Teledyne TV Converter Box T200U NOy User Manual

T200U-NOy ANALYZER  
ADDENDUM  
(for use with T200 Operators Manual, PN 06858  
and with the T200U Addendum, PN 06861)  
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION  
9480 CARROLL PARK DRIVE  
SAN DIEGO, CA 92121-5201  
USA  
Toll-free Phone: 800-324-5190  
Phone: 858-657-9800  
Fax: 858-657-9816  
Copyright 2011-2013  
Teledyne Advanced Pollution Instrumentation  
07303C DCN6646  
06 February 2013  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY MESSAGES  
Important safety messages are provided throughout this manual for the purpose of  
avoiding personal injury or instrument damage. Please read these messages carefully.  
Each safety message is associated with a safety alert symbol, and placed throughout this  
manual and inside the instrument. The symbols with messages are defined as follows:  
WARNING: Electrical Shock Hazard  
HAZARD: Strong oxidizer  
GENERAL WARNING/CAUTION: Read the accompanying message for specific  
information.  
CAUTION: Hot Surface Warning  
Do Not Touch: Touching some parts of the instrument without protection or proper tools  
could result in damage to the part(s) and/or the instrument.  
Technician Symbol: All operations marked with this symbol are to be performed by  
qualified maintenance personnel only.  
Electrical Ground: This symbol inside the instrument marks the central safety grounding  
point for the instrument.  
CAUTION  
This instrument should only be used for the purpose and in the manner described in  
this manual. If you use this instrument in a manner other than that for which it was  
intended, unpredictable behavior could ensue with possible hazardous consequences.  
NEVER use any gas analyzer to sample combustible gas(es)!  
Note  
For Technical Assistance regarding the use and maintenance of this instrument or any other Teledyne  
API product, contact Teledyne API’s Technical Support Department:  
Telephone: 800-324-5190  
07303C DCN6646  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
CONSIGNES DE SÉCURITÉ  
Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans  
le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire  
attentivement ces consignes. Chaque consigne de sécurité est représentée par un  
pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à  
l’intérieur des instruments. Les symboles correspondent aux consignes suivantes :  
AVERTISSEMENT : Risque de choc électrique  
DANGER : Oxydant puissant  
AVERTISSEMENT GÉNÉRAL  
/
MISE EN GARDE : Lire la consigne  
complémentaire pour des renseignements spécifiques  
MISE EN GARDE : Surface chaude  
Ne pas toucher : Toucher à certaines parties de l’instrument sans protection ou  
sans les outils appropriés pourrait entraîner des dommages aux pièces ou à  
l’instrument.  
Pictogramme « technicien » : Toutes les opérations portant ce symbole doivent  
être effectuées uniquement par du personnel de maintenance qualifié.  
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central  
de la mise à la terre sécuritaire de l’instrument.  
MISE EN GARDE  
Cet instrument doit être utilisé aux fins décrites et de la manière décrite dans  
ce manuel. Si vous utilisez cet instrument d’une autre manière que celle pour  
laquelle il a été prévu, l’instrument pourrait se comporter de façon imprévisible  
et entraîner des conséquences dangereuses.  
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz  
combustibles!  
ii  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY  
WARRANTY POLICY (02024F)  
Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne  
Instruments, Inc., provides that:  
Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment  
failure occur, TAPI assures its customers that prompt service and support will be available.  
COVERAGE  
After the warranty period and throughout the equipment lifetime, TAPI stands ready to  
provide on-site or in-plant service at reasonable rates similar to those of other manufacturers  
in the industry. All maintenance and the first level of field troubleshooting are to be  
performed by the customer.  
NON-TAPI MANUFACTURED EQUIPMENT  
Equipment provided but not manufactured by TAPI is warranted and will be repaired to the  
extent and according to the current terms and conditions of the respective equipment  
manufacturer’s warranty.  
PRODUCT RETURN  
All units or components returned to Teledyne API should be properly packed for  
handling and returned freight prepaid to the nearest designated Service Center. After the  
repair, the equipment will be returned, freight prepaid.  
The complete Terms and Conditions of Sale can be reviewed at http://www.teledyne-  
api.com/terms_and_conditions.asp  
CAUTION – Avoid Warranty Invalidation  
Failure to comply with proper anti-Electro-Static Discharge (ESD) handling and packing  
instructions and Return Merchandise Authorization (RMA) procedures when returning parts  
for repair or calibration may void your warranty. For anti-ESD handling and packing  
instructions please refer to “Packing Components for Return to Teledyne API’s Customer  
Service” in the Primer on Electro-Static Discharge section of this manual, and for RMA  
procedures please refer to our Website at http://www.teledyne-api.com under Customer  
Support > Return Authorization.  
