MaxTech Oxygen Equipment MAXO2 User Manual

medicalManual.RevD.qxd 9/9/04 4:18 PM Page 1  
BAT  
CAL  
®
MAXO2 +A  
®
MAXO2 +AE  
®
2
+
MAXO A /AE USER’S GUIDE & OPERATING INSTRUCTIONS  
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 4  
®
+
Before use, all individuals who will be using the MAXO2 must become thoroughly  
familiar with the information contained in this Operation Manual. Strict adherence  
to the operating instructions is necessary for safe, effective product performance.  
This product will perform only as designed if installed and operated in accordance  
with the manufacturer’s operating instructions.  
Classification  
Protection against electric shock:  
Protection against water:  
Mode of Operation:  
Sterilization:  
Internally powered equipment.  
IPX1  
Continuous  
See section 7.0  
Use only genuine Maxtec® accessories and replacement parts. Failure to do so  
may seriously impair the analyzer’s performance. Repair or alteration of the  
Safety Labeling  
®
+
The following symbols and safety labels are found on the MAXO2  
:
®
+
MAXO2 beyond the scope of the maintenance instructions, or by anyone other  
than an authorized Maxtec® service person, could cause the product to fail to  
perform as designed.  
Attention, Consult ACCOMPANYING DOCUMENTS  
®
+
Calibrate the MAXO2 weekly when in operation, or if environmental conditions  
change significantly. (ie, Elevation, Temperature, Pressure, Humidity — refer to  
Section 3.0 of this manual).  
Off/On  
®
+
Use of the MAXO2 near devices that generate electrical fields may cause erratic  
readings.  
Calibration  
®
+
If the MAXO2 is ever exposed to liquids (from spills or immersion) or to any  
other physical abuse, turn the instrument OFF and then ON. This will allow the  
unit to go through its self test to assure everything is operating correctly.  
Found to meet the requirements of U.S. and Canadian nationally  
recognized codes and standards listed or classified by ETL.  
®
+
Never autoclave, immerse or expose the MAXO2 (including sensor) to high  
temperatures (>70°C). Never expose the device to pressure, irradiation vacuum,  
steam, or chemicals.  
NOTE: Replace the batteries with recongnized high quality AA Alkaline or Lithium batteries.  
NOTE: If the unit is going to be stored (not in use for 1 month), we recommend that you  
remove the batteries to protect the unit from potential battery leakage.  
FAILURE TO COMPLY WITH THESE WARNINGS AND CAUTIONS COULD RESULT IN  
INSTRUMENT DAMAGE AND POSSIBLY JEOPARDIZE THE WELL BEING OF THE  
PATIENT AND/OR HEALTH CARE PROFESSIONAL.  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
iv  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 6  
TABLE OF CONTENTS  
1.0 SYSTEM OVERVIEW  
1.0 SYSTEM OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1  
1.1 Base Unit Description  
1.1 Base Unit Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1  
1.2 Component Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2  
1.3 Component Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2  
1.4 MAX-250+ Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
®
+
The MAXO2 analyzer provides unparalleled performance and reliability due  
to an advanced design that includes the following features and operational  
benefits.  
2.0 OPERATING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
2.1 Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
• Extra-life oxygen sensor of approximately 900,000 O2 percent hours  
(2-year warranty)  
2.1.1 Protect Tape . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
2.1.2 Remove the Battery Ribbon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
2.1.3 Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4  
®
2.2 Calibrating the MAXO2 + Oxygen Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4  
• Durable, compact design that permits comfortable, hand-held operation  
and easy to clean.  
2.2.1 In Line Calibration (Flow Diverter - Tee Adapter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5  
2.2.2 Direct Flow Calibration (Barb) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5  
• Operation using only two AA Alkaline batteries (2 x 1.5 Volts) for approxi-  
mately 5000 hours of performance with continuous use. For extra extend-  
ed long life, two AA Lithium batteries may be used.  
3.0 FACTORS INFLUENCING ACCURATE READINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6  
3.1 Elevation Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6  
3.2 Temperature Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6  
3.3 Pressure Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6  
3.4 Humidity Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7  
• Oxygen-specific, galvanic sensor that achieves 90% of final value in  
approximately 15 seconds at room temperature.  
