Teledyne Oxygen Equipment LXT 220 User Manual

OPERATING INSTRUCTIONS FOR  
LXT-220  
Dissolved Oxygen  
Controller  
P/N Mxxxxx  
ECO:  
DANGER  
Toxic gases and or flammable liquids may be present in this monitoring system.  
Personal protective equipment may be required when servicing this instrument.  
Hazardous voltages exist on certain components internally which may persist  
for a time even after the power is turned off and disconnected.  
Only authorized personnel should conduct maintenance and/or servicing.  
Before conducting any maintenance or servicing, consult with authorized  
supervisor/manager.  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Specific Model Information  
Instrument Serial Number: _______________________  
Instrument Range:  
Calibrated for:  
_______________  
_______________  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LXT-220 DO  
Safety Messages  
Your safety and the safety of others is very important. We have  
provided many important safety messages in this manual. Please read  
these messages carefully.  
A safety message alerts you to potential hazards that could hurt you  
or others. Each safety message is associated with a safety alert symbol.  
These symbols are found in the manual and inside the instrument. The  
definition of these symbols is described below:  
GENERAL WARNING/CAUTION: Refer to the  
instructions for details on the specific danger. These cautions  
warn of specific procedures which if not followed could cause  
bodily Injury and/or damage the instrument.  
CAUTION: HOT SURFACE WARNING: This warning is  
specific to heated components within the instrument. Failure  
to heed the warning could result in serious burns to skin and  
underlying tissue.  
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous  
voltages appear within this instrument. This warning is  
specific to an electrical hazard existing at or nearby the  
component or procedure under discussion. Failure to heed this  
warning could result in injury and/or death from  
electrocution.  
Technician Symbol: All operations marked with this symbol  
are to be performed by qualified maintenance personnel only.  
NOTE: Additional information and comments regarding a  
specific component or procedure are highlighted in the form  
of a note.  
No  
Symbol  
CAUTION:  
THE ANALYZER SHOULD ONLY BE USED FOR THE  
PURPOSE AND IN THE MANNER DESCRIBED IN  
THIS MANUAL.  
Teledyne Analytical Instruments  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
IF YOU USE THE ANALYZER IN A MANNER OTHER  
THAN THAT FOR WHICH IT WAS INTENDED,  
UNPREDICTABLE BEHAVIOR COULD RESULT  
POSSIBLY ACCOMPANIED WITH HAZARDOUS  
CONSEQUENCES.  
This manual provides information designed to guide you through  
the installation, calibration operation and maintenance of your new  
process controller. Please read this manual and keep it available.  
Occasionally, some instruments are customized for a particular  
application or features and/or options added per customer requests.  
Please check the front of this manual for any additional information in  
the form of an Addendum which discusses specific information,  
procedures, cautions and warnings that may be peculiar to your  
instrument.  
Manuals do get lost. Additional manuals can be obtained from  
TI/AI at the address given in the Appendix. Some of our manuals are  
available in electronic form via the internet. Please visit our website at:  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
LXT-220 DO  
Table of Contents  
List of Figures..............................................................................ix  
List of Tables ................................................................................x  
Introduction ................................................................................ 13  
1.1 Overview  
13  
14  
14  
16  
17  
1.2 Main Features of the Transmitter  
1.3 Optional Features  
1.4 Front Panel  
1.5 Internal Connections  
Installation .................................................................................. 19  
2.1 Unpacking the Instrument  
2.2 Mounting  
19  
19  
20  
21  
21  
22  
22  
2.3 Electrical Connections  
2.3.1 AC Power  
2.3.2 Sensor Wiring  
2.4 Configuring the LXT-220  
2.5 Calibration  
Operation .................................................................................... 23  
3.1 Using the Front Panel Keys  
3.1.1 Menu Selection Keys  
3.1.2 Calibrate Keys  
3.2 Menus  
23  
23  
23  
24  
25  
26  
26  
27  
27  
3.2.1 Contrast  
3.2.2 Main Menu  
3.2.3 Output Graph  
3.2.4 Parameter Selection Menu  
3.2.4.1 Calibration Submenu  
Teledyne Analytical Instruments  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
3.2.4.2 Setup Submenu  
3.2.4.3 Status Menu  
28  
29  
30  
30  
31  
33  
33  
34  
35  
36  
38  
40  
40  
41  
42  
43  
44  
44  
45  
46  
46  
47  
3.2.4.4 Exit/Stay Menu  
3.3 Configuring the LXT-220  
3.3.1 Password Protection  
3.3.1.1 Changing Protection Level  
3.3.1.2 Changing Operational Password  
3.3.1.3 Changing Configuration Password  
3.3.2 4-20 mA Output Assignments  
3.3.3 Manual Mode Assignment  
3.3.4 HART Output Assignments  
3.3.5 Relay Assignments  
3.3.6 Temperature Calibration  
3.3.7 4-20 Trim  
3.3.8 Input A/D Converter Trim  
3.3.9 PID Assignments  
3.4 Calibration  
3.4.1 Manual Output Mode  
3.4.2 Parameter Selection Menu  
3.4.2.1 Buffer Menu  
3.4.2.2 One-Point Buffer Calibration (Standardize)  
3.4.2.3 Two-Point Buffer Calibration (Span)  
3.4.2.4 Calibration Using a Grab Sample as a Standard 48  
3.4.2.5 Calibration Error Detection & Electrode  
Operational Guidelines  
49  
49  
3.4.2.6 Back-to-Factory Calibration  
Maintenance ................................................................................51  
4.1 Routine Maintenance  
4.1.1 Cleaning  
51  
51  
51  
4.1.2 Replacement of the Microprocessor:  
Appendix .....................................................................................53  
A.1 Specifications  
53  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LXT-220 DO  
A.4 Model Designation  
55  
Index............................................................................................ 59  
Teledyne Analytical Instruments  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
List of Figures  
Figure 1-1: LXT-220 Front Panel...................................................16  
Figure 1-2: Internal Layout of the LXT-220 DO Controller .............17  
Figure 2-1: Opening the Front Panel.............................................20  
Figure 2-2: LXT-220 Electrical Connections..................................21  
Figure 3-1: Available Menus and Submenus.................................25  
Figure 3-2: Numerical Entry Using Front Panel Keys ....................32  
Teledyne Analytical Instruments  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LXT-220 DO  
List of Tables  
Table 3-1: Protection Level Exceptions......................................... 33  
Table 4-1: Troubleshooting........................................................... 52  
Teledyne Analytical Instruments  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
This is a general purpose instrument designed for liquid  
measurements in a non-hazardous area. It is the customer's  
responsibility to ensure safety especially when hot or  
flammable liquids are being analyzed since the potential of  
leaks always exist.  
The customer should ensure that the principles of operating  
this equipment are well understood by the user. Misuse of  
this product in any manner, tampering with its components,  
or unauthorized substitution of any component may  
adversely affect the safety of this instrument.  
Since the use of this instrument is beyond the control of  
Teledyne Instruments/ Analytical Instruments, referred as  
TI/AI, no responsibility by TI/AI, its affiliates, and agents for  
damage or injury from misuse or neglect of this equipment is  
implied or assumed.  
Teledyne Analytical Instruments  
xi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LXT-220 DO  
Teledyne Analytical Instruments  
xii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Introduction  
Introduction  
1.1 Overview  
Teledyne’s Model LXT-220 Dissolved Oxygen Controller is a  
robust and versatile instrument for the measurement and process control  
of dissolved oxygen in a liquid component. Highly customizable, the  
Model LXT-220 comes standard with a single 4-20mA current output  
but can have as many as five. The current output is field selectable and  
may be expanded to any range within the transmitter’s full operating  
range. It can also be programmed as reverse acting; for example, a 4-20  
mA output for a range of 0 to 10 ppm can be reversed to 10 to 0 ppm.  
The standard instrument accepts a single input typically from a  
Teledyne SP1 or SP2 Dissolved Oxygen probe however a dual input  
option is available. This optional dual input extends the functionality of  
the instrument to accept conditioned input from a variety of sensors for  
specific process measurements such as:  
Second Dissolved Oxygen (DO) channel  
Resistivity/conductivity  
pH  
Oxidation-Reduction Potential (ORP)  
Specific Ion  
For process control, the microprocessor based LXT-220  
Transmitter employs a multi-bus system architecture for PID  
temperature control and sensor-specific process variable outputs such as  
DO, pH, ORP, Resistivity/Conductivity, and ion-specific element (ISE)  
output as a function of multiple inputs. Up to six field-configurable  
alarm or control relay outputs are available to the user to match  
individual process needs. These relays can be field configured as high or  
low alarms. Hysterisis (dead band) is defined by setting the ON and OFF  
relay values. These relays can be configured to any of the input  
parameters. (See Section 3.3.5).  
Teledyne Analytical Instruments  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
LXT-220 DO  
The LXT-220 can be upgraded with pre-pHault Diagnostics &  
HART Communications option which alerts the operator to a pending  
DO, pH, ORP, or ISE electrode failure.  
The LXT-220 is housed in a rugged NEMA-4X housing with all  
operator switches conveniently located on the front panel.  
1.2 Main Features of the Transmitter  
The LXT-220 Controller for dissolved oxygen applications  
incorporates the following standard features:  
Front panel membrane switches  
Menu driven backlit display with contrast adjustment  
Automatic Calibration. This allows calibration points to be  
defined upon initialization of the transmitter.  
