that was recorded when the cell was calibrated.  
Note that the Model 9840 has a switch on the back panel that allows you to select  
either a 30K shunt or a 60K shunt. Be sure that this switch is set correctly for the cell  
that you are using.  
Since this shunt check electronically applies approximately half the rated load or  
torque, it is important that the cell is not loaded when this check is run. If you see the  
“Cell Overloaded” message, remove any load or torque on the cell and try the check  
again.  
>>Cal by mV/volt (2-Point or 6-Point)  
This item is used to calibrate a cell using the manufacturer’s milli-volt per volt  
calibration factor (the second point of a two point calibration that automatically sets the  
first point at zero) or a series of six milli-volt per volt points with corresponding masses  
(usually listed on the load cell calibration certificate) or torques. Before you enter in  
the desired information, another menu will ask you if you want to calibrate by milli-volt  
per volt or 6 point milli-volt per volt. Pressing the arrow keys will allow you toggle  
between mV/volt and 6 point mV/volt modes. You will then be asked for the channel,  
the cell serial number, the current date, the calibration unit, the cell’s maximum rated  
load or torque, and the desired excitation voltage (5.0 or 10.0 volts). If milli-volt per  
volt calibration was selected, you are then prompted to enter the manufacturer’s milli-  
volt value. If 6 point milli-volt per volt calibration was selected, you are prompted for  
each mass or torque and milli-volt per volt value. A piece-wise linear interpolation is  
then performed between adjacent points. One of these points should be zero mass  
and zero milli-volts per volt. Enter each of these selections as prompted or press the  
escape button at any time to return to the setup mode main menu without changing  
the calibration. The cell should not be under load or torque when this calibration  
method is used since a shunt value will be measured and recorded for later use with  
“Cal Check” feature described above. The “No Masses Ready” or “No Torque Ready”  
message is intended to remind you of this fact. This general sequence of prompts  
and user responses also applies to the 5 point mass or torque calibration.  
>>Cal by Masses or Torque (2-Point or 5-Point)  
These items are used to calibrate a load or torque cell using known masses or  
torques. The appropriate choice will be available depending on which cell type was  
chosen at the beginning of calibration: a load cell will show Masses as an option and a  
torque cell will show Torque as an option. Before you enter in the desired information,  
another menu will ask if you want to calibrate by 2 Point or 5 Point. You will then be  
asked for the channel, the cell serial number, the date, the calibration unit, the cell’s  
maximum rated load or torque, and the desired excitation voltage (5.0 or 10.0 volts).  
You are then prompted to enter the value of the first mass or torque. The “Mass 1  
Ready?” or "Torque 1 Ready?" message will wait while you actually hang the mass or  
MODEL 9840  
PG 19 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
set the torque. When the Model 9840 is reading the cell it will display the “Reading...”  
message for about 10 seconds. During this time it is important that the masses or  
torque are not disturbed since all the readings taken are averaged together to obtain  
the calibration data. The second mass/torque is then entered and read. If 5 Point  
calibration was selected, masses/torque three through five will be entered and read.  
One of these readings should be at zero. A "No Masses Ready?” or "No Torque  
Ready" reading is also required. During this reading, a shunt check value is recorded.  
>>Cal by Shunt Val  
This item is used to calibrate a cell using the manufacturer’s shunt calibration factor  
expressed in pounds or pound-inches. You will be asked for the channel, the cell  
serial number, the date, the calibration unit, the cell’s maximum rated load or torque,  
the desired excitation voltage (5.0 or 10.0 volts), and the shunt calibration constant.  
The Model 9840 will then automatically perform the shunt calibration of the cell using  
its internal precision shunt resistor. Note that there is a back panel switch that allows  
you to select either a 30K or 60K resistor for shunt measurements (see Appendix B for  
a circuit diagram). Be sure to select the appropriate value for your cell. During this  
calibration the “Reading...” message will be displayed. Do not disturb the cell while it  
is being read. Like the “Cal by mV/volt” option, the cell should be free of load or  
torque during this calibration.  
>>Counts per Inch  
This entry allows you to set the number of counts per inch that the quadrature encoder  
input will use.  
>Limits  
This menu is used to set up the 4 contact closure switches. Instead of a submenu there  
are a series of selections. Pressing Enter will accept the displayed entry. Pressing  
Escape will return to the setup mode main menu without changing the limit setup. Note  
that changing a cell type on a channel will automatically erase limits that were set for that  
channel. These limits must then be reset by the user.  
1. Select Limit X -- The first selection chooses which of the limits you want to set up (1-  
4). Use the left/right or plus/minus buttons to change the number. Press enter to  
accept the displayed number. Press escape to return to the setup mode main menu.  
2. Normally Open / Closed -- This selection determines if the switch contacts are open  
or closed when the limit is de-activated.  
3. Enabled / Disabled -- Choosing enable allows the limit to operate as specified by the  
following selections. Choosing disable places the limit in its “normal” state and  
completes the limit setup.  
4. Source: -- This selection allows you to scroll through a list (use either the left/right or  
plus/minus buttons) to select the source of the data that will drive the limit. The  
standard list will include load or torque, peak, valley, position, and velocity. Press  
enter to select the displayed item.  
5. Units: -- This entry allows you to scroll through the list of available units for the data  
source you have chosen in the previous step.  
6. Set -- This is the numerical value of the source at which the limit will be activated (set  
point). It is specified in the units you chose in the previous step.  
MODEL 9840  
PG 20 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
7. Trip if source < / > Set -- Choosing source>set means that the limit will be activated  
when the source becomes larger than the set point. Choosing source<set means the  
limit will be activated when the source becomes less than the set point.  
8. Latching on/off -- Selecting latching off means that the limit will be de-activated  
automatically when the source passes the reset point (see next selection). Selecting  
latching on means that once the limit is activated you must manually reset it using the  
Reset button. If you select latching on then the limit setup is done.  
9. Rst -- This selection determines the numerical value of the source at which the limit  
will be deactivated (reset point). If you selected “source>set” in step 7 then the limit is  
deactivated when the source is less than the reset point. If you selected “source<set”  
in step 7 then the limit is deactivated when the source is greater than the reset point.  
Notes:  
The Set and Reset conditions for each limit are evaluated independently. If both  
conditions are met then Reset will take priority over Set.  
Limit status may show an asterisk if the limit is enabled but there is an error in the signal  
source. For example, if the limit source is load but there are no load cells currently  
calibrated, or the load cell has a rating of 0.0 Lb, or a millivolt per volt constant of zero.  
Another example would be if the limit source is position and zero has been entered for the  
encoder counts-per-inch constant. Limits that are in this error condition will be forced to  
the off position as if they were disabled.  
>System Options  
This menu contains some system wide parameters that are seldom changed. These are  
listed in a submenu that works exactly like the setup mode main menu.  
>>Printer Baud  
The front panel print button and the RS232 print commands will use this baud rate for  
communication with your printer. This setting has no effect on units that are not  
equipped with the printing option. 57.6 K baud is commonly used on older serial  
printers (Epson Stylus II), 230.4 K baud is used on more recent serial printers (Epson  
Stylus Color 600, Color 800).  
>>Auto Identify  
This allows you to turn automatic cell identification on or off. Usually you can leave  
auto-id on even if you never use automatic cell identification. In order to use this  
function you must either have an auto-id equipped cell or use an in-line auto-id module  
attached to your cell cable.  
>>Auto Zero Ch A/B  
This menu lets you enable or disable auto zeroing for measurements on Channel A or  
B. When enabled, this function will tare the load or torque to zero any time that it has  
remained within 10 counts of zero on the display for 10 seconds.  
MODEL 9840  
PG 21 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
>>Com Address  
This item allows you to set the address of this unit used for the RS232 ASCII  
command set. Valid values are 1 through 254. If the RS232 multi-drop option is  
installed you must be sure that all the attached units have different addresses to  
prevent data collisions.  
>>Com Baudrate  
This sets the baud rate for the RS232 ASCII command set. Available values are: 300,  
600, 1200, 2400, 4800, 9600, and 19.2k. Note that the “Special” entry is reserved for  
future use.  
>>Com Line Feed  
This controls whether a line feed is added to each carriage return that is sent out by the  
Model 9840 when the RS232 ASCII command set is being used.  
>>Retain Tare  
This controls whether the most recent tare value is retained through power down of the  
unit and used again immediately on power up. Choosing this option will retain tare on  
channel A in a single channel unit and both channels A and B in a two-channel unit.  
>>RS232 EOT Char  
The end of transmission character is a non-visible character (hexadecimal 0x04) that is  
sent at the end of each response by the unit to an RS232 command. The character is  
sent after the last carriage-return line-feed returned by the unit with one exception:  
when the user is viewing channel values with the RS232 command  
@123V(item)(unit)(repeat) and sets the value of repeat to 2 (streaming output), the  
character is sent immediately prior to the streaming output that the unit will send. The  
user can turn this option on or off.  
>Display Options  
This menu contains options that effect how load or torque, peak, and valley are displayed  
and printed. These are listed in a submenu that works exactly like the setup mode main  
menu.  
