RosemountModel 700 Breather Drain  
Optional Breather Drain  
Safety and information notices  
DANGER  
WILL CAUSE DEATH  
Failure to follow this warning will result in death or serious injury to personnel.  
WARNING  
DANGER TO PERSONNEL  
Failure to follow this warning may result in serious injury to personnel.  
CAUTION  
MAY CAUSE DAMAGE TO EQUIPMENT  
Failure to follow this warning may result in damage to the equipment.  
NOTICE  
Important messages will appear in this format.  
3
4
Contents  
Installation Instructions  
October 2019  
SK-07778  
vi  
Emerson.com/RosemountGasAnalysis  
Installation Instructions  
Certifications  
SK-07778  
October 2019  
1
Certifications  
1.1  
Certifications for the Rosemount Model 700  
Breather Drain  
These instructions provide guidance for the safe installation of the optional Ex D Breather  
Drain accessory for the Rosemount Model 700 Gas Chromatograph.  
The breather drain plug is the following type:  
Table 1-1: Breather/Drain  
Specification  
Description  
Manufacturer:  
Emerson P/N:  
Hoffman Enclosures Inc.  
2-3-0700-739 (ATEX and IEC configurations only)  
ATEX Certificate:  
ITS13ATEX17900X  
Ex d IIC Gb T6  
(Ta = -20 °C to 60 °C)  
IECEx Certificate:  
IECEx ITS 13.0049X  
Ex d IIC Gb T6  
(Ta = -20 °C to 60 °C)  
Manufacturer:  
Emerson P/N:  
Killark  
2-3-0700-740 (US and Canada configurations only)  
CSA/US:  
Class I, Zone 1, AEx d IIC, Enclosure Type 4. Class I, Division 1, Groups B,  
C, and D, IP66  
CSA/CANADA:  
Class I, Zone 1, Ex d IIC, Enclosure Type 4. Class I, Division 1, Groups B,  
C, and D, IP66  
NEMA  
Type 4, 4X  
IP66  
IEC 60529  
Model 700 Breather Drain  
7
   
Certifications  
Installation Instructions  
October 2019  
SK-07778  
8
Emerson.com/RosemountGasAnalysis  
Installation Instructions  
Specifications  
SK-07778  
October 2019  
2
Specifications  
2.1  
Rosemount Model 700 O-ring specifications  
Table 2-1: O-ring Materials and Temperature Limits  
O-ring Material(1)  
Nitrile  
Limiting Temperature  
-20 to 80 ℃  
EPDM  
-30 to 125 ℃  
-20 to 100 ℃  
-30 to 150 ℃  
-50 to 150 ℃  
Neoprene  
Silicone  
Fluorosilicone  
(1) It is the responsibility of the Emerson GC Service Team to ensure that the appropriate O-ring  
material is selected according to the type of application.  
Model 700 Breather Drain  
9
     
Specifications  
Installation Instructions  
October 2019  
SK-07778  
10  
Emerson.com/RosemountGasAnalysis  
Installation Instructions  
Installation  
SK-07778  
October 2019  
3
Installation  
3.1  
Install the breather drain  
This installation procedures give general instructions for the safe installation of the  
optional Ex d breather drain.  
Important  
The breather drain accessory option is available for GCs without the optional LSIV only.  
Figure 3-1: Rosemount Model 700 Breather Drain Installation Location  
A. Breather drain installation location  
Model 700 Breather Drain  
11  
   
Installation  
Installation Instructions  
October 2019  
SK-07778  
Procedure  
1. Install the breather drain in accordance with relevant local code and standards in  
the location as shown in Figure 3-2 .  
Figure 3-2: Rosemount Model 700 Right Side View  
A. Breather drain plug installation location  
2. De-energize the GC before commencing installation or maintenance procedures.  
3. Installation of the breather drain is carried out by a recognized Emerson Service  
Technician, familiar with hazardous area classifications and trained in GC  
maintenance practices.  
4. Remove dust, moisture, or debris from the area around the breather drain ¾ in. NPT  
opening (see Figure 3-2) to maintain the correct IP rating of the Rosemount Model  
700 Gas Chromatograph.  
12  
Emerson.com/RosemountGasAnalysis  
 
Installation Instructions  
Installation  
SK-07778  
October 2019  
5. Select the correct O-ring and fit it to the breather drain. Refer to Rosemount Model  
6. Ensure the O-ring is properly seated in the groove in the breather plug.  
Figure 3-3: Breather Drain O-rings for ATEX and IEC Applications  
A. Adapter with O-ring  
B. Breather drain with O-ring  
7. Thread the breather drain into the ¾ in. NPT opening in the GC enclosure.  
Model 700 Breather Drain  
13  
Installation  
Installation Instructions  
October 2019  
SK-07778  
8. Hand-tighten using a in. combination wrench or spanner and ensure the drain is  
properly seated and securely tightened.  
Figure 3-4: Breather Drain Assembly  
A. Tapped hole  
B. External wall of GC enclosure  
C. Breather drain  
D. O-ring seal  
9. The recommended installation torque for the breather drain is as follows:  
Thread Size  
Recommended Installation Torque  
¾ in. NPT  
Wrench tighten  
14  
Emerson.com/RosemountGasAnalysis  
Installation Instructions  
Maintenance  
SK-07778  
October 2019  
4
Maintenance  
4.1  
Maintaining the breather drain  
This section provides general maintenance for the Rosemount Model 700 breather drain.  
Procedure  
1. Check the breather drain during routine maintenance of the enclosure.  
2. Remove surface debris that may accumulate on the internal O-ring dust seal using  
compressed air.  
3. Clean the exterior drainage holes, if blocked, using compressed air.  
4. During maintenance or routine inspection, replace the breather drain if it is  
damaged or broken.  
5. Remove dust, moisture, or debris from the area around the breather drain ¾ in. NPT  
opening (see Figure 3-2) to maintain the correct IP rating of the Rosemount Model  
700 Gas Chromatograph.  
6. Follow the procedure, Install the breather drain, to replace the breather drain.  
Model 700 Breather Drain  
15  
   
SK-07778  
Rev. A  
2019  
AMERICAS  
Emerson Automation Solutions  
10241 West Little York, Suite 200  
Houston, TX 77040 USA  
Toll Free 866 422 3683  
+1 713 396 8880 (North America)  
+1 713 396 8759 (Latin America)  
+1 713 466 8175  
EUROPE  
Emerson  
MIDDLE EAST AND AFRICA  
Emerson  
ASIA-PACIFIC  
Emerson  
Neuhofstrasse 19a PO Box 1046  
CH-6340 Baar  
Switzerland  
Emerson FZE  
Jebel Ali Free Zone  
Dubai, United Arab Emirates, P.O. Box 17033  
1 Pandan Crescent  
Singapore 128461  
Republic of Singapore  
+41 (0) 41 768 6111  
+41 (0) 41 768 6300  
+971 4 811 8100  
+971 4 886 5465  
+65 6 777 8211  
+65 6 777 0947  
©
2019 Emerson. All rights reserved.  
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a  
mark of one of the Emerson family of companies. All other marks are the property of their  
respective owners.  

Tekkeon Healthy White Hx6711 User Manual
Sanyo Cr2016 User Manual
NEC MULTISYNC LCD1990SXI User Manual
MAYTAG MMV4205FW 02 User Manual
LG 55TNF5J User Manual
LG 50UL3J B User Manual
KENWOOD KVT 745DVD User Manual
EMERSON PD6810 User Manual
CRAFTSMAN 944.361361 User Manual
ACER DWX1907 User Manual