Lumex Syatems Camping Equipment 763400A User Manual

Low Air Loss Therapy  
Mattress Replacement System  
Model # 763400A  
Operators Manual  
PPlleeaassee rreeaadd tthhee mmaannuuaall bbeeffoorree uussee  
GF0500067REVA05  
Download from Www.Somanuals.com. All Manuals Search And Download.  
IMPORTANT SAFEGUARDS  
READ ALL INSTRUCTIONS BEFORE USING  
DANGER - To reduce the risk of electrocution:  
1. Always unplug this product immediately after using.  
2. Do not use while bathing.  
3. Do not place or store product where it can fall or be pulled into a tub or sink.  
4. Do not place in or drop into water or other liquid.  
5. Do not reach for a product that has fallen into water. Unplug immediately.  
WARNING - To reduce the risk of burns, electrocution, fire, or injury to persons:  
1. This product should never be left unattended when plugged in.  
2. Close supervision is necessary when this product is used by, on, or near children or  
invalids.  
3. Use this product only for its intended use as described in this manual. Do not use  
attachments not recommended by the manufacturer.  
4. Never operate this product if it has a damaged cord or plug, if it is not working properly, if  
it has been dropped or damaged, or dropped into water. Return the product to a service  
center for examination and repair.  
5. Keep the cord away from heated surfaces.  
6. Never block the air openings of this product or place it on a soft surface, such as a bed  
or couch, where their openings may be blocked. Keep the air opening free of lint, hair,  
and other similar particles.  
7. Never drop or insert any object into any opening or hose.  
8. This device is class II equipment with functional earth, no Protective Earth id provided.  
(For 120V system only)  
NOTE, CAUTION AND WARNING STATEMENTS:  
NOTE- Indicates some tips.  
CAUTION - Indicates correct operating or maintenance procedures in order to prevent  
damage to or destruction of the equipment or other property.  
WARNING - Calls attention to a potential danger that requires correct procedures or  
practices in order to prevent personal injury.  
SYMBOLS  
"BF"symbol, indicates this product is according to the degree of protecting against  
electric shock for type BF equipment.  
The Symbol for class II equipment, 2 concentric square indicating double insulation.  
Attention, should read the instructions !  
IPX0 Enclosed equipment without protection against ingress of water.  
SAVE THIS MANUAL FOR REFERENCE  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Introduction  
This manual should be used for initial set up of the system and for reference purposes.  
1.1 General Information  
LumexAir systems use most advanced technologies and mattress materials to deliver the  
best quality, comfort and whisper-quiet operations to you and your customers. Most systems  
in LumexAir are recommended for prevention and treatment of pressure ulcers up to and  
including stage IV pressure ulcers.  
This LumexAir range of system has been independently tested and successfully approved to  
the following standards:  
EN60601-1 / UL 60601-1  
EN 60601-1-2  
IEC 61000-4-2  
IEC 61000-4-3  
IEC 61000-4-4  
IEC 61000-4-5  
IEC 61000-4-6  
IEC 61000-4-8  
IEC 61000-4-11  
EN55011 CLASS B  
EN 61000-3-2  
EN 61000-3-3  
FOR US AND CANADA ONLY  
Medical Equipment-Air Pump  
E228589  
53DG  
with respect to electrical shock, fire and mechanical  
hazards only in accordance with UL60601-1 and  
CAN/CSA C22.2 No. 601.1  
1.2 Intended Use  
The system is a true low air loss replacement system suitable for high-risk patients. It is non-  
alternating mattress designed with very low average pressure dispersion. The pulsation  
function on the compact pump unit provides patients the message action to stimulate the  
blood circulation and reduce the edema.  
NOTE : Equipment not suitable for use in the presence of a flammable anesthetic  
mixture with air or with oxygen or nitrous oxide.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Product Description  
2.1 Pump and Mattress System  
The system consists of a high volume airflow blower and numerous large air loss holes on  
the mattress to achieve the true low air loss therapy. It also provides excellent ventilation  
and keeps patients dry from humidity. The 10 inch high mattress, which consists of 8 inch  
high air bladders at the top and a layer of 2" foam at the bottom, provide maximum comfort  
for patients. The 2" foam also functions as the emergency pressure dispersion in case of  
power failure. The air tube connector also operates as CPR facility can deflate the mattress  
within the shortest amount of time for emergency situation.  
