Teledyne TV Converter Box 200AU NOy User Manual

MANUAL ADDENDUM  
M200AU  
NOy CONVERTER OPTION  
(for use with M200AU Operators Manual, 02293)  
Teledyne Advanced Pollution Instrumentation  
9480 Carroll Park Dr.  
San Diego, CA 92121-5201  
TOLL-FREE: 800-324-5190  
FAX: 858-657-9818  
TEL: 858-657-9800  
02808  
REV. E  
Copyright 2007 API Inc.  
July 2007  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
Table of Contents  
02808 Rev. D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
02808 Rev. D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
1.0 Getting Started  
The M501Y has been designed to operate with the M200AU low level NOx analyzer. The M200AU  
manual, part number 02293, is a required reference to properly set up and operate this system.  
1.1 Unpacking  
1. The M200AU with NOy Option comes in three boxes.  
The M200AU analyzer  
The Bypass Pump Chassis  
The External Converter and umbilical cable assembly  
2. Verify that there is no apparent shipping damage. If damage has occurred please advise shipper  
first, then API.  
CAUTION  
To avoid personal injury, always use two persons to lift and  
carry the Model 501 Pump Pack Chassis.  
3. Before operation of the M200AU, remove the shipping screws as shown in the M200AU Manual.  
4. Please check the voltage and frequency label on the rear panel of the instrument for compatibility  
with the local power before plugging in the M501Y.  
1.2 Electrical and Pneumatic Connections  
1. Mount the converter on a suitable mast outside of the enclosure.  
2. Route the umbilical cable assembly into the shelter.  
3. Locate the M200AU and M501Y Bypass Pump Chassis in close proximity, preferably mounting  
one over the other in a 19” rack.  
4. Connect the electrical/pneumatic cable fittings to the rear of the Bypass Pump Chassis. Use the  
tags on each tube and Figure 1.1 to match up the correct tube with the rear panel bulkhead fitting.  
5. Connect the 7-pin power and signal cable to the connector on the rear of the Bypass Pump Chassis.  
6. Connect the power cord to the correct line voltage.  
02808 Rev. E  
Page 7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
WARNING  
Lethal voltages are present inside case  
Do not operate with cover off during normal operation  
Before operation, check for correct input voltage and frequency.  
Do not operate without proper chassis grounding  
Do not defeat the ground wire on power plug  
Turn off power before disconnecting electrical subassemblies  
02808 Rev. E  
Page 8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
Figure 1.1 - Rear Panel Pneumatic Connections  
02808 Rev. E  
Page 9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
1.3 Initial Operation  
1. After confirming proper supply voltage, turn on the instrument power. If you are unfamiliar with  
the M200AU with NOy Option, we recommend that you read the overview in Section 3 before  
proceeding.  
2. The power indicator light and display should immediately light, in addition the bypass pump  
should start up. The M501Y requires about 30 minutes for the converter to come up to  
temperature.  
3. While the converter is coming up to temperature, the M200AU should be powered-up also. Follow  
the instructions in Section 2 of the M200AU Manual to confirm proper operation of that  
instrument. Note: It will take 30 minutes for the ozone generator to start up in the M200AU.  
During that time the instrument will not respond to span gas.  
4. After 30 min, the display on the Bypass Pump Chassis should read 315oC, indicating that the  
converter is up to temperature.  
5. Proceed to Section 4 of this manual to do a zero/span check.  
Table 2.1 - Final Test and Calibration Values  
Test Values  
Observed Value  
Units  
oC  
Nominal Range  
315 ±7  
CONVERTER TEMP  
BYPASS FLOW – NO CHANNEL  
BYPASS FLOW – NOx CHANNEL  
CC/MIN  
CC/MIN  
1100 ± 200  
1100 ± 200  
Converter Serial # __________________________  
Date  
_______________________  
Technician __________________________  
02808 Rev. E  
Page 10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
2.0 Specifications, Warranty  
2.1 Specifications  
Converter  
Remotely mounted molybdenum converter with temperature  
controller in a NEMA – 4 enclosure.  
315oC ± 7oC with read-out on front panel of bypass pump  
chassis.  
