Teledyne Oxygen Equipment 9060L User Manual

OPERATING INSTRUCTIONS FOR  
Model 9060L  
Percent Oxygen Analyzer  
P/N M77713  
ECO:  
DANGER  
Toxic gases and or flammable liquids may be present in this monitoring system.  
Personal protective equipment may be required when servicing this instrument.  
Hazardous voltages exist on certain components internally which may persist  
for a time even after the power is turned off and disconnected.  
Only authorized personnel should conduct maintenance and/or servicing.  
Before conducting any maintenance or servicing, consult with authorized  
supervisor/manager.  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Specific Model Information  
Instrument Serial Number: _______________________  
Instrument Range:  
Calibrated for:  
Background Gas:  
Zero Gas:  
_______________  
_______________  
_______________  
_______________  
_______________  
Span Gas:  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060L  
Safety Messages  
Your safety and the safety of others is very important. We have  
provided many important safety messages in this manual. Please read  
these messages carefully.  
A safety message alerts you to potential hazards that could hurt you  
or others. Each safety message is associated with a safety alert symbol.  
These symbols are found in the manual and inside the instrument. The  
definition of these symbols is described below:  
GENERAL WARNING/CAUTION: Refer to the instructions  
for details on the specific danger. These cautions warn of  
specific procedures which if not followed could cause bodily  
Injury and/or damage the instrument.  
CAUTION: HOT SURFACE WARNING: This warning is  
specific to heated components within the instrument. Failure  
to heed the warning could result in serious burns to skin and  
underlying tissue.  
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous  
voltages appear within this instrument. This warning is  
specific to an electrical hazard existing at or nearby the  
component or procedure under discussion. Failure to heed this  
warning could result in injury and/or death from  
electrocution.  
Technician Symbol: All operations marked with this symbol  
are to be performed by qualified maintenance personnel only.  
NOTE: Additional information and comments regarding a  
specific component or procedure are highlighted in the form  
of a note.  
No  
Symbol  
CAUTION:  
THE ANALYZER SHOULD ONLY BE USED FOR THE  
PURPOSE AND IN THE MANNER DESCRIBED IN  
THIS MANUAL.  
Teledyne Analytical Instruments  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
IF YOU USE THE ANALYZER IN A MANNER OTHER  
THAN THAT FOR WHICH IT WAS INTENDED,  
UNPREDICTABLE BEHAVIOR COULD RESULT  
POSSIBLY ACCOMPANIED WITH HAZARDOUS  
CONSEQUENCES.  
This manual provides information designed to guide you through  
the installation, calibration operation and maintenance of your new  
analyzer. Please read this manual and keep it available.  
Occasionally, some instruments are customized for a particular  
application or features and/or options added per customer requests.  
Please check the front of this manual for any additional information in  
the form of an Addendum which discusses specific information,  
procedures, cautions and warnings that may be peculiar to your  
instrument.  
Manuals do get lost. Additional manuals can be obtained from  
TI/AI at the address given in the Appendix. Some of our manuals are  
available in electronic form via the internet. Please visit our website at:  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060L  
Table of Contents  
List of Figures ............................................................................ viii  
List of Tables................................................................................ ix  
Introduction ................................................................................. 11  
1.1 Overview  
11  
11  
12  
13  
1.2 Main Features of the Analyzer  
1.3 Front Panel Description  
1.4 Rear Panel Description  
Operational Theory ..................................................................... 15  
2.1 Introduction  
15  
15  
17  
18  
19  
20  
20  
20  
20  
21  
2.2 Oxygen Probe, Sensor, and Heater  
2.2.1 Probe Configuration  
2.2.2 The Zirconia Sensor  
2.2.3 Heater  
2.3 Applications Not at Atmospheric Pressure  
2.4 Sensor Impedance  
2.5 Electronics  
2.5.1 General  
2.5.2 Signal Processing  
Installation ................................................................................... 23  
3.1 Unpacking the Analyzer  
3.2 Control Unit Installation  
3.3 9060L Oxygen Probe Installation  
3.4 Electrical Connections  
3.4.1 Primary Input Power  
3.4.2 Analog Outputs  
23  
23  
23  
25  
25  
26  
26  
3.4.3 Alarm Relays  
Teledyne Analytical Instruments  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
3.4.1 Oxygen Probe Connections  
3.5 Installation Checklist  
28  
29  
Operation .....................................................................................31  
4.1 Introduction  
31  
32  
32  
33  
33  
33  
33  
34  
34  
34  
35  
35  
37  
37  
37  
38  
38  
38  
4.2 Using the Function and Data Entry Buttons  
4.3 Setting the Analysis Ranges  
4.3.1 HI Range  
4.3.2 LO Range  
4.4 Setting the Alarm Setpoints  
4.4.1 Set Alarm 1 (HI alarm)  
4.4.2 Set Alarm 2 (LO alarm )  
4.4.3 Sensor Fail Alarm  
4.5 Selecting a Fixed Range or Autoranging  
4.6 Oxygen Calibration  
4.6.1 Span Calibration  
4.6.2 Air Calibration  
4.7 Supplementary Information  
4.8 Probe Temperature  
4.8.1 Low Temperature  
4.8.2 Open Thermocouple  
4.9 Error Messages  
Maintenance.................................................................................41  
5.1 Replacing the Fuse 41  
Appendix......................................................................................43  
A.1 Specifications  
43  
44  
45  
45  
46  
46  
A.2 Probe Specifications  
A.3 Spare Parts List  
Email: [email protected] Reference Drawing  
A.4 Reference Drawing  
A.5 Probe or Sensor EMF Tables  
Index.............................................................................................49  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060L  
List of Figures  
Figure 1-1: Front Panel ................................................................ 12  
Figure 1-2 Rear Panel................................................................... 14  
Figure 2-1: Model 9060L Oxygen Probe ....................................... 16  
Figure 2-2: Model 9060L Probe..................................................... 17  
Figure 2-3: Internal Structure of the Zirconia Sensor and Probe... 18  
Figure 2-3: Signal Processing Block Diagram............................... 21  
Figure 3-1: Oxygen Probe Mounting ............................................. 25  
Figure 3-2: Model 9060L Rear panel............................................. 26  
Figure 3-3: Contact ID for FAILSAFE Relay Operation................. 27  
Figure 3-4:Model 9060L Heated Probe Cable Connection............ 28  
Figure 3-5: Connections to Probe ................................................. 29  
Figure 4-1: Front Panel Controls and Indicators............................ 31  
Figure 4-2: Calibration Port on Probe Head .................................. 35  
Figure 5-1: AC Fuse Replacement................................................ 41  
Teledyne Analytical Instruments  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
List of Tables  
Table 2-1: Probe Options ..............................................................17  
Teledyne Analytical Instruments  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060L  
DANGER  
COMBUSTIBLE GAS USAGE  
WARNING  
This is a general purpose instrument designed for use in a  
non-hazardous area. It is the customer's responsibility to  
ensure safety especially when combustible gases are being  
analyzed since the potential of gas leaks always exist.  