4.0 CALIBRATION ERRORS AND ERROR CODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7  
5.0 CHANGING THE BATTERIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8  
6.0 CHANGING THE OXYGEN SENSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9  
• Large, easy-to-read, 3 1/2-digit LCD display for readings in the 0-100%  
range.  
®
6.1 MAXO2 +A Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9  
• Simple operation and easy one-key calibration.  
• Self-diagnostic check of analog and microprocessor circuitry.  
• Low battery indication.  
®
6.2 MAXO2 +AE Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10  
7.0 CLEANING AND MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10  
8.0 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11  
8.1 Base Unit Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11  
8.2 Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12  
• Calibration reminder timer that alerts the operator, using a calibration icon  
on the LCD display, to perform a unit calibration.  
®
9.0 MAXO2 + SPARE PARTS AND ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12  
9.1 Included With Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12  
9.2 Standard Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13  
9.3 Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13  
9.3.1 Optional Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13  
9.3.2 Mounting Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13  
9.3.3 Carrying Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13  
10.0 WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
vi  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 8  
1.2 Component Identification  
1.4 MAX-250+ Oxygen Sensor  
The MAX-250+ oxygen sensor offers stability and extra life.  
The MAX-250+ is a galvanic, partial pressure sensor that is specific to oxygen. It  
consists of two electrodes (a cathode and an anode), a teflon membrane and an  
electrolyte. Oxygen diffuses through the teflon membrane and immediately reacts  
at a gold cathode. Concurrently, oxidation occurs electrochemically at the lead  
anode, generating an electrical current and providing a voltage output. Electrodes  
are immersed in a unique gelled weak acid electrolyte which is responsible for  
the sensors long life and motion insensitive characteristic. Since the sensor is  
specific to oxygen, the current generated is proportional to the amount of oxygen  
present in the sample gas. When no oxygen is present, there is no electrochemi-  
cal reaction and therefore, negligible current is produced. In this sense, the sen-  
sor is self-zeroing.  
FIGURE 1  
1.3 Component Description  
CAUTION: The Maxtec® MAX-250+ oxygen sensor is a sealed device containing a mild acid  
electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste  
constituents and should be disposed of properly, or returned to Maxtec® for proper  
disposal or recovery.  
1. 3 1/2-Digit Display - The 3 1/2 digit liquid crystal display (LCD) provides  
direct readout of oxygen concentrations in the range of 0 – 105.0%  
(100.1% to 105.0% used for calibration determination purposes). The  
digits also display error codes and calibration codes as necessary.  
CAUTION: Do not use ethylene oxide sterilization. Do not immerse the sensor in any  
cleaning solution, autoclave or expose the sensor to high temperatures.  
2. Low Battery Indicator - The low battery indicator is located at the top of the  
display and is only activated when the voltage on the batteries is below a  
normal operating level.  
CAUTION: Dropping sensor can adversely affect its performance.  
3. “%” symbol - The “%” sign is located to the right of the concentration  
number and is present during normal operation.  
2.0 OPERATING INSTRUCTIONS  
2.1 Getting Started  
4. Calibration symbol - The calibration symbol is located at the bottom of the  
display and is timed to activate when a calibration is necessary.  
5. ON/OFF Key - This key is used to turn the device on or off.  
2.1.1 Protect Tape  
6. Calibration Key - This key is used to calibrate the device. Holding the key  
Prior to turning on the unit, a protective film covering the threaded sensor  
face must be removed. After removing the film, wait approximately 20  
for more than three seconds will force the device to enter a calibration mode.  
minutes for the sensor to reach equilibrium.  
CAUTION: The device will assume a percent oxygen concentration when calibrating. Be  
sure to apply 100% oxygen, or ambient air concentration to the device during calibration  
or the device will not calibrate correctly.  
2.1.2 Remove the Battery Ribbon  
A ribbon has been inserted between the two case halves to prevent a battery  
connection. Remove the ribbon by pulling it completely out of the case. To  
energize the unit tighten all three screws with the included Phillips screwdriver.  
Sample Inlet Connection  
This is the port at which the device is connected to determine oxygen concentration.  
NOTE: If you do not tighten all three screws, the unit may not turn on or it may erratically turn  
on and off.  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
2
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 10  
• If you are unsure about the displayed O percentage.  
