Omega Thermometer CL24 User Manual

User’s Guide  
®
CL23A, 24, 25, 26, 27  
Digital Calibrator/Thermometer  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
General Information......................................................................... 2  
Common Specifications................................................................... 2  
Unique Specifications...................................................................... 5  
Features........................................................................................... 6  
Manual Addenda ............................................................................. 6  
Unpacking Your Unit........................................................................ 7  
Safety Information ........................................................................... 7  
Safety Precautions .......................................................................... 7  
Battery Installation/Replacement..................................................... 9  
Memory Backup.............................................................................. 10  
Operation with Rechargeable Battery............................................. 10  
Operating Instructions..................................................................... 11  
11. Diagnostics and Error Messages...................................... 11  
12. Initial Turn On ................................................................... 11  
3. Meter Mode Operation (Thermocouple) ........................... 13  
14. Meter Mode Operation (RTD, Thermistor, Ohms)............ 14  
15. Data Logging & Recall...................................................... 15  
16. Calibrator Mode Operation (Thermocouples)................... 17  
17. Calibrator Mode Operation (RTD, Thermistor, Ohms) ..... 18  
18. Storage and Recall of Calibrator Settings ........................ 20  
9. Step Function.................................................................... 22  
10. Ramp Function ................................................................. 22  
Service Information......................................................................... 23  
11. Calibration Procedures ..................................................... 23  
12. Calibration Verification...................................................... 32  
13. Mechanical Parts Diagram................................................ 38  
Warranty ......................................................................................... 45  
Limitation of Warranty..................................................................... 45  
Statement of Calibration................................................................. 45  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
GENERAL INFORMATION  
This manual provides operating instructions and maintenance information for  
five temperature instruments. These instruments are high performance cali-  
brator-thermometers capable of simulating and measuring a wide-variety of  
sensors. In addition, features such as data storage and ramping further  
enhance their versatility.  
It is recommended that you read this manual thoroughly, especially the sec-  
tions on safety, prior to operating these instruments.  
COMMON SPECIFICATIONS  
THERMOMETER INPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS,  
THERMISTOR; T-series instrumentation connector.  
CALIBRATOR OUTPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS,  
THERMISTOR; T-series instrumentation connector.  
THERMOCOUPLE TYPES: K, J, T, E, N, B, R, S, G, C, D.  
RTD TYPES: 1100platinum (= 0.00385)  
1000platinum (= 0.00385)  
THERMISTOR TYPE: 2252(YSI Series 400).  
RANGES:  
RTK; –200°C to 1372°C, –328°F to 2502°F  
RTJ; –210°C to 760°C, –346°F to 1400°F  
RTT; –200°C to 400°C, –328°F to 752°F  
RTE; –230°C to 1000°C, –382°F to 1832°F  
RTN; –200°C to 1300°C, –328°F to 2372°F  
RTB; 500°C to 1820°C, 932°F to 3308°F  
RTR; 0°C to 1768°C, 32°F to 3214°F  
RTS; 0°C to 1768°C, 32°F to 3214°F  
RTG; 300°C to 2316°C, 572°F to 4201°F  
RTC; 0°C to 2316°C, 32°F to 4201°F  
RTD; 0°C to 2316°C, 32°F to 4201°F  
RTD; –200°C to 850°C, –328°F to 1562°F  
THERMISTOR; –40°C to 150°C, –40°F to 302°F  
OHMS; 0-999.99Ω  
OHMS; 0-9,999.9Ω  
OHMS; 0-99,999Ω  
ACCURACY (18°C to 28°C ambient, 2 years)*:  
K, J, T, E, N; 0.5°F (rdg –50°F)  
K, J, T, E, N; 1.0°F (rdg < –50°F)  
B, R, S, G, C, D; 1.7°F  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
RTD; 0.2°F (rdg – 50°F)  
RTD; 0.5°F (rdg < – 50°F)  
RTD; 0.2°F 0.04% rdg (meter mode)  
THERMISTOR; 0.5F  
OHMS; 0.02% range  
* Exclusive of sensor errors, and lead resistance induced errors.  
RESOLUTION: TEMPERATURE; 0.1°/1° F/C  
RESOLUTION: OHMS; 0.01/0.1/1Ω  
REPEATABILITY (1 week at constant ambient temperature):  
K, J, T, E, N; 0.2°F typ.  
B, R, S, G, C, D; 1.0°F typ.  
RTD, THERMISTOR; 0.1°F typ.  
OHMS; 0.01% rng typ.  
TEMPERATURE COEFFICIENT:  
From 18°C to 28°C, included in accuracy specifications.  
From 0°C to 18°C, and 28°C to 50°C:  
K, J, T, E, N; (0.002% rdg + 0.03°C)/°C  
B, R, S, G, C, D; (0.002% rdg + 0.1°C)/°C  
RTD, THERMISTOR, OHMS; (0.002% rdg + 0.004% rng)/°C  
METER-MODE INPUT-CURRENT (THERMOCOUPLES): 50nA typ.  
METER-MODE READING-RATE: 1 reading/second typ.  
METER-MODE EXCITATION-CURRENTS:  
100RTD; <2mA  
1000RTD; <200µA  
999.99; <2mA  
9999.9; <200µA  
99999; <50µA  
THERMISTOR; T 113°F <50µA  
T> 113°F, <2mA  
CALIBRATOR MODE OUTPUT LOAD (THERMOCOUPLES): 500min.  
CALIBRATOR-MODE EXCITATION-CURRENTS:  
100RTD; 0.5 – 1mA (0.1 – 2mA typ.)  
1000RTD; 50 – 100µA (10-200µA typ.)  
999.99; Same as 100RTD  
9999.9; Same as 1000RTD  
99999; 10 – 20µA (1 – 20µA typ.)  
THERMISTOR; T 113°F, 10 – 20µA  
T> 113°F, 0.5 – 1mA  
MAXIMUM COMMON MODE VOLTAGE: 42V peak to earth.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
KEYPAD: 12 momentary switches with tactile feedback select;  
• On/Off  
• Change/Enter  
• Clear/–0  
• Sensor Select/1  
• °F/°C/2  
• Resolution/3  
• Operate/4  
• Store/5  
• Recall/6  
• Calibrator/Meter/Up Ramp/7  
• Calibrator Step/Down Ramp/8  
• Calibrator Ramp/Meter Hold/9  
DISPLAY: 5 digit LCD, 0.4" height, and decimal point. Annunciators;  
• Loaded Memory Locations (10 for data storage, 4 for ramp parameters)  
• Low Battery (BAT)  
• Number Change Mode (NUM)  
• Calibrator Mode (CALIB)  
• Meter Mode (METER)  
• Operate Mode (OPR)  
• Store Mode (STO)  
• Recall Mode (RCL)  
• Meter Hold (HOLD)  
• Scale (°F/°C)  
• Sensor Type  
POWER OFF CONFIGURATION RETENTION: Instrument retains last selected;  
• Sensor Type  
• °F/°C Scale  
• 0.1°/1° Resolution  
• Calibrator/Meter Mode  
• Memory Contents (data storage and ramp parameters).  
DIAGNOSTICS: Display codes indicate following conditions;  
• 'BAT': Low Battery  
• 'OPEN': Open Input Thermocouple, RTD or Thermistor.  
• 'E-1' (momentary): Invalid Keypad Entry  
• 'E-2' (momentary): Ramp Function Not Installed  
• 'HI': Temperature Reading or Keypad Entry Exceeds Rating  
• 'LO': Temperature Reading or Keypad Entry Too Low For Accuracy  
• LCD Test: During power-up, all segments and annunciators enabled momentarily  
ENVIRONMENTAL LIMITS FOR OPERATING: 0°C to 50°C, less than 90% relative  
humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 50°C.  
ENVIRONMENTAL LIMITS FOR STORAGE: –35°C to 60°C, less than 95% relative  
humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 60°C.  
POWER: 9 volt alkaline battery (NEDA 1604A).  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd  
(rechargeable).  
BATTERY INDICATOR: Display indicates BAT when less than 10% of life remains.  
DIMENSIONS, WEIGHT: 7.0" × 2.9 × 1.1". Net weight 12 oz.  
