15 Discovery Way, Acton, MA 01720
Phone: (978)263-3584, Fax: (978)263-5086
Web Site: www.piacton.com
Instructions for Mounting and Operation
of the Princeton Instruments Model XS-432
75 Watt Xenon Light Source System
Ver 2.0
1
Download from Www.Somanuals.com. All Manuals Search And Download.
ultraviolet light sources are in operation. Even when a monochromator output
is set to “non-ultraviolet” wavelengths, the output may contain “second order”
or stray ultraviolet radiation.
4. Ozone Hazard: All xenon lamps, other than ozone free lamps, allow ozone
gas to be generated while the lamp is in operation. Always provide for proper
ventilation to remove this ozone. High concentrations of ozone are dangerous
and can constitute a serious health hazard!
5. Other: Read manufacturers “Instructions for use of Xenon XBO”, attached,
before attempting to install or operate this lamp.
Mounting:
For proper operation, the lamp housing must always be mounted to the entrance slit
of the spectrometer and the lamp housing cover must always be in place.
To mount the lamp housing to the entrance slit, use the following procedure:
A1. Locate the two (2) shoulder screws and the two (2) 8-32 cap screws supplied
with the XS-432. Insert the shoulder screws into the two bottom outside
tapped holes of the entrance slit of the spectrometer and tighten. These are
used to mount the lamp housing.
A2. Remove the four (4) slotted screws from the top of the lamp housing, then
remove the cover.
Caution: Do not touch the surface of the focusing mirror.
A3. Carefully slide the two (2) “key hole” type slots of the lamp housing
mounting flange over the should screws in the entrance slit of the
spectrometer. Insure that the top two holes of the lamp housing mounting
flange align properly with the top tow tapped holes of the spectrometer
entrance slit.
A4. Insert the two (2) 8-23 cap screws in the top tow holes of the lamp housing
mounting flange and tighten.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
A5. If the lamp is to be installed at this time, proceed to stop #B2 below. If not,
replace the lap housing cover.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation of the Xenon Lamp in the Lamp Housing:
(Note: The lamp housing must be mounted to the entrance slit as described above)
-----------------------------------------------------------------------------------------------------
CAUTION
Never touch the xenon lamp (bulb) or the surface of the focusing mirror.
If the xenon lamp should become contaminated, refer to the manufacturer’s
handling precautions for proper cleaning procedure.
-----------------------------------------------------------------------------------------------------
To initially install the xenon lamp in the lamp housing, use the following procedure.
B1. Remove the four (4) screws from the lamp housing cover and remove the
cover.
B2. Locate the 1/16 and 7/64 Allen wrenches supplied.
B3. A shipping post is installed in place of the lamp for shipping purposes.
Inside the lamp housing, locate and become familiar with the shipping post,
the top terminal, the mirror locations and the base terminal with a split
clamp arrangement securing the shipping post.
Caution: Do not touch, talk or breathe on the mirrors.
B4. Note the orientation of the top terminal, as the lamp and top terminal must be
installed in the same orientation. Using the 1/16” Allen wrench, loosen the
set screw in the end of the top. Note: Do not remove the wire from the top
terminal. Remove the top terminal from the shipping post and set the
terminal inside the lamp housing, adjacent to the mirror.
B5. Insert the end of the 7/64” Allen wrench into the screw in the split clamp of
the base terminal and loosen the screw. Remove the shipping post and save
for future use.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
B6. Locate the section at the end of this manual tabbed “Lamp OEM
Instructions”. Read, become familiar with and follow the “Lamp OEM
Instructions”.
B7. Locate the lamp shipping box. Put on a protective mask, goggles (glasses),
and gloves and carefully remove the lamp from its shipping box. Do not
remove the protective cover from the lamp at this time.
B8. Become familiar with the lamp, noting in particular the “+” and “—”
terminals on each end of the lamp and the short black line on the “+” end.
Also, look at the top terminal in the lamp housing and note the small pin
protruding into the through hole in the top terminal. Insert the “+” terminal
on the lamp into the top terminal in the lamp housing. Orient the lamp so
that the clear part of the lamp envelope will face the source mounting flange
when installed with the top terminal in the original orientation. While gently
holding the lamp in the tope terminal, securely hold the top terminal and
tighten the terminal clamping screw in the top terminal using the Allen
wrench provided. Caution: Do not over tighten the clamping screw.
Caution: Do not apply any pressure to the lamp when tightening the set screw! Do
not over tighten the set screw.
B9. Very carefully remove the protective cover from the lamp. Gently insert the
lamp (with the top terminal connected) into the split clamp in the base
terminal until it rests on the small pin.
Caution: Do not apply any stress or strain to the lamp. Assure that the top terminal
is oriented approximately as noted in stop #B4.
B10. Tighten base terminal clamping screw using the Allen wrench provided
until the lamp is secure.
Caution: Do not over tighten the clamping screw.
B11. Replace the top cover.
The xenon lamp is now properly installed.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Operation:
-----------------------------------------------------------------------------------------------------
CAUTION
THE TOP COVER MUST BE IN PLACE BEFORE THE LAMP IS TURNED ON!
-------------------------------------------------------------------------------------------------------------------
Connections for the XS-432
1. Connect the DC power cable between the XS-432 lamp housing and the
Optiquip power supply.
2. Connect the line cord between the lamp housing fan and the outlet on the rear
of the Optiquip power supply.
3. With the power switch on the Optiquip power supply switched off, plug its
line cord into a 110 volt outlet.
To Turn Lamp On:
1. Turn on the switch on the Optiquip power supply to ignite the Xenon lamp.
This may take several seconds depending on the age and temperature of the
Xenon bulb.
2. The Optiquip power supply is preadjusted for a power of 75 watts to operate
the Xenon lamp by Princeton Instruments. If the power is not 75 watts during
operation, refer to the Optiquip power supply instruction manual.
-----------------------------------------------------------------------------------------------------
WARNING
IF LOUD SNAPPING SOUNDS ARE HEARD FROM THE LAMP HOUSING OR
POWER SUPPLY DIRNG IGNITION, TURN OFF THE POWER SUPPLY
IMMEDIATELY AND CHECK ALL CABLE CONNECTIONS.
FAINT CLICKING SOUNDS DURING IGNITION ARE NORMAL.
-------------------------------------------------------------------------------------------------------------------
To Turn Lamp Off:
1. Turn off power switch on the Optiquip power supply.
Refer to the attached manufacturer’s instruction manual for more details on
operation of the power supply and lamp.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
-----------------------------------------------------------------------------------------------------
CAUTION
High Voltage Light Sources such as the XS-432 May Cause Interference with
Computers and Microprocessors, Especially During start Up of the Lamp.
The Light Source Should be Turned On Before Starting and Computers or
Microprocessor Controlled Monochromators.
-----------------------------------------------------------------------------------------------------
8