No. MIIXR0002EAC
DICOM CONFORMANCE STATEMENT
MEDIA STORAGE
FOR
TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM
MODEL DFP-2000A
with XIDF-037A and XIDF-038A
or
SRS-1000A
(MIIXR0002EAC)
2001
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
Table of Contents
Page
REFERENCES..................................................................................................... 1
DEFINITIONS....................................................................................................... 1
IMPLEMENTATION MODEL............................................................................................. 3
APPLICATION DATA FLOW DIAGRAM ............................................................................... 3
FUNCTIONAL DEFINITIONS OF AE'S................................................................................. 3
DICOM Recording ............................................................................................... 3
DICOM Reader .................................................................................................... 4
SEQUENCING OF REAL WORLD ACTIVITIES ..................................................................... 4
Features............................................................................................................... 4
2.3.1.1 Transfer of X-ray Examination.......................................................................................4
2.3.1.2 Review of X-ray Examination ........................................................................................4
Operation ............................................................................................................. 5
2.3.2.1 Transfer of X-ray Examination.......................................................................................5
2.3.2.2 Review of X-ray Examination ........................................................................................5
3.1 DICOM RECORDING SPECIFICATION.............................................................................. 6
File Meta Information for the Application Entity................................................... 6
Real-World Activity .............................................................................................. 6
3.1.2.1 Real-World Activity - Transfer of X-ray Examination .....................................................6
Media Storage Application Profile........................................................................ 6
DICOM READER SPECIFICATION ................................................................................... 6
File Meta Information for the Application Entity................................................... 7
Real-World Activity .............................................................................................. 7
3.2.2.1 Real-World Activity - Review of X-ray Examination .......................................................7
Media Storage Application Profile........................................................................ 7
4.1 AUGMENTED APPLICATION PROFILES ............................................................................. 7
AUG-XABC-CD.................................................................................................... 7
4.1.1.1 SOP Class Augmentations............................................................................................7
4.1.1.2 Directory Augmentations...............................................................................................7
4.1.1.3 Other Augmentations ....................................................................................................7
PRIVATE APPLICATION PROFILES.................................................................................... 7
CONFIGURATION............................................................................................................. 8
ENTITY MODULE DEFINITIONS ........................................................................................ 9
- a -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
Media Storage Directory IOD Modules................................................................ 9
Media Storage XA IOD Modules.......................................................................... 9
XA IOD Modules ................................................................................................ 10
DICOM FILE META INFORMATION ................................................................................ 11
File-set Identification Module............................................................................. 11
Directory Information Module............................................................................. 12
DEFINITION OF SPECIFIC DIRECTORY RECORDS............................................................ 12
PATIENT keys ................................................................................................... 12
STUDY keys ...................................................................................................... 13
SERIES keys ..................................................................................................... 13
IMAGE keys....................................................................................................... 14
XA INFORMATION OBJECT DEFINISIONS........................................................................ 15
Patient Module................................................................................................... 15
General Study Module....................................................................................... 15
Patient Study Module......................................................................................... 15
General Series Module...................................................................................... 16
General Equipment Module............................................................................... 16
General Image Module ...................................................................................... 16
Image Pixel Module ........................................................................................... 17
Contrast/Bolus Module ...................................................................................... 17
Cine Module....................................................................................................... 17
8.5.10 Multi-frame Module ............................................................................................ 17
8.5.11 X-ray Image Module .......................................................................................... 18
8.5.12 X-ray Acquisition Module................................................................................... 18
8.5.13 XA Positioner Module ........................................................................................ 19
8.5.14 Curve Module..................................................................................................... 19
8.5.15 VOI LUT Module ................................................................................................ 19
8.5.16 SOP Common Module....................................................................................... 19
- b -
*
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
1 Introduction
This document is a DICOM Conformance Statement for Toshiba's Digital Fluorography
Systems. It is intended to provide the reader with the knowledge of how to integrate this product
within DICOM compliant equipments. It details the DICOM Service Classes, Information
Objects, and Application Profiles which are supported by this product.
