The HOLD function is activated by  
keeping pressed the HOLD key.  
The measured value is frozen on  
the display and the “HOLD” tag  
lights up.  
OPERATION:  
The meter is supplied complete with a  
9V battery. Slide off the battery compart-  
ment cover on the back of the meter  
and install the battery while paying at-  
tention to its polarity.  
M ilw aukee  
USER MANUAL  
PORTABLE uP-BASED  
pH METER  
After measurements, switch the  
meter off and store the electrode  
with a few drops of storage solu-  
tion in the protective cap.  
Always remove the electrode protective  
cap before taking any measurement. If  
the electrode has been left dry, soak  
the tip (bottom 2.5 cm) in rinse solution  
(M10000B) for a few minutes to reacti-  
vate it.  
MODEL: SM102  
Error message on display:  
Smart pH Meter  
Blinking full scale value: reading is  
over range;  
Connect the pH  
electrode and the  
temperature probe  
to the sockets on  
the meter’s top.  
The temperature  
probe can be used  
independently to  
take temperature  
measurements, or  
it can be used in  
conjunction with  
the pH electrode to utilize the meter’s  
ATC capability.  
“Eb”: battery low voltage indication;  
“Ec”: wrong buffer solution;  
“Clr”: the calibration data have been  
lost and recalibration is needed.  
CALIBRATION PROCEDURE:  
A) Preparation:  
Reference solutions:  
1. pH 7.01 (MA 9007)  
2. pH 4.01 (MA 9004) for measuring  
in the acid range (pH lower than 7)  
or pH 10.01 (MA 9010) for measur-  
ing in the alkaline range (pH higher  
than 7).  
Turn the instrument on by pressing the  
ON/OFF key.  
Use two beakers for each reference so-  
lution. One beaker for rinsing the  
electrode, the other for calibration. This  
way contamination between solutions  
during calibration is minimized.  
WARRANTY:  
Before taking any mea-  
surements, make sure  
that the meter has al-  
ready been calibrated.  
This instrument is warranted from all defects in  
materials and manufacturing for a period of two  
years from the date of purchase.  
The probe is warranted for a period of six months.  
B) Procedure:  
If during this period the repair or replacement of  
parts is required, where the damage is not due to  
negligence or erroneous operation by the user, please  
return the parts to either distributor or our office and  
the repair will be effected free of change.  
Immerse the tip (2.5 cm)  
of the pH electrode and  
the temperature probe  
into the sample and stir  
gently.  
Remove the electrode  
protective cap and rinse  
the tip with some  
pH7.01 solution, then  
immerse the electrode  
and the temperature  
probe into a pH7.01  
buffer solution.  
Note: We reserve the right to modify the design,  
construction and appearance of our products without  
advance notice.  
Take the pH reading when the hour-  
glass symbol stops blinking.  
To display the mea-  
sured temperature,  
press and hold the  
°C key.  
When the key is released, the display  
will return to the pH reading.  
Press the CAL key  
to enter the Cali-  
bration mode. The  
display flashes  
“pH 7.01” and the  
hourglass lights up.  
M ilw aukee  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miele Freezer K 122 Ui 1 User Manual
Miele Oven DG 4064 User Manual
Milwaukee Sander 6020 21 User Manual
MTD Snow Blower 31AH5C3F401 User Manual
NeumannBerlin Microphone KM 84 User Manual
New Holland Lawn Mower 1030 User Manual
N Tron Switch 608MFX User Manual
Panasonic DVD Player DV 686A s User Manual
ParaBody Home Gym 205101 User Manual
Patton electronic Modem 1008 I User Manual