Important Information
Warranty
The NI CVS-1450 Series compact vision system is warranted against defects in materials and workmanship for a period of one year from the
date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that
proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are
covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,
flood, accident, actions of third parties, or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
Trademarks
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section
on ni.com/legal for more information about National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no
agency, partnership, or joint-venture relationship with National Instruments.
Patents
For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file
on your CD, or ni.com/patents.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Download from Www.Somanuals.com. All Manuals Search And Download.
Compliance
Compliance with FCC/Canada Radio Frequency Interference
Regulations
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)
or Class B (for use in residential or commercial locations). All National Instruments (NI) products are FCC Class A products.
Depending on where it is operated, this Class A product could be subject to restrictions in the FCC rules. (In Canada, the
Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.) Digital
electronics emit weak signals during normal operation that can affect radio, television, or other wireless products.
All Class A products display a simple warning statement of one paragraph in length regarding interference and undesired
operation. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.
FCC/DOC Warnings
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions
in this manual and the CE marking Declaration of Conformity*, may cause interference to radio and television reception.
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department
of Communications (DOC).
Changes or modifications not expressly approved by NI could void the user’s authority to operate the equipment under the
FCC Rules.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user is required to correct the interference
at their own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Compliance with EU Directives
Users in the European Union (EU) should refer to the Declaration of Conformity (DoC) for information* pertaining to the
CE marking. Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance
information. To obtain the DoC for this product, visit ni.com/certification, search by model number or product line,
and click the appropriate link in the Certification column.
*
The CE marking Declaration of Conformity contains important supplementary information and instructions for the user or
installer.
Download from Www.Somanuals.com. All Manuals Search And Download.
Conventions
The following conventions are used in this manual:
»
The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a tip, which alerts you to advisory information.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
bold
Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes hardware
labels and parameter names.
italic
Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a word
or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames, and extensions.
monospace bold
Bold text in this font denotes the messages and responses that the computer
automatically prints to the screen. This font also emphasizes lines of code
that are different from the other examples.
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Hardware Overview.......................................................................................................1-1
Available Camera Bandwidth........................................................................................1-3
Software Overview ........................................................................................................1-4
NI-IMAQ for IEEE 1394 Cameras Driver Software.......................................1-4
Vision Builder for Automated Inspection.........................................1-5
Chapter 2
Setup and Configuration
Required Items...............................................................................................................2-1
Accessories ....................................................................................................................2-2
Hardware .........................................................................................................2-2
Vision Builder AI Documents.........................................................................2-3
Documents....................................................................................................2-3
Connecting a Camera and Monitor to the CVS-1450 Device.........................2-8
Wiring Power to the CVS-1450 Device ..........................................................2-9
Connecting to a Separate Main Supply ...........................................................2-10
Connecting the CVS-1450 Device to the Development Computer.................2-12
Vision Builder AI: Setting up the Development Computer...........................................2-13
Installing Vision Builder AI and NI-IMAQ for IEEE 1394 Cameras.............2-13
© National Instruments Corporation
vii
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Acquiring an Image in Vision Builder AI..................................................................... 2-15
Development Computer.............................................................................................. 2-15
Installing the LabVIEW Real-Time Module, Vision Development Module,
Configuring the IP Address using the LabVIEW Real-Time Module............ 2-16
Chapter 3
LEDs, DIP Switches, and Connectors
SAFE MODE Switch...................................................................................... 3-4
IEEE 1394....................................................................................................... 3-8
VGA................................................................................................................ 3-8
COM1.............................................................................................................. 3-10
TRIG 0 ............................................................................................................ 3-12
Chapter 4
Digital I/O Functionality
TTL and Isolated Inputs and Outputs............................................................................ 4-1
TTL Inputs and Outputs.................................................................................. 4-2
Isolated Inputs and Outputs............................................................................. 4-2
I/O for Normal Operation ............................................................................... 4-3
Trigger Inputs ................................................................................... 4-3
Timed Pulse Output.......................................................................... 4-4
NI CVS-1450 Series User Manual
viii
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Overheat............................................................................................4-11
Considerations When Connecting the Digital I/O .........................................................4-12
Wiring an Isolated Input to a Sourcing Output Device ...................................4-12
Protecting Inductive Loads..............................................................................4-13
Typical System Setup ....................................................................................................4-15
Chapter 5
Deployment
Connecting Multiple CVS-1450 Devices ......................................................................5-1
Appendix A
Troubleshooting
Appendix B
Specifications
Appendix C
Mounting Information
Appendix D
Technical Support and Professional Services
Glossary
Index
© National Instruments Corporation
ix
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
1
NI CVS-1450 Series Overview
This chapter provides an overview of the features and components of the
National Instruments CVS-1450 Series compact vision system.
About the NI CVS-1450 Series
NI CVS-1450 Series devices are easy-to-use, distributed, real-time imaging
systems that acquire, process, and display images from IEEE 1394 cameras
conforming to the IIDC 1394-based Digital Camera Specification,
Version 1.30. The CVS-1450 Series also provides multiple digital
input/output (I/O) options for communicating with external devices to
configure and start an inspection and to indicate results.
An Ethernet connection between the CVS-1450 device and a development
computer allows you to display measurement results and status information
and to configure the CVS-1450 device settings. When configured, the
CVS-1450 device can run applications without a connection to the
development computer.
Each CVS-1450 device ships with documentation and the NI-IMAQ
for IEEE 1394 Cameras driver software, which is licensed for one
development system and one deployment system.
Hardware Overview
The CVS-1450 device front panel consists of a VGA connector, an RS-232
serial port, a 10/100 Ethernet connector, and three IEEE 1394a ports.
The CVS-1450 device also includes LEDs for communicating system
status, DIP switches that specify startup options, TTL inputs and outputs
for triggering, and isolated inputs and outputs for connecting to external
devices, such as PLCs, sensors, LED indicators, and start/stop buttons. The
isolated inputs and outputs on the CVS-1450 device provide an easy means
for preventing ground loops that could degrade signal integrity.
Caution The isolation on the CVS-1450 device is not safety isolation.
© National Instruments Corporation
1-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
NI CVS-1450 Series Overview
Figure 1-1 shows the CVS-1450 Series front panel.
1
2
NI 1454
Compact Vision System
3
4
11
10
5
9
8
6
7
1
2
3
4
Power LED
Status LED
Isolated Digital Input
TTL Digital Outputs
5
6
7
8
IEEE 1394a Ports
TTL I/O and Isolated I/O 10 RS-232 Serial
Reset Button
DIP Switches
9
VGA
11 RJ-45 Ethernet Port
Figure 1-1. CVS-1450 Series Front Panel
NI CVS-1450 Series User Manual
1-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
NI CVS-1450 Series Overview
Available Camera Bandwidth
The IEEE 1394 bus provides a fixed amount of bandwidth that is shared
provide direct connection for up to three DCAM-compliant IEEE 1394
cameras, depending on the amount of bandwidth each camera requires.
Higher frame rates and larger image sizes require a higher data transfer rate
and use more bandwidth.
Table 1-1 shows the maximum number of cameras the CVS-1450 device
supports for three common video formats. Use this table as a guide when
Note Bandwidth calculations described in Table 1-1 do not include software decoding of
YUV images to RGB. This conversion time is dependent on processor speed.
By triggering your cameras at a frame rate slower than those rates listed in
Table 1-1, you can operate more cameras simultaneously. If the camera
combination exceeds the amount of available bandwidth, the software
returns an Insufficient Resources error.
Table 1-1. Available Camera Bandwidth
Maximum Number
Frames
of Cameras for
Video Format
per Second
Simultaneous Operation
640 × 480
8-bits/pixel mono
30
60
3
1
1
3
1
100
15
640 × 480 YUV
(4:2:2)
16-bits/pixel color
30
1024 × 768
16-bits/pixel mono
7.5
15
2
1
© National Instruments Corporation
1-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
NI CVS-1450 Series Overview
Software Overview
Programming the CVS-1450 device requires NI-IMAQ for IEEE 1394
Cameras 1.5 (or later) driver software to control the hardware and one of
the following application software packages to process images.
•
NI Vision Builder for Automated Inspection (AI) 2.0 or
later—Allows you to configure solutions to common inspection tasks
•
LabVIEW Real-Time Module 7.0 or later with the Vision
Development Module 7.0 or later—Provides customizable control
over acquisition hardware and algorithms
The following sections provide an overview of the driver software and the
application software. For detailed information about individual software
packages, refer to the documentation specific to the package.
NI-IMAQ for IEEE 1394 Cameras Driver Software
The NI-IMAQ for IEEE 1394 Cameras driver software ships with the
CVS-1450 device. NI-IMAQ for IEEE 1394 Cameras is the interface path
between the application software and the CVS-1450 device.
NI-IMAQ for IEEE 1394 Cameras includes an extensive library of VIs
you can call from LabVIEW. These VIs include routines for video
configuration, continuous and single-shot image acquisition, trigger
control, and register-level camera configuration.
The NI-IMAQ for IEEE 1394 Cameras driver software performs all
functions necessary for acquiring and saving images but does not perform
image analysis. For image analysis functionality, refer to the National
Instruments Application Software section.
NI-IMAQ for IEEE 1394 Cameras features both high-level and low-level
functions. A function that acquires images in multi-buffer, single-shot, or
continuous mode is an example of a high-level function. A function that
requires advanced understanding of the CVS-1450 device and image
acquisition, such as configuring an image sequence, is an example of a
low-level function.
The NI-IMAQ for IEEE 1394 Cameras software handles many of the
complex issues between the CVS-1450 device and the camera, such as
IEEE 1394 bus communication and camera control.
NI CVS-1450 Series User Manual
1-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
NI CVS-1450 Series Overview
National Instruments Application Software
This section describes National Instruments application software packages
you can use to analyze and process the images you acquire with the
CVS-1450 device.
Vision Builder for Automated Inspection
NI Vision Builder for Automated Inspection (AI) is configurable machine
vision software that you can use to prototype, benchmark, and deploy
machine vision applications. Creating applications in Vision Builder AI
does not require programming. However, Vision Builder AI applications
are scalable to the LabVIEW Real-Time Module.
Vision Builder AI allows you to easily configure and benchmark a
sequence of visual inspection steps, as well as deploy the visual inspection
system for automated inspection. With Vision Builder AI, you can perform
powerful visual inspection tasks and make decisions based on the results of
individual tasks. You can also migrate your configured inspection to
LabVIEW, extending the capabilities of your applications if necessary.
LabVIEW Real-Time Module with the Vision
Development Module
The LabVIEW Real-Time Module combines LabVIEW graphical
programming with the power of Real-Time (RT) Series hardware, such as
the CVS-1450 Series, enabling you to build deterministic, real-time
systems. You develop VIs in LabVIEW and embed the VIs on RT targets.
The RT target runs VIs without a user interface and offers a stable platform
for real-time VIs.
For more information about the LabVIEW Real-Time Module, refer to the
LabVIEW Help.
The Vision Development Module is an image acquisition, processing, and
analysis library of more than 270 functions for the following common
machine vision tasks:
•
•
•
•
•
Pattern matching
Particle analysis
Gauging
Taking measurements
Grayscale, color, and binary image display
© National Instruments Corporation
1-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
NI CVS-1450 Series Overview
You can use the Vision Development Module functions individually or in
combination. With the Vision Development Module you can acquire,
display, and store images as well as perform image analysis and processing.
Using the Vision Development Module, imaging novices and experts can
program the most basic or complicated image applications without
knowledge of particular algorithm implementations.
NI Vision Assistant, which is included with the Vision Development
Module, is an interactive prototyping tool for machine vision and scientific
imaging developers. With Vision Assistant, you can prototype vision
applications quickly and test how various vision image processing
functions work. Using the Vision Assistant LabVIEW VI creation wizard,
you can create LabVIEW VI block diagrams that perform the prototype
you created in Vision Assistant. You can use them in LabVIEW to add
functionality to the generated VI.
For information about how to use the Vision Development Module with the
LabVIEW Real-Time Module, refer to the IMAQ Vision for LabVIEW User
Manual.
NI CVS-1450 Series User Manual
1-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Setup and Configuration
This chapter provides instructions for setting up the NI CVS-1450 Series
hardware. This chapter also includes instructions for installing software,
configuring an IP address, and acquiring an initial image using the
application software.
