Teledyne Automobile Parts 203 User Manual

TELEDYNE  
HASTINGS  
INSTRUMENTS  
INSTRUCTION MANUAL  
201/203/205/207 SERIES  
FLOW METERS/CONTROLLERS  
I S O 9 0 0 1  
C E R T I F I E D  
Table of Contents  
1.  
2.  
GENERAL INFORMATION............................................................................................................................................ 4  
1.1.  
FEATURES .................................................................................................................................................................... 4  
SPECIFICATIONS HFM-201/205................................................................................................................................... 5  
SPECIFICATIONS HFC-203/207.................................................................................................................................... 5  
OPTIONAL 4 - 20 MA CURRENT OUTPUT...................................................................................................................... 6  
OTHER ACCESSORIES................................................................................................................................................... 6  
1.2.  
1.3.  
1.4.  
1.5.  
INSTALLATION AND OPERATION............................................................................................................................. 8  
2.1.  
RECEIVING INSPECTION ............................................................................................................................................... 8  
POWER REQUIREMENTS ............................................................................................................................................... 8  
OUTPUT SIGNAL........................................................................................................................................................... 8  
MECHANICAL CONNECTIONS....................................................................................................................................... 9  
ELECTRICAL CONNECTIONS......................................................................................................................................... 9  
OPERATION................................................................................................................................................................ 10  
RANGE CHANGING:.................................................................................................................................................... 14  
VALVE-OVERRIDE CONTROL..................................................................................................................................... 14  
2.2.  
2.3.  
2.4.  
2.5.  
2.6.  
2.7.  
2.8.  
3.  
4.  
THEORY OF OPERATION ........................................................................................................................................... 15  
3.1.  
OVERALL FUNCTIONAL DESCRIPTION: ...................................................................................................................... 15  
SENSOR:..................................................................................................................................................................... 15  
ELECTRONICS: ........................................................................................................................................................... 15  
SHUNT: ...................................................................................................................................................................... 16  
VALVE:...................................................................................................................................................................... 17  
3.2.  
3.3.  
3.4.  
3.5.  
MAINTENANCE.............................................................................................................................................................. 18  
4.1.  
AUTHORIZED MAINTENANCE..................................................................................................................................... 18  
TROUBLESHOOTING ................................................................................................................................................... 18  
ADJUSTMENTS ........................................................................................................................................................... 19  
END CAP REMOVAL:.................................................................................................................................................. 20  
PRINTED CIRCUIT BOARD REPLACEMENT.................................................................................................................. 20  
SENSOR REPLACEMENT: ............................................................................................................................................ 20  
ORIFICE CHANGES: .................................................................................................................................................... 20  
4.2.  
4.3.  
4.4.  
4.5.  
4.6.  
4.7.  
5.  
WARRANTY .................................................................................................................................................................... 23  
5.1.  
5.2.  
WARRANTY REPAIR POLICY ...................................................................................................................................... 23  
NON-WARRANTY REPAIR POLICY ............................................................................................................................. 23  
141-072011 - 201/203 Series  
Page 3 of 23  
1. General Information  
The Hastings HFM-201/HFM-205/HFC-203/HFC-207 series Mass flow meters (HFM-201/HFM-  
205) and controllers (HFC-203/HFC-207) are designed to accurately measure and control mass  
flow over the range of 30 slm to 2500 slm, without corrections or compensations for gas pressure  
and temperature with an accuracy of better than ±1% from the mean (±2% FS for 201/203 units in  
the ½” fitting size base with flows >500 slm). Hastings mass flow instruments do not require any  
periodic maintenance under normal operating conditions with clean gases. No damage will occur  
from the use of moderate overpressures (~500 psi/3.45 MPa) or overflows. Instruments are normally  
calibrated with the appropriate standard calibration gas (nitrogen) then a correction factor is used to  
adjust the output for the intended gas. Special calibrations for other gases, such as oxygen, helium  
and argon, are available upon special order.  
1.1. Features  
LINEAR BY DESIGN. The Hastings 200 series flow instrument is inherently linear (no linearization  
circuitry is employed). Should recalibration in the field be desired (a calibration standard is required), the  
customer needs to simply set the zero and span points. There will be no appreciable linearity change of the  
instrument when the flowing gas is changed.  
MODULAR SENSOR. The Hastings 200 series flow instrument incorporates a removable/replaceable  
sensor module. Field repairs to units can be achieved with a minimum of production line downtime.  
METER SETTLING TIME. Changes in flow rate for the HFM-201/HFM-205 are detected in less than 2  
seconds when using the speed up circuitry.  
LOW TEMPERATURE DRIFT. The temperature coefficient of span for the Hastings 200 series flow  
instrument is typically less than 0.05% of full scale/° C from 15° - 45° C. The temperature coefficient of  
zero is typically less than 0.1% of reading/° C from 0° - 50° C.  
FIELD RANGEABLE. The HFM-201/HFC-203 series is available in ranges from 30 slm to 1100 slm.  
The HFM-205/HFC-207 series is available in ranges from 1000 slm to 3000 slm. For HFC-203 & HFC-  
207 controllers, an orifice change is required as well. Calibration is required after all changes.  
CURRENT LOOP. The 4 - 20 mA option gives the user the advantages of a current loop output to  
minimize environmental noise pickup.  
141-072011 - 201/203 Series  
Page 4 of 23  
1.2. Specifications HFM-201/205  
Accuracy1 and Linearity.......................................................................................................±1% F.S.  
Repeatability ..................................................................................................................±0.05% F.S.  
Standard Pressure Rating......................................................................................................500 psig  
Pressure Coefficient ........................................................... +0.0067%/psi (0 - 1000 psig N2) typical  
High-Pressure Option ................................................................................Proof tested to 1500 psig  
Leak Integrity.................................................................................................................< 1x10-9 sccs  
Temperature Coefficient ........................................................... Zero: ±0.035% FS/° C (0° - 60° C)  
....................................................................................................Span: ±0.05% FS/° C (0° - 60° C)  
STP .....................................................................................................................0° C and 760 Torr  
Power (±15 Volt flow meter) ............................................. ±(14 - 16) VDC @ ±30 mA (< 1 Watt)  
Power (24 Volt flow meter) ...................................................................(14 - 32) VDC < 1.9 Watts  
Flow Signal .............................................(inherently linear) 0 - 5.00 VDC, 0 - 20 mA or 4 - 20 mA  
Wetted Material2 ............ 302 SS, 316 SS, Viton®, 82/18 Au/Ni Braze, Ni 200, Trace Silver Solder  
Connector .......................................................................15-pin subminiature D (9 pin for 24 Volt)  
Fittings½-in. Swagelok® - 201/203 small, 3/4-in. Swagelok® - 201/203 large, 1” Swagelok® - 205/207  
Weight (approx.) ...................201L: 3.55 lb (1610 g), 201H: 3.25 lb (1470g), 205: 8.1 lb (3650 g)  
1.3. Specifications HFC-203/207  
Accuracy1 and Linearity ......................................................................................................±1% F.S.  
