Teledyne 3160 User Manual

OPERATING INSTRUCTIONS FOR  
Model 3160  
Trace Oxygen Analyzer  
DANGER  
HIGHLYTOXICANDORFLAMMABLELIQUIDSORGASESMAYBEPRESENTINTHISMONITORINGSYSTEM.  
PERSONALPROTECTIVEEQUIPMENTMAYBEREQUIREDWHENSERVICINGTHISSYSTEM.  
HAZARDOUSVOLTAGESEXISTONCERTAINCOMPONENTSINTERNALLYWHICHMAYPERSISTFORA  
TIMEEVENAFTERTHEPOWERISTURNEDOFFANDDISCONNECTED.  
ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE  
CONDUCTINGANYMAINTENANCEORSERVICINGCONSULTWITHAUTHORIZEDSUPERVISOR/MANAGER.  
P/NM52972  
ECO# 99-0323  
08/06/99  
TELEDYNE ANALYTICAL INSTRUMENTS  
TELEDYNE ANALYTICAL INSTRUMENTS  
iii  
Table of Contents  
1Introduction  
1.1Features ....................................................................................... 1-1  
1.1.1 Fixed Features.............................................................. 1-1  
1.1.2 VariableFeatures ......................................................... 1-1  
1.2Components ................................................................................. 1-3  
1.3OptionsandModelNumbers........................................................ 1-3  
1.4Applications ................................................................................. 1-4  
2OperationalTheory  
2.1PrinciplesofOperation ........................................................... 2-1  
2.1.1Micro-FuelCellSensor ................................................. 2-3  
2.1.2SampleSystem .............................................................. 2-3  
2.1.3ElectronicsandSignalProcessing ............................. 2-5  
RangeIdentifierTable ....................................................... 2-7  
3Installation  
3.1UnpackingtheAnalyzer ................................................................ 3-1  
3.2 Front Panel ................................................................................. 3-1  
3.3 Rear Panel................................................................................... 3-2  
3.4GasLineConnections ................................................................... 3-4  
3.4.1 Span Gas In .................................................................. 3-4  
3.4.2InstrumentAir................................................................ 3-4  
3.4.3SampleGasIn ............................................................... 3-4  
3.4.4Vent .............................................................................. 3-4  
3.5Micro-FuelCellInstallationorReplacement .................................. 3-5  
3.6ElectricalConnections............................................................. 3-6  
3.6.1 VoltageSelection.......................................................... 3-7  
3.6.2 FuseChanging .............................................................. 3-8  
3.6.3 ParallelandSerialPorts .......................................... 3-9  
3.7InstallationChecklist ..................................................................... 3-10  
4Operations  
4.1FrontPanelControls..................................................................... 4-1  
4.1.1 SpinningWheel............................................................. 4-2  
4.1.2 Cell Output Factor........................................................ 4-2  
4.2ModesofOperation ..................................................................... 4-2  
ColdStart-Up ........................................................................ 4-5  
(continued)  
TELEDYNE ANALYTICAL INSTRUMENTS  
iv  
Table of Contents (continued)  
CalibrationZeroing ........................................................... 4-6  
CalibrationUsingSpanGas .............................................. 4-8  
SelectActiveSensor(TwoCellsOnly) ............................. 4-11  
InstallSensorandTestAlternateSensor.... (TwoCellsOnly) .. 4-12  
O2 Range Set-Up .................................................................. 4-14  
Set-upAlarms ........................................................................ 4-15  
LoggerSet-Up ....................................................................... 4-17  
Set-UpClockFunctions ......................................................... 4-19  
ChangingPasswordsForRemote  
MonitoringandControl ......................................... 4-21  
SystemStatistics ..................................................................... 4-23  
5Maintenance&Troubleshooting  
5.1RoutineMaintenance .................................................................... 5-1  
5.1.1 SensorMaintenance ............................................... 5-1  
5.1.2 ScrubberMaintenance .................................................. 5-1  
5.2Troubleshooting ............................................................................ 5-2  
Appendix  
A-1 SpecificationsSheet ................................................................ A-1  
A-2 RecommendedSparePartsList .............................................. A-2  
A-3 DrawingList ........................................................................... A-3  
A-4 MaterialSafetyDataSheet ..................................................... A-4  
A-5 MaterialSafetyDataSheet ..................................................... A-5  
TELEDYNE ANALYTICAL INSTRUMENTS  
v
TELEDYNE ANALYTICAL INSTRUMENTS  
vi  
Introduction 1  
Introduction  
1.1 Features  
1.1.1  
Fixed Features  
The Model 3160 is a microprocessor-based trace oxygen monitoring system that  
provides on-line monitoring of trace oxygen at parts-per-million (PPM) levels and can  
directly measure the level of purity in high-grade gases.  
The 3160 computer uses an Intel® 80188 microprocessor combined with 32  
kilobytes of random access memory (RAM) and 128 kilobytes of read-only memory  
(ROM). The total computer program resides in the ROM and is not affected by shut-  
down periods, loss of power, or battery failure. However, calibration parameters and  
stored measurement readings are retained in battery-backed-up RAM, which require  
continuous battery voltage in order to be retained. This information will be lost if  
battery power is interrupted, as in battery replacement.  
Two displays help you monitor trace oxygen levels: a red 4-digit light emitting  
diode (LED) display with one-inch numerals is easily read at a distance or even in  
bright daylight, and a liquid crystal display (LCD) helps you keep track of alarm  
setpoints, ranges, mode and system statistics.  
1.1.2  
Variable Features  
Various models are available. The following features describe the basic model,  
but the exact configuration depends on the standard options incorporated. Paragraph  
1.3 contains a list of standard variations on the basic model.  
The Class B-2 Micro-Fuel Cell measures trace O2 in a mixture of gases. Because  
the Micro-Fuel Cell sensor is a sealed electrochemical device, it is replaced as a unit.  
There are no electrolytes to change or electrodes to clean, making the sensor mainte-  
nance free. The analyzer can be configured with one or two sensor blocks. Each block  
can be isolated by pneumatically switched valves. The software automatically adjusts  
when a second cell block is present.  
TELEDYNE ANALYTICAL INSTRUMENTS  
1-1  
1 Introduction  
The 3160 features six full-scale linear ranges of analysis:  
0–1 PPM  
0–10 PPM  
0–100 PPM  
0–1000 PPM  
0–1%  
0–25%  
An AutoRanging feature automatically selects the range that is appropriate for a  
given reading. For example, if the O2 level exceeds 100% of the current range, the  
analyzer switches to the next higher range. When desired, a manual override feature  
allows a particular range of interest to be locked-in.  
With the AutoSpan feature, the analyzer automatically calibrates at scheduled  
intervals, and automatically performs electronic zeroing and sensor settling detection  
during calibration.  
Five user-programmable absolute-reading (PPM) alarm setpoints with assignable  
Form 1C (SPDT) relays are provided, along with a RS-232C bidirectional serial data  
port, and four signal outputs for chart recorders, etc. The  
RS-232 serial port can be used with or without a modem to connect with  
a personal computer loaded with the Teledyne Remote Analytical  
Control Software (TRACS)  
any terminal or terminal emulation software  
a custom computer program  
Several analyzer functions, including troubleshooting, can be run from your PC  
through the phone line to the analyzer.  
The analyzer has separate sample and span gas ports, which allow the installation  
of an external source of span gas for calibration without interfering with the sample  
gas line.  
An optional scrubber installed into the sample system allows sample gas to be  
used as a zero gas after treatment with the scrubber. If the scrubber is installed, pneu-  
matic valves on either side of the scrubber open automatically during zeroing.  
Note: Units equipped with a scrubber should only be used with inert gases  
and saturated hydrocarbons.  
The analyzer output is not affected by minor changes in the flow rate. Analyzer  
performance is stable under minor mechanical vibration.  
TELEDYNE ANALYTICAL INSTRUMENTS  
1-2  
Introduction 1  
1.2 Components  
The analyzer is designed to mount in a standard 19-inch rack. All of the opera-  
tional controls are easily accessed through the front panel. The analyzer contains each  
of the following:  
Analysis Section  
Single or Dual Electrochemical Micro-Fuel Cell  
Sample System  
Power Supply  
Microcontroller Module  
LED Display  
The front panel consists of:  
LED Display  
LCD Display with Touch Panel  
Micro-Fuel Cell Access Panel  
Flowmeter and flow set knob  
The back panel ports are:  
Gas Inlet/Outlet Ports  
Electrical Connections  
1.3 Options and Model Numbers  
The 3160 is available with the following standard options, which are designated  
by a descriptive suffix on the model number. Most of the options can be combined.  
3160SA  
Single fuel cell.  
3160SB  
Single fuel cell and oxygen scrubber  
Dual fuel cell  
3160DA  
3160DB  
3160SAS  
3160SBS  
3160DAS  
3160DBS  
Dual fuel cell and oxygen scrubber  
Single cell and stainless steel cell block  
Single cell, stainless steel cell block, and oxygen scrubber  
Dual cell and stainless steel cell block  
Dual cell, stainless steel cell block, and oxygen scrubber  
TELEDYNE ANALYTICAL INSTRUMENTS  
1-3  
1 Introduction  
1.4 Applications  
The analyzer is an invaluable tool in the following applications and industries:  
Inert and gaseous hydrocarbon stream monitoring.  
Measuring the purity of various gases in air separation plants.  
Prevention of oxidation by measuring the purity of blanketing gases in  
fiber and glass industries.  
Monitoring and controlling gas atmospheres in the heat treatment of metals  
in steel and other metal industries.  
Gas analysis and research in laboratories and research and development  
areas.  
