OPERATING INSTRUCTIONS FOR
Model 3000PA
Percent Oxygen Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS
MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PER-
SIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
P/NM64573
08/06/99
ECO:#99-0323
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.
BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED
SUPERVISOR/MANAGER.
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Specific Model Information
Theinstrumentforwhichthismanualwassuppliedmayincorporateoneor
moreoptionsnotincludedwiththestandardinstrument.Commonlyavailable
optionsarelistedbelow,withcheckboxes.Anythatareincorporatedinthe
instrumentforwhichthismanualissuppliedareindicatedbyacheckmarkinthe
box.
InstrumentSerialNumber _______________________
includesthefollowingoptions:
! 3000PA-C:
Inadditiontoallstandardfeatures,thismodelalsohas
separateportsforzeroandspangases,andbuilt-in
controlvalves.Theinternalvalvesareentirelyunderthe
controlofthe3000PAelectronics,toautomatically
switchbetweengasesinsynchronizationwiththe
analyzer’soperations
! 3000PA-S:
Inmodelswiththisoption,allwettedpartsaremade
from316stainlesssteel.
! 3000PA-M:
Inmodelswiththisoption,the4-20mAAnalogCurrent
outputisactive.(Inthestandardunits,itisnotactive.)
! 19" Rack Mnt: The19"RelayRackMountunitsareavailablewith
eitheroneortwo3000seriesanalyzersinstalledona
19"panel,andreadytomountinastandardrack.
! Cell Class:
___________________ See Maintenance for Specs.
Enter Class Designation.
Teledyne Analytical Instruments
iii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 3000PA
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-1
1.3 Main Features of the Analyzer ....................................... 1-1
1.4 Model Designations ....................................................... 1-2
1.5 Front Panel (Operator Interface) ..................................... 1-3
1.6 Recognizing Difference Between LCD & VFD............... 1-5
1.7 Rear Panel (Equipment Interface) .................................. 1-5
2 OperationalTheory
2.1 Introduction .................................................................... 2-1
2.2 Micro-Fuel Cell Sensor .................................................. 2-1
2.2.1 Principles of Operation ............................................ 2-1
2.2.2 Anatomy of a Micro-Fuel Cell .................................. 2-2
2.2.3 Electrochemical Reactions ...................................... 2-3
2.2.4 The Effect of Pressure.............................................. 2-4
2.2.5 Calibration Characteristics ...................................... 2-4
2.2.6 Micro-Fuel Cell “Class”............................................ 2-5
2.3 Sample System.............................................................. 2-6
2.4 Electronics and Signal Processing ................................ 2-8
3 Installation
3.1 Unpacking the Analyzer................................................. 3-1
3.2 Mounting the Analyzer ................................................... 3-1
3.3 Rear Panel Connections ................................................ 3-2
3.3.1 Gas Connections ................................................... 3-3
3.3.2 Electrical Connections ........................................... 3-4
3.3.3 Remote Probe Connector ...................................... 3-9
3.4 Installing the Micro-Fuel Cell ......................................... 3-9
3.5 Testing the System.........................................................3-11
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons ................... 4-2
4.3 The System Function ..................................................... 4-3
4.3.1 Setting the Display................................................. 4-4
4.3.2 Setting up an Auto-Cal........................................... 4-5
4.3.3 Password Protection .............................................. 4-5
4.3.3.1 Entering the Password ................................... 4-6
Teledyne Analytical Instruments
iv
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
4.3.3.2 Installing or Changing the Password ............. 4-7
4.3.4 Logout.................................................................... 4-8
4.3.5 System Self-Diagnostic Test .................................. 4-9
4.3.6 Version Screen ...................................................... 4-9
4.4 The Zero and Span Functions ....................................... 4-10
4.4.1 Cell Failure ............................................................ 4-12
4.4.2 Span Cal................................................................ 4-11
4.4.2.1 Auto Mode Spanning ..................................... 4-11
4.4.2.2 Manual Mode Spanning................................. 4-12
4.5 The Alarms Function...................................................... 4-12
4.6 The Range Function ...................................................... 4-15
4.6.1 Setting the Analog Output Ranges......................... 4-16
4.6.2 Autoranging Analysis............................................. 4-16
4.6.3 Fixed Range Analysis............................................ 4-16
4.7 The Analyze Function.................................................... 4-17
4.8 Signal Output ................................................................. 4-17
Maintenance
5.1 Routine Maintenance..................................................... 5-1
5.2 Cell Replacement .......................................................... 5-1
5.2.1 Storing and Handling Replacement Cells ............... 5-1
5.2.2 When to Replace a Cell ........................................... 5-2
5.2.3 Removing the Micro-Fuel Cell ................................. 5-3
5.2.4 Installing a New Micro-Fuel Cell.............................. 5-5
5.2.5 Cell Warranty ........................................................... 5-5
5.3 Fuse Replacement......................................................... 5-6
5.4 System Self Diagnostic Test........................................... 5-6
5.5 Major Internal Components............................................ 5-7
5.6 Cleaning ........................................................................ 5-8
5.7 Troubleshooting ............................................................. 5-9
Appendix
A-1 Model 3000PA Specifications ........................................ A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List ................................................................... A-4
A-4 19-Inch Relay Rack Panel Mount................................... A-4
A-5 Application Notes on Restrictors, Pressures & Flow...... A-5
A-6 Zero Functions............................................................... A-8
Teledyne Analytical Instruments
v
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 3000PA
DANGER
COMBUSTIBLE GAS USAGE WARNING
This is a general purpose instrument designed for usage in a
nonhazardous area. It is the customer's responsibility to ensure
safety especially when combustible gases are being analyzed
since the potential of gas leaks always exist.
The customer should ensure that the principles of operating of
this equipment is well understood by the user. Misuse of this
product in any manner, tampering with its components, or unau-
thorized substitution of any component may adversely affect the
safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne,no responsibility byTeledyne, its affiliates, and agents
for damage or injury from misuse or neglect of this equipment is
implied or assumed.
Teledyne Analytical Instruments
vi
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Introduction 1
Introduction
1.1 Overview
TheTeledyneAnalyticalInstruments
Model 3000PA Percent Oxygen Analyzer is a versatile microprocessor-
based instrument for detecting the percentage of oxygen in a variety of
background gases. This manual covers the Model 3000PA General Purpose
flush-panel and/or rack-mount units only. These units are for indoor use in a
nonhazardousenvironment.
1.2 Typical Applications
A few typical applications of the Model 3000PA are:
•
•
•
•
•
•
•
Monitoringinertgasblanketing
Airseparationandliquefaction
Chemicalreactionmonitoring
Semiconductormanufacturing
Petrochemicalprocesscontrol
Qualityassurance
Gasanalysiscertification.
1.3 Main Features of the Analyzer
The Model 3000PA Percent Oxygen Analyzer is sophisticated yet
simple to use. The main features of the analyzer include:
•
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•
High resolution, accurate readings of oxygen content from low
percent levels through 100 %. Large, bright, meter readout.
Teledyne Analytical Instruments
1-1
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 3000PA
•
Advanced Micro-Fuel Cell, designed for percent oxygen
analysis. Several options are available.
•
•
Versatile analysis over a wide range of applications.
Microprocessorbasedelectronics:8-bitCMOSmicroprocessor
with 32 kB RAM and 128 kB ROM.
•
Three user definable output ranges (from 0-1% through 0-100 %)
allow best match to users process and equipment.
•
•
Air-calibration range for convenient spanning at 20.9 %.
Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•
•
Twoadjustableconcentrationalarmsandasystemfailurealarm.
Extensive self-diagnostic testing, at startup and on demand, with
continuouspower-supplymonitoring.
•
•
Two way RFI protection.
RS-232 serial digital port for use with a computer or other digital
communicationdevice.
•
•
Analog outputs for percent-of-range and for range identification.
0–1 V dc. (Isolated 4–20 mA dc optional)
Convenient and versatile, steel, flush-panel orrack-mountable
case with slide-out electronics drawer.
1.4 Model Designations
3000PA:
Standardmodel.
3000PA-C: In addition to all standard features, this model also has
separate ports for zero and span gases, and built-in control
valves. The internal valves are entirely under the control of
the 3000PA electronics, to automatically switch between
gases in synchronization with the analyzer’s operations.
3000PA-M: This model has current output signals (4-20 mA) for percent-
of-range and range ID, in addition to voltage outputs.
3000PA-S: A Stainless Steel Probe and Probe Holder are used in this
model, for use where resistance to corrosion is important.
Teledyne Analytical Instruments
1-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Introduction 1
All of the above options are available in combination. For example, the
-C and -V options are combined as Model 3000PA-C-V.
Figure 1-1: Model 3000PA Front Panel
1.5 Front Panel (Operator Interface)
The standard 3000PA is housed in a rugged metal case with all controls
and displays accessible from the front panel. See Figure 1-1. The front panel
has thirteen buttons for operating the analyzer, a digital meter, an alphanu-
meric display, and a window for viewing the sample flowmeter.
Teledyne Analytical Instruments
1-3
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 3000PA
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the analyzer:
•
•
Analyze Perform analysis for oxygen content of a sample gas.
System Performsystem-relatedtasks(describedindetailin
chapter 4, Operation.).
•
•
•
•
Span
Zero
Span calibrate the analyzer.
Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
•
•
•
•
Left & Right Arrows
Selectbetweenfunctionscurrently
displayed on theVFD screen.
Up & Down Arrows
Increment or decrement values of
functionscurrentlydisplayed.