Required Items
The following items are necessary for setting up the CVS-1450 device.
Hardware
❑ CVS-1450 device
❑ Ethernet-equipped development computer running
Windows 2000/XP/Me/98
❑ DCAM-compliant IEEE 1394 camera
❑ IEEE 1394 jackscrew-to-latch cable (part number 778796-01) or any
standard IEEE 1394 cable—Facilitates plug-and-play connection from
the CVS-1450 device to up to three 1394 cameras. You can use a 4-pin
to 6-pin converter cable with cameras that have their own external
power supply and do not require power from the 1394 bus.
Note To maintain signal integrity, the IEEE 1394 cable length must be no longer
than 4.5 m.
❑ NI desktop power supply (part number 778794-01) or any
24 VDC 10%, 50 W power supply
–
Power supply cord—Connects the NI desktop power supply to an
outlet. Refer to ni.com for the part number specific to your region
and ordering information.
–
4-position power connector—Required if you do not use the
NI desktop power supply
© National Instruments Corporation
2-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
❑ 10 m 10/100Base-T Ethernet cable (part number
189174-10)—Standard Category 5 (CAT-5) 10/100Base-T Ethernet
cable that connects the CVS-1450 device to a network port. To connect
the CVS-1450 device directly to a local development computer, use an
Ethernet crossover cable.
Note To maintain signal integrity, the Ethernet cable length must be no longer than 100 m.
Software
❑ NI-IMAQ for IEEE 1394 Cameras driver software, which includes the
CVS-1450 Series support software
❑ One of the following application software packages:
–
–
NI Vision Builder for Automated Inspection (AI) 2.0 or later
NI LabVIEW Real-Time Module 7.0 or later with the Vision
Development Module 7.0 or later
Accessories
National Instruments offers the following accessories for use with a
CVS-1450 device.
Hardware
•
•
•
•
VGA-resolution camera (part number 778785-01)
12 mm fixed focal length lens (part number 778789-01)
LED ring light (part number 778787-01)
NI Vision I/O terminal block and prototyping accessory (part number
779166-01)
•
•
DIN rail/panel mount kit (part number 189154-01)
Digital I/O cable and horizontal DIN rail terminal block (part number
778790-01)
•
•
•
Digital I/O cable and vertical DIN rail terminal block (part number
778791-01)
75 Ω SMB 111 coaxial cable (part number 763422-01)—SMB to
BNC cable for connecting to triggers and light sources
10 m Ethernet crossover cable (part number 187375-10)—For
connecting the CVS-1450 device directly to the development
computer
NI CVS-1450 Series User Manual
2-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Documentation
The following documentation resources ship with the CVS-1450 device
and the application software. For more resources, including code examples
and tutorials, visit the NI Developer Zone at ni.com/zone.
Hardware Documents
•
•
•
NI CVS-1450 Series User Manual
NI 1450 Series Compact Vision System Digital I/O Help
NI 1450 Series Compact Vision System Quick Start Guide
Vision Builder AI Documents
•
•
•
NI Vision Builder for Automated Inspection Tutorial
NI Vision Builder for Automated Inspection: Configuration Help
NI Vision Builder for Automated Inspection: Inspection Help
NI-IMAQ for IEEE 1394 Cameras Driver Software Documents
•
•
•
•
Getting Started with NI-IMAQ for IEEE 1394 Cameras
NI-IMAQ for IEEE 1394 Cameras User Manual
NI-IMAQ for IEEE 1394 Cameras VI Reference Help
NI-IMAQ for IEEE 1394 Cameras Help
Vision Development Module and LabVIEW Real-Time Module
Documents
Vision Development Module Documents
•
•
•
•
•
NI Vision Assistant Help
NI Vision Assistant Tutorial
NI Vision Concepts Manual
NI Vision for LabVIEW User Manual
NI Vision for LabVIEW VI Reference Help
LabVIEW Real-Time Module Documents
•
Getting Started With the LabVIEW Real-Time Module
•
LabVIEW Real-Time Module Release Notes
© National Instruments Corporation
2-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
•
•
LabVIEW PID Control Toolset User Manual
LabVIEW Help
Safety Information
Caution The following paragraphs contain important safety information you must follow
when installing and operating the device.
Do not operate the device in a manner not specified in the documentation.
Misuse of the device may result in a hazard and may compromise the safety
protection built into the device. If the device is damaged, turn it off and do
not use it until service-trained personnel can check its safety. If necessary,
return the device to National Instruments for repair.
Keep away from live circuits. Do not remove equipment covers or shields
unless you are trained to do so. If signal wires are connected to the device,
hazardous voltages can exist even when the equipment is turned off. To
avoid a shock hazard, do not perform procedures involving cover or shield
removal unless you are qualified to do so. Disconnect all field power prior
to removing covers or shields.
If the device is rated for use with hazardous voltages (>30 Vrms, 42.4 Vpk,
or 60 Vdc), it may require a safety earth-ground connection wire. Refer to
the device specifications for maximum voltage ratings.
Because of the danger of introducing additional hazards, do not install
unauthorized parts or modify the device. Use the device only with the
chassis, modules, accessories, and cables specified in the installation
instructions. All covers and filler panels must be installed while operating
the device.
Do not operate the device in an explosive atmosphere or where flammable
gases or fumes may be present. Operate the device only at or below the
pollution degree stated in the specifications. Pollution consists of any
foreign matter—solid, liquid, or gas—that may reduce dielectric strength
or surface resistivity. Pollution degrees are listed below:
•
Pollution Degree 1—No pollution or only dry, nonconductive
pollution occurs. The pollution has no effect.
Pollution Degree 2—Normally only nonconductive pollution occurs.
Occasionally, nonconductive pollution becomes conductive because of
condensation.
NI CVS-1450 Series User Manual
2-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
•
Pollution Degree 3—Conductive pollution or dry, nonconductive
pollution occurs. Nonconductive pollution becomes conductive
because of condensation.
Clean the device and accessories by brushing off light dust with a soft,
nonmetallic brush. Remove other contaminants with a stiff, nonmetallic
brush. The unit must be completely dry and free from contaminants before
returning it to service.
You must insulate signal connections for the maximum voltage for which
the device is rated. Do not exceed the maximum ratings for the device.
Remove power from signal lines before connection to or disconnection
from the device.
Caution National Instruments measurement products may be classified as either
Measurement Category I or II. Operate this device at or below the Measurement Category
level specified in the hardware specifications.
Measurement Category1: Measurement circuits are subjected to working
voltages2 and transient stresses (overvoltage) from the circuit to which they
are connected during measurement or test. Installation Category establishes
standardized impulse withstand voltage levels that commonly occur in
electrical distribution systems. The following is a description of Installation
(Measurement3) Categories:
•
Measurement Category I is for measurements performed on circuits
not directly connected to the electrical distribution system referred to
as MAINS4 voltage. This category is for measurements of voltages
from specially protected secondary circuits. Such voltage
measurements include signal levels, special equipment, limited-energy
parts of equipment, circuits powered by regulated low-voltage sources,
and electronics.
•
Measurement Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category
refers to local-level electrical distribution, such as that provided by a
standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe).
1
2
3
4
Measurement Categories as defined in electrical safety standard IEC 61010-1.
Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.
Measurement Category is also referred to as Installation Category.
MAINS is defined as the (hazardous live) electrical supply system to which equipment is designed to be connected for the
purpose of powering the equipment. Suitably rated measuring circuits may be connected to the MAINS for measuring
purposes.
© National Instruments Corporation
2-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Examples of Installation Category II are measurements performed on
household appliances, portable tools, and similar products.
•
Measurement Category III is for measurement performed in the
building installation at the distribution level. This category refers to
measurements on hard-wired equipment such as equipment in fixed
installations, distribution boards, and circuit breakers. Other examples
are wiring, including cables, bus-bars, junction boxes, switches,
socket-outlets in the fixed installation, and stationary motors with
permanent connections to fixed installations.
•
Measurement Category IV is for measurements performed at the
primary electrical supply installation (<1,000 V). Examples include
protection devices and on ripple control units.
Setup Overview
Figure 2-1 illustrates the sequence for setting up and getting started with
the CVS-1450 device.
Set Up the Hardware
Set Up the Development
Set Up the Development
Computer Using LabVIEW
Real-Time with the Vision
Development Module
or
Computer Using
Vision Builder AI
Acquire an Image Using
LabVIEW Real-Time with the
Vision Development Module
Acquire an Image Using
or
Vision Builder AI
Figure 2-1. Setup Sequence
NI CVS-1450 Series User Manual
2-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
•
•
Setting Up the Hardware—This section explains how to connect a
camera, monitor, and power supply to a CVS-1450 device.
Setting Up the Development Computer—This section explains how
to use either Vision Builder AI or the LabVIEW Real-Time Module
with the Vision Development Module to perform the following tasks:
–
–
–
Connect the CVS-1450 device to the development computer
Install application and driver software
Obtain an IP address
•
Acquiring an Image—This section explains how to use either Vision
Builder AI or the LabVIEW Real-Time Module with the Vision
Development Module to acquire an image.
Before Getting Started: Connecting the CVS-1450 Device to a Network
Use a standard CAT-5 or CAT-6 Ethernet cable to connect the CVS-1450
device to an Ethernet network.
If the development computer is already configured on a network, you must
configure the CVS-1450 device on the same network. If the development
computer is not connected to a network, you can connect the two directly
using a CAT-5 or CAT-6 Ethernet crossover cable.
Caution To prevent data loss and to maintain the integrity of your Ethernet installation,
do not use a cable longer than 100 m. If you are using a 100 Mbps Ethernet, National
Instruments recommends using a CAT-5 or CAT-6 shielded twisted-pair Ethernet cable.
Subnet Considerations
To configure the CVS-1450 device, it must reside on the same subnet as the
development computer. Once the CVS-1450 device is configured, other
subnets can access and use it.
To use the CVS-1450 device on a subnet other than the one on which the
development computer resides, first connect and configure the CVS-1450
device on the same subnet as the development computer. Next, physically
move the CVS-1450 device to the other subnet and reassign an IP address.
Contact your network administrator for assistance in setting up the
development computer and CVS-1450 device on the same subnet.
© National Instruments Corporation
2-7
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Setting Up the Hardware
This section describes how to connect the basic hardware components of
the CVS-1450 device. When these basic components are connected for the
first time, the CVS-1450 device runs a program that acquires images. This
program verifies that all hardware components are properly connected and
functioning.
The following items are necessary for hardware setup.
❑ CVS-1450 device
❑ 24 VDC 10%, 50 W power supply
❑ DCAM-compliant IEEE 1394 camera
❑ IEEE 1394 cable
❑ Ethernet cable
❑ Monitor
Before connecting a camera and monitor to the CVS-1450 device, make
sure that all CVS-1450 device DIP switches are in the OFF position.
To connect an IEEE 1394 camera and a monitor to the CVS-1450 device,
refer to Figure 2-2 while completing the following steps:
1. Connect the VGA cable from the monitor to the VGA port on the
CVS-1450 device.
2. Plug the IEEE 1394 cable into one of the IEEE 1394a ports on the
CVS-1450 device. Plug the other end of the cable into the IEEE 1394
port on the camera.
If your camera requires an external power supply, connect it to the
camera, and verify that the camera is powered on.
3. Plug in and power on the monitor.
NI CVS-1450 Series User Manual
2-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
2
1
1
VGA Cable
2
IEEE 1394 Cable
Wiring Power to the CVS-1450 Device
This section describes how to connect the NI desktop power supply.
For instructions on how to connect a separate main supply, refer to the
Connecting to a Separate Main Supply section.
Caution Do not connect the CVS-1450 device main power to a source other than
24 VDC 10%. Do not connect the CVS-1450 device isolated power to a source less than
5 VDC or greater than 30 VDC. Doing so could damage the CVS-1450 device.
To connect power to the CVS-1450 device, refer to Figure 2-3 while
completing the following steps:
1. Plug the 4-position connector from the power supply into the power
receptacle on the CVS-1450 device.