If the reader is unfamiliar with DICOM, it is recommended that they read the DICOM
Specification (referenced below) prior to reading this conformance statement. Also note that
this document is formatted according to the DICOM Specification, Part 2: Conformance.
1.1 References
•
ACR-NEMA Digital Imaging and Communications in Medicine, DICOM V3.0.
1.2 Definitions
•
Information Object Definition (IOD) - An IOD is a data model which is an abstraction of
real-world information. This data model defines the nature and attributes relevant to the
class of real-world objects represented.
•
Service/Object Pair (SOP) Class - A SOP Class is defined by the union of an Information
Object Definition and a set of DIMSE Services. A DICOM Application Entity may support
one or more SOP Classes. Each SOP Class is uniquely identified by a SOP Class UID.
•
•
SOP Instance - A specific occurrence of a Information Object.
Transfer Syntax - The Transfer Syntax is a set of encoding rules that allow DICOM
Application Entities to negotiate the encoding techniques (e.g. data element structure, byte
ordering, compression) they are able to support.
•
Unique Identifier (UID) - A Unique Identifier is a globally unique, ISO compliant,
ASCII-numeric string. It guarantees uniqueness across multiple countries, sites, vendors
and equipment.
- 1 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
1.3 Acronyms, Abbreviations and Symbols
•
•
•
•
•
•
•
ACC
ACR
ASCII
AE
American College of Cardiology
American College of Radiology
American Standard Code for Information Interchange
Application Entity
ANSI
AP
American National Standards Institute
Application Profile
CEN TC251 Comite Europeen de Normalisation - Technical Committee 251 - Medical
Informatics
•
•
•
•
•
•
•
•
•
•
•
DICOM
FSC
FSR
FSU
IOD
Digital Imaging and Communications in Medicine
File-set Creator
File-set Reader
File-set Updater
Information Object Definition
International Standards Organization
Japan Industries Association of Radiological Systems
National Electrical Manufacturers Association
Real-World Activity
ISO
JIRA
NEMA
RWA
SOP
UID
Service-Object Pair
Unique Identifier
- 2 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
2 Implementation Model
2.1 Application Data Flow Diagram
Transfer of
X-ray
Examination
DICOM
Recording
CD-R
Medium
Review of
X-ray
Examination
DICOM
Reader
Legend
Application Entity
Local Real-World Activity
File-set Creator (FSC)
File-set Reader (FSR)
Local Relationship
Figure 1
2.2 Functional Definitions of AE's
2.2.1 DICOM Recording
The DICOM Recording supports the following tasks:
•
•
Builds DICOM XA Information Objects.
Creates a DICOMDIR file that represents the contents of DICOM XA Information
Objects to be recorded.
•
•
Records DICOM XA Information Objects and DICOMDIR from the local storage to
the CD-R media.
Duplicates the File-set of the CD-R media or the local storage onto the blank CD-R
media.
- 3 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
2.2.2 DICOM Reader
The DICOM Reader supports the following tasks:
•
•
Reads the DICOMDIR file that represents the contents of the data as recorded.
Displays the ordered list of patient and image identifying information or icon
images.
•
•
Reads the selected image SOP instance from the CD-R media and displays it on
the monitor of the DICOM Reader.
Reads the File-set of the CD-R media and writes it onto the local storage of DICOM
Reader.
2.3 Sequencing of Real World Activities
2.3.1 Features
2.3.1.1 Transfer of X-ray Examination
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
Operator requests to transfer images to the local storage after selecting the transferred
images from the Patient List or from the Image List.
Operator requests to record the images of the local storage onto the CD-R media after
transferring all data that are to be recorded.
When the size of images exceeds the capacity of a single CD-R, the request of
transferring images is canceled.
Each CD-R media has a DICOMDIR file that represents the contents of the data as
recorded.
Operator can request to duplicate the CD-R media from the recorded CD-R media to the
blank CD-R media
2.3.1.2 Review of X-ray Examination
ꢀ
ꢀ
Operator inserts the CD-R media to be read.
Operator requests to display the list of images or the catalog of icon images that are
recorded on the CD-R media.
ꢀ
ꢀ
Operator selects one image from the list or from the thumbnail images and requests to
review the image from the CD-R media.
