Agilent Technologies U2802A User Manual

Agilent U2802A  
31-Channel  
Thermocouple Input  
Device  
User’s Guide  
Agilent Technologies  
Safety Information  
The following general safety precautions must be observed during all phases of this instrument. Failure to comply  
with these precautions or with specific warnings elsewhere in this manual violates safety standards of design,  
manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the  
customer’s failure to comply with these requirements.  
Safety Symbols  
The following symbols indicate that precautions must be taken to maintain safe  
operation of the instrument.  
Direct current  
Alternating current  
Both direct and alternating current  
Three-phase alternating current  
Earth (ground) terminal  
Protective conductor terminal  
Frame or chassis terminal  
Equipotentiality  
On (Supply)  
U2802A 31-Channel Thermocouple Input User’s Guide  
III  
 
Off (Supply)  
Equipment protected throughout by double insulation or rein-  
forced insulation  
Caution, risk of electric shock  
Caution, hot surface  
Caution, risk of danger (See note.)  
In position of a bi-stable push control  
Out position of a bi-stable push control  
IV  
U2802A 31-Channel Thermocouple Input User’s Guide  
Regulatory Markings  
The CE mark is a registered trademark of the European  
community. This CE mark shows that the product complies with  
all the relevant European legal directive.  
ICES/NMB-001 indicates that this ISM device complies  
with Canadian ICES-001.  
The CSA mark is a registered trademark of the Canadian Stan-  
dards Association. A CSA mark with the indicators "C" and "US"  
means that the product is certified for both the U.S. and Canadian  
markets, to the applicable American and Canadian standards.  
The UL Mark is a registered trademark of Underwriters Laborato-  
ries Inc. UL listing mark with the indicators "C" and "US”  
indicates the product compliance with both Canadian and U.S.  
requirements.  
The C-tick mark is a registered trademark of the Spectrum Man-  
agement Agency of Australia. This signifies compliance with the  
Australian EMC Framework regulations under the terms of the  
Radio Communications Act of 1992.  
This product complies with the (2002/96/EC) marking  
equipment. The affixed product label indicates that you must not  
discard this electrical/electronic product in domestic household  
waste.  
U2802A 31-Channel Thermocouple Input User’s Guide  
V
General Safety Information  
Do not use the device if it is damaged. Before you use the device,  
inspect the case. Look for cracks or missing plastic. Do not operate the  
device around explosive gas, vapor or dust.  
WARNING  
Do not apply more than the rated voltage (as marked on the device)  
between terminals, or between terminal and external ground.  
Always use the device with the cables provided.  
Observe all markings on the device before connecting to the device.  
Turn off the device and application system power before connecting to  
the I/O terminals.  
When servicing the device, use only specified replacement parts.  
Do not operate the device with the removable cover removed or  
loosened.  
Do not connect any cables and terminal block prior to performing  
self-test process.  
Use only the power adapter supplied by the manufacturer to avoid any  
unexpected hazards.  
Do not load the input and output terminals above the specified  
operating limits. Input terminals should not exceed 10 V with respect  
to the module ground. Applying excessive voltage or overloading the  
device will cause irreversible damage to the circuitry.  
CAUTION  
Applying excessive voltage or overloading the input terminal will  
damage the device permanently.  
If the device is used in a manner not specified by the manufacturer, the  
protection provided by the device may be impaired.  
The U2802A can only be used with U2355A or U2356A DAQs and used  
with the SCSI cables provided.  
Always use dry cloth to clean the device. Do not use ethyl alcohol or  
any other volatile liquid to clean the device.  
Do not permit any blockage of the ventilation holes of the device.  
VI  
U2802A 31-Channel Thermocouple Input User’s Guide  
Waste Electrical and Electronic Equipment (WEEE) Directive  
2002/96/EC  
This instrument complies with the WEEE Directive (2002/96/EC) marking  
requirement. This affixed product label indicates that you must not discard this  
electrical/electronic product in domestic household waste.  
Product Category:  
With reference to the equipment types in the WEEE directive Annex 1, this  
instrument is classified as a “Monitoring and Control Instrument” product.  
The affixed product label is shown as below:  
Do not dispose in domestic household waste  
To return this unwanted instrument, contact your nearest Agilent office, or visit:  
for more information.  
U2802A 31-Channel Thermocouple Input User’s Guide  
VII  
Environmental Conditions  
The table below shows the general environmental requirements for the product.  
Environmental Conditions Requirements  
Temperature  
Humidity  
Operating temperature from 0 °C to +55 °C  
Relative humidity at 50% to 85% RH (Non-condensing)  
Altitude up to 2000 meters  
Altitude  
Storage compliance  
–40 °C to +70 °C  
VIII  
U2802A 31-Channel Thermocouple Input User’s Guide  
 
In This Guide...  
1
2
Getting Started  
This chapter introduces the new Agilent U2802A 31- channel  
thermocouple input device and provides quick start information. It  
also provides product outlook, installation configuration and  
troubleshooting guide.  
Features and Functions  
This chapter contains details of the product features, applications,  
system overview and theory of operation. From this chapter, you  
will understand the Agilent U2802A 31- channel thermocouple  
input system overview and functionality of this device.  
3
4
This chapter described the Agilent U2802A 31- channel  
thermocouple input device pin configurations and connector pinout  
for user’s reference.  
Product Specifications  
This chapter specifies the environmental conditions,  
characteristics, and specifications of the Agilent U2802A  
31- channel thermocouple input device. It also covers the system  
accuracy, typical performance and guidelines to make accurate  
temperature measurements.  
5
Calibration  
This chapter contains the calibration information and factory  
restore calibration procedure for the Agilent U2802A 31-channel  
thermocouple input device.  
U2802A 31-Channel Thermocouple Input User’s Guide  
IX  
 
                                                                                               
                                                                                                 
                                                                                                  
                                                                                                    
                                                                                                      
                                                                                                        
                                                                                                          
                                                                                                            
                                                                                                                                                                                                                           
                                                                                                                                                                                                                             
                                                                                                                                                                                                                              
                                                                                                                                                                                                                                 