Enter Moves VFD display on to the next screen in a series. If
none remains, returns to the Analyze screen.
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting
environment. It produces a continuous readout from 0-100 %. It is accurate
across all ranges without the discontinuity of analog range switching.
Alphanumeric Interface Screen: The VFD screen is an easy-to-use
interface from operator to analyzer. It displays values, options, and messages
that give the operator immediate feedback.
Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to
2.4 standard liters per minute (SLPM).
Standby Button: The
Sta ndby turns off the display and outputs,
butcircuitryisstilloperating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits.
Teledyne Analytical Instruments
1-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Introduction 1
Access Door: To provide access to the Micro-Fuel Cell, the front panel
swings open when the latch in the upper right corner of the panel is pressed
all the way in with a narrow gauge tool. Accessing the main circuit board
requires unfastening the rear panel screws and sliding the unit out of the
case.
1.6 Recognizing Difference Between LCD &
VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.7 Rear Panel (Equipment Interface)
The rear panel, shown in Figure 1-2, contains the gas and electrical
connectors for external inlets and outlets. The Zero and Span gas connectors,
and the Current signal outputs are optional and may not appear on your
instrument. The connectors are described briefly here and in detail in the
Installationchapterofthismanual.
Figure 1-2: Model 3000PA Rear Panel
•
•
Power Connection
Universal AC power source.
Gas Inlet and Outlet One inlet (must be externally valved)
and one exhaust out.
•
Analog Outputs
0-1 V dc concentration output, plus
0-1 V dc range ID.
Teledyne Analytical Instruments
1-5
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 3000PA
•
•
•
•
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
RS-232 Port
Serialdigitalconcentrationsignaloutput
and control input.
Remote Probe
Remote Span/Zero
Used in the 3000PA for controlling
external solenoid valves only.
Digital inputs allow external control of
analyzer calibration. (See Note, below.)
Calibration Contact To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
•
Range ID Contacts
Network
Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
For future expansion. Not implemented
atthisprinting.
Optional:
Calibration Gas Ports
•
Separate fittings for zero, span and
sample gas input, and internal valves
forautomaticallyswitchingthegases.
• Current Signal Output Additional isolated 4-20 mA dc plus
4-20 mA dc range ID.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000PA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Teledyne Analytical Instruments
1-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operational Theory 2
Operational Theory
2.1 Introduction
The analyzer is composed of three subsystems:
1. Micro-FuelCellSensor
2. SampleSystem
3. Electronic Signal Processing, Display and Control
The sample system is designed to accept the sample gas and transport it
through the analyzer without contaminating or altering the sample prior to
analysis. The Micro-Fuel Cell is an electrochemical galvanic device that
translates the amount of oxygen present in the sample into an electrical
current. The electronic signal processing, display and control subsystem
simplifies operation of the analyzer and accurately processes the sampled
data. The microprocessor controls all signal processing, input/output and
display functions for the analyzer.
2.2 Micro-Fuel Cell Sensor
2.2.1 Principles of Operation
The oxygen sensor used in the Model 3000P series is a Micro-Fuel Cell
designed and manufactured by Analytical Instruments. It is a sealed plastic
disposableelectrochemicaltransducer.
The active components of the Micro-Fuel Cell are a cathode, an anode,
and the 15% aqueous KOH electrolyte in which they are immersed. The cell
converts the energy from a chemical reaction into an electrical current in an
external electrical circuit. Its action is similar to that of a battery.
There is, however, an important difference in the operation of a battery
as compared to the Micro-Fuel Cell: In the battery, all reactants are stored
within the cell, whereas in the Micro-Fuel Cell, one of the reactants (oxygen)
comes from outside the device as a constituent of the sample gas being
Teledyne Analytical Instruments
2-1
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 3000PA
analyzed. The Micro-Fuel Cell is therefore a hybrid between a battery and a
true fuel cell. (All of the reactants are stored externally in a true fuel cell.)
2.2.2 Anatomy of a Micro-Fuel Cell
The Micro-Fuel Cell is a cylinder only 1¼ inches in diameter and 1
inch thick. It is made of extremely inert plastic, which can be placed confi-
dently in practically any environment or sample stream. It is effectively
sealed, although one end is permeable to oxygen in the sample gas. The
other end of the cell is a contact plate consisting of two concentric foil rings.
The rings mate with spring-loaded contacts in the sensor block assembly and
provide the electrical connection to the rest of the analyzer. Figure 2-1
illustratestheexternalfeatures.
Figure 2-1: Micro-Fuel Cell
Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which illus-
tratesthefollowinginternaldescription.
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)
At the top end of the cell is a diffusion membrane of Teflon, whose
thickness is very accurately controlled. Beneath the diffusion membrane lies
Teledyne Analytical Instruments
2-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operational Theory 2
the oxygen sensing element—the cathode—with a surface area almost 4 cm2.
The cathode has many perforations to ensure sufficient wetting of the upper
surface with electrolyte, and it is plated with an inert metal.
The anode structure is below the cathode. It is made of lead and has a
proprietary design which is meant to maximize the amount of metal available
forchemicalreaction.
At the rear of the cell, just below the anode structure, is a flexible
membrane designed to accommodate the internal volume changes that occur
throughout the life of the cell. This flexibility assures that the sensing mem-
brane remains in its proper position, keeping the electrical output constant.
The entire space between the diffusion membrane, above the cathode,
and the flexible rear membrane, beneath the anode, is filled with electrolyte.
Cathode and anode are submerged in this common pool. They each have a
conductor connecting them to one of the external contact rings on the contact
plate, which is on the bottom of the cell.
2.2.3 Electrochemical Reactions
The sample gas diffuses through the Teflon membrane. Any oxygen in
the sample gas is reduced on the surface of the cathode by the following
HALF REACTION:
O2 + 2H2O + 4e– → 4OH–
(cathode)
(Four electrons combine with one oxygen molecule—in the presence of
water from the electrolyte—to produce four hydroxyl ions.)
When the oxygen is reduced at the cathode, lead is simultaneously
oxidized at the anode by the following HALF REACTION:
Pb + 2OH– → Pb+2 + H2O + 2e–
(anode)
(Two electrons are transferred for each atom of lead that is oxidized.
Therefore it takes two of the above anode reactions to balance one cathode
reactionandtransferfourelectrons.)
The electrons released at the surface of the anode flow to the cathode
surface when an external electrical path is provided. The current is propor-
tional to the amount of oxygen reaching the cathode. It is measured and used
to determine the oxygen concentration in the gas mixture.
The overall reaction for the fuel cell is the SUM of the half reactions
above, or:
2Pb + O2 → 2PbO
Teledyne Analytical Instruments
2-3
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 3000PA
(These reactions will hold as long as no gaseous components capable of
oxidizing lead—such as iodine, bromine, chlorine and fluorine—are present
in the sample.)
The output of the fuel cell is limited by (1) the amount of oxygen in the
cell at the time and (2) the amount of stored anode material.
In the absence of oxygen, no current is generated.
2.2.4 The Effect of Pressure
In order to state the amount of oxygen present in the sample as a per-
centage of the gas mixture, it is necessary that the sample diffuse into the cell
under constant pressure.
If the total pressure increases, the rate that oxygen reaches the cathode
through the diffusing membrane will also increase. The electron transfer, and
therefore the external current, will increase, even though the oxygen concen-
tration of the sample has not changed. It is therefore important that the
sample pressure at the fuel cell (usually vent pressure) remain constant
betweencalibrations.
2.2.5 Calibration Characteristics
Given that the total pressure of the sample gas at the surface of the
Micro-Fuel Cell input is constant, a convenient characteristic of the cell is
that the current produced in an external circuit is directly proportional to the
rate at which oxygen molecules reach the cathode, and this rate is directly
proportional to the concentration of oxygen in the gaseous mixture. In other
words it has a linear characteristic curve, as shown in Figure 2-3. Measuring
circuits do not have to compensate for nonlinearities.
In addition, since there is zero output in the absence oxygen, the charac-
teristic curve has close to an absolute zero. In the percent ranges, the cell
itself does not need to be zeroed. In practical application zeroing is still used
to compensate for zero offsets in the electronics. (The electronics is zeroed
automatically when the instrument power is turned on.)
Teledyne Analytical Instruments
2-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operational Theory 2
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Cell
2.2.6 Micro-Fuel Cell “Class”
TBE manufactures Micro-Fuel Cells with a variety of characteristics to
give the best possible performance for any given sample conditions. A few
typical Micro-Fuel Cells are listed below with their typical use and electrical
specifications.
2.2.6.1 Class A-3 Cell
The class A-3 cell is for use in applications where it is exposed continu-
ously to carbon dioxide concentrations between 1 % and 100 % in the
samplegas.
Nominal output in air is 0.20 mA, and 90 % response time is 45 s.
Expected life in flue gas is 8 months.
2.2.6.2 Class A-5 Cell
The class A-5 cell is for use in applications where it is exposed intermit-
tently to carbon dioxide concentrations up to 100 % in the sample gas.
Teledyne Analytical Instruments
2-5
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 3000PA
Nominal output in air is 0.19 mA, and 90 % response time is 45 s.
Expected life in flue gas is 8 months.
2.2.6.3 Class B-1 Cell
The class B-1 cell is for use in applications where it is exposed to less
than 0.1 % of carbon dioxide, and where fast response is important.
Nominal output in air is 0.50 mA, and 90 % response time is 7 s.
Expected life in air is 8 months.
2.2.6.4 Class B-3 Cell
The class B-3 cell is for use in applications where a slightly longer
response time is acceptable in order to have a longer cell life.
