Operator's manual
Video Sweep Generator
MVG 10
Order no.: 208 320
Download from Www.Somanuals.com. All Manuals Search And Download.
Preface
Preface
Dear customer,
This handbook aims to help you use the various functions of the MVG10 in
the optimal manner. Please pay attention to all instructions. Kathrein-Werke
KG has made every effort to ensure the information and descriptions are
correct and complete.
We reserve the right to make changes to this handbook without prior notice.
In particular, this applies to changes made due to technical advancements.
We are always grateful to receive your comments and suggestions for
improvement.
Prior written consent from Kathrein-Werke KG is required for publishing,
copying, reprinting or electronically reproducing this handbook or parts
thereof.
All product names and trademarks in this handbook are the property of the
respective companies.
Please read the safety notes carefully!
Yours,
The KATHREIN team
Customer service
Please send the unit to the following address in case of defect or for
calibration:
Fa. ESC
Kathrein-Zentralkundendienst
Bahnhofstraße 108
83224 Grassau
Germany
Tel.: +49 8641 9545-25
Fax: +49 8641 9545-35
E-mail: [email protected]
3
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Contents
Preface .....................................................................................................................................................3
Contents ...................................................................................................................................................5
Safety notes..............................................................................................................................................6
Safety notes..............................................................................................................................................7
System description and usage .................................................................................................................8
Views, connections and controls........................................................................................................... 11
Button functions..................................................................................................................................... 12
System settings ..................................................................................................................................... 13
Noise generator..................................................................................................................................... 15
Signal generator .................................................................................................................................... 16
Sweep generator ................................................................................................................................... 18
Channel sweep generator ..................................................................................................................... 22
Recall..................................................................................................................................................... 26
Technical appendix................................................................................................................................ 27
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety notes
Safety notes
Validity of handbook
This handbook is valid for MVG 10, order no.: 208 320.
The following notes are important for operating the MVG 10 and
should be observed under all circumstances.
General safety notes
The MVG 10 was developed and produced in compliance with
the relevant harmonised guidelines, standards and additional
technical specifications. The product is state-of-the-art and
ensures the maximum level of safety.
However, this safety level can only be reached in practice if all
of the necessary measures are taken and is subject to the care
taken by the operator.
Symbols used
The following symbols are used in this operator's manual.
The main aim of these symbols is to bring the user's attention to
the text opposite the respective symbol.
Danger! Live component!
This symbol indicates danger to life and health.
Attention!
This symbol indicates that particular attention must be paid to
this section of the manual.
Example
This symbol indicates an example of the measurement function
being explained.
ꢀ
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety notes
Safety notes
Always observe the VDE safety regulations.
Observe the maximum permissible signal feed-in level.
Neither DC voltage nor low-frequency AC voltage may be applied
to the RF port.
Only use fuses with the same cut-out characteristics.
The unit is live even when not connected.
The unit may only be operated with all shielding covers fitted and
when closed to prevent electromagnetic interference. Only use
suitable shielded cable.
Improper use during mains operation is a risk to life!
Connections
Improperly connected connections can lead to operating faults or
defects in the unit.
Use in accordance with intended purpose
The operator must ensure that
the measuring instrument is only used in accordance with its intended
purpose.
the measuring instrument is only used when in good order and fully
functional.
the safety and warning notes on the measuring instrument are not
removed and remain legible.
Mechanical durability
The MVG 10 is designed for mobile usage and the mechanical
demands associated with this. The MVG 10 should not be exposed to
heavy mechanical stress such as being struck, knocked or dropped as
this can cause damage to the unit.
Electronic equipment must not be disposed of in domestic waste.
According to directive 2002/96/EC OF THE EUROPEAN
PARLIAMENT AND COUNCIL of 27 January 2003 on waste electrical
and electronic equipment, it must be disposed of professionally.
Please take this unit to a public collection point intended for this type
of disposal at the end of its useful life.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
System description and usage
System description and usage
The MVG 10 Video Sweep Generator is designed to selectively sweep
and measure the forward path and return path of television cable
installations without interfering with the connected subscribers. The
required sweep ranges, channels and signal levels (36…100 dBµV)
can be programmed step-by-step. The channels can be set to various
standards (e.g. B/G etc.).
Frequency ranges that are in use must be left out so that the
subscriber reception is not disrupted.
Make a frequency and signal level plan before performing the
measurement!
Before each frequency change, the MVG 10 transmits the next
reception frequency to the MSK 33 with a telemetric signal. In ranges
that are in use and which may not be sweeped, the MSK 33 can, as
an option, continue measurement in the same channel plan at
maximum scan speed. The MVG 10 pauses during this period.
