Omron Blood Pressure Monitor M10 IT User Manual

Digital Automatic Blood Pressure Monitor  
Model M10-IT  
EN  
FR  
DE  
IT  
• Instruction Manual  
• Mode d’emploi  
• Gebrauchsanweisung  
• Manuale di istruzioni  
• Manual de instrucciones  
• Gebruiksaanwijzing  
• ꢀYꢄꢂHꢂJWXHꢂ ꢁꢂ ꢃꢄWꢁQYFXF\NN  
ꢅꢆꢇꢈꢉꢊꢋꢆ ꢌꢍꢎꢇ •  
ES  
NL  
RU  
AR  
IM-HEM-7080IT-E-07-09/2012  
1663794-1I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Before using the unit  
Introduction  
Thank you for purchasing the OMRON M10-IT Intellisense Upper Arm  
Blood Pressure Monitor.  
The OMRON M10-IT Intellisense is a fully automatic blood pressure  
monitor, operating on the oscillometric principle. It measures your  
blood pressure and pulse rate simply and quickly. For comfortable  
controlled inflation without the need for pressure pre-setting or re-  
inflation the device uses its advanced “Intellisense” technology.  
The monitor stores measurement results for two people and features  
morning and evening averages. The auto mode allows for three  
consecutive measurements with timed intervals for an average  
reading.  
Omron’s blood pressure management software included with this unit  
allows you to view, manage and print data about blood pressure  
measured by the OMRON M10-IT Intellisense. Included software is in  
English only.  
EN  
Please read this instruction manual thoroughly before using  
the unit. For specific information about your own blood  
pressure, CONSULT YOUR DOCTOR.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Important Safety Information  
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis.  
Use the unit only to store the designated users measurement results.  
Please read this section carefully before using the unit.  
Caution:  
(General Usage)  
• Always consult your doctor. You should never stop or change the  
dosage of medication prescribed by your doctor.  
• Do not leave the unit unattended with infants or persons who cannot  
express their consent.  
• Do not use the unit for any purpose other than measuring blood  
pressure.  
• Use only the approved arm cuff for this unit. Use of other arm cuffs  
may result in incorrect measurement results.  
• Do not disassemble the unit or arm cuff.  
• Make sure that the air tube is not wrapped around other parts of your  
body when taking measurements at night. This could result in injury  
when the air pressure in the air tube is increased.  
• Do not wrap the arm cuff tightly when taking measurements at night.  
This could result in injury.  
• Do not inflate the arm cuff over 299 mmHg.  
• Do not allow objects to come into contact with the USB connector on  
the right of the unit when taking measurements.  
(Optional AC Adapter Usage)  
• Use only the original AC adapter designed for this unit. Use of  
unsupported adapters may damage and/or may be hazardous to the  
unit. Plug the AC adapter into the appropriate voltage outlet. Do not  
overload power outlets or use extension cords.  
• Do not use the AC adapter if the unit or the power cord is damaged.  
Turn off the power and unplug the power cord immediately.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Important Safety Information  
• Never plug in or unplug the power cord from the electric outlet with  
wet hands.  
(Battery Usage)  
• If battery fluid should get in your eyes, immediately rinse with plenty  
of clean water. Consult a doctor immediately.  
• If battery fluid should get on your skin or clothing, immediately rinse  
with plenty of clean water.  
• Use only four “AA” alkaline batteries with this unit. Do not use other  
types of batteries.  
• Do not insert the batteries with their polarities incorrectly aligned.  
• Replace old batteries with new ones immediately. Replace all four  
batteries at the same time.  
• Remove the batteries if the unit will not be used for three months or  
more. If the batteries have been removed for more than 30 seconds,  
the date and time setting will need to be reset.  
• Do not use new and used batteries together.  
• Read and follow the “Important information regarding Electro  
Magnetic Compatibility (EMC)” in the Technical Data Section.  
General Safety Precautions  
• Do not forcibly bend the arm cuff or bend the air tube excessively.  
• When removing the air tube, pull on the air plug at the connection with  
the main unit not the tube itself.  
EN  
• Do not apply strong shocks and vibrations to or drop the main unit  
and arm cuff.  
• Do not inflate the arm cuff when it is not wrapped around your arm.  
• Do not wash the arm cuff or immerse it in water.  
• Do not take measurements after bathing, drinking alcohol, smoking,  
exercising or eating.  
• Read and follow the “Correct Disposal of This Product” in the  
Technical Data Section when disposing of the device and any used  
accessories or optional parts.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Overview  
Main Unit  
A
H
G
F
E
B
C
J
D
I
A. Display  
H. Battery compartment  
I. Air jack  
J. AC adapter jack (for AC  
adapter)  
B. SET button  
C. Morning/Evening Average  
(
/
) buttons  
D. MEMORY button  
E. USB connector jack (USB 1.1  
compatible)  
F. O/I START button  
G. User ID selection switch  
(A & B)  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Overview  
Arm Cuff  
K
L
M
K. Arm cuff  
(Cuff: arm circumference 22-42 cm)  
L. Air Plug  
M. Air Tube  
EN  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Overview  
Display  
N
V
W
O
X
P
R
T
Y
Q
S
U
Z
AA  
AC  
AB  
AD  
N. Systolic blood pressure  
O. Diastolic blood pressure  
P. Heartbeat symbol  
V. User ID symbol (A or B)  
(Indicates user A or user B as  
selected by the user to take a  
measurement or when using  
the memory function.)  