Nominal output in air is 0.30 mA, and 90 % response time is 13 s.
Expected life in air is 12 months.
2.2.6.5 Class C-3 Cell
The class B-1 cell is for use in applications where it is exposed to less
than 0.1 % of carbon dioxide, and where a longer response time is accept-
able in order to have a longer cell life.
Nominal output in air is 0.20 mA, and 90 % response time is 30 s.
Expected life in air is 18 months.
2.2.6.6 Hydrogenand/orHeliumService
If the sample gas contains 10 % or more hydrogen and/or helium,
“clamp” cells are used. These Micro-Fuel cells are identified by the suffix -C
added to the cell class number.
2.3 Sample System
The sample system delivers gases to the Micro-Fuel Cell sensor from
the analyzer rear panel inlet. Depending on the mode of operation either
sample or calibration gas is delivered.
The Model 3000P sample system is designed and fabricated to ensure
that the oxygen concentration of the gas is not altered as it travels through the
sample system. The sample encounters almost no dead space. This mini-
Teledyne Analytical Instruments
2-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operational Theory 2
mizes residual gas pockets that can interfere with very low level oxygen
analysis.
The sample system for the standard instrument incorporates ¼ inch tube
fittings for sample inlet and outlet connections at the rear panel. For metric
system installations, 6 mm adapters are supplied with each instrument. The
sample or calibration gas flow through the system is monitored by a flow-
meter downstream from the cell. Figure 2-4 shows the piping layout for the
standardmodel.
Figure 2-4: Piping Layout and Flow Diagram for Standard Model
Figure 2-5 is the flow diagram for the sampling system. In the standard
instrument, calibration gases (zero and span) can be connected directly to the
Sample In port by teeing to the port with appropriate valves. The shaded
portion of the diagram shows the components added when the –C option is
ordered. The valving is installed inside the 3000PA-C enclosure and is
regulated by the instrument's internal electronics.
Teledyne Analytical Instruments
2-7
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 3000PA
Components in the shaded area are in
Span In
Zero In
the -C option (internal control valves)
only and are not shown in the piping
diagram above.
Cell
Sample In
Solenoid
Valves
In vacuum service the
Flowmeter
restrictor should be
placed here .
In normal service the
restrictor should be
placed here .
Exhaust Out
Restrictor
Figure 2-5: Flow Diagram
2.4 Electronics and Signal Processing
The Model 3000P Percent Oxygen Analyzer uses an 8031 microcon-
troller with 32 kB of RAM and 128 kB of ROM to control all signal pro-
cessing, input/output, and display functions for the analyzer. System power
is supplied from a universal power supply module designed to be compatible
with any international power source. Figure 2-6 shows the location of the
power supply and the main electronic PC boards.
Figure 2-6: Location of Electronic Components
Teledyne Analytical Instruments
2-8
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operational Theory 2
The signal processing electronics including the microprocessor, analog
to digital, and digital to analog converters are located on the motherboard at
the bottom of the case. The preamplifier board is mounted on top of the
motherboard as shown in the figure. These boards are accessible after re-
moving the back panel. Figure 2-7 is a block diagram of the Analyzer
electronics.
Figure 2-7: Block Diagram of the Model 3000P Electronics
Teledyne Analytical Instruments
2-9
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 3000PA
In the presence of oxygen the cell generates a current. A current to
voltage amplifier converts this current to a voltage, and then the voltage is
amplified in the second stage amplifier.
The second stage amplifier also supplies temperature compensation for
the oxygen sensor output. This amplifier circuit incorporates a thermistor,
which is physically located in the cell block. The thermistor is a temperature
dependent resistance that changes the gain of the amplifier in proportion to
the temperature changes in the block. This change is inversely proportional
to the change in the cell output due to the same temperature changes. The
result is a signal that is temperature independent. The output from the second
stage amplifier is sent to an 18 bit analog to digital converter controlled by
themicroprocessor.
The digital concentration signal along with input from the control panel
is processed by the microprocessor, and appropriate control signals are
directed to the display, alarms and communications port. The same digital
information is also sent to a 12 bit digital to analog converter that produces
the 0-1 V dc analog percent-of-range signal output and the analog range ID
output. Models with the –MA option also have a 4-20 mA dc percent-of-
range signal output and analog range ID output.
Signals from the power supply are also monitored by the microproces-
sor, and the system failure alarm is activated if a malfunction is detected.
Teledyne Analytical Instruments
2-10
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Installation 3
Installation
Installation of the Model 3000PA Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electricalconnections
5. InstallingtheMicro-FuelCell
6. Testing the system.
3.1 Unpacking the Analyzer
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the analyzer and inspect
it for damage. Immediately report any damage to the shipping agent.
3.2 Mounting the Analyzer
The Model 3000PA is for indoor use in a general purpose area. It is
NOT for use in hazardous environments of any type.
The standard model is designed for flush panel mounting. Figure 3-1 is
an illustration of the 3000PA standard front panel and mounting bezel. There
are four mounting holes—one in each corner of the rigid frame. Drawing
number D-62928, in the Drawings section in the rear of this manual, con-
tains a panel cutout diagram.
On special order, a 19" rack-mounting panel can be provided. For rack
mounting, one or two 3000 series analyzers are flush-panel mounted on the
rack panel. See Appendix for dimensions of the mounting panel.
Teledyne Analytical Instruments
3-1
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 3000PA
6.7”
10”
Figure 3-1: Front Panel of the Model 3000PA
All operator controls are mounted on the control panel, which is hinged
on the left edge and doubles as the door that provides access to the sensor
and cell block inside the instrument. The door is spring loaded and will
swing open when the button in the center of the latch (upper right corner) is
pressed all the way in with a narrow gauge tool (less than 0.18 inch wide),
such as a small hex wrench or screwdriver Allow clearance for the door to
open in a 90-degree arc of radius 7.125 inches. See Figure 3-2.
Figure 3-2: Required Front Door Clearance
3.3 Rear Panel Connections
Figure 3-3 shows the Model 3000PA rear panel. It contains all of the
gas and electrical inputs and outputs. Some ports are optional equipment.
Refer to page iii in the front of this manual for options included in your
instrument. Be sure to note the instrument serial number.
Teledyne Analytical Instruments
3-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Installation 3
Figure 3-3: Rear Panel of the Model 3000PA
3.3.1 Gas Connections
Before using this instrument, it should be determined if the unit will be
used for pressurized service or vacuum service and low pressure applica-
tions. Inspect the restrictor kit that came with the unit. The kit consist of two
restrictors and a union for 1/4” diameter tubing. Notice that the two 1 3/4”
long, 1/4” diameter tubing are restrictors. It has an open end and a closed
end with a small circular orifice. The restrictor without the blue sticker is for
;ow pressure and vacuum service. For high pressure (5 to 50 psig) applica-
tions, use the restrictor that has a blue sticker on the body.
For pressurized service, use the restrictor without the blue dot and union
from the restrictor kit and attach it to the Sample In port. The small circular
orifice should face away from the back of the unit (against the direction of
gas flow). Use the restrictor without the blue dot sticker in the same manner
for low pressure applications (less than 5 psig).
For vacuum service (5-10 in Hg), use the restrictor without the blue dot
sticker and union but attach it to the Exhaust Out port. The small circular
orifice should face toward the back of the unit (against the direction of gas
flow).
Remove the blue sticker from the restrictor before using.
WARNING:
Operating the unit without restrictors can cause damage to t
the micro-fuel cell.
Teledyne Analytical Instruments
3-3
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 3000PA
The unit is manufactured with 1/4 inch tube fittings. Six millimeter
adapters are supplied for metric system installations. For a safe connection:
1. Insert the tube into the tube fitting, and finger-tighten the nut until
the tubing cannot be rotated freely, by hand, in the fitting. (This
may require an additional 1/8 turn beyond finger-tight.)
2. Hold the fitting body steady with a backup wrench, and with
another wrench rotate the nut another 1 1/4 turns.
SAMPLE IN: In the standard model, gas connections are made at the
SAMPLE IN and EXHAUST OUT connections. Calibration gases must be
Tee'd into the Sample inlet with appropriate valves.
Ensure that the gas pressure is reasonably regulated. Pressures between
3 and 40 psig are acceptable as long as the pressure, once established, will
keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4
SLPM). Exact figures will depend on your process.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
Note: If the unit is for vacuum service, the above numbers apply
instead to the vacuum at the EXHAUST OUT connector, de-
scribed below, with minus signs before the pressure readings.
EXHAUST OUT: Exhaust connections must be consistent with the
hazard level of the constituent gases. Check Local, State, and Federal laws,
and ensure that the exhaust stream vents to an appropriately controlled area if
required.
Note: If the unit is for vacuum service, see Sample In, above, for gas
pressure/flow considerations.
ZERO IN and SPAN IN (Optional): These are additional input ports
for span gas and zero gas. There are electrically operated valves inside for
automatic switching between sample and calibration gases. These valves are
under control of the 3000P Electronics. They can be externally controlled
only indirectly through the Remote Cal Inputs, described below.
Pressure, flow, and safety considerations are the same as prescribed for
the SAMPLE IN inlet, above.
3.3.2 Electrical Connections
For safe connections, ensure that no uninsulated wire extends outside of
the connectors they are attached to. Stripped wire ends must insert com-
pletely into terminal blocks. No uninsulated wiring should be able to come in
contact with fingers, tools or clothing during normal operation.
Teledyne Analytical Instruments
3-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Installation 3
Primary Input Power: The power cord receptacle and fuse block are
located in the same assembly. Insert the female plug end of the power cord
into the power cord receptacle.