07303C DCN6646  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200EU-NOy Option Manual Addendum  
This page intentionally left blank.  
iv  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200EU-NOy Option Manual Addendum  
TABLE OF CONTENTS  
1.0 INTRODUCTION......................................................................................................7  
2.0 SPECIFICATIONS ...................................................................................................9  
2.1  
Specifications ........................................................................................................................9  
3.0 GETTING STARTED..............................................................................................11  
3.1  
3.2  
3.3  
Electrical and Pneumatic Connections.............................................................................12  
Initial Operation ...................................................................................................................15  
4.0 THE T200U-NOY EXTERNAL CONVERTER........................................................17  
4.1  
5.0 CALIBRATION AND ZERO/SPAN (OPTION) CHECKS .......................................19  
5.1  
Calibration or Cal Check Procedure..................................................................................20  
6.0 MAINTENANCE.....................................................................................................23  
6.1  
6.2  
6.3  
6.4  
6.6  
Replacing the Sample Particulate Filters..........................................................................24  
7.0  
8.0  
TROUBLESHOOTING, ADJUSTMENTS...........................................................31  
7.1  
Operation Verification – Diagnostic Techniques .............................................................31  
7.1.1 Pneumatic System .............................................................................................................31  
7.1.2 Leak Check ........................................................................................................................31  
7.1.3 501Y Bypass Pump Diagnostic Procedures......................................................................32  
7.1.4 Electrical Fault Isolation.....................................................................................................32  
T200U-NOY SPARE PARTS LIST.....................................................................37  
07303C DCN6646  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200EU-NOy Option Manual Addendum  
List of Figures  
Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom)......................13  
Figure 2. Rear Panel Pneumatic Connections............................................................................................14  
Figure 3. Replacing the Particulate Filter...................................................................................................25  
Figure 4. NOy Converter Guts Assembly Location....................................................................................27  
Figure 5. Pneumatics ..................................................................................................................................28  
Figure 7. Pneumatics ...................................................................................Error! Bookmark not defined.  
Figure 8 Electrical Diagram........................................................................................................................35  
List of Tables  
Table 1 Specifications for the External Converter and the 501Y Bypass Pump..........................................9  
Table 2 Final Test and Calibration Values.................................................................................................15  
Table 3 Manual Zero Calibration Procedure..............................................................................................20  
Table 4 Enter Expected Span Gas Concentrations Procedure For NO & NOy.........................................20  
Table 5 Span Calibration Procedure..........................................................................................................21  
Table 6 Preventative Maintenance Calendar.............................................................................................23  
Table 7 Temperature Controller Programming Guide ...............................................................................33  
vi  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.0 INTRODUCTION  
The T200U-NOy analyzer consists of an external Converter designed to support a  
NOx analyzer by converting multiple, unstable compounds grouped under the  
name NOy. The converter is mounted externally at the sample inlet to minimize  
flow time between sample in and converter, thereby optimizing measurement  
accuracy.  
This manual addendum is to be used in conjunction with two additional  
documents with operation instructions:  
T200 operation manual (part number 06858)  
T200U addendum (part number 06861).  
It is recommended that you read/familiarize yourself with all sets of instructions  
in order to use the analyzer and converter correctly.  
07303C DCN6646  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200EU-NOy Option Manual Addendum 1.0 Introduction  
This page intentionally left blank.  
8
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.0 SPECIFICATIONS  
2.1 Specifications  
Table 1 Specifications for the External Converter and the 501Y Bypass Pump  
Converter  
Remotely mounted molybdenum external Converter with  
temperature controller in a NEMA – 4 enclosure.  
315oC 7oC with read-out on front panel of bypass pump chassis.  
Converter Temperature  
Tube/Electrical Cable  
Assembly  
Up to 50 feet (15 m) maximum length  
Power, 501Y Bypass Pump  
Assembly  
120V~ 60 Hz, 230V~ 50 Hz 360 watts  
Power, Converter Assembly  
Converter Efficiency  
120V~ 50/60 Hz, 60 watts  
>96% for NO2  
Weight, 501Y Bypass Pump  
Assembly  
35 lbs (16 kg)  
Weight, Remote Converter  
30 lbs (11 kg)  
501Y Bypass Pump Chassis  
Dimensions HxWxD  
9" x17" x23.6" (23cm x 43cm x 61cm)  
Remote Converter Assy  
Dimensions HxWxD  
12"x19"x7" (31cm x 48cm x 18cm)  
Environmental  
Installation Category (Over-voltage Category) II  
Pollution Degree 2  
Bypass Flow rate  
Cal gas Flow rate  
800 cc/min 10%, each channel  
4000 cc/min (approx.)  
07303C DCN6646  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model T200EU-NOy Option Manual Addendum 2.0 Specifications  
This page intentionally left blank.  
10  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.0 GETTING STARTED  
The NOy converter has been designed to operate with the T200U low level NOx  
analyzer, with some additional modification.The combination is referred to as the  
T200U-NOy analyzer.  
3.1 Unpacking and Inspection  
CAUTION!  
Avoid personal injury: use two persons each to lift and carry the analyzer and the  
Model 501 Pump Pack chassis.  
The T200U- NOy comes in three boxes:  
The analyzer with AC power cord  
501Y Bypass Pump Chassis (typically referred to as “501Y herein) with AC  
power cord  
External Converter housed in a stainless steel shelter, and an umbilical cable  
assembly (herein referred to as “Converter” or “external Converter”)  
External Pump (herein referred to as “external Pump” not to be confused with  
the 501Y Bypass Pump)  
1. Verify that there is no apparent shipping damage. If damage has  
occurred please advise shipper first, then TAPI. (Keep original  
container and packaging for shipper’s inspection).  
2. Before operating the analyzer, remove the shipping screws as shown in  
the T200 Manual.  
3. Please check the voltage and frequency label on the rear panel of the  
instrument for compatibility with the local power before connecting  
the instruments to a power source.  
