INSTRUCTION MANUAL
Model 9110T
Nitrogen Oxides Analyzer
P/N M9110T
DATE 11/15/13
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
Telephone: (626) 934-1500
Fax: (626) 961-2538
Web: www.teledyne-ai.com
Teledyne Analytical Instruments
ii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Specific Configuration
INFORMATION ABOUT THE SPECIFIC CONFIGURATION OF YOUR
MODEL 9110T NOX ANALYZER
Note: All instruments must include the standard hardcopy manual.
SELECTED VERSIONS OF THE MODEL 9110T
Model 9110T— Standard Touch Screen Version
This Model 9110E NOx Analyzer is a touch screen version designed for analyzing the NOx
concentration in a background gas specified by the customer. It has a minimum settable range of 0-50
ppb and a maximum settable range of 0-10 ppm. The standard version is designed for ambient pressure
applications. The analyzer may have one or two analysis ranges with or without auto-ranging as listed
below. It includes an internal Moly converter and an external pump. Alarm relays are optional and if
included, that option will be checked below.
Model 9110TH — High Range Touch Screen Version
The Model 9110EH NOx Analyzer is a touch screen version designed for analyzing higher NOx
concentration (from 0-5 ppm to 0-5000 ppm) than the standard model. It includes an internal Hicon
converter and an external pump. Alarm relays are optional and if included, that option will be checked
below. There is no internal zero/span gas/oven option available for this model. The analyzer may have
one or two analysis ranges with or without autoranging as listed below.
This version is available with internal or external Hicon or Moly converters.
Model 9110TM — Mid-Range Touch Screen Version
The Model 9110EA NOx Analyzer is a touch screen version designed for analyzing a mid-range NOx
concentration. It has a minimum settable range of 0-1 ppm and a maximum settable range of 0-200 ppm.
t includes an internal Moly converter and an external pump. Alarm relays are optional and if included,
that option will be checked below. There is no internal zero/span gas/oven option available for this
model. The analyzer may have one or two analysis ranges with or without autoranging as listed below.
This model is similar to the EH version but does not have a sample bypass line. An optional
paramagnetic oxygen sensor is available for oxygen analysis in this version.
Converter Options
The 9110 is equipped with an internal Hicon thermal converter as standard equipment. Other
converters are available for the EH version as follows:
Internal Hicon (standard)
Internal Moly
External Hicon (2nd set of rack mounts req’d)
External Moly (2nd set of rack mounts req’d)
Teledyne Analytical Instruments
iii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Specific Configuration
Power Requirements
This Model 9110E is configured to operate from the following AC Power source:
100-120 VAC 60 Hz
100-120 VAC 50 Hz
220-240 VAC 60 Hz
220-240 VAC 50 Hz
100V 60 Hz
100V 50 Hz
Analog Output Signals
Analog output signals are available at A1, A2, and A3 on the rear panel. This instrument is configured
with the following analog outputs:
A1(NOx): 4-20 mA
A2 (NO): 0-5 V
A3 (NO2): 0-5 V
A2(NO): 4-20 mA
A2(NO2): 4-20 mA
Range Mode
The analyzer can be designed with a single or dual analysis ranges with auto-ranging or dual
independent ranges. This analyzer is configured with the following range mode:
Single Range:
Dual Range/Auto-ranging:
Dual Range/Independent:
Low Range:
Low Range:
High Range:
High Range:
Gas:
Selected Options for the Model 9110T
Mounting Options
19” rack mounting with 26” sliders with ears
19” rack mounting with ears only
Pump Mounting Options
None
Rack mounting hardware
Rear Panel Gas Fittings
1/4” SS Standard
6 mm SS Optional
Teledyne Analytical Instruments
iv
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Specific Configuration
Valve Options
No Valves
Internal SS Zero/Span Valves
Second Range Span Valve
Internal Zero/Span Valves with Oven (Not available for 9110EH/EM)
Internal Zero/Span Valves with Oven and Permeation Tube (Not available for 9110EH)
The permeation tube option installed depends on the sample gas and the effusion rate.
The specific permeation tube in this instrument is listed below:
Permeation tube Installed:
Alarm Relay Option
This option includes two concentration alarm relays.
Oxygen Sensor
This analyzer is equipped with a paramagnetic oxygen sensor for measuring the oxygen concentration
over the range of 0-25%
Gas Conditioner
A gas conditioner/dryer permeation gas exchange tube is installed for removing H2O and ammonia
from the sample stream. This item is required for EN certificatation.
Oxygenator
For applications where background gas less than 2% oxygen and using a Moly converter.
Remote Operation for Purged Enclosure
The operator interface is duplicated with switches mounted on the front door of the enclosure allowing
operation without compromising purge integrity.
Background Gas:
Notes:
Teledyne Analytical Instruments
v
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Specific Configuration
This page intentionally left blank..
Teledyne Analytical Instruments
vi
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Safety Messages
SAFETY MESSAGES
Important safety messages are provided throughout this manual for the purpose of avoiding personal
injury or instrument damage. Please read these messages carefully. Each safety message is associated
with a safety alert symbol, and are placed throughout this manual and inside the instrument. The symbols
with messages are defined as follows:
WARNING: Electrical Shock Hazard
HAZARD: Strong oxidizer
GENERAL WARNING/CAUTION: Read the accompanying message for
specific information.
CAUTION: Hot Surface Warning
Do Not Touch: Touching some parts of the instrument without protection or
proper tools could result in damage to the part(s) and/or the instrument.
Technician Symbol: All operations marked with this symbol are to be
performed by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central
safety grounding point for the instrument.
CAUTION
GENERAL SAFETY HAZARD
The T100 Analyzer should only be used for the purpose and in the
manner described in this manual. If you use the T100 in a manner other
than that for which it was intended, unpredictable behavior could ensue
with possible hazardous consequences.
NEVER use any gas analyzer to sample combustible gas(es).
Technical Assistance regarding the use and maintenance of the 9110T or
any other Teledyne product can be obtained by contacting Teledyne’s
Customer Service Department:
Note
Phone: 626-934-1500
Email: [email protected]
or by accessing various service options on our website at
http://www.teledyne-ai.com/.
Teledyne Analytical Instruments
vii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Safety Messages
CONSIGNES DE SÉCURITÉ
Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des
blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque
consigne de sécurité est représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent
dans ce manuel et à l’intérieur des instruments. Les symboles correspondent aux consignes suivantes :
AVERTISSEMENT : Risque de choc électrique
DANGER : Oxydant puissant
AVERTISSEMENT GÉNÉRAL
/
MISE EN GARDE : Lire la consigne
complémentaire pour des renseignements spécifiques
MISE EN GARDE : Surface chaude
Ne pas toucher : Toucher à certaines parties de l’instrument sans protection ou
sans les outils appropriés pourrait entraîner des dommages aux pièces ou à
l’instrument.
Pictogramme « technicien » : Toutes les opérations portant ce symbole doivent
être effectuées uniquement par du personnel de maintenance qualifié.
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central
de la mise à la terre sécuritaire de l’instrument.
MISE EN GARDE
Cet instrument doit être utilisé aux fins décrites et de la manière décrite dans
ce manuel. Si vous utilisez cet instrument d’une autre manière que celle pour
laquelle il a été prévu, l’instrument pourrait se comporter de façon imprévisible
et entraîner des conséquences dangereuses.
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz
combustibles!
Teledyne Analytical Instruments
viii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Safety Messages
ABOUT THIS MANUAL
This manual describes operation, specifications, and maintenance for the Model 9110T.
In addition this manual contains important SAFETY messages for this instrument. It is strongly
recommended that you read that operation manual in its entirety before operating the instrument.
ORGANIZATION
This manual is divided among three main parts and a collection of appendices at the end.
Part I contains introductory information that includes an overview of the analyzer, specifications,
descriptions of the available options, installation and connection instructions, and the initial calibration
and functional checks.
Part II comprises the operating instructions, which include initial functional checks and calibration,
basic, advanced and remote operation, advanced calibration, diagnostics, testing, and ends with specifics
of calibrating for use in EPA monitoring.
Part III provides detailed technical information, starting with maintenance, troubleshooting and service
with Frequently Asked Questions (FAQs), followed by principles of operation, and a glossary. It also
contains a special section dedicated to providing information about electro-static discharge and
protecting against its consequences.
The appendices at the end of this manual provide support information such as, version-specific software
documentation, lists of spare parts and recommended stocking levels, and schematics.
CONVENTIONS USED
In addition to the safety symbols as presented in the Important Safety Information page, this manual
provides special notices related to the safety and effective use of the analyzer and other pertinent
information.
Special Notices appear as follows:
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
This special notice provides information to avoid damage to your
instrument and possibly invalidate the warranty.
IMPORTANT
IMPACT ON READINGS OR DATA
Could either affect accuracy of instrument readings or cause loss of data.
Note
Pertinent information associated with the proper care, operation or
maintenance of the analyzer or its parts.
Teledyne Analytical Instruments
ix
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Safety Messages
This page intentionally left blank..
Teledyne Analytical Instruments
x
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Table of Contents
TABLE OF CONTENTS
Information About the Specific Configuration of Your Model 9110T NOX Analyzer ..............................................................iii
Mounting Options .............................................................................................................................................................. iv
Pump Mounting Options .................................................................................................................................................... iv
Rear Panel Gas Fittings.................................................................................................................................................... iv
Safety Messages....................................................................................................................................................................vii
About This Manual ................................................................................................................................................................. ix
PART I GENERAL INFORMATION....................................................................................................................................21
1. Introduction, Features and Options........................................................................................................................................23
1.1. Overview ........................................................................................................................................................................23
1.2. Features .........................................................................................................................................................................23
1.3. Documentation ...............................................................................................................................................................24
1.4. Options...........................................................................................................................................................................24
2. Specifications, Approvals, & Compliance...............................................................................................................................27
2.1. Specifications .................................................................................................................................................................27
2.2. EPA Equivalency Designation........................................................................................................................................28
2.3. Approvals and Certifications...........................................................................................................................................29
2.3.1. Safety .....................................................................................................................................................................29
2.3.2. EMC........................................................................................................................................................................29
2.3.3. Other Type Certifications........................................................................................................................................29
3. Getting Started
31
3.1. Unpacking the 9110T Analyzer ......................................................................................................................................31
3.1.1. Ventilation Clearance..............................................................................................................................................32
3.2. Instrument Layout...........................................................................................................................................................32
3.2.1. Front Panel.............................................................................................................................................................33
3.2.2. Rear Panel..............................................................................................................................................................37
3.2.3. Internal Chassis Layout ..........................................................................................................................................38
3.3. Connections and Setup ..................................................................................................................................................41
3.3.1. Electrical Connections ............................................................................................................................................41
3.3.2. Pneumatic Connections..........................................................................................................................................55
3.4. Startup, Functional Checks, and Initial Calibration.........................................................................................................74
3.4.1. Start Up ..................................................................................................................................................................74
3.4.2. Warning Messages.................................................................................................................................................75
3.4.3. Functional Checks ..................................................................................................................................................77
3.4.4. Initial Calibration.....................................................................................................................................................77
3.4.4.1. Interferents ..........................................................................................................................................................78
PART II – OPERATING INSTRUCTIONS............................................................................................................................85
4. Overview of Operating Modes................................................................................................................................................87
4.1. Sample Mode .................................................................................................................................................................88
4.1.1. Test Functions........................................................................................................................................................88
4.1.2. Warning Messages.................................................................................................................................................91
4.2. Calibration Mode ............................................................................................................................................................92
4.3. Setup Mode....................................................................................................................................................................92
4.3.1. Password Security..................................................................................................................................................93
4.3.2. Primary Setup Menu...............................................................................................................................................93
4.3.3. Secondary Setup Menu (SETUP MORE)............................................................................................................93
5. Setup Menu
95
5.1. SETUP CFG: Configuration Information ....................................................................................................................95
5.2. SETUP ACAL: Automatic Calibration Option..............................................................................................................95
5.3. SETUP DAS: Internal Data Acquisition System..........................................................................................................95
5.4. SETUP RNGE: Analog Output Reporting Range Configuration .................................................................................96
5.4.1. 9110T Physical Ranges..........................................................................................................................................96
5.4.2. 9110T Analog Output Reporting Ranges................................................................................................................96
5.4.3. SETUP RNGE MODE....................................................................................................................................98
5.5. SETUP PASS: Password Protection........................................................................................................................106
5.6. SETUP CLK: Setting the Internal Time-of-Day Clock ..............................................................................................108
5.6.1. Setting the Time of Day ........................................................................................................................................108
5.6.2. Adjusting the Internal Clock’s Speed ....................................................................................................................109
5.7. SETUP COMM: Communications Ports...................................................................................................................110
5.7.1. ID (Machine Identification)....................................................................................................................................110
5.7.2. INET (Ethernet) ....................................................................................................................................................111
5.7.3. COM1[COM2] (Mode, Baude Rate and Test Port) ...............................................................................................111
5.8. SETUP VARS: Variables Setup and Definition........................................................................................................111
Teledyne Analytical Instruments
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Table of Contents
5.9. SETUP Diag: Diagnostics Functions........................................................................................................................114
5.9.1. Signal I/O..............................................................................................................................................................116
5.9.2. Analog Output (DIAG AOUT)................................................................................................................................117
5.9.3. Analog I/O Configuration (DIAG AIO) ...................................................................................................................117
5.9.4. Test Chan Output (Selecting a Test Channel Function for Output A4) .................................................................132
5.9.5. Optic Test .............................................................................................................................................................133
5.9.6. Electrical Test.......................................................................................................................................................134
5.9.7. Ozone Gen Override.............................................................................................................................................134
5.9.8. Flow Calibration....................................................................................................................................................134
6. Communications Setup and Operation ................................................................................................................................135
6.1. Data Terminal / Communication Equipment (DTE DEC)..............................................................................................135
6.2. Communication Modes, Baud Rate and Port Testing...................................................................................................135
6.2.1. Communication Modes.........................................................................................................................................135
6.2.2. Com Port Baud Rate.............................................................................................................................................138
6.2.3. Com Port Testing..................................................................................................................................................138
6.3. RS-232 .........................................................................................................................................................................139
6.4. RS-485 (Option) ...........................................................................................................................................................140
6.5. Ethernet........................................................................................................................................................................140
6.5.1. Configuring Ethernet Communication Manually (Static IP Address).....................................................................140
6.5.2. Configuring Ethernet Communication Using Dynamic Host Configuration Protocol (DHCP)................................143
6.6. USB Port for Remote Access .......................................................................................................................................147
6.7. Communications Protocols...........................................................................................................................................149
6.7.1. MODBUS..............................................................................................................................................................149
6.7.2. Hessen .................................................................................................................................................................151
7. Data Acquisition System (DAS) and APICOM .....................................................................................................................163
7.1. DAS Structure ..............................................................................................................................................................164
7.1.1. DAS Channels......................................................................................................................................................164
7.1.2. Viewing DAS Data and Settings ...........................................................................................................................168
7.1.3. Editing DAS Data Channels..................................................................................................................................169
7.2. Remote DAS Configuration ..........................................................................................................................................181
7.2.1. DAS Configuration via APICOM ...........................................................................................................................181
7.2.2. DAS Configuration via Terminal Emulation Programs ..........................................................................................183
8. Remote Operation
185
8.1. Computer Mode............................................................................................................................................................185
8.1.1. Remote Control via APICOM................................................................................................................................185
8.2. Interactive Mode...........................................................................................................................................................185
8.2.1. Remote Control via a Terminal Emulation Program..............................................................................................185
8.3. Remote Access by Modem...........................................................................................................................................188
8.4. Password Security for Serial Remote Communications ...............................................................................................191
9. Calibration Procedures 193
9.1. Before Calibration.........................................................................................................................................................194
9.1.1. Required Equipment, Supplies, and Expendables................................................................................................194
9.1.2. Calibration Gases .................................................................................................................................................195
9.1.3. Data Recording Devices.......................................................................................................................................196
9.1.4. NO2 Conversion Efficiency (CE) ...........................................................................................................................197
9.2. Manual Calibration Checks and Calibration of the 9110T Analyzer in its Base Configuration......................................197
9.2.1. Setup for Basic Calibration Checks and Calibration of the 9110T analyzer..........................................................197
9.2.2. Performing a Basic Manual Calibration Check .....................................................................................................199
9.2.3. Performing a Basic Manual Calibration.................................................................................................................199
9.3. Manual Calibration with the Internal Span Gas Generator ...........................................................................................202
9.3.1. Performing “Precision” Manual Calibration when Internal Span Gas (IZS) Generator Option is Present .............202
9.3.2. Setup for Calibration with the Internal Span Gas Generator.................................................................................203
9.3.3. CAL On NO2 Feature............................................................................................................................................203
9.3.4. Performing a Manual Calibration Check with the Internal Span Gas Generator...................................................205
9.3.5. Performing a Manual Calibration with the Internal Span Gas Generator ..............................................................206
9.4. Manual Calibration and Cal Checks with the Valve Options Installed ..........................................................................209
9.4.1. Setup for Calibration Using Valve Options............................................................................................................209
9.4.2. Manual Calibration Checks with Valve Options Installed......................................................................................210
9.4.3. Manual Calibration Using Valve Options ..............................................................................................................211
9.5. Automatic Zero/Span Cal/Check (AutoCal) ..................................................................................................................213
9.5.1. SETUP ACAL: Programming and AUTO CAL Sequence ................................................................................216
9.6. Calibration Quality Analysis..........................................................................................................................................219
9.7. Gas Flow Calibration....................................................................................................................................................220
10. EPA Protocol Calibration....................................................................................................................................................221
10.1. 9110T Calibration – General Guidelines ....................................................................................................................221
Teledyne Analytical Instruments
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Table of Contents
10.2. Calibration Equipment, Supplies, and Expendables...................................................................................................222
10.2.1. Spare Parts and Expendable Supplies ...............................................................................................................222
10.2.2. Calibration Gas and Zero Air Sources ................................................................................................................223
10.2.3. Data Recording Device.......................................................................................................................................223
10.2.4. Record Keeping..................................................................................................................................................223
10.3. Calibration Frequency ................................................................................................................................................223
10.4. Level 1 Calibrations versus Level 2 Checks...............................................................................................................224
10.5. Gas Phase Titration (GPT).........................................................................................................................................225
10.5.1. GPT Principle of Operation.................................................................................................................................225
10.5.2. GPT Calibrator Check Procedure.......................................................................................................................225
10.6. GPT Multipoint Calibration Procedure........................................................................................................................228
10.6.1. Set Up for GPT Multipoint Calibration of the 9110T............................................................................................229
10.6.2. Zero Calibration..................................................................................................................................................230
10.6.3. Span Calibration.................................................................................................................................................230
10.7. GPT NO2 Check .........................................................................................................................................................231
10.8. Other Quality Assurance Procedures .........................................................................................................................232
10.8.1. Summary of Quality Assurance Checks .............................................................................................................232
10.8.2. Short Calibration Checks....................................................................................................................................233
10.8.3. Zero/Span Check Procedures ............................................................................................................................233
10.8.4. Precision Check..................................................................................................................................................234
10.9. Certification of Working Standards.............................................................................................................................234
10.10. References...............................................................................................................................................................235
PART III – Maintenance and Service ..................................................................................................................................237
11. Instrument Maintenance.....................................................................................................................................................239
11.1. Maintenance Schedule...............................................................................................................................................239
11.2. Predictive Diagnostics ................................................................................................................................................241
11.3. Maintenance Procedures............................................................................................................................................241
11.3.1. Replacing the Sample Particulate Filter..............................................................................................................242
11.3.2. Changing the O3 Dryer Particulate Filter.............................................................................................................243
11.3.3. Changing the Ozone Cleanser Chemical............................................................................................................244
11.3.4. Maintaining the External Sample Pump (Pump Pack) ........................................................................................247
11.3.5. Changing the Pump DFU Filter...........................................................................................................................247
11.3.6. Changing the Internal Span Gas Generator Permeation Tube...........................................................................249
11.3.7. Changing the External Zero Air Scrubber (OPT 86C).........................................................................................249
11.3.8. Changing the NO2 Converter..............................................................................................................................252
11.3.9. Cleaning the Reaction Cell .................................................................................................................................254
11.3.10. Replacing Critical Flow Orifices........................................................................................................................256
11.3.11. Checking for Light Leaks ..................................................................................................................................257
11.3.12. Checking for Pneumatic Leaks.........................................................................................................................258
12. Troubleshooting & Service.................................................................................................................................................261
12.1. General Troubleshooting............................................................................................................................................261
12.1.1. Fault Diagnosis with WARNING Messages........................................................................................................262
12.1.2. Fault Diagnosis With Test Functions ..................................................................................................................266
12.1.3. DIAG SIGNAL I/O: Using the Diagnostic Signal I/O Function .......................................................................267
12.2. Using the Analog Output Test Channel......................................................................................................................269
12.3. Using the Internal Electronic Status LEDs..................................................................................................................270
12.3.1. CPU Status Indicator ..........................................................................................................................................270
12.3.2. Relay PCA Status LEDs .....................................................................................................................................270
12.4. Gas Flow Problems....................................................................................................................................................272
12.4.1. Zero or Low Flow Problems................................................................................................................................272
12.5. Calibration Problems..................................................................................................................................................276
12.5.1. Negative Concentrations ....................................................................................................................................276
12.5.2. No Response......................................................................................................................................................277
12.5.3. Unstable Zero and Span.....................................................................................................................................277
12.5.4. Inability to Span - No SPAN Button (CALS)........................................................................................................278
12.5.5. Inability to Zero - No ZERO Button (CALZ).........................................................................................................278
12.5.6. Non-Linear Response.........................................................................................................................................279
12.5.7. Discrepancy Between Analog Output and Display .............................................................................................280
12.5.8. Discrepancy Between NO and NOX slopes........................................................................................................280
12.6. Other Performance Problems.....................................................................................................................................281
12.6.1. Excessive Noise .................................................................................................................................................281
12.6.2. Slow Response...................................................................................................................................................281
12.6.3. Auto Zero Warnings...........................................................................................................................................282
12.7. Subsystem Checkout..................................................................................................................................................283
12.7.1. AC Main Power...................................................................................................................................................283
Teledyne Analytical Instruments
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Table of Contents
12.7.2. DC Power Supply ...............................................................................................................................................283
12.7.3. I2C Bus ...............................................................................................................................................................285
12.7.4. LCD/Display Module...........................................................................................................................................285
12.7.5. Relay PCA..........................................................................................................................................................285
12.7.6. Motherboard .......................................................................................................................................................285
12.7.7. Pressure / Flow Sensor Assembly......................................................................................................................289
12.7.8. CPU....................................................................................................................................................................290
12.7.9. RS-232 Communications....................................................................................................................................290
12.7.10. NO2 NO Converter.......................................................................................................................................291
12.7.11. Simplified GPT Calibration................................................................................................................................296
12.7.12. Photomultiplier Tube (PMT) Sensor Module.....................................................................................................300
12.7.13. PMT Preamplifier Board ...................................................................................................................................302
12.7.14. PMT Temperature Control PCA........................................................................................................................303
12.7.15. O3 Generator ....................................................................................................................................................304
12.7.16. Internal Span Gas Generator and Valve Options .............................................................................................305
12.7.17. Temperature Sensor.........................................................................................................................................306
12.8. Service Procedures....................................................................................................................................................308
12.8.1. Disk-On-Module Replacement Procedure ..........................................................................................................308
12.8.2. O3 Generator Replacement ................................................................................................................................309
12.8.3. Sample and Ozone (Perma Pure®) Dryer Replacement.....................................................................................309
12.8.4. PMT Sensor Hardware Calibration .....................................................................................................................310
12.8.5. Replacing the PMT, HVPS or TEC .....................................................................................................................312
12.8.6. Removing / Replacing the Relay PCA from the Instrument ................................................................................315
12.9. Frequently Asked Questions ......................................................................................................................................316
12.10. Technical Assistance................................................................................................................................................318
13. Principles of Operation.......................................................................................................................................................319
13.1. Measurement Principle...............................................................................................................................................319
13.1.1. Chemiluminescence Creation in the 9110T Reaction Cell..................................................................................319
13.1.2. Chemiluminescence Detection in the 9110T Reaction Cell ................................................................................320
13.1.3. NOX and NO2 Determination...............................................................................................................................321
13.1.4. Auto Zero............................................................................................................................................................322
13.1.5. Measurement Interferences................................................................................................................................323
13.2. Pneumatic Operation..................................................................................................................................................326
13.2.1. Sample Gas Flow ...............................................................................................................................................326
13.2.2. Flow Rate Control - Critical Flow Orifices ...........................................................................................................330
13.2.3. Ozone Gas Generation and Air Flow..................................................................................................................332
13.2.4. Pneumatic Sensors.............................................................................................................................................335
13.3. Electronic Operation...................................................................................................................................................338
13.3.1. Overview.............................................................................................................................................................338
13.3.2. CPU....................................................................................................................................................................339
13.3.3. Motherboard .......................................................................................................................................................340
13.3.4. Relay PCA..........................................................................................................................................................345
13.4. Sensor Module, Reaction Cell....................................................................................................................................351
13.5. Photo Multiplier Tube (PMT).......................................................................................................................................352
13.5.1. PMT Preamplifier................................................................................................................................................353
13.5.2. PMT Cooling System..........................................................................................................................................355
13.6. Pneumatic Sensor Board............................................................................................................................................356
13.7. Power Supply/Circuit Breaker.....................................................................................................................................357
13.7.1. AC Power Configuration .....................................................................................................................................358
13.8. Front Panel Touchscreen/Display Interface................................................................................................................363
13.8.1. LVDS Transmitter Board.....................................................................................................................................364
13.8.2. Front Panel Touchscreen/Display Interface PCA ...............................................................................................364
13.9. Software Operation ....................................................................................................................................................364
13.9.1. Adaptive Filter.....................................................................................................................................................365
13.9.2. Temperature/Pressure Compensation (TPC) .....................................................................................................365
13.9.3. Calibration - Slope and Offset.............................................................................................................................365
14. A Primer on Electro-Static Discharge.................................................................................................................................367
14.1. How Static Charges are Created................................................................................................................................367
14.2. How Electro-Static Charges Cause Damage..............................................................................................................368
14.3. Common Myths About ESD Damage.........................................................................................................................369
14.4. Basic Principles of Static Control................................................................................................................................369
14.4.1. General Rules.....................................................................................................................................................369
14.5. Basic anti-ESD Procedures for Analyzer Repair and Maintenance............................................................................371
14.5.1. Working at the Instrument Rack .........................................................................................................................371
14.5.2. Working at an Anti-ESD Work Bench .................................................................................................................371
Teledyne Analytical Instruments
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Table of Contents
14.5.3. Transferring Components Between Rack and Bench.........................................................................................372
14.5.4. Opening Shipments from TAI Customer Service................................................................................................372
14.5.5. Packing Components for Return to TAI Customer Service.................................................................................373
Glossary..............................................................................................................................................................................375
APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION
APPENDIX B - SPARE PARTS
APPENDIX C - REPAIR QUESTIONNAIRE
APPENDIX D - ELECTRONIC SCHEMATICS
Teledyne Analytical Instruments
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Table of Contents
FIGURES
Figure 3-1:
Figure 3-2:
Figure 3-3:
Figure 3-4:
Figure 3-5:
Figure 3-6:
Figure 3-7:
Figure 3-8:
Figure 3-9:
Figure 3-10:
Figure 3-11:
Figure 3-12:
Figure 3-13
Figure 3-14:
Figure 3-15:
Figure 3-16:
Figure 3-17:
Figure 3-18:
Figure 3-19:
Figure 3-20:
Figure 3-21:
Figure 3-22:
Figure 3-23:
Figure 3-24:
Figure 3-25:
Figure 3-26:
Figure 3-27:
Figure 3-28:
Figure 3-29:
Figure 4-1:
Figure 4-2:
Figure 5-1:
Figure 5-2.
Figure 5-3.
Figure 5-4:
Figure 5-5:
Figure 5-6:
Figure 5-7:
Figure 5-8:
Figure 6-1.
Figure 6-2.
Figure 6-3.
Figure 6-4.
Figure 6-5.
Figure 6-6.
Figure 7-1:
Figure 7-2:
Figure 7-3:
Figure 7-4:
Figure 8-1:
Figure 9-1:
Front Panel Layout.......................................................................................................................33
Display Screen and Touch Control..............................................................................................34
Display/Touch Control Screen Mapped to Menu Charts .............................................................36
Rear Panel Layout – Base Unit ...................................................................................................37
Internal Layout – Top View with IZS Option ................................................................................39
Internal Layout - Top View Showing Other Options ....................................................................40
Analog In Connector....................................................................................................................42
Analog Output Connector ............................................................................................................43
Current Loop Option Installed on the Motherboard .....................................................................44
Status Output Connector .............................................................................................................45
Energizing the 9110T Control Inputs ...........................................................................................46
Concentration Alarm Relay..........................................................................................................47
Rear Panel Connector Pin-Outs for RS-232 Mode......................................................................50
Default Pin Assignments for CPU COMM Port Connector (RS-232). .........................................51
Jumper and Cables for Multidrop Mode.......................................................................................53
RS-232-Multidrop PCA Host/Analyzer Interconnect Diagram.....................................................54
Gas Line Connections from Calibrator – Basic 9110T Configuration..........................................59
Gas Line Connections from Bottled Span Gas – Basic 9110T Configuration.............................60
Pneumatics, Basic Configuration.................................................................................................62
Rear Panel Layout with Z/S Valve Options (OPT 50A) ...............................................................63
Gas Line Connections for 9110T with Z/S Valves Option (OPT 50A).........................................63
Pneumatics with Zero/Span Valves OPT 50A.............................................................................65
Rear Panel Layout with Ambient Zero/Pressurized Span Valves OPT 50B................................66
Gas Line Connection w/Ambient Zero/Pressurized Span Valves (OPT 50B) .............................67
Pneumatics with Ambient Zero/Pressurized Span Valves (OPT 50B) ........................................68
Rear Panel Layout with Internal Span Source (IZS) OPT 50G ...................................................70
Pneumatics with the Internal Span Gas Generator (OPT 50G) ..................................................71
Pneumatics for Sample Conditioner OPT 86A ............................................................................72
Pneumatics for External Zero Air Scrubber (OPT 86C) for Z/S Valves.......................................73
Front Panel Display......................................................................................................................87
Viewing 9110T Test Functions ....................................................................................................90
Analog Output Connector Pin Out ...............................................................................................97
SETUP – COMM Menu .............................................................................................................110
COMM– Machine ID ..................................................................................................................111
Accessing the DIAG Submenus ................................................................................................115
Accessing the Analog I/O Configuration Submenus..................................................................118
Setup for Checking / Calibrating DCV Analog Output Signal Levels.........................................123
Setup for Checking / Calibration Current Output Signal Levels Using an Ammeter .................125
Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels ................127
COMM – Communication Modes Setup ....................................................................................137
COMM – COMM Port Baud Rate ..............................................................................................138
COMM – COM1 Test Port .........................................................................................................139
COMM - LAN /Internet Manual Configuration............................................................................142
COMM – LAN / Internet Automatic Configuration (DHCP)........................................................145
COMM – Change Hostname ....................................................................................................146
Default DAS Channels Setup ....................................................................................................167
APICOM Remote Control Program Interface ............................................................................181
Sample APICOM User Interface for Configuring the DAS.........................................................182
DAS Configuration Through a Terminal Emulation Program ....................................................183
Remote Access by Modem........................................................................................................189
Set up for Manual Calibrations/Checks of 9110T’s in Base Configuration w/ a Gas Dilution
Calibrator ...................................................................................................................................198
Set up for Manual Calibrations/Checks of 9110T’s in Base Configuration w/ Bottled Gas .......198
Pneumatic Connections for 9110T Precision Calibration when IZS Generator Present...........202
Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas Generator203
Figure 9-2:
Figure 9-3:
Figure 9-4:
Teledyne Analytical Instruments
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Table of Contents
Figure 10-1:
Figure 11-1
GPT Calibration System ............................................................................................................229
Replacing the Particulate Filter..................................................................................................242
Particle Filter on O3 Supply Air Dryer ........................................................................................243
Ozone Cleanser Assembly ........................................................................................................245
Zero Air Scrubber Assembly......................................................................................................251
NO2 Converter Assembly...........................................................................................................253
Reaction Cell Assembly.............................................................................................................255
Critical Flow Orifice Assembly ...................................................................................................256
Example of Signal I/O Function .................................................................................................268
CPU Status Indicator .................................................................................................................270
Relay PCA Status LEDS Used for Troubleshooting..................................................................271
Location of DC Power Test Points on Relay PCA.....................................................................284
Typical Set Up of Status Output Test ........................................................................................287
Pressure / Flow Sensor Assembly.............................................................................................289
Setup for determining NO2 NO Efficiency – 9110T Base Configuration...............................293
Pre-Amplifier Board Layout........................................................................................................311
9110T Sensor Assembly............................................................................................................313
Relay PCA with AC Relay Retainer In Place.............................................................................315
Relay PCA Mounting Screw Locations.....................................................................................316
Reaction Cell with PMT Tube and Optical Filter........................................................................321
9110T Sensitivity Spectrum.......................................................................................................321
NO2 NO Conversion...............................................................................................................322
Pneumatic Flow During the Auto Zero Cycle.............................................................................323
Internal Gas Flow for Basic 9110T with External Pump ............................................................327
Basic Internal Gas Flow for Basic 9110T with Internal Pump ...................................................328
Vacuum Manifold, Standard Configuration................................................................................329
Flow Control Assembly & Critical Flow Orifice ..........................................................................330
Location of Flow Control Assemblies & Critical Flow Orifices...................................................331
Ozone Generator Principle ........................................................................................................333
Semi-Permeable Membrane Drying Process ............................................................................334
9110T Perma Pure® Dryer.........................................................................................................334
9110T Electronic Block Diagram ...............................................................................................338
CPU Board.................................................................................................................................340
Relay PCA Layout (P/N 045230100).........................................................................................345
Relay PCA P/N 045230100 with AC Relay Retainer in Place...................................................346
Status LED Locations – Relay PCA...........................................................................................348
Heater Control Loop Block Diagram..........................................................................................349
Thermocouple Configuration Jumper (JP5) Pin-Outs................................................................351
9110T Sensor Module Assembly...............................................................................................352
Basic PMT Design .....................................................................................................................353
PMT Preamp Block Diagram .....................................................................................................354
Typical Thermo-Electric Cooler .................................................................................................355
PMT Cooling System Block Diagram.........................................................................................356
Power Distribution Block Diagram .............................................................................................358
Location of AC power Configuration Jumpers...........................................................................359
Pump AC Power Jumpers (JP7)................................................................................................360
Typical Set Up of AC Heater Jumper Set (JP2) ........................................................................361
Typical Jumper Set (JP2) Set Up of Heaters............................................................................362
Front Panel and Display Interface Block Diagram.....................................................................363
Basic Software Operation..........................................................................................................364
Triboelectric Charging................................................................................................................367
Basic anti-ESD Workbench .......................................................................................................370
Figure 11-2:
Figure 11-3:
Figure 11-4:
Figure 11-5:
Figure 11-6:
Figure 11-7:
Figure 12-1:
Figure 12-2:
Figure 12-3:
Figure 12-4:
Figure 12-5:
Figure 12-6:
Figure 12-7:
Figure 12-8:
Figure 12-9:
Figure 12-10:
Figure 12-11:
Figure 13-1:
Figure 13-2:
Figure 13-3:
Figure 13-4:
Figure 13-5:
Figure 13-6:
Figure 13-7.
Figure 13-8:
Figure 13-9:
Figure 13-10:
Figure 13-11:
Figure 13-12:
Figure 13-13:
Figure 13-14:
Figure 13-15:
Figure 13-16:
Figure 13-17:
Figure 13-18:
Figure 13-19:
Figure 13-20:
Figure 13-21:
Figure 13-22:
Figure 13-23:
Figure 13-24:
Figure 13-25:
Figure 13-26:
Figure 13-27:
Figure 13-28:
Figure 13-29:
Figure 13-30:
Figure 13-31:
Figure 14-1:
Figure 14-2:
Teledyne Analytical Instruments
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Table of Contents
TABLES
Table 1-1.
Table 2-2:
Table 3-1:
Table 3-5:
Table 3-6:
Table 3-7:
Table 3-8:
Table 3-9:
Table 3-10:
Table 3-11:
Table 3-12:
Table 4-1:
Table 4-2:
Table 4-3:
Table 4-4:
Table 4-5:
Table 5-1:
Table 5-2:
Table 5-3:
Table 5-4:
Table 5-5:
Table 5-6:
Table 5-7:
Table 5-8:
Table 5-9:
Table 6-1:
Table 6-2:
Table 6-4:
Table 6-5:
Table 6-6:
Table 7-1:
Table 7-2:
Table 7-3:
Table 8-1:
Table 8-2:
Table 9-1:
Table 9-2:
Table 9-3:
Table 9-4:
Table 9-5:
Table 10-1:
Table 10-2:
Table 10-3:
Table 11-1:
Table 11-2:
Table 12-1:
Table 12-2:
Table 12-3:
Table 12-4:
Table 12-5:
Table 12-6:
Table 12-7:
Table 12-8:
Table 12-9:
Table 12-10:
Table 12-11:
Analyzer Options..........................................................................................................................24
Software Settings for EPA Equivalence.......................................................................................28
Ventilation Clearance...................................................................................................................32
Analog Output Pin Assignments..................................................................................................43
Status Output Pin Assignments...................................................................................................45
Control Input Pin Assignments ....................................................................................................46
NIST-SRM's Available for Traceability of NOx Calibration Gases ................................................57
Zero/Span Valves Operating States OPT 50A ............................................................................65
Valve Operating States OPT 50B installed..................................................................................69
Internal Span Gas Generator Valve Operating States OPT 50G................................................71
Possible Warning Messages at Start-Up.....................................................................................76
Analyzer Operating Modes ..........................................................................................................88
Test Functions Defined................................................................................................................88
Warning Messages Defined.........................................................................................................91
Primary Setup Mode Features and Functions.............................................................................93
Secondary Setup Mode Features and Functions ........................................................................93
IND Mode Analog Output Assignments.......................................................................................99
Password Levels........................................................................................................................106
Variable Names (VARS)............................................................................................................112
Diagnostic Mode (DIAG) Functions ...........................................................................................114
DIAG - Analog I/O Functions .....................................................................................................117
Analog Output Voltage Range Min/Max ....................................................................................119
Voltage Tolerances for the TEST CHANNEL Calibration..........................................................123
Current Loop Output Check.......................................................................................................127
Test Channels Functions available on the 9110T’s Analog Output...........................................132
COMM Port Communication Modes..........................................................................................135
Ethernet Status Indicators .........................................................................................................140
RS-232 Communication Parameters for Hessen Protocol ........................................................151
Teledyne's Hessen Protocol Response Modes.........................................................................154
Default Hessen Status Flag Assignments .................................................................................159
Front Panel LED Status Indicators for DAS...............................................................................163
DAS Data Channel Properties...................................................................................................164
DAS Data Parameter Functions ................................................................................................172
Terminal Mode Software Commands ........................................................................................186
Teledyne API's Serial I/O Command Types ..............................................................................186
IZS Option Valve States with CAL_ON_NO2 Turned ON..........................................................203
AUTOCAL Modes ......................................................................................................................213
AutoCal Attribute Setup Parameters..........................................................................................214
Example AutoCal Sequence......................................................................................................215
Calibration Data Quality Evaluation...........................................................................................219
Activity Matrix for EPA Calibration Equipment and Supplies.....................................................222
Definition of Level 1 and Level 2 Zero and Span Checks .........................................................224
Activity Matrix for Data Quality...................................................................................................233
9110T Maintenance Schedule...................................................................................................240
Predictive Uses for Test Functions............................................................................................241
Front Panel Warning Messages ................................................................................................264
Test Functions - Indicated Failures............................................................................................266
Test Channel Outputs as Diagnostic Tools ...............................................................................269
Relay PCA Watchdog LED Failure Indications..........................................................................270
Relay PCA Status LED Failure Indications................................................................................271
DC Power Test Point and Wiring Color Codes..........................................................................284
DC Power Supply Acceptable Levels ........................................................................................284
Relay PCA Control Devices.......................................................................................................285
Analog Output Test Function - Nominal Values Voltage Outputs .............................................286
Status Outputs Check................................................................................................................288
9110T Control Input Pin Assignments and Corresponding Signal I/O Functions......................288
Teledyne Analytical Instruments
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Table of Contents
Table 13-1:
Table 13-2:
Table 13-3:
Table 13-4:
Table 13-5:
Table 13-6:
Table 13-7:
Table 13-8:
Table 14-1:
Table 14-2:
List of Interferents ......................................................................................................................325
9110T Valve Cycle Phases........................................................................................................329
9110T Gas Flow Rates..............................................................................................................331
Relay PCA Status LED’s ...........................................................................................................347
Thermocouple Configuration Jumper (JP5) Pin-Outs................................................................350
AC Power Configuration for Internal Pumps (JP7)....................................................................360
Power Configuration for Standard AC Heaters (JP2)................................................................361
Power Configuration for Optional Heaters (JP6) .......................................................................362
Static Generation Voltages for Typical Activities.......................................................................367
Sensitivity of Electronic Devices to Damage by ESD................................................................368
Teledyne Analytical Instruments
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Table of Contents
This page intentionally left blank.
Teledyne Analytical Instruments
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Part I
PART I
GENERAL INFORMATION
Teledyne Analytical Instruments
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Part I
This page intentionally left blank.
Teledyne Analytical Instruments
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Introduction
1. INTRODUCTION, FEATURES AND OPTIONS
1.1. OVERVIEW
The Model 9110T Nitrogen Oxides Analyzer uses chemiluminescence detection, coupled with
state-of-the-art microprocessor technology to provide the sensitivity, stability and ease of use
needed for ambient or dilution CEM monitoring requirements of nitric oxide (NO), nitrogen
dioxide (NO2) and the total nitrogen oxides (NOx). The instrument:
Calculates the amount of NO present by measuring the amount of chemiluminescence given off
when the sample gas is exposed to ozone (O3).
Uses a catalytic-reactive converter to convert any NO2 in the sample gas to NO, which is then
measured as above (including the original NO in the sample gas) and reported as NOx.
Since the density of the sample gas effects the brightness of the chemiluminescence reaction, the 9110T
software compensates for temperature and pressure changes.
Stability is further enhanced by an Auto-Zero feature which periodically redirects the gas flow through
the analyzer so that no chemiluminescence reaction is present in the sample chamber. The analyzer
measures this “dark” condition and uses the results as an offset, which is subtracted from the sensor
readings recorded while the instrument is measuring NO and NOX. The result gives a sensitive, accurate,
and dependable performance under the harshest operating conditions.
The 9110T analyzer’s multi-tasking software gives the ability to track and report a large number of
operational parameters in real time. These readings are compared to diagnostic limits kept in the
analyzers memory and should any fall outside of those limits the analyzer issues automatic warnings.
Built-in data acquisition capability, using the analyzer's internal memory, allows the logging of multiple
parameters including averaged or instantaneous concentration values, calibration data, and operating
parameters such as pressure and flow rate. Stored data are easily retrieved through the rear panel serial
or Ethernet ports via our APICOM software or from the front panel, allowing operators to perform
predictive diagnostics and enhanced data analysis by tracking parameter trends. Multiple averaging
periods of one minute to 365 days are available for over a period of one year.
1.2. FEATURES
Some of the other exceptional features of your 9110T Nitrogen Oxides Analyzer are:
Ranges, 0-50 ppb to 0-20 ppm, user selectable
Independent ranges and auto ranging
Large, vivid, and durable graphics display with touch screen interface
Microprocessor controlled for versatility
Multi-tasking software to allow viewing test variables while operating
Continuous self checking with alarms
Permeation dryer on ozone generator
Bi-directional RS-232, optional USB and RS-485, and 10/100Base-T Ethernet ports for remote
operation
Front panel USB ports for peripheral devices and firmware upgrades
Digital status outputs to provide instrument operating condition
Adaptive signal filtering to optimize response time
Teledyne Analytical Instruments
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Introduction
Temperature and pressure compensation
Converter efficiency correction software
Catalytic ozone destruct
Comprehensive internal data logging with programmable averaging periods
Ability to log virtually any operating parameter
8 analog inputs (optional)
Internal zero and span check (optional)
1.3. DOCUMENTATION
In addition to this operation manual, two other manuals are available for download from Teledyne’s
website at http://www.teledyne-api.com/manuals/, to support the operation of the instrument:
APICOM Software Manual, part number 03945
DAS Manual, part number 02837
1.4. OPTIONS
The options available for your analyzer are present in with name, option number, a description and/or
comments, and if applicable, cross-references to technical details in this manual, such as setup and
calibration. To order these options or to learn more about them, please contact Teledyne Sales
Department at:
TOLL-FREE:
PHONE:
FAX:
888-789-8168
+1 626-934-1500
+1 626-961-2538
EMAIL:
http://www.teledyne-ai.com/
WEBSITE:
Table 1-1. Analyzer Options
Option
Number
Option
Pumps
Description/Notes
Reference
Pumps meet all typical AC power supply standards while exhibiting same
pneumatic performance.
11A
11B
12A
12B
12C
Ship without pump
N/A
N/A
N/A
N/A
N/A
Pumpless Pump Pack
Internal Pump 115V @ 60 Hz
Internal Pump 220V @ 60 Hz
Internal Pump 220V @ 50 Hz
Rack Mount
Kits
Options for mounting the analyzer in standard 19” racks
20A
20B
21
Rack mount brackets with 26 in. (660 mm) chassis slides
Rack mount brackets with 24 in. (610 mm) chassis slides
N/A
N/A
Rack mount brackets only (compatible with carrying strap, Option 29) N/A
23
Rack mount for external pump pack (no slides)
N/A
Teledyne Analytical Instruments
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Introduction
Option
Number
Option
Description/Notes
Reference
Carrying Strap/Handle
Side-mounted strap for hand-carrying analyzer
Extends from “flat” position to accommodate hand for carrying.
Recesses to 9mm (3/8”) dimension for storage.
Can be used with rack mount brackets, Option 21.
Cannot be used with rack mount slides.
29
N/A
CAUTION – GENERAL SAFETY HAZARD
THE 9110T ANALYZER WEIGHS ABOUT 18 KG (40 POUNDS).
TO AVOID PERSONAL INJURY WE RECOMMEND THAT TWO PERSONS LIFT AND CARRY THE
ANALYZER. DISCONNECT ALL CABLES AND TUBING FROM THE ANALYZER BEFORE MOVING IT.
Used for connecting external voltage signals from other instrumentation (such as
meteorological instruments).
Analog Input and USB port
64B
Current Loop Analog
Also can be used for logging these signals in the analyzer’s internal
DAS
Sections 3.3.1.2,
and 7
Adds isolated, voltage-to-current conversion circuitry to the analyzer’s analog
outputs.
Outputs
Can be configured for any output range between 0 and 20 mA.
May be ordered separately for any of the analog outputs.
Can be installed at the factory or retrofitted in the field.
Sections 3.3.1.4
and 5.9.3.7
41
Parts Kits
Spare parts and expendables
Expendables Kit includes a recommended set of expendables for
one year of operation of this instrument including replacement
sample particulate filters.
42A
Appendix B
Expendables Kit with IZS includes the items needed to refurbish
the zero air scrubber.
43
45
Appendix B
Appendix B
Spare Parts Kit includes spares parts for one unit.
Used to control the flow of calibration gases generated from external sources,
rather than manually switching the rear panel pneumatic connections.
Calibration Valves
AMBIENT ZERO AND AMBIENT SPAN VALVES
50A
50B
Zero Air and Span Gas input supplied at ambient pressure.
Gases controlled by 2 internal valves; SAMPLE/CAL & ZERO/SPAN.
Section 3.3.2.3
Section 3.3.2.4
AMBIENT ZERO AND PRESSURIZED SPAN VALVES
Span Gas input from external, pressurized source;
Span Gas flow rate maintained at 1 ATM by critical flow orifice & vented
through Vent port.
Shutoff valve stops flow of Span Gas when in sample mode to preserve
pressurized gas source.
Zero Air created via 2-stage scrubber & dry filter unit (DFU).
Gases controlled by 2 internal valves; SAMPLE/CAL & ZERO/SPAN.
ZERO SCRUBBER AND INTERNAL SPAN SOURCE (IZS)
Span Gas generated from internal NO2 permeation tube
Zero Air created by 2-stage scrubber & DFU.
Sections 3.3.2.5
and 3.3.2.6
50G
Gases controlled by 2 internal valves: Sample/Cal & Zero/Span.
NO2 Permeation Tubes
Replacement tubes; identical size/shape; different permeation rates.
Permeation Rate
Approximate NO2 Concentration @ 50°C
( 25%)
421 ng/min
842 ng/min
52B
52G
N/A
N/A
300ppb – 500 ppb 25%
0600 – 1000 ppb 25%
Each tube comes with a calibration certificate, traceable to a NIST standard,
specifying its actual effusion rate of that tube to within ± 5% @ 0.56 liters per
Section 3.3.2.5
Teledyne Analytical Instruments
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Introduction
Option
Number
Option
Description/Notes
Reference
minute, calibration performed at a tube temperature of 50°C.
Communication Cables
For remote serial, network and Internet communication with the analyzer.
Type
Description
Shielded, straight-through DB-9F to DB-25M cable, about
1.8 m long. Used to interface with older computers or
code activated switches with DB-25 serial connectors.
60A
RS-232
Section 3.3.1.8
Shielded, straight-through DB-9F to DB-9F cable of about
1.8 m length.
60B
60C
RS-232
Ethernet
USB
Section 3.3.1.8
Section 3.3.1.8
Section 3.3.1.8
Patch cable, 2 meters long, used for Internet and LAN
communications.
Cable for direct connection between instrument (rear
panel USB port) and personal computer.
60D
USB Port
For remote connection
For connection to personal computer. (Separate option only when
Option 64B, Analog Input and USB Com Port not elected).
Sections 3.3.1.8
and 6.6
64A
Concentration Alarm Relays
Issues warning when gas concentration exceeds limits set by user.
Four (4) “dry contact” relays on the rear panel of the instrument. This
relay option is different from and in addition to the “Contact Closures”
that come standard on all TAPI instruments.
Section 3.3.1.7
61
RS-232 Multidrop
62
Enables communications between host computer and up to eight analyzers.
Multidrop card seated on the analyzer’s CPU card.
Section 3.3.1.8
Each instrument in the multidrop network requres this card and a
communications cable (Option 60B).
Other Gas Options
Second gas sensor and gas conditioners
Oxygen (O2) Sensor
65A
86A
Figure 3-6
Ammonia removal sample conditioner (required for EN Certification)
3.3.2.6, 3.4.4.1
Figure 3-29,
Sections 3.3.2.6,
9.1.2.1, 11.3.7, and
11.3.7.1, Table 11-1
86C
External zero air scrubber
Special Features
Built in features, software activated
Maintenance Mode Switch, located inside the instrument, places
the analyzer in maintenance mode where it can continue sampling,
yet ignore calibration, diagnostic, and reset instrument commands.
This feature is of particular use for instruments connected to
Multidrop or Hessen protocol networks.
N/A
N/A
Call Customer Service for activation.
Second Language Switch activates an alternate set of display
messages in a language other than the instrument’s default
language.
Call Customer Service for a specially programmed Disk on Module containing
the second language.
N/A
N/A
N/A
Dilution Ratio Option allows the user to compensate for diluted
sample gas, such as in continuous emission monitoring (CEM) where
the quality of gas in a smoke stack is being tested and the sampling
method used to remove the gas from the stack dilutes the gas.
Call Customer Service for activation.
Section 5.4.3.5
Teledyne Analytical Instruments
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Specifications, Approvals, & Compliance
2. SPECIFICATIONS, APPROVALS, & COMPLIANCE
This section presents specifications for the 9110T, Agency approvals, EPA designation, and CE mark
and safety compliance..
2.1. SPECIFICATIONS
Table 2-1 presents the instrument’s parameters and the specifications that each meets.
Table 2-1:
9110T Basic Unit Specifications
Parameter
Specification
Min/Max Range
(Physical Analog Output)
Min: 0-50 ppb Full Scale
Max: 0-20,000 ppb Full Scale (selectable, independent NO, NO2, NOx ranges and
auto ranges supported)
Measurement Units
Zero Noise1
ppb, ppm, µg/m3, mg/m3 (selectable)
< 0.2 ppb (RMS)
Span Noise1
Lower Detectable Limit 2
Zero Drift
< 0.5% of reading (RMS) above 50 ppb or 0.2 ppb, whichever is greater
0.4 ppb
< 0.5 ppb (at constant temperature and voltage) /24 hours
< 0.5% of Full Scale (at constant temperature and voltage) /24 hours
20 seconds
Span Drift
Lag Time1
Rise/Fall Time1
<60 seconds to 95%
Linearity
1% of full scale / 24 hours
Precision
0.5% of reading above 50 ppb
Sample Flow Rate
AC Power
500 cm3/min ± 10%
100-120 V, 60 Hz (3.0 A); 220-240 V, 50 Hz (2.5 A)
Power, Ext Pump
100 V, 50/60 Hz (3.25 A); 115 V, 60 Hz (3.0 A);
220-240 V, 50/60 Hz (2.5 A)
10V, 5V, 1V, 0.1V (selectable)
All Ranges with 5% Under/Over Range
Analog Output Ranges
Analog Output Resolution
Recorder Offset
1 part in 4096 of selected full-scale voltage
± 10%
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs (7 defined, 1 spare)
6 opto-isolated digital control inputs (4 defined, 2 spar)
4 analog outputs
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
Dimensions H x W x D
Weight
Analyzer: 40 lbs (18 kg)
Teledyne Analytical Instruments
27
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Specifications, Approvals, & Compliance
Parameter
Specification
External Pump Pack: 15 lbs (7 kg)
Operating Temperature Range
5 - 40 C (with EPA equivalency)
Humidity Range
0-95% RH non-condensing
Environmental Conditions
1 As defined by the US EPA.
Installation Category (Over voltage Category) II Pollution Degree 2
2 Defined as twice the zero noise level by the US EPA.
2.2. EPA EQUIVALENCY DESIGNATION
TAI’s 9110T nitrogen oxides analyzer is designated as a reference method (Number RFNA-1104-099)
for NO2 measurement, as defined in 40 CFR Part 53, when operated under the following conditions:
Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).
Ambient temperature range of 5 to 40°C.
With 1-micron PTFE filter element installed in the internal filter assembly.
Equipped with ozone supply air filter
Gas flow supplied by External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute
(slpm) or better.
Software Settings, see Table 2-2:
Table 2-2: Software Settings for EPA Equivalence
Parameter
Setting
Dynamic Zero
OFF or ON
OFF
Dynamic Span
CAL-on-NO2
OFF
Dilution Factor
Temp/Pres compensation
AutoCal
1.0 or OFF
ON
ON or OFF
ON or OFF
ON or OFF
Independent range
Auto range
Under the designation, the Analyzer may be operated with or without the following options:
Rack mount with or without slides.
Rack mount for external pump.
4-20mA isolated analog outputs.
Zero/Span Valves option.
Internal Zero/Span (IZS) option with:
NO2 permeation tube - 0.4 ppm at 0.7 liter per minute; certified/uncertified.
NO2 permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.
Teledyne Analytical Instruments
28
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Specifications, Approvals, & Compliance
Note
Under the designation, the IZS option cannot be used as the source of
calibration.
2.3. APPROVALS AND CERTIFICATIONS
The TAI Model 9110T analyzer was tested and certified for Safety and Electromagnetic Compatibility
(EMC). This section presents the compliance statements for those requirements and directives.
2.3.1. SAFETY
IEC 61010-1:2001, Safety requirements for electrical equipment for measurement,
control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive
North American:
cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04
NEMKO-CCL (US): UL No. 61010-1 (2nd Edition)
2.3.2.
EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE: 2004/108/EC, Electromagnetic Compatibility Directive
2.3.3. OTHER TYPE CERTIFICATIONS
MCERTS: Sira MC 050068/05
For additional certifications, please contact Customer Service:
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: [email protected]
Teledyne Analytical Instruments
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Specifications, Approvals, & Compliance
This page intentionally left blank.
Teledyne Analytical Instruments
30
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3. GETTING STARTED
This section addresses the procedures for unpacking the instrument and inspecting for damage, presents
clearance specifications for proper ventilation, introduces the instrument layout, then presents the
procedures for getting started: making electrical and pneumatic connections, and conducting an initial
calibration check.
3.1. UNPACKING THE 9110T ANALYZER
CAUTION
GENERAL SAFETY HAZARD
To avoid personal injury, always use two persons to lift and carry the 9110T.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Printed Circuit Assemblies (PCAs) are sensitive to electro-static
discharges too small to be felt by the human nervous system. Failure to
use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 14 for more information on
preventing ESD damage.
CAUTION!
Do not operate this instrument until you’ve removed dust plugs from SAMPLE
and EXHAUST ports on the rear panel.
Note
TAI recommends that you store shipping containers/materials for future
use if/when the instrument should be returned to the factory for repair
and/or calibration service. See Warranty section in this manual and
shipping procedures on our Website at http://www.teledyne-api.com
under Customer Support > Return Authorization.
Verify that there is no apparent external shipping damage. If damage has occurred, please advise the
shipper first, then TAI.
Included with your analyzer is a printed record of the final performance characterization performed on
your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04490) is
an important quality assurance and calibration record for this instrument. It should be placed in the
quality records file for this instrument.
Teledyne Analytical Instruments
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
With no power to the unit, carefully removed the top cover of the analyzer and check for internal
shipping damage by carrying out the following steps:
1. Carefully remove the top cover of the analyzer and check for internal shipping
damage.
2. Remove the setscrew located in the top, center of the Front panel.
3. Slide the cover backwards until it clears the analyzer’s front bezel.
4. Lift the cover straight up.
5. Inspect the interior of the instrument to ensure all circuit boards and other
components are in good shape and properly seated.
6. Check the connectors of the various internal wiring harnesses and pneumatic hoses
to ensure they are firmly and properly seated.
7. Verify that all of the optional hardware ordered with the unit has been installed.
These are listed on the paperwork accompanying the analyzer.
WARNING – ELECTRICAL SHOCK HAZARD
Never disconnect PCAs, wiring harnesses or electronic subassemblies
while under power.
3.1.1. VENTILATION CLEARANCE
Whether the analyzer is set up on a bench or installed into an instrument rack, be sure to leave sufficient
ventilation clearance.
Table 3-1: Ventilation Clearance
MINIMUM REQUIRED
AREA
CLEARANCE
Back of the instrument
Sides of the instrument
10 cm / 4 in
2.5 cm / 1 in
2.5 cm / 1 in
Above and below the instrument
Various rack mount kits are available for this analyzer. Refer to Section 1.4 of this manual for more
information.
3.2. INSTRUMENT LAYOUT
Instrument layout shows front panel and display, rear panel connectors, and internal chassis layout.
Teledyne Analytical Instruments
32
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.2.1. FRONT PANEL
Figure 3-1 shows the analyzer’s front panel layout, followed by a close-up of the display screen in
Figure 3-2, which is described in Table 3-2. The two USB ports on the front panel are provided for the
connection of peripheral devices:
plug-in mouse (not included) to be used as an alternative to the touchscreen interface
thumb drive (not included) to download updates to instruction software (contact TAPI Customer
Service for information).
Figure 3-1:
Front Panel Layout
Teledyne Analytical Instruments
33
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-2:
Display Screen and Touch Control
The front panel liquid crystal display screen includes touch control. Upon analyzer start-up, the screen
shows a splash screen and other initialization indicators before the main display appears, similar to
Figure 3-2 above. The LEDs on the display screen indicate the Sample, Calibration and Fault states; also
on the screen is the gas concentration field (Conc), which displays real-time readouts for the primary
gases, NO, NO2, and NOx, and for the secondary gas if installed. The display screen also shows what
mode the analyzer is currently in, as well as messages and data (Param). Along the bottom of the screen
is a row of touch control buttons; only those that are currently applicable will have a label. Table 3-2
provides detailed information for each component of the screen.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Do not use hard-surfaced instruments such as pens to touch the control
buttons.
Teledyne Analytical Instruments
34
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Table 3-2:
Field
Display Screen and Touch Control Description
Description/Function
Status
LEDs indicating the states of Sample, Calibration and Fault, as follows:
Name
Color
State
Off
Definition
Unit is not operating in sample mode, DAS is disabled.
On
Sample Mode active; Front Panel Display being updated; DAS data
being stored.
SAMPLE Green
Unit is operating in sample mode, front panel display being updated,
DAS hold-off mode is ON, DAS disabled
Blinking
Off
Auto Cal disabled
CAL
Yellow
Red
On
Blinking
Auto Cal enabled
Unit is in calibration mode
Off
Blinking
No warnings exist
Warnings exist
FAULT
Displays the actual concentration of the sample gas currently being measured by the analyzer in the
currently selected units of measure.
Conc
Mode
Displays the name of the analyzer’s current operating mode
Displays a variety of informational messages such as warning messages, operational data, test function
values and response messages during interactive tasks.
Param
Control Buttons
Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.
Figure 3-3 shows how the front panel display is mapped to the menu charts illustrated in this manual.
The Mode, Param (parameters), and Conc (gas concentration) fields in the display screen are
represented across the top row of each menu chart. The eight touch control buttons along the bottom of
the display screen are represented in the bottom row of each menu chart.
Teledyne Analytical Instruments
35
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-3:
Display/Touch Control Screen Mapped to Menu Charts
Note
The menu charts in this manual contain condensed representations of the
analyzer’s display during the various operations being described. These
menu charts are not intended to be exact visual representations of the
actual display.
Teledyne Analytical Instruments
36
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.2.2. REAR PANEL
Figure 3-4:
Rear Panel Layout – Base Unit
Table 3-3 provides a description of each component on the rear panel.
Teledyne Analytical Instruments
37
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Table 3-3:
Component
cooling fan
Rear Panel Description
Function
Pulls ambient air into chassis through side vents and exhausts through rear.
Connector for three-prong cord to apply AC power to the analyzer.
AC power
connector
CAUTION! The cord’s power specifications (specs) MUST comply with the power
specs on the analyzer’s rear panel Model number label
Identifies the analyzer model number and provides power specs
Model/specs label
TO CONV
(not used)
(not used)
(not used)
(not used)
FROM CONV
MULTI
TO DRYER
Outlet for internal sample gas dryer; connect to external zero air scrubber (for IZS options
only).
FROM DRYER
Connect a gas line from the source of sample gas here.
SAMPLE
Calibration gases can also enter here on units without zero/span/shutoff valve options
installed.
Connect an exhaust gas line of not more than 10 meters long here that leads outside the
shelter or immediate area surrounding the instrument. The line must be ¼” tubing or
greater.
EXHAUST
On units with zero/span/shutoff valves option installed, connect a gas line to the source of
calibrated span gas here.
SPAN 1
SPAN2/VENT
ZERO AIR
On units with pressurized span valve option, used for venting.
Internal Zero Air: On units with zero/span/shutoff valves option installed but no internal
zero air scrubber attach a gas line to the source of zero air here.
LEDs indicate receive (RX) and transmit (TX) activity on the when blinking.
Serial communications port for RS-232 or RS-485.
RX TX
COM 2
RS-232
Serial communications port for RS-232 only.
Switch to select either data terminal equipment or data communication equipment during
RS-232 communication.
DCE DTE
For ouputs to devices such as Programmable Logic Controllers (PLCs).
For voltage or current loop outputs to a strip chart recorder and/or a data logger.
For remotely activating the zero and span calibration modes.
STATUS
ANALOG OUT
CONTROL IN
ALARM
Option for concentration alarms and system warnings.
Connector for network or Internet remote communication, using Ethernet cable
ETHERNET
Option for external voltage signals from other instrumentation and for logging these
signals
ANALOG IN
Connector for direct connection to laptop computer, using USB cable.
Includes voltage and frequency specifications
USB
Model Label
3.2.3. INTERNAL CHASSIS LAYOUT
Figure 3-5 and Figure 3-6 show internal chassis configurations with different options.
Teledyne Analytical Instruments
38
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Figure 3-5:
Internal Layout – Top View with IZS Option
Teledyne Analytical Instruments
39
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-6:
Internal Layout - Top View Showing Other Options
Teledyne Analytical Instruments
40
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.3. CONNECTIONS AND SETUP
This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) connections for setup
and preparing for instrument operation.
3.3.1. ELECTRICAL CONNECTIONS
Note
To maintain compliance with EMC standards, it is required that the cable
length be no greater than 3 meters for all I/O connections, which include
Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232,
and RS-485.