The customer should ensure that the principles of operating  
this equipment are well understood by the user. Misuse of  
this product in any manner, tampering with its components,  
or unauthorized substitution of any component may  
adversely affect the safety of this instrument.  
Since the use of this instrument is beyond the control of  
Teledyne Instruments/ Analytical Instruments, referred as  
TI/AI, no responsibility by TI/AI, its affiliates, and agents for  
damage or injury from misuse or neglect of this equipment is  
implied or assumed.  
Teledyne Analytical Instruments  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Introduction  
Introduction  
1.1 Overview  
The Teledyne Instruments/ Analytical Instruments (TI/AI) Model  
9060L is a microprocessor-based percent oxygen analyzer for real-time  
measurement of the percent of oxygen in inert gases, or in a wide variety  
of gas mixtures. It features simple operation, fast response, and a  
compact, rugged construction. Typical applications of the Model 9060L  
are monitoring kilns, boilers, and flue gas at temperature from ambient  
up to 1400°C (2550°F).  
1.2 Main Features of the Analyzer  
The main features of the analyzer include:  
High resolution, accurate readings of oxygen content from 0-1 %  
through 0-100 % O2. Large, bright, light-emitting-diode meter  
readout.  
Simple pushbutton controls.  
State of the art zirconia sensor.  
Longer life probes with greater resistance to thermal shock and  
mechanical damage during installation and start-up.  
Fast response and recovery time.  
Microprocessor based electronics: 8-bit CMOS microprocessor  
with on-board RAM and 16 KB ROM.  
Two user selectable ranges (from 0-1% through 0-100%) allow  
best match to user’s process and equipment.  
Air-calibration range for convenient spanning at 20.9 %.  
Operator can select autoranging, which allows the analyzer to  
automatically select the proper preset range for a given  
measurement, or lock the analyzer onto a single range.  
Teledyne Analytical Instruments  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Model 9060L  
Two concentration alarms with adjustable setpoints.  
A failure alarm.  
Three analog outputs: two for measurement (0–10VDC, and  
negative ground 4–20 mA DC) and one for range identification  
(0-10VDC).  
Compact and rugged control unit with flush-panel case.  
Designed for indoor use.  
1.3 Front Panel Description  
All controls and displays except the power switch are accessible  
from the front panel. See Figure 1-1. The front panel has seven  
pushbutton membrane switches, a digital meter, and an alarm indicator  
LED for operating the analyzer. These features are described briefly  
here and in greater detail in Chapter 4, Operation.  
Figure 1-1: Front Panel  
Function Keys: Seven pushbutton membrane switches are used to  
select the function performed by the analyzer:  
Set Alarm 1  
Sets the concentration at which alarm 1  
activates upon rising above the threshold  
(HI alarm).  
Teledyne Analytical Instruments  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Introduction  
Set Alarm 2  
Sets the concentration at which alarm 2  
activates upon dropping below the  
threshold (LO alarm).  
Set HI Range  
Set LO Range  
Span  
Set the high analysis range for the  
instrument (up to 0-100%).  
Set the low analysis range for the  
instrument (down to 0-1%).  
Span calibrate the analyzer.  
Data Entry Keys: Two pushbutton membrane switches are used to  
manually change measurement parameters of the instrument as they are  
displayed on the LED meter readout:  
Up Arrow  
Increment values of parameters upwards  
as they are displayed on the LED readout.  
Down Arrow  
Increment values of parameters  
downwards as they are displayed on the  
LED readout.  
Digital LED Readout: The digital display is a LED device that  
produces large, bright, 7-segment numbers that are legible in any  
lighting environment. It has two functions:  
Meter Readout As the meter readout, it displays the oxygen  
concentration currently being measured.  
Measurement Parameters Readout The parameter readout  
displays user-defined alarm setpoints, ranges, span calibration  
point when they are being checked or changed, and probe  
temperature on demand.  
1.4 Rear Panel Description  
The rear panel contains the electrical input and output connectors.  
The connectors are described briefly here and in detail in the Installation  
chapter of this manual.  
Teledyne Analytical Instruments  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Model 9060L  
Figure 1-2 Rear Panel  
Power Connection 110 VAC at 50/60 Hz. The connector  
housing includes the fuse holder and the  
power switch. (220VAC is optional via  
setting an internal switch).  
Fuse Holder: Replacing the fuse is  
described in Chapter 5, Maintenance.  
I/O Power Switch: Turns the instrument  
power ON (1) or OFF (0).  
Analog Outputs  
0–10 VDC concentration output.  
0–10 VDC range ID (or optional  
overrange) output.  
4–20 mA DC concentration output,  
negative ground.  
Alarm Connections Alarm 1 (HI alarm), Alarm 2 (LO alarm),  
and Probe Failure Alarm connections.  
Sensor Connector For connecting the probe’s Oxygen  
sensor.  
Heater Connector For connecting the probe’s heater  
Teledyne Analytical Instruments  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operational Theory  
Operational Theory  
2.1 Introduction  
The analyzer is composed of two subsystems:  
1. Oxygen probe  
2. Control unit with signal processing, display and controls  
The oxygen probe can be heated or unheated. It delivers the sample  
gas to a state-of-the-art zirconium sensor which translates the amount of  
oxygen present into a millivolt output.  
The control unit processes the sensor output and translates it into  
electrical concentration, range, and alarm outputs, and a percent oxygen  
meter readout. It contains a micro-controller that manages all signal  
processing, input/output, and display functions for the analyzer.  
2.2 Oxygen Probe, Sensor, and Heater  
TI/AI oxygen probes and sensors employ state-of-the-art zirconia  
sensors and advanced materials. They provide the following benefits:  
Improved control due to fast response time—typically less than  
four seconds.  
Cost-efficient design provides improved reliability.  
Longer-life probes with greater resistance to corrosion from  
sulphur and zinc contaminants in a flue gas.  
Low cost replacement reduces maintenance.  
Reduced probe breakage due to greater resistance to thermal  
shock and mechanical damage during installation and startup.  
The 9060L probes or sensors are simple to install and maintain. All  
models provide direct measurement of oxygen level. Figure 2-1 shows a  
typical oxygen probe. Probe specifications are listed in the Appendix of  
this manual and installation is covered in Section 3.  
Teledyne Analytical Instruments  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
Model 9060L  
All Teledyne oxygen probe or sensors are designed and  
manufactured to exacting standards of performance and reliability and are  
the result of extensive research and development. Teledyne Analytical  
Instruments provides worldwide application and after sales support for its  
oxygen probes, sensors and analyzers.  