An easy back to factory calibration menu allows the instrument  
to return to a predetermined factory calibration. The default  
parameters are: zero electrode offset and –157.5mV/FS slope.  
4-20mA current output fully expandable to any range within the  
transmitter’s full operating range. The output is field selectable  
and fully programmable. The current output can be configured to  
be reverse acting.  
Temperature Conversion. Temperature can be displayed in either  
Fahrenheit or Celsius. The units can be toggled from one to the  
other in the field with the conversion calculation performed  
automatically.  
1.3 Optional Features  
The Model LXT-220 DO Controller is a versatile instrument  
designed to meet the needs for liquid analytics and process control  
across a wide range of applications. There are many options available  
for the transmitter including:  
Dual Input: An optional second input can be provided for pH,  
ORP, resistivity/conductivity, or Specific Ion. The second input  
can also be used as a second analysis for dissolved oxygen or as  
specific compensation for the primary input.  
Teledyne Analytical Instruments  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Introduction  
PID Output: An optional PID output is available and provides a  
4-20 mA, three-mode control signal. Proportional band is  
adjustable from 0.1 to 1,000%. Reset and Rate are adjustable  
from 0.00 to 100 repeats per minute and 0.00 to 1,000 minutes,  
respectively. The PID output can be assigned and configured to  
any of the available outputs.  
Multiple Output: The standard LXT-220 DO Controller is  
provided with a single output. With the multiple output option,  
up to four additional outputs can be added. These outputs can be  
defined in the field as the process variable (DO, pH, ORP,  
resistivity, conductivity), temperature, or PID. The additional  
outputs are fully isolated from the inputs and the primary output  
and are "floating" so they can have either a positive or negative  
common.  
Note: These isolated outputs are floating and require an external  
24 VDC power source to operate. TI/AI can supply an  
internal 24 VDC power supply to power these outputs.  
Contact the factory for additional information.  
Alarm Relays: Up to six alarm or control relays are available for  
the primary channel, or divided between the two channels on  
dual output transmitters. Relay outputs can be field configured as  
high or low alarms. Hysteresis (dead band) is defined by setting  
the "ON" and "OFF" relay values. The relays can be configured  
to any of the input parameters.  
HART Output: This option provides both digital and analog  
outputs that is useful for monitoring and controlling process  
variables in a multiple output instrument.  
Optional Software  
Differential Output: Provides an output expressed as the  
difference between two inputs.  
Ratio Output: Provides an output expressed as the ratio of two  
inputs. This is applicable to percent rejection in some  
applications.  
Averaged Output: Provides an output expressed as the average  
of two inputs.  
Teledyne Analytical Instruments  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
LXT-220 DO  
1.4 Front Panel  
The operator interfaces with the instrument from the front panel.  
Figure 1-1 shows a standard single channel LXT-220 DO instrument  
front panel. The panel includes the LCD display capable of outputting  
information in both a textural and graphical format. There are 6  
membrane switches (keys) that control the operation of the unit. The  
keys are in two groups:  
MENU SELECT: UP/DOWN (/) These keys are used to  
change the display menu and move the cursor vertically. Any  
menu can be accessed using the appropriate MENU SELECT key.  
These keys are also used to exit the calibration mode and save  
calibration data.  
CALIBRATE KEYS: UP/DOWN/LEFT/RIGHT (///)  
The horizontal CALIBRATE keys are used to enter the calibrate  
mode and move the cursor horizontally. To enter the calibrate  
mode, both horizontal CALIBRATE keys must be pressed  
simultaneously. Once in the calibrate mode, the cursor can be  
positioned by pressing the appropriate individual CALIBRATE  
key. These keys are also used to select an item from a menu. See  
Section 3.1)  
The vertical CALIBRATE keys are used to perform numeric  
adjustments to displayed values. Pressing the CALIBRATE key  
will increase the value; pressing the CALIBRATE key will  
decrease the value.  
Figure 1-1: LXT-220 Front Panel  
Teledyne Analytical Instruments  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Introduction  
1.5 Internal Connections  
All user connections are made inside the unit on removable terminal  
blocks. Two conduit holes are provided on the NEMA-4X housing, one for  
low level DC signal and the other for 115VAC power wiring. Access to the  
terminal blocks is gained after removing the 4 captive jackscrews on the  
front of the instrument. Figure 2 shows the internal layout of the LXT-220  
Controller. A label on the enclosure cover shows the terminal block  
arrangement for all the wiring along with the model number and the factory  
assignment of the output signals. These can be changed in the field if the  
application prefers a different arrangement.  
Figure 1-2: Internal Layout of the LXT-220 DO Controller  
Teledyne Analytical Instruments  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
LXT-220 DO  
Teledyne Analytical Instruments  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Installation  
Installation  
Installation of the LXT-220 Controller involves:  
Unpacking the instrument  
Mounting  
Making Electrical Connections  
Configuring the Instrument  
Calibration  
2.1 Unpacking the Instrument  
The LXT-220 Controller has been carefully packaged to protect it  
from damage during shipment and dry storage. Upon receipt please  
follow the procedure outlined below:  
1. Before unpacking, inspect the condition of the shipping container  
to verify proper handling by the carrier. If damage is noted, save  
the shipping container as proof of mishandling for the carrier.  
2. Check the contents of the shipping container with the items and  
quantities shown on the packing list. Immediately report any  
discrepancies to TI/TAI.  
3. Save the original packing material until you are satisfied with the  
contents. In the event the transmitter must be returned to the  
factory, the packing material will allow you to properly ship it to  
TI/TAI.  
4. Familiarize yourself with the instrument before installation, and  
follow proper installation and wiring procedures.  
2.2 Mounting  
Three typical installation configurations are available for the LXT-  
220:  
Universal mounting plate  
Teledyne Analytical Instruments  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
LXT-220 DO  
Handrail mounting plate  
Panel mounting configurations.  
U-bolts may be ordered separately. Refer to the following  
application drawings for mounting details:  
4024011 Dimensions, wall & pipe mounting.  
The unit is supplied with a universal-mounting bracket. Two slots  
on each side will accommodate U-Bolts for pipe mounting.  
4024012 Dimensions, Panel Mounting. This installation uses 1/4"  
screws. Panel mounting is accommodated using the captive  
jackscrew that will cinch the instrument bezel against the panel.  
2000268 Dimensions, handrail mounting.  
These application drawings can be located in the back of this  
manual.  
2.3 Electrical Connections  
All electrical connections are made with connectors to a panel  
inside the NEMA-4X enclosure. To gain access to the panel, use a flat-  
blade screwdriver to remove the 4 captive screws on the front of the  
instrument and swing the panel out. See Figure 2-1.  
Figure 2-1: Opening the Front Panel  
Teledyne Analytical Instruments  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Installation  
2.3.1 AC Power  
The LXT-220 requires a 115 VAC 60 Hz power source. A 24VDC  
unit is available as an option.  
CAUTION:  
MAKE SURE THE CIRCUIT BREAKER PROTECTING  
THE AC INPUT LINE IS OFF BEFORE WIRING THE  
LXT-220.  
Application drawing 4024003 illustrates the proper power wiring  
for single channel LXT-220 as does the affixed label on the backside of  
the front panel. The wires are best fed through the right conduit as  
shown in Figure 2-2.  
Figure 2-2: LXT-220 Electrical Connections  
2.3.2 Sensor Wiring  
A conditioned input is required from the sensor or electrode to the  
LXT-220 for proper operation. Teledyne’s SP1 and SP2 sensors will  
provide the required conditioning. The application drawing 4024003  
Teledyne Analytical Instruments  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
LXT-220 DO  
illustrates the proper sensor wiring and is also shown on the affixed label  
on the backside of the front panel.  
2.4 Configuring the LXT-220  
The LXT-220 Controller is used for measurement and control for a  
wide range of liquid analytical applications. This manual describes the  
configuration for use in dissolved oxygen measurement and control  
applications. If your instrument included optional features such as dual  
channel input for handling other specific measurement processes, a  
second manual will be included to describe the setup and operational  
parameters specific to that function.  
Configuring the instrument is described in Section 3 of this manual.  
2.5 Calibration  
Prior to using the instrument for the first time, the LXT-220 must  
be calibrated. The specific calibration procedure used depends on the  
application. Among other things, the calibration routine establishes the  
correspondence between the 4-20mA output signal and the application  
dependant measurable parameter. This manual describes operation for  
dissolved oxygen measurements and process control for the LXT-220.  
Calibration can be performed without disturbing control or recorder  
functions using a manual output mode (see Section 3.4.1). Typically, the  
calibration is done using either a one point buffer calibration  
(standardization) or two point (span) calibration. There is also a Back to  
Factory calibration feature which resets the LXT-220 to “typical  
electrode” for electrode potentials. As the electrode ages it will  
deteriorate and the mV/FS will decay and eventually require  
replacement. If the membrane is torn or coated the calibration will be  
either sluggish or out of range. This feature restores the instrument to  
the predetermined factory calibration point of zero electrode offset  
and –157.5 mV/FS quickly.  
The calibration routines require a familiarity with the instrument  
operation and is described after fundamental operational parameters are  
discussed in Section 3.  
