| OPERATING INSTRUCTIONS FOR   Model 3160   Trace Oxygen Analyzer   DANGER   HIGHLYTOXICANDORFLAMMABLELIQUIDSORGASESMAYBEPRESENTINTHISMONITORINGSYSTEM.   PERSONALPROTECTIVEEQUIPMENTMAYBEREQUIREDWHENSERVICINGTHISSYSTEM.   HAZARDOUSVOLTAGESEXISTONCERTAINCOMPONENTSINTERNALLYWHICHMAYPERSISTFORA   TIMEEVENAFTERTHEPOWERISTURNEDOFFANDDISCONNECTED.   ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE   CONDUCTINGANYMAINTENANCEORSERVICINGCONSULTWITHAUTHORIZEDSUPERVISOR/MANAGER.   P/NM52972   ECO# 99-0323   08/06/99   TELEDYNE ANALYTICAL INSTRUMENTS   TELEDYNE ANALYTICAL INSTRUMENTS   iii   Table of Contents   1Introduction   1.1Features ....................................................................................... 1-1   1.1.1 Fixed Features.............................................................. 1-1   1.1.2 VariableFeatures ......................................................... 1-1   1.2Components ................................................................................. 1-3   1.3OptionsandModelNumbers........................................................ 1-3   1.4Applications ................................................................................. 1-4   2OperationalTheory   2.1PrinciplesofOperation ........................................................... 2-1   2.1.1Micro-FuelCellSensor ................................................. 2-3   2.1.2SampleSystem .............................................................. 2-3   2.1.3ElectronicsandSignalProcessing ............................. 2-5   RangeIdentifierTable ....................................................... 2-7   3Installation   3.1UnpackingtheAnalyzer ................................................................ 3-1   3.2 Front Panel ................................................................................. 3-1   3.3 Rear Panel................................................................................... 3-2   3.4GasLineConnections ................................................................... 3-4   3.4.1 Span Gas In .................................................................. 3-4   3.4.2InstrumentAir................................................................ 3-4   3.4.3SampleGasIn ............................................................... 3-4   3.4.4Vent .............................................................................. 3-4   3.5Micro-FuelCellInstallationorReplacement .................................. 3-5   3.6ElectricalConnections............................................................. 3-6   3.6.1 VoltageSelection.......................................................... 3-7   3.6.2 FuseChanging .............................................................. 3-8   3.6.3 ParallelandSerialPorts .......................................... 3-9   3.7InstallationChecklist ..................................................................... 3-10   4Operations   4.1FrontPanelControls..................................................................... 4-1   4.1.1 SpinningWheel............................................................. 4-2   4.1.2 Cell Output Factor........................................................ 4-2   4.2ModesofOperation ..................................................................... 4-2   ColdStart-Up ........................................................................ 4-5   (continued)   TELEDYNE ANALYTICAL INSTRUMENTS   iv   Table of Contents (continued)   CalibrationZeroing ........................................................... 4-6   CalibrationUsingSpanGas .............................................. 4-8   SelectActiveSensor(TwoCellsOnly) ............................. 4-11   InstallSensorandTestAlternateSensor.... (TwoCellsOnly) .. 4-12   O2 Range Set-Up .................................................................. 4-14   Set-upAlarms ........................................................................ 4-15   LoggerSet-Up ....................................................................... 4-17   Set-UpClockFunctions ......................................................... 4-19   ChangingPasswordsForRemote   MonitoringandControl ......................................... 4-21   SystemStatistics ..................................................................... 4-23   5Maintenance&Troubleshooting   5.1RoutineMaintenance .................................................................... 5-1   5.1.1 SensorMaintenance ............................................... 5-1   5.1.2 ScrubberMaintenance .................................................. 5-1   5.2Troubleshooting ............................................................................ 5-2   Appendix   A-1 SpecificationsSheet ................................................................ A-1   A-2 RecommendedSparePartsList .............................................. A-2   A-3 DrawingList ........................................................................... A-3   A-4 MaterialSafetyDataSheet ..................................................... A-4   A-5 MaterialSafetyDataSheet ..................................................... A-5   TELEDYNE ANALYTICAL INSTRUMENTS   v TELEDYNE ANALYTICAL INSTRUMENTS   vi   Introduction 1   Introduction   1.1 Features   1.1.1   Fixed Features   The Model 3160 is a microprocessor-based trace oxygen monitoring system that   provides on-line monitoring of trace oxygen at parts-per-million (PPM) levels and can   directly measure the level of purity in high-grade gases.   The 3160 computer uses an Intel® 80188 microprocessor combined with 32   kilobytes of random access memory (RAM) and 128 kilobytes of read-only memory   (ROM). The total computer program resides in the ROM and is not affected by shut-   down periods, loss of power, or battery failure. However, calibration parameters and   stored measurement readings are retained in battery-backed-up RAM, which require   continuous battery voltage in order to be retained. This information will be lost if   battery power is interrupted, as in battery replacement.   Two displays help you monitor trace oxygen levels: a red 4-digit light emitting   diode (LED) display with one-inch numerals is easily read at a distance or even in   bright daylight, and a liquid crystal display (LCD) helps you keep track of alarm   setpoints, ranges, mode and system statistics.   1.1.2   Variable Features   Various models are available. The following features describe the basic model,   but the exact configuration depends on the standard options incorporated. Paragraph   1.3 contains a list of standard variations on the basic model.   The Class B-2 Micro-Fuel Cell measures trace O2 in a mixture of gases. Because   the Micro-Fuel Cell sensor is a sealed electrochemical device, it is replaced as a unit.   There are no electrolytes to change or electrodes to clean, making the sensor mainte-   nance free. The analyzer can be configured with one or two sensor blocks. Each block   can be isolated by pneumatically switched valves. The software automatically adjusts   when a second cell block is present.   TELEDYNE ANALYTICAL INSTRUMENTS   1-1   1 Introduction   The 3160 features six full-scale linear ranges of analysis:   • • • • • • 0–1 PPM   0–10 PPM   0–100 PPM   0–1000 PPM   0–1%   0–25%   An AutoRanging feature automatically selects the range that is appropriate for a   given reading. For example, if the O2 level exceeds 100% of the current range, the   analyzer switches to the next higher range. When desired, a manual override feature   allows a particular range of interest to be locked-in.   With the AutoSpan feature, the analyzer automatically calibrates at scheduled   intervals, and automatically performs electronic zeroing and sensor settling detection   during calibration.   Five user-programmable absolute-reading (PPM) alarm setpoints with assignable   Form 1C (SPDT) relays are provided, along with a RS-232C bidirectional serial data   port, and four signal outputs for chart recorders, etc. The   RS-232 serial port can be used with or without a modem to connect with   • a personal computer loaded with the Teledyne Remote Analytical   Control Software (TRACS)   • • any terminal or terminal emulation software   a custom computer program   Several analyzer functions, including troubleshooting, can be run from your PC   through the phone line to the analyzer.   The analyzer has separate sample and span gas ports, which allow the installation   of an external source of span gas for calibration without interfering with the sample   gas line.   An optional scrubber installed into the sample system allows sample gas to be   used as a zero gas after treatment with the scrubber. If the scrubber is installed, pneu-   matic valves on either side of the scrubber open automatically during zeroing.   Note: Units equipped with a scrubber should only be used with inert gases   and saturated hydrocarbons.   The analyzer output is not affected by minor changes in the flow rate. Analyzer   performance is stable under minor mechanical vibration.   TELEDYNE ANALYTICAL INSTRUMENTS   1-2   Introduction 1   1.2 Components   The analyzer is designed to mount in a standard 19-inch rack. All of the opera-   tional controls are easily accessed through the front panel. The analyzer contains each   of the following:   • Analysis Section   • • Single or Dual Electrochemical Micro-Fuel Cell   Sample System   • • • Power Supply   Microcontroller Module   LED Display   The front panel consists of:   • • • • LED Display   LCD Display with Touch Panel   Micro-Fuel Cell Access Panel   Flowmeter and flow set knob   The back panel ports are:   • • Gas Inlet/Outlet Ports   Electrical Connections   1.3 Options and Model Numbers   The 3160 is available with the following standard options, which are designated   by a descriptive suffix on the model number. Most of the options can be combined.   • • • • • • • • 3160SA   Single fuel cell.   3160SB   Single fuel cell and oxygen scrubber   Dual fuel cell   3160DA   3160DB   3160SAS   3160SBS   3160DAS   3160DBS   Dual fuel cell and oxygen scrubber   Single cell and stainless steel cell block   Single cell, stainless steel cell block, and oxygen scrubber   Dual cell and stainless steel cell block   Dual cell, stainless steel cell block, and oxygen scrubber   TELEDYNE ANALYTICAL INSTRUMENTS   1-3   1 Introduction   1.4 Applications   The analyzer is an invaluable tool in the following applications and industries:   • • • Inert and gaseous hydrocarbon stream monitoring.   Measuring the purity of various gases in air separation plants.   Prevention of oxidation by measuring the purity of blanketing gases in   fiber and glass industries.   • • Monitoring and controlling gas atmospheres in the heat treatment of metals   in steel and other metal industries.   Gas analysis and research in laboratories and research and development   areas.   • • Semiconductor manufacturing.   Process monitoring of gaseous monomers—vinyl chloride, propylene,   butadiene, or ethylene.   • • • • Gas purity certification.   Glove box leak detection.   Natural gas treatment and transmission.   Inert gas welding of exotic metals.   TELEDYNE ANALYTICAL INSTRUMENTS   1-4   Operational Theory 2   Operational Theory   2.1 Principles of Operation   The sub-systems that make up the analyzer are:   1. Micro-Fuel Cell Sensor   2. Sample System   3. Electronic Signal Processing   The Micro-Fuel Cell sensor is an electrochemical galvanic device that translates   the amount of oxygen present in the sample into an electrical current. The sample   system delivers the sample gas in a form that is as unaltered as possible for testing,   and for leak-free transport of gases through the analyzer.   The electronic signal processing unit (control module) is designed to simplify the   operation of the analyzer and accurately process the signal from various components.   The control module incorporates a microprocessor which allows the operation of the   analyzer with a minimum of operator interaction. Figure 2-1 illustrates major   components.   TELEDYNE ANALYTICAL INSTRUMENTS   2-1   2 Operational Theory   I O Figure 2-1. Major Components.   TELEDYNE ANALYTICAL INSTRUMENTS   2-2   Operational Theory 2   2.1.1 Micro-Fuel Cell Sensor   The Micro-Fuel Cell is a sealed plastic disposable oxygen transducer that mea-   sures 1¼ inches in diameter and is ¾ inch thick. Inside of the cell are a cathode and   anode immersed in 15% aqueous KOH electrolyte. At one end of the sensor is a   Teflon diffusion membrane; the other end is sealed with a polyethylene membrane. At   the rear of the cell is a contact plate consisting of two concentric foils. The foils mate   with spring-loaded contacts in the sensor block assembly and provide the electrical   connection to the rest of the analyzer.   The sensing cathode, located beneath the diffusion membrane, has a surface area   of 2.48 cm2. The sample gas enters the sensor block through an inlet tube between the   cathode and the sensor block cap, diffuses through the Teflon membrane, and any   oxygen in the sample gas is reduced on the surface of the cathode by the following   mechanism:   O2 + 2H2O + 4e– → 4OH–   (cathode)   When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the   anode by the following mechanism.   2Pb → 2Pb+2 + 4e–   (anode)   The electrons released at the surface of the anode flow to the cathode surface via   an external circuit. This current is proportional to the amount of oxygen. It is mea-   sured and used to determine the oxygen concentration in the gas mixture.   The overall reaction for the fuel cell is:   2Pb + O2 → 2PbO   The output of the fuel cell is limited by the amount of oxygen in the cell at any   one time, and the amount of stored anode material. In the absence of oxygen, there is   no current generated.   2.1.2 Sample System   The sample system delivers gases to the Micro-Fuel Cell sensor from the ana-   lyzer rear panel inlet. Depending on the mode of operation either a sample or zero gas   is delivered.   The Model 3160 sample system is designed and fabricated to ensure that the   oxygen concentration of the gas is not altered as it travels through the sample system.   The following list of sample system and analyzer features are among those   available, but the exact configuration depends on the standard options incorporated.   Paragraph 1.3 contains a list of the standard variations on the basic model.   TELEDYNE ANALYTICAL INSTRUMENTS   2-3   2 Operational Theory   • Electropolished 316L-type stainless steel components.   To eliminate oxygen absorption and desorption from the internal   wetted surfaces of the sample system components, the sample system is   fabricated from electropolished 316L-type stainless steel.   • Welding/Metal Gasket-Type Fittings.   All of the joints upstream of the Micro-Fuel Cell oxygen sensor are   orbitally welded. Orbital welding is used in the sample system   wherever feasible. Orbital welding fuses the electropolished 316L   stainless steel components together, forming a smooth, clean internal   (wetted) weld junction and eliminating small spaces around the weld   junction where gases can get trapped or absorbed. All of the weld   junctions in the entire assembly are purged using an inert gas during   welding to ensure that there is no oxygen contamination.   Orbital welding is used where practical; otherwise, conventional   precision welding is used. For example, conventional precision welding   is used to fuse the tubes to the mounting blocks.   • Valves.   The analyzer sampling system utilizes three different types of valves.   Each valve is selected to prevent oxygen contamination of the sample   depending on its position and purpose in the circuit.   Air-Actuated Bellows Valves: These valves are normally closed in the   sample system. They are used to control the delivery through the sample   system of the sample or zero gas. The valve bodies are orbitally welded in   the system and the valve bonnets make a metal-to-metal seal to the body.   This valve system eliminates inboard and outboard gas leakage. The valves   are activated (open/closed) by computer-controlled solenoid valves.   Metering Valve: The metering valve is used to manually control the   gas flow rate to the sensor. The body of the metering valve is orbitally   welded, and the bonnet is sealed to the body with metal O-rings. This type of   valve eliminates inboard and outboard gas leakage.   Solenoid Valves: The solenoid valves control the air flow to the air-   activated bellows valves. The solenoid valves are controlled by the   microprocessor module. When de-energized, the valve outlet is open to   ambient air, allowing the air-activated bellows valve to close.   • Scrubber (optional). See Appendix for details.   In systems with a scrubber, oxygen-trapping media is used to remove   any trace oxygen from the sample gas. This occurs automatically in the   zero menu, where the pneumatic valves leading to the scrubber are   triggered by a certain menu sequence. This process enables sample gas   TELEDYNE ANALYTICAL INSTRUMENTS   2-4   Operational Theory 2   to be used to zero the analyzer, eliminating the need to switch external   tanks.   • Overall Design.   The design of the sample system minimizes the volume of dead space, which   can retain residual gas from another route or previous mode of analysis.   Figure 2-2: Model 3160 flow schematic.   2.1.3 Electronics and Signal Processing   The Model 3160 analyzer has an embedded microcomputer, which controls all   signal processing, input/output, and display functions for the analyzer. System power   is supplied from a power supply module designed to be compatible with any interna-   tional power source.   The microcomputer, a liquid crystal display (LCD), and all analog signal pro-   cessing electronics are located inside a replaceable control module. A light emitting   diode (LED) display is located outside the control module.   TELEDYNE ANALYTICAL INSTRUMENTS   2-5   2 Operational Theory   Electronics in the analyzer are grouped according to function:   1. Analog in   2. Analog out   3. Digital circuit   4. Power supply   Analog signal processing is accomplished in two plug-in circuit cards operating   under control of the microcomputer. All analog signals are converted to digital early   in the processing cycle to minimize analog processing and assure maximum system   accuracy, since digital processing is much more accurate than analog and immune to   many parameters such as drift and aging.   The initial processing and digitization of the signal from the Micro-Fuel Cell   takes place on two circuit boards: the six-decade programmable printed circuit board   (PCB), and the analog input board.   The first step in the chain takes place on the six-decade programmable PCB,   which is mounted directly on top of the cell block. The output of the sensor is a   current that is linearly proportional to the oxygen concentration. At low concentra-   tions, the current from the sensor is low, which makes the close physical proximity of   the pre-amp circuitry to the Micro-Fuel Cell necessary.   The sensor current is routed into a low-offset, low-bias current operational   amplifier configured as a current-to-voltage converter. The signal is sent through two   amplifiers to the analog input board. The pre-amp board provides sensor zeroing and a   connection to the thermistor in the sensor cell block.   The analog input PCB, which plugs into a DIN connector on the digital PCB,   performs three functions:   • Temperature compensation in the sensor signal   • Sensor offset correction   • Digitization   A thermistor inside the sensor block registers resistance and thus the temperature   of the sensor. Accordingly, an analog-to-digital converter configured as a digitally-   programmed attenuator (DPA) is used to reduce or increase the gain of the sensor   signal, so that the signal coming out of the DPA is independent of temperature.   If the software determines that the sensor needs offsetting, the sensor current is   back-calculated and an analog-to-digital converter on the analog board is driven by the   software to produce a voltage that is used to force a current into the current-to-voltage   converter located on the pre-amp board. The current injected is equal and opposite in   polarity to the portion of the sensor current that needs offsetting.   