Converter Temperature  
Tube/Electrical Cable  
Assembly  
Up to 50 feet (15 m) maximum length  
120V~ 60 Hz, 230V~ 50 Hz 360 watts  
Power, Bypass Pump  
Assembly  
Power, Converter Assembly  
Converter Efficiency  
Weight, Bypass Pump  
Assembly  
120V~ 50/60 Hz, 60 watts  
>96% for NO2  
35 lbs (16 kg)  
Weight, Remote  
30 lbs (11 kg)  
Converter  
Bypass Pump Chassis  
Dimensions HxWxD  
Remote Converter Assy  
Dimensions HxWxD  
Environmental  
9" x17" x23.6" (23cm x 43cm x 61cm)  
12"x19"x7" (31cm x 48cm x 18cm)  
Installation Category (Over-voltage Category) II  
Pollution Degree 2  
Bypass Flowrate  
Analog output  
1100 cc/min ± 20%, each channel  
Converter temp analog output. 3.15 VDC = 315oC  
02808 Rev. E  
Page 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
2.2 Warranty  
ADVANCED POLLUTION INSTRUMENTATION, INC.  
02024c  
Prior to shipment, API equipment is thoroughly inspected and tested. Should equipment failure occur, API assures  
its customers that prompt service and support will be available.  
COVERAGE  
After the warranty period and throughout the equipment lifetime, API stands ready to provide on-site or in-plant  
service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level  
of field troubleshooting are to be performed by the customer.  
NON-API MANUFACTURED EQUIPMENT  
Equipment provided but not manufactured by API is warranted and will be repaired to the extent and according to  
the current terms and conditions of the respective equipment manufacturers warranty.  
GENERAL  
API warrants each Product manufactured by API to be free from defects in material and workmanship under normal  
use and service for a period of one year from the date of delivery. All replacement parts and repairs are warranted for  
90 days after the purchase.  
If a Product fails to conform to its specifications within the warranty period, API shall correct such defect by, in  
API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product.  
The warranties set forth in this section shall be of no force or effect with respect to any Product:  
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other  
than in accordance with the instruction provided by API or (iii) not properly maintained.  
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE  
AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR  
PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE  
REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY  
WARRANTY CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR  
CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF API'S  
PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE.  
TERMS AND CONDITIONS  
All units or components returned to API should be properly packed for handling and returned freight prepaid to  
the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.  
02808 Rev. E  
Page 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
3.0 The M501Y NOy Converter  
3.1 Principle of Operation  
The API Model 200AU with NOy Option is designed to measure the concentration of NO, NO2, and  
other compounds that are too unstable to be measured when taken in through the normal ambient air  
sample inlet system. Please refer to Section 4 of the M200AU manual supplied with this system for a  
general discussion of the operation of a NOx analyzer. The suite of compounds known collectively as  
NOy is composed of roughly 30 compounds. There is some disagreement over whether certain  
compounds should be on the list. The NOy measurement is generally done in conjunction with a  
standard NOx measurement, with the difference between the two being the concentration of the  
unstable compounds, sometimes referred to as NOz.  
The system is composed of 3 modules:  
The M200AU, without a moly converter  
A Bypass pump chassis containing:  
Bypass pump  
Flow control  
Sample filtration  
Moly temperature control  
Pneumatic provisions for calibration  
An externally mounted molybdenum converter  
The M200AU w/NOy Option system allows the converter to be mounted at the sample inlet point.  
This configuration provides minimal time delay between the sample inlet port and the remotely  
mounted molybdenum converter. Minimizing the transit time between the sample inlet and converter  
enables the conversion of labile components of NOy. The equation for the conversion is:  
315oC  
3 NOY + Mo ---> 3 NO + MoO3  
02808 Rev. E  
Page 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
4.0 CALIBRATION AND ZERO/SPAN CHECKS  
Unlike most NOx analyzers, the M200AU with the NOy option does not have a sample inlet port on  
the rear panel; therefore zero/span calibration is different than a normal NOx instrument.  