Teledyne Analytical Instruments  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
Operation  
No one instrument has all the options described in this document.  
Some options must be specified at the time the instrument is purchased,  
and are not field-alterable. The instrument can be returned to the factory  
for re-programming if your requirements change.  
3.1 Using the Front Panel Keys  
Cursor positions or numeric adjustments are performed by pressing  
the appropriate keypad. The LXT-220 uses an "underscore" cursor in  
each of the menus. Holding down the keypad will automatically scroll  
the cursor or numeric values.  
Note: Simultaneous pressing both vertical keys or any  
combination of keys other than the two horizontal  
CALIBRATE keys is not recommended.  
3.1.1 Menu Selection Keys  
The MENU SELECTION keys are used to:  
1. Change the display menu and move the cursor vertically. Any  
menu can be accessed by the use of the appropriate up or down  
MENU SELECTION key.  
2. Exit the calibration mode and save calibration data.  
3.1.2 Calibrate Keys  
1. The horizontal CALIBRATE keys are used to enter the calibrate  
mode and move the cursor horizontally. To enter the calibrate  
mode, both horizontal CALIBRATE keys must be pressed  
simultaneously.  
2. Once in the calibrate mode, the cursor can be positioned by  
pressing the individual right-hand or left-hand horizontal  
CALIBRATE keys.  
Teledyne Analytical Instruments  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
3. With the cursor on a user selectable menu item, for instance  
SETUP in the Parameter Selection Menu, pressing both  
horizontal CALIBRATE keys will take you to the next lower  
level of screens: 4-20 mA screen, Relay Setup screen, etc.  
4. The vertical CALIBRATE keys are used to perform numeric  
adjustments to displayed values. To use these keys, the LXT-  
220 must be in the calibrate mode. Pressing the upper ()  
CALIBRATE key will increase the value; pressing the lower  
() CALIBRATE key will decrease the value.  
To exit the calibrate mode, press either of the MENU SELECTION  
keys.  
3.2 Menus  
The LXT-220 contains a series of menus and submenus from which  
different functions can be accessed. Even though the display only shows  
4 menus per screen, you can scroll through all available menus by using  
the MENU SELECT and keys.  
The menus available in the standard LXT-220 instrument  
configured for single channel dissolved oxygen controllers are shown in  
Figure 3-1.  
When the instrument is first powered up, a copyright display will  
appear for a few seconds. This screen displays the serial number and  
software version for your instrument. After a few seconds the display  
changes to the main menu. From the main menu, you can scroll up ()  
to the contrast menu or down () through the available menus for your  
controller.  
S/N 1004  
(C)TI/TAI  
V 1.00  
2002  
Copyright Display  
Some menus allow you to choose from several options. To select  
an option or item, scroll to the option using the MENU SELECT /▼  
keys. When the cursor is next to the item of interest select the item by  
simultaneously pressing both the left and right (/) CALIBRATE  
arrow keys on the right side of the front panel.  
Teledyne Analytical Instruments  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
There may be additional menus not shown in Figure 3-1 depending  
on the options you selected.  
Figure 3-1: Available Menus and Submenus  
In this manual, screens with a gray background are related to  
dissolved oxygen measurement or control while white background  
screens relate to a second channel process, for instance, ORP related  
screens. When there is only one gray screen the information applies to  
both process applications, for example, the contrast screen discussed  
next adjusts the contrast for all screens.  
3.2.1 Contrast  
Display contrast can be adjusted to allow for variations of ambient  
lighting and viewing angle by using this menu. The Contrast Menu is  
the same for all LXT-220 controllers and is always located as the  
uppermost menu. To access this menu from the Main Menu, press the  
MENU SELECT key.  
Contrast  
50  
Contrast Menu  
Teledyne Analytical Instruments  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
Procedure:  
1. Press the upper MENU SELECTION key to reach the Contrast  
Menu.  
2. Press both horizontal CALIBRATE keys simultaneously to enter  
the calibrate mode. Observe the cursor move to the last digit.  
3. Using the horizontal CALIBRATE keys, position the cursor  
under the digit to be adjusted.  
4. Increase or decrease the value by pressing the appropriate  
vertical CALIBRATE key.  
5. To save the contrast value and exit the calibrate mode, press  
either MENU SELECTION key.  
3.2.2 Main Menu  
This menu appears after the copyright display when power is first  
applied to the LXT-220. The MAIN MENU displays the measured  
process variable, temperature, and the current output in percent of full-  
scale. While in this menu, the current output can be adjusted and locked  
in a manual mode to provide an undisturbed output during sensor  
calibrations. The screens below are for a two-channel instrument. To  
access the following menus in a two-channel instrument scroll down for  
channel 1 and scroll up for channel 2. For a single channel instrument  
scrolling down accesses all menus.  
2 DO 8.40  
84.0%  
1 DO 9.28  
92.8%  
2 ORP 500  
50.0%  
25.0OC  
25.0OC  
25.0OC  
Main Menu, pH  
Main Menu, ORP  
3.2.3 Output Graph  
This screen graphically displays the output current that is set up in  
the Main screen. The X-axis of the graph display is factory set to ~150  
samples (~15min @ 6 sec or .1min/sample). The setup screen sets the  
Y-axis (the sensitivity). If the graph were assigned to a current output  
the sensitivity would be that of the 4 to 20 mA output.  
Teledyne Analytical Instruments  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
Graph  
3.2.4 Parameter Selection Menu  
Buffer  
Setup  
Status  
Parameter Selection Menu  
This menu allows you to select what function you would like to  
perform.  
With the cursor on BUFFER, if you press and together the  
LXT-220 will enter the calibration menus (standardize and span).  
Scrolling to and choosing SETUP brings you to where you can  
configure and setup the instrument. The STATUS section allows the  
user to view the calibration engineering data (mV input from the sensor  
and the zero potential from the last cal). Each section is entered by  
placing the cursor under the first letter of the section you wish to enter  
and pressing the two horizontal calibrate keys simultaneously. The  
Parameter Selection Menu is discussed in detail in Section 3.4.2.  
3.2.4.1 CALIBRATION SUBMENU  
Choosing BUFFER from the Parameter Selection Menu brings up a  
submenu that allows the user to perform a 1 or 2 point buffer calibration  
(standardize & span). This menu is described in detail in Section 3.4.2  
Calibration.  
1 DO  
Cal  
.00 ppm  
.0 mV  
Zero Calibration Menu DO  
Teledyne Analytical Instruments  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
3.2.4.2 SETUP SUBMENU  
Choosing SETUP from the Parameter Selection Menu brings up the  
first of a series of screens that allows the user to configure some initial  
parameters in the LXT-220 controller. These parameters are:  
4 to 20 mA outputs  
relay contacts  
PID control  
HART  
You can also setup how the graphic display plots the current output  
using the Plot function. The menu below shows the Plot function. In this  
case, the process variable assigned to Channel 1 is plotted and sampled  
every 6 seconds (.1 min).  
Plot  
Sample  
4-20 1  
.1 m  
(ch1 PV)  
Current output menu  
To set up additional parameters or to change to current preset  
parameters refer to Section 3.3 Configuring the LXT-220 in this manual.  
3.2.4.2.1 CURRENT OUTPUT SETUP  
This menu is used to set or adjust the current output range with the  
process variable. The top line defines the 4 mA point, while the bottom line  
defines the 20 mA point. Between these points, the output current is linear  
with respect to the sensor input. The standard LXT-220 provides a single  
output channel but options are available to expand the number of current  
outputs to 4. See Section 3.3 Configuring the LXT-220 in this manual for  
information on assigning other parameters to additional outputs.  
4mA1  
.00 ppm  
20mA1 10.00 ppm  
(now 18.84 ma)  
(ch1 PV)  
Current output menu  
Teledyne Analytical Instruments  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
3.2.4.2.2 RELAY CONTACTS  
Relay on/off set points are adjusted in this menu. The top line  
defines the point in which the relay is energized (on) and the bottom line  
defines the point in which the relay is de-energized (off). The LXT-220  
can contain up to 6 sets of relay contacts. Refer to Section 3.3.5 Relay  
Assignments to set up the number of contacts.  
1 On > 10.00 ppm  
1. Off < 9.85 ppm  
(ch1 PV)  
Relay set up menu  
3.2.4.2.3 DO LOCUS AND NOISE FILTER SETTINGS  
The Locus is the temperature where the different concentrations are  
equal and is referenced for temperature compensation. The noise filter  
is adjustable from 1 to 20 to dampen any distracting fluctuations on the  
sensor  
DO Locus  
-17.0 OC  
Noise filter  
10  
Electrode Data menu ORP  
Electrode Data menu ORP  
3.2.4.3 STATUS MENU  
This menu allows the user to view the time of day clock and the  
engineering information pertaining to the electrode as an informational  
screen only. The top line displays the real-time (active) absolute  
millivolt value (mVa) the electrode is generating, uncompensated for  
temperature variations. The bottom line displays the electrode zero  
potential from the last calibration.  
This information is extremely helpful in determining the diagnostic  
status of an electrode. In conjunction with regular calibrations, the  
Electrode Diagnostic Screen provides valuable information to track  
electrode performance.  