Figure 2-1: Model 9060L Oxygen Probe  
The probe assembly provides a means of exposing the zirconia  
sensor to the atmosphere to be measured with sensor. It also provides for  
means of interfacing the sensor, thermocouple and heater wires with the  
control unit. Reference air is fed via the top plug for unheated probes or  
by a separate gas thread connection for heated probes. Connections are  
provided on probes for an in-situ gas calibration check. If required, an  
air purge can be admitted via the calibration gas check entry. The outer  
sheath of probes is either metal or ceramic, depending on the  
application. Calibration check can be achieved on 9060LEX sensors  
using a three way solenoid which blocks the sample and at the same  
time admits a calibration gas to the sensor. As mentioned above, probe  
purging for removing any dust build up can be achieved in the same  
way.  
In-situ zirconia oxygen probes will give a lower oxygen reading  
than a sampled gas measurement on a chromatograph or paramagnetic  
analyzer because the flue gas contains a significant level of water vapor  
and a sampling system removes the water vapor through condensation.  
The oxygen content then appears as a higher percentage of the  
remaining gas. For example: If the gas contained five parts oxygen and  
fifteen parts moisture, removing the moisture would leave the oxygen at  
5.88%. This phenomena will depend on the fuel and the completeness  
of combustion. They are common to all zirconia oxygen sensors.  
Teledyne Analytical Instruments  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operational Theory  
Probes of 1000 mm (40”) normally have sufficient length for any  
installation. Customers requiring probes longer than 1500 mm (59”) are  
supplied with a flow guide tube that uses the flue velocity to pull flue  
gas through a filter at the sensing tip and exhaust it near the flue wall.  
2.2.1 Probe Configuration  
The probe can be configured in a variety of ways to suit the  
analytical environment. Figure 2-1 shows the most common  
configurable items of the probe. Not indicated is the actual  
thermocouple sensing element, however, see Table 2-1. You can choose:  
1. Probe insertion length (from process end of mounting thread to  
probe sensing tip). See Figure 2-2.  
2. Mounting thread (process connection), BST or NPT (for size of  
thread refer to specifications).  
3. Lagging extension length, if required.  
Table 2-1 indicates typical probe options.  
Table 2-1: Probe Options  
Probe Length  
Thermocouple  
Sheath  
Thread  
250 mm (10”)  
350 mm (14”)  
500 mm (20”)  
750 mm (30”)  
1000 mm (40”)  
Type K max 900°C (1560°F)  
316 SS max 850°C (1560°F)  
Inconel  
The Inconel option has all  
inconel wetted parts except for  
the ceramic sensor and viton  
‘o’ rings  
BST or NPT  
Figure 2-2: Model 9060L Probe  
Teledyne Analytical Instruments  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
Model 9060L  
2.2.2 The Zirconia Sensor  
Input to the analyzer is obtained from a solid electrolyte oxygen  
probe containing a zirconia element and thermocouple. The probe is  
designed to be inserted into a boiler or furnace exit gas flue or similar  
process. The 9060LEX sensor is designed to be installed outside of the  
flue or process. Sampling lines and filters are not required for in-situ  
probes but they are required for 9060LEX sensors. When a sampling  
line is required, the gas typically flows to the sensor under process  
pressure. In applications where the process pressure is negative or  
neutral, a suction pump will be required. A reference air pump is not  
provided in the Model 9060L Oxygen Analyzer. The internal  
construction of a probe or sensor is shown in Figure 2-3.  
Figure 2-3: Internal Structure of the Zirconia Sensor and Probe  
Heater control in the Model 9060L Oxygen Analyzer consists of a  
time proportioning temperature controller and solid state relay so that  
the thermocouple junction is controller to over 700°C (1300°F). Probes  
operating in a combustion environment above 650°C (1200°F) do not  
Teledyne Analytical Instruments  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operational Theory  
require a heater. When the outside and inside of the sensor are exposed  
to different oxygen partial pressures, an EMF (E) is developed which  
obeys the Nernst equation:  
RT  
4F  
(PO2) Inside  
E(millivolts) =  
log  
e
(PO2) Outside  
where T is the temperature (K) at the disc (>650°C (1200°F)), R is the  
gas constant. F is the Faraday constant and (PO2) Inside and (PO2)  
Outside are the oxygen partial pressures at the inner and outer electrodes  
respectively with the higher oxygen partial pressure being positive.  
If dry air at atmospheric pressure, (21% oxygen) is used as a  
reference at the inner electrode, the following equations are obtained:  
0.21  
2  
E(millivolts) = 2.154×10 T loge  
(PO2 ) Outside  
Transposing this equation:  
- 46.421E  
T
(%O ) Outside (ATM) = 0.21EXP  
2
The 9060L transmitter solves this equation which is valid above  
650°C (1200°F). Either the probe heater or the process maintains the  
sensor temperature at this level.  
2.2.3 Heater  
CAUTION:  
DANGEROUS VOLTAGE (120 VAC) EXISTS WITHIN  
THE PROBE OR HEATER AND PRESENTS AN  
ELECTRICAL SHOCK HAZARD TO MAINTENANCE  
PERSONNEL. ALWAYS DISCONNECT THE PROBE  
FROM THE ANALYZER BEFORE WORKING WITH  
THE PROBE, SENSOR, OR HEATER.  
THE EARTH CONNECTION (GREEN WIRE) FROM  
Teledyne Analytical Instruments  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
Model 9060L  
THE PROBE/SENSOR MUST ALWAYS BE  
CONNECTED TO EARTH.  
Power to the heater is supplied directly from the main power  
source, and the temperature is initially controlled above 700°C (1300°F)  
after power up.  
2.3 Applications Not at Atmospheric Pressure  
For high and low pressure applications where the pressure at the  
point of measurement is significantly above or below atmospheric  
pressure, compensation must be applied. The Model 9060L does not  
incorporate a pressure sensor thus no compensation is performed. This  
factor must be considered when reading concentration from the display  
under high or low pressure monitoring conditions.  
2.4 Sensor Impedance  
The zirconia sensor impedance is a basic measurement of the  
reliability of the oxygen reading. A probe or sensor with a high  
impedance reading will eventually produce erroneous signals. The  
analyzer checks the zirconia sensor impedance every 5 minutes and if  
the impedance is above the maximum level for a specific temperature  
then the impedance alarm will be activated. Typical sensor impedance  
is 1 Kto 8 Kat 720°C (1320°F).  
2.5 Electronics  
2.5.1 General  
The signal processing uses an Intel® micro-controller with on-board  
RAM and ROM to control all signal processing, input/output, and  
display functions for the analyzer. System power can be either 120 or  
220 VAC depending on the position of an internal switch.  
The power supply circuitry and temperature controller circuitry are  
on the Power Supply PCB, which is mounted vertically, just behind the  
rear panel of the control unit.  
