One For All Automobile Parts Hfc 302 User Manual

TELEDYNE  
HASTINGS  
INSTRUMENTS  
INSTRUCTION MANUAL  
HFM-300 FLOW METER,  
HFC-302 FLOW CONTROLLER  
Table of Contents  
1.  
2.  
GENERAL INFORMATION............................................................................................................................................ 4  
1.1.  
FEATURES .................................................................................................................................................................... 4  
SPECIFICATIONS........................................................................................................................................................... 5  
OPTIONAL 4-20 MA CURRENT OUTPUT............................................................................................................................ 5  
OTHER ACCESSORIES ...................................................................................................................................................... 6  
1.2.  
1.3.  
1.4.  
1.4.1. Hastings Power Supplies..................................................................................................................................... 6  
1.4.2. 300/302 Series Power Supply Interface Cables.............................................................................................. 6  
INSTALLATION AND OPERATION............................................................................................................................. 7  
2.1.  
RECEIVING INSPECTION ............................................................................................................................................... 7  
POWER REQUIREMENTS ............................................................................................................................................... 7  
OUTPUT SIGNAL........................................................................................................................................................... 7  
MECHANICAL CONNECTIONS....................................................................................................................................... 8  
2.2.  
2.3.  
2.4.  
2.4.1. Filtering.................................................................................................................................................................. 8  
2.4.2. Mounting ................................................................................................................................................................ 8  
2.4.3. Plumbing ................................................................................................................................................................ 8  
2.5.  
2.6.  
ELECTRICAL CONNECTIONS......................................................................................................................................... 8  
OPERATION.................................................................................................................................................................. 9  
2.6.1. Operating Conditions............................................................................................................................................. 9  
2.6.2. Zero Check ........................................................................................................................................................... 10  
2.6.3. High Pressure Operation ..................................................................................................................................... 10  
2.6.4. Blending of Gases................................................................................................................................................. 11  
2.7.  
2.8.  
2.9.  
2.10.  
2.11.  
2.12.  
OUTPUT FILTER ......................................................................................................................................................... 12  
CONTROLLING OTHER PROCESS VARIABLES ............................................................................................................. 12  
COMMAND INPUT....................................................................................................................................................... 13  
VALVE-OVERRIDE CONTROL..................................................................................................................................... 13  
GAIN POTENTIOMETER .............................................................................................................................................. 13  
TEMPERATURE COEFFICIENTS ................................................................................................................................... 14  
3.  
THEORY OF OPERATION ........................................................................................................................................... 15  
3.1.  
OVERALL FUNCTIONAL DESCRIPTION........................................................................................................................ 15  
SENSOR DESCRIPTION................................................................................................................................................ 15  
SENSOR THEORY........................................................................................................................................................ 15  
BASE.......................................................................................................................................................................... 17  
SHUNT DESCRIPTION .................................................................................................................................................. 17  
SHUNT THEORY ......................................................................................................................................................... 17  
CONTROL VALVE....................................................................................................................................................... 21  
ELECTRONIC CIRCUITRY............................................................................................................................................ 21  
3.2.  
3.3.  
3.4.  
3.5.  
3.6.  
3.7.  
3.8.  
4.  
MAINTENANCE.............................................................................................................................................................. 22  
4.1.  
4.2.  
TROUBLESHOOTING ................................................................................................................................................... 22  
ADJUSTMENTS ........................................................................................................................................................... 23  
4.2.1. Calibration Procedure ......................................................................................................................................... 23  
4.3.  
4.4.  
4.5.  
END CAP REMOVAL................................................................................................................................................... 23  
PRINTED CIRCUIT BOARD REPLACEMENT.................................................................................................................. 24  
SENSOR REPLACEMENT ............................................................................................................................................. 24  
5.  
6.  
7.  
8.  
GAS CONVERSION FACTORS.................................................................................................................................... 25  
VOLUMETRIC VS MASS FLOW................................................................................................................................. 27  
DRAWINGS AND REFERENCES ................................................................................................................................ 28  
WARRANTY .................................................................................................................................................................... 31  
8.1.  
8.2.  
WARRANTY REPAIR POLICY ...................................................................................................................................... 31  
NON-WARRANTY REPAIR POLICY ............................................................................................................................. 31  
300-302 Series  
Page 3 of 31  
1. General Information  
The Teledyne Hastings HFM-300 is used to measure mass flow rates in gases. In addition to flow rate  
measurement, the HFC-302 includes a proportional valve to accurately control gas flow. The Hastings mass flow  
meter (HFM-300) and controller (HFC-302), hereafter referred to as the Hastings 300 series, are intrinsically  
linear and are designed to accurately measure and control mass flow over the range of 0-5 sccm to 0-10 slm  
with an accuracy of better than ±0.75% F.S. at 3σ from the mean (versions >10 slm are ±1.0% F.S.) . Hastings  
mass flow instruments do not require any periodic maintenance under normal operating conditions with clean  
gases. No damage will occur from the use of moderate overpressures (~500 psi/3.45MPa) or overflows.  
Instruments are normally calibrated with the appropriate standard calibration gas (nitrogen) then a correction  
factor is used to adjust the output for the intended gas. Calibrations for other gases, such as oxygen, helium and  
argon, are available upon special order.  
1.1. Features  
LINEAR BY DESIGN. The Hastings 300 series is intrinsically linear (no linearization circuitry is  
employed). Should recalibration (a calibration standard is required) in the field be desired, the  
customer needs to simply set the zero and span points. There will be no appreciable linearity change  
of the instrument when the flowing gas is changed.  
NO FOLDOVER. The output signal is linear for very large over flows and is monotonically increasing  
thereafter. The output signal will not come back on scale when flows an order of magnitude over the  
full scale flow rate are measured. This means no false acceptable readings during leak testing.  
MODULAR SENSOR. The Hastings 300 series incorporates a removable/replaceable sensor module.  
Field repairs to units can be achieved with a minimum of production line downtime.  
LARGE DIAMETER SENSOR TUBE. The Hastings 300 sensor is less likely to be clogged due to its large  
internal diameter (0.026”/ 0.66mm). Clogging is the most common cause of failure in the industry.  
LOW P. The Hastings 300 sensor requires a pressure of approximately 0.25 inches of water (62 Pa) at  
a flow rate of 10 sccm. The low pressure drop across this instrument is ideal for leak detection  
applications since the pneumatic settling times are proportional to the differential pressure.  
FAST SETTLING TIME. Changes in flow rate are detected in less than 250 milliseconds when using the  
standard factory PC board settings.  
LOW TEMPERATURE DRIFT. The temperature coefficient of span for the Hastings 300 series is less  
than 0.03% of full scale/°C from 15-50°C. The temperature coefficient of zero is less than 0.1 % of  
reading/°C from 0-60°C.  
FIELD RANGEABLE. The Hastings 300 series is available in ranges from 0-5 sccm to 0-25 slpm. Each  
flow meter has a shunt which can be quickly and easily exchanged in the field to select different  
ranges. Calibration, however, is required.  
METAL SEALS. The Hastings 300 series is constructed of Stainless Steel. All internal seals are made  
with Ni 200 gaskets, eliminating the permeation, degradation and outgassing problems of elastomer O-  
rings.  
LOW SURFACE AREA. The shunt is designed to have minimal wetted surface area and no un-swept  
volumes. This will minimize particle generation, trapping and retention. CURRENT LOOP. The 4-20  
mA option gives the user the advantages of a current loop output to minimize environmental noise  
pickup.  
300-302 Series  
Page 4 of 31  
1.2. Specifications  
Accuracy .................................................................................... < ±0.75% full scale (F.S.) at 3σ  
(±1.0% F.S. for >10 slm versions)  
Repeatability.............................................................................±0.05% of reading + 0.02% F.S.  
Maximum Pressure........................................................................................500 psi [3.45 MPa]  
(With high pressure option) 1000 psi [6.9 MPa]  
Pressure Coefficient .................................................. <0.01% of reading/psi [0.0015%/kPa] (N2)  
See pressure section for higher pressure errors.  