Operator can request to read images from the CD-R media onto the local storage and then
operator can request to review the image from the local storage.
ꢀ
ꢀ
Operator can request to delete the images on the local storage.
The images on the local storage remain after being recorded onto the CD-R media, so
operators can display these images from the local storage also.
- 4 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
2.3.2 Operation
2.3.2.1 Transfer of X-ray Examination
ꢀ
The operation for manual transfer of X-ray examination is described below:
STEP-1: Select the images or the patient to be transferred.
STEP-2: Request transfer.
STEP-3: Insert the blank CD-R media.
STEP-4: Request recording the transferred images onto the CD-R media.
2.3.2.2 Review of X-ray Examination
ꢀ
The operation for manual review of X-ray examination is described below:
STEP-1: Insert the CD-R media to be review.
STEP-2: Select one image from the list or from the thumbnail images.
STEP-3: Request forward play, backward play, step by step, or other image processings.
2.4 File Meta Information for Implementation Class and Version
•
•
•
File Meta Information Version
Implementation Class UID
Implementation Version Name
0x0001
1.2.392.200036.9116.31
TM_XA_DCM_V12
- 5 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
3 AE Specifications
3.1 DICOM Recording Specification
The DICOM Recording AE provides Standard Conformance to the DICOM Media Storage
Service, File Format Class and the Media Storage Application Profile.
Table 1
Application Profile
Identifier
Real World Activity
Role
SC Option
Basic Cardiac X-ray
Angiographic Studies on
CD-R Media
Transfer of X-ray
Examination
STD-XABC-CD
FSC
Interchange
1024 X-ray Angiographic
Studies on CD-R Media
Transfer of X-ray
Examination
STD-XA1K-CD
FSC
Interchange
3.1.1 File Meta Information for the Application Entity
The Application Entity Title is registered into the DICOM File Meta Information header.
Source Application Entity Title TM_XA_DCM_V12
•
3.1.2 Real-World Activity
3.1.2.1 Real-World Activity - Transfer of X-ray Examination
DICOM Recording AE acts as a FSC when chosen by the operator. This AE writes the SOP
instances onto the CD-R media, creates a DICOMDIR and writes it onto the CD-R media.
3.1.3 Media Storage Application Profile
DICOM Recording AE uses the Application Profile defined in Table 1.
3.1.4 Information Object Definition and DICOMDIR keys
Information Object Definition and DICOMDIR keys are described in Chapter 8.
3.2 DICOM Reader Specification
The DICOM Reader AE provides Conformance to the DICOM Media Storage Service,
File Format Class and the Media Storage Application Profile.
Table 2
Application Profile
Identifier
Real World Activity
Role
SC Option
Basic Cardiac X-ray
Angiographic Studies on
CD-R Media
Review of X-ray
Examination
STD-XABC-CD
FSR
Interchange
Extended Cardiac X-ray
Angiographic Studies on
CD-R Media
Review of X-ray
Examination
AUG-XABC-CD
FSR
Interchange
- 6 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
3.2.1 File Meta Information for the Application Entity
The Application Entity Title is not relevant because this DICOM Reader AE only reads CD-R
media.
3.2.2 Real-World Activity
3.2.2.1 Real-World Activity - Review of X-ray Examination
DICOM Reader AE acts as a FSR when chosen by the operator. This AE reads the DICOMDIR
from the CD-R media and reads the SOP instances from the CD-R media.
3.2.3 Media Storage Application Profile
DICOM Reader AE uses the Application Profile defined in Table 2.
3.2.4 Information Object Definition and DICOMDIR keys
Information Object Definition and DICOMDIR keys are described in Chapter 8.
4 Augmented and Private Application Profiles
4.1 Augmented Application Profiles
DICOM Reader AE supports Augmented Application Profile: AUG-XABC-CD.
This is a extension of the standard Application Profile: STD-XABC-CD.
4.1.1 AUG-XABC-CD
4.1.1.1 SOP Class Augmentations
None
4.1.1.2 Directory Augmentations
None
4.1.1.3 Other Augmentations
The following attributes used within the X-ray Angiographic Image files have the specified
values.