Digitization is provided by a 12-bit analog-to-digital converter, which digitizes   the oxygen and temperature signals.   TELEDYNE ANALYTICAL INSTRUMENTS   2-6   Operational Theory 2   The analog output printed circuit board (PCB) generates the two 0–1 volt and the   two 4–20 mA analog signal outputs available on the rear panel of the analyzer. These   signals, generated in digital format by the microcomputer, are converted into analog   signals by the circuitry on this PCB. The output signals represent the following:   • 0–1V Signal (Oxygen Measurement): This output goes from 0 to 1, represent-   ing 0 to 100% of the scale that has been set; i.e., 0.6 volt is equal to 60% of the full   scale, or 6 PPM when on the 10 PPM scale. It is possible that the signal may go past   zero into the negative range up to -0.25, especially if the analyzer has been zeroed   with a gas that contains a significant concentration of oxygen. See Figure 2-3.   • 0-1V Range Identifier: This 0 to 1 volt output represents each range with a   particular voltage as shown in Table 2–1.   • Isolated 4–20 mA Signal (Oxygen Measurement): This is a 4 to 20 mA output   representing 0 to 100% of the scale, with 4 mA equal to 0%, and 20 mA equal to   100% of that range. This output may also range lower than 4 mA, especially if the   analyzer has been zeroed with a gas that contains a significant concentration of oxy-   gen. See Figure 2-3.   • Isolated 4–20 mA Range Identifier: This 4 to 20 mA output represents   individual ranges with discrete current output as shown in Table 2-1.   Table 2-1. Range Identifier   Identifier   Voltage (V)   0.0   Identifier   Current(mA)   4.0   Range   1 Scale   1 PPM   2 3 4 5 6 10 PPM   100 PPM   1000 PPM   1%   0.2   0.4   0.6   0.8   1.0   7.2   10.4   13.6   16.8   20.0   25%   Figure 2-3: Analog signal output offset   TELEDYNE ANALYTICAL INSTRUMENTS   2-7   2 Operational Theory   The digital printed circuit board (PCB) is a general purpose microcomputer used   to control all functions of the analyzer. The analog input PCB and the analog output   PCB plug directly into connectors located on the digital PCB. In addition to control-   ling these analog PCBs, the digital board performs the following functions:   1. Processes input from the control panel pushbuttons.   2. Provides signals for the selectable alarms.   3. Processes serial I/O functions (RS-232 data). The serial interface default   and programmable parameters are listed in Table 2-2.   Table 2-2. Default and Programmable Parameters   Defaults   1200 Baud   8 Bits   Programmable Options   1200 Baud   5, 6, 7, or 8 Bits   No Parity   1 Stop Bit   Even, Odd, or No Parity   1 or 2 Stop Bits   4. Controls the LCD and the LED displays.   LCD: This screen is a dot-matrix display located on the control module of   the analyzer. It is the user interface for system operations. It   displays the menus and command options available for each   function.   LED: This screen is a 7-segment display located on the front panel of the   analyzer, above the control module. It displays only the oxygen   concentration. It is large and bright to allow the operator to read it   at a greater distance. A dimmer switch for this display is located on   the display PCB behind the front panel.   The analyzer power supply module is a replaceable assembly containing four   power supplies and five alarm relays. Electronic circuitry used to drive and interface   the alarm relays to the output of the microcomputer is also located inside this module.   Note: This power supply contains an International Power Entry Module. This   feature allows operation on any of four international voltage ranges:   100V, 120V, 220V or 240V (50Hz or 60Hz). It also facilitates both North   American and European fusing arrangements. Programming this mod-   ule is described in the installation section of this manual.   TELEDYNE ANALYTICAL INSTRUMENTS   2-8   Installation 3   Installation   Installation of the analyzer includes:   1. Unpacking the system.   2. Recognizing the necessary precautions when installing the system.   3. Hooking up the sample/span gas and air supply to appropriate connections.   4. Installing the Micro-Fuel Cell sensor(s).   5. Hooking up electrical connections.   6. Testing the system.   3.1 Unpacking the Analyzer   The analyzer is shipped with all the materials and special items you need to install   and prepare the system for operation. Carefully unpack the analyzer and inspect it for   damage. Immediately report any damage to the shipping agent. Remove the packing   slip and verify that you have received all the components listed in Table 3-1.   Table 3-1. Accessory Kit for Model 3160   Qty   3 Part No.   G285   Description   Weld glands, VCR-type   3 G284   Gaskets, VCR-type   3 2 1 N284   W64   M52973   Nuts, female (Hoke 4NM316)   Wrench, Open End, 3/4"–5/8"   Instruction Manual   3.2 Front Panel   The Model 3160 front and rear panels are illustrated in Figure 3–1 on the follow-   ing page. The front panel contains the following displays and controls:   1. Main power switch.   2. Light Emitting Diode (LED) display: the O2 concentration is displayed in   bold lettering.   TELEDYNE ANALYTICAL INSTRUMENTS   3-1   3 Installation   3. Parts Per Million (PPM) and Percent Oxygen indicators: the PPM light will   light while the instrument is measuring O2 in units of PPM. The percent   light will light while the instrument is measuring O2 concentration in   percent.   4. Flow Indicator and Flow Set: a flowmeter and flow set knob are provided   for adjusting gas flow in standard cubic feet per hour (SCFH).   5. Liquid Crystal Display (LCD) with keypad: the LCD shows system menus   and data during instrument operation. The keypad is the primary user-input   device used to enter information in the system.   3.3 Rear Panel   Figure 3–1 also shows the rear panel of the 3160. Located on the rear panel are   the following electrical power and gas input ports, alarm relay outputs, and analog and   digital outputs:   1. Gas Input/Vent: three input fittings are provided for span gas, sample   gas, and compressed air for pneumatic valve operation. A single vent is used   for the sample and span gas outlet. Consult the Appendix for scrubber selec-   tion.   2. AC Power Input: 110, 120, 220 or 240 V ~ at 50/60 Hz 1.5 A MAX.   Use 250 V 1.6 A T Fuse for 110, 120 V ~   Use 250 V 0.8 A T Fuse for 220, 240 V ~   3. Alarm Circuit Connections: there are five contact closures/openings   provided for external alarms. The alarm functions are defined by the user via   keypad input within the LCD menu system.   4. Analog Outputs: four analog output connectors are provided for use   with a chart recorder. Two provide range and data in voltages; the other two   provide range and data for current-driven recorders.   5. RS-232 Serial Port: a 9–pin digital input/output connector is provided   for connecting either a serial printer or a serial link to an external computer.   Optional serial link software (such as TRACS) can be used with the instru-   ment for remote external computer monitoring and control of the instrument,   via modem or hardware link.   TELEDYNE ANALYTICAL INSTRUMENTS   3-2   Installation 3   In   mm   DIMENSIONS:   REMOVABLE   TOP COVER   LED   DISPLAY   17.406   (442)   POWER   SWITCH   19.000   (483)   PPM   INDICATOR   I 8.000   (203)   O PERCENT   INDICATOR   5.750   (146)   LIQUID   CRYSTAL   DISPLAY   (LCD)   18.000   (457)   COMPUTER   MODULE   1.50   (38)   KEYPAD   FLOWMETER   FLOW   SET   KNOB   SENSOR   ACCESS   PANEL   ANALOG   RS-232   SERIAL PORT   OUTPUT   CONNECTIONS   8.688   (221)   SPAN   GAS IN   INSTRUMENT   AIR   SAMPLE IN   AC POWER   VOLTAGE SELECT   ALARM   CIRCUIT   CONNECTIONS   SAMPLE/SPAN   VENT   Figure 3–1: Analyzer front and rear panels.   TELEDYNE ANALYTICAL INSTRUMENTS   3-3   3 Installation   3.4 Gas Line Connections   All of the gas lines to the system hook up at the back of the unit (see Figure 3–1).   All of the fittings on the removable back panel are ¼ " male VCR type fittings, with the   exception of the compressed air inlet fitting. Use the wrenches provided to connect the   gas lines. Insert the gasket between the fitting and tighten the female and male nuts   until finger-tight; then, by holding the male nut with the wrench, tighten the female nut   with the second wrench by ¼ turn.   CAUTION: Do not put any torque on the tubes welded on the sampling   system.   Check that each of the lines is hooked up to the correct connection. The lines   should be connected in the following order (from left to right):   3.4.1. Span Gas In   Provide at least one sample gas of a known oxygen concentration. Using   70-99 % of the range just one range above the range of interest is recommended. Any   range EXCEPT 0-1 ppm may be used to calibrate.   3.4.2. Instrument Air (Compressed Air Fitting)   The gas pressure (70–80 psi) needed to operate the pneumatic valves in the   analyzer can be supplied through this fitting.   CAUTION: Pressure higher than 80 psig can damage the solenoid   valves.   3.4.3. Sample Gas In   Hook up the sample gas to the sample gas inlet. When the sample gas flows   through the optional oxygen scrubber, the sample gas becomes the zero gas and can be   used for calibration.   Note: Sample and span gas pressure should be between 5 and 10 psig and   within ±2 psi of each other. A substantial difference in the span and   sample gas pressures will cause a sudden change in gas flow rate.   3.4.4. Vent   The vent transports the sample or span gas out of the system after exposure to   the Micro-Fuel Cell sensor.   TELEDYNE ANALYTICAL INSTRUMENTS   3-4   Installation 3   3.5 Micro-Fuel Cell Sensor Installation or Replacement   The 3160 comes with the Micro-Fuel Cell(s) already installed. When installing or   replacing the cells, remember to inspect the replacement sensor for leaks or damage. See   Figure 3-2.   Figure 3-2: Sensor Installation.   