Teledyne Outdoor Gas Burner 7500 User Manual

TELEDYNE  
ANALYTICAL INSTRUMENTS  
OPERATING INSTRUCTIONS  
Series Model 7500  
Infrared Gas Analyzer  
DANGER  
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS  
MONITORING SYSTEM.  
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.  
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST  
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.  
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.  
BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED  
SUPERVISOR/MANAGER.  
P/N  
06/20/06  
ECO #  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
SAFETY PRECAUTIONS  
Please read this section carefully to ensure using the analyzer in the correct way.  
The cautionary descriptions listed here contain important information about safety, so they should  
always be observed. Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and  
“PROHIBITED”.  
DANGER  
Wrong handling may cause a dangerous situation,  
in which there is a risk of death or heavy injury.  
CAUTION  
Wrong handling may invite a dangerous situation,  
in which there is a possibility of medium-level  
trouble or slight injury or only physical damage is  
predictable.  
PROHIBITED  
Items which must not be done are noted.  
Caution on installation and transport of gas analyzer  
DANGER  
CAUTION  
• This unit is not explosion-proof type. Do not  
use it in a place with explosive gases to prevent  
explosion, fire or other serious accidents.  
• For installation, observe the rule on it given in the  
instruction manual and select a place where the  
weight of gas analyzer can be endured. Installation  
at an unsuited place may cause turnover or fall  
and there is a risk of injury.  
• For lifting the gas analyzer, be sure to wear  
protective gloves. Bare hands may invite an injury.  
• Before transport, fix the casing so that it will not  
open. Otherwise, the casing may be separated  
and fall to cause an injury.  
• During installation work, care should be taken to  
keep the unit free from cable chips or other foreign  
objects. Otherwise, it may cause fire, trouble or  
malfunction of the unit.  
3
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Caution on piping  
DANGER  
In piping, the following precautions should be  
observed. Wrong piping may cause gas leakage. If  
the leaking gas contains a toxic component, there  
is a risk of serious accident being induced. Also, if  
combustible gas is contained, there is a danger of  
explosion, fire or the like occurring.  
• Connect pipes correctly referring to the  
instruction manual.  
• Exhaust should be led outdoors so that it will not  
remain in the locker and installation room.  
• Exhaust from the analyzer should be relieved in  
the atmospheric air in order that an unnecessary  
pressure will not be applied to the analyzer.  
Otherwise, any pipe in the analyzer may be  
disconnected to cause gas leakage.  
• For piping, use a pipe and a pressure reducing  
valve to which oil and grease are not adhering.  
If such a material is adhering, a fire or the like  
accident may be caused.  
Caution on wiring  
CAUTION  
• Wiring work must be performed with the main  
power set to OFF to prevent electric shocks.  
• Enforce construction of class-D rounding wire by  
all means. If the specified grounding construction  
is neglected, a shock hazard or fault may be  
caused.  
• Wires should be the proper one meeting the  
ratings of this instrument. If using a wire which  
cannot endure the ratings, a fire may occur.  
• Be sure to use a power supply of correct rating.  
Connection of power supply of incorrect rating may  
cause fire.  
Caution on use  
DANGER  
CAUTION  
• For correct handling of calibration gas or other  
reference gases, carefully read their instruction  
manuals beforehand. Otherwise, carbon monoxide  
or other hazardous gases may cause an  
intoxication particularly.  
• Before leaving unused for a long time or  
restarting after left at such a status for an  
extended length of time, follow the directions of  
each instruction manual because they are different  
from normal starting or shutdown. Otherwise, the  
performance may be poor and accidents or injuries  
may be caused.  
• Do not operate the analyzer for a long time with  
its door left open. Otherwise, dust, foreign matter,  
etc. may stick on internal walls, thereby causing  
faults.  
4
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Caution on use  
PROHIBITED  
• Do not allow metal, finger or others to touch the  
input/output terminals in the instrument. Otherwise,  
shock hazard or injury may occur.  
• Do not smoke nor use a flame near the gas  
analyzer. Otherwise, a fire may be caused.  
• Do not allow water to go into the gas analyzer.  
Otherwise, hazard shock or fire in the instrument  
may be caused.  
Caution on maintenance and check  
DANGER  
CAUTION  
• When doors are open during maintenance or  
inspection, be sure to purge sufficiently the inside  
of the gas analyzer as well as the measuring  
gas line with nitrogen or air, in order to prevent  
poisoning, fire or explosion due to gas leak.  
Be sure to observe the following for safe operation  
avoiding the shock hazard and injury.  
• Remove the watch and other metallic objects  
before work.  
• Do not touch the instrument wet-handed.  
• If the fuse is blown, eliminate the cause, and then  
replace it with the one of the same capacity and  
type as before. Otherwise, shock hazard or fault  
may be caused.  
• Do not use a replacement part other than  
specified by the instrument maker. Otherwise,  
adequate performance will not be provided.  
Besides, an accident or fault may be caused.  
• Replacement parts such as a maintenance part  
should be disposed of as incombustibles. For  
details, follow the local ordinance.  
Others  
CAUTION  
• If the cause of any fault cannot be determined  
despite reference to the instruction manual, be  
sure to contact the factory. If the instrument is  
disassembled carelessly, you may have a shock  
hazard or injury.  
5
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
CONTENTS  
PREFACE .............................................................................................................................2  
CAUTION ON SAFETY ..........................................................................................................3  
1. OVERVIEW.........................................................................................................................8  
2. NAME AND DESCRIPTION OF EACH PART ...................................................................9  
2.1 Description of each unit ......................................................................................9  
3. INSTALLATION ................................................................................................................10  
3.1 Installation conditions ........................................................................................10  
3.2 Installation of analyzer .......................................................................................11  
3.3 Piping .................................................................................................................11  
3.4 Sampling ............................................................................................................12  
3.5 Wiring method ....................................................................................................15  
4. OPERATION ................................................................................................... ..................24  
4.1 Preparation for operation ....................................................................................24  
4.2 Warm-up operation and regular operation ..........................................................24  
5. DESCRIPTION OF DISPLAY AND OPERATION PANEL .................................................25  
5.1 Name and description of operation panel ...........................................................25  
5.2 Overview of display and operation panel ............................................................26  
5.3 Overview of display screen .................................................................................27  
5.4 General operation................................................................................................30  
6. SETTING AND CALIBRATION...........................................................................................31  
6.1 Changeover of range ..........................................................................................31  
6.2 Calibration setting ...............................................................................................32  
6.2.1 Setting of calibration concentration......................................................32  
6.2.2 Setting of manual zero calibration ...................................................... 34  
6.2.3 Setting of calibration range ..................................................................36  
6.2.4 Setting of auto-calibration component .................................................38  
6.3 Alarm setting ........................................................................................................40  
6.3.1 Setting of alarm values ........................................................................40  
6.3.2 Hysteresis setting ................................................................................42  
6.4 Setting of auto calibration ....................................................................................43  
6.4.1 Auto calibration ....................................................................................43  
6.4.2 Forced stop of auto calibration ............................................................45  
6.5 Setting of auto zero calibration ............................................................................47  
6.5.1 Auto zero calibration ............................................................................47  
6.5.2 Forced stop of auto zero calibration ....................................................49  
6.6 Peak alarm setting ...............................................................................................51  
6.7 Parameter setting ................................................................................................53  
6
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.8 Maintenance mode …………………………………………………………………… 57  
6.9 Calibration …………………………………………………………………………….. 59  
6.9.1 Zero calibration …………………………………………………………… 59  
6.9.2 Span calibration ………………………………………………………….. 60  
7. MAINTENANCE ................................................................................................... 61  
7.1 Daily check ............................................................................................ 61  
7.2 Daily check and maintenance procedures ............................................ 61  
7.3 Cleaning of measuring cell .................................................................... 62  
7.3.1 Disassembly and assembly of measuring cell ....................... 62  
7.3.2 How to clean cell ……………………………………………….. 67  
8. TROUBLESHOOTING ......................................................................................... 68  
8.1 Error message .......................................................................................... 68  
9. SPECIFICATIONS ............................................................................................... 70  
9.1 Specifications ………………………………………………………………. 70  
9.2 Code symbols ………………………………………………………………. 73  
9.3 Outline diagram …………………………………………………………….. 78  
APPENDIX ………………………………………………………………………………… 80  
Drawing List ............................................................................................................. 80  
7
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
1. OVERVIEW  
The Model 7500 infrared gas analyzer measures the concentrations of CO2, CO, CH4, SO2 and O2  
contained in sample gas. CO2, CO, CH4 and SO2 are measured by non-dispersion infrared method,  
while O2 is measured by paramagnetic method. A maximum of 4 components including O2 (a maximum  
of 3 components for other than O2 measurement) are simultaneously measurable.  
A high-sensitivity mass flow sensor is used in the detector unit of infrared method. Due to use of single  
beam system for measurement, maintenance is easy and an excellent stability is ensured for a long  
period of time.  
In addition, a microprocessor is built in and a large sized liquid crystal display is provided for easier  
operation, higher accuracy and more functions.  
This analyzer is thus optimum for combustion control of various industrial furnaces, botanical study and  
global atmospheric research.  
8
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
2. NAME AND DESCRIPTION OF EACH PART  
2.1 Description of each unit  
Name  
Description  
1. Handle  
Draws the analyzer unit from the case.  
Turns ON/OFF this analyzer.  
Turn ON/OFF the back light of display.  
2. Power switch  
3. Switch for back light  
4. Display / Operation panel  
Liquid crystal display and keys for various  
operational settings are arranged.  
5. Sampling gas inlet  
6. Sampling gas outlet  
Port for connecting the sample gas injection pipe.  
Port for connecting the pipe for discharging the  
gas after analysis.  
7. Purge gas inlet  
8. Terminal block 1  
9. Terminal block 2  
10. Terminal block 3  
11. Terminal block 4  
12. Terminal block 5  
13. Connector 2  
Port for connecting the purge gas pipe.  
Analog output terminals  
Analog signal and contact input terminals  
Contact output terminals  
Contact output terminal  
Alarm output terminal  
Communication interface  
14. Power inlet  
Used to connect the power cable.  
9
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
3. INSTALLATION  
DANGER  
This unit is not explosion-proof type. Do not use it in a place with explosive gases to avoid explosion, fire  
or other serious accidents.  
CAUTION  
• For installation, observe the rule on it given in the instruction manual and select a place where the  
weight of gas analyzer can be endured. Installation at an unsuited place may cause turnover or fall and  
there is a risk of injury.  
• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an injury.  
• Before transport, fix the casing so that it will not open. Otherwise, the casing may be separated and fall  
to cause an injury.  
• The gas analyzer is heavy. It should be transported carefully by two or more persons if manually  
required. Otherwise, body may be damaged or injured.  
• During installation work, care should be taken to keep the unit free from cable chips or other foreign  
objects. Otherwise, it may cause fire, trouble or malfunction of the unit.  
