1. ALLGEMEINE INFORMATIONEN
Thank you for choosing this Soehnle Professional pro-
duct. All the features of this product were designed to
the state of the art and are optimised for simple and
straightforward use. If you have any queries or exper-
ience any problems with your appliance not addressed
in the operating manual, please contact your Soehnle
Professional service partner, or visit us on the web
Soehnle Professional will only assume responsibility
for the safety of the appliance if these instructions
are observed and the appliance is operated in com-
pliance with the operating instructions. The appliance
is a medical device and may only be used by persons
who are capable of handling it correctly as a result
of their training. Before using the appliance, the
user must ensure that the appliance is functional
and in proper working order. The operator must know
how to operate the appliance. The appliance is not
intended for use in medical rooms subject to explo-
sion hazards. Explosion-hazard areas may occur if
flammable anaesthetics, skin cleansing agents and
skin disinfectants are used.
■ Intended use
The appliance is designed for weighing people who
are lying down. The appliance may only be used for
this purpose. Any other use is prohibited. Messages
about errors or defects that could endanger the patient
or operator and errors that could lead to corruption of
the test results must be sent in writing to the manu-
facturer's medical product adviser (see address be-
low). These operating instructions as well as the
operating instructions for the 3020 Digital Indicator
are component parts of the appliance. They must
be kept near the appliance. Compliance with these
operating instructions is mandatory in order to en-
sure compliance with the intended use and correct
operation of the appliance. If the appliance fails to
function correctly, the appliance may be damaged.
It is essential that the appliance is then taken to an
authorised Soehnle Professional service centre. Only
genuine spare parts may be used if an authorised
service station carries out repairs. These genuine
parts are described in the service documentation
(Spare Parts List: 470.014.037 Scale / 470.630.273
Digital Indicator / 470.014.038 Weighing Device)
including Order Number.
■ CE sign
The product carries the CE sign in accordance with
the following directives:
•
•
•
•
89/336/EEC Electromagnetic Compatibility
73/23/EEC Low Voltage Directive
90/384/EEC Nonautomatic Weighing Instruments
93/42/EEC Medical
Products (Class I
with measurement
function)
Soehnle Professional GmbH & Co. KG.
Manfred-von-Ardenne-Allee 12 71522 Backnang Germany
■ Classification
•
•
Class I medical product with measuring function.
Electrical Safety Class II
(with double isolated function ground)
No applicator
Waterproof to IEC529:IPX2
•
•
3
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
■ Safety instructions
■ Applicator
Before putting the appliance into use,
The scale does not have an applicator. No part is in-
tended for direct contact with the patient. Leakage
currents were measured on the weighing devices in
compliance with classification of Type B applicators.
please read with care the information given in the
Operating Instructions. They contain important in-
structions for installation, intended use and mainte-
nance of the appliance. The manufacturer shall
not be liable if the following instructions are
not observed:
In case the electrical components are used under
increased safety requirements, compliance with the
appropriate regulations is compulsory. Never perform
work on the appliance while the power is switched
on. Improper installation will render the warranty null
and void. Ensure the voltage marked on the power
supply unit matches your mains power supply. The
device is designed for use within buildings. Observe
the permissible ambient temperatures for use (Tech-
nical Information). The device meets the requirements
for electromagnetic compatibility. Do not exceed the
maximum values specified in the applicable stand-
ards. If you have any problems, contact your local
Soehnle Professional service partner.
■ Operating concept
The appliance is operated by means of the keypad.
The keys are assigned with different functions de-
pending on the various modes.
■ Maintenance and service
The measuring inspection test was performed during
the initial calibration process in production. Further
regular measuring inspection tests (recalibration) must
be carried out by the calibration authorities respon-
sible in compliance with national regulations. Repairs
may only be carried out by a centre authorised by
Soehnle Professional using genuine spare parts.
The medicine device is equipped with a non-de-
tachable power line. The power line may not be ex-
changed by maintenance personal.
Warning: A change of the medicine device is pro-
hibited. This device may not be changed without the
manufacturer's authorisation. If the medicine device
is changed, appropriate analysis and examination
must be undertaken to ensure a safe continued use.
■ Cleaning
Before cleaning, disconnect the appliance from the
mains by removing the power plug from the wall
socket. Only clean the appliance using a moist cloth.
On no account may water be allowed to enter into
the appliance. Disinfectant may only be applied to
the membrane keyboard for the display. It is prohi-
bited to spray the appliance and the power plug.
The following disinfectants are permitted: methy-
lated spirits, Isopropanol; 2% Kohrsolin, 1% aqueous
Sokrena solution, 5% Sagrotan or 5% Gigasept.
■ Use of additional devices
Only additional appliances (printer, computer) may
be connected to the serial data port if they conform
to European Standard EN 60601-1 or a suitable dis-
connecting device is connected in series.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
■ Disposal
sunlight may have an effect on the accuracy of the
scale. The scale is protected to IP 54 protection class.
Therefore, avoid high levels of humidity, vapours,
aggressive liquid media and strong contamination.