UNIQUE SPECIFICATIONS  
NOTE: The specifications below are specific to the model(s) as identified by model  
number. The difference specifications are in addition to the OVERALL SERIES SPEC-  
IFICATIONS found on pages 2-5.  
MODEL CL23A/CL23EC:  
SENSOR SELECTION: K, J, T/K, J, T, L Thermocouples  
ACCURACY: (18°C to 28°C ambient, 2 years): 0.5°F (rdg –50°F).  
0.04% rdg (rdg > 1250°F)  
1.0°F (rdg < –50°F)  
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications.  
From 0°C to 18°C, and 28°C to 50°C; less than (0.006% rdg +0.03°C)/°C.  
RAMP FUNCTION: Not Included.  
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd  
(rechargeable).  
MODEL CL24:  
SENSOR SELECTION: K, J, T, E Thermocouples  
ACCURACY: (18°C to 28°C ambient, 2 years): 0.5°F (rdg –50°F).  
0.04% rdg (rdg > 1250°F)  
1.0°F (rdg < –50°F)  
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications.  
From 0°C to 18°C, and 28°C to 50°C; less than (0.006% rdg +0.03°C)/°C.  
RAMP FUNCTION: Not Included.  
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd  
(rechargeable).  
MODEL CL25/CL25EC:  
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C,  
D Thermocouples  
RAMP FUNCTION: Included.  
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd  
(rechargeable)  
MODEL CL26:  
SENSOR SELECTION: K, J, T, E Thermocouples  
100Platinum RTD  
2252Thermistor  
1,000/100,000Resistance  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
ACCURACY: (18°C to 28°C ambient, 2 years): 0.5 °F (rdg –50°F).  
0.04% rdg (rdg > 1250°F)  
1.0°F (rdg < –50°F)  
OHMS; 0.04% range  
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications.  
From 0°C to 18°C, and 28°C to 50°C.  
K, J, T, E; (0.006% rdg + 0.03°C)/°C  
RTD, THERMISTOR, OHMS; (0.006% rdg +0.012% rng)/°C  
RAMP FUNCTION: Not Included.  
MODEL CL27/CL27EC:  
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C,  
D Thermocouples  
100/1000Platinum RTD  
1,000/10,000Resistance  
RAMP FUNCTION: Included.  
FEATURES  
• Combination calibrator, thermometer and datalogger.  
• Function selection and numeric data entered via 12 keyswitch color coded  
keypad.  
• Keyswitches have tactile feedback.  
• 5 digit LCD includes annunciators for operating modes.  
• Up to 11 thermocouple types, 2 RTD types, and thermistor.  
• 1° or 0.1° resolution over temperature range of each sensor.  
• °F and °C scales.  
• Reading hold mode.  
• 10 memory locations for saving meter readings and/or calibrator outputs.  
• Step mode for quick sequential recall of calibrator outputs from memory.  
• Manual/automatic ramping for testing controller setpoints.  
• Conforms to ITS-90 thermocouple tables.  
• Dust proof, splash proof, drop proof construction.  
• Built-in tilt stand/hanger for bench use or hands free field measurements.  
• User friendly programming.  
• Retains data and programming, even when turned off.  
• Performs diagnostic tests and indicates fault conditions.  
• Low battery and open sensor indications.  
MANUAL ADDENDA  
Improvements or changes to this manual will be explained on an addendum  
included with the instrument. All change information should be incorporated  
immediately into the appropriate places in the manual.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
UNPACKING YOUR UNIT  
Remove the Packing List and verify that you have received all equipment. If  
you have any questions about the shipment, please call the OMEGA  
Customer Service Department at 1-800-622-2378 or (203) 359-1660.  
When you receive the shipment, inspect the container and equipment for any  
signs of damage. Note any evidence of rough handling in transit. Immediately  
report any damage to the shipping agent.  
NOTE: The carrier will not honor any claims unless all shipping  
material is saved for their examination. After examining and  
removing contents, save packing material and carton in the event  
reshipment is necessary.  
Make sure that you have the following items in the packing box.  
• 1 CL20 Series Calibrator and Holster.  
• 1 Operator’s Manual  
• 1 NIST Certificate  
• 1 Accessories Kit  
SAFETY INFORMATION  
SAFETY SYMBOLS AND TERMS  
!
The symbol L on the instrument denotes that the user should refer to the  
operating instructions.  
The WARNING used in this manual explains dangers that could result in  
personal injury or death.  
The CAUTION used in this manual explains hazards that could damage  
the instrument.  
SAFETY PRECAUTIONS  
WARNING  
These instruments are intended for use by qualified person-  
nel trained in the safe operation of electronic testing equip-  
ment. Read the instruction manual thoroughly before using,  
to become familiar with the instrument's operations and  
capabilities.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNING  
Do not touch a temperature probe sheath when measuring  
excessively high or low temperatures, or toxic substances.  
WARNING  
Do not attempt to measure temperatures beyond the range of  
the probe being used. Probe damage or personal injury  
could result from exceeding a probe's maximum temperature  
rating.  
WARNING  
The American National Safety Institute (ANSI) states that a  
shock hazard exists when probes or sensors are exposed to  
voltage levels greater than 42VDC or 42V peak AC. Do not  
use this instrument where voltages at the measurement sur-  
face exceed these levels.  
WARNING  
Do not substitute a metal part for the nylon screw in the rear  
case. Doing so will degrade electrical isolation of the case.  
WARNING  
The battery is accessible through a cover on the back of the  
instrument. To avoid electrical shock hazard, disconnect all  
temperature probes and sensors and turn the unit off before  
removing the cover.  
WARNING  
Never use this instrument or any probe or sensor inside a  
microwave oven.  
CAUTION  
Avoid making sharp bends in probe or sensor lead wires.  
Bending lead wires at a sharp angle can damage the wire  
causing probe failure.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAUTION  
Keep inputs electrically isolated from outputs. Do not con-  
nect an external voltage source between the two connectors,  
and avoid any external electrical pathways between input  
and output (T2 & T1).  
Failure to do so will disturb meter readings and calibrator  
output levels. Damage to the instrument is also possible.  
Where the meter and calibrator connectors are simultane-  
ously wired into a system lacking the above isolation, it is  
necessary to use an "ungrounded" (i.e. electrically isolated)  
probe at the thermometer input.  
BATTERY INSTALLATION/REPLACEMENT  
A 9V alkaline battery is supplied with the instrument but not installed. Read  
the following installation instructions before attempting to install or remove  
the battery. Replacement batteries should also be alkaline types.  
WARNING  
Turn the unit off and disconnect any input/output connec-  
tions before replacing the battery. Put the cover back into  
place on the battery compartment before resuming use of  
the instrument.  
1. Remove the cover from the battery compartment by sliding it off in the  
direction of the arrow located on the battery cover.  
Figure 1. Battery Installation  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Remove the old battery by lifting the battery extractor loop.  
3. Place the new battery in the battery compartment. Be sure to observe  
proper polarity.  
4. Re-install the battery cover before resuming use of the instrument.  
NOTES:  
• Less than 10% of battery life remains when the BAT annuncia-  
tor turns on.  
• When the battery is excessively discharged, the display is  
blanked.  
• If the instrument is going to be stored for a long period of time  
or in a high temperature environment, remove the battery to  
prevent leakage damage.  
After a new battery is installed, allow approximately 5 sec-  
onds for display turn-on the first time the unit is turned on.  
During this period, microcomputer initialization is per-  
formed.  
MEMORY BACKUP  
During battery replacement, the contents of user-programmed memory  
(data, operating modes, etc.) can be saved. Prior to removing the old battery,  
turn off the instrument, and connect an HH22-AC battery charger. Then  
exchange batteries, and disconnect the battery charger. Do not leave the bat-  
tery charger connected to instruments with non-rechargeable batteries.  
OPERATION WITH RECHARGEABLE  
BATTERY  
HH22-AC provides a 9-volt Ni-Cd battery and recharger suitable for use with  
the unit. This battery provides 3-4 hours of continuous operation. This dura-  
tion can be extended to several days by operating simultaneously from both  
the battery and recharger.  
Turn off the instrument to recharge the battery. Allow 7 hours to fully  
recharge the battery. Then unplug the recharger to avoid overcharging.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
1. DIAGNOSTICS AND ERROR MESSAGES  
Unit self-diagnostics provide fault condition readouts which are described  
below:  
PROBLEM:  
LIKELY CAUSE:  
Blank display, unit does not  
power-up.  
