National Instruments Graphics Tablet NI 5102 User Manual

USER GUIDE  
Connecting Your Calibrator, DMM, and NI 5102........................... 6  
Running the Calibration Executive Procedure................................. 6  
Viewing the Calibration Report....................................................... 7  
Technical Support Resources.................................................................. 8  
NI Web Support............................................................................... 8  
Worldwide Support.......................................................................... 8  
National Instruments, NI, and ni.comare trademarks of National Instruments Corporation. Product and company names mentioned herein are  
trademarks or trade names of their respective companies.  
323070A-01  
Copyright © 2001 National Instruments Corp. All rights reserved.  
April 2001  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Equipment and Other Test Requirements  
This section describes the equipment, documentation, software, and test  
conditions needed for calibration.  
Test Equipment  
Tables 1 and 2 list specifications for equipment you can use to calibrate  
your NI 5102 with Calibration Executive.  
Recommended Equipment  
Table 1 lists the equipment that National Instruments recommends for this  
procedure.  
Table 1. Required Equipment Specifications for NI 5102 Calibration  
Required  
Equipment  
Recommended  
Equipment  
Parameter  
Measured  
Necessary Specifications  
Vertical Offset  
Vertical Gain  
0 VDC 0.1 mV  
DC 40 mV to 4.5 V, 0.1%  
into 1 M Ω  
AC Coupling  
Bandwidth  
sine wave 9–13 Hz 100 ppm,  
1.8 Vpp 2% into 1 MΩ  
Scope Calibrator  
Wavetek 9500  
Scope Calibrator  
1.5% amplitude flatness  
for leveled sine wave  
100 kHz–15 MHz 50 ppm,  
0.2–1.9 Vpp 2% into 1 MΩ  
Input Impedance  
2-wire resistance accuracy of  
0.25% for 1 Mmeasurement  
Active Head  
Wavetek 9510  
Active Head or  
better  
If you do not have the recommended instruments, use the accuracy  
requirements listed above to select a substitute calibration standard.  
Refer to Table 2 for a list of alternative instruments you can use for the  
calibration procedure. Although these instruments are acceptable,  
National Instruments recommends that you use the instruments from  
Table 1.  
© National Instruments Corporation  
3
Calibrating the NI 5102 with Calibration Executive  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Alternative Equipment  
Table 2 contains a list of alternative equipment you can use to calibrate  
your NI 5102 with Calibration Executive.  
Table 2. Alternative Equipment Specifications for NI 5102 Calibration  
Required  
Equipment  
Recommended  
Equipment  
Parameter  
Measured  
Necessary Specifications  
Vertical Offset  
Vertical Gain  
0 VDC 0.1 mV  
Calibrator  
Fluke 5700  
Calibrator  
DC 40 mV to 4.5 V, 0.1%  
into 1 MΩ  
AC Coupling  
Bandwidth  
sine wave 913 Hz 100 ppm,  
1.8 Vpp 2% into 1 MΩ  
High-Speed Signal  
Generator  
Tektronix SG503  
1.5% amplitude flatness  
for leveled sine wave  
100 kHz15 MHz 50 ppm,  
0.21.9 Vpp 2% into 1 MΩ  
DMM  
HP 34401  
Input Impedance  
2-wire resistance accuracy of  
0.25% for 1 Mmeasurement  
Connectors  
If you are using the Wavetek 9500 and the Active Head, you do not need  
special connectors.  
If you are using the alternative instruments listed in Table 2, you need the  
following connection equipment:  
50 BNC cable  
50 BNC terminator  
BNC T-connector  
Documentation  
This section describes the documentation you need to calibrate your 5102  
device. In addition to this calibration procedure, you may need to refer to  
the following documents:  
NI-SCOPE Instrument Driver Quick Reference Guide  
Where to Start with Your NI 5102 Oscilloscope  
NI 5102 User Manual  
You can download these documents from the National Instruments Web site  
at ni.com/manuals  
Calibrating the NI 5102 with Calibration Executive  
4
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Software  
Complete the following steps to install Calibration Executive:  
1. Make sure that your computer and monitor are powered on and  
that you have installed Windows 2000/NT/Me/9x.  
2. Close all open applications.  
3. Insert the installation CD into the CD-ROM drive.  
4. Choose the Run option from the Start menu on the desktop task bar.  
5. In the command line box, type x:\setup.exe(where x is the letter  
of the CD-ROM drive you are using), and click OK.  
6. Follow the instructions that appear in the dialog boxes.  
The setup program installs Calibration Executive as well as  
the associated files listed in Table 3.  
Table 3. Calibration Executive Directories and Supporting Files  
Directory Name  
Contents  
CalExec\Procedures\NI 5102\Limits  
CalExec\Procedures\NI 5102\Support Files  
CalExec\Procedures\NI 5102\Support Files  
Test Conditions  
Microsoft Access database that  
stores the calibration limits.  
Directory structure that contains the  
calibration procedures.  
Required support files.  
