Made in USA
UserÕs Guide
®
e-mail: [email protected]
DP119-RTD
Miniature Panel
Thermometer
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE OF CONTENTS
SECTION
PAGE
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Models Available . . . . . . . . . . . . . . . . . . . . . . . . .iii
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Safety Considerations . . . . . . . . . . . . . . . . . . . . .2
About the Meter . . . . . . . . . . . . . . . . . . . . . . . . .3
Front of the Meter . . . . . . . . . . . . . . . . . . . . . . . .3
Back of the Meter . . . . . . . . . . . . . . . . . . . . . . . .4
Connector Description . . . . . . . . . . . . . . . . . . . . .4
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . .5
Main Board Power Jumpers . . . . . . . . . . . . . . . .5
Changing the Meter Parameters . . . . . . . . . . . . .7
Installation and Panel Mounting . . . . . . . . . . . . .10
Sensor Input Connections . . . . . . . . . . . . . . . . .12
Analog Output Connections . . . . . . . . . . . . . . . .14
Display Hold Connections . . . . . . . . . . . . . . . . .15
Main Power Connections . . . . . . . . . . . . . . . . .16
Disassembly/Assembly . . . . . . . . . . . . . . . . . . .18
Operation and Calibration . . . . . . . . . . . . . . . . .20
Equipment Required . . . . . . . . . . . . . . . . . . . . .20
Calibration Procedure . . . . . . . . . . . . . . . . . . . .21
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .23
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Section 1
1.1
1.2
Section 2
2.1
2.2
2.3
Section 3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
Section 4
4.1
4.2
Section 5
Section 6
i
Download from Www.Somanuals.com. All Manuals Search And Download.
PREFACE
Manual Objectives: This manual shows you how to set up and
use the RTD Panel Thermometer.
This meter is an economical Miniature Temperature Panel
Thermometer featuring a large display with a linearized analog
output that is supplied as a standard feature.
Each of the models* can be converted by the user to display in
degrees Fahrenheit or Celsius between one degree and a tenth
degree resolution.
RTD is an abbreviation for resistive temperature detector. The
meter is designed to measure 100 ohm platinum resistance with
0.00385 temperature coefficient, also known as the European
curve. The RTD will change resistance with a change in
temperature. When using a 2-wire measurement configuration,
lead wire impedance can contribute significant error to the
temperature measurement. When using long lead wires a 3- or
4-wire measurement configuration is recommended.
The meter is available in platinum RTD with 2-, 3-, or 4-wire
connections. All of the models listed in Table 1-1 come standard
with red LEDs, and can be ordered with a green LED display as
an option. The part numbers would then end with Ò-GR".
Also, these meters can be ordered with different power
configurations. Refer to Table 1-3 for available choices.
*Refer to Table 1-1 for the complete listing of models available.
ii
Download from Www.Somanuals.com. All Manuals Search And Download.
MODELS AVAILABLE
TABLE 1-1
RTD MODELS AVAILABLE
The following 3-1/2 digit mini RTD panel meters are
available and discussed in this OperatorÕs Manual.
MODEL
TYPE
¡C or ¡F
RESOLUTION
DP119-MF1
DP119-MF2
DP119-MC1
DP119-MC2
RTD
RTD
RTD
RTD
F
F
C
C
1.0¡F
0.1¡F
1.0¡C
0.1¡C
The accuracies and ranges are listed in Section 5.
Note
☞
TABLE 1-2
Power Options Available
MODEL
POWER
DP119-Mxx
115 Vac ±15%, 50/60 Hz
230 Vac ±15%, 50/60 Hz
9-26 Vdc @ 110mA max.
100 Vac ±15%, 50/60 Hz
24 Vac ±15%, 50/60 Hz
DP119-Mxx type-230
DP119-Mxx type-9/26
DP119-Mxx type-100
DP119-Mxx type-24
iii
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE 1-3
Other Models Available
The following 3-1/2 digit mini thermocouple panel
thermometers are discussed in a separate manual.