>>Filter Select  
There are two types of filters available on the Model 9840: Type I, and Type II. Type I  
filters are good for removing most kinds of noise but may leave some jitter on the end  
digits. Type II filters are more advanced and are optimized for the typical industrial  
environment. Both filter types allow you to select from 4 levels of filtering with Level 1  
being the least filtering and Level 4 being the most filtering.  
The settling time for each of the 4 levels of filtering is shown in the table below  
Level  
1
2
3
4
Settling Time (seconds) <1  
Read Rate (per second) 60  
2
60  
10  
30  
30  
10  
MODEL 9840  
PG 22 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
>>Decimals Ch A/B  
This lets you set the maximum number of digits to the right of the decimal that are  
displayed or printed for load or torque, peak, and valley. When large cells are selected  
the number of decimal digits shown may be limited, for example a 100 Lb cell will allow  
3 digits, while a 1000 Lb cell will only allow 2 digits. Also if a number is too large to fit  
on the display some decimal digits may be dropped (the display will “auto-range”). The  
number of digits shown for position data depends only on the number of counts per  
inch of the encoder and is not affected by this setting.  
>>Count By Ch A/B  
This allows you to choose the smallest unit of counting for the display of load or torque,  
peak, and valley. Counting by 1, 2, 5, 10, and 20 are the options.  
Note that the operation of the analog outputs and the limits is not effected by the  
settings of “decimal point” or “smallest count”. Thus if you set a limit to be activated at  
10.001 Lbs, and have “smallest count” set for 20, the limit is activated at 10.001 Lbs  
even though the display is only showing 10.000.  
>>Second Line  
This option allows you to choose what you want to have displayed on the second line.  
You can select limits, select a display, select text, or leave the line blank. Changing the  
type of cell attached to the unit from load to torque or back will reset the display such  
that both the top and bottom line will revert to the new cell type and its initial display -  
Load for a load cell, and Torq for a torque cell - for the channel on which the cell type  
was changed.  
>>Filter Window  
This option allows you to enable and set, or disable, a window to be used to turn the  
filters on or off. If there is a sudden change in the measurement that exceeds this set  
window, the filter is turned off and immediately turned back on so that this change of  
data will be visible to the user rather than filtered out. To turn the window on, press  
enter under the Filter Window option and toggle between on and off using the > and <  
buttons. Once the window is turned on, pressing enter and will display the current  
window's value and units (0 and Lbs. or LbI, respectively). Press enter again and you  
will be asked to set the units (toggle buttons take you through the list). Press enter  
once more and you reach the screen to set the value of the window. Once the value is  
set, press enter and you will return to the Filter Window option.  
To change the window, simply repeat the steps above. Instead of 0 and Lbs. in the  
current window display, you will see the most recently entered values. Turning the  
window off resets the window value to 0 and Lbs (or LbI for torque).  
Note that changing a cell type on a channel will automatically disable the filter window  
for that channel, reset the units to Lb or LbI, and set the window value to 0. The  
window must then be reset and re-enabled by the user.  
MODEL 9840  
PG 23 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
REMOTE OPERATION  
Digital Inputs  
There are 4 digital inputs available on the Model 9840. They are accessible on the Digital  
I/O connector on the back panel (pins 9 - 16). These inputs are individually opto-isolated  
and include current limiting resistors. An input voltage anywhere from +4 to +22 volts DC  
may be used to obtain the “on” state. See Appendix B for a full description of the Digital  
I/O connector.  
If you do not require isolation for your application you can use the +5 or +12 volt and GND  
available on the Digital I/O connector (pins 24-26).  
These 4 inputs are called IN1 through IN4 and have the following standard functions:  
IN1 will Tare the Load or Torq to zero while it is in the On state.  
IN2 will Reset the Peak and Valley measurements while it is in the On state.  
IN3 will Reset the Position measurement while it is in the On state.  
IN4 will send one set of measurements (load or torque, peak, valley, position, velocity,  
and limits) to the serial printer when it transitions from Off to On.  
Other functions may be assigned to these inputs by custom programming at the factory.  
All the features and settings of the Model 9840 can be accessed remotely using the  
Remote ASCII command set. These commands fall into two groups: commands that are  
similar Run Mode pushbutton commands; and commands are similar to Setup Mode  
pushbutton commands.  
Communications can be established via RS232/RS485 protocols through the serial port  
on the rear panel of the unit, or via USB communications through the USB port on the rear  
panel of the unit in conjunction with the USB driver installed on the host PC. The  
commands and command formats are the same for each method of communications.  
Pinouts for the RS232/RS485 serial port communications are listed in Appendix B of this  
manual. For older computers, USB communication may require the installation of the  
USB driver located on the CD-ROM that accompanies this product. If the host PC can  
communicate with the Model 9840 via USB, no driver is needed.  
USB Driver  
If the host PC cannot communicate with the Model 9840 via USB, install the USB driver  
that accompanies this product on CD-ROM. To install the USB Driver, place the CD in  
your CD-ROM drive. Select Run from your Windows Start menu. Type  
D:\USB_Driver\CDM_Setup.exe (“D” or the corresponding CD-ROM drive letter for your  
computer) in the box and hit Enter. You will be notified when the driver has finished  
installing.  
MODEL 9840  
PG 24 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Communications Settings  
The Model 9840 supports remote operation using a standard Remote interface consisting  
of ASCII characters and terminal emulation software package such as HyperTerminal or  
ProComm. The baud rate is set using the Com Baud rate item on the System Options  
menu. Supported rates are 300, 600, 1200, 2400, 4800, 9600, and 19.2k baud. There  
must be 8 data bits, no parity, and 1 stop bit.  
To support the multi-drop Remote option each unit is assigned a communications  
address, which is set using the Com Address item on the System Options menu.  
Communications Format  
All the commands have the following format:  
@123XYZ<CR>  
The @ symbol initiates the command, all commands must begin with this symbol. The  
number 123 is the address of the Model 9840 which should respond to this command.  
This MUST be 3 digits so type 001 for unit one, and 026 for unit twenty-six. The  
command itself will be a sequence of letters such as the XYZ shown above. If the  
command requires a number as one of its parameters the number must be followed by the  
pound sign “#”. Numbers that are less than one should include a leading zero so you  
should type 0.123# to enter one hundred-twenty three thousandths. Commands are not  
processed until a carriage return <CR> is sent.  
No Model 9840 will respond to address 000. All Model 9840’s will respond to address  
255. Throughout the rest of this section we will assume that you are using address 123.  
Whenever an Model 9840 executes a remote command it will issue an acknowledge that  
includes the address and any required data.  
RS232 RUN MODE COMMAND SUMMARY  
@(addr)  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
H
?
FV  
Hello fetches unit description  
Get item and unit numbers  
View item and unit settings  
Set displayed item and unit  
Alternate display, like view button  
Switches view pointer to display  
line 1 (upper)  
Switches view pointer to display  
line 2 (lower), or virtual display if  
line 2 is disabled  
FS (item)(unit)  
FA  
F1  
╠═  
F2  
╠═  
╠═  
╠═  
╠═  
╚═  
V
P
R
X
T
(item)(unit)(repeat)  
(repeat)  
(tare)(peak)(vall)(pos)  
Get Value of an item and unit  
Print a set of readings  
Reset readings  
Freeze display  
Place text on display  
(hold)(text)  
MODEL 9840  
PG 25 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
RS232 RUN MODE COMMANDS  
The commands listed in this section mirror (and in some cases extend) the functions that  
are available from the front panel of the Model 9840 when it is in Run Mode.  
Each entry will include the name of the command, followed by a short description that  
includes the command format, an example that shows what you would type (shown in  
plain font) and what the Model 9840 would return (shown in italics).  
Hello Command (H)  
This command fetches the model, version, and serial number of any unit at the specified  
address.  
Example:  
@123H  
Acknowledge:  
@123 Interface Model 9840 Version 6.0.0 Serial #-980500 Option #  
457977664  
Question Mark Command (?)  
This command returns a list of item numbers and unit numbers that are available on the  
Model 9840 that is addressed. These numbers are used as arguments for the many of  
the commands shown below**.  
Example:  
@123?  
Acknowledge:  
@123 These are the Item numbers:  
00 - Load A  
04 - Peak B  
13 - Limits  
01 - Peak A  
05 – Vall B  
14 - Grs A  
02 - Vall A  
09 - Pos  
15 - Grs B  
03 – Load B  
10 - Vel  
16 - Ch A+B  
**Alternatively torque units may be displayed or a combination of  
both depending upon which cell is loaded on which channel:  
17 - Torque A 18 - Peak A  
21 - Torque B 22 - Peak B  
25 - Ch A+B  
19 - Vall A  
23 - Vall B  
20 – Gross A  
24 – Gross B  
These items are for value display purposes only:  
50 – Cell* AB  
51 - Peak AB  
52 - Vall AB  
*
Note: If the channel is set as a load channel, the cell type implied is a LOAD  
CELL; if the channel is configured as a torque channel, the cell type implied is a  
TORQUE CELL.  