Follow these guidelines to optimize the connections and the environment  
during calibration:  
Keep connections to the NI 5102 short. Long cables and wires act as  
antennae, picking up extra noise that can affect measurements.  
Use a 50 BNC coaxial cable for all connections to the device.  
Keep relative humidity between 10 and 90%, noncondensing,  
or consult your device hardware manual for the optimum relative  
humidity.  
Maintain the temperature at 25 °C.  
Allow a warm-up time of at least 15 minutes to ensure that the  
measurement circuitry of the 5102 is at a stable operating temperature.  
© National Instruments Corporation  
5
Calibrating the NI 5102 with Calibration Executive  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Procedures  
This section explains how to set up and run the Calibration Executive  
procedure.  
Using the Wavetek 9500 in automated mode, the calibration procedure  
takes approximately ten minutes. In manual mode, the procedure takes  
approximately 30 minutes.  
Using the alternative instruments in automated mode, the calibration  
procedure takes approximately 20 minutes. In manual mode, the procedure  
takes approximately 40 minutes.  
Connecting Your Calibrator, DMM, and NI 5102  
The calibration procedure steps you through the connections between the  
calibrator, function generator, DMM, and NI 5102.  
Note You do not need a function generator or a DMM if you are using the Wavetek 9500.  
Running the Calibration Executive Procedure  
To run the Calibration Executive calibration procedure, complete the  
following steps:  
1. Launch Calibration Executive and follow the instructions in the in the  
Calibration Configuration Wizard to load the NI 5102 calibration  
procedure. Refer to Chapter 1, Introduction to Calibration Executive,  
in the Calibration Executive Software User Manual if you need more  
information on configuring and loading a calibration procedure.  
2. When prompted by Calibration Executive, enter information about the  
installed hardware, such as the calibrator, function generator, and  
DMM.  
Note If you are using the Wavetek 9500, you will not need a function generator or a  
DMM, so you can leave noneselected in the list boxes for Selected Generator  
Information, Selected DMM Information, and Available Addresses. You must also  
enter N/A in the appropriate fields for the tracking numbers and calibration due dates  
before you can proceed. The calibration report does not include any of this extra  
information if you are using the Wavetek 9500.  
3. When the procedure is loaded, click Run Procedure to begin. The  
procedure prompts you to enter the device number, which is the  
number assigned by Measurement & Automation Explorer (MAX).  
4. Follow any instructions you receive from Calibration Executive.  
Calibrating the NI 5102 with Calibration Executive  
6
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Viewing the Calibration Report  
When the procedure is finished, complete the following steps to view the  
calibration report:  
1. Click View»Reports.  
2. Select View to launch your browser and view your report. The  
calibration report appears as an printable HTML file.  
You have completed calibrating your NI 5102 with Calibration Executive.  
Note If your NI 5102 device fails after calibration, return the device to National  
Instruments for repair or replacement.  
© National Instruments Corporation  
7
Calibrating the NI 5102 with Calibration Executive  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Technical Support Resources  
NI Web Support  
National Instruments Web support is your first stop for help in solving  
installation, configuration, and application problems and questions. Online  
problem-solving and diagnostic resources include frequently asked  
questions, knowledge bases, product-specific troubleshooting wizards,  
manuals, drivers, software updates, and more. Web support is available  
through the Technical Support section of ni.com  
Worldwide Support  
National Instruments has offices located around the world to help address  
your support needs. You can access our branch office Web sites from the  
Worldwide Offices section of ni.com. Branch office Web sites provide  
up-to-date contact information, support phone numbers, e-mail addresses,  
and current events.  
If you have searched the technical support resources on our Web site  
and still cannot find the answers you need, contact your local office or  
National Instruments corporate. For telephone support in the United States,  
dial 512 795 8248. For telephone support outside the United States, contact  
your local branch office:  
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,  
Brazil 011 284 5011, Canada (Calgary) 403 274 9391,  
Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521,  
Canada (Toronto) 905 785 0085, China (Shanghai) 021 6555 7838,  
China (ShenZhen) 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11,  
France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427,  
Hong Kong 2645 3186, India 91805275406, Israel 03 6120092, Italy 02 413091,  
Japan 03 5472 2970, Korea 02 596 7456, Malaysia 603 9596711,  
Mexico 5 280 7625, Netherlands 0348 433466, New Zealand 09 914 0488,  
Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011,  
Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00,  
Switzerland 056 200 51 51, Taiwan 02 2528 7227,  
United Kingdom 01635 523545  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     

Miele Ventilation Hood 09 885 260 User Manual
Multiquip Pole Saw SP213H20 User Manual
Newcon Optik Binoculars NV66 G2 5x User Manual
Nikon Camera Lens 2129 User Manual
Oregon Scientific Thermometer RAR188A BK User Manual
Palsonic Stereo System PSML 826 User Manual
Panasonic Computer Monitor BT S950P User Manual
Panasonic DVR DMR EH65 User Manual
Panasonic Power Supply M9RZ90GB4LG User Manual
Panasonic Security Camera AK HRP900P User Manual