MODEL
TYPE
¡C or ¡F
RESOLUTION
DP119-JF1
DP119-JF2
DP119-JC1
DP119-JC2
J
J
J
J
F
F
C
C
1.0¡F
0.1¡F
1.0¡C
0.1¡C
DP119-KF1
DP119-KF2
DP119-KC1
DP119-KC2
K
K
K
K
F
F
C
C
1.0¡F
0.1¡F
1.0¡C
0.1¡C
DP119-TF1
DP119-TF2
DP119-TC1
DP119-TC2
T
T
T
T
F
F
C
C
1.0¡F
0.1¡F
1.0¡C
0.1¡C
DP119-EF1
DP119-EF2
DP119-EC1
DP119-EC2
E
E
E
E
F
F
C
C
1.0¡F
0.1¡F
1.0¡C
0.1¡C
iv
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by
these labels:
NOTE: provides you with information that is important
to successfully setup and use the Programmable
Digital Meter.
Note
☞
CAUTION or WARNING: tells you about the risk of
electric shock.
CAUTION, WARNING or IMPORTANT: tells you of
circumstances or practices that can effect the meter's
functionality and must refer to accompanying
documents.
v
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1 INTRODUCTION
1.1 UNPACKING
Remove the Packing List and verify that all equipment has been
received. If there are any questions about the shipment, use the
phone number for the Customer Service Department nearest
you.
Upon receipt of shipment, inspect the container and equipment
for any signs of damage. Take particular note of any evidence of
rough handling in transit. Immediately report any damage to the
shipping agent.
The carrier will not honor any claims unless all shipping
Note
☞
material is saved for their examination. After examining
and removing contents, save packing material and
carton in the event reshipment is necessary.
Verify that you received the following items in the shipping box:
QTY DESCRIPTION
1 Panel Thermometer with 3 small connectors plugged
into the rear of the meter.
1 OperatorÕs Manual
1
Download from Www.Somanuals.com. All Manuals Search And Download.
1.2 SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important to read this
manual before installing or commissioning this device as it contains important
information relating to Safety and EMC (Electromagnetic Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to
operate the unit if damage is found.
Note
☞
This instrument is a panel mount device protected in accordance with Class I of EN
61010 (115/230 AC power connections). Installation of this instrument should be done
by Qualified personnel. In order to ensure safe operation, the following instructions
should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in
the building installation as a disconnecting device. It shall be marked to indicate this function, and
it shall be in close proximity to the equipment within easy reach of the operator. The switch or
circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant
requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch
shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in
case of a fault in the equipment, an overcurrent protection device shall be installed.
• The Protective Conductor must be connected for safety reasons. Check that the power
cable has the proper Earth wire, and it is properly connected. It is not safe to operate
this unit without the Protective Conductor Terminal connected.
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
Note
☞
• Unit mounting should allow for adequate ventilation to ensure instrument does not
exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power
requirements. Install without exposing bare wire outside the connector to minimize
electrical shock hazards.
EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems
persist.
2
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 2 ABOUT THE METER
2.1 Front of the Meter
Figure 2-1 shows the panel thermometer.
Figure 2-1. Panel Thermometer
Features:
Display:3 1/2 Digit, 7-Segment Red or Green LED
Full-size:14.2 mm (0.56") LED Display
Analog Output Standard
1/8 DIN Standard Panel Cutout
Removable Screw-Clamp Cable Connector
Display Hold Capability
3
Download from Www.Somanuals.com. All Manuals Search And Download.
2.2 Back of the Meter
Figure 2-2 illustrates the rear of the meter.
