1. CONTROL AREA FEATURES
2. REAR SIDE AND BOTTOM FEATURES
HOME INDICATOR
• LIGHTS WHEN ACTIVITY TIMER IS ON
NOTE: THE CONSOLE IS
BACKUP
BATTERY
ACCESS DOOR
SHIPPED WITH REMOVABLE
LABELS COVERING
SOME OF THE REAR
CONNECTIONS
CONSOLE REAR
• LIGHT IS OFF WHEN ACTIVITY TIMER IS OFF
• BLINKS SLOWLY WHEN LESS THAN 15 MINUTES
REMAIN ON THE ACTIVITY TIMER
STATUS INDICATOR
• NORMALLY OFF
• LIGHTS WHEN ANY TRANSMITTERS ARE
MISSING (SUPERVISORY)
• BLINKS SLOWLY WHEN ANY TRANSMITTERS
• BLINKS QUICKLY WHEN THE ACTIVITY
TIMER HAS EXPIRED
HOME BUTTON
HAVE A LOW BATTERY
• PRESS TO RESET ACTIVITY TIMER
• PRESS FOR 5 SECONDS TO TURN
ACTIVITY TIMER AND REMINDER RPT. OFF
• IF ACTIVITY TIMER IS OFF,
PRESS FOR 5 SECONDS
TO TURN ACTIVITY TIMER AND
REMINDER REPORT ON
• USED IN TEST MODE TO
-
POWER
SWITCH
RECORD AND PLAYBACK
REMINDER MESSAGES
LOCAL
PHONE
JACK
TEST
BUTTON
EXTERNAL SPEAKER
AND RELAY OUTPUT
TERMINALS
EXTERNAL
SPEAKER
SENDING INDICATOR
EXTERNAL
MICROPHONE
TERMINALS
• NORMALLY OFF
• FLASHES WHEN CONSOLE
IS PLACING A CALL
TELEPHONE
LINE JACK
BYPASS JUMPER
(FACTORY INSTALLED)
POWER INDICATOR
• LIGHTS WHEN AC IS PRESENT
• BLINKS SLOWLY WHEN BATTERY IS CHARGING
• BLINKS QUICKLY WHEN AC IS ABSENT
(CONSOLE ON BATTERY POWER)
CONSOLE BOTTOM
HELP BUTTON
• PRESS AND HOLD TO TRIGGER
THE EMERGENCY ALARM
• PRESS DURING RING FOR
SPEAKERPHONE MODE
16VAC 1.25 AMP
PLUG-IN
TRANSFORMER
• OFF IN TEST AND SETUP MODES
• PRESS DURING SPEAKERPHONE
MODE TO DISCONNECT
• PRESS WHILE IN TEST MODE TO
CONNECT THE CONSOLE TO A PC
FOR LOCAL PROGRAMMING WITH
THE RA-2400 SOFTWARE
CANCEL BUTTON
• STOPS ALARM
• ADDS CANCEL TO AN ALARM
REPORT IN PROGRESS
• PRESS IMMEDIATELY AFTER
EMERGENCY TO PREVENT REPORT
• CLEARS INDICATORS
• STOPS TROUBLE BEEPER
• PRESS FOR 5 SECONDS TO ADD
A TRANSMITTER
HOME BUTTON COVER
• HIDES THE HOME BUTTON IF THE
ACTIVITY TIMER IS NOT USED
BATTERY FUSE COVER
FUSE TYPE: 2AG
FUSE RATING: 1AMP
TWO-WAY
AUDIO VOLUME
DIGITAL VOICE
VOLUME
(REMOVE IF INSTALLED TO RECORD
AND PLAYBACK REMINDER MESSAGES
• CLEARS REMINDER MESSAGE
3. TELEPHONE CONNECTIONS
4. BACKUP BATTERY INSTALLATION
REMOVE THE
BATTERY DOOR
1
2
BE SURE THAT
THE POWER
SWITCH IS OFF
REAR OF
CONSOLE
ON
OFF
PUSH THE RELEASE TABS IN
AND SLIDE THE DOOR UPWARD
CONNECT THE RED LEAD
TO POSITIVE AND THE
BLACK LEAD TO THE
NEGATIVE BATTERY
TERMINAL
3
CONNECT THIS JACK
TO THE LOCAL
TELEPHONE
CONNECT THIS JACK
TO THE TELEPHONE LINE
CAUTION: CONNECTING
THE BATTERY LEADS BACKWARDS
WILL BLOW THE BATTERY FUSE
FUSE TYPE: 1 AMP 2AG
RESIDENT'S
TELEPHONE
JACK
BATTERY TYPE:
12-VOLT, 1.2 AMP HOUR
SLIDE THE BATTERY
4
INTO THE CASE AND
REPLACE THE
BATTERY DOOR
IMPORTANT: WHEN REPLACING THE
BATTERY, THE OLD BATTERY MUST
BE RECYCLED. DELIVER THE BATTERY
TO AN APPROPRIATE RECYCLING FACILITY
RESIDENT'S
LOCAL TELEPHONE
(OPTIONAL)
1
Download from Www.Somanuals.com. All Manuals Search And Download.