These are the units for Load, Peak, and Valley:  
00 - Lb  
04 - MPa  
08 – mVv  
01 - kg  
05 - Klb  
09 - g  
02 - N  
03 - PSI  
07 - t  
06 – kN  
These are the units for Torque:  
00 – LbI 01 – NM  
02 - OzI  
03-mVv  
These are the units for Position:  
00 – In  
01 – Cm  
02 - %  
MODEL 9840  
PG 26 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
These are the units for Velocity:  
00 – I/M 01 – C/M  
Front Panel Display Command (F)  
This command is used to set the front panel display of the Model 9840 to a desired  
combination of item and unit. There are five subcommands: Front panel View (FV), Front  
panel Set (FS), Front panel Alternate(FA), Front panel pointer control to line 1 (F1), and  
Front panel pointer control to line 2 (F2) .  
Front Panel View Command (FV)  
The Front panel View (FV) command just tells you what is currently on the display, and  
what will be displayed when the View button on the front panel is pressed.  
Example:  
@123FV  
Acknowledge:  
@123 Active Display shows Load A in Lb  
Other Display shows Peak A in Lb  
Front Panel Set Command (FS)  
The Front panel Set (FS) command allows you to specify the item number and unit  
number for what is currently on the front panel display. The format is FS (item  
number)(unit number), where the first item and unit numbers are 2 digits each and may  
be found using the question mark command (?).  
Example:  
@123FS0200  
Acknowledge:  
@123 Active Display shows Vall A in Lb  
Other Display shows Peak A in Lb  
Front Panel Alternate Command (FA)  
The Front panel Alternate (FA) command allows you to switch between the currently  
displayed item and unit and the alternate item and unit just as if you had pressed the  
View button.  
Example:  
@123FA  
Acknowledge:  
@123 Active Display shows Peak A in Lb  
Other Display shows Vall A in Lb  
Front Panel Pointer Control to Line 1 (F1)  
The Front Panel Pointer Control to Line 1 (F1) assigns the view pointer control to  
display line 1 (upper line of the display) and shows what is on the display. This  
allows the operator to know which line will be configured or accessed.  
Example:  
MODEL 9840  
PG 27 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
@123F1  
Acknowledge:  
@123 Active Display shows Load A in Lb  
Other Display shows Peak A in Lb  
Front Panel Pointer Control to Line 2 (F2)  
The Front Panel Pointer Control to Line 2 (F2) assigns the view pointer control to  
display line 2 (lower line of the display, or the virtual display line if line 2 is  
disabled) and shows what is on the display. This allows the operator to know  
which line will be configured or accessed.  
Example:  
@123F2  
Acknowledge:  
@123 Active Display shows Peak A in Lb  
Other Display shows Load A in Lb  
Value Command (V) for one or both channels  
The value command for one channel is used to read the numerical value of a desired item  
and unit without altering the front panel display. The format is V(item number)(unit  
number)(repeat), where the item and unit numbers can be found using the question mark  
command (see above), and the repeat number must be 0,1, or 2. Choosing repeat 1 will  
make the Model 9840 send you the desired value once. Choosing repeat 2 will make the  
Model 9840 send the value at approximately 3 second intervals. Choosing repeat 0 will  
turn off any values that are being sent by a repeat 2 command. Note: when choosing to  
view Limits, use "00" for the unit part of the command.  
Example:  
@123V01021  
Acknowledge:  
@123 Peak A 55.676 N  
Example 2:  
@123V00002  
Acknowledge 2:  
@123 Load A 120.45 Lb  
@123 Load A 125.29 Lb  
.....  
@123 Load A 137.81 Lb  
The value command for both channels is similar to the command for one channel with the  
following exceptions:  
The item numbers available for two channel value display are 50 (Cells AB), 51 (Peak  
AB), and 52 (Valley AB).  
The format is V(item number)(unit number channel A) (unit number channel B)(repeat).  
where the item must be 50, 51, or 52 and unit numbers can be found using the question  
mark command (see above). Make sure to use the proper unit numbers for the proper cell  
type. A unit number of 08 for a torque cell will result in an unusable argument message.  
MODEL 9840  
PG 28 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
The repeat number is exactly like the one channel command. It must be 0, 1, or 2.  
Choosing repeat 1 will make the Model 9840 send you the desired value once. Choosing  
repeat 2 will make the Model 9840 send the value at approximately 3 second intervals.  
Choosing repeat 0 will turn off any values that are being sent by a repeat 2 command.  
Example 1:  
@123V5000002  
Acknowledge 2:  
@123 Load A 120.45 Lb  
@123 Load A 125.29 Lb  
.....  
Torq B 120.45 LbI  
Torq B 137.81 LbI  
@123 Load A 128.29 Lb  
Torq B 146.81 LbI  
Print Command (P)  
If the print option is purchased, the print command will fetch a full set of readings (like  
pressing the front panel print button). These will be in the units that are currently selected  
on the numerical display. The format is P(repeat number), where the repeat number must  
be 0, 1, 2, or 3. Repeat 1 will return 1 set of readings to the RS232 terminal. Repeat 2  
will return a set of readings approximately every three seconds (like the value command).  
Repeat 0 will turn off printing that was started with repeat 2. Repeat 3 will send one set of  
readings to the serial printer attached to the Model 9840 instead of to the RS232 terminal.  
Example:  
@123P1  
Acknowledge:  
@123 Load A 45.625 Lb  
Peak A 82.396 Lb  
Vall A -0.005 Lb  
Load B 32.415 Lb  
Peak B 47.483 Lb  
Vall B –0.0001 Lb  
Pos  
Vel  
3.500 In  
0.0 I/M  
Grs A 45.625 Lb  
Grs B 32.415 Lb  
Ch A+B 78.040 Lb  
Limits 0 0 0 1  
Reset Tare, Peak, Valley or Position Command (R)  
This command allows you to reset the tare, peak, valley, or position readings. The format  
is R(tare A)(peak A)(valley A) (tare B)(peak B)(valley B)(position select), where the selects  
are either a 0 (don’t reset) or a 1 (reset). Thus you can reset any combination of these  
readings with a single command.  
Example:  
MODEL 9840  
PG 29 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
@123R1000000  
Acknowledge:  
@123 Reset – Tare A  
Example 2:  
@123R0111000  
Acknowledge 2:  
@123 Reset – Peak A Valley A Tare B  
Freeze Display Command (X)  
In Run Mode, the freeze display command will freeze and unfreeze the run mode  
numerical display and the values sent from the unit for RS232 output using either the  
remote Value or Print command. Send the command once to freeze the display and  
output at the current value. Send again to return display and output to actual reading from  
sensor.  
Text Display Command (T)  
This command allows you to put a line of text on the front panel display of the Model  
9840. The format is T (text), where text is the message that you want to display. The  
text is limited to the number of characters that can be shown on the front panel display.  
Example:  
@123TStress Test Ready  
Acknowledge:  
@123 Text Displayed - Stress Test Ready  
MODEL 9840  
PG 30 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
RS232 SETUP MODE COMMAND SUMMARY  
@(addr)  
╠═  
U
User Data Entry  
View current settings  
Set Base Area  
╠═  
╠═  
╚═  
V
A
L
(channel)(number) #  
(number) #  
Set Base Length  
╠═  
A
╠═  
╚═  
Analog Output  
View current settings  
Set analog output  
V
S
(item)(unit)(full)#(zero)#  
╠═  
S
Sensor Select  
╠═  
╠═  
╠═  
╠═  
╚═  
A
B
V
S
D
View current sensor on A  
View current sensor on B  
View current sensors  
Select sensor by S/N  
Delete sensor by S/N  
(channel)(serial number)#  
(serial number)#  
╠═  
C
Calibration  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╚═  
C
B
E
V
Shunt Check load cell  
Begin load cell calibration  
Escape load cell calibration  
Calibrate by mV/V - 1 pt.  
Calibrate by mV/V - 6 pt.  
Calibrate by Mass 2&5 pt.  
Calibrate by Torque 2&5 pt.  
Calibration by Shunt Value  
Encoder counts per Inch  
View counts/inch  
(see text for format)  
(mv/v constant)#  
MV6 (see text for format)  
M
T
S
I
(see text for format)  
(see text for format)  
(shunt value Lb)#  
╠═  
╚═  
V
S(number)#  
Set counts/inch  
╠═  
L(lim)  
╠═  
╠═  
╠═  
╚═  
Limits  
View current settings  
Setup limit  
Escape from limit setup  
Manual reset for latch  
V
S
E
R
(see text for format)  
╠═  
O
System Options  
View current settings  
Printer baudrate  
Auto-Identify load cells  
Auto-Zero load  
Com Address  
Com Baudrate  
Com Linefeed  
Retain Tare  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╠═  
╚═  
V
P
I
Z
A
B
L
T
E
(baudrate code)  
(0 or 1)  
(channel)(0 or 1)  
(new address)#  
(baudrate code)  
(0 or 1)  
(0 or 1)  
(0 or 1)  
End of Transmission Char.  
╚═  
D
Display Options  
╠═  
╠═  
╠═  
╠═  
╠═  
╚═  
V
F
W
D
C
2
View current settings  
Filter select  
Filter Window  
Decimal digits max  
Smallest count  
2nd line shows a blank line, limits,  
text, or a display.  
(type)(level)  
(see text for format)  
(channel)(number digits)  
(channel)(countby code)  
(B, L, T, or D)  
╚═  
T
(text)  
Set the text displayed on 2nd line.  
