OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
Manufacturer
Check following components!
Vérifier les composants suivants !
Controleer de volgende onderdelen!
Prüfen Sie folgende Teile des Lieferumfangs! Проверьте следующие компоненты!
Controllare i componenti indicati di seguito!
¡Compruebe los siguientes componentes!
EU-representative
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
Production facility
Site de production
Planta de producción
Productiefaciliteit
OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.
Binh Duong Province, VIETNAM
Herstellungswerk
Stabilimento di produzione
Производственное подразделение
English
Français
Deutsch
Italiano
Subsidiary
Empresa filial
Dochteronderneming
Филиал
OMRON HEALTHCARE UK LTD.
Opal Drive
Succursale
Niederlassung
Consociata
Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a, 68163 Mannheim, GERMANY
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Uniquement pour le marché français:
OMRON Service Après Vente
N° Vert 0 800 91 43 14
Español
Nederlands
Русский
Digital Automatic Blood Pressure Monitor
Model M3W
Eng
lish
Français
Dig
Deu
tsc
ital
A
h
utom
atic
Mo
ct
B
del
io
Italia
loo
M3W
d
P
ress
no
ure
Moni
tor
Esp
Ned
añ
ol
erland
Ins
tru
n
Ma
nu
al
s
Ру
сский
IM-HEM
-7202
-E(V)
5334
-02-0
060-6
1/2012
B
Instruction Manual
IM-HEM-7202-E(V)-02-01/2012
5334060-6B
Download from Www.Somanuals.com. All Manuals Search And Download.
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff
inflation can cause internal bleeding.
(AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet hands.
(Battery Usage)
EN
• If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user
or patient or damage to the equipment or other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when taking measurements. This could
result in injury when the air pressure in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in
incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it can cause internal bleeding.
1
Download from Www.Somanuals.com. All Manuals Search And Download.
Important Safety Information
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of unsupported adapters may damage and/or
may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord
immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data
Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device
and any used accessories or optional parts.
2
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Overview
Open the rear cover page to read following:
The alphabet in the rear cover page correspond to those in the body page.
Main unit
Display
A
B
C
D
E
M
Display
Systolic blood pressure
N
O
O/I START ( ) button
User A/B button
Up/Down (
Diastolic blood pressure
Heartbeat symbol
(Flashes during measurement)
EN
) buttons
P
Q
R
S
T
Average value symbol
Deflation symbol
Buzzer and Date/Time setting
( ) button
Memory symbol
F
G
H
Air Jack
Irregular heartbeat symbol
Movement error symbol
Blood pressure level indicator
Pulse display and Memory
number
Battery compartment
AC adapter jack (for optional AC
adapter)
U
V
Arm cuff
I
Arm cuff (Wide Range Cuff: arm
W
X
Y
Z
Buzzer symbol
circumference 22-42cm)
Cuff wrapping guide
USER ID symbol (A/B)
Battery low symbol
Date/Time display
J
K
L
Marker
Air tube
Air plug
3
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
2.1 Installing/Replacingthe
Batteries
1. Remove the battery cover.
2. Insert four “AA” batteries as
indicated in the battery
compartment and then replace the battery cover.
Notes:
• If the battery low symbol (
time.
) appears on the display, turn the unit off then replace all batteries at the same
• The measurement values continue to be stored in memory even after the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the national
regulations for the disposal of batteries.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
2.2 Setting the Buzzer/ Date and Time
1. Press the button.
2. Set the Buzzer.
Push
select On or Off.
or
buttons to
1)
2)
EN
Push the
button to confirm.
Off
On
3. Set the monitor to the correct date and time before
taking a measurement for the first time.
Push
change.
or
buttons to
1)
Push the
Next setting appears.
button to confirm.
2)
Year
Month
Day
Hour
Minute
4. Press the O/I START button to store the setting.
Notes:
• If the batteries have been removed for 30 seconds or more, the date and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after measurement.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Open both the front and rear covers to read following:
The alphabet and number in the cover page correspond to those in the body page.
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your upper arm.
Do not place the cuff over thick clothes.
L
F
1. Insert the air plug into the air jack.
2. Put your arm through the cuff loop.
3. Position the arm correctly.
J
a
The bottom edge of the cuff should be
under tube) is centred on the middle of your inner arm. Close the fabric fastener FIRMLY.
1 to 2 cm above the elbow.
Marker (arrow
Notes:
• When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not
to rest your arm on the air tube. ---
b
• The blood pressure can differ between the right arm and the left arm, and therefore also the measured
blood pressure values can be different. OMRON recommends to always use the same arm for
measurement. If the values between the two arms differ substantially, please check with your physician
which arm to use for your measurement.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.2 How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under
comfortable room temperature. No eating, smoking or exercising 30 minutes before
taking a measurement.
• Sit on a chair with your feet flat on the floor.
c
• Sit upright with your back straight. ---
d
• The cuff should be at the same level as your heart. ---
EN
Cuff wrapping guide
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around
the arm. Even when the
is displayed, a blood pressure reading will be taken.
Note:
This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking
care to wrap it correctly and take the measurement again.