2. Plug the power cord into the power supply.
The CVS-1450 device ships with a factory-installed startup program that,
when the CVS-1450 device is connected to a camera and powered on,
acquires images and displays them on the monitor. If the images from the
camera display on the monitor, continue to the Connecting the CVS-1450
Device to the Development Computer section. If the images from the
© National Instruments Corporation
2-9
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
camera are not displayed on the monitor, refer to Appendix A,
Troubleshooting.
To Outlet
1
2
3
3
1
4-Position Power Connector
2
NI Desktop Power Supply
Power Supply Cord (to Outlet)
Figure 2-3. Wiring Power to the CVS-1450 Device
Connecting to a Separate Main Supply
If you use a power supply other than the NI desktop power supply, follow
the instructions in this section to connect power to the CVS-1450 device.
Note If you are not using the NI desktop power supply, use 0.75 mm2, 18 AWG ferrules
according to manufacturer specifications to terminate the wires leading to the 4-position
power connector, shown in Figure 2-4. Ferrules are available from Phoenix Contact
(part number 3200519).
NI CVS-1450 Series User Manual
2-10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
1
2
3
1
4-Position Power Connector
2
Ferrule
3
Power Supply Wires
Figure 2-4. Wiring a Third-Party Power Supply to the 4-Position Power Connector
Caution Do not connect the CVS-1450 device main power to a source other than
24 VDC 10%. Do not connect the CVS-1450 device isolated power to a source less than
5 VDC or greater than 30 VDC. Doing so could damage the CVS-1450 device.
The CVS-1450 device ships with a 4-position power connector that plugs
directly into the power input connector on the CVS-1450 device. To wire
power to the 4-position connector, complete the following steps:
1. Wire the voltage output of the 24 VDC 10% power supply to the main
voltage input, labeled V, on the 4-position connector.
2. Wire the common-mode signal (ground) output of the power supply
to the common-mode signal input, labeled C, on the 4-position
connector.
If you are using a separate power supply for the CVS-1450 device isolated
outputs, connect the voltage output on the power supply to the isolated
power (Viso) on the 4-position connector. Connect the common-mode
signal (ground) on the power supply to the isolated common-mode signal
(Ciso) on the connector.
Note If you do not require a separate power supply for the CVS-1450 device isolated
outputs, you can daisy-chain the V to the Viso and the C to the Ciso on the connector.
For information about grounding the CVS-1450 device chassis to earth
ground, refer to the Earth Ground Connection section of Chapter 3, LEDs,
DIP Switches, and Connectors.
© National Instruments Corporation
2-11
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Connecting the CVS-1450 Device to the Development Computer
The development computer communicates with the CVS-1450 device over
an Ethernet connection. Use a standard Ethernet cable to connect from the
Note If you are not connecting through a network, use an Ethernet crossover cable to
connect the CVS-1450 device directly to the development computer.
To connect the CVS-1450 device to the development computer, refer to
Figure 2-5 while completing the following steps:
1. Verify that the development computer is connected to the network and
is powered on.
2. Using a standard CAT-5 Ethernet cable, connect from the network port
to the Ethernet port on the CVS-1450 device.
3. Using a standard CAT-5 Ethernet cable, connect from the network port
to the Ethernet port on the development computer.
1
2
1
2
3
Standard Ethernet Cable Connecting from an Ethernet Hub to the Development Computer
Ethernet Hub or Other Network Port
Figure 2-5. Ethernet Connection
NI CVS-1450 Series User Manual
2-12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Vision Builder AI: Setting up the Development Computer
This section describes the sequence for installing Vision Builder AI and the
NI-IMAQ for IEEE 1394 Cameras software on the development computer,
configuring the CVS-1450 device to acquire an image using Vision
Builder AI.
Note This section applies only to Vision Builder AI users. If you are using the LabVIEW
Real-Time Module with the Vision Development Module, refer to the LabVIEW Real-Time
Module with the Vision Development Module: Setting up the Development Computer
section.
The following items are necessary for setting up the development
computer.
❑ Vision Builder AI, version 2.0 or later
❑ NI-IMAQ for IEEE 1394 Cameras driver, version 1.5 or later
Installing Vision Builder AI and NI-IMAQ for IEEE 1394 Cameras
This section describes how to install the Vision Builder AI software and the
NI-IMAQ for IEEE 1394 Cameras driver software onto the development
computer.
Note You must install Vision Builder AI before installing NI-IMAQ for IEEE 1394
Cameras.
Complete the following steps to install Vision Builder AI and NI-IMAQ for
IEEE 1394 Cameras onto the development computer.
1. Insert the Vision Builder AI CD into the CD-ROM drive.
2. When the installation splash screen appears, click Install NI Vision
Builder AI and follow the setup instructions.
3. Insert the NI-IMAQ for IEEE 1394 Cameras CD into the CD-ROM
drive.
Note When installing NI-IMAQ for IEEE 1394 Cameras, make sure to install the support
files for the LabVIEW Real-Time Module.
© National Instruments Corporation
2-13
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
4. When the installation splash screen appears, click Install NI-IMAQ
for IEEE 1394 Cameras, and follow the setup instructions.
5. Restart the development computer.
Configuring the IP Address on the CVS-1450 Device
Complete the following steps to set up an IP address and download
software onto the CVS-1450 device.
1. Launch Vision Builder AI on the development computer by navigating
to Start»Programs»National Instruments»Vision Builder AI.
2. Expand the Execution Target drop-down listbox, and click Select
Network Target. This process takes several seconds.
3. In the Select Remote Target window, click 192.168.10.12 to highlight
the row. This IP address is assigned to all unconfigured CVS-1450
devices.
Tip To uniquely identify unconfigured CVS-1450 devices, connect and configure one
CVS-1450 device at a time.
4. Click Configure to launch the Vision Builder AI Remote Target
Configuration Wizard.
5. In the Identification window, enter a name for the CVS-1450 device
in the Name field and a description of the CVS-1450 device in the
Description field.
Note Device names are limited to 15 characters with no spaces or special characters.
The first and last characters must be alphanumeric.
6. Click Next.
7. If the network is configured to issue IP addresses using DHCP, select
Obtain IP address from DHCP server. Otherwise, set the IP address
manually by selecting Edit IP Settings, Suggest Values, and OK.
8. Click Next. This window shows the status of the software installed on
the CVS-1450 device.
9. Select the Update Target Software checkbox.
10. Click OK to begin configuring the IP address and downloading
software onto the CVS-1450 device. This initialization process takes
several minutes.
NI CVS-1450 Series User Manual
2-14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Acquiring an Image in Vision Builder AI
Complete the following steps to acquire an image in Vision Builder AI.
1. Launch Vision Builder AI on the development computer.
2. In the Vision Builder AI Welcome screen, expand the Execution
Target drop-down listbox and click Select Network Target.
3. Select the CVS-1450 device you configured and click OK.
4. In the Vision Builder AI Welcome screen, click Configure
Inspection.
5. From the Acquire Images palette, click Acquire Image (IEEE 1394).
6. Click the Snap button to acquire a single image, or click the Grab
button to acquire continuous images.
7. When you have configured your acquisition, click OK to add the step.
You can now add inspection steps as documented in the NI Vision
Builder for Automated Inspection Tutorial.
LabVIEW Real-Time Module with the Vision
Development Module: Setting up the Development
Computer
This section describes the sequence for installing the application and driver
software on the development computer, obtaining an IP address, installing
software on the CVS-1450 device, and configuring the CVS-1450 device
to acquire an image using the LabVIEW Real-Time Module.
Note This section applies only to LabVIEW Real-Time Module users. If you are using the
Vision Builder AI software, refer to the Vision Builder AI: Setting up the Development
Computer section.
The following items are necessary for setting up the development
computer.
❑ LabVIEW and LabVIEW Real-Time Module software
❑ Vision Development Module software
❑ NI-IMAQ for IEEE 1394 Cameras driver software
© National Instruments Corporation
2-15
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Installing the LabVIEW Real-Time Module, Vision Development
Module, and NI-IMAQ for IEEE 1394 Cameras
Note You must install LabVIEW, the LabVIEW Real-Time Module, and the Vision
Development Module software before installing the NI-IMAQ for IEEE 1394 Cameras
driver software.
Complete the following steps to install LabVIEW, the LabVIEW
Real-Time Module, the Vision Development Module, and NI-IMAQ for
IEEE 1394 Cameras onto the development computer.
1. Insert the LabVIEW CD into the CD-ROM drive.
2. When the installation splash screen appears, click Install LabVIEW,
and follow the setup instructions.
3. Insert the LabVIEW Real-Time Module CD into the CD-ROM drive.
4. When the installation splash screen appears, click Install LabVIEW
Real-Time Module, and follow the setup instructions.
5. Insert the Vision Development Module CD into the CD-ROM drive.
6. When the splash screen appears, click Install Vision Development
Module, and follow the setup instructions.
7. Insert the NI-IMAQ for IEEE 1394 Cameras CD into the CD-ROM
drive.
Note If you select the custom software installation, make sure to install the support for the
CVS-1450 Series.
8. When the splash screen appears, click Install NI-IMAQ for
IEEE 1394 Cameras, and follow the setup instructions.
9. When prompted, click Yes to restart the development computer.
Configuring the IP Address using the LabVIEW Real-Time Module
To set up an IP address for the CVS-1450 device, complete the following
steps:
1. Open the Measurement & Automation Explorer (MAX) configuration
software by double-clicking the MAX icon on the desktop, or navigate
to MAX by selecting Start»Programs»National Instruments»
2. Expand the Remote Systems branch of the configuration tree, and
click 192.168.10.12 to display the Network Settings window.
This IP address is assigned to all unconfigured CVS-1450 devices.
NI CVS-1450 Series User Manual
2-16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
Tip To uniquely identify unconfigured CVS-1450 devices, connect and configure one
CVS-1450 device at a time.
3. In the Network Settings window, enter a name for the device in the
Name field and a description of the device in the Comment field.
Note Device names are limited to 15 characters with no spaces or special characters.
The first and last characters must be alphanumeric.
4. If the network is configured to issue IP addresses using DHCP, select
Obtain IP address from DHCP server. Otherwise, set the IP address
manually by selecting Edit the IP settings, Suggest Values, and OK.
5. Click Apply.
6. When prompted, click Yes to restart the CVS-1450 device. This
initialization process takes several minutes.
While the CVS-1450 device is restarting, an icon appears next to the device
name to indicate that the CVS-1450 device is disconnected. The MAX
status bar also indicates the connection status of the CVS-1450 device.
Downloading Software onto the CVS-1450 Device
1. In the MAX configuration window, click the Software tab. This
window displays the status of the software on the CVS-1450 device.
2. Click the Install Software button.
3. Select the software to download. For initial installation, make sure all
checkboxes are selected.
4. Click OK.
5. When prompted, click Yes to restart the CVS-1450 device. This
process takes several seconds.
Acquiring an Image Using the LabVIEW Real-Time
Module
Complete the following steps to acquire an image using the LabVIEW
Real-Time Module.
Instruments LabVIEW.
2. Expand the Execution Target drop-down listbox, and click Select
Target with Options.
© National Instruments Corporation
2-17
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Setup and Configuration
3. Enter the new IP address in the Machine Name/IP field, and
click OK.
4. Click the Open button.
5. Navigate to <LabVIEW>\examples\IMAQ.
6. Double-click IMAQ1394 examples.llb and select Grab.vi.
7. Click the Run button to begin acquiring images.
Now that you are acquiring images in LabVIEW, you can use the Vision
Development Module and the installed CVS-1450 device drivers to process
images and control inputs and outputs.
NI CVS-1450 Series User Manual
2-18
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
3
LEDs, DIP Switches, and
Connectors
This chapter provides information about the location and functionality of
the LED indicators, DIP switches, and connectors on the NI CVS-1450
each connector.
LED Indicators
Figure 3-1 shows the location of the POWER OK and STATUS LEDs on
the CVS-1450 device.
STATUS
NI 1454
Compact Vision System
POWER OK
Figure 3-1. POWER OK and STATUS LEDs
Refer to Appendix A, Troubleshooting, for information about
troubleshooting LEDs.
© National Instruments Corporation
3-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
POWER OK LED
Under normal operating conditions, the POWER OK LED remains green
while the CVS-1450 device is powered on. A green POWER OK LED
indicates that CVS-1450 device main power is receiving power and that the
CVS-1450 device is not in a fault state. A red POWER OK LED indicates
that the CVS-1450 device has shut down because of a fault state. A fault
state occurs when the user shutdown input is asserted, the processor
overheats, or the watchdog timer expires.