According to the present state of the art
the appliance contains no special environ-
mentally harmful substances. This product
is not to be treated as regular household
waste, but should be handed in to an electrical/elec-
tronic equipment recycling centre. You can obtain
further details from your local council, your munici-
pal waste disposal company or the firm from which
you purchased the product.
■ Warranty/liability
If the appliance is delivered with a fault or defect
which is within Soehnle Professional's scope of re-
sponsibility, Soehnle Professional is entitled to either
repair the fault or supply a replacement appliance.
Replaced parts remain the property of Soehnle Pro-
fessional. Should the fault repairs or replacement
delivery not be successful, the statutory provisions
shall apply. The period of warranty shall be 2 years,
beginning on the day of purchase. Please retain your
receipt as a proof of purchase. If your scale requires
servicing, please contact your dealer or Soehnle Pro-
fessional Customer Service. Only genuine spare parts
may be used if an authorised service station carries
out repairs. Soehnle will only assume responsibility
for the safety of the appliance if these instructions
are observed and the appliance is operated in com-
pliance with the operating instructions.
Battery disposal: Do not dispose of batteries in the
household rubbish. As consumer you are obliged by
law to return waste batteries. You can hand in your
old batteries as hazardous waste to the public col-
lection centres in your municipality or return them
to the firm where you purchased them. You will find
these labels on batteries containing hazardous sub-
stances: Pb = battery contains lead, Cd = battery
contains cadmium, Hg = battery contains mercury.
Lead-gel batteries are used to power the weighing
devices. They must also be disposed of as hazard-
ous waste and returned to the dealer when pur-
chasing new batteries.
■ Note
No responsibility shall be accepted for damage
caused by any of the following reasons: unsuitable
or improper storage or use, incorrect installation or
putting into operation by the owner or third parties,
natural wear and tear, changes or modifications, in-
correct or negligent handling, excessive use, chemi-
cal, electrochemical or electrical interference or hu-
midity, unless this is attributable to negligence on the
part of Soehnle Professional. If operating, climatic or
any other influences lead to a major change in con-
ditions or material quality, the warranty for perfect
Interference suppression of the unit corresponds to
the valid EU Directive 89/336/EEC. The display value
may be influenced by extreme electrostatic and electro-
magnetic interference, e.g. operation of a radio unit
or a mobile phone in the immediate vicinity of the
unit. When electrostatic interference is no longer
present, the appliance can again be used for its in-
tended purpose. You may need to switch it on again.
The appliance is a measuring instrument. Air currents,
vibrations, rapid changes in temperature and direct
5
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
unit functioning shall be rendered null and void. If
Soehnle Professional provides an individual warranty,
this means that the appliance supplied will be free
from faults for the duration of the warranty period.
Always keep the original packaging in case you have
to return the scale.
Soehnle Professional GmbH & Co. KG
Manfred-von-Ardenne-Allee 12 · DE-71522 Backnang
Phone +49 (0) 71 91 / 34 53-0 · Fax -211
6
Download from Www.Somanuals.com. All Manuals Search And Download.
2. PUTTING INTO OPERATION
■ Installing the scale elements
The indicator's placement must allow for a simple
disconnection from the power supply.
Insert rechargeable batteries in all four weighing
devices. When the batteries are fitted, the weighing
devices start to send and attempt to set up a radio
link to the digital indicator.
1
2
3
Caution (Fig 1–4 right):
Before switching on the scale, place the weighing
devices under the four wheels of the hospital bed.
Push the weighing devices as far as the stop under
the wheels. Press the lever down and raise the bed
wheel. (The weighing devices are designed to re-
ceive wheel diameters of 130–160 mm. Small diam-
eters are optional.)
4
The message "SOEHNLE PROFESSIONAL" appears in
the indicator display together with a search icon until
the radio link is set up to the weighing devices. This
normally takes 20 to 60 seconds.
■ Power-on / Power-off
Switch on or off the digital indicator
by pressing the On/off switch.
When the radio link to the weighing devices is set
up, the indicator display shows the current gross
weight value.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
■ Radio link
■ Charging the batteries
Open the cover of
the battery charger.
The charger has
four compartments
to recharge the four
The Bluetooth radio link operates over a max. radius
of 5m if there is no partition wall between the weig-
hing devices and the digital indicator. After short inter-
ruptions (e.g. to replace a battery or caused by the
loss of battery contact due to shocks), the link is
restored automatically within a few seconds. During
batteries simultane-
ously. The four LEDs
indicate the charge state for each battery. The fifth
LED on the charger is the mains power indicator for
the charger itself.
connection process appears the following symbol
.
LED status indicator on charger (for each charge
compartment):
•
•
LED lights up orange: battery still charging
LED lights up green: battery charged – trickle
charge
■ Power-off
If the sensors are overloaded or you
have started a menu, the scale swit-
ches off after you press the On/Off
button for 5 seconds.
LED status indicator on weighing device:
Digital indicator is switched on (radio link active)
It is not possible to switch off the sca-
le as long as it is waiting for an input in the editing
section of Setting mode. First exit the editing section
before switching the scale off.