TELEDYNE ANALYTICAL INSTRUMENTS   3-5   3 Installation   1. The sensor compartment is located in the lower right-hand corner of the   analyzer front panel. To remove the sensor compartment panel, loosen the   thumbscrews located left and right center of the panel.   2. Make sure that sample gas is flowing through the analyzer, then shut off   the flow completely. This insures that air does not diffuse into the analyzer.   3. Inside the compartment should be two (one) stainless steel cylindrical   sensor block(s), each with a clamp lever across the front. The bottom half   of the clamp face is bent outward. To release the clamp, pull the clamp   lever up and toward you slowly.   4. Releasing the clamp releases the stainless steel block cap on the bottom of   the sensor block. Pull the cap downward to remove it. There is a guide pin   to the rear of the sensor block that aligns the cap to the block.   Note: It is important to minimize the amount of time that the new cell is ex-   posed to air in order to reduce the time required for the reading to drop   to zero.   5. Remove the Micro-Fuel Cell from the plastic bag. The cell has a mesh   screen with a raised plastic dress ring on one side and flat gold contact   rings on the other. Place the screen side of the cell face down on the block   cap so that the dress ring locks onto the cap ridges and will not move side   to side. The gold contact rings will be facing up.   6. Carefully guide the sensor and cap into the block, putting the guide pin   through the hole in the back edge of the cap.   7. While holding the cap in place, squeeze the clamp down firmly until the   notches on the clamp lock onto the block side pins.   8. Immediately start the sample flow, and set to about 2 SCFH. The analyzer   now needs to be zeroed and calibrated.   3.6 Electrical Connections   All of the electrical connections are located on the analyzer back panel. The   analyzer is shipped with all of the electrical connections intact, with no assembly or   installation required. The power cord receptacle is located in the lower center of the   panel. The voltage selection terminal and the fuse block are in the same fitting directly   above the power cord receptacle. There are four output signal connectors with screw   terminals in the upper right-hand panel. There are two wires per output with the polari-   ty noted below each. The five alarm circuit connectors, located in the lower center   right-hand panel, are screw terminals for alarm relay contacts. These five provide a set   of Form C contacts for the user. The contacts are capable of switching up to 3 am-   peres at 115 V ac into a resistive load.   TELEDYNE ANALYTICAL INSTRUMENTS   3-6   Installation 3   Figure 3-3: 3160 Rear Panel.   3.6.1 Voltage Selection   WARNING:   Power cord must be disconnected before performing any voltage   selection!   The voltage setting and fuses of the analyzer can be changed to international   standards. To change the voltage setting or fuses:   1. Open the cover using a small blade screwdriver.   2. Remove the cover and the fuse block assembly.   3. Pull the voltage selector card straight out of the housing by pulling on the   indicator pin (see Figure 3–4).   TELEDYNE ANALYTICAL INSTRUMENTS   3-7   3 Installation   Figure 3-4: Removing the Voltage Card   4. Turn the card so that the desired voltage can be read at the bottom.   5. Slide the indicator pin around the card so that it is pointing up when the   voltage is read at the bottom (see Figure 3–5).   Figure 3–5: Voltage Selection   6. Place the voltage selector card back into the housing. The edge where the   desired voltage is printed should go in first and the printed side of the card   should be facing the IEC connector.   7. Replace the fuse block and the cover. Verify that the indicator pin is   pointing to the correct voltage.   3.6.2 Fuse Changing   NOTE:Spare fuses are located in a clip attached to the power supply enclosure   inside of the analyzer.   TELEDYNE ANALYTICAL INSTRUMENTS   3-8   Installation 3   WARNING:   Power cord must be disconnected before performing any   voltage selection!   To change from North American to European fuses, perform the following:   1. Use a small blade screwdriver to open the cover.   2. Loosen the Phillips screw two turns and remove the fuse block by sliding it   away from the screw and up from the pedestal.   3. Remove the North American fuse and turn the fuse block over.   4. Replace the fuse with European fuses. Two European fuses are required,   but the lower one may be replaced with a dummy fuse, or jumper bar (see   Figure 3–6).   European 1.6 Amp T Fuses   5 x 20 mm   North American 3 A T Fuses   Figure 3–6: Fuse Replacement   5. Slide the fuse block back on the Phillips screw and pedestal.   6. Tighten the Phillips screw and replace the cover.   NOTE:The fuses that go into the housing first are the active set.   TELEDYNE ANALYTICAL INSTRUMENTS   3-9   3 Installation   3.6.3 RS-232 Serial Digital Port   The RS-232 port is configured as data terminal equipment (DTE) and uses a 9-pin   D connector wired as follows:   1. No connection   2. Transmit Data   3. Receive Data   4. N/C   5. Ground   6. N/C   7. RTC   8. CTS   9. N/C   Connecting the analyzer to a modem requires a PC to modem cable. When con-   necting the analyzer directly to a PC, terminal or other data terminal equipment, a   “null-modem” cable is required.   The following serial interface default parameters are used:   Defaults   1200 Baud   8 Bits   Programmable Options   1200 Baud   5, 6, 7, or 8 Bits   Even, Odd, or None   1 or 2 Stop Bits   No Parity   1 Stop Bit   3.7 Installation Checklist   Have you:   • • • Checked for leaks from any of the rear panel connections?   Checked the sample and span gas pressures (5–10 psi)?   Checked the instrument air pressure (70–80 psi)?   TELEDYNE ANALYTICAL INSTRUMENTS   3-10   Operations 4   Operations   4.1 Front Panel Controls   The front panel of the analyzer, shown in Figure 4-1, contains indicators and displays   through which the computer module can be accessed. The upper left-hand side of the panel   has an LED screen that displays the oxygen content of the sample in one-inch high   numerals. This display can be brightened or dimmed by a potentiometer located directly   behind the panel inside of the analyzer case. To access it, remove the top cover of the   analyzer.   Figure 4.1: Analyzer front panel.   Below the LED screen are two red LED lights. Each will light to indicate whether the   oxygen is being displayed in parts per million (PPM) or percent (%).   To the right of the LED screen is a flowmeter in standard cubic feet per hour   (SCFH). The flow set knob adjusts the flow rate of the gas during calibration and zeroing.   In the panel below the LED screen is the LCD display. The five colored buttons   below it are used to interface with the computer module in selecting modes, with the LCD   TELEDYNE ANALYTICAL INSTRUMENTS   4-1   4 Operations   displaying the function of each button directly above it. The LCD screen displays the   current mode and any warning messages, instructions and button functions.   4.1.1 Spinning Wheel   A “spinning wheel” appears in the upper left-hand corner of the LCD display to   indicate that the alarms are enabled (any that are not defeated), or that scheduled zeroing   (AutoZeroing) or scheduled spanning (AutoSpanning) may take place. If the spinning   wheel does not appear, the alarms are disabled, and a scheduled zero or span is defeated.   Note that there is a delay of several minutes before the spinning wheel appears whenever   the Analyze mode is re-entered.   4.1.2 Cell Output Factor (Span Factor)   The expected life of the Micro-Fuel Cell sensor is about eight months. A guide to the   relative life left in the cell can be found on status page 2 (see System Statistics near the end   of this chapter) or retrieved from the serial port through the serial command SF. The span   factor ranges from 0.00 to 1.00, with 1.00 representing full life expectancy. A span factor   below 0.1 indicates that the Micro-Fuel Cell needs replacing.   4.2 Modes of Operation   To use the system and select displayed options, press the button directly underneath   the option you wish to select. There are eleven different menus:   Cold Start-Up: During the first startup, and when subsequent cold start-   ups are chosen by the user, initial values set at the factory for alarms, I/O,   calibration, zeroing and other data are used.   Upon cold start, zeroing and calibration should be performed before   accurate oxygen measurements are obtained. A system warm start is automati-   cally performed on each subsequent power-up. For a warm start, previous   user-input configuration data is preserved in RAM by battery-preserved   memory in the system control module.   Calibration Zeroing: For the highest possible accuracy, the analyzer   must be zeroed using an oxygen-free gas. By eliminating normal background   “noise” from the sensor reading, zeroing resets the level referred to as the   “zero” oxygen concentration.   If the analyzer is equipped with a scrubber, it is important to zero the   analyzer with a near-zero gas connected to the sample port. The zeroing   process automatically opens pneumatic valves on either side of the scrubber,   and a gas with a high concentration of oxygen will react with the scrubber and   necessitate replacement sooner.   TELEDYNE ANALYTICAL INSTRUMENTS   4-2   Operations 4   When waiting for manual sensor settling during calibration zeroing, wait   until the oxygen concentration has reached its lowest value and has remained   so for 10–15 minutes. It is important to make sure that the sensor is reading the   lowest possible value, since zeroing will reset the point at which the sensor   reads zero. Do not zero prior to reaching the lowest possible oxygen concentra-   tion. When initially installing the sensor(s), do a zero calibration, then a span   calibration. If desired, the instrument may be programmed for automatic   periodic zeroing (AutoZeroing) if the optional scrubber is installed. AutoZero-   ing uses the scrubber to generate a zero gas from a low O2 sample gas.   