Teledyne Home Theater Screen THCD 100 User Manual

TELEDYNE  
HASTINGS  
INSTRUMENTS  
INSTRUCTION MANUAL  
THCD-100  
CONFIGURABLE DISPLAY  
I S O 9 0 0 1  
C E R T I F I E D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
1.  
2.  
GENERAL INFORMATION............................................................................................................................................ 4  
1.1.  
1.2.  
FEATURES .................................................................................................................................................................... 4  
SPECIFICATIONS........................................................................................................................................................... 4  
INSTALLATION AND INTERCONNECTIONS........................................................................................................... 5  
2.1.  
INSTALLATION ............................................................................................................................................................. 5  
TRANSDUCER SOCKET PINOUT – 15 PIN 'D' TYPE......................................................................................................... 5  
INTERFACE SOCKET PIN OUT – MINIATURE 15 PIN 'D' TYPE ........................................................................................ 6  
SERIAL INTERFACE SOCKET PINOUT – 9 PIN 'D' TYPE.................................................................................................. 6  
POWER CONNECTION SOCKET ..................................................................................................................................... 7  
PANEL MOUNTING ....................................................................................................................................................... 7  
2.2.  
2.3.  
2.4.  
2.5.  
2.6.  
3.  
FEATURES......................................................................................................................................................................... 8  
3.1.  
INTRODUCTION ............................................................................................................................................................ 8  
ANALOG INPUTS .......................................................................................................................................................... 8  
ANALOG OUTPUTS....................................................................................................................................................... 8  
SETPOINT CONTROL..................................................................................................................................................... 8  
USER RE-ZERO............................................................................................................................................................. 9  
ADAPTIVE FILTERING .................................................................................................................................................. 9  
3.2.  
3.3.  
3.4.  
3.5.  
3.6.  
4.  
5.  
FRONT PANEL OPERATION....................................................................................................................................... 11  
4.1.  
4.2.  
4.3.  
INTRODUCTION .......................................................................................................................................................... 11  
COMMON SWITCH FUNCTIONS................................................................................................................................... 11  
FRONT PANEL SCREENS ............................................................................................................................................. 12  
SERIAL COMMUNICATION........................................................................................................................................ 17  
5.1.  
INTRODUCTION .......................................................................................................................................................... 17  
COMMAND/QUERY FORMAT...................................................................................................................................... 17  
MAIN COMMANDS ..................................................................................................................................................... 18  
SERIAL COMMANDS................................................................................................................................................... 20  
CHANNEL SETUP COMMANDS.................................................................................................................................... 21  
FILTERING COMMANDS.............................................................................................................................................. 21  
RELAY CONTROL COMMANDS ................................................................................................................................... 22  
5.2.  
5.3.  
5.4.  
5.5.  
5.6.  
5.7.  
OTHER COMMANDS.................................................................................................................................................................. 23  
6.  
SERVICE & CALIBRATION......................................................................................................................................... 24  
6.1.  
6.2.  
6.3.  
6.4.  
SERVICE..................................................................................................................................................................... 24  
ANALOG INPUTS ........................................................................................................................................................ 24  
CLEANING.................................................................................................................................................................. 24  
END OF LIFE............................................................................................................................................................... 24  
7.  
8.  
APPENDICES................................................................................................................................................................... 25  
7.1.  
7.2.  
7.3.  
APPENDIX A............................................................................................................................................................... 25  
APPENDIX B............................................................................................................................................................... 26  
APPENDIX C............................................................................................................................................................... 27  
WARRANTY .................................................................................................................................................................... 28  
8.1.  
8.2.  
WARRANTY REPAIR POLICY ...................................................................................................................................... 28  
NON-WARRANTY REPAIR POLICY ............................................................................................................................. 28  
165-072011 THCD-100 Power Supply  
Page 3 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. GENERAL INFORMATION  
The THCD-100 is a multi featured process display controller, capable of interfacing directly to a mass flow meter,  
mass flow controller or another process transducer such as the THPS-760S Vacuum Transducer. The THCD-100 is  
a single channel unit which can be used to power a bipolar (±15dc @ 250mA) device and/or display the output from  
a transducer that supplies voltage signals up to 10 VDC. There is also an input that can be used for external  
command (setpoint) control (+5v full scale). Digital interfaces to the THCD-100 include RS485 and RS232.  
1.1.Features  
16-bit Resolution  
±0.2% Reading + 1 Digit Accuracy  
0-5 Volt Set Point Via Front Panel Or Digitally (RS232/RS485)  
Dual Programmable Alarms with Relays  
RS232/485 and Analog Output (Standard)  
CE Approved  
1.2.Specifications  
For a list of specifications, please see the PB-185_rev-0711_THCD-100 Power Supply.  
165-072011 THCD-100 Power Supply  
Page 4 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Installation and Interconnections  
2.1.Installation  
The THCD-100 can be panel mounted, or used on a bench top and is a standard 96 x 48mm panel meter  
enclosure. Note that the instrument must always be situated in such a way as to enable adequate air  
circulation about the unit. If a transducer was supplied with the THCD-100 a suitable cable may have been  
included.  
See Section 2.6 for a cut out detail for panel mounting.  
2.2.Transducer socket Pinout – 15 Pin 'D' Type  
Pin Number  
Designation  
Signal Common  
Signal In (VDC)  
Case Gnd  
5
6
7
9
-15 VDC  
11  
+15 VDC  
12  
Valve Return  
+24 VDC  
13  
14  
Set Point Out (VDC)  
Not Connected  
All other pins  
If the transducer is being powered from another source (e.g. a mains powered supply) it will only be  
necessary to connect to the two signal pins. Note: The THCD-100 has a fully floating differential input. For  
single ended use (i.e. transducers with differential output) join 0V (pin5) & Signal Return (pin12) at the  
transducer. Ensure the common mode input range is not exceeded.  