(1) Improper battery installation.  
(1) Check battery polarity.  
(2) Dead battery.  
BAT Annunciator.  
(1) Low battery voltage, install a new  
battery. If problem persists, con-  
sult factory.  
Display reads OPEN.  
(1) No thermocouple, RTD or thermis-  
tor is plugged into the meter input.  
Display reads E1 momentarily.  
(1) This indicates that an invalid key-  
pad entry has been made. Review  
keystroke sequence, or consult  
manual for input instructions.  
Display reads E2 momentarily.  
Display reads HI.  
(1) Attempt made to activate non-  
installed ramp function. NOTE:  
RAMP Function enabled on  
Models CL25 and CL27 only.  
(1) Meter-mode input-temperature  
exceeds rating of selected sensor.  
(2) Calibrator-mode  
keypad-entry  
exceeds rating of sensor type.  
Display reads LO.  
(1) Meter-mode input-temperature too  
low for accurate measurement.  
(2) Calibrator-mode keypad-entry too  
low for accurate simulation.  
2. INITIAL TURN ON  
Turn on the instrument by depressing the ON-OFF keyswitch once (Figure  
2). At power-up, the unit first performs a display test. All segments and  
annunciators of the liquid crystal (LCD) are momentarily turned on for visual  
confirmation by the user (Figure 3).  
NOTE: In case of missing or poor-contrast segments, contact  
factory.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 2. Keypad  
Figure 3. Display Test  
If the unit is turning on for the first time after a new battery is installed, it will  
automatically default to the METER mode (ie. temperature measurement  
operation). It also defaults to K-type thermocouples, and °F readings with  
0.1° resolution. The corresponding LCD annunciators are enabled:  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
METER: (unit functioning as a thermometer)  
K:  
°F:  
OPR:  
(K-type thermocouple)  
(Fahrenheit readings)  
(Instrument is operational)  
If no input thermocouple is connected to the unit at this time, the numeric  
segments will indicate "OPEN".  
Plugging a thermocouple into the meter jack (connector T2, on the top right  
side) will give actual thermocouple temperature readings.  
NOTE: To obtain full accuracy, allow 1-2 minutes after connect-  
ing a thermocouple plug, for thermal setting.  
If the instrument is to be used as a thermometer or ohmmeter, refer to the  
Meter Mode Operation section for detailed instructions. For use as a tem-  
perature-calibrator or ohms simulator, refer to Calibrator Mode Operation.  
3. METER MODE OPERATION (Thermocouples)  
Set-Up:  
1. Turn on instrument by depressing the ON/OFF key.  
2. Check that the BAT annunciator turns off following the power-up LCD test.  
If it does not turn off, less than 10% of battery remains. Refer to Battery  
Installation/Replacement instructions.  
3. If the meter is already in the METER mode, the METER annunciator will  
be on. Otherwise the calibrator mode is enabled and the CALIB annunci-  
ator is on. Depress the METER key to return to meter mode.  
4. Connect an input thermocouple (TC) to connector T2. Select the appro-  
priate TC Type with the select-key. A display annunciator will indicate the  
selected TC-type.  
NOTE: Space and legibility constraints on the display preclude  
the use of dedicated annunciators for the less common tungsten-  
rhenium thermocouple types (G, C & D). Instead, these types are  
annunciated momentarily on the alpha-numeric display when first  
selected. Both T and R annunciators remain on to indicate a  
Tungsten-Rhenium selection. To recall the actual G, C, D selec-  
tion, turn the unit off, then on again for another momentary read-  
out.  
5. Select temperature scale with °F/°C key. The corresponding annunciator  
is enabled.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Select display resolution with 0.1°/1° key.  
NOTE: Set-up is retained during power-off.  
Hold:  
Meter readings can be put on hold at any time by depressing the HOLD key.  
At this time, the HOLD annunciator turns on, the reading is frozen, and all  
keys (except ON/OFF and HOLD) are locked out. Depress the HOLD key  
again to return to real-time readings, and re-enable the keypad.  
4. METER MODE OPERATION (RTD, Thermistor & Ohms)  
Set-Up:  
1. Connect a sensor or unknown resistance to the instrument as shown in  
Figures 4, 5, or 6.  
The 2-wire configuration is simplest, but includes lead-wire resistance in  
the measurement. 2-wire measurements are generally limited to high  
resistance sensors (e.g. 1000-ohm RTDs, 2252-ohm thermistors).  
Figure 4. 2-Wire Resistance Measurement  
Figure 5. 3-Wire Resistance Measurement  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WHT  
BLK  
GRN  
RED  
Figure 6. 4-Wire Resistance Measurement  
Lead resistance errors are compensated for in 3-wire configurations.  
However full compensation requires equal resistance in each lead. This  
configuration is common with 100-ohm RTDs.  
In 4-wire configurations, accuracy is unaffected by lead resistance, and  
resistance differences between leads.  
NOTE: These instruments measure input voltages on all 4 termi-  
nals, and are able to distinguish between 2-4 and 3-wire hookups.  
Switching between these measurement modes is automatic.  
2. Set instrument to METER mode.  
3. Use the sensor-select key to set the instrument to the appropriate func-  
tion and range.  
NOTE: Due to limited display area, 100and 1000RTD's are  
indicated on the display by a momentary readout of "RTD-100"  
(or "RTD-1000") when first selected.  
Thermistors are identified by a momentary alpha-numeric readout  
of "Y-400" (signifying YSI Series-400 type thermistor).  
Any of these readouts can be recalled by turning the unit off and  
back on.  
5. DATA LOGGING AND RECALL  
Operating as a meter, the unit can save and recall up to 10 readings (each  
with independent parameters: eg. °F/°C, TC-type and resolution) in random  
access memory.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
To Save Readings:  
1. While in the METER OPR mode, depress the STORE key. Note that the  
OPR annunciator is replaced by STO.  
2. Next depress one of the memory location numbers (keys 0, 1, . . ., 9). The  
corresponding memory annunciator turns on and the meter returns  
momentarily to the OPR mode, displaying thermocouple temperature.  
eg. STORE 7  
NOTES:  
The enabled location annunciator indicates that the meter read-  
ing at the instant that the location number key was hit, was  
stored in memory.  
• When data is stored in a location already filed with data, the  
new data replaces the old.  
• Stored readings are retained during power-off.  
To Recall Readings:  
1. While in the METER OPR mode, readings stored earlier can be displayed  
by first depressing the RECALL key. Note that the OPR annunciator is  
replaced by RCL, and all memory location annunciators are blanked.  
2. Next depress the keypad number corresponding to the desired location.  
Note that the annunciator for this location turns on, and the memory con-  
tents are displayed.  
3. Several stored readings can be randomly recalled by a key sequence  
such as:  
T1  
T2  
Calibrator  
(output)  
Thermometer  
(input)  
Figure 7.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
eg. RECALL 3  
eg. RECALL 5  
eg. RECALL 2  
eg. etc.  
4. To return the meter to the operate mode, depress OPR after recalling the  
last reading. All active location annunciators will turn on again, together  
with the OPR annunciator.  
NOTE: Recalling data from unused (ie, empty) locations will give  
a zero reading.  
To Erase Data Locations (Meter Mode):  
1. Individual data locations can be erased by recalling the location to be  
cleared, followed by a CLEAR key input. Note that the annunciator for the  
cleared location is turned off, and the meter returns to OPR.  
eg. RECALL 3  
eg. CLEAR  
NOTES: There is no need to clear a location if new data is ready  
for saving in this location. Just STORE the new data. It will write  
over and delete the old.  
6. CALIBRATOR MODE OPERATION (Thermocouples)  
Set-Up:  
1. Turn on the instrument by depressing the ON/OFF key.  
2. Check that the BAT annunciator turns off after display check. If not, refer  
to Battery Installation/Replacement instructions.  
3. When operating as a calibrator, the display indicates CALIB. Should the  
instrument be in the METER mode, hit the CALIB key to change to the cal-  
ibrator function.  