3.1 Installation conditions  
To install the analyzer for optimum performance, select a location that meets the following conditions:  
(1) Use this instrument indoors.  
(2) A vibration-free place  
(3) A place which is clean around the analyzer.  
(4) Power supply  
Rated voltage:  
Operating voltage:  
Rated frequency:  
Power consumption:  
Inlet:  
100V to 240VAC  
85V to 264VAC  
50/60 Hz  
70 VA max.  
Conformity to EN60320 class I type 3-pin inlet  
(5) Operation conditions  
Ambient temperature: -5 to 45°C  
Ambient humidity:  
90% RH or less, no condensation  
10  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
3.2 Installation of analyzer  
There are two methods of installing the analyzer.  
For detailed dimensions, see Chapter 9.3.  
Note: The mounting method should be selected to meet the installation requirements since the top cover  
must be detached from the gas analyzer for maintenance and check.  
Mounting method  
Slide rail  
Conditions  
Remarks  
No maintenance space is  
provided at the top.  
These methods must be rigid enough to  
withstand the mass (about 10 kg) of the  
gas analyzer.  
Guide rail  
Maintenance space is provided  
at the top  
Recommended slide rail: 305A-20, Accuride International Inc.  
3.3 Piping  
Caution on piping  
In piping, the following precautions should be observed.  
Wrong piping may cause gas leakage.  
CAUTION  
If the leaking gas contains a toxic component, there is a risk of serious accident being  
induced.  
Also, if combustible gas is contained, there is a danger of explosion, fire or the like  
occurring.  
• Connect pipes correctly referring to the instruction manual.  
• Exhaust should be led outdoors so that it will not remain in the locker and installation  
room.  
• Exhaust from the analyzer should be relieved in the atmospheric air in order that an  
unnecessary pressure will not be applied to the analyzer. Otherwise, any pipe in the  
analyzer may be disconnected to cause gas leakage.  
• For piping, use a pipe and a pressure reducing valve to which oil and grease are not  
adhering. If such a material is adhering, a fire or the like accident may be caused.  
11  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Observe the following when connecting the gas pipes.  
• The pipes should be connected to the gas inlet and outlet at the rear panel of the analyzer, respectively.  
• Connect the sampling system to the instrument by using corrosion-resistant tube such as Teflon,  
stainless steel, or polyethylene. In case where there is no danger of corrosion, donʼt use rubber or soft  
vinyl tube. Analyzer indication may become inaccurate due to the adsorption of gases.  
• Piping connections are Rc1/4 (NPT1/4) female-threaded. Cut the pipe as short as possible for quick  
response. Pipe of ø 4mm (inside diameter) is recommendable.  
• Entry of dust in the instrument may cause operation fault. Use clean pipes and couplings.  
Sampling gas inlet:  
Sampling gas outlet:  
Purge gas inlet:  
Connect the pipe so that zero/span calibration standard gas or measured gas  
pre-treated with dehumidification is supplied properly. The gas flow rate should  
be kept constant within the range of 1 L/min ±0.5 L/min.  
Measured gas is exhausted after measurement.  
Connect the pipe so that the gas may escape through the gas outlet into the  
atmosphere.  
It is used for purging the inside of the total gas analyzer. When the analyzer must  
be purged, refer to Item 3.4, Purging inside Analyzer. Use dry gas N2 or  
instrumentation air for purge gas. (flow rate of 1 L/min or more should be used  
and no dust or mist is contained).  
3.4 Sampling  
(1) Conditions of sample gas  
1. The dust contained in sample gas should be eliminated completely with filters. The filter at the  
final stage should be capable of eliminating dust of 0.3 microns.  
2.  
The dew point of sample gas must be lower than the ambient temperature for preventing  
formation of drain in the analyzer. If water vapor is contained in sample gas, its dew point should be  
reduced down to about 0°C through a dehumidifier.  
3
If SO3 mist is contained in sample gas, the mist should be eliminated with a mist filter, cooler, etc.  
Eliminate other mist in the same way.  
4. If a large amount of highly corrosive gas such as Cl2, F2 or HCl is contained in sample gas, the  
service life of analyzer will be shortened. So, avoid such gases.  
5. Sample gas temperature is allowed within a range from 0 to 50°C. Pay attention not to flow hot  
gas directly into the analyzer.  
12  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(2) Sampling gas flow rate  
A flow rate of sampling gas must be 1 L/min ±0.5 L/min. A flow meter should be provided as  
shown in Fig. 3-2 to measure flow rate values.  
(3) Preparation for standard gas  
Prepare the standard gas for zero/span calibration.  
No O2 meter  
N2 gas  
Built in O2 meter  
Externally mounted  
zirconia O2 meter  
Zero gas  
N2 gas  
Dry air or air (No  
calibration performed  
with CO2 meter)  
Span gas except for O2 Gas with concentration  
of 90% or more of full  
scale  
Gas with concentration  
of 90% or more of full  
scale  
Gas with concentration  
of 90% or more of full  
scale  
Span gas for O2  
Gas with concentration  
of 90% or more of full  
scale  
1 to 2% of O2 gas  
(4) Purging of instrument inside  
The inside of instrument need not be purged generally except for the following cases.  
1.  
2.  
3.  
A combustible gas component is contained in sample gas.  
Corrosive gas is contained in the atmospheric air at the installation site.  
The same gas as the sample gas component is contained in the atmospheric air at the installation  
site.  
In such cases as above, the inside of analyzer should be purged with the air for instrumentation or N2.  
Purging flow rate should be about 1L/min.  
If dust or mist is contained in purging gas, it should be eliminated completely in advance.  
(5) Pressure at sampling gas outlet  
Pressure at the sampling gas outlet should be set to atmospheric pressure.  
13  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(6) Example of sampling system configuration  
The system configuration may vary depending upon the nature of measured gas, coexistent gases or  
application. A typical configuration diagram is shown in Fig. 3-2. Since a system configuration depends  
upon measured gas, consult with Teledyne.  
Fig. 3-2 A typical example of sampling system  
14  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
3.5 Wiring method  
Caution on wiring  
CAUTION  
• Wiring work must be performed with the main  
power set to OFF to prevent electric shocks.  
• Enforce construction of class-D grounding wire  
by all means.  
If the specified grounding construction is  
neglected, a shock hazard or fault may be caused.  
• Wires should be the proper one meeting the  
ratings of this instrument. If using a wire which  
cannot endure the ratings, a fire may occur.  
• Be sure to use a power supply of correct rating.  
Connection of power supply of incorrect rating may  
cause fire.  
Each external terminal is provided on the rear panel of the analyzer. (See Fig. 3-3)  
Wire each terminal, referring to Fig. 3-3 and (1) to (7).  
Fig. 3-3 Rear panel  
(1) Power inlet  
When using supplied power cable, connect the female side to the power inlet at the rear panel of the  
analyzer, and insert the male side into a receptacle matching the rating.  
15  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
When a noise source is in the vicinity  
Do not install the analyzer near power noise generating electric equipment (such as high frequency  
furnace and electric welder). If the analyzer must be used near such equipment, a separate power line  
should be used for avoiding noise. In case noise may enter from a relay, solenoid valve, etc. through  
power supply, connect a varistor (such as ENA211-2 made by Fuji Electric) or spark killer (like S1201  
made by OKAYA) to the noise source as shown below. If the varistor or spark killer is located away from  
the noise source, no effect is obtainable. So, locate near the noise source.  
(2) Analog output signal (AO): terminal block 1 (1 to 10, 15 to 20)  
Output signal: 4 to 20 mADC or 0 to 1 VDC (selected when ordering)  
Non-insulated output  
Allowable load: 4 to 20 mADC, 550Ω or less  
0 to 1 VDC, 100kΩ or less  
• Analog output is provided from each terminal corresponding to the channel displayed in the  
measurement screen.  
Note: All of analog output signals for the instrument are not isolated. It is recommended to isolate  
signals individually to prevent interference from unnecessary signals or to prevent external interference,  
especially leading the cable of more than 30 meters or to outdoor.  
(3) O2 sensor input: terminal block 2 (1 – 2)  
Input signal:  
External zirconia O2 analyzer: Zirconia O2 sensor signal (Teledyne output)  
External O2 analyzer: 0 to 1 VDC (DC input resistor of 1MΩ or more)  
• It is used when the external zirconia O2 analyzer or external O2 analyzer is specified as order.  
To connect to the output of the external Zirconia analyzer or external O2 analyzer prepared separately.  
• In case of an external O2 analyzer, input a signal of 0 to 1 VDC with respect to O2 full scale of the  
analyzer.  
• In case of built-in O2 analyzer, do not use the terminals.  
Note: O2 sensor input is not isolated. It is recommended to isolate input signal to prevent interference  
from unnecessary signals or to prevent external interference. Zirconia O2 sensor should be installed at a  
location that is as close to this instrument as possible.  
16  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(4) Contact input (DI): terminal block 2 (13 to 20), terminal block 3 (5 to 10)  
• It is for a contact input at no voltage. An input is provided when switching to short circuit (on) or  
open (off).  
• No voltage is applied to the terminals  
(5) Contact output (DO): terminal block 3 (13 to 20), terminal block 4 and terminal block 5  
• Contact rating: 250VAC / 2A, load resistance  
• An output is for a relay contact output. An output is provided when switching to conductive (on)  
or open (off).  
Note: The wires of analog output signals, O2 sensor input and contact input should be fixed separately  
from power supply wiring and contact output wiring.  
(6) Communication interface: connector 2  
• Please refer to the manual for communication function.  
17  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(7) List of terminal blocks 1 to 5  
*1) Unused terminals are used for internal  
connection and should not be used as repeating  
terminals either.  
*2) O2 sensor input is used when an external O2  
analyzer is selected.  
18  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
8) Description on terminal block  
Terminal block 1 <TN1>  
Terminal block for analog output (non-isolated output)  
Output : 4 to 20 mA or 0 to 100V DC  
Between 1 – 2:  
Between 3 – 4:  
Between 5 – 6:  
Between 7 – 8:  
Between 9 – 10:  
Between 11 – 14:  
CH5 output  
CH4 output  
CH3 output  
CH2 output  
CH1 output  
For internal connection. Must not be  
wired. (Must not be used as junction  
terminal.)  
Between 15 – 16:  
Between 17– 18:  
Between 19 – 20:  
CH8 output  
CH7 output  
CH6 output  
Terminal block 2 <TN2>  
Between 1– 2:  
For O2 sensor input. (Input for our  
Zirconia oxygen sensor or external  
O2 sensor. Must not be used unless  
O2 meter is added.)  
Between 3 –12:  
For internal connection. Must not be  
wired. (Must not be used as junction  
terminal.)  