•
•
LED lights up green: battery charged
LED flashes green, mainly on, off for short
periods: battery still has sufficient capacity
for normal operation
The weighing devices have no on/off switch. They are
only switched off when the batteries are removed.
•
LED flashes green, normally off, only on for
short period : battery almost empty, battery
must be recharged
OFF display
When the scale is switched off but remains connected
to the power supply, the indicator display shows
"OFF" (in accordance with the Medical Equipment
Directive).
Digital indicator is off or in standby mode
•
LED lights up green for approx. 2 mins
and then starts to indicate standby mode
by flashing (mainly off)
The charging time for empty batteries is approx. 4
hours. Batteries should be charged and exchanged
in sets. The batteries may remain for any length of
time in the charger when it is on.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Display of link state:
Recommendation:
Press Info key + 1. The following appears in the
indicator display:
Always remove the batteries from the weighing de-
vices after you switch off the digital indicator and
place them in the battery box of in the charger.
This makes sure that the scale works reliably the
next time it is switched on.
Weighing device
State
OK
Error
OK
Battery voltage
6.3 V
1: 0080025005BB7
2: 0080025005C0D
3: 0080025005C70
4: 0080025005ADC
6.3 V
6.3 V
If the bed scale remains without use, remove the
batteries from the weighing devices to prevent
deep discharge. The scope of supply comprises
eight batteries.
OK
Charge run time:
A battery charge is sufficient for approx. 12 hrs con-
tinuous duty. When the scale is switched off for most
of the time (90% OFF), the batteries may remain in
the weighing devices for up to 125 hours before they
require a recharge.
Caution: Even when the digital indicator is switched
off, the Bluetooth link is activated at regular inter-
vals. For this reason, the batteries will discharge in
this state and must be recharged at the latest after
125 hours.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
3. INPUT AND CONTROL
■ Keypad for input and control
KEYPAD: To enter nume-
rical values or alphabetic
characters (alphanumeric
ORG key: Calls up
Organisation Data
INFO key: e.g. electro-
nic nameplate contai- keypad)
ning reference data
0 (zero) key:
Press several
times to display +
(plus), – (minus)
and the blank
character
F key: function key /
cursor up /
C key:
Deletes the
last numeral
or character
cursor left
ENTER key/
confirmation key
TARE key /
cursor down /
cursor right
ON/OFF key
ZEROING key
SPECIAL CHARACTER key
10
Download from Www.Somanuals.com. All Manuals Search And Download.
■ Alphanumeric input
racter pressed is accepted and the cursor moves
on to the next position.
It is possible to make alphanumeric inputs with the
combined numeric and alphabetic character keypad.
Numerical inputs can be made in direct succession
without any waiting period, as long as the same
number key is not pressed several times in successi-
on. A cursor flashes under the last item entered.
Press the F key and the TARE key to move the cursor
to the left or right. The C key clears the last character
on the left of the cursor or the character with the
cursor flashing underneath it.
Types of input box
•
•
•
Display fields, read only
Display/writing fields with numeric input option
Display/writing fields with alphanumeric input
option
Input procedure
Numeric fields: Type in the numerals 0 to 9 directly
into the KEYPAD. Input is right-flush. A cursor flashes
under the last digit entered. The C key clears the
numeral highlighted by the cursor. The ZEROING key
aborts the input without saving the new input values.
Existing values are retained. The Enter key saves the
value typed in and returns to the application pro-
gram. The setpoint, whether it is a dot, comma or
no decimal separator, is defined in Setting mode
(Setting mode > Terminal > Display > Decimal sepa-
rator, see p 14). Only the default number of characters
can be typed into an input box. Any other inputs
are neither displayed nor accepted.
The ZEROING key aborts the input without saving
the new input values. Existing values are retained.
The ENTER key saves the value typed in and returns
to the application program. Only the default number
of characters can be typed into an input box. Any
further input overwrites the last character.
Alphanumeric fields: The numeric keys are also as-
signed by multiple programming to alphanumeric
characters (upper case only) and characters. For
example, the 0 (zero) key is also a +, – and blank
character. The SPECIAL CHARACTER key is also avail-
able.
After pressing a numeric key twice within one se-
cond, this activates a rolling function to access a
character assigned to this key. If you do not press
the key a second time within one second, the cha-
11
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
■ Indicator display
The display shows a number of different views de-
pending on the situation. It is divided into display
panes.
■ Icons/characters
Icons in the icon bar
Alibi memory capacity 100%
Additional display
Icon bar
Alibi memory capacity 80%
Alibi memory capacity empty
Keylock
Display panel
Additional display
It displays BMI, Gross, Net, etc. If required, it can be
switched on with the F key according to the program
settings.
Multirange/Multidivision 1
Multirange/Multidivision 2
Multirange/Multidivision 3
Resolution x10 (10 times)
Zero
Icon bar
Shows information about the status of the active
measuring point.
Display panel
Shows weight value and unit characters.