165-072011 THCD-100 Power Supply  
Page 5 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.3.Interface Socket Pin out – Miniature 15 Pin 'D' Type  
Pin Number  
Designation  
1
Analog Output  
2
Relay 2 Common  
3
Relay 2 Normally Open contact  
Relay 2 Normally Closed contact  
0V (Signal Ground)  
External Setpoint Return  
N/C  
4
5
6
7
8
External Setpoint  
9
N/C  
10  
11  
12  
13  
14  
15  
0V (Signal Ground)  
TTL Input – Disable Menu  
TTL Input – Disable Zero  
Relay 1 Common  
Relay 1 Normally Open contact  
Relay 1 Normally Closed contact  
2.4.Serial  
Interface  
Socket Pinout – 9 Pin 'D' Type  
Pin Number  
Designation 1 2  
RS485 Tx +  
RS232 Rx  
RS232 Tx  
RS485 Tx –  
Common  
1
2
3
4
5
6
7
8
9
RS485 Rx +  
RS485 Rx –  
N/C  
Common  
1 For RS232 only, connect only the pins designated in bold font.  
2 For RS485 only, connect only the pins designated in normal font.  
See section 4 for instructions on changing the serial protocol.  
165-072011 THCD-100 Power Supply  
Page 6 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.5.Power Connection Socket  
The THCD-100 requires a 24V, 500mA supply to operate. This must be a rectified and regulated direct  
current supply, capable of 15W (recommended). Although the THCD-100 has an internal fuse and power  
supply overload shutdown protection the supply to the THCD-100 should be protected by fuse or other  
suitable electronic method. A standard 5.5mm diameter ‘dc power jack’ is required to connect to the  
THCD-100. Connect the positive conductor to the center pin. Connect the negative (0V) conductor to the  
outer of the jack.  
For convenience, Hastings offers a compatible supply (12-01-169) that, when matched with the US (12-01-  
160), Euro (12-01-164) or the UK (12-01-165) plug adapters, converts wall power to specified levels.  
Cautions: Ensure correct power supply rating and polarity. Do not use  
a partially assembled or faulty unit. Incorrect use of this equipment, or  
use in a manner not specified may impair the user’s protection.  
2.6.Panel Mounting  
A panel mount kit will have been supplied with the instrument. This consists of: 2 screws (M3x6), 2 brackets  
and 2 screws (50mm). Fit the M3x6 screws into the side of the unit into the diagonally opposed holes. Do  
not tighten these screws, to allow the bracket to slide over the screw head. Thread locking adhesive may be  
used but is not required as these screws cannot rotate once the bracket is tightened. Insert the meter into the  
panel. A cutout of 3.63” x 1.84” (92 x 45mm) is recommended. Slip the mounting bracket over the screw  
head and tighten the 50mm screws. The unit should now be secure.  
165-072011 THCD-100 Power Supply  
Page 7 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. FEATURES  
3.1.Introduction  
This section describes the principles used in the THCD-100. It details the various settings for the inputs and  
the outputs including setpoint control. It does not detail any servicing or calibration procedures.  
3.2.Analog Inputs  
The THCD-100 has two analog voltage input channels. The main input can be configured to accept any  
input voltage from 0V to a full scale of up to 10V. This input is calibrated using a straight line fit method and  
then can be displayed on the THCD with an applied range. The secondary input is used to accept an  
external command setpoint and has a non-configurable full scale of 5V. This input is used when the setpoint  
is set to slave source.  
The main input channel has a number of associated setup parameters:  
a) Range  
Each device attached to the THCD will be outputting a voltage that represents some form of units and  
the range is used with the full scale to determine what the user sees on the screen for any input voltage.  
b) Full Scale Voltage  
Each device attached to the THCD will have a full scale voltage which is the maximum voltage the  
device should output under normal operating conditions. Typical values are 10V, 5V and 1V. The full  
scale is used with the range to determine what the user sees on the screen for any input voltage. The  
THCD will show the over range error (“RANGE!”) for any channel whose voltage input is more than  
15% over the full scale voltage set for that channel.  
c) Units String  
To inform the user of what units are being dealt with, each channel has an associated units string. This is  
a free form text field (maximum 5 characters) that is displayed on the right side of the Main screen.  
Example A device has a full scale voltage of 10V and a range of 100 mbar.  
If the device now outputs 10V, the THCD-100 display will nominally show 100 (pending input  
calibration). If the device outputs 5V, the THCD-100 will show 50, and so on.  
3.3.Analog Outputs  
There are two analog outputs on the THCD-100, both voltage outputs. One is for setpoint control with its  
own set of setup parameters (see next section) and the other is a simple retransmission of the main input  
voltage and has no user setup control. Typically this output can be used for data logging purposes or for  
providing an external setpoint control to another THCD or similar instrument.  
3.4.Setpoint Control  
As mentioned above, there is one analog output that is used for command setpoint control that can be used  
to operate a mass flow controller or similar. The setpoint command output always has a voltage full scale of  
5V.  
Example If the input channel is setup for a 100 slpm 5V full scale device, and the setpoint value is 10.0  
165-072011 THCD-100 Power Supply  
Page 8 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(assuming setpoint is in AUTO mode and not a slave source), then the output voltage of the setpoint would  
be 0.5V.  
The setpoint does have some configuration parameters that can be used to alter the function of the setpoint  
control and they are detailed as follows:  
a) Setpoint Mode  
The setpoint can be configured in one of three modes – AUTO, OPEN and CLOSED.  