NOTE: 1. Zeroing in the 0-1 PPM range preserves the life of the scrubber.   In scheduled zeroing and spanning (AutoZeroing and AutoSpanning), a   time interval must be chosen. This interval determines how often the instru-   ment zeros or spans itself. To change the scheduled time, edit it after setting   the interval in the Set Up Clock Functions mode.   NOTE: 2. AutoZeroing and AutoSpanning may be delayed until the alarms have been   enabled.   Calibration Using Span Gas: After zeroing, the instrument can now   be calibrated using a span gas with a known oxygen concentration. The analyz-   er may be calibrated on any range other than 0–1 PPM, but it is best to cali-   brate one range up from the range where the sample is expected to be. The   span gas concentration should be within 70% to 99% of full scale of the range   selected.   NOTE: 3. Prior to calibration, allow the analyzer reading to come to a stable value with   zero or sample gas flowing through the analyzer.   Select Active Sensor (Two cell blocks only): One or two O2 sensor   blocks can be installed in the system. In dual sensor systems, either sensor can   be isolated with pneumatic valves, and the other sensor chosen as the active   sensor for O2 measurements. The system software is aware of how many   sensor blocks are present, but will not automatically switch sensors should one   fail.   This menu will not appear if only one cell block is installed.   Install Sensor and Test Alternate Sensor (Two cell blocks only): Pro-   cedures are provided for installing a new sensor and for testing the alternate O2   sensor.   This menu will not appear if only one cell block is installed.   O2 Range Set-Up: The measurement range may be manually chosen, or   the system may be set to automatically choose the range (AutoRanging). In the   AutoRanging mode, the range will step up at greater than 100% and step   down at 80% of full scale of the current range. The analyzer is programmed to   prevent oscillation between ranges during rapidly changing oxygen levels.   TELEDYNE ANALYTICAL INSTRUMENTS   4-3   4 Operations   Set-Up Alarms: Each of five alarms provided may be programmed to   actuate at any O2 trace level above or below a threshold. The fifth alarm may   also be programmed as a system alarm to monitor several system-level condi-   tions. The system alarm can be triggered by the following system operations:   • Scheduled zero (AutoZero) failure   • Scheduled span (AutoSpan) failure   • Battery failure   • Internal errors   Any condition triggering the system alarm can be determined from status   page 4 (see this chapter: System Statistics).   When the alarms are disabled, scheduled zeroing or spanning will not   take place.   Logger Set-Up: A system data logger keeps a record of past oxygen   measurements. The logger is presented in the form of a graphed chart on the   system LCD; gaps in time will not display, and all points shown on a single   chart have the same range and sensor number.   The logger will not record in the Test Alternate Sensor mode.   Set Up Clock Functions: An internal clock provides the time and date   for events scheduled by the user. The internal clock is also used by the logger   to record events scheduled at timed intervals.   Scheduled zeroing and spanning (AutoZeroing and AutoSpanning) are set   in this mode. The mode uses an electronic timer to set off the zeroing or   spanning sequence at pre-selected intervals. During an AutoZero or AutoSpan,   the alarms are defeated and the screen displays “PLEASE WAIT. SCHED-   ULED SPAN IN PROGRESS.”   Since AutoZeroing and AutoSpanning uses automatic settling detection   of the sensor, which can take up to fifteen minutes, oxygen monitoring will be   interrupted for the length of time it takes to zero or span, after which monitor-   ing will resume. The LCD will display the current background level during   zeroing or spanning.   Changing Passwords For Remote Monitoring and Control:   Programmable security passwords are provided to prevent unauthorized access   to remote monitoring and control. “Monitor” is the lowest authorization level   and only permits monitoring of analyzer functions, while “Control” allows   remote control of any function.   System Statistics: Menu screens are available to view various system   parameters, and for setting serial link operation parameters, including modem   operation.   TELEDYNE ANALYTICAL INSTRUMENTS   4-4   Operations 4   Cold Start-Up   1. A cold start will automatically occur the first time you   start the analyzer. In any other situation, you can cold   start by turning the analyzer off and holding down the   red key below the LCD screen while you turn the power   back on. Continue to hold the red key until the screen   comes on. There will be a two minute wait for system   initialization before the second screen.   A warm start will take you directly to step 5.   2. Set the time using the SELEC, UP and DOWN keys,   and press ACCPT.   3. Select sensor A or B as the active sensor.   (This menu is omitted if only one cell block is installed.)   4. If you have not done so, connect the sample gas to the   sample port and press ACK.   5. Using the valve and meter on the sample gas supply, set   the flow rate to 2 SCFH and press ACK.   If you have warm-started, this is the second screen you   will see.   6. The main Analyze mode menu will appear, and the   measured O2 level of the sample gas will display. The   system should first be zeroed, then spanned, to obtain   the most accurate O2 measurements.   TELEDYNE ANALYTICAL INSTRUMENTS   4-5   4 Operations   Calibration Zeroing   1. At the main Analyze mode screen press ZERO.   2. Press ACK to acknowledge that a low O2 gas has al-   ready been connected to the sample port. The sample   gas O2 level should be close to zero to preserve the life   of the scrubber (if one is installed).   3. Set the flow rate to 2 SCFH using the knob on the sys-   tem front panel and press ACK.   4. Press AUTO to let the system determine the settling   time before spanning or press MAN for manual.   5. If MAN was chosen in step 4, watch the logger or the   O2 value on the screen to assess when the output is con-   stant, and then press ACK.   6. After step 5, the system will automatically span to the   user-entered span value. If the zeroing is successful, the   main Analyze mode menu will appear as in step 10.   TELEDYNE ANALYTICAL INSTRUMENTS   4-6   Operations 4   7. If the O2 cell current is too large to be zeroed, an error   message will appear. Press ACK to retry or ESC to   abort the zeroing process.   8. If AUTO was chosen in step 4, wait for the O2 level to   settle before the system zeros. This may take from thirty   (30) seconds to fifteen (15) minutes, depending on the   sample gas oxygen level.   9. After a stable low oxygen level is detected by the sys-   tem, the system will automatically zero.   If the O2 cell current is too large to be zeroed, an error   message will appear (see step 7). You will then have to   press ESC to abort, or ACK to retry.   10. If the zero is successful, the system will return to the   main Analyzer mode menu.   TELEDYNE ANALYTICAL INSTRUMENTS   4-7   4 Operations   Calibration Using Span Gas   1. Enter the Calibration mode from the Analyze mode by   pressing SPAN.   2. Press ACK if zeroing has already been done. Make sure   to zero before calibrating to ensure correct O2 readings.   3. If the span gas tank is not connected to the span port,   connect it now and press ACK.   4. Enter the span gas concentration using the UP and   DOWN keys, and press ACCPT.   5. Set the flow rate to 2 SCFH and press ACK.   6. Select manual or automatic sensor settling detection by   pressing MAN or AUTO. Automatic sensor settling de-   tection lets the system determine when the sensor has   stabilized.   TELEDYNE ANALYTICAL INSTRUMENTS   4-8   Operations 4   7. If AUTO was chosen in step 6, wait for the O2 level to   settle before the system spans. This may take from fif-   teen (15) to thirty (30) minutes or longer, depending on   the sample gas oxygen level.   Go to step 11 if MAN was chosen in step 6.   8. If any key was pressed in step 7, the screen indicates   that the span has been aborted. Press ACK to retry   spanning, or ESC to abort the span.   9. After the system has automatically detected sensor sta-   bility, the system will automatically span. If the system   spans correctly, the next screen will be the Analyze   mode main menu (see step 14).   10. If the O2 cell current is too strong or too weak to be   spanned, an error message will appear. Insure that the   span gas concentration has been entered correctly. Press   ACK to retry, or ESC to abort the span.   11. If MAN was pressed in step 6, watch the logger or the   O2 value on the screen to determine when the output is   constant, and then press ACK.   12. After step 11, the system will automatically span to the   user-entered span value.   TELEDYNE ANALYTICAL INSTRUMENTS   4-9   4 Operations   13. If the O2 cell current is too strong or too weak to span,   an error message will appear. This will not occur if the   user-entered span gas concentration is correct. Press   ACK to retry, or ESC to abort the span.   14. If the span is successful, the system will return the main   Analyze mode menu. Check to see that the span value is   correct.   TELEDYNE ANALYTICAL INSTRUMENTS   4-10   Operations 4   Select Active Sensor (Two Cells Only)   1. Enter set-up from the Analyze mode main menu by   pressing SETUP.   If only one cell block is installed, the TEST and INSTL   options on the next screen are omitted.   2. Press SET to change the active sensor.   3. Enter your authorization code by using the SELEC, UP   and DOWN keys, and press ACCPT.   4. Press SENSR. (This option will not appear if only one   sensor block is installed.)   5. Select sensor A or B.   6. The main Analyze mode menu will appear, and will indi-   cate the chosen sensor.   TELEDYNE ANALYTICAL INSTRUMENTS   4-11   4 Operations   Install Sensor andTest Alternate Sensor (Two Cells Only)   1. Enter set-up from the Analyze mode by pressing   SETUP.   