1
2
3
1
2
3
1
2
3
4
P1
P2
P3
Figure 2-2. Rear View Showing P1, P2, P3 Locations
2.3 Connector Description
Connector Description
(AC) Earth Ground
(AC) Neutral
(AC) Line
(DC) -DC Return
Connector PIN #
P1
P1
P1
P1
P1
P1
P2
P2
P2
P3
P3
P3
P3
1
2
3
1
2
3
1
2
3
1
2
3
4
(DC) +DC
(DC) No Connection (Not used)
Display Hold (Active Low)
Analog Return
Analog Output
RTD Input -EXC
RTD Input Signal Lo
RTD Input Signal Hi
RTD Input +EXC
For RTD input wires, refer to Figure 3-3 thru 3-6.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 3 GETTING STARTED
3.1 Main Board Power Jumpers
Caution: The meter has no power-on switch, so it will
be in operation as soon you apply power.
The meter can be configured to operate on 115VAC or
230VAC by the proper combination of the soldered wire
jumpers that are located on the printed circuit board.
The meter is set at the factory to be powered by the
voltage specified at the time of ordering. The same
transformer is used for either configuration, so all you
need to do is to select the jumpers as described in this
section.
Important: These changes must be performed by a
qualified technician.
Note
☞
To change the Factory preset jumpers, do the following
steps:
Disconnect the power from the unit before proceeding.
1. Remove the main board from the case. Refer to
Disassembly/Assembly Section 3.8.
2. Locate the solder jumpers W1, W2, and W3 (located
near the edge of the main board alongside the
transformer).
3. If your power requirement is 115 V ac, solder jumpers
W1 and W3 should be wired, but jumper W2 should
not. If your power requirement is 230 Vac, solder
jumper W2 should be wired, but jumpers W1 and W3
should not.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Voltage Jumper Pin Settings
AC VOLTAGE
INSTALL
W1, W3
W2
REMOVE
W2
W1, W3
115VAC
230VAC
Figure 3-1 shows the location of solder jumpers W1 through W3.
S2
S4
ZERO
4
3
2
1
3
2
1
(COMPONENT SIDE)
P3
P2
S3
S5
TRANSFORMER
3
2
1
P1
W1 W2
W3
SPAN
Figure 3-1. Power Jumper Locations
6
Download from Www.Somanuals.com. All Manuals Search And Download.
3.2 Changing the Meter Parameters
The proper combinations of jumper plugs S1, S2, S3, S4, and S5
inside the meter allow you to change various parameters of the
meter. These parameters include resolution, range, units, RTD
sensor configuration and sensor break configuration. You can
change the parameters by moving the plugs on one or several of
the jumper locations. Keep in mind that the meter is factory set
for 3-wire RTD, and for the upscale position.
Figure 3-2 shows the location of the jumper plugs S1 through S5.
ZERO
SPAN
UNUSED PLUG POSITION
A
C
S1
B
S1
A
C
G
D
B
E
F
S2
S3
S4
A
S5
I
H
B
A
A
C
S2
S4
ZERO
4
3
2
1
3
2
1
(COMPONENT SIDE)
P3
P2
S3
S5
TRANSFORMER
3
2
1
P1
W1 W2
W3
SPAN
Figure 3-2. Jumper Plugs S1 through S5
7
Download from Www.Somanuals.com. All Manuals Search And Download.
The following tables summarizes what parameters you can
change.
TABLE 3-1 Range Parameters
Option
*-MF1
*-MF2
*-MC1
*-MC2
F/C Resolution
Range
S1 S4 S5
¡F
¡F
¡C
¡C
1¡
-328 to + 1562¡F
-199.9 to + 199.9¡F
-200 to + 850¡C
-199.9 to + 199.9¡C
--
C
--
A
--
A
--
A
C
B
B
0.1¡
1¡
0.1¡
C
* Refer to Table 1-1, for complete models (preface section).
TABLE 3-2 2-, 3-, 4-Wire, Upscale, Downscale Configuration
Configuration
S2
S3
S3
Upscale (Default)
A, B, E, F
Downscale
C, D, G, H
C, D, G, H
2- or 4-wire RTD
--
--
3-Wire RTD
A
I
A, B, E, F
Note
☞
3-Wire input connection is standard. For 2- or 4-Wire
remove S2-A and S3-I
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 3-3 gives an example of how to set S4 and S5 to configure
a MF1 meter option. There is a jumper plug across A on S4 and
another jumper plug across A on S5. B and C on S5 are not used.