The MSK measures the signal levels transmitted and presents them in
the spectrum. The measurement can also be printed out with the
integrated printer.
Measurement principle
MVG10
MSK33
Message FRQa+1
Message FRQy
FRQa
Transmission
range
FRQa
Recording
FRQx
Transmission-free
range
FRQx
quick
sweep
Message FRQz
FRQy
Transmission
range
FRQy
FRQz
Recording
Message FRQa
FRQz
Return path measurement
The MVG 10 is also suitable for measuring the return path – terminal
outlet to head-end – in the frequency range 4.0 MHz…80MHz.
For this, the MSK 33 must be equipped with the return path option.
The MVG 10 Video Sweep Generator must be modulated with a video
signal.
Although measurement can be performed in cables that are in use,
the ranges that are in use may not be sweeped. In addition, the
MVG 10 signal generator can sweep in several partial ranges i.e.
ranges that are in use can be left out by the sweep generator so that
TV reception is not interrupted. The MVG 10 sweep generator informs
the MSK 33 of the next respective reception frequency over the RF
channel.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
System description and usage
Downstream measurement: 47…860 MHz
Head-end (or transfer point)
Transfer point (or terminal outlet)
Cable installation
MVG 10
Sweep
MSK 33
receiver
generator
with return path-capable amplifiers
Upstream measurement: 4.0…80 MHz
Head-end (or transfer point)
Transfer point (or terminal outlet)
Cable installation
MSK 33
receiver
MVG 10
Sweep
generator
with return path-capable amplifiers
Comfortable upstream 4.0…80 MHz and
Downstream 47…860 MHz measurement
Cable installation
MVG10
Sweep
generator (1)
MSK33
receiver (1)
with return path-capable
amplifiers
ext CVBS
Graphics
RGB
MVG10 (2)
Signal
sweep
MSK33
receiver (2)
generator
Presentation of upstream spectrum at head-end
Measurement of downstream spectrum
Upstream measurement:
The MVG 10 (1) sweeps the return path in the free frequency ranges.
The MSK 33 (1) receives the signals in MVG 10 – tracking mode.
The graphics from the MSK33 (1) are provided to the MVG10 (2) in
RGB (60 Hz) and transmitted to the MSK33 (2) on a free downstream
channel (attention: double-sideband modulation).
9
Download from Www.Somanuals.com. All Manuals Search And Download.
System description and usage
The spectrum at the head-end can be printed out from the MSK 33 (1)
using the print command "Prt 999" on the MVG 10 (1).
The command "Clear" clears the spectrum presentation on the
MSK 33 (1).
Downstream measurement:
The MVG 10 (2) is switched from signal generator mode to
downstream sweep generator mode via the MSK 33 (1) using the
command "Fkt A 1" on the MVG 10 (1).
The MVG 10 (2) sweeps the forward path in the free frequency
ranges.
The MSK33 (2) receives the signals in MVG 10 – tracking mode.
The MVG 10 (2) is switched back to signal generator mode with the
command "Fkt A 0" on the MVG 10 (1). This is performed via the
turn-on voltage of the SCART line.
Uscart = 12 V effects recall no. 1 on MVG 10.
Uscart = 0 V effects recall no. 0 on MVG 10.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Views, connections and controls
Views, connections and controls
Left-hand side
Scart socket
Right-hand side
RS232 socket
Power supply
Frontal view
RF output
Display
Keypad
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Button functions
Button functions
The following information explains the operation of the MVG 10.
Please use the illustration of the operating concept for assistance.
Button
Function
Display
KATHREIN MVG10
1. Switch on unit when
depressed briefly
On
Off
The upper display appears first,
then the lower display with the
last function that was set
NG Store
nr: 0
2. Switch off unit when
depressed for an extended
period
Menu
Confirm selected main or subfunction (see below)
Select main and subfunctions
Select
Example: Setting Sys Backlight (display backlight).
▼ ▲
T
Press
button until "System…" appears in the display,
ꢀ
ꢀ
Menu
then press
, Sys FactSetup is displayed.
Select
T
Use
to set to Sys Backlight and
W X
use
to switch backlight on/off.
- +
W X
Set frequency, channel and signal level values
- +
Numerical entry of frequency, channel and signal level values
0
·
9
S-Ch
Preselection button for special channels
·
ꢀ
Example: Set special channel 25 =
S-Ch
2
5
ꢀ
"Enter" button for confirming numerical entries
12
Download from Www.Somanuals.com. All Manuals Search And Download.
System settings
System settings
We recommend checking the factory settings before initial start-up.