The signal processing electronics including the sensor amplifier,  
micro-controller, analog to digital, and digital to analog converters are  
Teledyne Analytical Instruments  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operational Theory  
located on the Main PCB, which is mounted vertically, just behind the  
front panel of the control unit.  
2.5.2 Signal Processing  
Figure 2-3 is a block diagram of the signal processing electronics  
described below.  
Figure 2-3: Signal Processing Block Diagram  
In the presence of oxygen the sensor generates a millivolt potential  
which is fed to a voltage amplifier. The output from the sensor amplifier  
is sent to an analog to digital converter (ADC), and the resulting digital  
concentration signal is passed to the micro-controller. The ADC has a  
second channel which monitors the temperature of the probe.  
Teledyne Analytical Instruments  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
Model 9060L  
The digital concentration signal along with input from the front  
panel buttons (KEYBOARD) is processed by the micro-controller, and  
appropriate output signals are directed to the display and alarm relays.  
The same digital information is also sent to a 12-bit digital to analog  
converter (DAC) that produces the 0-10 VDC analog concentration  
signal and the 0-10 VDC analog range ID output. A current to voltage  
converter (E–I CONV) produces the 4-20 mA DC analog concentration  
signal.  
Teledyne Analytical Instruments  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Installation  
Installation  
Installation of the analyzer includes:  
1. Unpacking the system.  
2. Mounting the control unit.  
3. Installing the probe.  
4. Making electrical connections.  
5. Testing the installation.  
CAUTION:  
READ THIS CHAPTER IN ITS ENTIRETY BEFORE  
INSTALLING THE SYSTEM.  
FOR INDOOR USE ONLY.  
3.1 Unpacking the Analyzer  
As soon as you receive the instrument, carefully unpack and inspect  
the control unit, and any included accessories for damage. Immediately  
report any damage to the shipping agent. The analyzer is shipped with  
all the materials you need to install and prepare the system for operation.  
3.2 Control Unit Installation  
The 9060L control unit is designed to be panel mounted in a  
general purpose, indoor area, away from moisture and the elements. The  
unit should be installed at viewing level in a sheltered area.  
Refer to the Outline Diagram D-77712 in the Appendix for the  
physical dimensions of the analyzer.  
3.3 9060L Oxygen Probe Installation  
To install the probe, weld a BSP or NPT socket to the flue in a  
suitable position for flue gas sensing. For the correct size of socket refer  
to probe specifications located in the Appendix.  
Teledyne Analytical Instruments  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 9060L  
Note: Mount the probe as close to the combustion source as  
possible. This reduces the lag time inherent in the  
sampling process and results in better control.  
The probe has a typical response time of less than four seconds, so  
most of the delay time is associated with the transit time of the gas from  
the point of combustion to the point of sensing.  
Probes can be mounted at any angle. When feasible, a vertical  
installation with the probe facing vertically downwards will eliminate  
the need for a filter. This is especially useful for processes with  
particulates in the flue gas. If a vertical mounting is not possible under  
these conditions, a filter may be necessary. This may require periodic  
filter replacement depending on the severity of the process.  
If a flow guide tube is used (heated probes only), it is important that  
the fin points directly downstream. If the exact flow direction is not  
known, use a wind vane that can be made from a piece of wire and flat  
metal. If the flow guide tube is installed facing the wrong direction for  
any period, the suction filter may block with flue gas dust particles.  
The maximum temperature for an unsupported flow guide tube is  
750°C (1380°F). Above this temperature, provide a support and flanged  
flexible rubber joint as shown in Figure 3-1. The maximum temperature  
of a supported flow guide tube is 900°C (1650°F).  
When installing a probe into a hot environment, slide the probe in  
slowly to avoid thermal shock to the internal ceramic parts. If the flue  
gas is 1000°C (1830°F), it should take approximately ten minutes to  
install a 500 mm (20”) probe, moving it in steps of about 20 mm (1”).  
CAUTION:  
IT IS IMPORTANT THAT THERE ARE NO AIR LEAKS  
UPSTREAM OF THE OXYGEN SENSING POINT,  
OTHERWISE THERE WILL BE A HIGH OXYGEN  
READING.  
If the probe is to be installed on a bend in the flue, it is best to  
locate it on the outer circumference of the bend to avoid dead pockets of  
flue gas flow. While the standard 9060L probe with a ‘U’ length of 250  
mm (10”) will suit most low temperature flue applications, it is  
occasionally necessary to have a longer probe with the sensing tip in the  
center of the flue gas stream.  
Although it is rare, occasionally a probe may sense oxygen vastly  
different from the average reading in the flue gas. If it occurs, move the  
Teledyne Analytical Instruments  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Installation  
probe or install a longer probe. This is normally caused by stratification  
of the flue gas.  
Figure 3-1: Oxygen Probe Mounting  
3.4 Electrical Connections  
Figure 3-1 shows the Model 9060L rear panel. For detailed  
pinouts, see the wiring/interconnection drawings in the Drawings section  
at the rear of this manual.  
3.4.1 Primary Input Power  
The power cord receptacle, fuse block and Power switch are  
located in the same assembly. A 6-foot, standard AC power cord is  
supplied with the control unit. Insert the female plug end of the power  
cord into the power cord receptacle.  
The power supply allows direct connection to any 110 VAC,  
50/60Hz power source. 220VAC is an optional configuration settable  
through an internal switch in the control unit. The fuse block, to the right  
of the power cord receptacle, accepts two 5x20mm IEC type T, 2 A,  
time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.)  
The Power switch is located on the right-hand end of the power  
source input receptacle assembly.  
Teledyne Analytical Instruments  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 9060L  
Figure 3-2: Model 9060L Rear panel  
3.4.2 Analog Outputs  
There are three DC output signal connectors with screw terminals  
on the panel. There are two wires per output with the polarity noted. See  
Figure 3-3. The outputs are:  
0–10V % Range:  
Voltage rises with increasing oxygen  
concentration, from 0V at 0 percent oxygen to  
10V at full scale percent oxygen. (Full scale =  
100% of programmed range.)  
0–10V Range ID:  
03.33V = Low Range, 06.66V = High Range,  
10V = Air Cal Range.  
4–20 mA % Range: Current increases with increasing oxygen  
concentration, from 4 mA at 0 percent oxygen to  
20 mA at full scale percent oxygen. (Full scale =  
100% of programmed range.)  
3.4.3 Alarm Relays  
The three alarm-circuit connectors are screw terminals for making  
connections to internal alarm relay contacts. There is one set of contacts  
for each type of alarm. Contacts are Form C, with normally open and  
normally closed contact connections capable of switching up to 1.0  
ampere at 125 VAC into a resistive load (2A for 30 VDC).  
Teledyne Analytical Instruments  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Installation  
The alarm relay circuits are designed for failsafe operation,  
meaning the relays are energized during normal operation. If power fails  
the relays de-energize (alarms activated).  