S4
A
S5
B
A C
JUMPER PLUG
COVERING BOTH PINS
JUMPER PLUG
COVERING
TWO PINS
OVER "A"
Figure 3-3. Example of How to Set Meter Option MF1
(jumpers S4 and S5)
SENSOR BREAK PROTECTION:
The proper combination of jumpers on S3 will determine the
direction in which the display and analog output will go if the
RTD sensor breaks (upscale or downscale). The meter is factory
set to upscale break protection. If the sensor breaks, the display
and analog signal will increment (upscale) and the 3 least
significant digits will then go blank. If you require downscale
break protection, refer to Table 3-2 to change the S3 jumper
configuration.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Section 3.3 Installation and Panel Mounting
Figure 3-4 shows the panel cutout dimensions, the dimensions
for the panel thickness, and the backplate for mounting the unit
in a panel. Connections will be done after mounting the unit.
1. Remove the 2 phillips screws from behind the display (you
don't have to go inside the meter) holding the rectangular
backplate to the meter. Remove the backplate and set aside.
2. Cut or punch a hole in the panel where you want the meter
to go. The panel can be as thick as 0.25" (6.4mm) to as thin
as 0.03" (0.8mm).
3. Insert the meter into the panel cutout.
4. From the rear of the panel, slide the backplate over the case
(smooth side out).
5. Install the 2 phillips screws to secure the meter in 1/8 DIN
mount. Center the meter in hole prior to tightening screws.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
PANEL THICKNESS
0.25 [6.4] MAX
0.03 [0.8] MIN
1/8 DIN PANEL CUTOUT
0.06
[1.5]
4PLCS
R
1.772 +0.024/-.000
[45.0 +0.61/-0.00]
3.622 +0.032/-.000
[92.00 +0.81/-0.00]
SCREWS
METER
BEZEL
LENS
BACKPLATE
PANEL WITH
1/8 DIN CUTOUT
Figure 3-4. Installation Information
11
Download from Www.Somanuals.com. All Manuals Search And Download.
3.4 Sensor Input Connections
The RTD sensor (see Figure 3-5), can be wired in three different
ways. They are 2-, 3-, and 4-wire configurations. The meter is
factory set internally to a 3-wire setup. If you wish to change it,
refer to Table 3-2.
2-Wire RTD:
Externally wire the P3 connector according to Figure 3-6. Keep
in mind that you will need to jumper pins 1 and 2 on P3 together,
and jumper pins 3 and 4 on P3 together. Use 24 AWG wire. The
black lead goes to Pin 1 or 2 and the red lead goes to Pin 3 or 4.
3-Wire RTD:
Externally wire the P3 connector according to Figure 3-7. Pin 4
on the P3 connector is not used.
4-Wire RTD:
Externally wire the P3 connector according to Figure 3-8.
1
2
3
4
Screws
underneath
Figure 3-5. RTD P3 Connector
See Figure 2-2 for the location of the P3 connector.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
3.4 Sensor Input Connections (Continued)
P3
RED*
+ Excitation 4
Signal Hi
Signal Lo
- Excitation 1
3
2
METER
RTD
BLACK*
Figure 3-6. 2-Wire RTD Wiring
P3
NOT USED
+ Excitation 4
RED*
BLACK*
BLACK*
Signal Hi
Signal Lo
3
2
METER
RTD
- Excitation 1
Figure 3-7. 3-Wire RTD Wiring
P3
+ Excitation 4
RED* (+Excitation)
RED* (Signal Hi)
BLACK* (Signal Lo)
BLACK* (-Excitation)
Signal Hi
Signal Lo
3
2
METER
RTD
- Excitation 1
Figure 3-8. 4-Wire RTD Wiring
Note
☞
Refer to Table 3-1 and Table 3-2 for internal jumper
settings.
