Invacare® TheraPure™Supine Whirlpool Tubs
Model 6302G
Model 6302GADH
User Manual
EN
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Download from Www.Somanuals.com. All Manuals Search And Download.
CONTENTS
General Guidelines............................................................................................................................................................................................................................. 6
Important Safety Information........................................................................................................................................................................................................... 6
Preparing the Work Area................................................................................................................................................................................................................. 14
Tub Dimensions and Connection Details..................................................................................................................................................................................... 15
Securing the Tub................................................................................................................................................................................................................................. 16
Leveling the Tub.................................................................................................................................................................................................................................. 16
2
Download from Www.Somanuals.com. All Manuals Search And Download.
CONTENTS
Cleaning and Disinfecting.................................................................................................................................................................................................................. 23
Maintenance for 6302GADH Adjustable Height Rub Panels and Glide Pads....................................................................................................................... 30
Troubleshooting.................................................................................................................................................................................................................................. 30
3
Download from Www.Somanuals.com. All Manuals Search And Download.
1 GENERAL
1 General
1.1 Symbols
Warnings
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. See the
information below for definitions of the signal words.
DANGER
ƽ
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
ƽ
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
ƽ
Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or
both.
IMPORTANT
!
Indicates a hazardous situation that could result in damage to property if it is not avoided.
Gives useful tips, recommendations and information for efficient, trouble-free use.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
1 GENERAL
1.2 Limited Warranty
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW
APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of our products.
This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state.
Invacare Continuing Care, Inc. (ICC1) warrants its bathing/whirlpool products to be free from defects in materials and workmanship for a period of three
years from the date of purchase. Warranty obligation is limited to parts replacement only. Warranty replacement parts are only covered for the duration of
the warranty period. All purchased replacement parts will carry a 30-day warranty from the date of shipment. If within such warranty period any such
product shall be proven to be defective, such product shall be repaired or replaced, at ICCI's option. This warranty does not include any labor or shipping
charges incurred in replacement part installation or repair of any such product. ICCI's sole obligation and your exclusive remedy under this warranty shall be
limited to such repair and/or replacement.
For warranty service, please contact the dealer from whom you purchased your ICCI product. In the event you do not receive satisfactory warranty service,
please write directly to ICCI at the address below. Provide dealer's name, address, model number, date of purchase, indicate nature of the defect and, if the
product is serialized, indicate the serial number.
ICCI will issue a return authorization. The defective unit or parts must be returned for warranty inspection using the serial number, when applicable, as
identification within thirty days of return authorization date. DO NOT return products to our factory without our prior consent. C.O.D. shipments will be
refused; please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO
ADHERE TO THE ENCLOSED INSTRUCTIONS. LABOR; EXPENDABLE ITEMS, I.E. SHOWER HOSES, VALVES, “O” RINGS, SEALS; PARTS BROKEN
DUE TO ABUSE OR IMPROPER USE; ITEMS THAT BECOME INOPERABLE DUE TO LACK OF MAINTENANCE, OR NON COMPATIBLE
CHEMICALS; STAINS DUE TO HARD WATER, IODINE SOLUTIONS, OR OTHER UNAPPROVED CHEMICALS; “SPIDER” CRACKS OR
“CRACKING” DUE TO MISHANDLING OR ABUSE. IN ADDITION, THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED
PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER
OPERATION, MAINTENANCE OR STORAGE; OR PRODUCTS MODIFIED WITHOUT ICCI'S EXPRESS WRITTEN CONSENT INCLUDING, BUT
NOT LIMITED TO: MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS: PRODUCTS DAMAGED BY
REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF ICCI; PRODUCTS DAMAGED BY
CIRCUMSTANCES BEYOND ICCI'S CONTROL; PRODUCTS REPAIRED BY ANYONE OTHER THAN AN ICCI DEALER, SUCH EVALUATION
SHALL BE SOLELY DETERMINED BY ICCI.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR
IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE
REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE
PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT
EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. THE MANUFACTURER SHALL NOT BE LIABLE FOR
ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
2 SAFETY
2 Safety
The Safety section contains important information for the safe operation and use of this product.
2.1 General Guidelines
WARNING
ƽ
DO NOT use this product or any available optional equipment without first completely reading and understanding these
instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with
this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare
professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
ACCESSORIES WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories
designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
ƽ
!
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.
2.2 Important Safety Information
DANGERS - ELECTRICAL SHOCK HAZARD
Grounding is required.