Between 13 – 14:  
Between 15 – 16:  
Between 17 – 18:  
Between 19 – 20:  
CH4 remote range changeover input  
CH3 remote range changeover input  
CH2 remote range changeover input  
CH1 remote range changeover input  
Note: High range is selected when open. Low range is  
selected when short-circuited. For details of action, refer to “6.7  
Parameters Setting, “Remote Range”.  
19  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Terminal block 3 <TN3>  
Between 1 – 4: For internal connection. Must not be  
wired. (Must not be used as junction  
terminal.)  
Between 5 – 6: Remote hold input. No hold when open.  
Output hold when short-circuited.  
Between 7 – 8: Average value reset input. Short-  
circuiting the contact input (for at 1.5 sec  
or more) resets O2 average and O2  
correction average simultaneously.  
Opening it restarts the average value.  
Between 9 – 10: Automatic calibration remote start  
input. Open input after strapping for  
at least 1.5 seconds starts the automatic  
calibration whether automatic calibration  
setting is ON or OFF.  
Between 11 – 12: Conductive when analyzer unit error  
is producted. Normally open.  
Between 13 – 14: CH4 range identification signal  
Between 15 – 16: CH3 range identification signal  
Between 17– 18: CH2 range identification signal  
Between 19– 20: CH1 range identification signal  
Note: Range identification signal is conductive at Low range  
or open at High range. In case of 1-range system, the signal  
remains open.  
20  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Terminal block 4 <TN4>  
Between 1 – 2: CH4 span calibration contact output  
Between 3 – 4: CH3 span calibration contact output  
Between 5 – 6: CH2 span calibration contact output  
Between 7 – 8: CH1 span calibration contact  
Between 9 – 10: Zero calibration contact output When the  
calibration contact output is measured with  
manual calibration, the calibration contact  
corresponding to calibration channel is  
conductive.  
For the automatic calibration, they are worked sequentially  
according to “6.4 Auto calibration setting”. If calibration is not  
performed, all of them are open.  
Between 11 – 12: For internal connection. Must not be wired.  
(Must not be used as junction terminal.)  
Between 13 – 14: Automatic calibration in progress,  
contact output.  
Conductive during automatic calibration. Open  
otherwise.  
Between 15 – 16: Calibration error contact output. Conductive  
when error is produced at zero or span  
calibration. Normally open.  
Between 17 – 18: Pump ON/OFF contact output. (Used for  
turning ON/OFF the pump. Conductive during  
measurement and open at zero span  
calibration.)  
21  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Terminal block 5 <TN5>  
1 and 11 – 14: For internal connection. Must not be  
wired. (Must not be used as junction terminal.)  
Between 2 - 3 and 3 - 4: CH3 alarm output. Conductive  
at 2 – 3 and open at 3 – 4 when set value is exceeded.  
Open at 2 – 3 and conductive at 3 – 4 otherwise.  
Between 5 - 6 and 6 - 7: CH2 alarm output. Conductive  
at 5 – 6 and open at 6 – 7 when set value is exceeded.  
Open at 5 – 6 and conductive at 6 – 7 otherwise.  
Between 8 - 9 and 9 - 10: CH1 alarm output. Conductive  
at 8 – 9 and open at 9 – 10 when set value is exceeded.  
Open at 8 – 9 and conductive at 9 – 10 otherwise.  
Between 15 - 16 and 16 - 17: Peak count alarm contact  
output. Conductive at 15 – 16 and open at 16 – 17 when  
preset peak count is exceed. Otherwise, open at 15 – 16  
and conductive at 16 – 17. For setting and action, refer  
to instruction manual “6.6 Peak Alarm Setting”.  
Between 18 – 19 and 19 – 20: CH4 alarm output.  
Conductive at 18 – 19 and open at 19 – 20 when set  
value is exceeded. Open at 18 – 19 and conductive at  
19 – 20 otherwise.  
22  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(9) Timing of calibration contact output  
1. Manual calibration (See “6.9 Calibration”)  
2. In case of automatic calibration (example shown in 6.4.1, Automatic calibration settings)  
23  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
4. OPERATION  
4.1 Preparation for operation  
(1) Check of gas sampling tube, exhaust tube and wiring  
Check that the pipes are correctly connected to the gas sampling port and drain port.  
Check that the analyzer is correctly wired as specified.  
4.2 Warm-up operation and regular operation  
(1) Operation procedure  
1. Turn ON the power switch at the left of the front panel.  
In one or two seconds, the measurement screen will appear at the front panel.  
2. About 2 hour warm-up operation  
About 2 hours are needed until the operating performance is stabilized. Warm-up  
operation should be continued with the power ON.  
3. Setting of various set values  
Set required set values according to Chapter 6, “Setting and calibration”.  
4. Zero and span calibration  
Perform zero calibration and span calibration after warm-up operation.  
See Chapter 6.9, “Calibration”.  
5. Introduction and measurement of measured gas  
Start measurement by introducing measured gas into the analyzer.  
Note: While in warm-up operation, the concentration reading may be beyond the upper limit of the range  
(range-over) or below the lower limit. But this is not an error.  
24  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5. DISPLAY AND OPERATION PANEL  
This section describes the display and operation panel of the gas analyzer. It also explains the name and  
description of function on the operation panel. See Fig. 5-1.  
5.1 Name and description of operation panel  
The measurement screen and the setting items are displayed.  
The configuration is as shown below.  
• Display:  
• Controls:  
Name  
Description  
Name  
Description  
1. MODE key  
Used to switch the mode.  
5. ESC key  
Used to return to a previous screen  
or cancel the setting midway.  
2. SIDE key  
Used to change the selected  
item (by moving the cursor) and  
numeral digit.  
6. ENT key  
Used for confirmation of selected  
items or values, and for execution  
of calibration.  
3. UP key  
Used to change the selected  
item (by moving the cursor) and  
to increase numeral value.  
7. ZERO key Used for zero calibration.  
8. SPAN key Used for span calibration.  
4. DOWN key  
Used to change the selected  
item (by moving the cursor) and  
to decrease numeral value.  
Fig. 5-1 Name and description of operation display and panel  
Note: The switch for back light is used for turning ON/OFF the back light of display. The life time of back  
light is 50,000 hours. Please turn OFF the switch when indicator is not necessary, so the life time will be  
longer.  
25  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5.2 Overview of display and operation panel  
Figure 5.2  
26  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5.3 Overview of display screen  
(1) Measurement mode screen  
On turning on the power switch, the Measurement Mode screen will appear.  
The measurement screen depends on the number of components. The following screen configuration  
shown as an example is for CO2, CO, O2 (Output at channel 6).  
When channel 5 or later is displayed, scroll up or down key to view another configuration which is beyond  
the screen.  
No. Name  
Function  
No. Name  
Description  
1
Component  
display  
Displays component  
of instantaneous  
value, converted  
instantaneous value,  
converted average  
value, etc.  
5
Peak alarm  
component  
display  
Displays peak alarm  
component.  
2
3
4
Concentration  
display  
Displays measured  
value of concentration.  
6
7
Peak alarm  
concentration  
display  
Displays peak alarm  
concentration (Upper  
limit value).  
Range display  
Unit display  
Displays range  
values.  
Peak alarm count Displays the alarm  
times exceeding the  
peak value.  
Displays unit with ppm 8  
and vol%.  
Peak alarm unit  
display  
Displays unit of peak  
alarm with times/h.  
Fig. 5-3 Name and function of measurement mode screen  
Instantaneous value and concentration value:  
The concentration display of CH (component) where sampling components such as “CO2”, “CO” or  
“O2 are displayed in the component display, indicates current concentration values of the measured  
components contained in gas that is now under measurement.  
• O2 correction concentration values:  
CH (components) where “cv**” is displayed as “cv CO” in the component display are calculated from the  
following equation, by setting sampling components, O2 instantaneous / concentration values and O2  
correction reference value (see item 6.8).  
Conversion output=  
On: The value of the O2 correction reference value (Value set by application)  
Os: Oxygen concentration (%)  
Cs: Concentration of relevant measured component  
K: The value of the fractional part is this equation.  
Where, K is When K ≥ 4, K=4. When K < 0, K=4. When Cs < 0, K=0.  
The correction components are CO and SO2 only.  
27  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
• O2 correction concentration average value:  
CH (component) where “AV / CV” is displayed as “AV / CV CO” in the component display and O2 average  
value, a value obtained by averaging O2 conversion concentration value or O2 concentration value in a  
fixed time is output every 30 seconds.  
Averaging time can be changed between 1 minute and 59 minutes or 1 hour and 4 hours according to the  
average time settings (See 6.7 Parameter setting).  
(The set time is displayed as “1h”, for instance, in the Range display.)  
* The measurement ranges of O2 correction concentration value and O2 correction concentration  
average value are the same as that of the measuring components. Also, the measurement range of O2  
average value is the same as that of O2.  
(2) Setting/selection screen  
The setting/selection screen is configured as shown below:  
• In the status display area, the current status is displayed.  
• In the message display area, messages associated with operation are displayed.  
• In the setting item and selection item display area, items or values to be set are displayed, as required.  
To work on the area, move the cursor to any item by using up, down, and right keys.  
Figure 5-4: Display screen  
28  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(3) Contents of measured channel (CH)  
The contents in each measured CH corresponding to the type are given below:  
Type code  
Contents  
5th  
digit  
6th digit 7th  
digit  
CH1  
O2  
CH2  
CH3  
CH4  
CH5  
CH6  
CH7  
Y
A
B
D
E
2
C
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
SO2  
CO  
Y
Y
CO2  
CH4  
CO2  
CH4  
CO2  
CO2  
SO2  
Y
Y
CO  
CO  
CH4  
CO  
O2  
3
Y
4
Y
5
Y
CH4  
A
B
D
E
2
A, B, C  
A, B, C Y, B CO  
O2  
A, B, C  
A, B, C  
Y
Y
CO2  
CH4  
O2  
O2  
A, B, C Y, B CO2  
A, B, C Y, B CH4  
CO  
CO  
CH4  
CO  
O2  
O2  
3
O2  
4
A, B, C  
A, B, C  
A, B, C  
Y
Y
A
CO2  
CO2  
SO2  
O2  
5
CH4  
O2  
Corrected O2  
SO2  
average  
A
Corrected  
SO2  
average  
B
2
3
5
A, B, C A, C CO  
A, B, C A, C CO2  
A, B, C A, C CH4  
A, B, C A, C CO  
O2  
CO  
CO  
CO  
Corrected  
CO  
Corrected O2  
CO  
average  
average  
O2  
Corrected Corrected O2  
CO  
CO  
average  
average  
O2  
Corrected Corrected O2  
CO  
CO  
average  
average  
CH4  
O2  
Corrected Corrected O2  
CO  
CO  
average  
average  
How to view the Table  
When SO2 is displayed in CH1, it means that display and output of CH1 correspond to SO2 component.  