* Wire colors are for OMEGA RTD wiring only.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
3.5 Analog Output Connections
The analog output is a linearized millivolt signal that is equivalent
to the displayed temperature. It comes as a standard feature
with every unit. The resolution of the analog output is no greater
than the resolution of the display.
Note
☞
The analog output is a non-isolated signal.
To access the analog signal, attach copper wire to the 3-pin
female connector that is supplied with the meter. Refer to
Figure 3-9 for wiring instructions. Plug the female connector into
the P2 (male) connector at the rear of the meter.
Figure 2-2 shows the location of the P2 connector on the meter.
Note that this connector shares a common return with the
ÒDisplay Hold" function.
+
–
Output
mV
RETURN
1
2
3
Screws
underneath
Figure 3-9. Analog Output Connections (P2)
14
Download from Www.Somanuals.com. All Manuals Search And Download.
3.6 Display Hold Connections
The ÒDisplay Hold" function can be used to freeze the display.
Connect a single pole, single throw (SPST) switch between pins
1 and 2 on the P2 connector. The display will stop updating when
this switch is on. When this switch is off, the display will update
itself 2.5 times per second.
To access the ÒDisplay Hold" function, attach copper wire to the
3-pin female connector that is supplied with the meter. Refer to
Figure 3-10 for wiring instructions. This connector shares a
common return with the Analog Output. Plug the female
connector into the P2 connector at the rear of the meter.
Figure 2-2 shows the location of the P2 connector on the meter.
Switch
Display Hold
RETURN
P2
1
2
3
Screws
underneath
Figure 3-10. Display Hold Connections
15
Download from Www.Somanuals.com. All Manuals Search And Download.
3.7 Main Power Connections
Connect the ac main power connections as shown below.
WARNING: Do not connect ac power to your meter until
you have completed all input and output connections.
Failure to do so may result in injury! This device must only
be installed electrically by specially trained electrician with
corresponding qualifications. The main power input to the
unit as well as the AC input signal to be measured must
agree with the wiring instruction.
The meter is factory set to the power specified by the customer
at the time of ordering. The voltage is printed on the Label under
Input power.
Refer to Section 3-2 to convert from one
configuration to another
Note
☞
To connect power to the unit, attach wires from the ~AC power
cord to the orange 3-pin female connector that is supplied with
the meter. Refer to Figure 3-11 for wiring instructions. Plug the
female connector into the orange P1 (male) connector at the
rear of the meter.
Figure 2Ð2 shows the location of P1 on the meter.
EARTH
NEUTRAL
LINE
GREEN WIRE
©
N
L
P1
1
FUSE
ORANGE
CONNECTOR
2
SWITCH
3
Check for proper Earth grounding in the
power distribution system (single phase).
Screws
underneath
Figure 3-11. ~AC Connector Wiring (P1)
16
Download from Www.Somanuals.com. All Manuals Search And Download.
DC WIRING
The meter can be configured at the factory to operate on a 9Ð26
VDC non-isolated power. The DC voltage option is specified at
the time of ordering. See Table 1-3 for model number listing. An
AC powered meter cannot be converted to a DC powered meter
or vice versa.
To connect power to the unit, attach the DC power wires to the
orange 3-pin female connector that is supplied with the meter.
Refer to Figure 3-12 for wiring instructions. Plug the female
connector into the orange P1 (male) connector at the rear of the
meter.
Figure 2-2 shows the location of P1 on the meter.
No Connection N/C
+ DC
External
Power Supply
– DC
P1
1
ORANGE
CONNECTOR
2
3
Screws
underneath
Figure 3-12.
=DC Power Connector Wiring (P1)
17
Download from Www.Somanuals.com. All Manuals Search And Download.