                                                                                                                                                                                                                                  
DECLARATION OF CONFORMITY  
According to EN ISO/IEC 17050-1:2004  
Generic example  
Manufacturer’s Name:  
Manufacturer’s Address:  
Agilent Technologies Microwave Products (M) Sdn. Bhd  
Bayan Lepas Free Industrial Zone,  
11900, Bayan Lepas, Penang, Malaysia  
Declares under sole responsibility that the product as originally delivered:  
Product Name:  
Models Number:  
Product Options:  
31-Channel Thermocouple Input Device  
U2802A  
This declaration covers all options of the above product(s)  
complies with the essential requirements of the following applicable European Directives, and  
carries the CE marking accordingly:  
Low Voltage Directive (2006/95/EC)  
EMC Directive (2004/108/EC)  
and conforms with the following product standards:  
EMC  
Standard  
Limit  
IEC 61326:2002 / EN 61326:1997+A1:1998+A2:2001+A3:2003  
CISPR 11:1990 / EN55011:1990  
Class A Group 1  
4 kV CD, 8 kV AD  
3 V/m, 80-1000 MHz  
0.5 kV signal lines, 1 kV power lines  
0.5 kV line-line, 1 kV line-ground  
3 V, 0.15-80 MHz  
IEC 61000-4-2:1995 / EN 61000-4-2:1995  
IEC 61000-4-3:1995 / EN 61000-4-3:1996  
IEC 61000-4-4:1995 / EN 61000-4-4:1995  
IEC 61000-4-5:1995 / EN 61000-4-5:1995  
IEC 61000-4-6:1996 / EN 61000-4-6:1996  
IEC 61000-4-11:1994 / EN 61000-4-11:1994  
1 cycle / 100%  
Canada: ICES-001:2004  
Australia/New Zealand: AS/NZS CISPR11:2004  
The product was tested in a typical configuration with Agilent Technologies test systems.  
IEC 61010-1:2001 / EN 61010-1:2001  
Canada: CAN/CSA-C22.2 No. 61010-1-04  
USA: ANSI/UL 61010-1:2004  
Safety  
This DoC applies to above-listed products placed on the EU market after:  
20-November-2007  
Date  
Mack Soh  
Quality Manager  
For further information, please contact your local Agilent Technologies sales office, agent or distributor,  
or Agilent Technologies Deutschland GmbH, Herrenberger Straße 130, 71034 Böblingen, Germany.  
Template: A5971-5302-2, Rev. E.00  
U2802A  
DoC Revision 1.0  
X
U2802A 31-Channel Thermocouple Input User’s Guide  
Product Regulations  
EMC  
Performance Criteria  
IEC 61326-1:2002 / EN 61326-1:1997+A1:1998+A2:2001+A3:2003  
CISPR 11:1990 / EN 55011:1990 – Group 1 Class A  
IEC 61000-4-2:1995 / EN 61000-4-2:1995 (ESD 4kV CD, 8kV AD)  
IEC 61000-4-3:1995 / EN 61000-4-3:1996 (3V/m, 80% AM)  
IEC 61000-4-4:1995 / EN 61000-4-4:1995 (EFT 0.5kV line-line, 1kV line-earth)  
IEC 61000-4-5:1995 / EN 61000-4-5:1995 (Surge 0.5kV line-line, 1kV line-earth)  
IEC 61000-4-6:1996 / EN 61000-4-6:1996 (3V, 0.15~80 MHz, 80% AM, power line)  
IEC 61000-4-11:1994 / EN 61000-4-11:1994 (Dips 1 cycle, 100%)  
Canada: ICES-001:2004  
A
A
B
B
A
B
Australia/New Zealand: AS/NZS CISPR11:2004  
IEC 61010-1:2001 / EN 61010-1:2001  
Canada: CAN/CSA-C22.2 No. 61010-1-04  
USA: ANSI/UL 61010-1:2004  
Safety  
Additional Information:  
The product herewith complies with the essential requirements of the Low Voltage Directive 2006/95/EC and the  
EMC Directive (2004/108/EC) and carries the CE Marking accordingly (European Union).  
1Performance Criteria:  
A Pass - Normal operation, no effect.  
B Pass - Temporary degradation, self recoverable.  
C Pass - Temporary degradation, operator intervention required.  
D Fail - Not recoverable, component damage.  
N/A – Not applicable due to the product is a battery operated device  
Notes:  
Regulatory Information for Canada  
ICES/NMB-001:2004  
This ISM device complies with Canadian ICES-001.  
Cet appareil ISM est confomre à la norme NMB-001 du Canada.  
Regulatory Information for Australia/New Zealand  
This ISM device complies with Australian/New Zealand AS/NZS CISPR11:2004  
U2802A 31-Channel Thermocouple Input User’s Guide  
XI  
XII  
U2802A 31-Channel Thermocouple Input User’s Guide  
Contents  
Notices II  
1 Getting Started  
Introduction to Agilent U2802A 31-Channel Thermocouple Input  
Standard Purchase Items Checklist  
2
6
Installations and Configurations  
7
2 Features and Functions  
Applications 27  
Theory of Operation 29  
4 Product Specifications  
General Specifications 48  
Product Characteristics 49  
System Accuracy Specifications 51  
System Typical Performance 57  
Making Accurate Temperature Measurements 59  
Contents 1  
5 Calibration  
Calibration Information 62  
Zeroing Function 62  
Restore Factory Calibration 63  
Contents 2  
List of Figures  
Figure 2-2 System functionality block diagram for U2802A 29  
U2802A 33  
Figure 3-1 U2802A pin assignment 38  
Figure 3-2 Connector 1 pin assignment for U2355A and U2356A 44  
Figure 3-3 Connector 2 pin assignment for U2355A and U2356A 45  
Figure 4-1 Thermoelectric characteristics for various thermocouple types 57  
Figure 4-2 U2802A measurement accuracy plot for various thermocouples type 58  
Contents 3  
List of Tables  
Table 4-1 U2802A measurement accuracy with U2355A or U2356A, at 23 °C 5 °C,  
with different number of averaging points. 51  
Table 4-2 U2802A measurement accuracy with U2355A, at 0 to 18 °C and 28 to 45 °C,  
with different number of averaging points. 52  
Table 4-3 U2802A measurement accuracy with U2356A, at 0 to 18 °C and 28 to 45 °C,  
with different number of averaging points. 53  
Contents 4  
Agilent U2802A 31-Channel Thermocouple Input  
User’s Guide  
1
Getting Started  
Introduction to Agilent U2802A 31-Channel Thermocouple Input  
2
Product Overview  
Product Outlook  
Installations and Configurations  
C. Install the Agilent IO Libraries Suite Version 14.2 or Higher 11  
E. Install the Agilent Measurement Manager 13  
F. Connect Your DAQ Device to the PC 17  
G. Hardware Verification 17  
H. Connect Your U2802A to DAQ 17  
I. Launch Your Agilent Measurement Manager 18  
IVI-COM Drivers 19  
This chapter introduces the new Agilent U2802A 31- channel  
thermocouple input device and provides quick start information. It also  
provides product outlook, installation configuration and troubleshooting  
guide.  
Agilent Technologies  
1
   
1
Getting Started  
Introduction to Agilent U2802A 31-Channel Thermocouple Input  
The Agilent U2802A 31- channel thermocouple input is a thermocouple  
input device that functions to convert low input voltage signal  
(< 100 mV) from a thermocouple into an output voltage range suitable for  
data acquisition (DAQ) device ( 10 V).  
The Agilent U2802A thermocouple signal conditioner is to be used in  
conjunction with the U2355A or U2356A model DAQ to enable  
temperature measurements using thermocouples.  
It works as a standalone device attached to a single DAQ. The U2802A  
thermocouple device is connected to the modular DAQ via SCSI cables.  
Agilent U2802A accepts eight standard thermocouple types defined in the  
NIST ITS-90 Thermocouple Database, which are Type B, E, J, K, N, R, S  
and T.  
It is ideal for a broad variety of temperature and voltage measurement  
applications in education, industrial and scientific environments. The  
U2802A comes with an on- board EEPROM features. Hence, it allows user  
to store calibration data in volatile memory. Therefore, the U2802A is  
robust, cost- effective, and user friendly device.  
For detailed product specifications, please refer to “General Specifications”  
on page 48.  
2
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Getting Started  
1
Product Overview  
Product Outlook  
Top View  
Connector 1  
Connector 2  
Cover  
Detachable  
Front View  
Railing Guide  
Strain Release Assembly  
U2802A  
31-Channel Thermocouple Input  
Footer  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
3
     
1
Getting Started  
Side View  
Ventilation  
Holes  
Bottom View  
4
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
Product Dimensions  
Top View  
Front View  
U2802A  
31-Channel Thermocouple Input  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
5
 