If only one cell is installed, the TEST and INSTL op-   tions on the next screen are omitted.   2. Press TEST to test the alternate sensor or INSTL to in-   stall a new sensor.   3. If TEST was chosen in step 2, the active sensor is dis-   played and you are directed to press ACK to test the al-   ternate sensor.   If INSTL was chosen in step 2, go to step 5.   4. After pressing ACK in step 3, the alternate sensor is   flushed for 10 minutes while the measured O2 level is   displayed. The 10 minute flush may be restarted by   pressing RESET, or aborted by pressing SKIP. Go to   step 8.   5. If INSTL was chosen in step 3, the screen will display   the active sensor. Press A or B for the sensor you are   installing.   6. Install the sensor (either A or B from step 5) into the   sensor block within two minutes, and press ACK.   TELEDYNE ANALYTICAL INSTRUMENTS   4-12   Operations 4   7. After pressing ACK in step 6, the new sensor is flushed   for 10 minutes while the measured O2 level is displayed.   The 10-minute flush may be restarted by pressing RE-   SET, or aborted by pressing SKIP.   8. When sensor installation or alternate sensor test are   complete, the system will return to the main Analyze   mode menu.   TELEDYNE ANALYTICAL INSTRUMENTS   4-13   4 Operations   O2 Range Set-Up   1. Enter Range Set-Up mode from the Analyze mode by   pressing RANGE.   2. Press AUTO for AutoRanging, or FIXED to manually   fix the O2 measurement range. In AutoRange mode, the   range will be automatically determined so that the O2   measurements are within a range which maximizes the   accuracy of the reading.   3. If AUTO is chosen in step 2, the main Analyze mode   menu will appear, and will indicate the automatically de-   termined range (see screen for step 1).   If FIXED is chosen in step 2, use the UP and DOWN   keys to choose the desired manual range, and press   ACCPT.   4. If FIXED was chosen in step 2 and the range was speci-   fied in step 3, the main Analyze mode screen will ap-   pear, and will indicate the new manually set range.   TELEDYNE ANALYTICAL INSTRUMENTS   4-14   Operations 4   Set-Up Alarms   1. Enter set-up from the Analyze mode by pressing   SETUP.   2. Press SET to set the alarms.   3. Enter the authorization code by using the SELEC, UP   and DOWN keys, and press ACCPT.   If this is a COLD START, the code you enter will be   the code set for every subsequent start-up until the next   COLD START.   If this is a WARM START, enter the code used during   the last COLD START.   4. Press ALRMS.   5. Press   • • • • AL1 for alarm 1   AL2 for alarm 2   MORE for alarms 3–5   ESC twice to return to the main menu   Note that alarm 5 may be programmed as a system   alarm.   6. If AL1 or AL2 was chosen in step 5, use the SELEC,   UP and DOWN keys to configure alarm 1 or 2, and   press ACCPT, which will take you back to step 5.   • LEVEL is the O2 concentration for the alarm   threshold.   • HIGH/LOW determines whether the alarm con-   dition is above or below the alarm threshold.   TELEDYNE ANALYTICAL INSTRUMENTS   4-15   4 Operations   • LATCHING actuates the alarm above a certain   setpoint even if the level falls back below the   setpoint.   • • • RELAY indicates the relay (on the rear panel)   actuated by the alarm, which is changeable.   FAILSAFE=YES enables the relay which actu-   ates the alarm during system failure.   DEFEAT=NO actuates the alarms for normal   use.   7. If MORE was pressed in step 5, choose alarms 3, 4, or 5   by pressing AL3, AL4, or AL5.   Choose AL5 to set-up a system alarm.   8. If AL5 was pressed in step 7, choose alarm 5 to be an   O2 or system alarm by pressing O2 or SYSTEM.   Configured as a system alarm, alarm 5 will ring when a   scheduled calibrator zeroing or a scheduled span calibra-   tion fails, when there is an internal system failure, or   when the RAM back-up battery fails.   9. If SYSTM was pressed in step 8, configure the system   alarm using the SELEC, UP and DOWN keys.   • LATCHING=YES actuates alarm 5 above a cer-   tain setpoint even if the level falls back below the   setpoint.   • • FAILSAFE=YES enables the relay which actu-   ates the alarm during system failure.   DEFEAT=YES actuates the alarm for normal   use.   Press ACCPT to accept the new system alarm configu-   ration. The next screen will ask you if there are more   alarms to set. If not, pressing ESC three times will take   you back to the Analyze mode main menu.   10. If O2 was pressed in step 8, the screen will resemble the   one in step 6. Use the SELEC, UP and DOWN keys to   configure alarm 5. When finished, press ACCPT. The   next screen will ask you if there are more alarms to set.   If not, pressing ESC three times will take you back to   the Analyze mode main menu.   TELEDYNE ANALYTICAL INSTRUMENTS   4-16   Operations 4   Logger Set-Up   1. Enter set-up from the Analyze mode by pressing   SETUP.   2. Press SET to set up the logger.   3. Enter the authorization code by using the SELEC, UP   and DOWN keys. Then press ACCPT.   4. Press LOG for the logger functions. Press ESC to return   to the Analyze mode main menu.   5. Press CLEAR if you want to clear the current logger   data. Press SETUP to continue setting up the logger.   6. If you chose SETUP in step 5, use SELEC to choose   logger ON/OFF, the time period between samples, or   the mode of measurement. Then choose EDIT to   change the values. After you are finished, press ACCPT,   which will take you to step 4.   The choices for the logger chart sampling mode (MODE   =) are on the next page.   TELEDYNE ANALYTICAL INSTRUMENTS   4-17   4 Operations   Note: The above screens use T to indicate two values: seconds and minutes. The first value is   the time period between samples on the logger, and is programmable. The second time   value is the width of the logger chart page, and is not programmable. There are 64   samples per logger chart page, and over 20 pages of past data which may be viewed.   Mode determines how to calculate the sample values shown on the logger chart:   Average:   Median:   The average of all measurements in the time period between samples.   The average of the minimum and the maximum values measured in the time   period between samples.   Maximum: The maximum of all values measured in the time period between samples.   Minimum: The minimum of all values measured in the time period between samples.   Sample:   The last sample measured in the time period between samples.   Press EDIT in step 6 and then use the UP and DOWN keys to select the above choices for the   logger chart mode.   TELEDYNE ANALYTICAL INSTRUMENTS   4-18   Operations 4   Set-Up Clock Functions   1. Enter set-up from the Analyze mode by pressing   SETUP.   2. Press SET to set the clock functions.   (If only one cell block is installed, TEST and INSTL   will be omitted.)   3. Enter the authorization code by using the SELEC, UP   and DOWN keys, and press ACCPT.   4. Press MORE.   5. Press CLOCK to select clock functions.   6. Enter the function that you would like to change.   SET: Sets the time.   ZERO:Sets the time for the next zero and the inter-   val times for automatic zeroing.   SPAN: Sets the time for the next span and interval   times for automatic spanning.   ALRMS: Displays alarm times and dates.   TELEDYNE ANALYTICAL INSTRUMENTS   4-19   4 Operations   7. If SET was pressed in step 1, the menu for setting the   time and date will appear. Use the SELEC, UP and   DOWN keys to set the time and date, and then press   ACCPT.   8. If ZERO was pressed in step 1, the menu for setting   AutoZeroing ON/OFF, time/date of the next zeroing,   and time interval for automatic zeroing will appear.   9. Use the SELEC, UP and DOWN keys to make selec-   tions, and press ACCPT.   10. If SPAN was pressed in step 1, the menu for setting   AutoSpanning ON/OFF, time/date for the next calibra-   tion, and interval for automatic calibration will appear.   11. Use the SELEC, UP and DOWN keys to make selec-   tions, and press ACCPT.   12. If ALRMS was pressed in step 1, a menu displaying the   most recent triggering time and date of each of the   alarms will appear. From here, press ESC three times to   return to the main menu screen.   TELEDYNE ANALYTICAL INSTRUMENTS   4-20   Operations 4   Changing Passwords for Remote Monitoring and Control   1. Enter set-up from the Analyze mode by pressing   SETUP.   2. Press SET to set the password(s).   3. Enter your authorization code by using the SELEC, UP   and DOWN keys, and press ACCPT. You must enter   the same authorization code set during the last COLD   START.   4. Press MORE. Press ESC to return to the Analyze mode   mainmenu.   5. Press PASS1 to change the remote monitoring pass-   word, or PASS2 to change the remote control pass-   word.   6. If you pressed PASS1 in step 5, the next screen will al-   low you to change the remote monitoring password.   Use SELEC, UP and DOWN to choose the new pass-   word, and then press ACCPT, which will take you to   step 4.   TELEDYNE ANALYTICAL INSTRUMENTS   4-21   4 Operations   7. If you pressed PASS2 in step 5, the next screen will al-   low you to change the remote control password. Use   SELEC, UP and DOWN to choose the new password,   and press ACCPT, which will take you to step 4.   TELEDYNE ANALYTICAL INSTRUMENTS   4-22   Operations 4   System Statistics   1. Enter set-up from the Analyze mode by pressing   SETUP.   ESC in any status page will take you to the last set-up   screen and eventually the main analyze mode.   PREV steps you back one status page.   NEXT moves forward one status page.   2. Press STAT for system status.   (If only one cell block is installed, TEST and INSTL   will be omitted.)   3. Status page one displays the active sensor (A or B), the   system time/date, and the system hardware configura-   tion. The number and date of the installed software ver-   sion are also shown.   Press PREV for the previous screen, or NEXT for the   next screen.   4. Status page two shows the O2 level set for spanning, the   cell strength factor, the type and time of the last span,   the time interval for scheduled spanning (AutoSpan),   and whether or not scheduled spanning is active. [The   cell strength factor is an indication of the amount of cell   life left in the cell, with 1.00 being the most and 0.00 be-   ing the least.]   