Operating Temperature ....................................................0-60°C in non-condensing environment  
Temperature Coefficient (zero) .................................... Maximum ±0.1%F.S./°C (from 0 to 60°C)  
Temperature Coefficient (span) ..................................Maximum ±300 ppm/°C (from 15 to 50 °C)  
Maximum ±450 ppm/°C (from 0 to 60 °C)  
Leak Integrity ...............................................................................................<1x10-9 std. cc/s.  
Flow Ranges .................................................................. 0-5 sccm to 0-25* slpm. (N2 Equivalent)  
Standard Output ........................................................................... 0-5 VDC. (load min 2k Ohms)  
Optional Output ............................................................................ 4 -20 mA. (load < 600 Ohms)  
Power Requirements..................................................................... ±(15) VDC @ 55 mA (meters)  
± (15) VDC @ 150 mA (controller)  
Class 2 power 150VA max  
Wetted Materials ................................................................................ stainless steel, nickel 200  
Attitude Sensitivity of zero..............................................< ±0.7% F.S. for 90° without re-zeroing  
{N2 at 19.7 psia (135 KPa)}  
Weight ............................................................................................................1.93 lb [0.88 kg]  
Electrical Connector............................................................................... 15 pin subminiature “D”  
Fitting Options................... ¼” Swagelok®, 1/8” Swagelok®, VCR®, VCO®, 9/16”-18 Female thread  
Face Seal to Face Seal Length................................................................1.88”(47.75 mm) VCR®  
(Specifications may vary for instruments with ranges greater than 10 slpm)  
1.3. Optional 4-20 mA Current Output  
An option to the standard 0-5 VDC output is the 4-20 mA current output that is proportional to flow.  
The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current loop output  
is useful for remote applications where pickup noise could substantially affect the stability of the  
voltage output.  
The current loop signal replaces the voltage output on pin 6 of the “D” connector. The current loop  
may be returned to either the signal common or the -15 VDC connection on the power supply. If the  
current loop is returned to the signal common, the load must be between 0 and 600 ohm. If it is  
returned to the -15VDC, the load must be between 600 and 1200 ohm. Failure to meet these  
conditions will cause failure of the loop transmitter.  
300-302 Series  
Page 5 of 31  
The 4-20 mA I/O option can accept a current input. The 0-5 VDC command signal on pin 14 can be  
replaced by a 4-20mA command signal. The loop presets an impedance of 75 ohms and is returned to  
the power supply through the valve common.  
1.4.Other Accessories  
1.4.1. Hastings Power Supplies  
Hastings power supplies are available in one or four channel versions. They convert 115 or 230 VAC to  
the ±15 VDC required to operate the flow meter. Interface terminals for the ±15 VDC input and the 0-5  
VDC linear output signal are located on the rear of the panel. Throughout this manual, when reference  
is made to a power supply, it is assumed the customer is using a Hastings supply. Hastings PowerPod-  
100 and PowerPod-400 power supplies are CE marked, but the Model 40 does not meet CE standards at  
this time. The Model 40 and PowerPod-100 are not compatible with 4–20 mA analog signals. With the  
PowerPod 400, individual channels’ input signals, as well as their commands, is become 4–20 mA  
compatible when selected. The PowerPod-400 also provides a totalizer feature.  
1.4.2. 300/302 Series Power Supply Interface Cables  
The Hastings 300 series normally comes with the standard “H” pin-out connector, which uses the AF-8-  
AM cable with grey backshells. “U” pin-out versions of the 300 series instruments require a different  
cable to connect to the Hastings Instruments power supply. This cable is identifiable by black backshell  
and is available as Hastings Instrument P/N 65-791.  
300-302 Series  
Page 6 of 31  
2. Installation and Operation  
This section contains the steps necessary to assist in getting a new flow meter/controller into  
operation as quickly and easily as possible. Please read the following thoroughly before attempting to  
install the instrument.  
2.1.Receiving Inspection  
Carefully unpack the Hastings unit and any accessories that have also been ordered. Inspect for any  
obvious signs of damage to the shipment. Immediately advise the carrier who delivered the shipment  
if any damage is suspected. Check each component shipped with the packing list. Insure that all parts  
are present (i.e., flow meter, power supply, cables, etc.). Optional equipment or accessories will be  
listed separately on the packing list. There may also be one or more OPT-options on the packing list.  
These normally refer to special ranges or special gas calibrations. They may also refer to special  
helium leak tests, or high pressure tests. In most cases, these are not separate parts, rather, they are  
special options or modifications built into the flow meter.  
Quick Start  
1. Insure flow circuit mechanical connections are leak free  
2. Insure electrical connections are correct (see label).  
3. Allow 30 min. to 1 hour for warm-up.  
4. Note the flow signal decays toward zero.  
5. Run ~20% flow through instrument for 5 minutes.  
6. Insure zero flow; wait 2 minutes, then zero the instrument.  
7. Instrument is ready for operation  
2.2.Power Requirements  
The HFM-300 meter requires +15 VDC @ 55 mA, -15 VDC @50 mA for proper operation. The HFC-302  
controller requires ±15 VDC @ 150mA. The supply voltage should be sufficiently regulated to no more  
than 50 mV ripple. The supply voltage can vary from 14.0 to 16.0 VDC. Surge suppressors are  
recommended to prevent power spikes reaching the instrument. The Hastings power supply described  
in Section 1.4.2 satisfies these power requirements.  
2.3. Output Signal  
The standard output of the flow meter is a 0-5 VDC signal proportional to the flow rate. In the  
Hastings power supply the output is routed to the display, and is also available at the terminals on the  
rear panel. If a Hastings supply is not used, the output is available on pin 6 of the “D” connector. It is  
recommended that the load resistance be no less that 2k. If the optional 4-20 mA output is used, the  
load impedance must be selected in accordance with Section 1.3.  
300-302 Series  
Page 7 of 31  
2.4.Mechanical Connections  
2.4.1. Filtering  
The smallest of the internal passageways in the Hastings 300 is the diameter of the sensor tube, which  
is 0.026”(0.66 mm), and the annular clearance for the 500 sccm shunt which is 0.006"(0.15 mm) (all  
other flow ranges have larger passages), so the instrument requires adequate filtering of the gas supply  
to prevent blockage or clogging of the tube.  
2.4.2. Mounting  
There are two mounting holes (#8-32 thread) in the bottom of the transducer that can be used to  
secure it to a mounting bracket, if desired.  
The flow meter may be mounted in any position as long as the direction of gas flow through the  
instrument follows the arrow marked on the bottom of the flow meter case label. The preferred  
orientation is with the inlet and outlet fittings in a horizontal plane.  
As explained in the section on operating at high pressures, pressure can have a significant affect on  
readings and accuracy. When considering mounting a flow meter in anything other than a horizontal  
attitude, consideration must be given to the fact that the heater coil can now set up a circulating flow  
through the sensor tube, thereby throwing the zero off. This condition worsens with denser gases or  
with higher pressures. Whenever possible, install the instrument horizontally.  
Always re-zero the instrument with zero flow, at its normal operating temperature and purged with its  
intended gas at its normal operating pressure.  
2.4.3. Plumbing  
The standard inlet and outlet fittings for the Hastings 300 Series are VCR-4, VCO-4 or 1/4" Swagelok. It  
is suggested that all connections be checked for leaks after installation. This can be done by  
pressurizing the instrument (do not exceed 500 psig unless the flow meter is specifically rated for  
higher pressures) and applying a diluted soap solution to the flow connections.  
2.5. Electrical Connections  
If a power supply from Hastings Instruments is used, installation consists of connecting the HFM-  
300/302 series cable from the “D” connector on the rear of the power supply to the “D” connector on  
the top of the flow meter /controller. The “H” pin-out requires cable AF-8-AM (grey molded  
backshell). The “U” pin-out requires cable # 65-791 (black molded backshell).  
