User Manual
Perfecto2™Series
Oxygen Concentrators without SensO2®,
Oxygen Concentrators with SensO2
HomeFill®Compatible
Perfecto2 with SensO2
Perfecto2
Model IRC5PO2
Model IRC5PO2W
Model IRC5PO2V
Model IRC5P
Model IRC5PW
Dealer: This manual MUST be given to the end
user of this product.
User: BEFORE using this product, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
TABLE OF CONTENTS
To Reduce the Risk of Burns, Electrocution, Fire or Injury to
Persons......................................................................................................... 7
Unpacking............................................................................................11
Inspection............................................................................................11
Storage.................................................................................................11
PARAMETERS ............................................... 12
Introduction........................................................................................15
Set Up ..................................................................................................16
Power Switch .....................................................................................18
Flowrate...............................................................................................19
3
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or
unsafe practices which could result in personal injury or
property damage. Refer to the following table for definitions
of the signal words.
SIGNAL WORD
MEANING
Danger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
DANGER
Warning indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
WARNING
Caution indicates a potentially hazardous situation
which, if not avoided, may result in property
damage or minor injury or both.
CAUTION
NOTICE
The information contained in this document is subject to
change without notice.
ꢁ DANGER
DO NOT SMOKE while using this device. Keep all
matches, lighted cigarettes or other sources of ignition
out of the room in which this product is located and away
from where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIAL NOTES
CAUTION
“Caution: Federal law restricts this device to sale or rental by
or on order of a physician, or any other practitioner licensed
by the law of the State in which he/she practices to use or
order the use of this device.”
Invacare recommends an alternate source of supplemental
oxygen in the event of a power outage, alarm condition or
mechanical failure. Consult your physician or equipment
provider for the type of reserve system required.
This equipment is to be used as an oxygen supplement and is
not considered life supporting or life sustaining.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
LABEL LOCATION
LABEL LOCATION
Perfecto2™
FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual)
DANGER - RISK OF FIRE
NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and
away from areas where oxygen is being delivered. Textiles, oil and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
DANGER - ELECTRIC SHOCK
DO NOT remove cover. Refer servicing to qualified service personnel.
Flowrates below 0.5 L/min. will trigger Low Flow Alarm (rapid audible alarm).
Set flowmeter at 3.0 L/min. or less when filling HomeFill® cylinders.
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
HomeFill®
Compatible
Serial Number Label is
located on the resonator
intake assembly
Specification Label is
located on the back
near the bottom
6
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
In order to ensure the safe installation, assembly and
operation of the Perfecto2 concentrator these instructions
MUST be followed.
ꢀ WARNING
This section contains important information for the safe
operation and use of this product.
To Reduce the Risk of Burns,
Electrocution, Fire or Injury to Persons
ꢁ DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing
to qualified service personnel. No user serviceable parts.
Avoid using while bathing. If continuous usage is required
by the physician’s prescription, the concentrator MUST be
located in another room at least 7 ft from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into
water or other liquid.
DO NOT reach for product that has fallen into water.
Unplug IMMEDIATELY.
If the concentrator has a damaged cord or plug, if it is not
working properly, if it has been dropped or damaged, or
dropped into water, call qualified technician for
examination and repair.
A spontaneous and violent ignition may occur if oil, grease
or greasy substances come in contact with oxygen under
pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all
other oxygen equipment. DO NOT use any lubricants
unless recommended by Invacare.
Avoid creation of any spark near medical oxygen
equipment. This includes sparks from static electricity
created by any type of friction.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
Operating Information
For optimum performance, Invacare recommends that each
concentrator be on and running for a minimum of 30 minutes
at a time. Shorter periods of operation may reduce maximum
product life.
Keep the oxygen tubing, cord, and unit out from under such
items as blankets, bed coverings, chair cushions, clothing and
away from heated or hot surfaces, including space heaters,
stoves and similar electrical appliances.
DO NOT move or relocate concentrator by pulling on the
power cord.
NEVER drop or insert any object or liquid into any opening.
Invacare recommends that Crush‐Proof oxygen tubing be
used with this product and not exceed 50 ft in length.
There are no user serviceable parts. This does not include
normal maintenance items. See maintenance section for user
maintenance items.