07303C DCN6646  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model T200U-NOy Option Manual Addendum  
3.2 Electrical and Pneumatic Connections  
NOTE  
To maintain compliance with EMC standards, it is required that the cable length (at the  
rear of the analyzer) be no greater than 3 meters for all I/O connections, which include  
Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232, and RS-485.  
1. Mount the Converter (Figure 1, bottom) on a suitable mast.  
2. Route the umbilical cable assembly into the shelter of the Converter.  
3. Locate the analyzer and 501Y Bypass Pump Chassis in close proximity,  
preferably mounting one over the other in a 19” rack.  
4. Connect the T200U-NOy analyzer rear panel electrical connectors per the  
T200 operation manual.  
5. Connect the pneumatic cable fittings from the analyzer rear panel (Figure 1,  
top) to the 501Y rear panel (Figure 1, center) and from the 501Y rear panel  
to the Converter (Figure 1, bottom).  
Refer to Figure 2 for an illustration of the pneumatic connections; refer to the  
tags on each tube to match up the correct tube with the appropriate rear panel  
fitting.  
6. Connect the pneumatic cable fitting from the calibrator to the 501Y rear panel  
(Figure 2).  
7. Also connect the EXHAUST port of the analyzer rear panel to the external  
Pump.  
8. Connect the 7-pin power and signal cable from the external Converter’s  
umbilical cable assembly to the CONVERTER POWER connector on the  
rear of the 501Y.  
9. Connect the analyzer and 501Y AC power cords from their AC power  
receptacles to the correct line voltage (refer to power specs on the  
respective rear panel label).  
WARNING!  
Lethal voltages are present inside the chassis.  
Do not operate with cover off during normal operation  
Before operation, check for correct input voltage and frequency.  
Do not operate without proper chassis grounding  
Do not defeat the ground wire on power plug  
Turn off power before disconnecting electrical subassemblies  
12  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom)  
07303C DCN6646  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
Figure 2. Rear Panel Pneumatic Connections  
Note: Pneumatic tubes and rear panel labels include numbers to facilitate matching connections  
between instruments.  
CONNECTIONS: Analyzer to Pump to Converter and Calibrator to Pump  
Calibrator  
Analyzer  
NO IN (1)  
NOY IN (2)  
Pump Chassis  
NO OUT (1)  
NOY OUT (2)  
CAL IN (3)  
Converter  
[cal gas out port]  
NOY IN (4)  
NO IN (5)  
CAL OUT (6)  
NO OUT (4)  
NOY OUT (5)  
CAL IN (6)  
14  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
3.3 Initial Operation  
1. After confirming proper supply voltage, turn on the power to the  
analyzer and the 501Y bypass pump, and plug in the external Pump. If  
you are unfamiliar with the T200U-NOy analyzer, we recommend that  
you read the overview in Section 3 of its operation manual before  
proceeding.  
The front panel displays should immediately light as the  
instruments start up.  
The 501Y requires about 30 minutes for the external Converter to  
come up to operating temperature.  
The T200U-NOy requires about 30 minutes for the ozone  
generator to start up. During that time the instrument will not  
respond to span gas.  
After 30 min, the display on the 501Y front panel should read  
315oC, indicating that the external Converter is at operating  
temperature.  
Follow the instructions in the T200 Manual and the T200U  
Addendum to confirm proper operation of the analyzer.  
2. Proceed to Section 4 of this manual to perform a Zero/Span check.  
Table 2 Final Test and Calibration Values  
Test Values  
Observed  
Value  
Units  
Nominal  
Range  
CONVERTER TEMP  
oC  
CC/MIN  
315 7  
800 80  
BYPASS  
FLOW  
NO  
CHANNEL  
BYPASS  
FLOW  
NOy  
CC/MIN 800 ± 80  
CHANNEL  
Converter Serial # __________________________  
Date  
_______________________  
Technician _____________________  
07303C DCN6646  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model T200U-NOy Option Manual Addendum  
This page intentionally left blank.  
16  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.0 THE T200U-NOY EXTERNAL CONVERTER  
4.1 Principle of Operation  
The T-API Model T200U-NOy is designed to measure the concentration of NO,  
NO2, and other compounds that are too unstable to be measured when taken in  
through the normal ambient air sample inlet system. Please refer to the T200U  
manual supplied with this system for a general discussion of the operation of a  
NOx analyzer. The suite of compounds known collectively as NOy is composed  
of roughly 30 compounds. The NOy measurement is generally done in  
conjunction with a standard NOx measurement, with the difference between the  
two being the concentration of the unstable compounds, sometimes referred to as  
NOz.  
The analyzer measures two independent gas streams. One bypasses the external  
Converter and measures NO in the sample. The other pulls the sample through  
the external Converter and is displayed as NOy. Under normal calibration  
conditions the NOy measurement is equivalent to the delivered NOx calibration  
gases.  
The system is composed of four modules:  
The T200U-NOy analyzer, without a Moly converter  
A 501Y Bypass Pump containing:  
bypass pump  
flow control  
sample filtration  
moly temperature control  
pneumatic provisions for calibration  
Externally mounted molybdenum converter (external Converter)  
External Pump  
The T200U-NOy system allows the point of sampling to be located in close  
proximity to the Converter. This configuration provides minimal time delay  
between the sample inlet and the remotely mounted (~10 meters above ground)  
external Converter. Minimizing the transit time between the sample inlet and  
Converter enables the conversion of labile components of NOy. The equation for  
the conversion is:  
315oC  
3 NOY + Mo ---> 3 NO + MoO3  
07303C DCN6646  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model T200U-NOy Option Manual Addendum  
This page intentionally left blank.  