4. Select the thermocouple type that will be used.  
NOTE: Space and legibility constraints on the display preclude  
the use of dedicated annunciators for the less common tungsten-  
rhenium thermocouple types (G, C, & D). Instead, these types are  
annunciated momentarily on the alpha-numeric display when first  
selected. Both T and R annunciators remain on to indicate a  
Tungsten-Rhenium selection. To recall the actual G, C, D selec-  
tion, turn the unit off, then on again for another momentary read-  
out.  
5. Select the desired temperature scale (°F/°C key).  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Select the resolution desired (0.1°/1° key).  
NOTE: With 1° resolution, one less key stroke will be required to  
enter calibrator output temperatures via the keypad.  
7. The unit accepts numeric inputs by first changing the keypad to its num-  
ber-entry mode (NUM) by depressing the CHANGE/ENTER key once.  
The numeric display will flash, the OPR annunciator will turn off, and the  
NUM annunciator will turn on. Key in desired temperature value, most sig-  
nificant digit first. After the desired temperature value is keyed in, hit  
CHANGE//ENTER to return to OPR mode.  
7. To enter a value of 212° with 0.1° resolution, the following keystrokes are  
used:  
CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER  
NOTES:  
Invalid numeric entries (eg. a temperature beyond the span of  
the selected TC type) will cause a corresponding error code to  
be displayed.  
• To enter negative (ie. minus) temperatures, hit the –0 key first  
when in the NUM mode. The first hit on this key enters the neg-  
ative sign. Subsequent hits enter zeros.  
• To correct a numeric entry while in the NUM mode, hit the  
CHANGE/ENTER key twice. This will return number entry to  
the beginning of a new number.  
• To return the calibrator output to 0° while in the OPR mode, hit  
the CLEAR key.  
• The unit can be used as a convenient °F to °C or °C to °F cal-  
culator.  
For example, to convert 77°F to a Celsius reading, go to the CALIB OPR  
mode, enter the reading to be converted, and change scale.  
eg. °F  
eg. CHANGE/ENTER  
eg. 77  
eg. CHANGE/ENTER  
eg. °C (Display reads 25°C)  
7. CALIBRATOR MODE OPERATION (RTD, Thermistor & Ohms)  
Set-Up:  
1. Connect measurement equipment under test (eg. RTD thermometer,  
ohmmeter, etc.) to the instrument as shown in Figures 8, 9, or 10.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
mA (current) source  
Calibrator (out)  
Thermometer (in)  
I out  
V out  
Figure 8. 2-Wire Resistance Simulation  
Figure 9. 3-Wire Resistance Simulation  
GRN  
RED  
BLK  
WHT  
NOTE: Colors indicated apply to  
Model 84937 Calibration Cable.  
Figure 10. 4-Wire Resistance Simulation  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. The same performance tradeoffs apply to 2, 3 and 4-wire simulations as  
to 2, 3, and 4-wire measurements. Refer to RTD, Thermistor and Ohms  
Meter-Mode section above for wiring guidelines.  
2. Set instrument to CALIBRATOR mode.  
3. Use the sensor-select key to set the instrument to the appropriate func-  
tion and range.  
NOTE: To ensure accurate calibrator outputs, observe specified  
limits to excitation currents. If in doubt, check current levels with  
a DMM in series with either source lead (1 or 3).  
NOTE: Due to limited display area, 100and 1000RTD's are  
indicated on the display by a momentary readout of "RTD-100"  
(or "RTD-1000") when first selected.  
Thermistors are identified by a momentary alpha-numeric readout  
of "Y-400" (signifying YSI Series-400 type thermistor).  
Any of these readouts can be recalled by turning the unit off and  
back on.  
8. STORAGE AND RECALL OF CALIBRATOR SETTINGS  
While operating as a calibrator, the unit can store and recall up to 10 output  
settings (each with independent parameters, eg. °F/°C, sensor type, and res-  
olution) in random access memory.  
To Store Outputs:  
1. While in the CALIB OPR mode, depress the STORE key. Note that the  
OPR annunciator is replaced by STO.  
2. Next depress one of the memory location numbers (keys 0, 1, 2, . . ., 9).  
The corresponding memory annunciator turns on, and the calibrator  
returns to the OPR mode.  
eg. STORE 2  
NOTES:  
• The enabled memory annunciator shows that the output set-up  
has been saved.  
• When a set up is stored in a location already in use, the new  
set-up replaces the old.  
• Saved set-ups are retained during power-off.  
• To store a reading of 0° just hit the CLEAR while in the CALIB  
OPR mode. Then depress STORE and a location number.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
To Recall Outputs:  
1. While in the CALIB OPR mode, set-ups saved earlier in memory can be  
recalled to program the calibrator output. First depress the RECALL key  
to recall a set-up.  
1. Note that the RCL annunciator turns on, OPR turns off, and all memory  
locations annunciators are blanked.  
2. Next hit the key with the number corresponding to the desired memory  
location. The corresponding annunciator turns on, and the LCD updates  
to show the new calibrator setting.  
eg. RECALL 8  
3. Several stored set-ups can be randomly recalled by a key sequence such  
as:  
eg. RECALL 6  
eg. RECALL 3  
eg. RECALL 9  
eg. etc.  
4. To return the unit to the CALIB OPR mode after recalling the last set-up,  
hit the OPR key.  
NOTES:  
Any other key except ON/OFF and CLEAR will also return the  
OPR mode.  
• Set-ups recalled from unused locations will display and output  
0°.  
• Calibrator set-ups can be recalled even when the unit is func-  
tioning in the meter mode. The same 10 storage locations are  
shared in both meter and calibrator modes.  
• Readings stored in memory while operating as a METER, can  
be recalled in the CALIB mode. In this unusual application, the  
unit generates calibrator outputs that simulate earlier stored  
meter inputs.  
To Erase Set-Up Locations (Calibrator Mode):  
1. Individual set-up locations can be erased by recalling the location followed  
by a hit to the CLEAR key.  
eg. RECALL 5  
eg. CLEAR  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. STEP FUNCTION  
When operating in the CALIBRATOR mode, the STEP key will sequentially  
recall data stored in memory (up to 10 steps). At each step, the calibrator will  
display and output these parameters. The STEP function is non-functional in  
METER mode.  
10. RAMP FUNCTION: CL25 and CL27  
This feature, available only on Models CL25 and CL27, allows the calibrator  
to automatically ramp up and down a user-defined staircase. Up and down  
keys allow for manual ramping by the user.  
To activate the RAMP function, the instrument must first be in the CALI-  
BRATE mode. Depress the RAMP key once to enter the ramp programming  
mode. The display reads "prog", with memory annunciators 1-4 flashing. A  
second hit to the RAMP key starts automatic ramping. A third hit turns off the  
RAMP function.  
User definable RAMP memory locations 1, 2, 3, & 4 are dedicated to storage  
of RAMP parameters; step size, staircase starting point, upper limit, and  
lower limit respectively. These memory locations are accessible only when in  
RAMP “prog” mode, and are independent of the other 10 memory locations.  
Numeric values for RAMP parameters can be set, stored, recalled, and  
cleared like other calibrator settings. (Refer page 17, CALIBRATOR MODE  
OPERATION).  
When automatic ramping is in progress, a hit to either the 7 or 8 key activates  
manual up-ramping or down-ramping respectively. Whenever the battery is  
replaced (and memory backup is not performed), the ramp parameters  
default to factory-set values:  
LOC 1 Step Size =10°F (K TC)  
LOC 2 Start Point = 500°F  
LOC 3 High Limit = 550°F  
LOC 4 Low Limit = 450°F  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SERVICE INFORMATION  
WARNING  
All service information is intended for qualified electronic mainte-  
nance personnel only.  
1. CALIBRATION PROCEDURES  
This is a two-part procedure. Products with thermocouple functions only, fol-  
low the Thermocouple Calibration procedure below. Products with resistance  
functions (RTD, thermistor, ohms) follow first the Thermocouple Calibration  
procedure, then the Resistance Calibration procedure.  
A. THERMOCOUPLE CALIBRATION  
Test Equipment Required:  
1. Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to  
ITS-90.  
2. Microvolt DMM (OMEGA OM7551, or equivalent)  
3. Calibration cables, per Figures 11, 12, 13.  
NOTE: The copper constantan cables used in Figures 12 and 13  
need to be calibrated. Voltage errors in these cables should be com-  
pensated at the simulator.  
