TELEDYNE
HASTINGS
INSTRUMENTS
INSTRUCTION MANUAL
HFM-301/305/306
FLOW METERS
HFC-303/307 FLOW
CONTROLLERS
I S O 9 0 0 1
C E R T I F I E D
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
1.
GENERAL INFORMATION............................................................................................................................................ 4
1.1.
FEATURES .................................................................................................................................................................... 4
SPECIFICATIONS........................................................................................................................................................... 5
OPTIONAL 4-20 MA CURRENT OUTPUT ....................................................................................................................... 6
OTHER ACCESSORIES................................................................................................................................................... 6
1.2.
1.3.
1.4.
2.
OPERATION...................................................................................................................................................................... 7
2.1.
2.2.
2.3.
2.4.
2.5.
2.6.
2.7.
2.8.
RECEIVING INSPECTION ............................................................................................................................................... 7
POWER REQUIREMENTS ............................................................................................................................................... 7
OUTPUT SIGNAL........................................................................................................................................................... 7
MECHANICAL CONNECTIONS....................................................................................................................................... 7
ELECTRICAL CONNECTIONS......................................................................................................................................... 8
4 – 20 MA CONNECTIONS........................................................................................................................................... 11
OPERATION................................................................................................................................................................ 12
RANGE CHANGING..................................................................................................................................................... 16
OUTPUT FILTER ......................................................................................................................................................... 16
CONTROLLING OTHER PROCESS VARIABLES ............................................................................................................. 16
COMMAND INPUT....................................................................................................................................................... 16
VALVE OVERRIDE CONTROL ..................................................................................................................................... 17
GAIN POTENTIOMETER .............................................................................................................................................. 17
TEMPERATURE COEFFICIENTS ................................................................................................................................... 18
2.9.
2.10.
2.11.
2.12.
2.13.
2.14.
3.
THEORY AND FUNCTION........................................................................................................................................... 19
3.1.
OVERALL FUNCTIONAL DESCRIPTION........................................................................................................................ 19
SENSOR DESCRIPTION................................................................................................................................................ 19
SENSOR THEORY........................................................................................................................................................ 19
BASE.......................................................................................................................................................................... 20
SHUNT DESCRIPTION .................................................................................................................................................. 21
SHUNT THEORY ......................................................................................................................................................... 21
CONTROL VALVE....................................................................................................................................................... 23
ELECTRONIC CIRCUITRY............................................................................................................................................ 24
INSTRUMENT PERFORMANCE ..................................................................................................................................... 24
3.2.
3.3.
3.4.
3.5.
3.6.
3.7.
3.8.
3.9.
4.
5.
MAINTENANCE.............................................................................................................................................................. 25
4.1.
AUTHORIZED MAINTENANCE..................................................................................................................................... 25
TROUBLESHOOTING ................................................................................................................................................... 25
ADJUSTMENTS ........................................................................................................................................................... 26
END CAP REMOVAL................................................................................................................................................... 27
PRINTED CIRCUIT BOARD AND SENSOR REPLACEMENT............................................................................................. 28
4.2.
4.3.
4.4.
4.5.
OUTLINE DRAWINGS .................................................................................................................................................. 29
5.1.
HFM-301................................................................................................................................................................... 29
HFC-303 ................................................................................................................................................................... 29
HFM-305................................................................................................................................................................... 30
HFM-306................................................................................................................................................................... 30
HFC-307 ................................................................................................................................................................... 31
5.2.
5.3.
5.4.
5.5.
6.
7.
WARRANTY .................................................................................................................................................................... 32
6.1.
6.2.
WARRANTY REPAIR POLICY ...................................................................................................................................... 32
NON-WARRANTY REPAIR POLICY ............................................................................................................................. 32
APPENDIX ....................................................................................................................................................................... 33
7.1.
7.2.
APPENDIX 1- VOLUMETRIC VERSUS MASS FLOW....................................................................................................... 33
APPENDIX 2 - GAS CONVERSION FACTORS ................................................................................................................ 34
Manual: 152-082010 301-305-306_303-307 Series
Page 3 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
1.
General Information
The Hastings 300 Series high range mass flow meters are designed to accurately measure mass flow at a
range between 25 slm and 10000 slm. The Hastings 300 Series high range mass flow controllers are
designed to accurately measure and control mass flow at a range between 25 slm and 2500 slm. These
instruments are intrinsically linear and have an accuracy of better than ±1% F.S. (full scale).
Hastings mass flow instruments do not require any periodic maintenance under normal operating
conditions with clean gases. No damage will occur from the use of moderate overpressures (~500
psi/3.45MPa) or overflows. Instruments are normally calibrated with the appropriate standard calibration
gas (nitrogen) then a correction factor is used to adjust the output for the intended gas. Calibrations for
other gases, such as oxygen, helium and argon, are available upon special order.
1.1.Features
•
LINEAR BY DESIGN. The HFM-301/305/306 and HFC-303/307 series is intrinsically linear (no
linearization circuitry is employed). Should recalibration (a calibration standard is required) in the
field be desired, the customer needs to simply set the zero and span points. There will be no
appreciable linearity change of the instrument when the flowing gas is changed.
•
•
NO FOLD OVER. The output signal is linear for very large over flows and is monotonically increasing
thereafter. The output signal will not come back on scale when flows an order of magnitude over
the full scale flow rate are measured. This means no false acceptable readings during leak testing.
MODULAR SENSOR. The HFM-301/305/306 and HFC-303/307 series incorporates a
removable/replaceable sensor module. Field repairs to units can be achieved with a minimum of
production line downtime. Calibration is required.
•
•
FAST SETTLING TIME. Changes in flow rate are detected in less than 500 milli-seconds when using
the standard factory PC board settings. Controller response is typically < 2 seconds.
LOW TEMPERATURE DRIFT. The temperature coefficient of span for the HFM-301/HFC-303 series is
less than 0.092% of full scale/°C from 0-60°C. The temperature coefficient of zero is less than 0.085
% of reading/°C from 0-60°C.
•
•
FIELD RANGEABLE. The HFM-305 and HFC-307 are available in ranges from 1000 slm to 2500 slm.
The HFM-301 and HFC-303 are available from 25-1000 slm. Each flow meter has a shunt and
transducer which can be exchanged in the field to select different ranges. Calibration, however, is
required.
CURRENT LOOP. The 4-20 mA option gives the user the advantages of a current loop output to
minimize environmental noise pickup.
Manual: 152-082010 301-305-306_303-307 Series
Page 4 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
1.2.Specifications
Accuracy..........................................................HFM-301/305 & HFC-303/307: ±1% full scale (F.S.)
...........................................................................................................HFM-306; ±3% full scale (F.S.)
Repeatability................................................................................................................ ±0.07% of F.S.
Maximum pressure................................................................................................500 psi [3.45 MPa]
..............................................................................................................................300 psi (HFM-306)
Maximum pressure................................................................................................1000 psi [6.9 MPa]
..................................................................... (with high pressure option on HFM-301/HFC303 only)
Pressure coefficient.................................................................................<0.026% of reading/psi (N2)
...............................................................................................See pressure section for pressure errors.
Operating temperature...........................................................0-60°C in non-condensing environment
Temperature coefficient (zero).......................................................maximum ±0.085%/°C (0-60 0C)
Temperature coefficient (span) ................................................... maximum ±0.092%/°C (15-50 0C)
Leak integrity........................................................................................................... <1x10-9 std. cc/s.