18  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.0 CALIBRATION AND CAL CHECKS  
Unlike most NOx analyzers, the T200U-NOy does not have a sample inlet port  
on its rear panel. The sample inlet is located on the external Converter.  
Follow the steps in the T200U manual to:  
Enter the expected NO and NOy span gas concentrations in the  
analyzer.  
Do the zero calibration procedure  
Do the span calibration procedure  
The analyzer always routes the zero/span gas through the external Converter. In  
standard configuration, the calibration gas needs to be input at the CAL IN (3)  
port of the 501Y bypass pump. In all cases, the calibration gas delivered must be  
under a small amount of pressure (2-5 psig) to overcome the resistance of the  
tubing.  
NOTE  
There must be no venting of the Zero or Span gas to atmosphere prior to the analyzer. The  
gases actually vent at the external Converter.  
Calibration gas must be provided for both the 501Y bypass system and the  
analyzer. Flow requirements are 0.80 LPM for NO bypass, 0.80 LPM for NOy  
bypass and 1 LPM for the analyzer. Calibration gas flow must be in sufficient  
excess (~4 LPM) to prevent any ambient air from entering the sample inlet of the  
external Converter.  
The maximum 50 ft length of tubing between the 501Y and the remote Converter  
has been proven to generate too much backpressure for some brands of  
calibrators. Teledyne calibrators are designed to overcome this potential issue.  
We strongly recommend that span calibration be done with NO span gas. Span  
checks can be done with one of: NO only, NO2 only, or a mixture of NO and NO2  
(GPT).  
Zero air used for all calibration procedures, including GPT, should have <1 ppt  
NO and NO2, less than 1 ppt of major interferents such as SO2, NH3,  
hydrocarbons and a dew point of -5oC or less. The calibration gasses should be  
from a reliable supplier, since the quality of the tank concentration values  
ultimately determines the accuracy of the analyzer.  
NOTE  
The T200U-NOy does NOT have equivalency approval.  
07303C DCN6646  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
5.1 Calibration or Cal Check Procedure  
Calibration of the analyzer can be checked or adjusted using gas introduced  
through the CAL IN (3) port on the 501Y rear panel. The calibration gas is  
routed to a tee fitting (see CAL IN (6) in Figure 4), near the sample inlet on the  
external Converter assembly.  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter  
the expected zero value. Table 3 details the zero calibration procedure.  
Table 3 Manual Zero Calibration Procedure  
Step  
Action  
Comment  
Number  
1.  
2.  
Press CAL  
The analyzer enters the calibrate mode.  
Wait for reading to stabilize at zero value.  
Wait For Stability  
Type STB < 0.1 PPB  
Press ZERO  
3.  
If you change your mind after pressing ZERO, you can still  
press EXIT here without zeroing the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations.  
The analyzer returns to sampling. Immediately after  
calibration, data is not added to the DAS averages.  
Table 4 Enter Expected Span Gas Concentrations Procedure For NO & NOy  
Step  
Action  
Comment  
1.  
Press  
CAL>CONC>NOy  
This menu sequence causes the analyzer to prompt for the  
expected NOy concentration.  
Enter the NOy span concentration value by pressing the  
applicable touchscreen buttons to set the expected value.  
2.  
3.  
Press ENTR  
ENTR stores the expected NOy span value.  
Press  
Now enter the expected NO span concentration as in step one.  
CAL>CONC>NO  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR stores the NO span value and returns the  
prompt to the CONC menu.  
Returns instrument to SAMPLE mode.  
If desired, compensation for converter efficiency (CE) can be included in the  
NOy concentration calculation. Prior to calibration, the converter efficiency  
should either be set to 1.0 or determined per the procedure in the T200 manual.  
20  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model T200U-NOy Option Manual Addendum  
Table 5 Span Calibration Procedure  
Step  
1.  
Action  
Press CAL  
Comment  
The analyzer enters the calibrate mode.  
Wait for reading to stabilize at span value.  
2.  
Wait For Stability  
Type STB < 2 PPB  
Press SPAN  
3.  
4.  
If you change your mind after pressing SPAN, you can still  
press EXIT here without spanning the instrument.  
Press ENTR  
Pressing ENTR actually changes the calculation equations and  
causes the instrument to read the NO and NOy span  
concentrations.  
5.  
Press EXIT  
The analyzer returns to sampling. Immediately after  
calibration, data is not added to the DAS averages.  
07303C DCN6646  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
This page intentionally left blank.  
22  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.0 MAINTENANCE  
6.1 Maintenance Schedule  
The schedule for preventative maintenance is presented in Table 6 below.  
