INSTRUCTION MANUAL
10 series Plus
Blood Pressure Monitor
with ComFitTM Cuff
Model BP791IT
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE OF CONTENTS
Display Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Downloading Instructions
®
TM
Downloading and Installing Microsoft HealthVault
. . . . . . . . .45
Downloading and Installing Omron Health
Care and Maintenance
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Before Using the Monitor
INTRODUCTION
®
Thank you for purchasing the Omron BP791IT 10 series Plus
TM
Automatic Blood Pressure Monitor with ComFit Cuff.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
• Staple your purchase receipt here
• Register your product on-line at
Your new blood pressure monitor uses the oscillometric method of
blood pressure measurement. This means the monitor detects your
blood’s movement through your brachial artery and converts the
movements into a digital reading. The monitor stores measurement
results for two people and features Morning and Evening Averages.
TM
TruRead allows for three consecutive measurements with timed
intervals for an average reading.
The BP791IT comes with the following components:
TM
• Monitor
• ComFit Cuff
• Storage Case
• AC Adapter
• USB Cable
• 4 “AA” Alkaline Batteries
• Instruction Manual printed in English and Spanish
• Quick Start Guide
The BP791IT 10 series Plus Automatic Blood Pressure Monitor is
intended for home use.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure,
CONSULT YOUR DOCTOR.
SAVE THESE INSTRUCTIONS
4
Download from Www.Somanuals.com. All Manuals Search And Download.
SAFETY INFORMATION
To assure the correct use of the product basic safety measures should
always be followed including the warnings and cautions listed in this
instruction manual.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
WARNING
CAUTION
Indicates a potentially hazardous situation
which, if not avoided, may result in minor or
moderate injury to the user or patient or
damage to the equipment or other property.
OPERATING THE DEVICE
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
DO NOT adjust medication based on measurement results from this
blood pressure monitor. Take medication as prescribed by your
physician. Only a physician is qualified to diagnose and treat High
Blood Pressure.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
SAFETY INFORMATION
OPERATING THE DEVICE (continued)
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
manual if your systolic pressure is known to be more than 220
mmHg. Inflating to a higher pressure than necessary may result in
bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any
other purpose.
Dispose of the device, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
Do not use a cellular phone near the device. It may result in an
operational failure.
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage the
unit.
Use only 1.5V alkaline batteries with this device.
Do not use other types of batteries. This may damage the unit.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
SAFETY INFORMATION
RISK OF ELECTRICAL SHOCK
Do not plug or unplug the power cord into the electrical outlet with
wet hands.
Do not overload power outlets. Plug the device into the appropriate
voltage outlet.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the
unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by Omron Healthcare will
void the user warranty. Do not disassemble or attempt to repair the
unit or components.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for
30 minutes prior to taking a measurement. Rest for at least 15
minutes prior to taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during
stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
5. Sit on a chair with your feet flat on the floor. Rest your arm on
a table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement may not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time. Try to measure your blood
pressure at about the same time each day for consistency.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating Instructions
KNOW YOUR UNIT
Main Unit:
Display
Calibration
Check System
SET Button
Morning/Evening Average
) Button
USER ID
Selection
Switch
(
Memory ( ) Button
Start/Stop Button
Up/Down
(
) Buttons
/
Battery Compartment
USB Port
Air Jack
AC Adapter Jack
9
Download from Www.Somanuals.com. All Manuals Search And Download.
KNOW YOUR UNIT
Components:
Housing
Air Tube
Cord
AC Adapter
Air Plug
(Connects to Air Jack)
Arm Cuff
4 “AA” Alkaline
Batteries
Storage Case
USB Cable
INSTRUCTION MANUAL
10 series
Blood Pressure Monitor
with ComFitTM Cuff
Model BP791IT
Instruction Manual
Quick Start Guide
10
Download from Www.Somanuals.com. All Manuals Search And Download.
UNIT DISPLAY
Morning Average Symbol
Evening Average Symbol
Date/Time Display
Irregular Heartbeat Symbol
USER ID Symbol (A or B)
Week Display
TruReadTM Symbol
Average Value Symbol
Memory Symbol
Systolic Blood Pressure
Diastolic Blood Pressure
Morning Hypertension Symbol
Movement Symbol
Blood Pressure Level Indicator
Deflation Symbol
Cuff Wrap Guide
Heartbeat Symbol
Low Battery Indicator
Pulse Display / Memory Number
11
Download from Www.Somanuals.com. All Manuals Search And Download.