Standard output................................................................................... 0-5 VDC (load min 2k Ohms)
Optional output ....................................................................................4 -20 mA. (load < 600 Ohms)
Power requirements............................................................................±(15) VDC @ 55 mA (meters)
.....................................................................................................± (15) VDC @ 150 mA (controller)
................................................................................................................... Class 2 power 150VA max
Wetted materials........................................................ 302/304 & 316 stainless steel, nickel 200, Viton
..................................................................................................................... Kalrez® (controller only)
Attitude sensitivity of zero .................................................. < ±0.25% F.S. for 90° without re-zeroing
..................................................................................................................{N2 at 19.7 psi (135 KPa)}
Attitude sensitivity of span...................................................................................< ±0.06% of reading
..................................................................................................................{N2 at 19.7 psi (135 KPa)}
Weight ...................................................................................................... HFM-301; 3.5 lbs. (1.6 kg)
...................................................................................................................HFC-303; 5.3 lbs. (2.4 kg)
.................................................................................................................. HFM-305; 8.0 lbs. (3.7 kg)
.............................................................................................................. HFM-306; 28.8 lbs. (13.2 kg)
...............................................................................................................HFC-307; 15.3 lbs. (6.94 kg)
Electrical connector ......................................................................................15 pin subminiature “D”
Fitting options..............................................HFM-301/HFC-303: 1/2” Swagelok®, 3/4” Swagelok®,
....................................................................................................................... 1/2”VCR®, 1/2”VCO®
...................................................................HFM-305/HFC-307: 1”Swagelok®, 1”VCR®, 1”VCO®
..........................................................HFM-306: 1” Swagelok®, 1”VCR®, 1”VCO®, 2” Swagelok®
Manual: 152-082010 301-305-306_303-307 Series
Page 5 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
1.3.Optional 4-20 mA Current Output
An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to flow.
The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current loop output is
useful for remote applications where pickup noise could substantially affect the stability of the voltage
output.
The current loop signal replaces the voltage output on pin 6 of the “D” connector. The load must be less
than 600Ω. Failure to meet this condition will cause failure of the loop transmitter.
1.4.Other Accessories
1.4.1. Hastings Power supplies
Hastings Power Pod power supply/display units are available in one and four channel versions. They
convert 100, 115 or 230VAC to the ±15 VDC required to operate the flow meter and provide a digital
indication of the flow rate. Interface terminals for the retransmission of the flow meter analog output
signal are located on the rear of the panel.
The Power Pod 100 and 400 models are built with controllers in mind but will work with meters as well.
The Model 40 is for flow meters only. Throughout this manual, when reference is made to a power
supply, it is assumed the customer is using a Hastings power supply. Hastings PowerPod-100 and
PowerPod-400 power supplies are CE marked, but the Model 40 does not meet CE standards at this time.
The Model 40 and PowerPod-100 are not compatible with 4–20 mA analog signals. With the PowerPod
400, individual channels’ input signals, as well as their commands, become 4–20 mA compatible when
selected. The PowerPod-400 also sports a Totalizer feature. More information about the Power Pods can
1.4.2. Interconnecting Cables
Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on the
back of the Power Pod directly to any of the 200 series and 300 series flow instruments (including digital
versions). More information about the available cables can be found in the Power Pod 400 bulletin on
The 300 series flow instruments normally come with the standard “H” pin-out connector. This type of
connector is supplied on the Hastings AF-8-AM series cables with grey backshells (such as #65-149). “U”
pin-out versions of the 300 series instruments require a different cable to connect to the power supply.
This cable is identifiable by black back-shells and is available as Hastings Instrument (#65-791)
Manual: 152-082010 301-305-306_303-307 Series
Page 6 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
2.
Operation
This section contains the necessary steps to assist in getting a new flow meter/controller into operation
as quickly and easily as possible. Please read the following thoroughly before attempting to install the
instrument.
2.1.Receiving Inspection
Carefully unpack the Hastings 300 Series instrument and any accessories that have also been ordered.
Inspect for any obvious signs of damage to the shipment. Immediately advise the carrier who delivered
the shipment if any damage is suspected. Check each component shipped with the packing list. Insure
that all parts are present (i.e., flow meter, power supply, cables, etc.). Optional equipment or
accessories will be listed separately on the packing list. There may also be one or more OPT-options on
the packing list. These normally refer to special ranges or special gas calibrations. They may also refer
to special helium leak tests, or high pressure tests. In most cases, these are not separate parts, but
special options or modifications built into the flow meter.
Quick Start
1. Insure flow circuit mechanical connections are leak free.
2. Insure electrical connections are correct (See label).
3. Power up. (Allow 30 minutes to 1 hour warm-up for best accuracy.)
4. Note the flow signal decays toward zero.
5. Run ~20% flow (~ 1 VDC) through instrument for 5 minutes.
6. Insure zero flow; wait 2 minutes, then zero the instrument.
7. Instrument is ready for operation.
2.2.Power Requirements
The HFM-301/305/306 requires ±15 VDC @ 55 mA. The HFC-303/307 requires ±15VDC @ 150 mA for
proper operation. The supply voltage should be sufficiently regulated to no more than 50 mV ripple.
The supply voltage can vary from 14.0 to 16.0 VDC. Surge suppressors are recommended to prevent
power spikes reaching the instrument. The Hastings power supply described in Section Error! Reference
source not found. satisfies these power requirements.
2.3.Output Signal
The standard output of the flow meter is a 0 - 5 VDC signal proportional to the flow rate. In the Hastings
power supply the output is routed to the display and is also available at the terminals on the rear panel.
If a Hastings supply is not used, the output is available on pin 6 of the “D” connector. It is recommended
that the load resistance be no less that 2 kΩ. If the optional 4 - 20 mA output is used, the load
impedance must be less than 600 Ω.
2.4.Mechanical Connections
The flow meter may be mounted in any position as long as the direction of gas flow through the
instrument follows the arrow marked on the bottom of the flow meter case label. The preferred
orientation is with the inlet and outlet fittings in a horizontal plane. If operating with a dense gas (e.g.
Manual: 152-082010 301-305-306_303-307 Series
Page 7 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
sulfur hexafluoride) or at high pressures (> 250 psig) the instrument must be installed horizontally.
When mounted in a different orientation the instrument should be re-zeroed at zero flow with the
system pressurized to the expected operating pressure.
One of the smallest of the internal passageways in the 300 Series is the diameter of the sensor tube,
which can be 0.026” (0.66 mm), 0.017” (0.43mm), or 0.014” (0.36mm), so the instrument requires
adequate filtering of the gas supply to prevent blockage or clogging of the tube.
The pressure regulator and the plumbing upstream must be of sufficient size to minimize changes in the
upstream pressure. When switching from full flow to zero flow, the inlet pressure of instrument should
rise to no more that 30% above the inlet pressure at full flow. In general, high capacity regulators and
large internal diameter plumbing help to make the system more stable. The pressure drop between the
regulator and the instrument due to line resistance should be minimized.
There are two threaded holes located on the bottom of the base that can be used to secure it to a
mounting bracket, if desired. Other holes for special mounting can be added if desired.
The optional inlet and outlet fittings for the 301/303 are 0.5”, 0.75” Swagelok, 0.5” VCR and 0.5” VCO
fittings. The O-rings for the end cap are Viton (optional Kalrez, Neoprene or Buna-N). The HFM-305 and
HFC-307 are provided with 1” Swagelok, VCR or VCO fittings. The HFM-306 can have 1” or 2” Swagelok
fittings. It is suggested that all connections be checked for leaks after installation. This can be done by
pressurizing the instrument (do not exceed pressure rating of the instrument) and applying a diluted
soap solution to the flow connections.
2.5.Electrical Connections
DANGER: Care must be taken to avoid high voltages that may be present when dealing
with power supplies.
If a power supply from Hastings Instruments is used, installation consists of connecting the 300 Series
cable from the “D” connector on the rear of the power supply to the “D” connector on the top of the
flow meter/controller. The “H” pin-out requires cable AF-8-AM (grey molded back shell). The “U” pin-
out requires cable # 65-791 (black molded back shell).
If a different power supply is used, follow the instructions below when connecting the flow meter and
refer to either table 2.1 or 2.2 for the applicable pin-out. The power supply used must be bipolar and
capable of providing ±15 VDC at 55 mA for flow meter applications and ±15 VDC at 150 mA for
controllers. These voltages must be referenced to a common ground. One of the “common” pins of the
Hastings instrument must be connected to the common terminal of the power supply. Case ground of
the instrument should be connected to the AC ground locally. The cable shield (if available) should be
connected to AC ground at the either the power supply end, or the instrument end of the cable, not at
both. Pin 6 is the output signal from the flow meter. The standard output will be 0 to 5 VDC, where 5
VDC is 100% of the rated or full scale flow.