If a different power supply is used, follow the instructions below when connecting the flow meter and  
refer to either table 2.1 or 2.2 for the applicable pin-out. The power supply used must be bipolar and  
capable of providing ±15 VDC at 55 mA for flow meter applications and ±15 VDC at 150 mA for  
controllers. These voltages must be referenced to a common ground. One of the “common” pins must  
be connected to the common terminal of the power supply. Case ground should be connected to the  
AC ground locally. The cable shield (if available) should be connected to AC ground at the either the  
power supply end, or the instrument end of the cable, not at both. Pin 6 is the output signal from the  
flow meter. The standard output will be 0 to 5 VDC, where 5 VDC is 100% of the rated or full scale  
flow.  
The command (set point) input should be a 0-5 VDC signal (or 4-20mA if configured as such), and must  
be free of spikes or other electrical noise, as these would generate false flow commands that the  
controller would attempt to follow. The command signal should be referenced to signal common.  
A valve override command is available to the flow controller. Connect the center pin of a single pole,  
three-position switch (center off) to the override pin. Connect +15 VDC to one end of the three  
position switch, and -15 VDC to the other end. The valve will be forced full open when +15 VDC is  
supplied to the override pin, and full closed when -15 VDC is applied. When there is no connection to  
the pin (the three-position switch is centered) the valve will be in auto control, and will obey the 0-5  
VDC commands supplied to command (set-point) input.  
300-302 Series  
Page 8 of 31  
Fig. 2.1  
Fig. 2.2  
Figures 2.1/2.2, and Tables 2.1/2.2, show the 300/302 pin out.  
Table 2.1  
Table 2.2  
"U" Pin-Out  
"H" Pin-Out  
Pin #  
Pin #  
1
2
3
4
5
6
7
8
9
Signal Common  
Do not use  
Do not use  
+15 VDC  
1
2
3
4
5
6
7
8
9
Do not use  
Do not use  
Do not use  
Do not use  
Signal Common  
Output 0-5 VDC (4-20mA)  
Case Ground  
Valve Override  
-15VDC  
Output 0-5 VDC (4-20mA)  
Signal Common  
Case Ground  
Valve Override  
10  
10 Do not use  
11 -15VDC  
11 +15VDC  
12 External Input  
13 Signal Common  
14 Signal Common  
15 Set Point 0-5 VDC (4-20mA)  
12 Signal Common  
13 External Input  
14 Set Point 0-5 VDC (4-20mA)  
15 Do not use  
2.6.Operation  
The standard instrument output is a 0 - 5 VDC out and the signal is proportional to the flow i.e.,  
0
volts = zero flow and 5 volts = 100% of rated flow. The 4 - 20 mA option is also proportional to flow, 4  
mA = zero flow and 20 mA = 100% of rated flow.  
2.6.1. Operating Conditions  
For proper operation, the combination of ambient temperature and gas temperature must be such that  
the flow meter temperature remains between 0 and 60°C. (Most accurate measurement of flow will  
be obtained if the flow meter is zeroed at operating temperature as temperature shifts result in some  
zero offset.) The Hastings 300 series instrument is intended for use in non-condensing environments  
only. Condensate or any other liquids which enter the flow meter may destroy its electronic  
components.  
300-302 Series  
Page 9 of 31  
2.6.2. Zero Check  
Turn the power supply on if not already energized. Allow for a 1 hour  
warm-up. Stop all flow through the instrument and wait 2 minutes.  
Caution: Do not assume that all metering valves completely shut off  
the flow. Even a slight leakage will cause an indication on the meter  
and an apparent zero shift. For the standard 0-5 VDC output, adjust  
the zero potentiometer located on the inlet side of the flow meter  
until the meter indicates zero (Fig 2.3). For the optional 4-20 mA  
output, adjust the zero potentiometer so that the meter indicates  
slightly more than 4 mA, i.e. 4.03 to 4.05 mA. This slight positive  
adjustment ensures that the 4-20 mA current loop transmitter is not in  
the cut-off region.  
The error induced by this adjustment is  
approximately 0.3% of full scale. This zero should be checked  
periodically during normal operation. Zero adjustment is required if  
there is a change in ambient temperature, or vertical orientation of  
the flow meter /controller.  
Fig. 2.3  
2.6.3. High Pressure Operation  
When operating at high pressure, the increased density of gas will cause natural convection to flow  
through the sensor tube if the instrument is not mounted in a level position. This natural convection  
flow will be proportional to the system pressure. This will be seen as a shift in the zero flow output  
that is directly proportional to the system pressure.  
Fig. 2.4  
Span Error vs Pressure  
(0.026" Sensor Tube)  
2.0%  
0.0%  
-2.0%  
-4.0%  
-6.0%  
-8.0%  
-10.0%  
-12.0%  
-14.0%  
Mean error  
-16.0%  
max  
min  
-18.0%  
-20.0%  
0
100  
200  
300  
400  
500  
600  
700  
800  
900  
1000  
y = 9.8877E-11x33.4154E-07x - 28.3288E-05x +  
Pressure(psig)  
300-302 Series  
Page 10 of 31  
Span Error Vs. Pressure  
0.017" Sensor  
Fig. 2.5  
5%  
4%  
3%  
2%  
1%  
0%  
-1%  
Mean  
Max  
Min  
-2%  
0
100  
200  
300  
400  
500  
600  
700  
800  
900  
1000  
Pressure (psig)  
If the system pressure is higher than 250 psig (1.7 MPa) the pressure induced error in the span reading  
becomes significant. The charts above show the mean error enveloped by the minimum/maximum  
expected span errors induced by high pressures. This error will approach 16% at 1000 psig. For  
accurate high pressure measurements this error must be corrected.  
The formulae for predicting mean error expressed as a fraction of the reading are:  
Error26 = (9.887*1011 )P3 (3.4154*107 )P2 + (8.3288*105 )P, (0.026" Sensor)  
Error = (1.533*10-10 )P3 (3.304 *107 )P2 + (1.8313*104 )P, (0.017" Sensor)  
17  
Where P is the pressure in psig and Error is the fraction of the reading in error.  
The flow reading can be corrected as follows:  
(
)
Corrected = IndicationIndication* Error  
Where the Indication is the indicated flow and Error is the result of the previous formula (or read from  
charts above).  
2.6.4. Blending of Gases  
This section describes two methods by which to achieve a controlled blending of different gasses. Both  
methods use the flow signal (Output) from one flow instrument as the Master to control the Command  
signal (Input) to a second unit.  
The first method requires that the two controllers use the same signal range (0 to 5 VDC or 4 to 20 mA)  
and that they be sized and calibrated to provide the correct ratio of gasses. Then, by routing the  
actual flow Output signal from the primary meter/controller through the secondary controller’s  
300-302 Series  
Page 11 of 31  
External Input pin (See Tables 2.1 & 2.2), the ratio of flows can be maintained over the entire range of  
gas flows.  
EXAMPLE: Flow controller A has 0-100 slpm range with a 5.00 volt output at full scale. Flow  
controller B has 0-10 slpm range with a 5.00 volt output at full scale. If flow controller A is set at 80  
slpm, its output voltage would be 4.00 volts (80 slpm/100 slpm x 5.00 volts = 4.00 volts). If the output  
signal from flow controller A is connected to the command Set Point of flow controller B, then flow  
controller B becomes a slave to the flow signal of controller A. The resultant flow of controller B will  
be the same proportion as the ratio of the flow ranges of the two flow controllers.  
If the set point of flow controller A is set at 50% of full scale, and the reference voltage from flow  
controller A is 2.50, then the command signal going to flow controller B would be 2.50 volts . The flow  
of gas through flow controller B is then controlled at 5 slpm (2.50 volts/5.00 volts x 10 slpm = 5 slpm).  
The ratio of the two gases is 10:1 (50 slpm/5slpm). The % mixture of gas A is 90.9090 (50slpm/55 slpm  
and the % mixture of gas B is 0.09091% (5 slpm/55 slpm).  
Should the flow of flow controller A drop to 78 slpm, flow controller B would drop to 3.9 slpm, hence  
maintaining the same ratio of the mixture. (78 slpm/100slpm x 5v = 3.90v x 50% = 1.95v; 1.95v/5.00v  
x 10 slpm = 3.9 slpm; 78 slpm: 3.9 slpm = 20:1)  
In the blending of two gases, it is possible to maintain a fixed ratio of one gas to another. In this case,  
the output of one flow controller is used as the reference voltage for the set point potentiometer of a  
second flow controller. The set point potentiometer then provides a control signal that is proportional  
to the output signal of the first flow controller, and hence controls the flow rate of the second gas as a  
percentage of the flow rate of the first gas.  