The installation and maintenance of the tub MUST be done by a qualified technician, certified electrician or plumber.
Disconnect electrical supply or turn off circuit breaker before installing or performing any maintenance on the tub.
The unit MUST be connected only to a supply circuit that is protected by a ground fault circuit-interrupter (GFCI).
Be certain ALL electrical work is performed by a qualified electrician and is in compliance with local electrical codes.
ALL electrical connections are to be water resistant
ƽ
DANGER - RISK OF SCALDING
The hot water supply to the tub MUST not exceed 110°F (43° C) to protect the user from scalding.
ƽ
6
Download from Www.Somanuals.com. All Manuals Search And Download.
2 SAFETY
WARNING - PLUMBING
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes.
ƽ
ƽ
WARNING - FLOOR LOAD CAPACITY
The installation area MUST be able to support 66 pounds per square inch (45 n/cm²) load capacity of the tub. Otherwise, damage
may occur.
WARNINGS - SAFE USE
ƽ
Use this unit only for its intended use as described in this manual.
If any part of the system is not functioning properly, DO NOT use the tub.
NEVER drop or insert any object into any opening.
DO NOT operate the tub without the jet assemblies installed.
The tub MUST be cleaned and disinfected after each use. Refer to Cleaning and Disinfecting on page 23.
Keep hair and body away from the jet assemblies when pump is running.
Hyperthermia occurs when the internal temperature of the body reaches a level several degrees above the normal body
temperature of 98.6° F. The symptoms of hyperthermia include an increase of the internal temperature of the body, dizziness,
lethargy, drowsiness, and fainting.
The tub is specifically designed to be used with a bath lift. ICCI recommends use of the Stretcher Lift Model 1200 ONLY. Bath lifts
designed by other manufacturers are NOT to be used with the tub. Use the Stretcher Lift Model 1200 to maintain resident safety
and product utility. Refer to the lift owner’s manual for complete lift operating and safety instructions.
To reduce the risk of injury, DO NOT permit children or infirm persons to use this tub UNLESS THEY ARE SUPERVISED AT
ALL TIMES.
Recommended length of time spent in a bath is 5 to 10 minutes. However, if you or the person you are caring for becomes
nauseated or dizzy or the person develops a headache, get out of the tub at once and safely bring the body temperature back to
normal (98.6°F).
Seek medical attention if symptoms or sickness persist. Check with the physician BEFORE using the tub again.
Anyone under a doctor’s supervision, pregnant, or with poor health should check with a physician BEFORE using this tub.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
3 OVERVIEW
3 Overview
3.1 Features
Tub Raise and Lower Buttons (Model 6302GADH ONLY)
Water Temperature Thermometer
Whirlpool ON/OFF Button
Fill Button
(Push-in once to auto-fill,
push-in and hold to add
additional water to tub)
Water Temperature Control
8
Download from Www.Somanuals.com. All Manuals Search And Download.
3 OVERVIEW
Shower/Massage ON/OFF Button
Shower/Massage Wand
Fill Spout
Overflow
Whirlpool Jet
Shower Spray
Adjustment
Flow Control
Valve
9
Download from Www.Somanuals.com. All Manuals Search And Download.
3 OVERVIEW
3.3 Typical Product Parameters
6302G
FIXED HEIGHT
6302GADH
ADJUSTABLE HEIGHT
Weight
Tub (shipping weight)
820 pounds (371 kg)
Tub (empty)
Tub w/103 gallons of water (max water capacity)
Floor Load Capacity
795 pounds (360 kg)
1,695 pounds (767 kg)
66 PSI (45 n/cm²)
Height
Tub Rim
Raised
Lowered
32 inches (813 mm)
N/A
N/A
41 inches (1041 mm)
27 inches (686 mm)
Overall Tub Height
Raised
56 inches (1422 mm)
N/A
N/A
65 inches (1651 mm)
51 inches (1295 mm)
Lowered
Width
Tub Exterior
38 inches (965 mm)
30 inches (762 mm)
25 inches (635 mm)
22 inches (559 mm)
Tub Interior at Shoulder
Tub Interior at Hip
Tub Interior at Foot
Length
Overall Tub Length
Tub Interior
Frame Base
94 inches (2388 mm)
69 inches (1753 mm)
49 inches (1245 mm)
Water Capacity
Maximum
103 gallons (390 liters)
with Tub occupied (average usage)
70-77 gallons maximum (265 to 292 liters)
Water Fill Rate
17 gallons per minute (65 liters per minute)
11
Download from Www.Somanuals.com. All Manuals Search And Download.