The command (set point) input should be a 0 - 5 VDC signal (or 4 - 20mA if configured as such), and must
be free of spikes or other electrical noise, as these would generate false flow commands that the
controller would attempt to follow. The command signal should be referenced to signal common.
A valve override command is available to the flow controller. Connect the center pin of a single pole,
three-position switch (center off) to the override pin. Connect +15 VDC to one end of the three position
switch, and -15 VDC to the other end. The valve will be forced full open when +15 VDC is supplied to the
override pin, and full closed when -15 VDC is applied. When there is no connection to the pin (the
three-position switch is centered) the valve will be in auto control, and will obey the 0 - 5 VDC
commands supplied to command (set-point) input.
Figure 2-1 and Table 2-1 show the 300 Series “H” pin out.
Figure 2-2 and Table 2-2 show the 300 Series U” pin out.
Manual: 152-082010 301-305-306_303-307 Series
Page 8 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 2-1
Table 2-1
“H” Pin-out
Pin
#
1
Do not use
2
3
Do not use
Do not use
4
Do not use
5
6
7
8
Signal Common
Output 0-5 VDC (4-20mA)
Case Ground
Valve Override
-15 VDC
9
10
Do not use
11 +15 VDC
12 Signal Common
13 External Input
14 Set Point 0-5 VDC (4-20mA)
15 Do not use
Manual: 152-082010 301-305-306_303-307 Series
Page 9 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 2-2
Table 2-2
“U” Pin-out
Pin
#
1
Signal Common
2
3
Do not use
Do not use
+15 VDC
4
5
6
Output 0-5 VDC (4-20mA)
Signal Common
7
8
Case Ground
9
Valve Override
10
11 -15 VDC
12 External Input
13 Signal Common
14 Signal Common
15 Set Point 0-5 VDC (4-20mA)
Manual: 152-082010 301-305-306_303-307 Series
Page 10 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
2.6.4 – 20 mA Connections
The 300 Series flow meters have a 4 - 20 mA current output option available as an alternative to the
standard 0 - 5 VDC output. This current output is useful for remote applications where noise pick-up
could substantially affect the stability of the voltage output. The 4 - 20 mA signal is produced from the 0
- 5 VDC output of the flow meter and replaces the voltage output on pin 6 of the “D” connector. The
current is sourced by the flow controller and the metering circuit must sink the current, such that the
current loop is completed through power supply common connections. The load must be less than 600
Ohms. Failure to meet this condition will cause failure of the loop transmitter.
The 300 Series flow controllers have a 4 – 20 mA current input option available as an alternative to the
standard 0 – 5 VDC command input. The controller sinks the input current through a 77 Ohm load to
power supply common.
−
−
+
Power
Supply
+
Hastings Flow controller
Pin−out H
PLC or Analog I/O Card
Pin 7
Power In
Pin 11
Pin 9
Pin 5,12
Analog Out
Analog In
Setpoint Input, Pin 14
77 Ohms
Analog Output, Pin 6
Analog Return
Figure 2-3
Manual: 152-082010 301-305-306_303-307 Series
Page 11 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
2.7.Operation
The standard instrument output is a 0 - 5 VDC and the signal is proportional to the flow i.e., 0 volts =
zero flow and 5 volts = 100% of rated flow. The 4 - 20 mA option is also proportional to flow, 4 mA =
zero flow and 20 mA = 100% of rated flow.
2.7.1. Operating Conditions
For proper operation, the combination of ambient temperature and gas temperature must be such that
the flow meter temperature remains between 0 and 60°C. Most accurate measurement of flow will be
obtained if the flow meter is zeroed at operating temperature as temperature shifts result in some zero
offset. The 300 Series is intended for use in non-condensing environments only. Condensate or any other
liquids which enter the flow meter may destroy its electronic components.
2.7.2. Zero Check
Turn the power supply on if not already energized. Allow for a 1 hour warm-up. Stop all flow through
the instrument and wait 2 minutes. Caution: Do not assume that all metering valves completely shut
off the flow. Even a slight leakage will cause an indication on the meter and an apparent zero shift. For
the standard 0 - 5 VDC output, adjust the zero potentiometer located on the lower inlet side of the flow
meter until the meter indicates zero. For the optional 4 - 20 mA output, adjust the zero potentiometer
so that the meter indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. This slight positive
adjustment ensures that the 4 - 20 mA current loop transmitter is not in the cut-off region. The error
induced by this adjustment is approximately 0.3% of full scale. This zero should be checked periodically
during normal operation. Zero adjustment is required if there is a change in ambient temperature or
orientation of the flow meter/controller.
2.7.3. High Pressure Operation
If the instrument is not mounted in a level position when operating at high pressure, the increased
density of the gas will cause natural convection to flow through the sensor tube. This natural convection
flow will be proportional to the system pressure and will be seen as a shift in the zero flow output that is
directly proportional to the system pressure.
If the system pressure is higher than 250 psig (1.7 MPa) the pressure induced error in the span reading
becomes significant. The following charts show the mean error and the minimum/maximum expected span
errors induced by high pressures. This error can approach 4.5% at 400 psig. for 0.017” sensors. For accurate
high pressure measurements, this error must be corrected. Consult your packing list to determine if you have a
0.026”, 0.017” or 0.014” sensor. After determining the sensor tube ID, use one of the formulae below to
determine the expected mean error, expressed as a fraction of the reading.
−11
3
−7
2
−5
Error = (9.887 *10
)P − (3.4154 *10 )P + (8.3288 *10 )P, (0.026" Sensor)
26
-10
3
−7 −4
2
Error = (1.533 *10
)P − (3.304 *10 )P + (1.8313 *10 )P, (0.017" Sensor)
−7 −5
17
-10
3
2
Error = (-1.692*10
)P + (1.776 *10 )P − (1.929 *10 )P, (0.014" Sensor)
14
Where P is the pressure in psig and Error is the fraction of the reading in error.
The flow reading can be corrected as follows:
Corrected = Indication − Full Scale * Error
Where the Indication is the indicated flow, Full Scale is the full scale reading and Error is the result of the
previous formula or read from charts below.
CAUTION: The HFM-306 does not have a high pressure option and is only rated for a
maximum working pressure of 300 psig.
Manual: 152-082010 301-305-306_303-307 Series
Page 12 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Span Error -vs- Pressure for 0.014" Sensors
3%
2%
1%
0%
-1%
-2%
-3%
Mean
Max
Min
0
100
200
300
400
500
600
700
800
900
1000
Pressure (psig)
Figure 2-4
Span Error -vs- Pressure for 0.017" Sensors
5%
4%
3%
2%
1%
0%
-1%
-2%
Mean
Max
Min
0
100
200
300
400
500
600
700
800
900
1000
Pressure (psig)
Figure 2-5
Manual: 152-082010 301-305-306_303-307 Series
Page 13 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Span Error vs Pressure for 0.026" Sensor
2.0%
0.0%
-2.0%
-4.0%
-6.0%
-8.0%
-10.0%
-12.0%
-14.0%
-16.0%
-18.0%
-20.0%
0
100
200
300
400
500
600
700
800
900
1000
Pressure(psig)
Figure 2-6
Manual: 152-082010 301-305-306_303-307 Series
Page 14 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Span Error vs Pressure
1.0%
0.8%
0.6%
0.4%
0.2%
0.0%
-0.2%
-0.4%
-0.6%
-0.8%
-1.0%
Mean error
max
min
0
50
100
150
200
250
300
Pressure(psig)
Figure 2-7
2.7.4. Blending of Gases
In the blending of two gases, it is possible to maintain a fixed ratio of one gas to another. In this case,
the output of one flow controller is used as the reference voltage for the set point potentiometer of a
second flow controller. The set point potentiometer then provides a control signal that is proportional
to the output signal of the first flow controller, and hence controls the flow rate of the second gas as a
percentage of the flow rate of the first gas.