1. Mattress  
2. Pump unit  
3. Air tube connector  
4. Control Panel  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Installation  
3.1 Unpacking  
The pump unit and mattress are packaged in a separate box to secure its contents inside.  
Unpacking these boxes to remove the pump unit and mattress and check for any damage,  
which may have occurred in shipping. If there are damages, please contact your dealer  
immediately.  
3.2 Setting Up  
1. Place the mattress on top of the bed frame. Please note for  
the foot end. Mattress can be secured firmly by fixing the  
elastic straps at the bottom bed frame.  
NOTE: Please cover the mattress with a cotton sheet to  
avoid direct skin contact and for patient's comfort.  
2. Hang the pump onto bed rail (foot-end), and adjust a suitable  
wideness of hangers.  
3. Connect air tube connector from air mattress to the pump unit  
When "click" is felt or heard, the connection is completed and  
secured.  
NOTE: Check and ensure the air tubes are not kinked or  
tucked under mattress.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Plug into main electrical outlet.  
NOTE  
1. Make sure the pump unit is suitable for the local power  
voltage.  
2. The plug also serves as the disconnect device  
CAUTION: The pump can only be applied to the mattress  
recommended by the manufacturer. Do not use it for any  
other purpose.  
5
Then turn the main power switch ON. The power switch is  
located on the right side panel of the pump unit.  
NOTE: Ensure the CPR connectors are sealed completely.  
Installation tips are listed below:  
After installation, the extra wire of the power cord, if any, should be neatly arranged to avoid  
any tripping accidents. The equipment should be firmly placed at position where  
users/doctors can access easily.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Operation  
NOTE: Always read the operating instructions before use.  
4.1 Function Description:  
Control Button  
Function Description  
POWER  
Two conditions:  
1. To turn on control unit. It takes approximately 1 minute to inflate the  
mattress. The default setting is at a comfort range, level 6, for the  
quickest inflation time.  
2. To turn off Power Failure Alarm in the event of a sudden power outage  
during normal use.  
ALARM MUTE  
Enable and disable the audible alarm feature. Press this button to  
silence the audible alarm. The alarm buzzer will sound again after 3  
minutes if the pressure has not yet returned to the normal condition.  
PANEL LOCK  
To lock the current setting and prevent others from changing. To unlock  
the panel, press this button and hold it for 3 seconds.  
LOW PRESSURE INDICATOR  
3 Conditions  
1. No light - normal condition; mattress is operating properly.  
2. Yellow light - pressure has not reached the preset setting.  
3. Flashing yellow light - low pressure condition.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Control Button  
Function Description  
AUTOFIRM  
For maximum inflation of mattress. All other functions are disabled  
unless this button is pressed again. This mode will last for 20 minutes  
and then it will return to default setting automatically.  
STATIC  
For true low air-loss therapy mode, similar to a regular air mattress  
with very low air pressure support and true low air-loss therapy.  
PULSATION  
For pulsating and true low air-loss therapy mode, it inflates and deflates  
the mattress between two comfort levels at 2 cycles per minute.  
COMFORT RANGE  
The mattress has 6 comfort ranges. Use UP/DOWN button to adjust  
the comfort levels. 1 is the softest & 6 is the firmest.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.2 Operating Instruction  
1. Press the POWER button on the display panel to start the system.  
2. For faster inflation, the AUTOFIRM function can be activated. This mode will last for 20  
minutes and then it will return to default setting automatically.  
3. After the mattress is fully inflated, user can change the setting according to their  
preference or recommended by physician.  
4.3 CPR  
When there is an emergency to perform CPR on the patient,  
quickly pull the CPR valves to release air from mattress. The  
CPR valves are located at the head-end, right-hand-side of  
the mattress. The air hose on the pump unit can be  
disconnected for even faster deflation.  