5. TRANSFORMER INSTALLATION
6. APPLY POWER - SELECT DIALING METHOD
PLUG TRANSFORMER
INTO AN UN-SWITCHED
AC OUTLET
1
SLIDE POWER
SWITCH ON
ON
OFF
NOTE: THE DEFAULT DIALING METHOD IS TONE DIALING
IF REQUIRED, CHANGE THE DIALING METHOD BY PRESSING
AND HOLDING THE HOME & CANCEL BUTTONS FOR 3 SECONDS
WHILE SWITCHING THE POWER ON
SECURE
TRANSFORMER
WITH SCREW
2
• 7 BEEPS SOUND FOR PULSE DIALING
• 8 BEEPS SOUND FOR TONE DIALING
7. TRANSMITTER INSTALLATION OPTIONS
8. INSTALLING SINGLE TRANSMITTERS
ADDS ONE TRANSMITTER AT A TIME
FOR ALL TWO-BUTTON TRANSMITTERS
1
2
PRESS CANCEL BUTTON
FOR 5 SECONDS
• CONSOLE ANNOUNCES
THAT IT IS READY TO
LEARN TRANSMITTERS
IT WILL
ACTIVATE
USING
EITHER
BUTTON
WHEN IT IS
INSTALLED
WITH THE
LEFT BUTTON
WHEN IT IS
INSTALLED
WITH THE
IT WILL ONLY
ACTIVATE
USING BOTH
BUTTONS
ACTIVATE TRANSMITTER
• CONSOLE ANNOUNCES
TRANSMITTER TYPE THEN
REVERTS TO STANDBY MODE
RIGHT BUTTON
NOTE: IF THE CONSOLE'S MEMORY
IS FULL, THE CONSOLE WILL ANNOUNCE
"ERROR ZONES FULL"
DXS-62A
FOR DXS-31 STATIONARY TRANSMITTERS
3
REPEAT STEPS 1 & 2 FOR ADDITIONAL
TRANSMITTERS (UP TO 16 TOTAL)
WHEN IT IS INSTALLED
WITH THE JUMPER SET
FOR DELAY
WHEN IT IS INSTALLED
WITH THE JUMPER SET
FOR INSTANT
NOTE: PRESS CANCEL BEFORE TRANSMITTING TO RETURN
TO STANDBY MODE WITHOUT ENTERING A TRANSMITTER
• THE CONSOLE WILL ANNOUNCE ITS STATUS
THE CONSOLE WILL RETURN TO STANDBY MODE AUTOMATICALLY
WHEN A TRANSMITTER IS LEARNED OR AFTER 30 SECONDS
HAS ELAPSED
IT WILL
FUNCTION
AS EMERGENCY
IT WILL
FUNCTION
AS ACTIVITY
NOTE: WRITE DOWN EACH TRANSMITTER'S TYPE AND ZONE NUMBER
WHEN THE TRANSMITTER IS ADDED. THE CENTRAL STATION WILL NEED
TO KNOW THE TYPE AND REPORTING ZONE FOR EACH TRANSMITTER.