4. Calibration cover (OMEGA 820-307-4).  
Ambient Conditions:  
Units should be calibrated at an ambient temperature of 23°C 1°C, with rel-  
ative humidity less than 80%.  
Preparation for Calibration:  
11. Remove battery from unit under test (U.U.T.). Refer to Battery  
Installation/Replacement section in the manual.  
12. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.  
13. Remove calibration-jumpers J1 and J2 from printed-circuit board.  
NOTE: Removal of J1, J2 causes partial loss of previously stored  
calibration data. Do not remove these jumpers unless recalibra-  
tion is intended.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTE: For calibrators with the older, single-jumper configura-  
tion, refer to calibration procedure beginning on page 39.  
14. Install calibration cover in place of bottom-cover.  
15. Re-install battery.  
16. Hook up test-equipment, calibration cables, and U.U.T. per Figure 10.  
17. Turn on DMM and TC simulator for warmup. Allow at least 30 minutes.  
18. Set the DMM to 200mV DC range.  
19. Set the TC simulator to its COPPER output mode, with a reference tem-  
perature of 32.0°F.  
10. Turn on U.U.T. and verify that "RTD" and "" annunciators are flashing  
on the display. This verifies removal of J1 and J2 respectively.  
NOTE: Flashing “RTD” indicates that zero-offset corrections can  
be stored in EEPROM. It also indicates that cold-junction com-  
pensation is disabled.  
Flashing '' indicates that error corrections for the cold-junction  
sensors can be stored or recalled from EEPROM.  
11. Clear the EEPROM locations used to store E1 (meter cold-junction sen-  
sor error) and E2 (calibrator cold-junction sensor error) as follows:  
a. Set U.U.T to CALIB, OPR, K, °F mode.  
b. Key in: CHANGE/0.0/ENTER/STO/1/OPR (E1 is set to 0.0°F)  
c. Key in: CHANGE/0.0/ENTER/STO/2/OPR (E2 is set to 0.0°F)  
NOTE: Both 1 & 2 memory annunciators should be turned on,  
but the “RTD” annunciator will have turned off.  
d. Perform RECALL 1 and RECALL 2 to verify storage of 0.0°F in both  
locations.  
NOTE: At the completion of calibration, E1 and E2 will be saved  
in EEPROM. Memory locations 1 and 2 will be free for normal  
use.  
e. Set U.U.T. to METER, OPR, K, °F.  
f. Turn U.U.T. OFF, then ON to re-enable 'RTD' and '' annunciators.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Adjustments:  
NOTE: Do not deviate from the calibration adjustment sequence  
that follows. This will ensure that adjustments to be stored in  
EEPROM go to the correct locations.  
1. Set meter-mode zero and gain adjustments per Table 1.  
2. Hit OPR key (RTD annunciator turns off).  
3. Set U.U.T. to CALIB, OPR, K, °F.  
4. Turn U.U.T. OFF, then ON, to re-enable RTD & annunciators.  
5. Set calibrator-mode zero and gain adjustments per Table 2. Hit  
CHANGE and ENTER at the completion of each step in Table 2.  
6. Hit OPR key (RTD annunciator turns off).  
7. Install J1 jumper.  
8. Set U.U.T. to METER, OPR, T, °F  
9. Change TC simulator output to Type T, 32°F, ALLOY mode.  
10. Reconfigure calibration set-up per Figure 12. Allow 2-3 minutes for ther-  
mal stabilization.  
11. Calculate E1 as follows: E1 = U.U.T. Reading — 32.0.  
12. Set U.U.T. to CALIB, OPR, T, °F mode.  
13. Store E1; CHANGE/(E1 value)/ENTER/STO/1/OPR.  
14. Go to METER mode, and verify U.U.T. reading of 32.0 0.1°F.  
15. Reconfigure calibration setup per Figure 12. Leave simulator output at  
32°F, Type T, ALLOY mode. Allow 2-3 minutes for thermal stabilization.  
16. Set U.U.T. to CALIB mode, 32°F, Type T.  
17. Calculate E2 as follows:  
DMM Reading (in µV)  
–22  
E2 =  
(For example with DMM reading of +11µV, E2 = –0.5°F)  
18. Store E2; CHANGE/(E2 value)/ENTER/STO/2/OPR.  
19. Install J2 jumper.  
20. Re-set U.U.T. output to 32°F, DMM reading should be 0 2µV.  
21. Turn OFF U.U.T.  
22. Thermocouple calibration is complete. Remove calibration cover. Re-  
install original back-cover, unless going on to Part B (Resistance  
Calibration).  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1: Meter-Mode Zero & Gain Adjustments  
Sensor  
Select  
U.U.T.  
Input  
U.U.T.  
Reading*  
Step  
Adjust  
1
2
K
K
0.000 mV  
54.856 mV  
P1  
P11  
32.0 0.1°F  
2500.0 0.1°F  
3
Repeat Steps 1 & 2 As Required  
4
5
6
7
8
J
T
E
N
B
R
S
**G**  
C
D
0.000 mV  
0.000 mV  
0.000 mV  
0.000 mV  
1.439 mV  
0.000 mV  
0.000 mV  
2.238 mV  
0.000 mV  
0.000 mV  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
32.0 0.1°F  
32.0 0.1°F  
32.0 0.1°F  
32.0 0.1°F  
1000.0 0.5°F  
32.0 0.5°F  
32.0 0.5°F  
600.0 0.2 °F  
32.0 0.2°F  
32.0 0.2°F  
9
10  
11  
12  
13  
** exclusive of noise.  
** on EC models, G replaced by L. Calibrate L per J.  
Table 2: Calibrator-Mode Zero & Gain Adjustments  
Sensor  
Select  
U.U.T.  
Output  
DMM  
Reading*  
Step  
Adjust  
1
2
K
K
32.0°F  
2500.0°F  
P1  
P41  
0.000 mV 1µV  
54.856 mV 1µV  
3
Repeat Steps 1 & 2 As Required  
4
5
6
7
8
J
T
E
N
B
R
S
**G**  
C
D
32.0°F  
32.0°F  
32.0°F  
32.0°F  
1000.0°F  
32.0°F  
32.0°F  
600.0°F  
32.0°F  
32.0°F  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
P1  
0.000 mV 1µV  
0.000 mV 1µV  
0.000 mV 1µV  
0.000 mV 1µV  
1.439 mV 1µV  
0.000 mV 1µV  
0.000 mV 1µV  
2.238 mV 1µV  
0.000 mV 1µV  
0.000 mV 1µV  
9
10  
11  
12  
13  
** exclusive of noise.  
** on EC models, G replaced by L. Calibrate L per J.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 11. Calibration Setup, Copper Mode.  
Figure 12. Calibration Setup, Meter Mode.  
Figure 13. Calibration Setup, Calibrator Mode.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
B. RESISTANCE CALIBRATION  
NOTE: Thermocouple and resistance calibrations are indepen-  
dent. However, potentiometer P1 is used in both thermocouple  
and resistance calibrations to set EEPROM constants. Follow cal-  
ibration procedures carefully to ensure that thermocouple and  
resistance calibrations do not disturb one another.  
Test Equipment Required:  
1. Resistance Decades (General Resistance RTD-100, RTD-500/1000, and  
RDS-54, or equivalent).  
2. Current Source (Fluke 5101B, or equivalent).  
3. Digital Multimeter (OMEGA OM7551, or equivalent).  
4. Calibration cables per Figure 14.  
Ambient Conditions:  
Unit should be calibrated at an ambient temperature of 23°C 1°C, with rela-  
tive humidity less than 80%.  
Preparation For Calibration:  
1. Turn off unit, and remove bottom-cover. Refer to Disassembly  
Instructions.  
2. Secure battery with a piece of adhesive tape.  
NOTE: Momentary loss of battery power during the following cal-  
ibration will invalidate previous adjustments, both resistance and  
thermocouple.  
3. Hook up test-equipment and unit per Figure 14.  
4. Set current-source to 1mA, STANDBY mode.  
5. Set DMM to 2V DC range.  
6. Turn on unit.  
7. Set unit to METER mode, RTD-100, °F.  
8. Turn off unit. Remove jumper J1. Turn on unit. 'RTD' annunciator should  
be flashing.  