4.4 Comfort Range Setup  
Users can adjust the pressure of air mattress by adjusting the comfort range from level 1 to 6.  
Please consult your physician for a suitable setting.  
When the Low Pressure Indicator is off, it indicates the mattress has reached to its preset  
softness. Patient can lie on the mattress.  
NOTE: Check to see if the suitable comfort range (or pressure setting) is selected  
by sliding one hand between the air cell and the patient to feel the patient buttock.  
Users should be able to feel the minimal contact.  
4.5 Pulsation Function  
The mattress will inflate to the preset pressure level and deflate to the next lower pressure  
level on a cycle of 2 times per minute. This means a 30 second period for one inflating and  
deflating cycle.  
For example, if the Pulsation function is activated and the preset comfort level is set at level  
2, then the pressure setting will increase and decrease 2 times per minute between comfort  
level 2 and 1.  
4.6 Low Pressure and Alarm Function  
When an abnormal low pressure occurs for a long time, the Low Pressure Indicator (flashing  
yellow LED) will light up with audible alarm. Check if all the connections are securely made  
and that they are correctly installed as per installation instructions.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTE: If the pressure level is consistently low, check for any leakage at air cells or  
connecting tubes. If necessary, replace any damaged cells or tubes or contact local  
qualified dealer for repair.  
If the control unit is equipped without audible alarm feature, the Low Pressure Indicator will  
flash until the low-pressure fault condition is solved. If the control unit is equipped with  
audible alarm feature, the Low Pressure indicator will flash and the alarm will sound when  
air pressure is below normal (preset pressure). To silence the alarm, simply press the alarm  
mute button, and it will be silent for 3 minutes; however, the Low Pressure Indicator still  
remains flashing. After 3 minutes, alarm buzzer will alert again if the pressure has not yet  
returned to normal.  
4.7 Power Failure Alarm  
If the pump unit has a sudden power outage during normal use, the power indicator LED on  
the control panel will flash " yellow" light, and alarm will sound. To silence the alarm, simply  
press Power button on the control panel.  
When power failure situation occurs, alarm will last until backup battery is low or it can be  
turned off by manual operation.  
Once power failure alarm situation is resolved, it is recommended to turn "OFF" the main  
power switch located at the lateral side of the pump. If the main power switch is not turned  
"OFF", once the main power comes back again, the pump will start again at "Standby" status.  
5. Cleaning  
CAUTION: Ensure the pump unit is disconnected from the main power before  
cleaning.  
5.1 Pump Unit  
Wipe with a damp cloth and a mild detergent and keep it away from dust. If other detergent  
is used, choose one that will have no chemical effects on the surface of the plastic case of  
the pump unit. All parts should be air dried thoroughly before use.  
CAUTION: Do not immerse or soak pump unit.  
5.2 Mattress  
Do not use Phenolic based products for cleaning.  
For suitable and locally available cleaner please ask your distributor.  
Wipe down with warm water containing a mild detergent. The cover may also be cleaned  
using sodium hypochlorite diluted in water. All parts should be air dried thoroughly before use.  
CAUTION: Dry the mattress on sunless area after cleaning.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Storage  
1. Pull the CPR tags and make sure all cells are deflated.  
2. Lay the mattress out flat and upside down.  
3. Protect the air tube coupler by putting it in the inner side of mattress.  
4. Roll the mattress from the head end towards the foot end.  
5. Foot-end straps can then be stretched around the rolled mattress to prevent unrolling.  
7. Maintenance  
7.1 General  
1. Check main power cord and plug if there is abrasions or excessive wear.  
2. Check mattress' cover for signs of wear or damage. Ensure mattress cover and tubes  
are stubbed together correctly.  
3. Check the air tube connections for any damage.  
4. Report any problems or malfunctions to local distributors.  
7.2 Air Filter Replacement  
The air filter should be checked and dusted periodically  
once per month.  
1. Two foam type filters are used. One is located on the  
back of pump and the other one is at the bottom of the  
Pump.  
2. The filter is reusable and can be washed gently with a  
mild detergent and water.  
3. Suggest to replace air filter regularly if the environment  
Is dirty. This will help to improve the airflow. Spare  
filters are available from your local distributor.  