2.7.Output Filter  
Fig. 2.6  
The output signal may have noise superimposed on the mean  
voltage levels. This noise may be due to high turbulence in  
the flow stream that the fast sensor is measuring or it could  
be electrical noise when the flow meter has a high internal  
gain. i.e. 5 sccm full scale meter. Varying levels of radio  
frequency noise or varying airflow over the electronics cover  
can also induce noise.  
JP-1  
Noise can be most pronounced when measuring the flow  
output with a sampling analog/digital (A/D) converter.  
When possible, program the system to take multiple samples  
and average the readings to determine the flow rate.  
If less overall system noise is desired, a jumper may be  
installed over the pins of JP-1 on the flow measurement  
card. See Figure 2.6. Covering the pins closest to the “D”  
connector will activate a resistor-capacitor (RC) filter that  
has a time constant of one second. This will increase the  
settling time of the indicated flow rate to approximately 4 seconds. Covering the other two pins will  
lower the response time to approx. 1 second. This adjustment will not affect the calibration of the  
flow meter circuit or the actual flow response to change in command signal (flow controllers). This  
will only slow down the indicated response (output voltage/current).  
2.8.Controlling Other Process Variables  
Normally, a flow controller is setup to control the mass flow. The control loop will open and close the  
valve as necessary to make the output from the flow measurement match the input on the command  
line. Occasionally, gas is being added or removed from a system to control some other process  
variable. This could be the system pressure, oxygen concentration, vacuum level or any other  
parameter which is important to the process. If this process variable has a sensor that can supply an  
analog output signal proportional to its value then the flow controller may be able to control this  
variable directly. This analog output signal could be 0-5 volts, 0-10 volts (or 4-20 ma for units with 4-  
20 ma boards) or any value in between.  
300-302 Series  
Page 12 of 31  
On the controller card there is a jumper that sets whether the control loop controls mass flow or an  
external process variable. See Figure 2.7. If the jumper is over the top two pins, the loop controls  
mass flow. If the jumper is over the bottom two pins, the loop controls an external process variable.  
This process variable signal must be supplied on pin 12 of the D connector (for U pin out units) of the  
measurement card. When the controller is set for external variable control it will open or close the  
valve as necessary to make the external process variable signal match the command signal. The  
command signal may be 0-5 volts, 0-10 volts (4-20 ma for 4-20 ma input/output cards) or any value in  
between. If the process variable has a response time that is much faster or slower than the flow  
meter signal it may be necessary to adjust the gain potentiometer.  
2.9.Command Input  
The flow controller will operate normally with any command input signal between 0-5 volts (4-20 ma  
for units with 4-20 ma input/output cards) If the command signal exceeds ±14 volts it may damage the  
circuit cards. During normal operation the control loop will open or close the valve to bring the output  
of the flow meter signal to within ± 0.001 volts of the command signal. The command signal will not  
match the flow signal if there is insufficient gas pressure to generate the desired flow. If the  
command signal exceeds 5 volts the controller will continue to increase the flow until the output  
matches the command signal. However, the flow output does not have any guaranteed accuracy  
values under these conditions.  
If the command signal is less than 2% of full scale (0.1 volts or 4.32 ma) the valve override control  
circuit will activate in the closed position. This will force the valve completely closed regardless of  
the flow signal.  
2.10.Valve-Override Control  
The valve override control line provides a method to override the loop controller and open or close the  
valve regardless of the flow or command signals. During normal operation this line must be allowed to  
float freely. This will allow the loop control to open and close the valve as it requires. If the valve  
override line is forced high (> +5 volts) the valve will be forced full open. If the valve-override line is  
forced negative (< -5 volts) the valve will be forced closed.  
Fig. 2.7  
2.11.Gain Potentiometer  
Gain Potentiometer  
On the top left of inlet side of the flow controller there is a hole  
through which the gain potentiometer is accessible (Fig 2.3). This  
gain potentiometer affects the gain of the closed loop controller.  
Normally this potentiometer will be set at the factory for good  
stable control. It may be necessary to adjust this potentiometer  
in the field if the system varies widely from the conditions under  
which the controller was setup. Turning this gain potentiometer  
clockwise will improve stability.  
Turning the potentiometer  
counter-clockwise will speed up the valve reaction time to  
changes in the command signal.  
Control Loop Jumper  
300-302 Series  
Page 13 of 31  
2.12.Temperature Coefficients  
As the ambient temperature of  
the instrument changes from  
the original calibration  
Fig. 2.8  
temperature, errors will be  
introduced into the output of  
the instrument. The  
Temperature Coefficient of  
Zero describes the change in  
the output that is seen at zero  
flow. This error is added to the  
overall output signal regardless  
of flow, but can be eliminated  
by merely adjusting the zero  
potentiometer of the flow  
meter/controller to read zero  
volts at zero flow conditions.  
The Temperature Coefficient of  
Span describes the change in  
output after the zero error is  
eliminated. This error cannot  
be eliminated, but can be  
compensated for  
mathematically if necessary.  
The curve pictured in Figure 2.8  
shows the span error in percent  
of point as a function of  
temperature assuming 230C is  
the calibration temperature.  
300-302 Series  
Page 14 of 31  
3. Theory of Operation  
This section contains an overall functional description of the Hastings 300 series of flow instruments.  
In this section and other sections throughout this manual, it is assumed that the customer is using a  
Hastings power supply.  
3.1.Overall Functional Description  
The Hastings 300 meter consists of a sensor, base, and a shunt. In addition to the components in a  
meter, The 300 controller includes a control valve and extra electronic circuitry. The sensor is  
configured to measure gas flow rate from 0 to 5 sccm, 0 to 10 sccm, or 0 to 20 sccm, depending on the  
customer’s desired overall flow rate. The shunt divides the overall gas flow such that the flow through  
the sensor is a precise percentage of the flow through the shunt. The flow through both the sensor  
and shunt is laminar. The control valve adjusts the flow so that the sensor’s flow measurement  
matches the set-point input. The circuit board amplifies the sensor output from the two RTD’s  
(Resistive Temperature Detectors) and provides an analog output of either 0-5 VDC or 4-20 mA.  
3.2.Sensor Description  
A cross section of the sensor is shown in Figure 3.1. The sensor consists of two coils of resistance wire  
with a high temperature coefficient of resistance (3500 ppm/oC) wound around a stainless steel tube  
with internal diameter of 0.6604 mm and 7.62 cm length. Each coil is 1.372 cm in length, and they are  
separated by 1.27 mm distance. These two identical resistance wire coils are used to heat the gas  
stream and are symmetrically located upstream and downstream on the sensor tube. Insulation  
surrounds the sensor tube and heater coils with no voids around the tube to prevent any convection  
losses. The ends of this sensor tube pass through an aluminum block and into the stainless steel sensor  
base. This aluminum block thermally shorts the ends of the sensor tube and maintains them at  
ambient temperature.  
There are two coils of resistance wire that are wound around the aluminum block. The coils are  
identical to each other, and are symmetrically spaced on the aluminum ambient block. These coils are  
wound from the same spool of wire that is used for the sensor heater coils so they have the same  
resistivity and the same temperature coefficient of resistance as the sensor heater coils. The number  
of turns is controlled to have a resistance that is 10 times larger than the resistance of the heater  
coils. Thermal grease fills any voids between the ambient temperature block and the sensor tube to  
ensure that the ends of the sensor tube are thermally tied to the temperature of this aluminum block.  
Aluminum has a very high thermal conductivity which ensures that both ends of the sensor tube and  
the two coils wound around the ambient block will all be at the same temperature. This block is in  
good thermal contact with the stainless steel base to ensure that the ambient block is at the same  
temperature as the main instrument block and, therefore, the same temperature as the incoming gas  
stream. This allows the coils wound on the aluminum block to sense the ambient gas temperature.  