MODEL 9840  
PG 31 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
RS232 SETUP MODE COMMANDS  
The commands in this section allow you to change the setup of the Model 9840. You will  
notice that these commands form a menu which approximately duplicates the Setup Mode  
menu used during setup of the Model 9840 from the front panel. There are additional  
commands that are for RS232 use only.  
If the optional Keylock is installed and locked you will not be to alter the setup until the unit  
has been unlocked. The acknowledgement will indicate this fact:  
@123 Setup is Locked - use Keylock to unlock  
User Data Entry Command (U)  
This command is used to view or set the options in the user data entry menu, namely the  
base area (used to calculate load in PSI, or MPa) and base length (used to calculate  
position in %). This command has three subcommands: User data View (UV), User data  
Area (UA), and User data Length (UL).  
User Data View Command (UV)  
The User data View (UV) command has the format UV.  
Example:  
@123UV  
Acknowledge:  
@123 Base Area Ch A is 1.00000 sq-in  
Base Area Ch B is 0.50000 sq-in  
Base Length is 3.0000 in  
User Data Area Command (UA)  
The User data Area (UA) command is used to set the base area. The format is  
UA(channel)(number)#.  
Example:  
@123UAA1.0025#  
Acknowledge:  
@123 Base Area Ch A is 1.0025 sq-in  
User Data Length Command (UL)  
The User data Length (UL) command is used to set the base length. The format is  
UL(number)#  
Example:  
@123UL2.5#  
Acknowledge:  
@123 Base Length is 2.5000 in  
MODEL 9840  
PG 32 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Analog Output Command (A)  
This command is used to view or set the analog output of the Model 9840. There are two  
subcommands Analog View (AV) and Analog Set (AS).  
Analog Output View Command (AV)  
The Analog View (AV) command has the format AV. It will return the current settings  
of the analog output.  
Example:  
@123AV  
Acknowledge:  
@123 Analog Output A is Load A in Lb  
10.0 Volts indicates 1000 Lb  
0.0 Volts indicates 0.00 Lb  
Analog Output Set Command (AS)  
The Analog Set (AS) command is used to change the settings of the analog output.  
The format is AS(item number)(unit number)(full scale)#(zero scale)#, where the item  
and unit numbers are used to determine the signal source for the analog output, the full  
scale number indicates what signal level is assigned to full scale analog output of 10.0  
volts, and the zero scale number indicates the signal level assigned to zero volts  
analog output. Note that when you change a cell type (from load to torque or back  
again) these settings must be reset for the new cell.  
Example:  
@123AS090010.0#5.0#  
Acknowledge:  
@123 Analog Output is Pos in In  
10.0 Volts indicates 10.00 In  
0.0 Volts indicates 5.00 In  
Sensor Command (S)  
This command is used to view or select calibration data from the list of calibrated load  
cells maintained by the Model 9840. There are 4 subcommands, Sensor View (SV),  
Sensor List (SL), Sensor Select (SS), and Sensor Delete (SD).  
Sensor View Channel A or B Command (SA or SB)  
The Sensor View Channel A or B (SA or SB) command has the format SA or SB. It  
returns information about the cell currently loaded on channel A (SA) or channel B(SB)  
including the serial number, the maximum load rating, the millivolt per volt calibration  
constant, the excitation voltage used, the latest date of calibration, and the shunt value  
recorded during that calibration. No other cells are listed.  
MODEL 9840  
PG 33 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Example:  
@123SA  
Acknowledge:  
@123 This is the list of cell calibration data:  
Ch A = S/N 123456, 100.00 Lb  
, 3.00150 mV/v,  
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt  
Sensor View Command (SV)  
The Sensor View (SV) command has the format SV. It returns information about the  
currently selected cell including the serial number, the maximum load rating, the  
millivolt per volt calibration constant, the excitation voltage used, the latest date of  
calibration, and the shunt value recorded during that calibration. All other cells  
currently in the unit are listed as well.  
Example:  
@123SV  
Acknowledge:  
@123 This is the list of cell calibration data:  
Ch A = S/N 123456, 100.00 Lb  
, 3.00150 mV/v,  
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt  
unused S/N 13368, 10.00 Lb  
, 3.00121 mV/v,  
10.00 V,  
Cal on Oct27-99, 4.9292 Lb Shunt  
unused S/N 63220, 100.00 Lb  
, 2.99984 mV/v,  
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb  
, 4.50015 mV/v,  
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt  
Sensor Select Command (SS)  
The Sensor Select (SS) command is used to select the calibration data for a stored cell  
using the cell serial number. The format is SS(channel)(serial number)#.  
Example:  
@123SSA13368#  
Acknowledge:  
@123 This is the list of load cell calibration data:  
unused S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V,  
49.532 Lb Shunt  
Ch A = S/N 13368, 10.00 Lb , 3.00121 mV/v, 10.00 V,  
4.9292 Lb Shunt  
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V,  
49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V,  
486.45 Lb Shunt  
Sensor Delete Command (SD)  
The Sensor Delete (SD) command is used to remove the calibration data for a stored  
cell from the list. The format is SD(serial number)#.  
MODEL 9840  
PG 34 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Example:  
@123SD13368#  
Acknowledge:  
@123 Deleted Sensor S/N 13368  
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V,  
49.532 Lb Shunt  
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V,  
49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V,  
486.45 Lb Shunt  
Calibration Command (C)  
This command has 8 subcommands. Calibration Check (CC) is used to run a shunt  
calibration check on a cell. Calibration Begin (CB) starts the calibration of a new or  
existing cell. Calibration by millivolt-per-Volt may be done using one point (CV) which  
runs a calibration using the cell millivolt-per-Volt constant or 6 point which uses cell  
manufacturer supplied mass and milli-volt per volt data (CMV6). Calibration by Masses  
(CM) does a two or five point mass calibration. Calibration by Torque (CT) does a two or  
five point torque calibration. Calibration by Shunt (CS) does calibration based on the  
shunt value of the cell. Calibration Escape (CE) cancels a calibration that is underway.  
Calibration counts per Inch View (CIV) displays the current setting of the encoder counts  
per inch constant, and Calibration counts per Inch Set (CIS) lets you set the encoder  
counts per inch constant.  
Cells must be calibrated for the channel on which they will be used.  
Calibration Check Command (CC)  
The Calibration Check (CC) command runs a calibration check on the currently  
selected cell. Note that the load cell should not be under load or torque when this  
check is run. During the 10 second shunt calibration the front panel will show “RS232  
Cal Check”. After the 10 seconds the unit will return to run mode and send the  
acknowledge line shown below. This indicates both the current reading of the shunted  
cell and the reading that was recorded when this cell was last calibrated.  
Example:  
@123CCA  
Acknowledge:  
@123 Cal Check Ch A, S/N 123456  
Shunt Value is 49.532 Lb  
Cal Value was 49.531 Lb  
Calibration Begin Command (CB)  
The Calibration Begin (CB) command is a 4 part command used to start a calibration  
by any method. You must issue a CB command before trying calibration by mv/volt,  
masses, or shunt (CV, CM, CT or CS). While the Model 9840 is being calibrated using  
the remote commands, the front panel will show “CAL STARTED”. Pressing the front  
panel Esc button will cancel the calibration and issue a warning to the RS232 terminal.  
If any remote commands other than calibration commands are issued, the calibration  
will also be canceled. The acknowledgement will indicate if this is a “New Cell” or if  
you are going to “Overwrite” existing data. If you do not wish to overwrite use the CE  
MODEL 9840  
PG 35 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
command to stop this calibration. Another method of overwriting existing data is to  
delete the sensor first and then calibrate. The CB1 Command has been modified to  
accept the cell type but is still usable by older software.  
The format for the CB commands is:  
For backward compatibility:  
CB1(space)(channel)(serial number)#  
Alternatively, to delineate between a Load and a Torque Cell:  
CB1(cell type)(channel)(serial number)#  
CB2(space)(month)(day)(year)  
CB3(space) (excitation)(calibration unit)  
CB4(space)(rated load)#  
where channel is A or B, cell type is 0 for load and 1 for torque, serial number is the  
serial number of the cell, not exceeding 6 digits, month, day, and year are 2 digit  
numbers to represent the current date, excitation should be a 0 for 5.00 volts or a 1 for  
10.00 volts, and calibration unit is the desired unit for this calibration, and rated load is  
the maximum load or torque rating for the cell.  
Example 1:  
@123CB1 A123456#  
Acknowledge 1:  
@123 Calibrate Begin 1 Command – New  
Load Cell S/N: 123456 – Channel A  
Example 1A:  
@123CB11A123456#  
Acknowledge 1:  
@123 Calibrate Begin 1 Command – New  
Torque Cell S/N: 123456 – Channel A  
Example 2:  
@123CB2 042298  
Acknowledge 2:  
@123 Calibrate Begin 2 Command – New  
Cal Date: Apr22-98  
Example 3:  
@123CB3 100  
Acknowledge 3:  
@123 Calibrate Begin 3 Command – New  
Excitation Voltage: 10.0 V, Calibration Unit: Lb  
Example 4:  
@123CB4 1000.0#  
Acknowledge 4:  
@123 Calibrate Begin 4 Command – New  
Rated Load: 1000.0 Lb  
MODEL 9840  
PG 36 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
Calibration Escape Command (CE)  
The Calibrate Escape (CE) command is used to cancel a calibration that was started  
with the Calibrate Begin (CB) command. This format is CE.  