CAUTION: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The switch on the front panel is for
switching power on or off to the displays and out-
puts only.
The universal power supply requires a 85–250 V ac, 47-63 Hz power
source.
Fuse Installation: The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the fuse in
whichever fuse receptacle is not used. The fuses are not installed at the
factory. Be sure to install the proper fuse as part of installation. (See Fuse
Replacement in chapter 5, maintenance.)
Analog Outputs: There are four DC output signal connectors with
spring terminals on the panel. (Two of them are optional, as explained
below.) There are two wires per output with the polarity noted. See Figure 3-
4. The outputs are:
0-1 V dc % Range:
Voltage rises linearly with increasing oxygen con-
centration, from 0 V at 0 percent to 1 V at full scale.
(Full scale = 100% of programmable range.)
0-1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.
4-20 mA dc % Range: –M option only. Current increases linearly with
increasing oxygen concentration, from 4 mA at 0
percent to 20 mA at full scale. (Full scale = 100% of
programmablerange.)
4-20 mA dc Range ID: –M option only. 8 mA = Low Range, 12 mA =
Medium Range, 16 mA = High Range, 20 mA = Air
Cal Range.
Teledyne Analytical Instruments
3-5
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 3000PA
Figure 3-4: Analog Output Connections
Alarm Relays: The three alarm-circuit connectors are spring terminals
for making connections to internal alarm relay contacts. Each provides a set
of Form C contacts for each type of alarm. Each has both normally open and
normally closed contact connections. The contact connections are indicated
by diagrams on the rear panel. They are capable of switching up to 3 am-
peres at 250 V ac into a resistive load. See Figure 3-5. The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentrationisbelowthreshold).
• Can be configured as failsafe or nonfailsafe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentrationisbelowthreshold).
• Can be configured as failsafe or nonfailsafe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
SystemAlarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be de-
feated. Actuates if self test fails.
(Reset by pressing
press again and any other button EXCEPT
Systemtoresume.
button to remove power. Then
Further detail can be found in chapter 4, section 4-5.
Figure 3-5: Types of Relay Contacts
Teledyne Analytical Instruments
3-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Installation 3
Digital Remote Cal Inputs: Accept 0 V (off) or 24 V dc (on) inputs
for remote control of calibration. (See Remote Calibration Protocol below.)
Zero:
Floating input. A 5 to 24 V pulse input across the + and –
terminals puts the analyzer into the Zero mode. Either side
may be grounded at the source of the signal. A synchronous
signal must open and close the external zero valve appropri-
ately. See 3.3.3 Remote Probe Connector. (The –C option
internalvalvesoperateautomatically.)
Span:
Floating input. A 5 to 24 V pulse input across the + and –
terminals puts the analyzer into the Span mode. Either side
may be grounded at the source of the signal. A synchronous
signal must open and close external span valve appropriately.
See 3.3.3 Remote Probe Connector. (The –C option internal
valvesoperateautomatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3000PA Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
closeagain.
Forexample:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Teledyne Analytical Instruments
3-7
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 3000PA
Note: The Remote Probe connector (paragraph 3.3.3) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the –C Internal valve
option—which includes additional zero and span gas inputs—
the 3000P automatically regulates the zero, span and sample
gas flow.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending order—Low range is as-
signed to Range 1 ID, Medium range is assigned to Range 2 ID, and High
range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal
Range (25%).
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. It requires a standard 9-pin D connector.
The data is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•
•
•
•
•
The concentration in percent
The range in use (HI, MED, LO)
The span of the range (0-10 %, etc)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Four input functions using RS-232 have been implemented to date.
They are described in Table 3-1.
Table 3-1: Commands via RS-232 Input
Command
as<enter>
az<enter>
co<enter>
Description
Immediatelystartsanautospan.
Immediatelystartsanautozero.
Reports "Raw Cell Output" (current output of the sensor
itself) in µA. For example—
RawCellOutput: 99 µA.
Teledyne Analytical Instruments
3-8
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Installation 3
st<enter>
Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-2 lists certain RS-232 values that are required by the 3000PA
implementation.
Table 3-2: Required RS-232 Options
Parameter
Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits
MessageInterval
1
2 seconds
3.3.3 Remote Probe Connector
The 3000PA is a single-chassis instrument, which has no Remote Probe
Unit. Instead, the Remote Probe connector is used as another method for
controlling external sample/zero/span gas valves. See Figure 3-6.
Figure 3-6: Remote Probe Connector Pinouts
Due to power supply limits, the maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
Teledyne Analytical Instruments
3-9
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 3000PA
Figure 3-7: FET Series Resistance
3.4 Installing the Micro-Fuel Cell
The Micro-Fuel Cell is not installed in the cell block when the
instrument is shipped. It must be installed before the analyzer is placed in
service.
Once it is expended, or if the cell is exposed to air for too long, the
Micro-Fuel Cell will need to be replaced. The cell could also require replace-
ment if the instrument has been idle for too long.
When the micro-Fuel Cell needs to be installed or replaced, follow the
procedures in chapter 5, Maintenance, for removing and installing cells.
3.5 Testing the System
Before plugging the instrument into the power source:
•
•
•
Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors.
Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 4, section
4.3.5.
Teledyne Analytical Instruments
3-10
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
Operation
4.1 Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
•
Setsystemparameters:
•
Establish a security password, if desired, requiring Operator
to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
•
Calibratetheinstrument.
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
•
Set alarm setpoints, and modes of alarm operation (latching,
failsafe,etc).
Before you configure your 3000PA these default values are in effect:
Ranges: LO = 1 %, MED = 5 %, HI = 10 %
Auto Ranging: ON
Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching
Zero: Auto, every 0 days at 0 hours
Span: Auto, at 20.9 %, every 0 days at 0 hours
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
Teledyne Analytical Instruments
4-1
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
4.2 Using the Data Entry and Function
Buttons
Data Entry Buttons: The < > arrow buttons select options from the
menu currently displayed on the VFD screen. The selected option blinks.
When the selected option includes a modifiable item, the ∆∇ arrow
buttons can be used to increment or decrement that modifiable item.
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that
are not yet accepted by use of the Enter button.
Figure 4-1 shows the hierarchy of functions available to the operator via
the function buttons. The six function buttons on the analyzer are:
• Analyze. This is the normal operating mode. The analyzer
monitors the oxygen content of the sample, displays the percent
of oxygen, and warns of any alarm conditions.
• System. The system function consists of six subfunctions that
regulate the internal operations of the analyzer:
•
•
•
•
•
•
•
Set LCD screen contrast
Setup Auto-Cal
Contrast Function is DISABLED
(Refer to Section 1.6)
Assign Password
InitiateSelf-Test
Check software version
View sensor output
Log out.
• Zero. Usedtosetupazerocalibration.
• Span. Used to set up a span calibration.
• Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/orfailsafe.
• Range. Used to set up three analysis ranges that can be switched
automatically with auto-ranging or used as individual fixed
ranges.
Any function can be selected at any time by pressing the appropriate
button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the following
procedures. The VFD screen text that accompanies each operation is repro-
Teledyne Analytical Instruments
4-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
Contrast Function is DISABLED
(Refer to Section 1.6)
Figure 4-1: Hierarchy of Functions and Subfunctions
duced, at the appropriate point in the procedure, in a Monospaced type
style. Pushbutton names are printed in Oblique type.
4.3 The System Function
The subfuctions of the System function are described below. Specific
procedures for their use follow the descriptions:
• Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.
• PSWD: Security can be established by choosing a 5 digit
password (PSWD) from the standard ASCII character set. (See
Installing or Changing a Password, below, for a table of ASCII
characters available.) Once a unique password is assigned and
activated, the operator MUST enter the UNIQUE password to
gain access to set-up functions which alter the instrument's
Teledyne Analytical Instruments
4-3
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
operation, such as setting the instrument span or zero setting,
adjusting the alarm setpoints, or defining analysis ranges.
After a password is assigned, the operator must log out to
activate it. Until then, anyone can continue to operate the
instrument without entering the new password.
Only one password can be defined. Before a unique password
is assigned, the system assigns TBEAI by default. This allows
access to anyone. After a unique password is assigned, to defeat
the security, the password must be changed back to TBEAI.
• Logout: Logging out prevents an unauthorized tampering with
analyzersettings.
• More: Select and enter More to get a new screen with additional
subfunctionslisted.
• Self–Test: The instrument performs a self-diagnostic test to
check the integrity of the power supply, output boards and
amplifiers.
• Version: Displays Manufacturer, Model, and Software Version
ofinstrument.
4.3.1 Setting the Display
Contrast Function is DISABLED
(Refer to Section 1.6)
Ifyoucannotreadanythingonthedisplayafterfirstpoweringup:
1. ObserveLEDreadout.
a. If LED meter reads all eights and periods, go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Press
button twice to turn Analyzer OFF and ON again.
LED meter should now read all eights and periods. Go to step 3.
Teledyne Analytical Instruments
4-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see chapter 3, installa-
tion), the Analyzer can cycle itself through a sequence of steps that automati-
cally zero and span the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000PA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose System from the Function buttons. The VFD will display five
subfunctions.
Contrast Auto—Cal
Contrast Function is DISABLED
(Refer to Section 1.6)
PSWD Logout More
Use < > arrows to blink Auto—Cal, and press Enter. A new screen for
Span/Zerosetappears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press < > arrows to blink Span (or Zero), then press Enter again. (You
won’t be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every...) screen appears.