Two identical Wheatstone bridges are employed, as shown in Figure 3.2. Each bridge utilizes an  
ambient temperature sensing coil and a heater coil. The heater coil and a constant value series  
resistor comprise the first leg of the bridges. The second leg of each bridge contains the ambient  
sensing coil and two constant value series resistors. These Wheatstone bridges keep each heater  
temperature at a fixed value of dT = 48°C above the ambient sensor temperature through the  
application of closed loop control and the proper selection of the constant value bridge resistors.  
3.3.Sensor Theory  
Consider the sensor design shown in Figure 3.1. The heat convected to or from a fluid is proportional  
to the mass flow of that fluid.  
Since the constant differential temperature sensor has 2 heater coils symmetrically spaced on the  
sensor tube, it is convenient to consider the upstream and downstream heat transfer modes  
separately. The electrical power supplied to either of the heater coils will be converted to heat, which  
300-302 Series  
Page 15 of 31  
can be dissipated by radiation, conduction, or convection. The radiation term is negligible due to the  
low temperatures used by the sensor, and because the sensor construction preferentially favors the  
conductive and convective heat transfer modes. The thermal energy of each heater will then be  
dissipated by conduction down the stainless steel sensor tube, conduction to the insulating foam, plus  
the convection due to the mass flow of the sensed gas.  
Because great care is taken to wind the resistive heater coils symmetrically about the midpoint of the  
tube, it is assumed that the heat conducted along the sensor tube from the upstream heater will be  
equal to the heat conducted through the tube from the downstream heater. Similarly, the heat  
conducted from the upstream and downstream coils to the foam insulation surrounding them is  
assumed to be equal, based on the symmetry of the sensor construction.  
Since the sensor tube inlet and outlet are linked by an aluminum ambient bar, the high thermal  
conductivity of the bar provides a ‘thermal short’, constraining the ends of the sensor tube to be at  
equal surface temperature. Moreover, the tube ends and the aluminum ambient bar have intimate  
thermal communication with the main flow passageway prescribed by the main stainless steel flow  
meter body. This further constrains each end of the sensor tube to be equal to the ambient gas  
temperature.  
Further, since the length of each heater section is nearly 21 times greater than the inside tube  
diameter, the mean gas temperature at the tubes axial midpoint is approximately equal to the tube  
surface temperature at that point. Recall that the outside of the sensor tube is well insulated from the  
surroundings; therefore the tube surface temperature at the axial midpoint is very close to the  
operating temperature of the heater coils. The mean temperature of the gas stream is then  
approximately the same as the heater temperature. Assuming the mean gas temperature is equal to  
the heater temperature, it can be shown that the differential pressure is:  
Pμ P = 2mCp (Theater Tambient  
)
(3.1)  
d
The value of the constant pressure specific heat of a  
gas is virtually constant over small changes in  
temperature. By maintaining both heaters at the  
same, constant temperature difference above the  
ambient gas stream temperature, the difference in  
heater power is a function only of the mass flow  
rate. Fluctuations in ambient gas temperature  
which cause errors in conventional mass flow  
sensors are avoided; the resistance of the ambient  
sensing coil changes proportionally with the ambient  
temperature fluctuations, causing the closed loop  
control to vary the bridge voltage such that the  
heater resistance changes proportionally to the  
ambient temperature fluctuation.  
Fig. 3.1  
The power supplied to each of the 2 heater coils is  
easily obtained by measuring the voltage across the  
heater, shown as V2 on Figure 3.2, and the voltage  
across the fixed resistor R1. Since R1 is in series with  
the heater RH they have the same current flowing  
through them. The electrical power supplied to a  
given heater is then calculated:  
(
)
V1 V2 V2  
Ρ =  
(3.2)  
R1  
With a constant differential temperature applied to each heater coil and no mass flow through the  
sensor the difference in heater power will be zero. As the mass flow rate through the sensor tube is  
increased, heat is transferred from the upstream heater to the gas stream. This heat loss from the  
300-302 Series  
Page 16 of 31  
heater to the gas stream will force the upstream bridge control loop to apply more power to the  
upstream heater so that the 48oC constant differential temperature is maintained.  
The gas stream will increase in temperature due to the heat it gains from the upstream heater. This  
elevated gas stream temperature causes the heat transfer at the downstream heater to gain heat from  
the gas stream. The heat gained from the gas stream forces the downstream bridge control loop to  
apply less power to the downstream heater coil in order to maintain a constant differential  
temperature of 48oC.  
The power difference at the RTD’s is a function of  
the mass flow rate and the specific heat of the  
gas. Since the heat capacity of many gases is  
relatively constant over wide ranges of  
temperature and pressure, the flow meter may be  
calibrated directly in mass units for those gases.  
Changes in gas composition require application of  
a multiplication factor to the nitrogen calibration  
to account for the difference in heat capacity.  
The sensor measures up to 20 sccm full scale flow  
rate at less than 0.75% F.S. error. The pressure  
drop required for a flow of 20 sccm through the  
Fig. 3.2  
sensor is approximately 0.5 inches of H2O (125 Pa).  
3.4. Base  
The stainless steel base has a 1.5" by 1.0” (38.1 mm by 25.4 mm) cross-section and is 3.64"(92.5 mm)  
long. The length from face seal fitting to face seal fitting is 4.88” (124.0 mm). The base has an  
internal flow channel that is 0.75"(19.1 mm) diameter. Metal to metal seals are used between the  
base and endcaps, as well as the base and sensor module. Gaskets made of nickel 200 are swaged  
between mating face seals machined into the stainless steel parts. All metal seals are tested at the  
factory and have leak rates of less than 1x10-9 std. cc/s. Because of this corrosion resistant, all metal  
sealed design, the Hastings 300 can measure corrosive gases, which would damage elastomer sealed  
flow meters.  
3.5.Shunt description  
The flow rate of interest determines the size of the shunt required. As previously indicated, 9  
separate shunts are required for the range of flow spanning 5 sccm to 10 slpm full scale. These shunts  
employ a patented method of flow division, which results in a more linear flow meter. As a result, the  
Hastings 300 flow meter calibration is more stable when changing between measured gases.  
For the 5 sccm, 10 sccm, and 20 sccm flow rates a solid stainless steel shunt is used. The shunt uses a  
close tolerance fit to block the main flow passage thereby directing all flow through the sensor tube.  
The 50 sccm flow range uses a stainless steel shunt which has been machined flat on an edge. The gap  
between the main flow passage and the flat machined on the shunt creates an alternate laminar flow  
passage such that the overall gas flow is split precisely between the sensor and the shunt. By  
increasing the number of flats and the size of the laminar shunt passageway, flow rates up to 200 sccm  
are accommodated.  
For flow rates above 200 sccm, the shunts are made so that an annular flow passage is formed between  
the shunt cylinder and the main flow passage. A stainless steel plug with an annular spacing of  
0.006"(0.15 mm) accommodates the 500 sccm flow range. Increased flow rates require larger gap  
dimensions. Eventually, a maximum annular gap dimension for laminar flow is obtained (~0.020"(0.5  
mm)). This patented shunt technology also includes inboard sensor ports which ensure laminar flow  
without the turbulence associated with end effects. This unique flow geometry provides an  
exceedingly linear shunt.  
3.6.Shunt Theory  
A flow divider for a thermal mass flow transducer usually consists of an inlet plenum, a flow  
restriction, shunt and an outlet plenum. (See Figure 3.3) Since stability of the flow multiplier is  
300-302 Series  
Page 17 of 31  
desired to ensure a stable instrument, there must be some matching between the linear volumetric  
flow versus pressure drop of the sensor and the shape of the volumetric flow versus pressure drop of  
the shunt. Most instruments employ  
Poiseuille’s law and use some sort of  
multi-passage device that creates  
laminar flow between the upstream  
sensor inlet and the downstream outlet.  
This makes the volumetric flow versus  
pressure drop curve primarily linear, but  
there are other effects which introduce  
higher order terms.  