29  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5.4 General operation  
• Measurement mode  
The measurement mode can be displayed for up  
to 5 channels in a single screen. When viewing  
a channel beyond the 5 channels, press the ( or  
) key and the screen can be scrolled one by one  
channel at a time.  
• User mode displays the following settings  
Changeover of range  
Calibration setting  
Alarm setting  
Setting of auto calibration  
Setting of auto zero calibration  
Peak alarm setting  
Parameter setting  
For settings, refer to “6, Setting and Calibration.”  
30  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6. SETTING AND  
CALIBRATION  
6.1 Changeover of range  
This mode is used to select the ranges of measured  
components.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Changeover of Range”. Press the  
ENT key.  
3 The “Channel Selection” screen appears. Press the UP  
or DOWN key until the cursor moved selects a desired  
CH (component).  
4. After selection, press the ENT key.  
Note: The range of O2 correction instantaneous values  
and O2 correction average values is automatically  
switched by changing the range of instantaneous value  
of each CH (component).  
5. In the Range Setting screen that appears, move the  
cursor by pressing the UP or DOWN key to select the  
range. (The range with a TRIANGLE mark is currently  
selected.)  
6. After selection, press the ENT key.  
7. Measurement is conducted within the selected range.  
The range identification signal (CO) is shorted with the  
low range (Range 1), and open with the high range  
(Range 2).  
Note: If the Remote Range is set to ON, the changeover  
of range cannot be performed on the screen.  
To close Changeover of range  
To close Changeover of range, or cancel the command  
midway, press the ESC key. A previous screen will  
return.  
31  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.2 Calibration setting  
This mode is used to set calibration concentration and  
actions. The calibration setting involves calibration  
concentration, zero calibration, calibration range and  
auto-calibration component.  
6.2.1 Setting of calibration concentration  
It allows you to set concentrations of the standard gas  
(zero and span) of each channel used for calibration.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting about Calibration” by  
pressing the UP or DOWN key. Press the ENT key.  
3. In the “Setting about Calibration” screen that appears,  
point the cursor to “Calibration Value” by pressing the UP  
or DOWN key. Press the ENT key.  
4. In the “Calibration Concentration CH Selection”  
screen that appears, point the cursor to CH you want to  
set by using the UP or DOWN key. Press the ENT key.  
32  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5. In the “Calibration Concentration Selection” screen  
that appears, select any concentration item you want to  
set by pressing the UP, DOWN, and RIGHT key.  
Note: Analyzers other than the zirconia O2 instrument  
cannot perform zero setting.  
6. In the “Calibration Concentration Value Setting”  
screen that appears, enter calibration gas concentration  
values (zero and span). For value entry, press the UP or  
DOWN key, and a 1-digit value increases or decreases.  
By pressing the RIGHT key, the digit moves.  
After setting, save the entry by pressing the ENT key.  
The saved value becomes valid from the next calibration  
process.  
Note: Enter the set values corresponding to each range.  
When the O2 measurement uses atmospheric air for the  
zero gas, set the concentration value to 21.00. When  
the cylinder air is used, set to the concentration value as  
indicated on the cylinder.  
To close the setting  
To close the calibration concentration value setting  
process or cancel this mode midway, press the ESC key.  
A previous screen will return.  
Setting range of values  
Paramagnetic O2, CO2, CO, SO2 and CH4  
measurement:  
Zero gas:  
Span gas:  
Fixed at 0  
Minimum digit, 1 to 100% of full  
scale  
(Full scale (FS) is the same as each range  
value.)  
Zirconia O2 measurement :  
Zero gas: 5 to 25 vol%  
Span gas: 0.01 to 5 vol%  
The setting cannot be performed beyond the range.  
33  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.2.2 Setting of manual zero calibration  
If zero calibration is to be made manually, select whether  
to calibrate all components at once or each of them  
separately upon selection.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting about Calibration” by  
pressing the UP or DOWN key. Press the ENT key.  
3. In the “Setting about Calibration” screen that appears,  
point the cursor to “About ZERO Calibration” by pressing  
the or key. Press the ENT key.  
4. In the “Manual Calibration CH Selection” screen that  
appears, point the cursor to CH you want to set by using  
the UP or DOWN key. Press the ENT key.  
34  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5. In the “Manual Calibration Selection” screen that appears, select “At once” or “Each” by pressing the  
UP or DOWN key. When selecting “Both”, the CH (components) to be set can be zero-calibrated at the  
same time. When selecting “Each”, either of the CH (components) to be selected is zero-calibrated. After  
setting, press the ENT key, and the calibration you specified is  
carried out.  
To close “About ZERO Calibration”  
To close “About ZERO Calibration” setting or to cancel this  
mode midway, press the ESC key. A previous screen will  
return.  
EXAMPLE  
Whether upon selection or at once can be determined for each  
CH (component).  
•Setting upon selection  
Select the CH (component) and then perform zero  
calibration on the manual zero calibration  
screen.  
•Setting at once  
At a manual zero calibration, CHs (components) for which “at once” was selected can  
simultaneously be calibrated.  
* When the zirconia O2 analyzer uses the cylinder air or atmospheric air for the zero gas, select “At once”.  
Manual Calibration screen  
35  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.2.3 Setting of calibration range  
This mode is used to set if the range of each CH  
(component) at the calibration manual calibration or auto  
calibration) should be calibrated with a single range or 2  
ranges.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting about Calibration” by  
pressing the UP or DOWN key. Press the ENT key.  
3. In the “Setting about Calibration” screen that appears,  
point the cursor to “About Calibration Range” by pressing  
the UP or DOWN key. Press the ENT key.  
4. In the “Calibration Range CH Selection” screen that  
appears, point the cursor to the CH you want to set by  
pressing the UP or DOWN key. Press the ENT key.  
36  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5. In the “Calibration Selection” screen that appears, select “Both” or “Current” by pressing the UP or  
DOWN key.  
• When selecting “both”, Range 1 and Range 2 of the set CH  
are calibrated together.  
• When selecting “Current”, the range alone displayed at the  
set CH is calibrated.  
To close “Setting of Calibration Range”  
To close “Setting of Calibration Range” or to cancel this mode  
midway, press the ESC key. A previous screen will return.  
EXAMPLE:  
CH1  
CO2  
Range 1: 0 to 10 vol%  
Range 2: 0 to 20 vol%  
both  
CH2  
CO  
Range 1: 0 to 500 ppm  
Range 2: 0 to 2000 ppm  
current  
CH1: Range 1 and Range 2 are calibrated together, with zero or span calibration.  
CH2: Only currently displayed range is calibrated, with zero or span calibration.  
Manual Calibration screen  
In case of “both” setting  
Two cursors will appear in both ranges.  
Note: When calibration is performed by the “Both” setting under the normal operating condition, prepare a  
span gas cylinder on the normal operating range side. It is recommend to perform span gas calibration in  
the normal operating range. The other range that is calibrated by “Both” may result in some error (max. of  
±5% of FS).  
37  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.2.4 Setting of auto-calibration component  
It sets the CH (component) to be calibrated in the auto-  
calibration.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting about Calibration” by  
pressing the UP or DOWN key. Press the ENT key.  
3. In the “Setting about Calibration” screen that appears,  
point the cursor to “Auto Calibration Components” by  
pressing the UP or DOWN key. Press the ENT key.  
4. In the “Auto Calibration Components” selection screen  
that appears, point the cursor to the CH you want to set  
by pressing the UP or DOWN key. Press the ENT key.  
38  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
5. In the “Auto Calibration Selection” screen that appears, select “enable” or “disable” by pressing the UP  
or DOWN key. After setting, press the ENT key.  
To close Auto Calibration Component setting  
To close “Setting of Auto Calibration Component” or to  
cancel this mode midway, press the ESC key. A previous  
screen will return.  
Example  
Auto calibration is made in the following rules according to the setting.  
1. Zero calibration at once of CHs (components) which were set to “enable”.  
2. The span of CHs (components) which were set to “enable” is calibrated in the  
ascending order of CH number.  
Example 1. In case all of CH1: CO2, CH2: CO, CH3: O2, were set to “enable”.  
Zero calibration (at once) of CH1 to CH3 n span calibration of CH1 (CO2) n  
span calibration of CH2 (CO) n span calibration of CH3 (O2)  
Example 2. In case, out of CH1: CO2, CH2: CO and CH3: O2, CH1 ( CO2) was set to “enable”,  
CH2 (CO) was set to “enable” an CH3 (O2) was set to “disable”.  
Zero calibration (at once) of CH1 and 2 n span calibration of CH1 (CO2) n  
span calibration of CH2 (CO)  
CAUTION  
The components which were set to “enable” is calibrated to zero at once at the time of auto calibration  
regardless of setting in “6.2.2 Setting of manual zero calibration”.  
39  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.3 Alarm setting  
6.3.1 Setting of alarm values  
This mode is used to set the upper and lower limit value  
to provide an alarm output during measurement. Before  
changing the alarm setting, set the ON/OFF to OFF.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Alarm Setting” by pressing the UP  
or DOWN key. Press the ENT key.  
3. After the alarm setting CH selection screen has  
appeared, operate the UP or DOWN key until the cursor  
is aligned with a desired CH and press the ENT key.  
4. After the alarm item selection screen has appeared,  
operate the or key until the cursor is aligned with a  
desired item and press the ENT key.  
CAUTION  
Set the values so that upper limit value > lower limit  
value and that (upper limit value - lower limit value) >  
hysteresis error.  
The alarm is activated in each range independently from  
the settings of each range.  
40  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
After setting, the alarm setting is now completed by pressing the ENT key.  
To close the “Alarm Setting”  
To close the “ Alarm Setting” or to cancel this mode midway,  
press the ESC key.  
A previous screen will return.  
Setting range  
0 to 100% FS (Settable in each range).  
Description of setting items  
Upper limit value:  
Lower limit value:  
Contact action:  
Sets the upper limit value (concentration) of alarm.  
Sets the lower limit value (concentration) of alarm.  
Selects one of upper limit alarm, lower limit alarm, and upper limit or lower limit  
alarm.  
Upper limit alarm ... Alarm contact operates when above upper limit value.  
Lower limit alarm ... Alarm contact operates when below lower limit value.  
Upper limit alarm or lower limit alarm ...Alarm contact operates when above upper limit value or  
below lower limit value.  
ON/OFF: Enables the alarm function if set at ON, or disables the alarm function if set at OFF.  
* The upper limit value cannot be set below the lower limit value, and the lower limit value cannot be set  
above the upper limit value.  
If it is desired to set the upper limit value below the lower limit value already stored in the memory, reduce  
the lower limit value beforehand, or vice versa.  
Typical on-screen display when an alarm occurs  
When an upper limit alarm has occurred, the “High alarm” message  
lights at CH (component)  
(“Low alarm” at lower limit alarm).  
CAUTION  
For 10 minutes after turning on power, the alarm judgment is inactive.  