This section presents the electrical connections for AC power and communications.
3.3.1.1. Connecting Power
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A
current at your AC voltage and that it is equipped with a functioning earth ground.
WARNING
ELECTRICAL SHOCK HAZARD
High Voltages are present inside the analyzers case.
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting electrical subassemblies.
Do not operate with cover off.
CAUTION
GENERAL SAFETY HAZARD
The 9110T analyzer can be configured for both
100-130 V and 210-240 V at either 47 or 63 Hz.
To avoid damage to your analyzer, ensure that the AC power voltage
matches the voltage indicated on the analyzer’s model/specs label (Figure
3-4) before plugging the 9110T into line power.
3.3.1.2. Connecting Analog Inputs (Option)
The Analog In connector is used for connecting external voltage signals from other instrumentation
(such as meteorological instruments) and for logging these signals in the analyzer’s internal DAS. The
input voltage range for each analog input is 0-10 VDC.
Teledyne Analytical Instruments
41
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-7:
Analog In Connector
Pin assignments for the Analog In connector are presented in Table 3-4.
Table 3-4: Analog Input Pin Assignments
DAS
PIN
DESCRIPTION
PARAMETER1
1
Analog input # 1
Analog input # 2
Analog input # 3
Analog input # 4
Analog input # 5
Analog input # 6
Analog input # 7
Analog input # 8
Analog input Ground
AIN 1
AIN 2
AIN 3
AIN 4
AIN 5
AIN 6
AIN 7
AIN 8
N/A
2
3
4
5
6
7
8
GND
1 See Section 7 for details on setting up the DAS.
3.3.1.3. Connecting Analog Outputs
The 9110T is equipped with several analog output channels accessible through a connector on the back
panel of the instrument.
Output channels A1, A2 and A3 are assigned to the NOx, NO and NO2 concentration signals of the
analyzer.
The default analog output voltage setting of these channels is 0 to 5 VDC with a reporting
range of 0 to 500 ppb.
An optional Current Loop output is available for each.
The output labeled A4 is special. It can be set by the user to output any one a variety of diagnostic test
functions (see Section 5.9.4).
The default analog output voltage setting of these channels is also 0 to 5 VDC.
See Section 5.9.4 for a list of available functions and their associated reporting range.
There is no optional Current Loop output available for Channel A4.
To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output
connections on the rear panel of the analyzer. Pin-outs for the analog output connector are:
Teledyne Analytical Instruments
42
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
ANALOG OUT
A2 A3
A1
A4
+
-
+
-
+
-
+
-
Figure 3-8:
Analog Output Connector
Table 3-5: Analog Output Pin Assignments
STANDARD
VOLTAGE OUTPUT LOOP OPTION
CURRENT
PIN
ANALOG OUTPUT
SIGNAL
1
2
3
4
3
4
7
8
V Out
Ground
V Out
I Out +
I Out -
NOx Concentration
A1
A2
I Out +
NO Concentration
NO2 Concentration
TEST CHANNEL
Ground
V Out
I Out -
I Out +
A3
Ground
V Out
I Out -
Not Available
Not Available
A41
Ground
To change the settings for the analog output channels, see Section 5.9.2.
3.3.1.4. Current Loop Analog Outputs (Option 41) Setup
If your analyzer had this option installed at the factory, there are no further connectons to be made.
Otherwise, it can be installed as a retrofit for each of the analog outputs of the analyzer . This option
converts the DC voltage analog output to a current signal with 0-20 mA output current. The outputs can
be scaled to any set of limits within that 0-20 mA range. However, most current loop applications call
for either 2-20 mA or 4-20 mA range. All current loop outputs have a +5% over-range. Ranges with the
lower limit set to more than 1 mA (e.g., 2-20 or 4-20 mA) also have a -5% under-range.
Figure 3-9 provides installation instructions and illustrates a sample combination of one current output
and two voltage outputs configuration. This section provides instructions for converting current loop
analog outputs to standard 0-to-5 VDC outputs. Information on calibrating or adjusting these outputs can
be found in Section 5.9.3.7.
CAUTION – AVOID INVALIDATING WARRANTY
Servicing or handling of circuit components requires electrostatic
discharge protection, i.e. ESD grounding straps, mats and containers.
Failure to use ESD protection when working with electronic assemblies will
void the instrument warranty. Refer to Section 14 for more information on
preventing ESD damage.
Teledyne Analytical Instruments
43
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-9:
Current Loop Option Installed on the Motherboard
CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD VOLTAGE OUTPUTS
To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation:
1. Turn off power to the analyzer.
8. If a recording device was connected to the output being modified, disconnect it.
9. Remove the top cover.
Remove the set screw located in the top, center of the rear panel.
Remove the screws fastening the top cover to the unit (one per side).
Slide the cover back and lift the cover straight up.
10. Remove the screw holding the current loop option to the motherboard.
11. Disconnect the current loop option PCA from the appropriate connector on the
motherboard (see Figure 3-9).
12. Each connector, J19 and J23, requires two shunts. Place one shunt on the two left
most pins and the second shunt on the two adjacent pins (see Figure 3-9).
13. Reattach the top case to the analyzer.
Teledyne Analytical Instruments
44
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
The analyzer is now ready to have a voltage-sensing, recording device attached to that output.
Calibrate the analog output as described in Section 5.9.3.2.
3.3.1.5. Connecting the Status Outputs
The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up to 50
mA of DC current. These outputs can be used interface with devices that accept logic-level digital
inputs, such as Programmable Logic Controllers (PLCs). Each Status bit is an open collector output that
can withstand up to 40 VDC. All of the emitters of these transistors are tied together and available at pin
D.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Most PLC’s have internal provisions for limiting the current that the input
will draw from an external device. When connecting to a unit that does
not have this feature, an external dropping resistor must be used to limit
the current through the transistor output to less than 50 mA. At 50 mA,
the transistor will drop approximately 1.2V from its collector to emitter.
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS
(Figure 3-4). Pin-outs for this connector are:
STATUS
1
2
3
4
5
6
7
8
D
+
+5V to external device
Figure 3-10:
Table 3-6: Status Output Pin Assignments
OUTPUT # STATUS DEFINITION CONDITION
Status Output Connector
1
On if no faults are present.
SYSTEM OK
CONC VALID
On if O3 concentration measurement is valid.
If the O3 concentration measurement is invalid, this bit is OFF.
2
3
4
5
On if unit is in high range of DUAL or AUTO Range Modes.
On whenever the instrument is in CALZ mode.
On whenever the instrument is in CALS mode.
HIGH RANGE
ZERO CAL
SPAN CAL
Teledyne Analytical Instruments
45
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
6
7-8
D
On whenever the instrument is in DIAGNOSTIC mode.
SPARE
DIAG MODE
Emitter BUS
The emitters of the transistors on pins 1 to 8 are bussed together.
SPARE
+
DC Power
+ 5 VDC, 300 mA source maximum
The ground level from the analyzer’s internal DC power supplies. This
connection should be used as the ground return when +5VDC power is used.
Digital Ground
3.3.1.6. Connecting the Control Inputs
The analyzer is equipped with three digital control inputs that can be used to remotely activate the zero
and span calibration modes (see Section 9.1.2.4). Access to these inputs is provided via a 10-pin
connector labeled CONTROL IN on the analyzer’s rear panel.
There are two methods for energizing the Control Inputs. The internal +5V available from the pin
labeled “+” is the most convenient method however, to ensure that these inputs are truly isolated; a
separate external 5 VDC power supply should be used.
CONTROL IN
CONTROL IN
A
B
C
D
E
F
U
+
A
B
C
D
E
F
U
+
+
-
5 VDC Power
Supply
External Power Connections
Local Power Connections
Figure 3-11:
Energizing the 9110T Control Inputs
Table 3-7: Control Input Pin Assignments
ON Condition
Status
Definition
Input #
REMOTE
ZERO CAL
The Analyzer is placed in Zero Calibration mode. The mode field of the display will
read ZERO CAL R.
A
B
REMOTE
SPAN CAL
The Analyzer is placed in Lo Span Calibration mode. The mode field of the display will
read SPAN CAL R.
C, D, E
& F
Spare
Teledyne Analytical Instruments
46
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
The ground level from the analyzer’s internal DC Power Supplies (same as chassis
ground).
Digital Ground
External Power
input
U
Input pin for +5 VDC required to activate pins A – F.
Internally generated 5V DC power. To activate inputs A – F, place a jumper between
this pin and the “U” pin. The maximum amperage through this port is 300 mA
(combined with the analog output supply, if used).
+
5 VDC output
3.3.1.7. Concentration Alarm Relay (Option 61)
The TAI “E” series analyzers have an option for four (4) “dry contact” relays on the rear panel of the
instrument. This relay option is different from and in addition to the “Contact Closures” that come
standard on all TAPI instruments. The relays have 3 pins that have connections on the rear panel (refer
Figure 3-12). They are a Common (C), a Normally Open (NO), & a Normally Closed (NC) pin.
Figure 3-12:
Concentration Alarm Relay
Alarm 1
“System OK 2”
Alarm 2“Conc 1”
Alarm 3“Conc 2”
Alarm 4“Range Bit”
“ALARM 1” RELAY
Alarm 1 which is “System OK 2” (system OK 1, is the status bit) is in the energized state when the
instrument is “OK” & there are no warnings. If there is a warning active or if the instrument is put into
the “DIAG” mode, Alarm 1 will change states. This alarm has “reverse logic” meaning that if you put a
meter across the Common & Normally Closed pins on the connector you will find that it is OPEN when
the instrument is OK. This is so that if the instrument should turn off or lose power, it will change states
& you can record this with a data logger or other recording device.
“ALARM 2” RELAY & “ALARM 3” RELAY
The “Alarm 2 Relay” on the rear panel, is associated with the “Concentration Alarm 1” set point in the
software & the “Alarm 3 Relay” on the rear panel is associated with the “Concentration Alarm 2” set
point in the software.
Alarm 2 Relay
NO Alarm 1 = xxx PPM
Teledyne Analytical Instruments
47
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Alarm 3 Relay
NO2 Alarm 2 = xxx PPM
NOX Alarm 1 = xxx PPM
NOX Alarm 2 = xxx PPM
Alarm 2 Relay
Alarm 3 Relay
The Alarm 2 Relay will be turned on any time the concentration set-point is exceeded & will return to its
normal state when the concentration value goes back below the concentration set-point.
Even though the relay on the rear panel is a NON-Latching alarm & resets when the concentration goes
back below the alarm set point, the warning on the front panel of the instrument will remain latched until
it is cleared. You can clear the warning on the front panel by either pushing the CLR button on the front
panel or through the serial port.
In instruments that sample more than one gas type, there could be more than one gas type triggering the
Concentration 1 Alarm (“Alarm 2” Relay). For example, the 9110T instrument can monitor both NO &
NO2 gas. The software for this instrument is flexible enough to allow you to configure the alarms so that
you can have 2 alarm levels for each gas.
NO Alarm 1 = 20 PPM
NO Alarm 2 = 100 PPM
NO2 Alarm 1 = 20 PPM
NO2 Alarm 2 = 100 PPM
In this example, NO Alarm 1 & NO2 Alarm 1 will both be associated with the “Alarm 2” relay on the
rear panel. This allows you do have multiple alarm levels for individual gases.
A more likely configuration for this would be to put one gas on the “Alarm 1” relay and the other gas on
the “Alarm 2” relay.
NO Alarm 1 = 20 PPM
NO Alarm 2 = Disabled
NO2 Alarm 1 = Disabled
NO2 Alarm 2 = 100 PPM
“ALARM 4” RELAY
This relay is connected to the “range bit”. If the instrument is configured for “Auto Range” and the
instrument goes up into the high range, it will turn this relay on.
3.3.1.8. Connecting the Communications Interfaces
The T-Series analyzers are equipped with connectors for remote communications interfaces: Ethernet,
USB, RS-232, RS-232 Multidrop and RS-485 (each described here). In addition to using the
Teledyne Analytical Instruments
48
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
appropriate cables, each type of communication method, must be configured using the SETUP>COMM
menu (see Sections 5.7 and 6).
ETHERNET CONNECTION
For network or Internet communication with the analyzer, connect an Ethernet cable from the analyzer’s
rear panel Ethernet interface connector to an Ethernet port. Although the analyzer is shipped with DHCP
enabled by default (Section 6.5.2), it should be manually assigned a static IP address.
Configuration: (manual, i.e., static) Section 6.5.1.
USB CONNECTION
The USB option can be used for direct communication between the analyzer and a PC; connect a USB
cable between the analyzer and computer USB ports. Baud rates must match: check the baud rate on
either the computer or the instrument and change the other to match (see Section 6.2.2). This USB
connection can only be used when the COM2 port is not in use except for RS-232 Multidrop
communication.
Configuration: Section 6.6.
Note
If this option is installed, the rear panel COM2 port cannot be used for
anything other than Multidrop communication.
RS-232 CONNECTION
For RS-232 communications with data terminal equipment (DTE) or with data communication
equipment (DCE) connect either a DB9-female-to-DB9-female cable (TAI part number WR000077) or a
DB9-female-to-DB25-male cable (Option 60A, Section 1.4), as applicable, from the analyzer’s rear
panel RS-232 port to the device. Adjust the DCE-DTE switch (Figure 3-4) to select DTE or DCE as
appropriate (Section 6.1).
Configuration: Section 6.3 and Section 6.7.2 (for Hessen protocol).
IMPORTANT
IMPACT ON READINGS OR DATA
Cables that appear to be compatible because of matching connectors
may incorporate internal wiring that makes the link inoperable. Check
cables acquired from sources other than TAI for pin assignments (Figure
3-13) before using.
Teledyne Analytical Instruments
49
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-13
Rear Panel Connector Pin-Outs for RS-232 Mode
The signals from these two connectors are routed from the motherboard via a wiring harness to two 10-
pin connectors on the CPU card, J11 and J12 (Figure 3-14).
Teledyne Analytical Instruments
50
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Figure 3-14:
Default Pin Assignments for CPU COMM Port Connector (RS-232).
TAI offers two mating cables, one of which should be applicable for your use.
P/N WR000077, a DB-9 female to DB-9 female cable, 6 feet long. Allows connection of the
serial ports of most personal computers.
P/N WR000024, a DB-9 female to DB-25 male cable. Allows connection to the most
common styles of modems (e.g. Hayes-compatible) and code activated switches.
Both cables are configured with straight-through wiring and should require no additional adapters.
Note
Cables that appear to be compatible because of matching connectors
may incorporate internal wiring that makes the link inoperable. Check
cables acquired from sources other than TAI for pin assignments before
using.
To assist in properly connecting the serial ports to either a computer or a modem, there are activity
indicators just above the RS-232 port. Once a cable is connected between the analyzer and a computer
or modem, both the red and green LEDs should be on.
If the lights are not lit, locate the small switch on the rear panel to switch it between DTE
and DCE modes.
If both LEDs are still not illuminated, ensure that the cable properly constructed.
Received from the factory, the analyzer is set up to emulate an RS-232 DCE device.
Teledyne Analytical Instruments
51
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
RS-232 (COM1): RS-232 (fixed) DB-9 male connector
Baud rate: 115200 bits per second (baud)
Data Bits: 8 data bits with 1 stop bit
Parity: None
COM2: RS-232 (configurable to RS 485), DB-9 female connector
Baud rate:19200 bits per second (baud)
Data Bits: 8 data bits with 1 stop bit
Parity: None
RS-232 MULTIDROP (OPTION 62) CONNECTION
When the RS-232 Multidrop option is installed, connection adjustments and configuration through the
menu system are required. This section provides instructions for the internal connection adjustments,
then for external connections, and ends with instructions for menu-driven configuration.
Note
Because the RS-232 Multidrop option uses both the RS232 and COM2
DB9 connectors on the analyzer’s rear panel to connect the chain of
instruments, COM2 port is no longer available for separate RS-232 or
RS-485 operation.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Printed Circuit Assemblies (PCAs) are sensitive to electro-static
discharges too small to be felt by the human nervous system. Failure to
use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 14 for more information on
preventing ESD damage.
In each instrument with the Multidrop option there is a shunt jumpering two pins on the serial Multidrop
and LVDS printed circuit assembly (PCA), as shown in Figure 3-15. This shunt must be removed from
all instruments except that designated as last in the multidrop chain, which must remain terminated. This
requires powering off and opening each instrument and making the following adjustments:
1. With NO power to the instrument, remove its top cover and lay the rear panel open
for access to the Multidrop/LVDS PCA, which is seated on the CPU.
2. On the Multidrop/LVDS PCA’s JP2 connector, remove the shunt that jumpers Pins
21 22 as indicated in. (Do this for all but the last instrument in the chain where
the shunt should remain at Pins 21 22).
3. Check that the following cable connections are made in all instruments (again refer
to Figure 3-15):
J3 on the Multidrop/LVDS PCA to the CPU’s COM1 connector
Teledyne Analytical Instruments
52
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
1. (Note that the CPU’s COM2 connector is not used in Multidrop)
J4 on the Multidrop/LVDS PCA to J12 on the motherboard
J1 on the Multidrop/LVDS PCS to the front panel LCD
Figure 3-15:
Jumper and Cables for Multidrop Mode
2. (Note: If you are adding an instrument to the end of a previously configured chain,
remove the shunt between Pins 21 22 of JP2 on the Multidrop/LVDS PCA in the
instrument that was previously the last instrument in the chain.)
4. Close the instrument.
5. Referring to Figure 3-16 use straight-through DB9 male DB9 female cables to
interconnect the host RS232 port to the first analyzer’s RS232 port; then from the
first analyzer’s COM2 port to the second analyzer’s RS232 port; from the second
analyzer’s COM2 port to the third analyzer’s RS232 port, etc., connecting in this
fashion up to eight analyzers, subject to the distance limitations of the RS-232
standard.
6. On the rear panel of each analyzer, adjust the DCE DTE switch so that the green
and the red LEDs (RX and TX) of the COM1 connector (labeled RS232) are both
lit. (Ensure you are using the correct RS-232 cables internally wired specifically for
RS-232 communication; see Table 1-1, “Communication Cables” and Section
3.3.1.8: Connecting the Communications Interfaces, “RS-232 Connection”).
Teledyne Analytical Instruments
53
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Female DB9
Male DB9
Host
RS-232 port
Analyzer
Analyzer
Analyzer
Last Analyzer
COM2
COM2
COM2
COM2
RS-232
RS-232
RS-232
RS-232
Ensure jumper is
installed between
JP2 pins 21
last instrument of
multidrop chain.
22 in
Figure 3-16:
RS-232-Multidrop PCA Host/Analyzer Interconnect Diagram
7. BEFORE communicating from the host, power on the instruments and check that
the Machine ID code is unique for each (Section 5.7.1).
3. a.
In the SETUP Mode menu go to SETUP>MORE>COMM>ID. The
default ID is typically the model number or “0”.
4. b.
to change the identification number, press the button below the digit to
be changed.
5. c.
Press/select ENTER to accept the new ID for that instrument.
8. Next, in the SETUP>MORE>COMM>COM1 menu (do not use the COM2 menu for
multidrop), edit the COM1 MODE parameter as follows: press/select EDIT and set
only QUIET MODE, COMPUTER MODE, and MULTIDROP MODE to ON. Do not
change any other settings.
9. Press/select ENTER to accept the changed settings, and ensure that COM1 MODE
now shows 35.
10. Press/select SET> to go to the COM1 BAUD RATE menu and ensure it reads the
same for all instruments (edit as needed so that all instruments are set at the same
baud rate).
The (communication) Host instrument can address only one instrument at a
time, each by its unique ID (see step 7 above).
Note
Teledyne Analytical Instruments
54
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
TAI recommends setting up the first link, between the Host and the first analyzer,
and testing it before setting up the rest of the chain.
Note
RS-485 CONNECTION
As delivered from the factory, COM2 is configured for RS-232 communications. This port can be
reconfigured for operation as a non-isolated, half-duplex RS-485 port. Using COM2 for RS-485
communication will disable the USB port. To reconfigure this port for RS-485 communication, please
contact the factory.
3.3.2. PNEUMATIC CONNECTIONS
This section provides not only pneumatic connection information, but also important information about
the gases required for accurate calibration (Section 3.3.2.1); it also illustrates the pneumatic layouts for
the analyzer in its basic configuration and with options.
Before making the pneumatic connections, carefully note the following cautionary and additional
messages:
CAUTION
GENERAL SAFETY HAZARD
Do not vent calibration gas or sample gas into enclosed areas.
CAUTION – GENERAL SAFETY HAZARD
In units with a permeation tube option installed, vacuum pump must be
connected and powered on to maintain constant gas flow though the
analyzer at all times. Insufficient gas flow allows gas to build up to levels
that will contaminate the instrument or present a safety hazard to
personnel.
Remove permeation tube when taking analyzer out of operation, and store
in sealed container (use original container that tube was shipped in).
(See Figure 3-6 for location and Section 11.3.6 for instructions on how to
remove the perm tube when the unit is not in operation).
IMPORTANT
IMPACT ON READINGS OR DATA
Sample and calibration gases should only come into contact with PTFE
tubing.
Teledyne Analytical Instruments
55
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Venting Pressurized Gas:
In applications where any gas (span gas, zero air supply, sample gas) is
received from a pressurized manifold, a vent must be provided to equalize
the gas with ambient atmospheric pressure before it enters the analyzer to
ensure that the gases input do not exceed the maximum inlet pressure of
the analyzer, as well as to prevent back diffusion and pressure effects.
These vents should be:
• at least 0.2m long
• no more than 2m long
• vented outside the shelter or immediate area surrounding the instrument.
Dust Plugs:
Remove dust plugs from rear panel exhaust and supply line fittings before
powering on/operating instrument. These plugs should be kept for reuse in
the event of future storage or shipping to prevent debris from entering the
pneumatics.
Leak Check:
IMPORTANT
Run a leak check once the appropriate pneumatic connections have
been made; check all pneumatic fittings for leaks using the
procedures defined in Section 11.3.12.1.
3.3.2.1. About Zero Air and Calibration (Span) Gas
Zero air and span gas are required for accurate calibration.
Note
Zero air and span gases must be supplied at twice the instrument’s
specified gas flow rate. Therefore, the 9110T zero and span gases should
be supplied to their respective inlets in excess of 1000 cc3/min (500
cc3/min x 2).
ZERO AIR
Zero air or zero calibration gas is defined as a gas that is similar in chemical composition to the
measured medium but without the gas to be measured by the analyzer.
For the 9110T this means zero air should be devoid of NO, NO2, CO2, NH3 or H2O vapor.
Note
Moderate amounts of NH3 and H2O can be removed from the sample gas
stream by installing the optional sample gas dryer/scrubber (see Section
3.3.2.6).
If your application is not a measurement in ambient air, the zero calibration gas should be matched
to the composition of the gas being measured.
Teledyne Analytical Instruments
56
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Pure nitrogen (N2) could be used as a zero gas for applications where NOX is measured in nitrogen.
If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero
air from ambient air.
For analyzers without the external zero air scrubber, a zero air generator such as the Teledyne
Model 701 can be used. Please visit the company website for more information.
CALIBRATION (SPAN) GAS
Calibration gas is a gas specifically mixed to match the chemical composition of the type of gas being
measured at near full scale of the desired reporting range. To measure NOX with the 9110T NOX
analyzer, it is recommended that you use a span gas with an NO concentration equal to 80% of the
measurement range for your application
EXAMPLE:
If the application is to measure NOX in ambient air between 0 ppm and 500 ppb, an appropriate
span gas would be 400 ppb.
If the application is to measure NOX in ambient air between 0 ppm and 1000 ppb, an appropriate
span gas would be 800 ppb.
Even though NO gas in nitrogen could be used as a span gas, the matrix of the balance gas is different
and may cause interference problems or yield incorrect calibrations.
The same applies to gases that contain high concentrations of other compounds (for example, CO2
or H2O).
The span gas should match all concentrations of all gases of the measured medium as closely as
possible.
Cylinders of the following types of calibrated NOx and NO gas traceable to NIST standards specifications
(also referred to as EPA protocol calibration gases or Standard Reference Materials) are commercially
available.
Table 3-8: NIST-SRM's Available for Traceability of NOx Calibration Gases
NOMINAL
CONCENTRATION
NIST-SRM4
TYPE
2627a
2628a
2629a
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
5 ppm
10 ppm
20 ppm
1683b
1684b
1685b
1686b
1687b
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
50 ppm
100 ppm
250 ppm
5000 ppm
1000 ppm
2630
2631a
2635
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
Nitric Oxide (NO) in N2
1500 ppm
3000 ppm
800 ppm
2636a
2000 ppm
2631a
1684b
Oxides of Nitrogen (NOx) in N2
Oxides of Nitrogen (NOx) in N2
2500 ppm
100 ppm
Note
The NO2 permeation tube included with the 9110T’s optional Internal
Teledyne Analytical Instruments
57
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Zero Air generator (IZS) has a limited accuracy of about ±5%.
While NO2 permeation tubes may be sufficient for informal calibration
checks, they are not approved by the US EPA as calibration sources for
performing actual calibration of the analyzer.
SPAN GAS FOR MULTIPOINT CALIBRATION
Some applications, such as EPA monitoring, require a multipoint calibration where span gases of
different concentrations are needed. We recommend using an NO gas of higher concentration combined
with a gas dilution calibrator such as a Teledyne Model 700E. This type of calibrator mixes a high
concentration gas with zero air to accurately produce span gas of the desired concentration. Linearity
profiles can be automated with this model and run unattended overnight.
If a dynamic dilution system such as the Teledyne Model 700 is used to dilute high concentration gas
standards to low, ambient concentrations, ensure that the NO concentration of the reference gas matches
the dilution range of the calibrator.
Choose the NO gas concentration so that the dynamic dilution system operates in its mid-range and not
at the extremes of its dilution capabilities.
EXAMPLE:
A dilution calibrator with 10-10000 dilution ratio will not be able to accurately dilute a 5000 ppm NO
gas to a final concentration of 500 ppb, as this would operate at the very extreme dilution setting.
A 100 ppm NO gas in nitrogen is much more suitable to calibrate the 9110T analyzer (dilution ratio
of 222, in the mid-range of the system’s capabilities).
Teledyne Analytical Instruments
58
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.3.2.2. Basic Connections from Calibrator, without and With Span Gas
Figure 3-17:
Gas Line Connections from Calibrator – Basic 9110T Configuration
Teledyne Analytical Instruments
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-18:
Gas Line Connections from Bottled Span Gas – Basic 9110T Configuration
For the 9110T nitrogen oxides analyzer in its basic configuration, attach the following pneumatic lines:
SAMPLE GAS SOURCE
Connect a sample gas line to the SAMPLE inlet
Use PTFE tubing; minimum OD ¼”
Sample Gas pressure must equal ambient atmospheric pressure (1.0 psig)
In applications where the sample gas is received from a pressurized manifold and the analyzer is
not equipped with one of the 9110T’s pressurized span options, a vent must be placed on the
sample gas line. This vent line must be:
No more than 10 meters long.
Vented outside the shelter or immediate area surrounding the instrument.
CALIBRATION GAS SOURCES
CAL GAS & ZERO AIR SOURCES: The source of calibration gas is also attached to the SAMPLE
inlet, but only when a calibration operation is actually being performed.
Use PTFE tubing; minimum OD ¼”.
Teledyne Analytical Instruments
60
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
VENTING
In order to prevent back diffusion and pressure effects, both the span gas and zero air supply lines should
be:
Vented outside the enclosure.
Minimum OD ¼”.
Not less than 2 meters in length.
Not greater than 10 meters in length.
EXHAUST OUTLET
Attach an exhaust line to the EXHAUST outlet fitting. The exhaust line should be:
Use PTFE tubing; minimum OD ¼”.
A maximum of 10 meters long.
Vented outside the 9110T analyzer’s enclosure
Note
Once the appropriate pneumatic connections have been made, check all
pneumatic fittings for leaks using the procedures defined in Sections
13.3.13.2 (or 13.3.13.3 for detailed check if leak suspected).
PNEUMATIC LAYOUT FOR BASIC CONFIGURATION
Teledyne Analytical Instruments
61
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-19:
Pneumatics, Basic Configuration
3.3.2.3. Connections w/Ambient Zero/Ambient Span (Z/S) Valves (OPT 50A)
This valve package includes:
Two solenoid valves located inside the analyzer that allow the user to switch either zero, span or
sample gas to the instrument’s sensor.
Two additional gas inlet ports (ZERO AIR and SPAN1).
Teledyne Analytical Instruments
62
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Figure 3-20:
Rear Panel Layout with Z/S Valve Options (OPT 50A)
Figure 3-21:
Gas Line Connections for 9110T with Z/S Valves Option (OPT 50A)
SAMPLE GAS SOURCE
Attach a sample inlet line to the SAMPLE inlet fitting.
Use PTFE tubing; minimum O.D ¼”.
Teledyne Analytical Instruments
63
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Sample Gas pressure must equal ambient atmospheric pressure (no greater than 1.0 psig).
In applications where the sample gas is received from a pressurized manifold, a vent must be
placed on the sample gas line. This vent line must be no more than 10 meters long.
CALIBRATION GAS SOURCES
SPAN GAS
Attach a gas line from the source of calibration gas (e.g. a Teledyne
M700E Dynamic Dilution Calibrator) to the SPAN1 inlet (see Figure
3-20). Use PTFE tubing; minimum O.D ¼”.
ZERO AIR
Zero air is supplied by the zero air generator such as a Teledyne
M701. Attach a gas line from the source of zero air to the ZERO AIR
inlet.
VENTING
In order to prevent back diffusion and pressure effects, both the span gas and zero air supply lines
should be:
Vented outside the enclosure.
Not less than 2 meters in length.
Not greater than 10 meters in length.
EXHAUST OUTLET
Attach an exhaust line to the EXHAUST OUTLET fitting. The exhaust line should be:
¼” PTFE tubing
maximum 10 meters long
Vented outside the 9110T analyzer’s enclosure
Note
Once the appropriate pneumatic connections have been made, check all
pneumatic fittings for leaks using the procedures defined in Section
13.3.12.
To find instructions on calibrating a 9110T with this option installed, see
section 10.4.
PNEUMATIC LAYOUT FOR AMBIENT ZERO/AMBIENT SPAN VALVES (OPT 50A)
Teledyne Analytical Instruments
64
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Figure 3-22:
Pneumatics with Zero/Span Valves OPT 50A
Table 3-9:
Zero/Span Valves Operating States OPT 50A
VALVE PORT
STATUS
MODE
VALVE
CONDITION
Sample/Cal
Zero/Span
Open to SAMPLE inlet
Open to ZERO AIR inlet
3 2
3 2
1 2
3 2
1 2
1 2
SAMPLE
Sample/Cal
Zero/Span
Open to ZERO/SPAN Valve
Open to ZERO AIR inlet
ZERO CAL
SPAN CAL
Sample/Cal
Zero/Span
Open to ZERO/SPAN Valve
Open to SPAN inlet
Teledyne Analytical Instruments
65
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
3.3.2.4. Connections w/Ambient Zero/Pressurized Span Valves (OpT 50B)
This calibration valve package is appropriate for applications where Span Gas is being supplied from a
pressurized source such as bottled NIST SRM gases. This option includes:
A critical flow orifice and vent that maintains the Span Gas supply at 1 ATM.
A SHUTOFF valve to preserve the Span Gas source when it is not in use.
Two solenoid valves located inside the analyzer that allow the user to switch either zero, span or
sample gas to the instrument’s sensor.
Three additional gas inlet ports (ZERO AIR, SPAN and VENT).
Figure 3-23:
Rear Panel Layout with Ambient Zero/Pressurized Span Valves OPT 50B
Teledyne Analytical Instruments
66
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Figure 3-24:
Gas Line Connection w/Ambient Zero/Pressurized Span Valves (OPT 50B)
SAMPLE GAS SOURCE
Attach a sample inlet line to the SAMPLE inlet fitting.
Use PTFE tubing; minimum O.D ¼”.
Sample Gas pressure must equal ambient atmospheric pressure (29.92 in-Hg).
In applications where the sample gas is received from a pressurized manifold, a vent must be
placed on the sample gas line. This vent line must be:
No more than 10 meters long.
Vented outside the shelter or immediate area surrounding the instrument.
CALIBRATION GAS SOURCES
SPAN GAS
Attach a gas line from the pressurized source of calibration gas (e.g. a
bottle of NISTSRM gas) to the SPAN1 inlet.. Use PTFE tubing;
minimum O.D ¼”.
ZERO AIR
(the dual-stage zero Air Scrubber makes zero air)
VENTING
Attach a line to the SPAN2/VENT outlet. It should be:
¼” PTFE tubing.
Teledyne Analytical Instruments
67
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Vented outside the enclosure.
Not less than 2 meters in length.
Not greater than 10 meters in length.
EXHAUST OUTLET
Attach an exhaust line to the EXHAUST outlet fitting. The exhaust line should be:
¼” PTFE tubing.
A maximum of 10 meters long.
Vented outside the 9110T analyzer’s enclosure.
PNEUMATIC LAYOUT FOR AMBIENT ZERO/PRESSURIZED SPAN (OPT 50B)
Figure 3-25:
Pneumatics with Ambient Zero/Pressurized Span Valves (OPT 50B)
Teledyne Analytical Instruments
68
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Table 3-10:
Valve Operating States OPT 50B installed
VALVE PORT
MODE
VALVE
CONDITION
STATUS
Sample/Cal
Zero/Span
Open to SAMPLE inlet
Open to ZERO AIR inlet
3 2
3 2
SAMPLE
Span Shutoff
Zero Air Shutoff
Closed
Closed
Sample/Cal
Zero/Span
Open to ZERO/SPAN Valve
Open to ZERO AIR inlet
OPEN
1 2
3 2
ZERO CAL
SPAN CAL
Span Shutoff
Zero Air Shutoff1
Closed
Sample/Cal
Open to ZERO/SPAN Valve
1 2
Zero/Span
Span Shutoff
Zero Air Shutoff
Open to SPAN inlet
Closed
1 2
OPEN
3.3.2.5. Zero Scrubber and Internal Span Source (IZS) (OPT 50G)
The 9110T nitrogen oxides analyzer can also be equipped with an internal NO2 span gas generator and
calibration valve option. This option package is intended for applications where there is a need for
frequent automated calibration checks without access to an external source of span gas.
This valve package includes:
A 2-stage external scrubber for producing zero air.
50% Purafil Chemisorbant® (for conversion of NO NO2).
50% charcoal (for removal of the NO2).
A heated enclosure for a NO2 permeation tube.
This option package DOES NOT contain an actual permeation tube. See
Section 1.4 (Options 52B and 52G) for information on specifying the correct
permeation tube for each application.
A special desorber that removes all HNO3 from the calibration gas stream.
One additional gas inlet port (ZERO AIR).
One additional gas outlet port (FROM DRYER).
Two internal valves for switching between the sample gas inlet and the output of the zero/span
subsystem.
Teledyne Analytical Instruments
69
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Figure 3-26:
Rear Panel Layout with Internal Span Source (IZS) OPT 50G
INTERNAL SPAN GAS GENERATION
The primary component of the internal span option is a permeation tube containing liquid NO2. As zero
air is passed over a permeable membrane on the end of the tube, molecules of NO2 slowly pass through
the membrane mixing with the zero air.
The resulting concentration of the NO2 span gas is determined by three factors:
Size of the membrane: The larger the area of the membrane, the more permeation occurs.
Temperature of the NO2: Increasing the temperature of the permeation tube increases the pressure
inside the tube and therefore increases the rate of permeation.
Flow rate of the zero air: If the previous two variables are constant, the permeation rate of the NO2
into the zero air stream will be constant. Therefore, a lower flow rate of zero air produces higher
concentrations of NO2.
In the Model 9110T the permeation tube enclosure is heated to a constant 50° C (10° above the
maximum operating temperature of the instrument) in order to keep the permeation rate constant. A
thermistor measures the actual temperature and reports it to the CPU for control feedback.
The flow rate of zero air across the permeation tube is maintained at 50 ± 10 cm³/min by a critical flow
orifice located in the analyzer’s exhaust manifold.
NITRIC ACID AND THE CHEMISTRY OF NO2 PERMEATION TUBES
The reaction of H2O with NO2 to form HNO3 (nitric acid) takes place whenever water and NO2 are
present in the same gas mixture. In the 9110T this is mitigated as much as possible by passing the air
supply for the span gas generator through a special dryer, however the permeable membrane of the NO2
tube will still allow H2O from the ambient environment to slowly collect in the tube at increasingly
higher concentrations. Over time this results in the presence of HNO3 in the permeation tube which is
exuded into the 9110T’s pneumatics along with NO2.
Teledyne Analytical Instruments
70
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
HNO3 is a liquid at room temperature, so once the HNO3 is released by the permeation tube it condenses
and collects along the 9110T’s wetted surfaces, While liquid HNO3 does not directly effect the quality of
NOx measurements of the Model 9110T, it does give off small amounts of gaseous HNO3 which is
converted into NO by the 9110T’s NOx NO converter resulting in an artificially high NO2
concentration by 8% to 12%. This is particularly bothersome when 9110T is attempting to measure a
zero point, such as during calibration, since the NO2 concentration will only reach a true zero point once
the majority of the HNO3 coating the wetted surfaces has reverted to NO2 and this can take a very long
time.
The 9110T includes a special HNO3 desorbed which eliminates any HNO3 given off by the permeation
tube before it can be converted into NO by the analyzer’s converter.
PNEUMATIC LAYOUT FOR ZERO SCRUBBER AND IZS (OPT 50G)
Figure 3-27:
Pneumatics with the Internal Span Gas Generator (OPT 50G)
Table 3-11: Internal Span Gas Generator Valve Operating States OPT 50G
VALVE PORT
STATUS
MODE
SAMPLE
VALVE
CONDITION
Sample/Cal
Open to SAMPLE inlet
3 2
Teledyne Analytical Instruments
71
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Zero/Span
Sample/Cal
Zero/Span
Sample/Cal
Zero/Span
Open to ZERO AIR inlet
Open to ZERO/SPAN Valve
Open to ZERO AIR inlet
Open to ZERO/SPAN Valve
Open to SPAN inlet
3 2
1 2
3 2
1 2
1 2
ZERO CAL
SPAN CAL
3.3.2.6. Gas Conditioner Options
AMMONIA REMOVAL SAMPLE CONDITIONER (OPT 86A)
The 9110T includes a Nafion® permeation gas exchange tube to remove H2O a from the ozone generator
supply gas stream to a dew point of about -20° C (~600 ppm H2O) and effectively remove concentrations
of ammonia (NH3) up to about 1 ppm.
An additional Sample Conditioner can be added to the 9110T’s sample gas stream.
Sample Gas
NH3 Dryer/Scrubber
NO/NOX
VALVE
SAMPLE
PRESSURE
SENSOR
VACUUM
PRESSURE
SENSOR
O3 FLOW
SENSOR
AUTOZERO
VALVE
PMT
Figure 3-28:
Pneumatics for Sample Conditioner OPT 86A
ZERO AIR SCRUBBER (OPT 86C), FOR Z/S VALVES
Teledyne Analytical Instruments
72
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
An external zero air scrubber for Z/S valves can be used in place of a zero air generator The following
pneumatic diagram illustrates the internal and external flow for a 9110T analyzer with a Z/S valve option
and the Zero Air Scrubber (Option 86C):
Figure 3-29:
Pneumatics for External Zero Air Scrubber (OPT 86C) for Z/S Valves
Teledyne Analytical Instruments
73
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
3.4. STARTUP, FUNCTIONAL CHECKS, AND INITIAL
CALIBRATION
CAUTION!
If the presence of ozone is detected at any time, power down the instrument
and contact TAI Customer Service as soon as possible:
(626) 934-1500 or email: [email protected]
If you are unfamiliar with the 9110T principles of operation, we recommend that you read Section 13.
For information on navigating the analyzer’s software menus, see the menu trees described in Appendix
A.
3.4.1. START UP
After the electrical and pneumatic connections are made, an initial functional check is in order. Turn on
the instrument. The pump and exhaust fan should start immediately. The display will show a splash
screen and other information during the initialization process while the CPU loads the operating system,
the firmware and the configuration data.
The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and
start monitoring the gas. However, there is an approximately one hour warm-up period before reliable
gas measurements can be taken. During the warm-up period, the front panel display may show messages
in the parameters (Param) field.
Teledyne Analytical Instruments
74
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.4.2. WARNING MESSAGES
Because internal temperatures and other conditions may be outside the specified limits during the
analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes after
power up. If warning messages persist after the 30 minutes warm up period is over, investigate their
cause using the troubleshooting guidelines in Section 12.1.
To view and clear warning messages, press:
Table 3-12 lists brief descriptions of the warning messages that may occur during start up.
Teledyne Analytical Instruments
75
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Table 3-12: Possible Warning Messages at Start-Up
MEANING
MESSAGE
SYSTEM RESET1
The computer has rebooted.
ANALOG CAL WARNING The A/D or at least one D/A channel have not been calibrated.
BOX TEMP WARNING
CANNOT DYN SPAN2
CANNOT DYN ZERO3
CONFIG INITIALIZED
DATA INITIALIZED
The temperature inside the 9110T chassis is outside the specified limits.
Contact closure span calibration failed while DYN_SPAN was set to ON.
Contact closure zero calibration failed while DYN_ZERO was set to ON.
Configuration storage was reset to factory configuration or erased.
DAS data storage was erased before the last power up occurred.
OZONE FLOW WARNING Ozone gas flow is too high or too low for accurate NOx, NO and NO2 readings.
Ozone generator is off.
This is the only warning message that automatically clears itself.
OZONE GEN OFF 4
It clears itself when the ozone generator is turned on.
Upon power up the Ozone generator will remain off for 30 minutes. This allows the perma-
pure dryer to reach its working dew point.
RCELL PRESS WARN
Reaction cell pressure is too high or too low for accurate NOx, NO and NO2 readings.
RCELL TEMP WARNING Reaction cell temperature is too high or too low for accurate NOx, NO and NO2 readings.
IZS TEMP WARNING 5
CONV TEMP WARNING
PMT TEMP WARNING
IZS temperature is too high or too low for efficient O3 production.
NO2 to NO Converter temperature too high or too low to efficiently convert NO2 to NO.
PMT temperature outside of warning limits specified by PMT_SET variable.
AZERO WARN [XXXX]
AutoZero reading too high. The value shown in message indicates auto-zero reading at
time warning was displayed.
MV
HVPS WARNING
High voltage power supply output is too high or too low for proper operation of the PMT.
CPU unable to communicate with motherboard..
REAR BOARD NOT DET
RELAY BOARD WARN
SAMPLE FLOW WARN
CPU is unable to communicate with the relay PCA.
The flow rate of the sample gas is outside the specified limits.
1
Clears 45 minutes after power up.
2
3
4
5
Clears the next time successful zero calibration is performed.
Clears the next time successful span calibration is performed.
Clears 30 minutes after power up.
Only Appears if the IZS option is installed.
Teledyne Analytical Instruments
76
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
3.4.3. FUNCTIONAL CHECKS
After the analyzer’s components have warmed up for at least 60 minutes, verify that the software
properly supports any hardware options that are installed.
For information on navigating through the analyzer’s software menus, see the menu trees described in
Appendix A.1.
Check to ensure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as their
expected values.
These functions are also useful tools for diagnosing problems with your analyzer.
The enclosed Final Test and Validation Data sheet (P/N 04409) lists these values before the
instrument left the factory.
To view the current values of these parameters press the following button sequence on the analyzer’s
front panel. Remember until the unit has completed its warm up these parameters may not have
stabilized.
3.4.4. INITIAL CALIBRATION
To perform the following calibration you must have sources for zero air and calibration (span) gas
available for input into the inlet/outlet fittings on the back of the analyzer (see Section 3.3.2.1).
Teledyne Analytical Instruments
77
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
Note
A start-up period of 4-5 hours is recommended prior to performing a
calibration on the analyzer.
The method for performing an initial calibration for the 9110T nitrogen oxides analyzer differs slightly
depending on the whether or not any of the available internal zero air or valve options are installed.
See Section 3.4.4.2 for instructions for initial calibration of the 9110T analyzers in their base
configuration.
See Section 9.3 for instructions for initial calibration of 9110T analyzers possessing an optional
Internal Span Gas Generator (OPT 51A).
See Section 9.4 for information regarding setup and calibration of 9110T analyzers with Z/S Valve
options.
If you are using the 9110T analyzer for EPA monitoring, only the calibration method described in
Section 10 should be used.
3.4.4.1. INTERFERENTS FOR NOX, NO AND NO2 MEASUREMENTS
The chemiluminescence method for detecting NOX is subject to interference from a number of sources
including water vapor (H2O), ammonia (NH3), sulfur dioxide (SO2) and carbon dioxide (CO2) but the
9110T has been designed to reject most of these interferences.
Ammonia is the most common interferent, which is converted to NO in the analyzer’s NO2 converter
and creates a NOX signal artifact.
If the 9110T is installed in an environment with high ammonia, steps should be taken to
remove the interferent from the sample gas before it enters the reaction cell.
TAI offers a sample gas conditioning option to remove ammonia and water vapor (Section
3.3.2.6).
Carbon dioxide (CO2) diminishes the NOX signal when present in high concentrations.
If the analyzer is used in an application with excess CO2, contact TAI's Customer Service
Department (see Section 12.10) for possible solutions.
Excess water vapor can be removed with one of the dryer options described in Section 3.3.2.6. In
ambient air applications, SO2 interference is usually negligible.
For more detailed information regarding interferents for NOx, NO and NO2 measurement, see Section
13.1.5.
3.4.4.2. Initial Calibration Procedure for 9110T Analyzers without Options
The following procedure assumes that:
The instrument DOES NOT have any of the available calibration valve or gas inlet options installed;
Teledyne Analytical Instruments
78
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer and;
The pneumatic setup matches that described in Section 3.3.2.
VERIFYING THE REPORTING RANGE SETTINGS
While it is possible to perform the following procedure with any range setting we recommend that you
perform this initial checkout using following reporting range settings:
Unit of Measure: PPB
Reporting Range: 500 ppb
Mode Setting: SNGL
While these are the default setting for the 9110T analyzer, it is recommended that you verify them before
proceeding with the calibration procedure, by pressing the following menu button sequence:
Teledyne Analytical Instruments
79
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
VERIFYING THE EXPECTED NOX AND NO SPAN GAS CONCENTRATION
IMPORTANT
IMPACT ON READINGS OR DATA
Verify the PRECISE Concentration Value of the SPAN gases
independently.
If you supply NO gas to the analyzer, the values for expected NO and
NOx MUST be identical.
The NOx and NO span concentration values automatically defaults to 400.0 PPB and it is recommended
that calibration gases of that concentration be used for the initial calibration of the unit.
Teledyne Analytical Instruments
80
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
To verify that the analyzer span setting is set for 400 PPB, press:
Teledyne Analytical Instruments
81
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Getting Started
INITIAL ZERO/SPAN CALIBRATION PROCEDURE
To perform an initial Calibration of the 9110T nitrogen oxides analyzer, press:
The 9110T Analyzer is now ready for operation.
Once you have completed the above set-up procedures, please fill out
the Quality Questionnaire that was shipped with your unit and return it
Note
to Teledyne. This information is vital to our efforts in continuously
improving our service and our products. THANK YOU.
Teledyne Analytical Instruments
82
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Getting Started
This page intentionally left blank.
Teledyne Analytical Instruments
83
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Part II
PART II
–
OPERATING INSTRUCTIONS
Teledyne Analytical Instruments
85
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Part II
This page intentionally left blank.
Teledyne Analytical Instruments
86
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Overview of Operating Modes
4. OVERVIEW OF OPERATING MODES
To assist in navigating the analyzer’s software, a series of menu trees is available for reference in
Appendix A of this manual.
Note
Some control buttons on the touch screen do not appear if they are not
applicable to the menu that you’re in, the task that you are performing, the
command you are attempting to send, or to incorrect settings input by the
user. For example, the ENTR button may disappear if you input a setting
that is invalid or out of the allowable range for that parameter, such as
trying to set the 24-hour clock to 25:00:00. Once you adjust the setting to
an allowable value, the ENTR button will re-appear.
The 9110T analyzer software has a variety of operating modes. The most common mode that the
analyzer will be operating in is the SAMPLE mode. In this mode, a continuous read-out of the NOx
concentrations can be viewed on the front panel and output as an analog voltage from rear panel
terminals.
The second most important operating mode is SETUP mode. This mode is used for
configuring the various sub systems of the analyzer such as for the DAS system, the reporting ranges,
or the serial (RS-232 / RS-485 / Ethernet) communication channels. The SETUP mode is also used for
performing various diagnostic tests during troubleshooting.
Figure 4-1:
Front Panel Display
The mode field of the front panel display indicates to the user which operating mode the unit is currently
running.
In addition to SAMPLE and SETUP, other modes the analyzer can be operated in are described in Table
7-1 below.
Teledyne Analytical Instruments
87
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Overview of Operating Modes
Table 4-1: Analyzer Operating Modes
MODE
EXPLANATION
DIAG
One of the analyzer’s diagnostic modes is active.
LO CAL A
Unit is performing LOW SPAN (midpoint) calibration initiated automatically by the analyzer’s
AUTOCAL feature
LO CAL R
Unit is performing LOW SPAN (midpoint) calibration initiated remotely through the COM ports or
digital control inputs.
M-P CAL
SAMPLE
This is the basic calibration mode of the instrument and is activated by pressing the CAL button.
Sampling normally, flashing text indicates adaptive filter is on.
SAMPLE A
SETUP X.#2
Indicates that unit is in SAMPLE mode and AUTOCAL feature is activated.
SETUP mode is being used to configure the analyzer. The gas measurement will continue during
setup.
SPAN CAL A1
SPAN CAL M1
SPAN CAL R1
Unit is performing SPAN calibration initiated automatically by the analyzer’s AUTOCAL feature
Unit is performing SPAN calibration initiated manually by the user.
Unit is performing SPAN calibration initiated remotely through the COM ports or digital control
inputs.
ZERO CAL A1
ZERO CAL M1
ZERO CAL R1
Unit is performing ZERO calibration procedure initiated automatically by the AUTOCAL feature
Unit is performing ZERO calibration procedure initiated manually by the user.
Unit is performing ZERO calibration procedure initiated remotely through the COM ports or digital
control inputs.
1 Only Appears on units with Z/S valve or IZS options.
2 The revision of the analyzer firmware is displayed following the word SETUP, e.g., SETUP G.3.
4.1. SAMPLE MODE
This is the analyzer’s standard operating mode. In this mode, the instrument is a calculating
NOx, NO and NO2 concentrations. These values are displayed in the CONC field of the analyzer's front
panel display. While the instrument is in SAMPLE mode, this field provides a readout of all the gas
concentrations being measured by the 9110T: NOx, NO and NO2.
When the analyzer is in sample mode the PARAM field will display warning messages and test
functions that give the user information about the operational status of the analyzer.
4.1.1. TEST FUNCTIONS
A variety of TEST functions are available for viewing at the front panel whenever the analyzer is at the
MAIN MENU. These functions provide information about the various functional parameters related to
the analyzer’s operation and its measurement of gas concentrations. This information is particularly
when troubleshooting a performance problem with the 9110T (see Section 13). Figure 4-2 will display
the Test Functions on the front panel screen. Table 4-2 lists the available TEST functions.
Table 4-2: Test Functions Defined
DISPLAY
PARAMETER
UNITS
DESCRIPTION
The Full Scale limit at which the reporting range of the analyzer’s ANALOG
OUTPUTS is currently set.
PPB,
PPM,
UGM
&
RANGE
RANGE
THIS IS NOT the Physical Range of the instrument. See Section 5.4.1 for
more information.
Teledyne Analytical Instruments
88
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Overview of Operating Modes
DESCRIPTION
DISPLAY
PARAMETER
UNITS
MGM
If AUTO Range mode has been selected, two RANGE functions will
appear, one for each range:
RANGE1
RANGE2
RANGE1: The range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.
If the IND Range mode has been selected, three RANGE functions will
appear, one for each range:
RANGE1
RANGE2
RANGE3
RANGE1: NOx concentration output un A1.
RANGE2: NO concentration output un A2.
RANGE2: NO2 concentration output un A3.
The standard deviation of concentration readings of the selected gas.
NOX STB
STABILITY
PPB
Data points are recorded every ten seconds. The calculation uses the
last 25 data points.
Gas flow rate of the sample gas into the reaction cell.
Gas flow rate of O3 gas into the reaction cell.
The raw signal output of the PMT.
SAMP FLW
OZONE FL
PMT
SAMPFLOW
OZONEFLOW
PMT
CC/M
CC/M
MV
The signal output of the PMT after is has been normalized for temperature,
pressure, auto-zero offset, but not range.
NORM PMT
NORMPMT
MV
The PMT signal with zero NOX, which is usually slightly different from 0 V.
This offset is subtracted from the PMT signal and adjusts for variations in
the zero signal.
AZERO
AUTOZERO
MV
The output power level of the high voltage power supply.
The temperature of the gas inside the reaction cell temperature.
The temperature inside the analyzer chassis.
HVPS
HVPS
V
C
C
C
C
C
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP1
MOLY TEMP
RCELLTEMP
BOXTEMP
PMTTEMP
IZSTEMP
The temperature of the PMT .
The temperature of the internal span gas generator's permeation tube.
The temperature of the analyzer's NO2 NO converter.
CONVTEMP
The current pressure of the sample gas in the reaction cell as measured at
the vacuum manifold.
RCEL
RCELLPRESS
SAMPPRESS
IN-HG-A
IN-HG-A
The current pressure of the sample gas as it enters the reaction cell,
measured between the NO/NOx and Auto-Zero valves.
SAMP
The slope calculated during the most recent NOx zero/span calibration.
The offset calculated during the most recent NOx zero/span calibration.
The slope calculated during the most recent NO zero/span calibration.
The offset calculated during the most recent NO zero/span calibration.
NOX SLOPE
NOX OFFS
NO SLOPE
NO OFFS
NOXSLOPE
NOXOFFSET
NOSLOPE
MV
NOOFFSET
MV
MV
Displays the signal level of the Test Function that is currently being
produced by the Analog Output Channel A4.
TEST
TESTCHAN
The current time. This is used to create a time stamp on DAS readings,
and by the AutoCal feature to trigger calibration events.
TIME
CLOCKTIME
HH:MM:SS
1Only appears if Internal Span Gas Generator option is installed.
To view these TEST functions, press,
Teledyne Analytical Instruments
89
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Overview of Operating Modes
Figure 4-2:
Viewing 9110T Test Functions
IMPORTANT
IMPACT ON READINGS OR DATA
A value of “XXXX” displayed for any of the TEST functions indicates an
out-of-range reading or the analyzer’s inability to calculate it. All
pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It
decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors
such as air conditioning and passing storms can cause changes in the
absolute atmospheric pressure.
Teledyne Analytical Instruments
90
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Overview of Operating Modes
4.1.2. WARNING MESSAGES
The most common and serious instrument failures will activate Warning Messages that are displayed on
the analyzer’s Front Panel. They are listed on Table 4-3 as follows:
Table 4-3: Warning Messages Defined
MESSAGE
ANALOG CAL WARNING
AZERO WARN
MEANING
The A/D or at least one D/A channel has not been calibrated.
Auto-zero reading above limit specified by AZERO_LIMIT variable. Value shown
in message indicates auto-zero reading at time warning was displayed.
BOX TEMP WARNING
CANNOT DYN SPAN
CANNOT DYN ZERO
CONFIG INITIALIZED
CONV TEMP WARNING
The temperature inside the 9110T chassis is outside the specified limits.
Contact closure span calibration failed while DYN_SPAN was set to ON.
Contact closure zero calibration failed while DYN_ZERO was set to ON.
Configuration storage was reset to factory configuration or erased.
NO2 NO converter temperature outside of warning limits specified by
CONV_SET variable.
DATA INITIALIZED
HVPS WARNING
DAS data storage was erased before the last power up occurred.
High voltage power supply output outside of warning limits specified by
HVPS_SET variable.
IZS TEMP WARNING 1
OZONE FLOW WARNING
OZONE GEN OFF
IZS temperature outside of warning limits specified by IZS_SET variable.
Ozone flow outside of warning limits specified by OFLOW_SET variable.
Ozone generator is off. This warning message clears itself when the ozone
generator is turned on.
PMT temperature outside of warning limits specified by PMT_SET variable.
PMT TEMP WARNING
RCELL PRESS WARN
Reaction cell pressure outside of warning limits specified by
RCELL_PRESS_SET variable.
Reaction cell temperature outside of warning limits specified by RCELL_SET
variable.
RCELL TEMP WARNING
REAR BOARD NOT DET
RELAY BOARD WARN
SAMPLE FLOW WARN
SYSTEM RESET
Motherboard was not detected during power up.
CPU is unable to communicate with the relay PCA.
The flow rate of the sample gas is outside the specified limits.
The computer has rebooted.
1
Only Appears if the Internal Span Gas Generator option is installed.
Teledyne Analytical Instruments
91
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Overview of Operating Modes
4.2. CALIBRATION MODE
Pressing the CAL button, switches the analyzer into calibration mode. In this mode the user can, in
conjunction with introducing zero or span gases of known concentrations into the analyzer, cause it to
adjust and recalculate the slope (gain) and offset of the its measurement range. This mode is also used to
check the current calibration status of the instrument.
If the instrument includes one of the available zero/span valve options, the SAMPLE mode display will
also include CALZ and CALS buttons. Pressing either of these buttons also puts the instrument into
calibration mode.
The CALZ button is used to initiate a calibration of the analyzer’s zero point using internally
generated zero air.
The CALS button is used to calibrate the span point of the analyzer’s current reporting range using
span gas.
Note
It is recommended that this span calibration be performed at 80% of full
scale of the analyzer’s currently selected reporting range.
EXAMPLES:
If the reporting range is set for 0 to 500 ppb, an appropriate span point
would be 400 ppb.
If the of the reporting range is set for 0 to 1000 ppb, an appropriate span
point would be 800 ppb.
Due to the critical importance and complexity, calibration operations are described in detail in other
sections of the manual:
Section 9 details setting up and performing standard calibration operations or checks.
Section 10 details setting up and performing EPA protocol calibrations.
For information on using the automatic calibrations feature (ACAL) in conjunction with the one of the
calibration valve options, see Sections 9.4.3 and 9.5.
IMPORTANT
IMPACT ON READINGS OR DATA
To avoid inadvertent adjustments to critical settings, activate calibration
security by enabling password protection in the SETUP – PASS menu
(Section 5.5).
4.3. SETUP MODE
The SETUP Mode contains a variety of choices that are used to configure the analyzer’s hardware and
software features, perform diagnostic procedures, gather information on the instruments performance and
configure or access data from the internal data acquisition system (DAS). For a visual representation of
the software menu trees, refer to Appendix A.
SETUP Mode is divided between Primary and Secondary Setup menus and can be protected through
password security.
Teledyne Analytical Instruments
92
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Overview of Operating Modes
4.3.1. PASSWORD SECURITY
SETUP Mode can be protected by password security through the SETUP>PASS menu (Section 5.5) to
prevent unauthorized or inadvertent configuration adjustments.
4.3.2. PRIMARY SETUP MENU
The areas accessed and configured under the primary SETUP Mode menu are shown in Table 4-4.
Table 4-4: Primary Setup Mode Features and Functions
CONTROL
BUTTON
LABEL
MANUAL
SECTION
MODE OR FEATURE
Analyzer Configuration
Auto Cal Feature
DESCRIPTION
CFG
Lists button hardware and software configuration information.
5.1
Used to set up and operate the AutoCal feature.
ACAL
5.2, 9.5
Only appears if the analyzer has one of the calibration valve
options installed.
Internal Data Acquisition
(DAS)
DAS
Used to set up the DAS system and view recorded data.
7
Analog Output Reporting
Range Configuration
Used to configure the output signals generated by the
instruments analog outputs.
RNGE
5.4
Calibration Password Security
Internal Clock Configuration
PASS
CLK
Turns the calibration password feature ON/OFF.
Used to set or adjust the instrument’s internal clock.
5.5
5.6
Secondary SETUP Mode
(Advanced SETUP features)
See
Table 4-5
MORE
This button accesses the instruments secondary setup menu.
4.3.3. SECONDARY SETUP MENU (SETUP MORE)
The areas accessed and configured under the secondary SETUP Mode menu are shown in Table 4-5.
Table 4-5: Secondary Setup Mode Features and Functions
CONTROL
BUTTON
LABEL
MANUAL
SECTION
MODE OR FEATURE
DESCRIPTION
Used to set up and operate the analyzer’s various external I/O
channels including RS-232; RS-485, modem communication
and/or Ethernet access.
External Communication
Channel Configuration
COMM
VARS
8
Used to view various variables related to the instruments
current operational status.
Changes made to any variable are not acknowledged and
recorded in the instrument’s memory until the ENTR button
is pressed.
System Status Variables
5.8
Pressing the EXIT button ignores the new setting.
If the EXIT button is pressed before the ENTR button, the
analyzer will beep alerting the user that the newly entered
value has been lost.
Used to access a variety of functions that are used to configure,
test or diagnose problems with a variety of the analyzer’s basic
systems.
System Diagnostic Features
and
DIAG
5.9, 5.9.2
Analog Output Configuration
Most notably, the menus used to configure the output signals
generated by the instruments’ analog outputs are located here.
Teledyne Analytical Instruments
93
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Overview of Operating Modes
IMPORTANT
IMPACT ON READINGS OR DATA
Any changes made to a variable during the SETUP procedures are not
acknowledged by the instrument until the ENTR button is pressed. If the
EXIT button is pressed before the ENTR button, the analyzer will make an
audible signal before exiting the menu, alerting the user that the newly
entered value had not been accepted.
Teledyne Analytical Instruments
94
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5. SETUP MENU
The SETUP Mode menu is used to set instrument parameters for performing configuration, calibration,
reporting and diagnostics operations according to user needs.
5.1. SETUP CFG: CONFIGURATION INFORMATION
Pressing the CFG button displays the instrument configuration information. This display lists the
analyzer model, serial number, firmware revision, software library revision, CPU type and other
information. Use this information to identify the software and hardware when contacting customer
service. Special instrument or software features or installed options may also be listed here.
5.2. SETUP ACAL: AUTOMATIC CALIBRATION OPTION
The menu button for this option appears only when the instrument has the zero span and/or IZS options.
See Section 9.5 for details.
5.3. SETUP DAS: INTERNAL DATA ACQUISITION SYSTEM
Use the SETUP>DAS menu to capture and record data. Refer to Section 7 for configuration and
operation details.
Teledyne Analytical Instruments
95
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.4. SETUP RNGE: ANALOG OUTPUT REPORTING RANGE
CONFIGURATION
Use the SETUP>RNGE menu to configure output reporting ranges, including scaled reporting ranges to
handle data resolution challenges. This section describes configuration for Single, Dual, and Auto Range
modes.
5.4.1. 9110T PHYSICAL RANGES
The 9110T NOx analyzer measures NOx, NO and NO2 concentrations from 2 to 20,000 ppb.
Electronically the 9110T analyzer converts the 0-5 volt analog signal output by the PMT into a digital
signal with 4096 counts of resolution. Since its measurement range is 0 ppb to 20,000 ppb, this only
allows about 3 ppb per count. While this might be acceptable for high concentration measurements
made in parts per million units (ppm), it is not good enough for lower level NOx measurements. To
overcome this limitation the 9110T is designed with two physical measurement ranges:
A LOW range that measures concentration from 0 ppb to 2,000 ppb with a resolution of 0.27 ppb per
count.
A HIGH range that measures the full 20,000 ppb range of the analyzer.
The analyzer’s CPU chooses the appropriate range based on how the user sets up the reporting ranges for
the instruments analog outputs when an analog range is selected with a lower limit between 0 and 2000
ppb the analyzer will utilize its low physical range. When an analog range is in use that has a reporting
range with an upper limit set between 2001 and 20,000 ppb the instrument will operate in its high
physical range.
Once both ranges have been using the same span gas values the analyzer’s front panel will accurately
report concentrations between 0 and 20,000 ppb, seamlessly switching between the low and high
physical ranges regardless of the selected analog reporting range.
5.4.2. 9110T ANALOG OUTPUT REPORTING RANGES
For applications using chart recorders or other analog recording devices, the 9110T's 20,000 ppb
physical range can cause resolution problems. For example, in an application where the expected
concentrations of NO, NO2 and NOx are typically less than 500 ppb, the full scale of expected values is
only 2.5% of the instrument’s 20,000 ppb physical range. The corresponding output signal would then
only be recorded across 2.5% of the range of the recording device.