3.8 Disassembly/Assembly
You must go through the following steps in order to
Note
☞
remove the printed circuit board from the case if you want
to change from 115VAC to 230VAC (or vice versa) or to
change from Fahrenheit to Celsius (or vice versa). These
changes must be performed by a qualified technician in
order to avoid damage to the unit.
WARNING: Do not connect ac power to your meter until
you have completed all input and output connections.
Failure to do so may result in injury!
1. Remove power and all wire connections from the unit.
2. Unplug the three connectors (P1, P2, and P3) from the
back of the meter. Grasp them firmly on the top and
bottom of the connector and pull. Figure 3-13 shows the
location of the three connectors.
3. Remove the friction fit lens by prying down and forward
from the top of the meter. Remove the mounting screws
from the rear of the panel backplate. Slide the unit out of
the panel (if it is installed in a panel).
4. The printed circuit board is held in place by a tab and
notch system located next to the P1 connector, pin 1
(refer to Figure 3-13). The tab is on the case and the
notch is on the edge of the PC Board. Hold the meter with
the display facing away from you. Gently pry the tab (on
the back corner of the case) out with the edge of your
finger, while pushing the back of the PC Board out
through the front of the meter. The PC Board will slide out
the front, free from the case.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
PULL TAB HERE
PC BOARD
PUSH HERE
1
2
3
1
2
3
1
2
3
4
P1
P2
P3
Figure 3-13. Removing the PC Board from the Case
When re-installing the printed circuit board:
1. Hold the case (with the two mounting holes facing upwards
and to the right side of the case) in the left hand.
2. Hold the printed circuit board with the components showing
(soldered side down) in the right hand.
3. Slip the component board into the case and slide in as far
as it will go. You will also have to release the tab at the rear
on the other side of the case (turn case over) to allow the
board to fit all the way inside the case.
4. Go through steps 3 thru 5 in Section 3.3 in order to
re-install the backplate and screws, and also install the lens.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4 OPERATION AND CALIBRATION
Important: These changes must be performed by a
qualified technician.
WARNING: Do not connect ac power to your meter until
you have completed all input and output connections.
Failure to do so may result in injury!
1. Wire the meter according to the wiring section. This
includes power, thermocouple and analog signal
connections.
2. Apply power to the meter and the display will show the
temperature of the probe.
The meter is shipped from the factory calibrated to the accuracy
as specified. No re-calibration is necessary. In the event that
there is a desire to check the calibration, the following procedure
is given.
4.1 EQUIPMENT REQUIRED
¥ Meter
¥ a small screwdriver
¥ a precision resistance source in 0.01¡
resolution or ÒDecade Box".
20
Download from Www.Somanuals.com. All Manuals Search And Download.
4.2 CALIBRATION PROCEDURE
1. Remove the front lens. Connect power to the meter,
observing proper wiring procedures.
2. Connect the precision resistance source to the P3
connector observing proper wiring configuration.
For 3Ðwire connections, use pins 1, 2, 3, and 4. Refer
to Figures 4-1 through 4-3.
3. Apply resistance input as indicated in Table 4-1.
4. Adjust either the ZERO or SPAN potentiometer as
shown in Figure 4-4.
5. After adjusting the post, go back and verify both
temperature readings. Repeat step 4 as required.
Note
☞
Refer to Table 3-1 and Table 3-2 for internal jumper
settings.
P3
RED
4
3
2
1
+ Excitation
Signal Hi
Signal Lo
- Excitation
PRECISION
RESISTANCE
SOURCE
RED
BLACK
BLACK
JUMPER WIRE
(24 AWG)
Figure 4-1. 2-Wire Configuration
P3
NOT USED
4
3
2
1
+ Excitation
Signal Hi
Signal Lo
- Excitation
PRECISION
RESISTANCE
SOURCE
RED
RED
BLACK
BLACK
BLACK
Figure 4-2. 3-Wire Configuration
21
Download from Www.Somanuals.com. All Manuals Search And Download.