3 OVERVIEW
6302G
6302GADH
FIXED HEIGHT
ADJUSTABLE HEIGHT
Water Pressure
Water Drain Rate
Number of Jets
65 PSI maximum (448 kPa)
25 gallons per minute (95 liters per minute)
4
Electrical
Classification
Power
Current Rating
Class I, Type B
115 VAC, 60 Hz
7 Amps
Tub Material:
Double-wall gel coated fiberglass
2 inch dial, 75° to 120° F (25° - 50° C)
Moen ExactTemp®
Thermometer:
Anti-Scald Water Regulator:
12
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4 Installation
WARNINGS
ƽ
ƽ
Check all parts for shipping damage. In case of damage, DO NOT use. Contact the carrier for further instruction.
The installation area MUST be able to support the 66 pounds per square inch (45 n/cm²) load capacity of the tub. Otherwise, injury or
damage might occur.
Access for servicing the tub MUST be provided. Install the tub in a location that will permit access to the service panels, plumbing and
electrical hookup.
CAUTIONS
DO NOT lift the tub by the plumbing.
DO NOT modify the tub in any way. Modifications to the system or any component may cause the system to malfunction and will void the
warranty.
4.1 Preparing the Work Area
Access for servicing the tub MUST be provided.
1. Carefully remove the tub from the carton and inspect for shipping damage. If damage is found, contact the carrier.
2. Remove the tub from the shipping pallet by removing the screws from the leg mounting pads.
Follow the recommended floor plan when installing the tub. Refer to Minimum Floor Space and Clearance Requirements on page 13.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4.2 Tub Dimensions and Connection Details
94”
29”
D
5”
25”
30”
38”
H
C
E
5.5”
Connection
Location
Type
C
D
E
10” (254 mm) from centerline, 5.5” (140 mm) up from floor
3/4” Cold water stub (with recommended shut off valve)
2” Drain (floor drain recommended)
5” (127 mm) from centerline, 29” (736 mm) from tub back edge
10.5” (267 mm) from centerline
115V AC, 60 Hz 15 A Input
H
6.5” (165 mm) from centerline, 5.5” (140 mm) up from floor
3/4” Hot water stub (with recommended shut off valve)
15
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4.3 Securing the Tub
SECURING LEGS TO FLOOR
WARNINGS
Protective eye wear MUST be worn when drilling to
avoid possible eye injury.
Leg
ƽ
Leg Mounting Pad
DO NOT adjust height of legs more than 1 inch. The leg
will disassemble from leg mounting pad and serious
injury may occur.
Lag Bolt
1. Set the tub in place.
2. Ensure the plumbing and electrical hook-ups have been installed.
Lag Bolt
3. Mark the locations of the anchor holes, using pre-drilled holes in leg
mounting pads as a template.
Anchors
4. Move the tub away from the mounting area.
5. Drill anchor holes in floor. Use jack-drill and carbide bit in concrete
or a wood bit for wood floors.
LEVELING SIDE TO SIDE
Level
6. Reposition the tub in the mounting area, aligning the holes in the leg
mounting pads with the holes drilled in STEP 4.
4.4 Leveling the Tub
1. Level the tub by adjusting the telescopic legs either up or down,
using a wrench, to arrive at the desired height.
LEVELING END TO END
2. Check to make sure the tub is level by placing a level on the front
and sides of the tub.
Level
3. Repeat STEPS 1 and 2 until the tub is level.
4. Secure the legs to the floor using fasteners appropriate to the floor
material and in keeping with local codes.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4.5 Electrical Installation
DANGERS - ELECTRICAL SHOCK HAZARD
Grounding is required.
The installation and maintenance of the tub MUST be done by a qualified technician, certified electrician or plumber.
Disconnect electrical supply or turn off circuit breaker before installing or performing any maintenance on the tub.
The National Electric Code requires the use of a ground fault circuit interrupter on all therapeutic pools and tubs (hydro therapeutic
tanks). Failure to follow this requirement could result in serious or fatal injury. Use a 15 AMP GFCI breaker only.
Be certain ALL electrical work is performed by a qualified electrician and is in compliance with local electrical codes.
ALL electrical connections MUST be water resistant.