In AUTO mode the setpoint output is dependant on the setpoint source and value  
settings (see below).  
In OPEN mode the setpoint outputs a voltage greater than the full scale of the  
device. Given that the setpoint command full scale is 5V, the output in this mode is 7V.  
In CLOSED mode the setpoint outputs a voltage less than the minimum output voltage  
of most devices, the setpoint output voltage is –0.25V.  
b) Setpoint Source  
The setpoint source dictates where the setpoint value comes from, assuming the mode is set to  
AUTO (see above). This can be one of two possibilities – Internal or Slave.  
Internal source – the setpoint uses the value set internally via the ‘& switches or via external  
communications.  
Slave source – in this case the setpoint uses a percentage of an externally produced value. The  
percentage is set via the ’&‘switches and the external value comes from the secondary input  
channel (as mentioned 5.2 in above)  
3.5.User Re-zero  
Over time it is possible that the input may ‘drift’ slightly due to various conditions (temperature changes,  
etc.). As such it may be necessary for the channel to be re-zeroed by the user.  
The user re-zero function is provided for this task. It simply takes the current reading (sampled and averaged  
over 3 seconds) and uses that as an additional offset for the channel in question, subtracting the value from  
all subsequent readings. Ensure any process value to be zeroed, is in fact truly zero before performing this  
function. This would mean isolating flow devices or fully pumping a pressure device.  
Note that the user re-zero via serial communications also provides the facility for clearing any user re-zero  
value that may have already been set. This should be used before any input calibration is performed to  
ensure that the calculated calibration points are not distorted by the user zero offset.  
3.6.Adaptive Filtering  
The THCD-100 includes an adaptive averaging filter on the display and serial communication readings  
output to aid in ‘smoothing out’ unwanted ‘noise’ on the displayed readings.  
a) Operational Band  
The filter can be configured to only operate within a certain band, meaning that excursions between  
subsequent readings that fall outside that band are shown as real (and not filtered) readings. The  
band is configurable between 0.01% and 1.00% of the full scale range of the input channel. It can  
also be set to OFF, which means that no filtering is performed (i.e. every reading is a ‘real’ one) and  
can also be set to ON which means that filtering is always shown no matter how big of an excursion  
occurs.  
165-072011 THCD-100 Power Supply  
Page 9 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
b) Filter Size  
The size of the filter is currently configurable between 0 and 6 seconds and simply indicates the  
amount of readings that are taken and buffered to calculate and show the averaged reading. A filter  
size of 0 will naturally turn off the filtering, irrespective of the band setting.  
It should be noted that this type of adaptive filtering can cause readings to be displayed abnormally,  
particularly if the buffer size is one of the larger available settings, where the displayed readings seem to  
‘jump about’ as the filter is disabled and then re-enabled when the data excursions fall into the preset band.  
This is because outside of the band, the display will show the ‘real’ unfiltered readings (whilst constantly  
taking an average in the background). When the excursions fall into the band then the display will switch to  
showing the filtered data. If this data is quite significantly slugged (due to a large buffer size) then the display  
will appear to nearly reach the new data value (due to ‘real’ values being shown) and then jump back to the  
filtered data which could be several seconds behind. Because of this, the setting of a buffer size greater than 5  
seconds will automatically change the band to ON, meaning that filtered data will always be shown.  
Selecting different filter band/size combinations will reduce these anomalies, turning the band OFF and ON  
will stop this anomaly.  
165-072011 THCD-100 Power Supply  
Page 10 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Front Panel Operation  
4.1.Introduction  
The front panel consists of a 112x16 dot matrix VFD display with 6 membrane key switches below it. Each  
switch has legends on and above or below it to indicate its function under different conditions, although  
there are certain common functions to each screen, as detailed below.  
For the purpose of the following subsections, the switches shall be  
referred to using the legends on the buttons.  
4.2.Common Switch Functions  
The switches on certain types of screens and in certain modes have the same functionality as detailed here:  
a. Menu Screens  
On all menu screens, the ‘’&‘’ switches are used to navigate through the menu options. The  
Func’ switch is used to go back to the parent menu, except on the Main Menu screen where it goes  
back to the normal display mode. The ‘Zero’ switch doubles up as an Enter switch and this actions  
the chosen menu selection. This may be to go to a submenu or to an editable or information screen  
(see HMI breakdown for more details).  
b. Editable Screens  
On any screen that has editable fields, switches ‘’&‘’ move the selection between the possible  
editable fields. ‘Func’ goes back to the parent screen which is usually the menu that was used to get  
to the screen in question. ‘Zero’ activates edit mode (see below).  
c. Edit Mode  
When in edit mode, the edit cursor appears on the first character in the editable field (the cursor is a  
horizontal bar below the character being edited). If the field is bigger than 1 character then switches  
’&‘’ are used to move left and right across the field. If the field is only one character (e.g. a  
Yes/No selection) or has a rolling selection (e.g. for selection of a percentage) then ‘’&‘’ do  
nothing. Switches ‘’&‘’ are used to change the character being edited within the valid range  
(field position dependant). Func’ cancels edit mode and returns the field to its previous value.  
Zero’ accepts any changes to the field and also comes out of edit mode.  
165-072011 THCD-100 Power Supply  
Page 11 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.3.Front Panel Screens  
Here follows a complete breakdown of all THCD-100 screens:  
a) Main Screen (not changing setpoint)  
The main screen comes in two forms. The first shows the main channel input data (with  
range, fullscale and calibration applied) and the associated units string. If an input is over  
range (more than 15% above full scale voltage) then the data value is replaced with the  
over range error – “RANGE!”. This screen also shows the current mode of the setpoint, if  
not in AUTO (i.e. OPEN or CLOSE). If either of the relays have been tripped then they  
are also indicated as such on the far right of the screen.  