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL (
)
When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol ( ) will appear on the
display with the measurement values.
Normal Heartbeat
An irregular heartbeat rhythm is defined as
a rhythm that varies by less than 25% from
Pulse
the average rhythm or 25% more than the
Blood
pressure
average rhythm detected while the monitor
is measuring the systolic and diastolic blood
Irregular Heartbeat
pressure.
Short Long
Pulse
If the irregular heartbeat symbol (
)
displays with your measurement results, we
recommend you consult your physician.
Follow the directions of your doctor.
Blood
pressure
MOVEMENT ERROR SYMBOL (
)
The Movement Error Symbol is displayed if you move your body during
the measurement. Please remove the arm cuff, and wait 2-3 minutes.
Take another measurement, remain still during inflation.
USER ID SYMBOL (
/
)
The USER ID Symbol indicates USER A or USER B as selected by
the user to take a measurement or when using the memory function.
The USER ID Symbol is not displayed when using the Guest Mode to
take a measurement.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
DISPLAY SYMBOLS
Internationally recognized guidelines for upper limit of normal home
blood pressure readings are as follows:
Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure
135 mmHg
Diastolic Blood Pressure
85 mmHg
This is a general guideline because blood pressure depends on
age and morbidity. A lower home blood pressure goal may be
recommended for certain patients, including diabetic patients,
pregnant women, and patients with renal failure.
The monitor is not intended to be a diagnostic device.
MORNING HYPERTENSION SYMBOL (
)
The Morning Hypertension Symbol is displayed if the morning average
reading for a week is above 135 for the Systolic Blood Pressure value
and/or 85 for the Diastolic Blood Pressure value when viewing the
Morning Averages and the Evening Averages in the memory.
Omron Healthcare recommends you contact your physician if the
MORNING HYPERTENSION SYMBOL appears on the display with
your measurement values.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
DISPLAY SYMBOLS
HEARTBEAT SYMBOL (
)
The Heartbeat symbol flashes on the display at every heartbeat during
the measurement.
AVERAGE VALUE SYMBOL (
)
The Average Value Symbol is displayed when you press the memory
button. The most recent average reading appears on the display screen.
MORNING AVERAGE SYMBOL (
)
The Morning Average symbol is displayed when viewing Morning
Averages using the Memory Function.
EVENING AVERAGE SYMBOL (
)
The Evening Average symbol is displayed when viewing Evening
Averages using the Memory Function.
CUFF WRAP GUIDE SYMBOL (
/
)
When the cuff is wrapped around the arm correctly, ( ) is displayed
while taking a measurement or using the memory function. If the cuff is
not wrapped around the arm correctly, (
) is displayed.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE AC ADAPTER
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
- DANGER -
TO REDUCE THE RISK OF
FIRE OR ELECTRIC SHOCK,
CAREFULLY FOLLOW THESE INSTRUCTIONS.
For connection to a power supply not in the U.S.A., use an
attachment plug adapter of the proper configuration for the power
outlet.
The power unit (AC Adapter) is intended to be correctly oriented
in a vertical or floor mount position.
NOTES: • Use only the authorized Omron AC Adapter that came with
this monitor.
• We recommend you install the batteries even if the AC
Adapter is used. If no batteries are installed, you may need to
reset the date and time if the AC Adapter is disconnected.
The measurement results will not be deleted.
CAUTION
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage the
unit.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE AC ADAPTER
TO CONNECT THE AC ADAPTER
① Insert the AC Adapter Plug into
the AC Adapter Jack on the back
side of the monitor as illustrated.
Housing
② Plug the AC Adapter into a 120V
AC outlet (60 cycles).
Cord
TO DISCONNECT THE AC ADAPTER
① Hold the Housing to remove the AC Adapter from the electrical
outlet.
② Remove the AC Adapter plug from the monitor.
WARNING
Do not plug or unplug the power cord into the electrical outlet with
wet hands.
WARNING
Do not overload power outlets. Plug the device into the appropriate
voltage outlet.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
BATTERY INSTALLATION
Battery Cover
1. Remove the battery cover by pulling the
hook up while pushing it in the direction
of the arrow.
