HASTINGS VACUUM GAUGES
MODEL NV-100
TELEDYNE BROWN ENGINEERING
HASTINGS INSTRUMENTS
P.O.BOX 1436
HAMPTON, VIRGINIA 23661 U.S.A.
TELEPHONE (804)723-6531
FAX (804)723-3925
TWX (710)882-0085
EASYLINK: 62300280
© May 1994
103A - PAGE 1
Download from Www.Somanuals.com. All Manuals Search And Download.
GENERAL
The Hastings Model NV-100 Vacuum Gauge is a complete self contained instrument with all circuits mounted on the
rear of the meter case. A remote type DV-100 gauge tube is used as the pressure sensor. The instrument measures
pressures over the range of 0.1 to 100 Torr utilizing the well known Hastings directly heated thermopile. Unlike other
Hastings gauges, this instrument operates in a constant temperature mode rather than constant applied voltage. The
changes in power to create a constant temperature rise above ambient are measured by the indicating meter as a
function of pressure. Corrections for gasses of differing thermal conductivities are necessary as in all thermal
conductivity type vacuum gauges.
INSTALLATION
The instrument is factory calibrated and ready for use when received. All that is necessary to place the gauge in
operation is to install the gauge tube into the system, plug in the gauge tube cable, and connect the power cord to
the AC line. Gauge tubes should be installed with the open end pointing downward so as to be self draining. On
dusty or dirty systems, a Hastings particle drop-out (protective) trap is recommended (see Product Bulletin #352).
OPERATION
The gauge is in operation once connected to the gauge tube and power. An initial check of the mechanical meter zero
may be necessary upon receipt. To do this, unplug the power cord and set (if necessary) the meter mechanical zero
adjustment the front of the meter to read "ATM". On the rear of the instrument are two calibration adjustments, they
should be correct as received. An occasional check of the ATM adjustment may be desirable. Also, anytime a new
gauge tube is replaced it is desirable to reset the ATM adjustment.
To do this, expose the tube to atmospheric pressure after the gauge is fully warmed up (about 30 minutes) and
adjust the trim potentiometer marked ATM until the needle reads ATM. Use of a trim potentiometer tuning tool will
facilitate this adjustment.
NOTE: When turned off, the meter needle rests at the right side of the meter (ATM mark).
When no tube is plugged in, the instrument reads off the scale below zero. If a broken gauge tube is plugged in it
will read off scale above ATM. Replace with a new gauge tube.
The circuit has a protective shut-down circuit. This shuts down the electronics whenever a gauge tube is not plugged
into the socket. This is to prevent burning out gauge tubes by application of excessive power surges when plugging
in or unplugging gauge tubes. A delay circuit applies power gently when connected. Pins 1 and 2 of the gauge tube
activate the circuit and act as the interlock switch. Thus the instrument will not operate unless the correct gauge tube
is used. Troubleshooting the circuit on a bench cannot be accomplished unless these two pins are jumpered at the
gauge tube socket, or a DV-100 gauge tube is plugged into the socket.
TROUBLE SHOOTING
If troubles develop, always try a new gauge tube before suspecting other problems. To check gauge tubes for
continuity of thermocouples, first make sure that the gauge tube is at atmospheric pressure, then using an
ohmmeter check for continuity between pins 3,5 and 7. Approximately 10 ohms resistance is read between all
combinations of these three pins on a good tube. If open, replace the tube. Never check while under vacuum as the
ohmmeter may burn out the gauge tube. Pins 1 and 2 should show zero ohms. Trouble shooting the electronics is
difficult because this is a closed loop feed-back circuit. If circuit problems develop, it is best to return it to the
factory for repair.
103A - PAGE 3
Download from Www.Somanuals.com. All Manuals Search And Download.
CALIBRATION
Calibration is relatively easy with this gauge, since a pressure of only 10 microns Hg (10 millitor) or lower is all that
are required. It need not be known accurately as long as it is this low or lower. A good clean, dry, compound
mechanical vacuum pump can achieve this pressure if the oil is clean and fresh and the volume of the system small.
1. Check mechanical meter zero with the gauge turned off. Set to read ATM if necessary with screw adjust in
lower left center of meter
2. Turn on and allow about 30 minutes warm-up. With gauge tube at atmospheric pressure, set the trim
potentiometer in the rear marked "ATM" until the meter needle reads ATM.
3. Pump down to 10 microns Hg or lower. Allow sufficient time to outgas the tube after exposure to ATMin (2)
above. Set "ZERO" adjust trim potentiometer in the rear until the meter reads "0" pressure.
Calibration is now complete. Repeat 2 and 3 if desired for a final trim.
EXTENSION CABLES
A maximum of 12 ft extension may be used provided recalibration of the ATM and span adjustments are
accomplished. Extension cables are of the type "OMS-(length)-OFS" and are available from Hastings Instruments.
A 5 wire cable of no less than #18 wire is required. Hastings provides an extension of 12 ft.
103A - PAGE 4
Download from Www.Somanuals.com. All Manuals Search And Download.
103A - PAGE 5
Download from Www.Somanuals.com. All Manuals Search And Download.
103A - PAGE 6
Download from Www.Somanuals.com. All Manuals Search And Download.
103A - PAGE 7
Download from Www.Somanuals.com. All Manuals Search And Download.
103A - PAGE 8
Download from Www.Somanuals.com. All Manuals Search And Download.
Service Form:
TBE / Hastings Instruments
RMA#
804 Newcombe Avenue
(contact factory)
Hampton, VA 23669 U.S.A.
Telephone (804)723-6531 Fax (804)723-3925
Model No.
Name
Company
Address
Serial No.
P.O. No.
Phone
Date
Zip
City
State
Describe problem:
Show a block diagram of your system including unit inlet and outlet pressures:
Calibration range
Power line voltage
Ambient temperature
Other
Gas
Cable length
Variation
Gas Temperature
(If special modifications have been made by user, please describe):
Application of product being returned
Has this product been used with a hazardous material?
YES
NO
If YES, please list types of gas, chemicals (common names, specific chemical), biological materials, or
other potentially harmful materials exposed to the product during its use (attach additional page if
necessary)
PRODUCTS EXPOSED TO RADIOACTIVE MATERIAL CANNOT BE ACCEPTED BY HASTINGS
INSTRUMENTS UNDER ANY CIRCUMSTANCES.
Signature of Hazards Control Officer:
For questions regarding use of the product and materials in contact with it:
Name
Phone
103A - PAGE 9
Download from Www.Somanuals.com. All Manuals Search And Download.
|