1
Getting Started  
Standard Purchase Items Checklist  
Inspect and verify that you have all the following items upon standard  
purchase of U2802A 31- channel thermocouple input device. If there are  
missing items, contact the nearest Agilent Sales Office.  
Agilent U2802A 31- channel thermocouple input device  
Power supply splitter  
Two 68- pin SCSI cables (1 m)  
One J- type thermocouple  
Quick Start Guide  
Product Reference CD- ROM  
Agilent Automation-Ready CD (contains the Agilent IO Libraries Suite)  
Certificate of Calibration  
6
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Getting Started  
1
Installations and Configurations  
The U2802A is used in conjunction with the U2355A or U2356A DAQ. If  
you are using the U2300A Series with the Agilent Measurement Manager,  
follow the step- by- step instructions as shown in the following flowchart.  
If you do not wish to specifically use the U2300A Series with the Agilent Measurement  
Manager software, and use the DAQ devices with Agilent VEE, LabVIEW or Microsoft  
Visual Studio only, you can skip steps E and I in the following flowchart.  
You need to install IVI-COM driver before using the U2300A Series with Agilent VEE,  
LabVIEW or Microsoft Visual Studio.  
NOTE  
If you have installed and currently using the U2300A Series, you may skip the  
initial steps and proceed to step H for U2802A Thermocouple Input  
installation in the following flowchart. Make sure Agilent Measurement  
Manager version 1.4 and above is installed before you proceed.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
7
   
1
Getting Started  
A. Check Your System Requirements  
i. If Agilent IO Libraries Suite version 14.1 or  
ii. If Agilent IO Libraries Suite 14.2 has  
been pre-installed  
(requires uninstalling of this version)  
iii. If Agilent IO Libraries Suite is not installed  
C. Install the Agilent IO Libraries Suite Version 14.2 or Higher  
D. Install the DAQ Hardware Driver  
E. Install the Agilent Measurement Manager  
G. Hardware Verification  
H. Connect Your U2802A to DAQ  
I. Launch Your Agilent Measurement Manager  
Figure 1-1 U2802A installation flowchart  
8
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Getting Started  
1
A. Check Your System Requirements  
Before installing the Agilent Measurement Manager software and DAQ  
hardware driver, make sure your PC meets the following minimum system  
requirements for installation and data acquisition.  
Processor 1.6 GHz Pentium IV or higher  
Operating system One of the following Microsoft Windows versions:  
Windows XP Professional or Home Edition (Service Pack 1 or later),  
Windows 2000 Professional (Service Pack 4 or later)  
Browser Microsoft Internet Explorer 5.01 or higher  
Available RAM 512 MB or higher is recommended  
Hard disk space 1 GB  
1
Prerequisite Agilent IO Libraries Suite 14.2 or higher, Agilent T&M  
2
Toolkit 2.1 Runtime version , Agilent T&M Toolkit Redistributable Package  
2
2
2.1 patch , Microsoft .NET Framework version 1.1 and 2.0  
1
2
Available on Agilent Automation-Ready CD.  
Bundled with Agilent Measurement Manager application software installer.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
9
 
1
Getting Started  
B. Check Your Pre-Installed IO Libraries Suite  
To check the version of your pre- installed IO Libraries Suite follow the  
instructions below.  
1 Right-click IO Control icon on your taskbar notification area and the  
context menu will appear as shown below. Select About Agilent IO Control.  
2 Agilent IO Control window will appear and the version of installed IO  
Libraries Suite will be displayed as shown below.  
If the Agilent IO Control icon is not visible on the taskbar notification area it may indicates  
that,  
NOTE  
you do not have the Agilent IO Libraries installed, or  
you have hidden the Agilent IO Control icon from the taskbar notification area.  
To activate the icon, go to Start > All Programs > Agilent IO Libraries Suite > Utilities >  
IO Control. You will now see the Agilent IO Control icon appear on your taskbar  
notification area.  
10  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Getting Started  
1
i. If Agilent IO Libraries Suite version 14.1 or lower has been pre-installed  
If you have Agilent IO Libraries Suite version 14.1 or lower installed on  
your PC, you are required to uninstall the IO Libraries Suite.  
1 To perform uninstallation, go to Start > Control Panel > Add or Remove  
Agilent IO Libraries Suite 14.1 or lower version.  
2 Click Change/Remove and select Remove when the instructions on the  
screen prompted to proceed uninstalling the IO Libraries Suite  
version 14.1 or lower from your PC.  
3 Proceed to C. Install the Agilent IO Libraries Suite Version 14.2 or Higher.  
ii. If Agilent IO Libraries Suite 14.2 has been pre-installed  
If you have the IO Libraries Suite 14.2 pre- installed on your PC, skip Step  
C and proceed to D. Install the DAQ Hardware Driver.  
iii. If Agilent IO Libraries Suite is not installed  
If you do not have IO Libraries Suite installed on your PC, go to C. Install  
the Agilent IO Libraries Suite Version 14.2 or Higher.  
C. Install the Agilent IO Libraries Suite Version 14.2 or Higher  
The Agilent IO Libraries Suite 14.2 is available in the Agilent  
Automation- Ready CD that comes with the standard purchase of Agilent  
U2300A Series USB Multifunction Data Acquisition Devices or U2802A  
31- Channel Thermocouple Input.  
If you do not have the Agilent Automation-Ready CD, obtain the Agilent IO Libraries  
NOTE  
Ensure that you do not have any USB DAQ device connected to your PC during  
installation of the Agilent IO Libraries Suite.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
11  
       
1
Getting Started  
1 Disconnect any USB DAQ device that is connected to your PC and close  
all other applications on your PC.  
2 Insert the Agilent Automation- Ready CD into your CD- ROM drive, and  
follow the instructions on your screen.  
3 If the installation does not start automatically, go to Start > Run (on the  
Windows start menu) and type <drive>:\autorun\auto.exe where  
drive is your CD- ROM drive.  
4 If you obtain the Agilent IO Libraries Suite from the web, save the  
self- extracting zip file (*.exe) to any location on your hard disk.  
5 Double- click the installation file to launch the installation.  
6 Follow the instructions on your screen to proceed with the installation.  
7 After the installation is completed, you will see the IO Control icon on  
the Windows taskbar notification area as shown below.  
For detailed installation instructions, refer to the Agilent IO Libraries Suite  
NOTE  
12  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
D. Install the DAQ Hardware Driver  
Refer to Agilent U2300A Series USB MultifunctionData Acquisition Devices  
User’s Guide for DAQ hardware driver installation procedure.  
NOTE  
E. Install the Agilent Measurement Manager  
Make sure Agilent Measurement Manager version 1.4 and above is installed  
before U2802A is used.  
NOTE  
1 Verify that you have the hardware driver installed. If not, close all other  
applications on your PC and insert the Product Reference CD- ROM into  
your CD- ROM drive.  
2 Click Measurement Manager on the Agilent Modular Products Installation  
Menu to begin the installation.  
3 If the installation menu does not appear after a few seconds, go to Start  
> Run and type <drive>:\Application\Modular Instruments  
Measurement Manager\setup.exe, where drive is your CD-ROM  
drive.  
4 Click OK to begin installation.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
13  
   