2. Install 4 “AA” size batteries so the +
(positive) and - (negative) polarities match
the polarities of the battery compartment
as indicated.
3. Replace the battery cover.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
BATTERY INSTALLATION
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display screen turn the
monitor off and remove all the batteries. Replace with four new batteries
at the same time. Longlife alkaline batteries are recommended.
WARNING
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
NOTES: • Turn the unit off before replacing the batteries. If the
batteries are removed while the unit is still on, the date and
time will be reset to that of the previous use. However, the
measurement results are not deleted.
• When the batteries are replaced, you may need to reset the
date and time. If "-:--" appears on the display, refer to
“Setting the Date and Time”.
• The supplied batteries may have a shorter life.
CAUTION
Dispose of the device, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a
measurement for the first time. If the date and time are not set correctly,
the measurement values stored in the memory and the average values
may not be correct.
If the date and time are not set, “-:--” appears during or after
measurement.
The Blood Pressure Monitor automatically stores up to 100
measurement values per user with the dates and times. The monitor also
calculates weekly averages for measurements in the morning and in the
evening. Morning and Evening averages are stored for this week and the
last seven weeks.
The monitor can also calculate an average reading based on the three
most recent measurements taken within 10 minutes of the last reading.
NOTE: • If only two readings are stored in the memory for the 10-
minute period, the average is based on the two readings.
• If one reading is stored in the memory for the 10-minute
period, this reading is displayed as the average.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING THE DATE AND TIME
1. TO START THE SETTING
Press and hold the SET button to
adjust the date and time.
The year flashes on the display.
2. SETTING THE YEAR
The year can be set between 2010 and
2040. When the display reaches 2040,
it will return to 2010.
Press the Up button (
by one year.
) to advance
Press the SET button to set the current
year.
The month flashes on the display.
NOTES: • Press the Down button (
) to decrease the date and time
values.
• Press and hold the Up or Down button to increase or
decrease the date and time values faster.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING THE DATE AND TIME
3. SETTING THE MONTH
Press the Up button (
by one month.
) to advance
Press the SET button to set the current
month.
The day flashes on the display.
4. SETTING THE DAY
Press the Up button (
by one day.
) to advance
Press the SET button to set the current
day.
The hour flashes on the display.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING THE DATE AND TIME
5. SETTING THE HOUR
The time is set using AM or PM.
Press the Up button (
by one hour.
) to advance
Press the SET button to set the current
hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button (
) to advance by one minute.
Press the SET button to set the minute setting.
7. TO TURN THE MONITOR OFF
Press the START/STOP button.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING TRUREADTM
TM
The TruRead Mode takes three consecutive measurements.
The monitor will inflate, take a measurement, and deflate three times
separated by a short interval of time between each measurement.
1. Select your USER ID (A or B).
2. Press the SET button.
TM
The TruRead symbol
on the display.
and the oFF setting appears
3. Press the Up or Down button (
/
) to select on.
NOTE: Press the Up or Down button
to alternate between on and
oFF.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING TRUREADTM
4. Press the SET button to confirm the change.
If on is selected, the number of seconds between each measurement is
displayed.
NOTE: The default value between each measurement is 60 seconds.
5. Press the Up or Down button (
time between each measurement.
/
) to change the interval of
The interval can be set to 15 seconds, 30 seconds, 60 seconds or 120
seconds. Continue to press the Up or Down button until the desired
interval appears on the display.
6. Press the SET button to confirm the setting.
The year flashes on the display.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
SETTING TRUREADTM
7. If the date and time are correct, press the START/STOP button to turn
the monitor off.
To change the date and time, refer to “Setting the Date and Time”.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
APPLYING THE ARM CUFF
CAUTION
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper left arm.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
APPLYING THE ARM CUFF
3. Sit on a chair with your feet flat on the floor. Place your left arm
on a table so the cuff is level with your heart.
Thumb Grip
4. Hold the thumb grip on the cuff
securely with your right hand.
Cloth Fastener
5. Turn the palm of your left hand upward.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
APPLYING THE ARM CUFF
6. Apply the cuff to your left upper arm so the blue stripe is on the
inside of your arm and aligned with your middle finger. The air tube
runs down the inside of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Blue Stripe
1/2 inch (1 - 2 cm)
Air Tube
7. Wrap the cuff firmly in place around your arm using the cloth
fastener.