A product should NEVER be left unattended when plugged
in. Make sure the Perfecto2 is Off when not in use.
Close supervision is necessary when this product is used near
children or physically‐challenged individuals.
Additional monitoring or attention may be required for
patients using this device who are unable to hear or see
alarms or communicate discomfort.
DO NOT connect the concentrator in parallel or series with
other oxygen concentrators or oxygen therapy devices.
Radio Frequency Interference
This equipment has been tested and found to comply with
EMC limits specified by IEC/EN 60601‐1‐2. These limits are
designed to provide a reasonable protection against
electromagnetic interference in a typical medical installation.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
Other devices may experience interference from even the low
levels of electromagnetic emissions permitted by the above
standards. To determine if the emissions from the Perfecto2 is
causing the interference, turn the Perfecto2 Off. If the
interference with the other device(s) stops, then the Perfecto2
is causing the interference. In such rare cases, interference
may be reduced or corrected by one of the following
measures:
• Reposition, relocate, or increase the separation
between the equipment.
• Connect the equipment into an outlet on a circuit
different from that to which the other device(s) is
connected.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 2—FEATURES
SECTION 2—FEATURES
Oxygen Outlet
Oxygen Purity Indicator
Lights / Fault and Power
Indicator Lights
Flowmeter
Circuit Breaker
Power Switch
Elapsed Time
Meter
ACCESSORIES (NOT SHOWN):
PreciseRX™ Pediatric Humidifier/Flowmeter
Accessory - IRCPF16 (Perfecto2 5 Only)
HomeFill home oxygen compressor - IOH200
REAR VIEW
*NOTE: This outlet fitting is to be
used only for filling oxygen cylinders
with the HomeFill home oxygen
compressor. The outlet fitting does not
affect concentrator performance.
Refer to the HomeFill ownerʹs
manual, part number 1100873, for
connection and operating
Cabinet Filter
instructions. When not in use, the
plug provided with the concentrator
should be inserted into the outlet
fitting. For more information about
the HomeFill, contact your Invacare
dealer.
Power
Cord
*Outlet
Fitting
10
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 3—HANDLING
SECTION 3—HANDLING
The concentrator should ALWAYS be kept in the upright
position to prevent cabinet damage while being transported.
If the concentrator is to be reshipped by common carrier,
additional cartons are available from Invacare.
Unpacking
NOTE: For this procedure, refer to FIGURE 3.1.
1. Check for any obvious
damage to the carton or
its contents. If damage is
evident, notify the
carrier, or your local
dealer.
2. Remove all loose packing
from the carton.
3. Carefully remove all the
components from the
carton.
NOTE: Unless the oxygen concentrator is to be used immediately,
retain containers and packing materials for storage until use of the
concentrator is required.
Inspection
Inspect/examine exterior of the oxygen concentrator for nicks,
dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
11
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
Sound Level:
3 A
43 ± 2 dBA Average for Perfecto2 Models and
Perfecto2V Model
40 ± 2 dBA Average for Perfecto2W Models
Altitude:
All Perfecto2 Models: Up to 8,000 ft (2438
meters) above sea level without degredation of
concentration levels. From 8,000 ft to 13,129
(4000 meters) below 90% efficiency.
All Perfecto2 Models while filling a
HomeFill System: Up to 6,000 ft (1828 meters)
above sea level without degradation of
concentration levels. From 6,000 ft (1828 meters)
to 13,129 ft (4000 meters) below 90% efficiency.
*Oxygen Output
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Concentration Levels:
Maximum Outlet Pres-
sure:
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range:
0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Low Flow Alarm:
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories
Connected).
Average Power
Consumption:
Perfecto2 and Perfecto2W Models: 280 W
(Operating @ 3 L/min)
Perfecto2V Model: 325 W
Pressure Relief Mecha-
nism Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
12
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Change in maximum
0.7 L/min
recommended flow when
back pressure of 7kPa is
applied:
Filters:
Cabinet, Outlet HEPA and Compressor Inlet
Safety System:
Current overload or line surge shutdown. High
temperature compressor shutdown. High
Pressure Alarm w/compressor shutdown. Low
Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO2 Oxygen
System (SensO2 Model) Low Flow Alarm.