41  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.3.2 Hysteresis setting  
To prevent chattering of an alarm output near the alarm  
setting values, set hysteresis.  
1. In the “Alarm Setting CH Selection” screen that  
appears, point the cursor to “Hysteresis” by pressing the  
UP or DOWN key. Press the ENT key.  
2. In the “Hysteresis Value Setting” screen that appears,  
enter hysteresis values. For the value entry, 1-digit value  
is increased or decreased by pressing the UP or DOWN  
key, and pressing the RIGHT key moves the digit. After  
setting, press the ENT key to carry out hysteresis.  
To close Hysteresis Setting  
To close the “Hysteresis Setting” or cancel the mode  
midway, press the ESC key. A previous screen will  
return.  
Setting range  
0 to 20% of full scale  
A full scale means each range provides a full scale of  
width.  
Caution  
The hysteresis is common to all CHs (components).  
Hysteresis  
If hysteresis values exceed the upper limit value as  
shown in graph, an alarm output is provided. Once the  
alarm output is turned ON, it remains ON until the value  
falls below the set lower limit of the hysteresis indication.  
42  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.4 Setting of auto calibration  
6.4.1 Auto calibration  
Auto calibration is automatically carried out when zero  
calibration and span calibration are set.  
Before changing the setting of auto calibration, set the  
ON/OFF to OFF.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting of Auto Calibration” by  
pressing the or key. Press the ENT key.  
3. In the “Setting of Auto Calibration” screen that  
appears, point the cursor to any item you want by  
pressing the UP or DOWN key. Press the ENT key.  
4. In the “Auto Calibration Parameter Setting” screen  
that appears, perform the value entry or the setting. For  
the value entry or setting change, use the UP or DOWN  
key. For the value entry and setting, use the UP or  
DOWN key. To change the setting, use the RIGHT key to  
move the cursor to the right.  
After setting, press the ENT key, and auto calibration is  
carried out by the entered setting value.  
Description of setting items  
• Start Time: Setting at the first calibration (day of the  
week, hour, minute)  
• Cycle: A period between the start time of one  
calibration and another (unit : hour/day)  
• Flow Time: The time required for the calibration gas to  
be replaced in the cell  
• ON/OFF: Auto zero calibration ON or OFF  
To close Setting of Auto calibration  
To close the “Setting of Auto calibration” or cancel this  
mode midway, press the ESC key. A previous screen will  
return.  
43  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Auto calibration status contact output is closed during auto calibration, and is open in other cases.  
Setting range  
Cycle:  
Flow time:  
1 to 99 hours or 1 to 40 days (initial value 7 days)  
60 to 599 sec (initial value 300sec)  
CAUTION  
When an auto calibration starts, the measurement screen automatically appears.  
Any operation other than forced stop of auto calibration (see Item 6.4.2) is not permitted during auto  
calibration. “Auto Calibration Cancel” cannot be performed with the key lock to ON. To cancel auto  
calibration forcedly, set the key lock to OFF and then execute “Auto Calibration Cancel”.  
Remote start  
Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the remote  
start input short-circuited for at least 1.5 seconds.  
44  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.4.2 Forced stop of auto calibration  
This mode is used to cancel the auto calibration forcedly.  
1. In the Menu mode that is displayed, point the cursor  
to “Setting of Auto Calibration” by pressing the UP or  
DOWN key. Press the ENT key.  
2. In the “Setting of Auto Calibration” item selection  
screen that appears, point the cursor to “Setting of Auto  
Calibration” by pressing the UP or DOWN key. Press the  
ENT key.  
3. “Stop Auto Calibration” is inverted. A message  
appears, prompting you to verify that you want to  
cancel or continue auto calibration. To cancel the auto  
calibration, press the ENT key. If you press the ESC key,  
auto calibration is not stopped.  
45  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
“Auto Calibration” screen  
Example  
In case where setting the auto calibration components  
(see Item 6.2.4) to “CH1: enable” and “CH2: enable”.  
• Zero calibration  
A message, “Zero calibration” blinks at CH1 and CH2.  
• CH1 span calibration  
A message, “Span calibration” blinks at CH1.  
• CH2 span calibration  
A message, “Span calibration” blinks at CH2.  
CAUTION  
During auto calibration, any key operation is not  
permitted other than operations such as key lock ON/  
OFF and “Stop Auto Calibration”. When the key lock is  
set at ON, even the “Stop Auto Calibration” cannot be  
used. To stop “Auto Calibration” forcedly, set the key lock  
to OFF and then execute “Stop Auto Calibration”.  
46  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.5 Setting of auto zero calibration  
6.5.1 Auto zero calibration  
Auto zero calibration is automatically carried out when  
zero calibration are set.  
Components for which a calibration is to be made are  
determined by setting of auto calibration component in  
6.2.4.  
Before changing the setting of auto zero calibration, set  
the ON/OFF to OFF.  
1. During measurement, press the MODE key to display  
the User mode.  
2. Point the cursor to “Setting of Auto Zero Calibration”  
by pressing the UP or DOWN key. Press the ENT key.  
3. In the “Setting of Auto Zero Calibration” screen that  
appears, point the cursor to any item you want by  
pressing the UP or DOWN key. Press the ENT key.  
4. In the “Auto Zero Calibration Parameter Setting”  
screen that appears, perform the value entry or the  
setting. For the value entry or setting change, use the  
UP or DOWN key. To change the setting, use the RIGHT  
key to move the cursor to the right.  
After setting, press the ENT key, and auto zero  
calibration is carried out by the entered setting value.  
Description of setting items  
• Start Time: Setting at the first calibration (day of the  
week, hour, minute)  
• Cycle: A period between the start time of one  
calibration and another (unit : hour/day)  
• Flow Time: The time required for the calibration gas to  
be replaced in the cell  
• ON/OFF: Auto zero calibration ON or OFF  
To close “Auto Zero Calibration”  
To close the “ Auto Zero Calibration “ or cancel this  
mode midway, press the ESC key. A previous screen will  
return.  
47  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Auto calibration status contact output is closed during auto zero calibration, and is open in other  
cases.  
Setting range  
Cycle:  
Flow time:  
1 to 99 hours or 1 to 40 days (initial value 7 days)  
60 to 599 sec (initial value 300 sec)  
Caution  
• When an auto zero calibration starts, the measurement screen automatically appears.  
• Any operation other than forced stop of auto zero calibration (see Item 6.5.2) is not permitted during auto  
zero calibration. “Auto Zero Calibration Cancel” cannot be performed with the key lock to ON. To cancel  
auto zero calibration forcedly, set the key lock to OFF and then execute “Auto Zero Calibration Cancel”.  
• If the auto calibration period and auto zero calibration period have overlapped, the auto calibration is  
retained, ignoring the auto zero calibration of that period.  
48  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.5.2 Forced stop of auto zero calibration  
This mode is used to cancel the auto zero calibration  
forcedly.  
1. In the Menu mode that is displayed, point the cursor to  
“Setting of Auto Zero Calibration” by pressing the UP or  
DOWN key. Press the ENT key.  
2. In the “Setting of Auto Zero Calibration” item selection  
screen that appears, point the cursor to “Setting of Auto  
Zero Calibration” by pressing the UP or DOWN key.  
Press the key.  
3. “Stop Auto Zero Calibration” is inverted. A message  
appears, prompting you to verify that you want to cancel  
or continue auto zero calibration. To cancel the auto zero  
calibration, press the ENT key. If you press the ESC key,  
auto zero calibration is not stopped.  
49  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
“Auto Zero Calibration” screen  
Example  
In case where setting the auto calibration components  
(see Item 6.2.4) to “CH1: enable” and “CH2: enable”.  
• Zero calibration  
A message, “Zero calibration” blinks at CH1  
and CH2.  
CAUTION  
During auto zero calibration, any key operation is not  
permitted other than operations such as key lock ON/  
OFF and “Stop Auto Zero Calibration”.  
When the key lock is set at ON, even the “Stop Auto  
Zero Calibration” cannot be used. To stop “Auto Zero  
Calibration” forcedly, set the key lock to OFF and then  
execute “Stop Auto Zero Calibration”.  
50  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.6 Peak alarm setting  
When the number of peaks of which CO concentration  
exceeds the upper limit value exceeds the setting time, a  
peak alarm is outputted.  
This section describes how to perform various settings of  
peak alarm.  
Note: The setting is optional and is valid only when peak  
alarm function is provided.  
1. Press the MODE key in the Measurement mode, and  
the User mode appears.  
2. Point the cursor to “Setting of Peak Alarm” by pressing  
the or key. Press the ENT key.  
3. In the “Peak Alarm Setting” item selection screen  
that appears, point the cursor to any item you want by  
pressing the UP or DOWN key. Press the ENT key.  
4. Then, enter numeric values and perform the setting.  
Entering or setting the numeric values should be carried  
out by using the UP or DOWN key.  
After setting, press the ENT key, and the set values you  
entered are saved.  
Description of setting items  
• Peak Alarm: Setting of peak alarm is performed with  
ON/OFF.  
• Alarm value: If peak concentrations exceed the set  
alarm value, a peak counter counts 1 time.  
• Alarm Count: When a peak in excess of the setting time  
occurs, a peak count alarm output is provided.  
• Hysteresis: To prevent possible chattering when the  
peak value may exceed the set peak concentration by  
only 1 time, the peak count has an allowance in the  
hysteresis width.  
51  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Setting range  
• Alarm Value: 0 to 1000 ppm 5 ppm step (initial value: 500 ppm)  
• Alarm Count: 1 to 99 times (initial value: 5 times)  
• Hysteresis:  
0 to 20% of full scale (initial value: 0% of full scale)  
* The hysteresis setting is made in terms of full scale.  
Action of peak alarm  
If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set  
times per hour, a peak count alarm becomes closed (ON). If it is less than the set times per hour, it is  
open (OFF). Since 5 times of peaks /hour is marked at 1 section from the above graph, the peak count  
alarm is turned ON. Since peaks of more than 5 times per 1 hour occur at the interval between 1 and 2 ,  
the peak count alarm remains ON. Since peaks are reduced to 4 times per hour, it is turned OFF.  
Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when measured gas  
is fluctuated near the alarm value.  
* For 10 minutes after the power is turned ON, a peak alarm counting is not carried out.  
Releasing peak alarm  
To release the peak alarm, set the peak alarm to OFF.  
Turning on the peak alarm initiates counting from 0.  
52  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.7 Parameter setting  
It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be set  
are as follows:  
Description of setting items  
• Current Time:  
• Key Lock:  
Sets the current day of the week, hour and time.  
Sets with ON/OFF so that any key operation except the key lock OFF cannot  
be performed.  
• Remote Range:  
• Output Hold:  
• Average Value Reset: Resets the average value.  
Sets with ON/OFF whether the Range Selection is made valid or invalid by  
external input.  