Table 3
Attribute Name
Modality
Tag
Value
(0008, 0060)
(0028, 0010)
(0028, 0011)
(0028, 0100)
(0028, 0101)
XA
1024
1024
8
Rows
Columns
Bits Allocated
Bits Stored
8
4.2 Private Application Profiles
Not applicable to this product.
- 7 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
5 Extensions, Specializations, and Privatizations of SOP
Classes and Transfer Syntaxes
Not applicable to this product.
6 Configuration
For the Digital Fluorography System, the configuration can be set.
Note: Settings and changes are performed by Toshiba Service Personnel at the time of
installation of the Digital Fluorography System.
7 Support of Extended Character Sets
This product supports the following character sets:
Table 4
Basic G0 Set
•
ISO-IR 6 (default)
- 8 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8 X-Ray Angiographic Information Object Definition
8.1 Entity Module Definitions
The information modules for the X-Ray Angiographic devices are defined below.
8.1.1 Media Storage Directory IOD Modules
Table 5
Object
Module / key
Reference
Usage1
DICOM File Meta
Information
M
Basic Directory
Information Object
Definition
File-set Identification Module
Directory Information Module
M
U
Definition of Specific PATIENT keys
M
M
M
M
Directory Records
STUDY keys
SERIES keys
IMAGE keys
1 M=Mandatory, C=Conditional, U=User option
8.1.2 Media Storage XA IOD Modules
Table 6
Object
Module / key
Reference
Usage1
DICOM File Meta
Information
M
XA Information
XA IOD Modules
M
Object Definition
1 M=Mandatory, C=Conditional, U=User option
- 9 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.1.3 XA IOD Modules
Table 7
Information Entity
Module
Reference
Usage1
Patient
Study
Patient Module
M
General Study Module
Patient Study Module
General Series Module
General Equipment Module
General Image Module
Image Pixel Module
Contrast/bolus Module
Cine Module
M
Study
U
Series
Equipment
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
Image
M
M
M
M
C
C
Multi-frame Module
Frame Pointers Module
Mask Module
C
Not Used
Not Used
Not Used
Not Used
Not Used
U
C
Display Shutter Module
Device Module
U
U
Therapy Module
U
X-ray Image Module
X-ray Acquisition Module
X-ray Collimator Module
X-ray Table Module
XA Positioner Module
Overlay Plane Module
Multi-Frame Overlay Module
Curve Module
M
8.5.12
M
Not Used
Not Used
U
C
M
Not Used
Not Used
8.5.14
U
C
C (Note)
C/U
U
Modality LUT Module
VOI LUT Module
Not Used
SOP Common Module
M
1 M=Mandatory, C=Conditional, U=User option
Note : Always set when ECG data is available
- 10 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.2 DICOM File Meta Information
8.2.1 DICOM File Meta Information of Directory IOD
Table 8
Attribute Name
File Preamble
Tag
Type
Attribute Description
No Tag
1
1
1
1
1
1
1
1
Always set
DICOM Prefix
No Tag
Always set ("DICM")
Always set
Group Length
(0002, 0000)
(0002, 0001)
(0002, 0002)
(0002, 0003)
(0002, 0010)
(0002, 0012)
File Meta Information Version
Media Storage SOP Class UID
Media Storage SOP Instance UID
Transfer Syntax UID
Always set (0x0001)
Always set ("1.2.840.10008.1.3.10")
Always set
Always set ("1.2.840.10008.1.2.1")
Always set
Implementation Class UID
8.2.2 DICOM File Meta Information of XA IOD
Table 9
Attribute Name
File Preamble
Tag
Type
Attribute Description
Always set
No Tag
1
1
1
1
1
DICOM Prefix
No Tag
Always set ("DICM")
Always set
Group Length
(0002, 0000)
(0002, 0001)