P3
RED
RED
BLACK
BLACK
+ Excitation
Signal Hi
Signal Lo
- Excitation
4
3
2
1
PRECISION
RESISTANCE
SOURCE
RED
BLACK
Figure 4-3. 4-Wire Configuration
TABLE 4-1
CALIBRATION VALUES WHEN YOU ARE USING
RESISTANCE SOURCE
MODEL RESISTANCE
OPTION READING
METER
DISPLAY
POT TO
ADJUST
*-MF1
*-MF2
*-MC1
*-MC2
93.01
00¡F
ZERO
SPAN
375.51
1472¡F
93.01
134.70
00.0¡F
194.0¡F
ZERO
SPAN
100
375.51
00¡C
800¡C
ZERO
SPAN
100.0¡C
174.0¡C
00¡C
195.0¡C
ZERO
SPAN
* Refer to Table 1-1, for complete models (preface section).
ZERO
SPAN
Figure 4-4. Calibration Potentiometers
22
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5 SPECIFICATIONS
INPUT TYPE:
RTD:....................................................100-ohm platinum with alpha of 0.00385
Calibration:.............................................................................IEC 751 (IPTS-68)
Configuration:...................................................2-, 3-, or 4-wire RTD connection
Polarity: ....................................................................................................Bipolar
Zero: ...........................................................................Adjustable ±20¡C (±36¡F)
Zero Tempco :...................................................................±0/05 degree/degree
Sensor-wire Resistance Tempco Effect per Conductor:
2-wire: ...........................................................20 mdeg/½/degree, up to 3.8 ½
3-wire: ............................................................50µdeg/½/degree, up to 100 ½
4-wire: ...........................................................5.0µdeg/½/degree, up to 100 ½
Sensor-excitation Current: .......................................................................0.5 mA
Sensor-break Indication:.............................................Meter displays overrange
(three least significant digits blank)
Overvoltage Protection: .............................................................................25 Vp
NOISE REJECTION
NMR , SIG HI to SIG LO:.........................................................70 dB, 50/60 Hz
CMR , Analog RTN to PWR GND:....................................120 dB, DC to 60 Hz
CMV , Analog RTN to PWR GND:........................Dielectric strength to 1500 V
transient per 260 Volt rms or DC working voltage.
ACCURACY at 25¡C
Range for rated accuracy:
MC1/MF1:...................................................-200¡ to +850¡C (-328¡ to +1562¡F)
Maximum Error: ......................................................±1.5¡C (±2.7¡F) ±1 count
MC2/MF2:.........................................-199.9¡ to +199.9¡C (-199.9¡ to +199.9¡F)
Maximum Error: ......................................................±0.5¡C (±0.5¡F) ±1 count
Resolution: ....................................0.1¡C, 0.1¡F, 1¡C or 1¡F (jumper-selectable)
Span Tempco : .................................................................±0.02% of reading/¡C
Full-Scale Step Response:.................................................................. 1 second
Warmup to Rated Accuracy:............................................................. 30 minutes
ANALOG OUTPUT (linearized)
Voltage: ..............................................................................................1mV/count
Calibration Error:
MC1/MF1:............................ ±1% of reading ± 1.5 mV on ¡C, ±2.7mV on ¡F
MC2/MF2: .............................................................. ±1% of reading ± 0.5 mV
Source Resistance: ...................................................................................100 ½
ANALOG TO DIGITAL CONVERSION
Technique: .................................................................Dual-slope, average value
Signal Integration Period:........................................................................100 ms
Read Rate: ..................................................................................................2.5/s
23
Download from Www.Somanuals.com. All Manuals Search And Download.