Switch Functions:  
Func  
No function *  
Go to Main Menu screen  
Enable setpoint mode override, active for 2 seconds  
Change setpoint mode to AUTO  
(override mode)  
Change setpoint value (main screen changes as below)  
Change setpoint mode to OPEN  
(override mode)  
(override mode)  
Zero  
Change setpoint value (main screen changes as below)  
Change setpoint mode to CLOSE  
Perform input rezero (if held for 3 seconds)  
* By default, the Func switch does nothing while this screen is displayed.  
b) Main Screen (changing setpoint)  
The other form of main screen is shown when changing the live setpoint (via an initial press of the  
’ or ‘’ switches). All the information shown on the first main screen is shown here (albeit some  
items are resized or repositioned) but in addition the current setpoint value is also shown. Note that  
this screen is not permanent – if the ‘’ or ‘’ switches are not pressed at all for 5 seconds, then the  
main screen reverts to the first variety shown previously.  
Switch Functions:  
Func  
No function*(See (a) above)  
Go to Main Menu screen  
Enable setpoint mode override, active for 2 seconds  
Change setpoint mode to AUTO  
Change setpoint value (hold for continuous change)  
Change setpoint mode to OPEN  
(override mode)  
(override mode)  
(override mode)  
Zero  
Change setpoint value (hold for continuous change  
Change setpoint mode to CLOSE  
Perform input rezero (if held for 3 seconds)  
165-072011 THCD-100 Power Supply  
Page 12 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
c) Main Menu Screen  
The Main Menu screen provides a path to select screens that are used for the configuration of the  
THCD-100 – the setpoint (value, mode, source), input (range, full scale, units string), digital  
communications, display filtering and relays can all be configured from screens selected via the main  
menu. It is also possible to reset settings to their factory defaults from the Main Menu screen. Only  
one option is shown at any one time, in the order shown above. Pressing Enter (Zero) accesses the  
next level.  
Descriptions of successive levels are described below.  
Switch Functions:  
Func  
Go back to Main screen  
Go to previous menu selection  
Go to next menu selection  
No function  
No function  
Zero  
Go to screen associated with menu selection  
d) Setpoint Screen  
From the setpoint screen you can edit the value of the command setpoint as well as change the mode  
(AUTO, OPEN, CLOSED) and the source (INTERNAL, SLAVE). If SLAVE is selected then the  
percentage sign (%) appears next to the SP value. For more information on the command setpoint,  
see the Principles section 5.4.  
Switch Functions:  
Func  
Go back to Main Menu screen  
Select previous character when in edit mode  
Select next character when in edit mode  
Move edit field selection up  
Move edit field selection down  
Zero  
Go to edit mode for currently selected field  
165-072011 THCD-100 Power Supply  
Page 13 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input Screen  
This screen allows you to change the input channel range and fullscale voltage (see the Principles  
section (5.2) for more information on this) and also assign a units label (up to 5, non-scaling  
characters) for the channel data. The number of decimal places selected on the range field dictates  
the number of decimal places shown for the input channel data on the Main screen. To reduce the  
number of decimal places, change the number under the edit cursor to a decimal point and the  
system will auto format the field when ‘Zero’ is pressed.  
Switch Functions:  
Func  
Go back to Main Menu screen  
Select previous character when in edit mode  
Select next character when in edit mode  
Move edit field selection up  
Zero  
Move edit field selection down  
Go to edit mode for currently selected field  
e) Filtering Screen  
This screen allows the user to modify the band and buffer size of the adaptive filter. The filter only  
applies to the display and serial communications output readings. The band is represented as a  
percentage of FS and can be set between 0.00% (OFF) and 1.00% and also to ON. The buffer size  
allows settings between 0 and 6 seconds of filtering. For more information on the adaptive filter, see  
the Principles section (5.6)  
Switch Functions:  
Func  
Go back to Main Menu screen  
Select previous character when in edit mode  
Select next character when in edit mode  
Move edit field selection up  
Move edit field selection down  
Zero  
Go to edit mode for currently selected field  
165-072011 THCD-100 Power Supply  
Page 14 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
f) Comms Menu Screen  
The Comms Menu screen provides options for setting the serial communication’s baud rate, address  
and RS232/RS485 protocol.  
Switch Functions:  
Func  
Go back to Main Menu screen  
Go to previous menu selection  
Go to next menu selection  
No function  
Zero  
No function  
Go to screen associated with menu selection  
g) Serial Screen  
The main configuration item on the serial screen allows you to select the baud rate of the THCD-  
100 for serial communication. The allowed baud rates are 9600, 19200 & 57600. There is also an  
address field that is mainly used for identifying a particular unit if several are connected together on  
an RS485 network. All commands sent to the unit must be prefixed with the address letter (see  
section 5.2 for more information). Valid letters are ‘a’ to ‘h’. The other editable field is used to tell  
the THCD-100 whether it is on an RS485 serial network.  
Switch Functions:  
Func  
Go back to Main Menu screen  
Select previous character when in edit mode  
Select next character when in edit mode  
Move edit field selection up  
Move edit field selection down  
Zero  
Go to edit mode for currently selected field  
165-072011 THCD-100 Power Supply  
Page 15 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
h) Relay Screen  
This screen allows you to set the trip point for when the relay switches. The relay also contains  
some hysteresis to stop it ‘chattering’. This is configurable from this screen as well and is  
represented as a percentage of full scale of the input channel, with limits of 0.0% (i.e. no  
hysteresis) to 10.0%. If the second relay option is factory fitted then the above fields are  
duplicated for the second relay and are changed to Trip 1, Hyst 1 & Trip 2, Hyst 2 as  
appropriate.  