Most flow transducers are designed such  
that the outlet plenum has a smaller  
diameter than the inlet plenum. This  
eases the insertion and containment of  
the shunt between the sensor inlet point  
and the sensor outlet point. If the shunt  
is removed, the energy of the gas must be conserved when passing from the inlet  
Fig. 3.3  
plenum to the outlet plenum. From Bernoulli’s equation, the sum of the kinetic  
energy and the pressure at each point must be a constant. Since all of the pressure  
drops are small, it can be assumed that the flow is incompressible.  
The pressure drop over the shunt can be shown to be:  
4
Di  
1
2
(3.3)  
ΔΡa = ρVi ⎢  
1⎥  
2
Do  
We can see that even with no effect from the shunt there will be a pressure drop between the sensor  
inlet and outlet points. This pressure drop will be a strong function of the ratio of the two diameters.  
Since the drop is a square function of the flow velocity the differential pressure will be non-linear with  
respect to flow rate. Note also that the pressure drop is a function of density. The density will vary as  
a function of system pressure and it will also vary when the gas composition changes. This will cause  
the magnitude of the pressure drop due to the area change to be a function of system pressure and gas  
composition.  
Most of the shunts used contain or can be approximated by many short capillary tubes in parallel.  
From Rimberg1 we know that the equation for the pressure drop across a capillary tube contains terms  
that are proportional to the square of the volumetric flow rate. These terms come from the pressure  
drops associated with the sudden compression at the entrance and the sudden expansion at the exit of  
the capillary tube. The end effect terms are a function of density which will cause the quadratic term  
to vary with system pressure and gas composition. The absence of viscosity in the second term will  
cause a change in the relative magnitudes of the two terms whenever the viscosity of the flowing gas  
changes.  
2
128μLQ 8ρQ  
(3.4)  
ΔΡa =  
+
(
KC + Ke  
)
πD4  
π 2 D4  
The end effect for a typical laminar flow element, in air, account for approximately 4% of the total  
pressure drop. For hydrogen, however, which has a density that is about 14 times less than air and  
has a viscosity that is much greater than air, the second term is completely negligible. For the heavier  
gasses, such as sulfur hexafluoride which has a density 5 times that of air, the end effects will become  
10% of the total. This fundamental property makes it a difficult to maintain accuracy specifications  
when calibrating an instrument using one gas for use with another gas.  
The pressure drop is linear with respect to the volumetric flow rate between a point that is  
downstream of the entrance area and another point further downstream but upstream of the exit  
region. From Kays & Crawford, that entrance length (L ) of a capillary tube in laminar flow is a  
function of the Reynolds number and the tube diameter. It can be shown that:  
300-302 Series  
Page 18 of 31  
Qρ  
5πμ  
Le =  
(3.5)  
For a typical flow divider tube the entry length is approximately 0.16 cm. From this it can be seen  
that if the sensor inlet pickup point is inside of the flow divider tube but downstream of the entrance  
length and if the sensor outlet point is inside the flow divider tube but upstream of the exit point then  
the pressure drop that drives the flow through the sensor would be linear with respect to volumetric  
flow rate. Since the pressure drop across the sensor now increases linearly with the main flow rate  
and the sensor has a linearly increasing flow with respect to pressure drop, there is now a flow through  
the sensor which is directly proportional to the main flow through the flow divider, without the flow  
division errors that are present when the sensor samples the flow completely upstream and  
downstream of the flow divider.  
Unfortunately, a typical shunt has an internal diameter on the order of 0.3 mm. This is too small to  
insert tap points into the tube. Also, the sample flow through the sensor is approximately 10 sccm  
while the flow through a shunt is approximately 25 sccm. This means the sample flow would be  
affecting the flow it was trying to  
measure. If the sensor tube is  
made large enough, and with  
enough flow through it to insert  
the sensor taps at these  
positions, then the pressure drop  
would be too small to push the  
necessary flow through the sensor  
tube.  
The solution is to use a different  
geometry for the flow tube. It  
must be large enough to allow  
the sample points in the middle  
yet with passages thin enough to  
create the differential pressures  
required for the sensor.  
annular passage meets these  
An  
Fig. 3.4  
requirements.  
The basic operation is similar to the operation of the tubular shunt but the equations for the entry  
length and pressure drop will be different.  
If we assume that the annular region is very small, (Δr << r):  
24  
C f =  
(3.6)  
Re  
Then it can be shown that the pressure drop is:  
12QLμ  
Pi P =  
(3.7)  
o
3
πτ  
(
Δτ  
)
The shunt must generate a pressure drop at the desired full scale flow which drives the proper flow  
through the sensor tube to generate a full scale output from the sensor. Since the full scale flow of  
the sensor is the same for all of the different full scale flows that may pass through the shunt, the  
geometry must vary for the different full scale flows in order to generate the same pressured drop for  
all of them. From Equation 3.7 it can be seen that if the width of the annular ring is varied slightly it  
can correct for very large changes in the full scale flow rate (Q).  
Below is a graph showing how the thickness of the annular ring must be changed to create a passage  
that will properly divide the flow for various full scale flows. This graph is based on the 75 Pa pressure  
drop required to push full scale flow through a particular sensor that has 2 cm spacing between the  
inlet and outlet taps. The flow divider has an outside diameter of 0.95 cm.  
300-302 Series  
Page 19 of 31  
Fig. 3.5  
Thickness of the annular ring as a function of flow rate  
for a sensor with a 75 Pa drop and a 2 cm spacing.  
0.18  
0.16  
0.14  
0.12  
0.1  
0.08  
0.06  
0.04  
0.02  
0
0
5
10  
15  
20  
25  
30  
35  
Flow (liter/min)  
Each shunt must have a section of the annular region upstream of the upstream sensor tap to allow the  
flow to become fully developed before reaching the first tap. The entry length for the annular passage  
is then:  
Qρ  
40πτμ  
(
Δr  
)
Le =  
(3.8)  
Below is a graph that demonstrates the entry length that would be required to design a flow divider for  
various full scale flows. The parameters on the sensor that the flow divider must match are the same  
as the ones on the previous graph.  
Entrance length as a function of flow rate  
for an annular ring of the size specified in figure 3.5.  
Fig. 3.6  
4.5  
4
3.5  
3
2.5  
2
1.5  
1
0.5  
0
0
5
10  
15  
20  
25  
30  
35  
Flow (liter/min)  
300-302 Series  
Page 20 of 31  
3.7.Control Valve  
Fig. 3.7  
The control valve is an “automatic metering solenoid” valve (see Figure 3.7). While most solenoid  
valves operate in either the fully open or closed positions, the automatic metering solenoid valve is  
designed to control flow. A spring is used to hold a magnetic plunger assembly tightly against an  
orifice, thereby shutting off the flow. The magnetic plunger assembly is surrounded by a coil of  
magnet wire. When the coil is energized the electric current passing through the wire coil produces a  
magnetic field which attracts the plunger. The plunger assembly moves away from the orifice allowing  
the gas flow to pass between the orifice and the plunger seat. The distance between the orifice and  
the plunger seat, and thus the flow through the valve, is controlled by the amount of current supplied  
to the coil.  
The valve seat is made of Kalrez (or equivalent) per fluoroelastomer. The valve orifice is made from  
Stainless Steel. The valve plunger and pole piece are made of nickel plated magnetic alloy (Hi-perm  
49) and the control springs are made of 302 stainless steel. Nickel gaskets seal all interfaces between  
the process gas and the outside environment, as described in section 3.4.  
3.8. Electronic Circuitry  
The Hastings 300 employs a thermal transfer principle (capillary tube described in section 3.2) to  
measure the flow through the sensor which is proportional to the total flow through the instrument.  
The sensor develops a differential voltage output signal proportional to flow, which is amplified to  
produce 5 VDC at full scale flow. The amplified output can be measured on the external “D”  
connector. If a Hastings power supply is employed, the 5 volt output is also sent to the terminals on  
the back and to the decoding circuitry in the display, the optional 4-20 mA analog output is available  
in lieu of an output voltage. The addition of a 4-20 mA current loop transmitter on a secondary PC  
board (mounted parallel to the main pc board) is required to provide this current loop. A jumper  
change is made on the secondary PC board to establish the selected output mode.  