Semiconductor manufacturing.  
Process monitoring of gaseous monomers—vinyl chloride, propylene,  
butadiene, or ethylene.  
Gas purity certification.  
Glove box leak detection.  
Natural gas treatment and transmission.  
Inert gas welding of exotic metals.  
TELEDYNE ANALYTICAL INSTRUMENTS  
1-4  
Operational Theory 2  
Operational Theory  
2.1 Principles of Operation  
The sub-systems that make up the analyzer are:  
1. Micro-Fuel Cell Sensor  
2. Sample System  
3. Electronic Signal Processing  
The Micro-Fuel Cell sensor is an electrochemical galvanic device that translates  
the amount of oxygen present in the sample into an electrical current. The sample  
system delivers the sample gas in a form that is as unaltered as possible for testing,  
and for leak-free transport of gases through the analyzer.  
The electronic signal processing unit (control module) is designed to simplify the  
operation of the analyzer and accurately process the signal from various components.  
The control module incorporates a microprocessor which allows the operation of the  
analyzer with a minimum of operator interaction. Figure 2-1 illustrates major  
components.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-1  
2 Operational Theory  
I
O
Figure 2-1. Major Components.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-2  
Operational Theory 2  
2.1.1 Micro-Fuel Cell Sensor  
The Micro-Fuel Cell is a sealed plastic disposable oxygen transducer that mea-  
sures 1¼ inches in diameter and is ¾ inch thick. Inside of the cell are a cathode and  
anode immersed in 15% aqueous KOH electrolyte. At one end of the sensor is a  
Teflon diffusion membrane; the other end is sealed with a polyethylene membrane. At  
the rear of the cell is a contact plate consisting of two concentric foils. The foils mate  
with spring-loaded contacts in the sensor block assembly and provide the electrical  
connection to the rest of the analyzer.  
The sensing cathode, located beneath the diffusion membrane, has a surface area  
of 2.48 cm2. The sample gas enters the sensor block through an inlet tube between the  
cathode and the sensor block cap, diffuses through the Teflon membrane, and any  
oxygen in the sample gas is reduced on the surface of the cathode by the following  
mechanism:  
O2 + 2H2O + 4e4OH–  
(cathode)  
When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the  
anode by the following mechanism.  
2Pb 2Pb+2 + 4e–  
(anode)  
The electrons released at the surface of the anode flow to the cathode surface via  
an external circuit. This current is proportional to the amount of oxygen. It is mea-  
sured and used to determine the oxygen concentration in the gas mixture.  
The overall reaction for the fuel cell is:  
2Pb + O2 2PbO  
The output of the fuel cell is limited by the amount of oxygen in the cell at any  
one time, and the amount of stored anode material. In the absence of oxygen, there is  
no current generated.  
2.1.2 Sample System  
The sample system delivers gases to the Micro-Fuel Cell sensor from the ana-  
lyzer rear panel inlet. Depending on the mode of operation either a sample or zero gas  
is delivered.  
The Model 3160 sample system is designed and fabricated to ensure that the  
oxygen concentration of the gas is not altered as it travels through the sample system.  
The following list of sample system and analyzer features are among those  
available, but the exact configuration depends on the standard options incorporated.  
Paragraph 1.3 contains a list of the standard variations on the basic model.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-3  
2 Operational Theory  
Electropolished 316L-type stainless steel components.  
To eliminate oxygen absorption and desorption from the internal  
wetted surfaces of the sample system components, the sample system is  
fabricated from electropolished 316L-type stainless steel.  
Welding/Metal Gasket-Type Fittings.  
All of the joints upstream of the Micro-Fuel Cell oxygen sensor are  
orbitally welded. Orbital welding is used in the sample system  
wherever feasible. Orbital welding fuses the electropolished 316L  
stainless steel components together, forming a smooth, clean internal  
(wetted) weld junction and eliminating small spaces around the weld  
junction where gases can get trapped or absorbed. All of the weld  
junctions in the entire assembly are purged using an inert gas during  
welding to ensure that there is no oxygen contamination.  
Orbital welding is used where practical; otherwise, conventional  
precision welding is used. For example, conventional precision welding  
is used to fuse the tubes to the mounting blocks.  
Valves.  
The analyzer sampling system utilizes three different types of valves.  
Each valve is selected to prevent oxygen contamination of the sample  
depending on its position and purpose in the circuit.  
Air-Actuated Bellows Valves: These valves are normally closed in the  
sample system. They are used to control the delivery through the sample  
system of the sample or zero gas. The valve bodies are orbitally welded in  
the system and the valve bonnets make a metal-to-metal seal to the body.  
This valve system eliminates inboard and outboard gas leakage. The valves  
are activated (open/closed) by computer-controlled solenoid valves.  
Metering Valve: The metering valve is used to manually control the  
gas flow rate to the sensor. The body of the metering valve is orbitally  
welded, and the bonnet is sealed to the body with metal O-rings. This type of  
valve eliminates inboard and outboard gas leakage.  
Solenoid Valves: The solenoid valves control the air flow to the air-  
activated bellows valves. The solenoid valves are controlled by the  
microprocessor module. When de-energized, the valve outlet is open to  
ambient air, allowing the air-activated bellows valve to close.  
Scrubber (optional). See Appendix for details.  
In systems with a scrubber, oxygen-trapping media is used to remove  
any trace oxygen from the sample gas. This occurs automatically in the  
zero menu, where the pneumatic valves leading to the scrubber are  
triggered by a certain menu sequence. This process enables sample gas  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-4  
Operational Theory 2  
to be used to zero the analyzer, eliminating the need to switch external  
tanks.  
Overall Design.  
The design of the sample system minimizes the volume of dead space, which  
can retain residual gas from another route or previous mode of analysis.  
Figure 2-2: Model 3160 flow schematic.  
2.1.3 Electronics and Signal Processing  
The Model 3160 analyzer has an embedded microcomputer, which controls all  
signal processing, input/output, and display functions for the analyzer. System power  
is supplied from a power supply module designed to be compatible with any interna-  
tional power source.  
The microcomputer, a liquid crystal display (LCD), and all analog signal pro-  
cessing electronics are located inside a replaceable control module. A light emitting  
diode (LED) display is located outside the control module.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-5  
2 Operational Theory  
Electronics in the analyzer are grouped according to function:  
1. Analog in  
2. Analog out  
3. Digital circuit  
4. Power supply  
Analog signal processing is accomplished in two plug-in circuit cards operating  
under control of the microcomputer. All analog signals are converted to digital early  
in the processing cycle to minimize analog processing and assure maximum system  
accuracy, since digital processing is much more accurate than analog and immune to  
many parameters such as drift and aging.  
The initial processing and digitization of the signal from the Micro-Fuel Cell  
takes place on two circuit boards: the six-decade programmable printed circuit board  
(PCB), and the analog input board.  
The first step in the chain takes place on the six-decade programmable PCB,  
which is mounted directly on top of the cell block. The output of the sensor is a  
current that is linearly proportional to the oxygen concentration. At low concentra-  
tions, the current from the sensor is low, which makes the close physical proximity of  
the pre-amp circuitry to the Micro-Fuel Cell necessary.  
The sensor current is routed into a low-offset, low-bias current operational  
amplifier configured as a current-to-voltage converter. The signal is sent through two  
amplifiers to the analog input board. The pre-amp board provides sensor zeroing and a  
connection to the thermistor in the sensor cell block.  
The analog input PCB, which plugs into a DIN connector on the digital PCB,  
performs three functions:  
Temperature compensation in the sensor signal  
Sensor offset correction  
Digitization  
A thermistor inside the sensor block registers resistance and thus the temperature  
of the sensor. Accordingly, an analog-to-digital converter configured as a digitally-  
programmed attenuator (DPA) is used to reduce or increase the gain of the sensor  
signal, so that the signal coming out of the DPA is independent of temperature.  
If the software determines that the sensor needs offsetting, the sensor current is  
back-calculated and an analog-to-digital converter on the analog board is driven by the  
software to produce a voltage that is used to force a current into the current-to-voltage  
converter located on the pre-amp board. The current injected is equal and opposite in  
polarity to the portion of the sensor current that needs offsetting.  
Digitization is provided by a 12-bit analog-to-digital converter, which digitizes  
the oxygen and temperature signals.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-6  
Operational Theory 2  
The analog output printed circuit board (PCB) generates the two 01 volt and the  
two 420 mA analog signal outputs available on the rear panel of the analyzer. These  
signals, generated in digital format by the microcomputer, are converted into analog  
signals by the circuitry on this PCB. The output signals represent the following:  
01V Signal (Oxygen Measurement): This output goes from 0 to 1, represent-  
ing 0 to 100% of the scale that has been set; i.e., 0.6 volt is equal to 60% of the full  
scale, or 6 PPM when on the 10 PPM scale. It is possible that the signal may go past  
zero into the negative range up to -0.25, especially if the analyzer has been zeroed  
with a gas that contains a significant concentration of oxygen. See Figure 2-3.  
0-1V Range Identifier: This 0 to 1 volt output represents each range with a  
particular voltage as shown in Table 21.  
Isolated 420 mA Signal (Oxygen Measurement): This is a 4 to 20 mA output  
representing 0 to 100% of the scale, with 4 mA equal to 0%, and 20 mA equal to  
100% of that range. This output may also range lower than 4 mA, especially if the  
analyzer has been zeroed with a gas that contains a significant concentration of oxy-  
gen. See Figure 2-3.  
Isolated 420 mA Range Identifier: This 4 to 20 mA output represents  
individual ranges with discrete current output as shown in Table 2-1.  