The 9110T solves this problem by allowing the user to select a reporting range for the analog outputs
that only includes that portion of the physical range that covers the specific application. This increases
the reliability and accuracy of the analyzer by avoiding additional gain-amplification circuitry.
Note
Only the reporting range of the analog outputs is scaled.
Both the DAS values stored in the CPU’s memory and the concentration
values reported on the front panel are unaffected by the settings chosen
for the reporting range(s) of the instrument.
Teledyne Analytical Instruments
96
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.4.2.1. Analog Output Ranges for NOx, NO and NO2 Concentration
The analyzer has three active analog output signals related to NOx, NO and NO2 concentration,
accessible through a connector on the rear panel.
ANALOG OUT
NOx Concentration
NO Concentration
NO2 Concentration
Test Channel
or O2 concentration
(if optional O2 sensor
is installed)
A1
A2
A3
A4
+
-
+
-
+
-
+
-
Figure 5-1:
Analog Output Connector Pin Out
The A1, A2 and A3 channels output a signal that is proportional to the NOx, NO and NO2 concentrations
of the sample gas, respectively. The 9110T can be set so that these outputs operate in one of the three
following modes: single range mode, independent range mode, or automatic range mode (Section 5.4.3).
Additionally, the signal levels of outputs A1, A2 and A3 outputs can be:
Configured full scale outputs of: 0 - 0.1 VDC; 0 – 1 VDC; 0 – 5 VDC or; 0 – 10 VDC.
Equipped with optional 0-20 mADC current loop drivers (see Section 3.3.1.4 ) and configured for any
current output within that range analog output (e.g. 0-20 mA, 2-20 mA, 4-20 mA, etc.).
Together these two set of parameters allow the user a great deal of flexibility in how the instrument
reports NOx , NO and NO2 concentration to external devices. For example, Using the IND mode the
following configuration could be created:
A1 OUTPUT: NOx Output Signal = 4 – 20 mA representing 0-1000 ppb concentration values
A2 OUTPUT: NO Output Signal = 0 – 10 VDC representing 0-500 ppb concentration values.
A3 OUTPUT: NO2 Output Signal = 0 – 5 VDC representing 0-500 ppb concentration values.
The user may also add a signal offset independently to each output (see Section 5.9.3.9) to match the
electronic input requirements of the recorder or data logger to which the output is connected.
IMPORTANT
IMPACT ON READINGS OR DATA
The instrument does not remember upper range limits settings associated
with the individual modes. Changes made to the range limits (e.g. 400 ppb
600 ppb) when in one particular mode will alter the range limit settings
for the other modes.
When switching between reporting range modes, ALWAYS check and
Teledyne Analytical Instruments
97
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
reset the upper range limits for the new mode selection..
5.4.2.2. Analog Output Reporting Range Default Settings
The default setting for these the reporting ranges of the analog output channels A1, A2 and A3 are:
SNGL mode
0 to 500.0 ppb
0 to 5 VDC
5.4.3. SETUP RNGE MODE
Single range mode (SNGL) reports all three of the NOx gas concentrations using the same reporting
range span (see Section 5.4.3.1).
Independent range mode (IND) allows the NOx, NO and NO2 analog outputs to be set with different
reporting range spans (see Section 5.4.3.2).
Automatic range mode (AUTO) allows the analyzer to automatically switch the reporting range between
two user upper span limits (designated LOW and HIGH) based on the actual concentrations being
measured for each (see Section 5.4.3.3). These are not the same as the analyzer’s low and high physical
ranges.
5.4.3.1. SETUP RNGE MODE SNGL: Configuring the 9110T Analyzer for Single
Range Mode
Note
When the single range mode is selected (SNGL), all analog NOx, NO and NO2 concentration outputs
(A1, A2 and A3) are slaved together and set to the same reporting range limits (e.g. 500.0 ppb). This
reporting range can be set to any value between 100 ppb and 20,000 ppb.
Although all three NOx outputs share the same concentration reporting range, the electronic signal ranges
of the analog outputs may still be configured for different values (e.g. 0-5 VDC, 0-10 VDC, etc; see
Section 5.9.3.1).
To select SNGL range mode and to set the upper limit of the range, press:
Teledyne Analytical Instruments
98
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.4.3.2. SETUP RNGE MODE IND: Configuring the 9110T Analyzer for
Independent Range Mode
The independent range mode (IND) assigns the three NOx, NO and NO2 concentrations to individual
analog output channels. In IND range mode the RANGE test function displayed on the front panel will
then be replaced by three separate functions:
Table 5-1:
IND Mode Analog Output Assignments
TEST
FUNCTION
CONCENTRATION
REPORTED
ANALOG OUTPUT
CHANNEL
RANGE1
RANGE2
RANGE3
NOx
NO
A1
A2
A3
NO2
Each can be configured with a different reporting range upper limit and analog signal span:
Teledyne Analytical Instruments
99
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
EXAMPLE:
NOx Concentration – RANGE1 Set for 0-800 ppb & output A1 set for 0-10 VDC
NO Concentration – RANGE2 Set for 0-200 ppb & output A2 set for 0-5 VDC
NO2 Concentration – RANGE3 Set for 0-400 ppb & output A3 set for 0-5 VDC
Setting analog range limits to different values does not affect the instrument’s calibration.
To select the IND range mode, press the following buttons:
SAMPLE
RANGE=500.0 PPM
NOX= XXXX
SETUP
<TST TST> CAL
Concentration field
displays all gases.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
EXIT
SETUP X.X
RANGE MODE MENU
MODE SET UNIT DIL
SETUP X.X
RANGE MODE:SNGL
SNGL IND AUTO
ENTR EXIT
ENTR EXIT
EXIT
SETUP X.X
RANGE MODE:IND
SNGL IND AUTO
SETUP X.X
RANGE MODE MENU
MODE SET UNIT DIL
Teledyne Analytical Instruments
100
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
To set the upper range limit for each independent reporting range, press:
.
Teledyne Analytical Instruments
101
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.4.3.3. SETUP RNGE MODE AUTO: Configuring the 9110T Analyzer for Auto
Range Mode
In AUTO range mode, the analyzer automatically switches the reporting range between two
user-defined ranges (LOW and HIGH). The same low and high span settings are applied equally to
NO, NO2 and NOX readings.
The unit will switch from LOW range to HIGH range when either the NO, or NOX concentration
exceeds 98% of the low range span.
The unit will return from HIGH range back to LOW range once both the NO and NOX concentrations
fall below 75% of the low range span.
IMPORTANT
IMPACT ON READINGS OR DATA
The LOW & HIGH ranges referred to here are NOT the same as the low &
high physical ranges referred to in Section 5.4.1.
Also the RANGE test function displayed on the front panel will be replaced by two separate functions:
RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.
The LOW/HIGH range status is also reported through the external, digital status bits (Section 3.3.1.4).
To set individual ranges press the following menu sequence.
Teledyne Analytical Instruments
102
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
Teledyne Analytical Instruments
103
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.4.3.4. SETUP RNGE UNIT: Setting the Reporting Range Units of Measure
The 9110T can display and report concentrations in ppb, ppm, ug/m3, mg/m3 units. Changing units
affects all of the COMM port values, and all of the display values for all reporting ranges. To change the
units of measure press:
IMPORTANT
IMPACT ON READINGS OR DATA
Concentrations displayed in mg/m3 and ug/m3 use 0C@ 760 mmHg for
Standard Temperature and Pressure (STP).
Consult your local regulations for the STP used by your agency.
(Example: US EPA uses 25C as the reference temperature).
Once the Units of Measurement have been changed from volumetric (ppb
or ppm) to mass units (ug/m3 or mg/m3) the analyzer MUST be
recalibrated, as the “expected span values” previously in effect will no
longer be valid.
Simply entering new expected span values without running the entire
calibration routine is not sufficient. This will also counteract any
discrepancies between STP definitions.
5.4.3.5. SETUP RNGE DIL: Using the Optional Dilution Ratio Feature
The dilution ratio feature is a software utility option designed for applications where the sample gas is
diluted before being analyzed by the 9110T. Typically this occurs in continuous emission monitoring
(CEM) applications where the quality of gas in a smoke stack is being tested and the sampling method
Teledyne Analytical Instruments
104
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
used to remove the gas from the stack dilutes the gas. Once the degree of dilution is known, this feature
allows the user to add an appropriate scaling factor to the analyzer’s NO, NO2 and NOx concentration
calculations so that the measurement range and concentration values displayed on the instrument’s front
panel display and reported via the instruments various outputs reflect the undiluted values.
Using the Dilution Ratio option is a 4-step process:
1. Select the appropriate units of measure (see Section 5.4.3.4).
2. Select the reporting range mode and set the reporting range upper limit (see
Section 5.4.2).
Ensure that the upper span limit entered for the reporting range is the maximum expected
concentration of the UNDILUTED gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1
part of sample gas):
1. Calibrate the analyzer.
Ensure that the calibration span gas is either supplied through the same dilution system as
the sample gas or has an appropriately lower actual concentration.
EXAMPLE: If the reporting range limit is set for 100 ppm and the dilution ratio of the sample gas is 20
gain, either:
Teledyne Analytical Instruments
105
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
a span gas with the concentration of 100 ppm can be used if the span gas passes through
the same dilution steps as the sample gas, or;
a 5 ppm span gas must be used if the span gas IS NOT routed through the dilution system.
5.5. SETUP PASS: PASSWORD PROTECTION
The 9110T provides password protection of the calibration and setup functions to prevent unauthorized
adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt
the user for a password anytime a password-protected function (e.g., SETUP) is selected. This allows
normal operation of the instrument, but requires the password (101) to access to the menus under
SETUP. When PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is enabled or disabled,
a password (default 818) is required to enter the VARS or DIAG menus in the SETUP>MORE menu.
There are three levels of password protection, which correspond to operator, maintenance and
configuration functions. Each level allows access to all of the functions in the previous level.
Table 5-2: Password Levels
PASSWORD
Null (000)
101
LEVEL
MENU ACCESS ALLOWED
Operation
All functions of the MAIN menu: TEST, GEN, initiate SEQ , MSG, CLR
Configuration/Maintenance Access to Primary Setup and Secondary SETUP Menus when PASSWORD is
enabled.
818
Configuration/Maintenance Access to Secondary SETUP Submenus VARS and DIAG whether PASSWORD
is enabled or disabled.
To enable passwords, press:
Teledyne Analytical Instruments
106
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
Example: If all passwords are enabled, the following touchscreen control sequence would be required to
enter the VARS or DIAG submenus:
Note
The instrument still prompts for a password when entering the VARS and
DIAG menus, even if passwords are disabled, but it displays the default
password (818) upon entering these menus. In this case, the user only
has to press ENTR to access the password-protected menus.
In order to disable the PASSWORD feature after it has been turned ON, the SETUP menu first requires a
password; once the password has been input and the ENTR button pressed, the PRIMARY SETUP
MENU appears, and now the PASS menu can be accessed, where pressing the ON button turns
PASSWORD ENABLE back to OFF, and pressing the ENTR button accepts the change (Table 5-2).
Teledyne Analytical Instruments
107
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.6. SETUP CLK: SETTING THE INTERNAL TIME-OF-DAY
CLOCK
The 9110T has an internal clock for setting the time and day; it’s speed can be adjusted to compensate
for faster or slower CPU clocks. Press SETUP>CLK to access the clock.
5.6.1. SETTING THE TIME OF DAY
The time-of-day feature of the internal clock supports the DURATION step of the automatic calibration
(ACAL) sequence feature, has a built-in clock for the AutoCal timer, for the time TEST function, and
for time stamps on COM port messages and on DAS data entries.
To set the clock’s time and date, press:
Teledyne Analytical Instruments
108
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.6.2. ADJUSTING THE INTERNAL CLOCK’S SPEED
In order to compensate for CPU clocks that run fast or slow, you can adjust a variable called
CLOCK_ADJ to speed up or slow down the clock by a fixed amount every day.
The CLOCK_ADJ variable is accessed via the VARS submenu: To change the value of this variable,
press:
Teledyne Analytical Instruments
109
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.7. SETUP COMM: COMMUNICATIONS PORTS
This section introduces the communications setup menu; Section 6 provides the setup instructions and
operation information. Press SETUP>ENTR>MORE>COMM to arrive at the communications menu.
Figure 5-2.
SETUP – COMM Menu
5.7.1. ID (MACHINE IDENTIFICATION)
In the SETUP>MORE>COMM menu press ID to display and/or change the Machine ID, which must be
changed to a unique identifier (number) when more than one instrument of the same model is used:
in an RS-232 multidrop configuration (Sections 3.3.1.8)
on the same Ethernet LAN (Section 6.5)
when applying MODBUS protocol (Section 6.7.1)
when applying Hessen protocol (Section 6.7.2)
The default ID for the Model 9110T is 0200. Press any button(s) in the MACHINE ID menu (Figure 5-3)
until the Machine ID in the Parameter field displays the desired identifier.
Teledyne Analytical Instruments
110
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
SETUP X.X
COMMUNICATIONS MENU
ID INET COM1 COM2
EXIT
Toggle to cycle
through the available
character set: 0-9
ENTR accepts the new
SETUP X.
MACHINE ID: 200 ID
settings
EXIT ignores the new
0
2
0
0
ENTR EXIT
settings
Figure 5-3.
COMM– Machine ID
The ID can be any unique 4-digit number and can also be used to identify analyzers in any number of
ways (e.g. location numbers, company asset number, etc.)
5.7.2. INET (ETHERNET)
Use SETUP>COMM>INET to configure Ethernet communications, whether manually or via DHCP.
Please see Section 6.5.2 for configuration details.
5.7.3. COM1[COM2] (MODE, BAUDE RATE AND TEST PORT)
Use the SETUP>COMM>COM1[COM2] menus to:
configure communication modes (Section 6.2.1)
view/set the baud rate (Section 6.2.2)
test the connections of the com ports (Section 6.2.3).
Configuring COM1 or COM2 requires setting the DCE DTE switch on the rear panel. Section 6.1
provides DCE DTE information.
5.8. SETUP VARS: VARIABLES SETUP AND DEFINITION
Through the SETUP>MORE>VARS menu there are several user-adjustable software variables that
define certain operational parameters. Usually, these variables are automatically set by the instrument’s
firmware, but can be manually re-defined using the VARS menu.
Table 5-3 lists all variables that are available within the 818 password protected level. See Appendix A2
for a detailed listing of all of the 9110T variables that are accessible through the remote interface.
Teledyne Analytical Instruments
111
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
Table 5-3: Variable Names (VARS)
DESCRIPTION
VARS
DEFAULT
VALUES
ALLOWED
VALUES
NO.
0
VARIABLE
Changes the Internal Data Acquisition System (DAS)
HOLDOFF timer:
May be set for
intervals
between
0.5 – 20 min
No data is stored in the DAS channels during situations
when the software considers the data to be questionable
such as during warm up of just after the instrument returns
from one of its calibration mode to SAMPLE Mode.
DAS_HOLD_OFF
MEASURE_MODE
15 min.
NO;
NOx;
Selects the gas measurement mode in which the
instrument is to operate. NOx only, NO only or NOx and NO
simultaneously.
NOx–NO
1
NOx–NO
Selects which gas measurement is displayed when the
STABIL test function is selected
NO; NOx;
NO2;
STABIL_GAS
TPC_ENABLE
NOx
2
3
Enables or disables the Temperature and Pressure
Compensation (TPC) feature (Section 13.9.2).
ON/OFF
ON/OFF
ON
Dynamic zero automatically adjusts offset and slope of the
NO and NOX response when performing a zero point
calibration during an AutoCal (see Section 9.5).
DYN_ZERO 1
DYN_SPAN 1
IZS_SET
OFF
OFF
4
5
6
7
8
Dynamic span automatically adjusts the offsets and slopes
of the NO and NOx response when performing a sp point
calibration during an AutoCal (see Section 9.5).
ON/OFF
Sets the internal span gas generator’s permeation tube
oven temperature. Changing this temperature will impact
the NO2 permeation rate (Section 3.3.2.5).
30C - 70C
51C
AUTO
0 sec
Allows the user to set the number of significant digits to the
CONC_PRECISION right of the decimal point display of concentration and
AUTO, 1, 2,
3, 4
stability values.
Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.
-60 to +60
s/day
CLOCK_ADJ
CAL_ON_NO2
Allows turning ON and OFF the ability to span the analyzer
with NO2, in which case the instrument acts as if NO and
NOX are spanned, even though it is supplied with NO2.
ON or OFF
OFF
9
The NO2 concentration is then zero by default.
1
Use of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency.
IMPORTANT
IMPACT ON READINGS OR DATA
There are more VARS available when using the password, 929, for
configuration. Use caution when pressing any buttons while in this
setup. Any changes made may alter the performance of the instrument
or cause the instrument to not function properly. Note that if there is an
accidental change to a setup parameter, press EXIT to discard the
changes.
Note
There is a 2-sec latency period between when a VARS value is changed
and the new value is stored into the analyzer’s memory. DO NOT turn
the analyzer off during this period or the new setting will be lost.
Teledyne Analytical Instruments
112
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
To access and navigate the VARS menu, use the following button sequence:
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
<TST TST> CAL
Concentration field
displays all gases.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
In all cases:
EXIT discards the new
setting.
COMM VARS DIAG
EXIT
ENTR accepts the
SETUP X.X
ENTER PASSWORD:818
new setting.
8
1
8
ENTR EXIT
Toggle these
buttons to enter
the correct
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
PASSWORD.
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
.0 ENTR EXIT
1
5
Toggle these buttons to set
the DAS HOLDOFF time
period in minutes
(MAX = 20 minutes).
SETUP X.X
1) MEASURE_MODE=NOXNO
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
MEASURE_MODE=NOX-NO
ENTR EXIT
PREV NEXT
Toggle these keys to
choose the gas(es) for
analyzer’s measurement
mode.
SETUP X.X
2) STABIL_GAS=NOX
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
STABIL_GAS=NOX
NO
NO2 NOX O2
ENTR EXIT
Use these buttons to select
which gas will be reported
by the STABIL test
function.
SETUP X.X
3) TPC_ENABLE
(O2 is only available if the
optional O2 sensor is
installed)
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
TPC_ENABLE=OFF
DYN_ZERO=OFF
OFF
ENTR EXIT
Toggle this button to turn
the Temperature /
Pressure compensation
feature
SETUP X.X
4) DYN_ZERO=OFF
ON/OFF.
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
OFF
ENTR EXIT
Toggle this button to turn
the Dynamic Zero
calibration feature ON/
OFF.
SETUP X.X
5) DYN_SPAN=OFF
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
DYN_SPAN=OFF
OFF
ENTR EXIT
Toggle this button to turn
the Dynamic Span
calibration feature ON/
OFF.
DO NOT CHANGE
these settings unless
specifically instructed to by
Teledyne API’s Customer
Service
SETUP X.X
6) IZS_SET=50.0 DegC
PREV NEXT JUMP
EDIT PRNT EXIT
personnel.
SETUP X.X
7) CONC_PRECISION=AUTO
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
CONC_PRECISION=AUTO
ENTR EXIT
AUTO
1
2
3
4
Use these buttons to select
the precision of the
SETUP X.X
8) CLOCK_ADJUST=0 Sec/Day
concentration display.
PREV
JUMP
EDIT ENTR EXIT
SETUP X.X
CLOCK_ADJUST=0 Sec/Day
+
0
0
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+).
SETUP X.X
9) CAL_ON_NO2=OFF
EDIT PRNT EXIT
PREV NEXT JUMP
SETUP X.X
CAL_ON_NO2=OFF
OFF
ENTR EXIT
Toggle this button to turn
ON/OFF the analyzer’s
ability to be calibrated
using NO2 span gas.
Teledyne Analytical Instruments
113
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.9. SETUP DIAG: DIAGNOSTICS FUNCTIONS
A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. The
parameters are dependent on firmware revision (see Appendix A). These tools can be used in a variety
of troubleshooting and diagnostic procedures and are referred to in many places of the maintenance and
troubleshooting sections of this manual.
The various operating modes available under the DIAG menu are:
Table 5-4: Diagnostic Mode (DIAG) Functions
Front Panel
Mode Indicator
MANUAL
SECTION
DIAG SUBMENU
SIGNAL I/O
SUBMENU FUNCTION
Allows observation of all digital and analog signals
in the instrument. Allows certain digital signals such
as valves and heaters to be toggled ON and OFF.
12.1.3
DIAG I/O
When entered, the analyzer performs an analog
12.7.6.1
ANALOG OUTPUT output step test. This can be used to calibrate a
DIAG AOUT
chart recorder or to test the analog output accuracy.
The signal levels of the instruments analog outputs
may be calibrated (either individually or as a group).
ANALOG I/O
5.9.2
Various electronic parameters such as signal span,
CONFIGURATION
DIAG AIO
and offset are available for viewing and
configuration.
TEST CHAN
OUTPUT
Selects one of the available test channel signals to
output over the A4 analog output channel.
5.9.4
0
DIAG TCHN
DIAG OPTIC
When activated, the analyzer performs an optic test,
which turns on an LED located inside the sensor module
near the PMT (Figure 13-20). This diagnostic tests the
response of the PMT without having to supply span gas.
OPTIC TEST
When activated, the analyzer performs an electrical test,
which generates a current intended to simulate the PMT
output to verify the signal handling and conditioning of the
PMT preamp board.
ELECTRICAL
TEST
0
DIAG ELEC
Allows the user to manually turn the O3 generator on or
off. During initial power up TMR (timer) is displayed while
the Ozone brick remains off for the first 30 minutes.
OZONE GEN
OVERRIDE 1
0
DIAG OZONE
DIAG FCAL
FLOW
This function is used to calibrate the gas flow output
signals of sample gas and ozone supply.
9.7
CALIBRATION 1
1
These settings are retained after exiting DIAG mode.
To access the various DIAG submenus, press the following buttons:
Teledyne Analytical Instruments
114
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
Figure 5-4:
Accessing the DIAG Submenus
Teledyne Analytical Instruments
115
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.9.1. SIGNAL I/O
The signal I/O diagnostic mode allows a user to review and change the digital and analog input/output
functions of the analyzer. Refer to Appendix A for a complete list of the parameters available for review
under this menu.
IMPORTANT
IMPACT ON READINGS OR DATA
Any changes of signal I/O settings will remain in effect only until the
signal I/O menu is exited. Exceptions are the ozone generator override
and the flow sensor calibration, which remain as entered when exiting.
Access the signal I/O test mode from the DIAG Menu (Figure 5-4), then press:
DIAG
SIGNAL I / O
Press NEXT & PREV to
move between signal
types.
PREV NEXT JUMP
ENTR EXIT
Press JUMP to go
directly to a specific
signal
DIAG I / O
0) EXT_ZERO_CAL=OFF
PREV NEXT JUMP
PRNT EXIT
See Appendix A-4 for
a complete list of
available SIGNALS
EXAMPLE
DIAG I / O
JUMP TO: 12
EXAMPLE:
1
2
ENTR EXIT
Enter 12 to Jump to
12) ST_SYSTEM_OK=ON
DIAG I / O
12) ST_SYSTEM_OK = ON
Exit to return
to the
DIAG menu
PREV NEXT JUMP
ON PRNT EXIT
Pressing the PRNT button will send a formatted
printout to the serial port and can be captured
with a computer or other output device.
Toggle ON/(OFF) button to
change status.
Teledyne Analytical Instruments
116
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.2. ANALOG OUTPUT (DIAG AOUT)
Analog Output is used to verify functionality and accuracy of the analog outputs. The test forces all
analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20%
increments. This test is useful to verify the operation of the data logging/recording devices attached to
the analyzer.
Section 12.7.6.1 presents instructions for use in troubleshooting and service.
5.9.3. ANALOG I/O CONFIGURATION (DIAG AIO)
The 9110T analyzer comes equipped with four analog outputs. The first three outputs (A1 A2, & A3)
carry analog signals that represent the currently measured concentrating of NOx, NO and NO2 (see
Section 5.4.2.1). The fourth output (A4) outputs a signal that can be set to represent the current value of
one of several test functions (see Table 5-9).
The following table lists the analog I/O functions that are available in the 9110T analyzer.
Table 5-5: DIAG - Analog I/O Functions
MANUAL
SECTION
SUB MENU
FUNCTION
Initiates a calibration of the A1, A2, A3 and A4 analog output channels that
determines the slope and offset inherent in the circuitry of each output.
5.9.3.1
AOUT CALIBRATED
These values are stored and applied to the output signals by the CPU
automatically.
Sets the basic electronic configuration of the A1 output (NOx Concentration).
There are four options:
RANGE1: Selects the signal type (voltage or current loop) and level of the
output
CONC_OUT_1
REC OFS: Allows them input of a DC offset to let the user manually adjust
the output level
5.9.2
AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature
CALIBRATED: Performs the same calibration as AOUT CALIBRATED, but
on this one channel only.
CONC_OUT_2
CONC_OUT_3
Same as for CONC_OUT_1 but for analog channel A2 (NO Concentration)
Same as for CONC_OUT_1 but for analog channel A3 (NO2 Concentration)
Same as for CONC_OUT_1 but for analog channel A4 (TEST CHANNEL)
Initiates a calibration of the A-to-D Converter circuit located on the Motherboard.
5.9.4
TEST OUTPUT
AIN CALIBRATED
5.9.3.10
XIN1
.
.
.
For each of 8 external analog inputs channels, shows the gain, offset,
engineering units, and whether the channel is to show up as a Test
function.
5.9.3.11
XIN8
Teledyne Analytical Instruments
117
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
To access the ANALOG I/O CONFIGURATION sub menu, press:
Figure 5-5:
Accessing the Analog I/O Configuration Submenus
Teledyne Analytical Instruments
118
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.3.1. Analog Output Voltage / Current Range Selection
In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals. Several other
output ranges are available (see Table 5-6). Each range is usable from -5% to +5% of the rated span.
Table 5-6: Analog Output Voltage Range Min/Max
RANGE NAME
RANGE SPAN
0-100 mVDC
0-1 VDC
MINIMUM OUTPUT
-5 mVDC
MAXIMUM OUTPUT
105 mVDC
0.1V
1V
-0.05 VDC
1.05 VDC
5V
0-5 VDC
-0.25 VDC
5.25 VDC
10V
0-10 VDC
-0.5 VDC
10.5 VDC
The default offset for all VDC ranges is 0-5 VDC.
CURR
0-20 mA
0 mA
20 mA
While these are the physical limits of the current loop modules, typical applications use 2-20 mA or 4-20 mA for the lower and upper
limits. Please specify desired range when ordering this option.
The default offset for all current ranges is 0 mA.
To change the output type and range, select the ANALOG I/O CONFIGURATION submenu (see
Figure 5-5) then press:
Teledyne Analytical Instruments
119
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.9.3.2. Calibration of the Analog Outputs
Analog output calibration should be carried out on first startup of the analyzer (performed in the factory
as part of the configuration process) or whenever recalibration is required. The analog outputs can be
calibrated automatically, either as a group or individually, or adjusted manually.
In its default mode, the instrument is configured for automatic calibration of all channels, which is useful
for clearing any analog calibration warnings associated with channels that will not be used or connected
to any input or recording device, e.g., data logger.
Manual calibration should be used for the 0.1V range or in cases where the outputs must be closely
matched to the characteristics of the recording device. The AUTOCAL feature must be disabled first for
manual calibration.
5.9.3.3. Enabling or Disabling the AutoCal for an Individual Analog Output
To enable or disable the AutoCal feature for an individual analog output, elect the ANALOG I/O
CONFIGURATION submenu (see Figure 5-5) then press:
Teledyne Analytical Instruments
120
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.3.4. Automatic Group Calibration of the Analog Outputs
IMPORTANT
IMPACT ON READINGS OR DATA
Manual calibration should be used for any analog output set for a 0.1V
output range or in cases where the outputs must be closely matched to
the characteristics of the recording device. (See Sections 5.9.3.2, 5.9.3.3,
and 5.9.3.6).
IMPORTANT
IMPACT ON READINGS OR DATA
Before performing this procedure, ensure that the AUTO CAL for each
analog output is enabled. (See Section 5.9.3.3).
To calibrate the outputs as a group with the AOUTS CALIBRATION command, select the ANALOG
I/O CONFIGURATION submenu (see Figure 5-5) then press:
5.9.3.5. Automatic Individual Calibration of the Analog Outputs
To use the AUTO CAL feature to initiate an automatic calibration for an individual analog output, select
the ANALOG I/O CONFIGURATION submenu (see Figure 5-5) then press:
Teledyne Analytical Instruments
121
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
Teledyne Analytical Instruments
122
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.3.6. Manual Calibration of the Analog Outputs Configured for Voltage Ranges
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Note
The menu for manually adjusting the analog output signal level will only
appear if the AUTO-CAL feature is turned off for the channel being
adjusted. (See Section 5.9.3.3).
Calibration is performed with a voltmeter connected across the output terminals and by changing the
actual output signal level using the front panel buttons in 100, 10 or 1 count increments. See Figure 3-8
for pin assignments and diagram of the analog output connector.
V
+DC Gnd
Figure 5-6:
Setup for Checking / Calibrating DCV Analog Output Signal Levels
Table 5-7: Voltage Tolerances for the TEST CHANNEL Calibration
MINIMUM
FULL
ZERO
SPAN
SPAN VOLTAGE
ADJUSTMENT
SCALE
TOLERANCE
TOLERANCE
(1 count)
0.1 VDC
1 VDC
±0.0005V
±0.001V
±0.002V
±0.004V
90 mV
900 mV
4500 mV
4500 mV
±0.001V
±0.001V
±0.003V
±0.006V
0.02 mV
0.24 mV
1.22 mV
2.44 mV
5 VDC
10 VDC
Teledyne Analytical Instruments
123
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
To adjust the signal levels of an analog output channel manually, select the ANALOG I/O
CONFIGURATION submenu (see Figure 5-5) then press:
Teledyne Analytical Instruments
124
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.3.7. Manual Adjustment of Current Loop Output Span and Offset
A current loop option may be purchased for the A1, A2 and A3 Analog outputs of the analyzer. This
option places circuitry in series with the output of the D-to-A converter on the motherboard that changes
the normal DC voltage output to a 0-20 milliamp signal (See Section 3.3.1.4).
1. The outputs can be ordered scaled to any set of limits within that 0-20 mA range,
however most current loop applications call for either 0-20 mA or 4-20 mA range
spans.
2. All current loop outputs have a +5% over range. Ranges whose lower limit is set above 1 mA also have
a –5% under range.
To switch an analog output from voltage to current loop, follow the instructions in Section 5.9.3.1 (select
CURR from the list of options on the “Output Range” menu).
Adjusting the signal zero and span levels of the current loop output is done by raising or lowering the
voltage output of the D-to-A converter circuitry on the analyzer’s motherboard. This raises or lowers the
signal level produced by the current loop option circuitry.
The software allows this adjustment to be made in 100, 10 or 1 count increments. Since the exact
amount by which the current signal is changed per D-to-A count varies from output-to-output and
instrument–to–instrument, you will need to measure the change in the signal levels with a separate,
current meter placed in series with the output circuit. See Figure 3-8 for pin assignments and diagram of
the analog output connector.
Figure 5-7:
Setup for Checking / Calibration Current Output Signal Levels Using an Ammeter
CAUTION – GENERAL SAFETY HAZARD
Do not exceed 60 V peak voltage between current loop outputs and instrument ground.
Teledyne Analytical Instruments
125
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
To adjust the zero and span signal levels of the current outputs, select the ANALOG I/O
CONFIGURATION submenu (see Figure 5-5) then press:
Teledyne Analytical Instruments
126
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
An alternate method for measuring the output of the Current Loop converter is to connect a 250 ohm
1% resistor across the current loop output in lieu of the current meter (see Figure 3-8 for pin
assignments and diagram of the analog output connector). This allows the use of a voltmeter connected
across the resistor to measure converter output as VDC or mVDC.
V
+DC Gnd
Figure 5-8:
Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following values:
Table 5-8: Current Loop Output Check
Voltage across
Resistor for 2-20 mA
Voltage across
Resistor for 4-20 mA
% FS
0
500 mVDC
1000 mVDC
5000 mVDC
100
5000 mVDC
5.9.3.8. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF
In its default configuration, a ± 5% over-range is available on each of the 9110T’s analog outputs. This
over-range can be disabled if your recording device is sensitive to excess voltage or current.
To turn the over-range feature on or off, select the ANALOG I/O CONFIGURATION submenu (see
Figure 5-5) then press:
Teledyne Analytical Instruments
127
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.9.3.9. ADDING A RECORDER OFFSET TO AN ANALOG OUTPUT
Some analog signal recorders require that the zero signal be significantly different from the baseline of
the recorder in order to record slightly negative readings from noise around the zero point. This can be
achieved in the 9110T by defining a zero offset, a small voltage (e.g., 10% of span).
To add a zero offset to a specific analog output channel, select the ANALOG I/O CONFIGURATION
submenu (see Figure 5-5) then press:
Teledyne Analytical Instruments
128
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
5.9.3.10. AIN Calibration
This is the submenu to conduct a calibration of the 9110T analyzer’s analog inputs. This calibration
should only be necessary after major repair such as a replacement of CPU, motherboard or power
supplies.
To perform an analog input calibration, select the ANALOG I/O CONFIGURATION submenu (see
Figure 5-5) then press:
Teledyne Analytical Instruments
129
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
From the
DIAG
PREV NEXT
EXIT
EXIT
<SET
CAL
Continue pressing
until you reach the
output to be configured
<SET
EXIT
Firmware automatically performs a zero point calibration
of the Motherboard’s analog Inputs
Firmware automatically performs a span point calibration
of the Motherboard’s analog Inputs
Perform
Troubleshooting or call
Teledyne API’s
<SET
EXIT
Customer Service
5.9.3.11. External Analog Inputs (XIN1…XIN8) Option Configuration
To configure the analyzer’s optional external analog inputs, define for each channel:
gain (number of units represented by 1 volt)
offset (volts)
engineering units to be represented in volts (each press of the touchscreen button scrolls the list of
alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions
Teledyne Analytical Instruments
130
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
These parameters can also be captured in the internal Data Acquisition System (DAS); refer to Appendix
A for Analog-In DAS parameters.
To adjust settings for the Analog Inputs option parameters press:
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
Press SET> to scroll to the first
channel. Continue pressing SET>
to view each of 8 channels.
< SET SET> CAL
DIAG AIO
XIN1:1.00,0.00,V,OFF
Press EDIT at any channel
< SET SET> EDIT
EXIT
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).
DIAG AIO
XIN1 GAIN:1.00V/V
SET> EDIT
EXIT
DIAG AIO
XIN1 OFFSET:0.00V
DIAG AIO
XIN1 GAIN:1.00V/V
< SET SET> EDIT
EXIT
+
0
0
1
.0
0
ENTR EXIT
DIAG AIO
XIN1 UNITS:V
Press to change
Gain value
< SET SET> EDIT
EXIT
EXIT
DIAG AIO
< SET
XIN1 DISPLAY:OFF
EDIT
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
Teledyne Analytical Instruments
131
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
5.9.4. TEST CHAN OUTPUT (SELECTING A TEST CHANNEL FUNCTION
FOR OUTPUT A4)
The test functions available to be reported are listed in Table 5-9:
Table 5-9: Test Channels Functions available on the 9110T’s Analog Output
TEST CHANNEL
NONE
DESCRIPTION
ZERO
FULL SCALE
TEST CHANNEL IS TURNED OFF
The output of the PMT detector
converted to a 0 to 5 VDC scale.
PMT DETECTOR
0 mV
5000 mV 1
The flow rate of O3 through the analyzer
as measured by the O3 flow sensor.
OZONE FLOW
SAMPLE FLOW
0 cm3/min
0 cm3/min
1000 cm3/min
1000 cm3/min
The calculated flow rate for sample gas
through the analyzer.
The pressure of the sample gas
measured upstream of the Auto Zero
Valve.
SAMPLE PRESSURE
RCELL PRESSURE
0 Hg-In-A
40 "Hg-In-A
The pressure of gas inside the reaction
cell of the sensor module.
0 Hg-In-A
40 Hg-In-A
The temperature of gas inside the
reaction cell of the sensor module.
RCELL TEMP
0 C
70 C
MANIFOLD TEMP
Not used in the Model 9110T.
The temperature of the permeation tube
oven of the optional internal span gas
generator.
IZS TEMP
0 C
70 C
CONV TEMP
PMT TEMP
The temperature NO2 NO converter.
0 C
500 C
The temperature inside PMT.
0 C
50 C
The temperature inside the 9110T’s
chassis.
BOX TEMP
0 C
70 C
Represents the output voltage of the
PMT's high voltage power supply.
HVPS VOLTAGE
0 mV
5000 mV 1
Once a function is selected, the instrument not only begins to output a signal on the analog output, but
also adds TEST to the list of test functions viewable via the front panel display.
Teledyne Analytical Instruments
132
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Setup Menu
To activate the TEST Channel and select a function, press:
5.9.5. OPTIC TEST
The OPTIC TEST function tests the response of the PMT sensor by turning on an LED located in the
cooling block of the PMT. The analyzer uses the light emitted from the LED to test its photo-electronic
subsystem, including the PMT and the current to voltage converter on the pre-amplifier board. To ensure
Teledyne Analytical Instruments
133
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Setup Menu
that the analyzer measures only the light coming from the LED, the analyzer should be supplied with
zero air. The optic test should produce a PMT signal of about 2000±1000 mV.
Section 12.7.12.1 presents instructions for use in troubleshooting and service.
IMPORTANT
IMPACT ON READINGS OR DATA
This is a coarse test for functionality and not an accurate calibration tool.
The resulting PMT signal can vary significantly over time and also
changes with low-level calibration.
5.9.6. ELECTRICAL TEST
The ELECTRICAL TEST function creates a current, which substitutes the PMT signal, and feeds it into
the preamplifier board. This signal is generated by circuitry on the pre-amplifier board itself and tests the
filtering and amplification functions of that assembly along with the A/D converter on the motherboard.
It does not test the PMT itself. The electrical test should produce a PMT signal of about 2000 ±1000 mV.
Section 12.7.12.2 presents instructions for use in troubleshooting and service.
5.9.7. OZONE GEN OVERRIDE
This feature is used to manually turn the ozone generator off and on. Read Section 13.2.3 to understand
the ozone generator, and refer to Section 12.7.15.1 for instructions on using the override feature in
troubleshooting and service.
5.9.8. FLOW CALIBRATION
This function is used to calibrate the gas flow output signals of sample gas and ozone supply. Section 9.7
presents instructions for flow calibration. Used to adjust the gas flow calculations made by the CPU
based on pressure and flow sensor readings.
Teledyne Analytical Instruments
134
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
6. COMMUNICATIONS SETUP AND OPERATION
This instrument’s rear panel connections include an Ethernet port, a USB port (option) and two serial
communications ports labeled RS232, which is the COM1 port, and COM2 (refer to Figure 3-4). These
ports give the user the ability to communicate with, issue commands to, and receive data from the
analyzer through an external computer system or terminal. Connection instructions were provided in
Section 3.3.1.8.
This section provides pertinent information regarding communication equipment, describes the
instrument’s communications modes, presents configuration instructions for the communications ports,
and provides instructions for their use, including communications protocol. Data acquisition is presented
in Section 7.
6.1. DATA TERMINAL / COMMUNICATION EQUIPMENT (DTE DEC)
RS-232 was developed for allowing communications between data terminal equipment (DTE) and data
communication equipment (DCE). Basic terminals always fall into the DTE category whereas modems
are always considered DCE devices. The difference between the two is the pin assignment of the Data
Receive and Data Transmit functions.
DTE devices receive data on pin 2 and transmit data on pin 3.
DCE devices receive data on pin 3 and transmit data on pin 2.
To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers (which can be
either), a switch mounted below the serial ports on the rear panel labeled DCE DTE (Figure 3-4) allows
the user to set the RS-232 configuration for one of these two data devices. This switch exchanges the
Receive and Transmit lines on RS-232 emulating a cross-over or null-modem cable. The switch has no
effect on COM2.
6.2. COMMUNICATION MODES, BAUD RATE AND PORT
TESTING
Use the SETUP>MORE>COMM menu to configure COM1 (labeled RS232 on instrument rear panel)
and/or COM2 (labeled COM2 on instrument rear panel) for communication modes, baud rate and/or
port testing for correct connection. If using a USB option communication connection, setup requires
configuring the COM2 baud rate (Section 6.2.2).
6.2.1. COMMUNICATION MODES
Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in
Table 6-1. As modes are selected, the analyzer sums the mode ID numbers and displays this combined
number on the front panel display. For example, if quiet mode (01), computer mode (02) and Multi-
Drop-Enabled mode (32) are selected, the analyzer would display a combined MODE ID of 35.
Table 6-1: COMM Port Communication Modes
MODE1
QUIET
ID
1
DESCRIPTION
Quiet mode suppresses any feedback from the analyzer (such as warning messages)
to the remote device and is typically used when the port is communicating with a
computer program where such intermittent messages might cause communication
Teledyne Analytical Instruments
135
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
problems.
Such feedback is still available but a command must be issued to receive them.
Computer mode inhibits echoing of typed characters and is used when the port is
communicating with a computer operated control program.
COMPUTER
2
HESSEN
PROTOCOL
The Hessen communications protocol is used in some European countries. TAPI P/N
02252 contains more information on this protocol.
16
When turned on this mode switches the COMM port settings from
No parity; 8 data bits; 1 stop bit to Even parity; 8 data bits; 1 stop bit.
When turned on this mode switches the COMM port settings from
No parity; 8 data bits; 1 stop bit to Even parity; 7 data bits; 1 stop bit.
E, 8, 1
E, 7, 1
8192
2048
Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence
over multidrop mode if both are enabled. Also, configuring for RS-485 disables the rear
panel USB port.
RS-485
1024
When enabled, the serial port requires a password before it will respond (see Section
5.5). The only command that is active is the help screen (? CR).
SECURITY
4
MULTIDROP
PROTOCOL
Multidrop protocol allows a multi-instrument configuration on a single communications
channel. Multidrop requires the use of instrument IDs.
32
ENABLE
MODEM
Enables to send a modem initialization string at power-up. Asserts certain lines in the
RS-232 port to enable the modem to communicate.
64
ERROR
Fixes certain types of parity errors at certain Hessen protocol installations.
CHECKING2
128
256
XON/XOFF
Disables XON/XOFF data flow control also known as software handshaking.
HANDSHAKE2
Enables CTS/RTS style hardwired transmission handshaking. This style of data
transmission handshaking is commonly used with modems or terminal emulation
protocols as well as by Teledyne Instrument’s APICOM software.
HARDWARE
HANDSHAKE
8
HARDWARE
FIFO2
Disables the HARDWARE FIFO (First In – First Out). When FIFO is enabled, it
improves data transfer rate for that COMM port.
512
COMMAND
PROMPT
Enables a command prompt when in terminal mode.
4096
1 Modes are listed in the order in which they appear in the
SETUP MORE COMM COM[1 OR 2] MODE menu
2 The default setting for this feature is ON. Do not disable unless so instructed by Teledyne Customer Service personnel.
Communication Modes for each COMM port must be configured independently. To turn on or off the
communication modes for either COM1 or COM2, access the SETUP>MORE.[COM1 OR COM2]
menu, and at the COM1 [2] Mode menu press EDIT.
Teledyne Analytical Instruments
136
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
Figure 6-1.
COMM – Communication Modes Setup
Teledyne Analytical Instruments
137
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
6.2.2. COM PORT BAUD RATE
To select the baud rate of either COMM Port, go to SETUP>MORE>COMM and select either COM1 or
COM2 as follows (use COM2 to view/match your personal computer baud rate when using the USB
port, Section 6.6):
SETUP X.X
COMMUNICATIONS MENU
Select which COM
port to configure.
(COM1 for example).
ID INET COM1 COM2
SETUP X.X
SET> EDIT
COM1 MODE:0
Press SET> until you
reach the COM1
BAUD RATE
EXIT
EXIT
EXIT
EXIT
EXAMPLE
SETUP X.X
COM1 BAUD RATE:19200
Use PREV and NEXT
to move between
available baud rates.
EXIT
ignores
the new
setting
<SET SET> EDIT
300
1200
4800
SETUP X.X
COM1 BAUD RATE:19200
ENTR
9600
ENTR
accepts
the new
setting
19200
38400
57600
115200
PREV NEXT
SETUP X.X
COM1 BAUD RATE:9600
ENTR
NEXT ON
Figure 6-2.
COMM – COMM Port Baud Rate
6.2.3. COM PORT TESTING
The serial ports can be tested for correct connection and output in the COMM menu. This test sends a
string of 256 ‘w’ characters to the selected COMM port. While the test is running, the red LED on the
rear panel of the analyzer should flicker.
To initiate the test press the following button sequence:
Teledyne Analytical Instruments
138
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
<TST TST> CAL
Concentration field
displays all gases.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
SETUP X.X
COMMUNICATIONS MENU
ID INET COM1 COM2
EXIT
EXIT
SETUP X.X
COM1 MODE:0
<SET SET> EDIT
Continue pressing <SET or SET> until ...
SETUP X.X
COM1: TEST PORT
<SET SET> TEST
ENTR EXIT
Test runs
automatically.
SETUP X.X
SETUP X.X
TRANSMITTING TO COM1
COM1: TEST PORT
PREV NEXT OFF
EXIT
Figure 6-3.
COMM – COM1 Test Port
6.3. RS-232
The RS232 and COM2 communications (COMM) ports operate on the RS-232 protocol (default
configuration). Possible configurations for these two COMM ports are summarized as follows:
RS232 port can also be configured to operate in single or RS-232 Multidrop mode (Option 62); refer
to Section 3.3.1.8.
Teledyne Analytical Instruments
139
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
COM2 port can be left in its default configuration for standard RS-232 operation including multidrop,
or it can be reconfigured for half-duplex RS-485 operation (please contact the factory for this
configuration).
Note that when the rear panel COM2 port is in use, except for multidrop communication, the rear panel
USB port cannot be used. (Alternatively, when the USB port is enabled, COM2 port cannot be used
except for multidrop).
A code-activated switch (CAS), can also be used on either port to connect typically between 2 and 16
send/receive instruments (host computer(s) printers, data loggers, analyzers, monitors, calibrators, etc.)
into one communications hub. Contact TAI Sales for more information on CAS systems.
To configure the analyzer’s communication ports, use the SETUP>MORE>COMM menu. Refer to
Section 5.7 for initial setup, and to Section 6.2 for additional configuration information.
6.4. RS-485 (OPTION)
The COM2 port of the instrument’s rear panel is set up for RS-232 communication but can be
reconfigured for RS-485 communication. Contact Customer Service. If this option was elected at the
time of purchase, the rear panel was preconfigured at the factory.
6.5. ETHERNET
When using the Ethernet interface, the analyzer can be connected to any standard 10BaseT or 100BaseT
Ethernet network via low-cost network hubs, switches or routers. The interface operates as a standard
TCP/IP device on port 3000. This allows a remote computer to connect through the network to the
analyzer using APICOM, terminal emulators or other programs.
The Ethernet connector has two LEDs that are on the connector itself, indicating its current operating
status.
Table 6-2: Ethernet Status Indicators
LED
amber (link)
green (activity
FUNCTION
On when connection to the LAN is valid.
Flickers during any activity on the LAN.
The analyzer is shipped with DHCP enabled by default. This allows the instrument to be connected to a
network or router with a DHCP server. The instrument will automatically be assigned an IP address by
the DHCP server (Section 6.5.2). This configuration is useful for quickly getting an instrument up and
running on a network. However, for permanent Ethernet connections, a static IP address should be used.
Section 6.5.1 below details how to configure the instrument with a static IP address.
6.5.1. CONFIGURING ETHERNET COMMUNICATION MANUALLY (STATIC
IP ADDRESS)
To configure Ethernet communication manually:
1. Connect a cable from the analyzer’s Ethernet port to a Local Area Network (LAN) or
Internet port.
Teledyne Analytical Instruments
140
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
From the analyzer’s front panel touchscreen, access the Communications Menu (SETUP>MORE>COMM,
see
2. Figure 5-2).
3. Enter the INET menu shown in Figure 6-4, turning DHCP mode to OFF and editing
the Instrument and Gateway IP addresses and Subnet Mask to the desired settings
(default settings showin in Table 6-3).
Alternatively, from the computer, enter the same information through an application such as
HyperTerminal.
Teledyne Analytical Instruments
141
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
Internet Configuration Button Functions
SETUP X.X
ID INET
COMMUNICATIONS MENU
BUTTON
[0]
FUNCTION
COM1 COM2
EXIT
ENTR EXIT
EXIT
Location of cursor. Press to cycle through the range of
numerals and available characters (“0 – 9” & “ . ”)
<CH CH> Moves the cursor one character left or right.
SAMPLE
ENTER SETUP PASS : 818
8
DEL
Deletes a character at the cursor location.
Accepts the new setting and returns to the previous
menu.
8
1
ENTR
Ignores the new setting and returns to the previous
menu.
EXIT
DHCP: ON is
default setting.
Skip this step
if it has been
set to OFF.
SETUP X.X
DHCP: ON
Some buttons appear only when relevant.
SET> EDIT
SETUP X.X
DHCP: OFF
SET> EDIT
EXIT
SETUP X.X INST IP: 000.000.000.000
<SET SET> EDIT
EXIT
Cursor
location is
indicated by
brackets
SETUP X.X INST IP: [0] 00.000.000
<CH CH>
DEL [0]
ENTR EXIT
SETUP X.X GATEWAY IP: 000.000.000.000
<SET SET> EDIT
EXIT
SETUP X.X GATEWAY IP: [0] 00.000.000
<CH CH> DEL [?] ENTR EXIT
SETUP X.X SUBNET MASK:255.255.255.0
<SET SET> EDIT
EXIT
SETUP X.X SUBNET MASK:[2]55.255.255.0
<CH CH> DEL [?] ENTR EXIT
SETUP X.X TCP PORT 3000
<SET
EDIT
EXIT
The PORT number must remain at 3000.
Do not change this setting unless instructed to by
Teledyne Instruments Customer Service personnel.
Pressing EXIT from
any of the above
display menus
causes the Ethernet
option to reinitialize
its internal interface
firmware
SETUP X.X
INITIALIZING INET 0%
…
INITIALIZING INET 100%
SETUP X.X
INITIALIZATI0N SUCCEEDED
SETUP X.X
INITIALIZATION FAILED
Contact your IT
Network Administrator
SETUP X.X
COMMUNICATIONS MENU
COM1 COM2
ID
INET
EXIT
Figure 6-4.
COMM - LAN /Internet Manual Configuration
Teledyne Analytical Instruments
142
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
Table 6-3. LAN/Ethernet Default Configuration Properties
PROPERTY
DHCP
DEFAULT STATE
DESCRIPTION
This displays whether the DHCP is turned ON or OFF. Press
EDIT and toggle ON for automatic configuration after first
consulting network administrator.
ON
INSTRUMENT
IP ADDRESS
This string of four packets of 1 to 3 numbers each (e.g.
192.168.76.55.) is the address of the analyzer itself.
Can only be edited when DHCP is set to OFF.
0.0.0.0
0.0.0.0
GATEWAY IP
ADDRESS
A string of numbers very similar to the Instrument IP address
(e.g. 192.168.76.1.) that is the address of the computer used by
your LAN to access the Internet.
Can only be edited when DHCP is set to OFF.
Also a string of four packets of 1 to 3 numbers each (e.g.
255.255.252.0) that identifies the LAN to which the device is
connected.
All addressable devices and computers on a LAN must have the
same subnet mask. Any transmissions sent to devices with
different subnets are assumed to be outside of the LAN and are
routed through the gateway computer onto the Internet.
SUBNET MASK
This number defines the terminal control port by which the
instrument is addressed by terminal emulation software, such as
Internet or Teledyne API’s APICOM.
3000
TCP PORT1
The name by which your analyzer will appear when addressed
from other computers on the LAN or via the Internet. To assign
or change, see Section 6.5.2.1.
[initially blank]
HOST NAME
1 Do not change the setting for this property unless instructed to by TAI’s Customer Service personnel.
6.5.2. CONFIGURING ETHERNET COMMUNICATION USING DYNAMIC
HOST CONFIGURATION PROTOCOL (DHCP)
The default Ethernet setting is DHCP.
Teledyne Analytical Instruments
143
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
1. Consult with your network administrator to affirm that your network server is running
DHCP.
Access the Communications Menu (SETUP>MORE>COMM, see
2. Figure 5-2).
3. Enter the INET menu and follow the setup sequence as shown in Figure 6-5.
Teledyne Analytical Instruments
144
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
SETUP X.X
COMMUNICATIONS MENU
From this point on,
EXIT returns to
COMMUNICATIONS
MENU
ID INET COM1 COM2
EXIT
SAMPLE
ENTER SETUP PASS : 818
8
1
8
ENTR EXIT
DHCP: ON is
SETUP X.X
DHCP: OFF
DHCP: ON
default setting.
If it has been
set to OFF,
press EDIT
and set to ON.
SETUP X.X
DHCP: ON
OFF
ENTR EXIT
SET> EDIT
EXIT
EXIT
EXIT
EXIT
SETUP X.X
ON
ENTR EXIT
SETUP X.X
<SET SET>
INST IP: 0.0.0.0
SETUP X.X GATEWAY IP: 0.0.0.0
EDIT button
disabled
<SET SET>
SETUP X.X SUBNET MASK: 0.0.0.0
<SET SET>
Do not alter unless
SETUP X.X
TCP PORT: 3000
directed to by Teledyne
Instruments Customer
Service personnel
<SET SET> EDIT
EXIT
EXIT
EXIT
SETUP X.X
TCP PORT2: 502
<SET SET> EDIT
SETUP X.X HOSTNAME:
<SET
EDIT
Figure 6-5.
COMM – LAN / Internet Automatic Configuration (DHCP)
Note
If the gateway IP, instrument IP and the subnet mask are all zeroes (i.e.,
“0.0.0.0”), the DCHP was not successful in which case you may have to
configure the analyzer’s Ethernet properties manually. See your network
administrator.
Teledyne Analytical Instruments
145
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
6.5.2.1. Changing the Analyzer’s HOSTNAME
The HOSTNAME is the name by which the analyzer appears on your network. The initial default
Hostname is blank. To assign or change this name (particularly if you have more than one 9110T
analyzer on your network, where each must have
a
different Hostname), enter the
SETUP>COMM>INET men and scroll to the HOSTNAME menu as in Figure 6-5; make the changes as
shown in Figure 6-6:
Figure 6-6.
COMM – Change Hostname
Teledyne Analytical Instruments
146
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
6.6. USB PORT FOR REMOTE ACCESS
The analyzer can be operated through a personal computer by downloading the TAPI USB driver and
directly connecting their respective USB ports.
1. Install the Teledyne T-Series USB driver on your computer, downloadable from the Teledyne API
website under Help Center>Software Downloads (www.teledyne-api.com/software).
2. Run the installer file: “TAPIVCPInstaller.exe”
3. Connect the USB cable between the USB ports on your personal computer and your analyzer. The
USB cable should be a Type A – Type B cable, commonly used as a USB printer cable.
4. Determine the Windows XP Com Port number that was automatically assigned to the USB
connection. (Start → Control Panel → System → Hardware → Device Manager). This is the com
port that should be set in the communications software, such as APIcom or Hyperterminal.
Refer to the Quick Start (Direct Cable Connection) section of the Teledyne APIcom Manual, PN 07463.
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following settings:
Teledyne Analytical Instruments
147
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode
ON
Computer Mode
ON
MODBUS RTUOFF
MODBUS ASCII
OFF
E,8,1 MODE OFF
E,7,1 MODE OFF
RS-485 MODE OFF
SECURITY MODE
OFF
MULTIDROP MODE OFF
ENABLE MODEM OFF
ERROR CHECKING ON
XON/XOFF HANDSHAKE
OFF
HARDWARE HANDSHAKE OFF
HARDWARE FIFO ON
COMMAND PROMPT OFF
6. Next, configure your communications software, such as APIcom. Use the COM port determined in
Step 4 and the baud rate set in Step 5. The figures below show how these parameters would be
configured in the Instrument Properties window in APIcom when configuring a new instrument. See
the APIcom manual (PN 07463) for more details.
USB configuration requires that the baud rates of the instrument and
the PC match; check the PC baud rate and change if needed.
Note
Teledyne Analytical Instruments
148
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
Using the USB port disallows use of the rear panel COM2 port except
for multidrop communication.
6.7. COMMUNICATIONS PROTOCOLS
Two communications protocols available with the analyzer are MODBUS and Hessen. MODBUS setup
instructions are provided here (Section 6.7.1) and registers are provided in Appendix A. Hessen setup
and operation instructions are provided in Section 6.7.2.
6.7.1. MODBUS
The following set of instructions assumes that the user is familiar with MODBUS communications, and
provides minimal information to get started. For additional instruction, please refer to the Teledyne API
MODBUS manual, PN 06276. Also refer to www.modbus.org for MODBUS communication protocols.
Minimum Requirements
Instrument firmware with MODBUS capabilities installed.
MODBUS-compatible software (TAPI uses MODBUS Poll for testing; see www.modbustools.com)
Personal computer
Communications cable (Ethernet or USB or RS232)
Possibly a null modem adapter or cable
Actions
Set Com Mode parameters
Comm
Ethernet:
Using the front panel menu, go to SETUP – MORE – COMM – INET; scroll
through the INET submenu until you reach TCP PORT 2 (the standard setting is
502), then continue to TCP PORT 2 MODBUS TCP/IP; press EDIT and toggle
the menu button to change the setting to ON, then press ENTR. (Change
Machine ID if needed: see “Slave ID”).
USB/RS232: Using the front panel menu, go to SETUP – MORE – COMM – COM2 – EDIT;
scroll through the COM2 EDIT submenu until the display shows COM2
MODBUS RTU: OFF (press OFF to change the setting to ON. Scroll NEXT to
COM2 MODBUS ASCII and ensure it is set to OFF. Press ENTR to keep the
new settings. (If RTU is not available with your communications equipment, set
the COM2 MODBUS ASCII setting to ON and ensure that COM2 MODBUS
RTU is set to OFF. Press ENTR to keep the new settings).
If your analyzer is connected to a network with at least one other analyzer of the same model,
a unique Slave ID must be assigned to each. Using the front panel menu, go to SETUP –
MORE – COMM – ID. The MACHINE ID default is the same as the model number. Toggle the
menu buttons to change the ID.
Slave ID
Reboot analyzer
For the settings to take effect, power down the analyzer, wait 5 seconds, and power up the
analyzer.
Make appropriate cable
connections
Connect your analyzer either:
via its Ethernet or USB port to a PC (this may require a USB-to-RS232 adapter for your PC;
if so, also install the sofware driver from the CD supplied with the adapter, and reboot the
computer if required), or
via its COM2 port to a null modem (this may require a null modem adapter or cable).
Specify MODBUS software
settings
1. Click Setup / [Read / Write Definition] /.
a. In the Read/Write Definition window (see example that follows) select a Function (what
(examples used here are for
Teledyne Analytical Instruments
149
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
Actions
MODBUS Poll software)
you wish to read from the analyzer).
b. Input Quantity (based on your firware’s register map).
c. In the View section of the Read/Write Definition window select a Display (typically Float
Inverse).
d. Click OK.
2. Next, click Connection/Connect.
a. In the Connection Setup window (see example that follows), select the options based on
your computer.
b. Press OK.
Read the Modbus Poll
Register
Use the Register Map to find the test parameter names for the values displayed (see example
that follows If desired, assign an alias for each.
Example Read/Write Definition window:
Example Connection Setup window:
Teledyne Analytical Instruments
150
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
Example MODBUS Poll window:
6.7.2. HESSEN
The Hessen protocol is a multidrop protocol, in which several remote instruments are connected via a
common communications channel to a host computer. The remote instruments are regarded as slaves of
the host computer. The remote instruments are unaware that they are connected to a multidrop bus and
never initiate Hessen protocol messages. They only respond to commands from the host computer and
only when they receive a command containing their own unique ID number.
The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments,
including the concentrations of all the gases measured; and to place remote instruments into zero or span
calibration or measure mode. TAI’s implementation supports both of these principal features.
The Hessen protocol is not well defined; therefore, while Teledyne-API’s application is completely
compatible with the protocol itself, it may be different from implementations by other companies.
6.7.2.1. HESSEN COMM PORT CONFIGURATION
Hessen protocol requires the communication parameters of the 9110T’s COMM ports to be set
differently than the standard configuration as shown in the table below.
Table 6-4:RS-232 Communication Parameters for Hessen Protocol
PARAMETER
Baud Rate
Data Bits
Stop Bits
Parity
STANDARD
HESSEN
1200
7
300 – 19200
8
1
2
None
Full
Even
Half
Duplex
To change the baud rate of the 9110T’s COMM ports, See Section 6.2.2.
To change the remaining COMM port parameters listed in the table above, see Section 6.2.1, Table 6-1.
Note
Ensure that the communication parameters of the host computer are also
properly set.
Also, the instrument software has a 200 ms latency period before it
responds to commands issued by the host computer. This latency should
Teledyne Analytical Instruments
151
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
present no problems, but you should be aware of it and not issue
commands to the instrument too frequently.
6.7.2.2. ACTIVATING HESSEN PROTOCOL
Once the COMM port has been properly configured, the next step in configuring the 9110T in order to
operate over a Hessen protocol network is to activate the Hessen mode for COMM ports and configure
the communication parameters for the port(s) appropriately.
To activate the Hessen Protocol, press:
Teledyne Analytical Instruments
152
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
6.7.2.3. Selecting a Hessen Protocol Type
Currently there are two versions of Hessen Protocol in use. The original implementation, referred to as
TYPE 1, and a more recently released version, TYPE 2 that has more flexibility when operating with
instruments that can measure more than one type of gas. For more specific information about the
difference between TYPE 1 and TYPE 2 download the Manual Addendum for Hessen Protocol from
the Teledyne API's web site: http://www.teledyne-api.com/manuals/.
Teledyne Analytical Instruments
153
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
To select a Hessen Protocol Type press:
NOTE
While Hessen Protocol Mode can be activated independently for COM1 and COM2, the
TYPE selection affects both Ports.
6.7.2.4. SETTING THE HESSEN PROTOCOL RESPONSE MODE
The Teledyne implementation of Hessen Protocol allows the user to choose one of several different
modes of response for the analyzer.
Table 6-5: Teledyne's Hessen Protocol Response Modes
MODE ID
CMD
MODE DESCRIPTION
This is the Default Setting. Reponses from the instrument are encoded as the traditional
command format. Style and format of responses depend on exact coding of the initiating
command.
Responses from the instrument are always delimited with <STX> (at the beginning of the
response, <ETX> (at the end of the response followed by a 2 digit Block Check Code
(checksum), regardless of the command encoding.
BCC
Responses from the instrument are always delimited with <CR> at the beginning and the end of
the string, regardless of the command encoding.
TEXT
To Select a Hessen response mode, press:
Teledyne Analytical Instruments
154
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
6.7.2.5. HESSEN PROTOCOL GAS LIST ENTRY FORMAT AND DEFINITIONS
The 9110T analyzer keeps a list of available gas types. Each entry in this list is of the following format.
[GAS TYPE],[RANGE],[GAS ID],[REPORTED]
WHERE:
GAS TYPE
RANGE
The type of gas to be reported (e.g. NOx, NO and NO2 etc.).
The concentration range for this entry in the gas list. This feature permits
the user to select which concentration range will be used for this gas list
entry. The 9110T analyzer has two ranges: RANGE1 or LOW &
RANGE2 or HIGH (see Section 5.4).
0 - The HESSEN protocol to use whatever range is currently active.
Teledyne Analytical Instruments
155
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
1 - The HESSEN protocol will always use RANGE1 for this gas list
entry
2 - The HESSEN protocol will always use RANGE2 for this gas list
entry
3 - Not applicable to the 9110T analyzer.
GAS ID
An identification number assigned to a specific gas. The 9110T analyzer
is a multiple gas instrument that measures NOx, NO and NO2. Their ID
numbers are as follows:
NOx 211
NO 212
NO2 213
REPORT
States whether this list entry is to be reported or not reported when ever
this gas type or instrument is polled by the HESSEN network. If the list
entry is not to be reported this field will be blank. It’s default gas list
consists of only reads:
NOX, 0, 211, REPORTED
NO, 0, 212, REPORTED
NO2, 0, 213, REPORTED
These default settings cause the instrument to report the concentration value of the currently active
range. If you wish to have just concentration value stored for a specific range, this list entry should be
edited or additional entries should be added to the list.