Example:  
@123CE  
Acknowledge:  
@123 Calibrate Command - Canceled, Calibration NOT Changed  
Calibrate by Millivolt-Per-Volt Command for One Point (CV)  
The Calibrate by milli-volt per Volt (CV) command is used to calibrate a load cell using  
its millivolt per volt calibration constant. You must have already started a calibration  
using the CB command for this command to function. The format is CV(millivolt per  
volt constant)# Note that a shunt check calibration will also be run at this time and so it  
is important that the load cell is not under load when this command is entered. Note  
that this command works for torque cell calibration as well.  
Example:  
@123CV4.5002#  
Acknowledge:  
@123 Calibrate Command - Reading for Shunt Check...  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 1000.0 Lb  
10.00 V , Cal on Oct27-99, 553.26 Lb Shunt  
unused S/N 63220, 100.00 Lb , 2.99984 mV/v,  
, 4.50020 mV/v,  
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb  
, 4.50015 mV/v,  
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt  
Calibrate by Millivolt-Per-Volt Command for Six Points (CMV6)  
The Calibrate by milli-volt per Volt for six points (CMV6) command is used to calibrate  
a load cell using 6 points of mass and milli-volt per volt each. You must have already  
started a calibration using the CB command for this command to function. The format  
is CMV6. Note that a shunt check calibration will also be run at the end of the  
calibration and so it is important that the load cell is set at 0 (or no masses) when the  
final command is entered.  
Example:  
@123CMV6  
Acknowledge:  
@123 Calibrate by mV/Volt - 6 Point  
Ready for Mass CMVM1 command  
Calibrate by mV/V Mass and Volt Commands (CMVM and CMVV)  
The Calibration by mV/V Mass (CMVM) and Volt (CMVV) commands can have up to  
seven parts each: CMVM1 through CMVM6 and CMVM0; and CMVV1 through CMVV6  
and CMVV0. You must have already started a calibration using the CB and CMV6  
commands for any of these commands to function. To input the first mass issue the  
MODEL 9840  
PG 37 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Calibration by mV/V Mass 1 (CMVM1) command. The mass is entered using the  
following format: CMVM1(mass value)#. Entering this command will result in the unit  
prompting the user for the first milli-volt per volt command.  
Example:  
@123CMVM15.0005#  
Acknowledge:  
@123 Calibrate Mass 1 Command entered  
Ready for mV/V Value CMVV1 or CE command  
The CE or Calibration Escape command will, if entered, end this calibration without  
calibrating the unit. The unit will remain calibrated according to the most recent  
calibration or, if this is the first calibration of the unit, the unit will remain uncalibrated.  
At this point the unit is prompting the user for the first milli-volt per volt value of the  
calibration. The mV/V value is entered using the Calibration by mV/V Volt Command  
with the following format: CMVV1(mV/V value)#.  
Example:  
@123CMVV11.9926#  
Acknowledge:  
@123 Calibrate mV/V 1 Command entered  
Ready for Mass Value CMVM2 or CE command  
The unit will continue to prompt the user for mass and milli-volt per volt values until the  
six points have been entered. Then the user will see the following acknowledgement:  
@123 Calibrate mV/V 6 Command entered  
Ready for Mass Value CMVM0 or CE command  
The user must now enter the CMVM0 command to finish the calibration. Note the load  
cell should have no masses on it at this point since a shunt calibration is run.  
Example:  
@123CMVM0  
Acknowledge:  
@123 Calibrate Command - Reading for Shunt Check…  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 1000.0 Lb  
, 4.50020 mV/v,  
4.50020 mV/v,  
4.50020 mV/V,  
4.50020 mV/V  
4.50020 mV/V  
5.00 V , Cal on Oct27-99,  
553.26 Lb Shunt  
Calibrate by mV/V Torque and Volt Commands (CMVT and CMVV)  
The Calibration by mV/V Torque (CMVT) and Volt (CMVV) works exactly like the  
calibrate by mV/V Mass and Volt commands. These commands can have up to seven  
parts each: CMVT1 through CMVT6 and CMVT0; and CMVV1 through CMVV6 and  
CMVV0. You must have already started a calibration using the CB and CMV6  
commands for any of these commands to function. To input the first torque issue the  
MODEL 9840  
PG 38 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Calibration by mV/V Torque 1 (CMVT1) command. The torque is entered using the  
following format: CMVT1(torque value)#. Entering this command will result in the unit  
prompting the user for the first milli-volt per volt command.  
Example:  
@123CMVT15.0005#  
Acknowledge:  
@123 Calibrate Torque 1 Command entered  
Ready for mV/V Value CMVV1 or CE command  
The CE or Calibration Escape command will, if entered, end this calibration without  
calibrating the unit. The unit will remain calibrated according to the most recent  
calibration or, if this is the first calibration of the unit, the unit will remain uncalibrated.  
At this point the unit is prompting the user for the first milli-volt per volt value of the  
calibration. The mV/V value is entered using the Calibration by mV/V Volt Command  
with the following format: CMVV1(mV/V value)#.  
Example:  
@123CMVV11.9926#  
Acknowledge:  
@123 Calibrate mV/V 1 Command entered  
Ready for Torque Value CMVT2 or CE command  
The unit will continue to prompt the user for mass and milli-volt per volt values until the  
six points have been entered. Then the user will see the following acknowledgement:  
@123 Calibrate mV/V 6 Command entered  
Ready for Torque Value CMVT0 or CE command  
The user must now enter the CMVT0 command to finish the calibration. Note the  
torque cell should have no torque on it at this point since a shunt calibration is run.  
Example:  
@123CMVT0  
Acknowledge:  
@123 Calibrate Command - Reading for Shunt Check…  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 1000.0 LbI  
, 4.50020 mV/v,  
4.50020 mV/v,  
4.50020 mV/V,  
4.50020 mV/V  
4.50020 mV/V  
5.00 V , Cal on Oct27-99,  
553.26 LbI Shunt  
Calibrate by Masses Command (CM)  
The Calibrate by Masses (CM) command tells the unit that you are doing a calibration  
by 2 point mass or by 5 point mass. The format is CM(number of masses).  
Example:  
@123CM5  
MODEL 9840  
PG 39 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Acknowledge:  
@123 Calibrate by Mass – 5 Point  
Ready for CMP1 command  
Calibrate by Masses Point Command (CMP)  
The Calibration by Masses Point (CMP) command can have up to six parts: CMP1  
through CMP5 and CMP0. You must have already started a calibration using the CB  
and CM commands for any of these commands to function. To read the first mass  
issue the Calibration Mass Point 1 (CMP1) command. The format is CMP1(mass)#.  
This will cause the Model 9840 to read 10 seconds of data from the load cell. These  
readings are averaged to obtain the first calibration mass reading. The example uses  
0.0 Lb for the first mass. Be sure you have placed the mass on the load cell before  
entering this command since reading will begin immediately.  
Example:  
@123CMP10.0#  
Acknowledge:  
@123 Calibrate Mass 1 Command - Reading...  
[10 seconds delay]  
Calibrate Mass 1 Command - Ready for CMP2 or CE command  
The Calibrate Mass 2 (CMP2) command has the format CMP2(mass)#. This command  
can only be issued after the CMP1 command has completed. The example uses 500  
Lb for the second calibration mass. Be sure you have placed the mass on the load cell  
before entering this command since reading will begin immediately  
Example:  
@123CMP2500#  
Acknowledge:  
@123 Calibrate Mass 2 Command - Reading...  
[10 seconds delay]  
Calibrate Mass 2 Command - Ready for CMP3 or CE command  
Masses three through five are done in a similar manner. The Calibrate Mass Point 0  
(CMP0) command completes the 2 or 5-point mass calibration. Remove the masses  
from the load cell before entering this command since it will perform the shunt  
calibration measurement that is recorded for use with the calibration check command.  
The format is CMP0.  
Example:  
@123CMP0  
Acknowledge:  
@123 Calibrate Command - Reading for Shunt Check...  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 100.00 Lb  
, 3.00150 mV/v,  
3.00230 mV/v,  
3.00238 mV/v,  
3.00353 mV/v,  
MODEL 9840  
PG 40 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt  
unused S/N 13368, 10.00 Lb  
, 3.00121 mV/v,  
10.00 V,  
Cal on Oct27-99, 4.9292 Lb Shunt  
unused S/N 63220, 100.00 Lb  
, 2.99984 mV/v,  
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb  
, 4.50015 mV/v,  
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt  
Calibrate by Torque Command (CT)  
The Calibrate by Torque (CT) command tells the unit that you are doing a calibration by  
2 point torque or by 5 point torque. The format is CT(number of torques).  