1
Getting Started  
5 If you do not have the Agilent T&M Toolkit 2.1 Runtime version and/or  
Microsoft .NET Framework version 1.1 and 2.0 installed, the  
InstallShield Wizard software pre- requisite will appear as shown below.  
6 Click OK to begin installation of the listed missing software.  
7 Once the above installation is completed, installation of the Agilent  
Measurement Manager software will proceed as normal  
8 The Agilent Measurement Manager InstallShield Wizard dialog box will  
appear as shown below. Click Next to begin.  
14  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
9 Read the License Agreement and select I accept the terms in the License  
Agreement to proceed. You may click Print to print a hardcopy of the  
Agilent License Terms for reference. Click Next to proceed.  
10 Fill in the Customer Information form, as shown below, accordingly and  
click Next.  
11 Click Next to install to the folder specified or click Change to install to  
a different folder.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
15  
1
Getting Started  
12 Click Install to begin the installation of Agilent Measurement Manager.  
13 Click Finish when the installation has completed.  
14 A shortcut to this software will be created on your desktop.  
USING THE LICENSED MATERIALS INDICATES YOUR ACCEPTANCE OF THE LICENSE  
TERMS. IF YOU DO NOT AGREE TO ALL OF THESE TERMS, YOU MAY RETURN ANY  
UNOPENED LICENSED MATERIALS FOR A FULL REFUND. IF THE LICENSED MATERIALS  
ARE BUNDLED OR PRE-LOADED WITH ANOTHER PRODUCT, YOU MAY RETURN THE  
ENTIRE UNUSED PRODUCT FOR A FULL REFUND.  
NOTE  
16  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
F. Connect Your DAQ Device to the PC  
Refer to Agilent U2300A Series USB MultifunctionData Acquisition Devices User’s Guide  
for “Connect Your DAQ Device to PC” procedure.  
NOTE  
During DAQ connectivity, the bundled power supply splitter in U2802A is used to supply  
power for both U2802A and U2355/56A DAQ, replacing the power supply cable.  
G. Hardware Verification  
Refer to Agilent U2300A Series USB MultifunctionData Acquisition Devices User’s Guide  
for “Hardware Verification” procedure.  
NOTE  
H. Connect Your U2802A to DAQ  
1 Connect U2802A to U2355/56A DAQ using two 68- pin SCSI cables.  
2 Connect 12 VDC power supply to the power jack on the rear panel of  
the U2802A and the U2355/56A DAQ with the bundled power supply  
splitter.  
3 Power on U2355A or U2356A DAQ.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
17  
     
1
Getting Started  
I. Launch Your Agilent Measurement Manager  
The Agilent IO Control will launch automatically when you start your PC.  
Launching Agilent Measurement Manager without Agilent IO Control running will  
cause Agilent Measurement Manager to fail from detecting or establishing any  
connection with the USB DAQ device connected to your PC.  
NOTE  
To run Agilent IO Control, go to Start > All Programs > Agilent IO Libraries Suite  
> Utilities > IO Control.  
1 Double- click the Agilent Measurement Manager software icon on your  
desktop or go to Start > All Programs > Agilent > Measurement Manager >  
Agilent Measurement Manager to launch the software.  
2 The Agilent Measurement Manager welcome screen will appear as  
shown below.  
3 The Select USB Device dialog box will appear and shows the connected  
DAQ devices. To start the application, select a DAQ device and click OK  
to establish the connection.  
For more information on how to use the Agilent Measurement Manager, refer to the  
Agilent Measurement Manager Help File.  
NOTE  
18  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Getting Started  
1
IVI-COM Drivers  
The Agilent IVI- COM drivers simplify instrument control when you are  
working in a COM- compatible environment. IVI- COM allows you to  
programmatically control your instrumentation and make measurements  
while providing a greater degree of instrument interchangeability and code  
reuse. The Agilent IVI- COM drivers support the use of IntelliSense for  
even greater ease- of- use within a Microsoft development environment.  
The Agilent IVI- COM driver supports all Agilent Series DAQs. The Agilent  
Firmware Revision: A.2006.10.10 is the minimum revision required for full  
driver functionality.  
An IVI- COM driver can program a particular set of instrument models. It  
implements an instrument- specific interface tuned to the capabilities of  
those models. The driver may also implement an IVI class- compliant  
interface which implements a limited set of functionality common to all  
instruments of the class. Instrument class-compliant interfaces are defined  
by the IVI Foundation. The application writer must choose whether to use  
the instrument- specific interface or the class- compliant interface.  
The IVI inherent capabilities, through the IIviDriver interface, are available  
in both the instrument- specific interface and class- compliant interface.  
The general programming techniques are also the same.  
Choosing Instrument-Specific Interface  
With this interface, you have the benefit of full access to the instrument's  
capabilities. All capabilities in the class- compliant interface are also  
covered by the instrument- specific interface, but you will find some  
capabilities in the instrument- specific interface that are not available  
through the class- compliant interface. You may also see some performance  
enhancements, as the driver can be tuned to use efficient programming  
methods for that particular instrument.  
Choosing Class-Compliant Interface  
By limiting your program to the class- compliant interface, you have the  
potential advantage of syntactic interchangeability. Hence, another  
IVI- COM driver (and instrument) which supports the same class could be  
substituted for the original driver, if the prior IVI- COM driver supports all  
the capability groups used in the original driver. In this case, the  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
19  
   
1
Getting Started  
application will compile, link, and execute without error. The test results,  
however, may be quite different because different instruments measure  
and generate signals differently. For more information on class- compliant  
Using Class-Compliant Interface  
Generally, you gain no advantage from using class-compliant interface over  
using just the instrument- specific interface. However, if you can isolate the  
usage of the instrument- specific interface, you may see some advantages.  
Replacing the IVI- COM driver then involves fixing the syntactic  
incompatibilities in the isolated code.  
IVI- COM drivers will be provided to users. The drivers can also be used in  
a variety of development environments. For more information on IVI, visit  
Below are the IVI- COM drivers provided:  
AgilentVEE support through COM mechanism using IVI- COM  
Visual Basic 6 support through COM mechanism using IVI- COM  
C++ support through COM mechanism using IVI- COM  
Visual Basic 7 support through COM Interop mechanism using IVI- COM  
C# support through COM Interop mechanism using IVI- COM  
National Instruments LabVIEW support through COM mechanism using  
IVI- COM  
The Agilent firmware update utility is provided to allow users to update  
firmware on instruments. Update is made available through Agilent  
Developer Network (ADN) website:  
20  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
Programming Environments  
An IVI- COM driver works well in a variety of application development  
environments (ADEs) below:  
Agilent VEE  
Microsoft® Visual Basic® 6  
Visual Studio C++  
Visual Basic 7  
C#  
National Instruments LabVIEW  
IVI-COM Driver Installation  
1 Verify that your PC meets the minimum system requirements. (See  
‘System Requirements’ on page 9.)  
2 Close all other applications on your PC.  
3 Insert the Product Reference CD into the CD- ROM drive of your PC.  
4 Wait for a few seconds for the auto- run window to appear.  
5 If the auto- run window does not appear automatically, click Start > Run,  
then type <drive>:\Autorun.exe, where <drive> is your CD drive  
alphabet.  
6 When the auto- run window appears, click Software Driver on the Agilent  
Modular Products Installation Menu.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
21  
1
Getting Started  
7 Click IVI-COM to open the IVI- COM Driver Installation Menu.  
8 Check on the U2300A Series and click Install and wait for the  
Installation Dialog to appear.  
9 When the Installation Dialog appears, click Next to begin the IVI Driver  
installation.  
10 Read the License Agreement(s). To accept the terms, click on the radio  
button labeled I accept the terms in the License Agreement then click Next to  
continue.  
11 When the Setup Type dialog box appears, as shown below, clicking Install  
will install all features for your configuration in standard locations on  
your PC.  
22  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Getting Started  
1
12 If you choose a Custom setup, the Select Features dialog box will appear.  
a Click on any feature in the list to see the feature’s description and  
space requirement. It is recommended that you install the sample  
programs if you plan to program with the IVI driver. However, you  
may omit this recommendation to save space.  
b Select the check box for each feature to be installed. Clear the check  
box to omit the feature selection.  
c Click Next.  
4
5
When the Ready to Install dialog box appears, click Install to confirm your  
choices and begin copying files.  
When the Complete dialog box appears, click Finish.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
23  
1
Getting Started  
24  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Agilent U2802A 31-Channel Thermocouple Input  
User’s Guide  
2
Features and Functions  
Features 26  
Applications 27  
System Overview 28  
Theory of Operation 29  
Functionality of the System 29  
Functional Block Diagram 30  
This chapter contains details of the product features, applications,  
system overview and theory of operation. From this chapter, you will  
understand the Agilent U2802A 31- channel thermocouple input system  
overview and functionality of this device.  
Agilent Technologies  
25  
   