Table 2-1. Range Identifier  
Identifier  
Voltage (V)  
0.0  
Identifier  
Current(mA)  
4.0  
Range  
1
Scale  
1 PPM  
2
3
4
5
6
10 PPM  
100 PPM  
1000 PPM  
1%  
0.2  
0.4  
0.6  
0.8  
1.0  
7.2  
10.4  
13.6  
16.8  
20.0  
25%  
Figure 2-3: Analog signal output offset  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-7  
2 Operational Theory  
The digital printed circuit board (PCB) is a general purpose microcomputer used  
to control all functions of the analyzer. The analog input PCB and the analog output  
PCB plug directly into connectors located on the digital PCB. In addition to control-  
ling these analog PCBs, the digital board performs the following functions:  
1. Processes input from the control panel pushbuttons.  
2. Provides signals for the selectable alarms.  
3. Processes serial I/O functions (RS-232 data). The serial interface default  
and programmable parameters are listed in Table 2-2.  
Table 2-2. Default and Programmable Parameters  
Defaults  
1200 Baud  
8 Bits  
Programmable Options  
1200 Baud  
5, 6, 7, or 8 Bits  
No Parity  
1 Stop Bit  
Even, Odd, or No Parity  
1 or 2 Stop Bits  
4. Controls the LCD and the LED displays.  
LCD: This screen is a dot-matrix display located on the control module of  
the analyzer. It is the user interface for system operations. It  
displays the menus and command options available for each  
function.  
LED: This screen is a 7-segment display located on the front panel of the  
analyzer, above the control module. It displays only the oxygen  
concentration. It is large and bright to allow the operator to read it  
at a greater distance. A dimmer switch for this display is located on  
the display PCB behind the front panel.  
The analyzer power supply module is a replaceable assembly containing four  
power supplies and five alarm relays. Electronic circuitry used to drive and interface  
the alarm relays to the output of the microcomputer is also located inside this module.  
Note: This power supply contains an International Power Entry Module. This  
feature allows operation on any of four international voltage ranges:  
100V, 120V, 220V or 240V (50Hz or 60Hz). It also facilitates both North  
American and European fusing arrangements. Programming this mod-  
ule is described in the installation section of this manual.  
TELEDYNE ANALYTICAL INSTRUMENTS  
2-8  
Installation 3  
Installation  
Installation of the analyzer includes:  
1. Unpacking the system.  
2. Recognizing the necessary precautions when installing the system.  
3. Hooking up the sample/span gas and air supply to appropriate connections.  
4. Installing the Micro-Fuel Cell sensor(s).  
5. Hooking up electrical connections.  
6. Testing the system.  
3.1 Unpacking the Analyzer  
The analyzer is shipped with all the materials and special items you need to install  
and prepare the system for operation. Carefully unpack the analyzer and inspect it for  
damage. Immediately report any damage to the shipping agent. Remove the packing  
slip and verify that you have received all the components listed in Table 3-1.  
Table 3-1. Accessory Kit for Model 3160  
Qty  
3
Part No.  
G285  
Description  
Weld glands, VCR-type  
3
G284  
Gaskets, VCR-type  
3
2
1
N284  
W64  
M52973  
Nuts, female (Hoke 4NM316)  
Wrench, Open End, 3/4"5/8"  
Instruction Manual  
3.2 Front Panel  
The Model 3160 front and rear panels are illustrated in Figure 31 on the follow-  
ing page. The front panel contains the following displays and controls:  
1. Main power switch.  
2. Light Emitting Diode (LED) display: the O2 concentration is displayed in  
bold lettering.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-1  
3 Installation  
3. Parts Per Million (PPM) and Percent Oxygen indicators: the PPM light will  
light while the instrument is measuring O2 in units of PPM. The percent  
light will light while the instrument is measuring O2 concentration in  
percent.  
4. Flow Indicator and Flow Set: a flowmeter and flow set knob are provided  
for adjusting gas flow in standard cubic feet per hour (SCFH).  
5. Liquid Crystal Display (LCD) with keypad: the LCD shows system menus  
and data during instrument operation. The keypad is the primary user-input  
device used to enter information in the system.  
3.3 Rear Panel  
Figure 31 also shows the rear panel of the 3160. Located on the rear panel are  
the following electrical power and gas input ports, alarm relay outputs, and analog and  
digital outputs:  
1. Gas Input/Vent: three input fittings are provided for span gas, sample  
gas, and compressed air for pneumatic valve operation. A single vent is used  
for the sample and span gas outlet. Consult the Appendix for scrubber selec-  
tion.  
2. AC Power Input: 110, 120, 220 or 240 V ~ at 50/60 Hz 1.5 A MAX.  
Use 250 V 1.6 A T Fuse for 110, 120 V ~  
Use 250 V 0.8 A T Fuse for 220, 240 V ~  
3. Alarm Circuit Connections: there are five contact closures/openings  
provided for external alarms. The alarm functions are defined by the user via  
keypad input within the LCD menu system.  
4. Analog Outputs: four analog output connectors are provided for use  
with a chart recorder. Two provide range and data in voltages; the other two  
provide range and data for current-driven recorders.  
5. RS-232 Serial Port: a 9pin digital input/output connector is provided  
for connecting either a serial printer or a serial link to an external computer.  
Optional serial link software (such as TRACS) can be used with the instru-  
ment for remote external computer monitoring and control of the instrument,  
via modem or hardware link.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-2  
Installation 3  
In  
mm  
DIMENSIONS:  
REMOVABLE  
TOP COVER  
LED  
DISPLAY  
17.406  
(442)  
POWER  
SWITCH  
19.000  
(483)  
PPM  
INDICATOR  
I
8.000  
(203)  
O
PERCENT  
INDICATOR  
5.750  
(146)  
LIQUID  
CRYSTAL  
DISPLAY  
(LCD)  
18.000  
(457)  
COMPUTER  
MODULE  
1.50  
(38)  
KEYPAD  
FLOWMETER  
FLOW  
SET  
KNOB  
SENSOR  
ACCESS  
PANEL  
ANALOG  
RS-232  
SERIAL PORT  
OUTPUT  
CONNECTIONS  
8.688  
(221)  
SPAN  
GAS IN  
INSTRUMENT  
AIR  
SAMPLE IN  
AC POWER  
VOLTAGE SELECT  
ALARM  
CIRCUIT  
CONNECTIONS  
SAMPLE/SPAN  
VENT  
Figure 3–1: Analyzer front and rear panels.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-3  
3 Installation  
3.4 Gas Line Connections  
All of the gas lines to the system hook up at the back of the unit (see Figure 31).  
All of the fittings on the removable back panel are ¼ " male VCR type fittings, with the  
exception of the compressed air inlet fitting. Use the wrenches provided to connect the  
gas lines. Insert the gasket between the fitting and tighten the female and male nuts  
until finger-tight; then, by holding the male nut with the wrench, tighten the female nut  
with the second wrench by ¼ turn.  
CAUTION: Do not put any torque on the tubes welded on the sampling  
system.  
Check that each of the lines is hooked up to the correct connection. The lines  
should be connected in the following order (from left to right):  
3.4.1. Span Gas In  
Provide at least one sample gas of a known oxygen concentration. Using  
70-99 % of the range just one range above the range of interest is recommended. Any  
range EXCEPT 0-1 ppm may be used to calibrate.  
3.4.2. Instrument Air (Compressed Air Fitting)  
The gas pressure (7080 psi) needed to operate the pneumatic valves in the  
analyzer can be supplied through this fitting.  
CAUTION: Pressure higher than 80 psig can damage the solenoid  
valves.  
3.4.3. Sample Gas In  
Hook up the sample gas to the sample gas inlet. When the sample gas flows  
through the optional oxygen scrubber, the sample gas becomes the zero gas and can be  
used for calibration.  
Note: Sample and span gas pressure should be between 5 and 10 psig and  
within ±2 psi of each other. A substantial difference in the span and  
sample gas pressures will cause a sudden change in gas flow rate.  
3.4.4. Vent  
The vent transports the sample or span gas out of the system after exposure to  
the Micro-Fuel Cell sensor.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-4  
Installation 3  
3.5 Micro-Fuel Cell Sensor Installation or Replacement  
The 3160 comes with the Micro-Fuel Cell(s) already installed. When installing or  
replacing the cells, remember to inspect the replacement sensor for leaks or damage. See  
Figure 3-2.  
Figure 3-2: Sensor Installation.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-5  
3 Installation  
1. The sensor compartment is located in the lower right-hand corner of the  
analyzer front panel. To remove the sensor compartment panel, loosen the  
thumbscrews located left and right center of the panel.  
2. Make sure that sample gas is flowing through the analyzer, then shut off  
the flow completely. This insures that air does not diffuse into the analyzer.  
3. Inside the compartment should be two (one) stainless steel cylindrical  
sensor block(s), each with a clamp lever across the front. The bottom half  
of the clamp face is bent outward. To release the clamp, pull the clamp  
lever up and toward you slowly.  
4. Releasing the clamp releases the stainless steel block cap on the bottom of  
the sensor block. Pull the cap downward to remove it. There is a guide pin  
to the rear of the sensor block that aligns the cap to the block.  
Note: It is important to minimize the amount of time that the new cell is ex-  
posed to air in order to reduce the time required for the reading to drop  
to zero.  
5. Remove the Micro-Fuel Cell from the plastic bag. The cell has a mesh  
screen with a raised plastic dress ring on one side and flat gold contact  
rings on the other. Place the screen side of the cell face down on the block  
cap so that the dress ring locks onto the cap ridges and will not move side  
to side. The gold contact rings will be facing up.  