5. Status page three shows the time of the last zero, the   time interval for scheduled zeroing (AutoZero), and   whether or not scheduled zeroing is active.   6. Status page four shows the settings for the five pro-   grammablealarms.   TELEDYNE ANALYTICAL INSTRUMENTS   4-23   4 Operations   7. Status page five shows the time and date of the last time   the alarm was triggered for each of the five program-   mablealarms.   8. Status page six shows the set-up for the serial port, and   allows you to reconfigure the serial port. Press EDIT to   bring up the following screen.   Press NEXT to go to status page 7.   9. Use the SELEC, UP and DOWN keys to select the de-   sired serial port configuration, and press ACCPT.   Baud rate: 300–2400   Parity:Odd, Even, or None   Mode: Monitor, Remote, or Off   ASCII Data Bits:7 or 8   Stop Bits: 1 or 2   Modem: On, Off or Auto   Output Interval for Monitoring: 1–32000 sec.   Transmit/Receive Line Terminators: CR or   CRLF   10. Status page seven allows for configuration of the mo-   dem/remote. Press EDIT to change the configuration.   • • • • • Modem Mode: AUTO/OFF/ON   Modem Status: NO (set by system)   Login Access: LOCAL (set by system)   Phone Jack: SINGLE/MULTI, RJ12   Comm. Std. for 1200B: BELL USA/CCITT   11. After pressing EDIT, a cursor will appear and you will   be able to edit the settings. Press ACCPT when done.   Pressing INIT in status page seven (step 10) takes you   to the screen at the top of the next page and gives you   the option of either checking for a modem, or setting up   the modem. Set-up sets up the modem for auto answer   and other communication protocols.   TELEDYNE ANALYTICAL INSTRUMENTS   4-24   Operations 4   12. After checking for the modem, pressing ESC returns   you to status page seven in step 10.   If INIT was pressed in status page 7 and then SETUP   pressed in step 11, pressing ACK will return you to this   screen. Press ESC to return to status page seven in step   10.   13. If you have purchased the parallel printer option, status   page eight allows for configuration of the port. Press   EDIT to change the configuration.   • • Mode: ON/OFF   Interval: 3–32000 seconds   NOTE: The printer report will always be in the same for-   mat—O2 Level, Range, Active Sensor, Time & Date.   14. After pressing EDIT, a cursor will appear and you will   be able to edit the settings.   15. Status page nine shows whether or not the last and next   start-ups were WARM or COLD. The next start-up may   be changed by pressing the COLD or WARM keys.   16. Status page ten displays logger data. Press VIEW to see   the logger data history. Press NEXT to return to status   page one. Press ESC twice to return to the Analyze   mode main menu.   17. If VIEW was pressed in step 13, logger data may be   viewed by using the +, –, REV and FWD keys. The O2   level, range, and time are shown for each data point.   NEXT takes you to status page two.   REV and FWD change the date by one page (or 64   points), while – and + go through the data one point at a   time within a single page. Press ESC to return to status   page ten.   TELEDYNE ANALYTICAL INSTRUMENTS   4-25   4 Operations   TELEDYNE ANALYTICAL INSTRUMENTS   4-26   Maintenance & Troubleshooting 5   Maintenance & Troubleshooting   5.1 Routine Maintenance   1. Calibrate the analyzer at least once each week during the first four weeks   and then once every two weeks during the next eight weeks. Afterwards,   the analyzer should be calibrated once a month.   2. Check the pressure of the air supply to the analyzer. Air pressure should be   maintained between 70 and 80 psig.   Note: The instrument air pressure must be greater than 70 psig in order to open   and close the pneumatic solenoid valves.   3. Check the sample pressure. Pressure should be maintained between 5 and   10 psig. Sample and span gas pressures should be within 2 psig of each   other.   5.1.1 Sensor Maintenance   Sensor maintenance in the 3160 is limited to replacing the sensor(s) when replace-   ment is indicated. Consult the troubleshooting section for symptoms that indicate that the   sensor needs replacing.   A guide to the relative life left in the cell can be found on status page 2. See “Cell   Strength Factor” at the beginning of Chapter 3.0.   In units with 2 cell blocks, if sensor A fails, sensor B will not automatically switch   online. This must be done manually through the Install Sensor and Test Alternate Sensor   menu.   5.1.2 Scrubber Maintenance   The oxygen scrubber is a replaceable component. With normal use, the scrubber   usually lasts for several years.   NOTE: Do not zero in ranges other than 0-1 ppm, as zeroing in higher ranges will   exhaust the scrubber.   TELEDYNE ANALYTICAL INSTRUMENTS   5-1   5 Maintenance & Troubleshooting   Analyzers equipped with a scrubber should only be used with inert gases and saturat-   ed hydrocarbons.   5.2 Troubleshooting   To troubleshoot the analyzer, identify the problem in the following list and then   follow the procedures to correct the problem. If you cannot identify the problem, or if you   cannot resolve the problem after following the corrective procedure, call a service repre-   sentative.   Symptom   Cause   Correction   Faulty LCD or LED.   Return the unit to factory.   LCD or LED display will not   light.   Remove analyzer cover. Adjust   potentiometer located behind   LCD display.   Display too bright.   Faulty electrical connection.   Remove analyzer cover. Check   PC board connections and   make sure that they are firmly   seated.   No analog output.   Bad sensor.   High O2 level.   Replace sensor.   Check zero gas inlet connec-   tion.   Zero gas signal >2 PPM.   Bad sensor.   Replace sensor.   Slow response/recovery time.   Re-zero sensor.   Displays “--” value.   a) The wrong span gas   concentration has been   entered.   a) Re-enter span gas concen-   tration.   Can't calibrate (with sensor   installed). Display reads “cell   too strong” or “cell too   weak.”   b) Replace or reinstall sensor.   a) The pneumatic air supply   pressure is too low, making   the valves stick.   a) Raise air supply pressure to   at least 70 psig.   No sample gas flow.   b) The sample supply pressure   is too low.   b) Check back panel connec-   tions. Raise sample supply   to at least 5 psig but within   2 psig of the span gas   pressure.   TELEDYNE ANALYTICAL INSTRUMENTS   5-2   Maintenance & Troubleshooting 5   Correction   Symptom   Cause   c) Make sure metering valve   (flow set knob on front   panel) is sufficiently open.   Back panel vent is blocked.   See a) above.   Clear and check connection to   back panel vent.   Will not switch modes.   See a) above.   Sensor output does not drop   to a stable, low value during   zeroing or spanning.   a) Sensor has been exposed   to high concentrations of O2   for an extended time period.   a) Replace sensor and/or   wait for stable reading.   b) Bad sensor.   b) Replace sensor.   c) Leak in system.   c) Check sensor block clamp.   Make sure the sensor block   is tightly closed. Check the   gas supply tank, line and   connections for leaks.   Unit will not turn on although   it is plugged in and the   Power Switch is ON (at "I"   position).   Fuse is blown.   Check fuse(s) in back panel   fuse block assembly located   above the power cord recep-   tacle.   TELEDYNE ANALYTICAL INSTRUMENTS   5-3   5 Maintenance & Troubleshooting   TELEDYNE ANALYTICAL INSTRUMENTS   5-4   Appendix   Appendix   A.1 Specifications   (Subject to change without notice.)   Ranges:   (Six ranges with AutoRanging on all ranges)   1:   2:   3:   4:   5:   6:   0–1 part per million (ppm) O2   0–10 ppm O2   0–100 ppm O2   0–1000 ppm O2   0–1 % O2   0–25 % O2   Accuracy:   2% of full scale when calibrated and operated at the same constant   temperatureandpressure.*   5% of full scale over the operating temperature range once tempera-   tureequilibriumisreached.*   * Additional 0.04 ppm on 0-1 ppm range, generated by internal scrub-   ber (which is necessary to maintain these specifications).   Response Time: 90 % of final reading in less than 61 seconds at 25 °C   OperatingTemperature:0-50°C (32-122 °F)   Analog Signal Outputs:   For percent of full scale indication:   one 0–1 V dc non-isolated signal   one 4–20 mA dc isolated signal   Forrangeidentification:   one 0–1 V dc non-isolated signal   one 4–20 mA dc isolated signal   Digital Data Lines:   One RS-232C serial interface for 2-way communications   with separate host computer—for remote monitoring and   control of all functions.   Alarms:   Fiveuser-programmableabsolute-readingalarmsetpoints,   with Form C, SPDT relays (3A resistive).   TELEDYNE ANALYTICAL INSTRUMENTS   A-1   Appendix   Mounting and Enclosure:   19" relay rack, 20.3 cm (12.25 inches) high,   for general purpose (nonhazardous) areas.   (Class I, Division II, hazardous area models   available on special order.)   Sensor: Class B-2, B2C, A2, or A2C, Micro-Fuel   Cells.   Altitude:   1,609 m   RelativeHumidy: Up to 99%   Power Requirements: 100, 120, 220, or 240~ 50-60 Hz 1.5 A MAX   Use 250 V 3.0 A T Fuse for 110, 120 V~   Use 250 V 1.6 A T Fuse for 220, 240 V~   TELEDYNE Analytical Instruments   16830 Chestnut Street   City of Industry, CA 91749-1580   Phone (626) 934-1500,   Fax (626) 961-2538   TWX (910) 584-1887 TDYANYL COID   Web:   A-2   TELEDYNE ANALYTICAL INSTRUMENTS   Appendix   A.2 Recommended Spare Parts List   Description P/N   Micro-FuelCellsensor C6689B2   C6689B2C   C6689A2   C6689A2C   O-ring   O165   Suctionpurifier P670   Solenoidvalve V462   Pneumatic valve upper body B435   Hose fitting tee T978   Fuse, 3A T Type F68   Fuse, 1.6A T Type (European) F768   Scrubber C58750   A.3 Drawing List   D-56534   C-64819   OutlineDrawing   InterconnectionDrawing   NOTE: The MSDS on this material is available upon request   through the Teledyne Environmental, Health and   Safety Coordinator. Contact at (626) 934-1592   TELEDYNE ANALYTICAL INSTRUMENTS   A-3   Appendix   A.4 Material Safety Data Sheet I   Section I – Product Identification   Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3,   and A-5.   