DISPLAY
Type:..........................................................................................7-segment, LED
Height: .......................................................................................0.56" (14.2 mm)
Symbols:..................................................................................................-1.8.8.8
Overrange Indication:....................................Three least-significant digits blank
Colors: .........................................................................................Red - standard
Green - optional
DIGITAL INPUTS
Hold:......................................................................TTL or 5V CMOS compatible
ENVIRONMENTAL
Operating Temperature: ...............................................0¡ to 60¡C(32¡ to 140¡F)
Storage Temperature:.............................................-40¡ to 85¡C (-40¡ to 185¡F)
Relative Humidity:..............................................95% at 40¡C (non-condensing)
MECHANICAL DIMENSIONS
Bezel:.................................................................... 01.89Ó x 3.78" (48 x 96 mm)
Depth Behind the Bezel:.................................... 4.32" (110 mm) w/ connectors
Panel Cutout: ................................. 1/8 DIN: 01.72" H x 3.62" W (45 x 92 mm)
Weight:............................................................................ 8 ounces (227 grams)
Panel Thickness ........................................................minimum: 0.03" (0.76mm)
maximum: 0.25" (6.25mm)
Case Material: ......................................................94V-1 UL-rated thermoplastic
POWER
AC Frequency:......................................................................................50/60 Hz
Max. Power: .................................................................................2.4 watts max.
AC Voltage: ................................................................115Vac ±15% 25mA max.
230Vac ±15% 14mA max.
100Vac ±15% 30mA max.
24Vac ±15% 127mA max.
DC Voltage:..........................................................................9-26 Vdc @ 110mA
External Fuse Protection needed:...................................32mA, 115Vac/230Vac
UL Slow-blow (per single unit)
40mA, 100Vac
160mA, 24Vac
125mA, 9-26Vdc
Installation Category: .................................................................. I per EN61010
Equipment Class:.........................................................................I per EN61010
Pollution Degree: ........................................................................2 per EN61010
Refer to the Glossary in Section 6.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
1.89
[48.0]
3.78
[96.0]
BEZEL
0.76 [19.3]
BACKPLATE
3.76
[95.5]
CASE
4.32
[109.7]
MAX
SIDE VIEW
TOP VIEW
PANEL THICKNESS
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL CUTOUT
0.06
[1.5]
4PLCS
R
1.772 +0.024/-.000
[45.01 +0.61/-0.00]
3.622 +0.032/-.000
[92.00 +0.81/-0.00]
1/8 DIN
Figure 5-1 Dimensions
25
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6 GLOSSARY
COMMON MODE REJECTION (CMR) - a measure of the affect
of a voltage on the indicated display. It is the ratio of the common
mode voltage to the measured voltage due to a common mode
voltage.
COMMON MODE VOLTAGE (CMV) - the average of the voltage
applied to both wires of a two wire or differential input.
NORMAL MODE REJECTION (NMR) - a measure of the
rejection of unwanted signal due to a voltage applied between
the wires of a two wire or differential input (shown as normal
mode voltage). NMR is the ratio of the normal mode voltage to
the measured voltage due to the normal mode voltage.
TEMPCO - abbreviation for temperature coefficient. It is the
effect of changes in ambient temperature on a particular
parameter, such as zero or span settings.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
27
Download from Www.Somanuals.com. All Manuals Search And Download.
MADE
IN
USA
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace
period to the normal one (1) year product warranty to cover handling and shipping time. This
ensures that OMEGAs customers receive maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGAs Customer
Service Department will issue an Authorized Return (AR) number immediately upon phone or written
request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced
at no charge. OMEGAs WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or
current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGAs control. Components which wear are not warranted, including but not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use of its products in accordance with information provided by
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be
as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a Basic
Component under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
FOR NON-WARRANTY REPAIRS, consult OMEGA
for current repair charges. Have the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGAs policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
' Copyright 1998 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior
written consent of OMEGA ENGINEERING, INC.
Download from Www.Somanuals.com. All Manuals Search And Download.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGAÉOf Course!
TEMPERATURE
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Transducers & Strain Gauges
Load Cells & Pressure Gauges
Displacement Transducers
Instrumentation & Accessories
FLOW/LEVEL
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
pH/CONDUCTIVITY
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
HEATERS
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
11434ML-98 Rev. A
M1488/1298
Download from Www.Somanuals.com. All Manuals Search And Download.
|