FOR MODEL DXS-32: CUT JUMPER (INSTANT TX) FOR USE
AS AN ACTIVITY TX. SEE DXS-32 INSTALLATION INSTRUCTIONS
2
Download from Www.Somanuals.com. All Manuals Search And Download.
9. INSTALLING MULTIPLE TRANSMITTERS
10. TESTING TRANSMITTERS
ADDS MULTIPLE TRANSMITTERS
1
PRESS TEST BUTTON
• CONSOLE ANNOUNCES
TEST MODE
1
PRESS TEST BUTTON
• CONSOLE ANNOUNCES
TEST MODE
2
PRESS TEST BUTTON AGAIN
• CONSOLE ANNOUNCES ITS
READY TO LEARN TRANSMITTERS
2
ACTIVATE EACH OF THE
CONSOLE'S TRANSMITTERS,
ONE AT A TIME
NOTE: EACH TIME THE TEST BUTTON IS PRESSED IN THE ADD
TRANSMITTER MODE, THE CONSOLE WILL BE READY TO INSTALL A
TRANSMITTER IN THE NEXT AVAILABLE SENSOR ZONE LOCATION.
THE CONSOLE WILL ANNOUNCE THE SENSOR ZONE LOCATION NUMBER.
3
ACTIVATE TRANSMITTER
• CONSOLE ANNOUNCES
THE TRANSMITTER TYPE
3
THE CONSOLE WILL ANNOUNCE
THE TRANSMITTER'S TYPE
AND ZONE NUMBER AND ANY
SUPERVISORY CONDITIONS
FOR THE TRANSMITTER
NOTE: IF THE CONSOLE'S MEMORY
IS FULL, THE CONSOLE WILL ANNOUNCE
"ERROR ZONES FULL"
REPEAT STEPS 2 & 3 FOR
ADDITIONAL TRANSMITTERS
(UP TO 16 TOTAL)
DXS-62A
4
5
PRESS CANCEL TO RETURN TO STANDBY MODE
• THE CONSOLE WILL ANNOUNCE ITS STATUS
THE CONSOLE WILL RETURN TO STANDBY MODE
AUTOMATICALLY AFTER 3 MINUTES
4
PRESS CANCEL TO RETURN TO STANDBY MODE
• THE CONSOLE WILL ANNOUNCE ITS STATUS
NOTE: WRITE DOWN EACH TRANSMITTER'S TYPE AND ZONE NUMBER
WHEN THE TRANSMITTER IS ADDED. THE CENTRAL STATION WILL NEED
TO KNOW THE TYPE AND REPORTING ZONE FOR EACH TRANSMITTER
THE CONSOLE WILL RETURN TO STANDBY MODE
AUTOMATICALLY AFTER 3 MINUTES
11. ERASING TRANSMITTERS
12. RECORDING REMINDER MESSAGES
RECORDING REMINDER MESSAGES
1. PRESS AND RELEASE THE TEST BUTTON
CAUTION: ALL TRANSMITTERS
WILL BE ERASED!