EXAMPLE: Changing the above NOx gas list entry to read:
NOX, 2, 211, REPORTED
Would only record the last NOx reading that occurred while RANGE2 (HIGH) range was active.
EDITING OR ADDING HESSEN GAS LIST ENTRIES
To add or edit an entry to the Hessen Gas List, press:
Teledyne Analytical Instruments
156
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
DELETING HESSEN GAS LIST ENTRIES
To delete an entry from the Hessen Gas list, press:
Teledyne Analytical Instruments
157
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
6.7.2.6. SETTING HESSEN PROTOCOL STATUS FLAGS
Teledyne’s implementation of Hessen protocols includes a set of status bits that the instrument includes
in responses to inform the host computer of its condition. Each bit can be assigned to one operational
and warning message flag. The default settings for these bit/flags are:
Teledyne Analytical Instruments
158
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
Table 6-6: Default Hessen Status Flag Assignments
DEFAULT BIT
ASSIGNMENT
STATUS FLAG NAME3
WARNING FLAGS
SAMPLE FLOW WARNING
0001
0002
0004
0008
0010
0020
0040
0080
8000
OZONE FLOW WARNING
RCEL PRESS WARN
BOX TEMP WARNING
RCELL TEMP WARNING
IZS TEMP WARNING1
PMT TEMP WARN
CONV TEMP WARNING
INVALID CONC
OPERATIONAL FLAGS
In MANUAL Calibration Mode
In ZERO Calibration Mode
In SPAN Calibration Mode
In WARMUP Mode
UNITS OF MEASURE FLAGS
UGM
0200
0400
0800
1000
0000
2000
4000
6000
0100
MGM
PPB
PPM
SPARE/UNUSED BITS
UNASSIGNED FLAGS (0000)
MANIFOLD TEMPERATURE2
HVPS WARNING
OZONE GEN OFF
FRONT PANEL WARN
ANALOG CAL WARNING
CANNOT DYN ZERO
SYSTEM RESET
RELAY BOARD WARNING
REAR BOARD NOT DETECTED
CANNOT DYN SPAN
AUTOZERO WARNING
Instrument is in MP CAL mode
1
Only applicable if the optional internal span gas generator is installed.
Only applicable if the 9110T is equipped with an oxygenator option.
2
3
It is possible to assign more than one flag to the same Hessen status bit. This
allows the grouping of similar flags, such as all temperature warnings, under the
same status bit.
Be careful not to assign conflicting flags to the same bit as each status bit will be
triggered if any of the assigned flags is active.
Teledyne Analytical Instruments
159
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Communications Setup and Operation
To assign or reset the status flag bit assignments, press:
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
<TST TST> CAL
Concentration field
displays all gases.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
SETUP X.X
HESSEN STATUS FLAGS
ID HESN COM1 COM2
<SET SET> EDIT
EXIT
SETUP X.X
HESSEN VARIATION:TYPE1
SETUP X.X
IZS TEMP WARNING:0020
<SET SET> EDIT
EXIT
PREV NEXT
EDIT PRNT EXIT
Continue pressing SET until ...
Continue pressing NEXT until desired
flag message is displayed
SETUP X.X
BOX TEMP WARNING:0008
PREV NEXT
EDIT PRNT EXIT
SETUP X.X
BOX TEMP WARNING:[0]008
DEL [0] ENTR EXIT
EXIT discards the
<CH CH>
The <CH and CH>
new setting.
buttons move the
cursor brackets “[ ]”
left and right along the
bit string.
ENTR accepts the
new setting.
Press the [?] button repeatedly to cycle
through the available character set: 0-9
INS Inserts a the
character at the
current location of the
cursor brackets.
NOTE: Values of A-F can also be set
but are meaningless.
6.7.2.7. Instrument ID
Each instrument on a Hessen Protocol network must have a unique identifier (ID number). Refer to
Section 5.7.1 for information and to customize the ID of each.
Teledyne Analytical Instruments
160
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Communications Setup and Operation
This page intentionally left blank.
Teledyne Analytical Instruments
161
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7. DATA ACQUISITION SYSTEM (DAS) AND APICOM
The 9110T analyzer contains a flexible and powerful, internal data acquisition system (DAS) that
enables the analyzer to store concentration and calibration data as well as a host of diagnostic
parameters. The DAS feature of the 9110T can store up to one million data points, which can,
depending on individual configurations, cover days, weeks or months of valuable measurements. The
data records are stored in non-volatile memory and are retained even when the instrument is powered
off. Data are stored in plain text format for easy retrieval and use in common data analysis programs
(such as electronic spreadsheets).
The DAS is designed to be flexible, users have full control over the type, length and reporting time of the
data. The DAS permits users to access stored data through the instrument’s front panel or remotely
through its communication ports.
The principal use of the DAS is logging data for trend analysis and predictive diagnostics, which can
assist in identifying possible problems before they affect the functionality of the analyzer. The
secondary use is for data analysis, documentation and archival in electronic format.
To support the DAS functionality, Teledyne offers APICOM, a program that provides a visual interface
for remote or local setup, configuration and data retrieval of the DAS. Using APICOM, data can even be
retrieved automatically to a remote computer for further processing. The APICOM manual, included
with the program, contains a more detailed description of the DAS structure and configuration and is
briefly described in this document.
The 9110T is configured with a basic DAS configuration already enabled. The data channels included in
this basic structure may be used as is or temporarily disabled for later or occasional use.
The green SAMPLE LED on the instrument front panel, which indicates the analyzer status, also
indicates certain aspects of the DAS status:
Table 7-1: Front Panel LED Status Indicators for DAS
LED STATE
DAS Status
System is in calibration mode. Data logging can be enabled or disabled for this mode. Calibration
data are typically stored at the end of calibration periods, concentration data are typically not
sampled, diagnostic data should be collected.
OFF
Instrument is in hold-off mode, a short period after the system exits calibrations. DAS channels can
be enabled or disabled for this period. Concentration data are typically disabled whereas diagnostic
should be collected.
BLINKING
ON
Sampling normally.
IMPORTANT
IMPACT ON READINGS OR DATA
DAS operation is suspended whenever its configuration is edited using
the analyzer’s front panel and therefore data may be lost. To prevent
such data loss, it is recommended to use the APICOM graphical user
interface for DAS changes (Section 7.2.1).
Please be aware that all stored data will be erased if the analyzer’s disk-
on-module or CPU board is replaced or if the configuration data stored
there is reset.
Teledyne Analytical Instruments
163
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
Note
The DAS can be disabled only by disabling or deleting its individual data
channels.
7.1. DAS STRUCTURE
The DAS is designed around the feature of a “record”. A record is a single data point. The type of data
captured in a record are defined by two properties:
PARAMETER type that defines the kind of data to be stored (e.g. the average of O3 concentrations
measured with three digits of precision). See Section 7.1.3.3.
A TRIGGER event that defines when the record is made (e.g. timer; every time a calibration is
performed, etc.). See Section 7.1.3.2.
The specific PARAMETERS and TRIGGER events that describe an individual record are defined in a
construct called a DATA CHANNEL (see Section 7.1.3). Each data channel relates one or more
parameters with a specific trigger event and various other operational characteristics related to the
records being made (e.g. the channels name, number or records to be made, time period between records,
whether or not the record is exported via the analyzer’s RS-232 port, etc.).
7.1.1. DAS CHANNELS
The key to the flexibility of the DAS is its ability to store a large number of combinations of triggering
events and data parameters in the form of data channels. Users may create up to 20 data channels and
each channel can contain one or more parameters. For each channel, the following are selected:
one triggering event
up to 50 data parameters, which can be the shared between channels.
several other properties that define the structure of the channel and allow the user to make
operational decisions regarding the channel.
Table 7-2: DAS Data Channel Properties
DEFAULT
SETTING
“NONE”
PROPERTY
NAME
DESCRIPTION
SETTING RANGE
Up to 6 letters or digits1.
The name of the data channel.
Any available event
(see Appendix A-5).
TRIGGERING
EVENT
The event that triggers the data channel to
measure and store the datum.
ATIMER
NUMBER AND
LIST OF
PARAMETERS
Any available parameter
(see Appendix A-5).
A user-configurable list of data types to be
recorded in any given channel.
1
(PMTDET)
000:00:01 to
366:23:59
(Days:Hours:Minutes)
The amount of time between each channel data
point.
000:01:00
(1 hour)
REPORT PERIOD
The number of reports that will be stored in the
data file. Once the limit is exceeded, the oldest
data is over-written.
1 to 1 million, limited by
available storage space.
NUMBER OF
RECORDS
100
Enables the analyzer to automatically report
channel values to the RS-232 ports.
RS-232 REPORT
OFF
OFF or ON
Teledyne Analytical Instruments
164
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
DEFAULT
PROPERTY
DESCRIPTION
Enables or disables the channel. Allows a channel
SETTING RANGE
SETTING
CHANNEL
ENABLED
ON
OFF or ON
OFF or ON
to be temporarily turned off without deleting it.
Disables sampling of data parameters while
CAL HOLD OFF
OFF
instrument is in calibration mode2.
1 More with APICOM, but only the first six are displayed on the front panel).
2 When enabled records are not recorded until the DAS_HOLD OFF period is passed after calibration mode. DAS_HOLD OFF SET in
the VARS menu (see Section 5.8).
7.1.1.1. Default DAS Channels
A set of default Data Channels has been included in the analyzer’s software for logging NOx, NO and
NO2 concentrations as well as certain predictive diagnostic data. For the software revision being shipped
with the 9110T at the time of this writing, these default channels are:
CONC: Samples NOx concentration at one minute intervals and stores an average every hour with a
time and date stamp. Readings during calibration and calibration hold off are not included in the data.
By default, the last 800 hourly averages are stored.
CALDAT: Logs new slope and offset of NOX and NO measurements every time a zero or span
calibration is performed and the result changes the value of the slope (triggering event: SLPCHG). The
NOX stability (to evaluate if the calibration value was stable) as well as the converter efficiency (for
trend reference) are also stored.
This data channel will store data from the last 200 calibrations and can be used to document
analyzer calibration and is useful for detect trends in slope and offset (instrument response) when
performing predictive diagnostics as part of a regular maintenance schedule (See Section 11.1).
The CALDAT channel collects data based on events (e.g. a calibration operation) rather than a
timed interval and therefore does not represent any specific length of time. As with all data
channels, a date and time stamp is recorded for every logged data point.
CALCHECK: This channel logs concentrations and the stability each time a zero or span check (not
calibration) is finished (triggered by exiting any calibration menu).
The data of this channel enable the user to track the quality of zero and span responses over time
and assist in evaluating the quality of zero and span gases and the analyzer’s noise specifications.
The STABIL parameter documents if the analyzer response was stable at the point of the calibration
check reading. The last 200 data points are retained.
DIAG: Daily averages of temperature zones, flow and pressure data as well as some other diagnostic
parameters (HVPS, AZERO).
This data is useful for predictive diagnostics and maintenance of the 9110T.
The last 1100 daily averages are stored to cover more than four years of analyzer performance.
HIRES: Records one-minute, instantaneous data of all active parameters in the 9110T. Short-term
trends as well as signal noise levels can be detected and documented.
Readings during calibration and the calibration hold off period are included in the averages.
The last 1500 data points are stored, which covers a little more than one day of continuous data
acquisition.
Teledyne Analytical Instruments
165
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
These default data channels can be used as they are, or they can be customized from the front panel to fit
a specific application. They can also be deleted to make room for custom user-programmed Data
Channels.
Appendix A lists the firmware-specific DAS configuration in plain-text format. This text file can either
be loaded into APICOM and then modified and uploaded to the instrument or can be copied and pasted
into a terminal program to be sent to the analyzer.
IMPORTANT
IMPACT ON READINGS OR DATA
Sending a DAS configuration to the analyzer through its COM ports will
replace the existing configuration and will delete all stored data. Back up
any existing data and the DAS configuration before uploading new
settings.
Teledyne Analytical Instruments
166
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
List of Channels
List of Parameters
STORE NUM
SAMPLES
PARAMETER
MODE
PRECISION
Name:
Event:
Parameters:
Report Period:
NOXCNC1
NOCNC1
N2CNC1
O2CONC
STABIL
AVG
AVG
AVG
AVG
AVG
4
4
4
4
4
OFF
OFF
OFF
OFF
ON
No. of Records:
RS-232 Report:
Channel Enabled:
Cal Hold OFF:
NXZSC1
NXSLP1
NXOFS1
NOZSC1
NOSLP1
NOOFS1
N2ZSC1
CNVEF1
STABIL
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
4
4
4
4
4
4
4
4
4
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Name:
Event:
Parameters:
No. of Records:
RS-232 Report:
Channel Enabled:
Cal Hold OFF:
NXCNC1
NOCNC1
N2CNC1
STABIL
AVG
AVG
AVG
AVG
4
4
4
4
OFF
OFF
OFF
OFF
Name:
Event:
Parameters:
No. of Records:
RS-232 Report:
Channel Enabled:
Cal Hold OFF:
SMPFLW
O3FLOW
RCPRES
SMPPRS
RCTEMP
PMTTMP
CNVTMP
MFTEMP
BOXTMP
O2TEMP
AZERO
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
2
2
2
2
2
2
2
2
2
2
2
1
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Name:
Event:
Parameters:
Report Period:
No. of Records:
RS-232 Report:
Channel Enabled:
Cal Hold OFF:
HVPS
NXCNC1
NOCNC1
N2CNC1
STABIL
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
AVG
4
4
4
4
2
2
2
2
2
2
2
2
2
2
2
1
1
1
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
SMPFLW
O3FLOW
RCPRES
SMPPRS
RCTEMP
PMTTMP
CNVTMP
MFTEMP
BOXTMP
O2TEMP
AZERO
Name:
Event:
Parameters:
Report Period:
No. of Records:
RS-232 Report:
Channel Enabled:
Cal Hold OFF:
HVPS
REFGND
REF4096
Figure 7-1:
Default DAS Channels Setup
Teledyne Analytical Instruments
167
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.1.2. DAS Configuration Limits
The number of DAS objects are limited by the instrument’s finite storage capacity. For information
regarding the maximum number of channels, parameters, and records and how to calculate the file size
for each data channel, refer to the DAS manual downloadable from the Teledyne website at
http://www.teledyne-api.com/manuals/ under Special Manuals.
7.1.2. VIEWING DAS DATA AND SETTINGS
DAS data and settings can be viewed on the front panel through the following menu sequence.
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
DAS VIEW – Touchscreen Functions
<TST TST> CAL
Button
FUNCTION
Concentration field
displays all gases.
Moves the VIEW backward 10 record
Moves the VIEW backward 1 records or channel
Moves the VIEW forward 1 record or channel
Moves the VIEW forward 10 records
PV10
PREV
NEXT
NX10
<PRM
PRM>
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
EXIT
SETUP X.X
DATA ACQUISITION
Selects the previous parameter on the list
Selects the next parameter on the list
VIEW EDIT
Buttons only appear when applicable.
SETUP X.X
CONC: DATA AVAILABLE
NEXT VIEW
EXIT
SETUP X.X
101:21:00 NXCNC1=59.0346 PPB
PV10 PREV NX10 NEXT <PRM PRM>
EXIT
SETUP X.X
101:22:00 NXCNC1=000.0000 PPB
SETUP X.X
101:21:00 NOCNC1=22.0934 PPB
EXIT
PV10 PREV NX10 NEXT <PRM PRM>
EXIT
PV10 PREV NX10 NEXT <PRM PRM>
SETUP X.X
CALDAT: DATA AVAILABLE
NEXT VIEW
EXIT
SETUP X.X
101:19:45 NXZSC1=401.0346
PV10 PREV NX10 NEXT <PRM PRM>
EXIT
SETUP X.X
102:04:55 NXZSC1=400.9868
SETUP X.X
101:19:45 NXSLP1=0.9987 PPB
PV10 PREV NX10 NEXT <PRM PRM>
EXIT
PV10 PREV NX10 NEXT <PRM PRM>
EXIT
Continue pressing NEXT to view remaining
DAS channels
Teledyne Analytical Instruments
168
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3. EDITING DAS DATA CHANNELS
DAS configuration is most conveniently done through the APICOM remote control program. The
following list of button strokes shows how to edit using the front panel.
When editing the data channels, the top line of the display indicates some of the configuration
parameters.
For example, the display line:
0) NXCNC1: ATIMER, 5, 800
Translates to the following configuration:
Channel No.: 0
NAME: NXCNC1
TRIGGER EVENT: ATIMER
PARAMETERS: Five parameters are included in this channel
EVENT: This channel is set up to store 800 records.
Teledyne Analytical Instruments
169
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.1. Editing DAS Data Channel Names
To edit the name of a DAS data channel, follow the instruction shown in Section 7.1.3 then press:
Teledyne Analytical Instruments
170
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.2. Editing DAS Triggering Events
Triggering events define when and how the DAS records a measurement of any given data channel.
Triggering events are firmware-specific and a complete list of Triggers for this model analyzer can be
found in Appendix A-5. The most commonly used triggering events are:
ATIMER: Sampling at regular intervals specified by an automatic timer. Most trending information is
usually stored at such regular intervals, which can be instantaneous or averaged.
EXITZR, EXITSP, and SLPCHG (exit zero, exit span, slope change): Sampling at the end of
(irregularly occurring) calibrations or when the response slope changes. These triggering events
create instantaneous data points, e.g., for the new slope and offset (concentration response) values
at the end of a calibration. Zero and slope values are valuable to monitor response drift and to
document when the instrument was calibrated.
WARNINGS: Some data may be useful when stored if one of several warning messages appears
such as WTEMPW (GFC wheel temperature warning). This is helpful for troubleshooting by
monitoring when a particular warning occurred.
To edit the list of data parameters associated with a specific data channel, follow the instruction shown
in Section 7.1.3 then press:
Note
A full list of DAS Trigger Events can be found in Appendix A-5 of this
manual.
Teledyne Analytical Instruments
171
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.3. Editing DAS Parameters
Data parameters are types of data that may be measured and stored by the DAS. For each TAI's analyzer
model, the list of available data parameters is different, fully defined and not customizable. Appendix
A-5 lists firmware specific data parameters for the 9110T. DAS parameters include data such as NOx,
NO and NO2 concentration measurements, temperatures of the various heaters placed around the
analyzer, pressures and flows of the pneumatic subsystem and other diagnostic measurements as well as
calibration data such as stability, slope and offset.
Most data parameters have associated measurement units, such as mV, ppb, cm³/min, etc., although
some parameters have no units (e.g. SLOPE). With the exception of concentration readings, none of
these units of measure can be changed. To change the units of measure for concentration readings, see
Section 5.4.3.4.
Note
DAS does not keep track of the units (i.e. PPM or PPB) of each
concentration value. Therefore, DAS data files may contain concentration
data recorded in more than one type of unit if the units of measure were
changed during data acquisition
Each data parameter has user-configurable functions that define how the data are recorded which are
listed in Table 7-3:
Table 7-3: DAS Data Parameter Functions
FUNCTION
EFFECT
PARAMETER
Instrument specific parameter name.
SAMPLE MODE
INST: Records instantaneous reading.
AVG: Records average reading during reporting interval.
SDEV: Records the standard deviation of the data points recorded during the reporting interval.
MIN: Records minimum (instantaneous) reading during reporting interval.
MAX: Records maximum (instantaneous) reading during reporting interval.
PRECISION
0 to 4: Sets the number of digits to the right decimal point for each record.
Example: Setting 4; “399.9865 PPB”
Setting 0; “400 PPB”
STORE NUM.
SAMPLES
OFF: Stores only the average (default).
ON: Stores the average and the number of samples in used to compute the value of the
parameter. This property is only useful when the AVG sample mode is used. Note that the
number of samples is the same for all parameters in one channel and needs to be specified
only for one of the parameters in that channel.
Users can specify up to 50 parameters per data channel (the 9110T provides about 40 parameters).
However, the number of parameters and channels is ultimately limited by available memory.
Data channels can be edited individually from the front panel without affecting other data channels.
However, when editing a data channel, such as during adding, deleting or editing parameters, all data for
that particular channel will be lost, because the DAS can store only data of one format (number of
parameter columns etc.) for any given channel. In addition, a DAS configuration can only be uploaded
remotely as an entire set of channels. Hence, remote update of the DAS will always delete all current
channels and stored data.
Teledyne Analytical Instruments
172
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
To modify, add or delete a parameter, follow the instruction shown in Section 7.1.3 then press:
Starting at the EDIT CHANNEL MENU
DAS EDIT – Touchscreen Functions
Button
PREV
NEXT
<SET
FUNCTION
SETUP X.X
0) CONC: ATIMER 5, 800
Selects the previous data channel or parameter
Selects the next data channel or parameter
Selects the previous property to be edited
Selects the next property to be edited
PREV NEXT
INS DEL EDIT PRNT EXIT
SETUP X.X
NAME: CONC
SET>
<SET SET> EDIT
EXIT
Inserts a new data channel or parameter into the list
BEFORE the selected channel
INS
Deletes the currently selected data channel or
parameter
DEL
Continue pressing <SET or SET> until ...
Enters EDIT mode
EDIT
Exports the configuration of all data channels to the
RS-232 interface
PRINT
Buttons only appear when applicable
SETUP X.X
PARAMETER:5
<SET SET> EDIT
EXIT
SETUP X.X
EDIT PARAMS (DELETE DATA)?
NO retains the
YES deletes all data
currently stored for
this data channel and
continues into EDIT
mode.
YES NO
data and
returns to the
previous
menu.
SETUP X.X
0) PARAM=NXCNC1, MODE=AVG
EXIT discards the new
PREV NEXT
INS DEL EDIT
EXIT
setting.
Toggle these
buttons to select a
different parameter.
ENTR accepts the
new setting.
SETUP X.X
PARAMETER:NXCNC1
<SET SET> EDIT
EXIT
SETUP X.X
PARAMETER:NXCNC1
PREV NEXT
ENTR EXIT
Toggle these buttons to
cycle through the list of
available parameters.
SETUP X.X
SAMPLE MODE:AVG
<SET SET> EDIT
EXIT
SETUP X.X
PARAMETER:NXCNC1
INST AVG SDEV MIN MAX
ENTR EXIT
Pressing <SET
returns to the
Press the desired
MODE button.
previous Function.
SETUP X.X
PRECISION:4
<SET SET> EDIT
EXIT
SETUP X.X
PRECISION:5
5
ENTR EXIT
Toggle this button to
set from 1 to 4.
SETUP X.X
STOR NUM SAMPLE:OFF
EDIT
<SET
EXIT
SETUP X.X
STOR NUM SAMPLE:OFF
OFF
ENTR EXIT
Toggle this button to
turn ON/OFF.
Note
When the STORE NUM SAMPLES feature is turned on, the instrument will
store the number of measurements that were used to compute the AVG,
SDEV, MIN or MAX value but not the actual measurements themselves.
Teledyne Analytical Instruments
173
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.4. Editing Sample Period and Report Period
The DAS defines two principal time periods by which sample readings are taken and permanently
recorded:
SAMPLE PERIOD: Determines how often DAS temporarily records a sample reading of the
parameter in volatile memory. SAMPLE PERIOD is only used when the DAS parameter’s sample
mode is set for AVG, SDEV, MIN or MAX.
The SAMPLE PERIOD is set to one minute by default and generally cannot be accessed from the
standard DAS front panel menu, but is available via the instruments communication ports by using
APICOM or the analyzer’s standard serial data protocol.
REPORT PERIOD: Sets how often the sample readings stored in volatile memory are processed,
(e.g. average, minimum or maximum are calculated) and the results stored permanently in the
instruments Disk-on-Module (DOM) as well as transmitted via the analyzer’s communication ports.
The Report Period may be set from the front panel. If the INST sample mode is selected the
instrument stores and reports an instantaneous reading of the selected parameter at the end of the
chosen report period.
Note
In AVG, SDEV, MIN or MAX sample modes (see Section 7.1.3.3), the
settings for the Sample Period and the Report Period determine the
number of data points used each time the parameters are calculated,
stored and reported to the COMM ports.
The actual sample readings are not stored past the end of the chosen
report period.
When the STORE NUM SAMPLES feature is turned on, the instrument will
store the number of measurements that were used to compute the AVG,
SDEV, MIN or MAX value but not the actual measurements themselves.
To define the REPORT PERIOD, follow the instruction shown in Section 7.1.3 then press:
Teledyne Analytical Instruments
174
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
The SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning and end
of the appropriate interval of the instruments internal clock.
If SAMPLE PERIOD is set for one minute the first reading would occur at the beginning of the next
full minute according to the instrument’s internal clock.
If the REPORT PERIOD is set for of one hour, the first report activity would occur at the beginning of
the next full hour according to the instrument’s internal clock.
EXAMPLE:
Given the above settings, if DAS parameters are activated at 7:57:35 the first sample would occur at 7:58
and the first report would be calculated at 8:00 consisting of data points for 7:58. 7:59 and 8:00.
During the next hour (from 8:01 to 9:00), the instrument will take a sample reading every minute and
include 60 sample readings.
7.1.3.5. Report Periods in Progress when Instrument Is Powered Off
If the instrument is powered off in the middle of a REPORT PERIOD, the samples accumulated during
that period are lost. Once the instrument is turned back on, the DAS restarts taking samples and
Teledyne Analytical Instruments
175
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
temporarily stores them in volatile memory as part of the REPORT PERIOD currently active at the
time of restart. At the end of this REPORT PERIOD, only the sample readings taken since the
instrument was turned back on will be included in any AVG, SDEV, MIN or MAX calculation.
The STORE NUM SAMPLES feature will also report the number of sample readings taken since the
instrument was restarted.
7.1.3.6. Editing the Number of Records
The number of data records in the DAS is limited to about a cumulative one million data points in all
channels (one megabyte of space on the DOM). However, the actual number of records is also limited
by the total number of parameters and channels and other settings in the DAS configuration. Every
additional data channel, parameter, number of samples setting etc. will reduce the maximum amount of
data points. In general, however, the maximum data capacity is divided amongst all channels (max: 20)
and parameters (max: 50 per channel).
The DAS will check the amount of available data space and prevent the user from specifying too many
records at any given point. If, for example, the DAS memory space can accommodate 375 more data
records, the ENTR button will disappear when trying to specify more than that number of records. This
check for memory space may also make an upload of a DAS configuration with APICOM or a terminal
program fail, if the combined number of records would be exceeded. In this case, it is suggested to
either try to determine what the maximum number of records available is using the front panel interface
or use trial-and-error in designing the DAS script or calculate the number of records using the DAS or
APICOM manuals.
To set the NUMBER OF RECORDS, follow the instruction shown in Section 7.1.3 then press:
Teledyne Analytical Instruments
176
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.7. RS-232 Report Function
The DAS can automatically report data to the communications ports, where they can be captured with a
terminal emulation program or simply viewed by the user using the APICOM software.
To enable automatic COMM port reporting, follow the instruction shown in Section 7.1.3 then press:
Teledyne Analytical Instruments
177
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
Starting at the EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 5, 800
PREV NEXT
INS DEL EDIT PRNT EXIT
Use the PREV and
NEXT buttons to
scroll to the DATA
CHANNEL to be
edited.
SETUP X.X
NAME: CONC
<SET SET> EDIT
EXIT
Continue pressing <SET or SET> until ...
SETUP X.X
RS-232 REPORT: OFF
<SET SET> EDIT PRNT
EXIT
SETUP X.X
RS-232 REPORT: OFF
EXIT discards the new
OFF
ENTR EXIT
setting.
Toggle these buttons
to turn the RS-232
REPORT feature
ON/OFF.
ENTR accepts the
new setting.
7.1.3.8. HOLDOFF Feature
The DAS HOLDOFF feature prevents data collection during calibration operations.
To enable or disable the HOLDOFF, follow the instruction shown in Section 7.1.3 then press:
Teledyne Analytical Instruments
178
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
HOLDOFF also prevents DAS measurements from being made at certain times when the quality of the
analyzer’s O3 measurements may be suspect (e.g. while the instrument is warming up). In this case, the
length of time that the HOLDOFF feature is active is determined by the value of the internal variable
(VARS), DAS_HOLDOFF.
To set the length of the DAS_HOLDOFF period, go to the SETUP>MORE>VARS menu and at the
DAS_HOLDOFF parameter (see Table 5-3), press the Edit button.
Teledyne Analytical Instruments
179
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.1.3.9. The Compact Report Feature
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting
each parameter in one channel on a separate line, up to five parameters are reported in one line.
The COMPACT DATA REPORT generally cannot be accessed from the standard DAS front panel
menu, but is available via the instrument’s communication ports by using APICOM or the analyzer’s
standard serial data protocol.
7.1.3.10. The Starting Date Feature
This option allows the user to specify a starting date for any given channel when the user wants to start
data acquisition only after a certain time and date. If the STARTING DATE is in the past (the default
condition), the DAS ignores this setting and begins recording data as defined by the REPORT PERIOD
setting.
The STARTING DATE generally cannot be accessed from the standard DAS front panel menu, but is
available via the instrument’s communication ports by using APICOM or the analyzer’s standard serial
data protocol.
7.1.3.11. Disabling/Enabling Data Channels
Data channels can be temporarily disabled, which can reduce the read/write wear on the Disk-on-Module
(DOM).
To disable a data channel, go to the DAS>EDIT menu as shown in Section 7.1.3 then continue as
follows:
Teledyne Analytical Instruments
180
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.2. REMOTE DAS CONFIGURATION
The DAS can be configured and operated remotely via either the APICOM interface or a terminal
emulation program. Once a DAS configuration is edited (which can be done offline and without
interrupting DAS data collection), it is conveniently uploaded to the instrument and can be stored on a
computer for later review, alteration or documentation and archival.
7.2.1. DAS CONFIGURATION VIA APICOM
Table 7-2 shows examples of APICOM’s main interface, which emulates the look and functionality of
the instrument’s actual front panel. Figure 7-3 shows an example of APICOM being used to remotely
configure the DAS feature.
The APICOM user manual (Teledyne P/N 039450000) is included in the APICOM installation file,
which can be downloaded at http://www.teledyne-api.com/software/apicom/.
Figure 7-2:
APICOM Remote Control Program Interface
Teledyne Analytical Instruments
181
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
Figure 7-3:
Sample APICOM User Interface for Configuring the DAS
Teledyne Analytical Instruments
182
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Data Acquisition System (DAS) and APICOM
7.2.2. DAS CONFIGURATION VIA TERMINAL EMULATION PROGRAMS
Although TAI recommends the use of APICOM, the DAS can also be accessed and configured through a
terminal emulation program such as HyperTerminal (see Figure 7-4 for example). It is best to start by
downloading the default DAS configuration, getting familiar with its command structure and syntax
conventions, and then altering a copy of the original file offline before uploading the new configuration.
Figure 7-4:
DAS Configuration Through a Terminal Emulation Program
See Section 8.2.1 for configuration commands and their strict syntax. Commands can be pasted in from
of an existing text file, which was first edited offline and then uploaded through a specific transfer
procedure.
IMPORTANT
IMPACT ON READINGS OR DATA
Whereas the editing, adding and deleting of DAS channels and
parameters of one channel through the front-panel control buttons can
be done without affecting the other channels, uploading a DAS
configuration script to the analyzer through its communication ports will
erase all data, parameters and channels by replacing them with the new
DAS configuration. Backup of data and the original DAS configuration is
advised before attempting any DAS changes.
Teledyne Analytical Instruments
183
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Data Acquisition System (DAS) and APICOM
This page intentionally left blank.
Teledyne Analytical Instruments
184
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Remote Operation
8. REMOTE OPERATION
This section provides information needed when using external digital and serial I/O for remote operation.
It assumes that the electrical connections have been made as described in Section 3.3.1.
The T100 can be remotely configured, calibrated or queried for stored data through the serial ports, via
either Computer mode (using a personal computer) or Interactive mode (using a terminal emulation
program).
8.1. COMPUTER MODE
Computer Mode is used when the analyzer is connected to a computer with a dedicated interface
program such as APICOM.
8.1.1. REMOTE CONTROL VIA APICOM
APICOM is an easy-to-use, yet powerful interface program that allows the user to access and control any
of TAI's’ main line of ambient and stack-gas instruments from a remote connection through direct cable,
modem or Ethernet. Running APICOM, a user can:
Establish a link from a remote location to the 9110T through direct cable connection via RS-232
modem or Ethernet.
View the instrument’s front panel and remotely access all functions that could be accessed manually
on the instrument.
Remotely edit system parameters and set points.
Download, view, graph and save data for predictive diagnostics or data analysis.
Retrieve, view, edit, save and upload DAS configurations (Section 7.2.1).
Check on system parameters for troubleshooting and quality control.
APICOM is very helpful for initial setup, data analysis, maintenance and troubleshooting. Refer to the
APICOM manual available for download from http://www.teledyne-api.com/software/apicom/.
8.2. INTERACTIVE MODE
Interactive mode is used with a terminal emulation programs or a “dumb” computer terminal.
8.2.1. REMOTE CONTROL VIA A TERMINAL EMULATION PROGRAM
Start a terminal emulation programs such as HyperTerminal. All configuration commands must be
created following a strict syntax or be pasted in from a text file, which was edited offline and then
uploaded through a specific transfer procedure. The commands that are used to operate the analyzer in
this mode are listed in Table 8-1 and in Appendix A.
Teledyne Analytical Instruments
185
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Remote Operation
8.2.1.1. Help Commands in Interactive Mode
Table 8-1: Terminal Mode Software Commands
COMMAND
Control-T
Function
Switches the analyzer to terminal mode (echo, edit). If mode flags 1 & 2 are OFF, the
interface can be used in interactive mode with a terminal emulation program.
Control-C
Switches the analyzer to computer mode (no echo, no edit).
CR
A carriage return is required after each command line is typed into the terminal/computer.
The command will not be sent to the analyzer to be executed until this is done. On
personal computers, this is achieved by pressing the ENTER button.
(carriage return)
BS
Erases one character to the left of the cursor location.
(backspace)
ESC
Erases the entire command line.
(escape)
?[ID] CR
This command prints a complete list of available commands along with the definitions of
their functionality to the display device of the terminal or computer being used. The ID
number of the analyzer is only necessary if multiple analyzers are on the same
communications line, such as the multi-drop setup.
Control-C
Control-P
Pauses the listing of commands.
Restarts the listing of commands.
8.2.1.2. Command Syntax
Commands are not case-sensitive and all arguments within one command (i.e. ID numbers, key words,
data values, etc.) must be separated with a space character.
All Commands follow the syntax:
X [ID] COMMAND <CR>
Where:
X
is the command type (one letter) that defines the type of command. Allowed
designators are listed in Table 8-2 and Appendix A-6.
[ID]
is the machine identification number (Section 5.7.1). Example: the
Command “? 200” followed by a carriage return would print the list of
available commands for the revision of software currently installed in the
instrument assigned ID Number 200.
COMMANDis the command designator: This string is the name of the command being
issued (LIST, ABORT, NAME, EXIT, etc.). Some commands may have
additional arguments that define how the command is to be executed.
Press ? <CR> or refer to Appendix A-6 for a list of available command
designators
<CR>
is a carriage return. All commands must be terminated by a carriage return
(usually achieved by pressing the ENTER button on a computer).
Table 8-2:
Teledyne API's Serial I/O Command Types
COMMAND
COMMAND TYPE
Calibration
C
D
Diagnostic
Teledyne Analytical Instruments
186
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Remote Operation
L
T
Logon
Test measurement
Variable
V
W
Warning
8.2.1.3. Data Types
Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean expressions and
text strings.
Integer data are used to indicate integral quantities such as a number of records, a filter length, etc.
They consist of an optional plus or minus sign, followed by one or more digits. For example, +1, -12,
123 are all valid integers.
Hexadecimal integer data are used for the same purposes as integers. They consist of the two
characters “0x,” followed by one or more hexadecimal digits (0-9, A-F, a-f), which is the ‘C’
programming language convention. No plus or minus sign is permitted. For example, 0x1, 0x12,
0x1234abcd are all valid hexadecimal integers.
Floating-point numbers are used to specify continuously variable values such as temperature set points,
time intervals, warning limits, voltages, etc. They consist of an optional plus or minus sign, followed by
zero or more digits, an optional decimal point and zero or more digits. (At least one digit must appear
before or after the decimal point.) Scientific notation is not permitted. For example, +1.0, 1234.5678, -
0.1, 1 are all valid floating-point numbers.
Boolean expressions are used to specify the value of variables or I/O signals that may assume only two
values. They are denoted by the key words ON and OFF.
Text strings are used to represent data that cannot be easily represented by other data types, such as data
channel names, which may contain letters and numbers. They consist of a quotation mark, followed by
one or more printable characters, including spaces, letters, numbers, and symbols, and a final quotation
mark. For example, “a”, “1”, “123abc”, and “()[]<>” are all valid text strings. It is not possible to
include a quotation mark character within a text string.
Some commands allow you to access variables, messages, and other items. When using these
commands, you must type the entire name of the item; you cannot abbreviate any names.
8.2.1.4. Status Reporting
Reporting of status messages as an audit trail is one of the three principal uses for the RS-232 interface
(the other two being the command line interface for controlling the instrument and the download of data
in electronic format). You can effectively disable the reporting feature by setting the interface to quiet
mode (Section 6.2.1, Table 6-1).
Status reports include warning messages, calibration and diagnostic status messages. Refer to Appendix
A-3 for a list of the possible messages, and this for information on controlling the instrument through the
RS-232 interface.
8.2.1.5. General Message Format
All messages from the instrument (including those in response to a command line request) are in the
format:
Teledyne Analytical Instruments
187
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Remote Operation
X DDD:HH:MM [Id] MESSAGE<CRLF>
Where:
X
is a command type designator, a single character indicating the
message type, as shown in the Table 8-2.
DDD:HH:MM is the time stamp, the date and time when the message was issued. It
consists of the Day-of-year (DDD) as a number from 1 to 366, the hour
of the day (HH) as a number from 00 to 23, and the minute (MM) as a
number from 00 to 59.
[ID]
is the analyzer ID, a number with 1 to 4 digits.
MESSAGE
is the message content that may contain warning messages, test
measurements, variable values, etc.
<CRLF>
is a carriage return / line feed pair, which terminates the message.
The uniform nature of the output messages makes it easy for a host computer to parse them into an easy
structure. Keep in mind that the front panel display does not give any information on the time a message
was issued, hence it is useful to log such messages for troubleshooting and reference purposes. Terminal
emulation programs such as HyperTerminal can capture these messages to text files for later review.
8.3. REMOTE ACCESS BY MODEM
The 9110T can be connected to a modem for remote access. This requires a cable between the
analyzer’s COMM port and the modem, typically a DB-9F to DB-25M cable (available from TAI with
P/N WR0000024).
Once the cable has been connected, check to ensure that:
The DTE-DCE is in the DCE position.
The 9110T COMM port is set for a baud rate that is compatible with the modem,
The Modem is designed to operate with an 8-bit word length with one stop bit.
The MODEM ENABLE communication mode is turned ON (Mode 64, see Section 6.2.1).
Once this is completed, the appropriate setup command line for your modem can be entered into the
analyzer. The default setting for this feature is:
AT Y0 D0 H0 I0 S0=0
This string can be altered to match your modem’s initialization and can be up to 100 characters long.
To change this setting press:
Teledyne Analytical Instruments
188
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Remote Operation
Figure 8-1:
To initialize the modem press:
Remote Access by Modem
Teledyne Analytical Instruments
189
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Remote Operation
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
<TST TST> CAL
Concentration field
displays all gases.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
EXIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
COM1 COM2
SETUP X.X
COM1 MODE:0
<SET SET> EDIT
Continue pressing <SET or SET> until ...
SETUP X.X
COM1: INITIALIZE MODEM
ENTR EXIT
<SET SET> INIT
SETUP X.X
SETUP X.X
INITIALIZING MODE
MODEM INITIALIZED
Test runs
automatically.
PREV NEXT OFF
EXIT
If there is a problem initializing the
modem the message,
“MODEM NOT INITIALIZED”
will appear.
Teledyne Analytical Instruments
190
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Remote Operation
8.4. PASSWORD SECURITY FOR SERIAL REMOTE
COMMUNICATIONS
In order to provide security for remote access of the 9110T, a LOGON feature can be enabled to require
a password before the instrument will accept commands. This is done by turning on the SECURITY
MODE (refer to Section 5.5). Once the SECURITY MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the
LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to
be disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given
LOGON FAILED - Password not given or incorrect
LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the 9110T analyzer with SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS-232_PASS issued
as follows:
V RS-232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
Teledyne Analytical Instruments
191
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Remote Operation
This page intentionally left blank.
Teledyne Analytical Instruments
192
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9. CALIBRATION PROCEDURES
This section contains information for calibrating a 9110T NOx Analyzer as well as other supporting
information. For information on EPA protocol calibration, please refer to Section 10. This section is
organized as follows:
SECTION 9.1 – Before Calibration
This section contains general information you should know before about calibrating the analyzer.
SECTION 9.2 – Manual Calibration Checks and Calibration of the 9110T Analyzer in its Base
Configuration
This section describes:
The procedure for checking the calibrating of the 9110T and calibrating the instrument with
no zero/span valves installed or if installed, not operating. It requires that zero air and span
gas be installed through the Sample port.
Instructions for selecting the reporting range to be calibrated when the 9110T analyzer is set
to operate in either the IND or AUTO reporting range modes.
SECTION 9.3 – Manual Calibration with the Internal Span Gas Generator
This section describes:
The procedure for manually checking the calibration of the instrument with optional internal
span gas generator installed.
The procedure for manually calibrating the instrument using the optional internal span gas
generator.
This practice is not approved by the US EPA.
SECTION 9.4 – Manual Calibration and Cal Checks with the Valve Options Installed
This section describes:
The procedure for manually checking the calibration of the instrument with optional
zero/span valves option installed.
The procedure for manually calibrating the instrument with zero/span valves and operating.
Instructions on activating the zero/span valves via the control in contact closures of the
analyzers external digital I/O.
SECTION 9.5 – Automatic Zero/Span Cal/Check (AutoCal)
This section describes:
The procedure for using the AutoCal feature of the analyzer to check or calibrate the
instrument.
The AutoCal feature requires that either the zero/span valve option or the internal span gas
generator option be installed and operating.
SECTION 9.6 – Calibration Quality Analysis
Teledyne Analytical Instruments
193
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
This section describes how to judge the effectiveness of a recently performed calibration.
SECTION 9.7 – Gas Flow Calibration
This section describes how to adjust the gas flow calculations made by the CPU based on pressure and
flow sensor readings.
Note
Throughout this Section are various diagrams showing pneumatic
connections between the 9110T and various other pieces of equipment
such as calibrators and zero air sources.
These diagrams are only intended to be schematic representations of
these connections and do not reflect actual physical locations of
equipment and fitting location or orientation.
Contact your regional EPA or other appropriate governing agency for
more detailed recommendations.
9.1. BEFORE CALIBRATION
The calibration procedures in this section assume that the range mode, analog range and units of measure
have already been selected for the analyzer. If this has not been done, please do so before continuing
(see Section 5.4.3 for instructions).
Note
If any problems occur while performing the following calibration
procedures, refer to Section 12.1 for troubleshooting tips.
9.1.1. REQUIRED EQUIPMENT, SUPPLIES, AND EXPENDABLES
Calibration of the 9110T NOx Analyzer requires:
Zero-air source.
Span gas source.
Gas lines - all gas line materials should be stainless steel or Teflon-type (PTFE or FEP).
High-concentration NO gas transported over long distances may require stainless steel to
avoid oxidation of NO due to the possibility of O2 diffusing into the tubing.
A recording device such as a strip-chart recorder and/or data logger (optional).
For electronic documentation, the internal data acquisition system (DAS) can be used.
Teledyne Analytical Instruments
194
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.1.2. CALIBRATION GASES
9.1.2.1. Zero Air
Zero air or zero calibration gas is defined as a gas that is similar in chemical composition to the
measured medium but without the gas to be measured by the analyzer.
For the 9110T, this means zero air should be devoid of NO, NO2, CO2, NH3 or H2O vapor.
Note
Moderate amounts of NH3 and H2O can be removed from the sample gas
stream by installing the optional sample gas dryer/scrubber (see Section
3.3.2.6).
If your application is not a measurement in ambient air, the zero calibration gas should be matched
to the composition of the gas being measured.
Pure nitrogen (N2) could be used as a zero gas for applications where NOX is measured in nitrogen.
If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero
air from ambient air.
For analyzers without the external zero air scrubber, a zero air generator such as the Teledyne Model 701
can be used. Please visit the company website for more information.
If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero air
from ambient air.
If your application is not a measurement in ambient air, the zero calibration gas should be matched
to the composition of the gas being measured.
Pure nitrogen could be used as a zero gas for applications where NOX is measured in nitrogen.
9.1.2.2. Span Gas
Calibration gas is a gas specifically mixed to match the chemical composition of the type of gas being
measured at near full scale of the desired reporting range. To measure NOX with the 9110T NOX
analyzer, it is recommended that you use a span gas with an NO concentration equal to 80% of the
measurement range for your application
EXAMPLE:
If the application is to measure NOX in ambient air between 0 ppm and 500 ppb, an appropriate
span gas would be 400 ppb.
If the application is to measure NOX in ambient air between 0 ppm and 1000 ppb, an appropriate
span gas would be 800 ppb.
We strongly recommend that span calibration be carried out with NO span gas. Alternatively it is
possible to use NO2 gas in a gas phase titration (GPT) calibration system (see Section 10.5).
Even though NO gas mixed into in nitrogen gas (N2) could be used as a span gas, the matrix of the
balance gas is different and may cause interference problems or yield incorrect calibrations.
Teledyne Analytical Instruments
195
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
The same applies to gases that contain high concentrations of other compounds (for example, CO2
or H2O).
Note
The span gas should match all concentrations of all gases of the
measured medium as closely as possible.
Cylinders of calibrated NOx and NO gas traceable to NIST-standards specifications (also referred to as EPA
protocol calibration gases or Standard Reference Materials) are commercially available. For a list of these
gases see Table 3-8).
9.1.2.3. Span Gas for Multipoint Calibration
Some applications, such as EPA monitoring, require a multipoint calibration where span gases of
different concentrations are needed. We recommend using an NO gas of higher concentration combined
with a gas dilution calibrator such as a Teledyne Model T700. For more information see Section 3.3.2.1
and Section 10.
9.1.2.4. NO2 Permeation Tubes
TAI offers an optional internal span gas generator that utilizes an NO2 permeation tube as a span gas
source. The accuracy of these devices is only about ±5%.
Whereas this may be sufficient for quick, daily calibration checks, we recommend using certified NO
gases for accurate calibration.
Note
The use of permeation tubes is not approved by the US EPA as calibration
sources for performing actual calibration of the analyzers used in EPA
mandated monitoring.
CAUTION!
Insufficient gas flow allows gas to build up to levels that will contaminate
the instrument or present a safety hazard to personnel.
In units with a permeation tube option installed, either the tube must be
removed and stored in sealed container (use original container that tube
was shipped in) during periods of non-operation, or vacuum pump must
be connected and powered on to maintain constant gas flow though the
analyzer at all times.
(See Figure 3-6 for location and Section 11.3.6 for instructions on how to remove the
perm tube when the unit is not in operation).
9.1.3. DATA RECORDING DEVICES
A strip chart recorder, data acquisition system or digital data acquisition system should be used to record
data from the serial or analog outputs of the 9110T.
If analog readings are used, the response of the recording system should be checked against a
NIST traceable voltage source or meter.
Teledyne Analytical Instruments
196
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
Data recording devices should be capable of bi-polar operation so that negative readings can be
recorded.
For electronic data recording, the 9110T provides an internal data acquisition system (DAS), which is
described in detail in Section 7.
APICOM, a remote control program, is also provided as a convenient and powerful tool for data
handling, download, storage, quick check and plotting (see Sections 7.2.1, and the APICOM software
manual downloadable from:
http://www,tekedtbe-api.com/manuals ).
9.1.4. NO2 CONVERSION EFFICIENCY (CE)
In order for the NO2 converter to function properly, oxygen must be present in the sample stream. In
addition, to ensure accurate operation of the 9110T, it is important to check the NO2 conversion
efficiency (CE) periodically and to update this value as necessary.
See Section 12.7.10 for instructions on checking or calculating the current NO2 NO converter
efficiency using 9110T’s onboard firmware.
See Section 12.7.11 for instructions on checking or calculating the current NO2 NO converter
efficiency using a simplified Gas Phase Titration Method.
9.2. MANUAL CALIBRATION CHECKS AND CALIBRATION OF
THE 9110T ANALYZER IN ITS BASE CONFIGURATION
IMPORTANT
IMPACT ON READINGS OR DATA
ZERO/SPAN CALIBRATION CHECKS VS. ZERO/SPAN CALIBRATION
Pressing the ENTR button during the following procedure resets the
stored values for OFFSET and SLOPE and alters the instrument’s
Calibration.
This should ONLY BE DONE during an actual calibration of the 9110T.
NEVER press the ENTR button if you are only checking calibration..
9.2.1. SETUP FOR BASIC CALIBRATION CHECKS AND CALIBRATION OF
THE 9110T ANALYZER.
Connect the sources of zero air and span gas as shown below in one of the following ways:
Teledyne Analytical Instruments
197
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
Figure 9-1: Set up for Manual Calibrations/Checks of 9110T’s in Base Configuration w/ a Gas Dilution
Calibrator
Figure 9-2:
Set up for Manual Calibrations/Checks of 9110T’s in Base Configuration w/ Bottled Gas
Teledyne Analytical Instruments
198
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.2.2. PERFORMING A BASIC MANUAL CALIBRATION CHECK
Note
If the ZERO or SPAN buttons are not displayed, the measurement made is
outside the allowable range for a reliable calibration.
See Section 12 for troubleshooting tips.
9.2.3. PERFORMING A BASIC MANUAL CALIBRATION
The following section describes the basic method for manually calibrating the 9110T NOX analyzer.
Teledyne Analytical Instruments
199
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which
range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.
9.2.3.1. Setting the Expected Span Gas Concentration
Note
The expected concentrations for both NOx and NO are usually set to the
same value unless the conversion efficiency is not equal to 1.000 or not
entered properly in the conversion efficiency setting.
When setting expected concentration values, consider impurities in your
span gas source (e.g. NO often contains 1-3% NO2 and vice versa).
The NO and NOx span gas concentrations should be 80% of range of concentration values likely to be
encountered in your application. The default factory reporting range setting is 500 ppb and the default
span gas concentration is 40.0 ppb.
To set the span gas concentration, press:
Teledyne Analytical Instruments
200
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.2.3.2. Zero/Span Point Calibration Procedure
Note
If the ZERO or SPAN buttons are not displayed, the measurement made
during is out of the allowable range allowed for a reliable calibration. See
Section 12 for troubleshooting tips.
Teledyne Analytical Instruments
201
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.3. MANUAL CALIBRATION WITH THE INTERNAL SPAN GAS
GENERATOR
IMPORTANT
IMPACT ON READINGS OR DATA
The internal span gas generator’s NO2 permeation tube has a limited
accuracy of about ±5%. In the USA this may be sufficient for informal
calibration checks (Section 9.3.4), but it is NOT approved by the US EPA
as a calibration source.
9.3.1. PERFORMING “PRECISION” MANUAL CALIBRATION WHEN
INTERNAL SPAN GAS (IZS) GENERATOR OPTION IS PRESENT
It is necessary to perform a precision calibration using more accurate zero and span gas standards prior
to IZS span calibration or cal check.
To perform a precision calibration of the 9110T, connect external sources of zero air and calibrated span
gas (Section 9.1.2) and temporarily disconnect the sample gas source as shown below; then follow the
procedures described in Section 9.2.3.
Figure 9-3: Pneumatic Connections for 9110T Precision Calibration when IZS Generator Present
IMPACT ON READINGS OR DATA
IMPORTANT
DO NOT USE THE CALZ or CALS buttons even though they will be visible,
as this will cause the instrument to use the internal zero air and span gas.
Instead, press the CAL button. This will cause the analyzer to use the
external calibration gas sources.
Teledyne Analytical Instruments
202
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.3.2. SETUP FOR CALIBRATION WITH THE INTERNAL SPAN GAS
GENERATOR
Connect the sources of zero air and span gas as shown in Figure 9-4.
Figure 9-4: Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas
Generator
9.3.3. CAL ON NO2 FEATURE
When using the IZS option to calibrate the 9110T, the analyzer’s CAL_ON_NO2 feature must be turned
on. This feature enables a continuous zero gas flow across the IZS permeation tube and through the NO2
converter. It also programs the analyzer to use the NO output from the NO2 converter to calibrate the
span value of both NO and NOX.
Note
This feature should only be enabled when a span calibration or calibration
check is performed.
While CAL_ON_NO2 is enabled, the NO2 concentration will always be reported as zero. This is
because the gas is continuously routed through the NO2 converter and the analyzer’s firmware simulates
calibration with NO gas.
Table 9-1: IZS Option Valve States with CAL_ON_NO2 Turned ON
Valve
Sample/Cal
Condition
Open to zero/span valve
Open to SPAN GAS inlet
Open to NO2 converter
Cycles normally
Valve Port Connections
1 2
1 2
1 2
N/A
Zero/Span
NO/NOx Valve
Auto Zero Valve
Since the instrument sees the same concentration of NO during both NO and NOX cycles, it reports an
NO2 concentration of zero.
TO turn the CAL_ON_NO2 feature ON/OFF, press:
Teledyne Analytical Instruments
203
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
SAMPLE
RANGE=500.0 PPB
NO=XXXX
<TST TST>
CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
Press EXIT
3 times.
8
1
8
ENTR EXIT
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
EDIT PRNT EXIT
<PREV NEXT> JUMP
Continue pressing NEXT until ...
SETUP X.X
09)CAL_ON_NO2=OFF
<PREV NEXT> JUMP
EDIT PRNT EXIT
MODE FLD
CAL_ON_NO2=OFF
CONC
ON
ENTR EXIT
Use this button to
turn this feature
ON/OFF.
Teledyne Analytical Instruments
204
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.3.4. PERFORMING A MANUAL CALIBRATION CHECK WITH THE
INTERNAL SPAN GAS GENERATOR
Note
If the ZERO or SPAN buttons are not displayed, the measurement made is
out of the allowable range for a reliable calibration. See Section 12 for
troubleshooting tips.
Teledyne Analytical Instruments
205
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.3.5. PERFORMING A MANUAL CALIBRATION WITH THE INTERNAL
SPAN GAS GENERATOR
If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which
range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.
9.3.5.1. Setting the Expected Span Gas Concentration
Note
The expected concentrations for both NOx and NO are usually set to the
same value unless the conversion efficiency is not equal to 1.000 or not
entered properly in the conversion efficiency setting.
When setting expected concentration values, consider impurities in your
span gas source (e.g. NO often contains 1-3% NO2 and vice versa).
When calibrating the instrument using the internal permeation tube as a span gas source, it is necessary
to know, as close as possible, the concentration value of the gas being outputted by the tube. To
determine this value:
1. Perform a precision calibration of the instrument as describes in Section 9.3.1.
3. Perform a calibration check as described in Section 9.3.4.
Record the value displayed for NO/NOx during the span check portion of the procedure.
This will be the concentration value used in subsequent calibrations using the internal span
gas source.
It is a good idea to measure the permeation tube output once every 4 to 6 months.
4. Ensure that the reporting range span point is set for a value at least 10% higher than the measured
value of the permeation tube output
To set the span gas concentration, press:
Teledyne Analytical Instruments
206
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
Teledyne Analytical Instruments
207
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.3.5.2. Zero/Span Point Calibration Procedure with Internal Span Gas Generator
If the ZERO or SPAN buttons are not displayed, the measurement made
during during this procedure is out of the range allowed for a reliable
Note
calibration. See Section 12 for troubleshooting tips.
Teledyne Analytical Instruments
208
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.4. MANUAL CALIBRATION AND CAL CHECKS WITH THE
VALVE OPTIONS INSTALLED
There are a variety of valve options available on the 9110T for handling calibration gases (see Section
1.4 for descriptions of each).
Generally performing calibration checks and zero/span point calibrations on analyzers with these options
installed is similar to the methods discussed in the previous sections. The primary differences are:
On instruments with Z/S valve options, zero air and span gas is supplied to the analyzer through
other gas inlets besides the sample gas inlet.
The zero and span calibration operations are initiated directly and independently with dedicated
buttons (CALZ & CALS).
9.4.1. SETUP FOR CALIBRATION USING VALVE OPTIONS
Each of the various calibration valve options requires a different pneumatic setup that is dependent on
the exact nature and number of valves present. Refer to the following diagrams for information on each
or these valve sets.
Teledyne Analytical Instruments
209
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.4.2. MANUAL CALIBRATION CHECKS WITH VALVE OPTIONS
INSTALLED
SAMPLE
Set the Display to show
the test
<TST
CAL CALZ CALS
SETUP
function.
Analyzer display
continues to cycle
through all of the
available gas
This function calculates
the stability of the NO/NOx
measurement.
Toggle
button until ...
measurements
throughout this
procedure.
SAMPLE
<TST TST> CAL
CALS
SETUP
Wait until
falls below 0.5 PPB.
This may take several
minutes.
Record NOX, NO or NO2 zero point readings
SAMPLE
<TST TST> CAL CALZ
SETUP
The
and/or
Wait until
falls below 0.5 PPB.
buttons will appear at various
points of this process.
This may take several
minutes.
It is not necessary to press
them.
Record NOX, NO, NO2 span point readings
Teledyne Analytical Instruments
210
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.4.3. MANUAL CALIBRATION USING VALVE OPTIONS
The following section describes the basic method for manually calibrating the 9110T NOX analyzer.
If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which
range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.
9.4.3.1. Setting the Expected Span Gas Concentration
Note
The expected concentrations for both NOx and NO are usually set to the
same value unless the conversion efficiency is not equal to 1.000 or not
entered properly in the conversion efficiency setting.
When setting expected concentration values, consider impurities in your
span gas source (e.g. NO often contains 1-3% NO2 and vice versa).
The NO and NOx span gas concentrations should be 80% of range of concentration values likely to be
encountered in your application. The default factory reporting range setting is 500 ppb and the default
span gas concentration is 400.0 ppb.
To set the span gas concentration, press:
Teledyne Analytical Instruments
211
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.4.3.2. Zero/Span Point Calibration Procedure
If the ZERO or SPAN buttons are not displayed, the measurement made
during is out of the allowable range allowed for a reliable calibration. See
Section 12 for troubleshooting tips.
Note
Teledyne Analytical Instruments
212
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.4.3.3. Use of Zero/Span Valve with Remote Contact Closure
Contact closures for controlling calibration and calibration checks are located on the rear panel
CONTROL IN connector. Instructions for setup and use of these contacts are found in Section 3.3.1.6.
When the contacts are closed for at least 5 seconds, the instrument switches into zero, low span or high
span mode and the internal zero/span valves will be automatically switched to the appropriate
configuration.
The remote calibration contact closures may be activated in any order.
It is recommended that contact closures remain closed for at least 10 minutes to establish a reliable
reading.
The instrument will stay in the selected mode for as long as the contacts remain closed.
If contact closures are being used in conjunction with the analyzer’s AutoCal (see Section 9.5) feature
and the AutoCal attribute “CALIBRATE” is enabled, the 9110T will not re-calibrate the analyzer
UNTIL when the contact is opened. At this point, the new calibration values will be recorded before the
instrument returns to SAMPLE mode.
If the AutoCal attribute “CALIBRATE” is disabled, the instrument will return to SAMPLE mode,
leaving the instrument’s internal calibration variables unchanged.
9.5. AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL)
The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by using
the 9110T’s internal time of day clock. AutoCal operates by executing SEQUENCES programmed by
the user to initiate the various calibration modes of the analyzer and open and close valves appropriately.
It is possible to program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each
sequence can operate in one of three modes, or be disabled.
Table 9-2: AUTOCAL Modes
MODE NAME
DISABLED
ZERO
ACTION
Disables the Sequence.
Causes the Sequence to perform a Zero calibration/check.
Causes the Sequence to perform a Zero point
calibration/check followed by a Span point
calibration/check.
ZERO-SPAN
SPAN
Causes the Sequence to perform a Span concentration
calibration/check only.
Teledyne Analytical Instruments
213
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
For each mode, there are seven parameters that control operational details of the SEQUENCE. They
are:
Table 9-3: AutoCal Attribute Setup Parameters
ATTRIBUTE
ACTION
TIMER
ENABLED
Turns on the Sequence timer.
STARTING DATE Sequence will operate after Starting Date.
STARTING TIME Time of day sequence will run.
Number of days to skip between each Sequence execution.
DELTA DAYS
DELTA TIME
If set to 7, for example, the AutoCal feature will be enabled once
every week on the same day.
Number of hours later each “Delta Days” Sequence is to be run.
If set to 0, the sequence will start at the same time each day. Delta
Time is added to Delta Days for the total time between cycles.
This parameter prevents the analyzer from being calibrated at the
same daytime of each calibration day and prevents a lack of data for
one particular daytime on the days of calibration
Number of minutes the sequence operates.
This parameter needs to be set such that there is enough time for the
concentration signal to stabilize.
The STB parameter shows if the analyzer response is stable at the
end of the calibration.
DURATION
CALIBRATE
This parameter is logged with calibration values in the DAS.
Enable to do a calibration – Disable to do a cal check only.
For analyzers with internal span gas generators installed and
functioning, when used in US EPA applications, this setting must be
set to OFF.
LOW calibrates the low range, HIGH calibrates the high range. Applies
RANGE TO CAL only to auto and remote range modes; this property is not available in
single and independent range modes.
IMPORTANT
IMPACT ON READINGS OR DATA
For US EPA controlled/related applications:
For analyzers used in US EPA controlled applications that have internal
span gas generators option installed, the CALIBRATE attribute must
always be set to OFF
Calibration of instruments used in US EPA related applications should
only be performed using external sources of zero air and span gas with an
accuracy traceable to EPA or NIST standards and supplied through the
analyzer’s sample port.
Teledyne Analytical Instruments
214
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
The following example sets sequence #2 to do a zero-span calibration every other day starting at 1:00
AM on September 4, 2011, lasting 15 minutes, without calibration. This will start ½ hour later each
iteration.
Table 9-4:Example AutoCal Sequence
MODE AND
ATTRIBUTE
VALUE
COMMENT
SEQUENCE
Define Sequence #2
2
Select Zero and
Span Mode
MODE
ZERO-SPAN
ON
TIMER ENABLE
STARTING DATE
Enable the timer
Start after
Sept 4, 2011
Sept. 4, 2011
First Span starts at
1:00AM
STARTING TIME
DELTA DAYS
DELTA TIME
DURATION
1:00 AM
2
Do Sequence #2
every other day
Do Sequence #2 ½
hr later each day
00:30
15.0
Operate Span valve
for 15 min
Calibrate at end of
Sequence
CALIBRATE
OFF
IMPORTANT
IMPACT ON READINGS OR DATA
The programmed STARTING_TIME must be a minimum of 5 minutes
later than the real time clock for setting real time clock (See Section
5.6).
Avoid setting two or more sequences at the same time of the day.
Any new sequence that is initiated whether from a timer, the COMM
ports or the contact closure inputs will override any sequence that is
in progress.
The CALIBRATE attribute must always be set to OFF on analyzers with
IZS Options installed and functioning.
Calibrations should ONLY be performed using external sources of
Zero Air and Span Gas whose accuracy is traceable to EPA standards.
Teledyne Analytical Instruments
215
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.5.1. SETUP ACAL: PROGRAMMING AND AUTO CAL SEQUENCE
To program the example sequence shown in Table 9-4, press:
SAMPLE
RANGE = 500.0 PPB
NOX=XXX.X
SETUP
< TST TST > CAL CALZ CZLS
SETUP X.X
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE
EXIT
EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE
SETUP X.X MODE: DISABLED
NEXT
ENTR EXIT
ENTR EXIT
SETUP X.X MODE: ZERO
PREV NEXT
SETUP X.X MODE: ZERO–SPAN
PREV NEXT
ENTR EXIT
SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT MODE SET
EXIT
SETUP X.X TIMER ENABLE: ON
SET> EDIT
EXIT
SETUP X.X STARTING DATE: 01–JAN–07
<SET SET> EDIT
EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0
4
SEP
0
8
ENTR EXIT
Toggle buttons to set
Day, Month & Year:
Format : DD-MON-YY
CONTINUE NEXT PAGE
With STARTING TIME
Teledyne Analytical Instruments
216
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
CONTINUED FROM PREVIOUS PAGE -
STARTING DATE
SETUP X.X STARTING DATE: 04–SEP–03
<SET SET> EDIT
EXIT
EXIT
SETUP X.X STARTING TIME:00:00
<SET SET> EDIT
Toggle buttons to set
time:
SETUP X.X STARTING TIME:00:00
Format : HH:MM
1
4
: 1
5
ENTR EXIT
This is a 24 hr clock . PM
hours are 13 – 24.