2
(see Factors influencing accurate readings.)  
2.1.3 Automatic Calibration  
After the unit is turned on it will automatically calibrate to room air. The  
display should be stable and reading 20.9%.  
A simple calibration may be made with the sensor open to static Ambient air. For  
optimum accuracy Maxtec® recommends that the Sensor be placed in a closed  
loop circuit where gas flow is moving across the sensor in a controlled manner.  
To check the oxygen concentration of a sample gas:  
(after the unit has been calibrated)  
2.2.1 In Line Calibration (Flow Diverter – Tee Adapter)  
1. Connect the Tygon tubing to the bottom of the  
analyzer by threading the barbed adapter onto  
the oxygen sensor. (figure 2, B)  
®
+
1. Attach the diverter to the MAXO2 by threading it on to the bottom of the sensor.  
®
+
2. Insert the MAXO2 in the center position of the tee adapter. (figure 2, A)  
FIGURE 2  
2. Attach the other end of the sample hose to  
the sample gas source and initiate flow of the  
sample to the unit at a rate of 1-10 liters per  
minute (2 liters per minute is recommended).  
3. Attach an open-ended reservoir to the end of the tee adapter. Then start  
the calibration flow of oxygen at two liters per minute.  
Six to 10 inches of corrugated tubing works well as a reservoir. A calibra-  
tion oxygen flow to the MAXO2 of two liters per minute is recommended  
to minimize the possibility of obtaining a "false" calibration value.  
3. Using the "ON/OFF" key, make sure the unit  
®
+
is in the power "ON” mode.  
4. Allow the oxygen reading to stabilize. This will normally take about 30  
4. Allow the oxygen to saturate the sensor. Although a stable value is usually  
observed within 30 seconds, allow at least two minutes to ensure that the  
sensor is completely saturated with the calibration gas.  
seconds or more.  
2.2 Calibrating the MAXO2®+ Oxygen Analyzer  
®
+
5. If the MAXO2 is not already turned on, do so now by pressing the  
®
+
NOTE: We recommend use of medical grade oxygen at 100% when calibrating the MAXO2  
.
analyzer “ON” button.  
®
®
+
+
The MAXO2 Analyzer should be calibrated upon initial power-up. Thereafter,  
6. Press the Cal button on the MAXO2 until you read the word CAL on the  
Maxtec® recommends calibration on a weekly basis. To serve as a reminder, a  
one week timer is started with each new calibration. At the end of one week a  
reminder icon “ ” will appear on the bottom of the LCD. Calibration is recom-  
analyzer display. This can take approximately 3 seconds. The analyzer will  
now look for a stable sensor signal and a good reading. When obtained,  
the analyzer will display the calibration gas on the LCD.  
mended if the user is unsure when the last calibration procedure was performed,  
or if the measurement value is in question.  
NOTE: Analyzer will read “Cal Err St” if the sample gas has not stablized.  
For hospital and home care a new calibration is required when  
2.2.2 Direct Flow Calibration (Barb)  
• The measured O percentage in 100% O is below 97.0% O  
2
2
2
®
+
1. Attach the Barbed Adapter to the MAXO2 by threading it on to the  
• The measured O percentage in 100% O is above 103.0% O  
2
2
2
bottom of the sensor.  
• The CAL reminder Icon is blinking at the bottom of the LCD  
• If you are unsure about the displayed O percentage. (see Factors  
influencing accurate readings.)  
2
2. Connect the Tygon tube to the barbed adapter. (figure 2, B)  
3. Attach the other end of the clear sampling tube to a source of oxygen with  
a known oxygen concentration value. Initiate flow of the calibration gas to  
the unit. Two liters per minute is recommended.  
For ID testing, (or optimum accuracy) a new calibration is required when  
• The measured O percentage in 100% O is below 99.0% O  
2
2
2
• The measured O percentage in 100% O is above 101.0% O  
2
2
2
• The CAL reminder Icon is blinking at the bottom of the LCD  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
4
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 12  
®
+
4. Allow the oxygen to saturate the sensor. Although a stable value is usually  
observed within 30 seconds, allow at least two minutes to ensure that the  
sensor is completely saturated with the calibration gas.  