Follow the steps in Section 7 of the M200AU manual to:  
a. Enter the expected NO and NOx span gas concentrations in the M200AU.  
b. Do the zero calibration procedure covered in Section 7.1.  
c. Do the span calibration procedure covered in Section 7.1.  
The M200AU / NOy can be calibrated only with zero/span gas entering near the remote  
sample port on the pole. There is a fitting on the Bypass Pump Module to input the  
calibration gas to this port via a third Teflon hose that is part of the umbilical.  
Calibration gas delivered in this manner must be under a small amount of pressure to  
overcome the resistance of the hose (pushed up the hose). The calibrator or gas  
bottle output must be directly connected to the calibration gas inlet on the back of the  
Bypass Pump module - there must be no other venting of the zero or span gas to  
atmosphere.  
The system contains no provision for calibration gas shutoff or flow control, therefore  
the calibration gas system must provide for both. Flow requirements are 1.1 LPM for  
NO, 1.1 LPM for NOx and 1 LPM for the M200AU. Calibration gas flow must be in  
sufficient excess (~4 LPM) to prevent any ambient air from entering the sample port on  
the pole.  
The maximum 50 ft length of tubing between the Bypass Pump Module and the remote moly converter  
has been proven to generate too much backpressure for some brands of calibrators. Experience to date  
is that while NO calibration is correct, GPT can be incorrect due to errors in dilution of the ozone.  
Although the M200AU manual shows several ways to check and adjust the calibration of the  
M200AU, there is only one way to calibrate or check the calibration of the M200AU/ NOy. This is  
calibration with zero gas coming in the sample port - covered in Section 7.1 of the M200AU manual.  
We strongly recommend that SPAN CALIBRATION be done with NO span gas. SPAN CHECKS can  
be done with either NO only, NO2 only or a mixture of NO and NO2 (GPT).  
Zero air used for all calibration procedures, including GPT, should have <1 ppt NO and NO2, less than  
1 ppt of major interferents such as SO2, NH3, hydrocarbons and a dew point of -5oC or less. The  
calibration gasses should be from a reliable supplier, since the quality of the tank concentration values  
ultimately determines the accuracy of the analyzer.  
02808 Rev. E  
Page 15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
NOTE  
The M200AU with the NOy Option does NOT have  
equivalency approval, and may not be used for EPA  
monitoring.  
Figure 4.1 - Calibration Setup  
02808 Rev. E  
Page 16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
4.1 Calibration or Cal Check Procedure  
The calibration of the instrument can be checked or adjusted using gas introduced through the  
calibration port on the Bypass Pump Module. The calibration gas is routed to a TEE near the sample  
inlet port on the remote converter assembly.  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero  
value. Table 4.1 details the zero calibration procedure.  
Table 4.1 - Manual Zero Calibration Procedure - Zero Gas Thru Sample Port  
Step  
Number  
Action  
Press CAL  
Comment  
1.  
The M200AU enters the calibrate mode and the adaptive  
filter is activated. This allows the instrument to respond  
rapidly to concentration changes regardless of their  
magnitude.  
2.  
3.  
Wait 10 min  
Press ZERO  
Wait for reading to stabilize at zero value.  
If you change your mind after pressing ZERO, you can  
still press EXIT here without zeroing the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation  
equations.  
M200AU returns to sampling. Immediately after  
calibration, data is not added to the DAS averages.  
02808 Rev. E  
Page 17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
Enter the expected NOx and NO span gas concentrations:  
Table 4.2 - Enter Expected Span Gas Concentrations Procedure  
Step  
Action  
Comment  
Number  
1.  
Press  
CAL-CONC-NOX  
This key sequence causes the M200AU to prompt for the  
expected NOx concentration.  
Enter the NOx span concentration value by pressing the  
key under each digit until the expected value is set.  
2.  
3.  
Press ENTR  
ENTR stores the expected NOx span value.  
Press  
CAL-CONC-NO  
Now enter the expected NO span concentration as in step  
one.  
4.  
5.  
Press ENTR  
Pressing ENTR stores the NO span value and returns the  
prompt to the CONC menu.  
Press  
EXIT  
Returns instrument to SAMPLE mode.  