Hold function active
Unit character
Symbol for standstill
After placing a load on the scale (bed, patient, etc.),
the scale requires time to finish oscillating. Until it
comes to a standstill the unit character (kg, lb) re-
mains blank. The unit character only appears when
the scale reaches standstill and a stable value is
obtained (clip off).
12
Download from Www.Somanuals.com. All Manuals Search And Download.
4. SETTING MODE
■ Keypad in Setting mode
ORG key:
Exits the Setting
mode
KEYPAD
F key: Moves the marking
arrow up the list or moves
the cursor to the left
C key:
PRINT/ENTER key:
Clears the
last numeral/
character
Calls up the marked
menu option or saves
an input or change
T key (TARE key):
Moves the marking arrow
down the list or moves the
cursor to the right
ZEROING key: Returns
ON/OFF key
to the next higher level
(back without saving
the change)
13
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
■ Overview of menu navigation
The following setting options are available:
Hold mode
Fixed tare value
Clear fixed tare value
Auto standby time
Plausibility check sensor
(p 20)
(p 17)
(p 19)
General
Program settings
(p 23/Error 12)
Provisional tare
Dialysis function
Hold mode
Second unit
BMI function
(p 19)
(p 19)
(p 20)
(p 14)
(p 20)
Weighing + taring
F key (p 22)
Version
Display
Keypad
Date/Time
(p 14)
(p 14)
(p 14)
(p 15)
Terminal
Calibrated counter status*
approvable*
Organisation data (p 15)
Filter setting*
User password
Reset terminal
(p 15)
(p 15)
Zero tracking*
Zero range at power-on*
Second unit*
Empty message*
Serial number MS*
Software version MS*
Scales
Scale parameters
(p 26)
Data transfer fixed values
Alibi memory
(p 18)
Data Processing
Application
Baud rate*
…*
(p 11)
(p 21)
(p 21)
Interface 1
Print template
Printout request
Printout when W+T=O
Decimal separator
I/O port settings*
Barcode*
Special thousands separator
Hide unit
Consec. no. print key
Factory setting
Service
(password-protected)*
* Only for service personnel authorised by Soehnle Professional
14
Download from Www.Somanuals.com. All Manuals Search And Download.
■ Calling up Setting mode
■ Settings in "Terminal" menu
In "Terminal" Setting mode, you will find the follow-
ing request and setting options.
Switch the appliance on. After the
power-on routine, the appliance starts
the last activated application program
automatically. To call
Version
up Setting mode first
press ENTER and then the F key
simultaneously. You are now in
the Setting mode selection menu.
AZG software version
AZG applications software
AZG serial number
Display only
Display only
Display only
Display only
+
CPU board serial number
■ Inputs in Setting mode
In most cases you can select from several different
settings available. With editable values, you can type
in values into the numerical or alphanumeric keypad.
Display
Backlight
Sets the luminosity, enter
as a percentage into the
numerical keypad; 0% = dark,
100% = fully luminosity;
factory setting: 80%
■ Settings in Setting mode
With the parameters in Setting mode, you can
adapt the weighing and taring application program
to your specific requirements.
Contrast
Contrast setting, enter as
a percentage into the nu-
merical keypad; 1% is low-
est contrast, 100% is full
contrast; factory setting 50%
Deutsch/English/François
Comma, point or none
Factory setting: comma
g/t/lb/mg/oz (additional
field in display)
Language
Decimal separator
■ Settings in "Program Settings"
menu
In the "Program Settings" Settings menu, you can
adapt the Application program of your scale to your
specific requirements – the bed scale mainly uses
"Weighing and taring".
Second unit
Keypad
Keylock
Individual, several or all
keys on the keypad can be
locked. Press the key to be
locked to change the status
of the key from "unlocked"
to "locked" and vice versa.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
To accept, press ENTER for
3 seconds.
mode is only accessible by
password. If you forget your
password, type in the word
CLEAR to release access.
Clears the password and
releases access
Keytone
Select "off" or "on". Confir-
mation tone when a key is
pressed. Factory setting: "off"
Select "off" or "on" Valida-
tion tone when a function
is executed. factory setting:
"off"
Clear
Validation tone
Reset terminal
Do not execute/
Executes
"Execute" clears all
all the individual terminal
settings and resets to fac-
tory settings
Date/time
Time
Enter into numerical key-
pad. Press ENTER to start
clock after confirmation
Enter into KEYPAD
Date
Date format
Select European or imperial
date; factory setting: Euro-
pean
Organisation data
Description of ORG A Enter the description of the
organisation data memory
A; numerals or text entered
into KEYPAD, e.g. "Bed
number", "Room number"
max. 12 characters
Description of ORG B/ Further input options
ORG C/ ORG D
for other organisation data
memories, see above
User password
New/change
Enter password with up to
8 characters into keypad;
after activation, the Setting
16
Download from Www.Somanuals.com. All Manuals Search And Download.
5. FUNCTIONS
■ Power-on zeroing
■ Weighing and taring
Display view
During the power-on procedure, the weight values
sent by the weighing devices are zeroed until they
are less than 50% of the max. weighing range. This
tares the empty bed and the weight display goes
to zero.