Span Every 0 d
Start 0 h from now
Use ∆∇ arrows to set an interval value, then use < > arrows to move to
the start-time value. Use ∆∇ arrows to set a start-time value.
To turn ON the Span and/or Zero cycles (to activate Auto-Cal): Press
System again, choose Auto—Cal, and press Enter again. When the Span/
Zero screen appears, use the < > arrows to blink the Span (or Zero) OFF/
ON field. Use ∆∇ arrows to set the OFF/ON field to ON. You can now turn
these fields ON because there is a nonzero span interval defined.
4.3.3 Password Protection
If a password is assigned, then setting the following system parameters
can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.
Teledyne Analytical Instruments
4-5
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
If you have decided not to employ password security, use the default
password TBEAI. This password will be displayed automatically by the
microprocessor. The operator just presses the Enter key to be allowed total
accesstotheinstrument’sfeatures.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4.3.3.1
EnteringthePassword
To install a new password or change a previously installed password,
you must key in and ENTER the old password first. If the default password
is in effect, pressing the ENTER button will enter the default TBEAI
password for you.
Press System to enter the System mode.
Contrast Auto—Cal
PSWD Logout More
Contrast Function is DISABLED
(Refer to Section 1.6)
Use the < > arrow keys to scroll the blinking over to PSWD, and press
Enter to select the password function. Either the default TBEAI password or
AAAAA place holders for an existing password will appear on screen
depending on whether or not a password has been previously installed.
T B E A I
Enter PWD
or
A A A A A
Enter PWD
The screen prompts you to enter the current password. If you are not
using password protection, press Enter to accept TBEAI as the default
password. If a password has been previously installed, enter the password
using the < > arrow keys to scroll back and forth between letters, and the ∆∇
arrow keys to change the letters to the proper password. Press Enter to enter
the password.
If the password is accepted, the screen will indicate that the password
restrictions have been removed and you have clearance to proceed.
PSWD Restrictions
Removed
In a few seconds, you will be given the opportunity to change this
password or keep it and go on.
Teledyne Analytical Instruments
4-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
Change Password?
<ENT>=Yes <ESC>=No
Press Escape to move on, or proceed as in Changing the Password,
below.
4.3.3.2
InstallingorChangingthePassword
If you want to install a password, or change an existing password,
proceed as above in Entering the Password. When you are given the oppor-
tunity to change the password:
Change Password?
<ENT>=Yes <ESC>=No
Press Enter to change the password (either the default TBEAI or the
previously assigned password), or press Escape to keep the existing pass-
word and move on.
If you chose Enter to change the password, the password assignment
screenappears.
T B E A I
<ENT> To Proceed
or
A A A A A
<ENT> To Proceed
Enter the password using the < > arrow keys to move back and forth
between the existing password letters, and the ∆∇ arrow keys to change the
letters to the new password. The full set of 94 characters available for pass-
word use are shown in the table below.
Characters Available for Password Definition:
A
K
U
_
i
s
}
)
B
L
V
`
j
t
→
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
E
O
Y
c
m
w
#
F
P
Z
d
n
x
$
.
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
-
7
3
=
6
@
8
When you have finished typing the new password, press Enter. A
verification screen appears. The screen will prompt you to retype your
passwordforverification.
\
Teledyne Analytical Instruments
4-7
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
A A A A A
Retype PWD To Verify
Wait a moment for the entry screen. You will be given clearance to
proceed.
A A A A A
<ENT> TO Proceed
Use the arrow keys to retype your password and press Enter when
finished. Your password will be stored in the microprocessor and the system
will immediately switch to the Analyze screen, and you now have access to
allinstrumentfunctions.
If all alarms are defeated, the Analyze screen appears as:
0.0
% Anlz
Range: 0 — 10
If an alarm is tripped, the second line will change to show which alarm
itis:
0.0
% Anlz
AL—1
Note: If you log off the system using the logout function in the
system menu, you will now be required to re-enter the
password to gain access to Span, Zero, Alarm, and Range
functions.
4.3.4 Logout
The Logout function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the instrument
off. By entering Logout, you effectively log off the instrument leaving the
system protected against use until the password is reentered. To log out,
press the System button to enter the System function.
Contrast Function is DISABLED
Contrast Auto—Cal
PSWD Logout More
(Refer to Section 1.6)
Use the < > arrow keys to position the blinking over the Logout func-
tion, and press Enter to Log out. The screen will display the message:
Protected Until
Password Reentered
Teledyne Analytical Instruments
4-8
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
4.3.5 System Self-Diagnostic Test
The Model 3000PA has a built-in self-diagnostic testing routine. Pre-
programmed signals are sent through the power supply, output board and
sensor circuit. The return signal is analyzed, and at the end of the test the
status of each function is displayed on the screen, either as OK or as a
number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation:
Press the System button to start the System function.
Contrast Function is DISABLED Contrast Auto—Cal
(Refer to Section 1.6)
PSWD Logout More
Use the < > arrow keys to blink More, then press Enter.
Version Self—Test
Cell Output: ### µA
Use the < > arrow keys again to move the blinking to the Self–Test
function. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right corner of
the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The results
screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.3.6 Version Screen
Move the < > arrow key to More and press Enter. With Version
blinking, press Enter. The screen displays the manufacturer, model, and
softwareversioninformation.
Teledyne Analytical Instruments
4-9
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
4.4 The Span Functions
The analyzer is calibrated using span gas.
NOTE: Zero is not necessary for Percent (%) level measurements.
Additional information on Zero functions is provided in the
Appendix A-6 of this manual.
Although the instrument can be spanned using air, a span gas with a
known oxygen concentration in the range of 70–90% of full scale of the
range of interest is recommended. Since the oxygen concentration in air is
20.9 %, the cell can take longer to recover if the instrument is used for less
than 1 % oxygen analysis immediately following calibration in air.
Connect the calibration gases to the analyzer according to the instruc-
tions given in Section 3.4.1, Gas Connections, observing all the prescribed
precautions.
Shut off the gas pressure before connecting it to the analyzer, and
be sure to limit the pressure to 40 psig or less when turning it back on.
Readjust the gas pressure into the analyzer until the flowrate (as read on
the analyzer’s SLPM flowmeter) settles between 0.5 and 2.4 SLPM (ap-
proximately1-5scfh).
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the instructions
in section 4.3.3 to enter your password. Once you have gained clearance to
proceed, you can enter the Zero or Span function.
4.4.1.
CellFailure
When the sensor in the 3000PA begins to fail, the analyzer will usually
require more and more frequent calibration. If the 3000PA analysis readings
driftdownwarduncharacteristically,tryrecalibration.Ifrecalibrationraises
the readings temporarily, the cell may be failing.
You can check the output of the cell itself by going to the System
function, selecting More, and pressing Enter. The cell output reading will be
on the second line of the display.
Version Self—Test
Cell Output: ### µA
The “good” reading depends on the class of cell your analyzer is using.
Teledyne Analytical Instruments
4-10
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
Although the B-1 cell is standard in the 3000PA, check Specific Model
Information in the Front Matter in this manual for the class of cell you
purchased. Then check Cell Replacement in chapter 5 Maintenance, and do
theprescribedcalculations.
If a weak cell is indicated, replace the cell as described there in chapter
5.
4.4.2 Span Cal
The Span button on the front panel is used to span calibrate the ana-
lyzer. Span calibration can be performed using the automatic mode, where
an internal algorithm compares consecutive readings from the sensor to
determine when the output matches the span gas concentration. Span cali-
bration can also be performed manual mode, where the operator determines
when the span concentration reading is acceptable and then manually exits
thefunction.
4.4.2.1
AutoModeSpanning
Press Span to enter the span function. The screen that appears allows
you to select whether the span calibration is to be performed automatically or
manually. Use the ∆∇ arrow keys to toggle between AUTO and MAN span
settling. Stop when AUTO appears, blinking, on the display.
Span: Settling: AUTO
<ENT> For Next
Press Enter to move to the next screen.
Span Val: 20.90
<ENT>Span <UP>Mod #
Use the ∆∇ arrow keys to enter the oxygen-concentration mode. Use
the < > arrow keys to blink the digit you are going to modify. Use the ∆∇
arrow keys again to change the value of the selected digit. When you have
finished typing in the concentration of the span gas you are using (20.90 if
youareusingair), press EntertobegintheSpancalibration.
####
Slope=####
%
Span
ppm/s
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates informa-
tion on Slope. Spanning automatically ends when the slope is less than 1/50
Teledyne Analytical Instruments
4-11
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
of the displayed value of the oxygen concentration (in ppm) for three min-
utes. Then the instrument automatically returns to the analyze mode.
4.4.2.2
ManualModeSpanning
Press Span to start the Span function. The screen that appears allows
you to select whether the span calibration is to be performed automatically or
manually.
Span: Settling:MAN
<ENT> For Next
Use the ∆∇ keys to toggle between AUTO and MAN span settling.
Stop when MAN appears, blinking, on the display. Press Enter to move to
the next screen.
Span Val: 20.90
<ENT>Span <UP>Mod #
Press ∆ (<UP>) to permit modification (Mod #) of span value.
Use the arrow keys to enter the oxygen concentration of the span gas
you are using (20.90 if you are using air). The < > arrows chose the digit,
and the ∆∇ arrows choose the value of the digit.
Press Enter to enter the span value into the system and begin the span
calibration.
Once the span has begun, the microprocessor samples the output at a
predeterminedrate. Itcalculatesthedifferencebetweensuccessivesamplings
and displays this difference as Slope on the screen. It takes several seconds
for the first Slope value to display. Slope indicates rate of change of the
Span reading. It is a sensitive indicator of stability.