NOTES: Flashing 'RTD' indicates that calibration adjustments  
will be written to memory. Avoid selecting thermocouple  
functions while in this mode.  
NOTES: Should it be necessary to scroll through thermocouple  
functions, first hit the OPR key (disables flashing 'RTD'),  
and start again at Step 7 above.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 14. Resistance Calibration Setup  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Adjustments:  
NOTE: Do not deviate from the calibration adjustment sequence  
that follows. This will ensure that adjustments to be stored in  
EEPROM go to the correct locations.  
1. Perform meter-mode adjustments and verifications per Table 3 (use Table  
5 for RTD-Thermistor models).  
2. Hit OPR key (Flashing 'RTD' annunciator turns off).  
3. Install J1 jumper.  
4. Set unit to CALIB mode, RTD-100, °F.  
5. Switch current source to OPERATE mode.  
6 Perform calibrator-mode adjustments and verifications per Table 4 (use  
Table 6 for RTD-Thermistor models).  
7. Turn off unit. Remove tape holding battery.  
8. Re-install back cover. Resistance-function calibration is complete.  
Table 3: Calibration of Meter-Mode Resistance Functions  
(RTD-100/1000)  
Sensor  
Select  
Unit  
Input  
Unit  
Reading*  
Step  
Adjust  
Verify  
1
2
3
4
5
6
7
8
9
RTD-100  
RTD-100  
RTD-100  
390.26Ω  
100.00Ω  
18.49Ω  
P1**  
1562.0°F 0.5°F  
32.0 0.1°F  
–328.0 0.2°F  
1562.0°F 0.5°F  
32.0+0.1°F  
900.00Ω  
RTD-1000 3,902.6Ω  
P1**  
P1  
RTD-1000  
1,000Ω  
1000.0Ω  
900.00Ω  
1.00Ω  
9,000.0Ω  
10.0Ω  
1,000Ω  
1.00 0.10Ω  
9000.0Ω  
10,000Ω  
10,000Ω  
P1  
10.0 1.0Ω  
* exclusive of noise.  
** adjust as necessary to obtain steps 2, 5.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 4: Calibration of Calibrator-Mode Resistance Functions  
(RTD-100/1000)  
Current  
Source  
Input  
Sensor  
Select  
Unit  
Output  
DMM  
Step  
Adjust Verify Reading*  
1
2
RTD-100  
RTD-100  
1562.0°F  
32.0°F  
1mA  
1mA  
P38  
0.39026v  
0.10000  
0.00003v  
0.01849  
0.00006v  
0.39026v  
0.10000  
0.00003v  
0.90000  
0.00010v  
0.00100  
0.00010v  
0.90000  
0.00010v  
0.00100  
0.00010v  
3
RTD-100  
–328.0°F  
1mA  
4
5
RTD-1000  
RTD-1000  
1562.0°F 100µA  
P51  
32.0°F  
900.00Ω  
1.00Ω  
100µA  
6
7
8
9
1,000Ω  
1,000Ω  
1mA  
1mA  
10,000Ω  
10,000Ω  
9,000.0100µA  
10.0100µA  
* exclusive of noise.  
Table 5: Calibration of Meter-Mode Resistance Functions  
(RTD-Thermistor)  
Sensor  
Select  
Unit  
Input  
Unit  
Reading*  
Step  
Adjust  
Verify  
1
2
3
4
5
6
7
8
9
RTD-100  
RTD-100  
RTD-100  
1,000Ω  
390.26Ω  
100.00Ω  
18.49Ω  
900.00Ω  
1.00Ω  
P1  
1562.0°F  
32.0 0.3°F  
–328.0 0.5°F  
900.00Ω  
P1  
P1  
P1  
1,000Ω  
1.00 0.20Ω  
90,000Ω  
100,000Ω  
100,000Ω  
Thermistor  
Thermistor  
Thermistor  
90,000Ω  
100Ω  
100 20Ω  
293.0°F  
77.0 0.2°F  
–40.0 0.2°F  
47.0Ω  
2252Ω  
10  
75,790Ω  
* exclusive of noise.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 6: Calibration of Calibrator-Mode Resistance Functions  
(RTD-Thermistor)  
Current  
Source  
Input  
Sensor  
Select  
Unit  
Output  
DMM  
Step  
Adjust Verify Reading*  
1
2
RTD-100  
RTD-100  
1562.0°F  
32.0°F  
1mA  
1mA  
P38  
0.39026v  
0.10000  
0.00006v  
0.01849  
0.00012v  
0.90000  
0.00020v  
0.00100  
0.00020v  
0.90000v  
0.00100  
0.00020v  
0.04700  
0.00020v  
0.02252  
0.00020v  
0.75790  
0.00500v  
3
4
5
RTD-100  
1,000Ω  
1,000Ω  
–328.0°F  
900.00Ω  
1.00Ω  
1mA  
1mA  
1mA  
6
7
100,000Ω  
100,000Ω  
90,000Ω  
100Ω  
10µA  
10µA  
P51  
8
9
Thermistor  
Thermistor  
Thermistor  
293.0°F  
77.0°F  
1mA  
10µA  
10µA  
10  
–40.0°F  
* exclusive of noise.  
2. CALIBRATION VERIFICATION  
Calibration is verified with the same instrument hookups as used for calibra-  
tion (ie. Figures 12 and 13 for thermocouple functions, Figure 14 for resis-  
tance functions).  
While one hookup serves to verify all resistance functions (RTD, thermistor,  
ohms), a full check-out of thermocouple functions requires a set of calibration  
cables for each thermocouple type.  
To simplify the verification of thermocouple calibration, software has been  
enhanced to reactivate the copper-mode used during calibration. This feature  
is available on products that use the following (and later) software revisions:  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model  
CL23A  
CL24  
CL25  
CL26  
CL27  
Software Revision  
6.7  
7.7  
3.7  
5.6  
4.16  
(Software revision is marked on a label attached to integrated-circuit U1 on  
the mother-board).  
"Copper-mode" is a non-compensated mode of operation during which  
cold-junction compensation of inputs and outputs is disabled. This permits  
the use of copper-wire interconnections between instruments, while avoid-  
ing the time and expense of multiple thermocouple-cable hookups. At the  
conclusion of calibration verification, cold-junction compensation is re-  
enabled, and operation returns to "alloy-mode".  
“Alloy-mode” refers to Thermocouple operations; for example, a Type-K  
thermocouple is made up of Nickel/Chromium and Nickel/Aluminum alloys.  
Referring to the OMEGA Calibrator(s) in TC operation, “Alloy-mode” means  
in effect “other than Copper-mode.” In order to make accurate measure-  
ments using a thermocouple device one must eliminate erroneous voltage  
signals. In “Alloy-mode”, the OMEGA calibrator automatically calculates the  
cold-junction compensation for the selected TC type, based on the value(s)  
from the ITS-90 table. This feature enables the user to make accurate tem-  
perature measurements without making manual calculations for error cor-  
rections.  
COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE  
FUNCTIONS  
Equipment Required:  
1. OMEGA Model to be verified.  
2. Voltmeter accurate to 1µV resolution.  
3. Cable assembly (Fig. 11).  
1. Start your calibration with a fresh battery.  
2. Press the ON function button to turn the unit on.  
3. Select the calibrator mode (“CALIB” on display) by pressing the  
CALIB/METER function button.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Select type K thermocouple (“K” on display) by pressing the SENSOR  
SELECT function button until type K is reached.  
5. Select the tenth degree display (with a digit appearing after the decimal  
point on the display) by pressing the 0.0.1 function button.  
6. Select °F (“°F” on display) by pressing the °F °C function button.  
7. Press the CHANGE function button. The numerical display should blink.  
8. Press 9 9 9 9 9. (“9999.9” should appear blinking on display).  
9. Press the ENTER function button. The display will blink “Hi” once or  
twice, and then will display “0.0 °F” blinking. There will also be an “”  
symbol blinking in the upper left. It will blink with the “0.0 °F” display.  