Teledyne Analytical Instruments  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
Input  
0 mV  
-150.4 mVa  
.00ppm  
Fri 12:00:37  
(volatile)  
Electrode Data menu ORP  
Time of day clock  
3.2.4.4 EXIT/STAY MENU  
This menu confirms that the user wishes to leave the section that  
they are currently in. This menu will show up each time that the LX-  
Exit Stay  
Exit / Stay menu  
220 is prompted to leave the current section before it goes back to the  
parameter selection menu. Press the appropriate menu selection key  
either to stay in the current section or exit and return to the parameter  
selection menu.  
3.3 Configuring the LXT-220  
To initially configure or reconfigure the instrument, use the left-  
side and MENU SELECT keys to scroll to the bottom of the  
menus list and stop when the CONFIGURE/TRIM option is displayed.  
Simultaneously depress theand keys to enter the configuration  
mode. This mode should be password protected if access to the keyboard  
is unrestricted. The new configuration remains in effect through  
indefinite power outages.  
The configuration mode is a submenu grouping of 9 or more  
functions that can be set by the user to tailor the instrument to a  
particular application. The following 7 functions appear in groups of 4  
for a standard instrument:  
PASSWORDS  
4-20 ASSIGN  
4-20 MANUAL MODE  
RELAY ASSIGN/TST  
°C/°F & TEMP CAL  
4-20 TRIM/TEST  
Teledyne Analytical Instruments  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
INPUT A/D CONVERTER TRIM  
Some instruments have two channels measuring two unrelated  
quantities, and in those cases both channels can be configured from the  
same configuration menus.  
Depending on options selected 2 additional submenus could appear:  
PID ASSIGN  
HART ASSIGN  
Scroll down to the item of interest, and then simultaneously depress  
the and keys to access the sub-menu for that item.  
3.3.1 Password Protection  
It is worthwhile to become familiar with the password capabilities  
of the instrument even if you decide not to use them, as it is possible that  
one gets set by mistake and you need to change or erase it.  
Password protection can be selectively turned on for the operational  
and configuration displays. The operational and configuration passwords  
can be different. Passwords are entered using only the right four keys,  
and the passwords are numbers consisting only of the digits 1 to 4. No  
digits are shown on the keyboard. It is necessary to remember that the  
top key is 1, the right key is 2, the bottom key is 3, and the left key is 4.  
See Figure 3-2.  
Teledyne Analytical Instruments  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
Figure 3-2: Numerical Entry Using Front Panel Keys  
After successfully entering the operational password it remains  
valid for 60 seconds after exiting the calibration mode on that display.  
During that time a second display can be accessed without re-entering  
the password. The 60 second timer starts again when exiting the cali-  
bration mode on that display also, so that any number of operational  
displays can be modified with a single password entry.  
If the wrong password is entered three consecutive times in  
accessing a protected display all password access is disabled, and the  
main display shows “Err PW”. The instrument continues to function  
normally, but none of the password-protected settings can be changed.  
Normal operation is restored by momentarily turning off power. No  
keyboard commands are required to restore normal password protection.  
The Err PW display is cleared by any keystroke, but when attempting to  
enter a password the display will still show disabled.  
If the password is lost and you want to gain access to protected  
screens and/or enter a new password, turn off the power, and then reapply  
it. Within one second of the time that text first appears on the screen,  
press the rightmost key on the keyboard. The screen will go blank to  
acknowledge the key press, but there will be no other visible indication  
that the key is recognized. After a few seconds the instrument will begin  
functioning normally except that the configuration password has been  
disabled for 60 seconds. The operational password remains in effect.  
Immediately enter the configuration mode and choose a new password.  
Note: This method will not work if the key is down at the time  
power is applied. The key must be depressed shortly after  
text first appears.  
The following displays control the password settings.  
Operate password level:  
3
no PW  
Teledyne Analytical Instruments  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
The protection level for the operational displays is the number on  
the left side of the second line. None of the operational displays are  
password protected if the protection level is 0. If it is not zero then, with  
the exceptions shown in Table 3-1, all displays that affect the operation  
of the instrument are protected. An N entry in the table means that the  
display is not protected.  
Table 3-1: Protection Level Exceptions  
level  
contrast  
man out  
1-point buf  
1
2
N
Y
N
Y
N
N
3
4
Y
Y
N
Y
Y
Y
3.3.1.1 CHANGING PROTECTION LEVEL  
To change the protection level, navigate using the right side  
(CALIBRATE) / keys to the old protection level number (3 in the  
above screen). Adjust the numerical value using the right side  
(CALIBRATE) /keys and then accept the input by pressing either  
of the left side MENU SELECT) /keys. The old operational  
password is erased if the protection level is changed to 0. Changing from  
0 to another value has no effect until a new password is entered. The  
display shows no PW if the password has been erased.  
3.3.1.2 CHANGING OPERATIONAL PASSWORD  
To enter a new operational password, navigate to the old password  
(or no PW) using the (CALIBRATE) / keys and select by pressing  
both (CALIBRATE) / keys simultaneously. This brings up the new  
operational password screen.  
New opr password  
Teledyne Analytical Instruments  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
Press both (CALIBRATE) / keys again to enter a new  
operational password. 1 to 8 digits may he entered. Press either of the  
menu keys to accept the new password when done.  
Note: A new password will not be accepted unless the protection  
level has already been set to a value greater than 0.  
3.3.1.3 CHANGING CONFIGURATION PASSWORD  
To change the configuration password, use the MENU SELECT  
/keys to navigate to the PASSWORD menu and select it by  
pressing both CALIBRATE / keys simultaneously. Use the MENU  
SELECT /keys to bring the cursor to the Configure password level  
screen.  
Config password level:  
1
no PW  
To enter a new configure password, navigate to the old password  
(or no PW) using the (CALIBRATE) / keys and select by pressing  
both (CALIBRATE) / keys simultaneously. This brings up the new  
configure password screen.  
New opr password  
Press both (CALIBRATE) / keys again to enter a new  
configure password. 1 to 8 digits may he entered. Press either of the  
menu keys to accept the new password when done.  
Note: A new configure password will not be accepted unless the  
protection level has already been set to 1.  
These two displays work the same as the corresponding operational  
displays, except that the protection level can only be 0 or 1.  
Teledyne Analytical Instruments  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
The configuration password is rarely used and the configuration  
settings are more sensitive, so the password should use more digits than  
the operational password. With eight base 4 digits the odds of breaking  
the code in three tries are 3 in 65536 if the user knows that there are 8  
digits. The user will not know the number of digits that you chose, so  
the odds are usually smaller than that. The maximum number of digits  
allowed in a password, or the number actually used, cannot be  
determined by entering trial passwords. A two or three digit password is  
therefore moderately effective if systematic attempts at breaking the  
code are not anticipated, but short passwords should nevertheless be  
avoided in a critical process.  
3.3.2 4-20 mA Output Assignments  
The next group of displays are used to assign the 4-20 mA outputs.  
Note: The LXT-220 has been configured at the factory for your  
specific application as specified at the time of the order.  
Using the MENU SELECT /keys, scroll down to the 4-20 mA  
ASSIGN option and select it by pressing both CALIBRATE / keys  
simultaneously. The screen will display 4-20 1: where . . . is one of the  
following selections:  
missing  
unused  
ch1 PV  
not installed  
Installed but not currently used  
Channel 1 process value  
Channel 1 temperature  
Channel 1 PID controller output  
Channel 2 process value  
Channel 2 temperature  
Channel 2 PID controller output  
ch1 temp  
ch1 PID  
ch2 PV  
ch2 temp  
ch2 PID  
The channel 2 options are only shown on two channel instruments,  
and specialized instruments may show additional options.  
To change the output assignment, first push both CALIBRATE /  
keys, causing the cursor to move to the right. Then use the right-side  
CALIBRATE /keys to scroll through the list of options. The list is  
Teledyne Analytical Instruments  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
cyclic. Use either of the MENU SELECT /keys to accept the  
selection and move the cursor back to the left. The displayed selection  
remains in effect until changed and the selection is remembered through  
power outages of indefinite length.  
Up to two 4-20 outputs can be assigned to the same source. There is  
no error indication if more than two are assigned to the same source, but  
there will be no span display for the third. A span display will otherwise  
be present in the main menu for each 4-20 output that is assigned to a  
source. If the source is associated with channel 2 then the span display  
will be in the channel 2 group. The new assignment is not functional  
until the setpoints in the span display are set. If a PID controller is  
selected as a source then its support displays will also be present.  
The next two displays are for the second and third 4-20mA outputs.  
Some units do not have all three outputs, but the displays will show  
three output options anyway, as the additional outputs can be retrofitted.  
The 'missing' assignment should be selected for uninstalled outputs. The  
software cannot determine how many 4-20 outputs are installed, so the  
'missing' selection might not reflect the actual configuration.  
Output #4 is dedicated to HART, and its assignment is discussed in  
the HART section. If the instrument is not connected to a HART  
network then the fourth output can be used like any of the other outputs.  
The fourth output is only present if the instrument was purchased with  
the HART option.  
A span display will be present in the setup menu for each 4-20  
output that is assigned to a source unless the source is a PID controller.  
PID controllers have different setup displays, and they are in the same  
section. If the source is associated with channel 2 then the display will be  
in the channel 2 group. The new assignment is not usable until the  
setpoints or PID parameters are specified in the operational displays.  
