Teledyne Marine Instruments HPM 4 6 User Manual

TELEDYNE  
HASTINGS  
INSTRUMENTS  
INSTRUCTION MANUAL  
MODEL HPM 4/6  
VACUUM GAUGE  
I S O 9 0 0 1  
C E R T I F I E D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
1.0  
2.0  
3.0  
QUICK START INSTRUCTIONS................................................................................................................... 4  
SPECIFICATIONS ............................................................................................................................................ 5  
INSTALLATION............................................................................................................................................... 6  
3.1.  
3.2.  
3.3.  
RECEIVING INSPECTION ................................................................................................................................. 6  
VACUUM SENSOR (DV-4 OR DV-6 SERIES).................................................................................................... 6  
BATTERY INSTALLATION ............................................................................................................................... 6  
4.0  
GENERAL INFORMATION............................................................................................................................ 7  
4.1.  
4.2.  
SENSOR ELECTRICAL CONNECTOR ............................................................................................................... 7  
PRESSURE MEASUREMENTS ........................................................................................................................... 7  
5.0  
CALIBRATION ................................................................................................................................................. 9  
5.1.  
5.2.  
CALIBRATION IN A VACUUM SYSTEM ............................................................................................................ 9  
CALIBRATION WITH A REFERENCE TUBE....................................................................................................... 9  
6.0  
6.1.  
MAINTENANCE............................................................................................................................................. 11  
AUTHORIZED MAINTENANCE ...................................................................................................................... 11  
7.0  
WARRANTY.................................................................................................................................................... 12  
7.1.  
7.2.  
WARRANTY REPAIR POLICY......................................................................................................................... 12  
NON-WARRANTY REPAIR POLICY................................................................................................................ 12  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 3 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.0 Quick Start Instructions  
1. Unpack unit and insure battery is installed  
2. Connect thermocouple tube.  
4. Indicated pressure is displayed on LCD  
display.  
3. Slide switch to DV4 or DV6 position  
according to tube type  
5. Slide switch to center/off position when not  
in use.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 4 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.0 Specifications  
Operating Range....................................................................DV-4 - 0.01 - 20 Torr (.01 to 20 mbar)  
........................................................................................... DV-6 - 0.001 – 1 Torr (.001 to 1.3 mbar)  
Operating temperature ......................................................................................................-20 to 50°C  
Digital Display ..................................................................................................................... 3 ½ Digit  
Input Voltage................................................................................................................. 9 Volt Battery  
Input Output ............................................................................................. 11mA @ 9 VDC (Battery)  
HPM 4/6 Dimensions In (mm) ............................................................................. 5.91” (150 mm) x  
........................................................................................................... 3.15” (80mm) x 1.18” (30mm)  
Weight with Case-Lbs (kg).......................................................................................... 1.8 lbs (0.8 kg)  
CE Compliance ........................................................................................... EN61326:1997, A1:1998  
................................................................................................................................ A2:2001, A3:2003  
Electromagnetic interferences (EMI) in the frequency range 80MHtz to 1000MHz at intensity levels of 3V/m may cause  
the HPM-4/6 unit to display up to a 21% deviation from the correct reading.  
Removing the source of the EMI or increasing the distance between the source and the HPM-4/6 will return the unit  
to its normal operating condition.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 5 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.0 Installation  
3.1.Receiving Inspection  
Unpack unit and inspect all items for obvious signs of damage. Immediately advise Teledyne  
Hastings and the carrier of any suspected damage.  
In the unlikely event that items need to be returned, first obtain an RMA (Return Material  
Authorization) number from our Customer Service Department at 1-800-950-2468.  
3.2.Vacuum Sensor (DV-4 or DV-6 series)  
Figure 3-1 shows a typical Hastings thermocouple vacuum tube. A standard 1/8” npt male stem  
is shown for illustration purposes only. If other type fittings are required, refer to Hastings product  
bulletin, PB-100A, for availability of other fittings. Product bulletins and other technical  
orientation however, the ideal installation is with the tube vertical and stem down. Install  
thermocouple tubes observing standard shop practices.  
Fig 3.1  
3.3.Battery Installation  
The battery cover is located on the bottom rear  
of the HPM 4/6 Instrument. When the battery  
requires replacing, it can be accessed by pressing in  
on the battery cover release latch (the small slot on  
the top edge of the cover) with a small object such as  
a key or small screwdriver and then rotating cover  
open. Plug 9-volt battery into battery connector  
terminals. Place battery and cable into the battery  
compartment. Reinstall battery cover by inserting  
the bottom edge of the cover first and then closing  
the cover until it snaps into the locked position.  
Note: A low battery will be indicated by “LOBAT”  
appearing in lower left corner of LCD display.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 6 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.0 General Information  
Operation of the Hastings gauge sensor (DV-4 or DV-6) is based on a low voltage AC bridge  
that heats a noble metal thermopile. A change in pressure in the gauge tube alters the molecular  
collision rate and therefore the thermal conduction of the gas or gas mixture surrounding the  
thermopile. This causes a temperature shift in the AC heated thermocouples A and B (Fig. 3-1).  
The resultant temperature shift causes a change in the DC output from couples A and B (which  
varies inversely with pressure). Thermocouple C provides compensation for transient changes in  
ambient temperature.  
4.1.Sensor Electrical Connector  
The thermocouple vacuum sensor tube interfaces with the HPM 4/6 Vacuum Gauge via the  
octal connector. The Octal connector is “keyed” so that proper connection is assured. Always grip  
the connector by its cover and not the cable when plugging or unplugging the connector.  