Width:
Height:
Depth:
Weight:
15 in ± 3/8 in (38.1 cm ± 1 cm)
23 in ± 3/8 in (58.4 cm ± 1 cm)
12 in ± 3/8 in (30.5 cm ± 1 cm)
43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto2 Models
45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto2W Models
40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto2V Model
Shipping Weight:
48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto2 Models
50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto2W Models
45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto2V Model
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity Perfecto2 Models and Perfecto2V Model
50°F - 90°F (10°C - 32°C) at 20-60% relative
humidity Perfecto2W Models
Exhaust:
Less than Ambient + 35°F (+19°C)
Less than Ambient + 6°F (+3°C)
Oxygen Output
Temperature:
Cabinet:
Impact Resistant flame-retardant plastic cabinet
that conforms to UL94V-0.
Regulatory Listing:
Double Insulated Product ETL certified
complying with UL1431 and UL1097 ETL certified
to CSA C22.2 No. 68.
Electrical:
No extension cords.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Placement:
Tubing:
No closer than three inches from any wall,
furniture, draperies, or similar surfaces.
7 ft cannula with a maximum 50 ft of Crush-Proof
Tubing (DO NOT pinch).
Relative Humidity:
Time of Operation:
20 to 60%
Up to 24 hours per day.
Recommended Storage
and Shipping Tempera-
ture:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity.
*NOTE: Stated concentration levels achieved after initial warm‐up
period (approximately 30 minutes).
14
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
Introduction
Your oxygen concentrator is intended for individual use
indoors. It is an electronically operated device that separates
oxygen from room air. It provides high concentration of oxygen
directly to you through a nasal cannula. Clinical studies have
documented that oxygen concentrators are therapeutically
equivalent to other types of oxygen delivery systems.
Your provider will show you how to use your oxygen
concentrator. He/She should be contacted with any questions
or problems regarding your oxygen concentrator. This
owner’s manual will tell you about your concentrator and will
serve as a reference as you use your concentrator.
Select a Location
ꢀ WARNING
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening may
be blocked. Keep the openings free from lint, hair and the like.
Keep unit at least 3 inches away from walls, draperies,
furniture, and the like.
You may select a room where using your oxygen concentrator
would be most convenient. Your concentrator can be easily
rolled from room to room on its casters.
Your oxygen concentrator will perform best when operated
under the following conditions. Refer to Typical Product
Parameters on page 12. Usage in environments other than
those described may result in the need for increased
equipment maintenance. The air intake of the unit should be
located in a well ventilated area to avoid smoke, airborne
pollutants and/or fumes. DO NOT place in a closet.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Set Up
Plug in Power Cord
Plug in power cord to an electrical outlet.
Connect Humidifier (If So Prescribed)
ꢀ WARNING
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections.
Water from the humidifier bottle will travel through the
cannula back to the patient.
1. Remove cap from bottle.
2. Fill humidifier with distilled water to the level indicated
by the manufacturer. Replace the humidifier cap and
securely tighten.
Humidifier
Bottle w/o Cap
Humidifier
Bottle with Cap
FIGURE 5.1 Filling the Humidifier
3. Insert a flathead screwdriver in the plate groove on the top
edge of the filter access door and gently pry the filter
access door off (FIGURE 5.2).
4. Pull up and remove the humidifier bottle adapter
16
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Humidifier
Bottle Adapter
Insert Flat Head
Screwdriver
Here
Cabinet Filter
Filter Access Door
FIGURE 5.2 Humidifer Bottle Adapter
5. Attach the humidifier bottle adapter to the humidifier
bottle by turning the wing nut on the humidifier bottle
counterclockwise until it is securely fastened. See Detail
6. Place the humidifier bottle/adapter assembly in the
humidifier compartment on the concentrator. See Detail
“B” in FIGURE 5.3.
7. Attach oxygen tubing from the humidifier bottle/adapter
assembly to oxygen outlet connector on the concentrator.
See Detail “B” in FIGURE 5.3.
8. Attach the cannula/patient supply tubing to the
humidifier bottle outlet. See Detail “B” in FIGURE 5.3.