ƽ
ƽ
WARNINGS
The unit MUST be connected only to a supply circuit that is protected by a ground fault circuit-interrupter (GFCI). Such a GFCI should
be provided by the installer and should be tested on a routine basis. To test the GFCI, push the test button. The GFCI should interrupt
power. Push the reset button. Power should be restored. If the GFCI fails to operate in this manner, GFCI is defective. If the GFCI
interrupts power to the bathtub without the test button being pushed, a ground current is flowing, indicating the possibility of an electric
shock. DO NOT use this tub. Disconnect the tub and have the problem corrected by a qualified service representative BEFORE use.
All electrical conduit, fittings and wire to unit are to be supplied and installed by a professional electrician in accordance with the National
electrical code as a minimum requirement, including the supply and installation of a Ground Fault Circuit Interrupter (GFCI).
For Canadian installation, ground fault interrupter and a 30-minute maximum timer MUST be installed in the electrical circuit. (Parts not
supplied by ICCI). The unit must be operated through the circuit. The 30-minute timer must be located in a position where a person in
the unit cannot reach the timer.
Use a 15 AMP GFCI breaker only.
Ensure a 15 AMP GFCI breaker is installed to protect the unit. Insert the plug into a pre-installed GFCI protected hospital grade receptacle.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4.6 Plumbing Connection Details
Wall
38 in.
Drain
MUST be a
minimum of 6 in
above the floor
3/4 in. Ball Valve
(965 mm)
Centerline of Tub
5 in.
(127 mm)
Tub
Back of Tub
Front of Tub
Elbow
Accommodate Reduced Pressure Zone (RPZ) if required.
2 in. PVC stub
Drain
4.7 Plumbing Installation
DANGER
ƽ
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes. Check local codes regarding
maximum water temperature allowed at hot water outlets for facility type.
WARNINGS
ƽ
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes.
Some state and local plumbing codes require the installation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and
cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing
code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be
provided by the customer if required.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
CAUTIONS
ƽ
Purge water lines. Hot and cold lines MUST be purged (blown down) to clear lines of debris prior to connection. Valves and spray heads
may become inoperable and/or damaged if lines are not properly cleared. Items damaged in this way are not covered under the warranty.
When operating water pressure exceeds 55 PSI, pressure regulators (similar to the ¾-inch Watts U5) MUST be installed.
1. Refer to Tub Dimensions and Connection Details on page 15 for the location of the hot water stub and cold water stub.
2. Measure the location of the drain on the unit. Install the drain in the floor.
3. Purge water lines. Hot and cold lines MUST be purged (blown down) prior to connection, to clear lines of debris.
4. Install ON/OFF valves at wall and use unions between valves and stubs.
5. Connect 2-inch drains in accordance with local plumbing codes.
6. Connect hot and cold supply lines to ¾-inch copper pipes found on back of tub in accordance with applicable building codes.
7. Attach the shower/massage wand and flow valve. Be sure to use the washers supplied.
8. Check for leaks in supply or drain lines and check operation of tub. Refer to Operation on page 21.
In House Water Service
HOT
COLD
Optional Pressure Regulator
Mixing Valve
Tub Spout
Shower Control Valve
Tub Fill Valve
Shower/Massage
Wand
Backflow Preventor
(located inside tub cabinet)
19
Download from Www.Somanuals.com. All Manuals Search And Download.
4 INSTALLATION
4.8 Post Installation Checklist
Before putting the tub into service, check the following:
❑
❑
❑
❑
❑
❑
Water lines have been purged of any foreign matter.
Both hot and cold water lines are properly connected to the tub.
All water lines are checked for leaks.
Power line is protected by a 15 AMP GFCI.
ICCI skin care and cleaning products have been ordered. To place your order, call 800-668-2337.
ICCI representative has been called to arrange for an In-Service on the tub. To schedule, call 800-668-2337.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
5 Operation
5.1 Overview
WARNINGS
ƽ
Read and understand ALL information in this manual BEFORE use.
Clean and disinfect the tub after EACH use to avoid resident infection and contamination of the tub.
DO NOT clean or disinfect the tub when it is occupied. Otherwise injury may occur.
To reduce the risk of injury, DO NOT permit children or infirm persons to use this tub UNLESS THEY ARE SUPERVISED ALL TIMES.
Recommended length of time spent in a bath is 5 to 10 minutes. However, if you or the person you are caring for becomes nauseated or
dizzy or the person develops a headache, get out of the tub at once and safely bring the body temperature back to normal (98.6°F). Seek
medical attention if symptoms or sickness persist. Check with the physician BEFORE using the tub again.