W. Pulse display  
1. (Flashes during measurement)  
2. (If flashing after measurement  
completed indicates blood  
pressure out of recommended  
range)  
X. Early morning hypertension  
symbol  
(Displayed if the morning  
average reading for a week is  
above 135 for the systolic  
blood pressure value and/or 85  
for the diastolic blood pressure  
value.)  
Q. Movement symbol  
(Displayed if you move your  
body during the measurement.)  
R. Irregular heartbeat symbol  
S. Deflation symbol  
T. Auto mode symbol  
(Displayed when Auto Mode is  
set.)  
Y. Morning average symbol  
(Displayed when viewing  
morning averages using the  
memory function.)  
U. Battery low symbol  
Z. Evening average symbol  
(Displayed when viewing  
evening averages using the  
memory function.)  
AA.Average value symbol  
(Displayed when viewing value  
for the last three  
measurements)  
AB.Memory symbol  
(Displayed when viewing  
values stored in memory)  
AC.Week display  
AD.Date/Time display  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Overview  
Package Contents  
AE  
AG  
AF  
AH  
AE.Soft storage bag  
• Instruction manual  
• Guarantee card  
AF.Four “AA” alkaline (LR6)  
batteries  
• Blood pressure pass  
AG.USB cable  
AH.CD-ROM (Blood pressure  
management software)  
• Blood Pressure Monitor unit  
• Installation guide (for Blood  
pressure management  
software)  
EN  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
2.1 Installing/Replacing the Batteries  
1. Turn the main unit upside down.  
2. Slide the battery cover in the  
direction of the arrow while  
pressing the ribbed part of the  
cover.  
3. Install or replace four AAsize  
batteries so that the + (positive)  
and - (negative) polarities  
match the polarities indicated  
on the battery compartment.  
4. Put the battery cover back in place.  
Slide the battery cover as indicated, until it clicks into place.  
Note: The measurement values continue to be stored in memory even  
after the batteries are replaced.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
Battery Life & Replacement  
If the battery low symbol ( ) appears on the display, replace all four  
batteries at the same time.  
- When the battery low symbol ( ) starts to blink, you will still be  
able to use the unit for a short while. You should replace the  
batteries with new ones ahead of time.  
- When the symbol ( ) remains lit, the batteries are exhausted. You  
should replace the batteries with new ones at once.Turn the unit off  
before replacing the batteries.  
• Remove the batteries if the unit will not be used for three months or  
more.  
• If the batteries have been removed for more than 30 seconds, the  
Date/Time setting will need to be reset. See Chapter 2.2 for details.  
• Dispose of batteries according to applicable local regulations.  
Four new AAalkaline batteries will last for approximately  
1400 measurements, when used to take six measurements a day.  
Since the supplied batteries are for monitoring use only, they may have  
a shorter life and not last for 1400 measurements.  
EN  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
2.2 Setting the Date and Time  
Your blood pressure monitor automatically stores up to 84 individual  
measurement values with the date and time.  
To make use of the memory and average values functions:  
• Set the unit to the correct date and time before taking a measurement  
for the first time.  
• If the batteries have been removed for more than 30 seconds, the  
Date/Time setting will need to be reset.  
1. When the O/I START button is  
pressed to turn the unit on for  
the first time after inserting  
batteries, the year digits  
(2007) will flash on the  
display.  
2. Press the MEMORY button to  
advance the digits one at a  
time.  
Notes:  
• The range for the year  
setting is 2007 to 2030. If the  
year reaches 2030, it will  
return to 2007.  
• If you hold down the  
MEMORY button, the digits  
will advance rapidly.  
3. Press the SET button to confirm the setting when the desired  
number appears on the display.  
The year is set and the month digits flash on the display.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
4. Repeat steps 2 and 3 to set  
the month.  
The month is set and the day  
digits flash on the display.  
5. Repeat steps 2 and 3 to set  
the day.  
The day is set and the hour  
digits flash on the display.  
6. Repeat steps 2 and 3 to set  
EN  
the hour.  
The hour is set and the  
minutes digits flash on the  
display.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
7. Repeat steps 2 and 3 to set  
the minutes.  
The minutes settings is set.  
The unit automatically turns  
itself off after the minute  
setting has been set.  
To adjust the date and time,  
press the SET button while  
the unit is in standby mode.  
Notes on Adjusting the Date and Time Setting  
If you need to adjust the date and time for some reason, or if the date  
and time has been reset after replacing the batteries, adjust the date  
and time between 10:00 and 18:59. This will avoid problems with  
morning and evening weekly averages stored in memory.  
To adjust the date or time at a later stage, start the unit in the off  
position. Follow the steps below.  
1. Press and hold the SET button.  
The auto mode symbol  
and the current setting (on or  
oFF) will appear on the display for the user ID selected.  
If the year is not flashing, press the SET button once or twice  
until the year is flashing.  