28
Download from Www.Somanuals.com. All Manuals Search And Download.
APPLYING THE ARM CUFF
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use this instruction
Apply the cuff to your right upper arm so the thumb grip is centered
on the inside of your inner arm. The bottom of the cuff should be
approximately 1/2'' above your elbow.
Thumb Grip
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff.
29
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
The monitor is designed to take measurements and store the
measurement values in the memory for two people using
USER ID A and USER ID B.
The monitor can also be used to take a single measurement for other
persons using the GUEST Mode. The USER ID symbol is not displayed
when using the GUEST Mode. Measurements taken using the GUEST
Mode are not stored in the memory.
WARNING
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
CAUTION
Operate the device only as intended. Do not use the device for any
other purpose.
CALIBRATION CHECK SYSTEM
When you press the START/STOP button, the Calibration Check
System light turns on and the unit starts monitoring your readings
using dual sensors. If the unit is accurate and functioning correctly,
the Calibration Check System light remains lit during the
measurement. If an error is detected, the Calibration Check System
light flashes and “ER” will appear on the display.
CUFF WRAP GUIDE
This monitor checks whether the arm cuff is wrapped around the arm
correctly during the inflation. When the cuff is wrapped correctly,
is
displayed while taking a measurement or using the memory function. If
the cuff is not wrapped correctly,
is displayed. If
is displayed,
refer to the section “Applying the Arm Cuff”, wrap the cuff firmly
around the arm and take another measurement.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
USING THE GUEST MODE
1. Press and hold the START/STOP button.
The A and B Symbols disappear from the display.
A and B Symbols disappear
2. Release the START/STOP button.
The cuff starts to inflate automatically.
NOTE: To stop the inflation press the START/STOP button.
The monitor will stop inflating, start deflation and turn off.
3. Inflation stops and the measurement is started. When the
measurement is completed, your blood pressure and pulse rate appear
on the display.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
SELECTING THE USER ID
Select your USER ID. Always use the same USER ID when taking a
measurement. The monitor stores the measurement values in the
memory by the USER ID selected. These measurement values are used
to calculate the Morning Averages and Evening Averages.
Slide the USER ID Selection Switch right or left to select USER A or B.
USING THE USER ID
After you select a USER ID, the monitor can be set to take one
measurement using the Single Mode or three consecutive measurements
TM
using the TruRead Mode.
NOTE: The default value is set to the Single Mode to take one
TM
measurement. Refer to “Setting TruRead ”.
32
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
USING THE SINGLE MODE
1. Select your USER ID (A or B).
2. Press the START/STOP button.
All symbols appear on the display. The USER ID Symbol you
selected flashes on the display.
The cuff starts to inflate automatically. As the cuff inflates, the monitor
automatically determines your ideal inflation level.
This monitor detects the pulse during inflation. Do not move your arm
and remain still until the entire measurement process is completed.
NOTE: To stop the inflation or measurement, press the START/STOP
button. The monitor will stop inflating, start deflating and will
turn off.
33
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display.
The Heartbeat Symbol ( ) flashes at every heartbeat.
4. When the measurement is complete, the arm cuff
completely deflates. Your blood pressure and pulse
rate are displayed.
5. Press the START/STOP button to turn the
monitor off.
NOTE: The monitor will automatically turn
off after two minutes.
34
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
TM
USING THE TRUREAD MODE
1. Select your USER ID (A or B).
2. Press the START/STOP button.
All symbols appear on the display. The USER ID Symbol you
selected flashes on the display. The cuff starts to inflate automatically.
NOTE: To stop the inflation or measurement, press the START/STOP
button. The monitor will stop inflating, start deflating and will
turn off.
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display.
The Heartbeat Symbol ( ) flashes at every heartbeat. When the
measurement is complete, the arm cuff completely deflates.
NOTE: Do not move your arm and remain still until the entire
measurement process is completed.
4. Check the progress of the consecutive measurements.
NOTE: If a measurement fails, the monitor will repeat taking
a measurement up to five times during the consecutive
measurement process.
The first measurement is complete, waiting to start the next
measurement. The interval of time set between each measurement will
appear on the display.