When the
is displayed, the cuff is correctly wrapped tightly enough on the arm and the reading is
accurate and reliable.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the O/I START button to turn off the unit and to release the air in the arm cuff.
• Remain still while taking a measurement.
The monitor is designed to take measurements and store the measurement values in
the memory for two people using USER ID A and USER ID B.
1. Press the O/I START button.
2. Select your USER ID (A or B).
3. Press the O/I START button again.
The cuff will start to inflate automatically.
The buzzer will sound when measurement is completed.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
If your systolic pressure is
more than 220 mmHg
START
INFLATE
DEFLATE
END
COMPLETE
After the cuff starts to inflate, press and hold
the O/I START button until the monitor
inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
EN
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Notes:
• During measurement, the buzzer (if set to “on”) will beep in rhythm with your heartbeat.
• Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the
arteries to return to the condition prior to taking the blood pressure measurement.
Warning:
Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of
your doctor.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
4. Remove the arm cuff.
5. Press the O/I START button to turn off the monitor.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after five minutes.
Important:
• Recent research suggests that the following values can be used as a guide to
high blood pressure for measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria, please refer to
Chapter 7.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement. The irregular heartbeat algorithm
automatically determines if the measurement is usable or needs to be repeated. If the
measurement results are affected by irregular heartbeats but the result is valid, the result is
shown together with the irregular heartbeat symbol (
). If the irregular heartbeats cause
the measurement to be invalid, no result is shown. If the irregular heartbeat symbol (
is shown after you have taken a measurement, repeat the measurement. If the irregular
)
heartbeat symbol (
) is shown frequently, please make your doctor aware of it.
• If you move during measurement, the movement error symbol ( ) will appear on the
display. Keep still and repeat the measurement.
EN
11
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.4 Using the Memory Function
The monitor automatically stores the result up to 60 sets for each user (A and B). It
can also calculate an average reading based on the measurements from the last
three readings taken within 10 minutes. If there are only two readings in memory for
that period, the average will be based on two readings. If there is one reading in
memory for that period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date
and time.
To View the Average Value
1. Press the or button.
2. Select your USER ID (A or B).
Note: If there are no measurements results stored in the memory, the
screen to the right is displayed.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
To View the Readings Stored in Memory
1. Press the button, while the average value is displayed.
The Memory number appears for a second
before the pulse rate is displayed. The newest
set is numbered “1”.
Alternating date/
time display
EN
2. Press the or button to view the readings stored in
memory.
: To the older readings
: To the more recent readings
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by USER ID.
When the memory symbol ( ) appears, first press
the
button. Then while holding it down, press
the O/I START button simultaneously for about 2-3
seconds.
Note: You cannot partially delete values stored in the memory.
All values for the user you select will be deleted.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Error Display
Cause
Remedy
Remove the arm cuff. Wait 2-3 minutes and then
take another measurement.
Irregular or weak pulses are detected.
Repeat the steps in section 3.3. If this error
continues to appear, contact your doctor.
Movement during measurement.
Cuff is not applied correctly.
Carefully read and repeat the steps in section 3.3.
Apply the arm cuff correctly. Refer to section 3.1.
You should replace them with new ones ahead of
time.
Refer to section 2.1.
The batteries are low.
Blink
Lit
You should replace them with new ones at once.
Refer to section 2.1.
The batteries are exhausted.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Troubleshooting and Maintenance
Error Display
Cause
Remedy
Carefully read and repeat the steps listed under
section 3.3.
Cuff is under inflated.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to section 3.3.
Movement during measurement.
Insert the air plug securely.
Refer to section 3.1.
EN
Air plug disconnected.
Apply the arm cuff correctly.
Refer to section 3.1.
Arm cuff not applied correctly.
Clothing is interfering with the arm cuff.
Air is leaking from the arm cuff.
Remove any clothing interfering with the arm cuff.
Refer to section 3.1.
Replace cuff with new one.
Refer to Chapter 5.
Do not inflate the arm cuff above 299 mmHg.
Refer to section 3.3.
The arm cuff was inflated above 299
mmHg when inflating the cuff manually.
Device error.
Contact your OMRON retail outlet or distributor.
Note:
The irregular heartbeat symbol (
) may also be displayed with error messages.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Troubleshooting and Maintenance
4.2 Troubleshooting
Problem
Cause
Remedy
Apply the arm cuff correctly.
Refer to section 3.1.
Arm cuff not applied correctly.
Remain still and do not talk during
measurement.
Refer to section 3.3.
Movement or talking during
measurement.
The reading is extremely low (or
high).
Remove any clothing interfering
with the arm cuff.
Refer to section 3.2.
Clothing is interfering with the arm
cuff.
Make sure that the air tube is
connected securely.
Refer to section 3.1.
The air tube is not securely
connected into the main unit.
Arm cuff pressure does not rise.
Arm cuff deflates too soon.
Replace the arm cuff with a new
one.
Refer to Chapter 5.
Air is leaking from the arm cuff.
The arm cuff is loose.
Apply the cuff correctly so that it is
firmly wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it is 30 to
40 mmHg above your previous
measurement result.
Cannot measure or readings are too
low or too high.
The arm cuff has not been inflated
sufficiently.
Refer to section 3.3.
16