TO ENTER TEST MODE
2. PRESS AND HOLD THE HOME BUTTON FOR
5 SECONDS (BE READY TO RECORD)
3. RECORD UP TO A 15 SECOND MESSAGE
(SPEAK AT NORMAL VOLUME, TWO FEET FROM CONSOLE)
4. PRESS THE HOME BUTTON TO STOP RECORDING
5. PRESS THE HOME BUTTON AGAIN TO PLAY THE
MESSAGE JUST RECORDED, PRESS THE HOME
BUTTON TWICE AND HOLD IT AFTER THE 2ND PRESS
TO RECORD ANOTHER MESSAGE IN THE NEXT
AVAILABLE MESSAGE NUMBER
1
PRESS AND HOLD THE TEST BUTTON FOR 5 SECONDS
• CONSOLE ANNOUNCES "ZONE RESET"
(ALL TRANSMITTERS ERASED)
6. REPEAT STEPS 2-5 FOR ADDITIONAL MESSAGES
(UP TO 8 MESSAGES ALLOWED)
7. PRESS AND RELEASE THE CANCEL BUTTON TO
EXIT TEST MODE
TEST PLAYING REMINDER MESSAGES
1. PRESS AND RELEASE THE TEST BUTTON
TO ENTER TEST MODE
2. PRESS AND RELEASE THE HOME BUTTON
3. THE CONSOLE WILL ANNOUNCE EACH MESSAGE
NUMBER, THEN PLAY THE RECORDED MESSAGE
4. PRESS THE HOME BUTTON TO ADVANCE TO
THE NEXT MESSAGE TO BE PLAYED
5. PRESS AND RELEASE THE CANCEL BUTTON TO
EXIT TEST MODE
NOTE: TIME SCHEDULING FOR PLAYING REMINDER
MESSAGES IS SET WITH THE RA-2400 SOFTWARE
ERASING ALL REMINDER MESSAGES
1. PRESS AND RELEASE THE TEST BUTTON
TO ENTER TEST MODE
2. PRESS AND HOLD THE HOME BUTTON FOR
MORE THAN 10 SECONDS
2
PRESS CANCEL TO RETURN TO STANDBY MODE
• THE CONSOLE WILL ANNOUNCE ITS STATUS
3. THE CONSOLE WILL ERASE ALL REMINDER MESSAGES
AND PLAYBACK SCHEDULES
4. PRESS AND RELEASE THE CANCEL BUTTON TO
EXIT TEST MODE
THE CONSOLE WILL RETURN TO STANDBY MODE
AUTOMATICALLY AFTER 3 MINUTES
3
Download from Www.Somanuals.com. All Manuals Search And Download.
13. LOCAL PROGRAMMING WITH RA-2400
PERFORM THE STEPS DETAILED IN THE
14. SYSTEM OPERATION
CONSOLE
STANDING BY
HOME LIGHT SHOWS
ACTIVITY TIMER STATUS
1
RA-2400/PERS REFERENCE GUIDE (P/N 224756)
TO SET UP THE RA-2400 PROGRAM
THE OTHER
POWER
LIGHT
IS ON
TO LOCALLY PROGRAM
2
CONSOLE
LIGHTS ARE
NORMALLY
OFF
THE CONSOLE, CONNECT
THE MODEM'S LINE JACK
TO THE CONSOLE'S LINE
JACK
LINE-TO-LINE
PRESS THE TEST BUTTON
MOMENTARILY
- IN CASE OF EMERGENCY -
3
• CONSOLE ANNOUNCES
TEST MODE
• START A LOCAL CONNECTION WITH
THE RA-2400 SOFTWARE
PRESS AND
MOMENTARILY
HOLD THE
HELP BUTTON
- OR -
ACTIVATE
AN EMERGENCY
TRANSMITTER
NOTE: CONSOLE WILL EXIT TEST MODE
AUTOMATICALLY AFTER 3 MINUTES
OR BY PRESSING THE CANCEL BUTTON
PRESS TRANSMITTER
BUTTON FOR AT LEAST
2 SECONDS
WHEN RA-2400 CONNECTION TIMER
STARTS COUNTING DOWN,
PRESS THE HELP BUTTON
• THE CONSOLE ANNOUNCES
"PHONE ON" WHEN THE CONSOLE
HAS CONNECTED
4
5
CONSOLE
IN ALARM
SENDING LIGHT
FLASHES DURING
COMMUNICATION
THE CONSOLE
ANNOUNCES
EMERGENCY
UNTIL THE CANCEL
BUTTON IS PRESSED,
CENTRAL STATION
REPORT IS COMPLETE,
OR THE SIREN TIME EXPIRES
PRESS CANCEL TO
STOP THE ALARM
(ADDS CANCEL TO