9. After assembly, ensure that oxygen is flowing through the
cannula.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Oxygen Outlet
Connector
DETAIL “A”
DETAIL “B”
Oxygen
Tubing
Humidifier
Bottle
Humidifier
Bottle
Adapter
Outlet
Humidifier
Bottle
Humidifier
Bottle
Humidifier
Compartment
FIGURE 5.3 Humidifier Compartment
Power Switch
NOTE:Forthisprocedure,referto
On/Off
(I/O)
Power
Switch
1. Press power switch to on
position. All the panel
lights and the audible
alarm will come on for
one second, indicating
thattheunitisfunctioning
properly.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Flowrate
NOTE: For this procedure, refer to FIGURE 5.5 on page 20.
1. Turn the flowrate knob to the setting prescribed by your
physician or therapist.
CAUTION
Never set the flow greater than 5 L/min.
ꢀ WARNING
DO NOT change the L/min setting on the flowmeter unless a
change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate
line on the flowmeter. Next, turn the flow knob until the ball rises to
the line. Now, center the ball on the L/min line prescribed.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min
for more than about one minute, the LOW FLOW alarm
will be triggered. This is a rapid beeping of the audible
alarm. Check your tubing or accessories for blocked or
kinked tubing or a defective humidifier bottle. After rated
flow is restored, between 0.5 L/min and 0.75 L/min, the
LOW FLOW audible alarm will turn off.
NOTE: The use of some accessories such as the pediatric flowstand
and the HomeFill compressor will deactivate the Low Flow Alarm.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Flow Knob
Flowmeter
0.5 L/min
Ball
Front Panel
Front Panel Indicators - Units with
SensO2
The SensO2 feature monitors the purity of oxygen generated
by the oxygen concentrator. If purity falls below factory preset
standards, indicator lights on the control panel will
illuminate.
OXYGEN PURITY
Auto Shut Down
RED
Use Backup
Call Supplier
Below
Normal
YELLOW
Normal
GREEN
O2
20
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up
time (approximately 30 min.) while waiting for the O2 purity to
reach maximum.
LABEL
O2 PURITY
INDICATOR LIGHTS (LED)
SYMBOL
SYSTEM OKAY
O2 over 85% (± 2%)
GREEN Indicator Light
O2
O2 Between 73% (± 3%) YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
to 85% (±2%)
Call a qualified technician. Have a
back-up supply of oxygen ready.
SYSTEM FAILURE
RED Indicator Light
O2 Below 73% (±3%)
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down. Call a qualified technician.
Switch to a back-up supply of oxy-
gen immediately.
Front Panel Indicators - Units without
SensO2
NOTE: For this procedure, refer to FIGURE 5.7.
Auto Shut Down
RED
Use Backup
Call Supplier
Normal
I/O
GREEN
21
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up
time (approximately 30 min). O2 purity will build to a maximum
during this period.
LABEL
SYMBOL
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
GREEN Indicator Light
I/O
SYSTEM FAILURE
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician. Switch
to a back-up supply of oxygen
immediately.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
ꢀ WARNING
The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. Only qualified
personnel should perform preventive maintenance on the
concentrator.
Power should be disconnected before beginning preventive
maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. In places
with high dust or soot levels, maintenance may need to be performed
more often. Refer to Preventive Maintenance Record on page 25.
Routine Maintenance
NOTE: The following routine maintenance should be performed
more frequently. See recommended intervals in each section and
preventive maintenance record.
Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 24.
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean at least once a week depending
on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust, air
pollutants, etc.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6—MAINTENANCE
2. Clean the cabinet filter with a vacuum cleaner or wash in
warm soapy water and rinse thoroughly.
3. Dry the filter thoroughly before reinstallation.
Cleaning the Cabinet
Clean the cabinet with a mild household cleaner and
non‐abrasive cloth or sponge.
Cleaning the Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions
provided by the manufacturer. If none are provided, follow these
STEPS:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten
parts water and one part vinegar.
3. Rinse thoroughly with hot water and refill with distilled
water to the level shown on the humidifier.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6—MAINTENANCE
MODEL NO. IRC
SERIAL NO.
FIGURE 6.2 Preventive Maintenance Record
25
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
1. Insert plug into outlet.
Short beeps, long 1. Power cord not
pause
plugged in.
2. No power at outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
Concentrator not
operating,
power switch On.
3. Tripped circuit
breaker.
3. Call Service Provider.
Beep......Beep......