■ Manual zeroing
Manual zeroing corrects minor devia-
tions from zero, e.g. caused by soiling
on the scale. The following character
appears in the icon bar when the dis-
play indicates zero: >0<.
■ Weighing with measured tare
Determine and tare a bed weight without patient.
Position the empty bed, including bedding, on the
scale. Press the TARE key. The tare value is saved
and the weight display goes to zero. Now the pati-
ent can be laid on the bed. The weight is displayed
with the message "Net".
Zero range approvable and non-approved: 1% to
+3% of the weighing range.
If zeroing is not possible, the error message "Value
above zero range limit" or "Value below zero range
limit" appears for two seconds in the display.
Displaying the value in the tare memory
Press the INFO key to display the tare memory and
then the TARE key. Press the ZEROING key to termi-
nate the display.
■ x10 (times 10) display
Shows the weight value with
an additional decimal place at
ten times the resolution. Switch
Changing the tare value manually later
Display the tare value by pressing the INFO key and
then the TARE key. Enter the new value into the
KEYPAD. The ZEROING key cancels the input. Exi-
sting values are retained or new tare values are sa-
ved by pressing the TARE key. Confirm by pressing
ENTER and return to the application program.
to the x10 display by pressing
+
ENTER and then the ZEROING
key simultaneously. According to the calibration
law, the x10 function may only be activated on ap-
provable scales for a short period of time. For this
reason, it s programmed so that the function is
switched off automatically after 5 seconds. The
message x10 appears in the icon bar as long as the
10-time higher display is selected. When the app-
liance is switched off, the function is deactivated.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
■ Weighing with manually
entered tare
If the patient is already on the bed and the
bed weight is known, then this can be tared by
manual input.
Entering tare weights into the fixed tare
value memory
•
Call up the input menu
(Setting mode > Program Settings > General
> Fixed tare values)
Enter the known bed weight (tare weight) using
the KEYPAD. The input appears in the display with
the current weight unit. Press the TARE key. The
nput is transferred into the tare memory and the
tare weight is deducted from the weight display.
The patient's weight measured in this way is dis-
played with the message "Net".
•
•
Press ENTER and F key simultaneously to access
Setting mode.
The arrow points to "Program Settings". Confirm
by pressing ENTER.
Confirm the message "General" by pressing ENTER.
Press TARE to browse to "Fixed tare values" and
accept by pressing ENTER.
•
•
Showing the value in the tare memory
Display the tare memory by pressing the INFO key,
then press the Tare key. Press the ZEROING key to
end the display.
Entering values
•
•
If no tare memory has yet been assigned, the
first free, or otherwise the last, selected fixed
tare value memory is displayed.
Press ORG followed by ENTER. The input field
for the description of a new tare memory is dis-
played.
Enter the description into the KEYPAD and con-
firm by pressing ENTER.
The arrow points to "Tare value". Press ENTER.
The input field is displayed to enter the tare value.
Enter the tare value and confirm by pressing
ENTER.
Go back by pressing the ZEROING key to select
the tare memory.
If you want to assign other tare memories, press
the ORG key to select the next free memory lo-
cation and repeat the procedure as described
above.
Changing the tare value manually later
Show the tare value by pressing the INFO key and
then the TARE key. Enter the new value into the
KEYPAD. Press the ZEROING key to cancel the input.
The existing value is retained or the new tare value
is saved by pressing the TARE key. Confirm by pres-
sing ENTER and return to the application program.
•
•
•
•
•
■ Weighing with stored tare values
Known bed weights can be stored as the tare value
in a power failsafe memory and called up as required
(100 different bed weights are possible).
Each memory location has a:
•
To return to weighing mode, press the ZEROING
key and then the ORG key.
•
•
•
Three-digit consecutive number
Description (text, 18 characters)
Tare value
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Transferring fixed tare values from a PC to the
digital indicator
•
•
If you want to change other tare memories,
press the F key of the TARE key to select the
next memory location to be changed and repeat
the procedure as described above.
To return to weighing mode, press the ZEROING
key and then the ORG key.
•
Here you require a special software. For more
information, please contact the Soehnle Profes-
sional Service personnel.
Changing or overwriting a fixed tare value memory
•
Call up the Input menu
(Setting mode > Program Settings > General
> Fixed tare values)
Press ENTER and the F key simultaneously to
access Setting mode.
The arrow points to "Program Settings". Confirm
by pressing ENTER.
Confirm the message "General" by pressing ENTER.
Press the TARE key to browse to "Fixed tare va-
lues" and accept by pressing ENTER.
Calling up the list of fixed tare value memories
for taring
The tare value from the fixed tare value memory
can be activated with the scale unloaded (i.e. be-
fore the weighing operation) or with the scale loa-
ded (during a weighing operation).
•
•
•
•
The display pane shows a weight value ≥ 0.000 kg.
First press ENTER and then the TARE key simul-
taneously. The fixed tare value memory last used
appears in the display.