Repeatability ..................................................................................................................±0.05% F.S.  
Std. Pressure Rating .............................................................................................................500 psig  
High Pressure Option ................................................................................Proof tested to 1500 psig  
Pressure Coefficient .......................................................... +0.0067%/psi (0 - 1000 psig N2) typical  
Control Valve DP* .............................................................................................Per customer order  
Leak Integrity ................................................................................................................< 1x10-9 sccs  
Temperature Coefficient ........................................................... Zero: ±0.035% FS/° C (0° - 60° C)  
.....................................................................................................Span: ±0.05% FS/° C (0° - 60° C)  
STP .....................................................................................................................0° C and 760 Torr  
Power (±15 Volt flow meter) ............................................. ±(14 - 16) VDC @ ±30 mA (< 1 Watt)  
Power (24 Volt flow meter) ...................................................................(14 - 32) VDC < 1.9 Watts  
Flow Signal ...............................................................(inherently linear) 0 - 5.00 VDC or 4 - 20 mA  
Command Signal ...................................................................................0 - 5.00 VDC or 4 - 20 mA  
Wetted Material 2 ........................................................... 316 SS, Nickel, Viton, 82/18 Au/Ni Braze,  
..............................................................................................................Trace Silver Solder, Kalrez®  
Connector .......................................................................15-pin subminiature D (9 pin for 24 Volt)  
Fittings½-in. Swagelok® - 201/203 small, 3/4-in. Swagelok® - 201/203 large, 1” Swagelok® - 205/207  
Weight (approx.) ...................203L: 5.2 lb (2360 g), 203K: 5.15 lb (2340g), 207: 14.9 lb (6760 g)  
1 Stated accuracy is for nitrogen or other gas specific calibration and using this gas only.  
2 Other materials are available. Viton is the standard O-ring option.  
* Consult factory for pressure ranges.  
141-072011 - 201/203 Series  
Page 5 of 23  
1.4.Optional 4 - 20 mA Current Output  
An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to  
flow. The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current  
loop output is useful for remote applications where pickup noise could substantially affect the  
stability of the voltage output or long cable runs where cable resistance would cause a voltage signal  
to decay.  
The current loop signal replaces the voltage output on pin 6 of the DA-15 connector (pin 2 of DE-9  
24 Volt). The current loop is sourced internally from the +15 VDC supply and must be returned to  
the signal supply common after (pin 5 [±15VDC] or pin 8 [24 VDC]) passing through the  
monitoring circuitry to complete the loop. The load must be between 0 and 600 Ohm.  
The 4 - 20 mA I/O option can accept a current input for flow controllers. The 0 - 5 VDC command  
signal on pin 14 (pin 3 for 24 Volts) can be replaced by a 4 - 20mA command signal. The loop  
presents an impedance of 50 Ohms and is returned to the signal common. On the 24 Volt units, the  
signal common will need to be tied to the power supply common external to the flow controller to  
complete the loop since the signal common is galvanically isolated internally from the supply  
common. This line must be current limited (<50 mA) to prevent short circuit currents from  
damaging the receiving circuit in the flow controller.  
24 Volt Supply  
±15 Volt Supply  
1.5. Other Accessories  
1.5.1. Hastings Power Supplies  
Hastings power supply/display units are available in one and four channel versions. They convert  
100, 115 or 230VAC to the ±15 VDC required to operate the flow meter and provide a digital  
indication of the flow rate. Interface terminals for the retransmission of the flow meter analog output  
signal are located on the rear of the panel.  
The THCD-100 and the THPS-100/400 models are built with controllers in mind but will work  
with meters as well. The Model 40 is for flow meters only. Throughout this manual, when  
reference is made to a power supply, it is assumed the customer is using a Hastings power supply.  
Hastings THCD and THPS power supplies are CE marked, but the Model 40 does not meet CE  
standards at this time. The Model 40, The THCD and THPS-100 are not compatible with 4 – 20  
mA analog signals. With the THPS-400, individual channels’ input signals, as well as their  
commands, become 4 – 20 mA compatible when selected. The THPS-400 also sports a Totalizer  
141-072011 - 201/203 Series  
Page 6 of 23  
feature. More information about Hastings Instruments’ power supplies can be found on the  
1.5.2.  
Interconnecting Cables  
Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on  
the back of Hastings power supplies directly to any of the 200 series and 300 series flow instruments  
(including digital versions). More information about the available cables can be found in the THPS  
400 bulletin on the Hastings web site. http://www.teledyne-hi.com/pdfs/bulletins.htm  
141-072011 - 201/203 Series  
Page 7 of 23  
2. Installation and Operation  
This section contains the necessary steps to get a new flow meter/controller into operation as quickly  
and easily as possible. Please read the following thoroughly before attempting to install the  
instrument.  
2.1. Receiving Inspection  
Carefully unpack the Hastings 201/203 series instrument and any accessories that have also been  
ordered. Inspect for any obvious signs of damage to the shipment. Immediately advise the carrier  
who delivered the shipment if any damage is suspected. Check each component shipped with the  
packing list. Insure that all parts are present (i.e., flow meter, power supply, cables, etc.). Optional  
equipment or accessories will be listed separately on the packing list. There may also be one or more  
OPT-options on the packing list. These normally refer to special ranges or special gas calibrations.  
They may also refer to special helium leak tests, or high pressure tests. In most cases, these are not  
separate parts, but special options or modifications built into the flow meter.  