EXAMPLE: Flow controller A has 0 - 100 slm range with a 5.00 volt output at full scale. Flow
controller B has 0 - 10 slm range with a 5.00 volt output at full scale. If flow controller A is set at
80 slm, its output voltage would be 4.00 volts (80 slm/100 slm x 5.00 volts = 4.00 volts). If the
output signal from flow controller A is connected to the command potentiometer of flow controller
B, it then becomes a variable reference voltage for flow controller B proportional to the flow rate
of flow controller A.
If the set point potentiometer of flow controller B is set at 50% of full scale, and the reference
voltage from flow controller A is 4.00, then the command signal going to flow controller B would be
2.00 volts (4.00 volts x 50.0% = 2.00 volts). The flow of gas through flow controller B is then
controlled at 4 slm (2.00 volts/5.00 volts x 10 slm = 4 slm).
The ratio of the two gases is 20:1 (80 slm/4slm). The % mixture of gas A is 95.2 (80slm/84slm and
the % mixture of gas B is 4.8% (4 slm/84 slm).
Should the flow of flow controller A drop to 78 slm, flow controller B would drop to 3.9 slm, hence
maintaining the same ratio of the mixture. (78 slm/100slm x 5v = 3.90v x 50% = 1.95v; 1.95v/5.00v
x 10 slm = 3.9 slm; 78 slm: 3.9 slm = 20:1)
Manual: 152-082010 301-305-306_303-307 Series
Page 15 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
2.8.Range Changing
Changing the range of a flow controller can be done in the field, but calibration is required. It is
recommended however, that the unit be sent back to the factory along with the new range desired, gas
and operating parameters. Consult factory for more information.
2.9.Output Filter
The flow output signal may have noise superimposed on the
mean voltage levels. This noise may be due to high
turbulence in the flow stream that the fast sensor is
measuring or it could be electrical noise when the flow meter
has a high internal gain (i.e. 5 sccm full scale meter).
Varying levels of radio frequency noise or varying airflow over
the electronics cover can also induce noise.
Noise can be most pronounced when measuring the flow
output with a sampling analog/digital (A/D) converter. When
possible, program the system to take multiple samples and
average the readings to determine the flow rate.
If less overall system noise is desired, a jumper may be
installed over the pins of JP-1 on the flow measurement card.
See Figure 2-8. Covering the pins closest to the “D” connector
will activate a resistor-capacitor (RC) filter that has a time
constant of one second. This will change the settling time of
the indicated flow rate to approximately 4 seconds. Covering
the other two pins will change the response time to approx. 1
second. This adjustment will not affect the calibration of the
flow meter circuit or the actual flow response to change in
Figure 2-8
command signal (flow controllers). This will only slow down
the indicated response (output voltage/current).
2.10.
Controlling Other Process Variables
Normally, a flow controller is setup to control the mass flow. The control loop will open and close the
valve as necessary to make the output from the flow measurement match the input on the command
line. Occasionally, gas is being added or removed from a system to control some other process variable.
This could be the system pressure, oxygen concentration, vacuum level or any other parameter which is
important to the process. If this process variable has a sensor that can supply an analog output signal
proportional to its value then the flow controller may be able to control this variable directly. This
analog output signal could be 0 - 5 Volts, 0 - 10 Volts (or 4 - 20 mA for units with 4 - 20 mA boards) or
any value in between.
On the CONTROLLER card there is a jumper (JP-1) that sets whether the control loop controls mass flow
or an external process variable. See Figure 2-9. If the jumper connects the top two pins, the loop
controls mass flow. If the jumper connects the bottom two pins, the loop controls an external process
variable. This process variable signal must be supplied on pin 13 of the D-connector (for H pin out units)
or pin 12 of the D-connector (for U pin out units) of the measurement card. When the controller is set
for external variable control it will open or close the valve as necessary to make the external process
variable signal match the command signal. The command signal may be 0 - 5 Volts, 0 - 10 Volts (4 - 20
mA for 4 - 20 mA input/output cards) or any value in between. If the process variable has a response
time that is much faster or slower than the flow meter signal it may be necessary to adjust the gain
potentiometer.
2.11.
Command Input
The flow controller will operate normally with any command input signal between 0 - 10 Volts (4 - 20 mA
for units with 4 - 20 mA input/output cards). If the command signal exceeds ±14 volts it may damage the
circuit cards. During normal operation the control loop will open or close the valve to bring the output
Manual: 152-082010 301-305-306_303-307 Series
Page 16 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
of the flow meter signal to within ± 0.001 Volts of the command signal. The command signal will not
match the flow signal if there is insufficient gas pressure to generate the desired flow. If the command
signal exceeds 5 Volts the controller will continue to increase the flow until the output matches the
command signal. However, the flow output may not maintain the published accuracy values under these
conditions.
If the command signal is less than 1% of full scale (0.05 Volts or 4.16 mA) the valve override control
circuit will activate in the closed position. This will force the valve completely closed regardless of the
flow signal.
2.12.
Valve Override Control
The valve override control line provides a method to override the loop controller and open or close the
valve regardless of the flow or command signals. During normal operation this line must be allowed to
float freely. This will allow the loop control to open and close the valve as it requires. If the valve
override line is forced high (> +5 Volts) the valve will be forced full open. If the valve overrides line is
forced negative (< -5 Volts) the valve will be forced closed.
Control Loop
Jumper
2.13.
Gain
Potentiometer
On the top left of the inlet side of the flow
controller there is a hole through which the gain
potentiometer is accessible. See Figure 2-9 and
Figure 4-2. This gain potentiometer affects the gain
of the closed loop controller.
Normally this
potentiometer will be set at the factory for good
stable control. It may be necessary to adjust this
potentiometer in the field if the system varies
widely from the conditions under which the
controller was setup.
Turning this gain
potentiometer clockwise will improve stability.
Turning the potentiometer counter-clockwise will
speed up the valve reaction time to changes in the
command signal.
Gain
Potentiometer
Figure 2-9
Manual: 152-082010 301-305-306_303-307 Series
Page 17 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
2.14.
Temperature Coefficients
As the temperature of the instrument changes from the calibration temperature, errors will be
introduced into the output of the instrument. The Temperature Coefficient of Zero describes the change
in the output that is seen at zero flow. This error is added in to the overall output signal regardless of
Span error -vs- Temperature
1.4%
1.2%
y = 5E-06x2 - 9E-05x - 0.0007
1.0%
0.8%
0.6%
0.4%
0.2%
0.0%
0
10
20
30
40
50
60
70
-0.2%
Temperature (°C)
Figure 2-10
flow, but can be eliminated by merely adjusting the zero potentiometer to read zero volts at zero flow
conditions. The Temperature Coefficient of Span describes the change in output after the zero error is
eliminated. This error cannot be eliminated, but can be compensated mathematically if necessary. The
curve pictured in Figure 2-10 shows the span error in percent of point as a function of temperature
assuming 23 0C is the calibration temperature.
Manual: 152-082010 301-305-306_303-307 Series
Page 18 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
3.
Theory and Function
3.1.Overall Functional Description
The 300 Series flow meters consist of a sensor, base, shunt, control valve (303/307) and electronic
circuitry. The sensor is configured to measure gas flow rate from 25-10000 slm, depending on the
customer’s desired overall flow rate. The shunt divides the overall gas flow such that the flow through
the sensor is a precise percentage of the flow through the shunt. The flow through both the sensor and
shunt is laminar. The control valve adjusts the flow so that the sensors measurement matches the set-
point input. The circuit board amplifies the sensor output from the two RTD’s (Resistive Temperature
Detectors) and provides an analog output of either 0 - 5 VDC or 4 - 20 mA.