ALARM REPORT IN
PROGRESS)
PERFORM PROGRAMMING IN THE RA-2400 SOFTWARE, SEND AND SAVE
THE ACCOUNT PROFILE, THEN DISCONNECT FROM THE CONSOLE
15. ACTIVITY TIMER OPERATION
16. AUDIO OPERATION
1
ENABLING THE ACTIVITY TIMER
PROGRAM THE CONSOLE FOR THE DESIRED AUDIO MODE:
1) LISTEN-ONLY
2) 2-WAY MANUAL
1
PRESS THE HOME BUTTON FOR
5 SECONDS TO TURN THE
ACTIVITY TIMER ON
3) 2-WAY VOICE ACTIVATED (DUPLEX)
NOTE: THE ACTIVITY TIMER CAN
BE ENABLED FOR 8, 12, OR
24 HOURS, OR DISABLED BY
PROGRAMMING
AFTER THE CENTRAL STATION OPERATOR ACKNOWLEDGES THE
COMMUNICATOR'S REPORT, THE CONSOLE WILL STOP ITS LOCAL
ALARM AND SWITCH TO THE SELECTED AUDIO MODE FOR 90 SECONDS
2
AUDIO OPERATION CAN ALSO BE ESTABLISHED BY CALLING THE
CONSOLE TWICE AND KEYING IN THE FOUR DIGIT REMOTE AUDIO
PASSWORD FOLLOWED BY THE # KEY (SEE SECTION #22)
NOTE: TO TURN THE ACTIVITY TIMER OFF, PRESS THE HOME BUTTON FOR FIVE
SECONDS, THIS WILL ALSO TURN ON/OFF REMINDER REPORTING IF REMINDER
SCHEDULE TIME OUT REPORT IS ENABLED IN THE RA-2400 SOFTWARE (ACTIVITY
TIMER CAN BE DISABLED SEPARATELY IN RA-2400 ALARMS PROGRAMMING)
3
CENTRAL STATION OPERATOR OR AUDIO CALLER:
USE THESE KEYS ON A TELEPHONE SET TO CONTROL THE CONSOLE'S
AUDIO MODE (IN CENTRAL STATION, CONNECT A TELEPHONE TO THE
SAME TELEPHONE LINE AS THE CENTRAL STATION RECEIVER)
2
3
ACTIVITY TIMER RUNNING
TIME RUNNING OUT
THE HOME INDICATOR WILL BLINK &
THE CONSOLE ANNOUNCES EVERY
10 SECONDS WHEN LESS THAN
15 MINUTES OF TIME REMAIN
THE HOME INDICATOR WILL LIGHT,
SHOWING THAT THE ACTIVITY
TIMER IS ENABLED
ALL
PRESS "9" TO
AUDIO
DISCONNECT THE CALL
MODES
PRESS "3" TO LISTEN
2-WAY
MANUAL
AUDIO
MODE
PRESS "1" TO TALK
4
RESETTING ACTIVITY TIMER
THE SUBSCRIBER NEEDS TO
PRESS THE HOME BUTTON OR
TRIGGER AN ACTIVITY
TRANSMITTER BEFORE THE
TIMER EXPIRES TO PREVENT
AN INACTIVITY REPORT
THE "1" AND "3" KEYS ARE NOT USED IN LISTEN-ONLY AUDIO MODE
THE "1" AND "3" KEYS CAN BE USED IN 2-WAY VOICE ACTIVATED MODE.
USING THEM WILL CAUSE THE SYSTEM TO SWITCH TO 2-WAY MANUAL
MODE. PRESSING "8" RESTORES 2-WAY VOICE ACTIVATED MODE.
PRESSING ANY KEY OTHER THAN "9" WILL EXTEND THE AUDIO TIME
90 MORE SECONDS
PRESSING THE HOME BUTTON
RESTARTS THE TIMER
4
Download from Www.Somanuals.com. All Manuals Search And Download.
17. SPEAKERPHONE OPERATION
PROGRAM THE CONSOLE TO ENABLE SPEAKERPHONE MODE
18. SUPERVISORY CONDITIONS
TRANSMITTER LOW BATTERY
WHEN A TRANSMITTER SENDS A LOW BATTERY SIGNAL THE
CONSOLE'S STATUS INDICATOR WILL BLINK SLOWLY
1
NOTE: SPEAKERPHONE MODE CAN ONLY BE USED ON INCOMING CALLS
AND WILL NOT FUNCTION WHEN THE CONSOLE IS OPERATING FROM
BACKUP BATTERY POWER. ATTEMPTING TO ANSWER WHILE THE CONSOLE
IS ON BACKUP POWER WILL CAUSE AN EMERGENCY ALARM.
