Quantum Medical Alarms 9412 9112 A User Guide

1 INTRODUCTION  
This manual is designed to help you quickly and easily learn to use your Quantum Composers  
Model 9412/9412A pulse generator. Rather than the usual array of knobs and switches, the  
9412/9412A uses a menu-driven user interface with "on-line" help -- once you learn the basics  
of operating your 9412/9412A, you may never need to refer to this manual again.  
The 9412/9412A has some unique features that are designed exclusively for use as a laser  
trigger signal generator, but can also be used as a general purpose pulse generator. Because  
it employs a flexible microprocessor-controlled architecture, the 9412/9412A can be custom-  
ized to fit your exact requirements -- contact Quantum Composers for details.  
Technical Support  
For questions or comments about operating the 9412/9412A, contact Quantum Composers via  
one of the following methods:  
P Phone - (406) 582-0227  
P Fax / Modem - (406) 582-0237  
Parts List  
The following parts are included with the 9412/9412A -- contact Quantum Composers if any  
parts are missing:  
P 9412/9412A Pulse Generator  
P AC Power Cord  
P User's Manual  
P Quick Reference Card  
Warranty  
In addition to a 30-day money back guarantee, the 9412/9412A has a one-year limited war-  
ranty from the date of delivery. This warranty covers defects in material and workmanship.  
Quantum Composers will repair or replace any defective unit. Contact us for information on  
obtaining warranty service.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Never work alone. Someone should always be nearby to render aid if necessary. Training in  
CPR first aid is highly recommended.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Item Button  
1
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 CONTROLS AND CONNECTORS  
This chapter explains the function of each control and connector on the 9412/9412A.  
Front Panel  
The front panel contains the most frequently used controls and connectors.  
Display  
The 9412/9412A features a backlit liquid crystal display, which displays menu selections and  
settings. The left half of the display shows the current menu selection, while the right half  
shows the current setting of the item associated with that menu.  
Control Buttons  
Toggles between the various items in a menu.  
Menu Button  
Toggles between the various sets of menus.  
Up Button  
Increments a number or increments the selection for the current item.  
Down Button  
Decrements a number or increments the selection for the current item.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Left Button (help)  
For numeric items, this button is used to select which digit in a numeric entry is selected for  
editing. For other items or when the left most digit is already selected for numeric items,  
pressing and holding this button down displays a brief description of the current item.  
Right Button (range)  
For numeric items, this button is used to select which digit in a numeric entry is selected for  
editing. For other items or when the right most digit is already selected for numeric items,  
pressing and holding this button down displays the range and units for the current item.  
Stop Button  
Disables output of pulses on all channels.  
Run Button  
Enables output of pulses.  
Back Panel  
QC Connectors  
There are three (3) QC connectors on the rear of the 9412/9412A, 2 outputs and 1 input.  
T1  
Channel 1 output -- a general-purpose pulse output, with programmable delay, pulse width, and  
pulse polarity. The maximum output voltage is adjustable and varies from 20 Volts open-  
circuit to 10 Volts when driving a 50 Ohm load. The maximum voltage is adjusted by a poten-  
tiometer located on the back of the 9412/9412A.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
T2  
Channel 2 output -- designed specifically for use as a solid-state laser Q-switch trigger and  
has all the capabilities of the other output channels, including several special modes which af-  
fect only the output of Channel 2.  
Ext/Gate  
External Trigger or Gate signal input -- used as either an external signal to trigger the genera-  
tion of pulses or as a gate signal to enable and disable the outputs.  
Output Voltage Adjustment  
The maximum output voltage of all output channels is adjusted by turning the potentiometer la-  
beled VOUT. Turn clockwise to increase the voltage, counter-clockwise to decrease the volt-  
age. The maximum voltage possible is 20 Volts into a high-impedance load or 10 Volts into a  
50 Ohm load.  
Serial Port  
This is a 9-pin D-sub connector for the RS232 interface that comes standard with the  
9412/9412A. It is pinned to be directly connected to a standard IBM-PC serial port -- use a  
straight-through cable (not a null modem cable) to connect to a PC.  
AC Input  
A standard IEC instrument-type cord should be used. The 9412/9412A is supplied with a  
power cord that fits the type of socket used in your country.  