2.2. Power Requirements  
The Hastings 200 series flow instruments (bipolar 15 Volt versions) require ±15 VDC @ ±30 mA/1  
Watt (HFM-201/205) +60 mA, -185 mA / 3 Watts (HFC-203/207) for proper operation. The  
supply voltage should be sufficiently regulated to no more than 50 mV ripple. The supply voltage  
can vary from 14.0 to 16.0 VDC. Surge suppressors are recommended to prevent power spikes  
reaching the instrument. The Hastings power supplies described in Section 1.4.1 satisfy these power  
requirements.  
The Hastings 200 series flow instruments (24 Volt versions) require 14 - 32 VDC @ 1.9 Watts  
(HFM-201/205) and 4.2 Watts (HFC-20203/207) for proper operation. The supply voltage should  
be reasonably regulated as power supply ripple may propagate to the output. The supply common  
is galvanically isolated from the signal common such that this instrument can be powered from a  
bipolar 15 Volt supply using only the positive and negative output terminal (30 Volts). Surge  
suppressors are recommended to prevent power spikes reaching the instrument.  
Attempting to operate the 24 volt versionfor any significant length of time  
Warning:  
at a voltage less than the specified minimimum voltage will cause  
the internal DC-DC convertors to fail.  
The Hastings 200 series flow instruments have an integral 5 VDC reference source. This stable  
voltage is on pin 15 of the DA-15 connector (pin 1 of the DE-9 for 24 Volt) and may be used for the  
command voltage.  
2.3. Output Signal  
The standard output of the flow meter is a 0 - 5 VDC signal proportional to the flow rate. In the  
Hastings power supplies, the output is routed to the display and is also available at the terminals on  
the rear panel. If a Hastings supply is not used, the output is available on pin 6 of the DA-15  
connector (pin 2 of the DE-9 for 24 Volt) and is referenced to pin 5 (pin 8 for 24 Volt). It is  
recommended that the load resistance be no less that 2kΩ. If the optional 4 - 20 mA output is used,  
the load impedance must be selected in accordance with Section 1.3.  
141-072011 - 201/203 Series  
Page 8 of 23  
2.4. Mechanical Connections  
The flow meter may be mounted in any position as long as the direction of gas flow through the  
instrument follows the arrow marked on the bottom of the flow meter case label. The preferred  
orientation is with the inlet and outlet fittings in a horizontal plane (if operating with a dense gas or at  
high pressures the instrument must be installed horizontally). When mounted in a different  
orientation the instrument should be re-zeroed at zero flow with the system pressurized to the  
expected operating pressure.  
The smallest of the internal passageways in the Hastings 200 series flow instrument is the diameter  
of the sensor tube, which is 0.0125”(0.31 mm), so the instrument requires adequate filtering of the  
gas supply to prevent blockage or clogging of the tube.  
The pressure regulator and the plumbing upstream must be of sufficient size to minimize changes in  
the upstream pressure. When switching from full flow to zero flow, the inlet pressure of the  
instrument should rise to no more that 30% above the inlet pressure at full flow. In general, high  
capacity regulators and large internal diameter plumbing help to make the system more stable. The  
pressure drop between the regulator and the instrument due to line resistance should be minimized.  
The differential pressure across the unit should be less than 6” of H2O at maximum flow.  
There are two 8-32 threaded holes, 0.25” deep, (1/4-28 threaded holes, 0.375” deep for 205/207)  
located on the bottom of the base that can be used to secure it to a mounting bracket, if desired  
(screws provided). Other holes for special mounting can be added to the end cap as desired.  
The standard inlet and outlet fittings for the 201/203 are 0.5” and 0.75” Swagelok (optional VCR,  
VCO, NPT and metric fittings are available). The standard inlet and outlet fittings for the 205/207  
are 1” Swagelok (optional VCR, VCO, and metric fittings are available).The O-rings for the end cap  
and the sensor are Viton (optional Kalrez , Buna-N (Nitrile) or Neoprene is available). It is  
suggested that all connections be checked for leaks after installation. This can be done by  
pressurizing the instrument (do not exceed 500 psig unless the flow meter is specifically rated for  
higher pressures) and applying a diluted soap solution to  
the tubing connections.  
2.5. Electrical Connections  
If a power supply from Hastings Instruments is used with  
a 15 Volt version of the HFM-200/HFC-202, installation  
consists of connecting the 200 series cable (#AF-8AM)  
from the “D” connector on the rear of the power supply  
to the “D” connector on the top of the flow meter. If a  
different power supply is used, follow the instructions  
below when connecting the flow meter.  
Bipolar (±15 Volt) Power Supply to Bipolar Connections  
Connecting the Hastings 200 series flow instrument with  
anything other than the prescribed cables and power  
supplies discussed above, can severely damage the  
instrument and void the warranty. The figure at the right  
of this page shows the schematic layout for connecting the  
instrument to an appropriate ±15 Volt power supply.  
The power supply used must be meet the power  
requirements delineated in section 2.2 (Power  
Requirements) above. If a bipolar supply is required, the  
voltages must be referenced to a common ground.  
Connect -15VDC to pin 9 of the DA-15 connector and  
+15VDC to pin 11. Pins 5 and 12 are both commons and they must be connected together and to  
141-072011 - 201/203 Series  
Page 9 of 23  
the ground connection at the power supply. Do not connect them together at the flow controller as  
the resulting crosstalk could result in flow instabilities.  
24 Volt Connections  
Refer to the diagram at right when connecting 24 Volt  
units.  
Connect the positive lead of the power supply to pin 7 of  
the DE-9 connector and negative lead to pin 4. The  
supply input is diode protected such that reversing the  
input polarity will not damage the instrument. The  
power supply is galvanically isolated from all other pins.  
General Connection Notes  
Pin 7 of the DA-15 (15 Volt), Pin 9 of the DE-9 (24 Volt)  
is the case ground. It should be connected to the cable  
shield if available and to the AC ground to the power  
supply.  
Pin 6 of the DA-15 (15 Volt), Pin 2 of the DE-9 (24 Volt)  
is the output signal from the flow controller. This output  
will be 0-5VDC/(4-20 mA), 5VDC/20mA being 100% of  
rated or full flow. Pin 14 of the DA-15 (15 Volt), Pin 3 of  
the DE-9 (24 Volt) is the command input. This should be  
a 0-5VDC or (4-20 mA) signal and must be free of spikes  
or other electrical noise, as these will generate false flow  
commands that the controller would attempt to flow. Pin  
15 of the DA-15 (15 Volt), Pin 1 of the DE-9 (24 Volt) is  
a regulated +5.00VDC output reference that is normally adjusted to 5.01VDC so that the user can  
achieve full scale command set point. The reference is designed to provide the command signal for  
pin 14 by connecting one end of a potentiometer to the voltage reference and the other end to  
ground. The center lead would then be connected to setpoint input.  