3.2.Sensor Description
The sensor consists of two coils of resistance wire with a high temperature coefficient of resistance
(3500 ppm/oC) wound around a stainless steel tube with internal diameter of 0.014”(0.36mm),
0.017”(0.43mm) or 0.026”(0.66mm) and 3.00” (7.62) cm length. These two identical resistance wire
coils are used to heat the gas stream and are symmetrically located upstream and downstream on the
sensor tube. Insulation surrounds the sensor tube and heater coils with no voids around the tube to
prevent any convection losses. The ends of this sensor tube pass through an aluminum block and into the
stainless steel sensor base. This aluminum block thermally shorts the ends of the sensor tube and
maintains them at ambient temperature.
There are two coils of resistance wire that are wound around the aluminum block. The coils are identical
to each other and are symmetrically spaced on the aluminum ambient block. These coils are wound
from the same spool of wire that is used for the sensor heater coils so they have the same resistivity and
the same temperature coefficient of resistance as the sensor heater coils. The number of turns is
controlled to have a resistance that is 10 times larger than the resistance of the heater coils. Thermal
grease fills any voids between the ambient temperature block and the sensor tube to ensure that the
ends of the sensor tube are thermally tied to the temperature of this aluminum block.
Aluminum has a very high thermal conductivity which ensures that both ends of the sensor tube and the
two coils wound around the ambient block will all be at the same temperature. This block is in good
thermal communication with the stainless steel base to ensure that the ambient block is at the same
temperature as the main instrument block and, therefore, the same temperature as the incoming gas
stream. This allows the coils wound on the aluminum block to sense the ambient gas temperature.
Two identical Wheatstone bridges are employed, as shown in Figure 3.2. Each bridge utilizes an ambient
temperature sensing coil and a heater coil. The heater coil and a constant value series resistor comprise
the first leg of the bridges. The second leg of each bridge contains the ambient sensing coil and two
constant value series resistors. These Wheatstone bridges keep each heater temperature at a fixed value
of ∆T = 48 oC above the ambient sensor temperature through the application of closed loop control and
the proper selection of the constant value bridge resistors.
3.3.Sensor Theory
The heat transferred by convection to or from a fluid is proportional to the mass flow of that fluid. Since
the constant differential temperature sensor has 2 heater coils symmetrically spaced on the sensor tube,
it is convenient to consider the upstream and downstream heat transfer modes separately. The electrical
power supplied to either of the heater coils will be converted to heat, which can be dissipated by
radiation, conduction, or convection. The radiation term is negligible due to the low temperatures used
by the sensor, and because the sensor construction preferentially favors the conductive and convective
heat transfer modes. The thermal energy of each heater will then be dissipated by conduction down the
stainless steel sensor tube, conduction to the insulating foam, plus the convection due to the mass flow
of the sensed gas.
Because great care is taken to wind the resistive heater coils symmetrically about the midpoint of the
tube, it is assumed that the heat conducted along the sensor tube from the upstream heater will be
equal to the heat conducted through the tube from the downstream heater. Similarly, the heat
Manual: 152-082010 301-305-306_303-307 Series
Page 19 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
conducted from the upstream and downstream coils to the foam insulation surrounding them is assumed
to be equal, based on the symmetry of the sensor construction.
Since the sensor tube inlet and outlet are linked by an aluminum ambient bar, the high thermal
conductivity of the bar provides a ‘thermal short’, constraining the ends of the sensor tube to be at
equal surface temperature. Moreover, the tube ends and the aluminum ambient bar have intimate
thermal communication with the main flow passageway prescribed by the main stainless steel flow meter
body. This further constrains each end of the sensor tube to be equal to the ambient gas temperature.
Further, since the length of each heater section is nearly 21 times greater than the inside tube diameter,
the mean gas temperature at the tubes axial midpoint is approximately equal to the tube surface
temperature at that point. Recall that the outside of the sensor tube is well insulated from the
surroundings; therefore the tube surface temperature at the axial midpoint is very close to the operating
temperature of the heater coils. The mean temperature of the gas stream is then approximately the
same as the heater temperature. Assuming the mean gas temperature is equal to the heater
temperature, it can be shown that the differential power is:
•
P − P = 2mC T
− T
)
u
d
p
heater
ambient
The value of the constant pressure specific heat of a gas is virtually constant over small changes in
temperature. By maintaining both heaters at the same, constant temperature difference above the
ambient gas stream temperature, the difference in heater power is a function only of the mass flow
rate. Fluctuations in ambient gas temperature which cause errors in conventional mass flow sensors are
avoided; The resistance of the ambient sensing coil changes proportionally with the ambient
temperature fluctuations, causing the closed loop control to vary the bridge voltage such that the heater
resistance changes proportionally to the ambient temperature fluctuation.
The power supplied to each of the 2 heater coils is easily obtained by measuring the voltage across the
heater, shown as V2 on Figure 3.2, and the voltage across the fixed resistor R1. Since R1 is in series with
the heater RH they have the same current flowing through them. The electrical power supplied to a
given heater is then calculated:
V −V V
)
1
2
2
P =
R
1
With a constant differential temperature applied to each heater coil and no mass flow through the
sensor the difference in heater power will be zero. As the mass flow rate through the sensor tube is
increased, heat is transferred from the upstream heater to the gas stream. This heat loss from the
heater to the gas stream will force the upstream bridge control loop to apply more power to the
upstream heater so that the 48 oC constant differential temperature is maintained.
The gas stream will increase in temperature due to the heat it gains from the upstream heater. This
elevated gas stream temperature causes the heat transfer at the downstream heater to gain heat from
the gas stream. The heat gained from the gas stream forces the downstream bridge control loop to apply
less power to the downstream heater coil in order to maintain a constant differential temperature of 48
oC.
The power difference at the RTD’s is a function of the mass flow rate and the specific heat of the gas.
Since the heat capacity of many gases is relatively constant over wide ranges of temperature and
pressure, the flow meter may be calibrated directly in mass units for those gases. Changes in gas
composition require application of a multiplication factor to the nitrogen calibration to account for the
difference in heat capacity.
3.4.Base
The 316 stainless steel base has a 2.0" by 2.0” (50.8 mm by 50.8 mm) cross-section and is 5.77"(146.6
mm) long on the HFC-303 and 4.32” (109.7 mm) long on the HFM-301. Lower flow (< 300 SLM) 301's are
0.2" shorter and lower flow 303's are 0.4" shorter. The base has an internal flow channel that is
Manual: 152-082010 301-305-306_303-307 Series
Page 20 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
1.75"(44.5 mm) diameter. The base and endcap seal is an O-ring gland configuration, which uses Viton
as a standard sealing material. The HFM-305/306 is made from a larger 3.0” by 3.0” (76.2 mm by 76.2
mm) cross section of 316 stainless steel and has a length of 5.3” (134.6 mm) while the 307 has a length
of 8” (203.2 mm). The 305/307 however has a larger 2.67” (67.8 mm) internal flow channel and knife
edge metal to metal seals on the end caps. Between the base and sensor module on the 300 Series is a
knife edge metal to metal seal. Gaskets made of nickel 200 are swaged between mating face seals
machined into the stainless steel parts. All seals are tested at the factory and have leak rates of less
than 1x10-9 std. cc/s.
The HFM-306 base is much larger. The base/end cap seal is an O-ring gland configuration, which uses
Viton as a standard sealing material.
3.5.Shunt description
The flow rate of interest determines the size of the shunt required. As previously indicated, the
301/305 series can be configured from 25 to 2500 slm (303/307 controller from 25-2500 slm) using
various base, shunt and transducer combinations. The shunts employ a patented method of flow
division, which results in a more linear flow meter/controller; further calibration is more stable when
changing between measured gases.
The shunts are comprised of a 316 stainless steel cylindrical shell concentrically located in the base that
forms an annular flow channel of precise dimension. This flow channel creates laminar flow by the
inboard sensor taps. The cylindrical shell encases a corrugated matrix of flow channels which serve as a
shunt. The size and number of these channels is consistent with the sensor ∆P and flow range.