The year flashes on the display.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
2. Refer to Chapter 2.2.2 to start setting the date and time.  
With this manual setting of date and time the unit needs to be  
manually shut off. It will not automatically turn off after the  
minutes have been set.  
Notes:  
• Please ensure that the date and time have been set correctly so that  
the weekly averages will be displayed correctly. If the Date/Time  
settings have been accidentally set incorrectly, press the SET button  
to circle through the settings in order to adjust to their correct values.  
• In the case where the time displays incorrectly after a long period of  
non-use of the unit, reset the Date and Time (refer to Chapter 2.2 for  
instructions on setting the clock values). Erase the recorded values  
when reusing the unit (refer to Chapter 3.5 for directions on erasing  
the recorded values).  
EN  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating instructions  
3. Using the Unit  
3.1 How to Sit Correctly When Taking a Measurement  
You can take a measurement on either your left or right arm.  
Notes:  
• Measurements should be taken in a quiet place and you should be in  
a relaxed, seated position. Make sure that the room is not too hot or  
cold.  
• Avoid eating, smoking, or exercising for at least 30 minutes before  
taking a measurement.  
• Do not move or talk during measurement.  
Correct  
Posture  
Remove tight fitting clothing from  
your upper arm, and any thick  
clothing such as a sweater.  
Do not place the cuff over thick  
clothes and do not roll up your sleeve  
if it is too tight.  
Sit upright  
with your  
back straight.  
Place your arm on a table so  
that the cuff will be at the same  
level as your heart.  
The gap between the  
chair and the top of the  
table should be between  
25cm to 30cm.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Notes:  
• Correct posture during measurement is necessary to get accurate  
results.  
• You should also try to measure your blood pressure at the same time  
each day. (Within 1 hour after waking up is recommended.)  
Incorrect Posture  
Arched back (leaning  
forwards)  
Sitting cross-legged  
Sitting on a sofa or at a low  
table so that you tend to lean  
forward  
These situations could lead to higher blood pressure values due to  
strain or the arm cuff being lower than the heart.  
If the arm cuff is at a lower position than your heart use cushions  
etc., to adjust the height of your arm.  
EN  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.2 Applying the Arm Cuff  
You can wrap the cuff either on your right or left arm.  
• Remove tight-fitting clothing from your  
upper arm.  
• Do not place the cuff over thick clothes and  
do not roll up your sleeve if it is too tight.  
1. Insert the air plug into the air jack (on the left side of the  
device).  
Air jack  
Air plug  
The cuff must be fully deflated when it is inserted into the air  
jack.  
2. Sit in a chair with your feet flat on the floor and place your arm  
on a table so that the cuff will be at the same level as your  
heart.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3. Hold the grip on the cuff securely  
Grip  
with your hand.  
Fabric fastener  
4. Turn the palm of your hand upward.  
5. Apply the cuff to your upper arm so that the blue strip is centred  
on the middle of your inner arm and points down the inside of  
the arm. The air tube should run down the inside of your  
forearm and be in line with your middle finger.  
The bottom of the cuff should be approximately 1 to 2 cm  
above the elbow.  
Bottom edge 1 to 2 cm  
Blue strip  
EN  
Air tube  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
6. When the cuff is positioned correctly,  
close the fabric fastener FIRMLY.  
Notes:  
• Make certain the cuff fits snugly around your arm.  
• The cuff should make good contact with your skin. You should  
be able to fit your index finger between the cuff and your arm  
easily, so you can pull the cuff off and on.  
• Make sure that there are no kinks in the air tubing.  
Taking measurements on the right arm  
Please note the following points when  
applying the cuff to your right arm. Apply the  
cuff so that the air tube is at the side of your  
elbow.  
Notes:  
• Be careful not to rest your arm on the air  
tube, or otherwise restrict the flow of air to  
the cuff.  
• Apply the arm cuff so that no part of the  
cuff is positioned over the elbow joint. The  
cuff should be 1 to 2 cm above the elbow.  
• If used 2 times per day, the cuff will last for 3 years.  
• The blood pressure can differ between the right arm and the left arm,  
and therefore also the measured blood pressure values can be  
different. Omron recommends to always use the same arm for  
measurement. If the values between the two arms differ substantially,  
please check with your physician which arm to use for your  
measurement.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.3 Taking a Reading  
The unit is designed to take measurements and store the  
measurement values in the memory for two people using user ID A and  
user ID B.  
The unit can also be used to take a single measurement for other  
persons using the guest mode. The user ID symbol is not displayed  
when using the guest mode. Measurements taken using the guest  
mode are not stored in the memory.  
Using the Guest Mode  
1. Press and hold the O/I START button until the A and B user  
symbols disappear from the display.  
A and B display  
goes out  
Press for at least 3 seconds  
EN  
2. Release the O/I START button.  
The cuff starts to inflate automatically.  
Note: To stop the inflation press and release the O/I START  
button. The unit will stop inflating, start deflation and turn  
off.  
3. Inflation stops and the measurement is started. When the  
measurement is completed, your blood pressure and pulse rate  
appear on the display.  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Selecting the User ID  
Select your user ID. Always use the same user ID when taking a  
measurement. The unit stores the measurement values in the memory  
by the user ID selected. These measurement values are used to  
calculate the morning averages and evening averages.  