Example 2:15 PM = 14:15
SETUP X.X STARTING TIME:14:15
<SET SET> EDIT
EXIT
SETUP X.X DELTA DAYS: 1
<SET SET> EDIT
EXIT
ENTR EXIT
EXIT
SETUP X.X DELTA DAYS: 1
Toggle buttons to set
number of days between
procedures (1-365).
0
0
2
SETUP X.X DELTA DAYS:2
<SET SET> EDIT
SETUP X.X DELTA TIME00:00
<SET SET> EDIT
EXIT
Toggle buttons to set
delay time for each
iteration of the sequence:
HH:MM
SETUP X.X DELTA TIME: 00:00
0
0
:3
0
ENTR EXIT
(0 – 24:00)
SETUP X.X DELTA TIME:00:30
<SET SET> EDIT
EXIT
CONTINUE NEXT PAGE
With DURATION TIME
Teledyne Analytical Instruments
217
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
SETUP
DURATION:15.0 MINUTES
<SET SET> EDIT
EXIT
Toggle buttons to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0.
SETUP
DURATION 15.0MINUTES
3
0
.0
ENTR EXIT
SETUP
DURATION:30.0 MINUTES
<SET SET> EDIT
EXIT
SETUP
CALIBRATE: OFF
<SET SET> EDIT
EXIT
ENTR EXIT
EXIT
SETUP
ON
CALIBRATE: OFF
Toggle button
Between Off and
ON.
SETUP X.X CALIBRATE: ON
<SET SET> EDIT
Display show:
SEQ 2) ZERO–SPAN, 2:00:30
SETUP X.X SEQ 2) ZERO–SPAN, 2:00:30
EXIT returns
to the SETUP
Menu.
Sequence
Delta Time
Delta Days
MODE
PREV NEXT MODE SET
EXIT
Note
If at any time an unallowable entry is selected (Example: Delta Days > 367)
the ENTR button will disappear from the display.
Teledyne Analytical Instruments
218
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110TH NOx Analyzer
Calibration Procedures
9.6. CALIBRATION QUALITY ANALYSIS
After completing one of the calibration procedures described above, it is important to evaluate the
analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response
curve of the analyzer, separately for NO and NOX. The values for these terms, both individually and
relative to each other, indicate the quality of the calibration.
To perform this quality evaluation, you will need to record the values of the following test functions (see
Section 4.1.1), all of which are automatically stored in the DAS channel CALDAT for data analysis,
documentation and archival.
NO OFFS
NO SLOPE
NOX OFFS
NOX SLOPE
Ensure that these parameters are within the limits listed in Table 9-5 and frequently compare them to
those values on the Final Test and Validation Data Sheet (P/N 04490) that came attached to your
manual, which should not be significantly different. If they are, refer to the troubleshooting Section 12.
Table 9-5: Calibration Data Quality Evaluation
Function
NOX SLOPE
NO SLOPE
NOX OFFS
NO OFFS
Minimum Value
-0.700
Optimum Value
1.000
Maximum Value
1.300
-0.700
1.000
1.300
-20.0 mV
-20.0 mV
0.0 mV
150.0 mV
150.0 mV
0.0 mV
The default DAS configuration records all calibration values in channel CALDAT as well as all
calibration check (zero and span) values in its internal memory.
Up to 200 data points are stored for up 4 years of data (on weekly calibration checks) and a lifetime
history of monthly calibrations.
Review these data to see if the zero and span responses change over time.
These channels also store the STB figure (standard deviation of NOX concentration) to evaluate if
the analyzer response has properly leveled off during the calibration procedure.
Finally, the CALDAT channel also stores the converter efficiency for review and documentation.
Teledyne Analytical Instruments
219
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Calibration Procedures
9.7. GAS FLOW CALIBRATION
Rate of sample gas and O3 flow through the 9110T is a key part of the NOx, NO and NO2 concentration
calculations. The FLOW CALIBRATION submenu located under the DIAG menu allows the
calibration/ adjustment of these calculations.
Note
To calibrate the flow of gas calculations made by the CPU, press.
: .
Teledyne Analytical Instruments
220
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
10. EPA PROTOCOL CALIBRATION
To ensure high quality, accurate measurements at all times, the 9110T analyzer must be calibrated prior
to use. A quality assurance program centered on this aspect and including attention to the built-in
warning features of the analyzer, periodic inspection, regular zero/span checks, regular evaluation of test
parameters for predictive diagnostics and data analysis and routine maintenance of the instrument are
paramount to achieving this goal.
The US EPA strongly recommends that you obtain a copy of the publication Quality Assurance
Handbook for Air Pollution Measurement Systems (abbreviated, Q.A. Handbook Volume II); USEPA
Order Number: EPA454R98004; or NIST Order Number: PB99-129876.
This manual can be purchased from:
EPA Technology Transfer Network (http://www.epa.gov/ttn/amtic)
National Technical Information Service (NTIS, http://www.ntis.gov/)
Special attention should be paid to Section 2.3 of the handbook8 which covers the measurement of NO2.
Specific regulations regarding the use and operation of ambient NOx analyzers can be found in
Reference 1 at the end of this Section.
If the 9110T is used for EPA compliance monitoring, it must be calibrated in accordance with the
instructions in this section.
A bibliography and references relating to NO2 monitoring are listed in Section 10.10.
10.1. 9110T CALIBRATION – GENERAL GUIDELINES
In general, calibration is the process of adjusting the gain and offset of the 9110T against a standard with
certified, traceable concentration. The reliability of data derived from the analyzer depends primarily
upon its state of calibration.
In this section, the term dynamic calibration is used to express a multipoint check against known
standards and involves introducing gas samples of known concentration into the instrument in order to
adjust the instrument to a predetermined sensitivity and to produce a calibration relationship. This
relationship is derived from the instrumental response to successive samples of different known
concentrations. As a minimum, three reference points and a zero point are recommended to define this
relationship. The instrument(s) supplying the zero air and Span calibration gases used must themselves
be calibrated and that calibration must be traceable to an EPA/ NIST primary standard (see Section 2.0.7
of the Q.A. Handbook and Table 3-8 of this instruction manual)
All monitoring systems are subject to some drift and variation in internal parameters and cannot be
expected to maintain accurate calibration over long periods of time. Therefore, it is necessary to
dynamically check the calibration relationship on a predetermined schedule. Zero and span checks must
be used to document that the data remain within control limits. These checks are also used in data
reduction and validation. The internal data acquisition system of the 9110T allows to store all calibration
checks (as well as full calibrations) over long periods of time for documentation.
Table 10-1 summarizes the initial quality assurance activities for calibrating equipment.
Table 10-2 contains a matrix for the actual, dynamic calibration procedure.
Teledyne Analytical Instruments
221
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Calibrations should be carried out at the field monitoring site. The analyzer should be in operation for at
least several hours (preferably overnight) before calibration so that it is fully warmed up and its
operation has stabilized. During the calibration, the 9110T:
Should be in the CAL mode,
The test atmosphere should be supplied through all components used during normal ambient
sampling and through as much of the ambient air inlet system as is practicable.
If the instrument will be used on more than one range, it should be calibrated separately on each
applicable range.
Details of documentation, forms and procedures should be maintained with each analyzer and also
in a central backup file as described in Section 12 of the Quality Assurance Handbook.
Personnel, equipment and reference materials used in conducting audits must be independent from
those normally used in calibrations and operations.
10.2. CALIBRATION EQUIPMENT, SUPPLIES, AND
EXPENDABLES
The measurement of NOx, NO and NO2 in ambient air requires a certain amount of basic sampling
equipment and supplemental supplies. These include, but are not limited to, the following:
Table 10-1: Activity Matrix for EPA Calibration Equipment and Supplies
Equipment/
Supplies
Frequency And Method Action If Requirements
Acceptance Limits
of Measurement
Are Not Met
Recorder
Compatible with output signal of
analyzer; min chart width of 150 mm (6
in) is recommended
Check upon receipt
Return equipment to supplier
Sample line and Constructed of PTFE, glass or stainless
Check upon receipt
Return equipment to supplier
manifold
steel
Calibration
equipment
Meets guidelines and Section 2.3.2 of
Q. A. Handbook
See Section 2.3.9 of Q. A. Return equipment/ supplies to
Handbook
supplier or take corrective
action
Working standard Traceable to NIST-SRM SRM. Meets Analyzed against NIST-SRM; Obtain new working standard
NO cylinder gas
limits in traceability protocol for
accuracy and stability. Section 2.0.7 of
Q. A. Handbook
see protocol in Section 2.0.7,
Q.A. Handbook
and check for traceability
Recording forms
Audit equipment
Develop standard forms
N/A
Revise forms as appropriate
Cannot be the same as used for
calibration
System must be checked out Locate problem and correct or
against known standards return to supplier
When purchasing these materials, a logbook should be maintained as a reference for future procurement
needs and as a basis for future fiscal planning.
10.2.1. SPARE PARTS AND EXPENDABLE SUPPLIES
In addition to the basic equipment described in the Q.A. Handbook, it is necessary to maintain an
inventory of spare parts and expendable supplies. Section 11 describes the parts that require periodic
replacement and the frequency of replacement. Appendix B contains a list of spare parts and kits of
expendables supplies.
Teledyne Analytical Instruments
222
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
10.2.2. CALIBRATION GAS AND ZERO AIR SOURCES
10.2.2.1. Zero Air
Devices that condition ambient air by drying and removal of pollutants are available on the commercial
market such as the Teledyne M701 zero air generator.
10.2.2.2. Span Gas
NO and NOx calibration gases supplied either directly from pressurized bottles of calibrated gas of
different concentrations (see Table 3-8 or from a single source of high concentration gas then mixed to
create various lower concentrations. In the latter case, we recommend using a gas dilution calibrator
such as a Teledyne M700E.
In ALL cases, the gases and instrument(s) used must be calibrated and that calibration must be traceable
to an EPA/NIST primary standard.
10.2.3. DATA RECORDING DEVICE
Either a strip chart recorder, data acquisition system, digital data acquisition system should be used to
record the data from the 9110T RS-232 port or analog outputs. If analog readings are being used, the
response of that system should be checked against a NIST referenced voltage source or meter. Data
recording device should be capable of bi-polar operation so that negative readings can be recorded. Strip
chart recorder should be at least 6” (15 cm) wide.
10.2.4. RECORD KEEPING
Record keeping is a critical part of all quality assurance programs. Standard forms similar to those that
appear in this manual should be developed for individual programs. Three things to consider in the
development of record forms are:
Does the form serve a necessary function?
Is the documentation complete?
Will the forms be filed in such a manner that they can easily be retrieved when needed?
10.3. CALIBRATION FREQUENCY
A system of Level 1 and Level 2 zero/span checks is recommended (see Section 10.4). Level 1 zero and
span checks should be conducted at least every two weeks. Level 2 checks should be conducted in
between the Level 1 checks at a frequency determined by the user. Span concentrations for both levels
should be between 70 and 90% of the reporting range.
To ensure accurate measurements of the NO, NOX, and NO2 concentrations, calibrate the analyzer at the
time of installation, and recalibrate it:
No later than three months after the most recent calibration or performance audit that indicated the
analyzer calibration to be acceptable.
An interruption of more than a few days in analyzer operation.
Teledyne Analytical Instruments
223
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Any repairs which might affect its calibration.
Physical relocation of the analyzer.
Any other indication (including excessive zero or span drift) of possible significant inaccuracy of the
analyzer.
Following any of the activities listed above, the zero and span should be checked to determine if a
calibration is necessary. If the analyzer zero and span drifts exceed the calibration limits in Section 12 of
the Q.A. Handbook6, a calibration should be performed.
10.4. LEVEL 1 CALIBRATIONS VERSUS LEVEL 2 CHECKS
All monitoring instruments are subject to some drift and variation in internal parameters and cannot be
expected to maintain accurate calibration over long periods of time the EPA requires a schedule of
periodic checks of the analyzer’s calibration be implemented. Zero and span checks must be used to
document that the data remains within required limits. These checks are also used in data reduction and
system validation.
A Level 1 Span check is used to document that the 9110T is within control limits and must be conducted
every 2 weeks. A Level 2 Span Check is to be conducted between the Level 1 Checks on a schedule to
be determined by the user.
Table 10-2: Definition of Level 1 and Level 2 Zero and Span Checks
LEVEL 1 ZERO AND SPAN CALIBRATION
LEVEL 2 ZERO AND SPAN CHECK
A Level 1 zero and span calibration is a simplified, A Level 2 zero and span check is an "unofficial" check of an
two-point analyzer calibration used when analyzer
linearity does not need to be checked or verified.
analyzer's response. It may include dynamic checks made with
uncertified test concentrations, artificial stimulation of the
(Sometimes when no adjustments are made to the analyzer's detector, electronic or other types of checks of a
analyzer, the Level 1 calibration may be called a
zero/span check, in which case it must not be
confused with a Level 2 zero/span check.) Since
most analyzers have a reliably linear or near-
linear output response with concentration, they
can be adequately calibrated with only two
concentration standards (two-point concentration).
Furthermore, one of the standards may be zero
concentration, which is relatively easily obtained
and need not be certified. Hence, only one
certified concentration standard is needed for the
two-point (Level 1) zero and span calibration.
Although lacking the advantages of the multipoint
calibration, the two-point zero and span
calibration--because of its simplicity--can be (and
should be) carried out much more frequently.
Also, two-point calibrations are easily automated.
Frequency checks or updating of the calibration
relationship with a two-point zero and span
calibration improves the quality of the monitoring
data by helping to keep the calibration relationship
more closely matched to any changes (drifts) in
the analyzer response.
portion of the analyzer, etc.
Level 2 zero and span checks are not to be used as a basis for
analyzer zero or span adjustments, calibration updates, or
adjustment of ambient data. They are intended as quick,
convenient checks to be used between zero and span
calibrations to check for possible analyzer malfunction or
calibration drift. Whenever a Level 2 zero or span check
indicates a possible calibration problem, a Level 1 zero and
span (or multipoint) calibration should be carried out before any
corrective action is taken.
If a Level 2 zero and span check is to be used in the quality
control program, a "reference response" for the check should be
obtained immediately following a zero and span (or multipoint)
calibration while the analyzer's calibration is accurately known.
Subsequent Level 2 check responses should then be compared
to the most recent reference response to determine if a change
in response has occurred. For automatic Level 2 zero and span
checks, the first scheduled check following the calibration
should be used for the reference response. It should be kept in
mind that any Level 2 check that involves only part of the
analyzer's system cannot provide information about the portions
of the system not checked and therefore cannot be used as a
verification of the overall analyzer calibration.
In addition, an independent precision check between 0.08 and 0.10 ppm must be carried out at least once
every two weeks. Table 10-1 summarizes the quality assurance activities for routine operations. A
discussion of each activity appears in the following sections.
Teledyne Analytical Instruments
224
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
To provide for documentation and accountability of activities, a checklist should be compiled and then
filled out by the field operator as each activity is completed.
For information on shelter and sample inlet system, an in-depth study is in Field Operations Guide for
Automatic Air Monitoring Equipment, Publication No. APTD-0736, PB 202-249 and PB 204-650, U.S.
Environmental Protection Agency, Office of Air Programs, October 1972
10.5. GAS PHASE TITRATION (GPT)
10.5.1. GPT PRINCIPLE OF OPERATION
Gas phase titration (GPT) is recommended during calibration of the 9110T. Those using a NO2
permeation tube should refer to the CFR1.
The principle of GPT is based on the rapid gas phase reaction between NO and O3 that produces
stoichiometric quantities of NO2 as shown by the following equation:
Equation 10-1
NOO3 NO O2 h
2
Given that the O3 concentration is known for this reaction, the resultant concentration of NO2 can be
determined. Ozone is added to excess NO in a dynamic calibration system as shown in Figure 10-1,
and the NO channel of the chemiluminescence analyzer detects the changes in NO concentration.
After the addition of O3, the observed decrease in NO concentration on the calibrated NO channel is
equivalent to the concentration of NO2 produced.
The amount of generated NO2 may be varied by adding varying amounts of O3 from a stable O3
generator.
All zero air used in this procedure should conform to the requirements stated in Section 10.2 .
Dynamic calibration systems based on this principle are commercially available, or may be assembled by
the user. A recommended calibration system is described in the Federal Register CFR1.
10.5.2. GPT CALIBRATOR CHECK PROCEDURE
It has been empirically determined that the NO-O3 reaction is complete (<1% residual O3) if the NO
concentration in the reaction cell (ppm) multiplied by the residence time (min.) of the reactants in the
chamber is >2.75 ppm min. The theory behind the development of this equation is in the Federal
Register1. The following procedures and equations should be used to determine whether an existing GPT
calibration system will meet required conditions for a specific calibration.
For calibrators that have known pre-set flow rates, use equations Equation 10-6 and Equation 10-7 below
to verify the required conditions.
If the calibrator does not meet specifications, follow the complete procedure to determine what flow
modifications must be made.
1. Select an NO standard gas that has a nominal concentration in the range of 50 to
100 ppm.
Determine the exact concentration [NO]STD by referencing against an NIST-
SRM SRM, as discussed in the Q.A. Handbook.
Teledyne Analytical Instruments
225
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
2. Determine the volume (cm3) of the calibrator reaction cell (VRC). If the actual
volume is not known, estimate the volume by measuring the approximate
dimensions of the chamber and using an appropriate formula.
3. Determine the required minimum total flow output (FT) :
Equation 10-2
FT = analyzer flow demand (cm3/min) x 110/100
If more than one analyzer is to be calibrated at the same time, multiply FT by
the number of analyzers.
4. Calculate the NO concentration [NO]OUT needed to approximate 90% of the URL of
the NO2 analyzer to be calibrated:
Equation 10-3
[NO]OUT = URL of analyzer (ppm) x 90/100
5. Calculate the NO flow (FNO) required to generate the NO concentration [NO]OUT
:
Equation 10-4
[NO x
]
FT
OUT
=
FNO
[NO
]
STD
6. Calculate the required flow through the ozone generator (FO):
Equation 10-5
[NO
X
X
]
FNO VRC
STD
=
-
Fo
FNO
2.75 ppm - min
7. Verify that the residence time (tR) in the reaction cell is <2 min:
Equation 10-6
VRC
t
R
2min
F
O
FNO
8. Verify that the dynamic parameter specification (PR) of the calibrator's reaction cell
is >2.75 ppm-min:
Teledyne Analytical Instruments
226
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Equation 10-7
FNO
VRC
P
R
NO
STD
2.75
F
O
FNO
F
O
FNO
Note
If TR is >2 minutes or if PR is <2.75 ppm min, changes in flow conditions
(FT, FO, FNO) or in the reaction cell volume (VRC), or both will
have to be made, and TR and PR will have to be recalculated.
9. After Equation 10-5 and Equation 10-6 are satisfied, calculate the diluent air flow
(FD):
Equation 10-8
FD
FT FO FNO
10.5.2.1. Example GPT Calculation
The following is an example calculation that can be used to determine whether an existing GPT
calibrator will meet the required conditions for a specific calibration. For this example, it is assumed that
only the volume of the reaction cell, VRC, and the concentration of the NO standard, [NO]STD, are known.
All flow settings (FNO, FO, FT, and FD) will be calculated. In many uses, these flow settings are known
and need only to be substituted in Equations 8-5 and 8-6 to verify the required conditions. Before doing
any calculations, the URL and flow demand of the analyzer being calibrated must be known. Operating
parameters are determined from the operations manual:
Upper range limit = 0.5 ppm, and
Flow demand = 500 cm3/min.
Volume of calibrator reaction cell is determined by physical measurement: VRC = 180 cm3
The concentration of the NO standard gas to be used is determined by reference against an NIST-
SRM SRM (Section 2.0.7, Q.A. Handbook):
[NO]STD = 50.5 ppm
1. Determine the minimum total flow (FT) required at the output manifold:
FT = 500 cm3/min (110/100) = 550 cm3/min
Because low flows are difficult to control and measure, it is often advantageous to set a higher total flow
than needed. In this example, we will set FT to 2750 cm3/min.
2. Determine the highest NO concentration, [NO]OUT, required at the output manifold,:
[NO]OUT = 0.5 ppm (90/100) = 0.45 ppm
3. Calculate the NO flow (FNO) required to generate the NO concentration [NO]OUT
:
Teledyne Analytical Instruments
227
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
0.45 ppm 2750
3 / min
cm
50.5ppm
F NO
cm3
/ min
=
24.5
4. Calculate the required flow rate through ozone generator (FO):
3
50.5 ppm x 24.5
3 /min x180
cm
cm
=
- 24.5
3 / min
cm
FO
2.75 ppm - min
6
2
= 80984
/
- 24.5 3 /min=260
cm
3 /min
cm
cm min
5. Verify that the residence time (tR) in the reaction cell is <2 min:
3
180
cm
tR
=
0.63min
260 3 /min 24.5 3 /min
cm cm
6. Verify the dynamic parameter specification (PR) of the calibrator reaction cell:
3
24 3 /min
cm
180
cm
PR
= 50.5ppm
2.75ppmmin
260 3 /min 24.5 3 /min 260 3 /min 24.5 3 /min
cm cm cm cm
7. Calculate the diluent airflow (FD) required at the mixing chamber:
FD = 2750 cm3/min - 260 cm3/min - 24.5 cm3/min = 2465.5 cm3/min
10.6. GPT MULTIPOINT CALIBRATION PROCEDURE
The procedure for calibration of chemiluminescence NOx analyzers by GPT is specified in the Federal
Register.1 This section applies the general procedure to the specific case of the 9110T.
Calibration must be performed with a calibrator that meets all conditions specified in the Q.A.
Handbook.
Note
The user should be sure that all flow meters are calibrated under the
conditions of use against a reliable standard.
All volumetric flow rates should be corrected to 25o C (78oF) and 760 mm
(29.92 in.) Hg. Calibrations of flow meters are discussed in the QA
Handbook, Vol. II, Part 1, Appendix 126.
Gas Phase Titration (GPT) requires the use of the NO channel of the analyzer to determine the amount
of NO2 generated by titration. Therefore it is necessary to calibrate and determine the linearity of the NO
channel before proceeding with the NO2 calibration. It is also necessary to calibrate the NOx channel.
This can be done simultaneously with the NO calibration. During the calibration the 9110T should be
operating in its normal sampling mode, and the test atmosphere should pass through all filters, scrubbers,
conditioners, and other components used during normal ambient sampling and as much of the ambient
air inlet system as is practicable. All operational adjustments to the 9110T should be completed prior to
the calibration.
Teledyne Analytical Instruments
228
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
10.6.1. SET UP FOR GPT MULTIPOINT CALIBRATION OF THE 9110T
The following 9110T features must be activated or set before calibration.
Calibrate the NO2 NO converter (see Section 9.1.4).
Set the reporting ranges for Independent mode (see Section 5.4.3.2).
Turn ON the automatic temperature/pressure compensation (TPC) feature (see Section 5.8).
Set the units of measure to ppb (see Section 5.4.3.4).
Figure 10-1:
GPT Calibration System
Teledyne Analytical Instruments
229
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
IMPORTANT
IMPACT ON READINGS OR DATA
The analyzer should be calibrated on the same range used for
monitoring.
Ensure that the GPT calibration system can supply the range of
concentrations at a sufficient flow over the whole range of
concentrations that will be encountered during calibration.
10.6.2. ZERO CALIBRATION
The zero point calibration procedure is identical to that described in Sections 9.1, 9.2, 9.4 or 9.5.
10.6.3. SPAN CALIBRATION
1. Adjust the NO concentration to approximately 80% of the URL of the NO channel.
The expected NO and NOX span concentrations can be determined by measuring the
cylinder and diluent flows and computing the resulting concentrations.
If there is any NO2 impurity in the NO standard gas it should be taken into account when the
NOX concentration is entered during the NO/NOX channel calibration.
This is done by ADDING the impurity concentration to the NO concentration to get the NOX
concentration for calibration. Calculate the exact NO and NOX concentrations as follows:
Equation 10-9
x [NO
FT
]
FNO
STD
[NO
=
]
OUT
2. Enter the respective concentrations using the procedure in Section 4.2. The
expected span concentrations need not be re-entered each time a calibration is
performed unless they are changed.
3. Enter the expected NOX and NO span gas concentrations:
4. Sample the generated concentration until the NO and the NOX responses have
stabilized.
5. Span the instrument by the following the same method as 9.1, 9.3 or 9.5.
The analog voltage output should measure 80% of the voltage range selected. (e.g.
4.00 VDC if 0-5V output is selected.)
The readings on the front panel display should be equal to the expected NO and NOX
concentrations.
Teledyne Analytical Instruments
230
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Note
See the Troubleshooting Section 12 if there are problems.
Also see the Calibration Quality Check procedure Section 9.6.
6. After the zero and the 80% upper range limit points have been set, generate five
approximately evenly spaced calibration points between zero and 80% upper range
limit without further adjustment to the instrument.
7. Allow the instrument to sample these intermediate concentrations for about 10
minutes each and record the instrument NO and NOX responses.
8. Plot the analyzer NO and NOX responses versus the corresponding calculated
concentrations to obtain the calibration relationships.
Determine the straight line of best fit (y = mx + b) determined by the method of
least squares.
9. After the best-fit line has been drawn for the NO and the NOX calibrations,
determine whether the analyzer response is linear.
To be considered linear, no calibration point should differ from the best-fit line
by more than 2% of full scale.
10.7. GPT NO2 CHECK
The 9110T computes the NO2 concentration by subtracting the NO from the NOX concentration. Unlike
analog instruments, this difference is calculated by the 9110T's internal computer software. It is
extremely unlikely that the NO2 concentration will be in error. Therefore this procedure is a confirmation
that the NO2 subtraction algorithm in the computer is operating correctly.
IMPORTANT
IMPACT ON READINGS OR DATA
Do not make any adjustments to the instrument during this procedure.
1. Generate an NO concentration near 90% of the upper range limit.
Dilution air and O3 generator air flows should be the same as used in the
calculation of specified conditions of the dynamic parameter according to
Section 10.5.
2. Sample this NO concentration until the NO and NOx responses stabilize. Record
the NO and NOx concentrations.
3. Turn ON and adjust the O3 generator in the calibrator to produce sufficient O3 to
decrease the NO concentration to about 10% of full scale.
This will be equivalent to 80% of the URL of the NO2 channel. After the analyzer
responses stabilize, record the resultant NO, NOX, and NO2 concentrations.
Note
If the NOX reading should drop to less than 96% of its starting value
during this step, it indicates the NO2 converter is in need of
troubleshooting or replacement. See Section 12.7.10 for further details.
Teledyne Analytical Instruments
231
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
4. While maintaining all other conditions, adjust the ozone generator to obtain several
other concentrations of NO2 evenly spaced between the 80% URL point and the
zero point. Record the NO, NOX, and NO2 concentrations for each additional point.
5. Calculate the resulting NO2 concentrations as follows:
Equation 10-10
* [
]
FNO NO2
IMP
[
= [NO
- [NO
REM
+
]
]
]
NO2
OUT
ORIG
FT
Where [NO]ORIG is the NO concentration before the GPT ozone is turned on, and [NO]REM is the NO
remaining after GPT.
Plot the NO2 concentration output by the instrument on the y-axis against the generated NO2 [NO2]OUT
on the x-axis. The plot should be a straight line within the ± 2% linearity criteria given for the NOx and
NO channels. If the plot is not linear, the most likely cause is that the converter needs replacing. See
Section 9.1.4 on NO2 converter efficiency and Section 11.3.8 for changing the converter if needed.
10.8. OTHER QUALITY ASSURANCE PROCEDURES
Precision is determined by a one-point check at least once every two weeks. Accuracy is determined by a
three-point audit once each quarter.
Essential to quality assurance are scheduled checks for verifying the operational status of the monitoring
system. The operator should visit the site at least once each week. Every two weeks a Level 1 zero and
span check must be made on the analyzer. Level 2 zero and span checks should be conducted at a
frequency desired by the user. Definitions of these terms are given in Table 10-2.
10.8.1. SUMMARY OF QUALITY ASSURANCE CHECKS
The following items should be checked on a regularly scheduled basis to assure high quality data from
the 9110T. See Table 10-3 for a summary of activities; also the QA Handbook should be checked for
specific procedures.
Teledyne Analytical Instruments
232
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Table 10-3: Activity Matrix for Data Quality
Frequency and Method
of Measurement
Action if Requirements
are not Met
Characteristic
Acceptance Limits
Shelter
temperature
Check thermograph chart
weekly for variations greater
than ± 2C (4F)
Mark strip chart for the
affected time period.
Mean temperature between 22C
and 28C (72 and 82 F), daily
fluctuations not greater than 2C
Repair or adjust
temperature control.
Sample
introduction
system
No moisture, foreign material,
leaks, obstructions; sample line
connected to manifold
Weekly visual inspection
Weekly visual inspection
Clean, repair, or replace as
needed.
Recorder
Adequate ink & paper
Legible ink traces
Replenish ink and paper
supply.
Adjust time to agree with
clock; note on chart.
Correct chart speed and range
Correct time
Analyzer
operational
settings
TEST measurements at nominal
values
Weekly visual inspection
Level 1 zero/span every
Adjust or repair as needed.
9110T in SAMPLE mode
Analyzer
operational
check
Zero and span within tolerance
Find source of error and
repair.
After corrective action, re-
calibrate analyzer.
limits as described in Section 12 of 2 weeks; Level 2 between
the Q.A. Handbook6
Level 1 checks at frequency
desired by user
Precision check Assess precision as described in
Sections 15&18 of the Q.A.
Every 2 weeks, Subsection
3.4.3 (Ibid.)
Calc, report precision,
Section 12 of the Q.A.
Handbook6.
Handbook6
10.8.2. SHORT CALIBRATION CHECKS
A system of Level 1 and Level 2 zero/span checks (Table 10-2) is recommended. These checks must be
conducted in accordance with the specific guidance given in Section 12 of the Q.A. Handbook6. Level 1
zero and span checks must be conducted every two weeks. Level 2 checks should be conducted in
between the Level 1 checks at a frequency desired by the user. Span concentrations for both levels
should be between 70 and 90% of the measurement range.
Zero and span data are to be used to:
Provide data to allow analyzer adjustment for zero and span drift;
Provide a decision point on when to calibrate the analyzer;
Provide a decision point on invalidation of monitoring data.
These items are described in detail in Sections 15 & 18 of the Q.A. Handbook6. Refer to Section 11 of this
manual if the instrument is not within the allowed margins. We recommend using APICOM and the DAS
for analysis and documentation of zero/span check data.
10.8.3. ZERO/SPAN CHECK PROCEDURES
The Zero and span calibration can be checked in a variety of different ways. They include:
Manual zero/span checks can be done from the front panel touchscreen. The procedure is in
Section 9.2.2 of this manual.
Automatic zero/span checks can be performed every night. See Section 9.5 of this manual for setup
and operation procedures.
Teledyne Analytical Instruments
233
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Zero/Span checks through remote contact closure can be initiated through remote contact closures
on the rear panel. See Section 9.4.3.3 of this manual.
Zero/span checks can also be controlled through the RS-232 port. See Section 8 of this manual for
more details on setting up and using the analyzer’s RS-232 port.
10.8.4. PRECISION CHECK
A periodic check is used to assess the data for precision. A one-point precision check must be carried out
at least once every 2 weeks on each analyzer at an NO2 concentration between 0.08 and 0.10 ppm.
The analyzer must be operated in its normal sampling mode, and the precision test gas must pass
through all filters, scrubbers, conditioners, and other components used during normal ambient
sampling.
The standards from which precision check test concentrations are obtained must be traceable to
NIST-SRM SRM. Those standards used for calibration or auditing may be used.
10.8.4.1. Precision Check Procedure
Connect the analyzer to a precision gas that has an NO2 concentration between 0.08 and 0.10 ppm. An
NO2 precision gas may be generated by either GPT or a NO2 permeation tube. If a precision check is
made in conjunction with a zero/span check, it must be made prior to any zero or span adjustments.
Allow the analyzer to sample the precision gas until a stable trace is obtained.
Record this value.
NO and NOX precision checks should also be made if those data are being reported. Information
from the check procedure is used to assess the precision of the monitoring data; see in Section 12
of the Q.A. Handbook6 for procedures for calculating and reporting precision.
10.9. CERTIFICATION OF WORKING STANDARDS
The NO content of the NO working standard must be periodically assayed against NIST-traceable NO or
NO2 standards. Any NO2 impurity in the cylinder must also be assayed. Certification of the NO working
standard should be made on a quarterly basis or more frequently, as required. Procedures are outlined
below for certification against NO traceable standard which is the simplest and most straightforward
procedure.
To assure data of desired quality, two considerations are essential:
The measurement process must be in statistical control at the time of the measurement and;
Any systematic errors, when combined with the random variation in the measurement process, must
result in a suitably small uncertainty.
Evidence of good quality data includes documentation of the quality control checks and the independent
audits of the measurement process by recording data on specific forms or on a quality control chart and
by using materials, instruments, and measurement procedures that can be traced to appropriate standards
of reference.
Teledyne Analytical Instruments
234
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
To establish traceability, data must be obtained routinely by repeated measurements of standard
reference samples (primary, secondary, and/or working standards). More specifically, working
calibration standards must be traceable to standards of higher accuracy.
10.9.1.1. Certification Procedures of Working Standards
This procedure requires the use of calibrated NO gas traceable to an NIST-SRM SRM and the gas phase
titration calibration procedure (Section 10.5) to calibrate the NO, NOX, and NO2 responses of the
analyzer. Also the efficiency of the analyzer’s NO2 converter must be determined (Section 9.1.4).
Generate several NO concentrations by diluting the NO working standard. Use the nominal NO cylinder
concentration, [NO]NOM, to calculate the diluted concentrations. Plot the analyzer NO response (in ppm)
versus the nominal diluted NO concentration and determine the slope, SNOM. Calculate the NO
concentration of the working standard [NO]STD from:
Equation 10-11
[NO]STD = [NO]NOM x SNOM
A more detailed procedure is presented in Reference 1 (Section 10.10).
10.9.1.2. Other Methods of Establishing Traceability
Methods of establishing traceability are:
Using a NO working standard traced to NIST NO2 standard.
Using a NO2 working standard traced to NIST NO2 standard.
Using a NO2 working standard traced to NIST NO standard.
IMPORTANT
IMPACT ON READINGS OR DATA
If the NO2 impurity in the NO cylinder, [NO2]imp is greater than the 1 ppm
value allowed in the calibration procedure, check that the NO delivery
system is not the source of contamination before discarding the NO
standard.
For further information on calibration by GPT and NO2 permeation
devices, refer to part 50 of Section 1, Title 40 CFR, Appendix F Reference
13 of that Appendix.
10.10. REFERENCES
1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50, Appendix F,
Measurement Principle and Calibration Procedure for the Measurement of Nitrogen Dioxide in the
Atmosphere (Gas Phase Chemiluminescence), Federal Register, 41 (232), 52688-52692, December
1976 (as amended at 48 FR 2529, Jan 20, 1983).
2. Ellis, Elizabeth C. Technical Assistance Document for the Chemiluminescence Measurement of Nitrogen
Dioxide, U.S. Environmental Protection Agency, Research Triangle Park, NC. 83 pages, December
1975. Available online at http://www.epa.gov/ttn/amtic/files/ambient/criteria/reldocs/4-75-003.pdf.
3. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 58, Appendix A,
Measurement Principle and Calibration Procedure for the Measurement of Nitrogen Dioxide in the
Teledyne Analytical Instruments
235
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
EPA Protocol Calibration
Atmosphere (Gas Phase Chemiluminescence), Federal Register, 41 (232), 52688-52692, December
1976 (as amended at 48 FR 2529, Jan 20, 1983).
4. Mavrodineanu, R., and Gills, T. E., Standard Reference Materials: Summary of Gas Cylinder and
Permeation Tube Standard Reference Materials Issued by the National Bureau of Standards, Document
SP260-108, May 1987.
And: Taylor, J. K., Standard Reference Materials: Handbook for SRM Users, Document number SP260-
100, February 1993. Available online at:
http://patapsco.nist.gov/srmcatalog/sp_publications/publications.htm
5. Quality Assurance Handbook for Air Pollution Measurement Systems - Volume I, “A Field Guide to
Environmental Quality Assurance," EPA-600/R-94/038a, April 1994. Available online at:
http://www.epa.gov/ttn/amtic/qabook.html.
6. Quality Assurance Handbook for Air Pollution Measurement Systems - Volume II, Ambient Air Specific
Methods. EPA-600/4-77/027a, December 1986. US EPA Order Number: 454R98004, available at the
National Technical Information Service (NTIS), 5285 Port Royal Rd Springfield, VA 22151. Portions are
also available at: http://www.epa.gov/ttn/amtic/qabook.html.
7. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 58, Appendix B,
Measurement Principle and Calibration Procedure for the Measurement of Nitrogen Dioxide in the
Atmosphere (Gas Phase Chemiluminescence), Federal Register, 41 (232), 52688-52692, December
1976 (as amended at 48 FR 2529, Jan 20, 1983).
8. Quality Assurance Guidance Document. Reference Method for the Determination of Nitrogen Dioxide in
the Atmosphere (Chemiluminescence). Draft document, 58 pages, February 2002. Office of Air Quality
Planning and Standards, Research Triangle Park NC 27711, draft document available at
http://www.epa.gov/ttn/amtic/qabook.html. Guidelines about the measurement of NO2 in this document
replace those in the old QA Handbook and should be consulted as the latest reference.
Teledyne Analytical Instruments
236
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Part III
PART III
–
MAINTENANCE AND SERVICE
Teledyne Analytical Instruments
237
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Part III
: This page intentionally left blank.
Teledyne Analytical Instruments
238
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11. INSTRUMENT MAINTENANCE
For the most part, the 9110T analyzer is maintenance free. However it is recommended that a minimal
number of simple procedures be performed regularly to ensure that the 9110T continues to operate
accurately and reliably over its lifetime. In general, the exterior can be wiped down with a lightly damp
cloth; avoid spraying anything directly onto any part of the analyzer.
Service and troubleshooting are covered in Section Troubleshooting & Service.
11.1. MAINTENANCE SCHEDULE
Table 11-1 shows a typical maintenance schedule for the 9110T. Please note that in certain
environments (i.e. dusty, very high ambient pollutant levels) some maintenance procedures may need to
be performed more often than shown.
IMPORTANT
IMPACT ON READINGS OR DATA
A span and zero calibration check (see CAL CHECK REQ’D Column of
Table 11-1, 9110T Maintenance Schedule) must be performed following
some of the maintenance procedures listed herein.
To perform a CHECK of the instrument’s Zero or Span Calibration, follow
the same steps as described in Section 9.3.
DO NOT press the ENTR button at the end of each operation. Pressing
the ENTR button resets the stored values for OFFSET and SLOPE and
alters the instrument’s Calibration.
Alternately, use the Auto Cal feature described in Section 9.5 with the
CALIBRATE attribute set to OFF.
WARNING – ELECTRICAL SHOCK HAZARD
Disconnect power before performing any of the following operations that require
entry into the interior of the analyzer.
CAUTION – QUALIFIED PERSONNEL
These maintenance procedures must be performed by qualified technicians only.
Teledyne Analytical Instruments
239
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
Table 11-1: 9110T Maintenance Schedule
DATE PERFORMED
CAL
CHECK
REQ’D
No
ITEM
ACTION
FREQ
TEST
functions
Review and
evaluate
Weekly
Weekly
Weekly
Particulate
filter
Change
particle filter
No
No
Zero/span
check
Evaluate
offset and
slope
Zero/span
calibration
Zero and
span
calibration
Every 3 months
Every 3 months
Yes
External zero
air scrubber
option
Exchange
chemical
No
External
dryer option
Replace
chemical
When indicator
color changes
No
Ozone
cleanser
Change
chemical
Annually
Yes
Yes
Reaction cell
window
Clean
Annually or as
necessary
(“optical
filter” in
Figure 11-6)
DFU filters
Change
Annually
No
particle filter
Pneumatic
sub-system
Check for
leaks in gas
flow paths
Annually or
after repairs
involving
Yes if a
leak is
repaired
pneumatics
Reaction cell
O-rings &
sintered
Replace
Annually
Yes
Yes
filters
PMT Sensor
Hardware
Calibration
Low-level
hardware
calibration
On PMT/
preamp
changes or if
slope is outside
of 1.0±0.3
Pump
Rebuild head
Replace
when RCEL
pressure
exceeds 10 in-
Hg-A (at sea
level)
Yes
Inline
Annually
No
Exhaust
Scrubber
Replace
converter
Every 3 years
or if conversion
efficiency drops
below 96%
Yes
NO2
converter
Desiccant
bags
Replace
Any time PMT
housing is
n/a
opened for
maintenance
Teledyne Analytical Instruments
240
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11.2. PREDICTIVE DIAGNOSTICS
Predictive diagnostic functions including failure warnings and alarms built into the analyzer’s firmware
allow the user to determine when repairs are necessary.
The Test Functions can also be used to predict failures by looking at how their values change over time.
Initially it may be useful to compare the state of these Test Functions to the values recorded on the
printed record of the Final Test and Validation Data Sheet, P/N 04490.
The following table can be used as a basis for taking action as these values change with time. The
internal data acquisition system (DAS) is a convenient way to record and track these changes. Use
APICOM (Section 8.1.1) to download and review this data from a remote location.
Table 11-2: Predictive Uses for Test Functions
FUNCTION
EXPECTED
ACTUAL
INTERPRETATION & ACTION
Fluctuating
Developing leak in pneumatic system. Check for leaks.
RCEL
(pressure)
Constant to within
± 0.5 in-Hg-A
Pump performance is degrading. Rebuild pump when pressure
is above 10 in-Hg-A.
Slowly increasing
Fluctuating
Developing leak in pneumatic system. Check for leaks.
Flow path is clogging up. Replace orifice filters.
Constant within
atmospheric
changes
SAMP
Slowly increasing
(pressure)
Developing leak in pneumatic system to vacuum (developing
valve failure). Check for leaks.
Slowly decreasing
Slowly decreasing
Constant to within
± 15
Flow path is clogging up. Replace orifice filters.
OZONE FL
AZERO
Developing AZERO valve failure. Replace valve.
PMT cooler failure. Check cooler, circuit, and power supplies.
Developing light leak.
Constant within
±20 of check-out
value
Significantly increasing
O3 air filter cartridge is exhausted. Change chemical.
Constant for
constant
concentrations
Slowly decreasing
signal for same
concentration
NO2
(Concentration)
Converter efficiency may be degrading. Replace converter
components.
Change in instrument response. Low level (hardware) calibrate
the sensor.
NO2
with IZS Option
installed
Decreasing over time
Constant
response from day
to day
Degradation of IZS permeation tube. Change permeation tube.
Heavily fluctuating from
day to day
Ambient changes in moisture are affecting the performance.
Add a dryer to the zero air inlet.
(Concentration)
Constant for
constant
concentration
NO
Drift of instrument response; clean RCEL window. Check for
flow leaks or irregularities.
Decreasing over time
(Concentration)
Note
11.3. MAINTENANCE PROCEDURES
The following procedures are to be performed periodically as part of the standard maintenance of the
9110T.
Teledyne Analytical Instruments
241
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11.3.1. REPLACING THE SAMPLE PARTICULATE FILTER
The particulate filter should be inspected often for signs of plugging or contamination. We recommend
that when you change the filter; handle it and the wetted surfaces of the filter housing as little as
possible. Do not touch any part of the housing, filter element, PTFE retaining ring, glass cover and the
o-ring with your bare hands. Teledyne API recommends using gloves or PTFE coated tweezers or
similar handling to avoid contamination of the sample filter assembly.
To change the filter:
1. Turn OFF the analyzer to prevent drawing debris into the instrument.
2. Open the 9110T’s hinged front panel and unscrew the retaining ring on the filter
assembly.
Figure 11-1
Replacing the Particulate Filter
3. Carefully remove the retaining ring, PTFE o-ring, glass filter cover and filter
element.
4. Replace the filter, being careful that the element is fully seated and centered in the
bottom of the holder.
5. Reinstall the PTFE o-ring with the notches up; the glass cover, then screw on the
retaining ring and hand tighten. Inspect the seal between the edge of filter and the
o-ring to assure a proper seal.
6. Restart the Analyzer.
Teledyne Analytical Instruments
242
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11.3.2. CHANGING THE O3 DRYER PARTICULATE FILTER
The air for the O3 generator passes through a Perma Pure© dryer, which is equipped with a small
particulate filter at its inlet. This filter prevents dust from entering the Perma Pure© dryer and degrading
the dryer’s performance over time. Change the filter according to the service interval in Table 11-1 as
follows:
1. Before starting the procedure, check and write down the average RCEL pressure
and the OZONE FLOW values.
2. Turn off the analyzer, unplug the power cord and remove the cover.
3. Unscrew the nut around the port of the filter using two 5/8” wrenches.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Ensure to use proper wrenches.
ATTENTION
Hold the main dryer fitting with a 5/8” wrench to ensure that it does not
turn against the Perma Pure© dryer.
Performing this procedure improperly or with incorrect tools creates a
risk of causing a significant leak.
4. Take off the old filter element and replace it with a suitable equivalent (TAI P/N FL-
3).
Figure 11-2:
Particle Filter on O3 Supply Air Dryer
Teledyne Analytical Instruments
243
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
5. Holding the main dryer fitting steady with a 5/8” wrench and tighten the nut with your
hands.
If necessary use a second wrench but do not over-tighten the nut.
6. Replace the cover, plug in the power cord and restart the analyzer.
7. Check the O3 flow rate; it should be around 80 cm³/min ± 15.
8. Check the RCEL pressure.
It should be the same value as before.
9. Refer to Section 13.3.12 to leak check after installing the new DFU particle filter.
11.3.3. CHANGING THE OZONE CLEANSER CHEMICAL
The ozone (O3) cleanser is located next to the O3 generator (see Figure 3-5) and cleans the O3 stream
from solid and liquid contaminants that are created inside the O3 generator. The content of the ozone
cleanser needs periodical exchange according to Table 11-1. A rebuild kit is available from the factory
(see Appendix B of this manual lists the part numbers).
To change the ozone cleanser chemical, follow these steps:
1. Turn of power to the analyzer and pump. Remove the analyzer cover and locate the
O3 filter in the front of the analyzer next to the O3 generator.
2. Use a 7/16” wrench to remove both pieces of 1/8” male nut with tubing from the
NPT fittings.
Teledyne Analytical Instruments
244
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
Figure 11-3:
Ozone Cleanser Assembly
3. Remove the integrated screws with a Phillips screw driver and remove the scrubber
manifold from the chassis.
4. Using a 9/16” wrench, remove both fittings from the cartridge.
5. Discard the glass wool.
6. Pour the contents of the scrubber manifold onto a sheet of white paper. If
necessary, remove the plug to ensure that all the contents are poured out.
Notice any discoloration of the contents, which is usually white and slightly
transparent.
The amount of discolored chemical (usually with yellow tint) may give you an
indication of the lifetime of the chemical in your application.
Teledyne Analytical Instruments
245
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
The maintenance cycle of this item is dependent on ambient moisture, sub-micron particle load and other
factors and may differ from that shown in Table 11-1.
7. Discard the used silica gel desiccant without touching it. It may contain nitric acid,
which is a corrosive and highly irritating substance.
CAUTION – GENERAL SAFETY HAZARD
Immediately wash your hands after contact with the silica gel disiccant.
8. Using a small powder funnel, fill the cartridge with about 10 g new silica gel
desiccant (TAI P/N CH43) so that it is level on both legs of the cartridge.
Slight vibration is required to settle the chemical into the cartridge and achieve
tightest packing, which increases performance and lifetime of the filter.
Ensure that the level of the chemical does not protrude farther than the first two
threads of the NPT fitting.
IMPORTANT
IMPACT ON READINGS OR DATA
Use only genuine, pre-conditioned Teledyne refill kits for this procedure.
Teledyne refill kits have been properly conditioned to prevent a significant
increase of the 9110T’s Auto Zero value which can cause large negative
offsets, which may take 2-3 weeks to disappear.
Do not leave this material uncovered for more than a few seconds, as it
will absorb contaminants from ambient air. Always store unused, well-
covered refill material in a cool dry place.
9. Seal the silica gel desiccant with 1 cm3 of glass wool on each well.
Ensure that the plug is large enough and compressed into the cartridge so that
the chemical is securely held in place.
10. Add new Teflon tape (P/N HN000036) to the NPT fittings.
11. Screw the NPT fittings back into the scrubber manifold.
12. Screw the cartridge back onto the chassis; orientation is not important.
13. Evaluate the ferrules on the tubing.
If the ferrules are too old, we recommend replacing them with new ferrules.
14. Reconnect the tubing using 7/16” and 9/16” wrenches.
Do not over-tighten the fittings.
15. If the service interval for this item has been exceeded, it may also be necessary to
clean the reaction cell as described in Section 11.3.9.
16. Leak check the system using the pressurized approach described in Section
11.3.12.2.
Teledyne Analytical Instruments
246
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
If necessary, tighten the fittings some more but do not over-tighten.
17. Restart the analyzer and pump and continue operation.
18. Recalibrate the analyzer after one hour (Section 9).
If Auto Zero is high or is changing/not constant, you may have to wait a day until
the silica gel is conditioned before recalibrating the instrument.
11.3.4. MAINTAINING THE EXTERNAL SAMPLE PUMP (PUMP PACK)
11.3.4.1. Rebuilding the Pump
The sample pump head periodically wears out and must be replaced when the RCEL pressure exceeds
10 in-Hg-A (at sea level, adjust this value accordingly for elevated locations).
A pump rebuild kit is available from the factory. Refer to the label on the pump for the part number.
Instructions and diagrams are included in the kit.
A flow and leak check after rebuilding the sample pump is recommended.
A span check and re-calibration after this procedure is necessary as the response of the analyzer
changes with the RCEL pressure.
11.3.4.2. Replacing the Scrubber
CAUTION!
Do NOT attempt to change the contents of the inline exhaust scrubber
cartridge; change the entire cartridge.
1. Through the SETUP>MORE>DIAG menu turn OFF the OZONE GEN OVERRIDE.
Wait 10 minutes to allow pump to pull room air through scrubber before proceeding
to step 2.
2. Disconnect exhaust line from analyzer.
3. Turn off (unplug) analyzer sample pump.
4. Disconnect tubing from (NOx or charcoal) scrubber cartridge.
5. Remove scrubber from system.
6. Dispose of according to local laws.
7. Install new scrubber into system.
8. Reconnect tubing to scrubber and analyzer.
9. Turn on pump.
10. Through the SETUP menu (per Step 1 above) turn ON the OZONE GEN
OVERRIDE.
11.3.5. CHANGING THE PUMP DFU FILTER
The exhaust air from the analyzer passes a small particle filter (Dry Filter Unit (DFU - filter), P/N FL3)
before entering the pump. It should be replaced when:
It becomes visibly dirty or;
Teledyne Analytical Instruments
247
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
The pressure differential between the test functions SAMP and RCEL increases significantly.
11.3.5.1. Procedure for Replacing Filters on External Pumps
1. Power down the analyzer and pump.
2. For internally mounted filters, skip the next two steps.
3. Remove the analyzer exhaust tube from the dust filter.
4. Remove the particle filter from the pump by pushing the white plastic ring into the
fitting and pulling the filter out of the fitting.
If necessary, use needle-nose pliers to pry the filter out of the fittings.
5. Push a new filter into the pump fitting and ensure that the arrow on the filter points
towards the pump.
6. Push the exhaust tubing onto the filter. Skip the next two steps.
7. For internally mounted filters at the inside rear panel, remove the chassis and locate
the filter between the vacuum manifold and the exhaust port fitting.
8. Disconnect the clear tubing from the filter body and change the filter with the arrow
pointing against the gas flow. To remove the hose clamps, slide the two clamp
ends in opposite directions with a needle-nose pliers until the clamp comes apart.
Reconnect the tubing by using the same or new clamps and pushing tightening
them until a good seal is achieved.
9. Restart the pump and clear any error warnings from the front panel display.
10. After about 5 minutes, check the RCEL pressure reading and ensure that it is
similar to its value before changing the filter but less than 10 in-Hg-A.
11.3.5.2. Procedure for Replacing Filters on Internal Pumps
1. Power down the analyzer and pump.
2. Remove the chassis top and locate the filter between the vacuum manifold and the
exhaust port fitting.
3. Disconnect the clear tubing from the filter body and change the filter with the arrow
pointing against the gas flow.
4. To remove the hose clamps, slide the two clamp ends in opposite directions with a
needle-nose pliers until the clamp comes apart.
5. Reconnect the tubing by using the same or new clamps and pushing tightening
them until a good seal is achieved.
6. Restart the pump and clear any error warnings from the front panel display.
7. After about 5 minutes, check the RCEL pressure reading and ensure that it is
similar to its value before changing the filter (but less than 10 in-Hg-A).
Teledyne Analytical Instruments
248
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11.3.6. CHANGING THE INTERNAL SPAN GAS GENERATOR PERMEATION
TUBE
1. Turn off the analyzer, unplug the power cord and remove the cover.
2. Locate the permeation tube (see Figure 3-5) oven in the rear left of the analyzer.
3. Remove the top layer of insulation if necessary.
4. Unscrew the black aluminum cover of the oven (3 screws) using a medium Phillips-
head screw driver.
Leave the fittings and tubing connected to the cover.
5. Remove the old permeation tube and replace it with the new tube.
Ensure that the tube is placed into the larger of two holes and that the open
permeation end of the tube (plastic) is facing up.
6. Re-attach the cover with three screws.
Ensure that the three screws are tightened evenly.
7. Replace the analyzer cover, plug the power cord back in and turn on the analyzer.
8. Carry out a span check to see if the new permeation device works properly (see
Section 9.3.4).
9. The permeation rate may need several days to stabilize.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Do not leave instrument turned off for more than 8 hours without
removing the permeation tube. Do not ship the instrument without
removing the permeation tube. The tube continues to emit NO2, even at
room temperature and will contaminate the entire instrument.
11.3.7. CHANGING THE EXTERNAL ZERO AIR SCRUBBER (OPT 86C)
The external zero air scrubber that is included with several of the 9110T’s optional calibration valve
packages contains two chemicals:
Pink Purafil© (P/N CH 9)that converts NO in the ambient air to NO2, and;
Black, charcoal (P/N CH 1) that absorbs the NO2 thereby creating zero air.
These chemicals need to be replaced periodically (see Table 11-1) or as needed.
Teledyne Analytical Instruments
249
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
CAUTION!
The following procedures apply only to the External Zero Air Scrubber and NOT
to the inline exhaust scrubber cartridge that is part of the pump pack assembly.
IMPORTANT
IMPACT ON READINGS OR DATA
This procedure can be carried out while the instrument is running,
however ensure that the analyzer is not in ZERO calibration mode.
1. Locate the scrubber on the outside rear panel; Figure 11-4 shows the exploded
assembly.
2. Remove the old scrubber by disconnecting the 1/4” plastic tubing from the DFU
particle filter using 9/16” and 1/2" wrenches.
3. Remove the DFU particle filter from the cartridge using 9/16” wrenches.
4. Unscrew the top of the scrubber canister and discard the Purafil© and charcoal
contents.
Ensure to abide to local laws about discarding these chemicals.
The rebuild kit (listed in Appendix B) comes with a Material and Safety Data
Sheet, which contains more information on these chemicals.
5. It is not necessary to remove the insert from the barrel, but if removed, perform the
following procedure:
Coat the threads of the insert with epoxy (TAI P/N CH32).
Hand tighten insert to barrel.
6. It is not necessary to remove the nylon tube fitting from the insert, but if removed,
apply Teflon tape (TAI P/N HW36) to the threads of the nylon tube fitting before
installing on the insert.
7. Refill the scrubber with charcoal at the bottom and the Purafil© chemical at the top.
Use three, white retainer pads to separate the chemicals as shown Figure 11-4
8. Replace the screw-top cap and tighten the cap; hand-tighten only.
9. If necessary, replace the filter with a new unit and discard the old. See Section
11.3.7.1.
The bottom retainer pad should catch most of the dust, the filter should not be
visibly dirty (on the inside).
10. Replace the scrubber assembly into its clips on the rear panel.
11. Reconnect the plastic tubing to the fitting of the DFU particle filter.
12. Adjust the scrubber cartridge such that it does not protrude above or below the
analyzer in case the instrument is mounted in a rack.
Teledyne Analytical Instruments
250
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
If necessary, squeeze the clips for a tighter grip on the cartridge.
Figure 11-4:
Zero Air Scrubber Assembly
11.3.7.1. Changing the External Scrubber’s DFU Filter
There is also a DFU filter on the inlet of the external zero air scrubber that is included in several of the
optional calibration valve packages.
To change this filter:
1. Disconnect the tube and fitting from one end and remove the filter from the scrubber
canister.
2. Insert a new filter and reattach the tubing.
3. Ensure that the small arrow embedded on the filter points in flow direction, i.e., to
analyzer.
Teledyne Analytical Instruments
251
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
11.3.8. CHANGING THE NO2 CONVERTER
The NO2 converter is located in the center of the instrument, Figure 3-5 for the location, and Figure 11-5
for the assembly.
The converter is designed for replacement of the cartridge only; the heater with built-in thermocouple is
to be reused.
CAUTION!
Wear gloves prior to changing the NO2 Converter to ensure that the fiberglass
insulation does not come into contact with your skin.
1. Turn off the analyzer power.
2. Remove the instrument cover and allow the converter to cool.
3. Remove the converter assembly cover as well as the Moly insulation (top layer and
corner cut out layers) until the Moly converter assembly can be seen.
CAUTION
HOT SURFACE HAZARD
The converter operates at 315º C. Severe burns can result if the assembly is not
allowed to cool.
Do not handle the assembly until it is at room temperature. This may take several
hours
4. Remove the tube fittings from the Moly converter assembly.
5. Disconnect the power and the thermocouple from the Moly converter assembly.
6. Unscrew the steel cable clamp (for the power leads) from the converter housing
with a Phillips-head screw driver.
7. Remove the Moly converter assembly (converter cartridge and band heater) from
the converter housing.
Make a note of the orientation of the tubes relative to the heater cartridge.
8. Unscrew the band heater and loosen it.
9. Remove the old converter cartridge.
Teledyne Analytical Instruments
252
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
Figure 11-5:
NO2 Converter Assembly
10. Wrap the band heater around the new replacement converter cartridge and tighten
the screws using a high-temperature anti-seize agent (TAI P/N CH42) such as
copper paste.
Ensure to use proper alignment of the heater with respect to the converter
tubes.
11. Replace the Moly converter assembly by routing the cables through the holes in the
converter housing and reconnecting them properly.
12. Reconnect the steel cable clamp around the power leads for safe operation.
13. Reattach the tube fittings to the converter and replace the Moly insulation (top layer
and corner cut out layers).
14. Reinstall the converter assembly cover.
Teledyne Analytical Instruments
253
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
15. Reinstall the instrument cover and power up the analyzer.
16. Allow the converter to burn-in for 24 hours, and then recalibrate the instrument.
11.3.9. CLEANING THE REACTION CELL
A dirty reaction cell will cause excessive noise, drifting zero or span values, low response or a
combination of all.
To clean the reaction cell, it is necessary to remove it from the sensor housing.
1. Turn off the instrument power and vacuum pump. Refer to Figure 11-6 for the
following procedure.
2. Disconnect the black 1/4" exhaust tube and the 1/8” sample and ozone air tubes
from the reaction cell. Disconnect the heater/thermistor cable.
3. Remove two screws (TAI P/N SN144) and two washers holding the reaction cell to
the PMT housing and lift the cell and manifold out.
Teledyne Analytical Instruments
254
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
Figure 11-6:
Reaction Cell Assembly
4. Remove two screws (Teledyne P/N SN150) and two washers.
5. The reaction cell will separate into two halves, the stainless steel manifold assembly
and the black plastic reaction cell with window gasket, stainless steel reaction cell
sleeve, optical filter and O-rings.
6. The reaction cell (both plastic part and stainless steel sleeve) and optical filter
should be cleaned with Distilled Water (DI - Water) and a clean tissue, and dried
thereafter.
7. Usually it is not necessary to clean the sample and ozone flow orifices since they
are protected by sintered filters.
Teledyne Analytical Instruments
255
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
If tests show that cleaning is necessary, refer to Section 11.3.10 on how to
clean the critical flow orifice.
8. Do not remove the sample and ozone nozzles. They are Teflon threaded and
require a special tool for reassembly. If necessary, the manifold with nozzles
attached can be cleaned in an ultrasonic bath.
9. Reassemble in proper order and re-attach the reaction cell to the sensor housing.
Reconnect pneumatics and heater connections, then re-attach the pneumatic
sensor assembly and the cleaning procedure is complete.
10. After cleaning the reaction cell, it is also recommended to exchange the ozone
supply air filter chemical as described in Section 11.3.3.
11. After cleaning, the analyzer span response may drop 10 - 15% in the first 10 days
as the reaction cell window conditions. This is normal and does not require another
cleaning.
11.3.10. REPLACING CRITICAL FLOW ORIFICES
There are several critical flow orifices installed in the 9110T (see Figure 13-9 for a pneumatic location of
each orifice). Despite the fact that these flow restrictors are protected by sintered stainless steel filters,
they can, on occasion, clog up, particularly if the instrument is operated without sample filter or in an
environment with very fine, sub-micron particle-size dust.
Figure 11-7:
Critical Flow Orifice Assembly
To clean or replace a critical flow orifice:
Teledyne Analytical Instruments
256
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
1. Turn off power to the instrument and vacuum pump.
2. Remove the analyzer cover and locate the reaction cell (Figure 11-5 and Figure
11-6).
3. Unscrew the 1/8” sample and ozone air tubes from the reaction cell.
4. For orifices on the reaction cell (Figure 11-6): Unscrew the orifice holder with a
9/16” wrench.
This part holds all components of the critical flow assembly as shown in Figure
11-7.
Appendix B contains a list of spare part numbers.
5. For orifices in the vacuum manifold: the assembly is similar to the one shown in
Figure 11-7, except:
Without the orifice holder, P/N 04090, and bottom O-ring, P/N OR34 and;
With an NPT fitting in place of the FT 10 fitting.
6. After taking off the connecting tube, unscrew the NPT fitting.
7. Take out the components of the assembly:
spring
sintered filter
two O-rings
the orifice
Note
For the vacuum manifold only, you may need to use a scribe or pressure
from the vacuum port to get the parts out of the manifold.
8. Discard the two O-rings and the sintered filter and install new ones.
9. Reassemble the parts as shown in Figure 11-7.
10. Reinstall the critical flow orifice assembly into the reaction cell manifold or the
vacuum manifold.
11. Reconnect all tubing, power up the analyzer and pump. After a warm-up period of
30 minutes, carry out a leak test as described in Section 13.3.12.
11.3.11. CHECKING FOR LIGHT LEAKS
When re-assembled or operated improperly, the 9110T can develop small gaps around the PMT, which
let stray light from the analyzer surrounding into the PMT housing. To find such light leaks, follow the
procedures below.
Teledyne Analytical Instruments
257
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
CAUTION – QUALIFIED PERSONNEL ONLY
This procedure is carried out with the analyzer running and its cover removed.
1. Scroll the front panel display to show then test function to PMT.
2. Supply zero gas to the analyzer.
3. With the instrument still running, carefully remove the analyzer cover.
WARNING – ELECTRICAL SHOCK HAZARD
Do NOT touch any of the inside wiring with the metal cover or with your body.
Do NOT drop screws or tools into a running analyzer.
4. Shine a powerful flashlight or portable incandescent light at the inlet and outlet
fitting and at all of the joints of the reaction cell as well as around the PMT housing.
The PMT value should not respond to the light, the PMT signal should remain
steady within its usual noise floor.
5. If there is a PMT response to the external light, symmetrically tighten the reaction
cell mounting screws or replace the 1/4” vacuum tubing with new, black PTFE
tubing (this tubing will fade with time and become transparent).
Note
Often, light leaks are also caused by O-rings being left out of the
assembly.
6. If, during this procedure, the black PMT housing end plate for the Sensor Assembly
is removed, ensure to replace the 5 desiccant bags inside the housing.
7. Carefully replace the analyzer cover. If tubing was changed, carry out a pneumatic
leak check.
11.3.12. CHECKING FOR PNEUMATIC LEAKS
CAUTION - TECHNICAL INFORMATION
Do not exceed 15 psi when pressurizing the system during either Simple or
Detailed checks.