2
Features and Functions  
Features  
The U2802A Thermocouple Input conditioning device is complete with the  
following features:  
Up to 31 differential input mode, or 31- single ended inputs in voltage  
input mode. Each of the 31 channels can be configured as either  
thermocouple or voltage input mode independently.  
×97.673 gain setting for thermocouple input mode.  
Built- in thermistor for cold junction compensation (CJC).  
Built- in zeroing function to compensate for overall system offset errors  
due to temperature drift.  
On- board EEPROM that allows user to restore back original factory  
calibration data.  
Open thermocouple detection that allows user to check for any loose or  
broken thermocouple connection before starting the data acquisition  
process.  
Supports thermocouple type J, K, R, S, T, N, E, and B.  
26  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Features and Functions  
2
Applications  
The U2802A Thermocouple Input conditioning device is designed for  
robust and demanding industrial applications. This product is suitable for  
a wide range of applications in various fields inclusive of:  
Consumer electronics  
Product thermal analysis and characterization  
Environmental testing (Eg: Temperature Cycle)  
Process monitoring (Eg: Oven or solder reflow temperature  
monitoring)  
Education  
Study of electronic cooling properties  
Material properties testing  
Container temperature profiling  
Appliances testing  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
27  
   
2
Features and Functions  
System Overview  
Figure 2-1 System overview of U2802A with DAQ  
The U2802A is essentially an amplifier module with a built- in temperature  
sensor (thermistor). In thermocouple mode, the U2802A input channel is  
used to amplify a differential voltage signal from a thermocouple (or any  
low voltage signal source in the range of 100 mV) by 100 times. The  
signal is then output as an analog voltage in the 10 V range into the DAQ  
for conversion to a digital voltage reading.  
The built- in thermistor in the U2802A can be read from Channel AI148 of  
the U2300A series DAQ. The conversion from voltage to temperature for  
this thermistor reading is done automatically by the AMM software. This  
temperature reading will subsequently be used as the Cold Junction  
Compensation (CJC) reference temperature.  
With the correct voltage reading from the thermocouple and the CJC  
temperature, the AMM software will then proceed to convert the  
thermocouple voltage reading into a temperature reading, based on the  
NIST ITS-90 Thermocouple Database. This reading is then corrected for  
both gain and offset errors due to the U2802A amplifiers using the  
calibration constants stored in the U2802A EEPROM, which are read by  
the PC via the DAQ's digital I/O lines.  
The U2802A also has a built- in zeroing function, which allows users to  
zero out the entire system's offset error, thus increasing the overall  
accuracy of the system.  
28  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
     
Features and Functions  
2
Theory of Operation  
Functionality of the System  
Figure 2-2 System functionality block diagram for U2802A  
1 Thermocouple voltage signals are detected at the U2802A thermocouple  
inputs.  
2 Signal is amplified with a gain of 97.673 by the U2802A.  
3 The U2355A or U2356A DAQ converts the analog voltage signals to  
digital voltage readings.  
4 The AMM software (or IVI- COM driver) reads the Gain and Offset  
calibration constants from the U2802A EEPROM via the DAQ DIO lines.  
The digital voltage readings will be calibrated based on these constants.  
5 The AMM software (or IVI- COM driver) converts the calibrated voltage  
readings to temperature readings using the ITS- 90 conversion  
polynomials.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
29  
       
2
Features and Functions  
Functional Block Diagram  
The block diagram below in Figure 2-3 illustrates the key functional  
components of the U2802A.  
Figure 2-3 Functional block diagram for U2802A  
30  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Features and Functions  
2
The major functional blocks of the U2802A module are:  
Analog input channel circuitry  
Cold junction sensor  
Digital control logic  
EEPROM  
Analog input channel circuitry  
The analog circuitry for each channel consists of an instrumentation  
amplifier with a fixed gain of 97.673, a 4 Hz RC low- pass filter, and an  
output buffer. The multiplexers at the input and output of each channel  
allows each channel to be configured for three modes of operation as  
listed below:  
Thermocouple input mode: In thermocouple mode, the thermocouples (or  
any floating voltage source) should be connected to the TCn+ and TCn–  
terminals as illustrated in Figure 2- 4. All TCnterminals are internally  
tied to module ground with a 10 MΩ resistor. The TCn+ and TCnsignals  
are routed to the differential inputs of the instrumentation amplifier.  
Differential voltage signals at the TCn+ and TCnterminals are amplified,  
filtered and driven out by single- ended output voltage to the  
corresponding AI channel on Rear Connector 1.  
Figure 2-4 Functional block diagram for thermocouple mode in U2802A  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
31  
 
2
Features and Functions  
Bypass mode: In bypass mode, the TCn+ input is routed directly to the  
corresponding AI channel on Rear Connector 1. The single- ended signals  
tied to TCn+ should be referenced to a GND pin, and not to the TCn–  
input, as it is not directly connected to GND. The signal connection will  
depend on the type of source used.  
For floating signal sources, all input signals are connected to the ground  
in the U2802A as illustrated in Figure 2- 3. However, it is not  
recommended to tie ground-referenced signal sources in this manner. Any  
potential differences between the signal source ground and the U2802A  
ground could potentially induce excessive current to flow through the  
ground wires causing the wires and module to be damaged.  
Figure 2-5 Floating signal source configuration in U2802A  
For ground- referenced signal sources and differential signal sources, the  
configuration in Figure 2- 6 is recommended. Take note that the  
corresponding DAQ channel will need to be configured as a DIFF input to  
enable this type of connection.  
32  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Features and Functions  
2
Figure 2-6 Ground-referenced and differential signal sources configuration in U2802A  
Zero mode: In zero mode, the positive and negative inputs of the  
instrumentation amplifier are shorted together. The output of the  
instrumentation amplifier is driven out to the corresponding AI channel.  
The voltage measured in this mode corresponds to the offset voltage of  
the channel. This voltage can be subtracted out of the subsequent  
thermocouple mode measurements in order to increase the measurement  
accuracy. Do take note that this mode only works for channels that have  
been configured to be in the thermocouple mode. Channels configured for  
bypass mode will not be affected when this mode is selected.  
Each channel is equipped with an open thermocouple detection feature,  
where the 10 MΩ resistor is tied to the +15 V power supply rail. This  
feature can only be globally enabled or disabled for all channels,  
regardless of the channel mode setting. When enabled, outputs of the  
channels are set to thermocouple mode where the inputs are left  
open- circuited. This causes the positive power supply rail voltage (above  
+10 V) to be saturated up, indicating that the channel either has a broken  
thermocouple or the thermocouple is not connected. For channels set to  
bypass mode, channels with an open- circuited input will also be saturated  
to the positive supply rail voltage.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
33  
 