Before plugging in, please make sure that the voltage selection switch is set  
to the proper setting for your country (either 110V or 220V).  
Line Voltage Selection  
This switch is used to select the voltage level used in your country -- make sure that it is on  
the proper setting or the 9412/9412A could be damaged.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 OPERATING INSTRUCTIONS  
This chapter presents detailed instructions on the operation of the 9412/9412A.  
Overview  
The 9412/9412A uses a single-line LCD to display menu selections. The front panel keys are  
used to "navigate" through sets of menus to set various options and control the pulse genera-  
tor. Some keys are only active at certain times or have different behavior depending on what  
state the menus are in -- once understood, the menus can be quickly and easily manipulated.  
The 9412/9412A also contains non-volatile memory for storing up to four (4) (eight (8) for the  
9412A) configurations, so you need not go through the entire setup process each time the in-  
strument is used.  
Selecting Menu Items  
The menu items in the 9412/9412A are arranged in sets. Pressing the  
key will toggle be-  
tween sets of menus, while the key will toggle between menu items within the current set.  
Hold down the Left arrow key to receive a short description of the current menu item. Press-  
ing and holding the Right arrow key will display the valid range for that item.  
Numeric Input Mode  
When the current item is numeric, the system enters the Numeric Input mode. In this mode,  
the Left and Right arrow keys are used to select a digit to edit. The selected digit blinks to  
identify itself as the active digit. The Up and Down arrow keys are then used to increment or  
decrement this digit. When the left most digit is active, pressing and holding the Left arrow  
key will display a brief description of the current item. When the right most digit is active,  
pressing and holding the Right arrow key will display the range and units for the current item.  
Item Edit Mode  
When the current item is non-numeric, the Up and Down arrow keys are used to select  
among different options for the entry. The Left and Right arrow keys are used to display a  
description of the item.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Main Menu  
The Main Menu consists of a set of menu selections used to set the operating mode, time pe-  
riod and frequency of the 9412/9412A.  
MODE - Pulse Modes  
There are five (5) different Pulse modes available.  
Continuous Mode  
While in Continuous mode, the 9412/9412A will generate pulses at the rate set by the internal  
To Period as set in the ToPer menu. The delay of Channel 1 is relative to To, while the delay  
for Channel 2 is from the active edge of Channel 1.  
The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or  
High. Pulses will then only be generated while the Ext/Gate input is active. Selecting Low  
will make the gate active low, while selecting High will make the gate active high. Set to Off  
to disable the gate function.  
Note: No pulses are generated until the Run key has been pressed.  
External Trigger Mode  
In the External Trigger mode, one pulse is output each time an active edge occurs on the  
Ext/Gate input. The external trigger polarity menu (TRIG) is used to determine whether the  
9412/9412A is triggered on the rising or falling edge of the external trigger.  
The delay of each output is relative to the active edge of the external trigger, except for  
Channel 2 which is delayed from the active edge of Channel 1.  
Since it is being used as a trigger source, the Ext/Gate input cannot be used as a gate in this  
mode.  
Single Shot Mode  
During Single Shot operation, one pulse is generated each time the Run key is pressed. The  
rate at which pulses are produced is totally dependent on how fast the Run key is pressed.  
The external trigger input may be used as a gate or interlock by setting the GATE menu to  
Low or High. Pulses will then only be generated while the external trigger input is active. Se-  
lecting Low will make the gate active low, while selecting High will make the gate active high.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Duty Cycle  
The Duty Cycle mode is similar to Continuous mode in that a continuous stream of pulses is  
output. The exception is the outputs are only active for a set number of seconds (DC:On  
menu), then disabled for another set time period (DC:Off menu) before being re-enabled. This  
cycle is repeated for a set number of times (CYCLES menu), after which all outputs cease.  
Setting the number of cycles to zero results in continuous cycling, until the Stop Key is  
pressed.  
The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or  
High. Pulses will then only be generated while the external trigger input is active. Selecting  
Low will make the gate active low, while selecting High will make the gate active high.  
The Run key must be pressed to start the Duty Cycle mode.  