(0002, 0002)
File Meta Information Version
Media Storage SOP Class UID
Always set (0x0001)
Always set
("1.2.840.10008.5.1.4.1.1.12.1")
Media Storage SOP Instance UID
Transfer Syntax UID
(0002, 0003)
(0002, 0010)
(0002, 0012)
(0002, 0013)
1
1
1
3
Always set
Always set ("1.2.840.10008.1.2.4.70")
Always set
Implementation Class UID
Implementation Version Name
Always set
8.3 Basic Directory Information Object Definitions
8.3.1 File-set Identification Module
Table 10
Attribute Name
Tag
Type
Attribute Description
Always set
File-set ID
(0004, 1130)
2
- 11 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.3.2 Directory Information Module
Table 11
Attribute Name
Tag
Type
Attribute Description
Offset of the First Directory Record of
the Root Directory Entity
(0004, 1200)
1
Always set
Offset of the Last Directory Record of
the Root Directory Entity
(0004, 1202)
1
Always set
File-set Consistency Flag
Directory Record Sequence
>Offset of the Next Directory Record
>Record In-use Flag
(0004, 1212)
(0004, 1220)
(0004, 1400)
(0004, 1410)
(0004, 1420)
1
Always set
2
Always set
1C
1C
1C
Always set when condition is met
Always set when condition is met
Always set when condition is met
>Offset of Referenced Lower-Level
Directory Entity
>Directory Record Type
(0004, 1430)
(0004, 1500)
(0004, 1510)
(0004, 1511)
(0004, 1512)
1C
1C
1C
1C
1C
Always set when condition is met
Always set when condition is met
Always set when condition is met
Always set when condition is met
Always set when condition is met
>Referenced File ID
>Referenced SOP Class UID in File
>Referenced SOP Instance UID in File
>Referenced Transfer Syntax UID in
File
8.4 Definition of Specific Directory Records
8.4.1 PATIENT keys
Table 12
Attribute Name
Patient's Name
Tag
Type
Attribute Description
Always set except for urgent patient
Always set
(0010, 0010)
(0010, 0020)
(0010, 0030)
(0010, 0040)
2
1
2
2
Patient ID
Patient's Birth Date
Patient's Sex
Length=0 when no entry is made
Always set
- 12 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.4.2 STUDY keys
Table 13
Attribute Name
Tag
Type
Attribute Description
Study Date
(0008, 0020)
(0008, 0030)
(0008, 1030)
(0020, 000D)
(0020, 0010)
(0008, 0050)
1
1
Always set
Study Time
Always set
Study Description
Study Instance UID
Study ID
2
Length=0 when no entry is made
Always set
1C
1
Always set
Accession Number
2
Length=0 when no entry is made
8.4.3 SERIES keys
Table 14
Attribute Name
Modality
Tag
Type
Attribute Description
Always set ("XA")
Always set
(0008, 0060)
(0020, 000E)
(0020, 0011)
(0008, 0080)
(0008, 0081)
(0008, 1050)
1
1
1
2
2
2
Series Instance UID
Series Number
Always set
Institution Name
Always set
Institution Address
Performing Physician's Name
Always set
Length=0 when no entry is made
- 13 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.4.4 IMAGE keys
Table 15
Attribute Name
Instance Number
Image Type
Tag
Type
1
Attribute Description
(0020, 0013)
(0008, 0008)
(0008, 0023)
(0008, 0033)
(0018, 1510)
(0018, 1511)
(0020, 4000)
(0028, 0008)
(0050, 0004)
(0088, 0200)
(0028, 0002)
(0028, 0004)
(0028, 0010)
(0028, 0011)
(0028, 0100)
(0028, 0101)
(0028, 0102)
(0028, 0103)
(7FE0, 0010)
Always set
1
Always set
Image Date
3
Always set
Image Time
3
Always set
Positioner Primary Angle
Positioner Secondary Angle
Image Comments
Number of Frames
Calibration Image
Icon Image Sequence
>Samples per Pixel
>Photometric Interpretation
>Rows
3