Qualified Maintenance Personnel Required  
Table 6 Preventative Maintenance Calendar  
Item  
Maintenance Interval  
Zero/Span Calibration  
Zero/Span Checks  
Particulate Filter  
Ozone Flow  
Annually or after maintenance or repairs  
Daily  
Weekly as needed  
Check every year and replace o-ring and orifice in reaction  
cell  
Ozone Dryer Filter-DFU  
Reaction Cell Window  
Sample Flow  
Replace every year  
Clean annually or as necessary (refer to T200 Manual  
Check every 2 months  
Converter  
Check efficiency every 2 months  
Check every 3 months  
Pneumatic Lines  
Factory Calibration  
Leak Check  
Calibrate each year or after repairs  
Check every 6 months  
Replace bypass flow orifices Every year / as needed  
Rebuild 501Ybypass pump Every year (check part number on pump label)  
07303C DCN6646  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model T200U-NOy Option Manual Addendum  
6.2 Replacing the Sample Particulate Filters  
The particulate filter should be inspected often for signs of plugging or  
contamination. It is also common for dirt particles to cause instrument drift, and  
affect accuracy.  
To check and change the filter:  
1. Fold down the 501Y front panel.  
2. Locate both filters on the left and right side of the 501Y front panel. See  
3. Visually inspect the filter through the glass window.  
4. If the filter appears dirty, unscrew the hold-down ring, remove the Teflon o-  
ring and then the filter.  
5. Replace the filter, being careful that the element is fully seated in the bottom  
of the holder. Replace the Teflon o-ring, then screw on the hold-down ring  
and hand tighten.  
NOTE:  
Handle the wetted surfaces as little as possible and clean with cloth prior to assembly  
24  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
Figure 3. Replacing the Particulate Filter  
07303C DCN6646  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
6.3 Checking Analyzer Flow Rate  
The external Pump is capable of maintaining the reaction cell pressure at less than  
5.0  
In-Hg-A. If a higher pressure is noted, the Pump may need servicing. Check the  
Pump and pneumatic system for leaks or rebuild the Pump.  
CAUTION!  
Never operate the analyzer without the ozone destruct component properly  
seated and connected within the pneumatic path. The ozone destruct is  
integrated into the converter case, at the exhaust of the reaction cell, inside  
the instrument.  
The sample flow as measured at ports NO IN (1) and NOy IN (2) at the rear of  
the analyzer should be 1000 100 cc/min.  
See Figure 1 for component locations.  
6.4 Replacing the Converter  
The heater, thermocouple, and converter guts assembly is designed to be replaced  
as a single unit. Check Figure 4 for the assembly location.  
CAUTION!  
The converter operates at 315°C. Severe burns can result if not enough time is  
allowed for the assembly to cool. Do not handle assembly until it is at room  
temperature.  
1. Turn off the power to the 501Y at its front panel ON/OFF switch.  
2. Allow the external Converter to cool.  
3. Disconnect the gas fittings and power cable from the internal box containing  
the heater, thermocouple and converter guts assembly (Figure 4).  
4. Remove the entire box from the Converter shelter.  
a. Remove the pneumatic fittings.  
b. Remove the guts assembly box from the chassis by loosening the 4  
captive screws that secures the assembly to the chassis.  
5. Remove bottom bracket and re-attach it to the replacement guts assembly.  
6. Install the box with guts assembly back into the analyzer.  
7. Re-attach the electrical and pneumatic fittings. Leak check the assembly  
when completed.  
26  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model T200U-NOy Option Manual Addendum  
8. Turn the 501Y power back on. The insulation can emit a burnt odor for the  
first 24 hours; this is normal. Allow the Converter to burn-in for 24 hours, then  
re-calibrate the instrument.  
Figure 4. NOy Converter Guts Assembly Location  
6.5  
Inspecting Pneumatic Lines  
Particulate matter in the pneumatic lines will affect both flow rate and response  
time. It is important that the pneumatic system be periodically inspected and  
thoroughly cleaned if necessary. Clean by disassembling and passing methanol  
through three times. Dry with clean zero air.  
Also inspect all pneumatic lines for cracks and abrasion on a regular basis.  
Replace as necessary.  
07303C DCN6646  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
Figure 5. Pneumatics  
28  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
6.6 Checking for Leaks  
If a leak checker is not available, it is possible to leak check the instrument using  
the sample pump plus a shut-off valve.  
1. Turn off power to the analyzer and the 501Y, and unplug the external Pump.  
2. Cap the sample inlet of the external Converter, the BYPASS OUT port of the  
501Y and the ozone generator air inlet (located inside the analyzer, Figure  
Figure 6. Location of O3 Generator Air Inlet  
3. Insert a shut-off valve between the external Pump and the EXHAUST port at  
the rear of the analyzer.  
4. Turn on the analyzer power and plug in the external Pump. Set the TEST  
function to RCEL, which measures the reaction cell pressure. Close the  
shutoff valve and monitor the cell pressure. The pressure should not  
increase more than 1"Hg (.5psi) in 5 minutes. If there is a leak, you’ll need to  
locate it by using a pressure leak checker described below.  
07303C DCN6646  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model T200U-NOy Option Manual Addendum  
If you have a leak checker:  
1. Turn off power to the analyzer and 501Y, and unplug the external Pump.  
2. Disconnect the external Pump from the analyzer’s rear panel. Cap the  
sample inlet of the external Converter, the BYPASS OUT port of the 501Y,  
the ozone generator air inlet (located inside the analyzer, Figure 6), and  
connect the leak checker to the instrument exhaust port.  
CAUTION!  
Pressure must be less than 15 PSI.  
3. Pressurize the system and check for leaks by watching overall pressure.  
The pressure should not drop more than 1"Hg (.5psi) in 5 minutes.  
If the instrument fails the pressure test, each fitting needs to be leak checked to  
find the location.  