The contact connections are indicated diagrammatically on the rear  
panel as Normally Closed, Common, and Normally Open. Figure 3-3  
explains how these act in failsafe operation.  
Alarm 1 and Alarm 2 are configured as HI and LO respectively. A  
HI alarm will activate when concentration is above threshold, while a  
LO alarm will activate when concentration is below threshold.  
Figure 3-3: Contact ID for FAILSAFE Relay Operation  
The specific descriptions for each type of alarm are as follows:  
Alarm #1  
Alarm #2  
Sensor Fail  
Programmable as high alarm. Can be set anywhere  
between 0 and 100 % but no lower than Alarm #2.  
Programmable as low alarm. Can be set anywhere  
from 0 to 100 % but no higher than Alarm #1.  
Actuates when the output of the sensor falls below a  
factory programmed acceptable level or the  
thermocouple is open.  
CAUTION:  
THERE COULD BE HAZARDOUS VOLTAGE AT THE  
ALARMS TERMINALS, EVEN WHEN POWER IS  
REMOVED FROM THE INSTRUMENT.  
Teledyne Analytical Instruments  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 9060L  
3.4.1 Oxygen Probe Connections  
The receptacle for the probe cable is located in the lower left-hand  
corner of the rear panel. The 6-pin polarized connector is keyed to fit  
only one way into the receptacle. Do not force it in. The other end of the  
cable connects to the probe assembly. See Figure 3-5. Connect the probe  
lead as shown in Figure 3-4 . The heater connector is on the right-hand  
corner of the rear panel. Six foot long cables are provided for the  
remote probe and the heater. Their Teledyne Analytical Instruments part  
numbers are listed on the spare parts list located in the appendix.  
Figure 3-4:Model 9060L Heated Probe Cable Connection  
Teledyne Analytical Instruments  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Installation  
Figure 3-5: Connections to Probe  
3.5 Installation Checklist  
Prior to powering up the unit make sure you have:  
Installed the unit correctly  
Checked the probe and welded probe socket for leaks  
Once the above checks have been made, you can connect the power  
source. Wait approximately 30 minutes for warm up, two hours wait is  
best. After that, the instrument is ready for operation.  
Teledyne Analytical Instruments  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 9060L  
Teledyne Analytical Instruments  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operation  
Operation  
4.1 Introduction  
Once the analyzer has been mounted, the probe installed, and the  
electrical connections made, the analyzer can be configured for your  
application. This involves setting the system parameters:  
Defining the user selectable analysis ranges.  
Setting alarm setpoints.  
Calibrating the instrument.  
All of these functions are performed via the front panel controls,  
shown in Figure 4-1.  
Analyzing for the percent oxygen level in the gas passing through  
the probe and sensor is the default mode of operation. As long as no  
front panel buttons are being pressed the analyzer is analyzing.  
Figure 4-1: Front Panel Controls and Indicators  
Teledyne Analytical Instruments  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Model 9060L  
4.2 Using the Function and Data Entry Buttons  
When no buttons on the analyzer are being pressed, the instrument  
is in the analyze mode. It is monitoring the percent of oxygen in the  
sample gas that is flowing through the remote probe.  
When one of the function buttons is being pressed, the analyzer  
enters the setup mode or the calibration mode.  
Setup Function Buttons:  
There are 4 setup function buttons on the analyze:  
SET ALARM 1  
SET ALARM 2  
SET HI RANGE  
SET LO RANGE  
Calibration Button:  
The calibration mode button is:  
SPAN  
Data Entry buttons:  
The data entry buttons (Δ and ) increment the values displayed on  
the PERCENT OXYGEN meter while one of the function buttons is  
being held down.  
• Δ : Increments the displayed value upwards.  
• ∇ : Increments the displayed value downwards.  
Any of the functions can be selected at any time by holding down  
the appropriate button.  
Each function will be described in the following sections. Although  
the operator can use any function at any time, the order chosen in this  
manual is appropriate for an initial setup.  
4.3 Setting the Analysis Ranges  
The two user definable analysis ranges are both capable of being  
adjusted for from 0-1 to 0-100% oxygen concentration. Whatever values  
are selected, the analyzer automatically switches from the LO range to  
the HI range when the oxygen concentration reaches 100% of the LO  
Teledyne Analytical Instruments  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operation  
range full scale value, and it switches back to the LO range when the  
oxygen concentration reaches 85% of the LO range full scale value  
Note: The HI Range setpoint MUST be set at a higher  
concentration percentage than the LO Range setpoint.  
4.3.1 HI Range  
Setting the HI Range full scale value defines the LEAST sensitive  
analysis range to be used. To set the HI Range:  
1. Press the SET HI RANGE function button once.  
2. Immediately (within 5 seconds) press either the Δ or button  
to raise or lower the displayed value, as required, until the  
display reads the desired full scale percent concentration.  
4.3.2 LO Range  
Setting the LO Range full scale value defines the MOST sensitive  
range to be used. To set the LO Range:  
1. Press the SET LO RANGE function button once.  
2. Immediately (within 5 seconds) press either the Δ or button  
to raise or lower the displayed value, as required, until the  
display reads the desired full scale percent concentration.  
4.4 Setting the Alarm Setpoints  
The alarm setpoints can be adjusted over the full range of the  
analyzer (0-100% oxygen content). They are set as a percent of oxygen  
content, so that an alarm set to indicate 9.6 on the display will activate at  
9.6% O2 on any O2 range.  
4.4.1 Set Alarm 1 (HI alarm)  
Alarm 1 is preset as a high alarm. To configure this alarm to your  
preferences:  
1. Press the SET ALARM 1 function button once.  
2. To change the setting at which the alarm will be actuated, press  
the SET ALARM 1 function button once more. The alarm  
setpoint will flash on the LED display. Press either the Up or  
Down keys to raise or lower the displayed value, as required,  
Teledyne Analytical Instruments  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Model 9060L  
until the display reads the desired percent concentration. If  
within 5 seconds no key is pressed, the instrument will return to  
the sample mode and display oxygen concentration.  
After setting the value wait for the unit to time out of this mode  
(approximately 5 seconds) and return to displaying oxygen  
concentration.  
4.4.2 Set Alarm 2 (LO alarm )  
Alarm 2 is preset as a low alarm. To configure this alarm to your  
preferences:  
1. Press the SET ALARM 2 function button once.  
2. To change the setting at which the alarm will be actuated, press  
the SET ALARM 2 function button once more. The alarm  
setpoint will flash on the LED display. Press either the Up or  
Down keys to raise or lower the displayed value, as required,  
until the display reads the desired percent concentration. If  
within 5 seconds no key is pressed, the instrument will return to  
the sample mode and display oxygen concentration.  
After setting the value wait for the unit to time out of this mode  
(approximately 5 seconds) and return to displaying oxygen  
concentration.  