If desired, compensation for converter efficiency (CE) can be included in the NOx concentration  
calculation. Prior to calibration, the converter efficiency should either be set to 1.0 or determined per  
the procedure in Section 7.6.6 of the M200AU manual.  
02808 Rev. E  
Page 18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
Table 4.3 - Span Calibration Procedure  
Step  
Number  
Action  
Press CAL  
Comment  
1.  
The M200AU enters the calibrate mode and the adaptive  
filter is activated. This allows the instrument to respond  
rapidly to concentration changes regardless of their  
magnitude.  
2.  
3.  
Wait 10 min  
Press SPAN  
Wait for reading to stabilize at span value.  
If you change your mind after pressing SPAN, you can  
still press EXIT here without spanning the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation  
equations and causes the instrument to read the NO and  
NOx span concentrations.  
M200AU returns to sampling. Immediately after  
calibration, data is not added to the DAS averages.  
02808 Rev. E  
Page 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
5.0 Maintenance  
5.1 Maintenance Schedule  
See Table 5.1 below.  
02808 Rev. E  
Page 21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
Table 5.1- Preventative Maintenance Calendar  
Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec  
Item  
Maintenance Interval  
Zero/Span Calibration  
Annually or after maintenance or  
repairs  
Zero/Span Checks  
Particulate Filter  
Ozone Flow  
Daily  
Weekly as needed  
Check every year and replace  
o’ring and orifice in reaction cell  
OzoneDryer Filter-DFU  
Sample Flow  
Replace every year  
Check every 2 months  
Check efficiency every 2 months  
Check every 3 months  
Converter  
Pneumatic Lines  
Factory Calibration  
Calibrate each year or after  
repairs  
Leak Check  
Check every 6 months  
Every year / as needed  
Replace bypass flow  
orifices  
Rebuildy bypass pump  
Every year  
02808 Rev. E  
Page 23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
5.2 Replacing the Sample Particulate Filter  
The particulate filter should be inspected often for signs of plugging or contamination. It is also  
common for dirt particles to cause instrument drift, and affect accuracy.  
To check and change the filter:  
1.  
2.  
Fold down the Bypass Pump Module front panel.  
Locate the filter on the left side of the analyzer front panel. See Figure 5.1 for an  
exploded view of the filter assembly.  
3.  
4.  
Visually inspect the filter through the glass window.  
If the filter appears dirty, unscrew the hold-down ring, remove the Teflon o-ring and then  
the filter.  
5.  
Replace the filter, being careful that the element is fully seated in the bottom of the  
holder. Replace the Teflon o-ring, then screw on the hold-down ring and hand tighten.  
NOTE:  
Handle the wetted surfaces as little as possible and clean with cloth prior to assembly  
02808 Rev. E  
Page 24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
Figure 5.1 - Replacing the Particulate Filter  
02808 Rev. E  
Page 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
5.3 Check Bypass Flow Rates  
The external sample pump is capable of maintaining the cell pressure at less than 3.0 "Hg-A. If a  
higher pressure is noted, the pump may need servicing. Check the pump and pneumatic system  
for leaks or rebuild pump. The ozone scrubber is integrated into the converter case inside the  
instrument.  
See Figure 1.1 for component locations.  
5.4 Replacing the Converter  
The heater, thermocouple, and converter assembly is designed to be replaced as a single unit.  
Check Figure 5.2 for component location.  
1.  
2.  
Turn off the power to the M501Y at the chassis front panel  
Allow the converter to cool.  
CAUTION!  
The converter operates at 315°C. Severe burns can result if  
not enough time is allowed for the assembly to cool. Do not  
handle assembly until it is at room temperature.  
3.  
4.  
Disconnect the gas fittings and power cable from the can.  
Remove the entire assembly from the NEMA chassis.  
A.  
B.  
Remove the pneumatic fittings.  
Remove the converter assembly from the chassis by loosening the 4 captive  
screws that secures the assembly to the chassis.  
5.  
6.  
7.  
Remove bottom bracket and re-attach it to the replacement converter assembly.  
Install the assembly back into the analyzer.  