300-302 Series  
Page 21 of 31  
4. Maintenance  
This section contains service and calibration information. Some portions of the instrument are  
delicate. Use extreme care when servicing the instrument. Authorized Maintenance  
With proper care in installation and use, the instrument will require little or no maintenance. If  
maintenance does become necessary, most of the instrument can be cleaned or repaired in the field.  
Most procedures may require recalibration. Do not attempt these procedures unless calibration  
references are available. Entry into the sensor or tampering with the printed circuit board will void  
warranty. Do not perform repairs on these assemblies while the unit is still under warranty.  
4.1.Troubleshooting  
Symptom: Output reads strong indication of flow with no flow present. Zero pot has no effect.  
Cause:  
Action:  
Power shorted out.  
Turn power supply off for a few seconds, and then turn it on again. If this is ineffective,  
disconnect the power supply from the unit. Check that the power supply voltages  
are correct. Incorrect voltages most likely signify a faulty regulator chip inside the  
supply. If the power supply display returns to zero after the instrument has been  
disconnected there may be a short from the unit to ground.  
Symptom: Hastings 300 output continues to indicate flow with no flow present, or indicates  
±14 volts. Power supply inputs are correct (see the above troubleshooting tip) and  
zero pot has no effect.  
Cause:  
Action:  
Faulty IC chip(s) on the main PC board.  
Replace main PC board. (See sections 4.5 and 6.1)  
Symptom: Output of flow meter is proportional to flow, but extremely small and not correctable  
by span pot.  
Cause:  
Action:  
Sensor is not being heated.  
Shut off gas supply and disconnect the power to the flow meter. Remove cover and  
PC board from unit. Check the resistance from pins 1 to 2, and 3 to 4 (refer to  
figures in section 6) of the sensor module. These pins should read 1650 Ω nominal  
resistance. Also check that the resistance from pins 5 to 6, and 7 to 8 are 400 Ω  
nominal value. Incorrect resistance values indicate that the sensor unit needs to be  
replaced.  
Symptom: Sensor has proper resistance readings, but little or no output with flow.  
Cause:  
Action:  
Plugged sensor.  
Shut off gas supply and disconnect the power to the flow meter. Remove cover and  
PC board from unit. Remove and inspect sensor. If sensor has evidence of clogging,  
clean or replace as applicable.  
Symptom: flow meter reads other than 0.00 VDC with no flow or there is a small flow when the  
flow meter reads 0.00 VDC.  
Cause:  
Zero pot is out of adjustment.  
300-302 Series  
Page 22 of 31  
Action:  
Shut off all flow. For the standard 0-5VDC output, adjust the zero potentiometer  
located on the upper right inlet side of the flow meter until the meter indicates zero.  
For the optional 4-20 mA output, adjust the zero potentiometer so that the meter  
indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. This slight positive  
adjustment ensures that the 4-20 mA transmitter is not in its cut-off region. The error  
induced by this adjustment is approximately 0.3% of full scale.  
Symptom: Flow meter is out of calibration and non-linear.  
Cause:  
Action:  
Leaks in the gas inlet or outlet fittings.  
Check all fittings for leaks by placing soap solution on all fittings between gas supply  
and final destination of gas. Check flow meter for leaks. Replace if required or  
recalibrate as necessary.  
Symptom: Little or no flow, even when the valve is in over-ride OPEN.  
Cause:  
Action:  
Blocked orifice or incorrect pressure across the Flowcontroller  
Verify that the pressure drop originally specified on the instrument is across the  
instrument. If the differential pressure across the instrument is correct, the orifice may  
be obstructed. Remove all gas pressure and shut off power supply. Remove the valve.  
4.2.Adjustments  
4.2.1. Calibration Procedure  
1.  
2.  
Calibration must take place with cover firmly in place.  
Fig. 4.1  
Connect power to “D” connector as specified in Section  
2.5. Allow the instrument to warm up for 60 minutes with  
10% of full scale flow.  
3.  
Completely shut off the flow and wait for 2 minutes. For  
the standard 0-5VDC output, adjust the zero  
potentiometer located on the lower inlet side of the flow  
meter until the meter indicates zero. For the optional 4-  
20 mA output, adjust the zero potentiometer so that the  
meter indicates slightly more than 4 mA, i.e. 4.03 to 4.05  
mA. This slight positive adjustment ensures that the 4-20  
mA transmitter is not in its cut-off region. The error  
induced by this adjustment is approximately 0.3% of full  
scale.  
4.  
5.  
6.  
Turn on gas supply to inlet of instrument and adjust the flow rate to the desired full scale  
flow as indicated by a reference flow meter/controller.  
Adjust Span pot until the indicated flow reads full scale (5.00VDC or 20 mA). Perform this  
step only if a calibrated reference flow meter is available.  
Record flow meter/controller and flow reference outputs for flow rates of 20%, 40%, 60%,  
80% and 100% and make sure data are within ± 0.75% 3σ of full scale.  
4.3.End Cap Removal  
The end cap on the inlet side must be removed to gain access to shunt assembly. First remove power  
and shut off the supply of gas to the instrument. Disconnect the fittings on the inlet and outlet sides  
of the transducer and remove it from the system plumbing. Remove the four Allen head screws  
holding the end cap to the instrument. Carefully remove the end cap, nickel gasket, spacer, and  
shunt, noting their order and proper orientation. The shunt can be severely damaged if dropped.  
Examine the shunt. If damaged, dirty or blocked, clean and replace as applicable. Reassemble in the  
300-302 Series  
Page 23 of 31  
reverse order of disassembly. A new nickel gasket will be required. Secure the endcap with 65 in lb.  
(7.3 N m) to 85 in lb (9.6 N m) of torque on each stainless steel socket head cap screw. Use of a  
fastener other than the one mentioned here may result in leakage at the seal. Recalibration of the  
Hastings 300 is necessary.  
4.4.Printed Circuit Board Replacement  
NOTE: This instrument contains static sensitive PC boards. Maintain static protection when handling  
the PC boards.  
In the event that any of the PC boards fail, they are easily removed from the instrument and replaced  
with a spare. This ease in disassembly and replacement substantially reduces instrument downtime.  
1.  
2.  
Replacement of the 4-20 mA option PC board: Unplug the power cable from the instruments “D”  
connector. Remove the fasteners and steel can. The 4-20 mA board is the PC board mounted by  
a single screw. Remove the screw and lift off the 4-20 mA board. Be careful not to damage the  
main board and 4-20 mA board connectors.  
Replacement of the main PC board: Unplug the power cable from the instruments “D”  
connector. Remove the fasteners and steel can. Remove the 2 screws which fasten the main PC  
board to the sensor module. Gently unplug the main board from the sensor (and from the  
4-20 mA board, if present).  
4.5.Sensor Replacement  
Follow instructions for removing the PC board(s) as described in Section 4.5. Remove the 4 Allen head  
cap screws that fasten the sensor to the main instrument base. Remove the sensor module from the  
base, discarding the used nickel gaskets. New nickel gaskets are required for re-assembly.  
To place an order or to obtain information concerning replacement parts, contact the factory  
representative in your area. See the last page in this manual for the address or phone number. When  
ordering, include the following information: Instrument model number, part description and Hastings  
part number.  
300-302 Series  
Page 24 of 31  
5. Gas Conversion Factors  
Gas conversion factors (GCF’s) for gasses metered using Hastings Instruments products, can be found by  
visiting the Hastings Instruments web site. The web address can be found at the end of this document.  
The gas conversion factors (GCF's) provided by Hastings Instruments (HI) fall into five basic accuracy  
domains that, to a large extent, are dependent on the method by which they are found. The following  
table summarizes the different methods used to determine the GCF's. The table lists the methods in  
decreasing order of the degree of accuracy that may be achieved when applying a conversion factor.  
Methods Used to Determine Gas Conversion Factors  
1. Determined empirically at Hastings Instruments  
2. Calculated From NIST tables  
3. Calculated using the virial coefficients of independent investigators' empirical  
data using both temperature and pressure as variables.  