NOTE  
Be careful that the system is always pressurized so as not to draw soap solution into the  
plumbing system. Make sure you dry off any accumulated bubble solution. Start by  
checking the external fittings first.  
30  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.0 TROUBLESHOOTING, ADJUSTMENTS  
CAUTION!  
The operations outlined in this section must be performed by qualified  
maintenance personnel only!  
This section of the manual contains information on diagnosing and repairing  
instrument performance problems. It provides troubleshooting procedures that  
address problems to the board level.  
As a guide to troubleshooting, think of the T200U-NOy as two systems:  
pneumatics and electronics.  
Pneumatics - Over 50% of all instrument problems are traced to pneumatic  
leaks. Suspect a leak first.  
Electronics – external Converter temp control board or DC power supply for  
the converter temperature readout.  
Another trouble-shooting tip has to do with incorrect span gas concentration:  
This could come either from the calibrator itself or from incorrectly  
entering the expected span gas concentration in the instrument.  
If the instrument does not respond to span gas, check troubleshooting  
section in the T200 manual.  
The above should get you started in diagnosing and repairing the most common  
faults. If these reasons have been eliminated and there are still problems with the  
unit, the next thing to do is a Factory Calibration covered in the T200 manual. If  
difficulties persist, contact our service department. The 800 telephone number is  
on the cover page of this manual.  
7.1 Operation Verification – Diagnostic Techniques  
This section provides guidance on diagnosing some possible problems.  
7.1.1 Pneumatic System  
Refer to Figure 5 for a diagram of this system’s pneumatics, to be used in  
conjunction with diagnostics and troubleshooting.  
7.1.2 Leak Check  
Refer to Section 6.6 for the leak check procedure.  
07303C DCN6646  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model T200U-NOy Option Manual Addendum  
7.1.3 501Y Bypass Pump Diagnostic Procedures  
The 501Y bypass pump is capable of maintaining a total bypass flow rate of ~2.2  
lpm. If flow checks indicate a fall off in bypass flow, the pump may need  
servicing. Check the pump, inlet fittings, and analyzer for leaks first. If other  
causes have been eliminated, rebuild the pump.  
7.1.4 Electrical Fault Isolation  
115 VAC is supplied to:  
the bypass pump  
the +5 volt and +24 volt switcher supplies  
the input power to the converter temperature controller  
Refer to Figure 7 to check the 115 VAC supply voltage to each of these  
components.  
The converter temperature is controlled by the FUJI controller located on the  
501Y front panel. The controller cycles power to the external converter, thereby  
maintaining the temperature at 315 OC, using the solid state relay.  
7.2 Setting the Converter Temperature  
A front panel-mounted programmable controller maintains the temperature of the  
remote external Converter. By pressing the PV/SV button in the lower left corner  
of the controller, you can see the Present Value (actual temperature) or the Set  
Value (temperature set point). The temperature and PID (proportional, integral  
and derivative) control parameters have been set at the factory and no further  
adjustment should be necessary. Should you need to adjust the temperature,  
follow these steps:  
1. Select SV with the PV/SV button.  
2. Press the "up-arrow" under the digit you want to change (the digit will  
flash),  
3. Press the "up-arrow" under the digit or the "down-arrow" at the left to  
scroll the digit to the desired value.  
4. Repeat for the other two digits.  
5. Press the ENTER button.  
6. Select PV with the PV/SV button to observe the actual temperature.  
7.2.1 Temperature Controller Setup  
The FUJI temperature controller has been programmed at the factory and should  
not need to be altered under normal usage.  
32  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model T200U-NOy Option Manual Addendum  
Indications that would require re-programming are:  
Failure / replacement of controller.  
Replacement of thermocouple element.  
Controller’s loss of PID parameters leading to instability of the temp  
setpoint.  
In the event that the control parameters are changed or in the event that a new  
controller is installed, it must be reprogrammed to suit the thermal characteristics  
of the instrument. The controller includes an Autotuning feature, which can be  
used to properly set the PID factors.  
To start the Autotune procedure:  
1. Press SEL until A7 (AT) is displayed  
2. Press Data until value = 1 (enable Autotune)  
While Autotuning, the decimal point LED will blink.  
NOTE: During the procedure, it is normal for the temp to  
overshoot up to 100°C.  
When the decimal point LED turns off, Autotune is complete. No  
further adjustment should be necessary.  
Table 7 Temperature Controller Programming Guide  
PRESS  
DISPLAY  
ACTION  
UP/DOWN SET TO 5.0 (ENTER)  
SET TO 50  
SEL  
P PROP BAND  
DATA  
SEL  
DATA  
SEL  
DATA  
SEL  
DATA  
SEL  
DATA  
SEL  
DATA  
SEL  
i INTEGRAL  
d DERIVATIVE TIME  
AL LOW ALARM SETP  
AH HIGH ALARM SETP  
SET TO 15  
SET TO 300  
SET TO 330  
7C CYCLE TIME  
HYS HYSTERESIS  
LOC LOCK  
SET TO 2 (FOR FAST SYSTEM)  
SET TO 0.3  
0 = OPEN 1 = LOCKED  
2 = SV ONLY OPEN  
SEL  
NOTE:  
DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC.  
07303C DCN6646  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
SECONDARY MENU (The fixed characteristics of the system).  