Slide the user ID selection switch up or down to select user A or B.  
Using the User ID  
When you select a user ID the unit can be set to take one  
measurement using the single mode or three consecutive  
measurements using the auto mode when you press the O/I START  
button.  
Note: The default value is set to the single mode to take one  
measurement.  
The auto mode takes three consecutive measurements. The unit will  
inflate the cuff and take three measurements separated by a short  
interval of time between each measurement.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Using the Single Mode  
1. Select your user ID (A or B).  
2. Press the O/I START button.  
All symbols appear on the display. The user ID symbol you  
selected flashes on the display.  
The cuff starts to inflate automatically. As the cuff inflates, the  
unit automatically determines your ideal inflation level. This unit  
detects the pulse during inflation. Do not move your arm and  
remain still until the entire measurement process is completed.  
EN  
Note: To stop the inflation or measurement, press and release  
the O/I START button. The unit will stop inflating, start  
deflating and will turn off.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3. Inflation stops automatically and the measurement is started.  
As the cuff deflates, decreasing numbers appear on the  
display. The heartbeat symbol ( ) flashes at every heartbeat.  
4. When the measurement is complete, the arm  
cuff completely deflates. Your blood pressure  
and pulse rate are displayed.  
5. Press the O/I START button to turn  
the unit off.  
Note: If you forget to turn the  
monitor off, it will  
automatically turn off after  
five minutes.  
Notes:  
• Self-diagnosis of measured results and treatment are  
dangerous. Please follow the instructions of your doctor.  
• Wait 2-3 minutes before taking another blood pressure  
measurement. Waiting between readings allows the arteries  
to return to the condition prior to taking the blood pressure  
measurement.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Selecting the Auto Mode  
Auto Mode may be selected for either A or B user.  
1. Make sure the unit is powered off.  
2. Select your user ID (A or B).  
3. Press and hold the SET button.  
The auto mode symbol (  
the display.  
) and the oFF setting appears on  
EN  
4. Press the MEMORY button ( )  
to select on.  
Note: Press the MEMORY  
button to alternate  
between on and oFF.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
5. Press the SET button to confirm the change.  
The number of seconds between each measurement is  
displayed.  
Note: The default value between each measurement is  
60 seconds.  
6. Press the MEMORY button ( ) to change the interval of time  
between each measurement.  
The interval can be set to 15 seconds, 30 seconds, 60 seconds  
or 120 seconds. Continue to press the MEMORY button until  
the desired interval appears on the display.  
7. Press the SET button to confirm the  
setting.  
The year flashes on the display.  
8. Press the O/I START button to turn  
the unit off.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Using the Auto Mode  
1. Select your user ID (A or B).  
2. Press the O/I START button.  
All symbols appear on the display. The user ID Symbol you  
selected flashes on the display. The cuff starts to inflate  
automatically.  
Note: To stop the inflation or measurement, press the O/I  
START button. The unit will stop inflating, start deflating  
and will turn off.  
3. Inflation stops automatically and the measurement is started.  
As the cuff deflates, decreasing numbers appear on the  
display. The heartbeat symbol ( ) flashes at every heartbeat.  
When the measurement is complete, the arm cuff completely  
deflates.  
Note: Do not move your arm and remain still until the entire  
measurement process is completed.  
EN  
4. Check the progress of the consecutive measurements.  
The first measurement is complete, waiting to start the next  
measurement. The interval of time set between each  
measurement will appear on the display.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
The second measurement starts automatically.  
The second measurement is complete, waiting to start the next  
measurement.  
The third measurement starts automatically.  
The consecutive measurements are complete. The average for  
the three measurements appears on the display.  
The three individual measurement results are not displayed  
while Auto Mode is taking the measurements. However, the  
results for each individual measurement can be displayed after  
all the measurements have been completed.  
Note: During the Auto Mode the unit will repeat the  
measurement in case of an error. After 3 errors the unit  
will stop taking further measurements. If only two  
measurements are successful the average of these two  
measurements will be displayed. If only one  
measurement is successful, this result will be displayed.  
5. Press the SET button to display the blood pressure and pulse  
rate for the first measurement. Press the SET button again to  
display the next measurement.  
6. Press the O/I START button to turn the unit off.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Important:  
• If your systolic or diastolic pressure is  
outside the standard range, the heartbeat  
symbol will blink when the measurement  
result is displayed.  
Recent research suggests that the following  
values can be used as a guide to high blood  
pressure for measurements taken at home.  
Systolic Blood Pressure  
Above 135 mmHg  
Diastolic Blood Pressure  
Above 85 mmHg  
This criteria is for home blood pressure measurement.  
For professional office blood pressure measurement criteria,  
please refer to Chapter 9.  