Example:  
@123CT5  
Acknowledge:  
@123 Calibrate by Torque – 5 Point  
Ready for CTP1 command  
Calibrate by Torque Point Command (CTP)  
The Calibration by Torque Point (CTP) command can have up to six parts: CTP1  
through CTP5 and CTP0. You must have already started a calibration using the CB  
and CT commands for any of these commands to function. To read the first torque,  
issue the Calibration Torque Point 1 (CTP1) command. The format is CTP1(torque)#.  
This will cause the Model 9840 to read 10 seconds of data from the torque cell. These  
readings are averaged to obtain the first calibration torque reading. The example uses  
0.0 LbI for the first mass. Be sure you have placed the torque on the cell before  
entering this command since reading will begin immediately.  
Example:  
@123CTP10.0#  
Acknowledge:  
@123 Calibrate Torque 1 Command - Reading...  
[10 seconds delay]  
Calibrate Torque 1 Command - Ready for CTP2 or CE command  
The Calibrate Torque 2 (CTP2) command has the format CTP2(torque)#. This  
command can only be issued after the CTP1 command has completed. The example  
uses 500 LbI for the second calibration torque. Be sure you have placed the torque on  
the cell before entering this command since reading will begin immediately.  
Example:  
@123CTP2500#  
Acknowledge:  
@123 Calibrate Torque 2 Command - Reading...  
[10 seconds delay]  
Calibrate Torque 2 Command - Ready for CTP3 or CE command  
Torques three through five are done in a similar manner. The Calibrate Torque Point 0  
(CTP0) command completes the 2 or 5-point torque calibration. Remove the torque  
MODEL 9840  
PG 41 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
from the cell before entering this command since it will perform the shunt calibration  
measurement that is recorded for use with the calibration check command. The format  
is CTP0.  
Example:  
@123CTP0  
Acknowledge:  
@123 Calibrate Command - Reading for Shunt Check...  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 100.00 LbI  
, 3.00150 mV/v,  
3.00230 mV/v,  
3.00238 mV/v,  
3.00353 mV/v,  
10.00 V , Cal on Oct27-99, 49.532 LbI Shunt  
unused S/N 13368, 10.00 Lb  
, 3.00121 mV/v,  
10.00 V,  
Cal on Oct27-99, 4.9292 Lb Shunt  
unused S/N 63220, 100.00 Lb  
, 2.99984 mV/v,  
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb  
, 4.50015 mV/v,  
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt  
Calibrate by Shunt Command (CS)  
The Calibrate by Shunt (CS) command is used to calibrate the cell using the shunt  
calibration factor expressed in pounds (or pound-inches for a torque cell). This  
command will only function if you have already issued a Calibrate Begin (CB)  
command. Note that the cell should not have any load or torque on it for this command  
to operate correctly. The format is CS(shunt value in Lb or LbI)#. The example shows  
calibration of a load cell using a shunt value of 553.26 Lb.  
Example:  
@123CS553.26#  
Acknowledge:  
@123 Calibrate Shunt Command - Reading...  
[10 seconds delay]  
@123 Calibrate Command - Reading for Shunt Check...  
[10 seconds delay]  
@123 Calibrate Command Completed  
Ch A = S/N 123456, 1000.0 Lb  
, 4.50014 mV/v,  
10.00 V , Cal on Oct27-99, 553.26 Lb Shunt  
unused S/N 63220, 100.00 Lb  
, 2.99984 mV/v,  
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt  
Ch B = S/N 89991, 1000.0 Lb  
, 4.50015 mV/v,  
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt  
Calibrate Counts-Per-Inch Command (CI)  
This command has two subcommands, a view command (CIV) and a set command  
(CIS).  
MODEL 9840  
PG 42 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
The Calibration counts per Inch View (CIV) command displays the current setting of the  
encoder counts per inch constant.  
Example:  
@123CIV  
Acknowledge:  
@123 Encoder is 32000 counts per inch  
The Calibration counts per Inch Set (CIS) command allows you to change the current  
setting of the encoder counts per inch constant.  
Example:  
@123CIS64000#  
Acknowledge:  
@123 Encoder is 64000 counts per inch  
Limits Command (L)  
This command allows you to view or set the 4 contact closure limit switches on the Model  
9840. There are three subcommands Limit View (LiV), Limit Set (LiS), and Limit Reset  
(LiR). Note that changing a cell type on a channel will automatically erase limits tat were  
set for that channel. They must then be reset by the user.  
Limit View Command (LiV)  
The Limit View (LiV) command is used to view the current settings of one of the limits.  
The format is L(limit number)V.  
Example:  
@123L1V  
Acknowledge:  
@123 Lim 1 NO Enabled Load A Lb Set 20.00 Trip>Set Latch Off  
Reset 1.000  
This line indicates that Limit 1 is a normally open switch (NO) rather than a normally  
closed switch (NC). The limit is Enabled rather than Disabled. The signal driving this  
limit is Load in Lb. The limit will be turned on when the load exceeds 20.00 Lb. It is  
not a Latching limit (latching requires a manual reset). The limit will automatically reset  
when the load is less than 1.00 Lb. Torque can be substituted for load for an  
appropriate cell.  
Limit Set Command (LiS)  
The Limit Set (LiS) command is broken up into 4 separate commands. They have the  
following format:  
L(limit number)SA(space) (normal position)(enable)(item number)(unit number)  
L(limit number)SB(space) (setpoint number)#  
L(limit number)SC(space) (< or >)(latching)  
L(limit number)SD(space) (reset point number)#  
MODEL 9840  
PG 43 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
The limit number chooses which limit (1-4) you want to setup. The normal position  
should be set to 0 for normally open or 1 for normally closed. The enable should be  
set to 0 for disabled or 1 for enabled. If the limit is disabled the rest of this command  
may be omitted (see example 3). The item and unit numbers indicate which signal is to  
drive this limit. The set point number indicates the signal level at which the limit is to  
be activated. The < or > sign indicates whether you want the limit to be activated when  
the signal is greater than the set point (>) or less than the set point (<). The latching  
should be set to 0 for latching off or 1 for latching on. If latching on is selected the rest  
of the command may be omitted (see example 2). Latching on will require a manual  
reset once the limit has been activated. The reset point indicates the signal level at  
which the limit should be de-activated.  
If you want to enable a limit without otherwise changing its settings you can enter  
L(limit number)SA(space)(normal position)1# instead of entering the full setup string  
(see example 4).  
Example 1:  
@123L1SA 010000  
Acknowledge 1:  
@123 Limit Setup Command A – Ready for Command B  
Example 2:  
@123L1SB 50.0#  
Acknowledge 1:  
@123 Limit Setup Command B – Ready for Command C  
Example 3:  
@123L1SC >0  
Acknowledge 3:  
@123 Limit Setup Command C – Ready for Command D  
Example 4:  
@123L1SD 10.0#  
Acknowledge 4:  
@123 Lim 1 NO Enabled Load A Lb Set 50.00 Trip>Set Latch Off  
Reset 10.00  
Example 5:  
@123L1SA 10  
Acknowledge 5:  
@123 Lim 1 NO Disabled Load A Lb Set 50.00 Trip>Set Latch Off  
Reset 10.00  
Example 6:  
@123L1SA 11#  
Acknowledge 6:  
@123 Lim 1 NC Enabled Load A Lb Set 50.00 Trip>Set Latch Off  
Reset 10.00  
Limit Setup Escape (LE)  
The Limit Escape (LE) command is used to cancel a limit setup that was started with  
the Limit Set (LiS) command. This format is LE.  
Example:  
MODEL 9840  
PG 44 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
@123LE  
Acknowledge:  
@123 Limit Setup Command Canceled  
Limit Reset Command (LiR)  
The Limit Reset (LiR) command is used to manually reset a latching limit that has been  
activated. The format is L(limit number)R.  
Example:  
@123L1R  
Acknowledge:  
@123 Reset Limit 1  
System Options Command (O)  
The system Options command has 8 subcommands that are used to view or set the  
options that appear on the system options menu. Option View (OV) displays all the  
settings in the system options menu, Option Printer (OP) is used to set the printer  
baudrate, Option Auto ID (OI) is used to turn auto-identification of load cells on or off,  
Option Auto Zero (OZ) is used to turn auto-zeroing of load on or off, Option Com Address  
(OA) is used to change the communications address of the unit, Option Baudrate (OB) is  
used to change the communications baudrate, Option Linefeed (OL) is used to turn the  
auto line feed option on or off, and OE which is used to turn the end of transmission  
character on or off. Each of these options is described below.  
Option View Command (OV)  
The Option View (OV) command returns a list of the current settings of all the system  
options. The format is OV.  
Example:  
@123OV  
Acknowledge:  
@123 Printer Baud Rate is 9600  
Auto Identify is off  
Auto Zero Channel A is off  
Auto Zero Channel B is off  
Com Address is 123  
Com Baud Rate is 9600  
Com Line Feed is off  
Retain Tare is off  
Freeze Display is off  
RS232 End of Transmission Character is off  
Option Printer Baudrate Command (OP)  
The Option Printer baudrate (OP) command is used to select the printer baudrate. The  
format is OP(code) where the code is from the table shown at the right.  