Always set [degree]
Always set [degree]
Length=0 when no entry is made
Always set
3
3
3
2
Always set
1
Always set
1
Always set (0x0001)
Always set ("MONOCHROME2")
Always set (0x0100)
Always set (0x0100)
Always set (0x0008)
Always set (0x0008)
Always set (0x0007)
Always set (0x0000)
Always set
1
1
>Columns
1
>Bits Allocated
1
>Bits Stored
1
>High Bit
1
>Pixel Representation
>Pixel Data
1
1
- 14 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.5 XA Information Object Definisions
8.5.1 Patient Module
Table 16
Attribute Name
Patient's Name
Tag
Type
Attribute Description
Always set except for urgent patient
Always set
(0010, 0010)
(0010, 0020)
(0010, 0030)
(0010, 0040)
(0010, 4000)
2
2
2
2
3
Patient ID
Patient's Birth Date
Patient's Sex
Length=0 when no entry is made
Always set
Patient Comments
Length=0 when no entry is made
8.5.2 General Study Module
Table 17
Attribute Name
Tag
Type
Attribute Description
Always set
Study Instance UID
Study Date
(0020, 000D)
(0008, 0020)
(0008, 0030)
(0008, 0090)
(0020, 0010)
(0008, 0050)
(0008, 1030)
1
2
2
2
2
2
3
Always set
Study Time
Always set
Referring Physician's Name
Study ID
Length=0 when no entry is made
Always set
Accession Number
Study Description
Length=0 when no entry is made
Length=0 when no entry is made
8.5.3 Patient Study Module
Table 18
Attribute Name
Patient's Size
Tag
Type
Attribute Description
Length=0 when no entry is made
Length=0 when no entry is made
(0010, 1020)
(0010, 1030)
3
3
Patient's Weight
- 15 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.5.4 General Series Module
Table 19
Attribute Name
Tag
Type
Attribute Description
Modality
(0008, 0060)
(0020, 000E)
(0020, 0011)
(0008, 0021)
(0008, 0031)
(0008, 1050)
(0008, 103E)
1
1
2
3
3
3
3
Always set ("XA")
Series Instance UID
Series Number
Series Date
Always set
Always set
Always set
Series Time
Always set
Performing Physician's Name
Series Description
Length=0 when no entry is made
Always set
8.5.5 General Equipment Module
Table 20
Attribute Name
Manufacturer
Tag
Type
Attribute Description
Always set
(0008, 0070)
(0008, 0080)
(0008, 0081)
(0008, 1010)
(0008, 1090)
(0018, 1000)
(0018, 1020)
2
3
3
3
3
3
3
Institution Name
Always set
Institution Address
Station Name
Always set
Always set
Manufacturer's Model Name
Device Serial Number
Software Versions
Always set
Always set
Always set
8.5.6 General Image Module
Table 21
Attribute Name
Tag
Type
2
Attribute Description
Always set
Image Number
Patient Orientation
Image Date
(0020, 0013)
(0020, 0020)
(0008, 0023)
(0008, 0033)
(0008, 0008)
(0008, 0022)
(0008, 0032)
(0020, 4000)
2C
2C
2C
3
Always Length=0
Always set
Image Time
Always set
Image Type
Always set
Acquisition Date
Acquisition Time
Image Comments
3
Always set
3
Always set
3
Length=0 when no entry is made
- 16 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.5.7 Image Pixel Module
Table 22
Attribute Name
Samples per Pixel
Photometric Interpretation
Rows
Tag
Type
Attribute Description
(0028, 0002)
(0028, 0004)
(0028, 0010)
(0028, 0011)
(0028, 0100)
(0028, 0101)
(0028, 0102)
(0028, 0103)
(7FE0, 0010)
1
1
1
1
1
1
1
1
1
Always set (0x0001)
Always set ("MONOCHROME2")
Always set (0x0400 or 0x0200)
Always set (0x0400 or 0x0200)
Always set (0x0008)
Always set (0x0008)
Always set (0x0007)
Always set
Columns
Bits Allocated
Bits Stored
High Bit
Pixel Representation
Pixel Data
Always set
8.5.8 Contrast/Bolus Module
Table 23