Anyone under a doctor’s supervision, pregnant, or with poor health should check with a physician BEFORE using this tub.
The Supine Tub is specifically designed to be used with a bath lift. ICCI recommends use of the Traverse Lift Model 1200 ONLY. Bath lifts
designed by other manufacturers are NOT to be used with the Supine Tub. Use the Traverse Lift Model 1200 to maintain resident safety
and product utility. Refer to the lift owner’s manual for complete life operating and safety instructions.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
Tub Raise and Lower Buttons (Model 6302GADH ONLY)
Water Temperature Thermometer
Whirlpool ON/OFF Button
Fill Button
(Push-in once to auto-fill,
push-in and hold to add
additional water to tub)
Water Temperature Control
22
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
Perform these procedures in the following order:
1. Make sure that the tub has been cleaned, disinfected and rinsed BEFORE each use. Refer to Cleaning and Disinfecting on page 23.
2. Use the stretcher lift to transfer the resident to the tub.
The resident will remain on the stretcher lift during bathing.
3. Bathe the resident.
4. Drain the tub.
5. Transfer the resident from the tub.
6. Clean, disinfect and rinse the tub.
5.2 Cleaning and Disinfecting
WARNINGS
ƽ
ƽ
Clean and disinfect the tub after EACH use to avoid resident infection and contamination of the tub.
DO NOT clean or disinfect the tub when it is occupied. Otherwise injury may occur.
Read and understand ALL information on disinfecting BEFORE use.
ALWAYS wear rubber gloves, an apron and a face shield when using disinfectant. In case of eye or skin contact, follow procedures on the
disinfectant container.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of these types of cleaners will
make the gel-coat finish of your tub scratched and dull. NEVER use cleaners containing Iodine, Household Bleach, Bromine, Betadine® or
Methylene Blue. The use of these types of cleaners will cause the tub to stain. Use of unapproved cleaners will dry out the rubber seals
and gaskets and the tub will not function properly.
1. Use the drain plug to close the drain.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
2. Remove and disassemble all jet assemblies.
A. Rotate each jet assembly counterclockwise and pull away from the jet casing. (Detail “A”).
B. Remove each impeller cap by rotating it clockwise (Detail “B”).
C. Remove each impeller (Detail “C”).
Detail “B”
Detail “A”
Detail “C”
Jet Casing
Impeller Cap
Impeller
Impeller
Jet Assembly
3. Clean the pieces with Dispatch™ Cleaner and Disinfectant. Lay the pieces in the bottom of the tub.
4. Spray all surfaces of the tub with Dispatch Cleaner and Disinfectant. Take a long handled brush and thoroughly clean the interior surfaces of the
tub and the jet casings.
5. Allow the Dispatch Cleaner and Disinfectant to sit on the surfaces for one minute.
6. Rinse all surfaces and pieces of the jet assemblies with water.
7. Use a clean towel to dry all tub surfaces.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
8. Replace the impeller and impeller cap on each jet assembly by
rotating them counterclockwise.
CAUTION
Jet Casing
ƽ
Cross-threading the jet assembly onto the casing may
cause damage to the whirlpool motor. After installation,
ensure that the jet assembly is flush against the jet
casing.
Sanijet™
Logo
9. Assemble each jet assembly to the jet casings.
A. Position the Sanijet® logo and red indicator at the 11 o’clock
position.
Jet
Assembly
B. Insert the jet assembly into the jet casing and rotate the jet
assembly clockwise until it locks into place.
Removing Soap Film and Hard Water Deposits
Use a commercially available cleaner such as Invacare Heavy Duty Cleaner. Remove the jet assemblies. Remove the impeller caps. Thoroughly spray
the exposed surfaces, including the jet casings. Use a long handled soft brush or similar product to wipe down all surfaces and allow the heavy duty
cleaner to remain in contact with the exposed surfaces for approximately ten minutes. Rinse off surfaces using the shower wand, adjusted to spray
position.