If a valve override switch is not desired, the unit is ready for use at this time. If the override switch is  
desired, connect the center pin of a single pole, three-position switch with the center off position to  
pin 8 of the DA-15 (15 Volt), Pin 6 of the DE-9 (24 Volt). Connect a voltage source >10 VDC to  
one end of the switch, and negative voltage to the other end. This will result in the valve being full  
open when the positive is supplied to the override pin, off when the negative voltage is supplied and  
auto-control when there is no connection to the valve override pin (OPEN-AUTO-CLOSE). This  
setup will be adequate for most purposes, but there will be a small delay for capacitors to charge  
between switch operation and control override.  
2.6. Operation  
The standard instrument output is a 0 - 5 VDC out and the signal is proportional to the flow i.e.,  
Volts = zero flow and 5 Volts = 100% of rated flow. The 4 - 20 mA option is also proportional to  
flow, 4 mA = zero flow and 20 mA = 100% of rated flow. It is suggested that all connections be  
checked for leaks after installation. This can be done by pressurizing the instrument (do not exceed  
0
500 psig unless the instrument is specifically rated for higher pressures) and applying a diluted soap  
solution to the connections.  
2.6.1. Operating Conditions  
For proper operation, the combination of ambient temperature and gas temperature must be such  
that the flow meter temperature remains between 0 and 50°C. (Most accurate measurement of flow  
will be obtained if the flow meter is zeroed at operating temperature as temperature shifts result in  
141-072011 - 201/203 Series  
Page 10 of 23  
some zero offset.) The 201/203 series is intended for use in non-condensing environments only.  
Condensate or any other liquids which enter the flow meter may destroy its electronic components.  
2.6.2. Zero Check  
Turn the power supply on if not already energized. Allow for a 1 hour warm-up. Stop all flow  
through the instrument and wait 2 minutes. Caution: Do not assume that all metering valves  
completely shut off the flow. Even a slight leakage will cause an indication on the meter and an  
apparent zero shift. For the standard 0 - 5 VDC output, adjust the zero potentiometer located on  
the lower outlet side of the flow meter until the meter indicates zero. For the optional 4 - 20 mA  
output, adjust the zero potentiometer so that the meter indicates slightly more than 4 mA, i.e. 4.03 to  
4.05 mA. This slight positive adjustment ensures that the 4 - 20 mA current loop transmitter is not  
in the cut-off region. The error induced by this adjustment is approximately 0.3% of full scale. This  
zero should be checked periodically during normal operation. Zero adjustment is required if there is  
a change in ambient temperature, or vertical orientation of the flow meter/controller.  
2.6.3. High Pressure/High Density Operation  
When operating at high pressure or with high density gases, the increased density of gas will cause  
natural convection to flow through the sensor tube if the instrument is not mounted in a level  
position. This natural convection flow will be proportional to the system pressure. This will be seen  
as a shift in the zero flow output that is directly proportional to the system pressure. This zero shift  
can usually be corrected by adjusting the zero potentiometer after mounting the flow controller in its  
final operating position. See section 2.4 and 2.6.2.  
Additionally, because the specific heat of a gas changes with pressure, an associated error is  
introduced with increased pressure. It is directly proportional to the change in specific heat versus  
pressure for that particular gas. For helium, there is virtually no change in indicated output; for  
nitrogen, the indicated output will increase at approximately 0.0067%/psi, which is the pressure  
coefficient. Thus,  
Actual flow = indicated flow / (1 + pressure * pressure coefficient)  
2.6.4. Blending of Gases  
In the blending of two gases, it is possible to maintain a fixed ratio of one gas to another. In this  
case, the output of one flow controller is used as the reference voltage for the set point potentiometer  
of a second flow controller. The set point potentiometer then provides a control signal that is  
proportional to the output signal of the first flow controller, and hence controls the flow rate of the  
second gas as a percentage of the flow rate of the first gas.  
EXAMPLE: Flow controller A has 0 - 100 slm range with a 5.00 Volt output at full scale. Flow  
controller B has 0 - 10 slm range with a 5.00 Volt output at full scale. If flow controller A is set at 80  
slm, its output voltage would be 4.00 Volts (80 slm/100 slm x 5.00 Volts = 4.00 Volts). If the output  
signal from flow controller A is connected to the command potentiometer of flow controller B, it  
then becomes a variable reference voltage for flow controller B proportional to the flow rate of flow  
controller A.  
If the set point potentiometer of flow controller B is set at 50% of full scale, and the reference voltage  
from flow controller A is 4.00, then the command signal going to flow controller B would be 2.00  
Volts (4.00 Volts x 50.0% = 2.00 Volts). The flow of gas through flow controller B is then  
controlled at 4 slm (2.00 Volts/5.00 Volts x 10 slm = 4 slm).  
The ratio of the two gases is 20:1 (80 slm/4slm). The % mixture of gas A is 95.2 (80slm/84slm and  
the % mixture of gas B is 4.8% (4 slm/84 slm).  
Should the flow of flow controller A drop to 78 slm, flow controller B would drop to 3.9 slm, hence  
maintaining the same ratio of the mixture. (78 slm/100 slm x 5 Volts = 3.90 Volts x 50% = 1.95  
Volts; 1.95 Volts /5.00 Volts x 10 slm = 3.9 slm;  
78 slm: 3.9 slm = 20:1)  
141-072011 - 201/203 Series  
Page 11 of 23  
2.6.5. Operation with a Hastings power supply.  
There are two controls for each flow controller connected to a Hastings power supply. A switch  
labeled “OPEN; AUTO; CLOSED” (valve override THPS 400 only) and a potentiometer labeled  
“COMMAND”. For normal operation, the valve override switch will be in the “AUTO” position.  
The “CLOSE” position removes all power from the valve, shutting off flow regardless of the  
command pot setting. The “OPEN” position applies full available valve voltage to the valve, causing  
it to open, regardless of the command pot setting. The “OPEN” position is useful for purging  
systems. It is recommended that the valve override switch not be left in this position for extended  
periods of time, with no flow through the controller, as a small positive zero shift may be observed.  