Note The POWER OK LED does not indicate the status of the isolated power, Viso.
STATUS LED
The orange STATUS LED remains off under normal operating conditions
and flashes a specific number of times to indicate error conditions or certain
DIP switch settings. The STATUS LED remains lit if the CVS-1450 device
detects an internal error.
Refer to the Hardware Errors section of Appendix A, Troubleshooting,
for information about LED error indications.
ACT/LINK LED
100 Mbps LED
The orange ACT/LINK LED blinks when the CVS-1450 device receives
data from or transmits data to the network through the Ethernet connection.
Unrelated network activity causes this LED to blink occasionally even
when the CVS-1450 device is inactive.
Figure 3-2 shows the location of the ACT/LINK LED on the
CVS-1450 device.
The green 100 Mbps LED is lit when the network provides 100 Mbps
support and the CVS-1450 device is communicating at 100 Mbps. If the
100 Mbps LED is not lit, the CVS-1450 device is not operating at
100 Mbps.
Figure 3-2 shows the location of the 100 Mbps LED on the
CVS-1450 device.
NI CVS-1450 Series User Manual
3-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
NI 1454
Compact Vision System
Figure 3-2. ACT/LINK and 100 Mbps LEDs
DIP Switches
This section describes the SAFE MODE, IP RESET, NO APP, and
USER 1 DIP switches on the CVS-1450 device.
To enable a DIP switch, move it to the ON (left) position and then reset the
CVS-1450 device by pressing the RESET button for at least two seconds.
Note You must reset the CVS-1450 device in order for the setting change to occur.
© National Instruments Corporation
3-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Figure 3-3 shows the location of the DIP switches on the CVS-1450 device.
NI 1454
Compact Vision System
ON
Figure 3-3. DIP Switches
SAFE MODE Switch
To start the CVS-1450 device in Safe mode, move the SAFE MODE
switch to the ON position and reset the CVS-1450 device. Use Safe mode
to reconfigure TCP/IP settings and to download or update software from
the development computer.
Downloading incorrect software to the CVS-1450 device may cause it to
hang during restart or become inaccessible over the network. Powering on
or resetting the CVS-1450 device in Safe mode starts the CVS-1450 device
but does not start the embedded LabVIEW RT engine. To resume normal
operations, restart the CVS-1450 device with the SAFE MODE switch in
the OFF position.
NI CVS-1450 Series User Manual
3-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
IP RESET Switch
To clear the CVS-1450 device IP settings, move the IP RESET switch to
the ON position and reset the CVS-1450 device. Use IP RESET to reset
the TCP/IP settings when moving the system from one subnet to another or
when the current TCP/IP settings are invalid.
Resetting the CVS-1450 device with the IP RESET switch in the ON
position resets the IP address to 0.0.0.0. You can then set up a new network
configuration for the CVS-1450 device from a development machine on the
same subnet, or you can use an Ethernet crossover cable to connect the
CVS-1450 device directly to the development computer.
NO APP Switch
To prevent the CVS-1450 device from automatically running VIs at
startup, move the NO APP switch to the ON position and reset the
CVS-1450 device. If the CVS-1450 device becomes inaccessible because
of a startup program, enable the NO APP switch and reset the CVS-1450
device.
Enable this switch to prevent the CVS-1450 device default startup program
or Vision Builder AI from running at startup.
USER 1 Switch (LabVIEW Real-Time Module Users)
The USER 1 switch is user-configurable and has no default functionality.
You can use the RT Read Switch VI to read the USER 1 switch state and
perform a custom action based on the current switch state position.
© National Instruments Corporation
3-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Connectors
This section describes the connectors on the CVS-1450 device and includes
pinouts and signal descriptions for each connector.
Table 3-1 summarizes the functions of the connectors on the CVS-1450
device.
Table 3-1. CVS-1450 Device Connectors Overview
Peripheral
External Connectors
Function
Power
4-position power
connector
Main power and power for isolated outputs
IEEE 1394a
VGA
6-pin IEEE 1394
Power and data connection to IEEE 1394
cameras
15-pin female DSUB
(standard VGA)
Video output
Serial
9-pin male DSUB
(standard RS-232 serial
port)
COM1
10/100 Ethernet
RJ-45 (standard
Ethernet port)
Ethernet network connection
TRIG 0
SMB receptacle
SMB receptacle
External isolated trigger input
External TTL output
TRIG 1 and TRIG 2
Digital Input/Output
44-pin female
External TTL I/O; External isolated I/O
high-density DSUB
Power Connector
The power connector on the CVS-1450 device accommodates two power
supplies. The terminals labeled V and C provide the voltage and
common-mode signal for the main power of the CVS-1450 device. The
terminals labeled Viso and Ciso provide the voltage and common-mode
signal to power the isolated output circuitry.
loops that could introduce noise into the system. This isolation does not provide safety
isolation.
NI CVS-1450 Series User Manual
3-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Figure 3-4 shows the power connector on the CVS-1450 device. Table 3-2
lists and describes each terminal on the power connector.
POWER
POWER
Ciso
Ciso
Viso
Viso
(5-30 VDC)
(5-30 VDC)
C
V
(24 VDC 10%)
C
V
(24 VDC 10%)
Figure 3-4. Power Connector
Table 3-2. Power Connector Terminals
Terminal
Description
Main power (24 VDC 10%)
Common-mode signal
V
C
Viso
Ciso
Isolated power (5 to 30 VDC)
Earth Ground Connection
Some system setups may require using the grounding lug on the CVS-1450
device to connect the chassis to earth ground. Connecting the grounding
lug, shown in Figure 3-5, to earth ground connects the common-mode
signal of the main power to earth ground through the CVS-1450 device
chassis.
Note An earth ground connection does not connect Ciso to earth ground.
© National Instruments Corporation
3-7
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
1
2
POWER
Ciso
Viso
(5-30VDC)
C
V
(24VDC 10%)
1
Grounding Lug
2
Power Connector
Figure 3-5. Grounding Lug on the CVS-1450 Device
IEEE 1394
The IEEE 1394 connectors on the CVS-1450 device provide a reliable,
high-frequency connection between the CVS-1450 device and up to
three DCAM-compliant IEEE 1394 cameras. For information about
the amount of bandwidth available for connecting cameras, refer to the
Available Camera Bandwidth section of Chapter 1, NI CVS-1450 Series
Overview.
To access the IEEE 1394 connectors on the CVS-1450 device, use any
standard 6-pin IEEE 1394 cable.
Note You can use a 4-pin to 6-pin converter cable with cameras that have their own
external power supply and do not require power from the IEEE 1394 bus.
VGA
The VGA connector on the CVS-1450 device provides connection between
the CVS-1450 device and a VGA monitor. Use any standard 15-pin VGA
cable to access the VGA connector. Figure 3-6 shows the location and
pinout of the VGA connector. Table 3-3 lists and describes the VGA
connector signals.
NI CVS-1450 Series User Manual
3-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
NI 1454
Compact Vision System
1
6
11
5
10 15
Figure 3-6. VGA Connector
Table 3-3 lists pin numbers, signal names, and signal descriptions for the
15-pin VGA connector on the CVS-1450 device.
Table 3-3. VGA Connector Signals
Pin
Number
Signal Name
Signal Description
1
2
3
4
5
R
G
Red
Green
Blue
B
NC
C
No Connect
Common-mode signal of the
CVS-1450 device main power
6
7
8
9
C
C
Common-mode signal of the
CVS-1450 device main power
Common-mode signal of the
CVS-1450 device main power
C
Common-mode signal of the
CVS-1450 device main power
+5 V
+5 V
© National Instruments Corporation
3-9
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Table 3-3. VGA Connector Signals (Continued)
Pin
Number
Signal Name
Signal Description
10
C
Common-mode signal of the
CVS-1450 device main power
11
12
13
14
15
NC
SD
No Connect
Serial Data
HSync
VSync
SC
Horizontal Sync
Vertical Sync
Serial Clock
COM1
serial devices, such as PLCs, scanners, and lighting devices.
Note The Serial Port VIs access COM1 as Port 0.
Figure 3-7 shows the location and pinout of the COM1 DSUB 9-pin
connector. Table 3-4 lists and describes the COM1 signals.
NI 1454
Compact Vision System
5 9
1 6
Figure 3-7. COM1 DSUB 9-Pin Connector
NI CVS-1450 Series User Manual
3-10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Table 3-4 lists pin numbers, signal names, and signal descriptions for the
9-pin COM1 connector on the CVS-1450 device.
Table 3-4. COM1 Connector Signals
Pin
Number
Signal Name
DCD
Signal Description
Data Carrier Detect
1
2
3
4
5
RXD
Receive Data
TXD
Transmit Data
DTR
Data Terminal Ready
C
Common-mode signal of the
CVS-1450 device main power
6
7
8
9
DSR
RTS
CTS
RI
Data Set Ready
Ready to Send
Clear to Send
Ring Indicator
Ethernet
The Ethernet port on the CVS-1450 device provides connection between
the CVS-1450 device and the development computer, either directly or
through a network port. The CVS-1450 device automatically detects the
speed of the connection and configures itself accordingly.
If you are connecting the CVS-1450 device to the development computer
through a network port, use a standard Ethernet cable. To connect the
CVS-1450 device directly to the development computer, use an Ethernet
crossover cable.
© National Instruments Corporation
3-11
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
TRIG 0
The TRIG 0 isolated input on the CVS-1450 device provides connection to
external devices, such as proximity sensors and start/stop buttons. For easy
coaxial cable (part number 763422-01).
Note Additional isolated inputs are available on the 44-pin DSUB connector.
Caution These isolated inputs are compatible with 5 V logic if the external circuit meets
the voltage and current requirements listed in Appendix B, Specifications.
TRIG 1 and TRIG 2
You can use the two TTL outputs available on the SMB connectors for
triggering cameras and external interfaces, such as lighting control units.
For easy connection to the TTL outputs, use the National Instruments
SMB 111 coaxial cable (part number 763422-01).
Note Additional TTL outputs are available on the 44-pin DSUB connector.
Caution Do not connect voltage or current sources to TTL outputs. Doing so could
damage the CVS-1450 device.
General-Purpose Digital I/O
The 44-pin DSUB connector, shown in Figure 3-8, provides access to the
general-purpose digital inputs and outputs. The general-purpose digital I/O
12 isolated inputs, and 4 isolated outputs. For easy connection to the digital
I/O connector, use the National Instruments digital I/O cable and terminal
block.
For detailed information about digital I/O functionality and recommended
use cases, refer to Chapter 4, Digital I/O Functionality.
Note Isolated inputs are compatible with 5 V logic if the external circuit meets the voltage
and current requirements listed in Appendix B, Specifications.
NI CVS-1450 Series User Manual
3-12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
NI 1454
Compact Vision System
44 30 15
31 16 1
Figure 3-8. 44-Pin DSUB Connector
Table 3-5 lists pin numbers, signal names, and signal descriptions for the
44-pin connector on the CVS-1450 device and the 37-pin terminal block.
Caution Do not draw more than 500 mA combined from the Viso pins on the 44-pin
DSUB connector. Do not draw more than 100 mA from 24 V or 30 V isolated outputs.
Do not draw more than 50 mA from 5 V isolated outputs.