####
% Span
Slope=#### ppm/s
When the Span value displayed on the screen is sufficiently stable,
press Enter. (Generally, when the Span reading changes by 1 % or less of
the full scale of the range being calibrated for a period of five minutes it is
sufficiently stable.) Once Enter is pressed, the Span reading changes to the
correct value. The instrument then automatically enters the Analyze func-
tion.
4.5 The Alarms Function
The Model 3000PA is equipped with 2 fully adjustable concentration
alarms and a system failure alarm. Each alarm has a relay with a set of form
Teledyne Analytical Instruments
4-12
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
“C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in
Chapter 3, Installation and/or the Interconnection Diagram included at the
back of this manual for relay terminal connections.
The system failure alarm has a fixed configuration described in chapter
3Installation.
The concentration alarms can be configured from the front panel as
either high or low alarms by the operator. The alarm modes can be set as
latching or non-latching, and either failsafe or non-failsafe, or, they can be
defeated altogether. The setpoints for the alarms are also established using
thisfunction.
Decide how your alarms should be configured. The choice will depend
upon your process. Consider the following four points:
1. Which if any of the alarms are to be high alarms and which if any
are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen
concentration rises above the setpoint. Setting an alarm as LOW
triggers the alarm when the oxygen concentration falls below the
setpoint.
Decide whether you want the alarms to be set as:
•
•
•
Both high (high and high-high) alarms, or
One high and one low alarm, or
Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in an
alarm condition. You can set either or both of the concentration
alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to non-
alarmconditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
Teledyne Analytical Instruments
4-13
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter the
Alarm function.
Press the Alarm button on the front panel to enter the Alarm function.
Make sure that AL–1 is blinking.
AL—1
AL—2
Choose Alarm
Set up alarm 1 by moving the blinking over to AL–1 using the < >
arrow keys. Then press Enter to move to the next screen.
AL—1 1.00 % HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen:
•
•
•
•
•
Value of the alarm setpoint, AL–1 #### % (oxygen)
Out-of-range direction, HI or LO
Defeated? Dft–Y/N (Yes/No)
Failsafe? Fs–Y/N (Yes/No)
Latching? Ltch–Y/N (Yes/No).
•
•
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the ∆∇ arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, the setpoint units are
percent-of-oxygen.)
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the ∆∇ arrow
keys to change the parameter.
•
•
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL–2).
To reset a latched alarm, go to Dft– and then press either ∆ two
times or ∇ two times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch– and then press either ∆ two times or ∇ two times.
(Toggle it to N and back to Y.)
Teledyne Analytical Instruments
4-14
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
4.6 The Range Function
The Range function allows the operator to program up to three concen-
tration ranges to correlate with the DC analog outputs. If no custom ranges
are defined by the user, the instrument defaults to:
Low = 0–1.00 %
Med = 0–5.00 %
High = 0–10.00 %.
The Model 3000PA is set at the factory to default to autoranging. In this
mode, themicroprocessorautomaticallyrespondstoconcentrationchanges
by switching ranges for optimum readout sensitivity. If the current range
limits are exceeded, the instrument will automatically shift to the next higher
range. If the concentration falls slightly below full scale of the next lower
range, the instrument will switch to that range. A corresponding shift in the
DC percent-of-range output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc.
However, the digital readout and the RS-232 output of the concentra-
tion are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in Chapter 1.
4.6.1 Setting the Analog Output Ranges
To set the ranges, enter the range function mode by pressing the
Range button on the front panel.
L—1.00 M—5.00
H—10.00 Mode—AUTO
Use the < > arrow keys to blink the range to be set: low (L), medium
(M), or high (H).
Use the ∆∇ arrow keys to enter the upper value of the range (all ranges
begin at 0 %). Repeat for each range you want to set. Press Enter to accept
the values and return to Analyze mode. (See note below.)
Note: The ranges must be increasing from low to high. For example,
if range 1 is set as 0–1 % and range 2 is set as 0–10 %, range 3
cannot be set as 0–5 % since it is lower than range 2.
Teledyne Analytical Instruments
4-15
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
4.6.2 Autoranging Analysis
Set your analysis ranges as in 4.6.1, above. Leave Mode in Auto, or
use the arrow buttons to change back to Auto.
When operating in autoranging, if the oxygen concentration in your
sample goes ABOVE your HIGHEST range setting, the analyzer will go
into the special 25 % cal range.
However, if one of your range settings is below 0-25 % and another is
set above 0-25 %, the special 0-25 % Air Cal range will NOT activate as the
oxygen level goes through 25 %. Nevertheless, if the oxygen concentration
in your sample goes ABOVE your HIGHEST range setting, the analyzer
will THEN drop back down into the special 25 % cal range.
Once the oxygen concentration drops back down into your highest
range setting, the analyzer will automatically switch back to that range.
CAUTION: While the analyzer is in the Air Cal range, the oxygen
reading cannot go over 25 %, even if the oxygen
concentration is higher than 25 %.
4.6.3 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing the
analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the Range button on the front panel.
Use the < > arrow keys to move the blinking over AUTO.
Use the ∆∇ arrow keys to switch from AUTO to FX/LO, FX/MED, or
FX/HI to set the instrument on the desired fixed range (low, medium, or
high).
L—1.00 M—5.00
H—10.00 Mode—FX/LO
or
L—1.00 M—5.00
H—10.00 Mode—FX/MED
or
L—1.00 M—5.00
H—10.00 Mode—FX/HI
Teledyne Analytical Instruments
4-16
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Operation 4
Press Escape to re-enter the Analyze mode using the fixed range.
NOTE:When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc. However, the digital readout and
the RS-232 output continue to read the true value of the oxy-
gen concentration regardless of the analog output range.
4.7 The Analyze Function
Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. Pressing the
Escape button in many cases also switches the analyzer back to the Ana-
lyze function. Alternatively, you can press the Analyze button at any time
to return to analyzing your sample.
4.8 Signal Output
The standard Model 3000PA Percent Oxygen Analyzer is equipped
with two 0–1 V dc analog output terminals accessible on the back panel (one
concentration and one range ID). The –MA option also has two isolated 4–
20 mA dc current outputs (one concentration and one range ID).
See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on range that was defined
as 0–10 % O2, then the output would be:
Voltage Signal Current Signal
% O2
Output (V dc)
Output (mA dc)
0
1
2
3
4
5
6
7
8
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
9
10
Teledyne Analytical Instruments
4-17
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 3000PA
The analog output signal has a voltage which depends on the oxygen
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autorangingmode.
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current if you
have current outputs) to represent a particular range. The following table
gives the range ID output for each analysis range:
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
0.75
1.00
12
16
20
HI
CAL (0-25%)
Teledyne Analytical Instruments
4-18
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Maintenance 5
Maintenance
5.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing Micro-Fuel cells and
fuses, and recalibration. For recalibration, see Section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
5.2 Cell Replacement
The Micro-Fuel Cell is a sealed electrochemical transducer with no
electrolyte to change or electrodes to clean. When the cell reaches the end
of its useful life, it is replaced. The spent fuel cell should be discarded
according to local regulations. This section describes fuel cell care as well
as when and how to replace it.
5.2.1 Storing and Handling Replacement Cells
To have a replacement cell available when it is needed, TAI recom-
mends that one spare cell be purchased 4-5 months after commissioning
the 3000PA, or shortly before the end of the cell warranty period.
CAUTION: Do not stockpile cells. The warranty period starts
on the day of shipment.
The spare cell should be carefully stored in an area that is not subject
to large variations in ambient temperature (75 °F nominal) or to rough
handling.
WARNING: THE SENSOR USED IN THE MODEL 3000PA PER-
CENT OXYGEN ANALYZER USES ELECTROLYTES
WHICH CONTAIN TOXIC SUBSTANCES, MAINLY
Teledyne Analytical Instruments
5-1
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 3000P
LEAD AND POTASSIUM HYDROXIDE, THAT CAN BE
HARMFUL IF TOUCHED, SWALLOWED, OR IN-
HALED. AVOID CONTACT WITH ANY FLUID OR
POWDER IN OR AROUND THE UNIT. WHAT MAY
APPEAR TO BE PLAIN WATER COULD CONTAIN
THESE TOXIC SUBSTANCES. IN CASE OF EYE
CONTACT, IMMEDIATELY FLUSH EYES WITH WA-
TER FOR AT LEAST 15 MINUTES. CALL PHYSICIAN.
(SEE APPENDIX, MATERIAL SAFETY DATA SHEET.)
CAUTION: Do not disturb the integrity of the cell package until
the cell is to actually be used. If the cell package is
punctured and air is permitted to enter, the cell will
require a longer time to reach zero after
installation.
5.2.2 When to Replace a Cell
When the sensor in the 3000PA begins to fail, the analyzer usually
requires more frequent calibration. If the 3000PA analysis readings drift
downward uncharacteristically, try recalibration. If recalibration raises the
readings temporarily, suspect the cell, but first check for leaks downstream
from the cell where gases may be leaking into the system.
You can check the output of the cell itself by going to the System
function, selecting More , and pressing Enter. The cell output reading will
be on the second line of the display.
Version
Self—Test
Cell Output: ### µA
The “good” cell output range depends on the class of cell your ana-
lyzer is using. The B-1 cell is standard in the 3000PA, but others can be
specified.
Check Specific Model Information in the Front Matter in this
manual for the class of cell you purchased. Then check Table 5-1, the
cell index table below, and do the simple calculation. If the resulting value
isbelowtheCe llOutputreading,replacethecell.