The “COMMAND” pot adjusts the Analog command signal sent to the flow controller. The setting  
for each controller connected to the power supply can be observed. (Depending on how the power  
supply was set up, the display could indicate in flow units or percent of full scale).  
Hastings Power Supply Feature Guide  
Feature  
Model 40 Power Pod 100 Power Pod 400  
9
9
9
9
9
9
9
9
9
9
9
Digital Readout  
±15 Volts  
Analog Outputs  
Controller  
9
9
Analog Control  
Front Panel Override  
Totalizer  
9
9
9
9
9
9
4 -20 mA  
9
Ratio Control  
Alarms  
9
9
9
9
Multi-Channel Display  
Conversion Factors  
Communications  
9
RS232  
RS232/RS485  
2.6.6. Operation with a power supply other than a Hastings.  
The flow controller must be connected to the power source as specified in section 2.6. In general, a  
0 - 5 VDC command signal proportional to the intended flow (0 Volts = zero flow; 5 Volts = 100%  
of rated flow) must be applied to pin 14 of the DA-15 (±15 Volt), Pin 3 of the DE-9 (24 Volt)  
connector. A 0 - 5 VDC signal proportional to the flow rate through the instrument will be present  
on pin 6 of the DA-15 (±15 Volt), Pin 2 of the DE-9 (24 Volt) connector. The control mode is  
selected via pin 8 of the DA-15 (±15 Volt), Pin 6 of the DE-9 (24 Volt) connector. Apply >10  
Volts for full open, < 0 Volts for closed and allow the override to float for flow proportional to the  
command voltage. Refer to your power supply manual for the specifics of implementing these  
parameters.  
141-072011 - 201/203 Series  
Page 12 of 23  
2.6.7. Operation with an external sensor. (Fig. 2.2)  
In some instances, it might be desirable to use an external sensor to provide process information to  
the control circuitry in the flow controller. For example, you might want to control the pressure in a  
vacuum system by adjusting the rate at which the system is backfilled with a gas. The new,  
enhanced HFC series of flow controllers have provision for accepting a 0 - 5VDC (or 0 - 10 VDC)  
output from an external sensor at pin 13 of the DA-15 (±15 Volt), Pin 1 of the DE-9 (24 Volt)  
connector. To activate this feature, the cover of the HFC must be removed to gain access to the  
electronics card and move a jumper.  
For the 15 Volts units JP1 is a three pin jumper block located just below the “D” connector. In the  
normal operating mode, the jumper covers the bottom two pins. To select “External Sensor”, move  
the jumper to the upper two pins. This swaps the flow input to the controller circuit from the flow  
meter output to pin 13 of the “DA-15” connector.  
Fig 2.2  
For the 24 Volt units JP6 is a six pin jumper block located  
just to the right of the sensor connector. Pin 1 is in the  
lower right corner. There are three jumpers installed  
horizontally. The bottom jumper ties the flow output to  
the valve control circuit and the middle jumper ties Pin 1  
of the DE-9 connector to reference voltage circuit.  
Remove both of the lower two jumpers and reinstall one  
jumper vertically between the outside two pins (pins 1 &  
3). This will tie Pin 1 of the DE-9 connector to the valve  
control circuit and allow the controller to maintain the  
external process variable equal to the desired setpoint.  
The 5 Volt reference is not available in this configuration.  
2.6.8. Response to Command Changes  
Fig 2.3  
The response of the control circuit to changes to the  
command signal is set at the factory for fast, stable response.  
If excessive overshoot is present, the response can be slowed  
removing the jumper labeled “JP4,” located in the center of  
PC-828 for the 15 Volt units or by removing the top jumper  
of JP6 for the 24 Volt boards (located to the right of the  
sensor connector.  
The fastest response to command changes is obtained when  
JP4/(JP6) is covered by the jumper. This setup allows large  
overshoot and undershoot swings in the actual flow rate while  
the control circuit is establishing control at the new command  
point. A slower response to command changes is obtained  
when JP4/JP6 is not covered by a jumper. This setup results  
JP6  
in no overshoot or undershoot in the actual flow rate as the  
controller circuit establishes control at the new command  
point. This jumper does not affect the system gain and  
will not dampen out oscillations.  
To adjust the response, you need a means of producing a  
step change in the command voltage from 10% of full  
scale to 100% of full scale. Follow the steps outlined  
below:  
External  
Variable  
Normal  
1) Cover the pins of JP4/JP6 with a jumper. (see fig. 2.2 & fig 2.3)  
2) Set the command voltage to 10% of full scale. Allow the flow to stabilize.  
141-072011 - 201/203 Series  
Page 13 of 23  
3) Step change the command voltage to 100%, and observe the flow through the controller. If the overshoot  
is too large, remove the jumper. Reset the command voltage to 10%, and allow the controller to stabilize.  
4) To prevent loss of the unused jumper, place it over one pin only on JP4/JP6.  
2.6.9. Operating Temperature  
For proper operation, the combination of ambient temperature and gas temperature must be such  
that the Flowcontroller temperature remains between 0 and 50°C. Most accurate measurement of  
flow will be obtained if the flow controller is re-zeroed at operating temperature, as temperature  
shifts result in some zero offset.  
2.7. Range Changing:  
The range of the flow controller can be changed in the field if recalibration facilities are available.  
The flow controller may require a different orifice, which can be purchased separately from the  
factory. A listing of the orifices available and their flow rates can be found in Section 5.0. The  
instructions to change the flow range can be found in Section 4.6.  
2.8. Valve-Override Control  
The valve override control line provides a method to override the loop controller and open or close  
the valve regardless of the flow or command signals. During normal operation this line must be  
allowed to float freely. This will allow the loop control to open and close the valve as it requires. If  
the valve override line is forced high (> +10 Volts) the valve will be forced full open. If the valve-  
override line is forced negative (< 0 Volts) the valve will be forced closed.  
141-072011 - 201/203 Series  
Page 14 of 23  
3. Theory of Operation  
This section contains an overall functional description of HFC Flow Controllers. Detailed  
schematics and parts lists can be found at the end of the manual in Section 6.0. In this section and  
other sections throughout this manual, when a power supply is mentioned, it is assumed that the  
customer has a Hastings Power Supply. These sections are not applicable if another type of power  
supply is used.  