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals
44-Pin
DSUB on
37-Pin
CVS-1450
Device
Terminal
Block
Pin Number Pin Number
Signal Name
TTL Input 0
C
Primary Function
Alternate Function
1
2
1
3
Pulse generator trigger input General-purpose input
Common-mode signal of the
CVS-1450 device main
power
—
3
4
4
5
TTL Output 0
TTL Output 1
Watchdog timer output
Pulse generator output
General-purpose output
General-purpose output
© National Instruments Corporation
3-13
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals (Continued)
44-Pin
DSUB on
37-Pin
CVS-1450
Device
Terminal
Block
Pin Number Pin Number
Signal Name
Primary Function
Alternate Function
5
6
C
Common-mode signal of the
CVS-1450 device main
power
—
6
7
8
7
8
6
TTL Output 2
TTL Output 3
C
Pulse generator output
Pulse generator output
General-purpose output
General-purpose output
—
Common-mode signal of the
CVS-1450 device main
power
9
NC
17
NC
No connect
—
—
—
10
11
Viso
Ciso
Isolated power
18
Isolated common-mode
signal
12
13
14
19
35
34
ISO Output 0
ISO Output 1
Ciso
General-purpose output
General-purpose output
—
—
—
Isolated common-mode
signal
15
16
17
9
2
3
ISO Input 0
TTL Input 1
C
Input port, Data(0)
—
Pulse generator trigger input General-purpose input
Common-mode signal of the
CVS-1450 device main
power
—
18
19
20
20
21
22
TTL Output 4
TTL Output 5
C
Pulse generator output
General-purpose output
General-purpose output
—
—
Common-mode signal of the
CVS-1450 device main
power
21
22
24
TTL Output 6
TTL Output 7
General-purpose output
General-purpose output
—
—
NI CVS-1450 Series User Manual
3-14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals (Continued)
44-Pin
DSUB on
CVS-1450
Device
37-Pin
Terminal
Block
Pin Number Pin Number
Signal Name
Primary Function
Alternate Function
23
22
C
Common-mode signal of the
CVS-1450 device main
power
—
24
25
26
NC
33
NC
No connect
—
—
—
Viso
Ciso
Isolated power
34
Isolated common-mode
signal
27
28
29
36
37
12
ISO Output 2
ISO Output 3
Ciso
General-purpose output
General-purpose output
—
—
—
Isolated common-mode
signal
30
31
32
33
10
11
13
16
ISO Input 1
ISO Input 2
ISO Input 3
Ciso
Input port, Data(1)
Input port, Data(2)
Input port, Data(3)
—
—
—
—
Isolated common-mode
signal
34
35
14
15
ISO Input 4
ISO Input 5
Input Port, Data(4)
—
Input port latch, Data(5)
Pulse generator trigger
input
36
28
Ciso
Isolated common-mode
signal
—
37
38
39
25
26
28
ISO Input 6
ISO Input 7
Ciso
Quadrature encoder Phase A General-purpose input
Quadrature encoder Phase B General-purpose input
Isolated common-mode
signal
—
40
41
27
29
ISO Input 8
ISO Input 9
General-purpose input
General-purpose input
© National Instruments Corporation
3-15
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
LEDs, DIP Switches, and Connectors
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals (Continued)
44-Pin
DSUB on
37-Pin
CVS-1450
Device
Terminal
Block
Pin Number Pin Number
Signal Name
Primary Function
Alternate Function
42
32
Ciso
Isolated common-mode
signal
—
43
44
30
31
ISO Input 10
ISO Input 11
General-purpose input
User shutdown
General-purpose input
General-purpose input
NI CVS-1450 Series User Manual
3-16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Digital I/O Functionality
This chapter describes the primary functions of the digital inputs and
outputs on the NI CVS-1450 device. This chapter also includes guidelines
for connecting the digital I/O and for setting up a typical CVS-1450 device.
Overview
The digital I/O functions are accessible through 2 TTL inputs, 10 TTL
outputs, 13 isolated inputs, and 4 isolated outputs.
You can use input signals as triggers, product selection ports, or to read
quadrature encoders. Uses for output signals include controlling camera
reset and exposure, controlling strobe lighting, outputting inspection
results, or communicating with PLCs. You can also define the functions of
digital input and output signals.
For information about how to use the LabVIEW Real-Time Module to
implement specific digital I/O functions, refer to the application software
documentation and examples in the following locations:
•
•
Program Files\National Instruments\NI 1450 Series\Docs
<LabVIEW>\examples\NI 1450
The NI 1450 Series Compact Vision System Digital I/O Help is available
for each example. This help file contains digital I/O reference information
and instructions for using the LabVIEW FPGA VIs.
Tip To quickly launch the digital I/O help from a LabVIEW example, press <F1>.
TTL and Isolated Inputs and Outputs
This section describes the TTL and Isolated I/O functions available on the
CVS-1450 device.
© National Instruments Corporation
4-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
TTL Inputs and Outputs
TTL is a fast-switching, 5 V, digital, signaling standard commonly used
TTL inputs and outputs do not require a separate power supply.
Caution Do not connect voltage or current sources to TTL outputs. Doing so could
damage the CVS-1450 device.
Table 4-1 summarizes the TTL inputs and outputs available on the
CVS-1450 device.
Table 4-1. TTL Inputs and Outputs
44-Pin DSUB
on CVS-1450
Device
Pin Number
37-Pin
Terminal Block
Pin Number
Primary
Function
Input or
Output
Number
Available
Signal Names
TTL Input 0
Trigger
Input
2
6
1
16
1
2
TTL Input 1
Timed Pulse
Output
TRIG 1, Pulse 5
TRIG 2, Pulse 6
TTL Output 1, Pulse 1
TTL Output 2, Pulse 2
TTL Output 3, Pulse 3
TTL Output 4, Pulse 4
—
—
4
6
7
—
—
5
7
8
18
20
Watchdog
Output
Output
1
3
TTL Output 0
3
4
General-Purpose
TTL Output 5
TTL Output 6
TTL Output 7
19
21
22
21
23
24
Isolated Inputs and Outputs
The isolated inputs and outputs on the CVS-1450 device have a separate
ground reference from the main CVS-1450 device supply, providing an
easy means to prevent ground loops that can introduce noise into a system.
You can apply signals up to 30 V to the isolated inputs. The voltage swing
of the isolated outputs is determined by the Viso you supply on the
connector.
Note The isolated outputs have current-limiting protection circuitry. If this circuitry is
the output state in the software.
NI CVS-1450 Series User Manual
4-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Table 4-2 summarizes the isolated inputs and outputs available on the
CVS-1450 device.
Table 4-2. Isolated Inputs and Outputs
44-Pin DSUB
on CVS-1450
Device
Pin Number
37-Pin
TerminalBlock
Pin Number
Input or
Output
Number
Available Signal Names
Primary Function
Trigger
Input
3
TRIG 0
ISO Input 5†
ISO Input 8
—
35
40
—
15
27
Quadrature Encoder
Input
Input
Input
1
1
1
ISO Input 6
ISO Input 7
37
38
25
26
External Shutdown
Control
ISO Input 11
44
31
Product Selection Port†
ISO Input 0
ISO Input 1
ISO Input 2
ISO Input 3
ISO Input 4
15
30
31
32
34
9
10
11
13
14
General-Purpose
General-Purpose
Input
2
4
ISO Input 9
ISO Input 10
41
43
29
30
Output
ISO Output 0
ISO Output 1
ISO Output 2
ISO Output 3
12
13
27
28
19
35
36
37
† ISO Input 5 can also function as a latch for the product selection port.
I/O for Normal Operation
The following sections describe I/O functions that are available on the
CVS-1450 device during normal operation.
Trigger Inputs
Trigger inputs are available from both TTL inputs and isolated inputs. You
can use these trigger inputs to synchronize the CVS-1450 device with an
external event, such as the assertion of a signal generated by a proximity
sensor or a PLC to indicate that an inspection item is passing in front of the
camera. The CVS-1450 device uses this input to initiate a timed pulse that
can be used for camera control, lighting control, encoder pulse counting,
and result output timing.
© National Instruments Corporation
4-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
For more information about creating a timed pulse output, refer to the
Timed Pulse Output section.
TTL Input 0, TTL Input 1, TRIG 0, and ISO Input 8 can alternatively
function as general purpose inputs. ISO input 5 can alternatively function
as a latch input for the product selection port.
Timed Pulse Output
The CVS-1450 device is capable of timed pulse output on six different
digital outputs, which provides precise control over time-critical signals,
such as camera exposure. This section describes the various uses for the
timed pulse output and the parameters you can set to control these outputs.
Uses for timed pulse output include controlling camera reset and exposure,
controlling strobe lighting, operating plungers on an assembly line, and
communicating with PLCs. You can configure the start of the pulse output
generation to occur from software or from a rising or falling edge of a
trigger input.
In addition to controlling the timing of pulse output, you can also configure
the polarity of the output signal, resulting in a high-true or low-true signal.
Based on the polarity setting, the output signal asserts after the appropriate
delay time and de-asserts after the configured pulse width. You can set the
delay time in microseconds or in quadrature encoder counts from the start
signal—either a hardware trigger or a software command. Width is always
configured in microseconds.
Initiating a Timed Pulse
Each timed pulse generator has a trigger input that specifies whether to wait
on a particular trigger input to generate the pulse or to immediately
generate the pulse when software sets the pulse mode to Start.
When the pulse generator is configured for a particular trigger input, after
generating a pulse, it waits for another trigger before generating another
pulse. When the pulse generator is configured to immediately generate the
pulse on a software start, after generating a pulse, it immediately generates
another pulse.
If the trigger input is set to Immediate, the pulse generation occurs as soon
as the pulse mode is set to Start. If the trigger input is set to one of the
hardware trigger inputs, the timed pulse output waits for an assertion edge
on the appropriate trigger input. The assertion edge is configurable based
on the trigger polarity parameter. It then generates one pulse and rearms to
NI CVS-1450 Series User Manual
4-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
wait for the next trigger. In either case, the pulse output generation stops
and resets if the pulse mode parameter is set to Stop.
Figure 4-1 shows an output pulse when a trigger is selected.
Trigger
Input
Output
Pulse
Figure 4-1. Output Pulse When Trigger Is Selected
Pulse Modes
Each pulse generator has a Start and a Stop mode. Configure the pulse
generator when in Stop mode and then set it to Start mode.
Pulse Delay
Pulse delay is the amount of time between a trigger and the first (assertion)
edge of an output pulse. The pulse delay is configurable in units of
microseconds or quadrature encoder counts. If configured for
microseconds, available values are between 10 µs and 4,294,967,295 µs,
which is 4,294 seconds, or approximately 71 minutes. If the delay is
configured for quadrature encoder counts, the range of choices is 0 counts
to 4,294,967,295 counts.
Pulse Width
Pulse width is the amount of time between the first (assertion) edge of a
pulse and the second (deassertion) edge. Pulse width is configurable only
in microseconds from 10 µs to 4,294,967,295 µs.
Trigger Polarity
Each pulse generator can be individually configured for rising or falling
edge triggering. Even if multiple pulse generators are using the same
trigger, each can have different polarities.
© National Instruments Corporation
4-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Figure 4-2 shows the output of a pulse generator configured to look for a
rising edge trigger and output a high pulse with a microsecond width and
delay.
Delay
Width
Trigger
Pulse
Figure 4-2. High Pulse with Rising Edge Trigger
Figure 4-3 shows how to create a high and low pulse train with a
microsecond delay and width.
High Pulse Train
Delay
Width
Delay
Width
Low Pulse Train
Software
Start
Figure 4-3. High and Low Pulse Trains
Quadrature Encoder
The quadrature encoder uses ISO Input 6 for its Phase A input and
ISO Input 7 for its Phase B input. Encoder speed is limited by the speed of
the isolated inputs. Each isolated input can change at a maximum rate of
100 kHz, making the maximum encoder rate 400,000 counts/s.
The quadrature encoder can also be used as a timebase for the pulse
generation delay.
NI CVS-1450 Series User Manual
4-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Figure 4-4 shows a rising edge trigger and a low pulse with a quadrature
encoder delay and a microsecond width.
Trigger
Low Pulse
Delay
Width
Phase A
Phase B
Figure 4-4. Rising Edge Trigger, Low Pulse with a Quadrature Encoder Delay
Product Selection Port
The product selection port consists of a group of five isolated digital inputs
that the software running on the CVS-1450 device reads simultaneously.
You can program the CVS-1450 device to switch between up to 32 (25)
inspection sequences for different parts on an assembly line.
Based on the input to the product selection port, you can configure the
application software to run the appropriate inspection sequence. For
example, an upstream CVS-1450 device programmed for part classification
might drive the product selection port of a downstream CVS-1450 device.
Alternatively, a PLC with information about which part is being inspected
can drive the product selection port of the CVS-1450 device.
Using ISO Input 5 as a Latch
You can configure the product selection port to use ISO Input 5 as a latch.
A rising edge on ISO Input 5 can latch the data into a data register on the
CVS-1450 device. Before each inspection, the software checks the status
of the product select inputs and reads the most recent value latched into the
register.