6. Carefully guide the sensor and cap into the block, putting the guide pin  
through the hole in the back edge of the cap.  
7. While holding the cap in place, squeeze the clamp down firmly until the  
notches on the clamp lock onto the block side pins.  
8. Immediately start the sample flow, and set to about 2 SCFH. The analyzer  
now needs to be zeroed and calibrated.  
3.6 Electrical Connections  
All of the electrical connections are located on the analyzer back panel. The  
analyzer is shipped with all of the electrical connections intact, with no assembly or  
installation required. The power cord receptacle is located in the lower center of the  
panel. The voltage selection terminal and the fuse block are in the same fitting directly  
above the power cord receptacle. There are four output signal connectors with screw  
terminals in the upper right-hand panel. There are two wires per output with the polari-  
ty noted below each. The five alarm circuit connectors, located in the lower center  
right-hand panel, are screw terminals for alarm relay contacts. These five provide a set  
of Form C contacts for the user. The contacts are capable of switching up to 3 am-  
peres at 115 V ac into a resistive load.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-6  
Installation 3  
Figure 3-3: 3160 Rear Panel.  
3.6.1 Voltage Selection  
WARNING:  
Power cord must be disconnected before performing any voltage  
selection!  
The voltage setting and fuses of the analyzer can be changed to international  
standards. To change the voltage setting or fuses:  
1. Open the cover using a small blade screwdriver.  
2. Remove the cover and the fuse block assembly.  
3. Pull the voltage selector card straight out of the housing by pulling on the  
indicator pin (see Figure 34).  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-7  
3 Installation  
Figure 3-4: Removing the Voltage Card  
4. Turn the card so that the desired voltage can be read at the bottom.  
5. Slide the indicator pin around the card so that it is pointing up when the  
voltage is read at the bottom (see Figure 35).  
Figure 35: Voltage Selection  
6. Place the voltage selector card back into the housing. The edge where the  
desired voltage is printed should go in first and the printed side of the card  
should be facing the IEC connector.  
7. Replace the fuse block and the cover. Verify that the indicator pin is  
pointing to the correct voltage.  
3.6.2 Fuse Changing  
NOTE:Spare fuses are located in a clip attached to the power supply enclosure  
inside of the analyzer.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-8  
Installation 3  
WARNING:  
Power cord must be disconnected before performing any  
voltage selection!  
To change from North American to European fuses, perform the following:  
1. Use a small blade screwdriver to open the cover.  
2. Loosen the Phillips screw two turns and remove the fuse block by sliding it  
away from the screw and up from the pedestal.  
3. Remove the North American fuse and turn the fuse block over.  
4. Replace the fuse with European fuses. Two European fuses are required,  
but the lower one may be replaced with a dummy fuse, or jumper bar (see  
Figure 36).  
European 1.6 Amp T Fuses  
5 x 20 mm  
North American 3 A T Fuses  
Figure 36: Fuse Replacement  
5. Slide the fuse block back on the Phillips screw and pedestal.  
6. Tighten the Phillips screw and replace the cover.  
NOTE:The fuses that go into the housing first are the active set.  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-9  
3 Installation  
3.6.3 RS-232 Serial Digital Port  
The RS-232 port is configured as data terminal equipment (DTE) and uses a 9-pin  
D connector wired as follows:  
1. No connection  
2. Transmit Data  
3. Receive Data  
4. N/C  
5. Ground  
6. N/C  
7. RTC  
8. CTS  
9. N/C  
Connecting the analyzer to a modem requires a PC to modem cable. When con-  
necting the analyzer directly to a PC, terminal or other data terminal equipment, a  
null-modemcable is required.  
The following serial interface default parameters are used:  
Defaults  
1200 Baud  
8 Bits  
Programmable Options  
1200 Baud  
5, 6, 7, or 8 Bits  
Even, Odd, or None  
1 or 2 Stop Bits  
No Parity  
1 Stop Bit  
3.7 Installation Checklist  
Have you:  
Checked for leaks from any of the rear panel connections?  
Checked the sample and span gas pressures (510 psi)?  
Checked the instrument air pressure (7080 psi)?  
TELEDYNE ANALYTICAL INSTRUMENTS  
3-10  
Operations 4  
Operations  
4.1 Front Panel Controls  
The front panel of the analyzer, shown in Figure 4-1, contains indicators and displays  
through which the computer module can be accessed. The upper left-hand side of the panel  
has an LED screen that displays the oxygen content of the sample in one-inch high  
numerals. This display can be brightened or dimmed by a potentiometer located directly  
behind the panel inside of the analyzer case. To access it, remove the top cover of the  
analyzer.  
Figure 4.1: Analyzer front panel.  
Below the LED screen are two red LED lights. Each will light to indicate whether the  
oxygen is being displayed in parts per million (PPM) or percent (%).  
To the right of the LED screen is a flowmeter in standard cubic feet per hour  
(SCFH). The flow set knob adjusts the flow rate of the gas during calibration and zeroing.  
In the panel below the LED screen is the LCD display. The five colored buttons  
below it are used to interface with the computer module in selecting modes, with the LCD  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-1  
4 Operations  
displaying the function of each button directly above it. The LCD screen displays the  
current mode and any warning messages, instructions and button functions.  
4.1.1 Spinning Wheel  
A spinning wheelappears in the upper left-hand corner of the LCD display to  
indicate that the alarms are enabled (any that are not defeated), or that scheduled zeroing  
(AutoZeroing) or scheduled spanning (AutoSpanning) may take place. If the spinning  
wheel does not appear, the alarms are disabled, and a scheduled zero or span is defeated.  
Note that there is a delay of several minutes before the spinning wheel appears whenever  
the Analyze mode is re-entered.  
4.1.2 Cell Output Factor (Span Factor)  
The expected life of the Micro-Fuel Cell sensor is about eight months. A guide to the  
relative life left in the cell can be found on status page 2 (see System Statistics near the end  
of this chapter) or retrieved from the serial port through the serial command SF. The span  
factor ranges from 0.00 to 1.00, with 1.00 representing full life expectancy. A span factor  
below 0.1 indicates that the Micro-Fuel Cell needs replacing.  
4.2 Modes of Operation  
To use the system and select displayed options, press the button directly underneath  
the option you wish to select. There are eleven different menus:  
Cold Start-Up: During the first startup, and when subsequent cold start-  
ups are chosen by the user, initial values set at the factory for alarms, I/O,  
calibration, zeroing and other data are used.  
Upon cold start, zeroing and calibration should be performed before  
accurate oxygen measurements are obtained. A system warm start is automati-  
cally performed on each subsequent power-up. For a warm start, previous  
user-input configuration data is preserved in RAM by battery-preserved  
memory in the system control module.  
Calibration Zeroing: For the highest possible accuracy, the analyzer  
must be zeroed using an oxygen-free gas. By eliminating normal background  
noisefrom the sensor reading, zeroing resets the level referred to as the  
zerooxygen concentration.  
If the analyzer is equipped with a scrubber, it is important to zero the  
analyzer with a near-zero gas connected to the sample port. The zeroing  
process automatically opens pneumatic valves on either side of the scrubber,  
and a gas with a high concentration of oxygen will react with the scrubber and  
necessitate replacement sooner.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-2  
Operations 4  
When waiting for manual sensor settling during calibration zeroing, wait  
until the oxygen concentration has reached its lowest value and has remained  
so for 1015 minutes. It is important to make sure that the sensor is reading the  
lowest possible value, since zeroing will reset the point at which the sensor  
reads zero. Do not zero prior to reaching the lowest possible oxygen concentra-  
tion. When initially installing the sensor(s), do a zero calibration, then a span  
calibration. If desired, the instrument may be programmed for automatic  
periodic zeroing (AutoZeroing) if the optional scrubber is installed. AutoZero-  
ing uses the scrubber to generate a zero gas from a low O2 sample gas.  
NOTE: 1. Zeroing in the 0-1 PPM range preserves the life of the scrubber.  
In scheduled zeroing and spanning (AutoZeroing and AutoSpanning), a  
time interval must be chosen. This interval determines how often the instru-  
ment zeros or spans itself. To change the scheduled time, edit it after setting  
the interval in the Set Up Clock Functions mode.  
NOTE: 2. AutoZeroing and AutoSpanning may be delayed until the alarms have been  
enabled.  
Calibration Using Span Gas: After zeroing, the instrument can now  
be calibrated using a span gas with a known oxygen concentration. The analyz-  
er may be calibrated on any range other than 01 PPM, but it is best to cali-  
brate one range up from the range where the sample is expected to be. The  
span gas concentration should be within 70% to 99% of full scale of the range  
selected.  
NOTE: 3. Prior to calibration, allow the analyzer reading to come to a stable value with  
zero or sample gas flowing through the analyzer.  
Select Active Sensor (Two cell blocks only): One or two O2 sensor  
blocks can be installed in the system. In dual sensor systems, either sensor can  
be isolated with pneumatic valves, and the other sensor chosen as the active  
sensor for O2 measurements. The system software is aware of how many  
sensor blocks are present, but will not automatically switch sensors should one  
fail.  
This menu will not appear if only one cell block is installed.  
Install Sensor and Test Alternate Sensor (Two cell blocks only): Pro-  
cedures are provided for installing a new sensor and for testing the alternate O2  
sensor.  
This menu will not appear if only one cell block is installed.  