The third line of the display will show a line such as (ch1 man  
mode) if an assignable manual mode control is associated with the  
output. The line will begin with a question mark if making the current  
selection permanent would result in a manual mode assignment conflict.  
Refer to Section 3.3.3 Manual Mode Assignment for more information.  
3.3.3 Manual Mode Assignment  
In the manual output mode, the current output is manually set to the  
desired level and saved until changed or released from the manual mode.  
The main menu display shows the process value and a 4-20 mA percent  
Teledyne Analytical Instruments  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
output field for each channel. The manual mode assignment allows the  
percent output display to be associated with any of the 4-20 outputs, in-  
cluding the HART output. When in the manual mode the 4-20 output is  
controlled from the keyboard. Zero percent corresponds to a current of 4  
mA, and 100 percent corresponds to a current of 20 mA. If a 4-20 output  
is used to control the process then the manual mode is normally assigned  
to that output, although the instrument does not require this choice.  
Cross-channel assignments are allowed in that the selected 4-20 can be  
assigned to something on the other channel. Cross-channel assignments can  
usually be avoided, and they should be avoided when possible.  
ch1 man mode on 4-20:  
3
(ch1 PV)  
To change the assignment simultaneously depress both  
CALIBRATE / keys, then use the right-side CALIBRATE /▼  
keys to change the selection. One of the selections is “none”, and with  
that selection the percent output field is not present in the operational  
display. The bottom line shows the signal that the selected 4-20 is  
assigned to. On two channel instruments the next display is the manual  
mode assignment for channel two.  
Each 4-20 output can have only one manual mode control assigned  
to it. If one already exists for the current selection it will be preceded by  
a question mark. Some instruments have more than two manual mode  
controls in the operational displays, and they must all he assigned to  
different 4-20 outputs if they are activated.  
The other manual mode controls are built-in in some instruments.  
An example would he the weight-percent display of a conductivity unit.  
The manual mode field appears in a separate display in these cases, and  
it is independent of the manual mode assignments for the main display.  
The manual mode capability is automatically activated when a 4-20 is  
assigned to one of these functions. The percent output field does not  
appear in the operational display unless a 4-20 is assigned to it.  
Conversely, when changing the 4-20 assignments a conflict can  
arise if a 4-20 with a pre-existing manual mode control is reassigned to  
Teledyne Analytical Instruments  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
a function with its own manual mode output. The 4-20 assignment  
display will show if the output is associated with an assignable manual  
mode. Regardless of the cause, the affected operational display manual  
mode fields will begin with a question mark when a conflict exists.  
When that happens refer to the manual mode configuration displays, and  
HART 1: …..  
the question marks there will show where the problem lies.  
No contention occurs unless two different displays are actually in  
the manual mode for the same 4-20, so the question marks in the  
displays can be ignored if the assignments are made for diagnostic  
purposes.  
3.3.4 HART Output Assignments  
If the instrument was purchased with the HART option then the  
next five displays are for assigning the HART digital and analog  
outputs. The first of these displays is:  
where… represents one of the items from the 4-20 mA assignment  
list in Section 3.3.2.  
HART command #3 requests four variables from the instrument.  
Each of these variables can be assigned to any of the sources shown in  
the 4-20 assignment section. Each variable has a units code, which  
identifies the variable as being pH, millivolts, degrees C, etc. The units  
code is supplied automatically depending on the assignment and the  
instrument configuration. If command 3 is not used then only the first  
output assignment will be reported. If any output is assigned to  
“missing” or “unused” then that value will be reported as “not a  
number”, and the units field will be “not used”.  
HART analog 1:  
The first digital output assignment is referred to as the primary  
variable in the HART system.  
Teledyne Analytical Instruments  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
HART analog 2:, HART analog 3:, HART analog 4:  
The next three displays assign the second, third, and fourth  
variables. The assignments are changed in the same way as for the 4-20  
outputs. See Section 3.3.2 4-20mA Output Assignments.  
It is not possible to assign more than one HART digital output to  
the same source. The current selection will be preceded by a < symbol  
on the display if making it permanent would result in duplication.  
HART analog: 5  
4-20 output #4 is dedicated to HART. This output normally carries  
the digital information superimposed on the 4-20 mA current. But if the  
HART system uses the multi-drop mode then the current output of 4-20  
#5 is disabled, with only the digital signal remaining. Several instruments  
are connected in parallel in the multi-drop mode, so analog signaling is  
not possible. The current that would have been output is still reported  
digitally, as a percentage, and that will be sufficient for some applications.  
But if an actual 4-20 current is required then one of the other 4-20 outputs  
must be used to supply it, and this display is used to make the assignment.  
The specified 4-20 must in turn be assigned to an appropriate source.  
Note: If the HART analog output is assigned to an output other  
than #5, then #5 cannot be assigned to anything. It must  
be set to “missing” or “unused”.  
The span settings for the selected 4-20 output correspond directly to  
the HART lower and upper range limits, and they can be set either from  
the keyboard or with command #35. If they are changed with the  
command then the new value will be shown on the instrument's span  
display for the selected output. If they are changed from the keyboard  
then those values are reported digitally over the HART network.  
If command 35 is to be used to set the limits then the HART 4-20  
output must be selected first because the output span does not exist until  
an assignment is made. The span settings must be readjusted or be  
resent when the HART analog output assignment is changed.  
The 4-20 supplying the HART analog output is normally assigned  
to the same source as the primary variable, and the HART displays show  
the engineering units for the upper and lower range limits as being the  
same as those of the primary variable. It is possible to assign the analog  
output to other sources, but other assignments are non-standard, and the  
units displayed by the remote system will not be correct.  
Teledyne Analytical Instruments  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
If the 4-20 selected for the HART analog output is assigned to  
“missing” or “unused” then the digital representation of the output  
current will be “not a number”, but with up to four digital variables still  
being transmitted.  
3.3.5 Relay Assignments  
The next 8 displays are for the relay assignments. The first of the  
relay displays is:  
Relay 1: ...  
Where ... represents an item from the same list that is used for the  
4-20 assignments. The procedure for selecting a relay assignment is the  
same as for the 4-20 outputs. See Section 3.3.2 4-20mA Output  
Assignments.  
Up to two relays can be assigned to the same source, so one relay  
can be used to control the process, and another relay can be used as an  
alarm if the process value is out of limits.  
The relay configuration menu provides for 8 relays, but most units  
do not have that many relays installed. Some units do not have any  
relays installed, but the relay menu will nevertheless be present as the  
relays can be retrofitted. The software cannot determine how many  
relays are actually present, so the “missing” selection might not reflect  
the actual configuration.  
A display showing the relay setpoints for each relay is present in the  
main menu. If the relay is assigned to a channel 2 function then the display  
will be in the channel 2 group. The display is not shown if the “missing” or  
“unused” selection is made. A new relay assignment is not necessarily  
functional until the setpoints in the span display are set. If a PID controller  
is selected as a source then its support displays will also be present.  
3.3.6 Temperature Calibration  
The next display is for calibrating the process temperature  
displayed by the instrument.  
Trim OC  
Unit  
.0  
25.0OC  
Temperature Calibration  
Teledyne Analytical Instruments  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
The bottom line shows the measured temperature. To change the  
displayed temperature so it agrees with a measured value, place the  
cursor on the top line then simultaneously depress both CALIBRATE  
/ keys. Use the right-side CALIBRATE /keys to adjust the  
number so that the displayed temperature agrees with that measured by a  
thermometer.  
The number shown on the top line is simply added to the  
temperature that the instrument measures. There is no slope correction.  
To toggle between degrees C and degrees F, place the cursor on the  
bottom line and simultaneously depress both CALIBRATE / keys.  
This will change the unit. Press both keys again to toggle back to the  
original unit.  
All temperature displays on both channels use the selected system  
of units. The scale and units code in the HART temperature output is  
also controlled by this selection.  
On two channel instruments the next display trims the channel 2  
temperature.  
3.3.7 4-20 Trim  
The next six displays are used to trim the 4-20 outputs. The 4-20  
outputs are calibrated at the factory and will never need recalibration in  
most applications. Periodic recalibration is not recommended.  
4mA1 trim  
.000  
4 to 20 mA trim  
As in the other menus, the cursor moves to the right when both  
CALIBRATE keys are depressed and the output current is set to 4.00  
mA, overriding the output assignment. Measure the 4-20 output #1  
current with an ammeter, and adjust the number on the top line until the  
meter reads 4.00 mA.  
Teledyne Analytical Instruments  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
The next display is for trimming the 20 mA point of output #1 and  
it works the same as the 4 mA trim. The 4 mA and 20 mA trims do not  
interact, and the endpoint trims are linearly interpolated for other output  
values.  
Note: The 4-20 #5 output must be trimmed by standard  
commands sent over the HART communication link.  
3.3.8 Input A/D Converter Trim  
This display is used to calibrate the analog to digital converter that is  
used to measure the process value. The bottom line shows the measured  
input voltage. The scale factor on the top line is adjusted until the  
displayed voltage is the same as that of a voltmeter connected to the input.  
If the process is stable then the voltage can be the output of a sensor, but it  
is better to replace the sensor with a fixed voltage source. For optimum  
accuracy, the calibration voltage should be a substantial portion of the full  
scale voltage. It can be of either polarity. The input zero point error is  
automatically removed, so there is no adjustment for it.  