2
Features and Functions  
For bypass mode channels that are connected to valid voltage sources, the  
10 MΩ pull- up resistor will cause additional current to flow through the  
voltage source. However, this additional current measurement is small and  
negligible for low impedance voltage sources.  
For thermocouple mode channels connected to valid thermocouples, the  
presence of the pull- up resistor introduces approximately 0.75 µA of  
current through the thermocouple wires. This current introduces  
additional errors when using thermocouples with high resistances, and the  
measurement accuracy could be affected.  
Cold junction sensor  
A thermistor (RT1) is placed in between the screw terminals to measure  
the temperature of the thermocouple junction for CJC. The output voltage  
from the sensor is fed through a 4 Hz RC low-pass filter and buffered to  
the AI148 pin on Rear Connector 1. The conversion from voltage to  
temperature is done automatically by the AMM software.  
Digital control  
The digital control circuit consists of registers that controls the mode of  
each channel and the open- thermocouple detect feature. The registers are  
addressed and clocked via the digital I/O pins on Rear Connector 2. This  
will be handled automatically by the AMM software.  
EEPROM  
The gain and offset calibration factors for each channel are stored in the  
EEPROM during factory calibration and will be retrieved prior to taking  
measurements. The EEPROM is tied to the digital I/O pins on Rear  
Connector 2. The communication between the EEPROM and host PC is  
automatically handled by the AMM software. In addition to the calibration  
factors, the EEPROM stores the module ID, serial number, date of  
calibration, which can also be retrieved before measurements are taken.  
34  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Features and Functions  
2
Open Thermocouple Detection  
The U2802A provides a built- in 10 MΩ resistor on each TC+ terminal,  
which is pulled up to the internal +15 V power supply rail. This resistor  
can be enabled or disabled via the digital I/O pins on Rear Connector 2.  
When enabled, this 10 MΩ pull- up resistor and the 10 MΩ pull- down  
biasing resistor will cause the output from any unconnected thermocouple  
input channels to saturate to the maximum output voltage. The U2355A  
and U2356A devices can read this saturated channel and detect that a  
particular channel has an open thermocouple input.  
Trigger, Counter, External Timebase, and Analog Output  
The U2802A provides a direct access to the analog and digital trigger  
lines, counter channels, external timebase input, and analog output  
channels from the U2355A and U2356A devices. These lines are routed  
directly from the Rear Connector 1 and 2 to the J60 screw terminal  
connector. Please refer to pin description for Connector J60 on page 43.  
Precautions should be taken when driving high slew rate and frequency  
clocks into the Counter and External Timebase lines to avoid excessive  
noise coupling into other analog and digital lines. If excessive coupling or  
crosstalk is observed, clock output drive strengths and slew rates should  
be lowered to reduce coupling while still maintaining proper digital  
function.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
35  
2
Features and Functions  
36  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Agilent U2802A 31-Channel Thermocouple Input  
User’s Guide  
3
Pin Configurations and Assignments  
Pin Configurations 38  
Pin Assignments 38  
Pin Description 39  
Connector Pinout 44  
This chapter described the Agilent U2802A 31- channel thermocouple  
input device pin configurations and connector pinout for user’s  
reference.  
Agilent Technologies  
37  
   
3
Pin Configurations and Assignments  
Pin Configurations  
Pin Assignments  
Figure 3-1 U2802A pin assignment  
38  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
       
Pin Configurations and Assignments  
3
Pin Description  
Connector J71  
Pin Pin name Description  
1
2
TC1+  
TC1–  
In thermocouple mode, TCx+ and TCx– are the thermocouple differential  
input. In voltage mode, single ended input at TCx+ and GND. TCx– is not  
connected.  
3
TC2+  
TC2–  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
4
5
TC3+  
TC3–  
6
7
TC4+  
TC4–  
8
9
TC5+  
TC5–  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
TC6+  
TC6–  
TC7+  
TC7–  
TC8+  
TC8–  
GND  
Module Ground  
TC17+  
TC17–  
TC18+  
TC18–  
TC19+  
TC19–  
TC20+  
TC20–  
TC21+  
TC21–  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
39  
 
3
Pin Configurations and Assignments  
28  
29  
30  
31  
32  
33  
34  
TC22+  
TC22–  
TC23+  
TC23–  
TC24+  
TC24–  
GND  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
Module Ground  
40  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Pin Configurations and Assignments  
3
Connector J50  
Pin Pin name Description  
1
GND  
Module Ground  
2
TC16–  
TC16+  
TC15–  
TC15+  
TC14–  
TC14+  
TC13–  
TC13+  
TC12–  
TC12+  
TC11–  
TC11+  
TC10–  
TC10+  
TC9–  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
3
4
5
6
7
8
9
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
30  
TC9+  
GND  
Module Ground  
GND  
Module Ground  
GND  
Module Ground  
TC31–  
TC31+  
TC30–  
TC30+  
TC29–  
TC29+  
TC28–  
TC28+  
TC27–  
TC27+  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
41  
3
Pin Configurations and Assignments  
31  
32  
33  
34  
TC26–  
TC26+  
TC25–  
TC25+  
TC input or voltage input (See TC1+/– description)  
TC input or voltage input (See TC1+/– description)  
42  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Pin Configurations and Assignments  
3
Connector J60  
Pin Pin name  
Description  
1
1
COUNT302_CLK  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Module Ground  
1
1
1
1
1
2
COUNT302_GATE  
COUNT302_UPDOWN  
COUNT302_OUT  
EXTD_AI_TRIG  
EXT_TIMEBASE  
GND  
3
4
5
6
7
1
8
AO_GND  
Directly connected to DAQ  
Directly connected to DAQ  
Module Ground  
1
9
AO_GND  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
GND  
GND  
Module Ground  
GND  
Module Ground  
1
1
1
1
1
COUNT301_CLK  
COUNT301_GATE  
COUNT301_UPDOWN  
COUNT301_OUT  
EXTD_AO_TRIG  
GND  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Module Ground  
1
1
1
1
AO201  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Directly connected to DAQ  
Module Ground  
AO202  
AO_EXT_REF  
EXTA_TRIG  
GND  
GND  
Module Ground  
1. Refer to the U2300A Series USB Multifunction Data Acquisition Devices User’s Guide for  
connectivity  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
43  
 
3
Pin Configurations and Assignments  
Connector Pinout  
Rear panel pinout for Connector 1  
Figure 3-2 Connector 1 pin assignment for U2355A and U2356A  
44  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
     
Pin Configurations and Assignments  
3
Rear panel pinout for Connector 2  
Figure 3-3 Connector 2 pin assignment for U2355A and U2356A  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
45  
 
3
Pin Configurations and Assignments  
46  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Agilent U2802A 31-Channel Thermocouple Input  
User’s Guide  
4
Product Characteristics 49  
System Accuracy Specifications 51  
Calculating System Accuracy 54  
System Typical Performance 57  
Making Accurate Temperature Measurements 59  
This chapter specifies the environmental conditions, characteristics, and  
specifications of the Agilent U2802A 31- channel thermocouple input  
device. It also covers the system accuracy, typical performance and  
guidelines to make accurate temperature measurements.  
Agilent Technologies  
47  
   
4
Product Specifications  
General Specifications  
POWER CONSUMPTION  
12 VDC, 750 mA maximum  
OPERATING ENVIRONMENT  
Operating temperature from 0 °C to 55 °C  
Relative humidity at 50% to 85% RH (non-condensing)  
Altitude up to 2000 meters  
STORAGE COMPLIANCE  
–40 °C to 70 °C  
SAFETY COMPLIANCE  
Certified with IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)  
EMC COMPLIANCE  
IEC 61326-1:2002 / EN 61326-1:1997+A1:1998+A2:2001+A3:2003  
CISPR 11:1990/EN55011:1990 – Group 1, Class A  
CANADA: ICES-001: 2004  
Australia/New Zealand: AS/NZS CISPR11:2004  
SHOCK & VIBRATION  
Tested to IEC/EN 60068-2  
IO CONNECTOR  
2 x 68-pin female SCSI connector  
2 x 34-pin screw terminal block  
1 x 24 pin screw terminal block  
DIMENSIONS (WxDxH)  
159.7 mm x 254.2 mm x 40.5 mm  
WEIGHT  
1.036 KG  
WARRANTY  
Three years  
48  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Product Specifications  
4
Product Characteristics  
GENERAL CHARACTERISTICS  
Number of channels  
31 differential and 1 CJC  
Input voltage range for voltage mode  
Input voltage (thermocouple mode)  
Sampling rate for thermocouple mode  
Sampling rate for overall module  
Thermocouple types  
10 V (signal + common mode)  
100 mV  
10 kSa/s total for all channels  
500 kSa/s  
J, K, R, S, T, N, E and B  
INPUT SPECIFICATIONS  
Accuracy (thermocouple mode)  
Overall gain error  
0.06% (23 °C 5 °C)  
15 µV (without zeroing) (23 °C 5 °C)  
6 µV (with zeroing)  
Overall offset error  
Nonlinearity  
< 0.005% of Full Scale Range  
System noise (rms)  
Gain (x1)  
100 µVrms  
5 µVrms  
Gain (x100)  
Common mode rejection ratio (60 Hz)  
Voltage mode  
> 60 dB  
> 80 dB  
Thermocouple mode  
Cold junction accuracy  
1.0 °C typical (23 °C 5°C)  
1.5 °C typical (0 °C to 18 °C, 28 °C to 55 °C)  
INPUT CHARACTERISTICS  
Bandwidth (voltage mode)  
Bandwidth (thermocouple mode)  
> 500 kHz  
4.0 Hz  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
49  
   