Switch Functions:  
Func  
Go back to Main Menu screen  
Select previous character when in edit mode  
Select next character when in edit mode  
Move edit field selection up  
Move edit field selection down  
Zero  
Go to edit mode for currently selected field  
i) Factory Defaults Confirmation Screen  
If Factory Defaults is selected from the Main Menu, this screen appears to ask for confirmation of  
the requested function. If confirmed then the screen changes to show a percentage of  
completion of the factory defaults and once completed the THCD-100 will restart.  
Switch Functions:  
Func  
Go back to Main Menu screen  
If held, starts factory default when Zero is pressed  
No function  
No function  
Zero  
No function  
Starts factory default if is held as well.  
165-072011 THCD-100 Power Supply  
Page 16 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. SERIAL COMMUNICATION  
5.1.Introduction  
The THCD-100 allows full control and feedback via serial communication protocols (RS232 or RS485).  
Calibration, in fact, all controls available through the front panel, are available as commands sent using serial  
communications.  
Serial communications is available using an RS232 or RS485 link between the client (typically a PC) and the  
THCD. A baud rate must be configured, along with the selection of protocol. Valid baud rates are 9600,  
19200 & 57600. Additional communication setup parameters (e.g. parity, stop bits, etc.) are always the same  
and nonconfigurable. By default the THCD is setup to communicate via RS232 with the following  
communication parameters:  
Baud: 57600  
Data bits: 8  
Parity: None  
Stop bits: 1  
Handshaking: None  
5.2.Command/Query Format  
The format of the commands and queries is common between all available communication protocols. That  
is, the address must be included whether using RS232 or 485.  
Format: accc[?] [p][,p][,p]  
… where ‘a’ = address, ‘ccc’ = command/query, ‘?’ = query identification; ‘p’ = parameter (separated by  
commas if more than one)  
The first letter to be entered on the command line, before any command or query, is the address identifier.  
This is a letter between a & h and is particularly important when communicating via RS485, if there are  
several units on the same RS485 network. On RS232 this address doesn’t really mean anything because  
RS232 is single point to point communication. However as mentioned above, the address letter is left in (and  
therefore required) between communication methods. By default the THCD has the letter ’a’ for its address.  
Most commands are identified by three letters (although this is not always the case) and a query is identified  
by following the command letters with a ‘?’ character. For all commands with parameters, you must separate  
the command from its parameters with a space, and separate multiple parameters with commas.  
Table 1 in Appendix A summarizes all commands and queries available. Note that where a command is also  
a query, you do not enter any parameters for the query ‘version’ unless explicitly stated in the table.  
For everything sent through the communications port, the THCD will reply by informing the user of the  
command/query it has just received, along with any parameters identified, any data requested (if it was a  
valid command/query) and a line indicating whether the command was accepted. This reply block is the  
same for anything sent to the THCD and is as follows . (‘’ indicates a carriage return/line feed pair):  
165-072011 THCD-100 Power Supply  
Page 17 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
For a command:  
For a query:  
*<a>*:<cmd>;<params>↵  
*<a>*:<cmd>;<params>↵  
!<a>!<response>!↵  
<data>↵  
!<a>!<response>!↵  
… where:  
<a> is the address letter  
<cmd> is the command/query (query identified by a ‘?’ after the three chars)  
<params> is the comma separated parameter list, if any  
<response> is the acceptance indicator.  
<data> is the data requested if a query was sent.  
NOTE: The data usually consists of some identification string, in CAPS, followed by a colon (:) and then  
the actual data. Some queries return multiple lines of information. See individual commands for details as  
appropriate.  
The acceptance indicator is a single character as follows:  
‘o’ = OK: if the command/query was recognized and accepted;  
‘b’ = BADCMD: if the command/query was not recognized or has incorrect/invalid parameters  
‘e’ = ERROR: if there was some internal communications error.  
‘w’ = BUSY: if the communications port is currently busy.  
5.3.Main Commands  
a. Output Current Readings – ‘r’  
This command tells the THCD to get the current reading of the input channel data. If the input  
voltage of the channel is more than 15% over the full scale voltage set for the channel, the  
<reading> below will be replaced with the over range error – “RANGE!”.  
The readings output format is as follows with the channel data followed by a setpoint mode  
indication, separated by a semicolon:  
READ:<reading>;<setpoint mode>↵  
The <setpoint mode> is a single number that represents the current mode of the setpoint: 0 =  
AUTO, 1 = OPEN, 2 = CLOSED. This number is mainly included for use with the THCD-100’s  
frontend software and is mainly useful when outputting repeated readings, where it effectively  
provides a constant update of the current mode of the setpoint.  
b. Output Current Readings Repeatedly – ‘rp’  
The THCD is also able to report the current readings repeatedly, at a given interval. This command  
starts and stops this repeat activity by setting the parameter accordingly:  
0 = Repeat off  
1 = Repeat every 100ms (not allowed if using a baud rate < 57600)  
2 = Repeat every 500ms (not allowed if using a baud rate < 57600)  
3 = Repeat every second  
4 = Repeat every minute The timing commences from the point at which the ‘rp’ command is  
entered. Use this command with care on a RS485 multidrop network! ‘r’ is recommended.  