• Your blood pressure monitor includes an  
irregular heartbeat feature. Irregular  
heartbeats can influence the results of the  
measurement. The irregular heartbeat  
algorithm automatically determines if the  
measurement is usable or needs to be  
repeated. If the measurement results are  
affected by irregular heartbeats but the result  
is valid, the result is shown together with the irregular  
heartbeat icon. If the irregular heartbeats cause the  
measurement to be invalid, no result is shown. If the irregular  
EN  
heartbeat symbol (  
) is shown after you have taken a  
measurement, repeat the measurement. If the irregular  
heartbeat icon is shown frequently, please make your doctor  
aware of it.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
What is Irregular Heartbeat?  
Normal Heartbeat  
Irregular Heartbeat  
An irregular heartbeat is a  
heartbeat rhythm that varies  
by more than 25% from the  
average heartbeat rhythm  
detected while the unit is  
measuring the systolic and  
diastolic blood pressure.  
Pulse  
Blood pressure  
Long  
Short  
Pulse  
If such an irregular rhythm is  
Blood pressure  
detected more than twice  
during measurement, the irregular heartbeat symbol (  
)
appears on the symbol when the measurement results are  
displayed.  
What is Arrhythmia?  
A heartbeat is stimulated by electrical signals that cause the  
heart to contract.  
Arrhythmia is a condition where the heartbeat rhythm is  
abnormal due to flaws in the bio-electrical system that drives  
the heartbeat. Typical symptoms are skipped heartbeats,  
premature contraction, an abnormally rapid (tachycardia) or  
slow (bradycardia) pulse. This can be caused by heart disease,  
aging, physical predisposition, stress, lack of sleep, fatigue etc.  
Arrhythmia can only be diagnosed by a doctor through a  
special examination.  
Whether the appearance of the irregular heartbeat symbol  
(
) in the results indicates arrhythmia or not can only be  
determined by an examination and diagnosis by your doctor.  
If the irregular heartbeat symbol (  
) is shown frequently,  
please make your doctor aware of it. Conducting self-diagnosis  
and treatment based on measurement results is dangerous. Be  
sure to follow the instructions of your doctor.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.4 Instructions for Special Conditions  
If your systolic pressure is known to be more than 220 mmHg, press  
and hold the O/I START button until the arm cuff inflates 30 to  
40 mmHg higher than your suspected systolic pressure.  
1. Press the O/I START button to turn  
the unit on.  
Measurement starts.  
2. When the cuff starts to inflate, press the O/I START button and  
keep it pressed until the pressure is 30 to 40 mmHg higher than  
your suspected systolic pressure.  
Note: You cannot inflate the cuff above 299 mmHg. (An error  
will be displayed if you try to inflate the cuff above  
300 mmHg.)  
3. Release the O/I START button when the cuff has been inflated  
to the desired pressure.  
EN  
The cuff starts to deflate and measurement starts.  
4. The rest of the procedure is the same as for normal  
measurement. Refer to Chapter 3.3.  
Note: Do not apply more pressure than necessary.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.5 Using the Memory Function  
The unit is designed to store the blood pressure and the pulse rate in  
the memory for two people (user A and user B) every time a  
measurement is completed.  
The unit automatically stores up to 84 sets of measurement values  
(blood pressure and pulse rate) for each user (A and B). When 84 sets  
of measurement values are stored, the oldest record is deleted to save  
the most recent values. The unit also stores 8 weeks of morning  
averages and 8 weeks of evening averages for each user (A and B).  
Average Function  
The unit calculates the average reading based on the three most  
recent sets of measurement values taken within 10 minutes of the most  
recent reading.  
Note: If two sets of measurement values are stored in the memory for  
the 10 minutes period, the average is based on the two sets of  
measurement values. If one set of measurement values is  
stored, this is displayed as the average.  
To Display the Measurement Values  
1. Select your user ID (A or B).  
2. Press the MEMORY button ( ).  
The average value is shown and the average symbol ( )  
appears above the memory symbol ( ) on the display.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3. Press the MEMORY button ( ) to display the most recent set  
of individual measurement values on the screen. The values  
are displayed from the most recent to the oldest.  
Note: The date and time are  
alternately displayed with the  
measurement values.  
Alternating display  
Press the MEMORY button to display the next set of values.  
Press and hold the MEMORY button to display the values  
faster.  
Auto mode  
The average for the consecutive measurements appears on  
the display with the auto mode symbol (  
). To display the  
values for the individual measurements press the SET button  
while the average is displayed.  
EN  
4. Press the O/I START button to turn the unit off.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Morning and Evening Averages  
The unit calculates and displays a weekly average for measurements  
taken in the morning ( ) and evening ( ). The unit stores 8 weeks of  
morning averages and 8 weeks of evening averages for each user (A  
and B).  
Note: The week begins Sunday at 4:00 AM  
Morning Averages  
Morning averages are based on the first average reading using the  
auto mode or the first reading using the single mode.  
Measurement times must be between:  
4:00 AM and 9:59 AM.  
Evening Averages  
Evening averages are based on the last average reading using the  
auto mode or the last reading using the single mode.  
Measurement times must be between:  
7:00 PM and 1:59 AM.  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
About the Weekly Averages  
Morning Weekly Average  
This is the average for measurements taken during the morning  
(4:00 AM-9:59 AM) between Sunday and Saturday. An average for  
each day is calculated for up to three measurements taken within 10  
minutes of the first measurement of the morning.  