To find out if your cell is too weak:
1. Flow span gas through the analyzer, and allow time to purge.
2. With span gas flowing, read the raw output of the cell from the
System function display.
Teledyne Analytical Instruments
5-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Maintenance 5
3. Divide the raw output reading by the percent oxygen
concentration of your span gas.
If the quotient is less than the Index value for the cell class you are
using, replace the cell.
Table 5-1: Cell Indices
Cell Class
A-3
Index
1.818
1.818
4.545
3.716
1.244
1.515
2.488
0.606
A-5
B-1
B-3
B-5
B-7
C-3
C-5
5.2.3 Removing the Micro-Fuel Cell
The Micro-Fuel cell is located inside the nylon Probe behind the front
panel. (See Figure 5-1.) To remove an existing cell:
1. Remove power to the instrument by unplugging the power cord
at the power source.
2. Open the front panel door by pressing the release button on the
top right corner of the door all the way in and releasing it.
3. Pull up on the nylon Probe, with a slight rocking motion, to
release it from the Probe Receptacle.
4. Do Not remove the O-rings unless they are worn and no longer
hold the Probe tightly. (If worn, replace them.)
5. When it is free, unscrew the Cap from the nylon Probe. Hold
the Probe vertically to prevent dropping the cell out of the
probe.
6. Remove the Cell from the Probe, and dispose of it in an
environmentally safe manner.
Teledyne Analytical Instruments
5-3
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 3000P
Figure 5-1: Removing or Installing a Micro-Fuel Cell
Teledyne Analytical Instruments
5-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Maintenance 5
5.2.4 Installing a New Micro-Fuel Cell
CAUTION: Do not touch the sensing surface of the cell. It is
covered with a delicate Teflon membrane that can
leak if punctured. The sensor must be replaced if
the membrane is damaged.
1. Place the Cell in the Probe with the sensing surface facing
outward (toward the screen in the Cap).
2. Screw the Probe Cap onto the Probe until it stops.
3. With the O-rings in place, push the assembled Probe down into
the Cell Holder—Cap Down—with a slight rocking motion
until it is seated on the bottom of the holder. This forces the
holder into position and forms a gas-tight seal.
5.2.5 Cell Warranty
The Class B-1 Micro-Fuel cell is standard in the Model 3000PA. This
cell is warranted for 6 months from the date of shipment. Check the Spe-
cific Model Information, and note any Addendum that might be attached to
the front of this manual for special information applying to the Cell in your
instrument.
The warranty period for spare cells begins on the date of shipment.
Do not purchase more than one spare cell per instrument. Do not stockpile
spare cells.
The B-1 cell is not designed for applications where CO2 is a major
component in the sample, however slight amounts will not adversely
effect the cell performance. Consult TAI for available options for either
intermittent or continuous CO2 exposure.
If a cell was working satisfactorily, but ceases to function before the
warranty period expires, the customer will receive credit toward the pur-
chase of a new cell.
If you have a warranty claim, you must return the cell in question to
the factory for evaluation. If it is determined that failure is due to faulty
workmanship or material, the cell will be replaced at no cost to you.
NOTE:Evidence of damage due to tampering or mishandling will
render the cell warranty null and void.
Teledyne Analytical Instruments
5-5
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 3000P
5.3 Fuse Replacement
1. Place small screwdriver in notch, and pry cover off, as shown in
Figure 5-2.
Figure 5-2: Removing Fuse Block from Housing
2. To change between American and European fuses, remove the
single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.
3. Replace fuse as shown in Figure 5-3.
4. Reassemble Housing as shown in Figure 5-2.
American Fuses
European Fuses
Figure 5-3: Installing Fuses
5.4 System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More , and press Enter.
3. Use the < > arrow keys to move to Se lf-Te st, and press Enter.
The following failure codes apply:
Teledyne Analytical Instruments
5-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Maintenance 5
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Preamp
0
1
2
3
OK
Zero too high
Amplifier output doesn't match test input
Both Failed
5.5 Major Internal Components
The Micro-Fuel cell is accessed by unlatching and swinging open the
front panel, as described earlier. Other internal components are accessed by
removing the rear panel and sliding out the entire chassis. See Figure 5-4,
below. The gas piping is illustrated in Figure 2-4, and the major electronic
components locations are shown in Figure 2-5, in chapter 2.
WARNINGS:See warnings on the title page of this manual.
The 3000PA contains the following major components:
•
Analysis Section
Micro Fuel Cell (B-1 standard—others available)
Nylon Probe and Holder
Sample System
•
•
•
Power Supply
Microprocessor
Displays
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
RS-232 Communications Port.
•
See the drawings in the Drawings section in back of this manual
for details.
Teledyne Analytical Instruments
5-7
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 3000P
X
X
X
X
X
X
X
X
Figure 5-4: Rear-Panel Screws
To detach the rear panel, remove only the eight screws marked with an X.
5.6 Cleaning
If instrument is unmounted at time of cleaning, disconnect the instru-
ment from the power source. Close and latch the front-panel access door.
Clean outside surfaces with a soft cloth dampened slightly with plain clean
water. Do not use any harsh solvents such as paint thinner or benzene.
For panel-mounted instruments, clean the front panel as prescribed in
the above paragraph. DO NOT wipe front panel while the instrument is
controlling your process.
Teledyne Analytical Instruments
5-8
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Maintenance 5
5.7 Troubleshooting
Problem:
Erratic readings of the Oxygen concentration as reported by the analyzer.
Possible Cause:
The analyzer may have been calibrated in an inaccurate fashion.
Solution:
Turn the analyzer off, then back on again. Press the System key when
prompted by the analyzer "Press System for default Values". This will
return the analyzer to its default settings in calibration and zero values. If
erratic behavior continues replace the sensor.
Possible Cause:
Atmospheric Oxygen may be diffusing in through the vent and affecting
the oxygen level which the sensor sees.
Solution:
Increase flow rate and/or length or vent tubing in order to dilute of mini-
mize the diffusion of oxygen from the vent back to the sensor.
Problem:
Inaccurate zero operation (i.e. the user has zeroed the analyzer accidentally
on gas much higher than one would normally use for a zero gas).
Solution:
Turn the analyzer off, then back on again. Press the System key when
prompted by the analyzer "Press System for default Values". This will
return the analyzer to its default settings in calibration and zero values.
Now proceed to carefully calibrate and zero the analyzer.
Teledyne Analytical Instruments
5-9
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 3000P
Teledyne Analytical Instruments
5-10
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Appendix
Appendix
A-1 Model 3000PA Specifications
Packaging: General Purpose
• Flush panel mount (Standard).
• Rack mount — Relay rack mounted to
contain either one or two instruments in one
19" relay rack mountable plate (Optional).
Sensor: Class B-1 Micro-Fuel Cell (standard). Others
available.
Cell Block: Nylon.
90 % Response Time: 10 seconds at 25 °C (77 °F).
Ranges: Three user definable ranges from 0-1 % to
0-100 %, plus air calibration range of 0-25 %.
Alarms: One system-failure alarm contact to detect
powerfailure.
Twoadjustableconcentrationthresholdalarms
withfullyprogrammablesetpoints.
Displays: 2 line by 20 character, alphanumeric, VFD
screen.
One 5 digit LED display.
DigitalInterface: Full duplex RS-232 communications port.
Power: Universal power supply 85-250 V ac, at
47-63 Hz.
OperatingTemperature: 0-50 °C (32-122 °F).
Teledyne Analytical Instruments
A-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 3000PA
Accuracy: ±2% of full scale at constant temperature.
±5% of full scale over operating temperature
range, on factory default analysis ranges, once
thermal equilibrium has been reached.
Analog outputs: 0-1 V dcpercent-of-range(Standard)
0-1 V dc range ID (Standard)
4-20 mA dc—isolated—percent-of-range
(Optional)
4-20 mA dc—isolated— range ID (Optional)
Dimensions: 19 cm high × 24.9 cm wide × 31 cm deep
(5.96″ × 8.7″ × 12.2″).
Teledyne Analytical Instruments
A-2
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Appendix
A-2 Recommended 2-Year Spare Parts List
Qty PartNumber Description
1
C62374
C62371-B
C62368-B
C62365-C
C62365-A
F9
Back Panel Board
1
Front Panel Board
1
PercentPreamplifierBoard
1*
1*
3
Main Computer Board (std)
Main Computer Board (4-20 mA)
Fuse, 1 A, 250 V, 3 AG, Slow Blow
Fuse, 1 A, 250 V, 5 × 20 mm, Slow Blow
Molex Connector for Remote Probe
3
F1275
1
R1460
1
T976
Molex Crimp Terminals for Remote Probe Connector
O-ring
2
O38
1** C6689-B1
Micro-FuelCell(foroptionssee**)
__________________
* Order -B or -C, not both.
** Check Specific Model Information for cell in your instrument.
A minimum charge of $20.00 US dollars is applicable to spare parts orders.
NOTE: Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 934-1500, Fax (626) 961-2538
TWX (910) 584-1887 TDYANYL COID
Web:
or your local representative.
Teledyne Analytical Instruments
A-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 3000PA
A-3 Drawing List
D-64573 FinalAssembly/OutlineDrawing
A-4 19-inch Relay Rack Panel Mount
Figure A-1: Single and Dual 19" Rack Mounts
Teledyne Analytical Instruments
A-4
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Appendix
3000 SERIES ANALYZERS
APPLICATIONNOTESONRESTRICTORS,
A-5
PRESSURES, AND FLOW RECOMMENDATIONS
3000seriesanalyzersrequirereasonablyregulatedsamplepressures.