Removing Calcium Deposits, Scale and Lime Build-up
Spray a commercially available cleaner (full strength) for the intended application on all surfaces to be cleaned. Allow the cleaner to remain in contact
with surfaces for at least five minutes to as long as 12 hours. Rinse off surfaces using the shower wand, adjusted to the spray position. Use the cleaner
according to the manufacturer's instructions.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
5.3 Water Temperature Setting
WARNING
ƽ
The causes, symptoms and effects of hyperthermia may be described as follows: Hyperthermia occurs when the internal temperature of
the body reaches a level several degrees above the normal body temperature of 98.6° F. The symptoms of hyperthermia include an
increase in the internal temperature of the body, dizziness, lethargy, drowsiness, and fainting.
The effects of hyperthermia include:
•
•
•
•
•
•
•
Failure to perceive heat
Failure to recognize the need to exit the spa or hot tub
Unawareness of impending hazard
Fetal damage in pregnant women
Physical inability to exit the spa or hot tub
Unconsciousness resulting in the danger of drowning
The use of alcohol, drugs, or medication can greatly increase the risk of fatal hyperthermia
Flow Water Thermometer - measures the temperature of water coming from the tub fill spout or shower/massage wand.
Water Temperature Control Valve - sets and maintains constant temperature for all incoming water.
Flow Water
Thermometer
Water
Temperature
Control
Valve
Cold (BLUE)
Hot (RED)
26
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
5.4 Filling the Tub
1. Use the drain plug to close the drain.
2. Position the resident (on the stretcher lift) in the tub.
3. Push the tub fill button once to start auto-fill.
The tub will fill to a predetermined level. To stop the tub from filling during auto-fill, press the fill button a second time (there will
be a five second delay). After the tub has auto-filled, push in and hold the tub fill button to add additional water.
4. Use the water temperature control valve to adjust the water temperature while watching the flow water thermometer. Test the water temperature
with your hand.
5.5 Using the Shower/Massage Wand
These instructions are to be done while holding the shower/massage wand over the tub.
1. Turn the shower/massage on by pushing the ON/OFF button.
2. Adjust the water temperature using the water temperature valve while watching the flow water thermometer. Test the temperature with your hand.
3. Adjust the flow of the water by turning the flow control.
4. Adjust the shower spray by rotating the shower wand head.
5. Turn the shower/massage off by pressing the ON/OFF button again.
Refer to the following chart for shower head adjustment settings.
Stream
Pulse Massage
Shower Spray
Mixture of Shower Spray and Pulse Massage
Flow Control serves as a volume control and water saver.
The use of the Flow Control permits the reduction of water flow without the need for readjustment of water temperature during bathing.
A trickle of water through the Flow Control when in the closed position is normal.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
5 OPERATION
5.6 Using the Whirlpool Function
1. If the water level does not cover the whirlpool jets, push the tub fill button to auto-fill the tub.
2. After the water level covers the whirlpool jets and after bathing the resident, add Invacare Defoamer to the bath water. Refer to the Defoamer
bottle for directions.
3. Push the whirlpool button to start whirlpool jets.
4. After use, press the whirlpool button again to stop the whirlpool jets.
5. Drain the tub completely.
6. Use the stretcher lift to transfer the resident from the tub.
5.7 Adjusting the Tub Height - 6302GADH Only
WARNING
ƽ
ƽ
DO NOT stand on the tub frame. Keep feet away from
tub and frame, otherwise injury may occur.
UP Button
CAUTION
DO NOT adjust the height more than three cycles (up
and down) within a ten minute period. Damage to the
actuator may result.
DOWN Button
DO NOT adjust the tub so it hits the stretcher lift. If
necessary, adjust the height of the stretcher lift.
Otherwise damage to the tub may occur.
The tub stops automatically when the maximum
height adjustment UP or DOWN is reached.
Perform one of the following:
•
Press and hold the UP button. Release the UP button when the desired height is reached.
•
Press and hold the DOWN button. Release the DOWN button when the desired height is reached.
28
Download from Www.Somanuals.com. All Manuals Search And Download.
6 INSPECTION, MAINTENANCE AND TROUBLESHOOTING
6 Inspection, Maintenance and Troubleshooting
If any part of the system is not functioning properly, DO NOT use the system. For parts and technical service information please contact ICC Technical
Service at (800) 668-2337, Extension 2.
6.1 Safety Inspection Checklists
Every six months or as necessary have a qualified technician perform a thorough inspection and servicing. Routine maintenance will
extend the life and efficiency of your tub.
Initially
❑
❑
❑
❑
❑
❑
❑
Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 23.
Visually inspect plumbing for leaks. Make repairs as necessary.
Ensure that the maximum water supply temperature does not exceed 110°F.