4. Calculated from virial coefficients using temperature only.  
5. Calculated from specific heat data at 0 C and 1 atmosphere  
1. The most accurate method is by direct measurement. Gases that can be handled safely, inert gases,  
gases common in the atmosphere, etc., can be run through a standard flow meter and the GCF determined  
empirically.  
2. The National Institute of Standards and Technology (NIST) maintains tables of thermodynamic  
properties of certain fluids. Using these tables, one may look up the necessary thermophysical property  
and calculate the GCF with the same degree of accuracy as going directly to the referenced investigator.  
3 and 4. Many gases that have been investigated sufficiently by other researchers, can have their molar  
specific heat (C'p) calculated. The gas conversion factor is then calculated using the following ratio.  
Cp' N 2  
GCF =  
Cp' GasX  
GCF's calculated in this manner have been found to agree with the empirically determined GCF's within a  
few tenths of a percent. Data from investigations that factor in pressure as well as temperature, usually  
supply a higher degree of accuracy in their predictions.  
5. For rare, expensive gases or gases requiring special handling due to safety concerns, one may look up  
specific heat properties in a variety of texts on the subject. Usually, data found in this manner applies only  
in the ideal gas case. This method yields GCF's for ideal gases but as the complexity of the gas increases,  
its behavior departs from that of an ideal gas. Hence the inaccuracy of the GCF increases.  
Hastings Instruments continually searches for better estimates of the GCF's of the more complex gases  
and regularly updates the list.  
Most Hastings flow meters and controllers are calibrated using nitrogen. The conversion factors published  
by Hastings are meant to be applied to these meters. To apply the GCF's, simply multiply the gas flow  
reading and the GCF for the process gas in use. For example, to calculate the actual flow of argon passing  
through a nitrogen-calibrated meter that reads 20 sccm, multiply the reading and the GCF for argon.  
20 x 1.4047 = 28.094  
Conversely, to determine what reading to set a nitrogen-calibrated meter in order to get a desired flow rate  
of a process gas other than nitrogen, you divide the desired rate by the GCF. For example, to get a desired  
flow of 20 sccm of argon flowing through the meter, divide 20 sccm by 1.4047.  
20 / 1.4047 = 14.238  
That is, you set the meter to read 14.238 sccm.  
Some meters, specifically the high flow meters, are calibrated in air. The flow readings must be corrected  
for the case where a gas other than air is flowing through the meter. In addition, there must be a  
correction for the difference in the GCF from nitrogen to air. In this case, multiply the reading and the  
ratio of the process gas' GCF to the GCF of the calibration gas. For example, a meter calibrated in air is  
300-302 Series  
Page 25 of 31  
being used to measure the flow of propane. The reading from the meter is multiplied by the GCF for  
propane divided by the GCF of air.  
20 * (0.3499/1.0015) = 6.9875  
To calculate a target setting (20 sccm) to achieve a desired flow rate of propane using a meter calibrated to  
air, invert the ratio above and multiply.  
20 * (1.0015/0.3499) = 57.2449  
Gas conversion factors can be found at the Hastings Instruments web site.  
Follow the link to Mass Flow Products and then to Gas Conversion Factors.  
300-302 Series  
Page 26 of 31  
6. Volumetric Vs Mass Flow  
Mass flow measures just what it says, the mass or number of molecules of the gas flowing through the  
instrument. Mass flow (or weight per unit time) units are given in pounds per hour (lb/hour), kilograms  
per sec (kg/sec) etc. When your specifications state units of flow to be in mass units, there is no  
reason to reference a temperature or pressure. Mass does not change based on temperature or  
pressure.  
However, if you need to see your results of gas flow in volumetric units, like liters per minute, cubic  
feet per hour, etc. you must consider the fact that volume DOES change with temperature and  
pressure. To do this, the density (grams/liter) of the gas must be known and this value changes with  
temperature and pressure.  
When you heat a gas, the molecules have more energy and they move around faster, so when they  
bounce off each other, they become more spread out, therefore the volume is different for the same  
number of molecules.  
Think about this:  
The density of Air at 0°C is 1.29 g/liter  
The density of Air at 25°C is 1.19 g/liter  
The difference is 0.1 g/liter. If you are measuring flows of 100 liters per minute, and you don’t use  
the correct density factor then you will have an error of 10 g/minute!  
Volume also changes with pressure. Think about a helium balloon with a volume of 1 liter. If you  
could scuba dive with this balloon and the pressure on it increases. What do you think happens to the  
weight of the helium? It stays the same. What would happen to the volume (1 liter)? It would shrink.  
Why is the word standard included with the volume terms liters and cubic feet in mass flow  
applications?  
A mass flow meter measures mass …and we know we can convert to volume.  
To use density we must pick one (or standard) temperature and pressure to use in our calculation.  
When this calculation is done, the units are called standard liters per minute (SLM) or standard cubic  
feet per minute (SCFM), for instance, because they are referenced to a standard temperature and  
pressure when the volume is calculated.  
Using the example to the left, we can see a standard  
liter can be defined differently. The first balloon  
contains 0.179 grams of Helium at 0 °C and 760  
Torr (density of 0.179 grams/liter). Heat up that  
balloon to room temperature and the volume  
increases, but the mass has not changed. The volume  
is not 1 liter anymore, it is 1.08 liters.  
So, to define a standard liter of Helium at 25 °C, we  
must extract only one liter from the second balloon  
and that liter weighs only 0.175 grams.  
If a mass flow meter is set up for STP at 0 °C and  
1.08 Liter  
1 Liter  
25 °C  
1 Liter  
0 °C  
0.179 grams/1  
liter  
760 Torr, when it measures 0.179 grams of He, it  
will give you results of 1 SLM. If a second meter is  
set up for STP at 25 °C and 760 Torr, when it  
measures 0.164 grams, it will give results of 1 SLM.  
25 °C  
0.179 g/1.08  
liters  
0.164 grams  
300-302 Series  
Page 27 of 31  
7. Drawings and References  
HFM-300 / HFC-302 Outline Drawing  
(All dim. Are approximate)  
300-302 Series  
Page 28 of 31  
Sensor and Main PC  
Board  
TRANSDUCER  
HFM-300  
Sectional View  
ENDCAP  
SHUNT  
ENDCAP  
300-302 Series  
Page 29 of 31  
TRANSDUCER  
VALVE  
ASSEMBLY  
CONTROLLER  
CARD  
SHUNT  
ENDCAP  
HFC-302  
Sectional View  
ENDCAP  
300-302 Series  
Page 30 of 31  
8. WARRANTY  
8.1.Warranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment to be  
free from defects in material and workmanship. This warranty does not apply to defects or failures  
resulting from unauthorized modification, misuse or mishandling of the product. This warranty does  
not apply to batteries or other expendable parts, or to damage caused by leaking batteries or any  
similar occurrence. This warranty does not apply to any instrument which has had a tamper seal  
removed or broken.  
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as  
to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential  
damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if  
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship during  
the warranty period. Defective instruments should be returned to Hastings Instruments, shipment  
prepaid, together with a written statement of the problem and a Return Material Authorization (RMA)  
number.  
Please consult the factory for your RMA number before returning any product for repair. Collect freight  
will not be accepted.  
8.2.Non-Warranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form  
and a written description of the problem with the instrument. If the repair cost is higher, you will be  
contacted for authorization before we proceed with any repairs. If you then choose not to have the  
product repaired, a minimum will be charged to cover the processing and inspection. Please consult  
the factory for your RMA number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
TOLL FREE  
FAX  
(757) 723-6531  
1-800-950-2468  
(757) 723-3925  
E MAIL  
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments web  
site.  
300-302 Series  
Page 31 of 31  

Lexmark 4549 User Manual
Mitsubishi Electronics Msz Ga22va User Manual
Panasonic Cu 4e27cbpg User Manual
Royal Sovereign Arp 1000es User Manual
Royal Sovereign Nothing Arp 1008 User Manual
Sanyo Spw K075xh User Manual
Schumacher 85 716 User Manual
Source Technologies Ac024x1222a User Manual
Toshiba Ras 16gavp E User Manual
White Rodgers Acm1000 User Manual