3.1. Overall Functional Description:  
The HFC Flow Controller consists of a sensor,  
electronic circuitry, a shunt and a valve. The  
sensor measures the flow rate from 0 to 10 sccm of  
the gas to be metered. The shunt divides the flow  
such that the flow through the sensor is a precise  
percentage of the flow through the shunt. The flow  
through the sensor and the shunt is always laminar.  
The circuit board amplifies the sensor output and  
uses this output to control the valve position. The  
2 stage valve employs an automatic metering  
solenoid, used to control the pressure differential  
across the main diaphragm seat assembly. All of  
these components working together result in a fast,  
stable flow controller.  
FIG 3.1  
3.2. Sensor:  
The Hastings HFM-201/HFC-203 series instruments operate on a unique thermal electric principle  
whereby a metallic capillary tube is heated uniformly by a resistance winding attached to the  
midpoint of the capillary (see Figure 3.1). Thermocouples TC-1 and TC-2 are welded at equal  
distances from the midpoint and develop equal outputs at zero flow.  
When flow occurs through the tubing, heat is transferred from the tube to the gas on the inlet side,  
and from the gas back to the tube on the outlet side creating an asymmetrical temperature  
distribution (sees Figure 3.2). The thermocouples sense this decrease and increase in the capillary  
tube temperature and produce a milliVolt output signal proportional to that change.  
For a constant power input, the differential thermocouple output is a function of the mass flow rate  
and the heat capacity of the gas. Since the heat capacity of many gases is relatively constant over  
wide ranges of temperature and pressure, the flow meter may be calibrated directly in mass units for  
those gases. Changes in gas composition usually only require application of a simple multiplier to  
the air calibration to account for the difference in heat capacity and thus the flow meter is capable of  
measuring a wide variety of gases. The HFM sensor measures approximately 10 sccm, full scale  
flow.  
3.3. Electronics:  
The HFM-201/HFC-203 series instruments use a thermal flow sensor to measure through a  
capillary tube, which is a fixed percentage of the total flow through the instrument. This sensor  
develops an output signal proportional to flow which is approximately 1 mv full scale magnitude.  
This signal is amplified by the meter circuitry until is 0-5.00 VDC. This 5 Volt output is sent back  
to the power supply and to the flow meter circuitry, if applicable. At the power supply the 5 Volt  
141-072011 - 201/203 Series  
Page 15 of 23  
output is sent to the terminals on the back  
and to the decoding circuitry in the display  
which converts it to a 3-digit output.  
The controller circuitry utilizes the  
Command and the Flow voltages as input  
signals. The 0 - 5VDC command signal is  
subtracted from the 0 - 5VDC flow signal  
creating an error signal. This signal is  
amplified and causes the solenoid valve to  
move. The amount and direction of the  
movement is dependent upon the value and  
the sign of the error signal, and tends to  
minimize the error signal.  
FIG 3.2  
3.4. Shunt:  
Measurement of flow rates higher than the  
10 sccm full scale is achieved by dividing the flow with a fixed ratio shunting arrangement, as is  
illustrated in Figure 3.3. This is accomplished by placing the measuring capillary tube parallel with  
one or more dimensionally similar channels, called a laminar flow element (LFE). Therefore, the  
sensor only needs to heat the gas passing through the capillary tube resulting in low power  
requirements, while retaining all the mass measuring characteristics.  
The HFC-203 uses corrugated and fused shunts similar to the shunts used in the lower flow range  
instruments. These high range shunts are factory adjustable from 0 - 30 slm to 0 - 500 slm (see  
Figure 3.4), using specific corrugated shunts that give the desired flow range.  
FIG 3.3  
141-072011 - 201/203 Series  
Page 16 of 23  
3.5. Valve:  
A solenoid valve is used as a pilot valve to control a much larger pneumatic diaphragm valve. The  
pilot valve is an “automatic metering solenoid” valve. While most solenoids operate in either the  
fully open or fully closed state, the automatic metering solenoid valve is designed to control flow (see  
Figure 3.5). A spring, connected to the plunger assembly, holds a magnetic plunger tightly against  
an orifice to shut off flow. The magnetic plunger is surrounded by an electrical coil, which when  
energized with electrical current lifts the plunger off the orifice and allows flow to pass between the  
orifice and the plunger seat. Controlling the current through the coil controls the distance between  
the orifice and the plunger seat, thus effectively controlling the flow through the valve.  
The main valve used in the HFC-203/207 is  
a pilot operated pneumatic valve. The main  
pneumatic valve uses a diaphragm to control  
Solenoid  
(F  
the height of the main seat above the orifice.  
The diaphragm position is controlled by  
balancing competing forces from the main  
spring and the pressure difference between  
the valve upstream pressure and the pressure  
in the bonnet chamber.  
Pilot  
Main  
Spring  
Bleed  
Screw  
The bonnet pressure is controlled by the  
solenoid operated pilot valve. When  
Diaphragm  
pressure is first applied and the pilot valve is  
closed, system pressure Pu is much higher  
Main  
Orifice  
than the bonnet pressure Pb. Pu will push up  
the diaphragm against Pb and the main flow  
will initiate. Flow through the restriction of the bleed screw Qb will also initiate. Since Qb is flowing  
into the bonnet chamber and the pilot flow Qp is shut off this will pressurize the bonnet chamber and  
Pb will rise. When the force due to the pressure difference across the diaphragm is balanced by the  
closing force from the main spring, the main valve will close.  
In order to open the valve, power is  
applied to the solenoid. The generated  
magnetic field attracts the plunger toward  
the pole against spring force. Flow  
initiates through the pilot valve Qp. When  
the pilot flow exceeds the bleed screw flow  
the pressure in the bonnet chamber will  
start to drop.  
Bleed  
Flow Qb  
Pb  
When the force due to the pressure  
Pilot  
Flow Qp  
difference across the diaphragm exceeds  
Main Flow Qm Pu  
the closing spring force then the  
diaphragm will start to lift. As the  
diaphragm moves up it will open the main  
valve and allow flow through the main  
orifice.  
Pd  
The control loop maintains sufficient power to the solenoid to keep the pilot valve flow equal to the  
bleed screw flow when the bonnet pressure is at the right level to hold the diaphragm high enough  
that the desired flow passes through the main valve.  
141-072011 - 201/203 Series  
Page 17 of 23  
4. Maintenance  
This section contains service and calibration information. Some portions of the instrument are  
delicate. Use extreme care when servicing the flow controller. The potentiometer positions and the  
electrical components referred to in the troubleshooting section can be found in Section 6.0 on the  
electrical component layout drawing.  