1 in gain  
400.0 mV  
1.0000  
Input A/D converter trim  
The normal input range is +/- 600 millivolts, but ORP and some  
specific ion units use a +/-1200 millivolt range. On conductivity units  
the analog voltage is not externally accessible, but conductivity systems  
do not have millivolt displays, so no calibration is required. On  
conductivity units this display is only used for factory testing.  
The input scale factor is calibrated at the factory and will never  
need recalibration in normal applications. Periodic recalibration is not  
recommended, as most units will remain accurate to within about 0.2  
percent for many years. Further, any inaccuracy in the displayed  
millivolts is calibrated out when buffer calibrations are performed, so  
there is no requirement that the displayed millivolts be highly accurate.  
If the scale is accidentally changed and it is not convenient to perform a  
voltmeter calibration then simply set the scale factor to 1.0000. Most  
units are accurate to 2 percent with that setting, which is adequate for  
most applications.  
Teledyne Analytical Instruments  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
On two channel units the next and last display is for trimming the  
channel 2 analog input.  
3.3.9 PID Assignments  
Two independent Proportional Integral Derivative (PID) controllers  
are available as options on all instruments, including one channel units.  
The sub-menu for assigning the controller inputs is:  
PID A: …..  
PID B: …..  
Where . . . . represents one of the items from the 4-20 mA  
assignment list in Section 3.3.2. The PID controller outputs in turn  
represent separate signal sources, and various outputs can be assigned to  
them.  
Not all signal sources are suitable as inputs to a PID processor, and  
an asterisk on the display marks the selections that are unavailable.  
Potentially useful sources, such as the process value, temperature, and  
quantities derived from the measured process value(s) are normally  
available. The assignment should be set to “unused” if no controller is  
needed. This will prevent the PID setup displays from appearing in the  
operational section.  
Either PID processor can be assigned to a signal from either  
channel, and they can both be assigned to signals on the same channel,  
but they cannot be assigned to the same signal on the same channel. The  
current selection will be preceded by a question mark on the display if  
making it permanent would result in duplication.  
Up to two 4-20 mA outputs and up to two relays can be assigned to  
each PID processor output. The output of the PID processor is discarded  
if nothing is assigned to it.  
If a 4-20 mA output is assigned to a PID processor there will be no  
display for it in the setup section of the operational displays, as the 4-20  
output scaling is fully specified by the PID setup parameters in the same  
section. If relays are assigned to a PID output there will be setpoint  
Teledyne Analytical Instruments  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
displays for them in the setup section, and they can be used for alarm  
purposes or otherwise. The bottom line of the display will identify the  
relay as being assigned to a PID output, and it will also show what the  
input to the PID processor is assigned to.  
A 0% relay setpoint refers to the minimum PID output, which  
corresponds to a current of 4 mA on any current loops assigned to the  
PID, and a 100% setpoint corresponds to a 20 mA current. When 4-20  
mA outputs are assigned to a PID output, it is helpful to assign one of  
the manual mode displays (Section 3.3.3) to one of the them so that the  
percent output of the controller will be visible in the operational  
displays.  
See Section 3-5 Proportional Control Setup and Configuration for  
configuring and setting up the optional PID control feature.  
3.4 Calibration  
Calibration involves performing either a one-point standardization  
calibration or two-point span calibration. These choices are available  
after choosing BUFFER from the Parameter Selection Menu  
3.4.1 Manual Output Mode  
In order to perform calibrations without interfering with control or  
recorder functions, the LXT-220 incorporates a manual output mode. In  
the manual output mode, the current output is set to the desired level and  
saved until changed or released from the manual mode. On LXT-220s  
with the optional PID controller output on channel 2, the current output  
displayed is the PID controller output. The following procedure  
demonstrates the use of the manual output.  
Note: Prior to any calibration, The LXT-220 controller should be  
placed into the manual mode  
Teledyne Analytical Instruments  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
DO  
M 50.0%  
7.80  
25.0OC  
Main Menu, DO  
Procedure:  
1. If not at the Main Menu, press the appropriate MENU  
SELECTION key to reach the Main Menu.  
2. Press both horizontal ◄►CALIBRATE keys simultaneously to  
enter the calibrate mode and observe the "M" appearing in front  
of the % current output value. The "M" signifies that the manual  
output is locked in the manual mode.  
3. Using the horizontal CALIBRATE keys one at a time, position  
the cursor under the digit to be adjusted.  
4. Increase or decrease the value by pressing the appropriate  
vertical /CALIBRATE key.  
To release manual output:  
1. Return to the Main Menu by pressing the appropriate MENU  
SELECTION key.  
2. Observe that the "M" appears in front of the % current output  
value. This "M" signifies that the transmitter output is in  
manual.  
3. Press only the left-hand CALIBRATE key and observe the  
"M" disappear, releasing the LXT-220 from the manual output  
mode. The % output value will return to the real-time current  
output.  
3.4.2 Parameter Selection Menu  
Scroll down from the main menu to reach the parameter set up  
menu. Once in the Parameter set up menu select the function you would  
like to enter.  
Teledyne Analytical Instruments  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
Buffer  
Setup  
Status  
BUFFER: Perform a 1 or a 2 point calibration  
SETUP: Set up the range of the 4-20 mA output  
STATUS: Information only. Shows the data stored from the last  
calibration.  
3.4.2.1 BUFFER MENU  
From the parameter selection menu chose the top line (the  
BUFFER command) and press the two horizontal ◄► CALIBRATE  
keys simultaneously to enter the BUFFER calibration submenu. Both  
the one-point calibration (standardize) and the two-point (span)  
calibrations are available within this menu. The procedures for each  
calibration function are given below.  
3.4.2.2 ONE-POINT BUFFER CALIBRATION (STANDARDIZE)  
Because all dissolved oxygen electrodes experience minor  
variations, buffer calibrations are necessary before installing the  
electrodes in service. Also, occasional calibrations are necessary to  
compensate for electrode degradation while in service. For  
measurement specifics refer to the dissolved oxygen sensor manual.  
1DO  
Cal  
.00 ppm  
.0 mV  
Standardization menu pH  
Procedure:  
1. Be sure that the cursor is located under the "1" as illustrated.  
2. For diagnostic reasons note the ppm and mV values from the  
prior calibration.  
Teledyne Analytical Instruments  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
3. Clean the electrode and insert the sensor into the desired buffer.  
4. Press both horizontal CALIBRATE keys simultaneously to enter  
the calibrate mode. Observe the diagnostic value on the bottom  
line change from historical data to a real-time value and that the  
term "CAL" has disappeared.  
5. To change the buffer point, position the cursor under the desired  
digit using the left-hand CALIBRATE key.  
6. Increase or decrease the value by pressing the appropriate  
vertical CALIBRATE key.  
7. Wait for the real-time mV value to stabilize.  
8. To save the calibration and exit the calibrate mode, press either  
MENU SELECTION key.  
A perfect electrode slope (efficiency) is typically greater than -150.  
mV/FS. If the value falls below -100.0 mV/FS, the electrode should be  
serviced or replaced.  
3.4.2.3 TWO-POINT BUFFER CALIBRATION (SPAN)  
When first installing an electrode, the second cal point  
establishes the slope variations or efficiency. The span is typically done  
in air which is the saturation point of the solution when in equilibrium.  
2DO  
Cal  
8.40 ppm  
-173.8 mV/FS  
Span calibration menu pH  
Procedure:  
1. Perform a one-point calibration as outlined in Section 3.4.2.2.  
2. Press the appropriate MENU SELECTION key to reach the  
Electrode Span Menu and locate the cursor under the "2" as  
illustrated.  
Teledyne Analytical Instruments  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
3. For diagnostic reasons note the DO and mV/FS values from the  
prior calibration.  
4. Press both horizontal CALIBRATE keys simultaneously to enter  
the calibrate mode. Observe the diagnostic value on the bottom  
line becomes a real-time value and that the term "CAL" has  
disappeared.  
5. To change the buffer point, position the cursor under the desired  
digit using the left-hand CALIBRATE key.  
6. Increase or decrease the value by pressing the appropriate  
vertical CALIBRATE key.  
7. Wait for the real-time mV value to stabilize.  
8. To file the calibration and exit the calibrate mode, press either  
MENU SELECTION key.  
A perfect electrode slope (efficiency) is typically greater than -  
150.0 mV/FS. If the value falls below -100.0 mV/FS, the electrode  
should be serviced or replaced.  
IMPORTANT:  
If a flashing asterisk appears during or after  
calibration, the electrode may require service, or  
the buffer solution may be contaminated.  
Note: Comparing the millivolt values before and after the  
calibration provides a valuable diagnostic tool in  
determining the degradation of the electrode. Large  
differences between calibrations may indicate coating or  
damage to the measurement half-cell.  
3.4.2.4 CALIBRATION USING A GRAB SAMPLE AS A STANDARD  
The LXT-220 allows easy standardization to a grab sample value  
by placing the process value into the first calibration point.  
Procedure:  
1. Record the O2 value of the process when the sample is extracted.  
2. Record the O2 value of the grab sample.  
3. Calculate the difference between the two values as follows:  
(Grab sample value) - (Displayed value) = differential  
Teledyne Analytical Instruments  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Operation  
4. Immediately before making the calibration adjustment, note the  
current value on the display and add the differential to the  
current value. This "total adjusted" value is entered as the buffer  
value.  