•
•
•
•
•
Press the F or TARE key to browse to the required
fixed tare value memory or type in the memory
location number and confirm by pressing ENTER.
Press ENTER to activate the fixed tare value.
The display shows the gross value minus the
tare value (net).
Correcting values
•
The last selected fixed tare value memory is dis-
played. Press the F or TARE key to select the tare
memory to be changed or enter the memory lo-
cation number and confirm by pressing ENTER.
Press ENTER. The input field for the tare location
description is displayed.
Change the description by typing it into the
KEYPAD and confirm by pressing ENTER.
The arrow points to "Tare value".
Press ENTER. The input field for the tare value
is displayed.
Change the tare value by typing it into the KEY-
PAD and confirm by pressing ENTER.
Go back to the tare memory selection by pressing
the ZEROING key.
•
•
Further tare values from the memory are activated
by repeating the procedure described above.
•
•
Clearing an existing fixed value
•
Call up the input menu
(Setting mode > Program Settings > General
> Fixed tare values)
Press ENTER and the F key simultaneously to
access Setting mode.
The arrow points to "Program Settings". Press
ENTER to confirm.
•
•
•
•
19
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
•
•
Confirm the message "General" by pressing ENTER.
Press the TARE key to browse to "Fixed tare va-
lues" and accept by pressing ENTER.
•
•
Exit Setting mode by pressing the ORG key. The
gross value is displayed.
Using the dialysis function
Clearing a tare value
Press the F key (with a load on the scale). Mes-
sage: "Place tare". Place the additional tare on
the scale. Confirm by pressing TARE. The addi-
tional tare is accepted into the tare memory and
the net value remains unchanged.
Repeat this procedure for other additional tares.
The weight is displayed with the message "Net".
When the scale is clipped off, the tare weight is
displayed with a negative sign.
•
The last selected fixed tare value memory is
displayed. Press the F or TARE key to browse
the required fixed tare value memory or enter
the memory location number and confirm by
pressing ENTER.
Clear the displayed fixed tare value memory
by pressing the C key.
The gap in the consecutive number scheme is
closed. The following fixed tare value memories
are given a number reduced by 1.
•
•
•
•
Clear the tare by pressing the ZEROING key.
Displaying the value in the tare memory
To display the tare memory, press the INFO key and
then the TARE key. Press the ZEROING key to exit
the display.
CAUTION: Under Setting mode > Program Set-
tings > General > Delete fixed tare values, all
fixed values are cleared from the memory!
Changing the tare value manually later
■ Provisional tare/dialysis function
Display the tare value by pressing the INFO key and
then the TARE key. Enter the new value into the
KEYPAD. Press the ZEROING key to abort the input.
The existing value is retained or the new tare value
is saved by pressing the TARE key. Confirm by pres-
sing ENTER and return to the application program.
Setting and using the operating mode
•
Setting in Setting mode
(Setting mode > Program Settings > Weig-
hing + Taring > F key > Dialysis function)
Press ENTER and the F key simultaneously to
access Setting mode.
The arrow points to "Program Settings". Press
ENTER to confirm.
•
•
•
Press TARE to browse down to "Weighing + Ta-
ring" and press ENTER to confirm.
•
•
Confirm the message "F key" by pressing ENTER.
Press TARE (down) or the F key (up) to browse
to "Dialysis function" and press ENTER to confirm.
There is an acoustic acknowledgement.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
■ Error message when taring
"Error 06 Taring not possible"
Possible causes:
Setting and using the operating mode
•
Setting in Setting mode(Setting mode > Pro-
gram Settings > Weighing + Taring > F key >
BMI function)
•
•
•
Scale empty
Scale in overload or underload
Manual tare input with tare value above
maximum load
•
•
•
Press ENTER and the F key simultaneously to
access Setting mode.
The arrow points to "Program Settings". Press
ENTER to confirm.
Press the TARE key to browse down to "Weig-
hing + Taring" and press ENTER to confirm.
Confirm the message "F key" by pressing ENTER.
The arrow points to the last function activated.
Press the TARE key to browse to "BMI function"
and press ENTER to accept. A beep tone confirms
acceptance.
■ Displaying the tare memory
To display the tare memory, press the INFO key and
then the TARE key. Press the ZEROING key to end
the display.
•
•
•
■ BMI function
•
Exit Setting mode by pressing the ORG key. The
weight value is displayed.
The Body Mass Index (BMI) is used to assess whe-
ther a person has a normal weight or is overweight.
The BMI is calculated according to the following
formula:
Using the BMI function
•
•
•
Press the F key. The following message appears:
"Enter body height". Enter your body height in
metres.
Confirm by pressing ENTER. Now, parallel to the
person's weight, the BMI value is displayed in
the additional display.
body weight (kg)
body height (m) x body height (m)
BMI =
The unit for the BMI is therefore kg/m2.
BMI classification separated according to gender:
The BMI display can be switched off by pressing
the F key again.