O2 Range Set-Up: The measurement range may be manually chosen, or  
the system may be set to automatically choose the range (AutoRanging). In the  
AutoRanging mode, the range will step up at greater than 100% and step  
down at 80% of full scale of the current range. The analyzer is programmed to  
prevent oscillation between ranges during rapidly changing oxygen levels.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-3  
4 Operations  
Set-Up Alarms: Each of five alarms provided may be programmed to  
actuate at any O2 trace level above or below a threshold. The fifth alarm may  
also be programmed as a system alarm to monitor several system-level condi-  
tions. The system alarm can be triggered by the following system operations:  
Scheduled zero (AutoZero) failure  
Scheduled span (AutoSpan) failure  
Battery failure  
Internal errors  
Any condition triggering the system alarm can be determined from status  
page 4 (see this chapter: System Statistics).  
When the alarms are disabled, scheduled zeroing or spanning will not  
take place.  
Logger Set-Up: A system data logger keeps a record of past oxygen  
measurements. The logger is presented in the form of a graphed chart on the  
system LCD; gaps in time will not display, and all points shown on a single  
chart have the same range and sensor number.  
The logger will not record in the Test Alternate Sensor mode.  
Set Up Clock Functions: An internal clock provides the time and date  
for events scheduled by the user. The internal clock is also used by the logger  
to record events scheduled at timed intervals.  
Scheduled zeroing and spanning (AutoZeroing and AutoSpanning) are set  
in this mode. The mode uses an electronic timer to set off the zeroing or  
spanning sequence at pre-selected intervals. During an AutoZero or AutoSpan,  
the alarms are defeated and the screen displays PLEASE WAIT. SCHED-  
ULED SPAN IN PROGRESS.”  
Since AutoZeroing and AutoSpanning uses automatic settling detection  
of the sensor, which can take up to fifteen minutes, oxygen monitoring will be  
interrupted for the length of time it takes to zero or span, after which monitor-  
ing will resume. The LCD will display the current background level during  
zeroing or spanning.  
Changing Passwords For Remote Monitoring and Control:  
Programmable security passwords are provided to prevent unauthorized access  
to remote monitoring and control. Monitoris the lowest authorization level  
and only permits monitoring of analyzer functions, while Controlallows  
remote control of any function.  
System Statistics: Menu screens are available to view various system  
parameters, and for setting serial link operation parameters, including modem  
operation.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-4  
Operations 4  
Cold Start-Up  
1. A cold start will automatically occur the first time you  
start the analyzer. In any other situation, you can cold  
start by turning the analyzer off and holding down the  
red key below the LCD screen while you turn the power  
back on. Continue to hold the red key until the screen  
comes on. There will be a two minute wait for system  
initialization before the second screen.  
A warm start will take you directly to step 5.  
2. Set the time using the SELEC, UP and DOWN keys,  
and press ACCPT.  
3. Select sensor A or B as the active sensor.  
(This menu is omitted if only one cell block is installed.)  
4. If you have not done so, connect the sample gas to the  
sample port and press ACK.  
5. Using the valve and meter on the sample gas supply, set  
the flow rate to 2 SCFH and press ACK.  
If you have warm-started, this is the second screen you  
will see.  
6. The main Analyze mode menu will appear, and the  
measured O2 level of the sample gas will display. The  
system should first be zeroed, then spanned, to obtain  
the most accurate O2 measurements.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-5  
4 Operations  
Calibration Zeroing  
1. At the main Analyze mode screen press ZERO.  
2. Press ACK to acknowledge that a low O2 gas has al-  
ready been connected to the sample port. The sample  
gas O2 level should be close to zero to preserve the life  
of the scrubber (if one is installed).  
3. Set the flow rate to 2 SCFH using the knob on the sys-  
tem front panel and press ACK.  
4. Press AUTO to let the system determine the settling  
time before spanning or press MAN for manual.  
5. If MAN was chosen in step 4, watch the logger or the  
O2 value on the screen to assess when the output is con-  
stant, and then press ACK.  
6. After step 5, the system will automatically span to the  
user-entered span value. If the zeroing is successful, the  
main Analyze mode menu will appear as in step 10.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-6  
Operations 4  
7. If the O2 cell current is too large to be zeroed, an error  
message will appear. Press ACK to retry or ESC to  
abort the zeroing process.  
8. If AUTO was chosen in step 4, wait for the O2 level to  
settle before the system zeros. This may take from thirty  
(30) seconds to fifteen (15) minutes, depending on the  
sample gas oxygen level.  
9. After a stable low oxygen level is detected by the sys-  
tem, the system will automatically zero.  
If the O2 cell current is too large to be zeroed, an error  
message will appear (see step 7). You will then have to  
press ESC to abort, or ACK to retry.  
10. If the zero is successful, the system will return to the  
main Analyzer mode menu.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-7  
4 Operations  
Calibration Using Span Gas  
1. Enter the Calibration mode from the Analyze mode by  
pressing SPAN.  
2. Press ACK if zeroing has already been done. Make sure  
to zero before calibrating to ensure correct O2 readings.  
3. If the span gas tank is not connected to the span port,  
connect it now and press ACK.  
4. Enter the span gas concentration using the UP and  
DOWN keys, and press ACCPT.  
5. Set the flow rate to 2 SCFH and press ACK.  
6. Select manual or automatic sensor settling detection by  
pressing MAN or AUTO. Automatic sensor settling de-  
tection lets the system determine when the sensor has  
stabilized.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-8  
Operations 4  
7. If AUTO was chosen in step 6, wait for the O2 level to  
settle before the system spans. This may take from fif-  
teen (15) to thirty (30) minutes or longer, depending on  
the sample gas oxygen level.  
Go to step 11 if MAN was chosen in step 6.  
8. If any key was pressed in step 7, the screen indicates  
that the span has been aborted. Press ACK to retry  
spanning, or ESC to abort the span.  
9. After the system has automatically detected sensor sta-  
bility, the system will automatically span. If the system  
spans correctly, the next screen will be the Analyze  
mode main menu (see step 14).  
10. If the O2 cell current is too strong or too weak to be  
spanned, an error message will appear. Insure that the  
span gas concentration has been entered correctly. Press  
ACK to retry, or ESC to abort the span.  
11. If MAN was pressed in step 6, watch the logger or the  
O2 value on the screen to determine when the output is  
constant, and then press ACK.  
12. After step 11, the system will automatically span to the  
user-entered span value.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-9  
4 Operations  
13. If the O2 cell current is too strong or too weak to span,  
an error message will appear. This will not occur if the  
user-entered span gas concentration is correct. Press  
ACK to retry, or ESC to abort the span.  
14. If the span is successful, the system will return the main  
Analyze mode menu. Check to see that the span value is  
correct.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-10  
Operations 4  
Select Active Sensor (Two Cells Only)  
1. Enter set-up from the Analyze mode main menu by  
pressing SETUP.  
If only one cell block is installed, the TEST and INSTL  
options on the next screen are omitted.  
2. Press SET to change the active sensor.  
3. Enter your authorization code by using the SELEC, UP  
and DOWN keys, and press ACCPT.  
4. Press SENSR. (This option will not appear if only one  
sensor block is installed.)  
5. Select sensor A or B.  
6. The main Analyze mode menu will appear, and will indi-  
cate the chosen sensor.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-11  
4 Operations  
Install Sensor andTest Alternate Sensor (Two Cells Only)  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
If only one cell is installed, the TEST and INSTL op-  
tions on the next screen are omitted.  
2. Press TEST to test the alternate sensor or INSTL to in-  
stall a new sensor.  
3. If TEST was chosen in step 2, the active sensor is dis-  
played and you are directed to press ACK to test the al-  
ternate sensor.  
If INSTL was chosen in step 2, go to step 5.  
4. After pressing ACK in step 3, the alternate sensor is  
flushed for 10 minutes while the measured O2 level is  
displayed. The 10 minute flush may be restarted by  
pressing RESET, or aborted by pressing SKIP. Go to  
step 8.  
5. If INSTL was chosen in step 3, the screen will display  
the active sensor. Press A or B for the sensor you are  
installing.  
6. Install the sensor (either A or B from step 5) into the  
sensor block within two minutes, and press ACK.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-12  
Operations 4  
7. After pressing ACK in step 6, the new sensor is flushed  
for 10 minutes while the measured O2 level is displayed.  
The 10-minute flush may be restarted by pressing RE-  
SET, or aborted by pressing SKIP.  
8. When sensor installation or alternate sensor test are  
complete, the system will return to the main Analyze  
mode menu.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-13  
4 Operations  
O2 Range Set-Up  
1. Enter Range Set-Up mode from the Analyze mode by  
pressing RANGE.  
2. Press AUTO for AutoRanging, or FIXED to manually  
fix the O2 measurement range. In AutoRange mode, the  
range will be automatically determined so that the O2  
measurements are within a range which maximizes the  
accuracy of the reading.  
3. If AUTO is chosen in step 2, the main Analyze mode  
menu will appear, and will indicate the automatically de-  
termined range (see screen for step 1).  
If FIXED is chosen in step 2, use the UP and DOWN  
keys to choose the desired manual range, and press  
ACCPT.  
4. If FIXED was chosen in step 2 and the range was speci-  
fied in step 3, the main Analyze mode screen will ap-  
pear, and will indicate the new manually set range.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-14  
Operations 4  
Set-Up Alarms  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
2. Press SET to set the alarms.  
3. Enter the authorization code by using the SELEC, UP  
and DOWN keys, and press ACCPT.  
If this is a COLD START, the code you enter will be  
the code set for every subsequent start-up until the next  
COLD START.  
If this is a WARM START, enter the code used during  
the last COLD START.  
4. Press ALRMS.  
5. Press  
AL1 for alarm 1  
AL2 for alarm 2  
MORE for alarms 35  
ESC twice to return to the main menu  
Note that alarm 5 may be programmed as a system  
alarm.  