Measurements of the morning  
Sunday  
Monday  
Tuesday Wednesday Thursday  
Friday  
Saturday  
4:00 AM  
First measurement of  
Measurement  
Measurement  
Measurement  
Measurement  
the morning  
Up to 3 measurements  
Within 10 minutes  
Morning Blood Pressure  
Measurement  
Measurement  
9:59 AM  
Evening Weekly Average  
This is the average for measurements taken during the evening  
(7:00 PM-1:59 AM) between Sunday and Saturday. An average for  
each day is calculated for up to three measurements taken within  
10 minutes of the last measurement of the evening.  
EN  
Measurements of the evening  
Sunday  
Monday  
Tuesday Wednesday Thursday  
Friday  
Saturday  
7:00 PM  
Measurement  
Measurement  
Within 10 minutes  
Measurement  
Measurement  
Measurement  
Measurement  
Up to 3 measurements  
Last measurement of  
the evening  
Evening Blood Pressure  
1:59 AM  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
To Display Morning and Evening Averages  
1. Select your user ID (A or B).  
2. Press the morning average button  
(
) or the evening average  
button (  
).  
The average for the current week  
“THIS WEEK” appears on the  
display.  
The morning and evening  
average value of the same week  
can be displayed by pressing the  
morning average button (  
or the evening average button  
)
(
). It is important to check both morning and evening  
average of the same week.  
3. Continue to press the morning average button (  
) or the  
evening average button (  
) to display the previous weeks.  
The unit displays “-1 WEEK” for the previous week to “-7  
WEEK” for the oldest set of averages.  
Note: If there are not enough measurements  
taken in a week to calculate the average  
the display will appear as illustrated.  
4. Press the O/I START button to turn the unit off.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
Notes on Display Combinations  
In addition to the morning average and evening average symbols, the  
unit may also display the morning hypertension symbol if the morning  
average for that week is above the home blood pressure guideline.  
(See Chapter 3 “Using the Unit - Important information” for details.)  
Depending on your measurement results, these may be displayed in  
the following combinations.  
Display  
Morning  
Average  
Morning average from This Week Morning average from 7 weeks  
with Morning Average Symbol  
ago with Morning Average  
Symbol + blinking heartbeat  
Symbol + Morning Hypertension  
Symbol  
EN  
Evening  
average  
Evening average from This Week  
with Evening Average Symbol  
Evening average from 7 weeks  
ago with Evening Average  
Symbol + blinking Heartbeat  
Symbol + Morning Hypertension  
Symbol  
Average within home blood  
pressure guideline  
Average above home blood  
pressure guideline + Morning  
Hypertension  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
The morning hypertension symbol ( ) appears if the weekly average  
for morning measurements is above 135/85.  
In this case the morning hypertension symbol ( ) is displayed when  
the evening average is displayed, regardless of the values for the  
evening average.  
To Delete All Values Stored in the Memory  
The values stored in the memory are deleted by user ID.  
You cannot partially delete values stored in the memory. All values for  
the user you select will be deleted.  
1. Select your ID (A or B).  
2. While holding the MEMORY button ( ) press the O/I START  
button simultaneously for more than 2 seconds and all values  
will be deleted.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.6 Downloading Data to a PC  
Omron’s blood pressure management software included with this unit  
allows you to view, manage and print data about blood pressure  
measured by the OMRON M10-IT Intellisense.  
Please see attached “Installation Guide” before installing software onto  
a PC. To operate the software application, please refer to the  
Installation Guide and use the help function in the software application.  
Use only the authorized OMRON USB cable sold with this unit. Use of  
any other cable may cause damage to the unit and will void the user  
warranty.  
1. Open the USB connector jack  
cover.  
2. Plug the normal USB connector  
Normal USB  
plug (large connector) into the  
personal computer.  
Connector  
EN  
Mini USB Connector  
3. Plug the mini USB connector  
plug (small connector) into the  
USB connector on the main unit.  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Quick Reference Guide  
This Quick Reference Guide is for reference only. Please  
read the entire manual before using the unit for the first time.  
To help ensure a reliable reading, avoid eating, smoking, or  
exercising for at least 30 minutes before taking a  
measurement.  
Note: Remove any tight-fitting clothing from your upper arm.  
1. Sit on a chair with your feet flat on the floor and place your arm  
on a table so that the arm cuff will be at the same level as your  
heart.  
2. Secure the cuff around your arm using the fabric fastener strip.  
3. Select your user ID (A or B).  
4. Press the O/I START button.  
When measurement is complete, the unit  
displays your blood pressure and pulse  
rate, and automatically deflates the cuff.  
Notes:  
• Always wait at least 2-3 minutes before taking another blood  
pressure measurement.  
• Please note that all measurement results are stored in the  
memory. If different people take measurements using the  
same device, make sure that you are aware of this fact.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Care and maintenance  
5. Handling Errors and Problems  
5.1 Error Messages  
Error Display  
Cause  
Remedy  
Cuff is under inflated.  
Movement during  
measurement.  
Carefully read and repeat the steps  
listed under Chapter 3.3.  
Cuff over inflated.  
Replace all four “AA” batteries with  
new ones.  