Classification
Underweight
Normal weight
Overweight
Adiposity
Severe adiposity
m
<20
20–25
25–30
30–40
>40
w
<19
19–24
24–30
30–40
>40
■ Hold function
•
Setting in Setting mode
(Setting mode > Program Settings > General
> Hold mode)
The Hold function "Hold mode" freezes the measu-
red value in the display dependent on the selected
setting. The following settings are possible:
21
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
not active: Hold mode is off.
Configuring the printer port
Still/Key: Value freezes.
•
Setting in Setting mode
Press the ON/OFF to cancel Hold mode.
Still/Empty: Value freezes.
(Setting mode > Data Processing > Interface
1 > Application > 2795.11)
Clip off the scale to cancel Hold mode.
Max/Key: Max. value freezes (at standstill).
Press the ON/OFF key to cancel Hold mode.
Max/Empty: Max. value freezes (at standstill).
Clip off the scale to cancel Hold mode.
Delay/Key: Max. value freezes (movement).
Press the ON/OFF key to cancel Hold mode.
Delay/Empty: Max. value freezes (movement).
Clip off the scale to cancel Hold mode.
•
•
Press ENTER and the F key simultaneously to
access Setting mode.
The arrow points to "Program Settings". Press
the TARE key to browse to "Data Processing".
Press ENTER.
The arrow points to "Data transfer". Press the
TARE key to browse to "Interface 1". Press ENTER.
The arrow points to "Application". Press ENTER.
The arrow points to "2795.11". Press the TARE
key or the F key to select the printer to be con-
nected.
•
•
•
■ Print
•
Press ENTER to confirm.
The 3020 Digital Indicator can be connected to a
printer via the serial port to print out scale chits. Up
to a max. of 15 print templates can be stored. You
can obtain a service software from Soehnle Profes-
sional service centres for the 3020/3030 to design
print templates with customised contents.
Connection options (factory setting):
none
2795.11
2795.12
(Soehnle Professional printer)
(Soehnle Professional printer)
Data Processing (PC)
Barcode
(barcode reader)
The parameters listed in the standard print templates
can be used to design user-defined print templates.
In addition, organisation data (identifiers) and user-
defined parameters (such as company names) can
be included. Please contact your Soehnle Professional
service partner about your print template requirements.
Start the print routine either by pressing ENTER or
by pressing another print command, e.g. the F key.
Large-format
display S302
(monitor)
Info about printer
Soehnle Professional printers are stored as standard.
However, you can also set up other printers using
the 3020/3030 Service Software. Other parameters
(baud rate, data bits, parity and XON/XOFF) can be
assigned as specified in the printer manual. Please
contact service personnel authorised by Soehnle
Professional. The parameters set at the factory are
configured for Soehnle Professional printers.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Default print templates, factory setting
•
•
ENTER/Accept key when ORG D is entered
Key combination INFO, F and ENTER
(dependent on settings)
ENTER/Accept key for manual input
Print alibi memory
The print-command keys can be assigned using the
print templates in the service software. In the fac-
tory setting, the print template below is already
stored and assigned. In other words, it can be prin-
ted out. This default print template is an example
showing what parameters are available.
•
•
■ Programming the F key
The F key is programmable to call up various func-
tions. It assumes different functions depending on
how it is assigned. The current function is selected
in Setting mode.
Date
Time
Scale no.
Gross
Tare
01.09.08
09:13:28
02
W
T
0.154 kg
0.015 kg
Access Setting mode by first pressing ENTER and
then the F key simultaneously.
In Setting mode you will find the F key program-
ming options under Program Settings > Weighing +
Taring > F key.
Select the required function by pressing the F or
the TARE key and confirm by pressing ENTER.
Print-command conditions
The print templates are assigned to print-command
conditions using the service software.
The following conditions can be assigned to
print templates:
•
•
•
Print key (ENTER)
The following functions are available:
Provisional tare
Automatic print/data record once immediately
Automatic print/data record once after weight
change at standstill
Dialysis function
Hold mode
•
•
•
Automatic print/data record after weight
change at standstill
Second unit
BMI function
Automatic print/data record after weight
change at standstill above the empty message
Automatic print/data record after weight change
at standstill above the empty message if the
active empty message was previously active.
Continuous transmission
ENTER/Accept key when ORG A is entered
ENTER/Accept key when ORG B is entered
ENTER/Accept key when ORG C is entered
Exit Setting mode by pressing the ORG key.
•
•
•
•
23
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
6. Error messages
■ Error numbers > Cause / Remedy / Info
05 Zeroing not possible
23 Print error
> Deviation from zero point too large
24 Print template memory corrupt
25 No print template(s)
06 Taring no possible
> Empty scale, scale underloaded or overloaded,
manual tare input with tare value above max. load
26 Printer update cancelled
30 Alibi memory full
07 Print not permitted
08 No mass unit switchover
09 Print signal buffer full
10 Load cell signal overflow
11 No measuring point signal
33 Alibi memory contents corrupt
34 Alibi memory empty
35 No or defective alibi memory
36 Cannot write alibi
12 Check sensor position
> Incorrect lever setting for at least one
weighing device or no weigh on at least one
weighing device (plausibility check active)
37 Alibi memory unreadable
40 General data format error
41 Input value too small
42 Input value too big
13 Converter failure
15 Measuring point write-protected
16 No measuring point detected
18 Measuring point not zeroed
20 Print format empty
44 Power-off not allowed
50 Value below zero range limit
> Zeroing not possible because measured value
is outside the set zero range limit
24
Download from Www.Somanuals.com. All Manuals Search And Download.