6. If AL1 or AL2 was chosen in step 5, use the SELEC,  
UP and DOWN keys to configure alarm 1 or 2, and  
press ACCPT, which will take you back to step 5.  
LEVEL is the O2 concentration for the alarm  
threshold.  
HIGH/LOW determines whether the alarm con-  
dition is above or below the alarm threshold.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-15  
4 Operations  
LATCHING actuates the alarm above a certain  
setpoint even if the level falls back below the  
setpoint.  
RELAY indicates the relay (on the rear panel)  
actuated by the alarm, which is changeable.  
FAILSAFE=YES enables the relay which actu-  
ates the alarm during system failure.  
DEFEAT=NO actuates the alarms for normal  
use.  
7. If MORE was pressed in step 5, choose alarms 3, 4, or 5  
by pressing AL3, AL4, or AL5.  
Choose AL5 to set-up a system alarm.  
8. If AL5 was pressed in step 7, choose alarm 5 to be an  
O2 or system alarm by pressing O2 or SYSTEM.  
Configured as a system alarm, alarm 5 will ring when a  
scheduled calibrator zeroing or a scheduled span calibra-  
tion fails, when there is an internal system failure, or  
when the RAM back-up battery fails.  
9. If SYSTM was pressed in step 8, configure the system  
alarm using the SELEC, UP and DOWN keys.  
LATCHING=YES actuates alarm 5 above a cer-  
tain setpoint even if the level falls back below the  
setpoint.  
FAILSAFE=YES enables the relay which actu-  
ates the alarm during system failure.  
DEFEAT=YES actuates the alarm for normal  
use.  
Press ACCPT to accept the new system alarm configu-  
ration. The next screen will ask you if there are more  
alarms to set. If not, pressing ESC three times will take  
you back to the Analyze mode main menu.  
10. If O2 was pressed in step 8, the screen will resemble the  
one in step 6. Use the SELEC, UP and DOWN keys to  
configure alarm 5. When finished, press ACCPT. The  
next screen will ask you if there are more alarms to set.  
If not, pressing ESC three times will take you back to  
the Analyze mode main menu.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-16  
Operations 4  
Logger Set-Up  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
2. Press SET to set up the logger.  
3. Enter the authorization code by using the SELEC, UP  
and DOWN keys. Then press ACCPT.  
4. Press LOG for the logger functions. Press ESC to return  
to the Analyze mode main menu.  
5. Press CLEAR if you want to clear the current logger  
data. Press SETUP to continue setting up the logger.  
6. If you chose SETUP in step 5, use SELEC to choose  
logger ON/OFF, the time period between samples, or  
the mode of measurement. Then choose EDIT to  
change the values. After you are finished, press ACCPT,  
which will take you to step 4.  
The choices for the logger chart sampling mode (MODE  
=) are on the next page.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-17  
4 Operations  
Note: The above screens use T to indicate two values: seconds and minutes. The first value is  
the time period between samples on the logger, and is programmable. The second time  
value is the width of the logger chart page, and is not programmable. There are 64  
samples per logger chart page, and over 20 pages of past data which may be viewed.  
Mode determines how to calculate the sample values shown on the logger chart:  
Average:  
Median:  
The average of all measurements in the time period between samples.  
The average of the minimum and the maximum values measured in the time  
period between samples.  
Maximum: The maximum of all values measured in the time period between samples.  
Minimum: The minimum of all values measured in the time period between samples.  
Sample:  
The last sample measured in the time period between samples.  
Press EDIT in step 6 and then use the UP and DOWN keys to select the above choices for the  
logger chart mode.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-18  
Operations 4  
Set-Up Clock Functions  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
2. Press SET to set the clock functions.  
(If only one cell block is installed, TEST and INSTL  
will be omitted.)  
3. Enter the authorization code by using the SELEC, UP  
and DOWN keys, and press ACCPT.  
4. Press MORE.  
5. Press CLOCK to select clock functions.  
6. Enter the function that you would like to change.  
SET: Sets the time.  
ZERO:Sets the time for the next zero and the inter-  
val times for automatic zeroing.  
SPAN: Sets the time for the next span and interval  
times for automatic spanning.  
ALRMS: Displays alarm times and dates.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-19  
4 Operations  
7. If SET was pressed in step 1, the menu for setting the  
time and date will appear. Use the SELEC, UP and  
DOWN keys to set the time and date, and then press  
ACCPT.  
8. If ZERO was pressed in step 1, the menu for setting  
AutoZeroing ON/OFF, time/date of the next zeroing,  
and time interval for automatic zeroing will appear.  
9. Use the SELEC, UP and DOWN keys to make selec-  
tions, and press ACCPT.  
10. If SPAN was pressed in step 1, the menu for setting  
AutoSpanning ON/OFF, time/date for the next calibra-  
tion, and interval for automatic calibration will appear.  
11. Use the SELEC, UP and DOWN keys to make selec-  
tions, and press ACCPT.  
12. If ALRMS was pressed in step 1, a menu displaying the  
most recent triggering time and date of each of the  
alarms will appear. From here, press ESC three times to  
return to the main menu screen.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-20  
Operations 4  
Changing Passwords for Remote Monitoring and Control  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
2. Press SET to set the password(s).  
3. Enter your authorization code by using the SELEC, UP  
and DOWN keys, and press ACCPT. You must enter  
the same authorization code set during the last COLD  
START.  
4. Press MORE. Press ESC to return to the Analyze mode  
mainmenu.  
5. Press PASS1 to change the remote monitoring pass-  
word, or PASS2 to change the remote control pass-  
word.  
6. If you pressed PASS1 in step 5, the next screen will al-  
low you to change the remote monitoring password.  
Use SELEC, UP and DOWN to choose the new pass-  
word, and then press ACCPT, which will take you to  
step 4.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-21  
4 Operations  
7. If you pressed PASS2 in step 5, the next screen will al-  
low you to change the remote control password. Use  
SELEC, UP and DOWN to choose the new password,  
and press ACCPT, which will take you to step 4.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-22  
Operations 4  
System Statistics  
1. Enter set-up from the Analyze mode by pressing  
SETUP.  
ESC in any status page will take you to the last set-up  
screen and eventually the main analyze mode.  
PREV steps you back one status page.  
NEXT moves forward one status page.  
2. Press STAT for system status.  
(If only one cell block is installed, TEST and INSTL  
will be omitted.)  
3. Status page one displays the active sensor (A or B), the  
system time/date, and the system hardware configura-  
tion. The number and date of the installed software ver-  
sion are also shown.  
Press PREV for the previous screen, or NEXT for the  
next screen.  
4. Status page two shows the O2 level set for spanning, the  
cell strength factor, the type and time of the last span,  
the time interval for scheduled spanning (AutoSpan),  
and whether or not scheduled spanning is active. [The  
cell strength factor is an indication of the amount of cell  
life left in the cell, with 1.00 being the most and 0.00 be-  
ing the least.]  
5. Status page three shows the time of the last zero, the  
time interval for scheduled zeroing (AutoZero), and  
whether or not scheduled zeroing is active.  
6. Status page four shows the settings for the five pro-  
grammablealarms.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-23  
4 Operations  
7. Status page five shows the time and date of the last time  
the alarm was triggered for each of the five program-  
mablealarms.  
8. Status page six shows the set-up for the serial port, and  
allows you to reconfigure the serial port. Press EDIT to  
bring up the following screen.  
Press NEXT to go to status page 7.  
9. Use the SELEC, UP and DOWN keys to select the de-  
sired serial port configuration, and press ACCPT.  
Baud rate: 3002400  
Parity:Odd, Even, or None  
Mode: Monitor, Remote, or Off  
ASCII Data Bits:7 or 8  
Stop Bits: 1 or 2  
Modem: On, Off or Auto  
Output Interval for Monitoring: 132000 sec.  
Transmit/Receive Line Terminators: CR or  
CRLF  
10. Status page seven allows for configuration of the mo-  
dem/remote. Press EDIT to change the configuration.  
Modem Mode: AUTO/OFF/ON  
Modem Status: NO (set by system)  
Login Access: LOCAL (set by system)  
Phone Jack: SINGLE/MULTI, RJ12  
Comm. Std. for 1200B: BELL USA/CCITT  
11. After pressing EDIT, a cursor will appear and you will  
be able to edit the settings. Press ACCPT when done.  
Pressing INIT in status page seven (step 10) takes you  
to the screen at the top of the next page and gives you  
the option of either checking for a modem, or setting up  
the modem. Set-up sets up the modem for auto answer  
and other communication protocols.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-24  
Operations 4  
12. After checking for the modem, pressing ESC returns  
you to status page seven in step 10.  
If INIT was pressed in status page 7 and then SETUP  
pressed in step 11, pressing ACK will return you to this  
screen. Press ESC to return to status page seven in step  
10.  
13. If you have purchased the parallel printer option, status  
page eight allows for configuration of the port. Press  
EDIT to change the configuration.  
Mode: ON/OFF  
Interval: 332000 seconds  
NOTE: The printer report will always be in the same for-  
matO2 Level, Range, Active Sensor, Time & Date.  
14. After pressing EDIT, a cursor will appear and you will  
be able to edit the settings.  
15. Status page nine shows whether or not the last and next  
start-ups were WARM or COLD. The next start-up may  
be changed by pressing the COLD or WARM keys.  
16. Status page ten displays logger data. Press VIEW to see  
the logger data history. Press NEXT to return to status  
page one. Press ESC twice to return to the Analyze  
mode main menu.  
17. If VIEW was pressed in step 13, logger data may be  
viewed by using the +, , REV and FWD keys. The O2  
level, range, and time are shown for each data point.  