11.3.12.1. Simple Vacuum Leak and Pump Check
Leaks are the most common cause of analyzer malfunction. This section presents a simple leak check,
whereas the next section details a more thorough procedure. The method described here is easy, fast and
detects, but does not locate, most leaks. It also verifies the sample pump condition.
1. Turn the analyzer ON, and allow at least 30 minutes for flows to stabilize.
2. Cap the sample inlet port (cap must be wrench-tight).
3. After several minutes, when the pressures have stabilized, note the SAMP (sample
pressure) and the RCEL (vacuum pressure) readings.
Teledyne Analytical Instruments
258
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
If both readings are equal to within 10% and less than 10 in-Hg-A, the
instrument is free of large leaks.
It is still possible that the instrument has minor leaks.
If both readings are < 10 in-Hg-A, the pump is in good condition.
A new pump will create a pressure reading of about 4 in-Hg-A (at sea level).
11.3.12.2. Detailed Pressure Leak Check
If a leak cannot be located by the above procedure, obtain a leak checker similar to Teledyne P/N 01960,
which contains a small pump, shut-off valve, and pressure gauge to create both over-pressure and
vacuum. Alternatively, a tank of pressurized gas, with the two-stage regulator adjusted to ≤ 15 psi, a
shutoff valve and a pressure gauge may be used.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Once tube fittings have been wetted with soap solution under a
pressurized system, do not apply or reapply vacuum as this will cause
soap solution to be sucked into the instrument, contaminating inside
surfaces.
1. Turn OFF power to the instrument and remove the instrument cover.
2. Install a leak checker or a tank of gas (compressed, oil-free air or nitrogen) as
described above on the sample inlet at the rear panel.
3. Disconnect the pump tubing on the outside rear panel and cap the pump port.
If IZS or zero/span valves are installed, disconnect the tubing from the zero and
span gas ports and plug them (Figure 3-3).
Cap the DFU particle filter on the Perma Pure dryer.
4. Pressurize the instrument with the leak checker or tank gas, allowing enough time
to fully pressurize the instrument through the critical flow orifice.
Check each tube connection (fittings, hose clamps) with soap bubble solution,
looking for fine bubbles.
Once the fittings have been wetted with soap solution, do not reapply vacuum
as it will draw soap solution into the instrument and contaminate it.
Do not exceed 15 psi pressure.
5. If the instrument has the zero and span valve option, the normally closed ports on
each valve should also be separately checked.
Connect the leak checker to the normally closed ports and check with soap
bubble solution.
Teledyne Analytical Instruments
259
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Instrument Maintenance
6. If the analyzer is equipped with an IZS Option Connect the leak checker to the Dry
Air inlet and check with soap bubble solution.
7. Once the leak has been located and repaired, the leak-down rate of the indicated
pressure should be less than 1 in-Hg-A (0.4 psi) in 5 minutes after the pressure is
turned off.
8. Clean surfaces from soap solution, reconnect the sample and pump lines and
replace the instrument cover.
9. Restart the analyzer.
11.3.12.3. Performing a Sample Flow Check
IMPORTANT
IMPACT ON READINGS OR DATA
Use a separate, calibrated flow meter capable of measuring flows between
0 and 1000 cm³/min to measure the gas flow rate though the analyzer. Do
not use the built in flow measurement viewable from the front panel of the
instrument.
This value is only calculated, not measured.
Sample flow checks are useful for monitoring the actual flow of the instrument, as the front panel
display shows only a calculated value. A decreasing, actual sample flow may point to slowly clogging
pneumatic paths, most likely critical flow orifices or sintered filters. To perform a sample flow check:
1. Disconnect the sample inlet tubing from the rear panel SAMPLE port.
2. Attach the outlet port of a flow meter to the sample inlet port on the rear panel.
Ensure that the inlet to the flow meter is at atmospheric pressure.
3. The sample flow measured with the external flow meter should be 500 cm³/min ±
10%.
If a combined sample/ozone air Perma Pure dryer is installed (optional
equipment), the flow will be 640 cm³/min ± 10% (500 cm³/min for the sample
and 80 cm³/min for the ozone generator supply air and 60 cm³/min for the purge
flow).
4. Low flows indicate blockage somewhere in the pneumatic pathway.
Teledyne Analytical Instruments
260
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12. TROUBLESHOOTING & SERVICE
This section contains a variety of methods for identifying the source of performance problems with the
analyzer. Also included in this section are procedures that are used in repairing the instrument.
Note:
To support your understanding of the technical details of maintenance,
Section 13, Principles of Operation, provides information about how the
instrument works.
CAUTION
The operations outlined in this section must be performed by qualified
maintenance personnel only.
WARNING
RISK OF ELECTRICAL SHOCK
Some operations need to be carried out with the analyzer open and
running.
Exercise caution to avoid electrical shocks and electrostatic or
mechanical damage to the analyzer.
Do not drop tools into the analyzer or leave those after your procedures.
Do not short or touch electric connections with metallic tools while
operating inside the analyzer.
Use common sense when operating inside a running analyzer.
The front panel of the analyzer is hinged at the bottom and may be
opened to gain access to various components mounted on the panel
itself or located near the front of the instrument (such as the particulate
filter).
Note
Remove the locking screw located at the right-hand side of the front
panel.
12.1. GENERAL TROUBLESHOOTING
The 9110T Nitrogen Oxide analyzer has been designed so that problems can be rapidly detected,
evaluated and repaired. During operation, it continuously performs diagnostic tests and provides the
ability to evaluate its key operating parameters without disturbing monitoring operations.
A systematic approach to troubleshooting will generally consist of the following five steps:
Teledyne Analytical Instruments
261
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
1. Note any WARNING MESSAGES and take corrective action as necessary.
2. Examine the values of all TEST functions and compare them to factory values.
Note any major deviations from the factory values and take corrective action.
3. Use the internal electronic status LEDs to determine whether the electronic
communication channels are operating properly.
Verify that the DC power supplies are operating properly by checking the
voltage test points on the relay PCA.
Note that the analyzer’s DC power wiring is color-coded and these colors match
the color of the corresponding test points on the relay PCA.
4. Suspect a leak first!
Customer service data indicate that the majority of all problems are eventually
traced to leaks in the internal pneumatics of the analyzer or the diluent gas and
source gases delivery systems.
Check for gas flow problems such as clogged or blocked internal/external gas
lines, damaged seals, punctured gas lines, a damaged / malfunctioning pumps,
etc.
5. Follow the procedures defined in Section 3.4.3 to confirm that the analyzer’s vital
functions are working (power supplies, CPU, relay PCA, touchscreen, PMT cooler,
etc.).
See Figure 3-5 or the general layout of components and sub-assemblies in the
analyzer.
See the wiring interconnect diagram and interconnect list in Appendix D.
12.1.1. FAULT DIAGNOSIS WITH WARNING MESSAGES
The most common and/or serious instrument failures will result in a warning message being displayed on
the front panel. Table 12-1 lists warning messages, along with their meaning and recommended
corrective action.
It should be noted that if more than two or three warning messages occur at the same time, it is often an
indication that some fundamental sub-system (power supply, relay PCA, motherboard) has failed rather
than an indication of the specific failures referenced by the warnings.
The analyzer will alert the user that a Warning Message is active by flashing the FAULT LED and
displaying the Warning message in the Param field along with the CLR button (press to clear Warning
message). The MSG button displays if there is more than one warning in queue or if you are in the TEST
menu and have not yet cleared the message.
The following display/touch screen examples provide an illustration of each:
Teledyne Analytical Instruments
262
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
The analyzer will also alert the user via the Serial I/O COMM port(s).
To view or clear the various warning messages press:
If a warning message persists after
several attempts to clear it, the message
may indicate a real problem and not an
artifact of the warm-up period.
Suppresses the
warning messages.
CAL
CAL
CAL
MSG CLR SETUP
CLR SETUP
returns the active
warnings to the message
field.
TEST
TEST
Press
to clear the current
message.
MSG
SETUP
SETUP
SETUP
If more than one warning is
active, the next message will take
its place.
Once the last warning has
been cleared, the analyzer’s
display will return to its
standard Sample Mode
configuration.
The display will continually
cycle between showing the
current NOX, NO and NO2
concentrations.
CAL
MSG
Teledyne Analytical Instruments
263
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Table 12-1: Front Panel Warning Messages
WARNING
FAULT CONDITION POSSIBLE CAUSES
AZERO valve not working
Valve control driver failed
Bad Relay PCA
Failed +12 VDC power supply
Gas leak across AZERO Valve ports
Dirty Reaction Cell
Auto-zero reading above 200 mV.
AZERO WARN
XXX.X MV
Value shown in message indicates
auto-zero reading at time warning
was displayed.
O3 flow problem to RCELL
Box Temperature typically runs ~7C warmer than ambient
temperature
Poor/blocked ventilation to the analyzer
Stopped Exhaust-Fan
Box Temp is < 7C
BOX TEMP WARNING
or > 48C.
Ambient Temperature outside of specified range
Measured concentration value is too high or low
Concentration Slope value to high or too low
Dynamic Span operation failed.
Dynamic Zero operation failed.
CANNOT DYN SPAN
CANNOT DYN ZERO
CONFIG INITIALIZED
Measured concentration value is too high
Concentration Offset value to high
Failed Disk on Module
User erased data
Configuration and Calibration data
reset to original Factory state.
Heater configured for wrong voltage type
Failed converter Temperature Sensor
Relay controlling the Heater is not working
Failed Relay Board
NO2 NO Converter temperature <
CONV TEMP WARNING
DATA INITIALIZED
305C or > 325C.
Failed Disk-on-Module
User cleared data.
Data Storage in DAS was erased.
No +15 VDC power supply to Preamplifier PCA
Drive voltage not adjusted properly
Failed PMT Preamplifier PCA
Dirty reaction cell
High voltage power supply output
outside of warning limits specified by
HVPS_SET variable.
HVPS WARNING
Bad pneumatic flow
Heater configured for wrong voltage type
Failed permeation tube Temperature Sensor
Relay controlling the Heater is not working
Failed Relay Board
Permeation tube oven temperature is
IZS TEMP WARNING
< 45C or > 55C.
Failed Sample Pump
Blocked O3 dryer
Blocked inlet/outlet to O3 purifier
Dirty O3 dryer DFU
OZONE FLOW
WARNING
O3 flow rate is < 50 cc/min or >
150 cc/min.
Leak downstream of RCELL
Failed O3 Flow Sensor
Ozone generator is off. This is the
only warning message that
automatically clears itself. It clears
itself when the ozone generator is
turned on.
O3 generator override is turned ON.
Electrical connection between motherboard and generator is
faulty.
OZONE GEN OFF
Bad +15VDC power supply
PMT fan not operating
Failed PMT Temperature Sensor
TEC not functioning
Sample temperature is < 5C or >
PMT TEMP WARNING
12C.
Failed PMT Preamp PCA
Teledyne Analytical Instruments
264
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
POSSIBLE CAUSES
WARNING
FAULT CONDITION
If Sample Pressure is < 15 in-HG:
Sample Pressure is <15 in-Hg or >
35 in-Hg
Blocked Particulate Filter
Blocked Sample Inlet/Gas Line
Failed Pressure Senor/circuitry
Normally 29.92 in-Hg at sea level
decreasing at 1 in-Hg per 1000 ft of
altitude (with no flow – pump
disconnected).
RCELL PRESS WARN
If Sample Pressure is > 35 in-HG:
Bad Pressure Sensor/circuitry
Pressure too high at Sample Inlet.
Heater configured for wrong voltage type
Failed RCELL Temperature Sensor
Relay controlling the heater is not working
Failed Relay Board
RCELL temperature is < 45C or >
RCELL TEMP WARNING
55C.
I2C Bus
This WARNING only appears on Serial I/O COMM Port(s) Front
Panel Display will be frozen, blank or will not respond.
Failure of Motherboard
Motherboard not detected on
power up.
REAR BOARD NOT DET
RELAY BOARD WARN
I2C Bus failure
The CPU cannot communicate with
the Relay Board.
Failed Relay Board
Loose connectors/wiring
Failed Sample Pump
Blocked Sample Inlet/Gas Line
Dirty Particulate Filter
Leak downstream of RCELL Critical Flow Orifice
Failed Sample Pressure Sensor
Failed Vacuum Pressure Sensor
Sample flow rate is < 350 cc/min or >
600 cc/min.
SAMPLE FLOW WARN
SYSTEM RESET
This message occurs at power on.
If it is confirmed that power has not been interrupted:
Failed +5 VDC power
The computer has rebooted.
Fatal Error caused software to restart
Loose connector/wiring
Note
A failure of the analyzer’s CPU, motherboard or power supplies can result
in any or ALL of the above messages.
Teledyne Analytical Instruments
265
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS
In addition to being useful as predictive diagnostic tools, the test functions viewable from the analyzers
front panel can be used to isolate and identify many operational problems when combined with a
thorough understanding of the analyzer’s principles of operation (see Section 13).
The acceptable ranges for these test functions are listed in the “Nominal Range” column of the analyzer
Final Test and Validation Data Sheet (P/N 04490) shipped with the instrument. Values outside these
acceptable ranges indicate a failure of one or more of the analyzer’s subsystems. Functions whose
values are still within acceptable ranges but have significantly changed from the measurement recorded
on the factory data sheet may also indicate a failure.
A worksheet has been provided in Appendix C to assist in recording the value of these test functions.
Note
Note
A value of “XXXX” displayed for any of these TEST functions indicates an
OUT OF RANGE reading.
Sample Pressure measurements are represented in terms of ABSOLUTE
pressure because this is the least ambiguous method reporting gas
pressure.
Absolute atmospheric pressure is about 29.92 in-Hg-A at sea level. It
decreases about 1 in-Hg per 1000 ft gain in altitude. A variety of factors
such as air conditioning systems, passing storms, and air temperature,
can also cause changes in the absolute atmospheric pressure.
Table 12-2: Test Functions - Indicated Failures
TEST FUNCTION
INDICATED FAILURE(S)
NOX STB
SAMP FlW
OZONE FL
PMT
Unstable concentrations; leaks
Leaks; clogged critical flow orifice
Leaks; clogged critical flow orifice
Calibration off; HVPS problem; no flow (leaks)
NORM PMT
AZERO
Auto Zero too high
Leaks; malfunctioning NO, NOx or Auto Zero valve; O3 air filter cartridge exhausted
HVPS
Calibration off; preamp board circuit problems
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP (option)
Malfunctioning heater; relay board communication (I2C bus); relay burnt out
Environment out of temperature operating range; broken thermistor
TEC cooling circuit broken; relay board communication (I2C bus); 12 V power supply
Malfunctioning heater; relay board communication (I2C bus); relay burnt out
Malfunctioning heater; disconnected or broken thermocouple; relay board communication
(I2C bus); relay burnt out; incorrect AC voltage configuration
MOLY TEMP
RCEL (pressure)
SAMP (pressure)
NOX SLOPE
Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices
Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices; sample inlet
overpressure
HVPS out of range; low-level (hardware) calibration needs adjustment; span gas
Teledyne Analytical Instruments
266
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
TEST FUNCTION
Troubleshooting & Service
INDICATED FAILURE(S)
concentration incorrect; leaks
NOX OFFset
NO SLOPE
NO OFFSet
TIME
Incorrect span gas concentration; low-level calibration off
HVPS out of range; low-level calibration off; span gas concentration incorrect; leaks
Incorrect span gas concentration; low-level calibration off
Internal clock drifting; move across time zones; daylight savings time?
12.1.3. DIAG SIGNAL I/O: USING THE DIAGNOSTIC SIGNAL I/O
FUNCTION
The signal I/O diagnostic mode allows access to the digital and analog I/O in the analyzer. Some of the
digital signals can be controlled through the touchscreen. These signals, combined with a thorough
understanding of the instrument’s principles of operation (Section 13), are useful for troubleshooting in
three ways:
The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs and
outputs.
Many of the components and functions that are normally under algorithmic control of the CPU can
be manually exercised.
The technician can directly control the signal level Analog and Digital Output signals.
This allows the technician to observe systematically the effect of directly controlling these signals on the
operation of the analyzer. Following is an example of how to use the Signal I/O menu to view the raw
voltage of an input signal or to control the state of an output voltage or control signal.
Teledyne Analytical Instruments
267
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Figure 12-1:
Example of Signal I/O Function
Note
Any I/O signals changed while in the signal I/O menu will remain in effect
ONLY until signal I/O menu is exited. The Analyzer regains control of
these signals upon exit.
See Appendix A for a complete list of the parameters available for review
under this menu.
Teledyne Analytical Instruments
268
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.2. USING THE ANALOG OUTPUT TEST CHANNEL
The signals available for output over the 9110T’s analog output channel can also be used as diagnostic
tools. See Section 5.9.2 for instruction on activating the analog output and selecting a function.
Table 12-3: Test Channel Outputs as Diagnostic Tools
TEST
CHANNEL
FULL
SCALE
CAUSES OF EXTREMELY
HIGH / LOW READINGS
DESCRIPTION
ZERO
0 mV
Failed PMT
PMT Temperature too High/Low
Bad PMT Preamp PCA
Failed HVPS
Misadjusted HVPS drive Voltage
Light Leak in reaction cell
The output of the PMT
detector converted to a 0
to 5 VDC scale.
PMT
DETECTOR
5000 mV
The flow rate of O3
through the analyzer as
measured by the O3 flow
sensor
OZONE
FLOW
0
1000
Check for Gas Flow problems in the O3 gas lines.
Check for Gas Flow problems in the sample gas lines.
Check for Gas Flow problems in the sample gas lines.
cm3/min
cm3/min
The calculated flow rate
for sample gas through
the analyzer.
SAMPLE
FLOW
0
1000
cm3/min
cm3/min
The pressure of the
sample gas measured
upstream of the Auto
Zero Valve
SAMPLE
PRESSURE
0 In-Hg-A
40 In-Hg-A
The pressure of gas
inside the reaction cell of
the sensor module
RCELL
PRESSURE
Check for Gas Flow problems in all gas lines.
0 In-Hg-A
40 In-Hg-A
The temperature of gas
inside the reaction cell of
the sensor module
Same as RCELL TEMP WARNING in Table 12-1.
RCELL TEMP
IZS TEMP
0 C
70 C
The temperature of the
permeation tube oven of
the optional internal span
gas generator.
Same as IZS TEMP WARNING in Table 12-1.
0 C
70 C
The temperature NO2
Same as CONV TEMP WARNING in Table 12-1.
Same as PMT TEMP WARNING in Table 12-1.
Same as BOX TEMP WARNING in Table 12-1.
CONV TEMP
PMT TEMP
BOX TEMP
0 mV
0 C
0 C
5000 mV
50 C
NO converter
The temperature inside
PMT
The temperature inside
the 9110T’s chassis
70 C
Represents the output
voltage of the PMT's high
voltage power supply
HVPS
VOLTAGE
Same as HVPSWARNING in Table 12-1.
0 mV
5000 mV
Teledyne Analytical Instruments
269
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.3. USING THE INTERNAL ELECTRONIC STATUS LEDS
Several LEDs are located inside the instrument to assist in determining if the analyzer’s CPU, I2C bus
and Relay PCA are functioning properly.
12.3.1. CPU STATUS INDICATOR
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU board,
flashes when the CPU is running the main program loop. After power-up, approximately 30 – 60
seconds, DS5 should flash on and off. If characters are written to the front panel display but DS5 does
not flash then the program files have become corrupted, contact TAI's Customer Service Department
(see Section 12.10) because it may be possible to recover operation of the analyzer. If after 30 – 60
seconds, neither DS5 is flashing nor have any characters been written to the front panel display then the
CPU is bad and must be replaced.
Motherboard
CPU Status LED
Figure 12-2:
CPU Status Indicator
12.3.2. RELAY PCA STATUS LEDS
There are sixteen LEDs located on the Relay PCA. Some are not used on this model.
12.3.2.1. I2C Bus Watchdog Status LEDs
The most important is D1, which indicates the health of the I2C bus.
Table 12-4: Relay PCA Watchdog LED Failure Indications
LED
Function
Fault Status
Indicated Failure(s)
Failed/Halted CPU
Faulty Motherboard, Touchscreen or Relay PCA
Faulty Connectors/Wiring between Motherboard,
Touchscreen or Relay PCA
Continuously ON
or
Continuously OFF
D1
(Red)
I2C bus Health
(Watchdog Circuit)
Failed/Faulty +5 VDC Power Supply (PS1)
If D1 is blinking, then the other LEDs can be used in conjunction with DIAG Menu Signal I/O to
identify hardware failures of the relays and switches on the Relay PCA.
Teledyne Analytical Instruments
270
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.3.2.2. Relay PCA Status LEDs
D10 (Green) – NO/NOx Valve
D9 (Green) – AutoZero Valve
D8 (Green) – Optional Sample/Cal Valve
D7 (Green) – Optional Zero/Span Valve
D3 (Yellow) NO2 NO Converter Heater
D2 (Yellow) Reaction Cell Heater
D5 (Yellow) – Optional Internal Span Gas Gen Heater
D11 (Green) – Optional Dual Span Select Valve
D12 (Green) – Optional Pressurized Span Shutoff Valve
D13 (Green) – Optional Pressurized Zero Shutoff Valve
D1 (RED)
Watchdog Indicator
Figure 12-3:
Relay PCA Status LEDS Used for Troubleshooting
Table 12-5: Relay PCA Status LED Failure Indications
FAULT
STATUS
LED
LED ROW 1
D2
COLOR
FUNCTION
INDICATED FAILURE(S)
Continuously
ON or OFF
Heater broken, thermistor broken
Yellow
Yellow
Green
Green
Green
Green
Reaction Cell heater
NO2 converter heater
Zero/Span valve status
Sample/Cal valve status
Auto-zero valve status
NO/NOx valve status
Continuously
ON or OFF
Heater broken, thermocouple broken
D3
D7
D8
D9
Continuously
ON or OFF
Valve broken or stuck, valve driver chip broken
Valve broken or stuck, valve driver chip broken
Valve broken or stuck, valve driver chip broken
Valve broken or stuck, valve driver chip broken
Continuously
ON or OFF
Continuously
ON or OFF
Continuously
ON or OFF
D10
LED ROW 2
D5
Internal span gas generator
perm tube heater
Continuously
ON or OFF
Heater broken, thermistor broken
Yellow
Green
Green
Continuously
ON or OFF
Valve broken or stuck, valve driver chip broken
Valve broken or stuck, valve driver chip broken
D11
D12
Dual span select valve
Continuously
ON or OFF
Pressurized Span shutoff valve
Continuously
ON or OFF
Valve broken or stuck, valve driver chip broken
N/A
D13
Green
Green
Pressurized Zero shutoff valve
Spare
N/A
D14- 16
Teledyne Analytical Instruments
271
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.4. GAS FLOW PROBLEMS
The 9110T has two main flow paths, the sample flow and the flow of the ozone supply air. With IZS or
zero/span valve option installed, there is a third (zero air) and a fourth (span gas) flow path, but either
one of those is only controlled by critical flow orifices and not displayed on the front panel or stored to
the DAS.
Flow is too high
Flow is greater than zero, but is too low, and/or unstable
Flow is zero (no flow)
When troubleshooting flow problems, it is essential to confirm the actual flow rate without relying on the
analyzer’s flow display. The use of an independent, external flow meter to perform a flow check as
described in Section 11.3.12.3 is essential. Refer to the pneumatic flow diagrams as needed for
reference.
12.4.1. ZERO OR LOW FLOW PROBLEMS
12.4.1.1. Sample Flow is Zero or Low
The 9110T does not actually measure the sample flow but rather calculates it from a differential pressure
between sample and vacuum manifold. On flow failure, the unit will display a SAMPLE FLOW
WARNING on the front panel display and the respective test function reports XXXX instead of a value
“0”. This message applies to both a flow rate of zero as well as a flow that is outside the standard range
(350-600 cm³/min).
If the analyzer displays XXXX for the sample flow, confirm that the external sample pump is operating
and configured for the proper AC voltage.
Whereas the 9110T can be internally configured for two different power regimes (100-120 V and
220-240 V, either 50 or 60 Hz), the external pump is physically different for each of three power
regimes (100 V / 50 Hz, 115 V / 60 Hz and 230 V / 50 Hz).
If the pump is not running, use an AC Voltmeter to ensure that the pump is supplied with the proper
AC power. If AC power is supplied properly, but the pump is not running, replace the pump.
Note
Sample and vacuum pressures mentioned in this chapter refer to
operation of the analyzer at sea level. Pressure values need to be
adjusted for elevated locations, as the ambient pressure decreases by
about 1 in-Hg per 300 m / 1000 ft.
If the pump is operating but the unit reports a XXXX gas flow, take the following three steps:
1. Check for actual sample flow.
To check the actual sample flow, disconnect the sample tube from the sample
inlet on the rear panel of the instrument.
Ensure that the unit is in basic SAMPLE mode.
Place a finger over the inlet and see if it gets sucked in by the vacuum or, more
properly, use a flow meter to measure the actual flow.
Teledyne Analytical Instruments
272
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
If there is proper flow of around 450-550 cm³/min, contact customer service.
If there is no flow or low flow, continue with the next step.
2. Check pressures.
Check that the sample pressure is at or around 28 in-Hg-A at sea level (adjust
as necessary when in elevated location, the pressure should be about 1” below
ambient atmospheric pressure) and that the RCEL pressure is below 10 in-Hg-
A.
The 9110T will calculate a sample flow up to about 14 in-Hg-A RCEL pressure
but a good pump should always provide less than 10 in.
If both pressures are the same and around atmospheric pressure, the pump
does not operate properly or is not connected properly. The instrument does
not get any vacuum.
If both pressures are about the same and low (probably under 10 in-Hg-A, or
~20” on sample and 15” on vacuum), there is a cross-leak between sample flow
path and vacuum, most likely through the Perma Pure dryer flow paths. See
troubleshooting the Perma Pure dryer later in this chapter.
If the sample and vacuum pressures are around their nominal values (28 and
<10 in-Hg-A, respectively) and the flow still displays XXXX, carry out a leak
check as described in Section 13.3.12.
If gas flows through the instrument during the above tests but goes to zero or is
low when it is connected to zero air or span gas, the flow problem is not internal
to the analyzer but likely caused by the gas source such as
calibrators/generators, empty gas tanks, clogged valves, regulators and gas
lines.
If an IZS or Zero/Span valve option is installed in the instrument, press CALZ
and CALS. If the sample flow increases, suspect a bad Sample/Cal valve.
3. If none of these suggestions help, carry out a detailed leak check of the
analyzer as described in Section 11.3.12.2.
12.4.1.2. Ozone Flow is Zero or Low
If there is zero or a low (<50 cm³/min) ozone flow, the unit displays an OZONE FLOW WARNING
message on the front panel and a value between 0.0 and 50 cm³/min for the actual ozone flow as
measured by the internal mass flow meter. In this case, carry out the following steps:
1. Check the actual flow rate through the ozone dryer by using an external flow
meter to the inlet port of the dryer.
This inlet port is inside the analyzer at the end of the plastic particle filter
(Section 11.3.2 for illustration).
If there is nominal flow (about 160 cm³/min from 80 cm³/min O3 flow and 80
cm³/min purge flow), consult customer service as there is a problem with the
firmware or electronics.
Teledyne Analytical Instruments
273
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
2. If the actual flow is low or zero, check if the pump operates properly. The
RCEL pressure should be below 10 in-Hg-A at sea level.
If it is above 10”, rebuild the pump (Section 11.3.4.1). Check the spare parts list
in Appendix B on how to order pump rebuild kits.
3. Check if the particle filter is clogged.
Briefly remove the particle filter to see if this improves the flow.
Be very cautious when handling the Perma Pure dryer fittings (see Section
11.3.2 on proper handling instructions).
If the filter is clogged, replace it with a new unit.
If taking off this filter does not solve the problem, continue to the next step.
Do not leave the Perma Pure dryer without filter for more than a few seconds,
as you may draw in dust, which will reduce the performance of the dryer.
4. A leak between the flow meter and the reaction cell (where the flow-
determining critical orifice is located) may cause a low flow (the system
draws in ambient air through a leak after the flow meter).
Check for leaks as described in Section 11.3.12.
Repair the leaking fitting, line or valve and re-check.
5. The most likely cause for zero or low ozone flow is a clogged critical flow
orifice or sintered filter within the orifice assembly.
The orifice that sets the ozone flow is located on the reaction cell.
Check the actual ozone flow by disconnecting the tube from the reaction cell
and measuring the flow going into the cell.
If this flow is correct (~80 cm³/min), the orifice works properly.
If this flow is low, replace the sintered filter.
The orifice holder assembly allows a quick and easy replacement of the filter
(see Section 11.3.5 and on for replacement procedures).
Appendix B lists a spare part kit with a complete orifice assembly that allows a
quick replacement with minimum instrument down-time.
12.4.1.3. High Flow
Flows that are significantly higher than the allowed operating range (typically ±10-11% of the nominal
flow) should not occur in the 9110T unless a pressurized sample, zero or span gas is supplied to the inlet
ports.
Ensure to vent excess pressure and flow just before the analyzer inlet ports.
Teledyne Analytical Instruments
274
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
When supplying sample, zero or span gas at ambient pressure, a high flow would indicate that one or
more of the critical flow orifices are physically broken (very unlikely case), allowing more than nominal
flow, or were replaced with an orifice of wrong specifications.
If the flows are within 15% higher than normal, we recommend measuring and recalibrating the flow
electronically using the procedure in Section 10, followed by a regular review of these flows over
time to see if the new setting is retained properly.
Also, check the flow assembly o-rings and replace as needed.
12.4.1.4. Sample Flow is Zero or Low but Analyzer Reports Correct Flow
Note that the 9110T analyzer can report a correct flow rate even if there is no or a low actual sample
flow through the reaction cell.
The sample flow on the 9110T is only calculated from the sample pressure and critical flow condition
is verified from the difference between sample pressure and vacuum pressure.
If the critical flow orifice assembly is partially or completely clogged, both the sample and vacuum
pressures are still within their nominal ranges (the pump keeps pumping, the sample port is open to
the atmosphere), but there is no flow possible through the reaction cell.
Although measuring the actual flow is the best method, in most cases, this fault can also be diagnosed by
evaluating the two pressure values.
Since there is no longer any flow, the sample pressure should be equal to ambient pressure, which
is about 1 in-Hg-A higher than the sample pressure under normal operation.
The reaction cell pressure, on the other hand, is significantly lower than under normal operation,
because the pump no longer has to remove 500 cm³/min of sample gas and evacuates the reaction
cell much better.
Those two indicators, taken together with a zero or low actual flow, indicate a clogged sample
orifice.
The 9110T features a new orifice holder, which makes switching sample and ozone flow orifices very
easy; refer to Section 11.3.10 on how to change the sample orifices and to Appendix B for part numbers
of these assemblies.
Again, monitoring the pressures and flows regularly will reveal such problems, because the pressures
would slowly or suddenly change from their nominal, mean values. Teledyne recommends to review all
test data once per week and to do an exhaustive data analysis for test and concentration values once per
month, paying particular attention to sudden or gradual changes in all parameters that are supposed to
remain constant, such as the flow rates.
Teledyne Analytical Instruments
275
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.5. CALIBRATION PROBLEMS
This section describes possible causes of calibration problems.
12.5.1. NEGATIVE CONCENTRATIONS
Negative concentration values can be caused for several reasons:
A slight, negative signal is normal when the analyzer is operating under zero gas and the signal is
drifting around the zero calibration point.
This is caused by the analyzer’s zero noise and may cause reported concentrations to be
negative for a few seconds at a time down to -20 ppb, but should randomly alternate with
similarly high, positive values.
The 9110T has a built-in Auto Zero function, which should take care of most of these
deviations from zero, but may yield a small, residual, negative value.
If larger, negative values persist continuously, check if the Auto Zero function was
accidentally turned off using the remote variables in Appendix A-2.
In this case, the sensitivity of the analyzer may be drifting negative.
A corruption of the Auto Zero filter may also cause negative concentrations.
If a short, high noise value was detected during the Auto Zero cycle, that higher reading will
alter the Auto Zero filter value.
As the value of the Auto Zero filter is subtracted from the current PMT response, it will
produce a negative concentration reading.
High Auto Zero readings can be caused by
a leaking or stuck Auto Zero valve (replace the valve),
by an electronic fault in the preamplifier causing it to have a voltage on the
PMT output pin during the Auto Zero cycle (replace the preamplifier),
by a reaction cell contamination causing high background (>40 mV) PMT
readings (clean the reaction cell),
by a broken PMT temperature control circuit, allowing high zero offset
(repair the faulty PMT cooler). After fixing the cause of a high Auto Zero
filter reading, the 9110T will take 15 minutes for the filter to clear itself, or
by an exhausted chemical in the ozone cleanser (see Section 11.3.3).
Calibration error is the most likely explanation for negative concentration values.
If the zero air contained some NO or NO2 gas (contaminated zero air or a worn-out zero air
scrubber) and the analyzer was calibrated to that concentration as “zero”, the analyzer may
report negative values when measuring air that contains little or no NOx.
The same problem occurs, if the analyzer was zero-calibrated using zero gas that is
contaminated with ambient air or span gas (cross-port leaks or leaks in supply tubing or
user not waiting long enough to flush pneumatic systems).
Teledyne Analytical Instruments
276
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
If the response offset test functions for NO (NO OFFS) or NOX (NOX OFFS) are greater than
150 mV, a reaction cell contamination is indicated.
Clean the reaction cell as described in Section 11.3.9.
12.5.2. NO RESPONSE
If the instrument shows no response (display value is near zero) even though sample gas is supplied
properly and the instrument seems to perform correctly.
1. Carry out an electrical test with the ELECTRICAL TEST procedure in the
diagnostics menu, see Section 12.7.12.2.
If this test produces a concentration reading, the analyzer’s electronic signal
path is correct.
2. Carry out an optical test using the OPTIC TEST procedure in the diagnostics
menu, see Section 12.7.12.1.
If this test results in a concentration signal, then the PMT sensor and the
electronic signal path are operating properly.
If the 9110T passes both ETEST and OTEST, the instrument is capable of
detecting light and processing the signal to produce a reading.
Therefore, the problem must be in the pneumatics or the ozone generator.
3. Check if the ozone generator is turned on.
Usually, the analyzer issues a warning whenever the ozone generator is turned
off.
Go to SETUP-MORE-DIAG-ENTR, then scroll to the OZONE GEN OVERRIDE
and see if it shows ON.
If it shows OFF, turn it ON and EXIT the DIAG menu.
If this is done and the ozone flow is correct, the analyzer should be properly
supplied with ozone unless the generator itself is broken.
4. Confirm the lack of response by supplying NO or NO2 span gas of about
80% of the range value to the analyzer.
5. Check the sample flow and ozone flow rates for proper values.
6. Check for disconnected cables to the sensor module.
7. If NO2 signal is zero while NO signal is correct, check the NO/NOx valve
and the NO2 converter for proper operation.
12.5.3. UNSTABLE ZERO AND SPAN
Leaks in the 9110T or in the external gas supply and vacuum systems are the most common source of
unstable and non-repeatable concentration readings.
Teledyne Analytical Instruments
277
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
1. Check for leaks in the pneumatic systems as described in Section 11.3.12.
2. Consider pneumatic components in the gas delivery system outside the
9110T such as a change in zero air source (ambient air leaking into zero air
line or a worn-out zero air scrubber) or a change in the span gas
concentration due to zero air or ambient air leaking into the span gas line.
3. Once the instrument passes a leak check, do a flow check (this chapter) to
ensure that the instrument is supplied with adequate sample and ozone air.
4. Confirm the sample pressure, sample temperature, and sample flow readings
are correct and steady.
5. Verify that the sample filter element is clean and does not need to be
replaced.
12.5.4. INABILITY TO SPAN - NO SPAN BUTTON (CALS)
In general, the 9110T will not display certain buttons whenever the actual value of a parameter is outside
of the expected range for that parameter. If the calibration menu does not show a SPAN button when
carrying out a span calibration, the actual concentration must be outside of the range of the expected
span gas concentration, which can have several reasons.
1. Verify that the expected concentration is set properly to the actual span gas
concentration in the CONC sub-menu.
2. Confirm that the NOx span gas source is accurate.
This can be done by comparing the source with another calibrated analyzer, or
by having the NOx source verified by an independent traceable photometer.
3. Check for leaks in the pneumatic systems as described in Section 11.3.12.
Leaks can dilute the span gas and, hence, the concentration that the analyzer
measures may fall short of the expected concentration defined in the CONC
sub-menu.
4. If the low-level, hardware calibration has drifted (changed PMT response) or
was accidentally altered by the user, a low-level calibration may be
necessary to get the analyzer back into its proper range of expected values.
One possible indicator of this scenario is a slope or offset value that is outside
of its allowed range (0.7-1.3 for slope, -20 to 150 for offsets). See Section
12.8.4 on how to carry out a low-level hardware calibration.
12.5.5. INABILITY TO ZERO - NO ZERO BUTTON (CALZ)
In general, the 9110T will not display certain buttons whenever the actual value of a parameter is outside
of the expected range for that parameter. If the calibration menu does not show a ZERO button when
carrying out a zero calibration, the actual gas concentration must be significantly different from the
actual zero point (as per last calibration), which may be for any of several reasons.
1. Confirm that there is a good source of zero air. If the IZS option is installed,
compare the zero reading from the IZS zero air source to a zero air source
Teledyne Analytical Instruments
278
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
using NOX-free air. Check any zero air scrubber for performance. It may
need to be replaced (Section 11.3.4.2).
2. Check to ensure that there is no ambient air leaking into zero air line. Check
for leaks in the pneumatic systems as described in Section 11.3.12.
12.5.6. NON-LINEAR RESPONSE
The 9110T was factory calibrated to a high level of NO and should be linear to within 1% of full scale.
Common causes for non-linearity are:
Leaks in the pneumatic system:
Leaks can add a constant of ambient air, zero air or span gas to the current sample gas
stream, which may be changing in concentrations as the linearity test is performed.
Check for leaks as described in Section 11.3.12.
The calibration device is in error:
Check flow rates and concentrations, particularly when using low concentrations.
If a mass flow calibrator is used and the flow is less than 10% of the full scale flow on either
flow controller, you may need to purchase lower concentration standards.
The standard gases may be mislabeled as to type or concentration.
Labeled concentrations may be outside the certified tolerance.
The sample delivery system may be contaminated.
Check for dirt in the sample lines or reaction cell.
Calibration gas source may be contaminated (NO2 in NO gas is common).
Dilution air contains sample or span gas.
Ozone concentration too low because of wet air in the generator.
Generator system needs to be cleaned and dried with dry supply air.
Check the Perma Pure dryer for leaks.
This mostly affects linearity at the low end.
Ozone stream may be contaminated with impurities.
An exhausted ozone cleanser chemical will let compounds such as HNO3 and ammonia
derivatives break through to the reaction cell.
Check the contents of the ozone cleanser and replace as necessary (Section 11.3.3).
This also will affect linearity mostly at the low level.
Sample inlet may be contaminated with NOx exhaust from this or other analyzers.
Teledyne Analytical Instruments
279
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Verify proper venting of the pump exhaust.
Span gas overflow is not properly vented and creates a back-pressure on the sample inlet port.
Also, if the span gas is not vented at all and does not supply enough sample gas, the
analyzer may be evacuating the sample line.
Ensure to create and properly vent excess span gas.
Diffusion of oxygen into Teflon-type tubing over long distances.
PTFE or related materials can act as permeation devices. In fact, the permeable membrane
of NO2 permeation tubes is made of PTFE.
When using very long supply lines (> 1 m) between high concentrations span gases and the
dilution system, oxygen from ambient air can diffuse into the line and react with NO to form
NO2.
This reaction is dependent on NO concentration and accelerates with increasing NO concentration,
hence, affects linearity only at high NO levels.
Using stainless steel for long span gas supply lines avoids this problem.
12.5.7. DISCREPANCY BETWEEN ANALOG OUTPUT AND DISPLAY
If the concentration reported through the analog outputs does not agree with the value reported on the
front panel, you may need to recalibrate the analog outputs.
This becomes more likely when using a low concentration or low analog output range.
Analog outputs running at 0.1 V full scale should always be calibrated manually.
See Section 5.9.3.2 for a detailed description of this procedure.
12.5.8. DISCREPANCY BETWEEN NO AND NOX SLOPES
If the slopes for NO and NOX are significantly different after software calibration (more than 1%),
consider the following three problems:
NO2 impurities in the NO calibration gas. NO gases often exhibit NO2 on the order of 1-2% of the
NO value.
This will cause differences in the calibration slopes. If the NO2 impurity in NO is known, it
can easily be accounted for by setting the expected values for NO and NO2 accordingly to
different values, e.g., 448 ppb NO and 450 ppb NOX.
This problem is worse if NO gas is stored in a cylinder with balance air instead of balance
gas nitrogen or large amounts of nitrous oxide (N2O).
The oxygen in the air slowly reacts with NO to yield NO2, increasing over time.
The expected concentrations for NO and NOX in the calibration menu are set to different values.
If a gas with 100% pure NO is used, this would cause a bias.
See Section 9.2.3.1 on how to set expected concentration values.
Teledyne Analytical Instruments
280
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
The converter efficiency parameter has been set to a value not equal to 1.000 even though the
conversion efficiency is 1.0.
The actual conversion efficiency needs to match the parameter set in the CAL menu.
See Section 9.1.4 for more information on this feature.
An instrument calibration with the IZS option (and expected concentrations set to the same amount) will
always yield identical slopes for NO and NOX, as the instrument measures only NOX and assumes NO to
be the same (with NO2 being zero).
12.6. OTHER PERFORMANCE PROBLEMS
Dynamic problems (i.e. problems that only manifest themselves when the analyzer is monitoring sample
gas) can be the most difficult and time consuming to isolate and resolve. The following section provides
an itemized list of the most common dynamic problems with recommended troubleshooting checks and
corrective actions.
12.6.1. EXCESSIVE NOISE
Excessive noise levels under normal operation usually indicate leaks in the sample supply or the analyzer
itself. Ensure that the sample or span gas supply is leak-free and carry out a detailed leak check as
described earlier in this chapter.
Another possibility of excessive signal noise may be the preamplifier board, the high voltage power
supply and/or the PMT detector itself.
Contact the factory on troubleshooting these components.
12.6.2. SLOW RESPONSE
If the analyzer starts responding too slow to any changes in sample, zero or span gas, check for the
following:
Dirty or plugged sample filter or sample lines.
Sample inlet line is too long.
Leaking NO/NOX valve. Carry out a leak check.
Dirty or plugged critical flow orifices. Check flows, pressures and, if necessary, change orifices
(Section 11.3.10).
Wrong materials in contact with sample - use glass, stainless steel or Teflon materials only. Porous
materials, in particular, will cause memory effects and slow changes in response.
Dirty reaction cell. Clean the reaction cell.
Insufficient time allowed for purging of lines upstream of the analyzer. Wait until stability is low.
Insufficient time allowed for NO or NO2 calibration gas source to become stable. Wait until stability
is low.
NO2 converter temperature is too low. Check for proper temperature.
Teledyne Analytical Instruments
281
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.6.3. AUTO ZERO WARNINGS
Auto Zero warnings occur if the signal measured during an Auto Zero cycle is higher than 200 mV.
Note
The Auto-Zero warning displays the value of the Auto Zero reading when
the warning occurs.
If this value is higher than 150 mV, check that the Auto Zero valve is operating properly.
To do so, use the SIGNAL I/O functions in the DIAG menu to toggle the valve on and off.
Listen if the valve is switching, see if the respective LED on the relay board is indicating
functionality.
Scroll the TST functions until PMT is displayed and observe the PMT value change between the two
valve states.
If the valve is operating properly, you should be able to hear it switch (once a minute under
normal operation or when manually activated from the SIGNAL I/O menu):
the PMT value should drop from span gas reading (e.g., 800-900 mV at
400 ppb NO) to less than 150 mV and;
the LED on the relay board should light up when the valve is activated.
If the PMT value drops significantly but not to less than 150 mV, the valve is probably
leaking across its ports.
In this case, replace the valve.
If the PMT value does not change at all, the valve is probably not switching at all.
Check the power supply to the valve (12 V to the valve should turn on and
off when measured with a voltmeter).
Note
It takes only a small leak across the ports of the valve to show excessive
Auto Zero values when supplying high concentrations of span gas.
Another reason for high (although not necessarily out-of-range) values for Auto Zero could be the ozone
air filter cartridge, if its contents have been exhausted and needs to be replaced.
This filter cartridge chemicals that can cause chemiluminescence and, if saturated, these chemicals
can break through to the reaction cell, causing an erroneously high Auto Zero value (background
noise).
A dirty reaction cell can cause high Auto Zero values.
Clean the reaction cell according to Section 11.3.9.
Finally, a high HVPS voltage value may cause excess background noise and a high AZERO value.
The HVPS value changes from analyzer to analyzer and could show nominal values between 450
and 800 V.
Check the low-level hardware calibration of the preamplifier board and, if necessary, recalibrate
exactly as described in Section 12.8.4 in order to minimize the HVPS.
Teledyne Analytical Instruments
282
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7. SUBSYSTEM CHECKOUT
The preceding sections of this manual discussed a variety of methods for identifying possible sources of
failures or performance problems within the analyzer. In most cases this included a list of possible
causes and, in some cases, quick solutions or at least a pointer to the appropriate sections describing
them. This section describes how to determine if a certain component or subsystem is actually the cause
of the problem being investigated.
12.7.1. AC MAIN POWER
The 9110T analyzer’s electronic systems will operate with any of the specified power regimes. As long
as system is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz it will turn on and after
about 30 seconds show a front panel display.
Internally, the status LEDs located on the Relay PCA, Motherboard and CPU should turn on as soon
as the power is supplied.
If they do not, check the circuit breaker built into the ON/OFF switch on the instruments front panel.
If the instrument is equipped with an internal pump, it will begin to run. If it does not:
Verify that the pump power configuration plug is properly wired (see Section 13.7.1.1 and
Figure 13-27)
If the configuration plug is set for 230 VAC and the instrument is plugged into 115 VAC or
100 VAC the sample pump will not start.
If the configuration plug is set for 115 or 100 VAC and the unit is plugged into a 230 VAC circuit, the
circuit breaker built into the ON/OFF Switch on the front panel will trip to the OFF position
immediately after power is switched on.
9110T’s without internal pumps that are configured for 230 V will still turn on at 115 V, but the
heaters may burn out or not heat up fast enough.
WARNING – ELECTRICAL SHOCK HAZARD
Should the AC power circuit breaker trip, investigate and correct the condition
causing this situation before turning the analyzer back on.
12.7.2. DC POWER SUPPLY
If you have determined that the analyzer’s AC mains power is working, but the unit is still not operating
properly, there may be a problem with one of the instrument’s switching power supplies. The supplies
can have two faults, namely no DC output, and noisy output.
To assist tracing DC Power Supply problems, the wiring used to connect the various printed circuit
assemblies and DC Powered components and the associated test points on the relay PCA follow a
standard color-coding scheme as defined in the following table.
Teledyne Analytical Instruments
283
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Table 12-6:
DC Power Test Point and Wiring Color Codes
NAME
DGND
+5V
TEST POINT#
COLOR
Black
Red
DEFINITION
Digital ground
1
2
3
4
5
6
AGND
+15V
-15V
Green
Blue
Analog ground
Yellow
Purple
+12R
12 V return
(ground) line
+12V
Orange
7
TP1 TP2 TP3 TP4 TP5 TP6 TP7
DGND +5V AGND +15V -15V +12R 12V
Figure 12-4:
Location of DC Power Test Points on Relay PCA
A voltmeter should be used to verify that the DC voltages are correct per the values in the table below,
and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if the supplies
are producing excessive noise (> 100 mV p-p).
Table 12-7: DC Power Supply Acceptable Levels
VOLTAGE
CHECK RELAY BOARD TEST POINTS
MIN V
MAX V
POWER
SUPPLY
FROM
Test Point
TO
Test Point
NAME
DGND
#
1
3
3
3
1
6
6
NAME
#
2
PS1
PS1
PS1
PS1
PS1
PS2
PS2
+5
+15
+5
+15
+4.85
+13.5
-13.5
-0.05
-0.05
+11.8
-0.05
+5.25
+16.0
-16.0
+0.05
+0.05
+12.5
+0.05
AGND
4
-15
AGND
-15V
5
AGND
Chassis
+12
AGND
DGND
Chassis
+12V
DGND
1
DGND
N/A
7
+12V Ret
+12V Ret
DGND
1
Teledyne Analytical Instruments
284
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.3. I2C BUS
Operation of the I2C bus can be verified by observing the behavior of D1 on the relay PCA & D2 on the
Valve Driver PCA. Assuming that the DC power supplies are operating properly, the I2C bus is operating
properly if D1 on the relay PCA and D2 of the Valve Driver PCA are flashing
There is a problem with the I2C bus if both D1 on the relay PCA and D2 of the Valve Driver PCA are
ON/OFF constantly.
12.7.4. LCD/DISPLAY MODULE
TOUCHSCREEN INTERFACE
Assuming that there are no wiring problems and that the DC power supplies are operating properly, the
display screen should light and show the splash screen and other indications of its state as the CPU goes
through its initialization process.
12.7.5. RELAY PCA
The Relay PCA can be most easily checked by observing the condition of the status LEDs on the Relay
PCA (see Section 12.3.2), and using the SIGNAL I/O submenu under the DIAG menu (see Section
12.1.3) to toggle each LED ON or OFF.
If D1 on the Relay PCA is flashing and the status indicator for the output in question (Heater power,
Valve Drive, etc.) toggles properly using the Signal I/O function, then the associated control device on
the Relay PCA is bad.
Several of the control devices are in sockets and can be easily replaced. The following table lists the
control device associated with a particular function:
Table 12-8:
Relay PCA Control Devices
FUNCTION
CONTROL DEVICE
SOCKETED
All valves
U5
K1
K2
Yes
Yes
Yes
Reaction Cell Heater
NO2 NO Converter heater
Permeation Tube Heater for
Optional Internal Span Gas Generator
K4
Yes
12.7.6. MOTHERBOARD
12.7.6.1. Test Channel / Analog Outputs Voltage
The ANALOG OUTPUT submenu, located under the SETUP MORE DIAG menu is used to
verify that the 9110T analyzer’s three analog outputs are working properly. The test generates a signal
on all three outputs simultaneously as shown in the following table:
Teledyne Analytical Instruments
285
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Table 12-9: Analog Output Test Function - Nominal Values Voltage Outputs
FULL SCALE OUTPUT OF VOLTAGE RANGE
(see Section 5.9.3.1)
100MV
1V
5V
10V*
STEP
%
0
NOMINAL OUTPUT VOLTAGE
1
2
3
4
5
6
0
0
0
1
2
3
4
5
0
2
20
40
60
80
100
20 mV
40 mV
60 mV
80 mV
100 mV
0.2
0.4
0.6
0.8
1.0
4
6
8
10
* For 10V output, increase the Analog Output Calibration Limits (AOUT CAL LIM in the
DIAG>Analog I/O Config menu) to 4% (offset limit) and 20% (slope limit).
For each of the steps the output should be within 1% of the nominal value listed except for the 0% step,
which should be within 0mV ±2 to 3 mV. Ensure you take into account any offset that may have been
programmed into channel (See Section 5.9.3.9).
If one or more of the steps fails to be within these ranges, it is likely that there has been a failure of the
either or both of the Digital-to-Analog Converters (DACs) and their associated circuitry on the
motherboard. To perform the test connect a voltmeter to the output in question and perform an analog
output step test as follows:
SETUP X.X
SETUP X.X
CFG DAS RNGE PASS CLK
EXIT
EXIT
EXIT
SETUP X.X
DIAG
COMM VARS
PREV
ENTR EXIT
DIAG AOUT
DIAG AOUT
Performs analog output step
test 0% to 100%
· Pressing the “x%” button pauses the
test. Brackets will appear around the
value: EXAMPLE: [10%]
· Pressing the “[x%]” button resumes the
test.
12.7.6.2. A/D Functions
The simplest method to check the operation of the A-to-D converter on the motherboard is to use the
Signal I/O function under the DIAG menu to check the two A/D reference voltages and input signals
that can be easily measured with a voltmeter.
Teledyne Analytical Instruments
286
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
1. Use the Signal I/O function (see Section 12.1.3 and Appendix A) to view the value
of REF_4096_MV and REF_GND.
If both are within 3 mV of nominal (4096 and 0), and are stable, ±0.2 mV then
the basic A/D is functioning properly. If not then the motherboard is bad.
2. Choose a parameter in the Signal I/O function list (see Section 12.1.3) such as
OZONE_FLOW .
Compare this voltages at its origin (see the interconnect drawing and
interconnect list in Appendix D) with the voltage displayed through the signal I/O
function.
If the wiring is intact but there is a large difference between the measured and
displayed voltage (±10 mV) then the motherboard is bad.
12.7.6.3. Status Outputs
V
+DC Gnd
Figure 12-5:
Typical Set Up of Status Output Test
To test the status output electronics:
1. Connect a jumper between the “D" pin and the “” pin on the status output
connector.
2. Connect a 1000 ohm resistor between the “+” pin and the pin for the status output
that is being tested.
3. Connect a voltmeter between the “” pin and the pin of the output being tested.
4. Under the DIAG Signal I/O menu (see Section 12.1.3), scroll through the inputs
and outputs until you get to the output in question.
Teledyne Analytical Instruments
287
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
5. Alternately, turn on and off the output noting the voltage on the voltmeter.
It should vary between 0 volts for ON and 5 volts for OFF.
Table 12-10: Status Outputs Check
PIN (LEFT TO RIGHT)
STATUS
1
2
3
4
5
6
7
8
ST_SYSTEM_OK
ST_CONC_VALID
ST_HIGH_RANGE
ST_ZERO_CAL
ST_SPAN_CAL
ST_DIAG_MODE
Not Used on 9110T
ST_O2_CAL
12.7.6.4. Control Inputs
The control input bits can be tested by applying a trigger voltage to an input and watching changes in the
status of the associated function under the SIGNAL I/O submenu:
EXAMPLE: to test the “A” control input:
1. Under the DIAG Signal I/O menu (see Section 12.1.3), scroll through the inputs
and outputs until you get to the output named EXT_ZERO_CAL.
2. Connect a jumper from the “+” pin on the appropriate connector to the “U” on the
same connector.
3. Connect a second jumper from the “” pin on the connector to the “A” pin.
4. The status of EXT_ZERO_CAL should change to read “ON”.
5. Connect a second jumper from the “” pin on the connector to the “B” pin.
6. The status of EXT_ZERO_CAL should change to read “ON”.
Table 12-11: 9110T Control Input Pin Assignments and Corresponding Signal I/O
Functions
INPUT
CORRESPONDING I/O SIGNAL
EXT_ZERO_CAL
A
B
EXT_SPAN_CAL1
NOT USED
C, D, E& F
Teledyne Analytical Instruments
288
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.7. PRESSURE / FLOW SENSOR ASSEMBLY
The flow and pressure sensors of the 9110T are located on a PCA just behind the PMT sensor (see
Figure 3-5) can be checked with a Voltmeter.
Figure 12-6:
Pressure / Flow Sensor Assembly
The following procedure assumes that the wiring is intact and that the motherboard and power supplies
are operating properly:
12.7.7.1. Basic PCA Operation Check:
Measure the voltage between TP2 and TP1 C1 it should be 10 VDC ± 0.25 VDC. If not then the
board is bad. Replace the PCA.
12.7.7.2. Sample Pressure Sensor Check:
1. Measure the pressure on the inlet side of S1 with an external pressure meter.
2. Measure the voltage across TP4 and TP1.
The expected value for this signal should be:
EXAMPLE: If the measured pressure is 20 Hg-in-A, the expected voltage level between TP4 and TP1
would be between 2870 mVDC and 3510 mVDC.
Teledyne Analytical Instruments
289
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
EXAMPLE: If the measured pressure is 25 Hg-in-A, the expected voltage level between TP4 and TP1
would be between 3533 mVDC and 4318 mVDC.
If this voltage is out of range, then either pressure transducer S1 is bad, the board is bad or there is
a pneumatic failure preventing the pressure transducer from sensing the absorption cell pressure
properly. Replace the PCA.
12.7.7.3. Vacuum Pressure Sensor Check
Measure the pressure on the inlet side of S2 with an external pressure meter.
Measure the voltage across TP5 and TP1.
Evaluate the reading in the same manner as for the sample pressure sensor.
12.7.7.4. O3 Flow Sensor Check
Measure the voltage across TP3 and TP1.
With proper flow (80 cc3/min through the O3 generator), this should be approximately 2V ±
0.25 (this voltage will vary with altitude).
With flow stopped (photometer inlet disconnected or pump turned OFF) the voltage should
be approximately 1V.
If the voltage is incorrect, the flow sensor S3 is bad, the board is bad (replace the PCA) or
there is a leak upstream of the sensor.
12.7.8. CPU
There are two major types of CPU board failures, a complete failure and a failure associated with the
Disk On Module (DOM). If either of these failures occurs, contact the factory.
For complete failures, assuming that the power supplies are operating properly and the wiring is intact,
the CPU is faulty if on power-on, the watchdog LED on the motherboard is not flashing.
In some rare circumstances, this failure may be caused by a bad IC on the motherboard, specifically
U57, the large, 44 pin device on the lower right hand side of the board. If this is true, removing U57
from its socket will allow the instrument to start up but the measurements will be invalid.
If the analyzer stops during initialization (the front panel display shows a fault or warning message),
it is likely that the DOM, the firmware or the configuration and data files have been corrupted.
12.7.9. RS-232 COMMUNICATIONS
12.7.9.1. General RS-232 Troubleshooting
TAI's analyzers use the RS-232 communications protocol to allow the instrument to be connected to a
variety of computer-based equipment. RS-232 has been used for many years and as equipment has
become more advanced, connections between various types of hardware have become increasingly
difficult. Generally, every manufacturer observes the signal and timing requirements of the protocol
very carefully.
Problems with RS-232 connections usually center around 4 general areas:
Teledyne Analytical Instruments
290
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Incorrect cabling and connectors. See Section 3.3.1.8, Figure 3-13 for connector and pin-out
information.
The BAUD rate and protocol are incorrectly configured. See Section 6.2.2.
If a modem is being used, additional configuration and wiring rules must be observed. See Section
8.3
Incorrect setting of the DTE – DCE Switch. See Section 6.1 to set correctly.
Verify that cable (P/N 03596) that connects the serial COMM ports of the CPU to J12 of the
motherboard is properly seated.
12.7.9.2. Troubleshooting Analyzer/Modem or Terminal Operation
These are the general steps for troubleshooting problems with a modem connected to a Teledyne
analyzer.
1. Check cables for proper connection to the modem, terminal or computer.
2. Check to ensure that the DTE-DCE is in the correct position as described in
Section 6.1.
3. Check to ensure that the set up command is correct (see Section 8.3).
4. Verify that the Ready to Send (RTS) signal is at logic high. The 9110T sets
pin 7 (RTS) to greater than 3 volts to enable modem transmission.
5. Ensure that the BAUD rate, word length, and stop bit settings between
modem and analyzer match. See Section 6.2.2.
6. Use the RS-232 test function to send “w” characters to the modem, terminal
or computer. See Section 6.2.3.
7. Get your terminal, modem or computer to transmit data to the analyzer
(holding down the space bar is one way); the green LED should flicker as the
instrument is receiving data.
8. Ensure that the communications software or terminal emulation software is
functioning properly.
Note
Further help with serial communications is available in a separate manual
“RS-232 Programming Notes” TAI P/N 01350.
12.7.10. NO2 NO CONVERTER
Provided that oxygen was present in the Sample stream during operation for the NO2 converter to
function properly, the NO2 converter assembly can fail in two ways:
An electrical failure of the band heater and/or the thermocouple control circuit and;
Teledyne Analytical Instruments
291
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
A performance failure of the converter itself.
12.7.10.1. NO2 NO Converter Electrical System
NO2 converter heater failures can be divided into two possible problems:
Temperature is reported properly but heater does not heat to full temperature.
In this case, the heater is either disconnected or broken or the power relay is broken.
Disconnect the heater cable coming from the relay board and measure
the resistance between any two of the three heater leads with a multi-
meter.
The resistance between A and B should be about 1000 Ω.
That between A and C should be the same as between B and C, about
500 Ω each.
If any of these resistances is near zero or without continuity, the heater is
broken.
Temperature reports zero or overload (near 500° C).
This indicates a disconnected or failing thermocouple or a failure of the thermocouple circuit.
Check that the thermocouple is connected properly and the wire does not
show signs of a broken or kinked pathway.
If it appears to be properly connected, disconnect the yellow
thermocouple plug (marked K) from the relay board and measure the
voltage (not resistance) between the two leads with a multi-meter
capable of measuring in the low mV range.
The voltage should be about 12 mV (ignore the sign) at 315° C and about
0 mV at room temperature.
Measure the continuity with an Ohm-meter.
It should read close to zero Ω. If the thermo-couple does not have continuity, it is broken.
If it reads zero voltage at elevated temperatures, it is broken.
To test the thermocouple at room temperature, heat up the converter can (e.g., with a heat gun) and
see if the voltage across the thermocouple leads changes.
If the thermocouple is working properly, the electronic circuit is broken.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
If the thermocouple is broken, do NOT replace the thermocouple without first
consulting the factory; using the wrong Type could cause permanent damage to
the instrument. The Type K thermocouple has a red and a yellow wire. If in
doubt, consult the factory.
Teledyne Analytical Instruments
292
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.10.2. NO2 Conversion Efficiency
The efficiency at which the NO2 NO converter changes NO2 into NO directly effects the accuracy of
the 9110T’s NOx, NO and NO2 measurements. The 9110T firmware includes a Converter Efficiency
(CE) gain factor that is multiplied by the NO2 and NOX measurements to calculate the final
concentrations for each. This gain factor is stored in the analyzer’s memory.
The default setting for the NO2 converter efficiency is 1.0000. Over time, the molybdenum in the NO2
NO converter oxidizes and it becomes less efficient at converting NO2 into NO.
To ensure accurate operation of the 9110T, it is important to check the NO2 conversion efficiency
periodically and to update this value as necessary.
For the analyzer to function correctly, the converter efficiency must be between 0.9600 and 1.0200
(96-102% conversion efficiency) as per US-EPA requirements.
If the converter’s efficiency is outside these limits, the NO2 converter should be replaced.
The current converter efficiency level is also recorded along with the calibration data in the DAS for
documentation and performance analysis (Section 7).
12.7.10.3. Calculating / Checking Converter Efficiency
The 9110T to automatically calculate the current NO2 conversion efficiency by comparing a known
starting concentration of NO2 gas to the measured NO output of the converter. There are three steps to
performing this operation.
Step 1:
Supply the analyzer with a known concentration of NO2 gas, to the analyzer.
Figure 12-7:
Setup for determining NO2 NO Efficiency – 9110T Base Configuration
Step 2:
Teledyne Analytical Instruments
293
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Input the starting NO2 concentration value into the 9110T by pressing:
SAMPLE
RANGE=500.0 PPB
CAL
NO=XXXX
<TST TST>
SETUP
This step only appears if the
analyzer’s reporting range is
set for AUTO range mode.
SAMPLE
RANGE TO CAL
Select LOW and press ENTR.
LOW HIGH
ENTR EXIT
Use these buttons to
select the appropriate
range.
Repeat entire procedure for
HIGH range.
Repeat entire procedure
for each range.
SAMPLE
RANGE=500.0 PPB
NO=XXXX
<TST TST> CAL
SETUP
SAMPLE
RANGE=500.0 PPB
NO=XXXX
<TST TST> ZERO
CONC
SETUP
M-P CAL
CONCENTRATON MENU
NOX NO CONV
EXIT
EXIT
Converter Efficiency Menu
M-P CAL
CONVERTER EFICIENCY MENU
NO2 CAL SET
The expected NO2 span
concentration value defaults
to 400.0 Conc.
M-P CAL
NO2 CE CONC: 500.0 Conc
0
4
0
0.
0
0
ENTR EXIT
Toggle these buttons
to change this value to
the concentration of
the NO2 gas being
used.
Make sure that you specify
the actual concentration value
of the NO2 gas.
Teledyne Analytical Instruments
294
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
STEP 3:
To cause the analyzer to calculate and record the NO2 NO converter efficiency, press:
Starting from
CONVERTER EFFICIENCY MENU
(see preceding steps)
M-P CAL
CONVERTER EFFICIENCY MENU
NO2 CAL SET
EXIT
M-P CAL
CE FACTOR:1000.0 Gain
1.