Note In Vision Builder AI, ISO Input 5 is always designated as a latch.
© National Instruments Corporation
4-7
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Table 4-3 lists the product selection ports.
Table 4-3. Product Selection Port
External Connection
ISO Input 5
Function
Data(5), rising edge latch
Data(4)
ISO Input 4
ISO Input 3
ISO Input 2
ISO Input 1
ISO Input 0
Data(3)
Data(2)
Data(1)
Data(0)
General-Purpose I/O
General-purpose inputs and outputs are available as both TTL and isolated
connections. At any time, the software running on the CVS-1450 device
can read the inputs and drive the outputs high or low.
General-Purpose Inputs
The primary difference between general-purpose inputs and trigger inputs
is that you cannot use general-purpose inputs to initiate a timed pulse
generator. In an application, use the general-purpose inputs to get the status
of the inputs at a given point and not to synchronize the CVS-1450 device
with an external event.
An example of how to use general-purpose inputs is reading the status of
a general-purpose input as the first step in your inspection sequence and
recording that value as part of your inspection.
General-Purpose Outputs
The primary difference between general-purpose outputs and timed pulse
outputs is that the timing of general-purpose outputs is controlled by
software rather than hardware. As a result, timing of general-purpose
outputs changes as the inspection algorithm changes, which makes
general-purpose outputs less appropriate than timed outputs for camera
control, strobe light control, and other applications that require precise
NI CVS-1450 Series User Manual
4-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
An example of using general-purpose outputs is driving a relay that turns
on an Inspection in Progress light for an operator to see while the inspection
sequence is running.
I/O for Fault Conditions
The CVS-1450 device recognizes the following fault conditions:
•
•
•
External shutdown, when Shutdown mode is enabled
Watchdog timer expiration
Overheat
The behavior of the CVS-1450 device in the event of a fault condition is
dependent on configuration settings of the SAFE MODE DIP switch and
the software-enabled Shutdown mode. Table 4-4 summarizes how user
configuration affects the behavior of the CVS-1450 device in the event of a
fault condition. To resume operation, address the fault condition and reset
the CVS-1450 device. The following sections describe each fault condition.
Table 4-4. Configuration Settings and Effects
Configuration Settings
Configuration Effects
Outputs Change to
Fault
Condition
Safe Mode
DIP
Shutdown
Enabled
Processing Halts
User-Defined States
External
Shutdown
On
Off
Off
On
Off
Off
On
Off
Off
N/A
On
N/A
Yes
No
No
Yes
No
No
Yes
No
No
Yes
No
Off
N/A
On
Watchdog
Overheat
N/A
Yes
Yes
Yes
Yes
Yes
Off
N/A
On
Off
© National Instruments Corporation
4-9
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Shutdown
Shutdown mode is a software-enabled feature, that when activated, allows
an external device to halt the CVS-1450 device processing operations.
Additionally, Shutdown mode allows you to specify user-defined
shutdown states for all fault conditions.
When Shutdown mode is enabled and the shutdown input signal,
ISO Input 11, turns off, the CVS-1450 device registers an external
shutdown condition. When this fault occurs, the POWER OK LED turns
red and all system operation halts. To resume operation, you must reset the
CVS-1450 device.
For all fault conditions, outputs operate according to user-defined
or 3-state, and each isolated output is on/off configurable.
Shutdown mode is unavailable when the SAFE MODE DIP switch is
turned on.
Refer to Table 4-4 for a summary of how user configuration affects the
shutdown condition.
Disabling Shutdown Mode
To disable Shutdown mode once it is enabled, remove Shutdown mode
from the software script and reset the CVS-1450 device. To reset the
system, press the RESET button on the CVS-1450 device front panel for
at least two seconds.
Note For prototyping when equipment is unavailable, you can wire from Viso to
ISO Input 11 to simulate external equipment that indicates to the CVS-1450 device to
operate normally.
Watchdog Timer
The watchdog timer is a software configurable feature that can monitor
software on the CVS-1450 device and take action if the software is
unresponsive. The millisecond counter on the watchdog timer is
configurable up to 65,534 ms, in 1 ms increments, before it expires.
NI CVS-1450 Series User Manual
4-10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Configure the watchdog timer to take one of the following actions when it
expires:
Caution Use the Indicator Only option only to test the watchdog timer. If software
becomes unresponsive, it cannot be relied upon to send notification to the host.
•
Indicator Only—This option sends the expiration signal back to the
development machine through software. True indicates an expired
watchdog timer. False indicates an unexpired watchdog timer. The
expiration signal that indicates an expired watchdog timer continues to
assert until the watchdog timer is disarmed. Disarming the watchdog
timer resets the I/O.
•
•
TTL Output 0—This option outputs a signal on TTL Output 0. A
High on TTL Output 0 indicates that the watchdog timer has expired.
Low indicates that it has not expired. If the watchdog timer has
timer is disarmed.
System Shutdown—This option halts the CVS-1450 device operation
and turns the POWER OK LED red. If Shutdown mode is enabled, the
outputs go to the user-defined shutdown states.
Table 4-4 summarizes how user configuration affects the shutdown
Overheat
Operating the CVS-1450 device outside of its temperature specifications
may cause the CVS-1450 device to overheat. Refer to Appendix B,
Specifications, for temperature specifications.
In the event of an overheat, all CVS-1450 device operation halts and the
POWER OK LED turns red. If Shutdown mode is enabled, the outputs
go to the user-defined shutdown states.
© National Instruments Corporation
4-11
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Considerations When Connecting the Digital I/O
and optically isolated. Figure 4-6 shows the trigger input interface circuit.
Wiring an Isolated Input to a Sourcing Output Device
You can wire an isolated input to a sourcing output device, as shown in the
example in Figure 4-5. Refer to Appendix B, Specifications, for switching
thresholds and current requirements.
Caution Do not apply a voltage greater than 30 VDC to the isolated inputs. Doing so could
damage CVS-1450 device.
Sourcing
Output
Device
Viso
Vcc
Input
Ciso
Current
Limiter
CVS-1450 device
Figure 4-5. Example of Connecting an Isolated Input to a Sourcing Output Device
Wiring an Isolated Output to an External Load
The digital output circuit sources current to external loads, as shown in
the example in Figure 4-6.
more than 50 mA from 5 V isolated outputs.
NI CVS-1450 Series User Manual
4-12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Viso
Vcc
Digital Output
Load
Ciso
CVS-1450 device
Figure 4-6. Example of Connecting an Isolated Output to an External Load
Protecting Inductive Loads
When an inductive load, such as a relay or solenoid, is connected to an
output, a large counter-electromotive force may occur at switching time
due to energy stored in the inductive load. This flyback voltage can damage
the outputs and the power supply.
To limit flyback voltages at the inductive load, install a flyback diode across
the load. Mount the flyback diode as close to the load as possible. Use this
protection method if you connect any of the isolated outputs on the
CVS-1450 device to an inductive load.
Figure 4-7 shows an example of using an external flyback diode to protect
inductive loads.
© National Instruments Corporation
4-13
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Viso
Vcc
Digital
Output
Load
Ciso
External
Flyback
Diode for
CVS-1450 device
Inductive Loads
Figure 4-7. Example of Using an External Flyback Diode for Inductive Loads
Transmission Line Effects
cause instability. To minimize transmission line effects, use twisted-pair
wires with a characteristic impedance of 118 Ω to connect external signals
to the 44-pin I/O DSUB connector. Use a 75 Ω coaxial cable, such as
RG-179, to connect to the SMB connectors.
Figure 4-8 shows connections to the 44-pin DSUB connector and the
TRIG 0 SMB connector that minimize transmission line effects.
NI CVS-1450 Series User Manual
4-14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
TRIG 1
SMB
RG-179
Coaxial Cable
75 Ω
TRIG 1
Receiving
Equipment
44-Pin
DSUB
118 Ω
TTL OUT(0)
3
2
Receiving
Equipment
+5 V
62 kΩ
RS
TTL IN(0)
16
17
Transmitting
Equipment
CVS-1450 device
Figure 4-8. Example Connections
When connecting to TTL inputs on the CVS-1450 device, match the output
impedance of the transmitting device to the characteristic impedance of the
cable. For example, if the cable characteristic impedance is 118 Ω, make
Rs equal to 118 Ω, as shown in Figure 4-8.
Typical System Setup
Digital I/O configuration varies depending on the number of cameras your
system setup requires. You can access the digital I/O through the 44-pin
DSUB connector or through the three SMB connectors.
The digital I/O capabilities of the SMB connectors on the CVS-1450 device
are ideal for typical single-camera applications. The SMB connectors are
labeled TRIG 0, TRIG 1, and TRIG 2.
© National Instruments Corporation
4-15
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 4
Digital I/O Functionality
Table 4-5 summarizes a typical single-camera configuration.
Table 4-5. Typical Single-Camera System Setup
Signal Type Purpose
Isolated input
Signal Name
TRIG 0
Trigger input from
proximity sensor or
external device
TRIG 1
TRIG 2
Timed pulse TTL output
Timed pulse TTL output
Exposure start and
control signal to camera
Strobe light control
Figure 4-9 shows a typical single-camera setup.
2
1
NI 1454
Compact Vision System
3
4
1
Lighting Control Unit
3
4
Inspection Items
Proximity Sensor
Figure 4-9. Typical Single-Camera Setup
NI CVS-1450 Series User Manual
4-16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Deployment
This chapter provides guidelines for connecting the NI CVS-1450 device
to a network.
Connecting Multiple CVS-1450 Devices
CVS-1450 device allows you to assign an IP address, configure the
CVS-1450 device, download inspection tasks, and remotely monitor an
ongoing inspection.
As with all Ethernet devices, you can connect multiple CVS-1450 devices
to the same network, as shown in Figure 5-1.
Figure 5-1. Multiple CVS-1450 Devices Connected to the Same Network
© National Instruments Corporation
5-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 5
Deployment
To connect multiple CVS-1450 devices to the same network, each
CVS-1450 device must have a unique IP address. By default, the
configuration utility running on the development computer displays each
CVS-1450 device that exists on the subnet. To add entries for CVS-1450
devices that exist on other subnets, you must know the IP address assigned
to each CVS-1450 device. This feature allows remote configuration,
a firewall or user password.
Use the CVS-1450 device serial number and media access control (MAC)
address to uniquely identify each unit. The serial number and MAC address
are printed on the top side of each CVS-1450 device unit, as shown in
Figure 5-2.
Figure 5-2. Serial Number and MAC Address Label on the CVS-1450 Device
The configuration environment on the host machine displays these values
in order to distinguish one CVS-1450 device from another. In addition to
distinguishing units based on serial number and MAC address, you can also
assign each CVS-1450 device a descriptive name in the configuration
environment.
When configured to run an inspection, the CVS-1450 device can run
autonomously without connection to the host machine. All image
acquisition, inspection, decision making, and I/O occurs on the CVS-1450
device itself. For local monitoring of the inspection, you can connect a
monitor directly to the Video Out connector on the CVS-1450 device, as
shown in Figure 5-3.
NI CVS-1450 Series User Manual
5-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 5
Deployment
Figure 5-3. CVS-1450 Device Connected to a Monitor
At any time, you can reconnect the host machine to the CVS-1450 device
and remotely monitor progress.
© National Instruments Corporation
5-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
A
Troubleshooting
This appendix provides instructions for troubleshooting the NI CVS-1450
device.
Software Configuration Problems
CVS-1450 Device Does Not Appear in MAX or in Vision Builder AI
Possible causes and solutions:
•
Verify that there is power to the CVS-1450 device and that both the
CVS-1450 device and the development computer are properly
connected to the network. The POWER OK LED should be lit. When
you click Browse, the ACT/LINK LED on the CVS-1450 device
should flash to indicate that communications are taking place.
•
Certain network devices, such as switches and routers, may filter out
the directed UDP broadcast packets the development computer uses to
communicate with unconfigured CVS-1450 devices, even if these
devices are on the same subnet. You can test this by connecting the
development computer and the CVS-1450 device directly to a network
port or by using a crossover cable between the CVS-1450 device and
the development computer. If the CVS-1450 device appears in the
software using this configuration, but not when connected to the larger
network, the network is not properly communicating with the
CVS-1450 device. Consult your network administrator.