Burst  
Burst mode is similar to Continuous, except only a certain number of pulses will be output  
(#/BURST menu), after which outputs are disabled.  
The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or  
High. Pulses will then only be generated while the external trigger input is active. Selecting  
Low will make the gate active low, while selecting High will make the gate active high.  
The Run key must be pressed to start Burst mode.  
GATE - Gate Select  
When not in External Trigger mode, the Ext/Gate input on the 9412/9412A can be used as a  
gate signal. When the gate is enabled, pulses will only be output while the signal into the  
Ext/Gate input is active -- thus, this signal will act as a safety interlock or external control sig-  
nal.  
The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or  
High. Pulses will then only be generated while the external trigger input is active. Selecting  
Low will make the gate active low, while selecting High will make the gate active high. Set-  
ting the GATE menu to Off will disable the Gate function.  
The setting of the GATE menu is used to determine when the signal is active. If the gate  
menu is set (active) High, pulses will only be generated if the Ext/Gate input signal is greater  
than 3.5 Volts. If set to (active) Low, outputs will be enabled as long as the Ext/Gate input  
signal is less than 2.0 Volts.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ToPer - To Period  
Except when in External Trigger or Single Shot mode, all pulse outputs are based on an inter-  
nal timer called To. This menu is used to select the rate at which To runs by selecting the  
time period in microseconds between timer pulses. When in Continuous, Burst, or Duty Cycle  
mode, pulses are output at the rate determined by To. The frequency of pulse outputs in these  
modes can be found by taking the reciprocal of the To Period setting.  
Width and Delay Menus  
This set of menus is used to set the delay and pulse width for each of the output channels.  
Note: Channel 2 delay is relative to the Channel 1 (T1) output.  
Tn:Dly - Tn Delay  
Each of the two channels has an independent delay setting which is set from the Tn:Dly,  
where n = 1 or 2.  
When in any mode but External Trigger, the Delay setting sets the delay from the time the in-  
ternal timer To goes high until a pulse is output on a particular channel.  
Note: In the case of Channel 2 (T2), the output is delayed from the active  
edge of T1 - not To as with the other channels.  
When in External Trigger mode, the delays are with respect to the active edge of the external  
trigger input signal.  
Tn:Wid - Tn Pulse Width  
Each of the two channels has an independent pulse width setting which is set from the Tn:Wid  
menu, where n = 1 or 2.  
The pulse width is used to specify the duration of the output pulse in mS. The setting of the  
output polarity (the Tn:Pol menu) for a particular channel determines whether the output is  
high for this duration or is low.  
Tn:Pol - Tn Polarity  
Each channel has an independent output polarity setting set from the Tn:Pol menu, where n =  
1 or 2.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
When set to Positive, the output will initially be at 0 Volts. When active, will go high for the  
specified pulse width time, after which it will return to 0. The voltage level while high will de-  
pend upon the load and the setting of the output voltage adjustment on the back panel.  
If set to Negative, the output will normally be at the high voltage level. When the output is  
activated, it will go to 0 Volts for the specified pulse width time, after which it will return to  
the high level again.  
T2 Menus  
Channel 2 (T2) has some special features designed to be particularly useful when used as a  
Q-switch trigger.  
T2:Enable  
This menu is used to separately enable or disable the output of Channel 2, independent of the  
other channels. If set to On, Channel 2's output will be active under the same conditions as  
the other channels. When set to Off, Channel 2's output will remain disabled under all condi-  
tions.  
T2:Gate  
When not in External Trigger mode, the Ext/Gate input can be used to gate the output of all  
output channels or just the output of Channel 2. In order to use the input to gate Channel 2,  
the GATE selection in the Main Menu set must be set to Off and the T2:Gate selection must  
be set to Low or High.  
The setting of the T2:Gate menu is used to determine when the signal is active. If the gate  
menu is set (active) High, Channel 2 pulses will only be generated if the Ext/Gate input signal  
is greater than 3.5 Volts. If set to (active) Low, outputs will be enabled as long as the  
Ext/Gate input signal is less than 2.0 Volts.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
T2:Wait  
The output of Channel 2 can be delayed for a number of pulses. This menu is used to set the  
number of pulses that must occur on the other channels before the output of Channel 2 is en-  
abled. This is particularly useful when using Channel 2 as a Q-switch trigger, as it allows the  
laser to thermally stabilize before activating the Q-switch.  