Whilethe3000analyzersarenotsensitivetovariationsofincomingpressure
(providedtheyareproperlyventedtoatmosphericpressure)Thepressuremust
bemaintainedastoprovideauseableflowratetroughtheanalyzer.Anyline
attachedtosampleventshouldbe1/4orlarger indiameter.
FLOWRATERECOMMENDATIONS:
Ausableflowratefora3000seriesanalyzerisonewhichcanbe
measuredontheflowmeter.Thisisbasically .2-2.4SLPM.Theoptimumflow
rateis1SLPM (midscale). Note:responsetimeisdependentonflowrate, a
lowflowratewillresultinslowresponsetoO2changesinthesamplestream.
Thespanflowrateshouldbetheapproximatelysameasthesampleflowrate.
CELLPRESSURECONCERNS:
Thesensorsusedin3000seriesanalyzersareoptimized tofunctionat
atmosphericpressure.Atpressuresotherthanatmosphericthediffusionrateof
O2willbedifferentthanoptimumvalue.HigherpressureswillproducefasterO2
diffusionrates resultinginhigherO2reading andshortercelllife.Tousea3000
seriesanalyzeratacellpressureotherthanatmospheric,theanalyzermustbe
calibratedwithaknowncalibrationgasatthenewcellpressuretoadjustforthe
differentdiffusionrate.Cellpressuresbelow2/3atmosphericarenot
recommendedbecauseastheytendtocauseexcessiveinternalexpansionwhich
mayresultinsealfailure.
Foroperationatcellpressuresotherthanatmosphericcaremustbe
takennottochangethesamplepressurerapidlyorcelldamagemayoccur.For
cellpressuresaboveatmospheric,cautionmustbeexercisedtoavoidover
pressuringthecellholder.(percentanalyzerswillrequiresometypeofcell
retainertopreventthecellfrombeingpushedoutbythepressure.) For
operationatpressuresbelowatmosphericpressureasuffixC(clamped)cellis
required.
RESTRICTIONDEVICES:
Forproperoperation,all3000seriesanalyzersrequireaflowrestriction
device.Thisdeviceistypicallyarestrictororavalve.Thisrestrictiondevice
servestwofunctionsinthesamplepath.Thefirstfunctionistolimittheflowrate
ofthesamplethroughtheanalyzer.Arestrictorischosentooperateoverarange
ofpressuresandprovideauseableflowrateoverthatrange.
Thesecond functionthattherestrictiondeviceprovidesisapressure
drop.Thisdeviceisselectedtoprovidetheonlysignificant pressuredropinthe
samplepath.
Teledyne Analytical Instruments
A-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 3000PA
RESTRICTORKIT
Thecurrentrevisionofthe3000seriesanalyzersaresuppliedwithakit
containingtworestrictorsandaunionwhichareuserinstalled.Theseparts
suppliedtogivetheendusermoreflexibilitywheninstallingtheanalyzer.The
restrictorkitissuitableforhighandlowpositivepressureapplicationsaswellas
vacuumservice(atmosphericpressuresample)applications(seemanualfor
installationinstructions).Thestandardrestrictor(BLUEDOT)isrecommended
for pressures between 5 PSIG and 50 PSIG. For positive low pressure
application ( 5 psig or less ) the un-marked restrictor is better suited . For
nonepressurizedsampleapplicationsthemarkedrestrictorshouldbeusedand
configuredforvacuumservice.Note:forextremelylowpositivepressure
applications(lessthen2psig)thevacuumserviceconfigurationshouldprovide
higherperformance(higherflowrates). Forvacuumservicetheendusermust
supplyavacuumpumpandaby-passvalveforthepump.Avacuumlevelof 5-
10inchesofmercuryshould providetheoptimumflowrate.CAUTION:flow
restrictors have very small orifices and may be plugged by small
particles ( .005” dia or larger) A sample filter must be included in the
samplelinepriortotherestrictor!(a60micronfilterisrecommended)
3000PAEXAMPLES:
Example1,withaincoming pressureof10psigthestdrestrictor(blue
dot) willprovideaflowrateof.76SLPM.Up-streamoftherestrictorthe
samplelinepressurewillbe10psig, whiledownstream(includingthecell)the
pressurewillbeatatmosphericpressure.(analyzerventedtoatmospheric
pressure) Note,allotherpressuredropsinthesamplepathareinsignificantat
theseflowrates.Thisinsuresthatthecelloperatesatatmosphericpressure.At
veryhighflowrates(offscaleofflow-meter),pressuredropsotherthanthe
restrictiondevicecouldbecomesignificant,andresultin pressurizingthecell.
Example2, A3000PAisconfiguredforvacuumserviceasfollows.The
un-markedrestrictorisplacedinthesampleventport.Thedownstreamendof
therestrictoristhenconnectedtoavacuumpumpandby-passvalve.Theby-
passvalveisadjustedtoprovideaflowrateof1SLPM.Thesamplepressure
betweenthepumpandtherestrictorwillbeapproximately -7inchesofmercury,
whilethepressureinthebalanceofthesamplesystemincludingthecellwillbe
approximatelyatatmosphericpressure.(providedthesampleflowintothe
analyzerisnotblocked.)
BY-PASS:
Toimprovethesystemresponse, aby-passcanbeaddedtoincrease
thesampleflowratetotheanalyzerbyafactoroften.Aby-passprovidesa
sampleflowpatharoundtheanalyzerof 2-18SCFH.typically.
CALIBRATION GAS:
3000seriesanalyzerrequirementsforunitswithAuto-Caloptions.The
Teledyne Analytical Instruments
A-6
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Appendix
customermustsupplyacontrolvalves(orrestrictors) foranySPANor ZEROgas
sourcewhichisattachedtotheAuto-Calports.Thevalveshouldbeadjustedtothe
sameflowrateasthesamplegas.Whenrestrictorsareused,thegaspressuremust
beadjustedtoachievetheproperflowrate.
OPERATIONWITHOUT ARESTRICTORDEVICE:
Operationwithoutarestrictordeviceisnotrecommendasmentioned
above. A3000PA withoutanyflowrestrictordevicewastestedon11-19-97.
This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the
standardsamplesys.
REFERENCE: FLOW_1.XLS & FLOW_2.XLS forinformationonflowrates
atvariouspressures.
TAI PART NUMBERS
RESTRICTORKIT:
UNION (SS)
LP.RESTRICTOR
STD..RESTRICTOR
NUT
A68729
U11
R2323 ( LOW PRESSURE / VAC. SERVICE )
R2324
N73
F73
BLUEDOT
FERRULE
FERRULE
F74
BOTHFERRULESARE
REQUIRED
CONVERSIONS:
1 PSI
1 SCFH
=
=
2.04 INCHES OF MERCURY (in. Hg.)
0.476 SLPM
Teledyne Analytical Instruments
A-7
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 3000PA
A-6
Zero Cal
The Zero button on the front panel is used to enter the zero calibration
function. Zero calibration can be performed in either the automatic or manual
mode. In the automatic mode, an internal algorithm compares consecutive
readings from the sensor to determine when the output is within the accept-
able range for zero. In the manual mode, the operator determines when the
reading is within the acceptable range for zero. Make sure the zero gas is
connected to the instrument. If zeroing is becoming more and more dificult,
skip to section 4.4.1.3 Cell Failure.
Auto Mode Zeroing
Press Zero to enter the zero function mode. The screen allows you to
select whether the zero calibration is to be performed automatically or manu-
ally. Use the ∆∇ arrow keys to toggle between AUTO and MAN zero
settling. Stop when AUTO appears, blinking, on the display.
Zero:Settling:AUTO
<ENT>ToBegin
Press Enter to begin zeroing.
#### % Zero
Slope=#### ppm/s
The beginning zero level is shown in the upper left corner of the dis-
play. As the zero reading settles, the screen displays and updates information
on Slope (unless the Slope starts within the acceptable zero range and does
not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown: 5 Le ft, 4 Le ft, and so fourth.
These are five steps in the zeroing process that the system must complete,
AFTER settling, before it can go back to Analyze.
#### % Zero
4Left=### ppm/s
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically
returns to the Analyze mode.
Manual Mode Zeroing
Press Zero to enter the Zero function. The screen that appears allows
you to select between automatic or manual zero calibration. Use the ∆∇ keys
Teledyne Analytical Instruments
A-8
Download from Www.Somanuals.com. All Manuals Search And Download.
Percent Oxygen Analyzer
Appendix
to toggle between AUTO and MAN zero settling. Stop when MAN appears,
blinking, on the display.
Zero:Settling:Man
<ENT>ToBegin
Press Enter to begin the zero calibration. After a few seconds the first
of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as Slope = a value in parts per million per second
(ppm/s).
#### % Zero
Slope=#### ppm/s
NOTE: It takes several seconds for the true Slope value to display. Wait about
10 seconds. Then, wait until Slope is sufficiently close to zero before
pressing Enter to finish zeroing .
Generally, you have a good zero when Slope is less than 0.05 ppm/s
for about 30 seconds. When Slope is close enough to zero, press Enter. In a
few seconds, the screen will update.
Once span settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the Analyze
mode.
NOTE: The MSDS on this material is available upon request
through the Teledyne Environmental, Health and
Safety Coordinator. Contact at (626) 934-1592
Teledyne Analytical Instruments
A-9
Download from Www.Somanuals.com. All Manuals Search And Download.
|