4.1. Authorized Maintenance  
With proper care in installation and use, the flow controller will require little or no maintenance. If  
maintenance does become necessary, most of the instrument can be cleaned or repaired in the field.  
Some procedures may require recalibration. Do not attempt these procedures unless facilities are  
available. Entry into the sensor or tampering with the printed circuit board will void warranty. Do  
not perform repairs on these assemblies while unit is still under warranty.  
4.2. Troubleshooting  
SYMPTOM: Override switch is in CLOSE position, but flow remains or 0.00 VDC is commanded  
and flow remains.  
CAUSE: Orifice out of adjustment or faulty op-amp  
ACTION: Check the valve voltage between pins 2 & 12 on the 15-pin D-connector for 15 Volt  
units and pins 5 & 8 on the DE-9 connector for 24 Volts. If the voltage is less than 3.00 VDC, then  
turn the orifice clockwise until flow stops.  
SYMPTOM: Output of unit is proportional to flow but extremely small and not correctable by span  
pot.  
CAUSE: Sensor is not being heated.  
ACTION: Unplug connector J2. Check the following resistance: The resistance between pins 2 &  
3 of the sensor should be approximately 2500 ohms (see Figure 3.1 on page 8). The resistance  
between pins 1 & 4 should be approximately 2.3 ohms. The resistance between pins 2 & 3 and the  
base of the sensor should be essentially infinite. If not, replace the sensor unit. If sensor reads O.K.,  
check the voltage output on pins 2 & 3 of the jack in the board. If it does not read approximately 22  
VDC then replace op-amp U2 (U3 on 24 Volt board).  
SYMPTOM: Sensor has proper resistance readings, but little or no output with flow.  
CAUSE: Plugged sensor.  
ACTION: Shut off gas supply and power supply. Remove cover and PC board from unit. Remove  
sensor assembly and examine. If sensor has evidence of plugging, clean or replace as applicable  
SYMPTOM: Flow controller oscillates.  
CAUSE: Flow controller not adjusted for the dynamics of the flow system.  
ACTION: Check upstream and downstream pressures. The gas supply regulator should not have  
excessive lockup when flow shuts off. Also ensure that there is not a large drop in pressure between  
the regulator and the instrument due to line resistance. Oscillations can also be caused if a large flow  
141-072011 - 201/203 Series  
Page 18 of 23  
restriction is pneumatically close to the downstream end of the flow controller. The differential  
pressure across the unit must be between 10-50 psig.  
SYMPTOM: Little or no flow, even with Valve Override switch OPEN.  
CAUSE: Plugged orifice.  
ACTION: Verify the presence of a 10-50 psig pressure across the instrument. If present, shut off  
gas supply and power supply. Remove orifice per Section 4.9. Examine orifice. If plugged, clean or  
replace as applicable. Reassemble valve.  
SYMPTOM: Flow meter reads other than 0.00 VDC with no flow, or there is a small flow when  
flow meter reads 0.00 VDC.  
CAUSE: ZERO potentiometer is out of adjustment.  
ACTION: Shut off all flow. Adjust ZERO potentiometer until output reads 0.00 VDC.  
SYMPTOM: Flow meter out of calibration and nonlinear.  
CAUSE: Leaks in gas inlet or outlet fittings.  
ACTION: Check all fittings for leaks by placing soap solution on all fittings between gas supply and  
final destination of gas. Check flow meter for leaks. Replace “O” rings if required or recalibrate as  
necessary.  
4.3. Adjustments  
4.3.1. Calibration Procedure: (Figure 4.1)  
NOTE: Adjusting the SPAN pot will require the use of a  
calibration reference in Step 5.  
1) Connect power cable to D connector as specified in Section 2.7.  
Allow instrument to warm up for 30 minutes with 10% flow and  
instrument in AUTO position.  
2) Set ZERO (R13 on 15 Volt, R20 for 24 Volt) potentiometer for  
0.000 VDC output.  
3) Turn on gas supply to inlet of instrument. Put Valve Override  
switch into CLOSE position. Adjust the orifice underneath  
controller to obtain zero flow. Put Valve Override switch into  
AUTO. Ensure that full range flow can still be obtained at  
minimum inlet pressure.  
4) Set command to 100%. Adjust SPAN (R18 on 15 Volt, R2 for 24 Volt) pot until the flow reference  
reads full scale flow (5.000 VDC). NOTE: Perform this step only if a calibrated reference flow  
meter is available.  
5) Record flow meter and flow reference outputs for flow rates of 20%, 40%, 60%, 80% and 100%.  
4.3.2. Miscellaneous adjustments  
Periodically, during normal operation, the ZERO should be checked and adjusted when required. If  
system parameters change, the RESPONSE pot may require a small adjustment for optimum  
141-072011 - 201/203 Series  
Page 19 of 23  
stability. If the instrument is not shutting completely off when Valve Override switch is in the  
CLOSE position, the orifice may require approximately 1/8 turn clockwise.  
4.4. End Cap Removal:  
The end cap on the inlet side must be removed to gain access to the filter or shunt assembly. First  
shut off the supply of gas to the instrument. Disconnect the Swagelok fitting on the inlet and outlet  
sides of the transducer, and remove it from the system plumbing. Remove the four hex bolts holding  
the end cap to the instrument (see Figure 4.1). Carefully remove the end cap, filter, wave spring (if  
present) and shunt, noting their order and proper orientation. The shunt can be severely damaged if  
dropped. Examine the filter and shunt. If either is dirty or blocked, clean or replace as applicable.  
Reassembly is the reverse of the removal procedure. Recalibration of the HFC is necessary.  
4.5. Printed Circuit Board Replacement  
In the unlikely event that the PC board fails, it is easily removed from the instrument and replaced  
with a spare to minimize instrument downtime. Replacement of the PC board will require the  
instrument to be recalibrated per Section 4.4.1.  
Unplug the power cable from the top of the transducer. Remove the two jackscrews next to the “D”  
connector and the two screws on the sides of the cover. Lift off the cover and unplug the four-wire  
sensor plug and the two wire valve plug, noting their orientation prior to removal.  
Remove the screw that holds the PC board to the sensor. Troubleshoot or replace as applicable.  