Please use the operating concept on page 2 for help with
understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
NG Store
nr: 0
then the lower display shortly
afterwards with the last function
that was set
Select "System" main function
and
System...
Accu 100 %
▼ ▲
show battery charge state
Sys FactSetup
Confirm selected function
Menu
Call up factory settings with
"Enter"
Select
Sys FactSetup
restored
ꢀ
Recall
Switches to display:
nr:
0
Sys FactSetup
Reselect "System" main
function
▼
▼
Menu
Select
Sys Backlight
on/off
Display backlight
on/off
W X
- +
Display system software
version
Sys Software
V1.3x 30.05.01
▼
▼
Sys SeriesNr.
XXX
Display serial number
13
Download from Www.Somanuals.com. All Manuals Search And Download.
System settings
Sys Unit
dBµV (dBmv, dBm)
▼
▼
W X
Display units
+
- +
Sys Baudr: 19.2k
System baud rate
setting not possible
Sys RTS/CTS
on
▼
Transmission readiness
Confirm entries
setting not possible
Menu
System..
Accu 100 %
Select
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Noise generator
Noise generator
The noise generator is used for measuring cable installations or
sections of cable installations that are not yet occupied with
programming.
It generates a broadband noise to evaluate the frequency response in
the range from 4 MHz to 1 GHz. The bandwidth can be set to 1 MHz
or 7/8 MHz.
Do not use the noise generator if programmes or services are fed
into the cable network and/or if subscribers are connected!
Please use the operating concept on page 2 for help with
understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
NG Store
nr: 0
then the lower display shortly
afterwards with the last function
that was set
Select the "Noise generator"
main function
NoiseGenerator..
Accu 100 %
▼ ▲
Show battery charge state
Confirm selected function
Show last signal level set
Menu
NG Level
XX.X dBµV
Select
W X
NG Level
60.0 dBµV
Set signal level
- +
or
Enter signal level directly
followed by "Enter"
ꢀ
see above
0
9
▼
W X
NG Bandwidth
1 MHz (7/8 MHz)
Set bandwidth
1 MHz or 7/8 MHz
+
- +
Store settings in memory at
addresses 0 to 9
NG Store
Nr: 0
▼
Menu
NoiseGenerator..
Accu 100 %
Confirm entries
Select
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Signal generator
Signal generator
The signal generator is needed to transmit test signals and the screen
view from the MSK 33 during "Comfortable Up/Downstream
measurement" (see system description). In addition to an internal test
picture of coloured bars it has an external modulation input
(CVBS/RGB + sound) and is double-sideband modulated.
The channels/frequencies and levels can be selected freely.
Only use the signal generator on free channels so that
programmes and services in the cable network are not
interrupted. Please use the operating concept on page 2 for help
with understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
then the lower display shortly
afterwards with the last function
that was set
NG Store
nr: 0
SignalGenerator..
Accu 100 %
Select "Signal generator"
main function and
▼ ▲
show battery charge state
SG Level
XX.X dBµV
Confirm selected function
Show last signal level set
Menu
Select
SG Level
60.0 dBµV
W X
Set signal level
- +
or
Enter signal level directly
followed by "Enter"
ꢀ
see above
0
9
SG Frequency
224.25 MHz
▼
W X
Set transmission frequency
in MHz
+
- +
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Signal generator
Enter transmission
or
frequency directly followed
by "Enter"
Channel frequency inputs with
incorrect entries after the point
are not accepted!
ꢀ
see above
0
9
Channel entry -
special channels with press
▼
W X
SG Channel
C __
+
·
- +
S-Ch
first!
or
Enter channel
number directly special
channels with press
ꢀ
see above
0
9
·
S-Ch
first!
·
Also switch back
to normal channel
entry with
S-Ch
▼
▼
W X
SG Standard
B/G
Set standards
B/G, D/K, I, M, Mj, H
+
- +
Set modulation
int-AV, ext-CVBS, ext-RGB,
ext-FM, int-FM
W X
SG Modulation
int-AV
+
+
- +
SG Store
Nr: 0
▼
W X
Store at addresses 0 to 9
- +
ꢀ
SignalGenerator..
Accu 100 %
Confirm entries
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Sweep generator
Sweep generator
The sweep generator is used to sweep in 10 preset frequency ranges
(measuring profiles). The profiles can be programmed to free ranges
depending on the channel allocation in the cable network, so that
measurement is possible without interrupting or interfering with
programmes. The frequencies and levels can be selected freely.
Only use the sweep generator on free channels so that
programmes and services in the cable network are not
interrupted.