4.4.3 Sensor Fail Alarm  
The SENSOR FAIL alarm is factory set to a reading less than  
0.05% O2. Should this alarm trigger, the ALARM indicator below the  
SET function buttons will blink, and the alarm relay contact dedicated to  
this function will change state.  
The SENSOR FAIL alarm will also activate when the temperature  
is below 650°C, or the thermocouple is sensed as open.  
4.5 Selecting a Fixed Range or Autoranging  
The Model 9060L can operate in fixed high, fixed low, or  
autoranging mode. To change modes:  
1. Press and then release the SET HI RANGE and the SET LO  
RANGE buttons simultaneously.  
Teledyne Analytical Instruments  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operation  
2. Immediately (within 5 seconds) press either the Δ or button  
until Auto, Lo, or Hi displays on the LCD, as desired.  
After about three seconds, the analyzer resumes monitoring in the  
selected range mode.  
4.6 Oxygen Calibration  
There are two types of calibration one is air calibration, and the  
other is span which uses some other gas other than air. Since the zirconia  
sensor uses air as reference, the output is nearly zero millivolts. So, in  
doing the air calibration, you are basically reading the offset of the  
sensor. The span calibration (using gas other than air) would make the  
sensor give an output other than zero, thus it is the high calibration.  
4.6.1 Span Calibration  
Connect a suitable span gas to the port on the side of the probe  
head. To do this, remove the plug on the gland on the side of the probe  
marked REF. See Figure 4.2 Attach the span gas at this port using a  
suitable fitting and use a flow rate between 1-5 lilters per minute. The  
span gas exits at the tip of the probe. A suitable span gas can be an  
oxygen mixture between 1 to 16 %, with the balance being nitrogen.  
Using a concentration between 1 to 5% is preferable, as the sensor will  
yield a higher output.  
Power up the instrument and allow the LED reading to stabilize.  
Set the alarm setpoints and the full scale ranges to the desired values.  
Figure 4-2: Calibration Port on Probe Head  
Span Procedure:  
1. Press the SPAN button once.  
Note: The numeric LED starts flashing and will continue to flash  
Teledyne Analytical Instruments  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Model 9060L  
for five minutes before timing-out. During this five minute  
interval, the LED will continue to track gas through the  
analyzer.  
2. Flow span gas of a known oxygen concentration (between 1 and  
16% oxygen) into the analyzer. The analyzer will begin to track  
the span gas.  
Note: The alarms will not change state during span calibration.  
3. When the meter stabilizes, use the UP/DOWN arrows to adjust  
the analyzer span value to the exact percent O2 concentration in  
the span gas. Remember that the acceptable range is 1 to 16% of  
adjustment. The display will freeze at 1% or 16% if you try to  
adjust beyond those limits.  
Note: When an arrow button is first pressed, the LED begins  
flashing slightly more rapidly and no longer tracks the span  
gas. Instead, it responds to the UP/DOWN keystrokes.  
Note: While the LED is flashing slightly more rapidly, the SPAN  
routine will time-out in five seconds (instead of five  
minutes), if no further keystrokes are entered.  
4. When the span value is set to the known concentration of O2 in  
the span gas, stop pressing the keys and wait for five seconds.  
Note: When you stop pressing the UP/DOWN keys, the rapid  
flashing will cease and the analyzer will acquire the new  
span value after five seconds.  
The alarms will only be non-responsive for 60 seconds.  
This time frame allows you to reintroduce sample gas into  
the analyzer.  
5. Shut off the span gas flow and immediately flow sample into the  
analyzer.  
6. After a few minutes, disconnect the calibration gas line and  
replace the plug to cap off the calibration port.  
Teledyne Analytical Instruments  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operation  
4.6.2 Air Calibration  
To perform air calibration, make sure you have air flowing through the  
same inlet you use for span calibration at a flow of 1- 5 liters per minute. Wait  
for readings to stabilize.  
Press both the SET ALARM 1 and SET ALARM 2 simultaneously. The  
LED display will show “OFSt” for a few seconds. After that the instrument  
should read near 20.9%.  
The sensor output is expected to be near zero millivolts, if the offset of  
the sensor is too high, the analyzer will not calibrate.  
4.7 Supplementary Information  
If, during the span procedure, you pressed the SPAN button by  
mistake, you must wait five minutes for the analyzer to resume analysis  
or you can press the UP button and then the DOWN button. (Pressing  
the UP and DOWN buttons causes the analyzer to time-out in five  
seconds instead of five minutes).  
If during the span procedure, you press the RANGE or ALARM  
buttons:  
- either the range or alarm routine will be activated.  
- any changes to span will be rejected.  
- the 60 second alarm delay will not occur, i.e., the alarms  
will be responsive immediately.  
CAUTION: TI/AI CONSIDERS THE ACTION OF PRESSING THE  
ALARM OR RANGE BUTTONS TO BE AT YOUR (THE  
USER’S) DISCRETION AND NOT UNDER  
GUARANTEE OF ALARM PROTECTION.  
4.8 Probe Temperature  
The probe temperature can be displayed momentarily and adjusted  
if needed. While the instrument is in normal mode, displaying oxygen  
concentration, press the UP and DOWN arrow keys simultaneously. The  
red LED display will show the probe temperature in degrees centigrade  
for five seconds.  
Teledyne Analytical Instruments  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Model 9060L  
Within the time the temperature is being displayed, one can use the  
UP and DOWN arrow keys to adjust the temperature, thus recalibrating  
the temperature readout.  
WARNING: TEMPERATURE READING IS CALIBRATED AT THE  
FACTORY. DO NOT ATTEMPT RECALIBRATION OF  
THE PROBE TEMPERATURE. ERROR OF A FEW  
DEGREES WOULD HAVE LITTLE IMPACT ON THE  
OXYGEN CALCULATIONS, BUT IF THE PROBE IS  
GROSSLY MISCALIBRATED IT WILL HAVE AN  
ADVERSE EFFECT ON THE OXYGEN READOUT.  
ONLY QUALIFIED PERSONNEL SHOULD RE-  
CALIBRATE TEMPERATURE.  
4.8.1 Low Temperature  
For the zirconia sensor to operate properly the temperature of the  
probe must be above 650 degrees Centigrade. While the temperature is  
below this threshold, the red LED alternates showing “t-Lo” and the  
oxygen concentration.  
4.8.2 Open Thermocouple  
The electronics checks if the thermocouple (TC) is open. If it  
detects that the TC is open the Sensor failure contacts will go into  
alarming position. Red LED display may alternate between oxygen  
concentration and the error code “-02-”. Refer to Section 4.9 for  
information on Error Codes.  
4.9 Error Messages  
The instrument will display an error code when it detects a  
malfunction or error condition. The possible codes and common causes  
are listed below:  
-01- This message indicates that there is possible failure on the  
Analog to Digital Converter, ADC. This may happen when the  
micro-controller cannot communicate with the ADC. In this  
state, the oxygen reading is unreliable.  