TRANSMITTER SUPERVISORY
IF A TRANSMITTER PROGRAMMED AS SUPERVISED FAILS
TO REPORT STATUS IN A 24 HOUR PERIOD, THE CONSOLE'S
STATUS INDICATOR WILL LIGHT
WHILE THE TELEPHONE IS RINGING, ACTIVATING AN EMERGENCY
TRANSMITTER OR PRESSING THE HELP BUTTON WILL SEIZE THE
TELEPHONE LINE
2
LOW CONSOLE BACKUP BATTERY
THE CONSOLE WILL TEST ITS BACKUP
BATTERY EVERY 8 HOURS. IF THE BATTERY
WHEN TROUBLE EXISTS,
TESTS LOW, THE POWER INDICATOR WILL
BLINK SLOWLY WHILE THE CONSOLE
CHARGES THE BATTERY.
THE CONSOLE WILL ANNOUNCE
THE CONDITION EVERY
HOUR UNTIL CLEARED
THE SYSTEM WILL BEGIN 2-WAY VOICE ACTIVATED MODE FOR AUDIO
COMMUNICATIONS
3
WARNING: THE CONSOLE WILL NOT BE ABLE
TO REPORT AN EMERGENCY WHILE THE
SPEAKERPHONE MODE IS ACTIVE
NOTE: TO REPORT AN EMERGENCY WHEN THE SPEAKERPHONE IS IN USE,
PRESS EITHER THE HELP BUTTON OR THE BUTTON ON AN EMERGENCY
TRANSMITTER TO END THE PHONE CALL AND THEN PRESS ONE OF THESE
BUTTONS AGAIN TO REPORT THE EMERGENCY.
CLEARING STATUS INDICATIONS
PRESS THE CANCEL BUTTON TO
CLEAR THE STATUS INDICATOR
AND STOP THE ANNUNCIATION
NOTE: A TROUBLE
CONDITION WILL LIKELY OCCUR AGAIN
UNTIL THE PROBLEM IS FIXED
THE CONSOLE WILL END AUDIO COMMUNICATIONS WHEN EITHER:
4
THE CONSOLE WILL REPORT THESE SUPERVISORY CONDITIONS
TO THE CENTRAL STATION IF ENABLED IN THE COMMUNICATOR
PROGRAMMING (AUTOMATIC FULL-SYSTEM CONSOLE SUPERVISORY
REPORTS CAN ALSO BE SENT AT 1, 7, OR 30 DAY INTERVALS)
• THE HELP BUTTON IS PRESSED
• ANY EMERGENCY TRANSMITTER IS ACTIVATED
• THE AUTO-DISCONNECT TIMER RUNS OUT (PROGRAMMABLE
FOR 15, 30, OR 60 MINUTES)
19. OPTIONAL OUTPUT CONNECTIONS
20. OPTIONAL REMOTE SPEAKER/MIC CONNECTIONS
USE A SMALL
SCREWDRIVER
TO PUSH ON THE
TERMINAL LOCKS
TO ALLOW THE
WIRE TO SLIDE
INTO THE TERMINAL
HOLE
REMOVE
JUMPER
WIRE
SHIELD
WIRE
NOTE: NOT TO BE USED IN
UL INSTALLATIONS
CONSOLE MAIN
CONNECT WIRES TO
RELAY TERMINALS
CONSOLE MIC TERMINALS
TERMINAL BLOCK
EXTERNAL MICROPHONE (+)
1 - +12 VDC @ 100 Ma
MICROPHONE COMMON (-)
2 - RELAY NORMALLY OPEN
NOTE: THE CONSOLE IS SHIPPED WITH
REMOVABLE LABELS COVERING SOME
OF THE REAR CONNECTIONS
3 - RELAY NORMALLY OPEN
4 -
GROUND
5 - EXTERNAL SPEAKER
NOTE: REMOVE
JUMPER WHEN
USING A UL
LISTED MODEL
RSM-1 OR OTHER
UL LISTED
CONSOLE
TERMINAL BLOCK
TERMINALS 2 & 3
ARE A FORM "A" DRY
CONTACT RELAY
RATED 24 VOLTS AT
1 AMP MAXIMUM
NOTE: NOT TO BE USED IN
UL INSTALLATIONS
USE
SHIELDED
TWISTED PAIR
WIRE FOR
MICROPHONE
SPEAKER
NOISEMAKER OR
VISUAL INDICATOR
TERMINALS 4 & 5
HAVE A FACTORY
INSTALLED JUMPER
DO NOT REMOVE
UNLESS USING
A UL LISTED MODEL
RSM-1 OR OTHER
UL LISTED SPEAKER
SPKR
SPKR
EXTERNAL
POWER SUPPLY
MIC (-)
MIC (+)
LINEAR
MODEL RSM-1
RSM-1
TERMINALS
RSM-1 TERMINAL BLOCK
5
Download from Www.Somanuals.com. All Manuals Search And Download.