Refer to Chapter 2.1.  
Batteries have run down.  
EN  
Blinks or appears  
continuously.  
Carefully read and repeat the steps  
listed under Chapter 3.3.  
Movement, sound.  
Cuff leaks air.  
Replace with a new cuff.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Handling Errors and Problems  
5.2 Troubleshooting  
Problem  
Cause  
Remedy  
No display appears  
when the O/I START  
button is pressed.  
The batteries are empty.  
Is the cuff wrapped  
Replace the batteries.  
Attach the arm cuff  
correctly around the arm? correctly.  
Cannot measure or  
The cuff pressure does  
not rise although the  
pump motor can be  
heard.  
Check that the air plug is  
readings are too high.  
properly connected to the  
unit. Push the air plug  
firmly into the air jack.  
The blood pressure is  
different each time. The  
reading is extremely low  
(or high).  
Blood pressure readings constantly vary with time of  
measurement and nervous condition. Take deep  
breaths to relax before taking a measurement.  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Maintenance and Storage  
To protect your unit from damage, please observe the  
following:  
• Do not subject your unit and cuff to extreme temperatures,  
humidity, moisture or direct sunlight.  
• Do not fold the cuff or tubing tightly.  
• Do not disassemble the unit.  
• Do not carry out repairs of any kind yourself. If a defect occurs,  
consult the OMRON distributor or Customer Services as  
mentioned on the packaging, or seek the advice of your surgical  
appliance stockist or pharmacist.  
• The OMRON M10-IT is a precision measuring device. It is  
important that the accuracy remains within the specifications. It is  
recommended that you have the unit inspected every two years  
to ensure it is accurate and functioning correctly. Consult the  
OMRON distributor or Customer Services as mentioned on the  
packaging for further details.  
• Do not subject the unit to strong shocks or vibrations (for  
example, dropping the unit on the floor.)  
• Do not use volatile liquids to clean the unit. THE UNIT SHOULD  
BE CLEANED WITH A SOFT, DRY CLOTH.  
• Do not wash the arm cuff or immerse it in water.  
EN  
Storage  
Keep the unit in its soft storage bag when not in  
use.  
Note: If you are using the optional AC adapter,  
make sure that the adapter is placed  
under the cuff and unit so that it does not  
damage the display.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Optional Parts  
AC Adapter S  
Arm Cuff  
Order No. 9515336-9  
Order No. 9956685-4  
Using the Optional AC Adapter  
Insert batteries into the battery compartment even when using the AC  
adapter.  
Notes:  
• Never plug in or unplug the power cord from the electric outlet with  
wet hands.  
• Use only the original AC adapter designed for this unit. Use of  
unsupported adapters may damage the unit.  
• When storing the AC adapter with the main unit, be careful not to  
damage the main unit or cuff.  
1. Insert the AC adapter plug into  
the AC adapter jack on the  
rear side of the main unit.  
2. Plug the AC adapter into an  
electrical outlet.  
To disconnect the AC adapter, unplug the AC adapter from the  
electrical outlet first and then immediately remove the AC  
adapter plug from the main unit.  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Technical Data  
Product Description  
Model  
Digital Automatic Blood Pressure Monitor  
OMRON M10-IT (HEM-7080IT-E)  
LCD Digital Display  
Display  
Measurement Method  
Measurement Range  
Oscillometric method  
Pressure: 0 mmHg to 299 mmHg  
Pulse: 40 to 180/min.  
Accuracy  
Pressure: 3 mmHg  
Pulse: 5% of display reading  
Inflation  
Deflation  
Memory  
Fuzzy-logic controlled by electric pump  
Automatic pressure release valve  
84 Measurements with date and time for each user  
(A and B)  
Power Source  
Battery Life  
4 “AA” batteries 1.5V or AC/ DC adapter (6V = 4W)  
Capacity of new alkaline batteries is approx. 1400  
measurements when used 6 times a day with 4 new  
alkaline batteries  
Operating temperature/  
Humidity  
+10°C to +40°C  
Maximum: 30 to 90% RH  
Storage temperature/  
-20°C to +60°C  
Maximum: 10 to 95% RH  
700 - 1060 hPa  
Humidity/Air Pressure  
Console Weight  
Cuff Weight  
Approximately 420g without batteries  
Approximately 240g  
Outer Dimensions  
Cuff Dimensions  
Approximately 155 (l) mm 131 (w) mm 84(h) mm  
Approximately 152 mm 600 mm  
(Cuff: arm circumference 22 to 42 cm)  
EN  
Package Content  
Cuff, instruction manual, soft storage bag, battery set,  
guarantee card, blood pressure pass, USB cable,  
CD-ROM, installation guide  
Note: Subject to technical modification without prior notice.  
• This OMRON product is produced under the strict quality system of OMRON  
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure  
monitors, which is the Pressure Sensor, is produced in Japan for assembly.  
= Type B  
This device fulfils the previsions of EC directive 93/42/EEC (Medical Device Directive).  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Technical Data  
Important information regarding Electro Magnetic Compatibility (EMC)  
With the increased number of electronic devices such as PC’s and mobile (cellular)  
telephones, medical devices in use may be susceptible to electromagnetic  
interference from other devices. Electromagnetic interference may result in incorrect  
operation of the medical device and create a potentially unsafe situation.  