35
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
The second measurement is complete, waiting to start the
next measurement.
The consecutive measurements are complete. The average for three
measurements appears on the display.
5. Press the SET button to display the blood pressure and pulse rate for
the first measurement. Press the SET button again to display the next
measurement.
6. Press the START/STOP button to turn the monitor off.
WARNING
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this
blood pressure monitor. Take medication as prescribed by your
physician. Only a physician is qualified to diagnose and treat High
Blood Pressure.
36
Download from Www.Somanuals.com. All Manuals Search And Download.
TAKING A MEASUREMENT
WARNING
The monitor is not intended to be a diagnostic device.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising
where the cuff is applied.
If your systolic pressure is known to be more than 220 mmHg, press
and hold the START/STOP button until the monitor inflates 30 to 40
mmHg higher than your expected systolic pressure.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
37
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
The monitor is designed to store the blood pressure and the pulse
rate in the memory for two people (USER A and USER B) every time
a measurement is completed.
The monitor automatically stores up to 100 sets of measurement values
(blood pressure and pulse rate) for each user (A and B). When 100 sets
of measurement values are stored, the oldest record is deleted to save
the most recent values. The monitor also stores 8 weeks of Morning
Averages and 8 weeks of Evening Averages for each user (A and B).
AVERAGING FUNCTION
The monitor calculates the average reading based on the three most
recent sets of measurement values taken within 10 minutes of the most
recent reading.
NOTES: • If two sets of measurement values are stored in the memory
for the 10 minutes period, the average is based on the two
sets of measurement values. If one set of measurement
values is stored, this is displayed as the average.
• When viewing the reading taken without setting the date and
time, “-:--” is displayed instead of the date and time.
38
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
3. Press the Down button (
) to display the most recent set of
measurement values on the screen. The values are displayed from the
most recent to the oldest.
NOTE: The date and time are alternately
displayed with the measurement
values.
Alternating display
Press the Up or Down button to display the
next set of values.
: To the older readings
: To the more recent readings
NOTE: Press and hold the Up or Down button to display the values
faster.
TM
TruRead Mode
The average for the consecutive measurements appears on the display
TM
with the TruRead symbol (
). To display the values for the
individual measurements press the SET button while the average is
displayed.
4. Press the START/STOP button to turn the power off.
40
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
MORNING AND EVENING AVERAGES
The monitor calculates and displays a weekly average for measurements
taken in the morning ( ) and evening ( ). The monitor stores 8
weeks of Morning Averages and 8 weeks of Evening Averages for each
user (A and B).
NOTE: The week begins Sunday at 4:00 AM.
MORNING AVERAGES
Morning averages are based on the first average reading using the
TM
TruRead Mode or the first reading using the Single Mode.
Measurement times must be between:
4:00 AM and 9:59 AM.
EVENING AVERAGES
Evening averages are based on the last average reading using the
TM
TruRead Mode or the last reading using the Single Mode.
Measurement times must be between:
7:00 PM and 1:59 AM.
41
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
TO DISPLAY MORNING AND EVENING AVERAGES
1. Select your USER ID (A or B).
2. Press the Morning/ Evening
Average button (
) to select
the Morning or Evening average.
The average for the current week
“THIS WEEK” appears on
the display.
3. Press the Down (
) button to display the previous weeks.
With each press of the Down (
from 1 week ago to the older.
With each press of the Up (
from 7 weeks ago to the recent.
) button, the values are displayed
) button, the values are displayed
NOTE: If no measurement is taken in a given week, the
display will appear as illustrated.
4. Press the START/STOP button to turn the monitor off.
42
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
DISPLAY COMBINATIONS
In addition to the Morning Average and Evening Average Symbols, the
monitor may also display the Morning Hypertension Symbol if the
morning average for that week is above the internationally recognized
results, these may be displayed in the following combinations.
Display
Morning
Average
Morning average from 7 weeks ago
Morning average from This Week
with Morning Average Symbol +
with Morning Average Symbol
Morning Hypertension Symbol
Evening
Average
Evening average from 7 weeks ago
with Evening Average Symbol +
Morning Hypertension Symbol
Evening average from This Week
with Evening Average Symbol
Average above internationally
recognized guidelines + Morning
Hypertension
Average within internationally
recognized guidelines
43
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE MEMORY FUNCTION
The Morning Hypertension Symbol ( ) appears if the weekly average
for morning measurements is above 135/85.