Electrochemical Oxygen Sensors, all classes except R-19.   Mini-Micro-Fuel Cells, all classes.   Manufacturer: Teledyne Analytical Instruments   Address:   16830 Chestnut Street, City of Industry, CA 91749   Phone: (818) 961-9221   Customer Service: Extension 222   Environmental Health   and Safety: Extension 230   Date Prepared : 2/12/96   Section II – Hazardous Ingredients/Composition   Material or   Component   TLV   C.A.S. # Quantity   OSHA PEL ACGIH   Lead (Pb)   7439-92-1 3–20 gms   0.05 mg/m3 0.15 mg/m3   Potassium Hydroxide1310-58-31–5 ml None   2 mg/m3   Solution 15% (KOH)   Section III – Physical/Chemical Characteristics   Material   Appearance   or Compo-   nent   Odor   Specific Vapor   Gravity Pres-   sure   11.34   na   Solubility   in Water   Melting Density Evap.   Boiling   Point (°C)   Point   Rate   (°C)   Solid, silver   gray, odorless   na   na   na   Insoluble   Complete   1744   1320   328   360   Lead   White or   slightly   yellow,   no odor   2.04   na   na   Potassium   Hydroxide   A-4   TELEDYNE ANALYTICAL INSTRUMENTS   Appendix   Section IV – Fire and Explosion Hazard Data   Flash Point:   na   Flammable Limits:   na   LEL: na   UEL:   na   Extinguishing Media:   Use extinguishing media appropriate to surrounding fire   conditions. No specific agents recommended.   Special Fire Fighting   Equipment:   Wear NIOSH/OSHA approved self-contained breathing   apparatus and protective clothing to prevent contact with   skin and eyes.   Unusual Fire and Explosion   Hazards:   Emits toxic fumes under fire conditions.   Section V – Reactivity Data   Stability:   Stable   Incompatibilities:   Aluminum, organic materials, acid chlorides, acid   anhydrides, magnesium, copper. Avoid contact with acids   and hydrogen peroxide > 52%.   Hazardous Decomposition of   Byproducts:   Toxic fumes   Hazardous Polymerization:   Will not occur.   Section VI – Health Hazard Data   Routes of Entry:   Inhalation:   Ingestion:   Highly unlikely   May be fatal if swallowed.   Skin:   The electrolyte (potassium hydroxide) is corrosive; skin   contact may cause irritation or chemical burns.   Eyes:   The electrolyte (potassium hydroxide) is corrosive; eye   contact may cause irritation or severe chemical burns.   Acute Effects:   The electrolyte is harmful if swallowed, inhaled or   adsorbed through the skin. It is extremely destructive to   tissue of the mucous membranes, stomach, mouth, upper   respiratory tract, eyes and skin.   Chronic Effects:   Prolonged exposure with the electrolyte has a destructive   effect on tissue.   Chronic exposure to lead may cause disease of the blood   and blood forming organs, kidneys and liver, damage to   the reproductive systems and decrease in fertility in men   and women, and damage to the fetus of a pregnant   woman. Chronic exposure from the lead contained in this   product is extremely unlikely.   TELEDYNE ANALYTICAL INSTRUMENTS   A-5   Appendix   Signs and Symptoms of   Exposure:   Contact of electrolyte with skin or eyes will cause a   burning sensation and/or feel soapy or slippery to touch.   Other symptoms of exposure to lead include loss of sleep,   loss of appetite, metallic taste and fatigue.   Carcinogenicity:   OSHA:   Lead is classified by the IARC as a class 2B carcinogen   (possibly carcinogenic to humans).   Where airborne lead exposures exceed the OSHA action   level, refer to OSHA Lead Standard 1910.1025.   NTP:   na   Lead exposure may aggravate disease of the blood and   blood forming organs, hypertension, kidneys, nervous   and possibly reproductive systems. Those with preexist-   ing skin disorders or eye problems may be more suscepti-   ble to the effects of the electrolyte.   Medical Conditions Generally   Aggravated by Exposure:   Emergency First Aid Procedures:   In case of contact with the skin or eyes, immediately   flush with plenty of water for at least 15 minutes and   remove all contaminated clothing. Get medical attention   immediately.   If ingested, give large amounts of water and DO NOT   INDUCE VOMITING. Obtain medical attention immedi-   ately.   If inhaled, remove to fresh air and obtain medical   attention immediately.   Section VII – Precautions for Safe Handling and Use   NOTE:The oxygen sensors are sealed, and under normal circum-   stances, the contents of the sensors do not present a health   hazard. The following information is given as a guide in the   event that a cell leaks.   Before opening the bag containing the sensor cell, check   Protective measures   during cell replacement:   the sensor cell for leakage. If the sensor cell leaks, do not   open the bag. If there is liquid around the cell while in   the instrument, wear eye and hand protection.   Cleanup Procedures:   Wipe down the area several times with a wet paper towel.   Use a fresh towel each time. Contaminated paper towels   are considered hazardous waste.   A-6   TELEDYNE ANALYTICAL INSTRUMENTS   Appendix   Section VIII – Control Measures   Eye Protection:   Chemical splash goggles   Rubber gloves   Hand Protection:   Other Protective Clothing:   Ventilation:   Apron, face shield   na   Section IX – Disposal   Both lead and potassium hydroxide are considered poisonous substances and are regulated under   TSCA and SARA Title III.   EPA Waste Number:   California Waste Number:   DOT Information:   D008   181   RQ Hazardous Waste Solid N.O.S. (Lead), 9, UN3077,   PG III   Follow all Federal, State and Local regulations.   Section X – References   Material Safety Data Sheets from J.T. Baker Chemical, Aldrich, Mallinckrodt, ASARCO   U.S. Department of Labor form OMB No. 1218-0072   Title 8 California Code of Regulations   TSCA   SARA Title III   CFR 49   CFR 29   CFR 40   NOTE:The above information is believed to be correct and is offered for   your information, consideration, and investigation. It should be   used as a guide. Teledyne Analytical Instruments shall not be held   liable for any damage resulting from handling or from contact with   the above product.   TELEDYNE ANALYTICAL INSTRUMENTS   A-7   Appendix   A.5 Material Safety Data Sheet II   Section I – Product Identification   Product Name:   Micro-FuelCells   Mini-Micro-FuelCells,allclasses   Super Cells, all classes except T–5F   Oxygen Sensors, all classes.   Manufacturer: TeledyneAnalyticalInstruments   Address:   16830 Chestnut Street, City of Industry, CA 91749   Phone: (818) 961-9221   Date Prepared or Last Revised: 08/08/91   Emergency Phone Number: (818) 961-9221   Section II – Physical and Chemical Data   Chemical and Common Names: Potassium Hydoxide (KOH), 15% (w/v)   Lead (Pb), pure   CAS Number: KOH 1310–58–3   Pb 7439–92–1   KOH (15% w/v)   MeltingPoint/Range: –10 to 0 °C   BoilingPoint/Range: 100 to 115 °C   Pb (pure)   328 °C   1744 °C   11.34   SpecificGravity: 1.09 @ 20 °C   pH:   SolubilityinWater: Completelysoluble   Percent Volatiles by Volume: None   Appearance and Odor: Colorless,odorlesssolution   >14   N/A   Insoluble   N/A   Grey metal, odor-   less   A-8   TELEDYNE ANALYTICAL INSTRUMENTS   Appendix   Section III – Physical Hazards   Potential for fire and explosion: The electrolyte in the Micro–Fuel Cells is not flam-   mable. There are no fire or explosion hazards associated with Micro–Fuel Cells.   Potential for reactivity: The sensors are stable under normal conditions of use. Avoid   contact between the sensor electrolyte and strong acids.   Section IV – Health Hazard Data   Primary route of entry:   Ingestion,eye/skincontact   Exposurelimits:OSHA PEL: .05 mg/cu.m. (Pb)   ACGIH TLV:   2 mg/cu.m. (KOH)   Effects of overexposure   Ingestion:   The electrolyte could be harmful or fatal if swal-   lowed.   Oral LD50 (RAT) = 3650 mg/kg   Eye: The electrolyte is corrosive; eye contact could result   in permanent loss of vision.   Dermal: The electrolyte is corrosive; skin contact could result   in a chemical burn.   Inhalation: Liquidinhalationisunlikely.   Signs/symptoms of exposure: Contact with skin or eyes will cause a burning sensa-   tion and/or feel soapy or slippery to touch.   Medicalconditions   aggravated by exposure: None   Carcinogenity: NTP Annual Report on Carcinogens: Not listed   LARC Monographs: Not listed   OSHA: Not listed   Other health hazards: Lead is listed as a chemical known to the State of   California to cause birth defects or other reproductive   harm.   TELEDYNE ANALYTICAL INSTRUMENTS   A-9   Appendix   Section V – Emergency and First Aid Procedures   Eye Contact:   Skin Contact:   Flush eyes with water for at least 15 minutes and get immediate   medicalattention.   Wash affected area with plenty of water and remove contaminated   clothing. If burning persists, seek medical attention.   Give plenty of cold water. Do not induce vomiting. Seek medical at-   tention. Do not administer liquids to an unconscious person.   Liquidinhalationisunlikely.   Ingestion:   Inhalation:   Section VI – Handling Information   NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents   of the sensors do not present a health hazard. The following information is given   as a guide in the event that a cell leaks.   Protectiveclothing:   Rubber gloves, chemical splash goggles.   Clean-up procedures: Wipe down the area several times with a wet paper towel.   Use a fresh towel each time.   Protectivemeasures   duringcellreplacement: Before opening the bag containing the sensor cell, check the   sensor cell for leakage. If the sensor cell leaks, do not open   the bag. If there is liquid around the cell while in the instru-   ment, put on gloves and eye protection before removing the   cell.   Disposal:   Should be in accordance with all applicable state, local and   federalregulations.   NOTE: The above information is derived from the MSDS provided by the manufacturer.   The information is believed to be correct but does not purport to be all inclusive   and shall be used only as a guide. Teledyne Analytical Instruments shall not be   held liable for any damage resulting from handling or from contact with the   aboveproduct.   A-10   TELEDYNE ANALYTICAL INSTRUMENTS   |