Ensure that the maximum water pressure does not exceed 55 PSI.
Verify the water flow temperature with a separate thermometer.
Visually inspect for cracks in the tub surface.
Check that all labels are present and legible. Replace if necessary.
Before Each Use
❑
❑
❑
❑
Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 23.
Ensure that the maximum water fill temperature does not exceed 110°F (43° C).
Ensure that the maximum water pressure does not exceed 55 PSI.
Visually inspect for cracks in the tub surface.
Every Six Months
❑
❑
❑
❑
❑
❑
Verify the water flow temperature with a separate thermometer.
Visually inspect plumbing for leaks. Make repairs as necessary.
Visually inspect the power cord for damage or breakage. Replace if necessary.
Check that all labels are present and legible. Replace if necessary.
Clean the Water Temperature Control Valve to remove lime deposits. Refer to Removing Calcium Deposits, Scale and Lime Build-up on page 25.
6302GADH ONLY - Apply white lithium grease to Glide Pads. Refer to Refer to Maintenance for 6302GADH Adjustable Height Rub Panels and Glide
29
Download from Www.Somanuals.com. All Manuals Search And Download.
6 INSPECTION, MAINTENANCE AND TROUBLESHOOTING
6.2 Maintenance for 6302GADH Adjustable Height Rub Panels and Glide Pads
The application of white lithium grease to the glide pads must be done every six months to keep the tub in good working order.
1. Remove the side panel to access the rub panels and glide pads.
2. Generously apply white lithium grease to the stainless steel rub panels and glide pads.
3. Repeat STEPS 1-2 for the opposite side.
4. Replace the side panels.
6.3 Troubleshooting
If problems are not remedied, contact technical support at 1-800-668-2337, Extension 2.
SYMPTOMS
FAULTS
SOLUTION
Hydromassage System
System will not start
No electrical power to tub
Switch air hose disconnected
Water deposits in whirlpool jets
Check and reset breaker
Reconnect hose to switch and controller
Disassemble, inspect and clean whirlpool jets
System is ineffective
System is noisy
Whirlpool jets are turned on without water Turn off the whirlpool jets. Add water to the tub to completely cover the
completely covering the jet assemblies jet assemblies and test the whirlpool function again.
Tub Operation
No water flow
Water supply turned off/no electrical con- Turn on water supply/check and reset breaker
nection
Restricted water flow
Water temperature control valve
Inspect and clean the water temperature control valve. Refer to the Moen
Temperature Control Valve, Service Manual Number 153768.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
6 INSPECTION, MAINTENANCE AND TROUBLESHOOTING
SYMPTOMS
FAULTS
SOLUTION
Height Adjustment (Model 6302GADH Only)
Tub will not raise or
lower
Glide pads and rub panels not greased
Apply grease to glide pads and rub panels. Refer to Maintenance for
Bad electrical connections, no electrical
power to tub
Unplug the tub. Check and reset GFCI. Check electrical cord, plug and
outlet for damage. Replace damaged parts as necessary per local codes.
Actuator overload
Wait five minutes and try to adjust the height again.
Contact your Invacare Dealer.
Actuator burned out
Obstacle in UP/DOWN travel path
Inspect for and remove any obstacle.
Maintenance Record
DATE
MAINTENANCE PERSON TYPE OF MAINTENANCE PERFORMED
NOTES
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Invacare Corporation
USA and Canada
One Invacare Way
Elyria, Ohio USA
44036-2125
Manufacturing Location
899 Cleveland Street
Elyria, Ohio 44036
© 2011 Invacare Corporation. All rights
reserved. Republication, duplication or
modification in whole or in part is prohibited
without prior written permission from Invacare.
Trademarks are identified by ™ and ®. All
trademarks are owned by or licensed to Invacare
Corporation or its subsidiaries unless otherwise
noted.
Customer Service
1-800-668-2337
Technical Support
1-800-668-2337, Extension 2
Dispatch is a registered trademark of Clorox
Industries, Inc.
Sanijet and Pipeless TM are registered
trademarks of Maple Industries, LLC, d.b.a.
Sanijet.
ExactTemp is a registered trademark of Moen
Incorporated.
Betadine is a registered trademark of Purdue
Pharma LP.
For more information regarding Invacare products, parts, and
Part No. 1167456
Rev C - 12/11
Invacare Continuing Care, Inc. is a subsidiary of Invacare Corporation
Download from Www.Somanuals.com. All Manuals Search And Download.
|