•
•
The CVS-1450 device may have been configured on another network
and then moved to the current network. Reconfigure the CVS-1450
device on the current network.
The grade of cable you are using may be insufficient for the speed of
your network, or you may be using the wrong type of Ethernet cable.
Use a straight-through cable when connecting to network hardware.
Use a crossover cable to connect directly to the development computer.
© National Instruments Corporation
A-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Troubleshooting
•
•
Another device on the network is using the IP address assigned to the
CVS-1450 device. Either remove or reconfigure the other device, or
reconfigure the CVS-1450 device IP address.
The DIP switch settings on the CVS-1450 device may be invalid, such
as all switches set to the ON position. Change the DIP switch settings
and reset the CVS-1450 device by pressing the RESET button on the
CVS-1450 device for at least two seconds.
Error When Configuring IP Settings for the CVS-1450 Device
The CVS-1450 device cannot be configured if the subnet mask or IP
address is incompatible with the network settings. Verify that the values
Status of CVS-1450 Device in MAX and Vision Builder AI is
NoSoftware Installed
Install application and driver software on the CVS-1450 device. Refer to
Chapter 2, Setup and Configuration, for installation instructions.
No Camera Found
Verify that the total power requirements of the cameras do not exceed the
power the CVS-1450 device provides on the IEEE 1394 bus. Verify that the
cameras comply with the IIDC Digital 1394-based Camera Specification,
Version 1.30.
Formatting the CVS-1450 Device Compact Flash Memory
Complete the following steps to format the Compact Flash memory and
restore the CVS-1450 device to its original state.
Caution Formatting the Compact Flash memory overwrites all of the files on the
CVS-1450 device. Back-up any user data files you want to preserve to another computer
before formatting the Compact Flash memory.
1. Start the CVS-1450 device in Safe mode. To start in Safe mode, enable
the SAFE MODE DIP switch on the front of the CVS-1450 device and
restart the device.
2. Launch MAX.
3. In the MAX configuration tree, expand Remote Systems.
NI CVS-1450 Series User Manual
A-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Troubleshooting
4. Right-click the name of the device you want to reformat and select
Format Disk.
of the format.
5. After the format completes, disable the SAFE MODE switch, and
restart the device.
6. Refer to the NI 1450 Series Compact Vision System Quick Start Guide,
or Chapter 2, Setup and Configuration, of this manual for information
about configuring the CVS-1450 device for your application.
Hardware Errors
POWER OK LED is Red
The CVS-1450 device is in one of the following fault conditions:
•
Shutdown enabled in software, but ISO Input 11 is not turned on
–
ISO Input 11 is used to indicate to the CVS-1450 device that all
external devices are ready for operation. When ISO Input 11 turns
off, the NI 1450 shuts down. Either turn on ISO Input 11 with an
external source and reset the CVS-1450 device, or disable
Shutdown in the software.
Note For prototyping when external equipment is unavailable, Viso can be wired to
ISO Input 11 to turn on the CVS-1450 device.
•
Watchdog timer expired while configured for the system shutdown
option
–
Determine if the interval between watchdog timer resets is
insufficient for the processing time. Check the software to
determine if it could monopolize the processor.
•
Processor overheat
–
Make sure you are operating the CVS-1450 device in compliance
with the temperature specifications in Appendix B,
Specifications.
Refer to Table 4-4, Configuration Settings and Effects, for more
information about fault conditions.
© National Instruments Corporation
A-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Troubleshooting
Cannot Drive Isolated Outputs
Verify that you have power connected to Viso and Ciso inputs on the
CVS-1450 device power connector and that the contact at the screw
terminals is solid. Because these outputs are electrically isolated from
the CVS-1450 device main supply, they require power in addition to the
CVS-1450 device main supply.
Runaway Startup Application
If a runaway startup application causes the CVS-1450 device to become
the NO APP switch or the SAFE MODE switch in the ON position.
Restarting the CVS-1450 device with the NO APP switch enabled prevents
any VIs from running at startup. Restarting with the SAFE MODE switch
enabled starts the CVS-1450 device in Safe mode and does not launch the
embedded LabVIEW RT engine. Refer to the DIP Switches section of
Chapter 3, LEDs, DIP Switches, and Connectors, for more information
about these DIP switches.
System Displays a MissingOperatingSystemError at Startup
The CVS-1450 device was powered on or reset with all DIP switches in the
ON position. Set the DIP switches to the off position, and reset the
CVS-1450 device by pressing the RESET button for at least two seconds.
LED Error Indications
The CVS-1450 device indicates specific error conditions by flashing the
orange STATUS LED a specific number of times. Table A-1 describes the
STATUS LED flashing sequences and the corresponding error condition.
NI CVS-1450 Series User Manual
A-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Troubleshooting
Table A-1. STATUS LED Flashes and Corresponding Error Conditions
Number
of Flashes
Error Condition
0 (stays lit)
The CVS-1450 device has detected an internal error.
Contact National Instruments for assistance.
1
2
The CVS-1450 device is unconfigured.
The CVS-1450 device has detected an error in its
software. This usually occurs when an attempt to
upgrade the firmware is interrupted. Repeat the
firmware upgrade process.
3
The CVS-1450 device is in Safe mode.
4 (or more) The CVS-1450 device has detected an unrecoverable
error. Contact National Instruments for assistance.
POWER OK LED Is Not Lit When the CVS-1450 Device
is Powered On
If the power supply is properly connected to the CVS-1450 device, but the
POWER OK LED does not light up, check that the power supply is
24 VDC 10% and within the specifications outlined in Appendix B,
Specifications. Using a power supply that is not within these specifications
might result in an unresponsive or unstable system and could damage the
CVS-1450 device.
© National Instruments Corporation
A-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
B
Specifications
This appendix lists the specifications of the NI CVS-1450 Series. These
specifications are typical at 25 °C, unless otherwise noted.
Power Requirements
Main supply voltage............................... 24 VDC 10%
Power (excluding cameras)............. 12 W, typical
22 W, maximum
IEEE 1394 bus power ..................... 18 W, maximum
(shared by all three ports)
Isolated supply1 ...................................... 5 to 30 VDC
Memory
SDRAM ................................................. 128 MB
Nonvolatile storage
CVS-1454 ....................................... 32 MB
CVS-1455 ....................................... 128 MB
CVS-1456 ....................................... 256 MB
Processor
CVS-1454............................................... Intel Celeron 400 MHz processor
CVS-1455............................................... Intel Celeron 650 MHz processor
CVS-1456............................................... Intel Celeron 733 MHz processor
1
Do not draw more than 500 mA combined from the Viso pins on the 44-pin DSUB connector. Do not draw more than 100 mA
from 24V or 30V isolated outputs. Do not draw more than 50 mA from 5V isolated outputs.
© National Instruments Corporation
B-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Specifications
Network
Network interface ...................................10BaseT and 100BaseTX
Ethernet
Compatibility..........................................IEEE 802.3
Communication rates..............................10 Mbps, 100 Mbps,
auto-negotiated
Maximum cabling distance.....................100 m/segment
TTL Inputs and Outputs
Digital logic levels
Level
Minimum
0 V
Maximum
0.5 V
5 V
Input low voltage (VIL)
Input high voltage (VIH)
2.2 V
—
Output low voltage (VOL), at 5 mA
Output high voltage (VOH), at 5 mA
0.4 V
—
2.4 V
TTL Inputs
Number of channels................................2
Maximum pulse rate ...............................2 MHz
Minimum pulse detected ........................500 ns
Power-on state ........................................Input (high-impedance)
61.9 kΩ pull-up to 5 V
TTL Outputs
Number of channels................................10
Output voltage range ..............................0 V to 5 V
Maximum pulse rate ...............................2 MHz
NI CVS-1450 Series User Manual
B-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Specifications
Optically Isolated Inputs and Outputs
Isolated (Current Sinking) Inputs
Number of channels ............................... 13
Input voltage range................................. 0 V to 30 V
Input ON voltage ............................ 3.5 V to 30 V
Input OFF voltage........................... 0 V to 2 V
Turn-on current ...................................... 7.1 mA, typical
14 mA, maximum
Maximum pulse rate............................... 100 kHz
Minimum pulse detected........................ 10 µs
Reverse polarity protection .................... Yes, –30 V
Isolated (Current Sourcing) Outputs
Number of channels ............................... 4
On-state voltage range ........................... 5 V to 30 V, maximum
Maximum on-state voltage
drop from Viso................................ 1.2 V at 100 mA
Output current
5 V isopower................................... 50 mA, maximum
24 V isopower................................. 100 mA, maximum
30 V isopower................................. 100 mA, maximum
Maximum pulse rate............................... 10 kHz (maximum load
resistance 100 kΩ)
Minimum pulse generated...................... 100 µs
Reverse polarity protection .................... Yes
© National Instruments Corporation
B-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Specifications
IEEE 1394
Number of ports......................................3
Camera Interface.....................................IEEE 1394a
Speed ......................................................100, 200, or 400 Mbps
Available bus power ...............................Refer to the Power Requirements
section
Physical Characteristics
Dimensions .............................................102 mm × 127 mm × 64 mm
(4 in. × 5 in. × 2.5 in.)
Weight ....................................................1 kg (2.2 lb)
Refer to Appendix C, Mounting Information, for dimensional drawings of
the CVS-1450 device.
Environmental
The CVS-1450 Series is intended for indoor use only.
Operating temperature
Vertical mounting position..............0 °C to 55 °C
All other positions ...........................0 °C to 45 °C
Maximum altitude...................................2,000 m
Storage temperature................................–20 °C to 85 °C
Humidity.................................................10% to 90% RH, noncondensing
Pollution Degree.....................................2
Operating shock (IEC 68-2-27) ..............50 g, 6 ms half sine,
3 shocks per side;
30 g 11 ms half sine,
3 shocks per side
Operating vibration
Random (IEC 60068-2-34)..............10 Hz to 500 Hz, 5 Grms
Swept Sine (IEC 60068-2-6) ...........10 Hz to 500 Hz, 5 g
NI CVS-1450 Series User Manual
B-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Specifications
Safety
The CVS-1450 Series meets the requirements of the following standards
for safety and electrical equipment for measurement, control, and
laboratory use:
•
•
•
EN 61010-1, IEC 61010-1
UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Note For UL and other safety certifications, refer to the product label or visit
ni.com/certification, search by model number or product line, and click the
appropriate link the Certification column.
Electromagnetic Compatibility
Emissions ............................................... EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity................................................ EN 61326:1997 + A2:2001,
Table 1
EMC/EMI............................................... CE, C-Tick, and FCC Part 15
(Class A) Compliant
Note For full EMC compliance, operate this device with shielded cabling.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
Low-Voltage Directive (safety) ............. 73/23/EEC
Electromagnetic Compatibility
Directive (EMC) .................................... 89/336/EEC
Note Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, visit
ni.com/certification, search by model number or product line, and click the
appropriate link in the Certification column.
© National Instruments Corporation
B-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
C
Mounting Information
This appendix provides the information necessary to create a custom mount
for the NI CVS-1450 Series.
If you do not want to create a custom mount, a panel and DIN rail mount
number 189154-01).
Mounting the CVS-1450 Series
If you want to create a custom mount, Figures C-1–C-4 provide
dimensional drawings and clearance information for the CVS-1450 Series.
Caution If you choose not to mount the CVS-1450 device on a DIN rail or flat surface, do
not position the CVS-1450 device with the heat sinks resting on any surface. Doing so may
cause the CVS-1450 device to overheat. Refer to Appendix B, Specifications, for
temperature specifications.
4X 4-40 UNC
Max Depth 5.08 mm (0.20 in.)
Mounting Holes for DIN Rail
and Panel Mount Kit
49.28 mm
(1.940 in.)
19.74 mm
(0.777 in.)
63.42 mm
(2.497 in.)
24.64 mm
(0.97 in.)
1.52 mm
(0.06 in.)
20.32 mm
(0.80 in.)
Figure C-1. Back View of the CVS-1450 Series with Dimensions
© National Instruments Corporation
C-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C
Mounting Information
127.25 mm
(5.010 in.)
103.12 mm
(4.060 in.)
107.89 mm
(4.248 in.)
112.09 mm
(4.413 in.)