Sets whether Calibration Output is held or not.  
• Response Speed: Sets the response time of electrical system.  
• Average Time: Sets the moving average time.  
• Maintenance mode: Enters passwords to switch to the Maintenance mode.  
* For the maintenance mode, see Item 6.8.  
1. To display the User mode, press the MODE key in the  
measurement mode.  
2. Point the cursor to “Parameter Setting” by pressing the or  
key. Press the ENT key.  
3. In the “Parameter Setting” screen that appears, point the  
cursor to any item you want by pressing the or key. Press  
the ENT key.  
53  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
4. In the Parameter Setting screen that appears, enter the  
numeric values and set the items. Entering the numeric  
values or setting the items should be carried out by using the  
or key. To move the cursor to the right, press the key.  
After setting, press the ENT key, that the parameter setting is  
carried out with the value you set.  
To close Parameter Setting screen  
To close the “Parameter Setting” screen or cancel this mode  
midway, press the ESC key. A previous screen will return.  
Setting Range  
• Response time:  
• Average time:  
1 to 60 sec. (Initial value: 3 sec)  
1 to 59 min or 1 to 4 hours (Initial  
value: 1 hour)  
When setting the unit of 1 to 59 minutes in terms of minute or 1 to 4 hours with hour  
• Maintenance mode: 0000 to 9999 (Initial value: 0000)  
Remote Range  
A range can be switched via an external input by setting the Remote Range to ON. (The switching action  
affects all of instantaneous value, O2 correction value, O2 correction average value and O2 average  
value.) If the Remote Range is set to OFF, the external input becomes invalid.  
Opening the input gives the High range, or short-circuiting the input gives the Low range.  
For the terminal input, refer to 3.4, Wiring.  
Switching the range cannot be performed by on-screen operation when the REMOTE RANGE is set to  
ON.  
Note: In case of 1 range system, this function is overridden.  
Output Hold  
By setting an output hold to ON, an output signal of each channel are held during the calibration (manual  
calibration and auto calibration) and for the gas flow time (refer to 6.4, Setting of Auto Calibration).  
Regardless of Hold ON/OFF setting, an output signal can be held via an external input.  
(1) Manual calibration  
54  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(2) Auto calibration  
(3) External hold  
(4) Screen display during holding  
The “Holding” message blinks on the measuring screen.  
Since the screen displays the process of calibration is displayed during the manual calibration, “Holding”  
is not displayed even if the screen is held, but the screen is displayed with the hold extending time.  
(5) If calibration is cancelled after the calibration gas is supplied regardless of during manual calibration  
or auto calibration, an output hold of the holding extending time will be performed.  
Average value reset  
This mode is used to clear O2 correction average value and O2 average value, and restarts averaging. All  
average values are reset at a time. The indication value and output value is 0 ppm, vol% or so at the time  
of the reset input.  
So long as short-circuited, resetting lasts.  
At the edge of changing from short circuit to opening, the  
average action restarts.  
Response time  
The response time of the electrical system can be changed.  
Setting is available by components.  
Note: It does not provide exact seconds for the setting time,  
but it gives a guide of the setting time. The setting value can be  
modified as requested by the customer.  
55  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Average period  
It allows you to set an average period of the average value of  
O2 correction.  
It enables you to set an average time of 1 to 59 minutes (1-  
minute step) or 1 to 4 hours (1-hour step).  
Changing the setting resets the average value of O2  
correction and O2 average value. (Pressing the ENT  
validates the resetting only for components whose setting  
was changed.)  
Example of average action  
Suppose the average period is 1 hour.  
• Sampling occurs every 30 seconds  
• Every 30 seconds, the average for last 1 hour (time setting) is output.  
• At the instant of resetting, zero is assumed for all past values. It means the average value will not be  
correct for 1 hour after resetting.  
Maintenance mode  
To open the maintenance mode, enter a password. After entering the password, press the ENT key.  
The password can be used for the Password Setting in the Maintenance mode. A password is set to  
“0000” before factory-shipment. This value is available for the Maintenance mode.  
56  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.8 Maintenance mode  
This mode is used for check of sensor input values,  
display of error log files or setting of passwords.  
First, enter a password and then use it from the next  
operation. This mode is displayed by selecting the  
Maintenance Mode from “6.7 Parameter Setting”  
1. Select the Maintenance mode from the Parameter  
Setting screen to display the Password Setting screen.  
2. Enter the password, and the Maintenance Mode item  
selection screen will be displayed. Point the cursor to the  
item you want by pressing the or key and press the ENT  
key.  
3. Next, each Maintenance screen is displayed.  
Note: “To Factory Mode” is used for our service  
engineers only. Refrain from using this mode.  
• Sensor Input Value screen  
Description of Sensor Input Value screen  
• CO2:  
• CO:  
• O2:  
CO2 sensor input value  
CO sensor input value  
O2 sensor input value  
Temperature : Temperature sensor input value  
• Error Log screen  
Description of Error Log screen  
Error history  
For error number, date and time (days, period) of  
occurrence, channel and other details of error, refer to  
“8.1 Error message”.  
57  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
• O2 correction reference value setting screen  
• Password Setting screen  
Description of Password Setting screen  
It enables you to set a password to be used when switching the  
parameter setting mode to the maintenance mode.  
Note: The password set herein must be managed for safety. Failure  
to enter the correct password will not open the Maintenance mode.  
• Station No. setting screen (option)  
Setting range  
• Station No.: 00 to 31 (Initial No.: 00)  
* Please refer to another manual (INZ-TN513327-E) about the  
communication function.  
58  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.9 Calibration  
6.9.1 Zero calibration  
It is used for zero point adjustment. For zero calibration  
gas, see 3.3 (3), Preparation for standard gas in  
Sampling. Use a gas according to application.  
1. Press the ZERO key on the Measurement screen to  
display the Manual Zero Calibration screen.  
2. Select the CH (component) to be calibrated by  
pressing the UP or DOWN key. After selection, press the  
ENT key, and zero gas will be supplied.  
CAUTION  
For the CH (components) that is set to “at once” in the  
“Zero Calibration” of the Calibration Setting mode, zero  
calibration is also carried out at the same time.  
3. Wait until the indication is stabilized with the zero gas  
supplied. After the indication has been stabilized, press  
the ENT key. Zero calibration in Range selected by the  
cursor is carried out.  
To close Zero Calibration  
To close the “Zero Calibration “ or cancel this mode  
midway, press the ESC key. A previous screen will  
return.  
59  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
6.9.2 Span calibration  
It is used to perform a span point adjustment. Supply  
calibration gas with concentration set to the span value  
to perform the span calibration. For the span calibration  
gas for the SO2, CO2, CO, H4 measurement, use the  
standard gas with a concentration of 90% or more of  
the range value. For the calibration gas for the O2  
measurement, use the standard gas of about 2 vol%.  
1. Press the SPAN key on the Measurement screen to  
display the Manual Span Calibration screen.  
2. Press the UP or DOWN key to select CH (component)  
to be calibrated. After selection, press the ENT key, and  
the span calibration contact output on corresponding to  
CH is turned ON. And supply span gas.  
CAUTION  
When Range Interlock from “Calibration Range” of  
the Calibration Setting mode is set, span calibration is  
performed together with 2 Ranges.  
3. Wait until the indication is stabilized in the state where  
the calibration gas is supplied. After the indication has  
been stabilized, press the ENT key. Span calibration of  
Range selected by the cursor is performed.  
To close Span Calibration  
To close the “ Span Calibration “ or cancel this mode  
midway, press the ESC key. A previous screen will  
return.  
60  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
7. MAINTENANCE  
7.1 Daily check  
(1) Zero calibration and span calibration  
1. It is used for zero point adjustment. For calibration, refer to 6.9.1, Zero calibration.  
2. After zero calibration, perform span calibration. For calibration, refer to 6.9.2, Span calibration.  
3. Zero calibration and span calibration should be performed once a week, if required.  
(2) Flow check  
1. Sampling gas flow rate and purge gas flow rate should be as follows;  
Sampling gas flow rate: 1.0 L / min ±0.5 L / min (stable)  
Purge gas flow rate: About 1L / min  
2. Maintenance and check should be carried out every day, if required.  
7.2 Daily check and maintenance procedures  
Table 7-1 Maintenance and check list  
Parts to be checked Phenomena  
Cause  
Remedy  
Every day  
Recorder indication Lower indication  
1. Dust is mixed in  
the sample cell  
1. Clean sampling  
cell and check for  
sampling device,  
especially gas filter  
2. Air is sucked in  
anywhere in the  
sampling tube  
2. Check for leak of  
the sampling line  
and repair if required  
Check for purge gas Standard flow rate  
Adjust the flow  
rate with flow rater  
needle valve  
flow if purging the  
sampling gas flow  
instrument  
is 1L / min. It is not  
within the range of  
the specified flow  
rate of 0.5 to 1.5 L /  
min.  
Replacement  
of Monitor filter  
(membrane filter)  
Much clogged  
Primary filter is  
damaged  
1. Replace primary  
filter  
2. Replace filter  
(paper)  
Every week Zero point of gas  
analyzer  
Out of zero point  
Zero calibration  
Span point of gas  
analyzer  
Out of the standard  
point  
Span calibration  
Replacement  
of monitor filter  
(membrane filter)  
Irrespective of  
phenomena  
Replace filter  
(paper)  
Every year  
Gas analyzer  
Irrespective of  
phenomena  
Overhaul  
61  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
7.3 Cleaning of measuring cell  
Entry of dust or water drops in the measuring cell contaminates the interior of the cell, thus resulting in  
a drift. Clean the inside if dirty. Then, check the sampling device, especially the filter, to prevent the cell  
from being contaminated by dust or mist.  
7.3.1 Disassembly and assembly of measuring cell  
There are two kinds of measuring cells, on block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and pipe  
cells (Cell length: 6.4 mm, 125 mm and 250 mm). 2-component analyzer may incorporate both measuring  
cells in optical unit. In such a case, detach the pipe cell and then block cell (See Fig. 7-1).  
(1) How to remove pipe cell (See Fig. 7-1)  
1. Stop measured gas. If it is harmful, purge in the measuring cell thoroughly with zero gas.  
2. Turn OFF the power switch.  
3. Detach the top cover.  
4. Remove the pipe connected to the measuring cell.  
5. Slide the infrared ray light source unit (No. 5) toward the front panel by loosening the screw  
(No. 1) fastened to the base plate to provide clearance between the pipe cell (No. 12)  
and light source unit.  
6. Loosen and remove a screw (No. 7) from the cell retainer (No. 11) fastening the pipe cell.  
7. Remove the cell from the measuring unit and unscrew the infrared transmission window  
(No. 14) at the both ends in the right direction.  
Note: The reflection plate in the cell is not closely attached to the cell.  
8. For assembly, reverse the disassembly procedure. Provide 0.5 mm clearance between the  
infrared ray light source unit and cell, and the cell and detector, respectively.  