3.6.Shunt Theory
A flow divider for a thermal mass flow transducer usually consists of an inlet plenum, a flow restriction,
shunt and an outlet plenum. (See Figure 3-1). Since stability of the flow multiplier is desired to ensure a
stable instrument, there must be some matching between the linear volumetric flow versus pressure
drop of the sensor and the shape of the volumetric flow versus pressure drop of the shunt. Most
instruments employ Poiseuille’s law and use some sort of multi-passage device that creates laminar flow
between the upstream sensor inlet and the downstream outlet. This makes the volumetric flow versus
pressure drop curve primarily linear, but there are other effects which introduce higher order terms.
Most flow transducers are designed such that the outlet plenum has a smaller diameter than the inlet
plenum. This eases the insertion and containment of the shunt between the sensor inlet point and the
sensor outlet point. However there will be a pressure drop between the sensor inlet and outlet points
created by the change in flow passage diameter. Since the drop is a square function of the flow velocity
the differential pressure will be non-linear with respect to flow rate. Note also that the pressure drop is
a function of density.
The density will vary as a function of system
pressure and it will also vary when the gas
composition changes. This will cause the
magnitude of the pressure drop due to the area
change to be a function of system pressure and
gas composition.
Most of the shunts used contain or can be
approximated by many short capillary tubes in
parallel. We know that the equation for the
pressure drop across a capillary tube contains
terms that are proportional to the square of the
volumetric flow rate. These terms come from
the pressure drops associated with the sudden
c
Figure 3-1
ompression at the entrance and the sudden expansion at the exit of the capillary tube. The end effect
terms are a function of density which will cause the quadratic term to vary with system pressure and gas
composition. The absence of viscosity in the second term will cause a change in the relative magnitudes
of the two terms whenever the viscosity of the flowing gas changes.
Manual: 152-082010 301-305-306_303-307 Series
Page 21 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
128μLQ 8ρQ2
ΔP =
+
K + K
(
)
πD4
π 2 D4
c
e
The end effects for a typical laminar flow element in air account for approximately 4% of the total
pressure drop. For hydrogen, however, which has a density that is about 14 times less than air and has a
viscosity that is much greater than air, the second term is completely negligible. For the heavier gasses
such as sulfur hexafluoride which has a density 5 times that of air the end effects will become 10% of the
total. These changes make it impossible to accurately calibrate an instrument on one gas and use it for
another gas.
The pressure drop is linear with respect to the volumetric flow rate between a point that is downstream
of the entrance area and another point further downstream but upstream of the exit region. For a
typical flow divider tube the entry length is approximately 0.16 cm. From this it can be seen that if the
sensor inlet pickup point is inside of the flow divider tube but downstream of the entrance length and if
the sensor outlet point is inside the flow divider tube but upstream of the exit point then the pressure
drop that drives the flow through the sensor would be linear with respect to volumetric flow rate. Since
the pressure drop across the sensor now increases linearly with the main flow rate and the sensor has a
linearly increasing flow with respect to pressure drop, there is now a flow through the sensor which is
directly proportional to the main flow through the flow divider, without the flow division errors that are
present when the sensor samples the flow completely upstream and downstream of the flow divider.
Unfortunately, a typical shunt has an internal diameter on the order of 0.3 mm. This is too small to
insert tap points into the tube. Also, the sample flow through the sensor is approximately 10 sccm while
the flow through a shunt is approximately 25 sccm. This means the sample flow would be affecting the
flow it was trying to measure. If the sensor tube is made large enough, and with enough flow through it
to insert the sensor taps at these positions, then the pressure drop would be too small to push the
necessary flow through the sensor tube.
The solution is to use a different geometry for the flow tube. It must be large enough to allow the
sample points in the middle yet with passages thin enough to create the differential pressures required
for the sensor. An annular passage meets these requirements. All of the 400 I series shunts have an
annular passageway that passes between the outside of the main shunt body and the inside edge of the
base where the sensor taps are located.
Figure 3-2
The shunt must generate a pressure drop at the desired full scale flow which drives the proper flow
through the sensor tube to generate a full scale output from the sensor. Since the full scale flow of the
sensor is the same for all of the different full scale flows that may pass through the shunt, the geometry
must vary for the different full scale flows in order to generate the same pressured drop for all of them.
This is accomplished by adding more channels to main shunt body. As the full scale flow rate increases
the flow meter body must get larger to allow room for the increased number of bypass passageways that
are required.
Manual: 152-082010 301-305-306_303-307 Series
Page 22 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
3.7.Control Valve
The control valve is an “automatic metering solenoid” valve (see Figure 3-3). While most solenoid valves
operate in either the fully open or closed positions, the automatic metering solenoid valve is designed to
control flow. A spring is used to hold a magnetic plunger assembly tightly against an orifice, thereby
shutting off the flow. The magnetic plunger assembly is surrounded by a coil of magnet wire. When the
coil is energized the electric current passing through the wire coil produces a magnetic field which
attracts the plunger. The plunger assembly moves away from the orifice allowing the gas flow to pass
between the orifice and the plunger seat. The distance between the orifice and the plunger seat, and
thus the flow through the valve, is controlled by the amount of current supplied to the coil.
The valve seat is made of Kalrez (or equivalent) perflouroelastomer. The valve orifice is made from 316
stainless steel. The valve plunger and pole piece are made of nickel plated magnetic alloy (Hi-perm 49)
and the control springs are made of 302 stainless steel. The Diaphragm and all seals are as specified per
the customer order (Viton, Neoprene, Buna-N or Kalrez).
Figure 3-3
Manual: 152-082010 301-305-306_303-307 Series
Page 23 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
3.8.Electronic Circuitry
The 300 Series flow meters employs a thermal transfer principle (capillary tube described in section 3.2)
to measure the flow through the sensor which is proportional to the total flow through the instrument.
The sensor develops a differential voltage output signal proportional to flow, which is amplified to
produce 5 VDC at full scale flow. The amplified output can be measured on pins 6 and 1 of the external
“D” connector. If a Hastings power supply is employed, the 5 Volt output is also sent to the terminals on
the back and to the decoding circuitry in the display, which converts it to a numeric output. The
optional 4 - 20 mA analog output is available in lieu of an output voltage. The addition of a 4 - 20 mA
current loop transmitter on a secondary PC board (mounted parallel to the main pc board) is required to
provide this current loop. A jumper change is made on the secondary PC board to establish the selected
output mode.
3.9.Instrument Performance
The combination of these principles embodied in the 300 Series give inherently linear and highly
accurate output and the typical response time from zero to full flow is also quite good. The meter
settling time of a 1000 slm full scale flow meter is typically under 0.5 seconds. Flow controller response
is less than 2 seconds.
Manual: 152-082010 301-305-306_303-307 Series
Page 24 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
4.
Maintenance
4.1.Authorized Maintenance
With proper care in installation and use, the flow meter will require little or no maintenance. If
maintenance does become necessary, most of the instrument can be cleaned or repaired in the field.
Some procedures may require recalibration. Do not attempt these procedures unless facilities are
available. Entry into the sensor or tampering with the printed circuit board will void warranty. Do not
perform repairs on these assemblies while the unit is still under warranty.
CAUTION: Some parts of the instrument are delicate and other parts are heavy enough
to cause an injury if dropped. Use extreme care when servicing the instrument. The
potentiometer positions and the electrical components referred to in the
troubleshooting section can be found in Section 4.3.
4.2.Troubleshooting
Symptom:
Cause:
Output reads strong indication of flow with no flow present. Zero pot has no effect.
Power shorted out.
Action:
Turn the power supply off for a few seconds then turn it on again. If this is ineffective,
disconnect the power supply from the unit. Check that the power supply voltages are
correct. Incorrect voltages most likely signify a faulty regulator chip inside the supply.
If the power supply display returns to zero after the instrument has been disconnected
there may be a short from the unit to ground.
Symptom:
300 Series instrument output continues to indicate flow with no flow present, or
indicates ±14 volts. Power supply inputs are correct (see the above troubleshooting tip)
and zero pot has no effect.
Cause:
Action:
Faulty IC chip(s) on the main PC board.
Replace main PC board. (See Section 4.5)
Symptom: Output of flow meter is proportional to flow,
but extremely small and not correctable by span pot.