This function is active only in Continuous, Burst, and Duty Cycle modes. Channel 2 is delayed  
at the start of each burst in the Burst mode or at the start of each cycle in the Duty Cycle  
mode.  
T2*n - T2 Period Factor  
The period at which Channel 2 (T2) is pulsed can be set to be a multiple of the other channels.  
This menu item is used to select the multiplier. For example, if the To period ToPer) is set to  
.100 sec (10Hz), setting T2*n to 2 will cause the channel 2 period to be .200 sec (5Hz).  
Store/Recall/Misc. Menus  
The 9412/9412A contains non-volatile memory for storing four (4) (eight (8) for the 9412A)  
different configurations. The settings of all menu selections are stored and can be quickly re-  
called.  
STORE Configuration  
Stores the current menu settings in the specified memory location (one (1) - four  
(4) (eight (8) for 9412A).  
Note: If you overwrite a particular location, its previous contents are destroyed.  
To store a configuration use the Up and Down arrow keys to select the memory location  
number and press the Left or Right arrow key. You will be prompted to press the  
key to  
verify that you want to overwrite the existing stored configuration. If you press any other key,  
the configuration will not be saved.  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RECALL Configuration  
Recalls a previously stored configuration from the specified memory location. To recall a  
configuration use the Up and Down arrow keys to select the memory location number and  
press the Left or Right arrow key. You will be prompted to press the  
key to verify that  
you want to overwrite the active configuration. If you press any other key, the configuration  
will not be recalled.  
Note: A zero configuration is available for recall. This configuration resets the instrument to  
the factory default settings.  
Shot Counter  
The 9412/9412A includes a nine (9) digit shot counter. The maximum count displayed is  
4,294,967,295 after which the count rolls over to zero. To manually reset the counter to zero,  
press the Left or Right arrow key, at which time you will prompted to press the  
key to ver-  
ify that you want to zero the counter. If you press any other key, the shot counter will not be  
reset.  
Fmark  
The character used as a decimal point can be set to either a "." (period) or a "," (comma).  
The period is generally used in North America, while the comma is used in Europe.  
Volume  
The key click volume may be adjusted from 0 (disabled) to 9. The default value is 2.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
4
3
2
1
9
8
7
6
1
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 COMPUTER INTERFACE  
The 9412/9412A comes standard with an RS232 serial interface. All menu settings can be set  
and retrieved over the computer interface using a simple command language.  
Serial (RS232) Pinout  
The serial port is located on the back of the 9412/9412A and uses a 9-pin D-type connector  
with the following pinout (as viewed from the back of the 9412/9412A):  
1. No connection  
2. Tx - Transmit (to computer)  
3. Rx - Receive (from computer)  
4. DTR - connected to pin 6  
5. Ground  
6. DSR - connected to pin 4  
7. RTS - connected to pin 8  
8. CTS - connected to pin 7  
9. No connection  
The voltage level of the Transmit signal is +/-10V, the Receive signal should be at least +/-5V.  
Command Language  
All menu items can be accessed and selected via the computer interface using a simple com-  
mand language.  
Command Protocol  
All commands have the following syntax:  
$<name> <##>CR  
where  
P $ indicates a command follows.  
P <name> is the command name. No spaces are allowed between the $ and the <name>. The  
name is case sensitive.  
P <##> is an unsigned, integer associated with the command. A space must be inserted between  
<name> and <##>.  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
P CR is a carriage return character (ASCII value 13).  
With echo enabled, all characters sent will be echoed back to the sender.  
Unidentified commands and most out of range conditions result in the pulse generator ignoring  
the command. If the command is properly identified, the pulse generator returns an ok fol-  
lowed by a carriage return and a line feed (ASCII 13, ASCII 10), otherwise it returns a ?1 (a  
question mark, followed by the number "one", a carriage return and a line feed). A ?2 is sent  
if a valid command is received but the parameter is out of range.  