Please use the operating concept on page 2 for help with
understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
NG Store
nr: 0
then the lower display shortly
afterwards with the last function
that was set
Select "Sweep generator"
main function and
show battery charge state
SweepGenerator..
Accu 100 %
▼ ▲
SwG Level
XX.X dBµV
Menu
Confirm selected main function
Show last level set
Select
SwG Level
60.0 dBµV
W X
Set signal level
- +
or
Enter signal level directly
plus "Enter"
see above
ꢀ
0 9
Enter number of measuring
profile to be pre-programmed,
0 to 9
▼
W X
SwG FreqRange
Nr: X
+
Incorrect entries are not
accepted.
- +
Enter profile directly
or
plus "Enter"
ꢀ
see above
0 9
Incorrect entries are not
accepted!
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Sweep generator
▼
W X
SwG FreqRange X
Start XXX.XX MHz
+
Enter start frequency
- +
or
or
or
or
Enter frequency directly
Enter end frequency
Enter frequency directly
see above
ꢀ
0 9
SwG FreqRange X
Stop XXX.XX MHz
▼
W X
+
- +
see above
ꢀ
0 9
▼
W X
50-kHz steps from one
SwG FreqRange X
frequency to next in MHz
Step
X.XX MHz
+
- +
Enter frequency directly
see above
ꢀ
0 9
Step duration from one
frequency to next in ms
(150 ms – 2500 ms)
▼
W X
SwG FreqRange X
XXX ms
+
- +
Enter time directly
see above
ꢀ
0 9
Tracking on/off
Switch sequential control
of MSK 33 on/off
▼
▼
W X
SwG Tracking
on/off
+
No other analysers
can be controlled.
- +
Select pre-programmed
measuring profiles (0…9) by
entering numbers without space
followed by "Enter" – the
sequence is started and can
be tracked on the MSK 33.
W X
ꢀ
SwG SelRange
123....... no run/run
+
+
+
-
+
0…9
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Sweep generator
▼
▼
X ꢀ
Clear screen presentation on
MSK 33 with "Enter"
SwG ControlData
+
+
Clear
no run
+
Command to clear spectrum
SwG ControlData
received by MSK33
(MVG10 tracking)
Clear
no run/run
Enter control printout number
(up to 3 digits), confirm with
"Enter" and run with "Enter"
X
ꢀ ꢀ
SwG ControlData
Prt XXX
no run
+
+
+
0…..9
Command to switch a second
MVG10 to stored position
"Recall 0" via MSK33
(MSK33 and MVG10
connected via SCART)
X
SwG ControlData
Fkt
A
0
no run
+
Command to switch a second
MVG10 to stored position
"Recall 1" via MSK33
SwG ControlData
Fkt no run
X
A
1
(MSK33 and MVG10
connected via SCART)
+
Command to switch the
MSK33 video signal from
internal graphics to ext. CVBS.
The respective signals are
relayed to the MVG10 via the
SCART socket for modulation.
X
SwG ControlData
Fkt no run
B
1
+
Command to switch the
MSK33 video signal from
ext. CVBS to internal graphics.
The respective signals are
relayed to the MVG10 via the
SCART socket for modulation.
X
SwG ControlData
Fkt no run
B
0
+
SwG Store
nr: 0
Store settings in memory at
addresses 0 to 9
▼
ChannelSweep...
Accu 100 %
ꢀ
Confirm entries
20
Download from Www.Somanuals.com. All Manuals Search And Download.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Channel sweep generator
Channel sweep generator
The channel sweep generator is used to sweep in 10 preset ranges.
The channels can be programmed to free ranges depending on the
channel allocation in the cable network, so that measurement is
possible without interrupting or interfering with programmes. The
channels and signal levels can be selected freely.
Only use the channel sweep generator on free channels so that
programmes and services in the cable network are not
interrupted.
Please use the operating concept on page 2 for help with
understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
NG Store
nr: 0
then the lower display shortly
afterwards with the last function
that was set
ChannelSweep...
Accu 100 %
Select "Channel sweep
generator" main function and
▼ ▲
show battery charge state
Confirm selected
main function
Show last level set
Menu
CSw Level
XX.X dBµV
Select
CSw Level
60.0 dBµV
W X
Set signal level
- +
or
Enter signal level directly
followed by "Enter"
see above
ꢀ
0 9
Enter pre-programmed
channel sequence (0…9)
CSw ChanRange
Nr: X
▼
W X
+
Incorrect entries are
not accepted!
- +
or
Enter pre-programmed
channel sequence
directly + "Enter"
ꢀ
see above
0 9
Incorrect entries are
not accepted!