3.4.2.5 CALIBRATION ERROR DETECTION & ELECTRODE OPERATIONAL  
GUIDELINES  
If the slope calibration performed in Section 3.4.2.3 is not within  
the predetermined limits of the instrument, a flashing asterisk (*) will  
appear.  
The presence of the asterisk indicates a potential calibration  
problem. In general, if the asterisk appears, the integrity of the  
electrode, the buffer solutions or the handling procedure should be  
questioned. See the dissolved oxygen sensor manual for specifics on the  
electrode.  
3.4.2.6 BACK-TO-FACTORY CALIBRATION  
This feature allows the LXT-220 to be reset to "typical electrode"  
for electrode potentials. As the electrode ages it will deteriorate and the  
mV/FS will decay and will eventually require replacement. If the  
membrane is torn or coated the calibration will be either sluggish or out  
of range.  
The following procedure for returning the LXT-220 to factory  
calibration can be used for both standardize and span calibrations.  
1 DO  
Cal  
.00 ppm  
.0 mV  
2 DO  
Cal  
10.0 ppm  
-157.5 mV/FS  
Standardization menu DO @ factory default  
settings  
Standardization menu @ factory default  
settings  
Procedure at the Standardization Menu:  
1. Position the cursor under the "C" in "Cal."  
2. Press both horizontal CALIBRATE keys simultaneously and  
observe the cursor briefly move to the right. The default setting  
is complete.  
Teledyne Analytical Instruments  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
LXT-220 DO  
3. This procedure is required at BOTH of the calibration points.  
Teledyne Analytical Instruments  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Maintenance  
Maintenance  
4.1 Routine Maintenance  
Aside from normal cleaning routine maintenance is limited to  
recalibration. For recalibration, see Section 3.4 Calibration. However, if  
service of the internal components is required always turn off the power  
to the instrument.  
4.1.1 Cleaning  
Although the LXT-220 NEMA 4X enclosure can withstand harsh  
environments, it may become necessary to clean the front panel if it  
becomes coated or dirty. Cleaning the front panel can be performed  
with a detergent and water.  
CAUTION:  
DO NOT USE ACETONE, ACIDS OR CAUSTIC  
SOLUTIONS ON THE ENCLOSURE SURFACE.  
Before cleaning, the integrity of the enclosure seals should be  
inspected. Also check the conduit fittings and seals to make sure  
moisture does not enter the enclosure while cleaning.  
If the enclosure cover must be removed, it is wise to clean and  
inspect the gasket seal. If the seal is damaged, replace the gasket.  
Always keep the gasket lightly lubricated with a silicone grease.  
4.1.2 Replacement of the Microprocessor:  
For replacement of the microprocessor, contact the factory.  
CAUTION:  
NEVER INTERCHANGE EPROMS FROM ONE  
TRANSMITTER TO ANOTHER. DOING SO WILL  
AFFECT TRANSMITTER PERFORMANCE.  
Teledyne Analytical Instruments  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
LXT-220 DO  
Table 4-1: Troubleshooting  
SYMPTOM  
POSSIBLE CAUSES  
SUGGESTED  
ACTION  
The LCD does not  
display.  
No power to the  
instrument.  
Check power supply  
to be sure the correct  
voltage is being  
supplied to the  
transmitter.  
Contrast level is set  
too low.  
Set contrast level to a  
higher value.  
Because the contrast  
menu is always the  
top menu, it can be  
reached by pressing  
the upper MENU  
SELECTION key at  
least 12 times. Enter  
the calibrate mode  
and increase the  
value.  
Sensor or signal  
conditioner has a  
short that draws too  
much power from the  
transmitter.  
Check the sensor for  
proper operation. To  
verify, disconnect all  
sensor wires and  
cycle power off, then  
on.  
The PROM has not  
been properly  
Check orientation of  
the PROM to the  
installed in the socket. socket. Make sure the  
pins are fully inserted  
and are not bent.  
Readings are not  
linear at the low end  
of the range.  
The sensor range is  
not compatible with  
the instrument.  
Verify the sensor  
range. Install a  
sensor with the  
correct range.  
Erratic readings.  
Air bubble in sensor.  
Check orientation of  
sensor.  
Teledyne Analytical Instruments  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Appendix  
Appendix  
A.1 Specifications  
Packaging: NEMA-4X weatherproof enclosure  
suitable for general purpose application  
Power Requirements: 115/220 VAC or optionally 24 VDC 0.25A  
Display: Menu driven Supertwist LCD.  
Simultaneously displays process ID,  
Process value in Eng. units, % output and  
temp in °C or °F.  
Sensor: Application specific  
Input Measurement Parameters:  
Conductivity, Resistivity, pH, Specific  
Ion, Dissolved Oxygen, ORP.  
Also, external events and analog inputs  
can be integrated logically or operational  
modifiers.  
Outputs: 4-20 mA, HART or RS232 up to a  
maximum of 6 analog outputs.  
Response Time: 90% of step change in 1 second  
Accuracy: ±0.10% of full scale  
Linearity: ±0.05% of full scale  
Sensitivity: ±0.05%  
Stability: ±0.2% per year @ 0 to 70°C  
Repeatability: ±0.1%  
Input/Output Isolation: Max 300V between process input and any  
4-20mA output  
Teledyne Analytical Instruments  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
LXT-220 DO  
Maximum Loop Impedance: 800Ω on 4-20 outputs with internal  
24VDC  
Digital Interface: HART option  
Operating Temperature: -4° to 158°F (-20° to 70°C)  
Dimensions: 5.67” H, 5.67” W, 6.97” D  
Temp. Compensation: Automatic, -30 to 140°C, RTD. Accuracy  
within ±1.0°C over a 0-100°C span.  
50/60Hz Noise Rejection: Greater than 70 db.  
Calibration: Auto calibration allows the definition of  
two calibration points, saved in memory,  
during the initial startup. This will allow  
subsequent standardization and span  
calibrations with only 2 keystrokes. Cal  
parameters initiated by configuration  
settings.  
Configurability: Parameter graph. Relays (up to 8) solid  
state or SPDT mechanical, 8 internal  
clocks and timers (one-shot and periodic),  
4 current outputs (4-20mA) and 2 PID  
current outputs, logical function (and/or  
gates) can all be setup as desired in the  
configuration menus.  
5A switching mechanical relays, .065-3A  
230VAC solid state with 600 W total load  
without additional heat sinking  
precautions.  
Password Protection: Multilevel password configuration allows  
selection of protected parameters in  
operational and/or configuration routines.  
Sentinel Compatible: 2 channels  
Conc. Curves: Built-in for conductivity measurements:  
HCl, H2SO4, NaOH, NaCl, HF, HNO3,  
KOH, etc.  
Teledyne Analytical Instruments  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Appendix  
A.2 Model Designation  
The standard LXT-220 controller is 110VAC powered unit in a 1/2  
DIN, NEMA 4X enclosure. It incorporates a backlit graphics display and  
a membrane keypad. This instrument requires an external 24VDC to  
power the outputs. An option is available for an internally generated  
24VDC powered instrument.  
The instrument is configured at the factory for the customer’s  
application and so there are many different model designators.  
The following codes will help in identifying a specific instrument  
configuration.  
Base Model:  
LXT-220  
LXT-220 controller  
Channel 1 Input:  
PH  
ORP Oxidation Reduction Potential (REDOX)  
DO Dissolved Oxygen (specify ppm or %)  
pH  
PION Specific ion  
CDL Conductivity, Low range (20uS or less)  
CDH Conductivity, High range (50uS to 50mS)  
CDT Conductivity Toroidal (electrodeless greater then 50mS)  
RS  
Channel 2 Input:  
PH pH  
ORP Oxidation Reduction Potential (REDOX)  
DO Dissolved Oxygen (specify ppm or %)  
Resistivity (0 to 50 megohms)  
PION Specific ion  
CDL Conductivity, Low range (20uS or less)  
CDH Conductivity, High range (50uS to 50mS)  
CDT Conductivity Toroidal (electrodeless greater then 50mS)  
RS  
Resistivity (0 to 50 megohms)  
Teledyne Analytical Instruments  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
LXT-220 DO  
Channel 3 Input:  
EX  
Input from an external 4 to 20mA device  
EX1 Input from an external relay contact device  
* Sentinel Diagnostic:  
DIAG1Sentinel Diagnostic on channel 1  
DIAG2Sentinel Diagnostic on channel 2  
** Number of 4 to 20 mA Outputs:  
1mA One 4 to 20mA Current Output  
2mA Two 4 to 20mA Current Outputs  
3mA Three 4 to 20mA Current Outputs  
4mA Four 4 to 20mA Current Outputs  
5mA Five 4 to 20mA Current Outputs  
6mA Six 4 to 20mA Current Outputs  
Hart  
HT  
Hart protocol option  
Type of relay contacts:  
SS  
C
SOLID STATE 240VAC 3amp  
Form C (SPDT)  
Number of relay contacts:  
/2  
/4  
/6  
/8  
Two relay contacts  
Four relay contacts  
Six relay contacts  
Eight relay contacts  
Mounting hardware:  
UM Universal mounting plate  
PM Panel mounting hardware  
HM Handrail mounting plate (including 2” hardware)  
Teledyne Analytical Instruments  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Appendix  
Optional power requirements:  
PS1 Standard powered by 120VAC  
PS2 24 VDC internally generated to power the 4 to 20 mA  
output cards  
PS3 LXT-220 controller powered by 24 VDC  
Special options:  
F2 Output = function of dual inputs: ratio, free chlorine,  
differential, average: SPECIFY!  