In this case the Morning Hypertension Symbol ( ) is displayed when
the Evening Average is displayed, regardless of the values for the
Evening Average.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
1. Select your USER ID (A or B).
2. While holding the Memory button ( ) and the START/STOP button
simultaneously for more than 2 seconds all values will be deleted.
44
Download from Www.Somanuals.com. All Manuals Search And Download.
Downloading Instructions
INSTALLING THE SOFTWARE
Optional Instructions for PC Connection
®
TM
To install software, choose either Microsoft HealthVault or Omron
Health Management software.
NOTE: Any measurement values taken before setting the date and time
TM
are not transferred to Microsoft HealthVault and Omron Health
Management software.
®
DOWNLOADING AND INSTALLING MICROSOFT
TM
HEALTHVAULT
HealthVault is a free Web-based platform designed to put people in
control of their health data. It helps them collect, store and share health
information with family members and participating healthcare providers,
and provides people with a choice of third-party applications and
devices to help them manage things such as fitness, diet and health.
2. Create a HealthVault account to store your blood
pressure data online.
3. Track your progress and manage your health goals
using HealthVault.
45
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLING THE SOFTWARE
OPERATING ENVIRONMENT REQUIREMENTS
The following operating environments are required to run Microsoft
HealthVault.
®
Supported OS*:
HealthVault works best with the following operating systems:
Either 32-bit or 64-bit versions of:
®
®
®
Microsoft Windows 7 Home Premium
®
®
Windows 7 Professional
®
®
Windows 7 Ultimate
Either 32-bit or 64-bit versions of:
®
®
®
Microsoft Windows Vista Home Basic
®
®
Windows Vista Home Premium
®
®
Windows Vista Business
®
®
Windows Vista Enterprise
®
®
Windows Vista Ultimate
32-bit versions of:
®
Windows XP Home Edition Service Pack 2 (SP2)
®
Windows XP Professional SP2
®
Windows XP Tablet PC Edition SP2
®
Windows XP Media Center Edition 2005 (Rollup Update 2)
®
Windows XP Service Pack 3 (SP3)
®
®
HealthVault won’t run on versions of Windows earlier than Windows
®
®
®
XP SP2, any 64-bit version of Windows XP, Windows Vista Starter
®
edition, or any version of Windows Server.
Memory capacity: 512 MB (minimum)/1 GB (recommended)
Processor:
1 GHz
Hard disc space:
200 MB
Internet connection: Broadband
Display resolution: 1024 × 768
* To view details about your computer:
1. Go to the Start menu.
2. Select Control Panel.
3. Select System and Maintenance.
4. Select System.
5. From this screen you can see your operating system, RAM, processor,
system type, and other details about your computer.
46
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLING THE SOFTWARE
DOWNLOADING AND INSTALLING OMRON HEALTH
MANAGEMENT SOFTWARE
Omron Health Management Software allows you to view, process and
print data about blood pressure and walking measured by Omron’s
compatible blood pressure monitor and pedometer.
• To install Omron Health Management Software onto your PC, you
must have administrative privileges.
• Close all applications currently running before you install the software.
• When using anti-virus software, disable anti-virus software before
installation. It may interfere with the set-up program of this software.
Enable the anti-virus software again after installation is complete.
1. Download the Omron Health Management Software at
www.software.omronhealthcare.com. Follow the installation
directions when prompted. Read the License Agreements thoroughly
and accept the terms if you agree. Once Omron Health Management
Software is installed, continue to the next section, “Using the
2. After downloading the software, extract a zip file and double-click the
[setup.exe] icon.
47
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLING THE SOFTWARE
OPERATING ENVIRONMENT REQUIREMENTS
The following operating environments are required to run the Omron
Health Management software.