• Calibrate the MAXO2 at the same pressure as the sample gas.  
• If sample gases flow through tubing, use the same apparatus and flow rates  
when calibrating as when measuring.  
®
+
5. If the MAXO2 is not already turned on, do so now by pressing the  
analyzer “ON” button.  
3.4 Humidity Effects  
®
+
6. Press the Cal button on the MAXO2 until you read the word CAL on the  
®
+
Humidity (non-condensing) has no effect on the performance of the MAXO2  
analyzer display. This can take approximately 3 seconds. The analyzer  
will now look for a stable sensor signal and a good reading. When  
obtained, the analyzer will display the calibration gas on the LCD.  
other than diluting the gas, as long as there is no condensation. Depending on  
the humidity, the gas may be diluted by as much as 4%, which proportionally  
reduces the oxygen concentration. The device responds to the actual oxygen  
concentration rather than the dry concentration. Environments where condensa-  
tion may occur are to be avoided since moisture may obstruct passage of gas to  
the sensing surface, resulting in erroneous readings and slower response time.  
For this reason, the following is recommended:  
3.0 FACTORS INFLUENCING ACCURATE READINGS  
3.1 Elevation Changes  
• Avoid usage in environments greater than 95% relative humidity.  
• Changes in elevation result in a reading error of approximately 1% of reading  
per 250 feet.  
• When used in a breathing circuit, place the sensor upstream of the humidifier.  
HELPFUL HINT: Dry sensor by lightly shaking moisture out, or flow a dry gas at two liters per  
minute across the sensor membrane.  
• In general, calibration of the instrument should be performed when elevation at  
which the product is being used changes by more than 500 feet.  
3.2 Temperature Effects  
4.0 CALIBRATION ERRORS AND ERROR CODES  
®
+
The MAXO2 will hold calibration and read correctly within 3% when at thermal  
®
+
equilibrium within the operating temperature range. The device must be thermally  
stable when calibrated and allowed to thermally stabilize after experiencing tem-  
perature changes before readings are accurate. For these reasons, the following  
is recommended:  
The MAXO2 analyzers have a self test feature built into the software to detect  
faulty calibrations, oxygen sensor failures, and low operating voltage. These are  
listed below, and include possible actions to take, if an error code occurs.  
E02: No sensor attached  
®
• For best results, perform the calibration procedure at a temperature close to  
the temperature where analysis will occur.  
MAXO2 +A: Open unit and disconnect and reconnect sensor. Unit should perform  
an auto calibration and should read 20.9%. If not, contact Customer Service for  
possible sensor replacement.  
• Allow adequate time for the sensor to equilibrate to a new ambient temperature.  
CAUTION: “CAL Err St” may result from a sensor that has not reached thermal equilibrium.  
• When used in a breathing circuit, place the sensor upstream of the heater.  
®
MAXO2 +AE: Disconnect and reconnect external sensor. Unit should perform an  
auto calibration, and should read 20.9%. If not, contact Customer Service for  
possible sensor replacement or cable replacement.  
E03: No valid calibration data available  
Make sure unit has reached thermal equilibrium. Press and hold the Calibration  
Button for three seconds to manually force a new calibration.  
3.3 Pressure Effects  
®
+
Readings from the MAXO2 are proportional to the partial pressure of oxygen.  
E04: Battery below minimum operating voltage  
Replace batteries.  
The partial pressure is equal to the concentration times the absolute pressure.  
Thus, the readings are proportional to the concentration if the pressure is held  
constant. Therefore, the following are recommended:  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
6
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 14  
CAL Err St: O2 Sensor reading not stable  
Wait for displayed oxygen reading to stabilize, when calibrating the device at  
100% oxygen.  
Carefully, bring the two halves of the case together while positioning the wires so  
they are not pinched between the two case halves. The gasket separating the  
halves will be captured on the back case half.  
Wait for unit to reach thermal equilibrium (Please note that this can take up to  
one half hour, if the device is stored in temperatures outside the specified operat-  
ing temperature range).  
Reinsert the three screws and tighten until the screws are snug. (figure 3)  
The device will automatically perform a calibration and begin displaying % of oxygen.  