51 Value above zero range limit
> Zeroing not possible because measured value
is outside the set zero range limit
52 Overload
> Measured value above weighing range
53 Underload
> Measured value below 0 (zero)
55 Place weight on scale
> Scale empty or no clip off setting made
according to set clip off factor
75 Replace backup battery
76 Charge batteries
80 Abort external command
81 Timeout external command
82 Number exists already
> The identification assigned to a fixed value
exists already
90 Fixed value memory occupied / full
> A total of 100 memory locations are available
valid for all programs
98 Wrong barcode
25
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
7. TEChNICAL INFORMATION
■ Technical data
Weighing range:
Division:
Display:
Dot matrix 240 x 64 pixels,
dimmable, backlighting
Fixed value memory for to-
tal of 100 fixed values, va-
lid for all programs, power
failsafe, alibi memory for 2
million inputs optional
1 x RS 232, other interfaces
optional, Bluetooth Quali-
fication
Digital Indicator: housing
plastic; weighing devices:
aluminium
130–160 mm ø,
max. width 71 mm
see nameplate
see nameplate
see nameplate
250 kg
Memory capacity:
Min/max load:
Max. bed weight:
Approval number:
D07-09-034-PTB (0102)
Ambient operating temperature:
–10°C to +40°C, rapid
Interfaces:
Materials:
Wheels
temperature changes, e.g.
in the event of cold drau-
ghts, may impair the test
result
Storage temperature:
Humidity:
–40°C to +70°C
20%–85% relative humidity,
non condensing
Digital indicator power supply:
power adapter/mains voltage
100–240 V / 50–60 Hz
Weighing device power supply:
■ Other technical data
Diagrams, lists of spare parts, and setting instruc-
tions can be requested through the manufacturer's
Customer Service.
From 6 V lead-gel battery,
battery capacity: 1200 mA
continuous duty approx. 12
hours
Transportation conditions:
■ Instructions for installing
the scale
The appliance may only be
transported in its original
packaging, comply with
storage temperature, avoid
vibrations and shocks
For design reasons, the scale is designed only for use
on level floors. For installation sites in Germany, this
condition is met when the floor complies with the
tolerance specifications for flat floors to DIN 18202,
Table 3, Row 4. For other countries the applicable
national standards should be taken as a basis.
Membrane keyboard: 19 keys, function key freely
programmable
Organisation data memory:
4 memories, alphanumeric
input via the numerical
keypad
26
Download from Www.Somanuals.com. All Manuals Search And Download.
■ Electronic nameplate and calibration counter
The appliance has an electronic nameplate.
Display by pressing the INFO key. Exit the display by pressing the INFO key again.
(sample display)
■ Explanations of labels and icons
Weighing ranges
Max
Maximum load of weighing range
Minimum load of weighing range
Calibration value (division)
07
XXXX
D07-09-034-PTB
Min
e=
SNR
Scale serial number (scale type, final number of
year of manufacture, incremental numeral)
CE mark of conformity with year of affixing and simul-
taneous commencement of validity of calibration
07
Precision class
D07-09-034-PTB
No. of EU approval
TYP
Manufacturer's product number
XXXX
E.g. 0102, Official Code of "Office Quoted"
(this body performed the initial calibration process)
27
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
Symbol for EU calibration
The calibrated counter status indicates how often
the scale was calibrated. The stored counter status
must match the calibrated counter status on the
inserted label secured by adhesive label.
Load Cell
SN
Serial number of weighing machine
Product number der Lastaufnehmer
SN:H2509xx
Typ:SEB46B-300-C3-MR-CC-6L
TYP
Soehnle Professional GmbH & Co. KG.
Manfred-von-Ardenne-Allee 12 71522 Backnang Germany
BT
IPX2
Bluetooth Address
Protection Class
Power Supply
Typ:7711
SN:771110070001 BT:008025005BBF
IP X2 6V 80 mA
6V DC
80mA
Soehnle Professional GmbH & Co. KG
Manfred-von-Ardenne-Allee 12 71522 Backnang Germany
Power Input
Attend to user manual, ISO 7010
Equipment for not ionized radiation, IEC 60417
Functional earth, IEC 60417
28
Download from Www.Somanuals.com. All Manuals Search And Download.
ENGLISH
Bed scale 7711
Soehnle Professional GmbH & Co. KG
Manfred-von-Ardenne-Allee 12
DE-71522 Backnang
Phone +49 (0) 71 91 / 34 53-0
Fax
+49 (0) 71 91 / 34 53-211
Scales, System Solutions And Special Developments
07/2010 470.051.103