62  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Number  
Name  
1
Screw (for fixing light source unit)  
Screw (for fixing detector)  
Screw (for fixing base plate)  
Base plate  
2
3
4
5
Light source unit  
6
Screw (for fixing support)  
Screw (for fixing cell retainer)  
Chopper motor connector  
Filter  
7
8
(9)  
10  
11  
12  
13  
14  
15  
16  
17  
(18)  
Support  
Cell retainer  
Pipe cell  
O-ring  
Infrared transmission window  
Detector  
Bridge PCB  
Bridge resistance  
Detector: Installed in 2-component analyzer only  
Fig. 7-1 Configuration of measuring unit (pipe cell)  
63  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(2) How to remove block cell (See Fig. 7-2)  
1. For Step 1 to 4, see Item 7.3.1, (1) How to remove pipe cell.  
5. Remove the connector to the detector output cord from the printed board. For the 2-component  
analyzer, remove the output cord connector of the 2-component analyzer detector (No. 13)  
from the printed board, and then remove the 2-component detector by unscrewing two mounting  
screws (No. 14) fastening the 2-component detector.  
6. Unscrew the two screws (No. 10) that hold the detector to the infrared ray light source unit to  
remove the detector from the measuring unit. The cell can be removed together with the detector.  
7. To remove the cell, unscrew the two screws (No. 6) holding the cell to the detector. The infrared  
transmission window is just sandwiched (not fixed) between the detector and block cell. Keep the  
detector facing up, when removing this window.  
8. For assembly, reverse the disassembly procedures.  
Note: The O-ring is placed between the window holder and cell. Take care about the O-ring position. With  
2-component analyzer, install 2-component detector last. Take care so that no space is left between the  
1-component and 2-component detectors. When inserting the detector output cord connector into the  
printed board, plug the connectors for 1-component detector and 2-component detector into position. The  
1-component connector should be plugged into CN11 and 2-component connector into CN1, respectively.  
64  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Number  
Name  
1
Screw (for fixing light source unit)  
Filter  
(2)  
3
Screw (for fixing detector)  
Base plate  
4
5
Light source unit  
6
Screw (for fixing block cell)  
Block cell  
7
8
Infrared transmission window (window holder)  
O-Ring  
9
10  
11  
12  
(13)  
(14)  
Screw (for fixing base plate)  
Chopper motors connector  
Detector  
Detector: Install in the 2-component analyzer  
Screws (for mounting the 2-component detector)  
Fig. 7-2 Configuration of measuring unit (block cell)  
65  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
3) How to remove measuring unit (See Fig. 7-3)  
1. For Step 1 to 4, see Item 7.3.1(1), How to remove pipe cell.  
2. Remove the detector output cord connector from the printed board.  
3. Remove wiring to the 2-pin terminals of the infrared ray light source assembly and chopper  
motor pin connector (No. 8) from the printed board.  
4. Detach the 4 screws (No. 3) fastening the base plate (No. 4) to remove the measuring unit.  
Note: Special care should be taken when assembling or disassembling the measuring cell to avoid the  
application of force to the detector pipe or infrared ray light source unit pipe. If the pipe is deformed or  
damaged by excessive force, there is a danger of gas leak, thus resulting in misoperation.  
Number  
Name  
1
Screw (for fixing light source)  
Screw (for fixing detector)  
Base plate  
2
3
4
Light source unit  
Screw (for fixing block cell)  
Block cell  
5
6
7
Infrared transmission window  
O-ring  
8
9
Detector  
10  
11  
12  
13  
14  
15  
16  
17  
18  
Screw (for fixing base plate)  
Support  
Screw (for fixing cell retainer)  
Cell retainer  
Pipe cell  
O-ring  
Infrared transmission window  
Screw (fixing detector)  
Detector  
Fig. 7-3 Configuration of measuring unit (2-component analyzer: block cell + pipe cell)  
66  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
7.3.2 How to clean cell  
1. To clean the cell inside or infrared ray transmission window, first clear large dirt of it with a soft  
brush and then wipe with soft cloth lightly.  
Donʼt use hard cloth.  
Note: Handle the fragile window with care. Use care not to rub off the dirt from the window roughly.  
2. If the window or the cell interior is very dirty, use a soft cloth moistened with absolute alcohol.  
3. If the window is corroded, rub off the scale from the window lightly with a soft cloth to which  
chrome oxide powder is applied. If it is excessively corroded, it should be replaced with  
new one.  
4. When cell or window cleaning is completed, assemble according to the cell disassembly and  
assembly procedures. Especially, the pipe should be closely connected without gas leak,  
and repair if the pipe is bent.  
67  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
8. TROUBLESHOOTING  
8.1 Error message  
If errors occur, the following contents are displayed.  
Error display  
Error No. 1  
Error contents  
Probable causes  
Sector motor rotation detector signal is faulty.  
• Infrared ray light source is faulty.  
• Sector motor rotation is faulty or stopped.  
• Motor rotation detector circuit is faulty.  
• Amplifier circuit is faulty.  
Error No. 3  
Error No. 4  
A/D conversion signal is faulty.  
• Circuit is faulty.  
Zero calibration is not within the allowable  
range.  
• Zero gas is not supplied.  
• Zero point is deflected much due to a dirty cell.  
• Detector is faulty.  
Error No. 5  
Error No. 6  
Error No. 7  
A amount of zero calibration (indication value) is  
over 50% of full scale.  
Span calibration is not within the allowable  
range.  
• Span gas is not supplied.  
• Calibrated concentration setting does not  
match cylinder concentration.  
• Zero calibration is not performed properly.  
• Span is deflected much due to dirty cell.  
• Detector sensitivity is deteriorated.  
A amount of span calibration (difference  
between indication value and calibrated  
concentration value) is over 50% of full scale.  
Error No. 8  
Error No. 9  
Error No. 10  
Measured values fluctuate to much during zero  
and span calibration.  
• Calibration gas is not supplied.  
• Time for supplying calibration gas is not short.  
Calibration is abnormal during auto calibration.  
• Error corresponding to No. 4 to No. 8 occurred  
during auto calibration.  
Output cable connection is improper.  
• Wiring is detached between analyzer and  
interface module.  
• Wiring is disconnected between analyzer and  
interface module.  
Note: When errors No. 1, No. 3 and No. 10 occur, instrument error output contacts are conductive.  
When errors No. 4 to No. 9 occur, calibration error output contacts are conductive.  
Screen display and operation at the occurrence of error.  
In case of Error No. 1 to No. 4, No. 6, No. 8 to No. 10  
68  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
In case of Error No. 5 and No. 7  
(2) Error log file  
If error occurs, the history is saved in an error log file. The error log file exists in the maintenance mode.  
* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every time  
a new occurs.  
* If the power display supply is turned OFF, the contents in the error log file will not be lost or damaged.  
Deletion of error history  
Press the ENT key on the above screen, and the “Error Log Clear” will be inverted. Further pressing the  
key will clear the error history.  
69  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Relay contact output:  
9. SPECIFICATIONS  
9.1 Specifications  
1a contact (250V AC/2A, resistive load) Instrument error,  
calibration error, range discrimination, auto calibration  
status, solenoid valve drive for auto calibration, pump  
ON/OFF.  
(1) Standard specifications  
Principle of measurement:  
CO2, CO, CH4, SO2; Non-dispersion infrared-ray  
absorption method; Single light source and single beam  
(single beam system)  
1c contact (250V AC/2A, resistive load) Upper/lower  
alarm contact output. Peak count alarm contact output.  
* All relay contacts are isolated mutually and from the  
internal circuit.  
O2: Paramagnetic method (O2 sensor built in) or  
zirconia sensor method (O2 sensor externally installed)  
Contact input:  
Non-voltage contact (ON/0V, OFF/5V DC, 5mA flowing at  
ON); Remote range changeover, auto calibration remote  
start, remote holding, average value resetting; Isolated  
from the internal circuit with a photocoupler. Contact  
inputs are not isolated from one another.  
Measurable gas components and measuring range:  
Minimum range Maximum  
range  
*Only M3.5 screw terminals are used for all signal inputs  
and outputs.  
CO2  
0 to 500 ppm  
0 to 200 ppm  
0 to 1000 ppm  
0 to 1000 ppm  
0 to 5 vol%  
0 to 100 vol%  
0 to 100 vol%  
0 to 100 vol%  
0 to 5000 ppm  
0 to 100 vol%  
0 to 25 vol%  
CO  
Power supply:  
Rated voltage:  
Operating voltage:  
Rated frequency:  
Power consumption:  
CH4  
100V to 240V AC  
85V to 264V AC  
50/60 Hz  
SO2  
O2 (built in)  
70 VA max.  
O2 (external  
zirconia)  
0 to 5 vol%  
Power inlet:  
Conformity to EN60320 Class I type  
Operation conditions:  
Ambient temperature: -5°C to 45°C  
• Max. 4 components measurement including O2.  
• 1 or 2 measuring range per component.  
• Measuring range ratio ≤ 1:5 (except built in O2) ≤ 1:20  
(built in O2)  
Ambient humidity:  
90% RH max.  
Storage conditions:  
Ambient temperature: -20°C to 60°C  
Max. 4 components and 2 ranges are selectable including  
an O2 measurement. For measurable components and  
possible combinations of measuring ranges, refer to  
Tables 1 to 6.  
Ambient humidity:  
100% RH max.noncondensing  
Dimensions (H x W x D):  
19-inch rack mounting type:  
Desk-top type:  
177 x 483 x 493mm  
194 x 483 x 493mm  
Measured value indication:  
Digital indication in 4 digits (LCD with back light)  
• Instantaneous value of each component  
• Instantaneous value after O2 correction (only in CO,  
SO2 measurement with O2)  
Mass: Approx. 10 kg  
Finish color:  
Front panel; Off-white (Munsell 10Y7.5 / 0.5 or equivalent)  
• Average value after O2 correction (only in CO, SO2  
measurement with O2)  
• O2 average value  
Casing: Steel-blue  
Enclosure: Steel casing, for indoor use  
Analog output signals:  
Material of gas-contacting parts:  
Gas inlet/outlet; SUS304  
Sample cell; SUS304/neoprene rubber  
Infrared-ray transmitting window; CaF2  
Internal tubing; Toaron tube  
4 to 20mA DC or 0 to 1V DC, non-isolated output. Analog  
output corresponds to measured value indication in 1:1.  
Permissible load; 550Ω max. for 4 to 20 mA DC100kΩ  
min. for 0 to 1V DC  
* Refer to Table 6, for the channel No. of displayed values  
and analog output signals.  
Gas inlet/outlet:  
Rc1/4 or NPT1/4 internal thread  
Analog input signal:  
Purge gas flow rate:  
For signal input from externally installed O2 sensor.  