Installation is the reverse of the above procedure. Recalibrate if any components were changed or if  
any potentiometers were adjusted.  
4.6. Sensor Replacement:  
If the sensor fails or becomes plugged it can be removed. Remove the cover and the PC board per  
Section 4.7 above. Remove the three bolts holding the sensor to the instrument base. Remove the  
sensor from the base noting the two O-rings (Parker 2-005, V884-75) between the sensor and the  
base. If the sensor is plugged it can be cleaned by running a fine wire (approximately 0.008"  
diameter) through the tube. If sensor needs replacement, obtain another from the factory and install  
it. Ensure that O-rings are clean and intact. Install O-rings on seating surface, then carefully place  
sensor over O-rings and tighten down the three screws evenly. Replacement of sensor will require  
recalibration per Section 4.3.1.  
4.7. Orifice Changes:  
The orifice may require replacement if a large flow range change is desired, if changing to a gas that  
has a specific gravity significantly different than the original gas, if a large change in the differential  
pressures across the valve is desired or in the event that a small orifice becomes plugged.  
Replacement orifices can be acquired from the factory. The diameter of the orifice can be calculated  
using the following procedure:  
Orifice Changes:  
A) Determine the minimum expected upstream pressure (Pu) in absolute pressure units (add  
atmospheric pressure – 14.7 psia) and the maximum expected downstream pressure (Pd) in absolute  
pressure units for full flow conditions.  
B) If Pu > 2Pd, use formula 1; otherwise use formula 2.  
C) Use a consistent set of units for pressure, flow, and density (i.e all lengths, masses, times in the  
same units, cm, ft, kg, sec etc.) 1 liter/minute = 1.667 x 10-5 m3/sec, 1 gm/liter = 1 kg/m3, 1 psia =  
6895 kg/m*sec2, 1 Pa = 1 kg/m*sec2,  
141-072011 - 201/203 Series  
Page 20 of 23  
D) This formula provides approximate results that tend to be undersized because it neglects  
pressure drops internal to the flow controller, compressible gas effects and temperature effects.  
Multiply the result by 1.5 to get the expected minimum orifice size that can reliably pass the  
desired flows at the expected pressures.  
Where:  
Formula 1:  
Formula 2:  
σ Q 2  
P P  
8ρ0 P  
Q2 σ 16ρ0 P  
0
0
4
D =  
4
D =  
P 2  
π 2  
P
(
)
π 2  
γ
d
u
d
u
D = Orifice Diameter  
Q = Flow rate in standard volumetric unit (slm, sccm, scfh)  
P0 = Standard Pressure (760 Torr, 101.325 kpa)  
Pu = Upstream pressure in absolute units (use minimum expected value)  
Pd = Downstream pressure in absolute units (use maximum expected value)  
γ
ρ0  
π
σ
= Ratio of specific heats, 1.2 for monatomic gases, 1.4 otherwise  
= Density of gas @ standard pressure and temperature of flow unit  
= Pi (3.1415…)  
= Specific gravity of gas (ratio of gas density to density of air)  
Choose the available orifice with the closest diameter that is larger than the calculated diameter.  
Orifice diameters (inches) available are 0.023, 0.032, 0.046, 0.086, 0.130, 0.156, 0.250, 0.312,  
0.375, 0.500 (large base only). Contact factory to order new orifice.  
As an example, if the maximum controlled flow will be 1000 slm of air with an upstream pressure of  
50 psig and exhausting to atmospheric pressure the minimum orifice diameter calculated from the  
previous equation would be 0.225 inches. The next larger orifice that has a diameter of 0.250 inches  
should be installed.  
4.7.1. HFC-203 Orifice  
To change the orifice in the HFC-203 unit, the valve must be dismantled. Remove the four 1/4"  
Allen head machine screws in the top of the main valve. Lift off the valve top, exposing the spring  
and diaphragm. Note that a small brazed ball bearing  
is on the down stream side of the valve top. Remove  
the spring and diaphragm assembly. The orifice is  
located in the bottom of the valve body and can be  
removed with a 9/16 socket wrench. See Figure 4.5.  
To reinstall an orifice, first install the gasket onto the  
orifice (replacement gaskets can be obtained from the  
factory). Next screw the orifice into the valve base.  
Snug up the orifice but do not over tighten. Place  
diaphragm assembly into the base. Line up the two  
small holes in the diaphragm with the two small holes  
in the valve base. Place the spring on top of the  
141-072011 - 201/203 Series  
Page 21 of 23  
diaphragm. Examine the o-ring on the valve top for damage if required. Install the valve top,  
ensuring that the ball bearing in the side is on the downstream side. Tighten down the valve top  
evenly to insure a proper seal at the diaphragm.  
141-072011 - 201/203 Series  
Page 22 of 23  
5. WARRANTY  
5.1. Warranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment to be  
free from defects in material and workmanship. This warranty does not apply to defects or failures  
resulting from unauthorized modification, misuse or mishandling of the product. This warranty does  
not apply to batteries or other expendable parts, or to damage caused by leaking batteries or any  
similar occurrence. This warranty does not apply to any instrument which has had a tamper seal  
removed or broken.  
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty  
as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or  
consequential damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if  
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship  
during the warranty period. Defective instruments should be returned to Hastings Instruments,  
shipment prepaid, together with a written statement of the problem and a Return Material  
Authorization (RMA) number.  
Please consult the factory for your RMA number before returning any product for repair. Collect  
freight will not be accepted.  
5.2. Non-Warranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA  
form and a written description of the problem with the instrument. If the repair cost is higher, you  
will be contacted for authorization before we proceed with any repairs. If you then choose not to  
have the product repaired, a minimum will be charged to cover the processing and inspection. Please  
consult the factory for your RMA number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
TOLL FREE  
FAX  
(757) 723-6531  
1-800-950-2468  
(757) 723-3925  
E MAIL  
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments  
web site.  
141-072011 - 201/203 Series  
Page 23 of 23  

Lexmark 138 User Manual
Panasonic Cu E9ckp5 User Manual
Samsung As18cm2n User Manual
Samsung Sh Eav Avmgh C User Manual
Sanyo Ch3682 User Manual
Schumacher 00 99 000886 User Manual
Sears Battery Charger 287123 User Manual
Sharp Ar 237m User Manual
Simplicity Sovereign Tp 400 2218 00 Sv Sma User Manual
Thermax 8510 User Manual