Welch Allyn Medical Diagnostic Equipment 12500 User Manual

Welch Allyn 12500  
Binocular Indirect Ophthalmoscope  
Power Source  
OWNER’S MANUAL  
1. Wall/Desk Power Source  
The Welch Allyn No. 74250 Wall/Desk Power Source is designed with  
versatility in mind. There are several mounting options.  
1. Wall mount  
2. Desk top mount  
3. Desk edge mount  
Features of the Wall/Desk Power Source include:  
Condensing lens  
storage area  
On/Off switch  
rheostat control  
Scope turns on  
when lifted from  
cradle  
to headset  
UL/IEC/CSA approved  
plug-in transformer  
Wall Mounting  
Fasten the mounting plate to the wall with the appropriate screws (wood  
screws are provided). Flat side of the plate should be facing the wall,  
long sides should be horizontal. When the screws are tightened, there  
will be a gap between the screw head and the surface of the plate.  
Mount the Wall/Desk bracket by fitting round holes in back plate over  
screw heads, then pushing downward on case. This slides the tapered  
slots in the back over the shoulders in the mounting plate.  
1-1  
Desk Top Usage  
The Welch Allyn Wall/Desk Power Source can be set on any flat sur-  
face. To prevent scratching and sliding, the small rubber pads should  
be attached to the bottom of the unit. Remove the pads from their  
backing and place them on the outermost corners of the unit’s base.  
Desk Top  
Mounting  
Desk Edge Mounting  
NOTE: Do not use rubber pads if mounting to the EDGE of a desk  
or table.  
The Welch Allyn Wall/Desk Power Source can also be mounted on  
the edge of a desk or cabinet. This will allow the Binocular Indirect  
Ophthalmoscope (BIO) to hang unobtrusively, close to your task area.  
To mount the Wall/Desk bracket on the edge of a desk or cabinet, it  
must be secure so the weight of the BIO does not tip it over. To secure  
the unit to the desk top, apply the two-sided tape found in the acces-  
sory bag to the back edge of the base of the Wall/Desk bracket.  
NOTE: Make sure the surface is clean before attaching. Locate the  
bracket so the cradle is hanging over the edge of the desk. When the  
bracket is in the desired position, apply firm pressure to it so it becomes  
properly and evenly seated.  
NOTE: If the Wall/Desk bracket ever needs to be relocated, it can be  
removed by prying up on one end. Any tape residue can be removed  
with an alcohol wipe.  
Desk Edge Mounting  
1-2  
Operation  
Once the Wall/Desk bracket is mounted, the unit is ready for use. Plug  
the Power Source into an appropriately-grounded outlet. Then plug the  
cord from the Power Source into the receptacle labeled “in” on the  
Wall/Desk bracket. To complete the connection, screw the external  
sleeve down until it stops — this will secure the plug in place. The cord  
from the BIO should be connected to the “out” receptacle in the same  
manner that the power source is connected. Be sure to complete the  
connection by screwing the external plug sleeve down until it stops.  
ON  
OFF  
IN  
OUT  
When the BIO is in the cradle of the Wall/Desk bracket, the unit is  
always off and the Light Emitting Diode (LED) on the front of the unit  
is turned off. When lifted from the cradle (with on/off rheostat in on  
position), power to the BIO is on and the LED will illuminate. The rheo-  
stat control/on-off switch controls power to the instrument as well as  
illumination intensity.  
Cleaning  
All surfaces may be cleaned with a soft cloth moistened with alcohol.  
Maintenance  
Periodic technical inspections should be performed by qualified per-  
sonnel. This equipment should be inspected at least once every six  
months, more often if used under adverse conditions.  
1-3  
2. Portable Power Source  
The Welch Allyn No. 74260 Portable Power Source provides complete  
portability and ease of use.  
Features of the Portable Power Source are:  
Charging indicator  
On/Off switch  
and rheostat  
control  
Belt clip  
In use/time  
remaining  
indicator  
Back  
Front  
UL/IEC/CSA approved  
direct plug-in recharger  
Operation  
When fully charged and operated at full  
intensity, the Portable Power Source has  
approximately fifty minutes of on-time. When  
the BIO is used at half intensity, the on-time  
is approximately seventy-five minutes.  
To recharge the Portable Power Source,  
plug the recharger into the appropriate outlet  
and into the proper receptacle on the  
battery pack. When the battery pack is re-  
charging, the charge indicator will glow yel-  
low. The battery cannot be overcharged.  
NOTE: Be sure the on/off switch-rheostat  
control is turned OFF during recharging.  
Do not attempt to plug the recharging  
cord into the same receptacle the BIO  
cord plugs into.  
To use the battery pack, plug the BIO cord  
into the proper receptacle. The external  
sleeve on the plug needs to be screwed  
down tightly to ensure a non-interrupted  
connection. Turn on the on/off switch-  
rheostat control and adjust to the desired  
intensity. When the battery pack is on,  
the power indicator will turn green. When  
approximately four to seven minutes of  
ON-time remain, this indicator will change  
to amber.  