Cause: Sensor is not being heated.
Action: Shut off gas supply and disconnect the power to
5 6 7 8
the flow meter. Remove cover and PC board from unit.
Check the resistance from pins 1 to 2 and 3 to 4 of the
sensor module. Both sets of pins should read
approximately the same value between 1 and 2 kΩ
4 3 2 1
nominal resistance. Also check that the resistance from
pins 5 to 6, and 7 to 8 are both nearly the same value
between 200 and 400 Ω. See Figure 4-1. Incorrect
resistance values indicate that the sensor unit needs to
be replaced.
Figure 4-1
Manual: 152-082010 301-305-306_303-307 Series
Page 25 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Symptom:
Cause:
Sensor has proper resistance readings, but little or no output with flow.
Plugged sensor.
Action:
Shut off gas supply and disconnect the power to the flow meter. Remove cover and PC
board from unit. Remove and inspect sensor. If sensor has evidence of clogging, clean
or replace as applicable.
Symptom:
Flow meter reads other than 0.00 VDC with no flow or there is a small flow when the
flow meter reads 0.00 VDC.
Cause:
Action:
Zero pot is out of adjustment.
Shut off all flow. For the standard 0 – 5 VDC output, adjust the zero potentiometer (see
Section 4.3) located on the upper right inlet side of the flow meter until the meter
indicates zero. For the optional 4 - 20 mA output, adjust the zero potentiometer so that
the meter indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. This slight positive
adjustment ensures that the 4 - 20 mA transmitter is not in its cut-off region. The error
induced by this adjustment is approximately 0.3% of full scale.
Symptom:
Cause:
Flow meter is out of calibration and non-linear.
Leaks in the gas inlet or outlet fittings.
Action:
Check all fittings for leaks by placing soap solution on all fittings between gas supply and
final destination of gas. Check flow meter for leaks. Replace if required or recalibrate as
necessary.
Symptom:
Cause:
Little or no flow, even when the valve is in override OPEN.
Blocked orifice or incorrect pressure across the flow controller
Action:
Verify that there is 10 to 15 psig pressure across the instrument. If the differential
pressure across the instrument is correct, the orifice may be obstructed. Remove all gas
pressure and shut off power supply. Remove the valve.
4.3.Adjustments
1. Connect power to “D” connector as specified in Section 2.5. Allow the instrument to warm up for at
least 30 minutes (60 minutes for best accuracy). Note the flow signal decays toward zero. Run ~20% of
flow range (~1 VDC) through instrument for 5 minutes.
2. Completely shut off the flow and wait for 2 minutes. For the standard 0 – 5 VDC output, adjust the
zero potentiometer located on the lower inlet side of the flow meter (see Figure 4-2) until the meter
indicates zero. For the optional 4 - 20 mA output, adjust the zero potentiometer so that the meter
indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. This slight positive adjustment ensures that the
4 - 20 mA transmitter is not in its cut-off region. The error induced by this adjustment is approximately
0.3% of full scale.
3. Turn on gas supply to inlet of instrument and adjust the flow rate to the desired full scale flow as
indicated by a reference flow meter/controller. If the instrument to be calibrated is a controller, adjust
the flow rate by setting the command to 100% of full scale.
4. Perform this step only if a calibrated reference flow meter is available. Adjust Span pot (see Figure
4-2) until the indicated flow reads full scale (5.00 VDC or 20 mA) and the reference meter/controller
reads the desired full scale flow rate.
5. Record flow meter/controller and flow reference outputs for flow rates of 20%, 40%, 60%, 80% and
100% and make sure data are within ± 1% of full scale.
Manual: 152-082010 301-305-306_303-307 Series
Page 26 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
Control
Loop Gain
Zero
Span
PC
Board
Figure 4-2
4.4.End Cap Removal
The end cap on the inlet side must be removed to gain access to shunt assembly. First remove power
and shut off the supply of gas to the instrument. Disconnect the fittings on the inlet and outlet sides of
the transducer and remove it from the system plumbing. On both the HFM-301 and HFC-303, remove the
four Allen head screws holding the end cap to the instrument. Carefully remove the “O”-ring sealed end
cap and shunt, noting their order and the proper annular orientation of the shunt. The shunt position is
critical to the flow dynamics and can substantially affect the instrument linearity. The shunt can be
severely damaged if dropped. Examine the shunt. If damaged, dirty or blocked, clean and replace as
applicable. Reassemble in the reverse order of disassembly. Secure the endcap with 65 in lb. (7.3 N.m)
to 85 in lb (9.6 N-m) of torque on each high-tensile socket head cap screw. Use of a fastener other than
the one mentioned here may result in leakage at the seal. Recalibration of the 301/303 series
instrument is necessary.
For the HFM-305/307 remove the four 12 point bolts securing the endcap while carefully holding the
endcap in place. Carefully remove the metal gasket sealed endcap and shunt, noting their order and the
proper annular orientation of the shunt. The shunt position is critical to the flow dynamics and can
substantially affect the instrument linearity. The shunt can be severely damaged if dropped. Examine
the shunt. If damaged, dirty or blocked, clean and replace as applicable. Reassemble in the reverse
order of disassembly replacing the used metal gasket with a new one. Secure the endcap with 35 ft lb. (
47.5 N.m) to 40 ft lb. (54.3 N.m) of torque on each 12 pt bolt. Use of a fastener other than the one
mentioned here may result in leakage at the seal. Recalibration of the 305/307 series instrument is
necessary.
For the HFM-306, remove the six hex socket bolts from the inlet end cap. Remove the end cap and
ensure that the o-ring is not damaged. Remove the flow conditioning screen assembly and if damaged,
dirty or blocked, clean and replace as applicable. Carefully remove the flow straightener and shunt
assembly. If damaged, dirty or blocked, clean and replace as applicable. Reassemble in the reverse order
of disassembly. Recalibration of the HFM-306 instrument is necessary.
Manual: 152-082010 301-305-306_303-307 Series
Page 27 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
4.5.Printed Circuit Board and Sensor Replacement
NOTE: This instrument contains static sensitive PC boards. Maintain static protection when handling the
PC boards.
In the event that any of the PC boards fail, they are easily removed from the instrument and replaced
with a spare. It is recommended that the metering printed circuit board and sensor be replaced as a unit
because they are electronically matched to optimize response time. The sensor can also be removed
easily. This ease in disassembly and replacement substantially reduces instrument downtime.
1. Replacement of the 4 - 20 mA option PC board: Unplug the power cable from the instruments “D”
connector. Remove the fasteners and steel can. The 4 - 20 mA board is the PC board mounted by a
single screw. Remove the screw and lift off the 4 - 20 mA board. Be careful not to damage the main
board and 4 - 20 mA board connector.
2. Replacement of the main PC board: Unplug the power cable from the instruments “D” connector.
Remove the fasteners and steel can. Remove the 2 screws which fasten the main PC board to the sensor
module. Gently unplug the main board from the sensor (and from the 4 - 20 mA board, if present).
3. Replacement of the flow sensor: Remove the PC board(s) as described above. Remove the four (4)
Allen head cap screws that fasten the sensor to the main instrument base. Remove the sensor module
from the base, discarding the used nickel gaskets. New nickel gaskets are required for re-assembly.
To place an order or to obtain information concerning replacement parts, contact the factory
representative in your area. See section 6 in this manual for the address or phone number. When
ordering, include the following information: Instrument model number, part description and Hastings
part number.
Manual: 152-082010 301-305-306_303-307 Series
Page 28 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
5.
Outline Drawings
5.1.HFM-301
5.2.HFC-303
Manual: 152-082010 301-305-306_303-307 Series
Page 29 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
5.3.HFM-305
5.4.HFM-306
Manual: 152-082010 301-305-306_303-307 Series
Page 30 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
5.5.HFC-307
Manual: 152-082010 301-305-306_303-307 Series
Page 31 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
6.