0
0
0
0
ENTR EXIT
Toggle these
buttons to initialize
the converter
efficiency at 1.0000.
M-P CAL
CONVERTER EFFICIENCY MENU
NO2 CAL SET
EXIT
SAMPLE
RANGE=500.0 PPB
NOX= XXXX
SETUP
< TST TST >
ENTR
Toggle TST> button until ...
Set the Display to show
the NOX STB test
function.
This function calculates
the stability of the NO/NOx
measurement.
SAMPLE
NO2 STB=XX.X PPB
<TST TST>
SETUP
Allow NO2 gas of the proper concetration to enter
the sample port at the rear of the analyzer.
Wait until NOX STB
falls below 0.5 ppb.
This may take several
minutes.
The analyzer
calculates the
converter’s
efficiency.
SAMPLE
NO2 STB=XX.X PPB
<TST TST> ENTR
SETUP
M-P CAL
CONVERTER EFICIENCY MENU
EXIT
NO2 CAL SET
Check the calculated efficiency
gain factor.
If the gain factor is NOT
between 0.9600 and 1.0200,
the
M-P CAL
0.
CE FACTOR=0.9852 Gain
ENTR EXIT
8
8
5
2
NO2
NO converter
needs to be replaced.
12.7.10.4. Evaluating NO2 NO Converter Performance
If the converter appears to have performance problems (conversion efficiency is outside of allowed range
of 96-102%), check the following:
Teledyne Analytical Instruments
295
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Recalculate the converter efficiency (see previous section)
Accuracy of NO2 source (GPT or gas tank standard).
NO2 gas standards are typically certified to only ±2% and often change in concentrations
over time. You should get the standard re-certified every year.
If you use the GPT calibration, check the accuracy of the ozone source.
Age of the converter.
The NO2 converter has a limited operating life and may need to be replaced every ~3 years
or when necessary (e.g., earlier if used with continuously high NO2 concentrations).
We estimate a lifetime of about 10000 ppm-hours (a cumulative product of the NO2
concentration times the exposure time to that concentration).
This lifetime heavily depends on many factors such as:
Absolute concentration (temporary or permanent poisoning of the
converter is possible).
Sample flow rate and pressure inside the converter.
Converter temperature.
Duty cycle.
This lifetime is only an estimated reference and not a guaranteed lifetime.
In some cases with excessive sample moisture, the oxidized molybdenum metal chips inside the
converter cartridge may bake together over time and restrict air flow through the converter, in which
case it needs to be replaced.
To avoid this problem, we recommend the use of a sample gas conditioner (Section
3.3.2.6).
Section 11.3.8 describes how to replace the NO2 converter cartridge.
With no NO2 in the sample gas and a properly calibrated analyzer, the NO reading is negative, while
the NO2 reading remains around zero.
The converter is destroying NO and needs to be replaced.
With no NO2 in the sample gas and a properly calibrated analyzer, the NOX reading is significantly
higher than the actual (gas standard) NO concentration.
The converter is producing NO2 and needs to be replaced.
12.7.11. SIMPLIFIED GPT CALIBRATION
This section describes how to determine the NO2 NO converter’s efficiency using a GPT
method where the actual concentration of ozone is not a factor in the accuracy of the calculation.
This procedure is based on the Code of Federal Regulations, Title 40, Chapter I, subchapter C, Part
50, Appendix F.
Teledyne Analytical Instruments
296
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
In the following example a reference point of 450 ppb NO gas will be used. This is only an example.
Any other reference points within measurement range of the instrument may be used.
For this procedure use a calibrated O3 generator, such as a Teledyne T700.
Note
There must be a minimum of 10% more NO than O3 produced.
Example, if the Ozone concentration used is 400 ppb then the NO
concentration must be used must be 440 ppb or more.
PART 1: PREPARATION
1. Leak check machine to ensure that there are no leaks in the analyzer.
2. Calibrate the instrument at the same NO span gas value as being used in this
method.
For this example 450 ppb NO span gas
3. If you have input a converter efficiency (CE) factor into the instrument firmware (see
12.7.10.3 ) other than 100%, change this back to 100% for the duration of this test.
(CAL>CONC>CONV>SET).
PART 2: DETERMINE THE AMOUNT OF NO OUTGASSED BY THE NO2 NO
CONVERTER.
4. Bypass the NO2 NO converter by placing a short piece of tubing in the gas
stream in place of the converter.
5. Perform a straight dilution with 445 ppb NO gas & air as a diluent gas.
6. Input the NO gas into the analyzer.
7. Allow the machine to stabilize & write down the NOx value on line 2 of GPT data
sheet (Section 12.7.11.1).
8. Remove the converter bypass so that the NO gas is flowing through the NO2 NO
converter
9. Allow the machine to stabilize.
10. Write down your NOx value on your data sheet on lines 3 AND line 5 of the GPT
data sheet.
11. Subtract line 2 from line 3 & write that number down on line 4. Also write the NO
value on line 8 of the data sheet.
The specification shown on the data sheet is the value that is used by TAI when performing
the procedure on new NO2 NO converters.
Older NO2 NO converters might outgas a bit more NO, therefore a slightly wider
specification might be in order.
If this value is a constant or changes only slightly over time, this is not a problem the
machine will calibrate this out.
Teledyne Analytical Instruments
297
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
PART 3: PERFORM THE SIMPLIFIED GPT CALCULATION.
12. Generate the same 450 ppb NO gas & input 400 ppb of O3 (or generate 450 ppb
NO & 400 ppb NO2, if that’s what your calibrator says).
13. Allow the instrument to stabilize for 10 minutes.
14. Write down the NOx value on line 6 & the NO value on line 9.
15. Subtract line 6 from line 6 & put that onto line 7.
16. Subtract line 8 from line 7 & put that onto line 10.
17. Write the number from line 7 into the blank next to letter A on line 11 & put the
number from line 10 into the blank next to letter B on line 11.
18. Divide A by B & multiply it by 100.
19. Write this value it into the blank next to letter C on lines 11 and 12.
20. Subtract that value from 100 & write it in the blank next to the letter D on line 12.
21. This is the converter efficiency.
This value should be >96%.
For CEMS applications, a converter efficiency of <96% might be acceptable,
depending on application & the guideline set up by the regulatory agency.
In any application, check with your regulatory agency (agencies) to see what
the minimum CE factor is before replacing the converter.
Teledyne Analytical Instruments
298
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.11.1. Simplified GPT Data Sheet
Line # TEST
RESULT
1
2
3
4
5
6
LEAK-CHECK (WHEN HOT)
YES / NO
NOx RESPONSE (MOLY BYPASSED)
NOx RESPONSE (MOLY IN-LINE)
OUT-GASSING (NO – NOX)
__________
__________
__________ (>-5 ppb, <5 ppb)
__________ ppb
__________ ppb
(NOx ORIG
)
(NOx mode, O3 off)
(NOx mode, O3 on)
(NOx REM
)
7
NOx LOSS
__________ (A)
(<4% of NOx ORIG
:
for 450PP 4% is 18 ppb)
__________ ppb
8
9
(NO ORIG
)
(NO mode, O3 off)
(NO mode, O3 on)
(NO REM
)
__________ ppb
10 NO2
__________ (B) (>300ppb)
11 Efficiency LOSS [ ( A / B ) x 100 ] = [ ( ____A / ____B ) x 100 ] = ____C%
12 Total Conv Eff [ 100% – C ] = 100% - ____C = _____D% ( > 96%)
Teledyne Analytical Instruments
299
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.12. PHOTOMULTIPLIER TUBE (PMT) SENSOR MODULE
The PMT detects the light emitted by the reaction of NO with ozone. It has a gain of about 500000 to
1000000. It is not possible to test the detector outside of the instrument in the field. The basic method to
diagnose a PMT fault is to eliminate the other components using ETEST, OTEST and specific tests for
other sub-assemblies.
12.7.12.1. Optic Test
The optic test function tests the response of the PMT sensor by turning on an LED located in the cooling
block of the PMT (see Figure 13-20). The analyzer uses the light emitted from the LED to test its photo-
electronic subsystem, including the PMT and the current to voltage converter on the pre-amplifier board.
To ensure that the analyzer measures only the light coming from the LED, the analyzer should be
supplied with zero air.
The optic test should produce a PMT signal of about 2000±1000 mV.
To activate the optics test, press:
This is a coarse test for functionality and not an accurate calibration tool.
The resulting PMT signal can vary significantly over time and also
changes with low-level calibration.
Note
Teledyne Analytical Instruments
300
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
12.7.12.2. Electrical Test
Troubleshooting & Service
The electrical test function creates a current, which is substituted for the PMT signal and feeds it into the
preamplifier board.
This signal is generated by circuitry on the pre-amplifier board itself and tests the filtering and
amplification functions of that assembly along with the A/D converter on the motherboard.
It does not test the PMT itself.
The electrical test should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test, press:
Teledyne Analytical Instruments
301
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.13. PMT PREAMPLIFIER BOARD
To check the correct operation of the preamplifier board, perform an the optics test (OTEST) and an
electrical test (ETEST) described in Sections 12.7.12.1 and 12.7.12.2 above.
If the instrument passes the OTEST but fails the ETEST, the preamplifier board may be faulty or
need a hardware calibration.
12.7.13.1. High Voltage Power Supply
The HVPS is located in the interior of the sensor module and is plugged into the PMT tube. It requires 2
voltage inputs.
The first is +15 V, which powers the supply.
The second is the programming voltage which is generated on the preamplifier board.
Adjustment of the HVPS is covered in the factory calibration procedure in Section 12.8.4.
This power supply has 10 independent power supply steps, one to each pin of the PMT. The following
test procedure below allows you to test each step.
1. Turn off the instrument.
2. Remove the cover and disconnect the 2 connectors at the front of the NOX sensor
module.
3. Remove the end cap from the sensor (4 screws).
4. Remove the HVPS/PMT assembly from the cold block inside the sensor (2 plastic
screws).
5. Disconnect the PMT from the HVPS.
6. Re-connect the 7 pin connector to the sensor end cap, and power-up the
instrument.
7. Scroll the front panel display to the HVPS test parameter.
8. Divide the displayed HVPS voltage by 10 and test the pairs of connector points as
shown in the figure below.
9. Check the overall voltage (should be equal to the HVPS value displayed on the front
panel and the voltages between each pair of pins of the supply
EXAMPLE
If the HVPS signal is 700 V the pin-to-pin voltages should be 70 V.
Teledyne Analytical Instruments
302
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
1. Turn off the instrument power, and reconnect the PMT, and then reassemble the sensor.
If any faults are found in the test, you must obtain a new HVPS as there are no user
serviceable parts inside the supply.
12.7.14. PMT TEMPERATURE CONTROL PCA
The TEC control PCA is located on the sensor housing assembly, under the slanted shroud, next to the
cooling fins and directly above the cooling fan.
If the red LED located on the top edge of this assembly is not glowing the control circuit is not receiving
power. Check the analyzers power supply, the relay board’s power distribution circuitry and the wiring
connecting them to the PMT temperature control PCA.
TEC Control Test Points
Four test points are also located at the top of this assembly they are numbered left to right start with the
T1 point immediately to the right of the power status LED. These test points provide information
regarding the functioning of the control circuit.
To determine the current running through the control circuit, measure the voltage between T1 and
T2. Multiply that voltage by 10.
To determine the drive voltage being supplied by the control circuit to the TEC, measure the voltage
between T2 and T3.
If this voltage is zero, the TEC circuitry is most likely open.
Or,
If the voltage between T2 and T3 = 0 VDC and the voltage measured between T1 and T2 =
0 VDC there is most likely an open circuit or failed op amp on control PCA itself.
If the voltage between T2 and T3 = 0 VDC and the voltage measured between T1 to T2 is
some voltage other than 0 VDC, the TEC is most likely shorted.
T4 is tied directly to ground. To determine the absolute voltage on any one of the other test points
make a measurement between that test point and T4.
Teledyne Analytical Instruments
303
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.7.15. O3 GENERATOR
The ozone generator can fail in two ways, electronically (printed circuit board) and functionally (internal
generator components). Assuming that air is supplied properly to the generator, the generator should
automatically turn on 30 minutes after the instrument is powered up or if the instrument is still warm.
See Section 13.2.3 for ozone generator functionality. Accurate performance of the generator can only be
determined with an ozone analyzer connected to the outlet of the generator. However, if the generator
appears to be working properly but the sensitivity or calibration of the instrument is reduced, suspect a
leak in the ozone generator supply air.
A leak in the dryer or between the dryer and the generator can cause moist, ambient air to leak into the
air stream, which significantly reduces the ozone output. The generator will produce only about half of
the nominal O3 concentration when run with moist, ambient air instead of dried air. In addition, moist
supply air will produce large amounts of nitric acid in the generator, which can cause analyzer
components downstream of the generator to deteriorate and/or causes significant deposit of nitrate
deposits on the reaction cell window, reducing sensitivity and causing performance drift. Carry out a
leak check as described earlier in this Section.
12.7.15.1. O3 Generator Override
This feature allows the user to manually turn the ozone generator off and on. This should be done before
disconnecting the generator, to prevent ozone from leaking out, or after a system restart if the user does
not want to wait for 30 minutes during warm-up time. To access this feature press the following buttons:
(Also note that the ozone generator does not turn on if the ozone flow conditions are out of specification
(e.g., if there is no flow through the system or the pump is broken).
Teledyne Analytical Instruments
304
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Note
This is one of the two settings in the DIAG menu that is retained after you
exit the menu.
12.7.16. INTERNAL SPAN GAS GENERATOR AND VALVE OPTIONS
The zero/span valves and Internal span gas generator options need to be enabled in the software (contact
the factory on how to do this).
Check for the physical presence of the valves or the IZS option.
Check front panel for correct software configuration. When the instrument is in SAMPLE mode, the
front panel display should show CALS and CALZ buttons in the second line of the display. The
presence of the buttons indicates that the option has been enabled in software. In addition, the IZS
option is enabled if the TEST functions show a parameter named IZS TEMP.
The semi-permeable PTFE membrane of the permeation tube is severely affected by humidity.
Variations in humidity between day and night are usually enough to yield very variable output results. If
the instrument is installed in an air-conditioned shelter, the air is usually dry enough to produce good
results. If the instrument is installed in an environment with variable or high humidity, variations in the
Teledyne Analytical Instruments
305
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
permeation tube output will be significant. In this case, a dryer for the supply air is recommended (dew
point should be –20° C or less).
The permeation tube of the internal span gas generator option is heated with a proportional heater circuit
and the temperature is maintained at 50°C ±1C. Check the front panel display or the IZS_TEMP
signal voltage using the SIGNAL I/O function under the DIAG Menu (Section 5.9.1). At 50° C, the
temperature signal from the IZS thermistor should be around 2500 mV.
12.7.17. TEMPERATURE SENSOR
12.7.17.1. Box Temperature Sensor
The box temperature sensor (thermistor) is mounted on the motherboard below the bottom edge of the
CPU board when looking at it from the front. It cannot be disconnected to check its resistance.
Box temperature will vary with, but will usually read about 5° C higher than, ambient (room)
temperature because of the internal heating zones from the NO2 converter, reaction cell and other
devices.
To check the box temperature functionality, we recommend checking the BOX_TEMP signal voltage
using the SIGNAL I/O function under the DIAG Menu (Section 12.1.3).
At about 30° C, the signal should be around 1500 mV.
To check the accuracy of the sensor, use a calibrated external thermometer / temperature sensor to
verify the accuracy of the box temperature by:
Placing it inside the chassis, next to the thermistor labeled XT1 (above connector J108) on
the motherboard.
Compare its reading to the value of the test function PMT TEMP.
12.7.17.2. PMT Temperature Sensor Control
The temperature of the PMT should be low and constant. It is more important that this temperature is
maintained at a constant level than it is to be a specific temperature.
The PMT cooler uses a Peltier, thermo-electric cooler element supplied with 12 V DC power from the
switching power supply PS2. The temperature is controlled by a proportional temperature controller
located on the preamplifier board.
Voltages applied to the cooler element vary from 0.1 to 12 VDC.
The temperature set point (hard-wired into the preamplifier board) will vary by ±2
The actual temperature will be maintained to within 0.1° C around that set point.
To check the operation of the PMT temperature control system:
1. Turn off the analyzer and let its internal components cool / heat to ambient
temperature.
2. Turn on the analyzer.
3. Set the front panel to show the PMT TEMP test function (see Section 4.1.1).
The temperature should fall steadily to 6-10° C.
Teledyne Analytical Instruments
306
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
If the temperature fails to reach this point after 60 minutes, there is a problem in
the cooler circuit.
If the control circuit on the preamplifier board is faulty, a temperature of –1° C
will be reported.
Teledyne Analytical Instruments
307
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
12.8. SERVICE PROCEDURES
This section contains some procedures that may need to be performed when a major component of the
analyzer requires repair or replacement.
Note
Maintenance procedures (e.g., replacement of regularly changed
expendables) are discussed in Section 11 (Instrument Maintenance) and
are not listed here).
Also, there may be more detailed service notes for some of the below
procedures. Contact Teledyne Customer Service Department.
WARNING – ELECTRICAL SHOCK HAZARD
Unless the procedure being performed requires the instrument be operating, turn it
off and disconnect power before opening the analyzer and removing, adjusting or
repairing any of its components or subsystems.
CAUTION – QUALIFIED TECHNICIAN
The operations outlined in this chapter are to be performed by qualified
maintenance personnel only.
12.8.1. DISK-ON-MODULE REPLACEMENT PROCEDURE
Note
Servicing of circuit components requires electrostatic discharge
protection, i.e. ESD grounding straps, mats and containers. Failure to use
ESD protection when working with electronic assemblies will void the
instrument warranty. Refer to Section 14 for more information on
preventing ESD damage.
Replacing the Disk-on-Module (DOM) will cause loss of all DAS data; it may also cause loss of some
instrument configuration parameters unless the replacement DOM carries the exact same firmware
version. Whenever changing the version of installed software, the memory must be reset. Failure to
ensure that memory is reset can cause the analyzer to malfunction, and invalidate measurements. After
the memory is reset, the A/D converter must be re-calibrated, and all information collected in Step 1
below must be re-entered before the instrument will function correctly. Also, zero and span calibration
should be performed.
1. Document all analyzer parameters that may have been changed, such as range,
auto-cal, analog output, serial port and other settings before replacing the DOM.
2. Turn off power to the instrument, fold down the rear panel by loosening the
mounting screws.
3. While looking at the electronic circuits from the back of the analyzer, locate the
Disk-on-Module in the right-most socket of the CPU board.
4. The DOM should carry a label with firmware revision, date and initials of the
programmer.
Teledyne Analytical Instruments
308
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
5. Remove the nylon standoff clip that mounts the DOM over the CPU board, and lift
the DOM off the CPU. Do not bend the connector pins.
6. Install the new Disk-on-Module, making sure the notch at the end of the chip
matches the notch in the socket.
7. It may be necessary to straighten the pins somewhat to fit them into the socket.
Press the chip all the way in.
8. Close the rear panel and turn on power to the machine.
9. If the replacement DOM carries a firmware revision, re-enter all of the setup
information.
12.8.2. O3 GENERATOR REPLACEMENT
The ozone generator is a black, brick-shaped device with printed circuit board attached to its rear and
two tubes extending out the right side in the front of the analyzer (see Figure 3-5). The board has a red
LED that, when lit, indicates ozone is being generated. To replace the ozone generator:
1. Turn off the analyzer power, remove the power cord and the analyzer cover.
2. Disconnect the 1/8” black tube from the ozone cleanser and the ¼” clear tube from
the plastic extension tube at the brass fitting nearest to the ozone generator.
3. Unplug the electrical connection on the rear side of the brick.
4. Unscrew the two mounting screws that attach the ozone generator to the chassis
and take out the entire assembly.
5. If you received a complete replacement generator with circuit board and mounting
bracket attached, simply reverse the above steps to replace the current generator.
Note
Ensure to carry out a leak check (11.3.12) and a recalibration after the
analyzer has warmed up for about 60 minutes.
12.8.3. SAMPLE AND OZONE (PERMA PURE®) DRYER REPLACEMENT
The 9110T standard configuration is equipped with a dryer for the ozone supply air. An optional dryer is
available for the sample stream and a combined dryer for both gas streams can also be purchased. To
change one or both of these dryers:
1. Turn off power to the analyzer and pump, remove the power cord and the analyzer
cover.
2. Locate the dryers in the center of the instrument, between sensor and NO2
converter (see Figure 3-5).
They are mounted to a bracket, which can be taken out when unscrewing the
two mounting screws (if necessary).
3. Disconnect all tubing that extends out of the dryer assembly.
Take extra care not to twist any of the white plastic fittings on the dryer.
Teledyne Analytical Instruments
309
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
These connect the inner drying tube to the outer purge tube and are delicate.
See Sections 13.3.1 and 11.3.2.
4. Note the orientation of the dryer on the bracket.
5. Cut the tie wraps that hold the dryer to the mounting bracket and take out the old
dryer.
If necessary, unscrew the two mounting screws on the bracket and take out the
entire assembly.
6. Attach the replacement dryer to the mounting bracket in the same orientation as the
old dryer.
7. Fix the dryer to the bracket using new tie wraps.
8. Cut off excess length of the wraps.
9. Put the assembly back into the chassis and tighten the mounting screws.
10. Re-attach the tubes to vacuum manifold, flow meter and/or NO/NOx valve using at
least two wrenches.
Take extra care not to twist the dryer’s white plastic fittings, as this will result in
large leaks that are difficult to trouble-shoot and fix.
11. Carry out a detailed leak check (see Section 11.3.12.2),
12. Close the analyzer.
13. Power up pump and analyzer and re-calibrate the instrument after it stabilizes.
12.8.4. PMT SENSOR HARDWARE CALIBRATION
The sensor module hardware calibration is used in the factory to adjust the slope and offset of the PMT
output and to optimize the signal output and HVPS.
If the instrument’s slope and offset values are outside of the acceptable range and all other more
obvious causes for this problem have been eliminated, the hardware calibration can be used to
adjust the sensor as has been done in the factory.
This procedure is also recommended after replacing the PMT or the preamplifier board.
To calibrate the PMT preamplifier PCA:
1. Perform a full zero point calibration using zero air (see Section 9).
2. Display the NOX STB test function on the front panel (Section 4.1.1).
3. Locate the preamplifier board (see Figure 3-5).
4. Locate the following components on the preamplifier board (Figure 12-8):
HVPS coarse adjustment switch (Range 0-9, then A-F).
HVPS fine adjustment switch (Range 0-9, then A-F).
Teledyne Analytical Instruments
310
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Gain adjustment potentiometer (Full scale is 10 turns).
5. Turn the gain adjustment potentiometer 12 turns clockwise or to its maximum
setting.
6. Feed NO gas into the analyzer.
This should be 90% of the upper limit setting for the 9110T’s reporting range:
EXAMPLE: if the reporting range is set at 500 ppb, use 450 ppb NO.
7. Wait until the STB value is below 0.5 ppb
Figure 12-8:
Pre-Amplifier Board Layout
8. Scroll to the NORM PMT test function on the analyzer’s front panel.
9. With the NO gas concentrations mentioned in Step 5 above, the norm pmt value
should be 900 mV.
10. Set the HVPS coarse adjustment to its minimum setting (0).
11. Set the HVPS fine adjustment switch to its maximum setting (F).
Set the HVPS coarse adjustment switch to the lowest setting that will give you
just above the target value for NORM PMT signal.
12. Adjust the HVPS fine adjustment such that the NORM PMT value is close to the
target value.
It may be necessary to go back and forth between coarse and fine adjustments
if the proper value is at the threshold of the min/max coarse setting.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
Do not overload the PMT by accidentally setting both adjustment switches
to their maximum setting. Start at the lowest setting and increment
Teledyne Analytical Instruments
311
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
slowly. Wait 10 seconds between adjustments.
Note
During these adjustments, the NORM PMT value will fluctuate as the
analyzer continues to switch between NO and NOx streams as well as
between measure and Auto Zero modes.
13. Perform a span point calibration (see Section 9) to normalize the sensor response
to its new PMT sensitivity.
14. Review the slope and offset values:
The slope values should be 1.000±0.300.
The offset values should be approximately 0.0 (-20 to +150 mV is allowed).
12.8.5. REPLACING THE PMT, HVPS OR TEC
The photo multiplier tube (PMT) should last for the lifetime of the analyzer, however, the high voltage
power supply (HVPS) or the thermo-electric cooler (TEC) components may fail. Replacing any of these
components requires opening the sensor module. This is a delicate assembly and it is recommend that
you ensure the PMT, HVPS or TEC modules are, indeed, faulty before unnecessarily opening of the
module.
CAUTION
QUALIFIED PERSONNEL
While the PMT or HVPS can be removed through the front panel without un-
mounting the entire sensor module, we recommend turning off the instrument,
opening its top cover and removing the entire assembly so that further repairs can
be carried out at an anti-ESD workstation.
Follow the guidelines defined in Section14for preventing electrostatic damage to
electronic components.
1. Turn OFF the analyzer and disconnect the power cord.
2. Remove the cover.
3. Disconnect all pneumatic and electrical connections from the sensor assembly.
4. Remove the sensor assembly.
5. If the TEC is to be replaced, remove the reaction cell assembly at this point by
unscrewing two holding screws.
This is necessary only if the repair being performed involves removing the PMT
cold block.
Teledyne Analytical Instruments
312
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Figure 12-9:
9110T Sensor Assembly
6. Remove the two connectors on the PMT housing end plate facing towards the front
panel.
7. Remove the end plate itself (4 screws with plastic washers).
Note
If the black PMT housing end plate for the Sensor Assembly is removed,
ensure to replace the 5 desiccant bags inside the housing.
8. Remove the dryer packages inside the PMT housing.
9. Unscrew the PMT assembly, which is held to the cold block by two plastic screws.
10. Discard the plastic screws and replace with new screws at the end of this procedure
(the threads get stripped easily and it is recommended to use new screws).
11. Along with the plate, slide out the OPTIC TEST LED and the thermistor that
measures the PMT temperature.
Thermistor will be coated with a white, thermal conducting paste.
Do not contaminate the inside of the housing with this grease, as it may
contaminate the PMT glass tube on re-assembly.
12. Carefully take out the assembly consisting of the HVPS, the gasket and the PMT.
Teledyne Analytical Instruments
313
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
13. Change the PMT or the HVPS or both, clean the PMT glass tube with a clean, anti-
static wipe and do not touch it after cleaning.
14. If the cold block or TEC is to be changed:
Disconnect the TEC driver board from the preamplifier board, remove the cooler
fan duct (4 screws on its side) including the driver board.
Disconnect the driver board from the TEC and set the sub-assembly aside.
15. Remove the end plate with the cooling fins (4 screws) and slide out the PMT cold
block assembly, which contains the TEC.
16. Unscrew the TEC from the cooling fins and the cold block and replace it with a new
unit.
17. Reassemble this TEC subassembly in reverse order.
Ensure to use thermal grease between TEC and cooling fins as well as between
TEC and cold block and that the side opening in the cold block will face the
reaction cell when assembled.
Evenly tighten the long mounting screws for good thermal conductivity.
CAUTION
QUALIFIED PERSONNEL
The thermo-electric cooler needs to be mounted flat to the heat sink.
If there is any significant gap, the TEC might burn out. Ensure to apply heat sink
paste before mounting it and tighten the screws evenly and cross-wise.
18. Reinsert the TEC subassembly in reverse order.
Ensure that the O-ring is placed properly and the assembly is tightened evenly.
19. Insert the LED and thermistor into the cold block, insert new drying packages and
carefully replace the end plate by making sure that the O-ring is properly in place.
Improperly placed O-rings will cause leaks, which – in turn – cause moisture to
condense on the inside of the cooler and likely cause a short in the HVPS.
20. Reinsert the PMT/HVPS subassembly in reverse order.
Don’t forget the gasket between HVPS and PMT.
Use new plastic screws to mount the PMT assembly on the PMT cold block.
21. Install new silica gel packets (desiccant bags).
22. Reconnect the cables and the reaction cell (evenly tighten these screws).
23. Replace the sensor assembly into the chassis and fasten with four screws and
washers.
24. Reconnect all electrical and pneumatic connections.
Teledyne Analytical Instruments
314
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
25. Leak check the system (see Section 13.3.12).
26. Turn ON the analyzer.
27. Verify the basic operation of the analyzer using the ETEST(12.7.12.2) and OTEST
features (12.7.12.1) or zero and span gases, then carry out a hardware calibration
of the analyzer followed by a zero/span point calibration (See Section 9.4.3.2).
12.8.6. REMOVING / REPLACING THE RELAY PCA FROM THE
INSTRUMENT
This is the most commonly used version of the Relay PCA. It includes a bank of solid state AC relays.
This version is installed in analyzers where components such as AC powered heaters must be turned ON
& OFF.
A retainer plate is installed over the relay to keep them securely seated in their sockets.
Retainer
Mounting
Screws
AC Relay
Retainer Plate
Figure 12-10: Relay PCA with AC Relay Retainer In Place
The Relay retainer plate installed on the relay PCA covers the lower right mounting screw of the relay
PCA. Therefore, when removing the relay PCA, the retainer plate must be removed first.
Teledyne Analytical Instruments
315
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
Mounting
Screws
AC Relay Retain Occludes
Mounting Screw on
P/N 045230200
Figure 12-11: Relay PCA Mounting Screw Locations
12.9. FREQUENTLY ASKED QUESTIONS
The following list was compiled from the TAI's Customer Service Department’s 10 most commonly
asked questions relating to the 9110T NOx Analyzer.
QUESTION
ANSWER
Why does the ENTR button
Sometimes the ENTR button will disappear if you select a setting that is
sometimes disappear on the front invalid or out of the allowable range for that parameter, such as trying to set
panel display?
the 24-hour clock to 25:00:00 or a range to less than 1 or more than 20000
ppb. Once you adjust the setting to an allowable value, the ENTR button
will re-appear.
Why is the ZERO or SPAN button The 9110T disables certain these buttons expected span or zero value
not displayed during calibration?
entered by the users is too different from the gas concentration actually
measured value at the time. This is to prevent the accidental recalibration
of the analyzer to an out-of-range response curve.
EXAMPLE: The span set point is 400 ppb but gas concentration being
measured is only 50 ppb.
How do I enter or change the
value of my Span Gas?
Press the CONC button found under the CAL or CALS buttons of the main
SAMPLE display menus to enter the expected NOx span concentration.
See Section 9.2.3.1 or for more information.
Can I automate the calibration of
my analyzer?
Any analyzer with zero/span valve or IZS option can be automatically
calibrated using the instrument’s AutoCal feature.
Teledyne Analytical Instruments
316
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
QUESTION
ANSWER
Can I use the IZS option to
calibrate the analyzer?
Yes. However, the accuracy of the IZS option’s permeation tube is only
±5%. Whereas this may be acceptable for basic calibration checks, the IZS
option is not permitted as a calibration source in applications following US
EPA protocols.
To achieve highest accuracy, it is recommended to use cylinders of
calibrated span gases in combination with a zero air source.
How do I measure the sample
flow?
Sample flow is measured by attaching a calibrated flow meter to the sample
inlet port when the instrument is operating. The sample flow should be 500
cm³/min 10%.
Section 13.3.12.3 includes detailed instructions on performing a check of
the sample gas flow.
Can I use the DAS system in
place of a strip chart recorder or
data logger?
Yes. Section 7 describes the setup and operation of the DAS system in
detail.
How often do I need to change
the particulate filter?
Once per week or as needed. Section 11 contains a maintenance schedule
listing the most important, regular maintenance tasks. Highly polluted
sample air may require more frequent changes.
How long does the sample pump
last?
The sample pump should last one to two years and the pump head should
be replaced when necessary. Use the RCEL pressure indicator on the front
panel to see if the pump needs replacement.
If this value goes above 10 in-Hg-A, on average, the pump head needs to
be rebuilt.
Why does my RS-232 serial
connection not work?
There are several possible reasons:
The wrong cable: please use the provided or a generic “straight-
through” cable (do not use a “null-modem” type cable) and ensure the
pin assignments are correct (Sections 3.3.1.8 and 6.3).
The DCE/DTE switch on the back of the analyzer is not set properly;
ensure that both green and red lights are on (Section 6.1).
The baud rate of the analyzer’s COMM port does not match that of the
serial port of your computer/data logger (Section 6.2.2).
How do I make the instrument’s
display and my data logger
agree?
This most commonly occurs when an independent metering device is used
besides the data logger/recorder to determine gas concentration levels
while calibrating the analyzer. These disagreements result from the
analyzer, the metering device and the data logger having slightly different
ground levels.
Use the data logger itself as the metering device during calibration
procedures.
Teledyne Analytical Instruments
317
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Troubleshooting & Service
QUESTION
ANSWER
Do the critical flow orifices of my
analyzer require regular
replacement?
No. The o-rings and the sintered filter associated with them require
replacement once a year, but the critical flow orifices do not.
See Section 11 for instructions.
See Section 3.3.1.6.
How do I set up and use the
Contact Closures (Control Inputs)
on the Rear Panel of the
analyzer?
12.10. TECHNICAL ASSISTANCE
If this manual and its troubleshooting & service section do not solve your problems, technical assistance
may be obtained from Teledyne’s Customer Service Department at:
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
Telephone: (626) 934-1500
Fax: (626) 961-2538
Web: www.teledyne-ai.com
or your local representative.
Before you contact Teledyne Customer Service, fill out the problem report form in Appendix C, which is
also available online for electronic submission at http://www.teledyne-api.com/manuals/.
Teledyne Analytical Instruments
318
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13. PRINCIPLES OF OPERATION
The 9110T Nitrogen Oxides Analyzer is a microprocessor controlled instrument that determines the
concentration of nitric oxide (NO), total nitrogen oxides (NOX, the sum of NO and NO2) and nitrogen
dioxide (NO2) in a sample gas drawn through the instrument.
It requires that sample and calibration gases be supplied at ambient atmospheric pressure in order
to establish a constant gas flow through the reaction cell where the sample gas is exposed to ozone
(O3), initiating a chemical reaction that gives off light (hv).
The instrument measures the amount of chemiluminescence to determine the amount of NO in the
sample gas.
A catalytic-reactive converter converts NO2 in the sample gas to NO which, along with the NO
present in the sample is reported as NOX. NO2 is calculated as the difference between NOX and
NO.
Calibration of the instrument is performed in software and usually does not require physical adjustments
to the instrument. During calibration, the microprocessor measures the sensor output signal when gases
with known amounts of NO or NO2 are supplied and stores these results in memory. The microprocessor
uses these calibration values along with the signal from the sample gas and data of the current
temperature and pressure of the gas to calculate a final NOX concentration.
The concentration values and the original information from which it was calculated are stored in the
unit’s internal data acquisition system (DAS Section 7) and are reported to the user through a vacuum
fluorescence display or several output ports.
13.1. MEASUREMENT PRINCIPLE
13.1.1. CHEMILUMINESCENCE CREATION IN THE 9110T REACTION CELL
The 9110T’s measures the amount of NO present in a gas by detecting the chemiluminescence which
occurs when nitrogen oxide (NO) is exposed to ozone (O3). This reaction is a two-step process:
In the first step, one molecule of NO and one molecule of O3 collide and chemically react to produce
one molecule of oxygen (O2) and one molecule of nitrogen dioxide (NO2). Some of the NO2
molecules created by this reaction retain excess energy from the collision and exist in an excited
state, where one of the electrons of the NO2 molecule resides in a higher energy state than normal
(denoted by an asterisk in the following equation).
Equation 13-1
NO O3 NO2* O2
The second step occurs because the laws of thermodynamics require that systems seek the lowest
stable energy state available, therefore the excited NO2 molecule quickly returns to its ground state,
releasing the excess energy. This release takes the form of a quantum of light (h). The distribution
of wavelengths for these quanta range between 600 and 3000 nm, with a peak at about 1200 nm.
Equation 13-2
Teledyne Analytical Instruments
319
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
NO2* NO h1200nm
2
All things being constant (temperature, pressure, amount of ozone present, etc.), the relationship between
the amount of NO present in the reaction cell and the amount of light emitted from the reaction is very
linear. If more NO is present, more IR light is produced. By measuring the amount of IR light produced
with a sensor sensitive in the near-infrared spectrum (see Figure 13-2) the amount of NO present can be
determined.
In addition, sometimes the excited NO2 collides with other gaseous molecules in the reaction cell
chamber or even the molecules of the reaction cell walls and transfers its excess energy to this collision
partner (represented by M in the equation 12-3 below) without emitting any light at all. In fact, by far
the largest portion of the excited NO2 returns to the ground state this way, leaving only a few percent
yield of usable chemiluminescence.
Equation 13-3
NO2* M NO2 M
The probability of a collision between the NO2* molecule and a collision partner M increases
proportionally with the reaction cell pressure. This non-radiating collision with the NO2* molecules is
usually referred to as third body quenching, an unwanted process further described in Section 13.1.5.2.
Even under the best conditions only about 20% of the NO2 that is formed by the reaction described in
equation 12-1 is in the excited state. In order to maximize chemiluminescence, the reaction cell is
maintained at reduced pressure (thereby reducing the amount of available collision partners) and is
supplied with a large, constant excess of ozone (about 3000-5000 ppm) from the internal ozone
generator.
13.1.2. CHEMILUMINESCENCE DETECTION IN THE 9110T REACTION
CELL
13.1.2.1. The Photo Multiplier Tube (PMT)
The 9110T uses a special kind of vacuum tube, called a photo-multiplier tube (PMT), to detect the
amount of light created by the NO and O3 reaction in the reaction cell.
Photons enter the PMT and strike a negatively charged photo cathode causing it to emit electrons. These
electrons are accelerated by an applied high voltage and multiplied through a sequence of similar
acceleration steps (dynodes) until a useable current signal is generated (see Section 13.5 for a more
detailed description). The more light present (in this case photons given off by the chemiluminescent
reaction described above), the more current is produced. Therefore the more NO present in the reaction
cell the more current is produced by the PMT.
The current produced by the PMT is converted to a voltage and amplified by the preamplifier board and
then communicated to the 9110T’s CPU via the A D converter circuitry on the analyzer.
13.1.2.2. Optical Filter
A high pass optical filter, only transparent to wavelengths of light above 645nm, placed between the
reaction cell and the PMT (see Figure 13-1) in conjunction with the response characteristics of the PMT
creates a very narrow window of wavelengths of light to which the 9110T will respond.
Teledyne Analytical Instruments
320
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-1:
Reaction Cell with PMT Tube and Optical Filter
The narrowness of this band of sensitivity allows the 9110T to ignore extraneous light and radiation that
might interfere with the 9110T’s measurement. For instance, some oxides of sulfur can also be
chemiluminescent emitters when in contact with O3 but give off light at much shorter wavelengths
(usually around 260nm to 480nm).
140
Visible
Infrared
Spectrum
120
100
80
NO + O3 Emission Spectrum
PMT
Response
60
40
20
0
Optical Hi-Pass Filter Performance
0.5
0.7
0.9
1.1
1.3
1.5
1.7
1.9
Wavelength (micrometers)
Area of Sensitivity
Figure 13-2:
9110T Sensitivity Spectrum
13.1.3. NOX AND NO2 DETERMINATION
The only gas that is actually measured by the 9110T is NO. NO2, and therefore NOx (which is defined
here as the sum of NO and NO2 in the sample gas), contained in the gas is not detected because NO2
does not react with O3 to create chemiluminescence.
Teledyne Analytical Instruments
321
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
In order to measure the concentration of NO2, and therefore the concentration of NOx, the 9110T
periodically switches the sample gas stream so that the pump pulls it through a special converter
cartridge filled with molybdenum (Mo, “moly”) chips that are heated to a temperature of 315°C.
Figure 13-3:
NO2 NO Conversion
The heated molybdenum reacts with NO2 in the sample gas and produces a NO gas and a variety of
molybdenum.
Equation 13-4
xNO yMoxNO MyOz (at 315C)
2
Once the NO2 in the sample gas has been converted to NO, it is routed to the reaction cell where it
undergoes the chemiluminescence reaction described in Equation 13-1 and Equation 13-2.
By converting the NO2 in the sample gas into NO, the analyzer can measure the total NOX) content of
the sample gas (i.e. the NO present + the converted NO2 present). By switching the sample gas stream
in and out of the “moly” converter every 6 - 10 seconds, the 9110T analyzer is able to quasi-
continuously measure both the NO and the total NOX content.
Finally, the NO2 concentration is not directly measured but calculated by subtracting the known NO
content of the sample gas from the known NOX content.
13.1.4. AUTO ZERO
Inherent in the operation of any PMT is a certain amount of noise. This is due to a variety of factors
such as black body infrared radiation given off by the metal components of the reaction cell, unit to unit
variations in the PMT units and even the constant universal background radiation that surrounds us at all
times. In order to reduce this amount of noise and offset, the PMT is kept at a constant 7° C (45° F) by a
Thermo-Electric Cooler (TEC).
While this intrinsic noise and offset is significantly reduced by cooling the PMT, it is not eradicated. To
determine how much noise remains, once every minute for about 8 seconds the 9110T diverts the sample
gas flow directly to the vacuum manifold without passing the reaction cell.
Teledyne Analytical Instruments
322
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
During this time, only O3 is present in the reaction cell, effectively turning off the chemiluminescence
reaction. Once the chamber is completely dark, the 9110T records the output of the PMT and keeps a
running average of these AZERO values. This average offset value is subtracted from the raw PMT
readings while the instrument is measuring NO and NOX to arrive at an Auto Zero corrected reading.
Figure 13-4:
Pneumatic Flow During the Auto Zero Cycle
13.1.5. MEASUREMENT INTERFERENCES
It should be noted that the chemiluminescence method is subject to interferences from a number of
sources. The 9110T has been successfully tested for its ability to reject interference from most of these
sources. Table 13-1 list the most common types of interferents that could affect the performance of your
9110T.
13.1.5.1. Direct Interference
Some gases can directly alter the amount of light detected by the PMT due to chemiluminescence in the
reaction cell. This can either be a gas that undergoes chemiluminescence by reacting with O3 in the
reaction cell or a gas that reacts with other compounds and produces excess NO upstream of the reaction
cell.
Teledyne Analytical Instruments
323
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.1.5.2. Third Body Quenching
*
As described by Equation 13-3, other molecules in the reaction cell can collide with the excited NO2 ,
*
causing the excited NO2 to return to its ground state without releasing a photon of light. This is known
as third party quenching.
Quenching is an unwanted phenomenon and the extent to which it occurs depends on the properties of
the collision partner.
Larger, more polarized molecules such as H2O and CO2 are the most significant quenching
interferents of NO chemiluminescence.
The influence of water vapor on the 9110T measurement can be eliminated with an optional,
internal sample gas dryer (see Section 3.3.2.6).
The interference of varying CO2 amounts at low concentrations (less that 0.5%) is
negligible.
In cases with excessively high CO2 concentrations (larger than 0.5%), the effect can be
calibrated out by using calibration gases with a CO2 content equal to the measured air.
Only very high and highly variable CO2 concentrations will then cause a measurable
interference. For those applications, it is recommended to use other analyzer models.
Please consult TAI Sales Department or our website (see Section 12.10).
Smaller less polar and electronically “harder” molecules such as N2 and O2 can cause interference
of this type as well, however, the concentrations of N2 and O2 are virtually constant in ambient air
measurements, hence provide a constant amount of quenching that is accounted for in the
calibration of the instrument .
13.1.5.3. Light Leaks
The 9110T sensitivity curve includes a small portion of the visible light spectrum (see Figure 13-2),
therefore it is important to ensure that the reaction cell is completely sealed with respect to light. To
ensure this:
All pneumatic tubing leading into the reaction cell is opaque in order to prevent light from entering
the cell.
Light penetration is prevented by stainless steel filters and orifices.
Teledyne Analytical Instruments
324
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Table 13-1: List of Interferents
Gas
Interference Type
Rejection Method
Dilution: Viscosity of CO2 molecules causes them to
collect in aperture of Critical Flow Orifice altering flow
rate of NO.
3rd Body Quenching: CO2 molecules collide with
NO2* molecules absorbing excess energy kinetically
and preventing emission of photons.
If high concentrations of CO2 are suspected,
special calibration methods must be performed to
account for the affects of the CO2.
CO2
Contact TAI’s Customer Service Department (see
Section 12.10) for details.
Some SOX variants can also initiate a
chemiluminescence reaction upon exposure to O3
producing excess light.
Wavelengths of light produced by
chemiluminescence of SOX are screened out by
the Optical Filter.
Chemically reacts with NH3, O2 and H2O in O3
generator to create (NH3)2SO4 (ammonium sulfate)
and NH3NO2 (ammonium nitrate) which form opaque
white deposits on optical filter window. Also forms
highly corrosive HNO3 (Nitric Acid)
Most of the ammonium sulfate and ammonium
nitrate produced is removed from the sample gas
by an air purifier located between the O3
Generator and the reaction cell.
SOX
If high concentrations of SOX are suspected,
special calibration methods must be performed to
account for the affects of the SO2.
3rd Body quenching: SOX molecules collide with NO2*
molecules absorbing excess energy kinetically and
preventing emission of photons.
Contact Teledyne Customer Service Department
(see Section 12.10) for details.
3rd Body quenching: H2O molecules collide with NO2* Analyzer’s operating in high humidity areas must
molecules absorbing excess energy kinetically and
preventing emission of light.
have some drying applied to the sample gas (see
Section 3.3.2.6 for more details).
Water also reacts with NH3 and SOX in the O3
generator to create (NH3)2SO4 (ammonium sulfate)
and NH3NO2 (ammonium nitrate) which form opaque
white deposits on the optical filter window. This also
forms highly corrosive HNO3 (nitric acid)
Water is effectively removed from the O3 gas
stream by the Perma Pure® Dryer (Section
13.2.3.2 for more details). We offer several
Perma Pure® dryers for the sample stream (see
Section 3.3.2.6 for more details).
H2O
Direct Interference: NH3 is converted to H2O and NO
by the NO2 converter. Excess NO reacts with O3 in
the reaction cell creating a chemiluminescence
artifact.
If a high concentration of NH3 is suspected, steps
must be taken to remove the NH3 from the sample
gas prior to its entry into the NO2 converter (see
Section 3.3.2.6 for more details).
NH3
NH3 also reacts with H2O, O2 and SOX in the O3
generator to create (NH3)2SO4 (ammonium sulfate)
and NH3NO2 (ammonium nitrate) which form opaque
white deposits on optical filter window. Also forms
highly corrosive HNO3 (nitric acid).
The Perma Pure® dryer built into the 9110T is
sufficient for removing typical ambient
concentration levels of NH3.
13.1.5.4. Reaction Cell Temperature Control
The stability of the chemiluminescence reaction between NO and O3 can be affected by changes in the
temperature and pressure of the O3 and sample gases in the reaction cell. In order to reduce temperature
effects, the reaction cell is maintained at a constant 50 C, just above the high end of the instrument’s
operation temperature range.
Two AC heaters, one embedded into the bottom of the reaction cell, the other embedded directly above
the chamber’s exhaust fitting, provide the heat source. These heaters operate off of the instrument’s
main AC power and are controlled by the CPU through a power relay on the relay board (see Section
13.3.4.4).
A thermistor, also embedded in the bottom of the reaction cell, reports the cell’s temperature to the CPU
through the thermistor interface circuitry of the motherboard (see Section 13.3.3.3).
Teledyne Analytical Instruments
325
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
13.2. PNEUMATIC OPERATION
IMPORTANT
Principles of Operation
IMPACT ON READINGS OR DATA
Could either affect accuracy of instrument readings or cause loss of data.
Note
The sample gas is the most critical flow path in the analyzer. At any point
before and in the reaction cell, the integrity of the sample gas cannot be
compromised. Therefore, it is important that the sample airflow system is
both leak tight and not pressurized over ambient pressure.
Regular leak checks should be performed on the analyzer as presented in the maintenance schedule,
Table 13-1. Procedures for correctly performing leak checks can be found in Section 13.3.12.
13.2.1. SAMPLE GAS FLOW
Note
In this section of the manual vacuum readings are given in inches of
mercury absolute (In-Hg-A). This pressure value is referenced against
zero (a perfect vacuum).
The gas flow for the 9110T is created by a pump that is pneumatically downstream from the rest of the
instrument’s components. This is either:
An external pump pneumatically connected to the analyzer’s exhaust port located on the rear panel.
This is the most common configuration for the 9110T (see Figure 13-5), or;
An optional internal pump pneumatically connected between the vacuum manifold and the exhaust
outlet. (see Figure 13-6).
In either case the pump creates a vacuum of approximately 5 in-Hg-A at one standard liter/minute,
which is provided to various pneumatic components by a vacuum manifold located just in front of the
rear panel (see Figure 3-5).
Gas flow is created by keeping the analyzer’s sample gas inlet near ambient pressure, usually by means
of a small vent installed in the sample line at the inlet, in effect pulling the gas through the instrument’s
pneumatic systems.
By placing the pump downstream from the analyzer’s reaction cell, several problems are avoided.
First, the pumping process heats and compresses the sample air complicating the measurement
process.
Additionally, certain physical parts of the pump itself are made of materials that might chemically
react with the sample gas.
Finally, in certain applications where the concentration of the target gas might be high enough to be
hazardous, maintaining a negative gas pressure relative to ambient means that should a minor leak
occur, no sample gas would be pumped into the atmosphere surrounding the analyzer.
Teledyne Analytical Instruments
326
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-5:
Internal Gas Flow for Basic 9110T with External Pump
Teledyne Analytical Instruments
327
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-6:
Basic Internal Gas Flow for Basic 9110T with Internal Pump
13.2.1.1. Vacuum Manifold
The vacuum created by the analyzer’s pump is supplied to all of the gas streams for the 9110T analyzer
through the vacuum manifold (also called the exhaust manifold).
Teledyne Analytical Instruments
328
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-7.
Vacuum Manifold, Standard Configuration
Configurations will vary depending on the optional equipment that is installed. For example:
An 9110T with the optional internal span gas generator installed will add another FT8 connector and
orifice assembly to the manifold where the FT28 fitting is shown in the above drawing.
An optional sample gas dryer will add a Tee-fitting so that two ¼” tubes can be connected to the
same port.
13.2.1.2. Sample Gas Flow Valves and Routing
As discussed in Section 13.1, the measurement of NOx, NO and NO2 requires that the sample gas flow
cycles through different routes that include and exclude various scrubbers and converters. There are
several valves that perform this function:
The NO/NOX valve directs the sample gas either directly to the reaction cell or through the unit’s
NO2 converter, alternating every ~8 sec.
The Auto Zero valve directs the sample gas stream to completely bypass the reaction cell for dark
noise measurement once every minute, which is then subtracted as a measurement offset from the
raw concentration signal.
Table 13-2: 9110T Valve Cycle Phases
NO/ NOX
Valve
Status
Auto Zero
Valve
Status
Time
Index
Phase
Activity
Figure
Wait period (NO dwell time). Ensures reaction cell has
been flushed of previous gas.
Open to
Auto Zero
valve
0 - 2 s
NO
Measure
Open to
reaction cell
Figure
13-3
2 - 4 s Analyzer measures chemiluminescence in reaction cell.
Wait period (NOX dwell time). Ensures reaction cell has
been flushed of previous gas.
4 – 6 s
Open to
NO2
converter
NOX
Measure
Open to
reaction cell
Figure
13-3
Analyzer measures NO + O3 chemiluminescence in
reaction cell.
6 – 8 s
Cycle repeats every ~8 seconds
Teledyne Analytical Instruments
329
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Wait period (AZERO dwell time). Ensures reaction cell
0 – 4 s has been flushed of sample gas and chemi-
luminescence reaction is stopped.
Open to
Auto Zero
valve
Open to
vacuum
manifold
Auto
Zero
Figure
13-4
Analyzer measures background noise without sample
4 - 6 s
gas
Cycle repeats every minute
13.2.2. FLOW RATE CONTROL - CRITICAL FLOW ORIFICES
Sample gas flow in the 9110T analyzer is created via the use of several flow control assemblies (see
Figure 13-8 for an example) located in various places in the gas streams of the instrument. These
assemblies consist of:
a critical flow orifice
two o-rings, Located just before and after the critical flow orifice, the o-rings seal the gap between
the walls of assembly housing and the critical flow orifice
a sintered filter
a spring (applies mechanical force needed to form the seal between the o-rings, the critical flow
orifice and the assembly housing)
Figure 13-8:
Flow Control Assembly & Critical Flow Orifice
13.2.2.1. Critical Flow Orifice
The most important component of each flow control assembly is the critical flow orifice. Critical flow
orifices are a simple means to regulate stable gas flow rates. They operate without moving parts by
taking advantage of the laws of fluid dynamics. By restricting the flow of gas through the orifice, a
pressure differential is created. This pressure differential, created by the analyzer’s external pump,
draws the gas through the orifice.
As the pressure on the downstream side of the orifice (the pump side) continues to drop, the speed that
the gas flows though the orifice continues to rise. Once the ratio of upstream pressure to downstream
pressure is greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound. As
long as that ratio stays at least 2:1, the gas flow rate is unaffected by any fluctuations, surges, or changes
in downstream pressure because such variations only travel at the speed of sound themselves and are
therefore cancelled out by the sonic shockwave at the downstream exit of the critical flow orifice.
Teledyne Analytical Instruments
330
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the size and
shape of the aperture in the orifice. The larger the hole, the more gas molecules (moving at the speed of
sound) pass through the orifice.
In addition to controlling the gas flow rates into the reaction cell, the two critical flow orifices at the
inlets of the reaction cell also maintain an under-pressure inside it, effectively reducing the number of
molecules in the chamber and the corresponding incidence of third body quenching (see Section
13.1.5.2) and therefore increasing the chemiluminescence yield.
The 9110T reaches its peak sensitivity at about 2 in-Hg-A, below which the sensitivity drops due to
there being too few molecules present and a corresponding decrease in chemiluminescence.
13.2.2.2. Locations and Descriptions of Critical Flow Orifices Inside the 9110T
The 9110T uses several of the following critical flow orifices (Figure 13-9) to create and maintain the
proper flow rate of gas through its various components. (Please note that not all features displayed in
Figure 13-9 are standard components of 9110T analyzers).
NO/NOX
VALVE
SAMPLE/ CAL
VALVE
ZERO/SPAN
VALVE
SAMPLE
PRESSURE
SENSOR
VACUUM
PRESSURE
SENSOR
O3 FLOW
SENSOR
AUTOZERO
VALVE
PMT
PURE
DRYER
Figure 13-9:
Location of Flow Control Assemblies & Critical Flow Orifices
Table 13-3: 9110T Gas Flow Rates
Flow rate
(nominal)
500 cm³/min
Orifice
Diameter
Location
Purpose
Sample gas inlet of reaction cell
0.010” (0.25 mm)
Controls rate of flow of sample gas into the
Teledyne Analytical Instruments
331
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
reaction cell.
Controls rate of flow of ozone gas into the
reaction cell.
O3 supply inlet of reaction cell
0.004” (0.10 mm)
80 cm³/min
80 cm³/min
Controls flow rate of dry air return / purge
air of the dryer.
Dry air return of Perma Pure® dryer
0.004” (0.10 mm)
0.010” (0.25 mm)
Controls rate of sample gas flow when
bypassing the reaction cell during the Auto
Zero cycle.
Vacuum manifold, Auto Zero port.
500 cm³/min
60 cm³/min
Controls rate of flow of zero purge gas
through the optional Internal span gas
generator when it is installed.
Vacuum manifold, Internal span gas
generator exhaust port
0.003” (0.10 mm)
The necessary 2:1 ratios across the critical flow orifices is largely exceeded by the pumps supplied with
the analyzer which are designed to accommodate a wide range of possible variability in atmospheric
pressure and age related degradation of the pump itself. Once the pump does degrade the ratio between
sample and vacuum pressures may fall to less than 2:1. At this point, the instrument will display an
invalid sample flow rate measurement (XXXX).
Note
The diameter of a critical flow orifice may change with temperature
because of expansion of the orifice material and, hence, the most crucial
critical flow orifices in the 9110T (those controlling the sample gas and O3
flow into the cell itself) are located in the reaction cell where they can be
maintained at a constant temperature.
13.2.3. OZONE GAS GENERATION AND AIR FLOW
The excess ozone needed for reaction with NO in the reaction cell is generated inside the analyzer
because of the instability and toxicity of ozone. Besides the ozone generator itself, this requires a dry air
supply and filtering of the gas before it is introduced into the reaction cell.
Due to its toxicity and aggressive chemical behavior, O3 must also be removed from the gas stream
before it can be vented through the exhaust outlet.
CAUTION
GENERAL SAFETY HAZARD
Ozone (O3) is a toxic gas.
Obtain a Material Safety Data Sheet (MSDS) for this gas. Read and
rigorously follow the safety guidelines described there.
Always ensure that the plumbing of the O3 generation and supply system
is maintained and leak-free.
13.2.3.1. The O3 Generator
The 9110T uses a dual-dielectric, Corona Discharge (CD) tube for creating its O3, which is capable of
producing high concentrations of ozone efficiently and with low excess heat (see Figure 13-10). The
primary component of the generator is a glass tube with hollow walls of which the outermost and
innermost surfaces are coated with electrically conductive material.
Air flows through the glass tube, between the two conductive coatings, in effect creating a capacitor with
the air and glass acting as the dielectric. The layers of glass also separate the conductive surfaces from
Teledyne Analytical Instruments
332
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
the air stream to prevent reaction with the O3. As the capacitor charges and discharges, electrons are
created and accelerated across the air gap and collide with the O2 molecules in the air stream splitting
them into elemental oxygen.
Some of these oxygen atoms recombine with O2 to O3. The quantity of ozone produced is dependent on
factors such as the voltage and frequency of the alternating current applied to the CD cells. When
enough high-energy electrons are produced to ionize the O2 molecules, a light emitting, gaseous plasma
is formed, which is commonly referred to as a corona, hence the name corona discharge generator.
Figure 13-10: Ozone Generator Principle
13.2.3.2. Ozone Generator Dry Air Supply
Ambient air usually contains enough water vapor to greatly diminishes the yield of ozone produced by
the ozone generator. Water also reacts with chemicals inside the O3 Generator to produce caustic
substances such as ammonium sulfate or highly corrosive nitric acid that will damage the optical filter
located between the reaction cell and the PMT.
To prevent this, the air supply for the O3 generator is dried using a special Perma Pure® single tube
permeation dryer. The dryer consists of a single tube of Nafion® that is mounted within an outer,
flexible plastic tube. Nafion® is a co-polymer that absorbs water very well but not most other chemicals.
As gas flows through the inner Nafion® tube, water vapor is absorbed into the membrane walls. The
absorbed water is transported through the membrane wall and evaporated into the dry purge gas flowing
through the outer tube, countercurrent to the gas in the inner tube.
Teledyne Analytical Instruments
333
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-11: Semi-Permeable Membrane Drying Process
The process by which the water vapor molecules are collected and transported through Nafion® material
is called per-evaporation and is driven by the humidity gradient between the inner and outer tubes as
well as the flow rates and pressure difference between inner and outer tubing. Unlike micro-porous
membrane permeation, which transfers water through a relatively slow diffusion process, per-
evaporation is a simple kinetic reaction. Therefore, the drying process occurs quickly, typically within
milliseconds.
Because this chemical reaction is based on hydrogen bonds between the water molecule and the Nafion®
material most other chemical components of the gas to be dried are usually unaffected. Specifically, the
gases of interest for the 9110T, NO and NO2, do not get absorbed and pass the dryer unaltered.
On the other hand, other small polar gases that are capable of hydrogen bonds such as ammonia (NH3)
can be absorbed this way, too. This is an advantage since gases such as NH3 can cause interference for
the measurement of NOx, NO and NO2 (see Table 13-1).
Figure 13-12: 9110T Perma Pure® Dryer
Teledyne Analytical Instruments
334
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
To provide a dry purge gas for the outer side of the Nafion tube, the 9110T returns some of the dried air
from the inner tube to the outer tube. This means that any time the analyzer is turned on after having
been OFF for 30 minutes or more, the humidity gradient between the inner and outer tubes is not very
large and the dryer’s efficiency is low. Since it takes a certain amount of time for the humidity gradient
to become large enough for the Perma Pure® Dryer operate efficiently, in such cold start cases the O3
Generator is not turned on until 30 minutes has passed in order to ensure that it is not operating until its
air supply is properly dry. During this 30 minute duration the O3 GEN OVERRIDE menu displays
“TMR” on the front panel screen.
Note
When rebooting the instrument within less than 30 minutes of power-
down, the generator is turned on immediately.
The Perma Pure® Dryer used in the 9110T is capable of adequately drying ambient air to a dew point of
≤ -5˚C (~4000 ppm residual H2O) at a flow rate of 1 standard liter per minute (slpm) or down to ≤ -15˚C
(~1600 ppm residual H2O) at 0.5 slpm. The Perma Pure® Dryer is also capable of removing ammonia
from the sample gas up to concentrations of approximately 1 ppm.
13.2.3.3. Ozone Supply Air Filter
The 9110T uses ambient air as the supply gas for the O3 generator and may produce a variety of
byproducts. Small amounts of water, ammonia and various sulfur oxides can combine to create
ammonium sulfate, ammonium nitrate, nitric acid and other compounds. Whereas sulfates and nitrates
can create powdery residues inside the reaction cell causing sensitivity drift, nitric acid is a very
aggressive compound, which can deteriorate the analyzer’s components. In order to remove these
chemical byproducts from the O3 gas stream, the output of the O3 generator flows through a special filter
between the generator and the reaction cell.
The small amount of NOX produced in the generator (from the reaction of O2 or O3 and N2 in the air) will
not affect the 9110T’s ability to measure NOx, NO and NO2 as it is accounted for and removed from the
concentration calculations by the analyzer’s Auto Zero feature (see Section 13.1.4).
13.2.3.4. Ozone Destruct
Even though ozone is unstable and typically reacts to form O2, the break-down is not quite fast enough to
ensure that it is completely removed from the exhaust gas stream of the 9110T by the time the gas exits
the analyzer. Due to the high toxicity and reactivity of O3, a highly efficient catalytic converter scrubs or
converts all of the O3 from the gas exiting the reaction cell. The conversion process is very safe. It only
converts ozone to oxygen and does not produce any toxic or hazardous gases.
The O3 destruct is located just inside the NO2 converter. As this is a true catalytic converter, there are no
maintenance requirements as would be required for charcoal-based ozone destructs.
A certain amount of fine, black dust may exit the catalyst, particularly if the analyzer is subjected to
sudden pressure drops (for example, when disconnecting the running pump without letting the analyzer
properly and slowly equilibrate to ambient pressure). To prevent the dust from entering the reaction cell
or the pump, the ozone destruct is equipped with a quartz wool filter material.
13.2.4. PNEUMATIC SENSORS
Note
The 9110T displays all pressures in inches of mercury absolute (in-Hg-A),
i.e. absolute pressure referenced against zero (a perfect vacuum).
Teledyne Analytical Instruments
335
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
The 9110T uses three pneumatic sensors to verify the flow and pressure levels of its gas streams. They
are located on a printed circuit assembly, called the pneumatic pressure/flow sensor board, located just
behind the sensor assembly. The measurements made by these sensors are used for a variety of
important calculations and diagnostics.
13.2.4.1. Sample Pressure Sensor
An absolute pressure transducer connected to the input of the NO/NOX valve is used to measure the
pressure of the sample gas before it enters the analyzer’s reaction cell.
In conjunction with the measurement made by the vacuum pressure sensor, this “upstream”
measurement is used to compute the sample gas sample flow rate and to validate the critical flow
condition (2:1 pressure ratio) through the sample gas critical flow orifice (Section 13.2.2).
If the Temperature/Pressure Compensation (TPC) feature is turned on (Section 13.9.2), the output
of this sensor is also used to supply pressure data for that calculation.
The actual pressure value is viewable through the analyzer’s front panel display as the test function
SAMP.
The flow rate of the sample gas is displayed as SAMP FLW and the SIGNAL I/O function
SAMPLE_PRESSURE.
13.2.4.2. Vacuum Pressure Sensor
An absolute pressure transducer connected to the exhaust manifold is used to measure the pressure
downstream from and inside the instrument’s reaction cell.
The output of the sensor is used by the CPU to calculate the pressure differential between the gas
upstream of the reaction cell and the gas downstream from it and is also used as the main
diagnostic for proper pump operation.
If the ratio between the upstream pressure and the downstream pressure falls below 2:1, a warning
message (SAMPLE FLOW WARN) is displayed on the analyzer’s front panel (see Section 4.1.2)
and the sample flow rate will display XXXX instead of an actual value.