CAL Err lo: Sensor voltage too low  
HELPFUL HINT: If unit does not function, verify that the screws are tight to allow proper  
electrical connection.  
Press and hold the Calibration Button for three seconds to manually force a new  
calibration. If unit repeats this error more than three times, contact Customer  
Service for possible sensor replacement.  
HELPFUL HINT (MAXO2+ AE): Before closing the two case halves together, verify that the keyed  
slot on top of the coiled cable assembly is engaged on the small tab located on the back case.  
This is designed to position the assembly in the correct orientation and prevent it from rotat-  
ing. Improper positioning could hinder the case halves from closing and prevent operation  
when tightening the screws.  
CAL Err hi: Sensor voltage too high  
Press and hold the Calibration Button for three seconds to manually force a new  
calibration. If unit repeats this error more than three times, contact Customer  
Service for possible sensor replacement.  
CAL Err Bat: Battery voltage too low to recalibrate  
Replace batteries  
6.0 CHANGING THE OXYGEN SENSOR  
®
6.1 MAXO +A Model  
2
Should the oxygen sensor require changing, the device will indicate this by  
presenting “Cal Err lo” on the display after initiating a calibration.  
5.0 CHANGING THE BATTERIES  
Should the batteries require changing, the device will indicate this in one of two ways:  
To change the oxygen sensor, begin by removing  
the three screws from the back of the device.  
A #1 phillips screwdriver is required to  
remove these screws.  
• The battery icon on the bottom of the display will begin to flash. This icon  
will continue to flash until the batteries are changed. The unit will continue  
to function normally for approx. 200 hours.  
Lock Lever  
Once the screws are removed, gently separate  
the two halves of the device.  
• If the device detects a very low battery level, an error code of “E04” will be pres-  
ent on the display, and the unit will not function until the batteries are changed.  
Disconnect the oxygen sensor from the printed  
circuit board by pressing the un-lock lever first and  
then pulling the connector out of the receptacle.  
To change the batteries, begin by removing the three  
screws from the back of the device. A #1 Phillips  
screwdriver is required to remove these screws.  
FIGURE 4  
The oxygen sensor can now be replaced from the back half of the case.  
Once the screws are removed, gently separate  
the two halves of the device.  
HELPFUL HINT: Be sure to orient the new sensor by aligning the red arrow on the sensor with  
the arrow in the back case. A small tab is located on the back case that is designed to engage  
the sensor and prevent it from rotating within the case. (figure 4)  
The batteries can now be replaced from the back half of the  
case. Be sure to orient the new batteries as indicated in  
the embossed polarity on the back case.  
NOTE: If the oxygen sensor is installed incorrectly, the case will not come back together and  
the unit may be damaged when the screws are reinstalled.  
NOTE: If the batteries are installed incorrectly the  
batteries will not make contact and the device will not operate.  
FIGURE 3  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
8
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 16  
Reconnect the oxygen sensor to the connector on the printed circuit board.  
Carefully bring the two halves of the case together while positioning the wires to  
ensure they are not pinched between the two case halves. Make sure the sensor is  
fully inserted and in the proper orientation.  
exhaled breath or secretions, unless you intend to dispose of the sensor, flow diverter  
and tee adapter after use.  
• Clean the sensor with a cloth moistened with isopropyl alcohol  
(65% alcohol/water solution).  
Reinsert the three screws and tighten until the screws are snug. Verify the unit  
operates properly.  
• Maxtec® does not recommend use of spray disinfectants because they can con-  
tain salt, which can accumulate in the sensor membrane and impair readings.  
The device will automatically perform a calibration and begin displaying % of oxygen.  
• The oxygen sensor is not intended for steam, ethylene oxide or radiation sterilization.  
®
6.2 MAXO +AE Model  
2
Accessories  
Should the oxygen sensor require changing, the device will indicate this by  
• The flow diverter and tee adapter may be disinfected by washing them  
with isopropyl alcohol or Cidex (per manufacturer's instructions). The parts  
must be thoroughly dry before they are used.  
presenting “Cal Err lo” on the display.  
Unthread the sensor from the cable by rotating the thumbscrew connector  
counterclockwise and pull the sensor from the connection.  
Replace the new sensor by inserting the electrical plug from the coiled cord into  
the receptacle on the oxygen sensor. Rotate the thumbscrew clockwise until snug.  