MODEL 9840  
PG 45 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Example:  
Option Printer  
Command  
@123OP9  
Acknowledge:  
@123 Printer Baud Rate is 230.4K  
Code  
Baudrate  
4
5
6
8
9
4800  
9600  
Option Auto-Identify Command (OI)  
The Option auto-Identify (OI) command is used to turn  
auto-identification of load cells on or off. The format is  
OI(0 or 1), where 0 turns auto-id off, and 1 turns auto-  
id on.  
19.2K  
57.6K  
230.4K  
Example:  
@123OI0  
Acknowledge:  
@123 Auto Identify is off  
Option Auto-Zeroing Command (OZ)  
The Option auto-Zero (OZ) command is used to turn auto-zeroing of load on or off.  
The format is OZ(channel)(0 or 1) where channel is A or B and 0 & 1 turn auto-zeroing  
off and on, respectively.  
Example:  
@123OZA1  
Acknowledge:  
@123 Auto Zero Channel A is on  
Auto Zero Channel B is off  
Option Com Address Command (OA)  
The Option com Address (OA) command is used to change the communications  
address assigned to this unit. The format is OA(new address)#. Note that valid  
addresses are 1 through 254.  
Example:  
@123OA003#  
Acknowledge:  
@123 Com Address is 003  
MODEL 9840  
PG 46 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Option Com Baudrate Command (OB)  
Option Baudrate Command  
The Option com Baudrate (OB) command is used to  
change the communications baudrate for this unit.  
The format is OB(code), where the baudrate codes  
are given in the table. Code 7 is a special baudrate  
that is reserved for future expansion. Note that the  
acknowledgement is given at the current baudrate  
and then the baudrate is changed.  
Code  
Baudrate  
300  
0
1
2
3
4
5
6
7
600  
1200  
2400  
Example:  
@123OB2#  
4800  
Acknowledge:  
@123 Changing Communications Baudrate  
to 1200  
9600  
19.2K  
Special  
Option Linefeed Command (OL)  
The Option com Linefeed (OL) command turns on or off the generation of linefeed  
characters when the Model 9840 sends information to your RS232 terminal. The  
format is OL(0 or 1), where 0 turns linefeeds off, and 1 turns them on.  
Example:  
@123OL1  
Acknowledge:  
@123 Com Linefeed is on  
Option Tare Retain Command (OT)  
The Retain Tare (OT) command enables and disables the retention of the current tare  
value in memory for use after the unit has been powered down and then powered back  
up. The format for the OT command is: OT(0 or 1), where 0 is disable and 1 is enable.  
Once enabled, the tare values are retained until this option is disabled. Disabling the  
tare retention option resets the tare values for both channels to 0.  
Example:  
@123OT1  
Acknowledge:  
@123 Retain tare is on  
Option End of Transmission Character (OE)  
The end of transmission character is a non-visible character (hexadecimal 0x04) that is  
sent at the end of each response by the unit to an RS232 command. The character is  
sent after the last carriage-return line-feed returned by the unit with one exception:  
when the user is viewing channel values with the @123V(item)(unit)(repeat) and sets  
the value of repeat to 2 (streaming output), the character is sent immediately prior to  
the streaming output that the unit will send. The OE command enables or disables  
this character and its format is: OE(0 or 1), where 0 is disable and 1 is enable.  
Example:  
@123OE1  
Acknowledge:  
MODEL 9840  
PG 47 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
@123 RS232 EOT is on.  
Display Options Command (D)  
The Display options command has 5 subcommands used to set options that effect the  
front panel display of the Model 9840.  
Display View Command (DV)  
The Display View (DV) command is used to return the current values of all the display  
options. The format is DV.  
Example:  
@123DV  
Acknowledge:  
@123 Filter is Type II Level 2  
Filter Window A is off (if on, the window value is shown)  
Filter Window B is off (if on, the window value is shown)  
Channel A shows 4 decimal digits  
Channel B shows 4 decimal digits  
Channel A counts by 1  
Channel B counts by 1  
Second line shows limit status  
(if second line is Display, the text is shown here)  
Display Filter Command (DF)  
The Display Filter (DF) command sets the filter type and level. The format is  
DF(type)(level), where type is either 1 for type I filters, or 2 for type II filters, level must  
be 1, 2, 3, or 4 depending on the level of filtering desired. (4 is most filtering and  
slowest rise time).  
Example:  
@123DF13  
Acknowledge:  
@123 Filter is Type I Level 3  
Display Decimal Command (DD)  
The Display Decimal (DD) command sets the maximum number of decimal digits that  
are shown for Load, Peak, and Valley. The format is DD(channel)(number), where  
channel is A or B and the number must be between zero and 5.  
Example:  
@123DDA0  
Acknowledge:  
@123 Channel A shows 0 decimal digits  
MODEL 9840  
PG 48 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Display Count By Command (DC)  
The Display Count by (DC) command sets the  
smallest number that the display will count by  
for the Load or Torq, Peak, and Valley  
measurements. The format is  
DC(channel)(code) where the code is selected  
from the table at the right.  
Display Count by Command  
Code  
Count By  
0
1
2
3
4
1
2
Example:  
@123DCA3  
Acknowledge:  
@123 Channel A counts by 10  
5
10  
20  
Display Second Line Command (D2)  
The Display Second Line (D2) command is used to set the second line to be blank, or  
to show the limit status. The format is D2(B, L, D, or T), where B makes the line blank,  
L makes the line show limit status, D makes the line show another display, and T  
makes the line show text.  
Example:  
@123D2B  
Acknowledge:  
@123 Second Line set to Blank  
Display Second Line Text (DT)  
The Display Second Line Text (DT) command is used to select the text that would be  
displayed on the second line if text was set to go on the second line.  
Example:  
@123DTModel 9840  
Acknowledge:  
@123 Text Message – Model 9840  
Display Filter Window Commands (DW)  
The Display Filter Window (DW) command set enables and disables the filter window  
and allows the user to enter the unit and value of the filter window to use. There are  
two commands used to enable and set the filter windows. The format to enable or  
disable the Filter Window is DW1(channel)(0 or 1) where channel is A or B, 1 enables  
the window and 0 disables the window. The format to set the window unit and value is  
DW2(channel)(window unit)(window value)# where window unit is the desired unit for  
the filtering window, and window value is the desired value of the window. The units  
available will depend on the sensor that is currently active for the channel. It is a good  
idea to issue the @123? command and find out what the active sensor is on the  
channel with the @123SA or @123SB commands to make sure that the appropriate  
unit will be entered. The window must be enabled before a value can be entered.  
MODEL 9840  
PG 49 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Disabling the window sets this value to 0. The window, once enabled and set, will take  
effect immediately and will remain in effect until the window is disabled.  
Note that changing a cell type on a channel will automatically disable the filter window  
for that channel, reset the units to Lb or LbI, and set the window value to 0. The  
window must then be reset and re-enabled by the user.  
Example:  
@123DW1A1  
Acknowledge:  
@123 Filter Window A is On  
Example:  
@123DW2A000.0002#  
Acknowledge:  
@123 Filter Window A Unit = Lb  
@123 Filter Window A = 0.0002 Lb  
MODEL 9840  
PG 50 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
SYSTEM CALIBRATION MODE  
The following procedure should be followed to calibrate the Model 9840 itself. This  
calibration requires the use of a precision millivolt per volt reference which will be attached  
on the load cell connector. The resulting scale factors are recorded in the non-volatile  
memory of the Model 9840.  
ATTENTION: System Calibration should ONLY be performed by trained, QUALIFIED  
TECHNICIANS utilizing the proper precision devices to ensure the most accurate  
performance of the unit.  
1. To enter system calibration mode, turn off the Model 9840. Hold down the two end  
buttons ( “<“ and “>“ ) while turning power back on. As soon as the display comes on  
you can release the buttons.  
2. The unit will go through its regular start up sequence but rather than entering run  
mode and showing the load in pounds a password prompt will be displayed. Enter the  
number 777 as the system calibration password.  
3. The unit will now display “Unit mV / V Cal” to indicate that it is ready to be calibrated.  
A five minute warm-up period with the millivolt-per-volt reference attached is  
recommended at this time if the unit has just been turned on or the millivolt-per-volt  
reference has just been attached. Press any button to clear this display.  
4. You are now prompted to enter the first of two millivolt per volt references. (We  
recommend using -3.0 mv/volt for the first and +3.0 mv/volt for the second.) Set your  
reference and enter the appropriate number. As soon as you press Enter the Model  
9840 will start reading the reference.  
5. Next you are prompted for the second mv/volt value. Set your reference and enter the  
appropriate number. Again reading will start as soon as Enter is pressed.  
6. When the read is complete, the unit will display “Calibration Done”. Pressing any  
button at this time will place the unit back into its normal run mode.  
MODEL 9840  
PG 51 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
APPENDIX A -- MODEL 9840 SPECIFICATIONS  
Transducer Interface  
Excitation  
5VDC or 10 VDC software/auto selectable  
Current Drive  
180 mA at 5VDC or 10 VDC  
Yes  
Push Button Shunt  
Internal Shunt Resistor  
Calibration Method  
Push Button Tare  
Sensitivity Adjust  
Accuracy  
1 standard, 2 optional. Rear panel selectable  
Shunt, mV/V, Known Load  
100% of range--display and analog output  
1mV/V to 4.5 mV/V  
0.01% of full scale +/- 1 digital count  
115 dB  
CMRR  
Additional Transducer Channel  
Auxiliary Channel  
optional  
D/A Analog Output  
Full Scale Output  
Linearity  
+/- 10 VDC nominal +/- 0.5 VDC  
0.02 % of full scale  
Scaling  
Gain and offset selected by software.  