21. NON-VOICE PROMPT BEEP LEGEND
WHEN VOICE PROMPTS ARE DISABLED,THE CONSOLE WILL SOUND BEEPS
TO INDICATE VARIOUS CONDITIONS
23. WIRELESS INSTALLATION TIPS
WALLBOARD AND
WOOD STUDS
CONSTRUCTION
1 BEEP...................TRANSMITTER ADDED TO A SENSOR ZONE
1 LONG BEEP........SENSOR ZONES FULL
MATERIALS WILL
AFFECT RADIO RANGE
90% - 100%
OF FULL POWER
2 BEEPS.................CONSOLE IN STANDBY MODE, ACTIVITY TIMER ON
3 BEEPS.................CONSOLE IN STANDBY MODE, ACTIVITY TIMER OFF
4 BEEPS.................CONSOLE IN TEST MODE
CONCRETE WITH STEEL
REINFORCEMENT OR
METAL LATH AND PLASTER
LIGHT CONCRETE
OR BRICK
5 BEEPS.................CONSOLE IN LEARN TRANSMITTER MODE
6 BEEPS.................CONSOLE HAS ERASED ALL TRANSMITTERS
7 BEEPS.................PULSE DIALING HAS BEEN SELECTED FOR THE CONSOLE
8 BEEPS.................TONE DIALING HAS BEEN SELECTED FOR THE CONSOLE
65% - 95%
OF FULL POWER
10% - 70%
OF FULL POWER
22. REMOTE AUDIO COMMUNICATIONS
TO ESTABLISH REMOTE AUDIO COMMUNICATIONS WITH THE CONSOLE:
CENTRALLY LOCATE CONSOLE
1. DIAL THE NUMBER OF THE PHONE LINE THAT THE CONSOLE IS CONNECTED TO,
ALLOW IT TO RING THREE TIMES, THEN HANG UP
2. WAIT ABOUT 10 SECONDS, THEN CALL THE CONSOLE A SECOND TIME
CONSOLE LOCATION WILL
AFFECT RADIO RANGE
RIGHT
3. WHEN THE CONSOLE ANSWERS, A TONE WILL SOUND FOR ABOUT 8 SECONDS
4. WHEN THE TONE ENDS, ENTER THE REMOTE AUDIO PASSWORD (DEFAULT 1234)
FOLLOWED BY THE # KEY
THE CONSOLE WILL BEGIN AUDIO COMMUNICATIONS (2-WAY DUPLEX, 2-WAY MANUAL,
OR LISTEN-ONLY) DEPENDING ON THE CONSOLE’S PROGRAMMING
TRANSMITTERS AT THE
OTHER END OF THE HOUSE
MIGHT BE TOO FAR AWAY
SEE SECTION 16 FOR AUDIO COMMUNICATION OPERATION DETAILS
NOTE: IFTHE CONSOLE AUDIO IS SETTO “NONE” REMOTE AUDIO COMMUNICATIONS
WILL NOT FUNCTION
WRONG!