NEXT takes you to status page two.  
REV and FWD change the date by one page (or 64  
points), while and + go through the data one point at a  
time within a single page. Press ESC to return to status  
page ten.  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-25  
4 Operations  
TELEDYNE ANALYTICAL INSTRUMENTS  
4-26  
Maintenance & Troubleshooting 5  
Maintenance & Troubleshooting  
5.1 Routine Maintenance  
1. Calibrate the analyzer at least once each week during the first four weeks  
and then once every two weeks during the next eight weeks. Afterwards,  
the analyzer should be calibrated once a month.  
2. Check the pressure of the air supply to the analyzer. Air pressure should be  
maintained between 70 and 80 psig.  
Note: The instrument air pressure must be greater than 70 psig in order to open  
and close the pneumatic solenoid valves.  
3. Check the sample pressure. Pressure should be maintained between 5 and  
10 psig. Sample and span gas pressures should be within 2 psig of each  
other.  
5.1.1 Sensor Maintenance  
Sensor maintenance in the 3160 is limited to replacing the sensor(s) when replace-  
ment is indicated. Consult the troubleshooting section for symptoms that indicate that the  
sensor needs replacing.  
A guide to the relative life left in the cell can be found on status page 2. See Cell  
Strength Factorat the beginning of Chapter 3.0.  
In units with 2 cell blocks, if sensor A fails, sensor B will not automatically switch  
online. This must be done manually through the Install Sensor and Test Alternate Sensor  
menu.  
5.1.2 Scrubber Maintenance  
The oxygen scrubber is a replaceable component. With normal use, the scrubber  
usually lasts for several years.  
NOTE: Do not zero in ranges other than 0-1 ppm, as zeroing in higher ranges will  
exhaust the scrubber.  
TELEDYNE ANALYTICAL INSTRUMENTS  
5-1  
5 Maintenance & Troubleshooting  
Analyzers equipped with a scrubber should only be used with inert gases and saturat-  
ed hydrocarbons.  
5.2 Troubleshooting  
To troubleshoot the analyzer, identify the problem in the following list and then  
follow the procedures to correct the problem. If you cannot identify the problem, or if you  
cannot resolve the problem after following the corrective procedure, call a service repre-  
sentative.  
Symptom  
Cause  
Correction  
Faulty LCD or LED.  
Return the unit to factory.  
LCD or LED display will not  
light.  
Remove analyzer cover. Adjust  
potentiometer located behind  
LCD display.  
Display too bright.  
Faulty electrical connection.  
Remove analyzer cover. Check  
PC board connections and  
make sure that they are firmly  
seated.  
No analog output.  
Bad sensor.  
High O2 level.  
Replace sensor.  
Check zero gas inlet connec-  
tion.  
Zero gas signal >2 PPM.  
Bad sensor.  
Replace sensor.  
Slow response/recovery time.  
Re-zero sensor.  
Displays --value.  
a) The wrong span gas  
concentration has been  
entered.  
a) Re-enter span gas concen-  
tration.  
Can't calibrate (with sensor  
installed). Display reads cell  
too strongor cell too  
weak.”  
b) Replace or reinstall sensor.  
a) The pneumatic air supply  
pressure is too low, making  
the valves stick.  
a) Raise air supply pressure to  
at least 70 psig.  
No sample gas flow.  
b) The sample supply pressure  
is too low.  
b) Check back panel connec-  
tions. Raise sample supply  
to at least 5 psig but within  
2 psig of the span gas  
pressure.  
TELEDYNE ANALYTICAL INSTRUMENTS  
5-2  
Maintenance & Troubleshooting 5  
Correction  
Symptom  
Cause  
c) Make sure metering valve  
(flow set knob on front  
panel) is sufficiently open.  
Back panel vent is blocked.  
See a) above.  
Clear and check connection to  
back panel vent.  
Will not switch modes.  
See a) above.  
Sensor output does not drop  
to a stable, low value during  
zeroing or spanning.  
a) Sensor has been exposed  
to high concentrations of O2  
for an extended time period.  
a) Replace sensor and/or  
wait for stable reading.  
b) Bad sensor.  
b) Replace sensor.  
c) Leak in system.  
c) Check sensor block clamp.  
Make sure the sensor block  
is tightly closed. Check the  
gas supply tank, line and  
connections for leaks.  
Unit will not turn on although  
it is plugged in and the  
Power Switch is ON (at "I"  
position).  
Fuse is blown.  
Check fuse(s) in back panel  
fuse block assembly located  
above the power cord recep-  
tacle.  
TELEDYNE ANALYTICAL INSTRUMENTS  
5-3  
5 Maintenance & Troubleshooting  
TELEDYNE ANALYTICAL INSTRUMENTS  
5-4  
Appendix  
Appendix  
A.1 Specifications  
(Subject to change without notice.)  
Ranges:  
(Six ranges with AutoRanging on all ranges)  
1:  
2:  
3:  
4:  
5:  
6:  
01 part per million (ppm) O2  
010 ppm O2  
0100 ppm O2  
01000 ppm O2  
01 % O2  
025 % O2  
Accuracy:  
2% of full scale when calibrated and operated at the same constant  
temperatureandpressure.*  
5% of full scale over the operating temperature range once tempera-  
tureequilibriumisreached.*  
*
Additional 0.04 ppm on 0-1 ppm range, generated by internal scrub-  
ber (which is necessary to maintain these specifications).  
Response Time: 90 % of final reading in less than 61 seconds at 25 °C  
OperatingTemperature:0-50°C (32-122 °F)  
Analog Signal Outputs:  
For percent of full scale indication:  
one 01 V dc non-isolated signal  
one 420 mA dc isolated signal  
Forrangeidentification:  
one 01 V dc non-isolated signal  
one 420 mA dc isolated signal  
Digital Data Lines:  
One RS-232C serial interface for 2-way communications  
with separate host computerfor remote monitoring and  
control of all functions.  
Alarms:  
Fiveuser-programmableabsolute-readingalarmsetpoints,  
with Form C, SPDT relays (3A resistive).  
TELEDYNE ANALYTICAL INSTRUMENTS  
A-1  
Appendix  
Mounting and Enclosure:  
19" relay rack, 20.3 cm (12.25 inches) high,  
for general purpose (nonhazardous) areas.  
(Class I, Division II, hazardous area models  
available on special order.)  
Sensor: Class B-2, B2C, A2, or A2C, Micro-Fuel  
Cells.  
Altitude:  
1,609 m  
RelativeHumidy: Up to 99%  
Power Requirements: 100, 120, 220, or 240~ 50-60 Hz 1.5 A MAX  
Use 250 V 3.0 A T Fuse for 110, 120 V~  
Use 250 V 1.6 A T Fuse for 220, 240 V~  
TELEDYNE Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500,  
Fax (626) 961-2538  
TWX (910) 584-1887 TDYANYL COID  
Web:  
A-2  
TELEDYNE ANALYTICAL INSTRUMENTS  
Appendix  
A.2 Recommended Spare Parts List  
Description P/N  
Micro-FuelCellsensor C6689B2  
C6689B2C  
C6689A2  
C6689A2C  
O-ring  
O165  
Suctionpurifier P670  
Solenoidvalve V462  
Pneumatic valve upper body B435  
Hose fitting tee T978  
Fuse, 3A T Type F68  
Fuse, 1.6A T Type (European) F768  
Scrubber C58750  
A.3 Drawing List  
D-56534  
C-64819  
OutlineDrawing  
InterconnectionDrawing  
NOTE: The MSDS on this material is available upon request  
through the Teledyne Environmental, Health and  
Safety Coordinator. Contact at (626) 934-1592  
TELEDYNE ANALYTICAL INSTRUMENTS  
A-3  
Appendix  
A.4 Material Safety Data Sheet I  
Section I Product Identification  
Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3,  
and A-5.  
Electrochemical Oxygen Sensors, all classes except R-19.  
Mini-Micro-Fuel Cells, all classes.  
Manufacturer: Teledyne Analytical Instruments  
Address:  
16830 Chestnut Street, City of Industry, CA 91749  
Phone: (818) 961-9221  
Customer Service: Extension 222  
Environmental Health  
and Safety: Extension 230  
Date Prepared : 2/12/96  
Section II Hazardous Ingredients/Composition  
Material or  
Component  
TLV  
C.A.S. # Quantity  
OSHA PEL ACGIH  
Lead (Pb)  
7439-92-1 320 gms  
0.05 mg/m3 0.15 mg/m3  
Potassium Hydroxide1310-58-315 ml None  
2 mg/m3  
Solution 15% (KOH)  
Section III Physical/Chemical Characteristics  
Material  
Appearance  
or Compo-  
nent  
Odor  
Specific Vapor  
Gravity Pres-  
sure  
11.34  
na  
Solubility  
in Water  
Melting Density Evap.  
Boiling  
Point (°C)  
Point  
Rate  
(°C)  
Solid, silver  
gray, odorless  
na  
na  
na  
Insoluble  
Complete  
1744  
1320  
328  
360  
Lead  
White or  
slightly  
yellow,  
no odor  
2.04  
na  
na  
Potassium  
Hydroxide  
A-4  
TELEDYNE ANALYTICAL INSTRUMENTS  
Appendix  
Section IV Fire and Explosion Hazard Data  
Flash Point:  
na  
Flammable Limits:  
na  
LEL: na  
UEL:  
na  
Extinguishing Media:  
Use extinguishing media appropriate to surrounding fire  
conditions. No specific agents recommended.  
Special Fire Fighting  
Equipment:  
Wear NIOSH/OSHA approved self-contained breathing  
apparatus and protective clothing to prevent contact with  
skin and eyes.  