Alarm:
Continuous
System Failure:
1. Unit overheating due
to blocked air intake.
1a. Remove and clean cabinet
filters.
1b. Move oxygen
Concentrator not
operating, Power
Switch On.
concentrator at least three
inches away from walls,
draperies or furniture.
2. Insufficient power at
outlet.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
Beeeeeeeep....
3. Internal repairs
required.
3. Call Service Provider.
Alarm:
Rapid
Low Flow Alarm:
1. Kinked or blocked
tubing, cannula or
humidifier.
1. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
Beep....Beep...
Beep...Beep
2. Flowmeter set at 0.5
L/min or less.
2. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only
with Perfecto2 5).
26
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
YELLOW or RED 1. Low oxygen purity.*
1. Clean or Replace filters.
light Illuminated.
2. Kinked or blocked
tubing, cannula or
humidifier.*
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
Alarm:
Continuous
On RED light
only.
3. Flowmeter set at 0.5
* Only applies to
IRC5PO2.
3. Check flowmeter is set to
1.0 L/min or more.
L/min or less.*
4. Unit overheating due
to blocked air intake.
4a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three inches from walls,
draperies, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN light with 1. Internal repairs
1. Call Service Provider.
YELLOW Light
Flashing.
required.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
NOTES
28
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
NOTES
29
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
NOTES
30
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITED WARRANTY
LIMITED WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY
WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who purchases
this product when new and unused from Invacare Corporation or a
dealer. This warranty is not extended to any other person or entity and it
is not transferable or assignable to any subsequent purchaser or owner.
Coverage under this warranty will end upon any such subsequent sale or
other transfer of title to any other person. This warranty gives you specific
legal rights and you may also have other legal rights which may vary from
state to state.
Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when
purchased new and unused to be free from defects in materials and
workmanship for a period of three years from date of purchase from
Invacare or a dealer, with a copy of the seller’s invoice required for
coverage under this warranty.
DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE
BODY. This will void any and all warranty applicable to the valve body. If
within such warranty periods any such product shall be proven to Invacare
Corporation's satisfaction to be defective, such product shall be repaired
or replaced, at Invacare Corporation's option. This warranty only applies
to the labor for repairs performed by the Invacare Service Department or
Invacare Authorized Service Centers. It does not apply to the labor
performed by the purchaser or user. This warranty does not include
normal wear and tear or shipping charges. Invacare Corporation's sole
obligation and your exclusive remedy under this warranty shall be limited
to such repair or replacement. Routine maintenance items, such as filters,
are excluded from this warranty.
For warranty service, please contact Invacare Corporation's Service
Department at the toll free number on the back cover. Upon receiving
notice of an alleged defect in a product, Invacare Corporation will issue a
serialized return authorization. It shall then be the responsibility of the
purchaser to return the entire unit or remove, at purchaser's cost, the
defective component part(s) identified, pack the component part(s) in a
manner to avoid shipping damage and to ship the component part(s) to
either Invacare Corporation's plant or service center as specified by
Invacare Corporation in advance. Defective component part(s) MUST be
returned for warranty inspection using the serial number as identification
within thirty days of return authorization date. DO NOT return products
to our factory without prior consent. C.O.D. shipments will be refused;
please prepay shipping charges.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY
SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE,
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE,
SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN
NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS
MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS
WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO,
MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS
OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF
REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC
CONSENT OF INVACARE CORPORATION OR TO PRODUCTS
DAMAGED BY CIRCUMSTANCES BEYOND INVACARE
CORPORATION'S CONTROL.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN
LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER
EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT
OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS
CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED
WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE
DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN.
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR
INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION
OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION
OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE
EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Invacare Corporation
USA
© 2009 Invacare Corporation. All rights
reserved. Republication, duplication or
modification in whole or in part is
prohibited without prior written
permission from Invacare. Trademarks are
identified by ™ and ®. All trademarks are
owned by or licensed to Invacare
Corporation or its subsidiaries unless
otherwise noted.
One Invacare Way
Elyria, Ohio USA
44036-2125
800-333-6900
Technical Services
800-832-4707
Part No 1143482
Rev E - 11/09
Download from Www.Somanuals.com. All Manuals Search And Download.
|