PRESS  
DISPLAY  
ACTION  
SEL hold until  
SEL hold until  
SEL  
SET TO 0 – REVERSE ACTION  
SET TO 2 = TYPE J THERMOCOUPLE  
SET TO 5  
p-n1  
p-n2  
p-dF DIGITAL FILTER  
P-SL LOWER LIMIT  
P-SU UPPER LIMIT  
P-AH ALARM  
SEL  
SET TO -40 (-40ºC)  
SET TO 350 (350ºC)  
SET TO 0  
SEL  
SEL  
SEL  
SET TO 5  
P-An HYTERESIS  
P-dP PROG DECIMAL  
PVOF PROCESS OFFSET  
SVOF SET POINT OFFS  
P-F  
SEL  
SET TO 0  
SEL  
LEAVE AT 0  
SEL  
LEAVE AT 0  
SEL  
SET TO 0  
(0 = oC, 1 = oF)  
SEL  
ALREADY CONFIGURED  
P-n1  
SEL hold until RETURNS TO MAIN MENU  
SEL  
PV/SV  
SET DISPLAY TO 315 C  
ALREADY CONFIGURED  
SEL hold until  
P PROP BAND  
34  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model T200U-NOy Option Manual Addendum  
Figure 7 Electrical Diagram  
07303C DCN6646  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
This page intentionally left blank.  
36  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8.0 T200U-NOY SPARE PARTS LIST  
The following spare parts list supplements the spare parts list for the T200U analyzer.  
Note  
Use of replacement parts other than those supplied by API may result in non-compliance with  
European Standard EN 61010-1.  
Note  
Due to the dynamic nature of part numbers, please refer to the TAPI Website at  
http://www.teledyne-api.com or call Customer Service at 800-324-5190 for more recent updates  
to part numbers.  
Reference: 06848 11/8/2010 1:13:10 PM  
PARTNUMBER  
000940100  
000940400  
000941100  
001330001  
001761800  
002270100  
002730000  
004330000  
005960000  
005970000  
009690200  
009690300  
011310000  
011630000  
013140000  
014030000  
014080200  
016290000  
016300900  
024600100  
024930000  
037860000  
039700100  
040010000  
040030800  
040400000  
040410500  
040420200  
041510200  
041800300  
DESCRIPTION  
CD, ORIFICE, .003 GREEN  
CD, ORIFICE, .004 BLUE  
CD, ORIFICE, .014 ORG/VIOLET  
SLEEVE, GOLD PLATED, NOX  
ASSY, FLOW CTL, 90CC, 1/4" TEE-TMT, B  
AKIT, GASKETS, WINDOW, (12 GASKETS = 1)  
CD, FILTER, 665NM (KB)  
ZERO AIR SCRUBBER (NO/NO2)  
AKIT, EXP, 6LBS ACT CHARCOAL (2 BT=1)  
AKIT, EXP, 6LB PURAFIL (2BT=1)  
AKIT, TFE FLTR ELEM (FL19,100=1) 47mm  
AKIT, TFE FLTR ELEM (FL19, 30=1) 47mm  
ASSY, DRYER, NOX  
HVPS INSULATOR GASKET (KB)  
ASSY, COOLER FAN (NOX/SOX)  
AKIT, NOX EXPENDABLES, IZS  
ASSY, HVPS, NOX  
WINDOW, SAMPLE FILTER, 47MM (KB)  
ASSY, SAMPLE FILTER, 47MM, ANG BKT, 5UM  
ASSY, PREREACTOR NO/NOX VALVE  
CD, PMT, NOX  
ORING, TEFLON, RETAINING RING, 47MM (KB)  
HEATER, BAND, TYPE K, DUAL VOLTAGE(KB)  
ASSY, FAN REAR PANEL  
PCA, PRESS SENSORS (2X), FLOW, (NOX)  
ASSY, HEATERS/THERMAL SWITCH, RX CELL  
ASSY, VACUUM MANIFOLD  
ASSY, O3 GEN BRK, HIGH-O/P  
ASSY, PRESS/FLOW SENSOR, NOX  
PCA, PMT PREAMP, (OBS)  
07303C DCN6646  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model T200U-NOy Option Manual Addendum  
Reference: 06848 11/8/2010 1:13:10 PM  
PARTNUMBER  
041800600  
041920000  
DESCRIPTION  
PCA, PMT PREAMP, VR,  
ASSY, THERMISTOR  
042680100  
043420000  
044340000  
044540000  
044600000  
045230100  
045230200  
045430000  
04550010A  
04550030A  
045500400  
046030000  
046480000  
046490000  
047150000  
049310100  
049760300  
050700200  
051210000  
051990000  
052820000  
052930200  
055300000  
055740000  
055740100  
055740200  
056490000  
058021100  
058230000  
059940000  
062390000  
064540000  
064540100  
064540200  
066970000  
067240000  
067300000  
067300100  
067300200  
067900000  
068250100  
068810000  
069500000  
072150000  