• Sterilizing -The flow diverter and tee adapter may be sterilized using  
Cidex, steam or ethylene oxide (per manufacturer's instructions). Due  
to the varying conditions imposed on the materials during sterilization it  
is not possible to determine the exact number of times the sterilization  
processes can be carried out. Therefore, Maxtec® recommends that oper-  
ators carefully examine the flow diverter and tee adapter after sterilization  
and prior to use to verify that the item is fit for use. The operator should  
verify that there are no cracks and the item does not show any indication  
of material changes or physical damage that may compromise its effective  
use. Both flow diverter and tee adapter should be free of any chemical  
residue remaining from the sterilization process.  
The device will automatically perform a calibration and begin displaying % of oxygen.  
7.0 CLEANING AND MAINTENANCE  
®
+
Store the MAXO2 analyzer in a temperature similar to its ambient environment of  
daily use.  
Because of the variability of the cleaning, disinfecting and sterilizing processes,  
Maxtec® cannot provide specific sterilization instructions, nor can the sterility of  
the item be ensured. Therefore, we highly recommend referring to the manufac-  
turer's instructions on the details of method.  
The instruction given below describes the methods to clean and disinfect the  
instrument, sensor and its accessories (e.g. flow diverter, tee adapter):  
Instrument  
®
+
• When cleaning or disinfecting the exterior of the MAXO2 analyzer, take  
appropriate care to prevent any solution from entering the instrument. Do  
not immerse unit in fluids.  
8.0 SPECIFICATIONS  
8.1 Base Unit Specifications  
®
+
• The MAXO2 analyzer surface may be cleaned using a mild detergent and  
a moist cloth.  
Measurement Range:  
Resolution:  
Accuracy and Linearity:  
0-100%  
0.1%  
®
+
• The MAXO2 analyzer is not intended for steam, ethylene oxide or radiation  
1% of full scale at constant temperature, R.H. and  
pressure when calibrated at full scale.  
sterilization.  
Total Accuracy:  
Response Time:  
3% actual oxygen level over full operating  
temperature range.  
90% of final value in approximately 15 seconds at 23°C.  
Oxygen Sensor  
WARNING: Never install the sensor in a location that will expose the sensor to patient's  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
10  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 18  
BAT  
CAL  
Warm-up Time:  
none required  
9.2 Standard Replacement Parts  
Operating Temperature:  
Storage Temperature:  
Atmospheric Pressure  
Humidity:  
Power Requirements:  
Battery Life:  
Low Battery Indication:  
Sensor Type:  
Expected Sensor Life:  
15°C - 40°C (59°F - 104°F)  
-15°C - 50°C (5°F - 122°F)  
800-1013 mBars  
A Model AE Model  
Part Number  
R125P02-011  
R125P03-002  
R115P85  
Item  
MAX-250+ Oxygen Sensor  
MAX-250E Oxygen Sensor  
MAX-250ESF Oxygen Sensor  
Gasket  
X
X
X
0-95% (non-condensing)  
2, AA Alkaline batteries (2 x 1.5 Volts)  
approximately 5000 hours with continuous use  
"BAT" icon displayed on LCD  
Maxtec® MAX-250+ galvanic fuel cell  
>900,000 O percent hours minimum 2-years in  
2
typical medical applications  
R217P08  
X
X
X
X
X
X
X
X
X
X
RP06P25  
#4-40 Pan Head Stainless Steel Screw  
Front Assembly (Includes Board & LCD)  
Back Assembly  
R217P16-001  
R217P11-002  
R217P24  
Coiled Cable Assembly  
Overlay  
R217P09-001  
X
A Model Dimensions:  
A Weight:  
3.0"(W) x 4.0"(H) x 1.5"(D) [76mm x 102mm x 38mm]  
0.4 lbs. (170g)  
AE Model Dimensions:  
3.0"(W) x 36.0"(H) x 1.5"(D) [76mm x 914mm x 38mm]  
Height includes external cable length (retracted).  