NOTE for 100ms repeat readings: The data output is effectively lots of small data packets. On  
TCP networks, this can cause problems when communicating with front end software running on  
MS Windows. It has been found that the Winsock control, commonly used in TCP communications  
development on Windows, does not handle lots of small data packets very well, coming in at speeds  
of around 10Hz or faster. As such, it has been decided that, when 100ms repeated output is used,  
165-072011 THCD-100 Power Supply  
Page 18 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
the data will be taken every 100ms and buffered up, to be sent out every 500ms in blocks of 5  
readings. This has proven to be a more robust output method.  
c. Setpoint Value – ‘spv’  
This command is used to set the value of the command setpoint. The command takes one parameter  
which is the actual value to set.  
The current value of the setpoint can be retrieved using the query ‘spv?’. Returned data is:  
SP VALUE: <setpoint value>  
d. Setpoint Mode – ‘spm’  
This command sets the mode for the command setpoint. The command takes one parameter which  
is a number between 0 & 2, identifying the mode, as follows:  
0 = AUTO mode  
1 = OPEN mode  
2 = CLOSED mode  
For more information on setpoint modes, see the Principles section 5.4(a).  
The current mode of the setpoint can be retrieved by using the query ‘spm?’. The returned data is:  
SP MODE: (<x>) <mode string>  
… where …  
<x> = 0, 1, or 2 and <mode string> = AUTO, OPEN, or CLOSED respectively.  
e. Setpoint Source – ‘sps’  
The setpoint source can be an internally set value or a percentage of the secondary THCD input  
(known as slave input). This source is set using this command and as with the other setpoint  
commands, it takes just one parameter which is as follows:  
0 = Internal source  
1 = Slave source  
For more information on setpoint sources, see the Principles section 5.4(b). The current source of  
the setpoint can be retrieved by using the query ‘sps?’. The returned data is:  
SP SOURCE: (<x>) <source id>  
…where…  
<x> = 0 or 1 as above, <source id> = string identifier for <x> : INTERNAL or SLAVE.  
f. Setpoint Initial Value – ‘siv’  
The setpoint value set with the ‘spv’ command (above) is not nonvolatile. This command can be  
used to set an initial setpoint value on system startup. It takes one parameter which is the initial  
setpoint value.  
The current value of the setpoint can be retrieved using the query ‘siv?’. Returned data is:  
SP INIT VAL: <initial setpoint value>  
165-072011 THCD-100 Power Supply  
Page 19 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
g. Setpoint Initial Mode – ‘sim’  
As with the value, the setpoint mode set using the ‘spm’ command (above) is a volatile setting. Again  
this command provides a way of setting an initial startup mode for the setpoint on power up of the  
THCD. It takes one parameter which is a number between 0 & 2, identifying the initial mode, as  
follows:  
0 = AUTO mode 1 = OPEN mode 2 = CLOSED mode  
The current mode of the setpoint can be retrieved by using the query ‘spm?’. The returned data is:  
SP INIT MODE: (<x>) <mode string>  
… where …  
<x> = 0, 1, or 2 and <mode string> = AUTO, OPEN, or CLOSED respectively.  
5.4.Serial Commands  
a) Serial Baud Rate – ‘bra’  
The serial baud rate (BR) can be set using this command. Any value can be entered as the parameter  
and the THCD decides on the actual baud rate using the following criteria:  
Parameter value (BR) Actual baud rate  
BR < 14400  
14400 BR < 28800  
BR 28800  
9600  
19200  
57600  
When changing baud rate using serial commands, the THCD will give the usual response to the  
command at the current baud, then change baud rate and give the !<address>!o!response at the  
new baud rate. Host front ends can therefore monitor for this to confirm that the baud rate has  
changed. If this command is used with  
To retrieve the current baud, use the query ‘bra?’. Returned data is:  
BAUD: <baud rate>  
b) Serial Communication Protocol – ‘pro’  
This command switches the serial protocols between RS232 & RS485. The parameter passed should  
be one of the following:  
0 = RS485  
1 = RS232  
To retrieve the current protocol setting, use the query ‘pro?’. Returned data is:  
PROTOCOL: <protocol>  
… where  
<protocol> = 0 or 1 as above.  
c) Serial Communication Address – ‘add’  
As mentioned previously (section 4.2), all commands are prefixed with an address letter. This  
command allows you to change that address. Valid addresses are a through h. When the THCD  
replies to the command, the first line will contain the old address and the response line will contain  
165-072011 THCD-100 Power Supply  
Page 20 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
the new address. All subsequent communications must use the new address.  
The current communication address can be obtained using the query ‘add?’. The returned data is:  
ADDR: <address>  
5.5.Channel Setup Commands  
a) Input Channel Units String – ‘uiu’  
The input channel has an associated freeform text field that can be used to identify the units being  
used. This command is used to set this and its only parameter is a max of 5 characters for the units  
string.  
The current units string can be retrieved using the query ‘uiu?’. The data returned is as follows:  
INPUT UNITS STR: <units>  
… where… <units> = the exact units string as displayed on the front panel  
b) Input Channel Range – ‘uir’  
Use this command to set the range of the input channel. The single parameter is the actual value for  
the range. The number of decimal places passed in the range value dictates the number of decimal  
places shown for the main data output (up to a max of 4 decimal places  
– any more will get chopped off)  
The current range of the input channel is retrieved using the query ‘uir?’.  
The data returned is as follows:  
INPUT RANGE: <range>  
… where…  
<range> = the range of the channel (engineering units)  
c) Input Channel Full Scale – ‘uif’  
This command is used to set the full scale voltage for the input channel. The single parameter is the  
actual full scale value. The full scale voltage setting for the input channel is retrieved using the query  
‘uif?’. The data returned is as follows:  
INPUT FULLSCALE: <fs>  
… where…  
<fs> = the full scale voltage of the channel.  
For more information on the use of input range and full scale, see the Principles section 5.2.  