10. Press the ENTER function button. The “0.0 °F” display will stop blinking,  
and the “” will continue to blink. You are now in the copper mode. If or  
when you turn off the power (or remove the battery) the unit will no longer  
be in the copper mode and you would have to repeat the above steps to  
return to the copper mode. As long as you see the “” blinking you  
are in the copper mode.  
11. While in the copper mode attach the Copper Mode Calibrator Verification  
Cable to the connector on the top of the CL25 that is labeled T1. Attach  
the other end of the cable to a precision voltmeter.  
12. Press the CHANGE function button. The “0.0 °F” display will blink, and  
“NUM” will appear on the display.  
13. Press 3 2 0 buttons and then press the ENTER function button. “NUM”  
will disappear and “32.0 °F” will appear, not blinking. The voltmeter  
should read 0.000 mV.  
14. Press the CHANGE function button. The “32.0 °F” display will blink, and  
“NUM” will appear on the display.  
15. Press 2 5 0 0 0 buttons and then press ENTER function button. “NUM”  
will disappear and “2500.0 °F” will appear, not blinking. The voltmeter  
should read 54.845 mV.  
16. Press the CHANGE function button. The “2500.0 °F” display will blink,  
and “NUM” will appear on the display.  
17. Press 3 2 0 buttons and then press the ENTER function button. “NUM”  
will disappear and “32.0 °F” will appear, not blinking. The voltmeter  
should read 0.000 mV.  
18. Press the SENSOR SELECT function button. The sensor type will  
change to type J. The “K” will disappear from the display and a “J” will  
appear on the display. The voltmeter should read 0.000 mV.  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
19. Press the SENSOR SELECT function button. The sensor type will  
change to type T. The “J” will disappear from the display and a “T” will  
appear on the display. The voltmeter should read 0.000 mV.  
20. Press the SENSOR SELECT function button. The sensor type will  
change to type E. The “T” will disappear from the display and an “E” will  
appear on the display. The voltmeter should read 0.000 mV.  
21. Press the SENSOR SELECT function button. The sensor type will  
change to type N. The “E” will disappear from the display and a “N” will  
appear on the display. The voltmeter should read 0.000 mV.  
22. Press the SENSOR SELECT function button. The sensor type will  
change to type B. The “N” will disappear from the display and a “B” will  
appear on the display. “LO” will also appear on the display.  
23. Press the CHANGE function button. The “32.0 °F” display will blink, and  
“NUM” will appear on the display.  
24. Press 1 0 0 0 0 buttons and then press ENTER function button. “NUM”  
will disappear and “1000.0 °F” will appear, not blinking. The voltmeter  
should read 1.438 mV.  
25. Press the SENSOR SELECT function button. The sensor type will  
change to type R. The “B” will disappear from the display and a “R” will  
appear on the display.  
26. Press the CHANGE function button. The “1000.0 °F” display will blink,  
and “NUM” will appear on the display.  
27. Press 3 2 0 buttons and then press ENTER function button. “NUM” will  
disappear and “32.0 °F” will appear, not blinking. The voltmeter should  
read 0.000 mV.  
28. Press the SENSOR SELECT function button. The sensor type will  
change to type S. The “R” will disappear from the display and a “S” will  
appear on the display. The voltmeter should read 0.000 mV.  
29. Press the SENSOR SELECT function button. The sensor type will change to  
type G. The “R” will disappear from the display and a “g” will appear momen-  
tarily in the center of the display. “LO” will then appear in the center of the dis-  
play and both the “T” and “R” will appear along the top row of the display.  
30. Press the CHANGE function button. The “32.0 °F” display will blink, and  
“NUM” will appear on the display.  
31. Press 6 0 0 0 buttons and then press ENTER function button. “NUM” will dis-  
appear and “600.0 °F” will appear, not blinking. The voltmeter should read  
2.238 mV.  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
32. Press the SENSOR SELECT function button. The sensor type will change to  
type C. A “C” will appear momentarily in the center of the display. “600.0” will  
then appear in the center of the display and both the “T” and “R” will appear  
along the top row of the display.  
33. Press the CHANGE function button. The “600.0 °F” display will blink, and  
“NUM” will appear on the display.  
34. Press 3 2 0 buttons and then press ENTER function button. “NUM” will dis-  
appear and “32.0 °F” will appear, not blinking. The voltmeter should read  
0.000 mV.  
35. Press the SENSOR SELECT function button. The sensor type will change to  
type D. A “d” will appear momentarily in the center of the display. “32.0” will  
then appear in the center of the display and both the “T” and “R” will appear  
along the top row of the display. The voltmeter should read 0.000 mV.  
36. Press the SENSOR SELECT function button. The sensor type will change to  
type K. The “T” and the “R” will disappear from the display and a “K” will  
appear on the display. The voltmeter should read 0.000 mV.  
37. Press the OFF function button. The unit will turn off.  
38. Press the ON function button. The unit will power up and return to the cali-  
brator mode. “CALIB” will appear on the display along with the “K” enuncia-  
tor, the “OPR” enunciator, and “32.0 °F” will be displayed. The blinking ”  
symbol will have completely disappeared indicating that you have exited the cop-  
per mode.  
39. Remove the Copper Mode Calibrator Verification Cable from both the voltmeter  
and the CL25.  
40. The copper mode verification is now complete. Press the OFF function button to  
turn the unit off.  
ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE  
FUNCTIONS  
Equipment Required:  
1. OMEGA Model to be verified.  
2. Precision temperature simulator / voltage source.  
3. Voltmeter accurate to 1µV resolution.  
4. Cable assembly.  
1. Start your calibration with a fresh battery.  
2. Press the ON function button to turn the unit on.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Select the calibrator mode (“CALIB” on display) by pressing the  
CALIB/METER function button.  
4. Select type K thermocouple (“K” on display) by pressing the SENSOR  
SELECT function button until type K is reached.  
5. Select the tenth degree display with a digit appearing after the decimal  
point on the display) by pressing the 0.0.1 function button.  
6. Select °F (“°F” on display) by pressing the °F °C function button.  
7. Attach the Alloy Mode Calibrator Verification Cables per Fig. 12 & 13. If  
you are using a Type K Thermocouple Temperature Simulator ITS-90  
you will use the regular set up. If you are using a Precision Voltage  
Source use the alternate set up. When performing the following proce-  
dure using a Precision Voltage Source with the alternate set up, any-  
where a temperature is transmitted (output) by the Thermocouple  
Temperature Simulator, you should adjust the Voltage Source to the  
equivalent voltage as listed by a Type K N.I.S.T. ITS-90 Thermocouple  
Reference Table. (Example: 32.0 °F would equal 0.000 mV)  
8. Press the CHANGE function button and the numerical display will blink  
and “NUM” will appear on the display.  
9. Press 3 2 0 buttons and then press the ENTER function button. “NUM”  
will disappear and “32.0 °F” will appear, not blinking.  
10. Adjust the Thermocouple Temperature Simulator to 32.0 °F.  
00. Note: The reading on the voltmeter should read null. 0.000 mV. Adjusting  
the CL25 to any temperature and adjusting the Temperature Simulator to  
the same temperature should result in a null, or 0.000 mV on the volt-  
meter.  
11. Remove the Alloy Mode Calibrator Verification Cable from all equipment.  
12. The Alloy Mode verification is now complete. Press the OFF function but-  
ton to turn the unit off.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. MECHANICAL PARTS DIAGRAM  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CALIBRATION PROCEDURES  
Test Equipment Required:  
1. Thermocouple simulator (OMEGA CL25, or equivalent).  
2. Microvolt null-meter (Keithley 155, or equivalent)  
3. Calibration cover (OMEGA 820-307-4).  
4. Calibration cables, T.C. types K, J, T & E (as required) per Figure1 below.  
5. Low-thermal (eg. copper) shorting-link for null-meter.  
NOTE: The (2) alloy-copper junctions in the calibration cables  
can be either solder-joints (cleaned to remove flux), or crimp-con-  
nections. Insulate each junction with heat-shrink tubing, then  
wrap them close together in copper-foil to ensure that they main-  
tain identical temperatures (ambient).  
Ambient Conditions:  
Units should be calibrated at an ambient temperature of 23°C 1°C, with rel-  
ative humidity less than 80%. Test equipment and the unit-under-test  
(U.U.T.) should have 1 hour to reach thermal equilibrium with ambient-tem-  
perature.  