Signal requirement; (1) Signal from Teledyneʼs Zirconia  
O2 sensor (TYPE: ZFK7) (2) 0 to 1V DC from an O2  
sensor Input section is not isolated. This feature is  
effective when an O2 sensor is not built in.  
1 L / min (when required)  
70  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(2) Standard functions  
Auto zero calibration:  
Auto zero calibration is carried out periodically at  
the preset cycle. This cycle is independent on “Auto  
calibration” cycle.  
When zero calibration gas and solenoid valve for  
opening/closing the calibration gas flow line are prepared  
externally by the customer, zero calibration will be carried  
out with the solenoid valve drive contact for zero  
calibration turned on/off at the set auto zero calibration  
timing.  
Output signal holding:  
Output signals are held during manual and auto  
calibrations by activation of holding (turning on its  
setting). The values to be held are the ones just before  
start calibration mode. Indication values will not be held.  
Remote output holding:  
Output signal is held at the latest value by short-circuiting  
the remote output holding input terminals. Holding  
is maintained while the terminals are short-circuited.  
Indication values will not be held.  
Auto zero calibration cycle setting:  
Auto zero calibration cycle is set. Setting is variable  
within 1 to 99 hours (in increments of 1 hour) or setting is  
variable within 1 to 40 days (in increments of 1 day).  
Remote range changeover:  
Measuring range can be changed according to an  
external signal when remote range changeover input  
is received. Changeover is effective only when remote  
range setting is turned on. In this case, measuring range  
cannot be changed manually.  
Gas flow time setting:  
The timing for flowing zero gas in auto zero calibration  
is set. Settable 60 to 599 seconds (in increments of 1  
second)  
When the contact input terminals for each component  
are short-circuited, the first range is selected, and it is  
changed over to the second range when the terminals are  
open.  
Upper/lower limit alarm:  
Alarm contact output turns on when the preset upper or  
lower limit alarm value is reached.  
Contacts close when the instantaneous value of each  
component becomes larger than the upper alarm limit  
value or smaller than the lower alarm limit value.  
Range identification signal:  
The present measuring range is identified by a contact  
signal. The contact output terminals for each component  
are short-circuited when the first range is selected, and  
when the second range is selected, the terminals are  
open.  
Instrument error contact output:  
Contacts close at occurrence of analyzer error No. 1, 3 or  
10.  
Calibration error contact output:  
Contacts close at occurrence of manual or auto  
calibration error (any of errors No. 4 to 9).  
Auto calibration:  
Auto calibration is carried out periodically at the preset  
cycle. When a standard gas cylinder for calibration and  
a solenoid valve for opening/closing the gas flow line are  
prepared externally by the customer, calibration will be  
carried out with the solenoid valve drive contacts for  
zero calibration and each span calibration turned on/off  
sequentially at the set auto calibration timing.  
Auto calibration status contact outputs:  
Contacts close during auto calibration.  
Pump ON/OFF contact output:  
During measurement, this contact close. While calibration  
gas is flowing, this contact open. This contact is  
connected in power supply of pump, and stop the sample  
gas while calibration gas flowing.  
Auto calibration cycle setting:  
Auto calibration cycle is set. Setting is variable within 1  
to 99 hours (in increments of 1 hour) or 1 to 40 days (in  
increments of 1 day).  
Gas flow time setting:  
(3) Optional functions  
O2 correction:  
Conversion of measured CO and SO2 gas concentrations  
into values at standard O2 concentration  
The time for flowing each calibration gas in auto  
calibration is set. Settable within 60 to 599 seconds (in  
increments of 1 second)  
21-On  
Correction formula: C = ––––––––- x Cs  
21-Os  
Auto calibration remote start:  
Auto calibration is carried out only once according to an  
external input signal. Calibration sequence is settable in  
the same way as the cyclic auto calibration. Calibration  
starts when a non-voltage rectangular wave is applied to  
the auto calibration remote start input terminals (opened  
after short-circuiting for 1.5 seconds or longer). Auto  
calibration is started when the contacts open.  
C: Sample gas concentration after O2 correction  
71  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
CS: measured concentration of sample gas  
Os : Measured O2 concentration  
Interference from other gases:  
On: Standard O2 concentration (value changeable by  
setting)  
*The upper limit value of the fractional part in this  
calculation is 4. The result of calculation is indicated and  
output in an analog output signal.  
Interference  
component  
CO2  
ana-  
lyzer  
CO  
analyzer  
CH4  
ana-  
lyzer  
SO2  
ana-  
lyzer  
O2  
ana-  
lyzer  
CO 1000  
ppm  
≤1%  
FS  
≤1%  
FS  
≤1%  
FS  
CO2 15%  
≤1% FS (for ≤1%  
≤1%  
FS  
≤2%  
FS  
Average value after O2 correction and O2 average  
value calculation:  
200 ppm  
analyzer,  
≤2.5% FS)  
FS  
The result of O2 correction or instantaneous O2 value  
can be output as an average value in the determined  
period of time. Used for averaging is the moving average  
method in which sampling is carried out at intervals of 30  
seconds.  
H2O satura- ≤1%  
tion at 20°C FS  
≤1% FS (for ≤1%  
200 ppm  
analyzer,  
≤2.5% FS)  
FS  
(Output is updated every 30 seconds. It is the average  
value in the determined period of time just before the  
latest updating.) Averaging time is settable within 1 to 59  
minutes (in increments of 1 minute) or 1 to 4 hours (in  
increments of 1 hour).  
H2O satura-  
tion at 2°C  
≤2.5% FS  
(for 200  
ppm ana-  
lyzer)  
≤5%  
FS  
Average value resetting:  
(6) Standard Requirements for Sample Gas  
Flow rate:  
The above-mentioned output of average value is started  
from the initial state by applying a non-voltage rectangular  
wave to the average value resetting input terminals  
(opened after short-circuiting for 1.5 seconds or longer).  
Output is reset by short-circuiting and restarted by  
opening.  
1L / min ±0.5L / min  
0 to 50°C  
Temperature:  
Pressure:  
10 kPa or less (Gas outlet side should  
be open to the atmospheric air.)  
Dust:  
Mist:  
Moisture:  
100 µg/Nm3 in particle size of 0.3 µm or less  
Unallowable  
Below a level where saturation occurs  
CO concentration peak count alarm:  
at room temperature (condensation unallowable). Below  
a level to cause saturation at 2°C for CO measurement  
with 0 to 200 ppm range and SO2 measurement.  
(added only for CO/O2 measurement) Alarm output turns  
on according to the preset concentration and count.  
Whenever the instantaneous value of CO exceeds the  
preset concentration value, count increments. If the count  
exceeds the preset value in one hour, the alarm contacts  
close.  
Corrosive component: HCl 1 ppm or less  
Standard gas for calibration:  
Zero gas; Dry N2  
Span gas; Each sample gas having concentration 90  
to 100% of its measuring range (recommended). Gas  
beyond concentration 100% is unusable.  
In case a zirconia O2 analyzer is installed externally  
and calibration is carried out on the same calibration gas  
line:  
Zero gas; Dry air or atmospheric air  
Span gas; For other than O2 measurement, each  
sample gas having concentration 90 to 100% of its  
measuring range. For O2 measurement, O2 gas of 1 to 2  
vol%.  
Communication function:  
RS-232C (9pins D-sub)  
Half-duplex bit serial  
Start-stop synchronization  
ModbusTM protocol  
Contents: Read/Write parameters; Read measurement  
concentration and instrument status.  
Remark : When connecting via RS-485 interface, an RS-  
232C / RS-485 converter should be used.  
(4) Applicable standards  
Product safety: EN61010-1 :2001  
EMC: EN61326-1 :1997, A1 :1998, A2 :2001  
(7) Installation conditions  
Use this instrument indoors.  
(Select a place where the equipment does not receive  
direct sunshine, draft/rain or radiation from hot  
substances. If such a place cannot be found, a roof or  
cover should be prepared for protection. )  
A vibration-free place.  
(5) Performance  
Repeatability:  
Linearity:  
Zero drift:  
±0.5% of full scale  
±1% of full scale  
±2% of full scale/week  
±2% of full scale/week  
Span drift:  
A place which is clean around the analyzer.  
Response time: (for 90% FS response, including  
replacement time of sample gas)  
1 or 2 component measurement: Within 15 seconds  
3 or 4 component measurement : Within 30 seconds  
72  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
9.2 Code symbols  
73  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
74  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
<Code specification for ordering>  
(1) Code symbols should be specified.  
(2) Range combination should be the one allowed in separate tables.  
(note 1)  
“Y” should be specified when only O2 measurement is necessary.  
(note 2)  
the 6th digit.  
External O2 sensor signal should be 0-1VDC linear of full scale, when “ B ” is specified at  
(note 3)  
(note 4)  
(note 5)  
(note 6)  
(note 7)  
“C” should be specified when “ Y ” is specified at the 5th digit.  
Allowable combinations of ranges are specified in the Table 1 to 6 in page 9-7 and 9-8.  
Allowed only for CO analyzer.  
Allowed only for CO and CO2 analyzers.  
O2 correction is carried out for CO and SO2. At the same time, average value output  
after O2 correction and O2 average value output are added. Peak count alarm is carried out for CO.  
(note 8) “Y” should be specified when without O2 measurement (when 6th digit is Y).  
(note 9)  
A gas composition table should be issued.  
(note 10) Rated voltage of supplied power cable and plug configuration differ when “E” and “U” are  
specified at the 10th digit. Select one from three power voltages. Plug configuration is North America type  
when specifying “E”, and European type when specifying “U”.  
75  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Measurable component and range - availability check table -  
76  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Table 5: Triple component analyzer (CO2, CO, CH4)  
Covered in combination of Table 2, Table 3 and Table 4.  
77  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
9.3 Outline diagram  
(1) Main unit  
78  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
(2) Accessory slide rail (unit: mm)  
* The slide rails are attached to this equipment when designated.  
Model : 305A-20/Accuride International Inc.  
19-inch rack mounting method:  
The mass of the instrument should be supported at the bottom of the unit (or the side of the unit when  
mounted with the slide rails).  
Also, for facilitate maintenance, a structure which allows extraction of the main unit by using the slide rail  
is recommended.  
79  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne Analytical Instruments  
Appendix  
Drawing List  
D79104  
Outline diagram  
80  
Model 7500 Instruction Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tasco Digital Camera 119256CW User Manual
Thermos Gas Grill 461644604 User Manual
Toshiba DVD Recorder D R160SB User Manual
Toshiba Sprinkler 6F3B0253 User Manual
Toshiba Two Way Radio MMU AP0091H User Manual
TRENDnet Switch TE100S80g User Manual
Triarch Indoor Furnishings 60162 User Manual
Tripp Lite Power Supply 120V Input Output User Manual
Uniden Radio GMR1088 2CK User Manual
Vantec Network Card CB USBARC User Manual