Figure C-2. Side View of the CVS-1450 Series with Dimensions
6-32 UNC
Grounding Lug
19.79 mm
(0.779 in.)
62.56 mm
(2.463 in.)
Figure C-3. Bottom View of the CVS-1450 Series with Dimensions
Note You can ground the CVS-1450 device to your mount by connecting a grounding wire
to the grounding lug on the CVS-1450 device. Refer to the Earth Ground Connection
section of Chapter 3, LEDs, DIP Switches, and Connectors, for more information about the
grounding lug.
NI CVS-1450 Series User Manual
C-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C
Mounting Information
Caution Your installation must meet the following space and cabling clearance
•
Allow 76.2 mm (3.0 in.) on the top and bottom of the CVS-1450 device for air
circulation.
•
•
Allow 50.8 mm (2.0 in.) on the sides of the CVS-1450 device for air circulation.
Allow enough space in front of the CVS-1450 device to connect cables.
Refer to Figure C-4 for a illustration of the space and cabling requirements for the
CVS-1450 Series.
76.2 mm
(3.00 in.)
50.8 mm
(2.00 in.)
50.8 mm
(2.00 in.)
76.2 mm
(3.00 in.)
© National Instruments Corporation
C-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C
Mounting Information
Complete the following steps to secure the CVS-1450 device to your
mount:
1. Align the screw holes of your mounting bracket with the four holes on
the back of the CVS-1450 device. Figure C-1 shows the location of the
mounting holes on the CVS-1450.
2. Insert four 4-40 UNC screws and tighten them until they are secure
(3.5 in.-lb). Make sure that the heads of the screws are flush with the
mounting bracket.
NI CVS-1450 Series User Manual
C-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
D
Technical Support and
Professional Services
Visit the following sections of the National Instruments Web site at
ni.com for technical support and professional services:
•
Support—Online technical support resources at ni.com/support
include the following:
–
Self-Help Resources—For answers and solutions, visit the
award-winning National Instruments Web site for software drivers
and updates, a searchable KnowledgeBase, product manuals,
step-by-step troubleshooting wizards, thousands of example
programs, tutorials, application notes, instrument drivers, and
so on.
–
Free Technical Support—All registered users receive free Basic
Service, which includes access to hundreds of Application
Engineers worldwide in the NI Developer Exchange at
ni.com/exchange. National Instruments Application Engineers
make sure every question receives an answer.
For information about other technical support options in your
area, visit ni.com/services or contact your local office at
ni.com/contact.
•
•
•
Training and Certification—Visit ni.com/training for
self-paced training, eLearning virtual classrooms, interactive CDs,
and Certification program information. You also can register for
instructor-led, hands-on courses at locations around the world.
System Integration—If you have time constraints, limited in-house
technical resources, or other project challenges, National Instruments
Alliance Partner members can help. To learn more, call your local
NI office or visit ni.com/alliance.
Declaration of Conformity (DoC)—A DoC is our claim of
compliance with the Council of the European Communities using
the manufacturer’s declaration of conformity. This system affords
the user protection for electronic compatibility (EMC) and product
safety. You can obtain the DoC for your product by visiting
ni.com/certification.
© National Instruments Corporation
D-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix D
Technical Support and Professional Services
•
Calibration Certificate—If your product supports calibration,
you can obtain the calibration certificate for your product at
ni.com/calibration.
If you searched ni.com and could not find the answers you need, contact
your local office or NI corporate headquarters. Phone numbers for our
worldwide offices are listed at the front of this manual. You also can visit
the Worldwide Offices section of ni.com/niglobal to access the branch
office Web sites, which provide up-to-date contact information, support
phone numbers, email addresses, and current events.
NI CVS-1450 Series User Manual
D-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
Symbol
Prefix
micro
milli
Value
10–6
10–3
103
µ
m
k
kilo
M
mega
106
B
buffer
Temporary storage for acquired data.
D
DCAM-compliant
Cameras that comply with the IIDC 1394-based Digital Camera
Specification, Version 1.30.
DIP switch
Dual Inline Package switch.
E
Ethernet cable,
standard
CAT-5 or CAT-6 Ethernet cable used to connect a Real-Time target to a
network port.
Ethernet crossover
cable
CAT-5 or CAT-6 cable used for direct connection between the development
computer and the Real-Time target.
F
falling edge
The digital signal transition from the high state to the low state.
I
IEEE
Institute of Electrical and Electronics Engineers. A standard-setting body.
© National Instruments Corporation
G-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
L
LED
Light-emitting diode.
M
MAC
Media access control. The MAC address uniquely identifies each unit
connected to a network.
N
NI-IMAQ
Driver software for National Instruments image acquisition (IMAQ)
hardware.
P
PLC
Programmable Logic Controller. An industrial computer used for factory
automation, process control, and manufacturing systems.
pulse train
A signal consisting of a series of continuous pulses.
R
rising edge
The digital signal transition from the low state to the high state.
Standard electrical interface for serial data communications.
RS-232
S
subnet
A set of systems whose IP addresses are configured such that they can
communicate directly with one another. Data will not flow through an
intermediate router.
T
TCP
Transmission Control Protocol. A set of standard protocols for
communicating across a single network or interconnected set of networks.
TCP is for high-reliability transmissions.
trigger
Any event that causes or starts some form of data capture.
NI CVS-1450 Series User Manual
G-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
U
UDP
User Datagram Protocol. A set of standard protocols for communicating
across a single network or interconnected set of networks. UDP is for
low-overhead transmissions.
V
VDC
Volts direct current.
VI
Virtual Instrument. A combination of hardware and/or software elements,
typically used with a PC, that has the functionality of a classic stand-alone
instrument.
© National Instruments Corporation
G-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
SMB, 3-6
TRIG 0, 3-12
Numerics
100 Mbps LED, 3-2
A
conventions used in the manual, v
CVS-1450 hardware
connecting multiple CVS-1450 systems
(figure), 5-1
accessories, 2-2
application software
installing LabVIEW, 2-16
installing Vision Builder AI, 2-13
installing Vision Development
Module, 2-15
connecting to development computer, 2-12
connectors (table), 3-6
D
B
DCAM specification, 1-1, G-1
Declaration of Conformity (NI resources), D-1
deployment, 5-1
bandwidth, available (table), 1-3
C
development computer
connecting to CVS-1450, 2-13
installing application software, 2-13
installing LabVIEW, 2-16
installing Vision Development
Module, 2-15
cables
crossover, 2-2
digital I/O, 2-2
Ethernet crossover, 2-2
Ethernet standard, 2-2
SMB to BNC, 2-2
diagnostic tools (NI resources), D-1
digital camera specification, 1-1
characteristic impedance, 4-15
calibration certificate (NI resources), D-2
camera
44-pin DSUB, 3-12
connector pinout (figure), 3-13
connector signals (table), 3-13
examples (LabVIEW Real-Time
Module), 4-1
available bandwidth, 1-3
connecting to the CVS-1450, 2-8
digital camera specification, 1-1
video formats, 1-3
COM1, connector signals, 3-11
connecting multiple systems, subnet, 5-2
connectors, 3-6
external load, wiring, 4-12
general-purpose, 3-12
COM1, 3-10
Ethernet, 3-11
IEEE 1394, 3-8
© National Instruments Corporation
I-1
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
isolated I/O, 4-2
table, 4-3
protecting inductive loads, 4-13
signal functionality, 4-1
sourcing output device, wiring
(figure), 4-12
transmission line effects, 4-14
TTL, 4-2
accessories, 2-2
connecting a camera, 2-8
connecting the CVS-1450 to the
development computer, 2-12
overview, 2-6
DIP switches
IP RESET, 3-5
location, 3-4
NO APP, 3-5
overview, 3-3
SAFE MODE, 3-4
required items, 2-1
subnets, 2-7
documentation
conventions used in the manual, v
NI resources, D-1
help, technical support, D-1
drivers (NI resources), D-1
IEEE 1394
E
cable, 2-1
connector, 3-8
licensing, 1-1
electromagnetic compatibility
specifications, B-5
environmental specifications, B-4
error indications (table), A-4
Ethernet
IEEE 1394 bus, available power, B-1
IEEE 1394 camera
connecting to CVS-1450, 2-8
digital camera specification, 1-1
IEEE 1394a (table), 3-6
image acquisition
LabVIEW Real-Time Module, 2-17
Vision Builder AI, 2-15
inductive loads, protecting, 4-13
inputs
connecting CVS-1450 (figure), 2-12
connector, 3-11
crossover cable, 2-2
LINK LED, 3-2
port (table), 3-6
examples (NI resources), D-1
external load, wiring, 4-12
general-purpose, 4-8
TTL, 4-2
F
instrument drivers (NI resources), D-1
IP address
flyback diode, inductive loads (figure), 4-14
flyback voltage, flyback diode, 4-13
configuring with Vision Builder AI, 2-14
configuring with Vision Development
Module, 2-16
IP RESET DIP switch, 3-5
NI CVS-1450 Series User Manual
I-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
isolated input, sourcing output device, 4-12
isolated output, external load, 4-12
IEEE 1394 connector, 3-8
optional hardware, 2-2
overview, 1-1
parts locator, 1-2
K
NI support and services, D-1
NI-IMAQ for IEEE 1394 Cameras
driver software, 1-4
KnowledgeBase, D-1
installing, 2-13
L
label, MAC address, serial number, 5-2
LabVIEW Real-Time Module
acquiring images, 2-17
description, 1-5
outputs
installing, 2-15
LabVIEW, installing, 2-16
latch, 4-7
general-purpose, 4-8
TTL, 4-2
LEDs
figure, 3-1
POWER OK, 3-2
STATUS, 3-2
parts locator, 1-2
physical characteristic specifications, B-4
POWER OK LED, 3-2
power requirements specifications, B-1
power supply
M
MAC address, 5-2
connector, 3-6, 3-7
table, 3-6
MAX, 2-16
Measurement & Automation Explorer
(MAX), 2-16
desktop, 2-1
earth ground, 3-7
media access control address, 5-2
memory, specifications, B-1
mounting kits, part numbers, 2-2
separate main supply, 2-10
terminals (table), 3-7
wiring power, 2-9
figure, 2-10
product selection port, 4-7
table, 4-8
programming examples (NI resources), D-1
pulse modes, 4-5
pulse width, 4-5
N
National Instruments support and
services, D-1
network
connecting, 2-7
specifications, B-2
subnet, 2-7
NI 1450 hardware
DIP switches, overview, 3-3
Q
quadrature encoder, 4-6
© National Instruments Corporation
I-3
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
R
required hardware, 2-1
T
S
technical support, D-1
terminal block, 37-pin (table), 3-13
timed pulse output, initiating, 4-4
training and certification (NI resources), D-1
transmission line effects, 4-14
SAFE MODE DIP switch, 3-4
safety specifications, B-5
serial number, 5-2
serial port, connector (table), 3-6
setup
multiple CVS-1450s, 5-1
typical single-camera, 4-15
setup and configuration, 2-1
shutdown, disabling, 4-10
software
cable, 3-12
connector (table), 3-6
cable, 3-12
connector (table), 3-6
application, 1-4, 1-5
installing on development computer, 2-13
NI resources, D-1
cable, 3-12
connector (table), 3-6
trigger input lines, 4-3
troubleshooting
NI-IMAQ for IEEE 1394 Cameras, 1-4
software choices, 1-4
sourcing output device
wiring, 4-12
hardware, A-3
NI resources, D-1
software, A-1
STATUS LED error indications
(table), A-4
figure, 4-12
specifications
electromagnetic compatibility, B-5
environmental, B-4
memory, B-1
network, B-2
optically isolated
inputs, B-3
outputs, B-3
physical characteristics, B-4
power requirements, B-1
safety, B-5
connector
figure, 3-9
table, 3-6
connector signals (table), 3-9
video formats, 1-3
Vision Builder AI
configuring IP address, 2-14
description, 1-5
image acquisition, 2-15
installing, 2-13
TTL
inputs, B-2
outputs, B-2
error indications (table), A-4
NI CVS-1450 Series User Manual
I-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
Vision Development Module
configuring IP address, 2-16
description, 1-5
image acquisition, 2-17
installing, 2-15
W
Web resources, D-1
© National Instruments Corporation
I-5
NI CVS-1450 Series User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
|