WARRANTY
6.1.Warranty Repair Policy
Hastings Instruments warrants this product for a period of one year from the date of shipment to be free
from defects in material and workmanship. This warranty does not apply to defects or failures resulting
from unauthorized modification, misuse or mishandling of the product. This warranty does not apply to
batteries or other expendable parts, or to damage caused by leaking batteries or any similar occurrence.
This warranty does not apply to any instrument which has had a tamper seal removed or broken.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as
to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential
damages.
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship during
the warranty period. Defective instruments should be returned to Hastings Instruments, shipment
prepaid, together with a written statement of the problem and a Return Material Authorization (RMA)
number.
Please consult the factory for your RMA number before returning any product for repair. Collect freight
will not be accepted.
6.2.Non-Warranty Repair Policy
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form
and a written description of the problem with the instrument. If the repair cost is higher, you will be
contacted for authorization before we proceed with any repairs. If you then choose not to have the
product repaired, a minimum will be charged to cover the processing and inspection. Please consult the
factory for your RMA number before returning any product repair.
TELEDYNE HASTINGS INSTRUMENTS
804 NEWCOMBE AVENUE
HAMPTON, VIRGINIA 23669 U.S.A.
ATTENTION: REPAIR DEPARTMENT
TELEPHONE
TOLL FREE
FAX
(757) 723-6531
1-800-950-2468
(757) 723-3925
E MAIL
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments web
site.
Manual: 152-082010 301-305-306_303-307 Series
Page 32 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
7.
Appendix
7.1.Appendix 1- Volumetric versus Mass Flow
Mass flow measures just what it says, the mass or number of molecules of the gas flowing through the
instrument. Mass flow (or weight per unit time) units are given in pounds per hour (lb/hour), kilograms
per sec (kg/sec) etc. When your specifications state units of flow to be in mass units, there is no reason
to reference a temperature or pressure. Mass does not change based on temperature or pressure.
However, if you need to see your results of gas flow in volumetric units, like liters per minute, cubic feet
per hour, etc. you must consider the fact that volume DOES change with temperature and pressure. To
do this, the density (grams/liter) of the gas must be known and this value changes with temperature and
pressure.
When you heat a gas, the molecules have more energy and they move around faster, so when they
bounce off each other, they become more spread out, therefore the volume is different for the same
number of molecules.
Think about this:
The density of Air at 0°C is 1.29 g/liter
The density of Air at 25°C is 1.19 g/liter
The difference is 0.1 g/liter. If you are measuring flows of 100 liters per minute, and you don’t use the
correct density factor then you will have an error of 10 g/minute!
Volume also changes with pressure. Think about a helium balloon with a volume of 1 liter. If you could
scuba dive with this balloon and the pressure on it increases. What do you think happens to the weight
of the helium? It stays the same. What would happen to the volume (1 liter)? It would shrink.
Why is the word standard included with the volume terms liters and cubic feet in mass flow applications?
A mass flow meter measures mass …and we know we can convert to volume.
To use density we must pick one (or standard) temperature and pressure to use in our calculation.
When this calculation is done, the units are called standard liters per minute (SLM) or standard cubic
feet per minute (SCFM), for instance, because they are referenced to a standard temperature and
pressure when the volume is calculated.
Using the example to the left, we can see a
standard liter can be defined differently. The first
balloon contains 0.179 grams of Helium at 0 °C
and 760 Torr (density of 0.179 grams/liter). Heat
up that balloon to room temperature and the
volume increases, but the mass has not changed.
The volume is not 1 liter anymore, it is 1.08
liters.
So, to define a standard liter of Helium at 25 °C,
we must extract only one liter from the second
balloon and that liter weighs only 0.175 grams.
If a mass flow meter is set up for STP at 0 °C and
760 Torr, when it measures 0.179 grams of He, it
will give you results of 1 SLM. If a second meter
is set up for STP at 25 °C and 760 Torr, when it
measures 0.164 grams, it will give results of 1
SLM.
1.08 Liter
1 Liter
25 °C
1 Liter
0 °C
0.179 grams/1
liter
25 °C
0.179 g/1.08
liters
0.164 grams
Manual: 152-082010 301-305-306_303-307 Series
Page 33 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
7.2.Appendix 2 - Gas Conversion Factors
Gas conversion factors (GCF’s) for gasses metered using Hastings Instruments products, can be found by
visiting the Hastings Instruments web site. The web address can be found at the end of this document.
The gas conversion factors (GCF's) provided by Hastings Instruments (HI) fall into five basic accuracy
domains that, to a large extent, are dependent on the method by which they are found. The following
table summarizes the different methods used to determine the GCF's. The table lists the methods in
decreasing order of the degree of accuracy that may be achieved when applying a conversion factor.
Methods Used to Determine Gas Conversion Factors
1. Determined empirically at Hastings Instruments
2. Calculated From NIST tables
3. Calculated using the virial coefficients of independent investigators' empirical
data using both temperature and pressure as variables.
4. Calculated from virial coefficients using temperature only.
5. Calculated from specific heat data at 0 C and 1 atmosphere
1. The most accurate method is by direct measurement. Gases that can be handled safely, inert gases,
gases common in the atmosphere, etc., can be run through a standard flow meter and the GCF
determined empirically.
2. The National Institute of Standards and Technology (NIST) maintains tables of thermodynamic
properties of certain fluids. Using these tables, one may look up the necessary thermophysical
property and calculate the GCF with the same degree of accuracy as going directly to the referenced
investigator.
3 & 4. Many gases that have been investigated sufficiently by other researchers, can have their molar
specific heat (C'p) calculated. The gas conversion factor is then calculated using the following ratio.
Cp' N 2
Cp' GasX
GCF =
GCF's calculated in this manner have been found to agree with the empirically determined GCF's
within a few tenths of a percent. Data from investigations that factor in pressure as well as
temperature, usually supply a higher degree of accuracy in their predictions.
5. For rare, expensive gases or gases requiring special handling due to safety concerns, one may look up
specific heat properties in a variety of texts on the subject. Usually, data found in this manner
applies only in the ideal gas case. This method yields GCF's for ideal gases but as the complexity of
the gas increases, its behavior departs from that of an ideal gas. Hence the inaccuracy of the GCF
increases.
Hastings Instruments continually searches for better estimates of the GCF's of the more complex gases
and regularly updates the list.
Most Hastings flow meters and controllers are calibrated using nitrogen. The conversion factors
published by Hastings are meant to be applied to these meters. To apply the GCF's, simply multiply the
gas flow reading and the GCF for the process gas in use. For example, to calculate the actual flow of
argon passing through a nitrogen-calibrated meter that reads 20 sccm, multiply the reading and the GCF
for argon.
20 x 1.4047 = 28.094
Conversely, to determine what reading to set a nitrogen-calibrated meter in order to get a desired flow
rate of a process gas other than nitrogen, you divide the desired rate by the GCF. For example, to get a
desired flow of 20 sccm of argon flowing through the meter, divide 20 sccm by 1.4047.
Manual: 152-082010 301-305-306_303-307 Series
Page 34 of 35
Download from Www.Somanuals.com. All Manuals Search And Download.
20 / 1.4047 = 14.238
That is, you set the meter to read 14.238 sccm.
Some meters, specifically the high flow meters, are calibrated in air. The flow readings must be
corrected for the case where a gas other than air is flowing through the meter. In addition, there must
be a correction for the difference in the GCF from nitrogen to air. In this case, multiply the reading and
the ratio of the process gas' GCF to the GCF of the calibration gas. For example, a meter calibrated in
air is being used to measure the flow of propane. The reading from the meter is multiplied by the GCF
for propane divided by the GCF of air.
20 * (0.3499/1.0015) = 6.9875
To calculate a target setting (20 sccm) to achieve a desired flow rate of propane using a meter
calibrated to air, invert the ratio above and multiply.
20 * (1.0015/0.3499) = 57.2449
Gas conversion factors can be found at the Hastings Instruments web site.
Follow the link to Mass Flow Products and then to Gas Conversion Factors.
Manual: 152-082010 301-305-306_303-307 Series
Page 35 of 35