Disconnecting gauge tube by pulling on the cable may result in damage to the octal socket  
and/or cable.  
4.2.Pressure Measurements  
The HPM 4/6 is shipped from the factory fully calibrated, and will provide immediate and  
accurate pressure measurements.  
1. Plug octal socket on to DV4 or DV6 thermocouple tube.  
2. Turn unit on by positioning the slide switch on the side on of instrument case to DV-4 or  
DV-6 position as required depending on which thermocouple tube is being monitored. The  
DV4 or DV6 switch position indicators are labeled on the bottom of the unit. Once the slide  
switch is in the proper position, the display will indicate one of the following:  
a. The correct pressure. (Value within range of sensor selected).  
b. Pressure over range “1”, where the sensor is at a pressure greater than full scale or the  
sensor not connected.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 7 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
c. Pressure under range “-0.00 to -0.05” (negative display values) indicating that the  
instrument requires low-pressure adjustment. Refer to Section 4.  
3. Always place slide switch to the off position when not in use. This will help prolong the life of  
the battery.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 8 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.0 Calibration  
5.1.Calibration in a Vacuum System  
1
Install a DV-4 or DV-6 sensor in your vacuum system and evacuate vacuum chamber to a  
pressure less than 1.00E-04 Torr.  
NOTE: Refer to Fig 3 for locations of calibration pots. Adjust only the pot that  
corresponds to the particular type of sensor tube installed (DV4 or DV6).  
2
DV-4 or DV6 zero adjustment pot:  
a. Remove battery compartment cover.  
b. Adjust DV-4 or DV6 (as required for particular type of tube installed) calibration pot, until  
display reads 0.00.  
3
4
The HPM4/6 is now calibrated. Slide switch to OFF position and disconnect gauge cable from  
sensor.  
Re-Install battery compartment cover.  
5.2.Calibration with a Reference Tube  
NOTE: Refer to Fig 3 for locations of calibration pots. Adjust only the pot that  
corresponds to the particular type of sensor tube installed (DV4 or DV6).  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 9 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
DV-4 low pressure adjustments:  
a. Connect instrument gauge cable to a DB-16D Reference Tube and position slide switch to  
DV-4.  
b. Remove battery compartment cover, and then adjust DV-4 calibration pot until display reads  
pressure value listed on reference tube label.  
c. Set slide switch to OFF position and disconnect gauge cable from sensor.  
d. Reinstall battery compartment cover.  
2
DV-6 low pressure adjustment:  
a. Connect instrument gauge cable to a DB-20 Reference Tube and position slide switch to  
DV-6.  
b. Remove battery compartment cover, and then adjust DV-6 calibration pot until display reads  
pressure value listed on reference tube label.  
c. Slide switch to OFF position and disconnect gauge cable from sensor.  
d. Reinstall battery compartment cover.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 10 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.0 Maintenance  
6.1.Authorized Maintenance  
The HPM 4/6 Power Supply/Display does not require maintenance other than routine battery  
replacement and calibration.  
Note: A low battery will be indicated by “LOBAT” appearing in lower left corner of LCD  
display.  
In the unlikely event that the HPM4/6 fails or is damaged, only qualified Teledyne Hastings  
personnel should attempt repair. If the HPM4/6 should require service, the unit may be returned by  
first obtaining an RMA (Return Material Authorization) number from our Customer Service  
Department at 1-800-950-2468.  
Refer to Section 7 last page.  
Instrument repair should only be performed by Teledyne Hastings service technicians.  
Contact with harsh solvents and chemicals to any plastic parts of the HPM 4/6 Vacuum  
Gauge can damage the unit and may void the warranty.  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 11 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.0 Warranty  
7.1.Warranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment  
to be free from defects in material and workmanship. This warranty does not apply to defects or  
failures resulting from unauthorized modification, misuse or mishandling of the product. This  
warranty does not apply to batteries or other expendable parts, nor to damage caused by leaking  
batteries or any similar occurrence. This warranty does not apply to any instrument which has had a  
tamper seal removed or broken.  
This warranty is in lieu of all other warranties, expressed or implied, including any implied  
warranty as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or  
consequential damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product  
if Hastings Instruments determines, in good faith, that it is defective in materials or workmanship  
during the warranty period. Defective instruments should be returned to Hastings Instruments,  
shipment prepaid, together with a written statement of the problem and a Return Material  
Authorization (RMA) number. Please consult the factory for your RMA number before returning  
any product for repair. Collect freight will not be accepted.  
7.2. Non-Warranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order,  
RMA form and a written description of the problem with the instrument. If the repair cost is higher,  
you will be contacted for authorization before we proceed with any repairs. If you then choose not to  
have the product repaired, a minimum will be charged to cover the processing and inspection. Please  
consult the factory for your RMA number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
(757) 723-6531  
1-800-950-2468  
FAX  
(757) 723-3925  
E MAIL  
INTERNET ADDRESS  
Repair Forms may be obtained from the “Information Request” section of the Hastings  
Instruments  
Manual: 153-082010_HPM 4-6 Vacuum Gauge  
Page 12 of 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Technics Home Theater System SC DV170 User Manual
Technics Universal Remote EUR643809 User Manual
Telect Network Card MIX 56 User Manual
Timex Watch W 191 NA User Manual
Topcom Telephone 2900 C User Manual
Toshiba DVD Player SD 330ESB User Manual
Trane Ventilation Hood Tracer Unit Ventilator User Manual
Tripp Lite Network Cables P753 010 User Manual
Unitech Stereo System PA966 User Manual
VTech IP Phone DS6472 6 User Manual