Re-attach the electrical and pneumatic fittings. Leak check the assembly when  
completed. 8.  
hours, this  
Turn the power back on. The insulation can emit a burnt odor for the first 24  
is normal. Allow the converter to burn-in for 24 hours, then re-calibrate the  
instrument.  
02808 Rev. E  
Page 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
NOy Out  
NO Out  
Sample In  
Figure 5.2 - Converter Assembly  
02808 Rev. E  
Page 27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
5.5 Pneumatic Line Inspection  
Particulate matter in the pneumatic lines will affect both flow rate and response time. It is  
important that the pneumatic system be periodically inspected and thoroughly cleaned if  
necessary. Clean by disassembling and passing methanol through three times. Dry with nitrogen  
or clean zero air.  
Also inspect all pneumatic lines for cracks and abrasion on a regular basis. Replace as  
necessary.  
02808 Rev. E  
Page 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
Figure 5.3 - Converter Assembly  
02808 Rev. E  
Page 29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
5.6 Leak Check Procedure  
If a leak checker is not available, it is possible to leak check the instrument using the M501Y's  
pump plus a shut-off valve.  
1.  
2.  
Turn off instrument power and pump power.  
Cap the sample inlet port, ozone generator air inlet, and zero air inlet (if Z/S valve option  
present).  
3.  
4.  
Insert a shut-off valve between the sample pump and the vacuum manifold at the rear of  
the instrument.  
Turn on the sample pump and set the TEST function to RCEL, which measures the  
reaction cell pressure. Close the shutoff valve and monitor the cell pressure. The  
pressure should not drop more than 1"Hg(.5psi) in 5 minutes. If there is a leak, it is not  
possible by this method to tell where it is located. You can locate the leak by using a  
pressure leak checker described below.  
If you have a leak checker:  
1.  
2.  
Turn off instrument power and pump power.  
Disconnect pump at rear panel. Cap the sample inlet port, ozone generator air inlet, and  
zero air inlet (if Z/S valve option present) and connect the leak checker to the exhaust  
port.  
CAUTION: PRESSURE MUST BE LESS THAN 15 PSI.  
3.  
Pressurize system and check for leaks by watching overall pressure. The pressure should  
not drop more than 1"Hg (.5psi) in 5 minutes.  
If the instrument fails the pressure test, each fitting needs to be leak checked to find the location.  
Be careful that the system is always pressurized so as not to draw soap solution into the  
plumbing system. Make sure you dry off any accumulated bubble solution.  
The Sensor module can be leak checked as a unit using a 1/8" tubing fitting on top of the  
assembly. The same rules as above apply.  
02808 Rev. E  
Page 30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
6.0 Troubleshooting, Adjustments  
NOTE  
THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE  
PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL  
ONLY!  
This section of the manual contains information on diagnosing and repairing converter  
performance problems.  
This Section provides troubleshooting procedures that address problems to the board level. For  
component level troubleshooting, consult the schematics for the appropriate board in the  
Appendix.  
6.0.1 General Troubleshooting Hints  
1.  
Think of the M501Y as two sections:  
A.  
Section 1 - Pneumatics - Over 50% of all instrument problems are traced to leaks  
in the M200AU, pump assembly, pole mounted converter assembly, sample filter,  
or zero/span gas delivery system (calibrator). Suspect a leak first.  
Section 2 - Electronics – converter temp control board or DC power supply for  
the converter temperature readout.  
B.  
2.  
3.  
Incorrect span gas concentration - this could come either from the calibrator or entering  
the expected span gas concentration in the M501Y incorrectly, see Table 7.3 in the  
M200AU manual.  
If the instrument does not respond to span gas, check Section 9.2.3 in the M200AU  
manual.  
The above should get you started in diagnosing and repairing the most common faults. If these  
reasons have been eliminated, the next thing to do is a Factory Calibration covered in M200AU  
manual. If difficulties persist, contact our service department. The 800 telephone number is on  
the cover page of this manual.  
6.1 Operation Verification - M501Y Diagnostic Techniques  
6.1.1 Pneumatic System  
The pneumatic system is diagramed in Figure 5.3.  