Unusual Fire and Explosion  
Hazards:  
Emits toxic fumes under fire conditions.  
Section V Reactivity Data  
Stability:  
Stable  
Incompatibilities:  
Aluminum, organic materials, acid chlorides, acid  
anhydrides, magnesium, copper. Avoid contact with acids  
and hydrogen peroxide > 52%.  
Hazardous Decomposition of  
Byproducts:  
Toxic fumes  
Hazardous Polymerization:  
Will not occur.  
Section VI Health Hazard Data  
Routes of Entry:  
Inhalation:  
Ingestion:  
Highly unlikely  
May be fatal if swallowed.  
Skin:  
The electrolyte (potassium hydroxide) is corrosive; skin  
contact may cause irritation or chemical burns.  
Eyes:  
The electrolyte (potassium hydroxide) is corrosive; eye  
contact may cause irritation or severe chemical burns.  
Acute Effects:  
The electrolyte is harmful if swallowed, inhaled or  
adsorbed through the skin. It is extremely destructive to  
tissue of the mucous membranes, stomach, mouth, upper  
respiratory tract, eyes and skin.  
Chronic Effects:  
Prolonged exposure with the electrolyte has a destructive  
effect on tissue.  
Chronic exposure to lead may cause disease of the blood  
and blood forming organs, kidneys and liver, damage to  
the reproductive systems and decrease in fertility in men  
and women, and damage to the fetus of a pregnant  
woman. Chronic exposure from the lead contained in this  
product is extremely unlikely.  
TELEDYNE ANALYTICAL INSTRUMENTS  
A-5  
Appendix  
Signs and Symptoms of  
Exposure:  
Contact of electrolyte with skin or eyes will cause a  
burning sensation and/or feel soapy or slippery to touch.  
Other symptoms of exposure to lead include loss of sleep,  
loss of appetite, metallic taste and fatigue.  
Carcinogenicity:  
OSHA:  
Lead is classified by the IARC as a class 2B carcinogen  
(possibly carcinogenic to humans).  
Where airborne lead exposures exceed the OSHA action  
level, refer to OSHA Lead Standard 1910.1025.  
NTP:  
na  
Lead exposure may aggravate disease of the blood and  
blood forming organs, hypertension, kidneys, nervous  
and possibly reproductive systems. Those with preexist-  
ing skin disorders or eye problems may be more suscepti-  
ble to the effects of the electrolyte.  
Medical Conditions Generally  
Aggravated by Exposure:  
Emergency First Aid Procedures:  
In case of contact with the skin or eyes, immediately  
flush with plenty of water for at least 15 minutes and  
remove all contaminated clothing. Get medical attention  
immediately.  
If ingested, give large amounts of water and DO NOT  
INDUCE VOMITING. Obtain medical attention immedi-  
ately.  
If inhaled, remove to fresh air and obtain medical  
attention immediately.  
Section VII Precautions for Safe Handling and Use  
NOTE:The oxygen sensors are sealed, and under normal circum-  
stances, the contents of the sensors do not present a health  
hazard. The following information is given as a guide in the  
event that a cell leaks.  
Before opening the bag containing the sensor cell, check  
Protective measures  
during cell replacement:  
the sensor cell for leakage. If the sensor cell leaks, do not  
open the bag. If there is liquid around the cell while in  
the instrument, wear eye and hand protection.  
Cleanup Procedures:  
Wipe down the area several times with a wet paper towel.  
Use a fresh towel each time. Contaminated paper towels  
are considered hazardous waste.  
A-6  
TELEDYNE ANALYTICAL INSTRUMENTS  
Appendix  
Section VIII Control Measures  
Eye Protection:  
Chemical splash goggles  
Rubber gloves  
Hand Protection:  
Other Protective Clothing:  
Ventilation:  
Apron, face shield  
na  
Section IX Disposal  
Both lead and potassium hydroxide are considered poisonous substances and are regulated under  
TSCA and SARA Title III.  
EPA Waste Number:  
California Waste Number:  
DOT Information:  
D008  
181  
RQ Hazardous Waste Solid N.O.S. (Lead), 9, UN3077,  
PG III  
Follow all Federal, State and Local regulations.  
Section X References  
Material Safety Data Sheets from J.T. Baker Chemical, Aldrich, Mallinckrodt, ASARCO  
U.S. Department of Labor form OMB No. 1218-0072  
Title 8 California Code of Regulations  
TSCA  
SARA Title III  
CFR 49  
CFR 29  
CFR 40  
NOTE:The above information is believed to be correct and is offered for  
your information, consideration, and investigation. It should be  
used as a guide. Teledyne Analytical Instruments shall not be held  
liable for any damage resulting from handling or from contact with  
the above product.  
TELEDYNE ANALYTICAL INSTRUMENTS  
A-7  
Appendix  
A.5 Material Safety Data Sheet II  
Section I Product Identification  
Product Name:  
Micro-FuelCells  
Mini-Micro-FuelCells,allclasses  
Super Cells, all classes except T5F  
Oxygen Sensors, all classes.  
Manufacturer: TeledyneAnalyticalInstruments  
Address:  
16830 Chestnut Street, City of Industry, CA 91749  
Phone: (818) 961-9221  
Date Prepared or Last Revised: 08/08/91  
Emergency Phone Number: (818) 961-9221  
Section II Physical and Chemical Data  
Chemical and Common Names: Potassium Hydoxide (KOH), 15% (w/v)  
Lead (Pb), pure  
CAS Number: KOH 1310583  
Pb 7439921  
KOH (15% w/v)  
MeltingPoint/Range: 10 to 0 °C  
BoilingPoint/Range: 100 to 115 °C  
Pb (pure)  
328 °C  
1744 °C  
11.34  
SpecificGravity: 1.09 @ 20 °C  
pH:  
SolubilityinWater: Completelysoluble  
Percent Volatiles by Volume: None  
Appearance and Odor: Colorless,odorlesssolution  
>14  
N/A  
Insoluble  
N/A  
Grey metal, odor-  
less  
A-8  
TELEDYNE ANALYTICAL INSTRUMENTS  
Appendix  
Section III Physical Hazards  
Potential for fire and explosion: The electrolyte in the MicroFuel Cells is not flam-  
mable. There are no fire or explosion hazards associated with MicroFuel Cells.  
Potential for reactivity: The sensors are stable under normal conditions of use. Avoid  
contact between the sensor electrolyte and strong acids.  
Section IV Health Hazard Data  
Primary route of entry:  
Ingestion,eye/skincontact  
Exposurelimits:OSHA PEL: .05 mg/cu.m. (Pb)  
ACGIH TLV:  
2 mg/cu.m. (KOH)  
Effects of overexposure  
Ingestion:  
The electrolyte could be harmful or fatal if swal-  
lowed.  
Oral LD50 (RAT) = 3650 mg/kg  
Eye: The electrolyte is corrosive; eye contact could result  
in permanent loss of vision.  
Dermal: The electrolyte is corrosive; skin contact could result  
in a chemical burn.  
Inhalation: Liquidinhalationisunlikely.  
Signs/symptoms of exposure: Contact with skin or eyes will cause a burning sensa-  
tion and/or feel soapy or slippery to touch.  
Medicalconditions  
aggravated by exposure: None  
Carcinogenity: NTP Annual Report on Carcinogens: Not listed  
LARC Monographs: Not listed  
OSHA: Not listed  
Other health hazards: Lead is listed as a chemical known to the State of  
California to cause birth defects or other reproductive  
harm.  
TELEDYNE ANALYTICAL INSTRUMENTS  
A-9  
Appendix  
Section V Emergency and First Aid Procedures  
Eye Contact:  
Skin Contact:  
Flush eyes with water for at least 15 minutes and get immediate  
medicalattention.  
Wash affected area with plenty of water and remove contaminated  
clothing. If burning persists, seek medical attention.  
Give plenty of cold water. Do not induce vomiting. Seek medical at-  
tention. Do not administer liquids to an unconscious person.  
Liquidinhalationisunlikely.  
Ingestion:  
Inhalation:  
Section VI Handling Information  
NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents  
of the sensors do not present a health hazard. The following information is given  
as a guide in the event that a cell leaks.  
Protectiveclothing:  
Rubber gloves, chemical splash goggles.  
Clean-up procedures: Wipe down the area several times with a wet paper towel.  
Use a fresh towel each time.  
Protectivemeasures  
duringcellreplacement: Before opening the bag containing the sensor cell, check the  
sensor cell for leakage. If the sensor cell leaks, do not open  
the bag. If there is liquid around the cell while in the instru-  
ment, put on gloves and eye protection before removing the  
cell.  
Disposal:  
Should be in accordance with all applicable state, local and  
federalregulations.  
NOTE: The above information is derived from the MSDS provided by the manufacturer.  
The information is believed to be correct but does not purport to be all inclusive  
and shall be used only as a guide. Teledyne Analytical Instruments shall not be  
held liable for any damage resulting from handling or from contact with the  
aboveproduct.  
A-10  
TELEDYNE ANALYTICAL INSTRUMENTS  

Toshiba Computer Drive GMAA0020901T User Manual
Teac CD Player CD RW901 User Manual
Sun Microsystems StorEdge 6900 Series User Manual
Samsung SPINPOINT MP2 HM200JJ User Manual
Samsung Spinpiont F1 HD321HJ User Manual
Samsung SGH B220 User Manual
Samsung SCH i600 User Manual
Research In Motion Blackberry Cell Phone Curved 8520 User Manual
Panasonic KXL CB30AU User Manual
Palm TreoTM 750 User Manual