***%C Percent (%) ion concentration display only  
(Available for Conductivity only)  
PA  
Signal conditioner mounted inside transmitter enclosure.  
AV  
T1  
Field selectable time average (2 to 32 seconds)  
Periodic timer  
T2  
One-Shot timer  
LG  
PT  
Logic Gates Field configurable  
Special temperature input from sensor 1000 ohm pt  
PID 3 mode proportional control  
TPC Time Proportional Control (Requires 2 solid state relay  
contacts)  
GO2 Gaseous Oxygen (O2)  
For options not listed, consult factory for price and availability.  
*
Diagnostic Feature is only available on pH, ORP, and  
Pion. If the diagnostic option is selected the sensor and  
electrode must also contain the diagnostic option for the  
feature to function.  
**  
4 to 20mA Current outputs are field configurable. If the  
outputs are to be configured by Teledyne at time of  
shipment the output parameters must be specified. If not  
specified the outputs will be configured as standard 4 to  
20mA outputs corresponding to the specified input.  
***  
If an output is required along with the display a second  
Teledyne Analytical Instruments  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
LXT-220 DO  
output must be specified.  
Output configuration options:  
MA 4-20 mA, standard  
TMP 4-20 mA, Temperature (adjustable -100 to +200°C)  
PID 4-20 mA, 3 mode control  
% (ion) mA % concentration for display and 2nd output  
(Conductivity only)  
DIAG Sentinel Reference Diagnostic  
As an example:  
MODEL NUMBER EXAMPLE:  
LXT-220-PH/MA-PS2-UM  
This indicates the LXT-220 Controller for pH applications with a  
standard single 4-20mA output. The controller has an internal 24 VDC  
power supply to power the output card. The instrument is equipped with  
a universal mounting plate.  
Teledyne Analytical Instruments  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Index  
Index  
A to D converter trim, 42  
AC power. See power  
accuracy, 42, 53  
address, 55  
air, 47  
current output, 26  
configure, 28  
decreasing a value, 16  
dedicated output, 39  
default parameters, 14  
diagnostic feature, 59  
digital interface, 54  
digital signal, 39  
alarm, 13, 15  
HI, 13  
LO, 13  
alarm relay, 40  
dimensions, 54  
analog to digital converter, 42  
ASSIGN, 35  
display, 14, 16, 53  
display contrast, 25  
assigning parameters, 28  
assignment conflict, 36, 37  
auto calibration, 54  
automatic calibration, 14  
back to factory calibration, 14, 22, 49  
backlit graphics display, 57  
BUFFER, 27, 44, 46  
CALIBRATE key, 16, 23  
numerical equivalence, 31  
calibrate mode, 23  
calibration, 22, 44  
calibration menu, 27  
calibration voltage, 42  
caution sign, iv  
change the protection level, 33  
changing output assignment, 37  
cleaning, 51  
dissolved oxygen, 13, 14, 24, 53  
dissolved oxygen electrode, 46  
dual channel input, 22  
dual input. See input  
dual input option, 13  
efficiency, 47  
electrical connections, 20  
electrode, 29  
electrode degradation, 22, 46  
electrode diagnostic screen, 29  
electrode failure, 14  
electrode performance, 29  
electrode potential, 22  
electrode slope, 47, 48  
electrode variation, 46  
enclosure. See NEMA-4X  
Err PW, 32  
command #3, 38  
erratic reading, 52  
command #35, 39  
error detection, 49  
concentration curves, 54  
conductivity, 13, 14, 53  
conduit, 21  
EXIT/STAY MENU, 30  
external event, 53  
factory calibration point, 22  
feature codes, 57  
conduit hole, 17  
configuration password. See password  
CONFIGURE/TRIM, 30  
configuring LXT-220, 30  
connections, 17  
features, 14  
figures listing, ix  
flashing asterisk, 49  
front panel, 16  
contrast, 52  
gasket, 51  
contrast menu, 24, 25  
copyright, ii  
grab sample, 48  
gray screen, 25  
copyright display, 24  
cross-channel assignment, 37  
HART, 14, 15, 28, 36, 38, 40, 53, 54  
HART analog output, 39  
Teledyne Analytical Instruments  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
LXT-220 DO  
HART assignment, 38, 39  
hysteresis, 15  
hysterisis, 13  
opening the front panel, 20  
operating temperature, 54  
operation, 23  
impedance, 54  
incrementing a value, 16  
input, 13, 27  
operational password, 32  
options, 14  
ORP, 14, 22, 42, 53  
output, 13, 28  
conditioned, 21  
dual, 14  
analog, 53  
input range, 42  
assignment, 35  
installation, 19  
averaged, 15  
internal clock, 54  
configure, 28  
internal layout, 17  
internal power supply, 15  
isolation, 53  
keyboard control, 37  
LCD. See display  
correspondence, 22  
current, 14, 36, 54  
current, adjust, 26  
dirrerential, 15  
floating, 15  
linear, 28  
linearity, 53  
graphic display, 26  
isolated, 15  
locus, 29  
multiple, 15  
low level DC signal, 17  
main menu, 24, 26  
manual mode, 26, 37  
manual mode assignment, 37  
manual output mode, 22, 36, 44  
manuals, additional, v  
measurable parameter, 22  
membrane, 22, 49  
membrane switche. See switch  
menu options, 24  
PID, 15, 43, 54  
ratio, 15  
signal factory assignment, 17  
output configuration options, 60  
oxidation-reduction potential. See ORP  
panel mounting, 20  
parameter selection menu, 27, 45  
password, 31, 54  
configuration, 34  
operation, 33  
MENU SELECT key, 16, 23  
menus, 24  
percent output display, 37  
pH, 14, 53  
microprocessor, 13  
microprocessor replacement, 51  
missing assignment, 36  
model designation, 57  
mounting bracket, 20  
mounting plate  
PID, 13, 15, 28, 36, 40, 43  
minimum output, 44  
PID assignment, 43  
plot function, 28  
power, 17, 21, 53  
pre-pHault diagnostics, 14  
primary variable, 38  
probe, 13  
handrail, 20  
universal, 19  
multi-bus architecture, 13  
multi-drop mode, 39  
NEMA 4X, 57  
NEMA-4X, 14, 17, 20, 53  
no display, 52  
process control, 13, 40  
process variable display, 26  
proportional band, 15  
protection level, 32, 34  
question mark, 36, 37, 43  
rate, 15  
noise filter, 29  
noise rejection, 54  
non-linear reading, 52  
numeric adjustments, 16, 23  
one point calibration, 22  
one-point calibration, 44, 46  
relay, 15, 29, 54  
configure, 28  
relay assignment, 40  
relay configuration menu, 40  
relay output, 13  
Teledyne Analytical Instruments  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dissolved Oxygen Controller  
Index  
relay setpoints, 40  
release manual output, 45  
repeatability, 53  
STATUS, 27, 46  
STATUS menu, 29  
submenus, 24  
reset, 15  
Supertwist, 53  
resistivity, 13, 14, 53  
response time, 53  
restore password, 32  
reverse acting, 13, 14  
routine maintenance, 51  
RS232, 53  
safety information, iv  
saturation point, 47  
scale factor, 42  
switch, 14, 16  
tables listing, x  
technician symbol, iv  
Teledyne address, 55  
temperature calibration, 40  
temperature compensation, 54  
temperature control, 13  
temperature conversion, 14, 41  
terminal block, 17  
timer, 32, 54  
sensitivity, 53  
sensor  
trim, 41  
SP1 & SP2, 21  
troubleshooting, 52  
two point calibration, 22  
two-point calibration, 44, 46  
typical electrode, 22, 49  
U-bolt, 20  
undisturbed output, 26  
warning sign, iv  
warranty, ii  
sensor orientation, 52  
sensor wiring, 22  
serial number, iii, 24  
SETUP, 27, 28, 46  
shipping damage, 19  
slope variation, 47  
software, 15  
software version, 24  
span calibration, 22, 44  
span display, 36  
weatherproof, 53  
web address, 55  
website address, v  
white screen, 25  
span setting, 39  
specific compensation, 14  
specific ion, 13, 14, 42, 53  
specifications, 53  
stability, 53  
wiring, 21  
wiring label, 17, 21  
zero electrode offset, 14, 22  
zero potential, 27, 29  
standardization, 22, 44, 46  
Teledyne Analytical Instruments  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Technicolor Thomson DVD Recorder DTH8555 X User Manual
Toshiba CD Player SR C8002 User Manual
Toshiba Projector TLP B2 User Manual
Toshiba TV DVD Combo MD20FM1 User Manual
Trane Air Conditioner TAM7A0B30H21SB User Manual
Trane Humidifier TCONT800 User Manual
Tricity Bendix Refrigerator FD 792 User Manual
Troy Bilt Trimmer TB2000 User Manual
Uniden Headphones DCX170 User Manual
Wayne Dalton CRT Television 9100 9400 9600 User Manual