-02- This message is displayed when the input voltage to the ADC  
rises above the upper limit of the ADC. This could occur if the  
probe is exposed to zero oxygen concentration, making the  
Teledyne Analytical Instruments  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Operation  
output of the sensor higher than the top scale of the ADC. It may  
also occur if the thermocouple should open.  
-03- This message is displayed when the input voltage to the ADC  
has fallen below the lower negative limit of the ADC.  
Teledyne Analytical Instruments  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Model 9060L  
Teledyne Analytical Instruments  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Maintenance  
Maintenance  
Aside from normal cleaning and checking for leaks at the probe  
connection, the Model 9060L should not require any maintenance  
beyond replacement of a blown fuse. Routine maintenance includes  
occasional recalibration, as described in Chapter 4, Operation.  
5.1 Replacing the Fuse  
CAUTION:  
REMOVE POWER TO UNIT BEFORE REPLACING  
THE FUSE.  
When a fuse blows, check first to determine the cause, then replace  
the fuse using the following procedure:  
1. Disconnect the AC power and place the power switch located on  
the rear panel in the O position. Remove the power cord from the  
receptacle.  
2. The fuse receptacle is located in the power cord receptacle  
assembly in the upper left-hand corner of the rear panel. See  
Figure 5-1.  
Figure 5-1: AC Fuse Replacement  
Teledyne Analytical Instruments  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 9060L  
3. Insert a small flat-blade screwdriver into the slot in the receptacle  
wall nearest the fuse and gently pry open the fuse receptacle. The  
fuse holder will slide out. There are two fuses in use and are  
visible in the clip.  
4. Remove the bad fuse and replace it with a 5x20mm 2A,  
250 VAC, IEC time lag (T) fuse (P /N F1296).  
5. Replace the fuse holder into its receptacle, pushing in firmly  
until it clicks.  
Teledyne Analytical Instruments  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Appendix  
Appendix  
A.1 Specifications  
Ranges: 0-3 % and 0-10 % oxygen (Standard  
Ranges), and 0-25 % (nominal)  
Calibration Range. User selectable %  
Ranges can be set between 1% and 100 %  
oxygen .  
Signal Output: Voltage: 0–10 VDC, negative ground  
Current: 4-20 mA, negative ground  
Range ID: 0-10 VDC.  
Display: Light emitting diode display.  
Alarms: Two customer selectable fully adjustable  
HI and LO alarms.  
One sensor failure alarm.  
Alarm relays form "C" contacts, dry  
contacts rated at 2A for 30VDC, 1A for  
115VAC.  
System Operating Temp: 0-50 °C  
Accuracy: ±2% of full scale at constant temperature  
±5% of full scale through operating temp.  
range once temp. equilibrium is reached.  
(At 3 % and higher user defined ranges.)  
Response Time: 90% in less than 4 seconds at 25 °C.  
System Power Requirements: 120 VAC, 50/60Hz (optional 220VAC),  
110W.  
System Enclosure: Panel mount. Dimensions (Approx.) 7.0  
W x 6.75 H x 4.26 D.  
Sensor Type: Zirconia, solid state  
Teledyne Analytical Instruments  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 9060L  
A.2 Probe Specifications  
Application: Combustion flue below 700°C (1290°F),  
above 700°C (1290°F) 900°C (1650°F) — no  
contaminants.  
With contaminants see notes 1 &2  
Temperature Range: 0-900°C  
800-1400°C 800-1200°C  
(32-1650°F) (1470-2550°F) (1470-2190°F)  
Length: 250-1000 mm 500-1160 mm 457-1219 mm  
(10” to 40”) (20” to 46”) (18” to 48”)  
Process Connection: 1 1/2" BSP 3/4" BSP 1" BSP or equivalent  
NPT  
Electrical Connection: Weatherproof plug-in connector or  
optional screw terminals. Plug connector  
supplied with the cable. Ex d heads have  
screw terminals.  
Cable: Order a specific length with the analyzer  
except for hazardous installations where  
the cable is supplied by the customer.  
Heater: Yes  
Flue Gas Thermocouple: K  
Response Time: Typically < 4 secs.  
Head Temperature: 125°C (250°F)  
Reference Gas: Ambient air 50 to 500 cc/min (6 to 60  
scfm). Pump supplied with analyzer  
Ref Air Connection: 1/4" NPT integral air line in probe cable.  
Barbed fitting to 3/16" ID PVC tube.  
Filter: Removable sintered stainless steel  
particulate filter, 15 micron.  
Calibration Gas Connection: 1/8" NPT female 1/8" NPT female 1/8"  
NPT female  
Weight : 0.6 kg (1.32 lbs.)  
Teledyne Analytical Instruments  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Appendix  
Notes:  
1. Care must be taken to avoid contact with explosive or  
inflammable gases with 9060L heated probes when hot.  
2. Please contact factory for corrosives other than sulphur or zinc.  
We can provide test materials to try in your atmosphere.  
A.3 Spare Parts List  
QTY. P/N  
DESCRIPTION  
1
1
2
B-76751A  
PC Board, Main  
B-76513A  
F-1296  
PC Board, Power Supply  
Fuse (AC), 2A, 250VAC, IEC Type T,  
5x20mm  
1
---  
Oxygen Probe (call TI/AI to specify U  
Length)  
1
1
B77926A  
B77926B  
Cable Assembly, Probe Heater  
Cable Assembly, Remote Probe  
Note :Orders for replacement parts should include the part  
number and the model and serial number of the system for  
which the parts are intended.  
Send orders to:  
TELEDYNE INSTRUMENTS  
Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Telephone: (626) 934-1500  
Fax: (626) 961-2538  
or your local representative.  