1-4  
To Change the Battery  
Unscrew the two screws which hold the housings  
together. Be certain not to touch the green circuit  
board or any of the components on it. Disconnect  
the old battery pack from the connector and  
attach a new battery pack. Install the battery  
pack with label facing you, as shown. Be  
sure the battery pack connector and wire  
lies across the bottom edge of the housing,  
as shown. Use only Welch Allyn No. 72240  
rechargeable battery pack.  
BATTERY  
Cleaning  
DISCONNECT  
CONNECT  
All external surfaces may be cleaned with  
a soft cloth moistened with alcohol.  
Symbol Descriptions  
Attention. Read Operating Manual for  
Cautions and Instructions for Use.  
TYPE B Equipment.  
TYPE BF Equipment.  
Double Insulation.  
Indoor Use Only.  
Licensed under TÜV to conform with IEC 601-1.  
Recognized component UL 2601-1.  
+
Internally Powered.  
The CE mark on this product indicates it has been  
tested to and conforms with the provisions noted  
within the 93/42/EEC Medical Device Directive.  
Authorized European Representative Address:  
European Regulatory Manager  
Welch Allyn, Ltd.  
Kells Road, Navan,  
County Meath, Republic of Ireland  
Tel. 353 46 28122  
Fax 353 46 28536  
1-5  
Specifications No. 74260  
Portable Power Source  
Power Pack:  
Model Number:  
Output Voltage:  
Output Current:  
On-time:  
74201  
4.8 Vdc  
2.55 A  
42-50 minutes @ high setting.  
On-time not to exceed 10 minutes,  
off-time not less than 15 minutes.  
Low Voltage Indication Time: 4-7 minutes  
Recharge Time:  
Recharge Current:  
Approvals for:  
10 hrs minimum  
225 ± 50 mA  
UL-2601-1, CAN/CSA C22.2  
No. 601.1, IEC 601-1,  
IEC 601-1-2, CE  
Charger:  
Model Number:  
Input:  
74280  
95-130 Vac, 60 Hz  
0.08 A @ 120 Vac, 60 Hz  
Output:  
5.3 Vdc, 0.24 A @ 120 Vac  
SPT, 2 conductor, AWG #22  
Output Cable:  
Approvals for:  
UL-2601-1, CAN/CSA C22.2  
No. 601.1, IEC 601-1,  
IEC 601-1-2, CE  
Rechargeable Battery:  
Model Number:  
Type of Cells:  
72240  
Nickel-cadmium battery (NiCad)  
4.8 Vdc, 1.8 Ahr.  
Nominal Voltage:  
Approvals for:  
UL-2601-1, CAN/CSA C22.2  
No. 601.1, IEC 601-1,  
IEC 601-1-2, CE  
1-6  
Specifications No. 74250  
Wall/Desk Power Source  
Plug-In Power Transformer:  
Model Number:  
Input:  
74290  
95-130 Vac, 60 Hz  
0.18 A @ 120 Vac, 60 Hz  
Output:  
5.15 Vac, 2.60 A @ 120 Vac  
Output Cable:  
Approvals for:  
SJT/SVT, 2-conductor, AWG #18  
UL-2601-1, CAN/CSA C22.2 No. 601.1,  
IEC 601-1, IEC 601-1-2, CE  
Wall/Desk Bracket:  
Model Number:  
Input:  
74241  
5.15 Vac, 60 Hz, 2.60 A  
Output at Output Jack: 4.8 Vac, 2.55 A  
Approvals for: UL-2601-1, CAN/CSA C22.2 No. 601.1,  
IEC 601-1, IEC 601-1-2, CE  
1-7  
Service and Warranty Information  
Service  
For customers in North America, please return instruments requiring  
service to a Welch Allyn Technical Service Department listed below or  
to an authorized Welch Allyn distributor.  
Technical Service Department  
Welch Allyn, Inc.  
4341 State Street Road  
Skaneateles Falls, NY 13153-0220  
U.S.A.  
Telephone: 800-669-9771  
Fax: 315-685-4653  
For customers outside of North America, return your unit to a local,  
authorized Welch Allyn distributor or to your nearest Welch Allyn  
service center.  
Rechargeable Battery Warranty  
Welch Allyn nickel-cadmium batteries are guaranteed by Welch Allyn  
for two years from date of manufacture (when used in Welch Allyn  
instruments only). Defective batteries will be replaced on a pro rata  
basis should failure occur prior to expiration date on battery.  
1-8  
1-9  
Welch Allyn, Inc.  
4341 State Street Road  
P.O. Box 220  
Skaneateles Falls, NY 13153-0220  
Telephone: 315-685-4560  
1-800-535-6663 (only in U.S.A.)  
FAX: 315-685-3361  
Part No. 741916-3 Rev. A  
Printed in U.S.A.  

Viking VFP12 User Manual
Texas Instruments El Dorado JA BT2040 User Manual
Sony TCM 343 User Manual
Sony Ericsson P910 User Manual
Sony DSR PD100 User Manual
Sony Camcorder CD TRV35 User Manual
Sony Camcorder BRC Z300 User Manual
Sharp LL 171A LL 171G User Manual
Sharp Car Video System PN S655 User Manual
Samsung SCL860 User Manual