Medical devices should also not interfere with other devices.  
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with  
the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has  
been implemented. This standard defines the levels of immunity to electromagnetic  
interferences as well as maximum levels of electromagnetic emissions for medical  
devices.  
This medical device manufactured by OMRON HEALTHCARE conforms to this  
EN60601-1-2:2007 standard for both immunity and emissions.  
Nevertheless, special precautions need to be observed:  
• Do not use mobile (cellular) telephones and other devices, which generate strong  
electrical or electromagnetic fields, near the medical device. This may result in  
incorrect operation of the unit and create a potentially unsafe situation.  
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of  
the device in case the distance is shorter.  
Further documentation in accordance with EN60601-1-2:2007 is available at  
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction  
manual.  
Correct Disposal of This Product  
(Waste Electrical & Electronic Equipment)  
This marking shown on the product or its literature, indicates that it  
should not be disposed of, with other household wastes at the end of  
its working life. To prevent possible harm to the environment or human  
health from uncontrolled waste disposal, please separate this from  
other types of wastes and recycle it responsibly to promote the  
sustainable reuse of material resources.  
Household users should contact either the retailer where they purchased this product,  
or their local government office, for details of where and how they can take this item  
for environmentally safe recycling.  
Business users should contact their supplier and check the terms and conditions of  
the purchase contract. This product should not be mixed with other commercial  
wastes for disposal.  
This product does not contain any hazardous substances.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Some Useful Information about Blood Pressure  
What is Blood Pressure?  
Blood pressure is a measure of the force of blood flowing against the  
walls of the arteries. Arterial blood pressure is constantly changing  
during the course of the heart's cycle.  
The highest pressure in the cycle is called the Systolic Blood Pressure;  
the lowest is the Diastolic Blood Pressure.  
Both pressure readings, the Systolic and Diastolic, are necessary to  
enable a doctor to evaluate the status of a patient's blood pressure.  
Why is it a Good Thing to measure Blood Pressure at Home?  
Having your blood pressure measured by a doctor can cause anxiety  
which is itself a cause of high blood pressure. As a variety of conditions  
affect blood pressure, a single measurement may not be sufficient for  
an accurate diagnosis.  
Many factors such as physical activity, anxiety, or the time of day, can  
influence your blood pressure. Thus it is best to try and measure your  
blood pressure at the same time each day, to get an accurate  
indication of any changes in blood pressure. Blood pressure is typically  
low in the morning and increases from afternoon to evening. It is lower  
in the summer and higher in the winter.  
Blood pressure is measured in millimetres of mercury (mmHg) and  
measurements are written with the systolic pressure before the  
diastolic e.g. A blood pressure written as 140/90, is referred to as  
140 over 90 mmHg.  
EN  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Some Useful Information about Blood Pressure  
Upper curve: systolic blood pressure  
Lower curve: diastolic blood pressure  
200  
150  
100  
50  
6
12  
18  
24  
Time of day  
Example: fluctuation within a day (male, 35 years old)  
Classification of Blood Pressure by the World Health  
Organization  
The World Health Organization (WHO) and the International Society of  
Hypertension (ISH) developed the Blood Pressure Classification  
shown in this figure.  
Systolic blood pressure  
(mmHg)  
Severe hypertension  
180  
Moderate hypertension  
160  
Mild hypertension  
140  
Normal systolic value  
130  
Normal blood  
pressure  
120  
Optimal blood  
pressure  
(target value)  
Diastolic  
blood  
pressure  
(mmHg)  
80  
85  
90  
100 110  
This classification is based on the blood pressure values measured on  
people in a sitting position in outpatient departments of hospitals.  
Note: There is no universally accepted definition of hypotension.  
However, those having the systolic pressure below 100 mmHg  
are assumed as hypotensive.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Some Useful Information about Blood Pressure  
Manufacturer  
OMRON HEALTHCARE Co., Ltd.  
53, Kunotsubo, Terado-cho, Muko, Kyoto,  
617-0002 JAPAN  
EU-representative  
OMRON HEALTHCARE EUROPE B.V.  
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS  
OMRON (DALIAN) CO., LTD.  
Production facility  
Subsidiary  
Dalian, CHINA  
OMRON HEALTHCARE UK LTD.  
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.  
OMRON MEDIZINTECHNIK  
HANDELSGESELLSCHAFT mbH  
John-Deere-Str. 81a, 68163 Mannheim, GERMANY  
OMRON SANTÉ FRANCE SAS  
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE  
Made in China  
EN  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  

MTD Tiller 410 422 User Manual
Multiquip Water Pump qp402h User Manual
Murphy Battery Charger S10 Series User Manual
Nintendo Video Games 45496740993 User Manual
Nokia Cell Phone 3110 User Manual
Omega Juicer VRT400 User Manual
Optimus Portable CD Player CD 3330 User Manual
OWI Speaker V6810 User Manual
Palm Personal Computer 406 4555A US User Manual
Panasonic Answering Machine KX F2710NZ User Manual