4
Product Specifications  
i
ii  
Overvoltage protection  
TC Mode  
Common mode: 17 V (TC+ and TC– with  
respect to GND)  
Differential mode: 7 V (Differential voltage  
between TC+ and TC–)  
Bypass mode  
20 V (TC+ input with respect to GND)  
Power Off Mode  
11 V (TC+, TC– input with respect to GND)  
Input impedance  
Input bias current  
Input offset current  
Gain drift  
> 1 GΩ  
2.5 nA max  
1.5 nA max  
60 ppm / °C max  
1 µV / °C max  
Offset drift  
Filter cutoff frequency (–3 dB) (thermocouple  
mode)  
4.0 Hz  
Filter type (thermocouple mode)  
Low Pass RC Filter  
OTHER FEATURES  
Recommended warm up time  
30 minutes  
i
The overvoltage protection levels specified above indicate the maximum voltage each input pin can  
tolerate without resulting in any damages. However, prolonged exposure to these levels may affect  
device safety and reliability. Hence, it should be avoided where possible.  
ii On the channels configured for thermocouple mode, the TC+ and TC– pins can tolerate up to 17 V  
of differential voltage for a few minutes. However, exceeding 100 mV voltage range on these  
channels can cause additional current to be drawn from the device’s power supply regulators,  
which may damage the device if multiple channels are overdriven for prolonged periods. This  
applies to the case where a voltage source is tied across the TC + and TC – pin. Voltage sources  
n
n
greater than 100 mV should be tied to TC + and GND (floating source), or TC + and TC +  
n
n
n+1  
(grounded source), and have the channels set for bypass mode. Refer to Figure 2-5 on page 32.  
50  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Product Specifications  
4
System Accuracy Specifications  
The Agilent U2802A thermocouple input measurement accuracy with the  
U2355A and U2356A is as shown in Table 4- 1, Table 4-2, and Table 4- 3.  
Assume a 1 °C error in the CJ measurement due to sensor error and temperature  
gradient error in the accuracy numbers in Table 4-1, Table 4-2, and Table 4-3 below.  
Table 4-1, Table 4-2, and Table 4-3 are derived from the U2802A and DAQ input accuracy  
specifications without including the thermocouple error. Refer to “Calculating System  
Accuracy” on page 54 for calculation methodology.  
NOTE  
Thermocouple Measurements Accuracy  
(U2355A, U2356A @ 23 °C 5°C)  
ITS-90  
Temperature  
Range (°C)  
Optimum  
Measurement  
Range(°C)  
Without  
averaging  
50 points  
averaging  
500 points  
averaging  
T/C  
Type  
(
°C)  
(
°C)  
(
°C)  
Low  
High  
Low  
1100  
400  
High  
1820  
1100  
1000  
–150  
1200  
–150  
1200  
–100  
1300  
–100  
1760  
300  
B
E
J
0
1820  
1.9  
1.2  
1.0  
4.4  
1.7  
2.4  
1.6  
2.7  
1.5  
2.7  
1.5  
3.0  
2.0  
5.0  
2.1  
4.5  
1.5  
2.7  
2.5  
1.6  
2.3  
1.5  
2.6  
1.4  
2.6  
1.3  
2.7  
1.4  
3.1  
1.6  
2.8  
1.4  
2.5  
2.0  
1.6  
2.3  
1.5  
2.5  
1.4  
2.6  
1.3  
2.6  
1.3  
2.6  
1.4  
2.4  
1.4  
2.5  
–270  
–210  
–270  
–270  
–50  
1000  
1200  
1372  
1300  
1768  
1768  
400  
–150  
–200  
–150  
–210  
–100  
–200  
–100  
–200  
300  
K
N
R
S
T
–50  
–50  
400  
1760  
400  
–50  
–270  
–100  
–200  
400  
–100  
Table 4-1 U2802A measurement accuracy with U2355A or U2356A, at 23 °C 5 °C, with  
different number of averaging points.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
51  
     
4
Product Specifications  
Thermocouple Measurements Accuracy  
(U2355A @ 0 °C to 18 °C and 28 °C to 45 °C)  
ITS-90  
Temperature  
Range (°C)  
Optimum  
Without  
50 points  
averaging  
500 points  
averaging  
Measurement  
averaging  
Range(°C)  
T/C  
Type  
(
°C)  
(
°C)  
(
°C)  
Low  
High  
Low  
1100  
400  
High  
1820  
1100  
1000  
–150  
1200  
–150  
1200  
–100  
1300  
–100  
1760  
300  
B
0
1820  
3.4  
2.4  
2.2  
7.5  
2.7  
3.8  
2.5  
4.2  
2.9  
4.3  
2.6  
4.9  
3.8  
8.5  
4.2  
7.7  
2.4  
4.3  
3.6  
2.6  
3.6  
2.4  
4.0  
2.8  
4.0  
2.5  
4.2  
3.1  
4.6  
3.4  
4.2  
2.2  
4.0  
2.2  
2.5  
3.6  
2.4  
3.9  
2.8  
3.9  
2.5  
4.0  
3.0  
3.3  
3.2  
3.1  
2.2  
3.9  
E
J
–270  
–210  
–270  
–270  
–50  
1000  
1200  
1372  
1300  
1768  
1768  
400  
–150  
–200  
–150  
–210  
–100  
–200  
–100  
–200  
300  
K
N
R
S
T
–50  
–50  
400  
1760  
400  
–50  
–270  
–100  
–200  
400  
–100  
Table 4-2 U2802A measurement accuracy with U2355A, at 0 to 18 °C and 28 to 45 °C,  
with different number of averaging points.  
52  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Product Specifications  
4
Thermocouple Measurements Accuracy  
(U2356A @ 0 °C to 18 °C and 28 °C to 45 °C)  
ITS-90  
Temperature  
Range (°C)  
Optimum  
Without  
50 points  
averaging  
500 points  
averaging  
Measurement  
averaging  
Range(°C)  
T/C  
Type  
(
°C)  
(
°C)  
(
°C)  
Low  
High  
Low  
1100  
400  
High  
1820  
1100  
1000  
–150  
1200  
–150  
1200  
–100  
1300  
–100  
1760  
300  
B
E
J
0
1820  
6.1  
3.1  
2.4  
14.4  
3.0  
4.2  
2.9  
4.9  
3.3  
5.3  
3.4  
6.8  
6.2  
15.7  
6.4  
14.2  
3.0  
5.3  
6.3  
2.6  
3.7  
2.5  
4.1  
2.9  
4.2  
2.7  
4.6  
3.7  
7.2  
4.0  
6.6  
2.4  
4.2  
2.7  
2.6  
3.6  
2.5  
4.0  
2.9  
4.0  
2.6  
4.1  
3.2  
3.8  
3.4  
3.4  
2.2  
3.9  
–270  
–210  
–270  
–270  
–50  
1000  
1200  
1372  
1300  
1768  
1768  
400  
–150  
–200  
–150  
–210  
–100  
–200  
–100  
–200  
300  
K
N
R
S
T
–50  
–50  
400  
1760  
400  
–50  
–270  
–100  
–200  
400  
–100  
Table 4-3 U2802A measurement accuracy with U2356A, at 0 to 18 °C and 28 to 45 °C,  
with different number of averaging points.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
53  
   