7.3 Fuse Replacement  
1. Disconnect the plug from main power when a fuse is to  
be replaced.  
2. Remove the cover of the fuse holder by means of a  
small screwdriver.  
3. Insert a new fuse of the correct rating in, and replace  
the cover of the fuse holder back. Check the technical  
specification for correct fuse rating.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Troubleshooting  
Problem  
Solution  
• Check if the plug is connected to main power.  
• Check if there is any blown fuse.  
Power is not ON  
• Check if the connection between air tube connector to  
pump unit is tightly secured.  
• Check if the CPR is sealed.  
• Check if there are any leakage on air cells.  
• Check if all tubing connections along mattress are secured.  
Alarm is on  
(audible & visual)  
• Pressure setting might be inadequate for the patient,  
adjust comfort range 1 to 2 levels higher and wait for a  
few minutes.  
Patient is bottoming out  
• This may happen when the AUTOFIRM is activated for  
too long. Release some air by disconnecting the air tube  
connector and change to preferred setting.  
Mattress becomes too firm and  
pressure can’t be reduced.  
• Check if all the snap buttons or straps of mattress are  
all securely fastened.  
• Check if the mattress is fixed to the bed frame by  
elastic straps.  
Air Mattress is loose  
• Press handles on both sides of air tube connector  
slightly before pushing the connector into the pump unit.  
Air tube connector won’t fit  
into the pump unit  
• Make sure the connector is tightly secured by hearing  
“click” while pushing it into pump unit.  
Large airflow from air  
bladders  
• This is normal since this system has high airflow and  
numerous air loss holes on all bladders.  
If your questions can't be answered with above information, it might require a technician to  
take care the problem. Please contact your local dealer directly.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Technical Description  
Specifications:  
Pump  
Specification  
120V AC 60Hz, 1A (For 120V system)  
230VAC 50Hz, 7W ( For 230V system)  
Power Supply  
(Note: See rating label on the product)  
Fuse Rating  
Dimension  
Weight  
5A (For 120V system), T3A ( For 230V system)  
37 x 27 x 13 (cm) or 14.4" x 10.5" x 5"  
4.5 kg or 9.9 lb  
Operation:10° C to 40° C (50° F to 104° F)  
Temperature Storage: -10° C to 50° C (14° F to 122° F)  
Shipping: -10° C to 70° C (14° F to 158° F)  
Operation: 30% to 75% non-condensing  
Storage: 10%to 90% non-condensing  
Shipping: 10 % to 90 % non-densing  
Environment  
Classification  
Humidity  
Atmosphere 700 hPa to 1060 hPa  
• Class II, Type BF, IPX0  
• Applied part: Air Mattress  
• Not suitable for use in the presence of a flammable  
anesthetic mixture(No AP or APG protection)  
• Continuous operation  
Mattress  
Specification  
Dimension (LxWxH)  
Weight *  
2.. x 90 x 20.3 (cm) or 78.7" x 35.4" x 8"  
7 kg or 15.4 lb  
* The weight and dimension of mattress is measured without foam.  
Consult the distributor or EU representative for further technical documents.  
NOTE: The specifications, also apply to those areas operating with the same  
power supply.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
U.S.A., Corporate Headquarters:  
Graham-Field Health Products  
2935 Northeast Parkway  
Atlanta, Georgia 30360  
telephone: 800-347-5678, 770-447-1609  
fax: 800-726-0601, 678-291-3232  
Download from Www.Somanuals.com. All Manuals Search And Download.  

KTM Motorcycle 990 SUPER DUKE User Manual
Leica Projection Television CM1950 User Manual
Lenovo Network Router TS140 User Manual
LOREX Technology Security Camera CVC6994CL User Manual
Lucent Technologies IP Phone CIB 3020 User Manual
Magnadyne Conference Phone BT HFKP10 User Manual
Makita Saw 5057KB User Manual
Maxon Telecom Radio SP 300 Series Synthesized Scanning Radio User Manual
Melissa Electric Grill 651 003 User Manual
Michelin GPS Receiver Navigation User Manual