Attribute Name
Tag
Type
Attribute Description
Length=0 when no entry is made
Length=0 when no entry is made
Length=0 when no entry is made
Length=0 when no entry is made
Contrast/Bolus Agent
Contrast/Bolus Route
Contrast/Bolus Volume
Contrast Flow Rate(s)
(0018, 0010)
(0018, 1040)
(0018, 1041)
(0018, 1046)
2
3
3
3
8.5.9 Cine Module
Table 24
Attribute Name
Tag
Type
Attribute Description
Frame Time
(0018, 1063)
1C
Always set
8.5.10 Multi-frame Module
Table 25
Attribute Name
Number of Frames
Tag
Type
Attribute Description
Always set
(0028, 0008)
(0028, 0009)
1
1
Frame Increment Pointer
Always set (0x00181063)
- 17 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.5.11 X-ray Image Module
Table 26
Attribute Name
Frame Increment Pointer
Image Type
Tag
Type
1C
1
Attribute Description
(0028, 0009)
(0008, 0008)
(0028, 1040)
(0028, 0002)
(0028, 0004)
(0028, 0100)
(0028, 0101)
(0028, 0102)
(0028, 0103)
(0008, 1140)
Always set (0x00181063)
Always set
Pixel Intensity Relationship
Samples per Pixel
Photometric Interpretation
Bits Allocated
1
Always set ("DISP" or "LIN")
Always set (0x0001)
Always set ("MONOCHROME2")
Always set (0x0008)
Always set (0x0008)
Always set (0x0007)
Always set
1
1
1
Bits Stored
1
High Bit
1
Pixel Representation
Reference Image Sequence
1
1C
Not set when Image type(0008,0008)
Value 3 is "SINGLE PLANE"
>Reference SOP Class UID
(0008, 1150)
(0008, 1155)
1C
1C
Always set when Reference Image
Sequence is present
>Reference SOP Instance UID
Always set when Reference Image
Sequence is present
8.5.12 X-ray Acquisition Module
Table 27
Attribute Name
Tag
Type
2
Attribute Description
Always set [kV]
KVP
(0018, 0060)
(0018, 1155)
(0018, 1151)
(0018, 1150)
(0018, 1154)
(0018, 1162)
(0018, 1147)
(0018, 1149)
(0018, 1164)
Radiation Setting
1
Always set
X-ray Tube Current
Exposure Time
2C
2C
3
Always set [mA]
Always set [msec]
Always set [msec]
Always set [mm]
Average Pulse Width
Intensifier Size
3
Field of View Shape
Field of View Dimension(s)
Imager Pixel Spacing
3
Always set("ROUND ")
Always set [mm]
3
3
Always set [mm]
- 18 -
Download from Www.Somanuals.com. All Manuals Search And Download.
No. MIIXR0002EAC
8.5.13 XA Positioner Module
Table 28
Attribute Name
Tag
Type
Attribute Description
Distance Source to Detector
Estimated Radiographic Magnification
Factor
(0018, 1110)
(0018,1114)
3
3
Always set [mm]
Always set
Positioner Motion
(0018, 1500)
(0018, 1510)
(0018, 1511)
2C
2
Always set ("STATIC")
Always set [degree]
Always set [degree]
Positioner Primary Angle
Positioner Secondary Angle
2
8.5.14 Curve Module
Table 29
Attribute Name
Curve Dimensions
Number of Points
Tag
Type
1
Attribute Description
(5000, 0005)
(5000, 0010)
(5000, 0020)
(5000, 0030)
(5000, 0103)
(5000, 0110)
(5000, 0112)
(5000, 0114)
(5000, 3000)
Always set
Always set
Always set
Always set
Always set
Always set
Always set
Always set
Always set
1
Type of Data
1
Axis Units
3
Data Value Representation
Curve Data Descriptor
Coordinate Start Value
Coordinate Step Value
Curve Data
1
1C
1C
1C
1
Note : This module is set only if ECG data is available.
8.5.15 VOI LUT Module
Table 30
Attribute Name
Window Center
Window Width
Tag
Type
3
Attribute Description
(0028, 1050)
(0028, 1051)
Always set
Always set
1C
8.5.16 SOP Common Module
Table 31
Attribute Name
Tag
Type
Attribute Description
SOP Class UID
(0008, 0016)
(0008, 0018)
1
1
Always set
Always set
SOP Instance UID
E