Example - send the following command to set the number of cycles:  
$CYCLES 100  
The pulse generator will return the following:  
Echo Enabled: CYCLES 100<cr><lf>  
ok<cr><lf>  
Echo Disabled: ok<cr><lf>  
Only one command will be processed per message. Commands are not processed until the  
carriage return is sent. If an error is made and identified prior to sending the carriage return,  
sending a new $ will reset the input buffer and allow a corrected command to be sent.  
To determine the current value of a parameter the syntax is:  
$<name> ?CR  
where name = a valid command name.  
This will return an integer followed by a space, carriage return, and a line feed character  
(ASCII 10).  
Example - the following command would retrieve the current Cycles setting:  
$CYCLES ?  
The pulse generator will return the following:  
Echo Enabled: CYCLES ?<cr><lf>  
100<cr><lf>  
Echo Disabled: 100<cr><lf>  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Communication Protocol  
Communication parameters are set at 9600 baud, 8 bits, parity = none, and 1 stop bit. To help  
establish RS232 communications, the pulse generator emits the characters QC approximately  
every 200 ms after power up and until the 9412/9412A receives a $.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command List  
The following table summarizes all the commands.  
Note: The name of each command is identical to how it appears on the display of the  
9412/9412A.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9412/9412A Communication Commands  
Command Name Parameters  
Description  
MODE  
0 = Continuous mode  
1 = Burst mode  
Sets the pulse generator mode.  
2 = Duty Cycle mode  
3 = Single Shot mode  
4 = External Trigger mode  
0 = off  
1 = active low  
2 = active high  
GATE  
ToPer  
Enables the external trigger input to function as a gate  
controlling all outputs. Cannot be used with external  
trigger mode.  
1000 - 99999999 ms  
Sets the To internal sync period, used only in Continu-  
500 - 99999999 ms (400B) ous, Burst and Duty Cycle modes.  
DC:On  
DC:Off  
TRIG  
1 - 10000 Sec  
Sets the on time, in seconds, for the Duty Cycle mode.  
1 - 10000 Sec  
Sets the off time, in seconds, for the Duty Cycle mode.  
Sets the active edge of the Ext/Gate signal when used as  
an external trigger.  
0 = falling edge  
1 = rising edge  
0 - 10000 cycles  
1 - 50000 pulses  
1 - 99999999 ms  
0 - 99999999 ms  
CYCLES  
#/BURST  
T1:Wid  
T1:Dly  
Sets the number of cycles for the Duty Cycle mode.  
Sets the number pulses in the Burst mode.  
Sets the pulse width, in microseconds, for Channel 1.  
Sets the delay, in microseconds, for Channel 1.  
T1:Pol  
0 = Negative (active low) Sets the polarity of the output for Channel 1.  
1 = Positive (active high)  
T2:Wid  
T2:Dly  
1 - 99999999 ms  
0 - 99999999 ms  
Sets the pulse width, in microseconds, for Channel 2.  
Sets the delay, in microseconds, for Channel 2. Note:  
The delay is relative to Channel 1.  
T2:Pol  
0 = Negative (active low) Sets the polarity of the output for Channel 2.  
1 = Positive (active high)  
T2:Enable  
T2:Gate  
0 = Off  
1 = On  
Turns the output of Channel 2 on and off.  
0 = Off  
Enables the external input to act as a gate for Channel 2's  
output.  
1 = Active low  
2 = Active high  
0 - 10000 pulses  
T2:Wait  
Set the number of pulses to wait after starting T1 before  
enabling T2. Note: This is used to allow time for the la-  
ser to stabilize when T1 is used as a flashlamp trigger  
and T2 is the Q-switch trigger.  
T2*n  
BEEP  
RUN  
1 - 10000  
1 - 1000  
Sets the period of T2 relative to T1. The period of T2 is  
the period of T1 multiplied by n.  
Beeps the buzzer inside the 9412/9412A the specified  
number of times.  
0 = Disable  
1 = Enable  
Enable / disables the pulse generator output.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9412/9412A Communication Commands  
RECALL  
STORE  
0 - 4 (9412)  
0 - 8 (9412A)  
1 - 4 (9412)  
Recall a stored configuration.  
Store a configuration.  