02808 Rev. E  
Page 31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
6.1.2 Leak Check  
Refer to Section 5.6 for the leak check procedure.  
6.1.3 Bypass Pump Diagnostic Procedures  
The bypass pump is capable of maintaining a total bypass flowrate of 4 lpm. If flow checks  
indicate a fall off in bypass flow, the pump may need servicing. Check the pump, inlet fittings,  
and analyzer for leaks first. If other causes have been eliminated, rebuild the pump.  
6.1.4 Electrical Fault Isolation  
115 VAC is supplied to:  
1. The bypass pump.  
2. The +5 volt switcher supply.  
3. The input power to the converter temperature controller.  
Use Figure 6.1 to check the 115 VAC supply voltage to each of these components.  
The switcher supply produces 5 VDC to power the front panel display. Use Figure 6.1 to check  
the 5 VDC supply voltage to the display.  
The converter temperature is controlled by the 002560100 assembly in the NEMA chassis. The  
00257 schematic is located at the back of this manual.  
The converter temperature controller produces ±15 VDC to power the controller logic. The  
OPTO ISOLATOR component switches the 115VAC to the heater. Check the heater element  
for continuity. The DS1 LED is on when power is applied to the heater. Check for appropriate  
voltages. The temperature is sensed by a thermocouple attached to TB1. Typical thermocouple  
voltage at the 315oC setpoint is about 16 mV. U1 provides cold junction compensation and  
linearizes the output of the thermocouple. The output of pin 8 on U1 is 1/100 of the temperature.  
In other words 3.15 VDC represents 315oC.  
6.2 Setting the Converter Temperature  
A front panel-mounted programmable controller maintains the temperature of the remote  
molybdenum converter. By pressing the PV/SV button in the lower left corner of the controller,  
you can see the Present Value (actual temperature) or the Set Value (temperature set point). The  
temperature and PID control parameters have been set at the factory and no further adjustment  
should be necessary. Should you need to adjust the temperature, follow these steps:  
1. Select SV with the PV/SV button,  
2. Press the "up-arrow" under the digit you want to change,  
3. (the digit will flash),  
02808 Rev. E  
Page 32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
4. Press the "up-arrow" under the digit or the "down-arrow" at the left to scroll  
the digit to the desired value,  
5. Repeat for the other two digits,  
6. Press the ENTER button.  
Select PV with the PV/SV button to observe the actual temperature.  
6.1.4 Temperature Controller Setup  
The temperature controller has been programmed at the factory and should not need to be altered  
under normal usage.  
Indications that would require re-programming are:  
Failure / replacement of controller  
Replacement of thermocouple element  
Controller’s loss of PID parameters leading to instability of the temp setpoint.  
In the event that the control parameters are changed or in the event that a new controller is  
installed, it must be reprogrammed to suit the thermal characteristics of the instrument. The  
controller includes an Autotuning feature, which can be used to properly set the PID factors. To  
start Autotune procedure:  
Press Sel until A7 (AT) is displayed  
Press Data until value = 1 (enable Autotune)  
While Autotuning, the decimal point will blink.  
NOTE: During the procedure, it is normal for the temp to overshoot up to 100°C.  
When the decimal point LED turns off, Autotune is complete. No further adjustment  
should be necessary.  
Table 6.1 – Temperature Controller Programming Guide  
PRESS  
DISPLAY  
ACTION  
SEL X2  
DATA  
SEL  
DATA  
SEL  
UP/DOWN SET TO 5.0 (ENTER)  
P PROP BAND  
SET TO 50  
SET TO 5  
i INTEGRAL  
d DERIVATIVE TIME  
DATA  
SEL  
DATA  
SET TO 15 (FOR FAST SYSTEM)  
7C CYCLE TIME  
30 DEFAULT  
SEL  
SEL  
SEL  
SET TO 0.3  
SET TO 0  
SET TO 1 = ON  
HYS HYSTERESIS  
5rr RAMP RATE  
A7 AUTOTUNE  
DATA  
SEL  
0 = OPEN 1 = LOCKED  
2 = SV ONLY OPEN  
LOC LOCK  
02808 Rev. E  
Page 33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
SEL END RETURNS TO SV.  