072820000  
ASSY, VALVE (SS)  
ASSY, HEATER/THERM, O2 SEN  
ASSY, HTR, BYPASS MANIFOLD  
ASSY, THERMISTOR, NOX  
AKIT, SPARES, NOX  
PCA, RELAY CARD  
PCA, RELAY CARD  
ASSY, DRYER, O3, W/RELAYS & BRKT *  
ASSY, ORIFICE HOLDER, 4 MIL  
ASSY, ORIFICE HOLDER, 10 MIL, (NOX) (KB)  
ASSY, ORIFICE HOLDER, 3 MIL  
AKIT, CH-43, 3 REFILLS  
ASSY, DILUTION MANIFOLD, (KB)  
OPTION, DILUTION MANIFOLD (KB)  
AKIT, EXPENDABLES, NOX  
PCA,TEC DRIVER,PMT,(KB)  
ASSY, TC PROG PLUG, MOLY,TYP K, TC1  
KIT, RELAY BD NOX CONFIGURATION  
DESTRUCT w/FTGS, O3 *  
ASSY, SCRUBBER, INLINE EXHAUST, DISPOS  
ASSY, IZS, HEATER/THERM, NOX  
ASSY, BAND HEATER TYPE K, NOX  
AKIT, PUMP REBUILD, THOMAS 2688, DUAL HD  
ASSY, PUMP, NOx PUMP PACK, 115V/60HZ  
ASSY, PUMP, NOx PUMP PACK, 220V/60HZ  
ASSY, PUMP, NOx PUMP PACK, 220V/50HZ  
FILTER, DFU, DESORBER (SOAKED)  
PCA, MOTHERBD, GEN 5-ICOP  
ASSY, O3 CLEANSER, ALUMINUM  
OPTION, SAMPLE GAS CONDITIONER, NOX*  
ASSY, MOLY GUTS w/WOOL  
ASSY, PUMP NOX INTERNAL, 115V/60HZ  
ASSY, PUMP NOX INTERNAL, 230V/60HZ  
ASSY, PUMP NOX INTERNAL, 230V/50HZ  
PCA, INTRF. LCD TOUCH SCRN, F/P  
CPU, PC-104, VSX-6154E, ICOP *  
PCA, AUX-I/O BD, ETHERNET, ANALOG & USB  
PCA, AUX-I/O BOARD, ETHERNET  
PCA, AUX-I/O BOARD, ETHERNET & USB  
LCD MODULE, W/TOUCHSCREEN  
DOM, w/SOFTWARE, STD, T200U *  
PCA, LVDS TRANSMITTER BOARD  
PCA, SERIAL & VIDEO INTERFACE BOARD  
ASSY. TOUCHSCREEN CONTROL MODULE  
KIT, T200U MANUAL  
38  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model T200U-NOy Option Manual Addendum  
Reference: 06848 11/8/2010 1:13:10 PM  
PARTNUMBER  
CN0000073  
FL0000001  
FL0000003  
FM0000004  
FT0000010  
HW0000005  
HW0000020  
HW0000030  
HW0000036  
HW0000099  
HW0000101  
HW0000453  
KIT000051  
KIT000095  
KIT000115  
KIT000156  
KIT000253  
KIT000254  
OR0000001  
OR0000002  
OR0000025  
OR0000027  
OR0000034  
OR0000039  
OR0000044  
OR0000046  
OR0000083  
OR0000086  
OR0000094  
PU0000047  
PU0000048  
PU0000083  
RL0000015  
SW0000025  
WR0000008  
DESCRIPTION  
POWER ENTRY, 120/60 (KB)  
FILTER, SS (KB)  
FILTER, DFU (KB)  
FLOWMETER (KB)  
CONNECTOR-ORING, SS, 1/8" (HK)  
FOOT  
SPRING  
ISOLATOR  
TFE TAPE, 1/4" (48 FT/ROLL)  
STANDOFF, #6-32X.5, HEX SS M/F  
ISOLATOR  
SUPPORT, CIRCUIT BD, 3/16" ICOP  
AKIT, REBUILD, RX CELL AMBIENT  
AKIT, REPLACEMENT COOLER  
OPTION, SAMPLE GAS COND, (KB)  
OPTION, CARRYING HANDLE  
ASSY & TEST, SPARE PS37  
ASSY & TEST, SPARE PS38  
ORING, 2-006VT *(KB)  
ORING, 2-023V  
ORING, 2-133V  
ORING, 2-042V  
ORING, 2-011V FT10  
ORING, 2-012V  
ORING, 2-125V  
ORING, 2-019V  
ORING, 105M, 1MM W X 5 MM ID, VITON  
ORING, 2-006, CV-75 COMPOUND(KB)  
ORING, 2-228V, 50 DURO VITON(KB)  
PUMP, THOMAS 2688VE44, 2 HEAD, 115V/60HZ  
PUMP, THOMAS 2688VGH144, 230V 50/60HZ CE  
KIT, REBUILD, PU80, PU81, PU82  
RELAY, DPDT, (KB)  
SWITCH, POWER, CIRC BREAK, VDE/CE *(KB)  
POWER CORD, 10A(KB)  
07303C DCN6646  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model T200U-NOy Option Manual Addendum  
This page intentionally left blank.  
40  
07303C DCN6646  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Timex Watch W 102 User Manual
Toshiba CRT Television 32ZT29B User Manual
Toshiba Telephone DK24 User Manual
Toshiba VCR W 704C User Manual
Tricity Bendix Kitchen Entertainment Center HG210 User Manual
Troy Bilt Edger 769 02047 User Manual
Tru Test Battery Charger 1 J User Manual
Vector Automobile Accessories VEC012AP User Manual
Velocity Micro Computer Accessories SEAL IT 100 User Manual
ViewSonic Mouse VS102127 User Manual