1 - 8 (9412A)  
0 = Reset to zero  
SHOTS  
(9412A Only)  
Resets shot counter. SHOT ? will return the number of  
shots.  
ECHO  
0 = Disable  
1 = Enable  
0 = , (comma)  
1 = . (period)  
0 - 9  
Enables/Disables the RS232 echo. The result string is  
always sent back.  
Fmark,  
Sets the character used as a decimal point on the display.  
VOLUME  
Sets the key click volume level.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 SPECIFICATIONS  
DELAYS  
P CHANNELS Two (2) independent outputs, with digitally controlled delay and pulsewidth.  
P DELAY  
0 to 99.999999 sec  
10 ms to 99.999999 sec  
1 ms  
P PULSEWIDTH  
P RESOLUTION  
P ACCURACY100 ns + .0002 x delay  
P TIMEBASE 8 MHz, 50 PPM crystal oscillator  
P RMS JITTER 50 ns + .0002 x period, 200 ns max  
P TRIG DELAY  
Ext Trig to To < 100 ms  
EXTERNAL TRIG / GATE  
P RATE Dc to 1/ (100 ms + largest delay + period)  
P THRESHOLD  
P TRIGGER SLOPE  
P GATE  
3 VDc (3 mA into optoisolator)  
Rising or falling edge  
Active low or active high  
1000 W  
P IMPEDANCE  
INTERNAL RATE GENERATOR  
P MODES  
Single shot, burst, duty cycle, continuous, external trigger, exter-  
nal gate  
P RATE (To period)  
1000 ms to 99.999999 sec (.01 Hz to 1 KHz) (400A) 500 ms to  
99.999999 sec (.01 Hz to 2 KHz) (400B)  
P ACCURACY100 ns + .0002 x period  
P RMS JITTER 50 ns + .0002 x period  
BURST MODE  
1 to 50,000 pulses  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTPUTS  
(T1,T2)  
P IMPEDANCE  
50 W  
P SLEW RATE > .2 V/ns  
P OVERSHOOT  
< 100 mV + 10% of pulse amplitude  
P AMPLITUDE  
load manually adjusted,  
1-10V into 50W load  
2-20V into high impedance  
all outputs at same level  
P PEAK CURRENT  
400 mA per channel  
P AVERAGE CURRENT  
200 mA average (total for all channels)  
COMPUTER INTERFACE  
P RS232 9600 Baud. All instrument functions and settings may be con-  
trolled over the interface bus.  
GENERAL  
P STORAGE Four (4) complete configurations (8 - 9412A),  
from the front panel or the computer  
may be stored and recalled  
interface.  
P DIMENSIONS  
7.5" x 9.0" x 4.0"  
5 lbs.  
P WEIGHT  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 MAINTENANCE  
Except for fuse replacement, the 9412/9412A should require no maintenance.  
Replacing the Fuse  
The fuse is mounted on the circuit board located inside the 9412/9412A. Remove the top  
cover to access and follow the instructions given below to replace the fuse:  
1. Unplug the 9412/9412A from power source and remove power cord from the back panel.  
2. Remove the two screws located on the sides of the 9412/9412A (they hold the front plastic  
bezel in place) and remove bezel.  
3. Remove the two screws located on the sides of the 9412/9412A (they hold the back plastic  
bezel in place) and remove the bezel.  
4. Lift the top cover to remove.  
5. Locate the fuse on the circuit board (small glass cylinder located in the right rear corner) at-  
tached to the circuit board by metal spring clips.  
6. Remove faulty fuse and replace with one of a similar rating (5 x 20 mm, .125 Amp).  
7. Replace top cover and bezels.  
8. Reattach the power cord.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 CUSTOM FEATURES  
None.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Radica Games Video Games 75038 User Guide
Radio Shack Range PRO 2018 User Guide
Radio Shack Video Gaming Accessories 26 445 User Guide
RayTek Camcorder ThermoView Ti30 User Guide
Remington Blower RM180B User Guide
Riccar Sewing Machine R552 User Guide
Rolls Portable Stereo System RA2100B User Guide
Samsung Printer ML 1520 User Guide
Sanyo Rice Cooker ECJ 5205SN User Guide
Seagate Switch KSD 541 User Guide