| Olympus Suction Pump Suction Cleaning Adapter MAJ 222 User Manual | 
| To use this chart as a wallchart, cut   off the top of the chart as illustrated.   To insert this chart into a binder, punch out   the holes and fold the chart as illustrated.   BRONCHOSCOPE DISINFECTION GUIDE   EXERA™ BF-160 Series and EVIS™ BF-240 Series Bronchovideoscopes   Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information   WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the   detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.   HIGH-LEVEL DISINFECTION   2 1 3 4 Suction Cleaning Adapter   Fill basin with disinfectant. Attach a   suction cleaning adapter (MAJ-222)   to the previously cleaned broncho-   scope. Immerse completely in   disinfectant solution.   Attach a syringe to the adapter.   While immersed, detach the adapter   Soak all equipment according to   the time and temperature   recommended by the disinfectant   manufacturer. Use a timer to   measure the exact contact time.   Pull the plunger of the syringe to   fill the instrument/suction channels   and channel cleaning adapter   with disinfectant solution.   from the bronchoscope. Use a clean,   lint-free cloth to wipe away any bub-   bles clinging to external surfaces.   RINSING   To Suction Pump   Suction Cleaning   Adapter (MAJ-222)   5 6 7 8 To Suction Pump   Alcohol   Connect the suction cleaning   adapter and syringe. Inject air   to remove disinfectant from the   instrument/suction channels,   then remove the bronchoscope   from the disinfectant.   Disconnect the suction cleaning   adapter and immerse the   bronchoscope and adapter in   sterile water. Using a sterile,   lint-free cloth, thoroughly rinse   and wipe all surfaces.   Connect suction cleaning adapter   to bronchoscope and to a tube   from a suction pump. Aspirate   water through channels for 30   seconds. Remove from water and   aspirate air through channels for   60 seconds.   Immerse the bronchoscope's distal   end in 70% alcohol and aspirate   alcohol for 5 seconds. Remove the   distal end from the alcohol and   aspirate air for 20 seconds.   DRYING   TRANSPORTATION   9 10   Turn the suction pump off. Hold the control   section with the openings pointing down and   disconnect the suction cleaning adapter.   Dry the bronchoscope and adapter with an   alcohol-moistened cloth. Make sure that the   equipment is completely dry prior to storage.   Hold all parts of the endoscope when carrying. Do   not let the ends of the bronchoscope swing freely.   NOMENCLATURE   IMPORTANT   • Wear appropriate personal protective   equipment during reprocessing.   Single-Use Suction Valve (MAJ-209)   Suction Cylinder   Light Guide   Universal Cord   • Autoclave all autoclavable accessories.   • Store the bronchoscope hanging vertically   with all valves and caps removed.   Remote Switches   • Use a compatible disinfectant and test   it for proper concentration as recom-   mended by the manufacturer.   Electrical Connector   Venting Connector   Single-Use Biopsy Valve (MAJ-210)   Instrument/Suction Channel Port   • Do not use the carrying case to store   "patient-ready" endoscopes. Any broncho-   scope placed in the carrying case must be   reprocessed prior to patient use.   Water-Resistant Cap   (MH-553)   Insertion Tube   Control Section   Boot   Bending Section   Clean the bronchoscope   Distal Tip   UP/DOWN Angulation Control Lever   meticulously prior to disinfection.   |