0.6 lbs. (285g)  
9.3 Optional Accessories  
9.3.1 Optional Adapters  
AE Weight:  
8.2 Sensor Specifications  
Part Number  
RP16P02  
R103P90  
RP16P12  
RP16P05  
RP16P10  
R207P17  
Item  
Blue Tee Adapter  
Perfusion Tee Adapter  
Long-Neck Tee Adapter  
Pediatric Tee Adapter  
MAX-Quick Connect  
Threaded Adapter with Tygon Tubing  
Type:  
Life:  
Galvanic fuel sensor (0-100%)  
2-years in typical applications  
+
BAT  
9.0 MAXO SPARE PARTS AND ACCESSORIES  
2
CAL  
9.3.2 Mounting Options (requires dovetail)  
9.1 Included With Your Unit  
Part Number  
R206P75  
R205P86  
R213P31  
R100P10  
R217P23  
Item  
A Model AE Model  
Part Number  
R217P43-002  
R217P44-002  
R217M40  
Item  
Pole Clamp  
Wall Mount  
Swivel Mount  
Rail Mount  
Dovetail Assembly  
MAXO2+A Medical Unit with Sensor  
MAXO2+AE Medical Unit with Sensor  
User’s Guide and Operating Instructions  
Quick Reference Calibration Card  
Lanyard  
X
X
X
X
X
X
X
X
X
X
X
X
X
X
R217P27  
RP76P06  
9.3.3 Carrying Options  
R207P17  
Threaded Adapter with Tygon Tubing  
Screw Driver  
Part Number  
RP77P01  
R217P22  
R213P02  
R213P56  
R217P32  
R213P90  
Item  
Jump Bag  
Belt Clip and Pin  
RP18P01  
X
X
X
R110P10  
Flow Diverter  
Zipper Carrying Case with Shoulder Strap  
Deluxe Carrying Case, Water Tight  
Soft Case, Tight Fit Carrying Case  
RP16P02  
Blue Tee Adapter  
R217P10-001  
R217P15-001  
R217P10-002  
R217P15-002  
Quick Reference Guide  
Getting Started Guide  
Quick Reference Guide  
Getting Started Guide  
®
Handi/MAXO2 + Water Tight Box  
X
X
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
12  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 20  
10.0 WARRANTY  
®
The MAXO2 + Analyzer is designed for medical oxygen delivery equipment and systems. Under normal  
®
operating conditions, Maxtec® warrants the MAXO2 + Analyzer to be free from defects of workmanship or  
materials for a period of 2-years from the date of shipment from Maxtec®, provided that the unit is prop-  
erly operated and maintained in accordance with Maxtec®’s operating instructions. Based on Maxtec®’s  
product evaluation, Maxtec®'s sole obligation under the foregoing warranty is limited to making replace-  
ments, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the  
buyer purchasing the equipment directly from Maxtec® or through Maxtec®'s designated distributors and  
agents as new equipment.  
®
Maxtec® warrants the MAX-250+ oxygen sensor in the MAXO2 + Analyzer to be free from defects in mate-  
®
rial and workmanship for a period of 2-years from Maxtec®'s date of shipment in a MAXO2 + unit. Should  
a sensor fail prematurely, the replacement sensor is warranted for the remainder of the original sensor  
warranty period.  
Routine maintenance items, such as batteries, are excluded from warranty. Maxtec® and any other sub-  
sidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or  
equipment that has been subject to abuse, misuse, mis-application, alteration, negligence or accident.  
These warranties are exclusive and in lieu of all other warranties, expressed or implied, including warranty  
of merchantability and fitness for a particular purpose.  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
14  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 22  
M A N U FA C T U R E D B Y M A X T E C , I N C .  
WWW. M A X T E C I N C . C O M  
16  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 24  
Download from Www.Somanuals.com. All Manuals Search And Download.  

KWC Plumbing Product 28243234000 User Manual
Leica Binoculars BL 42 User Manual
Lennox Hearth Electric Heater EBVI25 User Manual
Leviton Video Gaming Accessories MRS4 User Manual
LG Electronics Cell Phone TU500F User Manual
LG Electronics Computer Monitor 72WX70MF User Manual
LOREX Technology Security Camera CVS1000 User Manual
Makita Planer 1805N User Manual
Meyer Sound Speaker CQTM Series User Manual
Midland Radio Two Way Radio LXT360 Series User Manual