Approx. 15 Hz (plus filtering)  
Any displayed quantity  
Frequency Response  
Output Source  
Output Channels  
1 std, 2 optional  
Direct Analog Output  
Full Scale Output  
Linearity  
-/+ 8.14 VDC +/- 0.25 VDC  
0.02 % of full scale  
Scaling  
Fixed with no tare, transducer channel only  
Approx. 1.5 kHz  
Frequency Response  
Peak/Hold  
Peak/Valley/Hold  
Digital 1/60 second capture  
MODEL 9840  
PG 52 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Digital I/O  
Digital Inputs  
Interface  
4 optically isolated inputs  
Function  
remote tare, peak/valley reset, position reset, print  
Digital Outputs  
Function  
4 independent limits  
Quad-Limits Setup  
Front panel, latching or non-latching set points  
normally open/closed, enable/disable  
Quad-Limits Output  
Quad-Limits Source  
4 isolated solid state switches  
Any displayed quantity  
Serial Interface  
RS-232 Serial Setup/Output  
RS-485 Serial Setup/Output  
Multi-drop RS-232  
USB  
Standard  
Optional  
Optional  
Optional  
Optional  
Printing  
Interface Type  
Quantities Printed  
Serial  
All displayed items  
Optional  
Encoder Position Channel  
Encoder Type  
Excitation  
Quadrature, 4X number of lines  
12 VDC std., 5 VDC optional  
Single ended or differential  
User adjustable counts/inch  
Interface  
Calibration  
Auto Transducer Identification  
Auto Identification and Setup  
Yes, standard  
Push Buttons  
Smart Push Buttons  
Adjustments  
6
Displayed item, displayed units, tare, peak/valley  
reset alternate view, print. Alternate functions for  
programming.  
MODEL 9840  
PG 53 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
Digital Display  
Display Characters  
Display Update  
Scaling  
2 lines by 20 characters  
> 4/Sec  
Automatic or manual  
Maximum Display Count  
Decimal Points. Selection  
Display Type  
± 999,999  
0 to 5 software selectable.  
Vacuum fluorescent  
24 bit  
Resolution  
Software Filters  
8, software selectable  
Electrical Characteristics  
Power Requirements  
110 VAC, 60 Hz, Optional 220VAC, 50 Hz  
12 W (unloaded)  
Power Consumption  
Transducer Excitation Current  
Absolute Max Transducer Input  
Shutdown mode when shorted to ground  
30 VDC without damage  
Physical Characteristics  
Package Size  
7.29” X 2.52 “ X 9.0” (W X H X D)  
Case Material  
Aluminum  
Available  
Mounting Hardware  
Environmental  
Operating Temp  
Storage Temp  
0 C to 50 C  
-10 C to 60 C  
Relative Humidity  
95 % without condensation  
The Model 9840 is  
compliant.  
MODEL 9840  
PG 54 PUB. 2856-16  
INTERFACE MODEL 9840 VER 6.0.2  
APPENDIX B -- CABLES AND CONNECTORS  
Load Cell Connectors  
Notes:  
DB9 Female  
Pin Signal  
1. If the sense lines are not used, SENSE - HI must  
be tied to EXCITE - HI and SENSE - LO must be tied  
to EXCITE - LO.  
1
2
3
4
5
6
7
8
9
EXCITE - HI  
SENSE - HI1  
CELL OUTPUT - HI  
CELL OUTPUT - LO  
SENSE - LO 1  
EXCITE - LO  
AUTO ID - A  
AUTO ID - B  
2. Incorrect wiring of these ports can cause damage  
to the internal circuitry of the unit. Please contact  
Interface if you have questions or need assistance  
with these configurations.  
CHASSIS GND  
Serial Port Connector  
Notes:  
DB 15 Male  
Pin  
1
2
3
4
5
6
7
8
Signal  
TXD  
RXD  
1. Pins 1 through 4 are used for RS232 ASCII  
command set communications. Use 8 data bits, no  
parity, 1 stop, and set the baud rate using the Com  
baud rate entry on the System Options menu.  
No Connect  
SERIAL GND  
TXD+ (Printer)  
TXD- (Printer)  
RXD- (RS485)  
RXD+ (RS485)  
No Connect  
SERIAL GND  
2. The printer uses RS422 differential signaling on  
pins 5, 6 and 10. Baud rate for the printer is  
selectable using the Printer Baud entry on the  
System Options menu.  
9
10  
3. Pins 7, 8, 14 and 15 are used for RS485 ASCII  
command set communications. Use 8 data bits, no  
parity, 1 stop, and set the baud rate using the Com  
baud rate entry on the System Options menu.  
11-13 No Connect  
14  
15  
TXD+ (RS485)  
TXD- (RS485)  
MODEL 9840  
PG 55 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Digital I/O Connector  
Notes:  
High Density 26 Female  
1. Limits are optically isolated solid state switches  
that can control AC or DC voltage. Peak blocking  
voltage is 350V, maximum continuous current is 120  
mA.  
Pin  
1
Signal  
LIM 1 A1  
2
LIM 1 B1  
3
4
5
6
7
8
9
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
LIM 2 A  
LIM 2 B  
LIM 3 A  
LIM 3 B  
LIM 4 A  
LIM 4 B  
2. Digital inputs are opto-isolated with internal  
current limiting resistors. Voltage range is +4 to +22  
VDC on the “HI” input with respect to the “LO” input.  
Other voltage ranges are possible with added  
external resistors, contact Interface for details. If  
isolation is not required these inputs may be  
operated using the +12 VDC supplied on pin 26 and  
the GND on pin 24.  
ISO IN 1 - HI2  
ISO IN 1 - LO2  
ISO IN 2 - HI  
ISO IN 2 - LO  
ISO IN 3 - HI  
ISO IN 3 - LO  
ISO IN 4 - HI  
ISO IN 4 - LO  
KEYLOCK3  
GND  
3. The keylock input will be in the “unlocked” state if  
left unconnected. Tie this pin to GND to “lock” it.  
When locked the setup of the Model 9840 cannot be  
changed (you cannot enter Setup mode).  
4. The encoder inputs are intended for use with a  
+12 VDC quadrature encoder. For single ended  
signaling use the CHA and CHB pins, the CHA-L  
and CHB-L pins are provided for differential signals.  
The encoder can be supplied with the +12 VDC and  
GND available on pins 26 and 24.  
The Model 9840 will also accept +5 VDC encoder  
inputs but these MUST be differential. Single ended  
+5 VDC inputs are not supported on the standard  
unit. Options for +5 VDC encoders are available.  
Pin 26 will supply +5 VDC for this option and must  
be ordered accordingly.  
ENCODER CHA-L4  
ENCODER CHA  
GND  
ENCODER CHB-L  
ENCODER CHB  
GND  
GND  
+12 VDC  
(+5 VDC optional4)  
Chart Rec. Connector  
Notes:  
DB9 Male  
1. The source and scaling of ANALOG -  
OUT is selected through the menus.  
2. The HIGH BANDWIDTH - OUT is a  
buffered analog output permanently  
connected to the load cell channel.  
3. If the 4-20 mA Option is purchased.  
Pin Signal  
2
3
4
ANALOG - OUT 1  
ANALOG - GND  
HIGH BANDWIDTH - OUT 2  
CHANNEL A  
5
6
7
8
9
4-20 mA – OUT 3  
CHANNEL A  
2nd ANALOG-OUT1  
(Optional)  
4-20 mA – OUT 3  
CHANNEL B  
HIGH BANDWIDTH - OUT 2  
CHANNEL B  
No Connect  
MODEL 9840  
PG 56 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
USB Connector  
Pin  
Name  
VCC  
D -  
Description  
1
+5 VDC  
Data –  
Data +  
Ground  
2
3
D+  
4
GND  
MODEL 9840  
PG 57 PUB. 2856-16  
 
INTERFACE MODEL 9840 VER 6.0.2  
Precision Shunt Calibration Resistors  
This diagram shows the connection of the precision internal shunt resistors.  
The back panel switch (shown to the left of the two resistors) selects between 30 KΩ  
and 60 KΩ.  
The shunt check relay (shown to the right of the two resistors) connects the selected  
resistor between the Excite High and the Cell Output High.  
Sense High  
Excite High  
30 K  
60 K  
Ω
Ω
LOAD  
CELL  
Cell Output High  
Excite Low  
Sense Low  
Cell Output Low  
MODEL 9840  
PG 58 PUB. 2856-16  
 

Hitachi CML152XW User Manual
Gigabyte 8S648FX RZ User Manual
GE JTP70CM User Manual
GE 169214 User Manual
GE 27920 User Manual
Asus P8h61 M Lx3 R2 0 Desktop Motherboard P8H61MLX3R20 User Manual
Asus Computer Hardware P8H77 I User Manual
Asus A8N E User Manual
AG Neovo 17 LCD Display X 17 User Manual
Acer B223W User Manual