®
®
Supported OS:
Microsoft Windows 2000 Professional (SP4)
®
®
Microsoft Windows XP Home Edition (SP2)
®
®
Microsoft Windows XP Professional (SP2)
®
®
®
Microsoft Windows Vista Home Basic
®
®
®
Microsoft Windows Vista Home Premium
®
®
®
Microsoft Windows Vista Business
®
®
®
Microsoft Windows Vista Enterprise
®
®
®
Microsoft Windows Vista Ultimate
®
®
®
Microsoft Windows 7 Starter
®
®
®
Microsoft Windows 7 Home Basic
®
®
®
Microsoft Windows 7 Home Premium
®
®
®
Microsoft Windows 7 Professional
®
®
®
Microsoft Windows 7 Enterprise
®
®
®
Microsoft Windows 7 Ultimate
PC:
800 MHz or higher CPU, PC/AT compatible machine
Memory capacity: 512 MB or more
Hard disc space:
Display:
32 MB or more*
Super VGA (800 × 600 pixel) or higher resolution video
adapter and monitor
Others:
USB connector (USB Revision 1.1 compliant)
Adobe Acrobat Reader 5.0 or higher
NOTE: If the OS environmental requirements are above those listed,
then additional resources may be required.
*Additional 200 MB is necessary in the case that Microsoft.
NET or MDAC is not installed.
Trademark and Registered trademark
®
®
• Microsoft and Windows are registered trademarks in the United
Sates and other countries.
®
®
®
• Adobe , Adobe logo, Acrobat and Reader are registered trademarks
or trademarks of Adobe Systems Incorporated in the United States and
other countries.
48
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE SOFTWARE
For help using the software, double-click [Omron Health
Management Software Users Manual.pdf] shortcut on
desktop, or click [Help] in this software after starting up.
1. Double-click the [Omron Health Management Software]
shortcut icon on the desktop.
If no icon appears on the desktop: From the [Start]
menu, select [All programs] or [Programs] [Omron
Health Management Software], and click the [Omron
Health Management Software] shortcut.
2. If you are using Omron Health
Management Software for the
first time, click [Add User] on
the User Management screen.
49
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE SOFTWARE
3. In the New User Registration
box, enter your name
(e.g., Chuck) and click [OK].
NOTES:
• Omron Health Management
Software can manage multiple users’ data. Data is controlled by user
name.
• The number of users who can register for this software varies
depending on the amount of free disc space in your PC.
• The user name can be up to 28 characters.
4. Click the user name, and click
[Graph Screen].
5. Click the [Blood Pressure
Monitor] icon.
50
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE SOFTWARE
Use only the authorized Omron USB cable, Model HHX-CABLE-
USB2, included with this unit.
CAUTION
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage
the unit.
1. Open the USB Port cover.
Normal USB
Connector
2. Plug the normal USB Connector
(Large Connector) into the
personal computer.
Mini USB Connector
3. Plug the mini USB connector
(Small Connector) into the
USB Port.
NOTE: You cannot take a
measurement while
connecting the USB cable.
51
Download from Www.Somanuals.com. All Manuals Search And Download.
USING THE SOFTWARE
4. Click [Download Data] icon to import the data.
5. The Confirm dialog
appears. Make sure your
cable is connected then
click [Yes].
6. Select the blood pressure
monitor being used and
click [OK].
7. Start importing data. Click
[Cancel] if you want to
cancel.
8. The dialog appears when complete. Click [OK]
and view your data on the graph.
52
Download from Www.Somanuals.com. All Manuals Search And Download.
Care and Maintenance
CARE AND MAINTENANCE
To keep your blood pressure monitor in the best condition and protect
the unit from damage, follow the directions listed below:
Keep the monitor in the storage case
when not in use. Make sure the AC
adapter is placed under the main unit so
that it does not damage the display.
Avoid kinking or sharply bending the AC
Adapter cord.
Do not forcefully bend the arm cuff or
air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth. Do not use any abrasive
or volatile cleaners. Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water.
Do not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the
unit on the floor.
Remove the batteries if the unit will not be used for three months or
longer. Always replace all the batteries with new ones at the same time.
53
Download from Www.Somanuals.com. All Manuals Search And Download.
CARE AND MAINTENANCE
Use the unit consistent with the instruction provided in this manual.
CAUTION
Changes or modification not approved by Omron Healthcare will
void the user warranty. Do not disassemble or attempt to repair the
unit or components.
CAUTION
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage
the unit.
54
Download from Www.Somanuals.com. All Manuals Search And Download.
ERROR INDICATORS AND TROUBLESHOOTING TIPS
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Insert the plug securely. Refer to “Applying the
Arm Cuff”.