5.6.Filtering Commands  
a. Filter Band – ‘flb’  
This command is used to set the point at which the adaptive filtering kicks in. It takes one parameter  
which is the percentage of full scale of the input. Valid values are between 0.01% and 1.00%.  
Additionally the parameter can be OFF which will mean the filtering will never kick in, or ON which  
means the filtering is always used. Note that if the filter size is set to greater than 5 seconds, the filter  
band command will return a Bad Command acknowledgement because the filter is always on in this  
165-072011 THCD-100 Power Supply  
Page 21 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
situation and it therefore makes no sense to attempt to change the band setting. For more  
information on this see the Principles section (5.6).  
To retrieve the current filter band, use the query ‘flb?’. The returned data is:  
FILTERING BAND: <band>%  
b. Filter Size – ‘fls’  
Use this command to set the size of the adaptive filtering buffer. The one parameter this command  
takes is the buffer size in seconds, between 0 and 6 seconds (integer values only).  
(0 can be used to effectively turn the adaptive filtering off).  
To retrieve the current filter size, use the query ‘fls?’. The returned data is:  
FILTERING SIZE: <x> sec  
… where  
<x> = number of seconds of filtering  
[Note if current size is 0 then returned data is: FILTERING SIZE: 0 (NO FILTER) ]  
5.7.Relay Control Commands  
a) Relay Trip Point – ‘rlt’  
This command sets the point at which the relay switches from one state to the other. When the  
source data is below this point then the relay is CLOSED, so therefore when the source data rises  
above this point then the relay is OPEN. The command takes two parameters – the first is the relay  
whose trip point is to be set (1 or 2) and the second is the trip point in engineering units, as  
configured with the Input Channel range command (see 5.2(a)). Note that the first parameter is  
always required even if the second relay has not been factory fitted.  
The current trip point values can be retrieved using the query ‘rlt?’. The returned data is one line for  
each relay, as follows:  
RELAY <x>,TRIP POINT: (comma req’d, no space)  
… where ...  
<x> = the relay number and TRIP POINT is in terms of the set full-scale display value.  
b) Relay Hysteresis – ‘rlh’  
This command configures the hysteresis setting of the relay. It takes two parameters – the first if the  
relay in question, the second is the percentage hysteresis, represented as a percentage of full scale of  
the source channel, with limits of 0.0% (i.e. no hysteresis) to 10.0%.  
The current hysteresis setting can be retrieved using the query ‘rlh?’. The returned data is one line  
for each relay, as follows:  
RELAY <x>,HYSTERESIS: (comma req’d, no space)  
… where ...  
<x> = the relay number and HYSTERESIS is in percent.  
165-072011 THCD-100 Power Supply  
Page 22 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Other Commands  
a) User Input Re-zero – ‘irz’  
To perform a user input re-zero function, use this command. The command needs no parameters.  
To find the current user re-zero values for all input channels, use the query ‘irz?’. The data returned  
is as follows:  
CH<x> REZERO: <re-zero>  
…where…  
<x> = the input channel number <re-zero> = the current re-zero value  
b) Date of Last Factory Calibration – ‘dlc?’  
This query is used to retrieve the date of the last factory calibration run. The data is returned as  
follows:  
LAST CAL DATE: <yymmdd>  
…where…  
yy = year, mm = month & dd = day.  
E.g. 051201 means the last calibration run was 1st December 2005.  
c) Retrieve All Settings – ‘ras’  
As well as issuing some of the above commands as queries to retrieve individual settings, this  
command has been provided which will return a comma separated list of most of the settings for the  
input channel.  
The exact settings provided by this command are detailed in Table 3 in Appendix C.  
165-072011 THCD-100 Power Supply  
Page 23 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. SERVICE & CALIBRATION  
6.1.Service  
There are no user serviceable parts inside the instruments. Should any difficulties be encountered in the use  
of the THCD-100, it is recommended that you contact Hastings Instruments for advice and instructions.  
6.2.Analog Inputs  
There is no user calibration for the THCD-100. It is recommended that the instrument be returned to  
Hastings Instruments whenever such service is required.  
There are no user adjustments in the instrument, indeed, the presence of lethal voltages within the  
instrument means that the user is strictly forbidden from removing the covers without invalidating Hastings  
Instrument’s obligations under warranty.  
6.3.Cleaning  
A dirty instrument may be wiped clean with a soft cloth that has been sprayed with a mild foaming cleaner  
then wiped dry immediately.  
Under no circumstances should the instrument be wetted directly or  
left damp  
6.4.End of Life  
Hastings Instruments complies with the WEEE legislation and is registered as a manufacturer and importer  
of Electrical and Electronic Equipment. Once this equipment reaches its end of life, or is no longer required  
it may be returned to Hastings Instruments to be reused or recycled in accordance with the legislation. Please  
contact Hastings Instruments for full details. Terms and conditions apply.  
165-072011 THCD-100 Power Supply  
Page 24 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Appendices  
7.1.Appendix A  
Table 1 – THCD-100 Command Set Summary  
Purpose  
Cmd Query  
Parameters  
Notes  
Output current readings  
r
Output current readings  
repeatedly  
rp  
0 - 4  
P1: 0 = Repeat off  
1 = Repeat every 100ms  
2 = Repeat every 500ms  
3 = Repeat every second  
4 = Repeat every minute  
(1 & 2 not allowed when using serial  
communication at baud < 57600)  
Setpoint value  
Setpoint mode  
spv?  
spv  
#
P1: Real value to set  
spm  
spm?  
0 - 2  
P1: 0 = AUTO mode  
1 = OPEN mode  
2 = CLOSED mode  
Setpoint source  
sps  
sps?  