SET TO 315 (315oC SET VALUE)  
NOTE:  
DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC.  
SECONDARY MENU (The fixed characteristics of the system).  
PRESS  
DISPLAY  
ACTION  
SET TO 0 – REVERSE ACTION  
SEL HOLD TILL p-n1  
SEL HOLD TILL p-n2  
SET TO 2 = TYPE J THERMOCOUPLE  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SEL  
SV  
SET TO 5  
p-dF DIGITAL FILTER  
P-SL LOWER LIMIT  
P-SU UPPER LIMIT  
P-Ab ALARM  
P-An HYTERESIS  
P-dP PROG DECIMAL  
P-48 COLD JUNCTION  
r5 RAMP TO SET POINT  
PVOF PROCESS OFFSET  
SVOF SET POINT OFFS  
P-F  
DEFAULT = 32  
SET TO 350 (350ºC)  
NONE  
NONE  
SET TO 0  
SET TO 2 (TYPE J)  
LEAVE AT 0  
LEAVE AT 0  
LEAVE AT 0  
SET TO 0  
(0 = oC, 1 = oF)  
P-n1  
HOLD TILL P  
SCROLL TO A7  
SET VALUE  
SET TO 1 (AUTOTUNE )  
SHOULD BE 315°C  
02808 Rev. E  
Page 34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 200AU-NOy Option Manual Addendum  
Figure 6.1 – Electrical Diagram  
02808 Rev. E  
Page 35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 200AU-NOy Option Manual Addendum  
7.0 M501Y Spare Parts List  
Note: Use of replacement parts other than those supplied by API may result in non-  
compliance with European Standard EN 61010-1.  
PART NO.  
00094-4  
00094-10  
01630-04  
02315  
01872  
02293A  
01357  
DESCRIPTION  
ORIFICE, 4 MIL 80 CC, O3 FLOW Rx CELL  
ORIFICE, 13 MIL 1100 CC, Rx CELL  
SAMPLE FILTER ASSY  
SENSOR ASSY, M501Y, EXCHANGE  
WELDED CONVERTER W/ O3 KILLER  
OPERATORS MANUAL FOR M501Y  
THERMOCOUPLE ASSY  
02317  
02318  
00969  
00969-1  
02230  
PS004  
01086  
M501Y EXPENDABLES KIT  
M501Y LEVEL 1 SPARES KIT  
FILTER, TFE, 47 mm Quantity 100  
FILTER, TFE, 47 mm Quantity 25  
DC POWER SUPPLY BOARD  
5, ±15 VDC, SWITCHING POWER SUPPLY  
STATUS/TEMP BOARD  
00728  
01198  
01236  
01314  
DVM DISPLAY  
ASSY, THERMOCOUPLE  
FAN, BYPASS PUMP CHASSIS  
FAN, PMT COOLER  
FL001  
FL003  
01068  
HW020  
HW036  
HW037  
OR058  
01631  
SINTERED FILTER  
FILTER, DFU  
HEATER, CONVERTER  
SPRING, FLOW CONTROL  
TFE THREAD TAPE (48 FT)  
TIE, CABLE  
O-RING, SAMPLE FILTER  
RETAINING RING, SAMPLE FILTER  
O-RING, FITTING  
OR034  
PU028  
PU030  
SAMPLE PUMP,M501Y  
SAMPLE PUMP REBUILD KIT  
02808 Rev. E  
Page 37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

Toshiba Blu ray Player BDX2000KU User Manual
Toshiba Flat Panel Television 62HM14 User Manual
Toshiba Network Card MK6026GAX User Manual
Toshiba Projection Television 52HM84 User Manual
Toshiba Tablet Accessory P50 B User Manual
Tripp Lite Switch B034 002 R User Manual
True Fitness Treadmill Z91 User Manual
Universal Security Instruments Smoke Alarm MPI305 User Manual
Vanguard Heating Oven SDVBNA User Manual
VocoPro Musical Instrument PKJ 9090PRO User Manual