Calibration Adjustments:  
11. Remove battery from unit under test (U.U.T.). Refer to Battery  
Installation/Replacement instructions in the manual.  
12. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.  
13. Remove calibration-jumper J1 from printed-circuit board.  
NOTE: When installed, J1 protects the contents of EEPROM  
locations where zero-offset adjustments are stored. Once J1 is  
removed, this data is erased when the U.U.T. is turned on.  
14. Install calibration cover in place of bottom-cover.  
15. Re-install battery.  
16. Connect U.U.T. to test equipment per Figure 2 using the K-type calibra-  
tion cable.  
17. Turn on U.U.T. and allow 2-3 minutes for thermal re-stabilization of the  
TC mini-plug and input-jack.  
NOTE: • U.U.T. will be in the Type K, °F meter-mode. The FUNC  
annunciator is on, indicating that new constants will be  
stored in EEPROM.  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTE: • With J1 removed, the digital-filters in the U.U.T. are dis-  
abled. This is to speed up U.U.T. response to calibra-  
tion adjustments. In this mode 0.1°F of reading noise  
is normal.  
NOTE: • Do not deviate from the calibration adjustment  
sequence that follows. This will ensure that zero adjust-  
ments for each TC type will be stored in the appropriate  
memory locations.  
18. Apply 32°F (Type K) to U.U.T. Adjust potentiometer P1 for a reading of  
32.0°F (exclusive of noise).  
19. Apply 2500°F (Type K) to U.U.T. Adjust potentiometer P11 for a reading  
of 2500.0°F.  
NOTE: Adjustment steps 8 & 9 are interactive. Repeat these  
steps as necessary to get the desired readings.  
10. Hit OPR key to store K-meter zero-adjustment in EEPROM. Note that  
“FUNC” annuciator turns off.  
11. Select Type J thermocouple (KJT key on CL23A, SENSOR SELECT key  
on CL24).  
12. Turn off U.U.T.  
13. Replace K calibration-cable with J calibration-cable.  
14. Turn on U.U.T. Note that “FUNC” annunciator is on again. Allow 2-3 min-  
utes for thermal restabilization.  
15. Apply 32°F (Type J). Adjust P1 for a reading of 32.0°F (exclusive of  
noise).  
16. Hit OPR key to store J-meter zero-adjustment.  
17. Select Type T thermocouple.  
18. Turn off U.U.T.  
19. Replace J calibration-cable with T calibration-cable.  
20. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.  
21. Apply 32°F (Type T). Adjust P1 for a reading of 32.0°F (exclusive of  
noise).  
22. Hit OPR key to store T-meter zero-adjustment.  
NOTE: Steps 23 thru 28 apply to Model CL24 only. For Model  
CL23A, go to step 29 next.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
23. Select Type-E thermocouple.  
24. Turn off U.U.T.  
25. Replace T calibration-cable with E calibration-cable.  
26. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.  
27. Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of  
noise).  
28. Hit OPR key to store E-meter zero-adjustment.  
29. Select Type-K thermocouple.  
30. Select CALIB mode.  
31. Turn off U.U.T.  
32. Connect U.U.T. to test-equipment per Figure 3 using the K calibration-  
cable.  
33. Turn-on U.U.T. and allow 2-3 minutes, for thermal stabilization  
34. Set both the U.U.T. and TC-simulator to output 32.0°F (Type K). Adjust  
P2 for a null-meter reading of 0 3µV.  
35. Set both the U.U.T. and T-simulator to output 2400°F (Type K). Adjust  
P41 for a null-meter reading of 0 3µV.  
NOTE: Steps 34 and 35 are interactive. Repeat each as neces-  
sary to get the desired null readings.  
36. Hit OPR key to store K-calib zero adjustment.  
37. Select J thermocouple.  
38. Turn off U.U.T.  
39. Replace K calibration-cable with J calibration-cable.  
40. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.  
41. Set both the U.U.T. and TC-simulator to output 32.0°F (Type J). Adjust  
P1 for a null-meter reading of 0 3µV.  
42. Hit OPR Key to store J-calib zero adjustment.  
43. Select Type-T thermocouple.  
44. Turn off U.U.T.  
45. Replace J calibration-cable with T calibration-cable.  
46. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
47. Set both the U.U.T. and TC-simulator to output 32.0°F (Type T). Adjust  
P1 for a null-meter reading of 0 3µV.  
48. Hit OPR key to store T-calib zero adjustment.  
NOTE: Steps 49 thru 54 apply to Model CL24 only. For Model  
CL23A go next to step 55.  
49. Select Type-E thermocouple.  
50. Turn off U.U.T.  
51. Replace T calibration-cable with E calibration-cable.  
52. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.  
53. Set both the U.U.T. and thermocouple-simulator to 32.0°F (Type E).  
Adjust P1 for a null-meter reading of 0 3µV.  
54. Hit OPR Key to store E-calib zero adjustment  
55. Turn off U.U.T.  
56. Remove calibration cover. Be careful not to disturb battery connections.  
57. Re-install J1 jumper.  
58. Re-install original back cover. Calibration is complete.  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
IN  
USA  
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a  
period of 37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace  
period to the normal three (3) year product warranty to cover handling and shipping time. This  
ensures that OMEGA’s customers receive maximum coverage on each product.  
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s  
Customer Service Department will issue an Authorized Return (AR) number immediately upon  
phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will  
be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting  
from any action of the purchaser, including but not limited to mishandling, improper interfacing,  
operation outside of design limits, improper repair, or unauthorized modification. This  
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence  
of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration;  
improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s  
control. Components which wear are not warranted, including but not limited to  
contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However,  
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any  
damages that result from the use of its products in accordance with information provided by  
OMEGA, either verbal or written. OMEGA warrants only that the parts  
manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER  
WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED,  
EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MER-  
CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMI-  
TATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total  
liability of OMEGA with respect to this order, whether based on contract, warranty, negligence,  
indemnification, strict liability or otherwise, shall not exceed the purchase price of the compo-  
nent upon which liability is based. In no event shall OMEGA be liable for consequential, incidental  
or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic  
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical  
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or  
activity, medical application, used on humans, or misused in any way, OMEGA assumes no  
responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser  
will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out  
of the use of the Product(s) in such a manner.  
RETURN REQUESTS / INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.  
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED  
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID  
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return  
package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent  
breakage in transit.  
FOR WARRANTY RETURNS, please have the fol-  
lowing information available BEFORE  
contacting OMEGA:  
1. P.O. number under which the product was  
PURCHASED,  
FOR NON-WARRANTY REPAIRS, consult  
OMEGA for current repair charges. Have the fol-  
lowing information available BEFORE  
contacting OMEGA:  
1. P.O. number to cover the COST  
of the repair,  
2. Model and serial number of the product under  
warranty, and  
2. Model and serial number of product, and  
3. Repair instructions and/or specific  
problems relative to the product.  
3. Repair instructions and/or specific problems  
relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.  
This affords our customers the latest in technology and engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,  
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of  
OMEGA ENGINEERING, INC.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
TEMPERATURE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
PRESSURE, STRAIN AND FORCE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Transducers & Strain Gauges  
Load Cells & Pressure Gauges  
Displacement Transducers  
Instrumentation & Accessories  
FLOW/LEVEL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Rotameters, Gas Mass Flowmeters & Flow Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
pH/CONDUCTIVITY  
ߜ
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
ߜ
 
ߜ
 
ߜ
 
DATA ACQUISITION  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Datalogging Systems  
Recorders, Printers & Plotters  
HEATERS  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Laboratory Heaters  
ENVIRONMENTAL  
MONITORING AND CONTROL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen Instruments  
M1554/499  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miele Ventilation Hood DA 420 V User Manual
Milwaukee Cordless Drill 2602 20 User Manual
Milwaukee Home Safety Product 2235 20 User Manual
MPI Technologies Printer D60425 06 User Manual
Nextar GPS Receiver M30408EH02 User Manual
Nilfisk Advance America Pressure Washer C 1005 User Manual
Nokia Cell Phone X200 User Manual
Nortel Networks Power Supply Passport 8002PS User Manual
Oeuf Baby Carrier CR001 User Manual
Olympus MP3 Player DS 20 User Manual