SPECIFICATIONS
CODING FORMATS
LINEAR LIMITED WARRANTY
This Linear product is warranted against defects in material and workmanship for twelve
(12) months.This warranty extends only to wholesale customers who buy direct from Linear
or through Linear’s normal distribution channels. Linear does not warrant this product to
consumers. Consumers should inquire from their selling dealer as to the nature of the
dealer’s warranty, if any. There are no obligations or liabilities on the part of Linear LLC
for consequential damages arising out of or in connection with use or performance
of this product or other indirect damages with respect to loss of property, revenue, or
profit, or cost of removal, installation, or reinstallation. All implied warranties, including
implied warranties for merchantability and implied warranties for fitness, are valid only until
the warranty expires. This Linear LLCWarranty is in lieu of all other warranties express
or implied.
DX CODE 24-BIT A1D PPM 1,048,567 CODES
DXS CODE 24-BIT A1D PPM 1,048,567 CODES
POWER REQUIREMENTS
PRIMARY SUPPLY 16 VAC 1.25 AMP PLUG-IN TRANSFORMER
BACKUP SUPPLY 12 VOLT 1.2 AMP/HOUR BATTERY
NORMAL OPERATING CURRENT 120 mA
ALARM OPERATING CURRENT 190 mA
BACKUP OPERATION DURATION 30 HOURS MINIMUM
BATTERY LIFE EXPECTED 3 YEARS
BACKUP SUPPLY OPERATING CURRENT 40 mA
BATTERY FUSE 1 AMP, TYPE 2AG
All products returned for warranty service require a Return Product Authorization Number
(RPA#). Contact Linear Technical Services at 1-800-421-1587 for an RPA# and other
important details.
RF CHARACTERISTICS
RECEIVER TYPE DUAL-DIVERSITY SUPERHETERODYNE
RF SENSITIVITY -100 dBm TYPICAL
RADIO FREQUENCY 315 Mhz
MAXIMUM RANGE 400 FEET LINE-OF-SIGHT, OPEN RANGE
UL LISTED FOR 150 FEET
IMPORTANT WIRELESS INFORMATION
Linear radio controls provide a reliable communications link and fill an important need in
portable wireless signaling. However, there are some limitations which must be observed.
COMMUNICATOR
COMMUNICATOR FORMATS 4 BY 2 (GENERAL OR POINT-TO-POINT),
ADEMCO CONTACT ID
•
For U.S. installations only: The radios are required to comply with FCC Rules and
Regulations as Part 15 devices. As such, they have limited transmitter power and
therefore limited range.
UL TESTED WITH SURGARD SGMLRG-DG
CENTRAAL STATION RECEEIVER
DIALING TONE OR PULSE
DIALING DELAY 0 OR 15 SECONDS
REMOTE PROGRAMMING PASSWORD 6 DIGITS
•
A receiver cannot respond to more than one transmitted signal at a time and may be
blocked by radio signals that occur on or near their operating frequencies, regardless of
code settings.
NOTE:THE INSTALLER MUST
REMOTE COMMUNICATIONS PASSWORD 4 DIGITS
VERIFY COMPATIBILITY
ACCOUNT NUMBER LENGTH 4 DIGITS
PRIMARY TELEPHONE NUMBER 20 DIGITS
BETWEEN THE PERS-3600B AND
•
•
Changes or modifications to the device may void FCC compliance.
THE UL LISTED DIGITAL ALARM
SECONDARY TELEPHONE NUMBER 20 DIGITS
SUPERVISORY TELEPHONE NUMBER 20 DIGITS
CONSOLE STATUS REPORT OPTIONAL 1, 7 OR 30 DAY REPORT
COMMUNICATOR RECEIVER
Infrequently used radio links should be tested regularly to protect against undetected
interference or fault.
•
A general knowledge of radio and its vagaries should be gained prior to acting as a
wholesale distributor or dealer, and these facts should be communicated to the ultimate
users.
CONSOLE
OPERATING TEMPERATURE 32°-120° F (0-49° C)
HELP ALARM DURATION 1-30 MINUTES
ACTIVITY TIMER PERIOD 8, 12, OR 24 HOURS
AUDIO LISTEN-ONLY, 2-WAY MANUAL, 2-WAY VOICE
ACTIVATED (DUPLEX) SPEAKERPHONE
RELAY OUTPUT N.O. @ 24 VOLT AC/DC 1 AMP
MAXIMUM, NEC CLASS II CIRCUITRY