If this pressure exceeds 10 in-Hg-A, an RCEL PRESS WARN is issued, even though the analyzer
will continue to calculate a sample flow up to ~14 in Hg.
If the Temperature/Pressure Compensation (TPC) feature is turned on (see Section 13.9.2), the
output of this sensor is also used to supply pressure data for that calculation.
This measurement is viewable through the analyzer’s front panel as the test function RCEL and the
SIGNAL I/O function RCELL_PRESSURE.
13.2.4.3. Sample Gas Flow Calculation
Sample gas flow in the 9110T analyzer is not a directly measured value, but is rather calculated based on
the measured pressure differential across the sample gas critical flow orifice. Specifically, the upstream
reading of the sample pressure sensor is compared to the downstream pressure reading of the vacuum
pressure sensor and this differential is used, by the analyzer’s CPU, to derive the gas flow rate through
the reaction cell.
The results of this calculation are viewable from the instruments front panel via the test function
SAMP FLW.
Teledyne Analytical Instruments
336
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Since this is a calculated value and not a measured reading there is no corresponding SIGNALI/O
function.
13.2.4.4. O3 Supply Air Flow Sensor
In contrast to the sample gas flow, the ozone flow is measured with a mass flow sensor, which is
mounted on the flow/pressure sensor PCA just behind the PMT sensor assembly. Pneumatically, it lies
between the Perma Pure® dryer and the O3. This mass flow sensor has a full scale range of 0-1000
cm³/min and can be calibrated through software to its span point (Section 9.7).
Since the flow value displayed on the front panel is an actual measurement (and not a calculated value),
short term variability in the measurement may be higher than that of the sample flow, which is based on
a calculation from (more stable) differential pressures. On the other hand, any sustained drift, i.e. long-
term change, in the ozone flow rate may usually indicate a flow problem.
This information is used to validate the O3 gas flow rate.
If the flow rate exceeds ±15% of the nominal flow rate (80 cm³/min), a warning message OZONE
FLOW WARNING is displayed on the analyzer’s front panel (see Section 4.1.2) and the O3
generator is turned off.
A second warning, OZONE GEN OFF is also displayed.
This flow measurement is viewable through instrument’s front panel display as the test function
OZONE FL and the SIGNAL I/O function OZONE_FLOW.
As with all other test parameters, we recommend to monitor the ozone flow over time for predictive
diagnostics and maintenance evaluation.
Teledyne Analytical Instruments
337
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.3. ELECTRONIC OPERATION
13.3.1. OVERVIEW
Figure 13-13 shows a block diagram of the major electronic components of the analyzer.
(I2C Bus)
Touchscreen
Display
Figure 13-13: 9110T Electronic Block Diagram
Teledyne Analytical Instruments
338
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
The core of the analyzer is a microcomputer/central processing unit (CPU) that controls various internal
processes, interprets data, makes calculations, and reports results using specialized firmware developed
by Teledyne. It communicates with the user as well as receives data from and issues commands to a
variety of peripheral devices via a separate printed circuit assembly onto which the CPU is mounted: the
motherboard.
The motherboard is directly mounted to the inside rear panel and collects data, performs signal
conditioning duties and routes incoming and outgoing signals between the CPU and the analyzer’s other
major components.
Data are generated by the sensor module which outputs an analog signal corresponding to the amount of
chemiluminescence present in the reaction cell. This signal is converted into digital data by a unipolar,
analog-to-digital converter, located on the motherboard.
A variety of sensors report the physical and operational status of the analyzer’s major components, again
through the signal processing capabilities of the motherboard. These status reports are used as data for
the various concentration calculations and as trigger events for certain warning messages and control
commands issued by the CPU. This information is stored in memory by the CPU and in most cases can
be viewed by the user via the front panel display.
The CPU issues commands via a series of relays and switches (also over the I2C bus) located on a
separate printed circuit assembly, called the relay PCA, to control the function of key electromechanical
devices such as heaters and valves. It also issues some commands directly to the Sensor module (e.g.
initiate Electric Test or Optical Test).
By controlling the state of various valves the CPU directs the flow of sample gas through the various gas
paths of the analyzer (NO measurement path; NOx measurement path; Auto Zero Path). Based on which
path is active, the CPU interprets the sensor output to derive raw data representing concentrations for
NOx, NO and zero (dark condition), accesses the operational data stored in memory then calculates final
concentrations for NOx, NO and NO2.
The CPU communicates with the user and the outside world in several ways:
Through the analyzer’s front panel LCD touch-screen interface
Through the serial I/O channels
Various analog voltage and current outputs
Several sets of Digital I/O channels
Ethernet
13.3.2. CPU
The unit’s CPU card, installed on the motherboard located inside the rear panel, is a low power (5 VDC,
720mA max), high performance, Vortex86SX-based microcomputer running Windows CE. Its operation
and assembly conform to the PC 104 specification.
Teledyne Analytical Instruments
339
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-14:
CPU Board
The CPU includes two types of non-volatile data storage: a Disk-on-Module (DOM) and an embedded
flash chip.
13.3.2.1. Disk-On-Module
The DOM is a 44-pin IDE flash drive with storage capacity to 128 MB. It is used to store the computer’s
operating system, the Teledyne firmware, and most of the operational data generated by the analyzer’s
internal data acquisition system (DAS).
13.3.2.2. Flash Chip
This non-volatile, embedded flash chip includes 2MB of storage for calibration data as well as a backup
of the analyzer configuration. Storing these key data on a less heavily accessed chip significantly
decreases the chance of data corruption.
In the unlikely event that the flash chip should fail, the analyzer will continue to operate with just the
DOM. However, all configuration information will be lost, requiring that the unit be recalibrated.
13.3.3. MOTHERBOARD
This PCA provides a multitude of functions including, A/D conversion, digital input/output, PC-104 to
I2C translation, temperature sensor signal processing and is a pass through for the RS-232 and RS-485
signals.
Teledyne Analytical Instruments
340
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
13.3.3.1. A to D Conversion
Principles of Operation
Analog signals, such as the voltages received from the analyzers various sensors, are converted into
digital signals that the CPU can understand and manipulate by the analog to digital converter (A/D).
Under the control of the CPU, this functional block selects a particular signal input and then coverts the
selected voltage into a digital word.
The A/D consists of a Voltage-to-Frequency (V-F) converter, a Programmable Logic Device (PLD),
three multiplexers, several amplifiers and some other associated devices. The V-F converter produces a
frequency proportional to its input voltage. The PLD counts the output of the V-F during a specified
time, and sends the result of that count, in the form of a binary number, to the CPU.
The A/D can be configured for several different input modes and ranges but in the 9110T it is used in
unipolar mode with a +5V full scale. The converter includes a 1% over and under-range. This allows
signals from –0.05V to +5.05V to be fully converted.
For calibration purposes, two reference voltages are supplied to the A/D converter: Reference ground
and +4.096 VDC. During calibration, the device measures these two voltages, outputs their digital
equivalent to the CPU. The CPU uses these values to compute the converter’s offset and slope and uses
these factors for subsequent conversions. See Section 5.9.3.10 for instructions on performing this
calibration.
13.3.3.2. Sensor Inputs
The key analog sensor signals are coupled to the A/D through the master multiplexer from two
connectors on the motherboard. 100K terminating resistors on each of the inputs prevent cross talk from
appearing on the sensor signals.
PMT DETECTOR OUTPUT: The PMT detector output from the PMT preamplifier is used in the
computation of the NO, NOX and NO2 concentrations displayed on the front panel display and reported
through the instrument’s analog outputs and COMM ports.
This information is available in several forms:
As a raw voltage signal via the test function PMTDET and the SIGNAL I/O function PMT_SIGNAL,
or;
Normalized for temperature, pressure and auto-zero offset via the front panel test function NORM
PMT.
It is recorded by the DAS system in the following parameters:
PMTDET – The same as the test function PMT DET.
RAWNOX – The raw PMT output when the instrument is measuring NOX.
RAW NO – The raw PMT output when the instrument is measuring NO.
HIGH VOLTAGE POWER SUPPLY LEVEL: The PMT high voltage is based on the drive voltage
from the preamplifier board. It is digitized and sent to the CPU where it is used to calculate the voltage
setting of the HVPS.
The value of this signal is viewable via the front panel test function HVPS and the SIGNAL I/O
function HVPS_VOLTAGE.
It is recorded by the DAS system as the parameter HVPS.
Teledyne Analytical Instruments
341
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
PMT TEMPERATURE: The PMT temperature is measured with a thermistor inside the PMT cold
block. Its signal is amplified by the PMT temperature feedback circuit on the preamplifier board and is
digitized and sent to the CPU where it is used to calculate the current temperature of the PMT.
The value of this signal is viewable via the front panel test function PMT TEMP and the SIGNAL I/O
function PMT_TEMP.
It is recorded by the DAS system as the parameter PMTTMP.
SAMPLE GAS PRESSURE SENSOR: This sensor, located on the flow/pressure sensor PCA,
measures the gas pressure in the sample chamber upstream of the sample gas stream flow control
assembly. Its functions are described in Section 13.2.4.1.
The value of this signal is viewable via the front panel test function SAMP and the SIGNAL I/O
function SAMPLE_PRESSURE.
It is recorded by the DAS system as the parameter SMPPRS.
VACUUM PRESSURE SENSOR: This sensor, also located on the flow/pressure sensor PCA, is
pneumatically located downstream from the reaction cell and measures the pressure of the gas mixture
inside the reaction cell . Its functions are described in Section 13.2.4.2.
The value of this signal is viewable via the front panel test function RCEL and the SIGNAL I/O
function RCEL_PRESSURE.
It is recorded by the DAS system as the parameter RCPRES.
O3 FLOW SENSOR: This sensor, located on the flow/pressure sensor PCA, measures the flow rate of
the O3 gas stream as it is supplied to the reaction cell. Its functions are described in Section 13.2.4.4.
The value of this signal is viewable via the front panel test function OZONE FLOW and the SIGNAL
I/O function OZONE_FLOW.
It is recorded by the DAS system as the parameter O3FLOW.
13.3.3.3. Thermistor Interface
This circuit provides excitation, termination and signal selection for several negative coefficient,
thermistor temperature sensors located inside the analyzer. They are:
REACTION CELL TEMPERATURE SENSOR: The reaction cell temperature sensor is a thermistor
embedded in the reaction cell manifold. This temperature is used by the CPU to control the reaction cell
heating circuit and as a parameter in the temperature/pressure compensation algorithm.
The value of this signal is viewable via the front panel test function RCEL TEMP and the SIGNAL
I/O function RCELL_TEMP.
It is recorded by the DAS system as the parameter RCTEMP.
BOX TEMPERATURE SENSOR: A thermistor is attached to the motherboard. It measures the
analyzer’s inside temperature. This information is stored by the CPU and can be viewed by the user for
troubleshooting purposes through the front panel display. It is also used as part of the NO, NOx and NO2
calculations when the instrument’s Temperature/Pressure Compensation feature is enabled.
The value of this signal is viewable via the front panel test function BOX TEMP and the SIGNAL I/O
function BOX_TEMP.
It is recorded by the DAS system as the parameter BOXTMP.
Teledyne Analytical Instruments
342
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
INTERNAL SPAN GAS GENERATOR OVEN TEMPERATURE: This thermistor reports the
temperature of the optional internal span gas generator’s NO2 permeation source to the CPU as part of a
control loop that keeps the tube at a high constant temperature (necessary to ensure that the permeation
rate of NO2 is constant). It is stored and reported as test function IZS TEMP.
The value of this signal is viewable via the front panel test function IZS TEMP and the SIGNAL I/O
function IZS_TEMP.
It is recorded by the DAS system as the parameter IZTEMP.
Note
There are two thermistors that monitor the temperature of the PMT
assembly:
One is embedded in the cold block of the PMT’s TEC. Its signal is
conditioned by the PMT preamplifier PCA and reported to the CPU via
the motherboard (see Section 13.3.3.2).
The second is located on the PMT Preamplifier PCA and is used only
as a reference for the preamplifier circuitry. Its output is neither
reported nor stored.
13.3.3.4. Analog Outputs
The analyzer comes equipped with four analog outputs. On the instrument’s rear panel analog connector
(see Figure 3-4), they are labeled A1, A2, A3 and A4.
CONCENTRATION OUTPUTS: Outputs labeled A1, A2 and A3 carry the concentration signals of
NOX, NO and NO2, respectively. A variety of scaling measurement and electronic factors apply to these
signals.
See Sections 3.3.1.3 and 5.4 for information on setting the reporting range type and measurement
range scaling factors for these output channels.
See Sections 5.9.3.2 for instructions calibrating and scaling the electronic output of these channels.
TEST OUTPUT: The fourth analog output, labeled A4 is special. It can be set by the user (see Section
5.9.4) to carry the current signal level of most of the parameters accessible through the TEST menu of
the unit’s software.
In its standard configuration, the 9110T comes with all four of these channels set up to output a DC
voltage. However, 4-20mA current loop drivers can be purchased for the first three of these outputs,
A1, A2 and A3.
OUTPUT LOOP-BACK: All of the functioning analog outputs are connected back to the A/D
converter through a Loop-back circuit. This permits the voltage outputs to be calibrated by the CPU
without need for any additional tools or fixtures (see Section 5.9.3.4).
13.3.3.5. External Digital I/O
This external digital I/O performs two functions.
STATUS OUTPUTS: Logic-Level voltages (0-5 VDC) are output through an optically isolated 8-pin
connector located on the rear panel of the analyzer (see Figure 3-4). These outputs convey good/bad and
on/off information about certain analyzer conditions. They can be used to interface with certain types of
programmable devices.
Teledyne Analytical Instruments
343
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
For information on setting up the status outputs (see Section 3.3.1.4).
CONTROL INPUTS: By applying 5V DC power to these digital inputs from an external source such as
a PLC or Data logger zero point and span point calibrations can be remotely initiated. .
For information on setting up the status inputs (see Section 3.3.1.6).
13.3.3.6. Internal Digital I/O
There are several internal digital control signals that are generated by the motherboard under CPU
control and used to control subsystems of the analyzer.
ELECTRICAL TEST CONTROL: When the CPU sets this control signal to high (ON) the electric
test feature (ETEST) is initiated (see Section 8.3).
The ETEST can be initiated by following the procedure in Section 12.7.12.2, or by setting the
SIGNAL I/O Function ELEC_TEST to ON.
OPTICAL TEST (OTEST) CONTROL: When the CPU sets this control signal to high (ON) the
optical test feature is initiated (see Section 8.3).
The OTEST can be initiated by following the procedure in 12.7.12.1, or by setting the SIGNAL I/O
Function OPTIC_TEST to ON.
PMT PREAMPLIFIER RANGE CONTROL: The CPU uses this control switch the instrument
between its LOW and HIGH physical ranges (see Section 5.4.1).
The instrument can be forced into its HIGH physical range setting the SIGNAL I/O function
PREAMP_RANGE_HI to ON.
O3 GEN STATUS: The CPU uses this control signal to turn the O3 generator ON/OFF by setting it to
HIGH/LOW respectively. The CPU turns OFF the O3 generator if there is if there is no or low air flow
to it as measured by the O3 flow sensor or if the instrument has been turned off for more than 30 minutes.
The O3 generator can be manually turned ON/OFF by using the OZONE GENERATOR OVERIDE
feature (See Section 12.7.15.1) or by setting the SIGNAL I/O function O3GEN_STATUS to ON or
OFF.
Note
Any I/O signals changed while in the signal I/O menu will remain in effect
ONLY until signal I/O menu is exited.
The analyzer regains control of these signals upon exit and returns them
to their normal value/setting.
13.3.3.7. I2C Data Bus
I2C is a two-way, clocked, bi-directional digital serial I/O bus that is used widely in commercial and
consumer electronic systems. A transceiver on the Motherboard converts data and control signals from
the PC-104 bus to I2C format. The data is then fed to the relay board, optional analog input board and
valve driver board circuitry.
Teledyne Analytical Instruments
344
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
13.3.3.8. Power-Up Circuit
Principles of Operation
This circuit monitors the +5V power supply during start-up and sets the analog outputs, external digital
I/O ports, and I2C circuitry to specific values until the CPU boots and the instrument software can
establish control.
13.3.4. RELAY PCA
The CPU issues commands via a series of relays and switches located on a separate printed circuit
assembly, called the relay PCA, to control the function of key electromechanical devices such as heaters
and valves. The relay PCA receives instructions in the form of digital signals over the I2C bus, interprets
these digital instructions and activates its various switches and relays appropriately.
The relay PCA is located in the right-rear quadrant of the analyzer and is mounted vertically on the
backside of the same bracket as the instrument’s DC power supplies.
Status LED’s
(D2 through D16)
Thermocouple
Signal Output
Watchdog
Status LED (D1)
DC Power Supply
Test Points
(JP5)
Thermocouple
Configuration
Jumpers
I2C Connector
J15
J16
TC1 Input
NO2 NO Converter
J21
J19
J14
Temp Sensor
Power
Connection
for DC
Heater AC Power
Configuration
Jumpers
J17
Heaters
Pump AC
Configuration
Jumper
Valve Control
Drivers
JP2
JP7
R16
J12
J11
J10
J9
J4
Pump Power
Output
Valve Control
Connector
AC Relay K4
(OPT Internal Span Gen Heater)
AC Power
IN
J18
J2
J8
AC Relay K2
(NO2 NO Converter Heater)
Connector for
AC Relays
K1 & K2
AC Relay K1
J7
(Reaction Cell Heater)
DC Power
Distribution
Connectors
Connector for AC Relays K4 & K5
Figure 13-15: Relay PCA Layout (P/N 045230100)
CAUTION
ELECTRICAL SHOCK HAZARD
Only those relays actually required by the configuration of the 9110T are populated.
A protective retainer plate is installed over the ac power relay to keep them securely
Teledyne Analytical Instruments
345
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
seated in their sockets and prevent accidental contact with those sockets that are not
populated see Figure 13-16).
Never remove this retainer while the instrument is plugged in and turned on. The
contacts of the AC relay sockets beneath the shield carry high AC voltages even
when no relays are present.
Retainer
Mounting
Screws
AC Relay
Retainer Plate
Figure 13-16: Relay PCA P/N 045230100 with AC Relay Retainer in Place
13.3.4.1. Status LED’s
Eight LED’s are located on the analyzer’s relay PCA to show the current status on the various control
functions performed by the relay PCA (see Figure 13-17). They are:
Teledyne Analytical Instruments
346
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Table 13-4: Relay PCA Status LED’s
Status When Lit
Status When Unlit
LED
Color
Function
(Energized State)
(Default State)
Cycles ON/OFF every 3 Seconds
under direct control of the analyzer’s CPU.
D1
Red
Watchdog Circuit
D2
D3
D4
Yellow
Yellow
Reaction Cell Heater
Heating
Heating
Not Heating
Not Heating
NO2 NO Converter Heater
SPARE
Heating
SPARE
Internal Span Gas Generator
Perm Tube Oven Heater
D51
Yellow
Not Heating
D6
D7
Green
Green
Zero/Span Valve
Sample/Cal Valve
Valve OPEN to span gas flow
Valve OPEN to zero gas flow
Valve OPEN to
calibration gas flow
Valve OPEN to
sample gas flow
D8
D9
Sample gas flow BYPASSES
Sample gas flow is routed
THROUGH the reaction cell
Green
Green
Green
Green
Green
Auto Zero Valve
NO/NOx Valve
the reaction cell
Gas flow routed THROUGH
Gas Flow BYPASSES
NO2 NO converter
D10
D112
D123
NO2 NO converter
Dual Span Gas
Select Valve
Valve OPEN to SPAN 1
Valve OPEN to SPAN2 inlet
Span gas flow OPEN
Zero gas flow OPEN
gas inlet
Pressurized Span
Shutoff Valve
Span gas flow SHUTOFF
Pressurized Zero
Shutoff Valve
D134
Zero gas flow SHUTOFF
D14 - 16
SPARE
1
Only active when the optional internal span gas generator is installed.
Only active when the dual pressurized span option is installed.
Only active when one of the pressurized span gas options is installed.
Only active when one the pressurized zero gas option is installed.
2
3
4
Teledyne Analytical Instruments
347
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
D10 (Green) – NO/NOx Valve
D9 (Green) – AutoZero Valve
D8 (Green) – Optional Sample/Cal Valve
D7 (Green) – Optional Zero/Span Valve
D3 (Yellow) NO2 NO Converter Heater
D2 (Yellow) Reaction Cell Heater
D5 (Yellow) – Optional Internal Span Gas Gen Heater
D11 (Green) – Optional Dual Span Select Valve
D12 (Green) – Optional Pressurized Span Shutoff Valve
D13 (Green) – Optional Pressurized Zero Shutoff Valve
D1 (RED)
Watchdog Indicator
Figure 13-17: Status LED Locations – Relay PCA
13.3.4.2. Watchdog Circuitry
The most important of the status LED’s on the relay board is the red I2C bus watch-dog LED. It is
controlled directly by the analyzer’s CPU over the I2C bus. Special circuitry on the relay PCA watches
the status of D1. Should this LED ever stay ON or OFF for 30 seconds, indicating that the CPU or I2C
bus has stopped functioning, this Watchdog Circuit automatically shuts all valves and turns off all
heaters.
13.3.4.3. Valve Control
The relay board also hosts two valve driver chips, each of which can drive up four valves. The main
valve assembly in the 9110T is the NO/NOX - Auto-zero solenoid valve component mounted right in
front of the NO2 converter housing (see Figure 3-5).
These two valves are actuated with 12 V supplied from the relay board and under the control of the
CPU through the I2C bus.
Additional valve sets also controlled by the CPU via the I2C bus and the relay PCA can be included in
the 9110T (see Table 1-1 and Sections 3.3.2.3, 3.3.2.4, and 3.3.2.5 for descriptions of these valve sets).
Teledyne Analytical Instruments
348
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
13.3.4.4. Heater Control
Principles of Operation
For a variety of reasons such as, efficiency of certain chemical reactions, stabilization of sample gas
temperature and pressure, etc., various subcomponents of the 9110T are heated/cooled.
Two types of sensors are used to gather temperature data for the CPU:
THERMISTORS: These are used in areas where the temperature control point is at or near ambient
temperature (e.g. the reaction cell temperature, internal chassis temperate).
Thermistors change resistance as they heat up and cool down. A DC signal is sent from the
Mother board of a sent voltage and current. As the thermistor changes resistance, the
returning voltage rises and falls in direct relationship to the change in temperature.
The output signal from the thermistors is received by the motherboard, converted into digital
data which is forwarded to the CPU.
THERMOCOUPLES: These are used where the target temperature is high such as the NO2 NO
converter.
Thermocouples generate DC voltage that rises and falls as the thermocouple heats up and
cools down.
This DC signal interpreted, conditioned and amplified by the Relay PCA then transmitted to
the motherboard where it is also converted into digital data and forwarded to the CPU.
All of the heaters used in the 9110T are AC powered which are turned ON/OFF by AC Relays located
on the relay PCA in response to commands issued by the CPU.
Figure 13-18: Heater Control Loop Block Diagram.
Teledyne Analytical Instruments
349
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Note
The PMT temperature is maintained by a separate control loop that does
not involve the relay PCA (see Section 13.5.2).
13.3.4.5. Thermocouple Inputs and Configuration Jumper (JP5)
Although the relay PCA supports two thermocouple inputs, the current 9110T analyzers only utilize one.
It is used to sense the temperature of the NO2 NO converter.
This single thermocouple input is plugged into the TC1 input (J15).
TC2 (J16) is currently not used (see Figure 13-15 for location of J15 and J16).
The type and operating parameters of this thermocouple are set using a jumper plug (JP5).
The default configuration for this thermocouple is:
Type-K
Temperature compensated for Type-K
Isolated
Table 13-5: Thermocouple Configuration Jumper (JP5) Pin-Outs
JUMPER
PAIR
TC INPUT
DESCRIPTION
FUNCTION
Selects preamp gain factor for J or K TC
Gain Selector
OUT = K TC gain factor;
IN = J TC gain factor
1 – 11
Selects preamp gain factor for J or K TC
OUT = 10 mV / °C; IN = 5 mV / °C
Output Scale Selector
2 – 12
When present, sets Cold Junction
Compensation for J type Thermocouple
3 – 13
Type J Compensation
TC1
Type K
Compensation
When present, sets Cold Junction
Compensation for K type Thermocouple
4 – 14
5 – 15
Selects between Isolated and grounded TC
IN = Isolate TC;
OUT = Grounded TC
Termination
Selector
TC2
NOT USED
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
ATTENTION
The correct Thermocouple Type must be used if there is ever the need for
replacement. If in doubt please consult Teledyne Customer Service.
Teledyne Analytical Instruments
350
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
TC2
TC1
Not Used
Purple Jumpers
Figure 13-19: Thermocouple Configuration Jumper (JP5) Pin-Outs
13.4. SENSOR MODULE, REACTION CELL
The 9110T sensor assembly consists of several subassemblies, each with different tasks:
The Photo Multiplier Tube (PMT) detects the intensity of the light from the chemiluminescence
reaction between NO and O3 in the reaction cell. It outputs a current signal that varies in relationship
with the amount of light in the reaction cell.
The PMT Preamplifier PCA converts the current output by the PMT into a voltage and amplifies it to
a signal strong enough to be usable by the motherboard’s A D converter.
It also supplies the drive voltage and gain adjustment for the PMT’s High Voltage Power Supply
(HVPS)
The Thermo-Electric Cooler (TEC) controls the temperature of the PMT to ensure the accuracy and
stability of the measurements.
Teledyne Analytical Instruments
351
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
PMT Housing End Plate
This is the entry to the PMT Exchange
PMT Output
Connector
PMT Preamp PCA
PMT Power Supply
& Aux. Signal
Connector
High voltage Power Supply
(HVPS)
PMT
O-Test LED
PMT Cold Block
Connector to PMT
Pre Amp PCA
12V Power
Connector
Insulation Gasket
Light from Reaction
Chamber shines
through hole in side
of Cold Block
PMT Temperature
Sensor
Thermo-Electric Cooler
(TEC)
PMT Heat Exchange Fins
TEC Driver PCA
Cooling Fan
Housing
Figure 13-20: 9110T Sensor Module Assembly
13.5. PHOTO MULTIPLIER TUBE (PMT)
The 9110T uses a photo multiplier tube (PMT) to detect the amount of chemiluminescence created in the
Reaction Cell.
A typical PMT is a vacuum tube containing a variety of specially designed electrodes. Photons from the
reaction are filtered by an optical high-pass filter, enter the PMT and strike a negatively charged photo
cathode causing it to emit electrons. A high voltage potential across these focusing electrodes directs the
electrons toward an array of high voltage dynodes.
The dynodes in this electron multiplier array are designed so that each stage multiplies the number of
emitted electrons by emitting multiple, new electrons. The greatly increased number of electrons
emitted from one end of electron multiplier are collected by a positively charged anode at the other end,
which creates a useable current signal. This current signal is amplified by the preamplifier board and
then reported to the motherboard.
Teledyne Analytical Instruments
352
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Figure 13-21: Basic PMT Design
A significant performance characteristic of the PMT is the voltage potential across the electron
multiplier. The higher the voltage, the greater the number of electrons emitted from each dynode of the
electron multiplier, in effect, making the PMT more sensitive and responsive to smaller variations in
light intensity but also more noisy (this is referred to as “dark noise”).
The gain voltage of the PMT used in the 9110T is usually set between 400 V and 800 V.
This parameter is viewable through the front panel as test function HVPS (see Section 4.1.1).
For information on when and how to set this voltage, see Section 12.8.4.
The PMT is housed inside the PMT module assembly (see Figure 13-20). This assembly also includes
the high voltage power supply required to drive the PMT, an LED used by the instrument’s optical test
function, a thermistor that measures the temperature of the PMT and various components of the PMT
cooling system including the TEC.
13.5.1. PMT PREAMPLIFIER
The PMT preamplifier board provides a variety of functions:
It amplifies the PMT signal into a useable analog voltage (PMTDET) that can be processed by the
motherboard into a digital signal to be used by the CPU to calculate the NO, NO2 and NOx
concentrations of the gas in the sample chamber.
It supplies the drive voltage for the HVPS.
It includes the circuitry for switching between the two physical ranges.
It amplifies the signal output by the PMT temperature sensor and feeds it back to the thermoelectric
cooler driver PCA. This amplified signal is also sent to the Motherboard to be digitized and
forwarded to the CPU. It is viewable via the front panel as the test function PMT TEMP.
It provides means for adjusting the electronic signal output from the PMT by:
Adjusting this voltage directly the sensitivity of the PMT’s dynode array and therefore the
strength of the signal output by the PMT through the use of two hexadecimal switches.
Directly adjusting the gain of the output signal.
Teledyne Analytical Instruments
353
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Note
These adjustments should only be performed when encountering
problems with the software calibration that cannot be rectified otherwise.
See Section 12.8.4 for more information about this hardware calibration.
Figure 13-22: PMT Preamp Block Diagram
The PMT preamplifier PCA also operates two different tests used to calibrate and check the performance
of the sensor module.
The electrical test (ETEST) circuit generates a constant, electronic signal intended to simulate the
output of the PMT (after conversion from current to voltage). By bypassing the detector’s actual
Teledyne Analytical Instruments
354
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
signal, it is possible to test most of the signal handling and conditioning circuitry on the PMT
preamplifier board. See section 12.7.12.2 for instructions on performing this test.
The optical test (OTEST) feature causes an LED inside the PMT cold block to create a light signal
that can be measured with the PMT. If zero air is supplied to the analyzer, the entire measurement
capability of the sensor module can be tested including the PMT and the current to voltage
conversion circuit on the PMT preamplifier board. See Section 12.7.12.1 for instructions on
performing this test.
13.5.2. PMT COOLING SYSTEM
The performance of the analyzer’s PMT is significantly affected by temperature. Variations in PMT
temperature are directly reflected in the signal output of the PMT. Also the signal to noise ratio of the
PMT output is radically influenced by temperature as well. The warmer the PMT is, the noisier its
signal becomes until the noise renders the concentration signal useless.
To alleviate this problem a special cooling system exists utilizing a type of electronic heat pump called a
thermo-electric cooler (TEC). A TEC is a solid-state active heat pump which transfers heat from a heat
absorbing “cool” side to a heat releasing “hot” side via a series of DC powered semiconductor junctions.
The effectiveness of the pump at moving heat away from the cold side is reliant on the amount of current
flowing through the semiconductor junctions and how well the heat from the hot side can be removed.
HEAT ABSORBING COLD JUNCTION
ELECTRICAL
INSULATOR
ELECTRICAL
CONDUCTOR
N
P
N
P
N
P
N
P
HEAT RELEASING HOT JUNCTION
+
―
DC POWER
SOURCE
Figure 13-23: Typical Thermo-Electric Cooler
In the case of the 9110T, the current flow is controlled by the TEC Control PCA which adjusts the
amount of current applied to the TEC based on the temperature sensed by a thermistor embedded in the
PMT’s cold block. The higher the temperature of the PMT, the more current is pumped through the
TEC. The “hot” side of the TEC is cooled by a constant flow of ambient air that is directed across a set
of heat sinks by a fan.
Teledyne Analytical Instruments
355
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Preamp PCA sends
buffered and
amplified thermistor
signal to TEC PCA
TEC PCA sets
appropriate
drive voltage for
cooler
TEC
Control
PCA
PMT
Preamp
PCA
Thermo-Electric Cooler
PMT Temperature Sensor
Thermistor
outputs temp of
cold block to
preamp PCA
PMT
Cold Block
Heat form PMT is absorbed
by the cold block and
transferred to the heat sink
via the TEC then bled off
into the cool air stream.
Cooling Fan
Figure 13-24: PMT Cooling System Block Diagram
The target temperature at which the TEC system keeps the PMT is approximately 8.0ºC. Arriving at this
temperature may take up to 30 minutes after the instrument is turned on.
The actual temperature of the PMT can be viewed via the front panel as the test function PMT TEMP
(see Section 4.1.1).
13.5.2.1. TEC Control Board
The TEC control PCA is located on the sensor housing assembly, under the slanted shroud, next to the
cooling fins and directly above the cooling fan. Using the amplified PMT temperature signal from the
PMT preamplifier board (see Section 10.4.5), it sets the drive voltage for the thermoelectric cooler. The
warmer the PMT gets, the more current is passed through the TEC causing it to pump more heat to the
heat sink.
A red LED located on the top edge of this circuit board indicates that the control circuit is receiving
power.
Four test points are also located at the top of this assembly.
For the definitions and acceptable signal levels of these test points see 12.7.14.
13.6. PNEUMATIC SENSOR BOARD
The flow and pressure sensors of the 9110T are located on a printed circuit assembly just behind the
PMT sensor. Refer to Section 12.7.6.1 for a figure and on how to test this assembly. The signals of this
board are supplied to the motherboard for further signal processing. All sensors are linearized in the
firmware and can be span calibrated from the front panel.
Teledyne Analytical Instruments
356
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.7. POWER SUPPLY/CIRCUIT BREAKER
The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50 Hz or 60Hz. Individual
instruments are set up at the factory to accept any combination of these five attributes. A 6.75 amp
circuit breaker is built into the ON/OFF switch. In case of a wiring fault or incorrect supply power, the
circuit breaker will automatically turn off the analyzer.
Under normal operation, the 9110T draws about 1.5 A at 115 V and 2.0 A during start-up.
WARNING
ELECTRICAL SHOCK HAZARD
Should the AC power circuit breaker trip, investigate and correct the condition
causing this situation before turning the analyzer back on.
Power enters the analyzer through a standard International Electrotechnical Commission (IEC) 320
power receptacle located on the rear panel of the instrument. From there it is routed through the
ON/OFF Switch located in the lower right corner of the front panel. AC Line power is stepped down
and converted to DC power by two DC power supplies (PS).
One PS provides +5 VDC (3 A) and 15 VDC (1.5/0.5 A) for logic and analog circuitry as well as the
power for the O3 generator.
A second PS provides +12 VDC (5 A), for the PMT’s thermoelectric cooler, fans and as well as the
various gas stream valves (both standard and optional).
All AC and DC Voltages are distributed via the relay PCA.
Teledyne Analytical Instruments
357
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
SENSOR MODULE
Pre-Amplifiers
KEY
HVPS
PMT
& Amplifiers
ANALOG
AC POWER
DC POWER
SENSORS
(e.g. Temp
Sensor Control
& I/O Logic
Sensors, Flow
Sensors)
LOGIC DEVICES
(e.g. CPU, I2C bus,
MotherBoard, etc.)
O3
Generator
PS 1
+5 VDC
±15 VDC
PUMP
Configuration
(Internal Only)
Jumpers
ON / OFF
SWITCH
AC HEATERS
NO2
NO
Configuration
Jumpers
(Converter &
Reaction Cell)
Optional
AC HEATERS
( Internal Span
Generator Perm Tube
Heater)
PS 2
(+12 VDC)
Solenoid
Drivers
RELAY PCA
AC
POWER IN
MODEL SPECIFIC
Fans:
TEC and
Chassis
OPTIONAL
VALVES
(e.g. Sample/Cal,
Zero/Span, Shutoff,
etc.)
VALVES
(e.g. NOX – NO Valves,
Auto-zero valves, etc.)
Figure 13-25: Power Distribution Block Diagram
13.7.1. AC POWER CONFIGURATION
The 9110T analyzer’s digital components will operate with any of the specified power regimes. As long
as instrument is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz,. Internally, the
status LEDs located on the Relay PCA, Motherboard and CPU should turn on as soon as the power is
supplied.
However, some of the analyzer’s non-digital components, such as the various internal pump options or
the AC powered heaters for the NO2 NO converter the reaction cell and some of the 9110T’s must be
properly configured for the type of power being supplied to the instrument.
Configuration of the power circuits is set using several jumper sets located on the instruments relay
PCA.
Teledyne Analytical Instruments
358
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
RELAY PCA
JP6
Configuration Jumpers
for Optional AC Heaters
(O2 Sensor, Internal Perm
Tube Oven Heater)
JP2
JP7
Pump
Configuration
(Internal Pump
Options Only)
Configuration Jumpers
for AC Heaters
(NO2 NO converter,
Reaction Cell)
Figure 13-26: Location of AC power Configuration Jumpers
Teledyne Analytical Instruments
359
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.7.1.1. AC Configuration – Internal Pump (JP7)
If your 9110T includes an internal pump the following table, jumper set JP7 is used to configure the
power supplied to it as shown in Figure 13-27.
Table 13-6: AC Power Configuration for Internal Pumps (JP7)
JUMPER
BETWEEN
PINS
LINE
POWER
LINE
FREQUENCY
JUMPER
COLOR
FUNCTION
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pins 2 & 4 to Neutral
2 to 7
3 to 8
4 to 9
60 HZ
WHITE
BLACK
110VAC
115 VAC
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pins 2 & 4 to Neutral
2 to 7
3 to 8
4 to 9
50 HZ1
Connects pump pins 3 and 4 together
Connects pump pin 1 to 220 / 240VAC power line
Connects pump pins 3 and 4 together
1 to 6
3 to 8
1 to 6
3 to 8
60 HZ
BROWN
BLUE
220VAC
240 VAC
50 HZ1
Connects pump pin 1 to 220 / 240VAC power line
1 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the Model 9110T.
110 VAC /115 VAC
220 VAC /240 VAC
1
2
3
4
5
1
2
3
4
5
6
7
6
7
8
8
9
9
10
10
May be present on 50 Hz version of
jumper set, but is not functional on
the T200
Figure 13-27: Pump AC Power Jumpers (JP7)
Teledyne Analytical Instruments
360
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.7.1.2. AC Configuration – Standard Heaters (JP2)
Power configuration for the AC the standard heaters is set using Jumper set JP2 (see Figure 13-28 for the
location of JP2).
Table 13-7: Power Configuration for Standard AC Heaters (JP2)
JUMPER
JUMPER
COLOR
LINE VOLTAGE
HEATER(S)
BETWEEN
PINS
FUNCTION
Common
1 to 8
Reaction Cell / Sample
Chamber Heaters
Neutral to Load
Neutral to Load
Common
2 to 7
4 to 9
110 VAC / 115 VAC
50Hz & 60 Hz
WHITE
3 to 10
4 to 9
Moly Converter
Neutral to Load
Neutral to Load
6 to 11
Reaction Cell / Sample
Chamber Heaters
Load
Load
1 to 7
3 to 9
220 VAC / 240 VAC
50Hz & 60 Hz
BLUE
Moly Converter
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
Reaction Cell
Heaters
Reaction Cell
Heaters
9
9
NO2 NO
Converter Heaters
NO2 NO
Converter Heaters
10
10
11
12
11
12
220 VAC / 240 VAC
110 VAC /115 VAC
Figure 13-28: Typical Set Up of AC Heater Jumper Set (JP2)
13.7.1.3. AC Configuration – Heaters for Option Packages (JP6)
The IZS valve option includes an AC heaters that maintain an optimum operating temperature for key
components of those options. Jumper set JP6 is used to connect the heaters associated with those options
to AC power. Since these heaters work with either 110/155 VAC or 220/240 VAC, there is only one
jumper configuration.
Teledyne Analytical Instruments
361
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
Table 13-8:
Power Configuration for Optional Heaters (JP6)
JUMPER
BETWEEN
PINS
JUMPER
COLOR
HEATER(S)
FUNCTION
Common
1 to 8
2 to 7
Internal Permeation Tube
Oven Heater
RED
Neutral to Load
10
9
IZS
Permeation Tube
12
11
8
7
Heater
6
5
4
3
2
1
Figure 13-29: Typical Jumper Set (JP2) Set Up of Heaters
Teledyne Analytical Instruments
362
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.8. FRONT PANEL TOUCHSCREEN/DISPLAY INTERFACE
Users can input data and receive information directly through the front panel touchscreen display. The
LCD display is controlled directly by the CPU board. The touchscreen is interfaced to the CPU by means
of a touchscreen controller that connects to the CPU via the internal USB bus and emulates a computer
mouse.
Figure 13-30: Front Panel and Display Interface Block Diagram
Teledyne Analytical Instruments
363
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.8.1. LVDS TRANSMITTER BOARD
The LVDS (low voltage differential signaling) transmitter board converts the parallel display bus to a
serialized, low voltage, differential signal bus in order to transmit the video signal to the LCD interface
PCA.
13.8.2. FRONT PANEL TOUCHSCREEN/DISPLAY INTERFACE PCA
The front panel interface PCA controls the various functions of the display and touchscreen. For driving
the display it provides connection between the CPU video controller and the LCD display module. This
PCA also contains:
Power supply circuitry for the LCD display module
A USB hub that is used for communications with the touchscreen controller and the two
front panel USB device ports
The circuitry for powering the display backlight
13.9. SOFTWARE OPERATION
The 9110T NOx Analyzer has a high performance, VortexX86-based microcomputer running
WINDOWS CE. Inside the WINDOWS CE shell, special software developed by Teledyne interprets
user commands via the various interfaces, performs procedures and tasks, stores data in the CPU’s
various memory devices and calculates the concentration of the sample gas.
Figure 13-31: Basic Software Operation
Teledyne Analytical Instruments
364
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
13.9.1. ADAPTIVE FILTER
The 9110T NOX analyzer software processes sample gas concentration data through a built-in adaptive
filter. Unlike other analyzers that average the output signal over a fixed time period, the 9110T averages
over a defined number of samples, with samples being about 8 seconds apart (reflecting the switching
time of 4 s each for NO and NOX). This technique is known as boxcar filtering. During operation, the
software may automatically switch between two different filters lengths based on the conditions at hand.
During constant or nearly constant concentrations, the software, by default, computes an average of the
last 42 samples, or approximately 5.6 minutes. This provides smooth and stable readings and averages
out a considerable amount of random noise for an overall less noisy concentration reading.
If the filter detects rapid changes in concentration the filter reduces the averaging to only 6 samples or
about 48 seconds to allow the analyzer to respond more quickly. Two conditions must be
simultaneously met to switch to the short filter. First, the instantaneous concentration must differ from
the average in the long filter by at least 50 ppb. Second, the instantaneous concentration must differ
from the average in the long filter by at least 10% of the average in the long filter
13.9.2. TEMPERATURE/PRESSURE COMPENSATION (TPC)
The 9110T software includes a feature that compensates for some temperature and pressure changes
that might affect measurement of NO and NOX concentrations.
When the TPC feature is enabled (default setting), the analyzer divides the value of the PMT output
signal (PMTDET) by a value called TP_FACTOR, which is calculated using the following four
parameters:
BOX TEMP: The temperature inside the analyzer’s case measured in K. This is typically about 5 K
higher than room temperature.
RCELL TEMP: The temperature of the reaction cell, measured in K.
RCEL: The pressure of the gas in the vacuum manifold, measured in in-Hg-A.
SAMP: The pressure of the sample gas before it reaches the reaction cell, measured in in-Hg-A.
This measurement is ~1 in-Hg-A lower than atmospheric pressure.
As RCEL TEMP, BOX TEMP, RCELL and SAMP pressure increase, the value of TP_FACTOR
increases and, hence, the PMTDET value decreases. These adjustments are meant to counter-act
changes in the concentrations caused by these parameters.
The current value of all four of these measurements are viewable as TEST FUNCTIONS through the
instrument’s front panel display (see Section 4.1.1).
The preset gain parameters are set at the factory and may vary from analyzer to analyzer. The TPC
feature is enabled or disabled by setting the value of the variable TPC_ENABLE (see Section 5.8).
13.9.3. CALIBRATION - SLOPE AND OFFSET
Calibration of the analyzer is performed exclusively in software. During instrument calibration, (see
Sections 9 and 10) the user enters expected values for zero and span via the front panel touchscreen
control and commands the instrument to make readings of calibrated sample gases for both levels.
Teledyne Analytical Instruments
365
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Principles of Operation
The readings taken are adjusted, linearized and compared to the expected values.
With this information, the software computes values for instrument slope and offset and stores these
values in memory for use in calculating the NOx, NO and NO2 concentrations of the sample gas.
The instrument slope and offset values recorded during the last calibration can be viewed via the
instrument’s front panel (see Section 4.1.1).
Teledyne Analytical Instruments
366
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
14. A PRIMER ON ELECTRO-STATIC DISCHARGE
TAI considers the prevention of damage caused by the discharge of static electricity to be extremely
important part of making sure that your analyzer continues to provide reliable service for a long time.
This section describes how static electricity occurs, why it is so dangerous to electronic components and
assemblies as well as how to prevent that damage from occurring.
14.1. HOW STATIC CHARGES ARE CREATED
Modern electronic devices such as the types used in the various electronic assemblies of your analyzer,
are very small, require very little power and operate very quickly. Unfortunately, the same
characteristics that allow them to do these things also make them very susceptible to damage from the
discharge of static electricity. Controlling electrostatic discharge begins with understanding how electro-
static charges occur in the first place.
Static electricity is the result of something called triboelectric charging which happens whenever the
atoms of the surface layers of two materials rub against each other. As the atoms of the two surfaces
move together and separate, some electrons from one surface are retained by the other.
Materials
Makes
Contact
Materials
Separate
+
+
+
+
PROTONS = 3
ELECTRONS = 2
PROTONS = 3
ELECTRONS = 4
PROTONS = 3
ELECTRONS = 3
PROTONS = 3
ELECTRONS = 3
NET CHARGE = -1
NET CHARGE = +1
NET CHARGE = 0
NET CHARGE = 0
Figure 14-1:
Triboelectric Charging
If one of the surfaces is a poor conductor or even a good conductor that is not grounded, the resulting
positive or negative charge cannot bleed off and becomes trapped in place, or static. The most common
example of triboelectric charging happens when someone wearing leather or rubber soled shoes walks
across a nylon carpet or linoleum tiled floor. With each step, electrons change places and the resulting
electro-static charge builds up, quickly reaching significant levels. Pushing an epoxy printed circuit
board across a workbench, using a plastic handled screwdriver or even the constant jostling of
StyrofoamTM pellets during shipment can also build hefty static charges.
Table 14-1: Static Generation Voltages for Typical Activities
MEANS OF GENERATION
Walking across nylon carpet
Walking across vinyl tile
Worker at bench
65-90% RH
1,500V
250V
10-25% RH
35,000V
12,000V
6,000V
100V
Poly bag picked up from bench
1,200V
20,000V
Moving around in a chair padded
with urethane foam
1,500V
18,000V
Teledyne Analytical Instruments
367
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
14.2. HOW ELECTRO-STATIC CHARGES CAUSE DAMAGE
Damage to components occurs when these static charges come into contact with an electronic device.
Current flows as the charge moves along the conductive circuitry of the device and the typically very
high voltage levels of the charge overheat the delicate traces of the integrated circuits, melting them or
even vaporizing parts of them. When examined by microscope the damage caused by electro-static
discharge looks a lot like tiny bomb craters littered across the landscape of the component’s circuitry.
A quick comparison of the values in with the those shown in the , listing device susceptibility levels,
shows why Semiconductor Reliability News estimates that approximately 60% of device failures are the
result of damage due to electro-static discharge.
Table 14-2:
Sensitivity of Electronic Devices to Damage by ESD
DAMAGE SUSCEPTIBILITY VOLTAGE
RANGE
DEVICE
DAMAGE BEGINS
OCCURRING AT
CATASTROPHIC
DAMAGE AT
MOSFET
VMOS
10
30
100
1800
100
NMOS
60
GaAsFET
EPROM
60
2000
100
100
140
150
190
200
300
300
300
500
500
500
JFET
7000
500
SAW
Op-AMP
2500
3000
2500
3000
7000
500
CMOS
Schottky Diodes
Film Resistors
This Film Resistors
ECL
SCR
1000
2500
Schottky TTL
Potentially damaging electro-static discharges can occur:
Any time a charged surface (including the human body) discharges to a device. Even simple
contact of a finger to the leads of a sensitive device or assembly can allow enough discharge to
cause damage. A similar discharge can occur from a charged conductive object, such as a metallic
tool or fixture.
When static charges accumulated on a sensitive device discharges from the device to another
surface such as packaging materials, work surfaces, machine surfaces or other device. In some
cases, charged device discharges can be the most destructive.
A typical example of this is the simple act of installing an electronic assembly into the connector or
wiring harness of the equipment in which it is to function. If the assembly is carrying a static charge,
as it is connected to ground a discharge will occur.
Whenever a sensitive device is moved into the field of an existing electro-static field, a charge may
be induced on the device in effect discharging the field onto the device. If the device is then
momentarily grounded while within the electrostatic field or removed from the region of the
electrostatic field and grounded somewhere else, a second discharge will occur as the charge is
transferred from the device to ground.
Teledyne Analytical Instruments
368
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
14.3. COMMON MYTHS ABOUT ESD DAMAGE
I didn’t feel a shock so there was no electro-static discharge: The human nervous system isn’t
able to feel a static discharge of less than 3500 volts. Most devices are damaged by discharge
levels much lower than that.
I didn’t touch it so there was no electro-static discharge: Electro Static charges are fields
whose lines of force can extend several inches or sometimes even feet away from the surface
bearing the charge.
It still works so there was no damage: Sometimes the damages caused by electro-static
discharge can completely sever a circuit trace causing the device to fail immediately. More likely,
the trace will be only partially occluded by the damage causing degraded performance of the device
or worse, weakening the trace. This weakened circuit may seem to function fine for a short time, but
even the very low voltage and current levels of the device’s normal operating levels will eat away at
the defect over time causing the device to fail well before its designed lifetime is reached.
These latent failures are often the most costly since the failure of the equipment in which the
damaged device is installed causes down time, lost data, lost productivity, as well as possible failure
and damage to other pieces of equipment or property.
Static Charges can’t build up on a conductive surface: There are two errors in this statement.
Conductive devices can build static charges if they are not grounded. The charge will be equalized
across the entire device, but without access to earth ground, they are still trapped and can still build
to high enough levels to cause damage when they are discharged.
A charge can be induced onto the conductive surface and/or discharge triggered in the presence of
a charged field such as a large static charge clinging to the surface of a nylon jacket of someone
walking up to a workbench.
As long as my analyzer is properly installed, it is safe from damage caused by static
discharges: It is true that when properly installed the chassis ground of your analyzer is tied to
earth ground and its electronic components are prevented from building static electric charges
themselves. This does not prevent discharges from static fields built up on other things, like you and
your clothing, from discharging through the instrument and damaging it.
14.4. BASIC PRINCIPLES OF STATIC CONTROL
It is impossible to stop the creation of instantaneous static electric charges. It is not, however difficult to
prevent those charges from building to dangerous levels or prevent damage due to electro-static
discharge from occurring.
14.4.1. GENERAL RULES
Only handle or work on all electronic assemblies at a properly set up ESD station. Setting up an ESD
safe workstation need not be complicated. A protective mat properly tied to ground and a wrist strap are
all that is needed to create a basic anti-ESD workstation.
Teledyne Analytical Instruments
369
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
Protective Mat
Ground Point
Wrist Strap
Figure 14-2:
Basic anti-ESD Workbench
For technicians that work in the field, special lightweight and portable anti-ESD kits are available from
most suppliers of ESD protection gear. These include everything needed to create a temporary anti-ESD
work area anywhere.
Always wear an Anti-ESD wrist strap when working on the electronic assemblies of your
analyzer. An anti-ESD wrist strap keeps the person wearing it at or near the same potential as
other grounded objects in the work area and allows static charges to dissipate before they can build
to dangerous levels. Anti-ESD wrist straps terminated with alligator clips are available for use in
work areas where there is no available grounded plug.
Also, anti-ESD wrist straps include a current limiting resistor (usually around one meg-ohm) that
protects you should you accidentally short yourself to the instrument’s power supply.
Simply touching a grounded piece of metal is insufficient. While this may temporarily bleed off
static charges present at the time, once you stop touching the grounded metal new static charges
will immediately begin to re-build. In some conditions, a charge large enough to damage a
component can rebuild in just a few seconds.
Always store sensitive components and assemblies in anti-ESD storage bags or bins: Even
when you are not working on them, store all devices and assemblies in a closed anti-Static bag or
bin. This will prevent induced charges from building up on the device or assembly and nearby static
fields from discharging through it.
Use metallic anti-ESD bags for storing and shipping ESD sensitive components and
assemblies rather than pink-poly bags. The famous, pink-poly bags are made of a plastic that is
impregnated with a liquid (similar to liquid laundry detergent) which very slowly sweats onto the
surface of the plastic creating a slightly conductive layer over the surface of the bag.
While this layer may equalizes any charges that occur across the whole bag, it does not prevent the
build up of static charges. If laying on a conductive, grounded surface, these bags will allow
charges to bleed away but the very charges that build up on the surface of the bag itself can be
transferred through the bag by induction onto the circuits of your ESD sensitive device. Also, the
liquid impregnating the plastic is eventually used up after which the bag is as useless for preventing
damage from ESD as any ordinary plastic bag.
Anti-Static bags made of plastic impregnated with metal (usually silvery in color) provide all of the
charge equalizing abilities of the pink-poly bags but also, when properly sealed, create a Faraday
cage that completely isolates the contents from discharges and the inductive transfer of static
charges.
Storage bins made of plastic impregnated with carbon (usually black in color) are also excellent at
dissipating static charges and isolating their contents from field effects and discharges.
Teledyne Analytical Instruments
370
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
Never use ordinary plastic adhesive tape near an ESD sensitive device or to close an anti-
ESD bag. The act of pulling a piece of standard plastic adhesive tape, such as Scotch® tape, from
its roll will generate a static charge of several thousand or even tens of thousands of volts on the
tape itself and an associated field effect that can discharge through or be induced upon items up to
a foot away.
14.5. BASIC ANTI-ESD PROCEDURES FOR ANALYZER REPAIR
AND MAINTENANCE
14.5.1. WORKING AT THE INSTRUMENT RACK
When working on the analyzer while it is in the instrument rack and plugged into a properly grounded
power supply:
1. Attach your anti-ESD wrist strap to ground before doing anything else.
Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the
instrument chassis.
This will safely connect you to the same ground level to which the instrument and all of its
components are connected.
2. Pause for a second or two to allow any static charges to bleed away.
3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing of
your analyzer has isolated the components and assemblies inside from any conducted or
induced static charges.
4. If you must remove a component from the instrument, do not lay it down on a non-ESD
preventative surface where static charges may lie in wait.
5. Only disconnect your wrist strap after you have finished work and closed the case of the
analyzer.
14.5.2. WORKING AT AN ANTI-ESD WORK BENCH
When working on an instrument of an electronic assembly while it is resting on a anti-ESD workbench:
1. Plug your anti-ESD wrist strap into the grounded receptacle of the work station before touching
any items on the work station and while standing at least a foot or so away. This will allow any
charges you are carrying to bleed away through the ground connection of the workstation and
prevent discharges due to field effects and induction from occurring.
2. Pause for a second or two to allow any static charges to bleed away.
3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies after
you have plugged your wrist strap into the workstation.
Lay the bag or bin on the workbench surface.
Before opening the container, wait several seconds for any static charges on the outside
surface of the container to be bled away by the workstation’s grounded protective mat.
4. Do not pick up tools that may be carrying static charges while also touching or holding an ESD
sensitive device.
Teledyne Analytical Instruments
371
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your
workstation. Never lay them down on any non-ESD preventative surface.
5. Place any static sensitive devices or assemblies in anti-static storage bags or bins and close the
bag or bin before unplugging your wrist strap.
6. Disconnecting your wrist strap is always the last action taken before leaving the workbench.
14.5.3. TRANSFERRING COMPONENTS BETWEEN RACK AND BENCH
When transferring a sensitive device from an installed TAI analyzer to an Anti-ESD workbench or back:
1. Follow the instructions listed above for working at the instrument rack and workstation.
2. Never carry the component or assembly without placing it in an anti-ESD bag or bin.
3. Before using the bag or container allow any surface charges on it to dissipate:
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at an anti-ESD workbench, lay the container down on the conductive work
surface.
In either case wait several seconds.
4. Place the item in the container.
5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.
Folding the open end over isolates the component(s) inside from the effects of static fields.
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag
from forming a complete protective envelope around the device.
6. Once you have arrived at your destination, allow any surface charges that may have built up on
the bag or bin during travel to dissipate:
Connect your wrist strap to ground.
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at a anti-ESD workbench, lay the container down on the conductive work surface.
In either case wait several seconds.
7. Open the container.
14.5.4. OPENING SHIPMENTS FROM TAI CUSTOMER SERVICE
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient generators of static
electric charges. To prevent damage from ESD, TAI ships all electronic components and assemblies in
properly sealed ant-ESD containers.
Teledyne Analytical Instruments
372
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
Static charges will build up on the outer surface of the anti-ESD container during shipping as the packing
materials vibrate and rub against each other. To prevent these static charges from damaging the
components or assemblies being shipped ensure that you:
Always unpack shipments from Teledyne Customer Service by:
1. Opening the outer shipping box away from the anti-ESD work area.
2. Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area.
3. Follow steps 6 and 7 of Section 14.2.3 above when opening the anti-ESD container at the work
station.
4. Reserve the anti-ESD container or bag to use when packing electronic components or
assemblies to be returned to TAI.
14.5.5. PACKING COMPONENTS FOR RETURN TO TAI CUSTOMER
SERVICE
Always pack electronic components and assemblies to be sent to Teledyne Customer Service in anti-
ESD bins, tubes or bags.
CAUTION
ESD Hazard
DO NOT use pink-poly bags.
NEVER allow any standard plastic packaging materials to touch
the electronic component/assembly directly.
This includes, but is not limited to, plastic bubble-pack,
Styrofoam peanuts, open cell foam, closed cell foam, and
adhesive tape.
DO NOT use standard adhesive tape as a sealer. Use ONLY anti-
ESD tape.
Never carry the component or assembly without placing it in an anti-ESD bag or bin.
1. Before using the bag or container allow any surface charges on it to dissipate:
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at an anti-ESD workbench, lay the container down on the conductive work
surface.
In either case wait several seconds.
2. Place the item in the container.
3. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.
Folding the open end over isolates the component(s) inside from the effects of static fields.
Teledyne Analytical Instruments
373
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Primer on ESD
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag
from forming a complete protective envelope around the device.
Note
If you do not already have an adequate supply of anti-ESD bags or
containers available, Teledyne Customer Service department will supply
them (see Section 12.10 for contact information). Follow the instructions
listed above for working at the instrument rack and workstation.
Teledyne Analytical Instruments
374
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
GLOSSARY
Term
Glossary
Description/Definition
10BaseT
100BaseT
APICOM
ASSY
an Ethernet standard that uses twisted (“T”) pairs of copper wires to transmit at
10 megabits per second (Mbps)
same as 10BaseT except ten times faster (100 Mbps)
name of a remote control program offered by Teledyne-API to its customers
Assembly
CAS
Code-Activated Switch
CD
Corona Discharge, a frequently luminous discharge, at the surface of a
conductor or between two conductors of the same transmission line,
accompanied by ionization of the surrounding atmosphere and often by a power
loss
CE
Converter Efficiency, the percentage of the total amount that is actually
converted (e.g., light energy into electricity; NO2 into NO, etc.)
CEM
Continuous Emission Monitoring
Chemical formulas that may be included in this document:
CO2
C3H8
CH4
H2O
HC
carbon dioxide
propane
methane
water vapor
general abbreviation for hydrocarbon
HNO3
H2S
nitric acid
hydrogen sulfide
NO
nitric oxide
NO2
NOX
NOy
nitrogen dioxide
nitrogen oxides, here defined as the sum of NO and NO2
nitrogen oxides, often called odd nitrogen: the sum of NOX plus other compounds such as HNO3
(definitions vary widely and may include nitrate (NO3), PAN, N2O and other compounds as well)
NH3
O2
ammonia
molecular oxygen
ozone
O3
Teledyne Analytical Instruments
375
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Glossary
Term
Description/Definition
SO2
sulfur dioxide
cm3
metric abbreviation for cubic centimeter (replaces the obsolete abbreviation
“cc”)
CPU
DAC
DAS
DCE
DFU
Central Processing Unit
Digital-to-Analog Converter
Data Acquisition System
Data Communication Equipment
Dry Filter Unit
DHCP
Dynamic Host Configuration Protocol. A protocol used by LAN or Internet
servers to automatically set up the interface protocols between themselves and
any other addressable device connected to the network
DIAG
DOM
Diagnostics, the diagnostic settings of the analyzer.
Disk On Module, a 44-pin IDE flash drive with up to 128MB storage capacity for
instrument’s firmware, configuration settings and data
DOS
Disk Operating System
DRAM
DR-DOS
DTE
Dynamic Random Access Memory
Digital Research DOS
Data Terminal Equipment
EEPROM
Electrically Erasable Programmable Read-Only Memory also referred to as a
FLASH chip or drive
ESD
Electro-Static Discharge
Electrical Test
ETEST
Ethernet
a standardized (IEEE 802.3) computer networking technology for local area
networks (LANs), facilitating communication and sharing resources
FEP
Fluorinated Ethylene Propylene polymer, one of the polymers that Du Pont
markets as Teflon®
Flash
FPI
non-volatile, solid-state memory
Fabry-Perot Interface: a special light filter typically made of a transparent plate
with two reflecting surfaces or two parallel, highly reflective mirrors
GFC
Gas Filter Correlation
I2C bus
a clocked, bi-directional, serial bus for communication between individual
analyzer components
Teledyne Analytical Instruments
376
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Glossary
Term
Description/Definition
IC
Integrated Circuit, a modern, semi-conductor circuit that can contain many basic
components such as resistors, transistors, capacitors etc in a miniaturized
package used in electronic assemblies
IP
Internet Protocol
IZS
Internal Zero Span
Local Area Network
Liquid Crystal Display
Light Emitting Diode
Liters Per Minute
LAN
LCD
LED
LPM
MFC
M/R
Mass Flow Controller
Measure/Reference
NDIR
Non-Dispersive Infrared
MOLAR MASS
the mass, expressed in grams, of 1 mole of a specific substance. Conversely,
one mole is the amount of the substance needed for the molar mass to be the
same number in grams as the atomic mass of that substance.
EXAMPLE: The atomic weight of Carbon is 12 therefore the molar mass of
Carbon is 12 grams. Conversely, one mole of carbon equals the amount of
carbon atoms that weighs 12 grams.
Atomic weights can be found on any Periodic Table of Elements.
NDIR
Non-Dispersive Infrared
NIST-SRM
PC
National Institute of Standards and Technology - Standard Reference Material
Personal Computer
PCA
PC/AT
PCB
PFA
Printed Circuit Assembly, the PCB with electronic components, ready to use
Personal Computer / Advanced Technology
Printed Circuit Board, the bare board without electronic component
Per-Fluoro-Alkoxy, an inert polymer; one of the polymers that Du Pont markets
as Teflon®
PLC
Programmable Logic Controller, a device that is used to control instruments
based on a logic level signal coming from the analyzer
PLD
Programmable Logic Device
PLL
Phase Lock Loop
PMT
Photo Multiplier Tube, a vacuum tube of electrodes that multiply electrons
collected and charged to create a detectable current signal
P/N (or PN)
Part Number
Teledyne Analytical Instruments
377
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 9110T NOx Analyzer
Glossary
Term
Description/Definition
Prevention of Significant Deterioration
PSD
PTFE
Poly-Tetra-Fluoro-Ethylene, a very inert polymer material used to handle gases
that may react on other surfaces; one of the polymers that Du Pont markets as
Teflon®
PVC
Poly Vinyl Chloride, a polymer used for downstream tubing
Rdg
Reading
RS-232
specification and standard describing a serial communication method between
DTE (Data Terminal Equipment) and DCE (Data Circuit-terminating Equipment)
devices, using a maximum cable-length of 50 feet
RS-485
specification and standard describing a binary serial communication method
among multiple devices at a data rate faster than RS-232 with a much longer
distance between the host and the furthest device
SAROAD
SLAMS
SLPM
Storage and Retrieval of Aerometric Data
State and Local Air Monitoring Network Plan
Standard Liters Per Minute of a gas at standard temperature and pressure
Standard Temperature and Pressure
STP
TCP/IP
Transfer Control Protocol / Internet Protocol, the standard communications
protocol for Ethernet devices
TEC
TPC
USB
Thermal Electric Cooler
Temperature/Pressure Compensation
Universal Serial Bus: a standard connection method to establish communication
between peripheral devices and a host controller, such as a mouse and/or
keyboard and a personal computer or laptop
VARS
V-F
Variables, the variable settings of the instrument
Voltage-to-Frequency
Z/S
Zero / Span
Teledyne Analytical Instruments
378
Download from Www.Somanuals.com. All Manuals Search And Download.
|