0 - 1  
P1: 0 = Internal source  
1 = Slave (External) source  
Setpoint initial value  
Setpoint initial mode  
siv  
siv?  
#
P1: Real value to set  
sim  
sim?  
0 - 2  
P1: 0 = AUTO mode  
1 = OPEN mode  
2 = CLOSED mode  
Retrieve all settings  
Serial baud rate  
ras  
bra  
Comma separated string returned  
See Table 3 for actual settings  
bra?  
pro?  
9600, 19200,  
57600  
BR< 14400 = 9600  
14400 BR < 28800 = 19200  
BR 28800 = 57600  
Serial protocol  
pro  
0 or 1  
P1: 0 = RS485  
1 = RS232  
Communications address  
Input channel units str  
Input channel range  
add  
uiu  
uir  
add?  
uiu?  
uir?  
a through h  
P1: Max length 5 chars  
ccccc  
#
P1: Real value for range to set  
P1: Real value for full scale to set  
P1: Band percentage (0.011.00) or OFF, or ON  
P1: Buffer size in seconds  
Input channel full scale  
Adaptive filter band  
uif  
flb  
uif?  
flb?  
#
#
Adaptive filter size  
fls  
rlt  
fls?  
rlt?  
0 - 6  
Relay switch threshold  
1 - 2,  
#
P1: Relay number  
P2: Real number for switching point  
Relay hysteresis  
rlh  
irz  
rlh?  
irz?  
1 - 2,  
0.01 - 0.0  
[0]  
P1: Relay number  
P2: Hysteresis percentage value  
P1 (optional): clears current user input rezero  
setting  
User input rezero function  
Date of last calibration  
dlc?  
Returned date format: yymmdd  
165-072011 THCD-100 Power Supply  
Page 25 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.2.Appendix B  
Table 2 – Factory Defaults  
Setting  
Default Value  
Additional Information  
Repeat Rate (for output readings)  
Serial Communications Baud  
Serial Communications Protocol  
Communicationss Address  
Input Channel Units String  
Input Channel Range  
0
Repeat off  
57600  
1
RS232  
a
“”  
No units strings set  
10.00  
10.000  
2
Input Channel Full Scale  
Input Channel Display Precision  
Input Channel User Rezero  
Setpoint Value  
2 decimal places  
No rezero set  
0.0  
0.0  
0.0%  
0
Setpoint Slave Value  
Setpoint Source  
Internal source  
Setpoint Mode  
0
AUTO mode  
Date of Last Factory Calibration  
Adaptive Filter Band  
010101  
0.2%  
2
Adaptive Filter Buffer Size  
2 seconds  
Relay Trip Point (for both relays, if factory fitted)  
Relay Hysteresis (for both relays, if factory fitted)  
10.0  
2.0%  
165-072011 THCD-100 Power Supply  
Page 26 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.3.Appendix C  
Table 3 – Settings in ‘ras’ command string  
Type and size  
Description  
Input channel units string  
5 char string  
8 char string representation of a float  
8 char string representation of a float  
8 char string representation of a float  
8 char string representation of a float  
1 byte (02 as for spm? query)  
range  
full scale  
Setpoint value  
slave value  
mode  
1 byte (01 as for sps? query)  
source  
initial value  
8 char string representation of a float  
8 char string representation of a float  
1 byte (02 as for sim? query)  
initial slave value  
initial mode  
Adaptive filter band  
size  
4 char string representation of a float  
1 byte (06 as for fls? query)  
Relay 1 trip point  
hysteresis value  
8 char string representation of a float  
4 char string representation of a float  
Relay 2 trip point (if factory fitted)  
hysteresis value (if factory fitted)  
Last factory calibration date  
8 char string representation of a float  
4 char string representation of a float  
6 char string (format as for dlc? query)  
165-072011 THCD-100 Power Supply  
Page 27 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. WARRANTY  
8.1.Warranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment to be free  
from defects in material and workmanship. This warranty does not apply to defects or failures resulting from  
unauthorized modification, misuse or mishandling of the product. This warranty does not apply to batteries  
or other expendable parts, or to damage caused by leaking batteries or any similar occurrence. This warranty  
does not apply to any instrument which has had a tamper seal removed or broken.  
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to  
fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential  
damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if Hastings  
Instruments determines, in good faith, that it is defective in materials or workmanship during the warranty  
period. Defective instruments should be returned to Hastings Instruments, shipment prepaid, together  
with a written statement of the problem and a Return Material Authorization (RMA) number.  
Please consult the factory for your RMA number before returning any product for repair. Collect freight will  
not be accepted.  
8.2.Non-Warranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form and a  
written description of the problem with the instrument. If the repair cost is higher, you will be contacted for  
authorization before we proceed with any repairs. If you then choose not to have the product repaired, a  
minimum will be charged to cover the processing and inspection. Please consult the factory for your RMA  
number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
TOLL FREE  
FAX  
(757) 723-6531  
1-800-950-2468  
(757) 723-3925  
E MAIL  
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments web  
site.  
165-072011 THCD-100 Power Supply  
Page 28 of 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Technicolor Thomson CD Player DCD SA1 User Manual
Timex Watch W 46 User Manual
Toshiba DVD Player SD 190ESE User Manual
Toshiba IP Phone CIX AG NP VA User Manual
Toshiba Telephone DK24 User Manual
TRENDnet Computer Accessories TKEX4 User Manual
Tricity Bendix Microwave Oven im900 User Manual
VeriFone Video Gaming Accessories Vx670 User Manual
Vermont Casting Stove CDW244 User Manual
Walker Indoor Furnishings V42Y79B User Manual