4
Product Specifications  
Calculating System Accuracy  
The overall measurement system comprises of three major components:  
1 DAQ (U2355A or U2356A)  
2 Signal Conditioner (U2802A), which includes CJ Sensor error  
3 Sensor (Thermocouples)  
Errors introduced by each of the above components has to be accounted  
for when calculating the total system accuracy. Since errors from each  
component are not correlated with each other, the total system error will  
be the root- sum- square (RSS) of all the errors:  
2
2
2
2
E
= E  
+ E  
+ E  
THERMOCOUPLE  
TOTAL  
DAQ  
SIG_COND  
Example:  
Assume the following conditions:  
DAQ: Agilent U2355A  
Signal Conditioning: Agilent U2802A  
Ambient temperature: 23 °C  
Thermocouple type: J- type, standard limits of error  
Temperature to measure: 600 °C  
Assume the following error specifications:  
U2355A: Gain error = 0.02% of reading  
Offset error = 1 mV  
U2802A gain = 97.673  
Gain error = 0.06% of reading  
Offset error = 15 µV (with respect to input)  
Zeroing error = 6 µV (with respect to input)  
CJ measurement accuracy = 1 °C  
Thermocouple = greater than 2.2 °C or 0.75% error  
Noise error has been omitted to simplify the example  
54  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
 
Product Specifications  
4
With zeroing, the offset errors from the DAQ and the U2802A can be  
removed, and replaced with the zeroing error.  
Based on the ITS- 90 Thermocouple table, a J- type thermocouple will  
output 33.102 mV at 600 °C, and changes at a rate of approximately  
59 µV/°C. This corresponds to (33.102 mV × 97.673) or 3.2332 V at the  
input of the DAQ.  
Hence, E  
E
and E  
are calculated as follows:  
DAQ, SIG_COND,  
ZEROING  
E  
DAQ  
= 0.02% × 33.102 mV  
= 6.62 µV  
= 6.62 µV ÷ 59 µV/°C  
= 0.112 °C  
E  
SIG_COND  
= (0.06% × 33.102 mV) ÷ 59 µV/ °C  
= 0.337 °C  
E  
ZEROING  
= 6 µV ÷ 59 µV/°C  
= 0.102 °C  
Next, the cold junction sensor error is calculated.  
At 23 °C, a J- type thermocouple output voltage changes at a rate of  
52 µV/ °C. Thus, the CJ sensor error of 1 °C at 23 °C corresponds to  
52uV/ °C × 1°C = 52 µV.  
At 600 °C,  
E  
CJC  
= 52 µV ÷ 59 µV/°C  
= 0.88 °C  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
55  
4
Product Specifications  
Therefore,  
E
= 0.75% × 600 °C = 4.5 °C  
THERMOCOUPLE  
Using the above individual component errors, the total measurement  
system accuracy is calculated as below.  
System accuracy without thermocouple sensor error:  
E
TOTAL  
2
2
2
2
= SQRT(E  
= 0.95 °C  
+ E  
+ E  
+ E  
CJC  
)
DAQ  
SIG_COND  
ZEROING  
System accuracy with thermocouple sensor error:  
E
TOTAL  
2
2
2
2
2
= SQRT(E  
= 4.6 °C  
+ E  
+ E  
+ E  
+ E  
THERMOCOUPLE  
)
DAQ  
SIG_COND  
ZEROING  
CJC  
56  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Product Specifications  
4
System Typical Performance  
Thermoelectric Characteristics  
The thermoelectric characteristics for various thermocouple types is shown  
in Figure 2- 1.  
Figure 4-1 Thermoelectric characteristics for various thermocouple types  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
57  
     
4
Product Specifications  
Typical Error vs. Measurement Performance  
The U2802A measurement error with U2355A or U2356A at 23 °C 5 °C  
is shown in Figure 2- 2.  
Figure 4-2 U2802A measurement accuracy plot for various thermocouples type  
Assume a 1 °C error in the CJ measurement due to sensor error and temperature  
gradient error in the accuracy numbers in Figure 4-2.  
NOTE  
58  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
   
Product Specifications  
4
Making Accurate Temperature Measurements  
Thermocouple measurement accuracy is very sensitive to cold junction  
sensor errors and temperature gradients across the terminals. Keep the  
module away from any heat sources and drafts to minimize any variation  
between channels.  
The channels located closest to the center near the reference thermistor  
will have the best accuracy. It is important to use channels that are  
physically close together on the screw terminals when taking relative  
measurements. Channels that are closest together will have the best  
agreement.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
59  
   
4
Product Specifications  
60  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Agilent U2802A 31-Channel Thermocouple Input  
User’s Guide  
5
Calibration  
Calibration 62  
Calibration Information 62  
Zeroing Function 62  
Restore Factory Calibration 63  
This chapter contains the calibration information and factory restore  
calibration procedure for the Agilent U2802A 31- channel thermocouple  
input device.  
Agilent Technologies  
61  
   
5
Calibration  
Calibration  
Calibration Information  
The Agilent U2802A is factory calibrated and the calibration constants are  
stored in the EEPROM. During initial setup, the calibration constants are  
read from the EEPROM before any measurements are taken.  
Zeroing Function  
The Agilent U2802A thermocouple input device operating in thermocouple  
mode can be set to zero mode, where the differential inputs of each  
channel are shorted together. This zeroing function is used to measure the  
total system offset errors due to initial offset error, temperature drift  
error, and long term drift error from the DAQ (U2355A or U2356A) and  
the U2802A. This measurement can then be subtracted from subsequent  
measurements in order to remove the system offset error.  
62  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
         
Calibration  
5
Restore Factory Calibration  
The Restore Factory Calibration function in the Agilent U2802A is used to  
restore calibration data from user’s settings to factory original settings. To  
perform factory restore calibration, follow the step- by- step instructions  
shown below:  
1 Click Restore Factory Calibration in the thermocouple form.  
2 A dialog box will appear as shown below.  
3 Click OK to start the factory restore calibration process. Click Cancel  
to not perform the restore factory calibration process.  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
63  
   
5
Calibration  
64  
Agilent U2802A 31-Channel Thermocouple Input User’s Guide  
Contact us  
To obtain service, warranty or technical support  
assistance, contact us at the  
following phone numbers:  
United States:  
(tel) 800 829 4444  
Canada:  
(tel) 877 894 4414  
China:  
(tel) 800 810 0189  
Europe:  
(fax) 800 829 4433  
(fax) 800 746 4866  
(fax) 800 820 2816  
(tel) 31 20 547 2111  
Japan:  
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840  
Korea:  
(tel) (080) 769 0800  
Latin America:  
(tel) (305) 269 7500  
Taiwan:  
(fax) (080) 769 0900  
(tel) 0800 047 866  
(fax) 0800 286 331  
Other Asia Pacific Countries:  
(tel) (65) 6375 8100  
(fax) (65) 6755 0042  
Or visit Agilent World Wide Web at:  
Product specifications and descriptions in this  
document subject to change without notice.  
© Agilent Technologies, Inc. 2008  
First Edition, 7 January 2008  
U2802-90003  
Agilent Technologies  

Hotpoint GQ74SI User Manual
Electrolux SIG 305 User Manual
Electrolux 922090 User Manual
Diamond Multimedia Viper X1300PCIE256 User Manual
Dell PowerVault K086P User Manual
Dell Computer Monitor SX2210T User Manual
Cadco LINE CHEF XAF 115 User Manual
Asus P5VDC X User Manual
Asus Computer Monitor MS236 User Manual
Acer X203H User Manual