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Channel sweep generator
Enter start channel
Special channels with press
▼
W X
CSw ChanRange X
Start C 12
·
+
S-Ch
first!
- +
or
Enter channels directly
see above
ꢀ
0 9
Enter end channel
Special channels with press
▼
W X
·
CSw ChanRange X
Stop C 30
+
S-Ch
first!
- +
or
Enter channels directly
ꢀ
0 9
see above
·
Also switch back to normal
channel entry with
S-Ch
▼
W X
Channel steps
(skip a previously defined
number of channels)
CSw StepChan X
+
Step
X
- +
or
Enter channel steps directly
see above
ꢀ
0 9
▼
W X
CSw StepTime
XXX ms
Step duration from one
channel to next in ms
+
- +
or
Enter time directly
see above
ꢀ
0 9
▼
W X
Modulation
CSw Modulation
Tracking
+
Tracking/int-AV/ext-CVBS/
ext-RGB/ext-FM/int-FM
- +
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Channel sweep generator
▼
W X
CSw Standard
B/G
Set standards
B/G, D/K, I, M, Mj, H
+
- +
Set the pre-programmed
channel ranges to be
processed (0…9)
CSw SelRange
……… no run/run
▼
Command to clear spectrum
CSw ControlData
Clear
received by MSK33 (MVG10
tracking)
run
▼
Enter control printout number
(up to 3 digits), confirm with
"Enter" and run with "Enter"
X
ꢀ ꢀ
CSw ControlData
Prt XXX
no run
+
+
+
0…..9
Command to switch a second
MVG10 to stored position
"Recall 0" via MSK33
CSw ControlData
X
Fkt
A
0
no run
(MSK33 and MVG10
connected via SCART)
+
Command to switch a second
MVG10 to stored position
"Recall 1" via MSK33
CSw ControlData
Fkt no run
X
A
1
(MSK33 and MVG10
connected via SCART)
+
Command to switch the
MSK33 video signal from
internal graphics to ext. CVBS.
The respective signals are
relayed to the MVG10 via the
SCART socket for modulation.
X
CSw ControlData
Fkt no run
B
1
+
Command to switch the
MSK33 video signal from
ext. CVBS to internal graphics.
The respective signals are
relayed to the MVG10 via the
SCART socket for modulation.
X
CSw ControlData
Fkt no run
B
0
+
Store settings in memory at
addresses 0 to 9
CSw Store
nr: 0
▼
ChannelSweep...
Accu 100 %
ꢀ
Confirm entries
Enter
24
Download from Www.Somanuals.com. All Manuals Search And Download.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Recall
Recall
The "Recall" setting is used to call up the setting stored at addresses
0…9 and start the measuring process.
Please use the operating concept on page 2 for help with
understanding the operating sequence.
Function
Button actuation
Display
KATHREIN MVG10
Switch on unit when depressed On
briefly
Off
The upper display appears first,
NG Store
nr: 0
then the lower display shortly
afterwards with the last function
that was set
Select "Recall" main function
and
Recall
nr: 0
▼ ▲
no.: X
Use the arrow buttons to set the
settings stored at an address
(0…9).
W X
- +
or
Enter number directly and
confirm with "Enter"
see above
ꢀ
0 9
ꢀ
Confirm entry
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Technical appendix
Technical appendix
Technical data
Frequency range:
Frequency adjustment:
Frequency resolution:
Display:
4.0 MHz...1000 MHz
50 kHz
62.5 kHz
LCD alphanumeric 2 x 16
characters, bar chart; illuminated
Adjustment:
direct frequency and channel
entry; +/- step buttons
Sweep ranges:
Channel hopping:
10 ranges with start/stop
and frequency step entry
10 ranges with start/stop
and channel step entry
Output signal level:
Signal level accuracy:
Signal level resolution:
Interference level:
36 dBµV ...100 dBµV
2dB
0.1 dB
≤40 dBµV (to 90 dBµV -
output signal level)
Noise generator:
4.0 MHz...1000 MHz,
Standing wave ratio 2dB
Noise level:
85 dBµV (BW 1MHz)
1 dB
Noise level resolution:
Output:
75 Ω BNC socket
Modulation (double-sideband): Test picture, SCART-CVBS
SCART RGB
FM sound modulation
Power supply:
Internal lead-acid battery
12 V/2.8 Ah
External mains adaptor
for operating and charging
230 V ~ /50-60 Hz
Weight:
approx. 3 kg (incl. leather bag)
Dimensions, H x W x D:
90(115) x 162 x 235 mm³
() = incl. bag accessory
EMC/EMD:
CE
Included in delivery:
Leather bag with carry strap,
charger
27