Teledyne Analytical Instruments  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 9060L  
A.4 Reference Drawing  
D-77712  
D77713  
C77925  
Outline Diagram  
Control Unit Final Assembly  
Interconnection Diagram  
A.5 Probe or Sensor EMF Tables  
ZIRCONIA OXYGEN SENSOR OUTPUT (mV)  
PROBE TYPE 9060L, SENSOR TYPE 9060LEX  
% OXYGEN  
21.0  
20.5  
20.0  
19.5  
19.0  
18.5  
18.0  
17.5  
17.0  
16.5  
16.0  
15.5  
15.0  
14.5  
14.0  
13.5  
13.0  
12.5  
12.0  
11.5  
mV at 720°C (1320°F)  
0.00  
0.46  
0.99  
1.53  
2.09  
2.66  
3.25  
3.85  
4.47  
5.11  
5.77  
6.45  
7.15  
7.87  
8.62  
9.40  
10.21  
11.05  
11.92  
12.83  
Teledyne Analytical Instruments  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Appendix  
11.0  
10.5  
10.0  
9.5  
9.0  
8.5  
8.0  
7.5  
7.0  
6.5  
6.0  
5.5  
5.0  
4.5  
4.0  
3.5  
3.0  
2.5  
2.0  
1.5  
1.0  
0.5  
0.2  
13.78  
14.78  
15.82  
16.92  
18.08  
19.30  
20.60  
21.98  
23.45  
25.04  
26.75  
28.61  
30.65  
32.90  
35.42  
38.28  
41.58  
45.48  
50.25  
56.41  
65.08  
79.91  
99.51  
'K’ TC mV 29.212 at 720 °C (1320°F)  
These tables are based on the Nernst equation:  
Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen),  
where T = ° K (° C + 273), e.m.f. is in mV’s  
Teledyne Analytical Instruments  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 9060L  
Teledyne Analytical Instruments  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Index  
Index  
accuracy, 43  
copyright, ii  
address, 45  
air calibration, 35, 37  
air purge, 16  
current to voltage converter, 22  
data entry keys, 13, 32  
default mode, 31  
alaram delay, 36  
alarm, 43  
digital concentration signal, 22  
digital meter. See display  
concentration, 12  
HI, 27  
digital to analog converter, 20, 22  
dimensions, 43  
LO, 27  
display, 11, 13, 43  
display control, 20  
display, frozen, 36  
Down Arrow, 13  
E–I converter, 22  
EMF table, 46  
sensor fail, 27, 34  
sensor failure, 12  
setpoint, 33  
threshold, 12, 27  
Alarm 1, 27  
Alarm 2, 27  
error code, 38  
alarm connection, 26  
alarm connections, 14  
alarm delay, 37  
failsafe, 27  
features, 11  
figures listing, viii  
filter, 17, 24, 44  
flow guide tube. See guide tube  
flow rate, 37  
alarm indicator, 12  
alarm relay circuit, 27  
alarm setpoint, 13  
analog concentration signal, 22  
analog to digital converter, 20, 21, 38  
analysis range, 32  
analyze mode, 32  
applications, 11  
flow rate, span gas, 35  
flue velocity, 17  
form "C" contacts, 43  
Form C relay contacts, 26  
front panel, 12, 31  
function keys, 12  
fuse, 25, 42  
autoranging, 11  
block diagram, 21  
BSP socket, 23  
fuse block, 25  
cable, 44  
calibration, 35  
fuse replacement, 41  
gas connection, 44  
general purpose area, 23  
guide tube, 17, 24  
heated probe, 16  
heater, 19  
heater control, 18  
heater power, 20  
HI range, 32  
calibration check, 16  
calibration mode, 32  
calibration mode keys, 32  
calibration range, 11  
caution sign, iv  
CMOS, 11  
combustible gas warning, x  
concentration output. See output  
contact ID, 27  
contaminants, 15  
control unit, 15, 16  
impedance, 20  
impedance alarm, 20  
increment the display, 32  
initial setup, 32  
Teledyne Analytical Instruments  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
Model 9060L  
input connector, 13  
insertion length, 17  
instasllation checklist, 29  
instrument range, 11  
internal power switch, 20, 25  
lagging extension, 17  
leak check, 29  
probe mounting angle, 24  
probe options, 17  
probe temperature, 13, 21, 37  
process pressure. See pressure  
pump, 18, 44  
RAM, 11, 20  
range, 13, 43  
LED, 13  
range ID, 22, 43  
LED flashing, 35  
LO range, 33  
main PCB, 21  
maintenance, 41  
range identification, 12  
range switch-over point, 33  
rear panel, 13, 14, 25  
REF, 35  
manuals, additional, v  
Measurement Parameters Readout, 13  
membrane switches, 12  
Meter Readout, 13  
micro-controller, 15, 20, 21  
millivolt potential, 21  
mounting thread, 17  
Nernst equation, 19, 47  
NPT socket, 23  
reference air, 16  
reference air pump, 18  
reference gas, 44  
relay, 18, 22, 43  
relay contacts, 26  
response time, 15, 24, 43, 44  
ROM, 11, 20  
safety information, iv  
sampling line, 18  
OFSt, 37  
sensor, 11, 15, 16, 18, 20  
sensor amplifier, 20, 21  
sensor connector, 14  
sensor offset, 35  
open thermocouple, 27, 34, 38  
operating temperature, 43  
operation, 31  
output, 15, 43  
sensor temperature, 19  
serial number, iii  
analog, 12, 14  
current, 26  
Set Alarm 1, 12  
range ID, 26  
voltage, 26  
SET ALARM 1, 32, 33  
Set Alarm 2, 13  
output connector, 13  
oxygen partial pressure, 19  
oxygen reading reliability, 20  
oxygen reading unreliable, 38  
panel mount, 23  
SET ALARM 2, 32, 34  
Set HI Range, 13  
SET HI RANGE, 32, 33  
Set LO Range, 13  
SET LO RANGE, 32, 33  
setup function keys, 32  
setup mode, 32  
parts list, 45  
power, 20  
power connection, 14  
power cord, 25  
sheath, 16  
signal processing, 15, 20  
solenoid, 16  
solid electrolyte, 18  
Span, 13  
SPAN, 32, 35  
span calibration. See calibration  
span calibration point, 13  
span gas, 35  
span gas concentration, 35  
span value acquisition time, 36  
stratification, 25  
power cord receptacle, 25, 41  
power requirements, 43  
power supply PCB, 20  
power switch, 14, 25  
pressure, 18  
pressure compensation, 20  
probe, 15, 16, 18, 20  
probe cable connection, 28  
probe configuration, 17  
probe installation, 23  
probe length, 17, 44  
system parameters, 31  
Teledyne Analytical Instruments  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Oxygen Analyzer  
Index  
tables listing, ix  
transit time, 24  
technician symbol, iv  
temperature, 44  
U length, 24  
Up Arrow, 13  
temperature controller, 18  
temperature maximum, 24  
temperature too low, 38  
thermal shock, 15, 24  
thermocouple, 18  
thermocouple type, 44  
time out, 34  
voltage amplifier, 21  
warning sign, iv  
warranty, ii  
water vapor, 16  
website address, v  
zero oxygen concentration, 38  
zirconia element, 18  
zirconia sensor. See sensor  
time-out, 37  
t-Lo, 38  
Teledyne Analytical Instruments  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Texas Instruments Network Card TMS320C6000 User Manual
Toro Marine Radio Network 8000 User Manual
Toshiba Printer e STUDIO500P User Manual
Tribest Video Gaming Accessories FL 3000 User Manual
Tyco Home Security System Power608 User Manual
Uniden Baby Monitor UBR223 User Manual
Univex Refrigerator Refrigerator Freezer User Manual
ViewSonic Mouse KU709 User Manual
ViewSonic Network Card WPCC100 User Manual
Vivotek Security Camera FD8365EHV User Manual