Air plug disconnected.
Apply the arm cuff correctly. Refer to “Applying
the Arm Cuff”.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Replace the cuff with the new one.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
Cuff over or under inflated.
If “E2” appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to “Taking a Measurement”.
The arm cuff was inflated above 299
Do not inflate the cuff above 299 mmHg. Refer to
mmHg when inflating the cuff manually. “Taking a Measurement”.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
Movement during measurement.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering with the arm cuff.
Refer to “Applying the Arm Cuff”.
Device error.
Contact Customer Service.
Sensor may be out of alignment.
Contact Customer Service for repair information.
Cuff is not applied correctly.
Refer to “Applying the Arm Cuff”.
55
Download from Www.Somanuals.com. All Manuals Search And Download.
ERROR INDICATORS AND TROUBLESHOOTING TIPS
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Replace the four batteries.
Refer to “Battery Installation”.
Batteries are worn.
TROUBLESHOOTING TIPS
PROBLEM
CAUSES AND SOLUTIONS
No power.
Replace all four batteries with new
ones.
No display appears on the unit. Check the battery installation
for proper placement of the
battery polarities.
Blood pressure varies constantly.
Many factors including stress, time
of day, and how you wrap the cuff,
Measurement values appear
too high or too low.
may affect your blood pressure.
Review the sections “Before Taking
a Measurement” and “Taking a
Measurement”.
56
Download from Www.Somanuals.com. All Manuals Search And Download.
FCC STATEMENT
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If the product does cause harmful interference to radio or television
reception, which can be determined by turning the product on and off, the
user is encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing equipment
standard entitled “Digital Apparatus”, ICES-003 of the Canadian Department of
Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables
aux appareils numériques de Clase B prescrites dans la norme sur le materiel
brouilleur: “Appareils Numériques”, ICES-003 édictée par le minister des
communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
57
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITED WARRANTY
®
Your Omron BP791IT 10 series Plus Automatic Blood Pressure Monitor,
excluding the arm cuff and batteries, is warranted to be free from defects in materials
and workmanship appearing within 5 years from the date of purchase, when used
in accordance with the instructions provided with the monitor. The arm cuff is
warranted to be free from defects in materials and workmanship appearing within
one year from the date of purchase when the monitor is used in accordance with
the instructions provided with the monitor. The above warranties extend only to
the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility
and your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for
the address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to
prevent damage in transit. Because of possible loss in transit, we recommend
insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON
IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY
DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES
AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE
LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS
WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other
rights that vary by jurisdiction. Because of special local requirements, some of
the above limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:
Call toll free:
1-800-634-4350
58
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIFICATIONS
Model:
Display:
BP791IT (HEM-7222-ITZ)
LCD Digital Display
Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Accuracy/Calibration: Pressure: 3mmHg or 2% of reading
Pulse:
5% of reading
Inflation:
Deflation:
Rapid Air Release:
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Measurement Method: Oscillometric method
Power Source:
Battery Life:
120V 60Hz AC adapter or 4 “AA” batteries
Approx. 300 uses when used 6 times a day
with 4 new alkaline batteries
Operating Temperature 50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
/Humidity:
Storage Temperature
/Humidity/Air Pressure: /700 - 1060 hPa
Main Unit Weight: Approximately 15 1/2 oz. (440 g)
not including batteries
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
Main Unit Dimensions: Approx. 6 1/4" (l) × 4 7/8" (w) × 3 5/16" (h)
(158 mm × 124 mm × 86 mm)
Cuff Size:
Approximately 6" (w) (152 mm)
Cuff tube 23 5/8" (l) (600 mm)
Fits arm circumferences 9" to 17"
(220 mm to 420 mm)
Cuff Circumference:
Memory:
Contents:
Up to 100 per user
TM
Main Unit, ComFit Cuff, AC Adapter,
4 “AA” Alkaline Batteries, USB Cable,
Instruction Manual, Quick Start Guide, Storage Case
= Type B
NOTE: These specifications are subject to change without notice.
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Manufactured by OMRON HEALTHCARE Co., Ltd.
Made in China
Distributed by:
Omron Healthcare, Inc.
1200 Lakeside Drive
Bannockburn, IL 60015
Copyright © 2011 Omron Healthcare, Inc.